Measures the thickness of non-magnetic materials (example: paint,
plastic, porcelain enamel, copper, zinc, aluminum, chrome, ect.) on
magnetic materials (example: iron, nickel, ect.). Common uses include
measuring thickness of galvanizing layer, lacquer layer, porcelain layer,
phosphide layer, copper tile, aluminum tile, some alloy tile, and paper.
N Probe Description
Measures the thickness of non-magnetic coatings on non-magnetic
metals. Common uses include measuring the layer of varnish, paint,
enamel, plastic coatings, or powder applied to aluminum, brass,
non-magnetic stainless steel.
www
reedinstruments
www.GlobalTestSupply.com
com
4
Operating Instructions
Measuring Procedure
1. Plug in the F-probe or NF-probe and place it away from any
substrate or other metal materials.
2. Press the power button to turn on the meter and to perform the
auto calibration, which takes 3 seconds.
3. The meter will recognize the probe and display F or NF on the
LCD screen.
4. Select the measurement unit by pressing µm/mil button, which
will appear on the LCD screen.
5. Place the probe on a coating layer to be measured.
6. The LCD screen will display the thickness of the coating layer.
7. The reading can be corrected by removing the probe from the
layer and by pressing the plus or minus button.
8. To take another measurement, lift the probe to more than 1
centimeter and then repeat steps 5-7.
9. Turn the meter off by pressing the Power button. The meter will
also turn off after 2 minutes of inactivity.
Zero Adjustment
1. Plug in the F-probe or NF-probe and place it away from any
substrate or other metal materials.
2. Press the power button to turn on the meter and to perform the
auto calibration, which will last 3 seconds.
3. Place the probe on a substrate.
4. Press the Zero button and ‘0' will appear on the display.
5. Do not press the Zero button if the probe is not placed on
a substrate.
6. Select an appropriate calibration foil according to your
measurement range.
7. Place the standard foil onto the substrate. Place the sensor gently
onto the standard foil and lift.
www
reedinstruments
www.GlobalTestSupply.com
com
5
8. The reading on the LCD screen is the value measured.
9. The reading can be corrected by removing the probe from the
layer and by pressing the plus or minus button.
10. Repeat step 7-10 until the result is correct.
Calibration Foils
This meter includes different foil sets for different ranges.
Please see the following table for reference.
Range (µm)CM25CM50CM100 CM200 CM500 CM1000
0-200••••
0-500••••
0-1000•••••
0-2000•••••
Battery Replacement
1. When the low battery symbol appears in the display, it is time to
replace the batteries.
2. Turn the meter off, open the battery compartment, and remove
the batteries.
3. Install 4 new AA batteries verifying you are following the
correct polarity.
For service or information on this or any other REED product, contact
REED Instruments at info@reedinstruments.com.
www
reedinstruments
www.GlobalTestSupply.com
com
6
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.