Raindance Raindrop Sense User Manual

Page 1
RainDrop® Sense
Operator’s Manual
Page 2
Patent, Trademarks, & Copyrights
Patent
www.raindancetechnologies.com/patents
Trademarks
Copyrights
© 2018 RainDance Technologies, Inc. All rights reserved. RainDance Technologies, Inc. owns the copyright to the RainDance software.
Reproduction Disclaimer
You may make a reasonable number of copies of this manual for internal use only and only for use in connection with the RainDrop Sense Instrument. All such copies shall contain the copy right info set forth on Page ii of this manual and this reproduction disclaimer. Except as explicitly provided by this section, neither this manual nor any part of it may be reproduced, photocopied, or electronically transmitted in any way without the advanced written permission of RainDance.
-
Contact
RainDance Technologies, Inc. 749 Middlesex Turnpike Billerica, MA 01821 Fax: (978) 528-6090 (support) Web: www.raindancetech.com/support Email: customercare@raindancetech.com
For Research Use Only. Not for use in diagnostic procedures.
Page 3
RainDrop Sense Operator’s Manual

Table of Contents

Chapter 1: Foreword 1-1
About this Manual ............................................................................................. 1-2
Intended Use..................................................................................................... 1-2
Data Entry and Collection ................................................................................. 1-2
Design Change Disclaimer................................................................................ 1-3
Reproduction Disclaimer ................................................................................... 1-3
RainDance Support........................................................................................... 1-3
Microsoft Windows Familiarity........................................................................... 1-4
Notes, Warnings, and Cautions ........................................................................ 1-4
Notes ........................................................................................................... 1-4
Warnings...................................................................................................... 1-4
Cautions....................................................................................................... 1-5
Biohazard Safety Warnings ......................................................................... 1-5
Laser Safety Warnings ................................................................................ 1-6
Safety Information............................................................................................. 1-7
Electrical Safety Warnings........................................................................... 1-7
Mechanical Safety ....................................................................................... 1-8
Gas Pressure Safety.................................................................................... 1-8
Product Labels .................................................................................................. 1-8
Safety Interlock ................................................................................................. 1-8
Packing and Transport ...................................................................................... 1-8
Chapter 2: Introduction 2-1
System Overview .............................................................................................. 2-2
RainDrop System Major Components............................................................... 2-2
RainDrop Sense Major Components ................................................................ 2-3
RainDrop Sense Layout.................................................................................... 2-4
RainDrop Sense Barcode Reader ............................................................... 2-5
RainDrop Sense Chip .................................................................................. 2-6
Handling the Sense Chip............................................................................. 2-6
RainDrop Sense Chip Compression Plate................................................... 2-7
Specifications.................................................................................................... 2-8
Chapter 3: Installation and Setup 3-1
Space Requirements......................................................................................... 3-2
Gas Connectivity.......................................................................................... 3-2
Connecting RainDrop Sense and Instrument Controller ................................... 3-3
Connecting Cables to the RainDrop Sense Instrument ............................... 3-3
Connecting Cables to the Instrument Controller.......................................... 3-4
Gas Pressure Quality ........................................................................................ 3-6
Starting Up the RainDrop Sense Instrument..................................................... 3-6
Relocating RainDrop ......................................................................................... 3-7
LCN 50-04344 Rev. D iii
Page 4
RainDrop Sense Operator’s Manual
Chapter 4: Launching the Instrument Control Software (ICS) 4-1
Introduction ....................................................................................................... 4-2
The Microsoft Windows 7 Login ........................................................................ 4-2
Starting the ICS Application .............................................................................. 4-2
ICS Login Levels.......................................................................................... 4-3
Changing the ICS Password........................................................................ 4-5
The ICS Account Types............................................................................... 4-6
ICS Initialization ........................................................................................... 4-7
Status Indicators ............................................................................................. 4-10
Using Tabs in the ICS ..................................................................................... 4-10
Navigating Through the Tabs .................................................................... 4-12
Canceling a Process.................................................................................. 4-12
Completing a Process................................................................................ 4-13
Using Scan Buttons ................................................................................... 4-13
Alarms, Log, and Status Tabs......................................................................... 4-14
Chapter 5: Performing a Sense Run 5-1
Introduction ....................................................................................................... 5-2
Gathering Supplies............................................................................................ 5-2
Starting a Sense Run........................................................................................ 5-3
Completing the ICS Setup................................................................................. 5-9
Checking the Status Indicators .................................................................... 5-9
Completing the Run Data Tab......................................................................... 5-11
Completing the Run Info Tab..................................................................... 5-14
Completing the Run Tabs Using the Import Function................................ 5-15
Using Email Notification............................................................................. 5-15
Performing the Run......................................................................................... 5-17
Canceling a Run.............................................................................................. 5-18
Handling Data ................................................................................................. 5-19
Exporting Data........................................................................................... 5-20
Moving Data Files...................................................................................... 5-21
Chapter 6: Managing the System 6-1
Introduction ....................................................................................................... 6-2
Managing Users................................................................................................ 6-2
Adding a New User...................................................................................... 6-3
Editing a User .............................................................................................. 6-5
Deleting a User............................................................................................ 6-7
Resetting Password or Unlocking an Account............................................. 6-7
Viewing Past Runs ............................................................................................ 6-8
Viewing Runs............................................................................................... 6-8
LCN 50-04344 Rev. D iv
Page 5
RainDrop Sense Operator’s Manual
Chapter 7: Support, Customer Maintenance, and Troubleshooting 7-1
RainDance Support........................................................................................... 7-2
RainDance Planned Maintenance..................................................................... 7-2
Customer Maintenance ..................................................................................... 7-2
Restarting the IC Workstation...................................................................... 7-2
Cleaning the RainDrop Sense ..................................................................... 7-2
Refilling the Oil Reservoir.................................................................................. 7-3
Service and Diagnostic Requirements .............................................................. 7-6
Powering Off the System .................................................................................. 7-6
Powering On or Restarting the System............................................................. 7-6
Restoring the System After a Failure ................................................................ 7-6
Support Agreement Information ........................................................................ 7-6
Moving the System............................................................................................ 7-7
Storing the System............................................................................................ 7-7
Reinstalling/Updating ICS Application............................................................... 7-7
Error Messages................................................................................................. 7-7
Troubleshooting ................................................................................................ 7-8
Chapter 8: Waste Management 8-1
Introduction and Disclaimer............................................................................... 8-2
Classification of Waste...................................................................................... 8-2
Anticipated Typical Waste Stream Composition ............................................... 8-3
Recommended Waste Stream Disposal ........................................................... 8-4
Chapter 9: Customer Care Information 9-1
General Care Information.................................................................................. 9-2
Consumables .................................................................................................... 9-2
Chapter 10: Regulatory Information 10-1
Sense Instrument Declaration of Conformity .................................................. 10-2
Chapter 11: Symbol Glossary 11-1
Appendix A: RainDrop
®
Consumables and Equipment A-1
LCN 50-04344 Rev. D v
Page 6
Chapter 1:
CHAPTER 1
Chapter 1:

Foreword

This chapter covers the following topics:
About this Manual page 1-1
Intended Use page 1-2
Data Entry and Collection page 1-2
Design Change Disclaimer page 1-2
Reproduction Disclaimer page 1-3
RainDance Support page 1-3
Microsoft Windows Familiarity page 1-3
Notes, Warnings, and Cautions page 1-3
Notes page 1-4
1
Warnings page 1-4
Cautions page 1-5
Biohazard Safety Warnings page 1-5
Laser Safety Warnings page 1-6
Safety Information page 1-7
Electrical Safety Warnings page 1-7
Mechanical Safety page 1-8
Gas Pressure Safety page 1-8
Product Labels page 1-8
Safety Interlock page 1-8
Packing and Transport page 1-8

About this Manual

This manual is designed to serve the operators of the RainDance Technologies®, Inc.
(hereafter called “RainDance”) RainDrop
instructions, product illustrations, screen graphics, troubleshooting, error messages, and
other relevant information are contained in this manual. It is the operator's responsibility to
ensure that all safety instructions in this manual are strictly applied.
®
Sense instrument. All material operating
Where there is any conflict between the RainDance Software End User L
included in the software, and the terms and conditions of sale, the order of precedence shall
LCN 50-04344 Rev. D 1-1
icense Agreement
Page 7
be as follows: (1) the RainDance Software End User License Agreement and (2) the terms
and conditions.

Intended Use

For Research Use Only. Not for use in Diagnostic Procedures.
RainDance Technologies RainDrop Sense is the detection instrument that enables
identification and counting of target molecules in each picoliter droplet following PCR
amplification. Together the RainDance Source and RainDrop Sense instruments make up the
RainDrop Digital PCR System delivering quantitative results.

Data Entry and Collection

RainDance requires that you do not enter health information identifiable to a particular
person, or other information that is subject to regulatory or contractual protection, into any
RainDance instrument. Such information is not protected from disclosure to service or other
personnel and, as noted below, may be included in information provided to RainDance for
purposes of servicing or evaluating the performance of your instrument.
RainDrop Sense Operator’s Manual
Further, by requesting service of your RainDance instrument, you give RainDance consent to
access and use data stored in your RainDance instrument. RainDance uses the data to
troubleshoot and service the instrument in response to your service requests. RainDance
also may aggregate data collected from your instrument in response to service requests with
similar data from other customers and may use the aggregated data to improve instrument
performance.
Your instrument from RainDance is enabled with an automatic email functionality that allows
RainDance to receive performance data from the instrument from time to time without your
knowledge. You may choose to disable that functionality. Disabling that functionality prevents
the instrument from sending such data to RainDance. The data collected with this automatic
email functionality consists of the number of samples processed, the success rate of the
samples processed, data specific to the performance of your instrument, and data that is
entered by your operators. RainDance collects this data solely so that it may better
understand and improve instrument performance. By allowing the functionality to remain
enabled, you give RainDance permission to use the automatic email functionality.

Design Change Disclaimer

Due to design changes and product improvements, information in this manual is
subject to change without notice. RainDance reserves the right to change the product design, including illustrations, screen shots, and diagrams, at any time without notice to anyone, which may subsequently affect the contents of this manual.
LCN 50-04344 Rev. D 1-2
Page 8
RainDance assumes no responsibility for any errors that may appear in this manual.
RainDance will make every commercially reasonable effort to ensure that this manual is up-to-date and corresponds with the shipped RainDrop Sense Instrument.
The depictions of the workstation screens in this manual are representative only.
Depending on the hardware and software versions of the system, minor differences may appear between the actual displays and those shown in this manual.

Reproduction Disclaimer

You may make a reasonable number of copies of this manual for internal use only and only
for use in connection with the RainDrop Sense Instrument. Except as provided by the
previous sentence, neither this manual nor any part of it may be reproduced, photocopied, or
electronically transmitted in any way without the advanced written permission of RainDance.

RainDance Support

If you have any difficulty running the RainDance Source instrument, contact your RainDance
Support representative.
solutions. The information given is general. Some applications may require additional
procedures or equipment modification. Contact RainDance Support via email:
support@raindancetech.com, Monday through Friday, excluding holidays.
Chapter 7 of this manual includes a list of problems and possible
RainDrop Sense Operator’s Manual

Microsoft Windows Familiarity

The supplied Instrument Control Software (ICS) is designed for the Microsoft®
Windows® Embedded 7 operating system and uses the standard Windows interface. This manual assumes that you are familiar with these standard techniques: clicking the mouse, highlighting an item, right-clicking, dragging, using the keyboard, interacting with dialog boxes, keyboard commands, selection techniques, and other Windows basics.
If you require assistance with these aspects of the software user interface, refer to the
Microsoft Windows documentation.

Notes, Warnings, and Cautions

The symbols described in this section are used throughout this document to draw your
attention to specific situations involving safety and proper use of the equipment. Symbols
found on the instruments are listed in
Chapter 11: “Symbol Glossary.”
LCN 50-04344 Rev. D 1-3
Page 9

Notes

Warning: A Warning indicates a potentially hazardous situation that, if not avoided,
could result in serious personal injury or death to you or others. Warnings are set
within boxed rules and indented.
RainDrop Sense Operator’s Manual
Notes are set apart from the body of the document with either the word Note or Important.
They are meant to draw your attention to reminders and information that is important to
successful operation of the RainDrop Sense Instrument.

Warnings

Note: A
Important: An
communicate greater weight to the issue at hand.
Warnings are represented by a red triangle with an exclamation point inside of it. Warnings
are reserved for situations that indicate the possibility of personal injury if the instructions are
not followed.
Note indicates a useful piece of information.
Important note is similar to a regular Note, except that it is meant to
LCN 50-04344 Rev. D 1-4
Page 10

Cautions

Caution: Do not operate the System in a manner other than specified in this manual. Doing so could damage the instrument or produce erroneous results.
Caution: RainDance service representatives are trained in the safe operation of the RainDrop Sense Instrument, including accessing those areas reserved for trained service representatives.
Warning: Follow your laboratory's procedures and regulations when handling
biohazardous materials.
Warning: RainDance suggests that users be professional and conscientious and
take the appropriate safety measures when preparing, handling, and disposing of
any biohazardous samples or waste.
Warning: When working with biohazardous samples, or waste, always follow
standard universal safety procedures (lab coats, safety glasses, gloves, mask,
etc.).
Cautions are represented by a blue triangle with an exclamation point inside of it. They
indicate that you must follow a particular procedure in order to correctly operate the RainDrop
Sense Instrument and to avoid any possible damage to the equipment or loss of
consumables.

Biohazard Safety Warnings

A biohazard warning indicates that the user must use precautions, for example, gloves, when
handling biological samples, to prevent any infections or hazards that can be caused by
them.
RainDrop Sense Operator’s Manual
In addition, users should take precautions in accordance with local, state, and national
requirements.
LCN 50-04344 Rev. D 1-5
Page 11

Laser Safety Warnings

Warning: A Class 3B laser is hazardous if the eye is exposed directly, but diffuse
reflections such as those from paper or other matte surfaces are not harmful. The
AEL for continuous lasers in the wavelength range from 315 nm to far infrared is
0.5 W. For pulsed lasers between 400 and 700 nm, the AEL is 30 mW. Other limits apply to other wavelengths and to ultrashort lasers. Protective eye wear is typically required where direct viewing of a class 3B laser beam may occur. Class­3B lasers must be equipped with a key switch and a safety interlock.
The RainDrop Sense, classified as a Class 2 System, contains a Class 3B laser and Class 2
barcode laser. Avoid direct exposure to the Class 3B laser radiation when the unit is open
and interlocks are defeated. Only a qualified service person should remove the cover or
attempt to service this device, due to possible eye injury.
RainDrop Sense Operator’s Manual
A laser warning indicates that the user must use precautions whe
to avoid injury from the lasers.
In addition, users should take precautions in accordance with local, state, and national
requirements.
n operating the instrument
LCN 50-04344 Rev. D 1-6
Figure 1-1: Laser Safety Warnings
Page 12

Safety Information

Warning: This equipment is operated with hazardous voltages that can shock,
burn, or cause death. To reduce the possibility of electrical shock, do not remove any fixed panels. Ensure that all service to the system is performed only by qualified RainDance service personnel.
Warning: Remove the wall plug before servicing the equipment. Never pull cords to remove the power cord from the outlet. Grasp the power cord plug and pull it from the outlet to disconnect it.
Warning: Do not operate the equipment with a damaged power cord.
Warning: Ensure that there is easy and adequate clearance to the power cord so
that it can be disconnected in the event of a problem.
Warning: Position the power cord so that it cannot be tripped over, walked on, rolled over, crimped, bent, pinched, or accidentally pulled from the wall outlet.
Warning: Connect the equipment only to a grounded outlet.
Warning: This instrument is not disconnected from the AC power source (mains)
as long as it is still connected to the wall outlet, even if the instrument is turned off. The primary means of disconnecting the instrument is to remove the plug from the
wall outlet. Do not locate the instrument such that access to this plug is difficult.

Electrical Safety Warnings

RainDrop Sense Operator’s Manual
LCN 50-04344 Rev. D 1-7
Page 13

Mechanical Safety

Warning: The RainDrop Sense instrument weighs approximately 65 lbs (29 kg). To
avoid injury and damage to the instruments, do not move it without assistance from RainDance Technologies.
Warning: The potential for serious injury to hands or fingers exists as a result of
rotating or clamping motion. Watch your hands and fingers when opening and
closing the cover. Keep your hands away from moving parts.
To move the instrument to another location, contact RainDance Support.

Gas Pressure Safety

Do not exceed a maximum input pressure of 120 psi at the instrument.
RainDrop Sense Operator’s Manual

Product Labels

Figure 1-2 shows the RainDrop Sense product label.
Figure 1-2: RainDrop Sense Product Label

Safety Interlock

The RainDrop Sense is equipped with an automatic safety interlock, which prevents it from operating unless its door is properly closed.

Packing and Transport

RainDance provides initial shipping and installation. If you are considering moving the RainDrop Sense instrument within your laboratory, refer to Chapter 3 for the space requirements. Do not move the instrument without assistance from
LCN 50-04344 Rev. D 1-8
RainDance Technologies.
Page 14
Chapter 1:
CHAPTER 2
Chapter 2:

Introduction

This chapter covers the following topics:
RainDrop® System Overview page 2-2
RainDrop System Major Components page 2-2
RainDrop Sense Major Components page 2-3
RainDrop Sense Layout page 2-4
RainDrop Carrier and Drive Oils page 2-4
RainDrop Sense Barcode Reader page 2-5
RainDrop Sense Chip page 2-6
2
Handling the Sense Chip page 2-6
RainDrop Sense Chip Compression Plate page 2-7
Specifications page 2-8
LCN 50-04344 Rev. D 2-1
Page 15
RainDrop Sense Operator’s Manual

RainDrop® System Overview

The RainDance RainDrop® System comprises two instruments: the RainDance Source and the RainDrop Sense.

RainDrop System Major Components

RainDance Source Instrument
RainDrop Sense Instrument Two Instrument Control (IC) Workstations (computers), one for each instrument
Two Instrument Control (IC) Monitors, one for each instrument
Instrument Control Cables and Peripherals
IC Workstation Power Cables
IC Monitor Power Cables
IC Keyboards IC Mouses
IC USB Cables
Instrument Control Software (ICS) RainDrop Analyst II™ Software
Figure 2-1: RainDrop System Major Components
LCN 50-04344 Rev. D 2-2
Page 16
RainDrop Sense Operator’s Manual
Instrument control workstation monitor RainDrop Sense instrument

RainDrop Sense Major Components

Figure 2-2 shows the major components of RainDrop Sense instrument.
Figure 2-2: RainDrop Sense Major Components
LCN 50-04344 Rev. D 2-3
Page 17

RainDrop Sense Layout

Chip alignment pins
Tube strip nest
Objective lens
Carrier oil reservoir fill port
Carrier oil low indicator light
Drive oil reservoir fill port
Drive oil low indicator light
Figure 2-3 shows the RainDrop Sense layout and oil reservoir fill ports.
RainDrop Sense Operator’s Manual
Figure 2-3: RainDrop Sense Layout
RainDrop Carrier and Drive Oils
Aqueous samples are encapsulated within each droplet on the Source instrument, surrounded by an immiscible carrier oil. The droplets are stabilized with surfactants allowing for robust manipulation both on and off the chip. Precise control of carrier and drive oil flow rates on the Sense instrument allows for control of droplet spacing. Use only RainDance­supplied oils to avoid damage to the RainDrop System. The carrier oil and drive oil ports can be filled with the oil provided in the Carrier Oil Syringe.
LCN 50-04344 Rev. D 2-4
Page 18

RainDrop Sense Barcode Reader

The barcode reader is used to scan barcodes that identify the various components used in the RainDrop Sense (see Figure 2-4). The following lists the components to be scanned:
RainDrop Sense Chip(s)
RainDrop Carrier Oil
Tube Strips (May have a user-generated barcode. Scanning tube strips is optional.)
Items bearing a user-generated barcode for sample tracking (within the limits of
supported barcode formats)
RainDrop Sense Operator’s Manual
Figure 2-4: RainDrop Sense Barcode Reader
Scanning Barcodes
Scan items as follows:
1. Click Sc
2. Hold
the barcode, it automatically fills in the barcode information. The ICS accepts a barcode in the 2D Data Matrix format, which allows up to 96 characters of text. You may hear a subtle beep tone from the instrument when the barcode is accepted. If the scanner does not read the barcode, the ICS software displays a message and allows you to enter the barcode manually.
LCN 50-04344 Rev. D 2-5
an in the ICS software. The red scanner light pulses for 10 seconds.
the barcode on the item in front of the barcode reader. When the scanner reads
Page 19

RainDrop Sense Chip

Microfluidic chip
Droplet detection region of interest (ROI)
Fluid lanes
Chip alignment hole
Chip alignment holeWaste collection area
Oil input gaskets
The RainDrop Sense Chip is placed into the instrument on top of the PCR tube strip. As part of the Sense process, the chip punctures the tube strip caps and transfers the emulsified sample through its lanes where the Sense instrument identifies and counts the droplets.
RainDrop Sense Operator’s Manual

Handling the Sense Chip

Figure 2-5: RainDrop Sense Chip
Handle the Sense Chip as follows:
Grasp the chip by its sides.
Do not touch the clear microfluidic chip and its droplet detection region of interest.
Save the bag the chip comes in for later disposal of the chip.
LCN 50-04344 Rev. D 2-6
Page 20

RainDrop Sense Chip Compression Plate

The RainDrop Sense Compression Plate is needed for smooth operation of the RainDrop Sense Instrument. Position the plate over the Sense chip prior to start of the run. When the plate is properly positioned, its RainDance logo should be facing you, the plate should be flat, and it should not be movable left to right or front to back.
RainDrop Sense Operator’s Manual
Figure 2-6: RainDrop Sense Chip Compression Plate
LCN 50-04344 Rev. D 2-7
Page 21

Specifications

Table 2-1 shows the specifications for the RainDrop Sense instrument. Table 2-2 shows the
specifications for the Instrument Control (IC) Workstation.
Table 2-1: RainDrop Sense Hardware Specifications
Requirement or Spec
ification Description
RainDrop Sense Operator’s Manual
Width
Depth 17 inches (43.2 cm)
Height
Weight ~65 lbs (~29 kg)
Power Source (Electrical) Voltage: 100 to 240 VAC, 35 W max
Clearance 2 inches (5 cm) in rear
Fuses 2 fuses, 2 Amperes Medium Acting
Cable Connections USB: 1 USB 3.0 and 1 USB 2.0
Required compressed gas input pressure range at the instrument
Capacity 8 lanes per chip
Maximum Sample Throughput per 8 hour shift
11 inches (27.9 cm)
15 inches (38.1 cm)
1 standard grounded outlet, within 6 ft (2 m) of the instrument
90-120 PSI (0.62 MPa – 0.82 MPa)
24 samples per shift in Standard Mode 56 samples per shift in Fast Mode
Detection Time ~8 lanes in 4 hours for Standard Mode (for 25 µL
sample) ~8 lanes in < 2 hours for Fast Mode (for 25 µL sample)
Droplet Identification Precise control of relative pressure and oil flow rate
creates ideal spacing of the thermal cycled emulsion to allow for proper droplet identification and analysis.
Droplet Processing Droplets are processed on a chip that has no moving
parts or valves. Samples have minimal contact with either walls or air. In addition, one-time-only chip use prevents sample cross-contamination.
LCN 50-04344 Rev. D 2-8
Page 22
RainDrop Sense Operator’s Manual
Table 2-2: Instrument Control (IC) Workstation Specifications
Requirement or Specification
Instrument Control (IC) Workstation
Power Voltage: 100-240 VAC
Dimensions Width: 1.36 inches (3.45 cm)
Weight
Clearance 2 inches (5 cm) in rear
Network Connection
Description
Intel Computer: 19 V, 65 W Standard grounded outlet within 6 ft (2 m) of the instrument
Depth: 4.41 inches (11.20 cm) Height: 4.59 inches (11.66 cm)
1.4 lb (520 g)
TCP/IP connectivity (i.e., Internet) using a standard Ethernet connector. Instruments must be placed near an active Ethernet port so the computers can be networked. The Instrument Control Software has the ability to email notifications to customers if it is networked. In addition, RainDance has software that allows us, with customer permission, to access the computers to look at log files to assist in troubleshooting and to assess instrument performance. Without Internet access, support is compromised.
Instrument Control (IC) Monitor
Power Voltage: 100 to 240 VAC
View Sonic Monitor: 22 W max, 20 W typical grounded outlet within 6 ft (2 m) of the instrument
Dimensions Width: 16.4 inches (41.62 cm)
Depth: 7.8 inches (19.91 cm) Height: 15.8 inches (40.22 cm)
Weight
Clearance 0.5 inches (1.2 cm) in all directions
6.6 lb (3.0 kg)
LCN 50-04344 Rev. D 2-9
Page 23
Table 2-3: Operating Environmental Requirements
RainDrop Sense Operator’s Manual
Requirement or Specification
Laboratory Ambient Temperature Range
Ambient Temperature F
luctuation
Relative Humidity 5-85% RH, non-condensing
Atmospheric Pressure
Safety Compliance IEC 61010-1:2010, the international safety standard for
Data Backup RainDance suggests using a standard backup
Description
15-30°C
<0.3°C per minute
101-81 kPa (1010-810 mbar) or sea level to 1800 m
laboratory equipment, including USA (UL), Canadian (CSA), and European (CENELEC/EN) differences
dure for any computer system.
proce
LCN 50-04344 Rev. D 2-10
Page 24
Chapter 1:
CHAPTER 3
Chapter 3:

Installation and Setup

This chapter covers the following topics:
Space Requirements page 3-2
Connecting RainDrop Sense and Instrument Controller page 3-3
Connecting Cables to the RainDrop Sense Instrument page 3-3
Connecting Cables to the Instrument Controller page 3-4
Gas Pressure Quality page 3-7
Starting Up the RainDrop Sense Instrument page 3-8
Relocating RainDrop page 3-9
3
Important: Only train
RainDrop
®
Sense instrument from its packaging and install it.
ed and certified RainDance Service Representatives may remove the
LCN 50-04344 Rev. D 3-1
Page 25

Space Requirements

Caution: The RainDrop Sense instrument is sensitive to vibration and should be placed on a bench top that is free from vibration.
The RainDrop® Sense instrument is designed to be placed on a standard laboratory bench. The Instrument Controller (IC) Workstation can be located on either the left or right side of the instrument.
Table 3-1: RainDrop Sense Dimensions (Including the IC Workstation)
Minimum Height (door open) 28 inches (71.1 cm)
RainDrop Sense Operator’s Manual
Minimum Depth
Bench Loading Weight 73 lb (26.4 kg)
Minimum Width 29 inches (74 cm)
Note: Refer to

Gas Connectivity

Gas source should be located in proximity to the instrument because the gas supply connection is located on the back of the instrument. Use 4mm OD polyurethane tubing with an in-line shutoff valve. See Appendix A for information on suppliers.
19 inches (48.3 cm)
the Pre-Site Guide for any additional installation details.
LCN 50-04344 Rev. D 3-2
Page 26
RainDrop Sense Operator’s Manual
Power switch
AC power cable
Gas pressure
USB connections

Connecting RainDrop Sense and Instrument Controller

Before powering on the RainDrop Sense instrument and its IC Workstation, make sure all cables are properly connected as described in “Connecting Cables to the RainDrop Sense
Instrument” on page 3-3 and “Connecting Cables to the Instrument Controller” on page 3-4.

Connecting Cables to the RainDrop Sense Instrument

The RainDrop Sense instrument has three cable connections (AC Power Cable and two USB) and a gas pressure connection as shown in Figure 3-1.
Figure 3-1: Cable Connections on RainDrop Sense
LCN 50-04344 Rev. D 3-3
Page 27
RainDrop Sense Operator’s Manual
To plug cables into the back of the Sense instrument:
1. Plug one end of the AC Power Cable into the Power Cable Connection and the other
end into a standard grounded wall outlet.
Important: The Detachable Main AC Power Cable for North America or other countries with 120 VAC 50/60 Hz supply voltage provided for the RainDrop Sense instrument is a 3-wire cord, 18 AWG, 10 A, 125 VAC. Never substitute a Detachable Main AC Power Cable with a lower current or voltage rating.
2. Plug the two USB Cables into the USB Connections on the Sense instrument as
shown in
Figure 3-1.

Connecting Cables to the Instrument Controller

The RainDrop IC Workstation has USB connections used for:
The mouse The keyboard
The RainDrop Sense instrument
Both the Source and Sense instruments connect to their IC Workstations via USB connections.
Figure 3-2 illustrates the USB cable connections on the IC Workstation.
Note: It is important that all the USB cables on the instruments are connected to USB ports on the workstations. Failure to connect the USB cable correctly can result in a loss of communication with the RainDrop instrument. In addition, the USB connections on your workstation may differ from those shown in
Figure 3-2.
LCN 50-04344 Rev. D 3-4
Page 28
RainDrop Sense Operator’s Manual
USB port that connects to the left output from the Sense
USB port that connects to the right output from the Sense
HDMI port that connects to the monitor
The left USB/output connects to the USB port furthest from the gas inlet and the right USB/output
Figure 3-2: Cable Connections to the IC Workstation
connects to the other available USB port.
LCN 50-04344 Rev. D 3-5
Page 29
RainDrop Sense Operator’s Manual
USB receiver
LCN 50-04344 Rev. D 3-6
Figure 3-3: USB Receiver
1. With the IC Workstation turned off, plug the USB receiver for the keyboard and mouse
into the IC Workstation.
2. Plu
3. Plu
g the right USB output from the Sense instrument into the right USB in Figure 3-2. g the left USB output from the Sense instrument into left USB in Figure 3-2.
Page 30

Gas Pressure Quality

You may use compressed argon or nitrogen. Ensure that the compressed gas system you use meets the following standards.
RainDrop Sense Operator’s Manual
The input pressure must be regulated to a value, at the system, within the 120 psi.
In addition, the quality must meet the ISO/DIS8573 compressed gas qu Class 2 for instrument quality gas, as follows:
Table 3-2:
Oil (ppm) 0.1
Water (dew point)
Dirt (µ) 1.0
GasPressure Quality
-40°F
ality standard,
range of 90-
LCN 50-04344 Rev. D 3-7
Page 31
RainDrop Sense Operator’s Manual

Starting Up the RainDrop Sense Instrument

Use the following startup sequence for the RainDrop Sense instrument:
1. With the
have been connected properly as described in “Connecting Cables to the RainDrop
Sense Instrument” on page 3-3 and “Connecting Cables to the Instrument Controller” on page 3-4.
2. T
urn on the IC Workstation (on the top of the workstation).
3. Lo
g on to the Windows 7 operating system using the RDT account.
Note: Logg
User name = RDT
No password
4. T
urn on the RainDrop Sense instrument (located on the back of the instrument).
5. W
ait 1 minute for the RainDrop Sense instrument to perform internal initialization. As the instrument initializes, the lights on the panel illuminate to indicate its status (see
Figure 3-4).
RainDrop Sense and IC Workstation turned off, ensure that all the cables
ing on using any other account results in ICS failures.
Figure 3-4: RainDance LED Lights
6. Launch the Instrument Control Software (ICS) application as described in Chapter 4.
Table 3-3 lists the LED lights that indicate the status of the instrument.
Table 3-3: LED Lights Within the RainDance Logo
LED Name and Illumination Pattern Meaning
LCN 50-04344 Rev. D 3-8
Page 32
RainDrop Sense Operator’s Manual
Caution: The RainDrop Sense instrument weighs approximately 65 lbs (29 kg). To avoid injuries and damage to the instrument, do not move it without assistance from RainDance Technologies. If you want to move your instrument either within your laboratory or to another location, contact RainDance Technologies.
Table 3-3: LED Lights Within the RainDance Logo
Main Power LED (Solid) Instrument is powered on.
All LEDs except Main Power are rotating
and off.
on
All LEDs are flashing simultaneously. Instrument has detected an error.
All LEDs are on.
All LEDs except Main Power are off. Instrument is disconnected from the ICS.

Relocating RainDrop

Instrument is operating normally.
Instrument is connected to the ICS.
LCN 50-04344 Rev. D 3-9
Page 33
Chapter 1:
CHAPTER 4
Chapter 4:

Launching the Instrument Control Software (ICS)

This chapter covers the following topics:
Introduction page 4-2
The Microsoft Windows 7 Login page 4-2
Starting the ICS Application page 4-2
ICS Login Levels page 4-3
Changing the ICS Password page 4-5
4
The ICS Account Types page 4-6
ICS Initialization page 4-7
Status Indicators page 4-10
Using Tabs in the ICS page 4-11
Navigating Through the Tabs page 4-12
Canceling a Process page 4-12
Completing a Process page 4-13
Using Scan Buttons page 4-13
Alarms, Log, and Status Tabs page 4-14
LCN 50-04344 Rev. D 4-1
Page 34

Introduction

The Instrument Control Software (ICS) application controls the operation of the RainDrop® Sense.

The Microsoft Windows 7 Login

The Instrument Control Workstation is configured with a single Microsoft Windows 7 Operating System login. Do not edit this account or add additional accounts. Failure to comply will result in ICS failures. The default login is an administrative account. The system differentiates between upper and lower case characters.
RainDrop Sense Operator’s Manual
Note: Do not
use the RainDance Support Windows Logon. This will result in ICS failure.

Starting the ICS Application

The startup sequence works as follows:
1. Power on
Power on the IC Controller and log in to Windows. Power on the RainDrop Sense. Wait one minute.
2. Click the ICS icon on
Note: You can also start the ICS application by selecting Windows 7 Start Program Menu,
and then selecting Start > All Programs > RainDance Technologies > Launch ICS.exe.
each component of the RainDrop Sense in the following sequence:
your desktop:
Note: The designated application and software uses.
LCN 50-04344 Rev. D 4-2
IC Workstation must be dedicated to ICS Operation. Use it only for RainDance
Page 35
RainDrop Sense Operator’s Manual
The Login window opens (see Figure 4-1).
Figure 4-1: Login Window
3. Enter your Login information in the appropriate fields. Enter the Username followed by the Password and then click Login. See ICS Login Levels for more information.
Note: Microsoft
application, both usernames and passwords are case sensitive.

ICS Login Levels

The ICS application supports two levels of access to the ICS program: User and Administrator. The ICS application is configured with two default login IDs: one for User and one for Administrator. RainDance recommends that you change the default password on each of these accounts after initial use.
Table 4-1: ICS Default User and Administrator Login Information
Requirement or Specification Description
User Login Username User
User Login Password
Administrator Login User Name Admin
Administrator Login Password admin
Windows 7 usernames are not case sensitive, but passwords are. In the ICS
user
LCN 50-04344 Rev. D 4-3
Page 36
Caution: After three failed attempts to login with the wrong password, the ICS application locks the account. You will not be able to login to your account. If this happens, you must contact someone in your facility with an Administrator account who can either unlock or reset your password. For details, see “Resetting Password
or Unlocking an Account” on page 6-7.
Figure 4-2 shows the message for a failed login.
RainDrop Sense Operator’s Manual
Figure 4-2: Invalid Username Message
LCN 50-04344 Rev. D 4-4
Page 37

Changing the ICS Password

This section describes how to change an ICS password. RainDance recommends that you change the default User and Administrator passwords after initial use.
RainDrop Sense Operator’s Manual
1. Select Chan
(Figure 4-3).
2. Enter the old password in the Old Password field, the new password in the New Password field, and the new password again in the Verify Password field. Click Save
when you are done to save the new password. Click Cancel to dismiss the dialog without saving your changes.
ge Password from the Home menu. The Change Password dialog opens
Figure 4-3: Change Password Window
LCN 50-04344 Rev. D 4-5
Page 38

The ICS Account Types

Your account type defines the features you have access to (User and Administrator) within the ICS application.
Table 4-2: ICS Account Type
RainDrop Sense Operator’s Manual
Feature User
Change Passwords X X
Export Instrument Files X
View Runs X X
Maintenance Actions
Add New Users X
Edit Users
Reset Passwords X
Unlock Passwords
Start Runs X X
Stop Runs X X
The pull-down menu selections differ depending on your access righ different for a User and an Administrator account.
Administrator
X
X
X
X
ts; they are slightly
LCN 50-04344 Rev. D 4-6
Page 39

ICS Initialization

When you login to the instrument, the Initialization screen appears (see Figure 4-4).
RainDrop Sense Operator’s Manual
Figure 4-4: Initialization Screen
From the Initialization screen, click Start Initialization to initialize the system. A progress bar shows the progress of the initialization.
Figure 4-5: Initialization Progress Bar
LCN 50-04344 Rev. D 4-7
Page 40
RainDrop Sense Operator’s Manual
If initialization fails, the following message appears:
Figure 4-6: Initialization Failure Message
Click OK and contact RainDance Support. See “RainDance Support” on page 1-3.
This error is often the result of starting ICS right after you turn on recommends the following steps:
1. Exit the
2. T
urn the instrument off.
3. T
urn the instrument back on.
4. W
ait 1 minute.
5. Start t
ICS.
he ICS.
If the message appears again, restart the ICS.
the instrument. RainDance
LCN 50-04344 Rev. D 4-8
Page 41
RainDrop Sense Operator’s Manual
If initialization is successful, the ICS home page is displayed (see Figure 4-7).
Note: All
menu items are inactive until the initialization is finished.
Figure 4-7: Home Page
From the home page, you can Setup a Run (see Chapter )
You can perform additional tasks by ma
View Runs (see Chapter 6)
Maintenance Actions (see Chapter 7)
Add New Users (Administrator only) (see Chapter 6) Edit Users (Administrator only) (see Chapter 6)
king selections from the drop down menus:
LCN 50-04344 Rev. D 4-9
Page 42

Status Indicators

A common set of status indicators appears throughout the system indicating the status of an item or process being performed. In general, the green check mark ( everything is working as expected and you can proceed. The orange exclamation point (!) indicator is a warning that means that you can continue, but that some optional fields are blank. The red X indicator means there is a problem that needs to be resolved before you can continue.
RainDrop Sense Operator’s Manual
?
) indicator means that
Table 4-3 shows the status indicators a
Table 4-3: Status Icons
Icon
Meaning
Ready – If all icons are green check marks, the system is ready to start a
run.
Warning (Run Will Proceed) – If there is one or more orange exclamation point icons on the tabs, the run can start. These represent optional information.
Error (Run Will Not Proceed) – If there is one or more red X icons on the tabs, there is an issue that may prevent the run from being performed.
nd describes their meanings.
LCN 50-04344 Rev. D 4-10
Page 43

Using Tabs in the ICS

There are two tabs in the ICS that appear after you click Setup a Run: Run Data and Run Info. To perform a run, you will need to complete some of the fields on each tab.
RainDrop Sense Operator’s Manual
To ensure that all of the steps are completed correctly, RainDrop the right side of the tabs to indicate the state of each of the required items. As you complete each step in the process the status icon for that step is updated. This section provides general information on what appears on the tabs and how to complete these processes.
Details on how to complete each of these pro descriptions assume you are familiar with how to use the tabs described h
Figure 4-8 shows an example of the Sense interface.
process are shown on the left side of the screen:
Run Data
Run Info
cesses is provided in Chapter . Those
The tabs required to complete the
provides a System Status on
ere.
Figure 4-8: Sense Tabs
Only one tab is visible at a time. The selected (or active) tab has a white background and the inactive tab has a transparent background (Run Data is the active tab in Figure 4-8).
LCN 50-04344 Rev. D 4-11
Page 44
RainDrop Sense Operator’s Manual
To prepare for a run on the Sense instrument, you must complete the following fields in the Sense ICS:
Run Data Tab:
Run Name
Sense Chip barcode lot and serial number At least one selected lane (indicated by a green check mark)
In addition, before you can proceed with the run on the Sense instrument, you
must have clicked the Insert PCR Tube button to indicate that you have inserted the closed tube strip into the instrument.
Select Standard or Fast Mode detection.
Run Info Tab:
Carrier Oil field and Drive Oil field and only if the instrument is out of one of those
oils
As you complete each tab, the software updates the System Status fo green check mark is displayed, the item is ready to go. If an orange exclamation point is displayed, it means that optional fields are blank. If a red X is displayed, there are errors or missing data required for the run (see Table 4-3).
For a description of the meaning of the Status Indicators used on system, see “Status Indicators” on page 4-10.

Navigating Through the Tabs

You can switch back and forth between the two tabs at will. Click the tab name to move to a tab.

Canceling a Process

The Run Data tab has a red Cancel button in the lower left corner (Figure 4-8). If you click Cancel while a
run is active, the application asks you to confirm that you want to cancel.
r various items. If the
the tabs and throughout the
Figure 4-9: Cancel Confirmation
When you click Yes, the information you entered when you started the process is deleted, the run is canceled and you are returned to the RainDrop main screen.
LCN 50-04344 Rev. D 4-12
Page 45

Completing a Process

When you have completed each step to prepare for a run, review the System Status indicators. Ideally, all the status indicators will be green check marks, although you can still continue with orange exclamation points since they indicate only that optional fields are blank. The process cannot be started if there is even one red X (error indications) on any of the tabs.
RainDrop Sense Operator’s Manual
When you are satisfied that the steps for the process are complete bottom right side of the Run Data tab. This starts the process on the RainDrop Sense instrument.

Using Scan Buttons

The ICS allows you to scan barcodes to enter information into various text fields. There are two types of Scan buttons available in the system as shown in Table 4-4.
Table 4-4: Scan Buttons
Button Description Use
, click Start Run on the
Figure 4-10: Start Run Button
Manual Scan Button The instrument requires that you
manually enter a barcode.
Scan Button The system allows you to scan any
type of barcode.
Clicking either type of scan butt to scan the barcode.
If you scan the wrong type of barcode, an error message appears. ICS will not a invalid barcode.
LCN 50-04344 Rev. D 4-13
on activates the barcode reader on the system and allows you
ccept an
Page 46

Alarms, Log, and Status Tabs

The Alarms, Log, and Status tabs are displayed on the bottom of the screen (see Figure 4-
11). Each of these tabs is described below.
Figure 4-11: Example of Screen Showing Alarms, Log, and Status Tab at Bottom
The Alarms tab is displayed by default. Click Log or Status to display that tab.
Alarms Tab – This tab lists instrument errors that exist in the system. If an error
appears, the system is in a warning or error state. Different errors occur during the setup of the instrument. If an error does not clear when you close the door and click Start Run, check “Troubleshooting” on page 7-8. If the issue is still not resolved, contact supp
Log Tab – This tab is used for service purposes. It is used to check on the system
automation.
Status Tab – This tab is used for service purposes. It is used to determine the status of
the instrument at different points during operation.
ort@raindancetech.com for assistance.
RainDrop Sense Operator’s Manual
LCN 50-04344 Rev. D 4-14
Page 47
Chapter 1:
CHAPTER 5
Chapter 5:

Performing a Sense Run

This chapter covers the following topics:
Introduction page 5-2
Gathering Supplies page 5-2
Starting a Sense Run page 5-3
Completing the ICS Setup page 5-10
Checking the Status Indicators page 5-10
Completing the Run Data Tab page 5-12
Completing the Run Info Tab page 5-15
5
Completing the Run Tabs Using the Import Function page 5-16
Using Email Notification page 5-16
Performing the Run page 5-18
Canceling a Run page 5-20
Handling Data page 5-21
Exporting Data page 5-22
Moving Data Files page 5-23
LCN 50-04344 Rev. D 5-1
Page 48

Introduction

Perform the operations described in this chapter on the RainDrop® Sense instrument. This
procedure requires a number of consumables (available from RainDance Technologies other manufacturers).
Note: For the list of required consumables, see “RainDrop® Consumables and Equipment”
on page A-1.
This chapter describes how to start a run on the RainDrop Sense instrument. Some of the steps involve preparing the physical components of the system and some of them take place within the RainDrop ICS.
For detailed information the ICS software, see Chapter 4 and Chapter 6.
Note: Extraneous clusters have been noted in 50ul samples run in the FAST mode on the Sense instrument. For the highest quality data we recommend limiting sample volumes to 25ul if the samples are to be run in the FAST mode on the Sense instrument.
RainDrop Sense Operator’s Manual
®
and

Gathering Supplies

To start a run on the RainDrop Sense instrument, have the following equipment and supplies available:
Gloves, nitrile or latex Sense Chip, P/N 30-06086
Sense Chip Compression Plate, P/N 30-06423
Drive Oil, shipped in syringes, P/N 30-07117
PCR tube strip containing emulsion generated on the RainDance Source instrument
and following thermal cycling.
For Standard Mode detection, standard PCR Tube Strip Caps (P/N 40-06087)
For Fast Mode detection, High Speed Caps (P/N 40-08286)
LCN 50-04344 Rev. D 5-2
Page 49

Starting a Sense Run

This procedure assumes you have prepared your sample and that you have assembled all the necessary equipment and supplies. It also assumes that you have completed thermal cycling the sample. It shows you how to start a run on the RainDrop Sense instrument.
unch the ICS software for the Sense instrument and log in.
1. La
RainDrop Sense Operator’s Manual
Figure 5-1: Login Screen
2. Click Start Initialization.
Figure 5-2: Initialization Screen
When you click Start Initialization, the system automatically prepares the instrument for normal operation. It sets valves and pressures and turns the laser in the barcode reader off. In addition, it homes the motion control subsystem. Initialization takes approximately four minutes.
LCN 50-04344 Rev. D 5-3
Page 50
RainDrop Sense Operator’s Manual
Note: If initialization fails, the ICS returns to the Initialization screen after about fifteen minutes. Contact RainDance Support for assistance. See “RainDance Support”
on page 1-3.
3. Click Se
tup a Run.
Figure 5-3: Instrument Status and Setup a Run Screen
LCN 50-04344 Rev. D 5-4
Page 51
4. Select the Run Info tab.
RainDrop Sense Operator’s Manual
Figure 5-4: Run Info Tab
5. On the Run Info tab, click Open Door. The door on the Sense instrument unlocks
allowing you to lift it to the upright, open position.
6. If ru
nning in Fast Mode, use High Speed Caps. Following thermal cycling, carefully remove the domed caps (Axygen) and replace with High Speed Caps. Rock the caps back and forth to seat properly, then press into place. Ensure the cap is secured and the strip is tightly covered.
A properly covered PCR tube strip has:
A flat cap all the way across
Its tab oriented to the left (over number 1 on the tube strip)
An improperly covered PCR tube strip has:
A bump or buckle
A loosely fitting cap
Its tab to right (over number 8 on the tube strip)
LCN 50-04344 Rev. D 5-5
Page 52
A bump in the cap
7. Lift the door to the fully open position.
.
RainDrop Sense Operator’s Manual
Figure 5-5: Improperly Covered PCR Tube Strip
LCN 50-04344 Rev. D 5-6
Page 53
RainDrop Sense Operator’s Manual
Figure 5-6: Sense Door in Fully Open Position
LCN 50-04344 Rev. D 5-7
Page 54
RainDrop Sense Operator’s Manual
8. Insert a tube strip into the tube strip nest (see Figure 2-3). Orient the tabbed end to the
left.
Figure 5-7: Insert a Tube Strip
9. Remove a Sense Chip from its packaging; for more information on the chip, see see
“RainDrop Sense Chip” on page 2-6. Handle the Sense Chip as follows:
With gloved hands, grasp the chip by its sides.
Do not touch the clear microfluidic chip and its droplet detection region of
interest.
Save the bag the chip comes in for later disposal of the chip.
Do not touch the objective lens.
10. Scan the
chip by clicking Scan in the ICS software. The red scanner light pulses for
10 seconds. Hold the Sense Chip in front of the barcode reader. If the scanner reads the barcode, it automatically fills in the barcode information. The ICS accepts a barcode in the 2D Data Matrix format, which allows up to 96 characters of text. You may hear a subtle beep tone from the Sense instrument.
If the scanner does not read the barcode, the ICS software displays a
message and
allows you to enter the barcode manually. To do so, click the Manual button and enter the barcode numbers from the label of the chip. In manual format, the ICS accepts 99 characters of text.
11. In
sert the chip into the Sense instrument. Orient the Sense Chip over the alignment pins and ensure that the pins are coming through the alignment holes on the chip. Do not insert it at an angle. The instrument is designed so that the Sense Chip punctures the caps of the tube strip. Do not apply pressure to the chip after inserting it. It is important that the instrument is able to make a seal and then puncture the tube caps. The Sense instrument will not allow you to proceed until the chip is oriented properly.
LCN 50-04344 Rev. D 5-8
Page 55
RainDrop Sense Operator’s Manual
Figure 5-8: Inserting a Sense Chip
12. Locate the Chip Compression Plate; for more information on the plate, see see
“RainDrop Sense Chip Compression Plate” on page 2-7.
13. Place t
When the plate is properly positioned, its RainDance logo should be
he Chip Compression Plate over the Sense chip as shown in Figure 5-9.
facing you, the
plate should be flat, and it should not be movable left to right or front to back.
Figure 5-9: Placing the Chip Compression Plate
After you insert the Chip Compression Plate, you are done setting up the physical
components.
14. Clo
se the door on the Sense instrument.
LCN 50-04344 Rev. D 5-9
Page 56

Completing the ICS Setup

Before starting a run on the Sense instrument, you must complete a number of setup steps. The steps ensure that the instrument is ready for operation and provide data about the run. They include:
Check the status of various components.
Complete the Run Data.
Complete the Run Info.

Checking the Status Indicators

Status indicators tell you about the readiness of the instrument.
RainDrop Sense Operator’s Manual
Figure 5-10: System Statuses
Observe the status indicators in the software.
Table 5-1: System Status Indicators
Indicator
Pressure Ready
Carrier Oil
Drive Oil Indicates the status of the drive oil reservoir. When there is sufficient oil to
Meaning
Indicates the status of the instrument gas pressure. When the pressure is sufficient, the ready indicator is displayed.
Indicates the status of the carrier oil reservoir. When there is sufficient oil to complete a run, the ready indicator is displayed. A valid carrier oil lot number is required for the green check mark to be displayed.
complete a run, the ready indicator is displayed. A valid drive oil lot number is required for the green check mark to be displayed.
LCN 50-04344 Rev. D 5-10
Page 57
Table 5-1: System Status Indicators
RainDrop Sense Operator’s Manual
Chip Inserted
Indicates the presence of the Sense Chip. When the Sense Chip is present and inserted properly, the ready indicator is displayed.
Note: You must scan the chip prior to use; otherwise, the instrument will not detect it. If you have scanned the chip before inserting it and it is still not detected, wiggle the chip and try again.
Door Closed Indicates the status of the instrument door. When the door is properly
closed, the ready indicator is displayed.
System Ready
Indicates the overall status of the instrument. When all the requirements of the instrument are present and ready for a run, the ready indicator is displayed.
In addition to satisfying the Status Indicators, in order to start a run you must also have:
Scanned or input the barcode number from a Sense Chip Clicked Insert Tube Strip to affirm that you have inserted the tube strip
LCN 50-04344 Rev. D 5-11
Page 58

Completing the Run Data Tab

The Run Data tab opens after you click Start Run on the startup screen and contains the details about the contents of the run. You can enter this information manually or you can import it from a file you saved on the Source instrument.
RainDrop Sense Operator’s Manual
Figure 5-11: Run Data Tab
Only the following fields on the Run Data tab are required for the run to proceed:
Run Name Sense Chip barcode lot and serial number
At least one selected lane (indicated by a green check mark)
Your lab procedures determine how much of the optional information specific situation.
In addition, before you can proceed with the run, you must have button to indicate that you have inserted the closed tube strip into the instrument.
is necessary for your
clicked the Insert PCR Tube
LCN 50-04344 Rev. D 5-12
Page 59
Table 5-2: Run Data Fields
RainDrop Sense Operator’s Manual
Field Name
Run Name The name of the run. Run name is required and can contain up to 20
Lanes A - H
Lane Description
Sense Chip Lot & Serial Number
Meaning
characters. This is required.
The identifiers for each tube (or lane) on the chip. These fields are not required. The lane letter identifiers are hard-coded and cannot be edited.
The text description that you enter. This is not required.
The identifiers for each Sense Chip. completed when you scan the chip and is required.
This information is automatically
Run Filenames
When the ICS saves data at the end of a run, it uses the information you enter in the fields in the above table to name the file. It creates one file per lane in a folder named for the Run ID, as follows:
The filename for one lane of a run looks like this:
<Run ID>-<Lane Letter>-<Lane Description>.fcs
For example: 1302181535-A-ExampleDescription.fcs
The Run ID is the numeric run number from the upper right corner of appears on both the Run Data tab and the Run Info tab. See “Handling Data” on page 5-21 for more information.
the ICS screen. It
Entering the Run Data
Enter the run data as follows:
1. T
ype the name of the run into the Run Name field.
2. All lan
3. If n
es are selected by default. Deselect any lanes you are not using. At least one
lane must be selected in order for the run to proceed.
ecessary, enter lane information in one of the following ways:
Type the name for the lane manually. When you enter the name manually, the field
limits entry to 99 characters.
LCN 50-04344 Rev. D 5-13
Page 60
RainDrop Sense Operator’s Manual
Scan the barcode from a printout. To do so, place the printout in front of the
barcode reader and click Scan next to the correct lane. When you enter the name using the scanner, the field limits entry to 96 characters.
Import the lane information as described in see “Completing the Run Tabs Using
the Import Function” on page 5-16.
4. Select the number of droplets you want from the Droplet drop-down.
5. Click Insert PCR Tube to confirm that you have inserted the tube strip into the
instrument.
Note: The Sense Chip field should already display the information you scanned earlier in see
“Starting a Sense Run” on page 5-3.
LCN 50-04344 Rev. D 5-14
Page 61

Completing the Run Info Tab

The fields on the Run Info tab contain information about the run. The only fields on this tab that are required are the Carrier Oil and Drive Oil fields and only if the instrument is out of one of these types of oil.
RainDrop Sense Operator’s Manual
Figure 5-12: Run Info Tab
Your lab procedures determine how much of the information is necessary for your specific situation. Enter the run information as follows:
Table 5-3: Run Info Fields
Field Name
Operator The name of the person performing the run.
PCR Tube ID
Run Comments
LCN 50-04344 Rev. D 5-15
Meaning
The identifier of the sample to be run.
Any detailed descriptive information you want stored with the run information.
Page 62
RainDrop Sense Operator’s Manual

Completing the Run Tabs Using the Import Function

To complete the run tabs using the import function:
1. Click Import at the top of the Run Data tab. The import works only with XML files
generated by the Source instrument. The page displays a list of runs.
2. Locate the file to import and click on it.
3. ICS imports only the following fields:
On the Run Data tab:
Run Name
Lane information
On the Run Info tab:
Operator
PCR Tube ID
Run Comments
4. Complete any other of the optional fields, as needed.

Using Email Notification

The RainDrop Sense instrument is located in a lab space which may be a distance from your desk or usual work area. Email Notification allows you to receive messages when a run is finished, so that you do not have to continually check the instrument. Select Send Run Emails on the Run Info tab. If email addresses have been entered by an Administrator, they appear in the Email Notification field. You can leave these addresses as is, edit them, or add to them. Notifications are sent to those addresses.
Note: Email notification does not support text messaging.
Setting Up Email Notification (Administrator)
Setup Email Notification as needed. You do not have to edit these settings for each run. You can establish email settings and then not change them unless you need to change who is notified at the end of a run.
To enter email addresses:
1. Select Setup Email from the Actions menu. The Email dialog opens.
2. Enter email addresses in the Email Notification field, separated by semi-colons.
These email addresses will appear on the Run Info tab by default.
3. Select Send Run Emails, if you want this to be the default action.
LCN 50-04344 Rev. D 5-16
Page 63
RainDrop Sense Operator’s Manual
4. Click Save to save the new addresses. Click Close to dismiss the dialog.
Email Messages
Email messages contain the following information:
Table 5-4: Email Message Data
Field Name What is reported
RainDance ID
Run Name The name the user entered in the Run Name field prior to
Description
Start Time The date and time the run was started
End Time
Sample Info For each sample, it lists: Lane, Sample Description, Droplets,
Total Samples The number of samples attempted
Samples Succe
Samples Failed The number of samples that generated an error, including any
Errors All errors encountered during the run
ssful
The serial number of the instrument
beginning the run. If the user does not enter a Run Name, this lists a name provided by the RainDrop Sense instrument.
The comments entered by the user at run time, if any
The date and time the run ended
CV (coefficient of variance), Read Length, and Status
The number of samples that successfully ran without errors
samples not processed because the run was stopped because of a prior/fatal error
LCN 50-04344 Rev. D 5-17
Page 64

Performing the Run

After you have entered the required run data, you are ready to begin the run.
RainDrop Sense Operator’s Manual
1. Click the Start Run bu
progress as shown in the following image.
tton. The screen displays information about the Sense run in
Figure 5-13: Sense Run Screen
The center of the screen displays information about which sample is being processed. The right side of the screen displays information about the progress of the run.
2. Whe
3. Lift the
4. T
LCN 50-04344 Rev. D 5-18
n the sample is done, the door unlocks automatically.
door to its full open position.
o minimize the potential for contaminating the lab environment with the amplicon, return the used chip to the bag it came in and discard it in a solid waste disposal container.
Page 65
The ICS displays the Run Complete screen.
RainDrop Sense Operator’s Manual
Figure 5-14: Sense Run Complete Screen
5. Click View Run Report to review the run data.
6. Click Sa
ve and Close. ICS saves the data from the run in a file in FCS format, which
can be read by the RainDrop Analyst software.
Note:
As stated above, the instrument door unlocks after the run is complete. If you
accidentally close the door and still need access, click Open Door to open it.
LCN 50-04344 Rev. D 5-19
Page 66

Canceling a Run

If you want to stop a run after you have started filling out the Run Data tab, but before clicking Start Run, click Cancel. This brings you back to the Instrument Status and Setup a Run Screen (see
If you want to stop a run after clicking Start Run, click the Abort button. This brings you to the Run Completion screen.
Figure 5-3) and blanks out any data you have entered for this run.
RainDrop Sense Operator’s Manual
LCN 50-04344 Rev. D 5-20
Page 67

Handling Data

The data from Sense runs is collected on the hard drive of the ICS Workstation. Files are stored in folders on the D drive named for their runs within a RainDance folder as follows:
D:\RainDance\RunLogs\<RunID>
The <RunID> is displayed in the upper right corner of the ICS screen on both the Run Data tab and the Run Info tab.
The Run ID is built on the date and time as follows: YYMMDDHHMM. In the above image, the number 1302181529 represents 3:29 pm on February 18, 2013. The Sense ICS creates the folder when you click Start Run.
Upon completing the run, the Sense ICS creates one FCS file per lane within the Run ID folder. The filename for one lane of a run looks like this:
<RunID>-<Lane Letter>-<Lane Description>.fcs
For example: 1302181529-A-ExampleDescription.fcs
RainDrop Sense Operator’s Manual
The lane letter and lane description come from data you enter on the Run Data tab. See
“Completing the Run Data Tab” on page 5-12 for more information.
LCN 50-04344 Rev. D 5-21
Page 68

Exporting Data

The Export Data function allows you to search for previous runs and to save that run data to either the same or a new location. This allows you to move your FCS files from one location to another to make them available for use by other applications such as analysis software. To export data:
1. La
RainDrop Sense Operator’s Manual
unch the ICS as you normally do. See “Starting the ICS Application” on page 4-2.
2. Click Export Data.
The Export Data screen opens.
Figure 5-15: Export Data Screen
3. Use the Start Date and End Date fields to specify a range for the data. Click Update List to tell ICS to display only data that fall between the dates you enter.
4. Select o runs or Ctrl+Click to select non-contiguous runs.
5. T
ype the name of an Export Directory or click Browse to locate a folder for the
export.
6. Click Ex the message.
LCN 50-04344 Rev. D 5-22
ne or more runs by clicking on them. Use Shift+Click to select contiguous
port. When the export is done, ICS displays a message. Click OK to dismiss
Page 69

Moving Data Files

The data files created by the Sense instrument from just one run can easily exceed several gigabytes. You can use the Export process described above to locate your FCS files in a location that is accessible by your analysis software. In addition, RainDance offers the following instructions for moving your files when disk space becomes an issue on the ICS Workstation. Keep in mind that the network configuration at your location is unique. You may need to seek assistance from your in-house IT department to learn their recommendations.
Because RainDrop files are large, RainDance has several recommendations for managing your data files.
Never attempt to move files while a run is in process. Move files at a time when network activity is at a minimum so that file transfer is as
efficient as possible.
Consider developing a file management protocol. The disk drive on the ICS is large,
but not infinite. You may need to save files to an external storage device and remove some from the internal hard drive from time to time. Consult with your internal IT department to develop a strategy for managing large amounts of data.
RainDrop Sense Operator’s Manual
To copy files using a flash drive:
1. Insert the flash drive into a USB port on the ICS Workstation.
2. Launch Windows Explorer and navigate to the D drive.
3. Open the RainDance folder on the D drive.
4. Open the Run folder whose data you want to copy.
5. Click the Type column to sort the data files by type. The only files you need to move to
the computer running RainDrop Analyst II are the FCS files. They should appear at the top of the sorted list.
6. Select only the FCS files. Click and drag them to the flash drive icon within Windows Explorer.
7. When Windows Explorer is done copying the files to the flash drive, bring the drive to the computer running RainDrop Analyst.
To copy files over a network:
1. Launch Windows Explorer and navigate to the D drive.
LCN 50-04344 Rev. D 5-23
Page 70
RainDrop Sense Operator’s Manual
2. Open the RainDance folder on the D drive.
3. Open the Run folder whose data you want to copy.
4. Click the Type column to sort the data files by type. The only files you need to move to
the computer running RainDrop Analyst II are the FCS files. They should appear at the top of the sorted list.
5. Select only the FCS files. Right-click on the collection of files.
6. From the pop-up menu, select Copy.
7. Navigate to the network drive where you want to copy the files.
8. Locate the folder where you want to move the FCS files. Right-click and from the pop- up menu, select Paste.
LCN 50-04344 Rev. D 5-24
Page 71
Chapter 1:
CHAPTER 6
Chapter 6:

Managing the System

This chapter covers the following topics:
Introduction page 6-2
Managing Users page 6-2
Adding a New User page 6-3
Editing a User page 6-5
Deleting a User page 6-7
Resetting Password or Unlocking an Account page 6-7
Viewing Past Runs page 6-8
6
Viewing Runs page 6-8
LCN 50-04344 Rev. D 6-1
Page 72

Introduction

You must have a username and password before you can login to the ICS application. There are two types of user accounts: Administrator and User. Chapter 4 provides detailed information on logins and account types.
RainDrop Sense Operator’s Manual
Important:
This chapter describes several soft
Managing Users
Viewing Past Runs
To manage users, you must be logged in with Administrator privileges.

Managing Users

The ICS software lets you manage user accounts from the Actions menu (see Figure 6-1).
ware management functions. These include:
Figure 6-1: Actions Menu
LCN 50-04344 Rev. D 6-2
Page 73

Adding a New User

To add a new user:
RainDrop Sense Operator’s Manual
1. From the Ac The Add New User dialog
tions menu, select Add New User (see Figure 6-1).
opens (see Figure 6-2).
Figure 6-2: Add New User Dialog
2. Enter the information for the new user as required (see Table 6-1) and click Add to save it or Cancel to dismiss the dialo
LCN 50-04344 Rev. D 6-3
g without saving your entries.
Page 74
Table 6-1: Add New User Dialog
RainDrop Sense Operator’s Manual
Field or Table Element
User Name Specifies the name you use to login. Case sensitive.
Password
First Name, Last Name Specifies the user's first and last names. Required.
Email Address
Address Specifies the user's address. Required.
Contact Number
Privilege Specifies the user's Privilege level (Admin or User)
Add, Cancel buttons Adds the user information, or cancels the process. If
Description
Required.
Specifies the password; any alphanumeric combination is acceptable. Case sensitive. Required.
Specifies the user's email address. Required.
Specifies the user’s phone number. Required.
selected from a drop-down menu. Required.
added
successfully, a confirmation window displays.
LCN 50-04344 Rev. D 6-4
Page 75

Editing a User

The Edit Users dialog allows a user with administrator privileges to edit the information for an existing user or remove an existing user from the database. An administrator can also add a new user from this dialog.
RainDrop Sense Operator’s Manual
1. From the Ac opens (see Figure 6-3).
tions menu, select Edit Users (see Figure 6-1). The Edit Users dialog
Figure 6-3: Edit Users Dialog Box
LCN 50-04344 Rev. D 6-5
Page 76
RainDrop Sense Operator’s Manual
2. To edit the data for an existing user, click Edit in the row for that user. The Edit User Information dialog opens for the selected user (see Figure 6-4).
Figure 6-4: Edit User Dialog Box
3. Editing user information is essentially the same as adding new user information. Edit
the user information as required (see Table 6-1) and th
en click Save. Click Cancel
button to dismiss the dialog without saving any changes.
LCN 50-04344 Rev. D 6-6
Page 77
RainDrop Sense Operator’s Manual

Deleting a User

To delete a user:
1. Select Edit Users from the Actions menu (see Figure 6-1). The Edit Users dialog opens as in Figure 6-3.
2. Click Remove in the same row as the user to be deleted. The system asks you to confirm that you want to delete the user. Click Yes to confirm or No to cancel the deletion.

Resetting Password or Unlocking an Account

After three failed attempts to login with the wrong password, the ICS application locks the account and you are not able to login. If this happens, you must contact someone in your facility with Administrator privileges who can either reset or unlock your password.
To reset an account:
1. Select the Reset Password check box in the Edit User dialog (see Figure 6-4). This changes the user's password to the default password, abc123.
2. To save the changes to the account, click Save. The application displays a message confirming that the changes have been saved.
To unlock an account:
1. Select the Unlock Account check box in the Edit User dialog (see Figure 6-4). This allows the user to login again with his/her original password.
2. To save the changes to the account, click Save.The application displays a message confirming that the changes have been saved.
LCN 50-04344 Rev. D 6-7
Page 78

Viewing Past Runs

This section describes how to view previous runs. It also describes how to generate and view reports and how to export run results.

Viewing Runs

Certain run parameters are saved by the ICS application. The ICS application lets you search for and view past run parameters by entering a specific Start Date and an End Date (inclusive) on which the run was completed.
To view a run report:
RainDrop Sense Operator’s Manual
1. Select V
iew Runs from the Actions menu (see Figure 6-1). The View Run Report
window opens (see Figure 6-5).
Figure 6-5: View Run Report Window
2. Enter a date range to locate the run of interest and click Search. The default start and
end dates are today’s date, so if you want to see the run you just completed, you can just click Search without changing the date.
All runs that match the selected date range appear in a list below the white title b
LCN 50-04344 Rev. D 6-8
ar.
Page 79
RainDrop Sense Operator’s Manual
3. Clicking or selecting a row in the list of runs opens the report for that run. This may take 2-5 seconds.
Figure 6-6: Example of a Sense Run Report
Reports are divided into three sections. The first section contains:
Run Name Operator
RainDance ID (machine name)
Tube Strip
The second section contains a summary of how the run performed. It
Channel Pass/Fail
Sample Information
The last section lists the following:
Start Time End Time
Chip Lot
Chip Serial Number Carrier Oil Lot
Drive Oil Lot
Run Comments
includes:
LCN 50-04344 Rev. D 6-9
Page 80
RainDrop Sense Operator’s Manual
4. From the menu bar (see Figure 6-7) you can print the report or export it to various formats, such as a PDF file or an Excel file.
Figure 6-7: Report Menu
5. When you are done viewing the report, close it by clicking on the Close box in the
upper right corner of the screen. You return to the View Runs window so you can view additional reports if necessary.
LCN 50-04344 Rev. D 6-10
Page 81
Chapter 1:
CHAPTER 7
Chapter 7:

Support, Customer Maintenance, and Troubleshooting

This chapter covers the following topics:
RainDance Support page 7-2
RainDance Planned Maintenance page 7-2
Customer Maintenance page 7-2
Restarting the IC Workstation page 7-2
Cleaning the RainDrop Sense page 7-2
7
Refilling the Oil Reservoir page 7-3
Service and Diagnostic Requirements page 7-6
Powering Off the System page 7-6
Powering On or Restarting the System page 7-6
Restoring the System After a Failure page 7-6
Support Agreement Information page 7-7
Moving the System page 7-7
Storing the System page 7-7
Reinstalling/Updating ICS Application page 7-7
Error Messages page 7-7
Troubleshooting page 7-8
LCN 50-04343 Rev. D 7-1
Page 82
Maintenance refers to non-technical procedures you must perform to keep the system
working properly. Service refers to tasks that are intended to be performed only by a qualified
RainDance-authorized service representative. The RainDrop® System comprises two instruments: RainDance Source and RainDrop Sense. The sections in this chapter describe them separately only when each instrument requires different maintenance or service.

RainDance Support

RainDrop Sense Operator’s Manual
RainDance Technologies
Some troubleshooting information is provided within this chapter (see
“Troubleshooting” on page 7-8.). Additional self-service technical and troubleshooting
information is available at www.raindancetech.com/support.
To request technical assistance from RainDance support personnel, send an email to
support@raindancetech.com.
®
provides full support for their products.

RainDance Planned Maintenance

A Planned Maintenance (PM) schedule and tasks are designed to be performed by RainDance Support personnel. A number of PMs are conducted as part of the service contract agreement. Contact RainDance Support for a recommended PM schedule.

Customer Maintenance

This section lists the steps you should follow for proper maintenance of the RainDrop Sense system.

Restarting the IC Workstation

Good operating practice for any Windows operating system includes restarting the operating system on a weekly basis. To perform a shutdown, select the Start icon the lower left corner of the screen and select Shut down. When the system has completed its power off routine, reboot the Windows 7 operating system on the IC Workstation as described in
or Restarting the System” on page 7-6.
“Powering On

Cleaning the RainDrop Sense

There is no extraordinary routine maintenance required for the RainDrop instruments. Clean the chip dock of spills and residues as follows:
Always use protective gloves.
Use only a soft, lint-free towel to clean the chip dock; wet with either isopropanol or
HFE-7100.
For DNA decontamination, RainDance recommends using HFE-7100 or DNA-
ExitusPlus.
LCN 50-04343 Rev. D 7-2
Page 83
Keep cleaning fluids away from the objective lens.
Caution: Use caution while cleaning the chip dock, specifically in the area of the lenses of each instrument.
Caution: Never refill an oil reservoir unless the red light is illuminated. Failure to comply can result in an internal overflow and potential damage to the instrument.
DO NOT submerge the system in liquid or pour or spray liquid onto the system.
Clean the IC workstation monitor window as described in its manufacturer's
documentation.
All technical maintenance or service must be performed by RainDance Support
personnel.

Refilling the Oil Reservoir

The oil reservoir on the RainDrop Sense instrument must be refilled from time to time. You will know that it is time to refill the reservoir when the oil light on the front of the instrument glows red.
RainDrop Sense Operator’s Manual
Carrier oil reservoir
fill port
Carrier oil low
indicator light
Drive oil reservoir
fill port
Drive oil low
indicator light
Figure 7-1: RainDrop Sense Oil Reservoir Fill Ports
LCN 50-04343 Rev. D 7-3
Page 84
To refill the oil reservoir:
RainDrop Sense Operator’s Manual
1. Obtain
2. Sca
a pre-measured syringe of oil (P/N 30-07117).
n the barcode on the oil syringe. Refer to “Scanning Barcodes” on page 2-5 for
more information on how to scan.
3. Remove t
4. Pla
ce a gloved finger over the end of the syringe to ensure that it does not squirt oil
he cap from the tip of the syringe.
while you are moving it.
Figure 7-2: Gloved Finger over Syringe
5. Unscrew the cap that covers the oil reservoir fill port and place it in a safe place.
LCN 50-04343 Rev. D 7-4
Page 85
RainDrop Sense Operator’s Manual
Caution: Do not overtighten the syringe onto the fill port.
6. Place the tip of the syringe over the fill port and begin to turn the entire syringe
clockwise. The syringe screws onto the fill port to create a tight fit.
Figure 7-3: Syringe Connected to the Fill Port
7. Continue to turn the syringe until it fits snugly.
8. Press the end of the syringe and inject its contents until the syringe is empty.
9. Unscrew
the empty syringe and discard it in a lab-approved solid waste stream. When
you are done refilling the reservoir, the oil low indicator light turns off.
10. Replace
the cover on the fill port. Do not overtighten.
LCN 50-04343 Rev. D 7-5
Page 86
RainDrop Sense Operator’s Manual

Service and Diagnostic Requirements

The system checks its status at start-up automatically. The ICS application prevents
runs when it detects a system failure or not ready condition.
The hardware is designed to facilitate field replacement of subsystems and/or
subsystem components as required.
There are no user serviceable parts inside the RainDrop instrument. For safety
reasons, those components have limited access.
All technical maintenance or service is provided by RainDance Support.

Powering Off the System

Follow this sequence to turn the system off:
1. Exit the ICS application. Wait 10 seconds.
2. Power off the instrument by turning off the switch on the rear of the instrument.
3. Power off the IC Workstation by selecting the Start icon and then selecting Shut down.
4. Press the Power button on the monitor to turn it off.

Powering On or Restarting the System

Refer to “Starting the ICS Application” on page 4-2.

Restoring the System After a Failure

In most cases, the RainDrop instruments automatically reset after a failure. You can start subsequent runs without manual intervention.
LCN 50-04343 Rev. D 7-6
Page 87
Note: Never start another run until the root cause of the failure is identified and resolved.
Warning: The RainDrop Sense instrument weighs approximately 65 lbs (29 kg). To
avoid injury and damage to the instruments, do not move them without assistance from RainDance Technologies.

Support Agreement Information

For detailed information on Support Agreements, go to
www.raindancetech.com/support and click
on the Support Agreement link.

Moving the System

To move the instrument to another location, contact RainDance Support.
RainDrop Sense Operator’s Manual

Storing the System

For short term storage (less than 30 days) of the instruments:
Remove any RainDrop Chips and PCR tube strips from the instruments.
Leave the Carrier Oil and Drive Oil from the last run in the instruments.
Close the door. Power off the instruments.
For long term storage of the instruments contact RainDance Support.

Reinstalling/Updating ICS Application

The ICS application is installed and tested during installation. You are not required to perform any installation or maintenance of ICS. Any new version of ICS that is user-installable includes instructions for installation. If the ICS application files lose their integrity for any reason (for example, they get accidentally moved or deleted), you can re-install the ICS application after contacting RainDance.

Error Messages

For current error messages and actions, refer to www.raindancetech.com/support.
LCN 50-04343 Rev. D 7-7
Page 88

Troubleshooting

The following table lists some problems that you may observe while working with the RainDrop Sense instrument. The table is divided into three columns as follows:
Observed Problem
Possible Cause
Corrective Actions and Comments
Table 7-1: Troubleshooting the RainDrop Sense Instrument
Observed Problem Possible Cause
RainDrop Sense Operator’s Manual
Corrective Actions and Comments
There is a red X on the
Instrument Status and Setup a Run screen.
The ready indicator next to the Chip Inserted System Status displays a red X even after I insert a chip.
The gas pressure is not connected or is not adequate.
The instrument does not detect a properly inserted chip. The chip is not inserted properly, may not be seated well, or is not aligned with the chip alignment pins.
Ensure that the gas pressure cable is connected properly and that the gas pressure is sufficient. Click Start Run. A dialog box opens listing the reason(s) the run cannot start. If you continue to have problems, check the Alarms Tab first; see
Alarms, Log, and Status Tabs on page 4-14.
Then contact RainDance Support. See
Support on page 1-3.
The chip sensor will not detect the chip if there is an error in the barcode name. If you have entered the barcode try scanning the chip, then placing it in the instrument. Shut down the software and instrument, then restart the instrument, wait 1 minute, and restart the software and scan the barcode again, then place it in the instrument. If the chip is still not recognized, try another chip and contact RainDance support. See
RainDance Support on page 1-3.
RainDance
LCN 50-04343 Rev. D 7-8
Page 89
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
The ICS does not display droplets during a run or displays an unstable pattern of droplets.
One of the lanes fails, displays erratic droplet formation, or uneven droplet spacing.
There is a white sheen on the tube strip when I remove it from the instrument and there is liquid in the tube strip nest and the emulsion is not in the tube as expected.
You have under-filled one or more sample wells. This means that there is less than 25 µL in the problematic well. The lane will fail.
Sample may be improperly sheared, may have too much DNA, or may have contaminants such as fibers that interfere with droplet formation. Refer to the RainDrop Assay dPCR Guidelines for information in preparing the sample.
Inspect the tube strip to see if there is a hole in the bottom of one of the tubes. This is a rare, but known, manufacturing defect and the sample may have leaked out of the tube.
The run may time out after 50 minutes or it may finish properly.
The sample in that lane is lost. Deselect the corresponding lane on the Sense run.
The sample is lost. Make sure the DNA is properly sheared. Try diluting the sample and generating droplets. Filter materials through a 0.2 micron filter to remove debris and contaminants.
Visually inspect tube strips before using them.
The sample is lost in tube strip nest and is unrecoverable. In addition, you must contact RainDance Support (see
RainDance Support on page 1-3) to clean the
instrument to ensure that the tube strip sensor is working properly.
The ICS displays unstable droplet generation, a lower than expected number of droplets, or the lane fails.
LCN 50-04343 Rev. D 7-9
The sample was improperly pipetted and either is mixed with too much air or is a bead on the side wall of the sample input well.
Sample may be lost. When you pipette sample, ensure that you place the tip at the bottom of the well to avoid leaving the sample on the wall of the sample input well.
Page 90
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
One of the lanes does not display any droplets. In addition, the run may stop prematurely if emulsion is not detected.
The ICS displays a message telling me that the chip I inserted has been used before.
The ICS displays a message telling me that the barcode for samples is invalid.
Emulsion is not detected because the emulsion sensor is tripped. This may be due to one of the following:
--the tube strip is dirty
--the tube strip contains writing in the conical area
--the tube strip was partially filled prior to your using it
--there is a static problem that pulls the emulsion down the side of the tube
--tube strip is not installed properly
You are attempting to run a chip that has already been run.
You are attempting to use a barcode format (for samples) that is not supported by the barcode reader.
The sample is lost. In some cases, you may be able to recover the sample. Contact RainDance Support (see
Support on page 1-3) for
more information.
You must use a new chip for each run.
You must use a supported barcode format or manually enter the barcode information for samples. The ICS accepts a barcode in the 2D Data Matrix format, which allows up to 96 characters of text.
RainDance
I am not receiving emails, even though I have entered an email address.
You have entered an invalid email address. The system does not detect undelivered emails.
There may be a connectivity issue with your network.
If you are not receiving email messages at the end of the run, check the email addresses you have entered for accuracy.
Check with your IT department to be sure you are properly connected.
When setting up a run, ensure that the Send Run Emails check box is selected.
Click the Send Test Email button to see if you have solved the problem.
LCN 50-04343 Rev. D 7-10
Page 91
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
I tried to fill the oil reservoir before the oil light came on.
There is dirt, debris, or oil residue on the instrument.
The sample has evaporated. The sample volume in one or more wells is reduced or totally gone. The droplet count is lower than expected.
The tube strip or the tube strip cap was deformed during thermal cycling and resulted in a leak into the PCR tube strip nest of the Sense.
It is possible to overfill the oil reservoir and to spill oil into internal areas of the instrument. You may not notice it, even when you have done this.
The instrument has not been properly cleaned.
The tube strip cap was not seated properly or the tube strip was not covered within 30 minutes of Source run completion.
For this reason, do not refill the oil reservoir until the oil light comes on.
Clean the instrument according to the instructions in
RainDrop Sense on page 7-2.
Ensure that the tube strip cap is seated properly. It should be even along the entire length of the tube strip. An extra cap is supplied in each kit; if you find one that does not fit properly, discard it and use another.
Do not reuse tube strip caps. Caps are pierced by the Sense instrument as part of the process. Pierced tube strip caps allow evaporation and introduce the potential for cross contamination.
Cleaning the
Clean the PCR tube strip nest on the Sense instrument. Use the recommended cleaning protocol. See
RainDrop Sense on page 7-2.
One lane or the entire run has failed.
LCN 50-04343 Rev. D 7-11
There may be dirt, debris, scratches, or smudges on the imaging region of interest on the Sense Chip.
Handle chip as recommended in
the Sense Chip on page 2-
6.
Keep the chip dock clean as described in
the RainDrop Sense on page 7-2.
Cleaning the
Handling
Cleaning
Page 92
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
The Alarms tab or Log tab displays an error message. The ICS displays an error message in a dialog box.
The ready indicator next to the Chip Inserted System Status displays a red X even after I insert a chip.
A known error has occurred. The ICS may suggest a way for you to deal with the error.
The instrument does not detect a properly inserted chip. The chip is not inserted properly, may not be seated well, or is not aligned with the chip alignment pins.
Follow the instructions provided by the error message. If no instructions are provided, verify that the instrument has power and verify that all cable connections are secure. See
Connecting RainDrop Sense and Instrument Controller on page 3-3.
Contact RainDance Support if the problem is not resolved. See
RainDance Support on page 1-3.
The chip sensor will not detect the chip if there is an error in the barcode name. If you have entered the barcode try scanning the chip then placing it in the instrument. Shut down the software and instrument, then restart the instrument, wait 1 minute, and restart the software and scan the barcode again, then place it in the instrument. If the chip is still not recognized, try another chip and contact RainDance support. See
RainDance Support on page 1-3.
There is a large leak from the PCR tube.
The ICS reports a minimal number of droplets being read. In addition, you may receive an email notification that includes failure information.
LCN 50-04343 Rev. D 7-12
You are using an incorrect PCR tube strip, meaning that it is not an Axygen tube strip.
There is no sample in one or more lanes. The assay may be too dim for non­amplified droplets to be detected.
The sample is lost.
The lane or lanes will fail or timeout after 30 minutes. Multiplex your assay or reorder with a different quencher.
Page 93
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
Some of the expected lanes were not processed. The sample remains in the Sense Chip sample wells.
The data does not appear where I expect it. The expected results appear in another lane or are not detectable at all.
You did not select the lanes on the Run Data tab of the ICS (see
the Run Info Tab on page -
15) that represent the
lanes on the Sense Chip that contain sample.
You may have confused the lane name with the physical position of the sample in the tube strip.
Completing
The sample may be recoverable, but you cannot reuse the chip. Contact RainDance Support (see
Support on page 1-3) for
help recovering the sample. In addition, there is a risk of contaminating the work area with amplicon. See
the RainDrop Sense on page 7-2.
The run proceeds as normal. Make a note of which samples have been switched and address this issue in downstream analysis. Use a multichannel pipettor in the future to reduce the risk of mixing up samples.
RainDance
Cleaning
The Sense ICS displays a lower or higher than expected droplet count.
The ICS displays a message telling me that the chip I inserted has been used before.
The ICS displays a message telling me that the barcode for samples is invalid.
You have selected a droplet count, but the Sense Instrument is not delivering what you requested.
You are attempting to run a chip that has already been run.
You are attempting to use a barcode format (for samples) that is not supported by the barcode reader.
The run proceeds as normal. It is not always possible to achieve an exact number of droplets.
The system is designed to produce a number of droplets higher than what you select to account for filtering losses.
You must use a new chip for each run.
You must use a supported barcode format or manually enter the barcode information for samples. The ICS accepts a barcode in the 2D Data Matrix format, which allows up to 96 characters of text.
LCN 50-04343 Rev. D 7-13
Page 94
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
I am not receiving emails, even though I have entered an email address.
I am not receiving a text message at the end of a run even though I have entered my phone number into the email field.
You have entered an invalid email address. The system does not detect undelivered emails.
There may be a connectivity issue with your network.
The RainDrop Sense instrument does not support text messages.
If you are not receiving email messages at the end of the run, check the email addresses you have entered for accuracy.
Check with your IT department to be sure you are properly connected.
When setting up a run, ensure that the Send Run Emails check box is selected.
Click the Send Test Email button to see if you have solved the problem.
You must use a valid email address for email notification.
I tried to fill the oil reservoir before the oil light came on.
There is dirt, debris, or oil residue on the instrument.
It is possible to overfill the oil reservoir and to spill oil into internal areas of the instrument. You may not notice it, even when you have done this.
The instrument has not been properly cleaned.
For this reason, do not refill the oil reservoir until the oil light comes on.
Clean the instrument according to the instructions in
RainDrop Sense on page 7-2.
Cleaning the
LCN 50-04343 Rev. D 7-14
Page 95
Table 7-1: Troubleshooting the RainDrop Sense Instrument
RainDrop Sense Operator’s Manual
The sample has evaporated. The sample volume in one or more wells is reduced or totally gone. The droplet count is lower than expected.
The tube strip or the tube strip cap was deformed during thermal cycling and resulted in a leak into the PCR tube strip nest of the Sense.
The tube strip cap was not seated properly or the tube strip was not covered within 30 minutes of Source run completion.
Ensure that the tube strip cap is seated properly. It should be even along the entire length of the tube strip. An extra cap is supplied in each kit; if you find one that does not fit properly, discard it and use another.
Do not reuse tube strip caps. Caps are pierced by the Sense instrument as part of the process. Pierced tube strip caps allow evaporation and introduce the potential for cross contamination.
Clean the PCR tube strip nest on the Sense instrument. Use the recommended cleaning protocol. See
RainDrop Sense on page 7-2.
Cleaning the
The Alarms tab or Log tab displays an error message. The ICS displays an error message in a dialog box.
A known error has occurred. The ICS may suggest a way for you to deal with the error.
Follow the instructions provided by the error message. If no instructions are provided, verify that the instrument has power and verify that all cable connections are secure. See
Connecting RainDrop Sense and Instrument Controller on page 3-3.
Contact RainDance Support if the problem is not resolved. See
RainDance Support on page 1-3.
LCN 50-04343 Rev. D 7-15
Page 96
Chapter 1:
CHAPTER 8
Chapter 8:

Waste Management

This chapter covers the following topics:
Introduction and Disclaimer page 8-2
Classification of Waste page 8-2
Anticipated Typical Waste Stream Composition page 8-3
Recommended Waste Stream Disposal page 8-4
8
LCN 50-04344 Rev. D 8-1
Page 97

Introduction and Disclaimer

The waste stream from the RainDrop® platform includes fluorocarbon oil-emulsified aqueous samples and wetted solid materials (PCR tubes, chips, etc.). The following guidelines are intended to assist the customer in communicating with Environmental Health and Safety staff and/or commercial waste disposal service providers. The following guidelines are not definitive. The Customer is solely responsible for complying with applicable laws and regulations with respect to waste management and disposal.
Important: The Customer must dispose of all waste in accordance with all applicable environmental laws and regulations.

Classification of Waste

Depending on applicable environmental laws and regulations, the waste stream may
be treated as hazardous chemical waste.
Depending on applicable environmental laws and regulations, the waste stream may
not be classified as mixed hazardous waste.
RainDrop Sense Operator’s Manual
Important: If the DNA sample is classified as infectious by the institution that operates the RainDrop, it is not considered the normal waste stream. Refer to the institution's EH&S regulations.
The customer may be required to certify to the chemical waste disposal service
provider that the normal waste stream is non-infectious.
Consult applicable laws and regulations to determine if fluorinated materials present in
the waste stream are regulated.
Customers should consult with local authorities to determine specific state and local
requirements.
LCN 50-04344 Rev. D 8-2
Page 98
RainDrop Sense Operator’s Manual

Anticipated Typical Waste Stream Composition

During normal workflow operation, for every 96 samples run on the RainDrop Sense system, approximately 570 grams (20.11 oz) of waste is generated. The following approximate composition of the waste stream is expected if all consumables and waste reagents are combined into a single waste stream:
Table 8-1: Solids Appearing in Waste Stream
Sense Waste
Material
Fluorocarbon oil (from Carrier and Drive Oil) 15.3
Amount (wt %)
Fluorosurfactant (from Carrier and
Water 0.8
DNA, polymerase, DNTPs, buffer salts
Commercial polymers (for example, poly(etheretherketone), polyethylene, polycarbonate, polypropylene, COC)
Figure 8-1 and Figure 8-2 show images of solid waste.
Drive Oil) 0.4
Trace
83.4
Figure 8-1: Syringe
LCN 50-04344 Rev. D 8-3
Page 99
RainDrop Sense Operator’s Manual
Figure 8-2: Spent Sense Chip with Tube Strip Attached

Recommended Waste Stream Disposal

Depending on applicable environmental laws and regulations, the waste stream, which
contains fluorinated chemicals, should be incinerated in an industrial or commercial facility in the presence of a combustible material.
Combustion products will include hydrofluoric acid.
The incineration facility must be capable of handling halogenated – particularly
fluorinated – materials.
Depending on applicable environmental laws and regulations, the non-infectious
aqueous samples and the solid materials (chips, tube strips, and caps, etc.) do not need to be separated from the fluorinated oil.
Depending on applicable environmental laws and regulations, all components can be
collected in a single container, and the container can be incinerated as miscellaneous lab debris.
Customers should consult with local authorities to determine specific state and local
requirements.
LCN 50-04344 Rev. D 8-4
Page 100
Chapter 1:
CHAPTER 9
Chapter 9:

Customer Care Information

This chapter covers the following topics:
General Care Information page 9-2
Consumables page 9-2
9
LCN 50-04344 Rev. D 9-1
Loading...