RADWAG AS series, X series User Manual

User manual no.: LMI-31-08/04/13/ENG
Analytical balances AS/X
series
BALANCES AND SCALES
RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com
www.radwag.com
- 2 -
APRILE 2013
- 3 -
Table of contents
1. UNPACKING AND ASSEMBLING................................................................................. 5
1.1. Balance AS/X series ............................................................................................. 5
2. STARTUP ....................................................................................................................... 6
2.1. Intended use ......................................................................................................... 6
2.2. Leveling................................................................................................................. 6
2.3. Balance temperature stabili z atio n period .......................................................... 6
3. APPLICATION ............................................................................................................... 7
4. BALANCE DESCRIPTION ............................................................................................. 8
4.1. Graphic display .................................................................................................... 8
4.2. Keyboard............................................................................................................... 8
4.3. Sockets and interfaces ........................................................................................ 9
5. USER MENU ................................................................................................................ 10
5.1. Moving through the menu ................................................................................. 14
5.2. User menu - content .......................................................................................... 16
6. WEIGHING ................................................................................................................... 18
6.1. User log in........................................................................................................... 19
7. ADJUSTMENT ............................................................................................................. 22
7.1. Internal automatic adjustment .......................................................................... 22
7.2. Adjustment test .................................................................................................. 24
7.3. Manual adjustment ............................................................................................. 25
7.4. Adjustment report printout ................................................................................ 27
8. DETERMINING CONTENT OF A PRINTOUT FOR GLP PROCEDURES ................... 28
9. DATE AND TIME SETTINGS ....................................................................................... 29
10. SETTING BALANCE OPERATING PARAMETERS.................................................... 32
10.1. Filter settings ...................................................................................................... 32
10.2. Value release ...................................................................................................... 32
10.3. Time interval of display refreshment ................................................................ 32
10.4. Autozero function .............................................................................................. 33
10.5. Last digit ............................................................................................................. 33
10.6. Negative .............................................................................................................. 33
10.7. Air buoyancy correction .................................................................................... 33
10.7.1. Means of operation ................................................................................... 35
10.7.2. Activating air buoyancy correction ......................................................... 35
10.7.3. Determining the coefficient value for known density of air and weighed sample 36
10.7.4. Coefficient determining process using a set of mass standards ......... 37
10.8. Operating conditions ......................................................................................... 40
11. RS 232 FUNCTIONS .................................................................................................... 41
12. PRINTOUTS ................................................................................................................. 42
13. SETTING ACCESSIBILITY OF MEASURING UNITS ......................................... 42
14. SETTING ACCESSIBILITY OF WORKING MODES ................................................... 43
15. OTHER PARAMETERS ............................................................................................... 43
- 4 -
16. WORKING MODES ...................................................................................................... 46
16.1. Parts counting of the same mass ..................................................................... 46
16.2. Checkweighing ................................................................................................... 50
16.3. Filling (dosing) ................................................................................................... 54
16.4. Percent setup ..................................................................................................... 55
16.5. Animal weighing ................................................................................................. 58
16.6. Density determination of solids and liquids .................................................... 59
16.7. Formulation ........................................................................................................ 61
16.8. Statistics ............................................................................................................. 66
16.9. Calibration of pipettes ....................................................................................... 68
17. PRINTOUTS ................................................................................................................. 72
17.1. Standard printout ............................................................................................... 72
17.2. Non-standard printouts ..................................................................................... 73
18. COOPERATION WITH A PRINTER OR A COMPUTER ............................................. 79
18.1. Cross-section through connecting cables ....................................................... 79
19. COOPERATION WITH A CITIZEN LABEL PRINTER ................................................. 80
20. COOPERATION WITH EPSON RECEIPT PRINTER .................................................. 85
21. UNDER HOOK WEIGHING .......................................................................................... 86
22. CONNECING SCHEME OF EXTERNAL BUTTONS ................................................... 87
23. LIST OF COMMANDS COMPUTER - BALANCE ....................................................... 87
24. TECHNICAL DATA ...................................................................................................... 90
24.1. Balance AS/X series ........................................................................................... 90
25. ERROR MESSAGES.................................................................................................... 91
- 5 -
1. UNPACKING AND ASSEMBLING
1.1. Balance AS/X series
Cut the protective tape of balance’s factory packaging. Carefully remove the balance from its packaging. Take out from the box all the accessories needed for correct operation of a balance. Place the balance in its inteded place of use and assembly its components as specified in below scheme. The balance is powered from mains through a power adapter 13,5-16V DC.
Open side glass door of the weighing
chamber,
Inside the weighing chamber place
bottom metal plate (5),
Next place a centering ring (3) onto
the bottom metal plate,
Inside the centering ring assembly
the weighing pan (2),
Inside the weighing chamber and onto the centering ring place an anti­draft shield (1),
Close the side glass doors of the weighing chamber,
Plug the balance to mains using the dedicated power adapter. The power socket for plugging the power adapter’s pin is located at the back of balance’s housing.
Fig. 1. Assembling compon ents of a balance AS/X series
- 6 -
2. STARTUP
2.1. Intended use
The balance should be located on a stable wall console desk or a stable working table
which is not affected by vibrations and distant from heat sources.
Balance should be used in locations free of gusts of air and drafts.
Operate the balance in a room with stabilized temperature and humidity conditions.
Ambient air temperature should not exceed the range of +10°C ÷ +40°C
In order to avoid influence of static electricity on the measurement process, ground
the balance’s housing. The grounding bolt is located at the back of balance’s housing,
The balance should be set to level using the level indicator located at the back of
balance’s housing. Leveling the balance is one of the conditions for ensuring required measuring accuracy.
Before unplugging the balance from mains always switch of the display by pressing
ON/OFF key
2.2. Leveling
Before plugging to mains, level the device using two adjusting feet located at the back of balance’s housing. Turn the adjusting feet in a way that the air bubble of the level is placed centrally. The level is located at the back wall of balance’s housing.
2.3. Balance temperatu re st abili z at ion period
Before start of measuring processes, it is necessary to wait until the balance is thermally stabilized. It is a period of so called balance self-heating time.
In case of analytical balance AS/X series self-heating period takes approximately 1 hour. The specified time interval refers to balances that have been stored in room temperature before plugging to mains. For balances that were stored in much lower temperatures before plugging to mains (e.g. during winter period) thermal stabilization should last approximately 8 hours. During temperature stabilization time the indications on balance’s display may change.
- 7 -
3. APPLICATION
Balance featuring a graphic display is intended to precise determining mass in laboratory conditions. The balance enables zeroing the indication in whole measuring range.
Balance AS/X series enables determining mass in the following measuring units:
Fig. 2. Measuring units
Apart from determining mass of weighed object using different measuring units, the software balance AS/X series comprises the following working modes:
parts counting
checkweighing
filling (doing)
percent setup
animal weighing
determining density of solids and liquids
mixtures from accepted formulation and ingredients
formula making
pipette calibration
statistics
Both measuring units and working modes can be set as inaccessible in the user menu. This option is implemented for the purpose of adjusting balance to user needs and requirements, i.e. providing access only to those functions and units which are required by a user.
Determining accessibility attribute of a working mode / measuring unit is set in balance’s menu and it is described further in this user manual.
- 8 -
4. BALANCE DESCRIPTION
4.1. Graphic display
Fig. 3. Balance display
1. Mass indication of weighed load or counted parts,
2. Measuring unit,
3. Pictogram denoting whether measurement result is stable,
4. „BARGRAPH” presenting which part of accessible measuring range is in use,
5. Text informing on enabled working mode,
6. Current date,
7. Current time,
8. Pictogram denoting, that the indication is in precise ZERO.
4.2. Keyboard
Each button of the balance’s keyboard operates as a dual-function key, i.e. it can either carry out a specific function or be used to move in balance’s menu structure.
- 9 -
ON/OFF
key enables switching on and off balance’s display. If
switched off balance components other than the display are powered, and balance is in stand-by mode.
F
key. Function key, which enables quick entering the settings
of an active working mode
MODE key for selecting balance’s working mode.
UNITS key, changes measuring units.
PRINT/ENTER key -
Sends current display status to a peripheral device (PRINT) or accepts selected value of a parameter or function (ENTER).
ESC/ZERO TARE key –
zeroing / tarring of balance’s
indication.
Adjustment –
function key of immediate initiating the
adjustment / calibration process.
Function key for entering the main menu of a balance.
Navigating arrows for moving in balance’s menu or changing parameter value
4.3. Sockets and interfac es
1. Power supply socket
2. PS/2 keyboard socket
3. USB port (only in non legalization
balances)
4. RS 232 socket
5. Additional display socket
Fig. 4. Interfaces of a balance AS/X series
- 10 -
5. USER MENU
User menu of a balance AS/X series consists of 9 main menu groups named using letter P and a corresponding number. The name and content of the menu is presented below.
P1 Adjustment
01 Internal adjust. |
* * * * * * * *
| function
02 External adjust. |
* * * * * * * *
| function
03 User adjustment |
* * * * * * * *
| function
04 Adjustment test |
* * * * * * * *
| function 05 Weight correction | * * * * **0.0 | 06 Automatic adjust. | * * * * **0.3 | both 07 Auto adjust. time | * * * * *0.3 | 3 hours 08 Report result | * * * * *0.1 | yes
P2 GLP
01 User | Smith John | 02 Project | AR – 65/04 | 03 Print time | * * * * *0.0 | no 04 Print date | * * * * *0.0 | no 05 Print user | * * * * *0.0 | no 06 Print project | * * ** * *0.0 | no 07 Print Id | * * ** * *0.0 | no 08 Print adjustment | * * * * *0.0 | no 09 Print adjust diff. | * * * * * *.1 | yes
P3 Date/Time
01 Date format | * * * * * *0.0 | D/M/R 02 Time format | * * * * * *0.0 | 24 hours 03 Time |
* * * * * * * *
| function
04 Date |
* * * * * * * *
| function
05 Display time | * * * * * *0.1 | yes 06 Display date | * * * * * *0.1 | yes
P4 Readout
01 Filter | * * * * * *0.3 | normal (average) 02 Value release | * * * * * *0.1 | fast+reliable 03 Display refresh | * * * * * *0.1 | 0.08 s 04 Autozero | * * * * * *0.1 | yes 05 Last digit | * * * * * *0.1 | always 06 Negative
| * * * * * *0.0 | no 07 Air buoyancy corr. | * * * * * *0.0 | no 08 Environment | * * * * *0.1 | stable
- 11 -
P5 RS - 232
01 Interface | * * * * * *0.0 | RS 232 02 Baud rate | * * * * * *0.1 | 4800 03 Parity | * * * * * *0.0 | none 04 Data bits | * * * * * *0.2 | 8 bits 05 Stop bits | * * * * * *0.1 | 1 bit 06 Automatic printout | * * * * * *0.0 | no 07 Interval | * * * * * *0.1 | * 0.1 s 08 Min. mass | * * * * * *0.4 | 10 d 09 Print stable | * * * * * *0.1 | yes 10 Printer type | * * * * * *0.0 | standard 11 Printout cut | * * * * * *0.0 | no 12 Erase statistics | * * * * * *0.0 | no
P6 Printout
01 Printout no. | * * * * * *0.0 | standard 02 Printout 1 start | * * * * * *0.1 | 03 Printout 1 stop | * * * * * *0.1 | 04 Printout 2 start | * * * * * *0.1 | 05 Printout 2 stop | * * * * * *0.1 |
... . . . . . . . . . . | * * * * * *0. |
10 Printout editing |
* * * * * * * *
| function 11 String 1 | * * * * * *0. | 11 String 2 | * * * * * *0. |
... . . . . . . . . . . | * * * * * *0. |
89 String 80 | * * * * * *0. |
P7 Units
01 Grams | * * * * * *0.1 | yes 02 Milligrams | * * * * * *0.1 | yes 03 Carats | * * * * * *0.1 | yes 04 Pounds | * * * * * *0.1 | yes 05 Ounce | * * * * * *0.1 | yes 06 Ounce troy | * * * * * *0.1 | yes 07 Dwt | * * * * * *0.1 | yes 08 Taele Hk. | * * * * * *0.1 | yes 09 Taele S. | * * * * * *0.1 | yes 10 Taele T. | * * * * * *0.1 | yes 11 Mommes | * * * * * *0.1 | yes 12 Grains | * * * * * *0.1 | yes 13 Newtons | * * * * * *0.1 | yes 14 Tical’ e | * * * * * *0.1 | yes 15 Custom unit | * * * * * *0.1 | yes
- 12 -
P8 Working modes
01 Parts counting | * * * * * *0.1 | yes 02 Checkweighing | * * * * * *0.1 | yes 03 Filling (Dosing) | * * * * * *0.1 | yes 04 Percent setup | * * * * * *0.1 | yes 05 Animal weighing | * * * * * *0.1 | yes 06 Density | * * * * * *0.1 | yes 07 Formulation | * * * * * *0.1 | yes 08 Pipette calibration | * * * * * *0.1 | yes 09 Statistics | * * * * * *0.1 | yes
P9 Other
01 ID setting |
* * * * * * * *
| function 02 Autom. ID print | * * * * * *0.0 | no 03 Beep | * * * * * *0.1 | yes 04 Language | * * * * * *0.1 | Polish 05 Backlight | * * * * * *0.1 | yes 06 Brightness |
* * * * * * * *
| function 07 Contrast |
* * * * * * * *
| function 08 Screen saver | * * * * * *0.0 | no 09 Temperature |
* * * * * * * *
| function 10 Factory no. | 114493 *0 .| 11 Software no. | xxxxxxxxx | 12 Parameter printout |
* * * * * * * *
| function 13 Upload parameters |
* * * * * * * *
| function 14 Password protect. |
* * * * * * * *
| function
Parameters type in the user menu:
function – having a specific operation, e.g. balance adjustment
selectable – enables selecting one of a few available values, which are
permanently set in balance’s memory, like: display refreshing, screen saver, determining availability of a measuring unit or a working mode.
Enabling data entering – balance user can enter a value of a parameter, e.g. set date, time, user no, strings (texts) in a printout.
- 13 -
Preview of balance menu – graphic presentation
While in the weighing mode press SETUP key. The display opens balance’s main menu (display I). Press UP or DOWN navigating arrows on the balance’s overlay to move the cursor upwards or downwards in the menu content. Place the cursor next to a menu option to be previewed. Press RIGHT ARROW navigating key on balance’s overlay to open the submenu content (display II).
Fig. 5. Preview of balance menu
1 – balance menu no. 2 – cursor for selecting a menu option 3 – menu content name 4 – name of selected menu option (setting) 5 – submenu number 6 – submenu name 7 – attribute set for a submenu option 8 – value (description) of an attribute set a submenu option
- 14 -
5.1. Moving through the menu
Moving in the user menu can be carried out using:
- Balance keyboard,
- External PC keyboard PS/2 type connected to balance’s socket,
- Commands sent from a connected computer to a balance
5.1.1.
Moving in the user menu using balance keyboard
Setup key. Entering balance’s main menu
Moving the cursor down in the menu list
Moving the cursor up in the menu list
Selecting submenu for activating. On pressing the key, the display
indicates the content of a selected group.
Exit to previous menu level, e.g. to main menu
Esc/TARE key. Abandon parameter changes
- 15 -
5.1.2. Return to weighing mode
Changes introduced in balance memory will be saved on returning to weighing with procedure of saving changes. Press ESC key for a few times until the display shows a question: Save? As displayed, select one of available options:
ENTER – save changes and go back to menu; ESC – abandon changes and go back to menu.
Fig. 6. Return to weighing mode
5.1.3.
Moving in the user menu using external computer keyboard PS/2 type
Al keys and buttons located on balance’s overlay have their equivalents on a computer keyboard PS/2 type. See below table for reference:
- equivalents of function keys
Description
Key on balance’s
overlay
ON/OFF key enables switching on and off balance’s display
Function key for entering the main menu of a balance
Selecting balance’s working mode, e.g.: animal weighing
Selecting measuring unit
PRINT key
TARE key
- 16 -
- equivalents of navigating arrows
Moving the cursor up in the menu list
Exit to higher level in menu structure , e.g. to main menu
Entering settings of a selected submenu.
Moving the cursor down in the menu list
- equivalents of ENTER / PRINT key and ESC key
Accepting entered value of a parameter
Abandon parameter changes and exit to to main menu
5.1.4. Moving in the user menu using virtual keyboard, via RS 232
interface
Most of the functions controlled or set using balance’s overlay or an external computer keyboard PS/2 type can be carried out by a set of commands sent from a computer to a balance. The commands enable moving through user menu, setting parameters or controlling balance operation. The list of commands is provided at the end of this user manual.
5.2. User menu - content
Menu structure of a balance AS/X series is described in point 5 of this user manual. On order to enter software’s main menu, when in the main weighing window press Setup key. The display opens a list with main menu. By pressing up and down navigating arrows move the cursor and place it next to a submenu to be edited.
- 17 -
Fig. 7. Balance main menu – submenu select ion
In order to edit a submenu, press RIGHT ARROW key, which opens the content of selected submenu. When inside the submenu structure, the user can select an option to be edited (modified) by placing the cursor next to submenu name (use up and down navigating arrows). When the cursor is placed next to desired option press RIGHT ARROW key to enter submenu settings.
Balance reactions for above procedure:
- A specific process (e.g. balance adjustment) which is carried out in a submenu described as a function;
- Editing an attribute of a submenu (flickering digit of a submenu setting enables changing parameter value of entering a sequence of characters)
Fig. 8. Balance submenu – selection buttons
- 18 -
6. WEIGHING
Basic working conditions for obtaining reliable measurement results:
- Stable and constant temperature in a weighing room,
- Stable foundation of a balance,
- Selecting adequate balance settings adjusted to ambient conditions at a
workstation.
Before start of weighing process or in case of essential change of ambient conditions at a workstation (e.g. ambient temperature change at a workstation more than 1°C/h) the balance requires adjusting. The procedure of balance adjustment is described in point 7.1. of this user manual.
Before start of weighing procedure, it is recommended to load the balance’s weighing pan a few times with mass close to balance max capacity. Check if unloaded balance indicates “precise zero” (the pictogram is visible on in the upper left corner of balance’s display) (and only if parameter P4 06 Autozero is set to 1: yes) and whether measurement is stable – (the pictogram is visible in the upper right corner of the display. If the mass indication is other than zero, press zeroing key:
If the working conditions are unfavourable (i.e. unstable measurement result), then the display previews dashes (horizontal lines). After exceeding a preset amount of time for zeroing the indication, the balance returns to weighing mode without zeroing the indication. In such case the user should wait for stabilization of working conditions and once again press Esc key.
Press Units key to set a measuring unit. Place weighed object on balance’s weighing pan and read the result only on stabilization of the measurement. If the measuring unit is not displayed on pressing the Units key, then go to the corresponding submenu and check the accessibility attribute of the measuring unit.
Mass indication of a load placed on balance’s weighing pan can be zeroed for multiple times. Pay attention not to exceed maximal capacity of a balance by applying multiple zeroing function.
During times between carrying out the following measurement series do not unplug the balance from mains. It is recommended to switch off balance’s display by pressing ON/OFF key. On repeated pressing of the ON/OFF key the balance is ready for operation and does not require thermal stabilization.
Recommended balance settings while weighing small mass ( ≤0,6g) and with reference to ambient conditions at a workstation:
- filtering level AuE: the slowest
- value release ConF: reliable
- 19 -
6.1. User log in
The users of a balance AS/X series can have their specific access code to the balance’s menu. The password system is determined by balance’s administrator, i.e. a user of the higher order in relation to the other balance users. The access password can comprise up to 6 digits.
Balance software enables determining:
A single Administrator, who has access to all balance settings and software
functions, including changing the password of the administrator and other users.
A single User who is authorized to access balance settings and functions, as set
by the balance administrator
Setting passwords and access levels
Remember, that after the first entering the passwor d sett ing s (see parameter P9
13 Password protection), the user should set a password for the balance Administrator.
The software requires inserting an administrator password when entering
parameter P9 13 Password protection only if the administrator password is other than “0”.
On the following entering this parameter, the softw are wi ll require inserting the
administrator’s password. Access to parameter settings of submenu P9 13 Password protection will be granted only on entering correct password. Unless the password is correct, the balance displays a message on incorrect password and goes back to displaying previous screen.
Depending on the settings, the inserted password is previewed either as
a sequence of digits or as asterisks (the initial value of each entered digit is always = 0).
Following the point 5.1.1 menu P9 Other contains the following options
Fig. 9. Password – fu nct ion ac tivat ing
- 20 -
Fig. 9-1. Password protection – menu content
- Administrator
the field for inserting administrator’s password. Balance administrator has access to all balance functions and settings.
- User
the field for inserting user’s password. Balance user has access to the functions and settings which attribute is set to NO (i.e. no password protection).
- Startup
if the option is set to YES, then on balance startup the software requires entering a password (of the administrator or user).
- Functions
if the option is set to NO (i.e. no password protection), balance user can use all working modes implemented in the balance software.
- Settings
if the option is set to NO (i.e. no password protection), balance user can change balance settings.
- Cal + GLP only
if the option is set to YES, then balance user has access only to carrying out balance adjustment /calibration and generating a report from adjustment process.
- Asterisk
if the option is set to YES , then on balance startup the entered password is previewed in a form of asterisks.
Inserting Administrator’s password
Enter a password for the balance administrator (a sequence of 6 digits) and for the user. Balance administrator has full access to balance menu. The user access is limited to the one described in previous point (balance menu, startup, adjustment, etc. options can be attributed YES/NO). It is very important to remember the password, as if option “password on startup” is enabled (set to YES), then on the following plugging the balance to mains, the software will request for entering the password. If it is entered incorrectly, then balance operating will be blocked.
While entering the password use balance keys presented on Fig. 2 or use external keyboard PS/2 type connectable to balance’s socket. Set availability to other balance settings and functions, depending on access level required by the balance user.
- 21 -
Fig. 10. Keys on balance overlay – entering values in editing fields of balance menu
- 22 -
7. ADJUSTMENT
In order to ensure the highest measuring accuracy, it is recommended to periodically introduce to balance memory a corrective factor of indications in relation to a mass standard – i.e. balance adjustment.
Adjustment should be carried out:
- Before the beginning of a weighing procedure,
- If long breaks between following measuring series occur
- If temperature inside the balance changes more than: 0,8°C
Types of adjustment
- Internal automatic adjustment
* triggered by temperature change * triggered by elapsing time
- Manual internal adjustment
* initiated from balance’s keyboard
- Adjustment with external weight
* with declared mass which cannot be modified * with optional mass which needs to be specified before process initiation
(only in non-verified balances)
Caution:
In case of verified balances only the automatic internal adjustment and manual internal adjustment systems are available for a user.
Remember to Perform the adjustment when there is no load on the pan!
7.1. Internal automatic adjustment
Activation of automatic internal adjustment is triggered on:
- Elapsing a specified amount of time from last carried out adjustment
process, or
- Ambient temperature changes by a value specified by the balance
manufacturer.
* in case of balance AS/X series it is 0,8°C, On recognizing any of the above case, balance’s display shows the following message box.
Fig. 11. Automatic internal adjustment – display content
- 23 -
The time delay enables the user to take the weighed load of the weighing pan, if a weighing process is in progress. Pressing T/O key causes temporary delay of the adjustment process initiation.
Automatic adjustment settings
Fig. 12. Settings of automatic balance adjustment
1 – main menu number 2 – marker of selected function 3 – function name 4 – name of an active function / mode / process 5 – selecting factor triggering auto-adjustment process (time / temperature) 6 – determining time interval between the following auto-adjustment processes 7 – value of set auto-adjustment triggering factors 8 – value of set time interval between the following automatic adjustment processes
Changing the values of automatic adjustment triggering factor and automatic adjustment time causes changes in description of the above fields (fields in fig no. 9 and 10).
01 Internal automatic adjustment
Initiates internal automatic adjustment process, which is carried out fully automatically with no operator’s activity. If balance’s weighing pan is loaded, then the balance displays a command ordering unloading it.
02 External adjustment
Adjustment process carried out with an external weight, which value is saved in balance’s memory. The function is disabled in verified balances.
03 User adjustment
Adjustment process carried out with an optional weight. Mass of the weight is specified before process initiation. The function is disabled in verified balances.
- 24 -
04 Adjustment test
mass comparison of internal adjustment weight with its value saved in balance memory.
05 Internal weight correction
The function enables correcting the value of internal adjustment weight. The function is disabled in verified balances.
06 Automatic adjustment
Determination of factors triggering start of automatic internal adjustment: 0 no – none of factors will cause start of adjustment 1 time – adjustment triggered by time interval set in point 07 2 temperature – adjustment triggered by change of temperature 3 both – adjustment triggered by time and temperature
07 Time of automatic adjustment
Determination of time interval, after which automatic adjustment process is initiated.
Return to weighing mode
Changes introduced in balance memory will be saved on returning to weighing with procedure of saving changes. Press ESC key for a few times until the display shows a question: Save? As displayed, select one of available options: ENTER – save changes and go back to menu; ESC – abandon changes and go back to menu.
(see Fig. 9. Return to weighing mode. point. 5.1.2. Return to weighing mode).
7.2. Adjustment test
Adjustment test is a comparison of internal adjustment weight with its value stored in balance’s memory. This process is carried out automatically and its result is shown on the display.
Fig. 13. Adjustment test
Cal. – the value of internal adjustment weight Curr. – weighing result of the internal adjustment weight Dev. – deviation calculated between the both values
- 25 -
Return to weighing mode
Changes introduced in balance memory will be saved on returning to weighing with procedure of saving changes. Press ESC key for a few times until the display shows a question: Save? As displayed, select one of available options: ENTER – save changes and go back to menu; ESC – abandon changes and go back to menu.
(see Fig. 9. Return to weighing mode. point. 5.1.2. Return to weighing mode).)
7.3. Manual adjustment
7.3.1.
Internal adjustment
1. Go to submenu P1 – Adjustment.
2. Place the marker next to a function 01 Internal adjustment.
3. Press RIGHT ARROW KEY.
4. The balance automatically carries out internal adjustment process. While
adjustment process is in progress do not load the weighing pan with any weight.
5. On completing the internal adjustment procedure the balance saves adjustment
data in its memory and returns to weighing mode.
Caution:
- In order to abort adjustment process press ESC key.
- If during internal adjustment process the weighing pan is loaded, then the balance
displays an error message. The adjustment process is automatically stopped. On taking off the load from the weighing pan, the process is resumed and completed.
- I f the DRH function is enabled in balance sett ing s, then bala n ce user can not abor t the adjustment process once initiated.
7.3.2.
External adjustment
The external adjustment in balances AS/X series should be carried out with an external mass standard / weight class: E
2
.
List of weights / mass standards required for adjusting balances is specified in the technical data provided at the end of this user manual.
1. Go to menu P1 – Adjustment.
2. Set the marker next to a function 02 External adjustment.
3. Press RIGHT ARROW KEY.
4. The software displays a command to take off any load from the weighing pan (the weighing pan must by empty). On unloading the weighing pan, press ENTER key.
5. The balance determines mass of an empty weighing pan.
6. Load a weight / mass standard which mass is given on the display and press ENTER key.
- 26 -
7. On completing adjustment process the balance returns to displaying submenu P1
- Adjustment
8. Return to weighing mode – in accordance with point 5.1.2.
If the DRH function is enabled in balance settings, then external adjustment process is disabled. The DRH function is enabled in verified balances (which are subject to conformity assessment).
7.3.3.
User adjustment
The external adjustment in balances AS/X series should be carried out with an optional mass standard / weight class: E
2
.
- Go to menu P1 – Adjustment and set the marker next to a function 03 User
adjustment.
- Press RIGHT ARROW KEY.
- The balance displays a command to enter mass of an adjustment weight. The first
digit of the weight value is flicering, and it is ready for editing.
Fig. 14. User adjustment – declaring weig ht valu e
- Use function kwys (as specified in point 5.1.1 of this user manual) to enter the value
of the external weight/mass standard.
- Accept the weight’s value as entered. The balance initiates adjustment process by
indicating process commands on the display.
- The balances shows a command on determining mass of the empty weighing pan,
which is followed by a command to place a weighed with pre-determined mass.
- On placing the determined weight on the weighing pan accept it by pressing Enter
key.
- On completing adjustment process the balance returns to displaying submenu P1 -
Adjustment.
- Return to weighing mode – in accordance with point 5.1.2.
It is recommended that the mass of an external adjustment weight is approximately ¾ of balance’s maximum capacity.
If the DRH function is enabled in balance settings, then external adjustment process is disabled.
- 27 -
7.4. Adjustment report printout
On completion of any type of adjustment process, the balance enables preparing a report from adjustment process. The report can be printed on a connected printer and sent to a computer and saved in a form of file for records.
P1 08 Report printout : 1: yes – report printout enabled P1 08 Report printout : 0: no – report printout disabled
Remember, that if the parameter is set for YES, then a report is generated and sent automatically.
Fig. 15. Submenu: Adjustment
The content of the report from adjustment process depends on settings of GLP parameters. Any option in the GLP submenu which attribute is YES is included in a report from adjustment process.
Fig. 16. GLP submenu - settings
- 28 -
Apart from information set in menu group the report contains: Mass of adjustment weight stored in balance’s memory from last carried out adjustment (description: Old:) Mass of adjustment weight determined in current adjustment process (description: Adjustment:) Adjustment deviation, i.e. difference between the two mass records (description: Deviation:).
Fig 17. An example of a report from adjustment process
8. DETERMINING CONTENT OF A PRINTOUT FOR GLP PROCEDURES
Menu P2 GLP is group of the parameters which enables declaring variables that are present on a printout from adjustment process. Fields referring to:
- user (max 8 alphanumeric characters)
- project (max 8 alphanumeric characters)
are editable are enable entering a text using balance’s keyboard or connectable external computer keyboard PS/2 type. The other fields listed in the GLP are set:
- 1 yes (print on a report)
- 0 no (do not print on a report)
GLP submenu is presented in figure 16. If a user applies an external keyboard PS/2 type, then the relation between the balance’s keyboard and the computer keyboard are as specified in point 5.1.3 of this user manual.
Loading...
+ 64 hidden pages