RADWAG WPT-8B 300C User Manual

User manual
Medical scales
User manual no.:
ITKU-16-08-10-13-EN
BALANCES AND SCALES
RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com www.radwag.com
- 2 -
27 August 2012
- 3 -
Table of contents
1. INTENDED USE ................................................................................................................ 5
2. PRECAUTIONATY MEASURES ....................................................................................... 5
2.1. Maintenance .............................................................................................................. 5
2.2. Battery pack ............................................................................................................... 5
2.2.1. Power supply of scales featuring an indicator in plastic housing ......................... 6
2.2.2. Replacement of used batteries ........................................................................... 7
3. WARRANTY CONDITIONS ............................................................................................... 8
4. MAIN D IMENSIONS .......................................................................................................... 9
5. UNPACKING AND ASSEMBLY ...................................................................................... 12
5.1. Personal scale ......................................................................................................... 12
5.2. Bed scale WPT/4B series ......................................................................................... 13
6. CLEANING ...................................................................................................................... 15
7. START UP ....................................................................................................................... 16
8. KEYBOARD .................................................................................................................... 16
9. KEYS AND FUNCTIONS ................................................................................................. 17
10. DI SPLAY INDICATIONS ................................................................................................ 17
11. USER MENU .................................................................................................................. 18
11.1. Sc al e’s paramet er groups....................................................................................... 18
11.2. Moving through user m enu ..................................................................................... 18
11.2.1. Keyboard ........................................................................................................ 18
11.2.2. Return to weighing mode ................................................................................ 19
12. WEIGHING..................................................................................................................... 19
12.1. Tarring f unct i on ...................................................................................................... 19
12.2. Manual tare insertion .............................................................................................. 20
12.3. Zeroing f unct i on ..................................................................................................... 21
12.4. Weighing on dual range scales............................................................................... 21
12.5. Selection of basic measuring unit ........................................................................... 22
12.6. Select i ng an instant aneous m easuring unit ............................................................ 23
13. MAIN SCALE PARAMETERS ....................................................................................... 24
13.1. Setting filtering level ............................................................................................... 24
13.2. Median fil ter ........................................................................................................... 25
13.3. Aut ozero f unct i on ................................................................................................... 26
14. RS 232 PARAMETERS ................................................................................................. 27
14.1. Determining minimum mass for function operation ................................................. 27
14.2. Baud rat e ............................................................................................................... 28
14.3. Serial transmission parameters .............................................................................. 29
15. OTHER PARAMETRS ................................................................................................... 30
15.1. Displ ay back l i ght .................................................................................................... 30
15.1.1. Display backlight for supply from mains .......................................................... 30
15.1.2. Display backlight for supply from batteries or accumulators ............................ 31
15.2. “beep” sound – sc al e reaction to pressi ng a key ..................................................... 32
15.3. Scale automatic switch off ...................................................................................... 32
15.4. Charging and disc hargi ng bat teries / accumulators ................................................ 34
15.4.1. Battery / accumulator voltage level check ....................................................... 34
15.4.2. Operation of battery / accumulator charge / discharge indication .................... 34
15.4.3. Accumulator charging options ......................................................................... 35
15.4.4. Accumulator charging process ........................................................................ 36
16. WORKING MODES ....................................................................................................... 37
- 4 -
16.1. Means of select i ng a working mode........................................................................ 37
16.2. Weighing with lock of measurement result on the display - hoLd ............................ 37
16.3. Determining BMI factor ........................................................................................... 39
17. SCALE ADJUSTMENT .................................................................................................. 41
17.1. Adjustment ............................................................................................................. 41
17.2. Start mass adjustment ............................................................................................ 43
18. COOPERATION WITH A PRINTER ............................................................................... 44
19. COOP ER ATION WITH A COMPUTER .......................................................................... 45
20. COM M UNICATION PROTOCOL ................................................................................... 46
20.1. General information ................................................................................................ 46
20.2. A set of commands recognized by the indic at or ...................................................... 47
20.3. Responses format for commands sent from computer level ................................... 47
20.4. Command’ s descri ption .......................................................................................... 48
20.4.1. Zero scale ....................................................................................................... 48
20.4.2. Tare scale ....................................................................................................... 48
20.4.3. Give tare value................................................................................................ 48
20.4.4. Send stable measurement result in basic measuring unit ................................ 49
20.4.5. Immediately send measurement result in basic measuring unit ....................... 49
20.4.6. Send stable measurement result in current weighing unit................................ 50
20.4.7. Immediately send measurement result in current measuring unit .................... 51
20.4.8. Switch on continuous transmission in basic measuring unit ............................ 51
20.4.9. Switch off continuous transmission in basic measuring unit ............................ 52
20.4.10. Switch on continuous transmission in current measuring unit ........................ 52
20.4.11. Switch off continuous transmission in current measuring unit ........................ 52
20.4.12. Send all implemented commands ................................................................. 52
20.5. Manual printout ...................................................................................................... 53
20.6. Continuous transmission ........................................................................................ 54
20.7. Printout configuration ............................................................................................. 54
21. ERROR MESSAGES ..................................................................................................... 54
22. SCALE TECHNICAL PARAMETERS ............................................................................ 55
22.1. Chair scale ............................................................................................................. 55
22.2. Personal scales ...................................................................................................... 55
22.3. Bed scales ............................................................................................................. 56
22.4. Baby scal es ............................................................................................................ 56
23. TROUBLE SHOOTING .................................................................................................. 57
24. ADDI TIO NAL EQUIPMENT ........................................................................................... 57
25. EMC INFORMATION ..................................................................................................... 58
26. APPENDIX A ................................................................................................................. 62
26.1. Pict ograms on the data plat e .................................................................................. 62
26.2. CE marking ............................................................................................................ 62
- 5 -

1. INTENDED USE

RADWAG medical scales are designed to measuring mass of weighed patients. In case of personal scales, the mass determining process is carried out when a patient stands on the scale’s weighing platform, on chair scales – when a patient is sitting, and on bed scales a patient is weighed while lying on a bed. Baby scales enable determining body mass of a baby who is lying on scale’s platform. Each of medical scales enables tarring in the whole measuring range.
Functions:
Display backlight,
Digital filters (adjusting scale’s operation to ambient and operating
conditions of the place of use),
Median filter,
Autozero function,
Manual setting of tare value,
Baud rate settings,
Determining minimum mass for operating of automatic functions,
Automatic scale’s switch off,
User adjustment (available in non-verified scales),
Peak hold – weighing wit h locking the measurement result on the
display - HOLD,
Determining BMI factor.

2. PRECAUTIONATY MEASURES

2.1. Maintenance

A. Read the user manual carefully before commissioning and using the
scale according to its intended use;
B. Instruments to be decommissioned should be decommissioned in
accordance with valid legal regulat ions.

2.2. Battery pack

Medical scales featuring an indicator PUE C/31 series (plastic housing) are designed for buffer powering by NiMH batteries / accumulators) with rated voltage 1,2V, size R6 (AA) and capacity from 1800 to 2800mAh.
- 6 -
Indicators plugged to mains automatically monitor and maintain batteries / accumulators charging status.
In case of scale’s / indicator’s elongated storage in low temperatures, it is not allowed to fully discharge the accompanied batteries.
The equipment including accumulators does not belong to your regular household waste. The Europe an legislation requires that electric and electronic equipment be collected and disposed separately from other communal waste with the aim of being recycled.
Hint: Some symbols on accumulators / batteries identify harmful compounds:
Pb = lead, Cd = cadmium, Hg = mercury.

2.2.1. Power supply of scales featuring an in d icator in plastic housing

New NiMH accumulators, which come standard with the indicator PUE C/31 series in plastic housing require forming in accordance with the procedure contained in point 15.4.4 of this user manual.
Alternatively, the indicator can be powered by standard R6 / AA batteries. Should an indicator be powered from standard batteries, not the NiMH accumulators, then:
Before installing the batteries inside the indicator’s housing, power the indicator from mains only, and go to menu parameter <5.5.CHr6> and change its value to <no>,
On completing the above step it is possible to assemble standard batteries to indicator’s batteries compartment.
- 7 -
Installing standard batteries without setting the parameter <5.5.CHr6> to value <no> may, on powering the indicator from mains, damage the batteries and the indicator.
Powering the scale from mains is permissible using only the power adapter that comes standard with the scale.

2.2.2. Replacement of used batteries

Scales featuring indicator PUE C/31 series (plastic housing) enable replacing used batteries with new ones .
Procedure:
Open the lid of the battery compartment located in the bottom section of indicator’s housing:
Remove used (discharged) accumulators / batteries and replace them with new ones, pay attention to correct polarity (+/-)marked in the battery compartment:
- 8 -
Close the battery compartment’s lid:

3. WARRANTY CONDITIONS

A. RADWAG feels obliged to repair or exchange all elements that
appear to be faulty by production or their construction,
B. Defining defects of unclear origin and means of their elimination can
only be realized with assistance of manufacturer and user representatives,
C. RADWAG does not bear any responsibility for defects, losses or
accidents resulting from unauthorized or inadequate performing of production or service processes,
D. Warranty does not cover:
Mechanical defects caused by product exploitation other than intended, defects of thermal and chemical origin, defects caused by lightning, overvoltage in the power network or other random event,
Maintenance activities (cleaning).
E. Loss of warranty takes place if:
A repair is carried out outside RADWAG sales office or authorized service point,
Service claims intrusion into mechanical or electronic construction by unauthorized people,
The scale does not bear company’s protective stickers.
F. Accumulators are covered by 6 months warranty period. G. Detailed warranty conditions are listed on a service card.
- 9 -

4. MAIN DIMENSIONS

Personal scale WPT xx OW series
Personal scale WPT xx O series
A. Personal scales are available in two versions
– with a height meter (indicator is attached to the height meter); – without a height meter (indicator is attached to the scale’s weighing platform).
- 10 -
B. Chair scale features a moveable handle and wheels aiding scale’s
transport to intended place of use.
Chair scale – dimensions
C. Bed scale WPT/4B series features four measuring blocks, which are
placed under the wheels of a bed using a dedicated hoist. The design of the WPT/4B series enables its using independently on bed type (beds with wheel diameter from φ100 to φ200) and applied brake system.
Bed scale WPT/4B series – dimensions
- 11 -
D. Bed scale WPT/8B series comprises two weighing beams, which are
unfolded maximall y to 2,5 m eters . A gentle entrance onto the beam enables easy locating the bed onto the scale’s beams. On positioning onto the beams, the bed is weighed. The design of bed scale WPT/8B series enables its using independently on bed type and applied brake system.
Bed scale WPT/8B series – dimensions
E. Baby scales WPT 6/15D and WPT 10/20D are precise m eas uring
instruments designed in the accuracy class III, and dedicated to weighing infants. Weighing process can be carried out with indication lock on stabilization of measurement result, i.e. peak hold function. In such case, infant’s mass can be precisely read independently on baby’s movement on scale’s weighing platform.
Baby scale – dimensions
- 12 -

5. UNPACKING AND ASSEMBLY

5.1. Personal scale

A. Carefully remove the personal scale from its packaging, B. In order to ensure precise indications, place the scale on a firm
foundation, distant from heat and cold sources,
In case of bed scales, place the beams on a firm, stable and even foundation, and unfold the beams to adjust them to a type of bed that is weighed,
In case of personal scales with a height meter, assembly the height meter to scale’s base using an allen key that comes standard with the scale,
Assembly of persona l sca le WP T xx OW series
- 13 -

5.2. Bed scale WPT/4B series

A. Carefully remove the bed scale from its packaging
General view
B. Assembly the scale’s components:
Screw down the indicator,
Push in the castors,
C. Move the assembly to its operation place and place it next to a bed to
be weighed, D. Lock the castors’ brakes to set the trolley in a determined position, E. Lock the castors of the bed, F. Move one of the measuring blocks of the WPT/4B scale to one of the
bed’s wheels, G. Assemble the hoist:
- 14 -
Step 1
Step 2
Step 3
Step 4
H. Use the hoist to place bed’s wheel in the scale’s measuring block.
See below figures:
I. Apply the same procedure to the following bed’s wheels until all of the
bed’s wheels are located on the measuring blocks, as shown below:
- 15 -
Caution:
The foundation under the measuri ng block s should b e even and leveled. Additionally, make sure, that the connecting cable is not clenched under the measuring block, as it may be source of incorrect measurements.

6. CLEANING

Scales intended for weighing infants The scale requires periodical cleaning of the weighing platform. Clean the
weighing platform using disinfectants and cleaning agents dedicated for medical equipment. While cleaning avoid pressing hard the weighing platform as it may damage the scale’s measuring system.
Chair scales The scale requires periodical cleaning of the seat and the arms. Clean the
scale parts using disinfectants and cleaning agents dedicated for medical equipment. While cleaning avoid turning over the scale or carrying out other activities which may result in damaging the scale .
Bed scales Clean the measuring blocks and measuring beams (depending on scale
type) using disinfectants and cleaning agents dedicated for medical equipment.
Personal scales Clean the weighing platform and scale’s body using disinfectants and
cleaning agents dedicated for medical equipment.
Caution:
Do not drench the indicator of medical scales in any way (use dry cloth or alternatively a damp cloth to clean the indicator).
- 16 -

7. START UP

First unpack and assemble the scale (see point 5 of this user manual). Then:
Level the scale using its adjustable feet located in the bottom base of the medical scale. Monitor the level status on the level located in the base of the medical scale. Leveling is correct if the air bubble of the level indicator is located in the central location:
Switch on the scale by pressing 0,5 second,
On turning on wait until the scale completes the auto test procedure,
key – press and hold the key for
Then the zero mass indication appears on the display, with the following pictogram s:
- precise zero indication
- stable measurement result
kg - measuring unit
If measurement result is other than zero, press zeroin g ke y on indicator’s overlay.

8. KEYBOARD

- 17 -
No.
Indication
Description
1.
Fil
2.
bAud
3.
Auto
4.
t1
9.
Net
10.
hold

9. KEYS AND FUNCTIONS

Switching on/off scale’s display – press and hold the key for about 1 second
Function key (working mode selection)
Sending measurement result to a connected printer or computer
Zeroing
Tarring
Caution: Simultaneous pressing of
and keys changes functionality of the keys for the time of function programming. Means of their use is described further in this user manual.

10. DISPLAY INDICATIONS

Filtering level of weighing result Baud rate Control and correcting scale’s zero indicat ion Power save mode – time defined scale’s switch off
5.
6.
7. kg (g) Scale in working mode weighing
8.
11. bmi
Indication in autozero mode (indication = zero) Stable measurement result (ready to read)
Discharged batteries / accumulators / defected accumulator or defected power adapter
Tare function applied Weighing with peak hold function applied Determining BMI factor
- 18 -
P1 rEAd
P 1.1 Fil
|
4
P 1.2 Auto
|
YES
P 1.4 Fnnd
|
no
P2 Prnt
P2.2 S_Lo
|
P2.3 bAud
|
9600
P2.4 S_rS
|
8d1SnP
P3 Unit
P3.1 StUn
|
kg
P5 othr
P5.1 bL
|
Auto
P5.2 bLbt
|
70
P5.3 bEEP
|
YES
P5.4 t1
|
Auto
P5.5 CHr6
|
YES
P6 CAL
P6.1 St_u
|
* FUNCTION *
P6.2 uCAL
|
* FUNCTION *
+
Entering main menu
Manual tare inserting in working m ode: weigh in g
+
Check battery / accumulator status
+
Toggling between gross / net values

11. USER MENU

11.1. Scale’s parameter groups

Scale menu is divided into 5 basic submenus. Each group has its own name preceded by a capital letter P:

11.2. Moving through user menu

The user moves in the menu by pressing keys on indicator’s overlay.

11.2.1. Keyboard

+
Change of digit value by “1” up Scrolling the menu upwards
- 19 -
Selecting a parameter on one menu level
Entering a selected submenu
Accepting carried out changes
Leaving a function without saving changes
Changing the value of an active parameter
Activating a parameter for modification
Moving one level “up” in menu structure

11.2.2. Return to weighing mode

The changes made in scale’s memory are permanently saved on returning to weighing mode with procedure of saving carried out changes.
Press
key for a few times until the display indicates
message <SAuE?>. Then press :
– to save changes or
– to abandon changes. On pressing one of the above keys, the scale automatically returns to weighing mode.

12. WEIGHING

For patients moving on the weighing platform it is recommended to set higher filtering level (see point 13.1 of this user manual). In such case the weighing time is longer, but the obtained measurement result is reliable.

12.1. Tarring function

Place a nappy or a blanket on the scale’s weighing platform.
After indication stabilizing, signaled by accompanying pictograms
press
key (mass indication returns to zero, and Net pictogram is
lit up in the upper left corner of the display).
- 20 -
Place a weighed baby / infant on scale’s weighing platform (baby
scale)
Stand on scale’s weighing platform (personal scale)
Sit on the chair (of a chair scale)
Lie on the bed (bed scale)
After a few seconds, after measurement stabilization, it is possible to read the result.
Caution:
A. Baby scale is designed to weighing infants in lying position. If it is
used to weight infants / babies in sitting position, pay special attention to place the infant / baby in the center of weighing platform. In other case, the weighing platform may rest on the protective bumpers, thus causing incorrect and unreliable measurement results.
B. While weighing babies in sitting position pay special attention that
the baby does not loose balance and falls off the weighing platform.
C. Tarring is impossible if the display indicates negative or zero mass
value. In such case the scale signals an error message Err3 and emits a short beep sound.

12.2. Manual tare insertion

The sale enables manual determining of tare value. While in weighing mode:
Simultaneously press
and keys
The display indicates the following:
- 21 -
Use and keys to set tare value,
Press key,
The scale returns to weighing mode and the display indicates value
of inserted tare with ”–” sign
Tare value can be inserted in the optional moment of weighing process.

12.3. Zeroing function

Zeroing the indication is carried out by pressing key, which is followed by zero mass indication and appearing pictograms:
and .
Zeroing is equal to determining new zero point, which is recognized by the scale as a precise zero. Zeroing option is available only with stable measurement status.
Caution:
Zeroing of display indication is possible only within the ±2% range from scale’s maximal capacity. If zeroed value is more than the ±2% of scale’s maximal capacity the display indicates <Err2> message and short beep sound is emitted.

12.4. Weighing on dual range scales Transition from weighing in the I weighing range to weighing in the II

weighing range takes place automatically without operator’s interference
(i.e. on exceeding the Max of the I weighing range). Weighing in the II weighing range is signaled by displaying a
corresponding pictogram in the upper left corner of the display. On unloading, the scale’s indication returns to zero. Weighing is carried out
with the accuracy of II weighing range until the indication returns to zero.
Return from weighing in the II weighing range to weighing in the I weighing range takes place automatically on unloading the weighing platform and
- 22 -
indication returning to AUTOZERO zone – signaled by pictogram on the display.
As the indication returns to zero, the pictogram of the II weighing range is switched off, and the scale measures mass with the accuracy of the I
weighing range.

12.5. Selection of basic measuring unit

The scale enables setting a measuring unit which is activated on scale’s switch on.
Procedure:
Enter submenu <P3.Unit> in accordance with point 11.2 of this user manual, and:
Press
key for a few times, and the display indicates available
measuring units in sequence:
Available settings:
A. If the main measuring unit is [kg], the user can select between the
following unit s : [kg, lb, N] in case of verified scales [lb] unit is disabled,
B. If the main measuring unit is [g], the user can select between the
following unit s : [g, ct, lb] in case of verified scales [lb] unit is disabled.
- 23 -
On selecting a measuring unit press key, the scale returns to displaying a window:
Return to weighing with procedure o f saving changes:
See point - 11.2.2. – return to weighing.
Caution:
On setting a new basic measuring unit the scale will switch on with the new measuring unit set as default one.

12.6. Selecting an instantaneous measuring unit

The function enables setting an instantaneous measuring unit used for indicating mass. The measuring is active from the moment of its selection until it is changed or the scale is switched off and on.
Procedure:
Press
key, and:
- 24 -
On accepting a selected measuring unit, go back to weighing mode with new measuring unit.
Available settings:
A. If the main measuring unit is [kg], the user can select between the
following units: [kg, lb – not available in verified scales, N],
B. If the main measuring unit is [g], the user can select between the
following units: [g, ct, lb – not available in verified scales

13. MAIN SCALE PARAMETERS

The scale enables adjusting its operation to external and ambient conditions (filtering level) or user needs (autozero operation, tare value memory). These parameters are grouped in menu <P1.rEAd>.

13.1. Setting filtering level Procedure:

Enter submenu <P1.rEAd> in accordance w ith point 11.2 of this user manual, and:
- 25 -
1 - 4 - filtering level set in relation to current ambient and operating conditions
Return to weighing:
See point - 11.2.2. – return to weighing.
Caution:
The higher filtering level, the longer stabilization time of measurement result.

13.2. Median filter

The median filter is to eliminate short impulse disturbances (like mechanical shocks).
Procedure:
Enter submenu <P1.rEAd> in accordance with point 11.2 of this user manual, and:
Fnnd no - median filter operation disabled Fnnd YES - median filter operation enabled
Return to weighing:
See point - 11.2.2. – return to weighing.
- 26 -

13.3. Autozero function

The AUTOZERO function is implemented to assure precise scale’s indications. The function automatically controls and corrects scale’s zero indication. When the function is enabled, the following measurement results are compared in constant time intervals. If the following values differ less than declared AUTOZERO range, e.g. 1 division, then the scale automatically zeroes its indication and displays pictograms of stable measurement result
and precise zero – .
– When the AUTOZERO function is enabled, each measurement result starts from precise zero point. There are, however, cases when autozero function may disturb weighing process, like while very slow loading / pouring weighed object on scale’s weighing platform. In such case, the zero correcting system can also correct the actual mass indication of weighed object.
Procedure:
Enter submenu <P1.rEAd> in accordance with point 11.2 of this user manual, and:
AUTO no - autozero disabled AUTO YES - autozero enabled
Return to weighing:
See point - 11.2.2. – return to weighing.
- 27 -

14. RS 232 PARAMETERS

14.1. Determining minimum mass for function operation

Scale software enables weighing with peak hold function (HOLD) and determining BMI* (Body Mass Index) factor of weighed patients. The functions are enabled if the measurement result is above a value of minimum mass declared for function operation S_Lo. If the measurement result is below the declared minimum mass value, then:
The display shows a message -Lo- (in BMI function).
The measurement result is not locked (peak hold (HOLD) option
disabled)
* BMI mode is not available in baby scales.
Procedure:
Enter submenu <P2.Prnt> in accordance with point 11.2 of this user manual, and:
Return to weighing:
See point - 11.2.2. – return to weighing.
- 28 -

14.2. Baud rate Procedure:

Enter submenu <P2.Prnt> in accordance with point 11.2 of this user manual, and:
Return to weighing:
See point - 11.2.2. – return to weighing.
- 29 -

14.3. Serial transmission parameters

Enter submenu <P2.Prnt> in accordance with point 11.2 of this user manual, and:
7d2SnP - 7 data bits; 2 stop bits, no parity control 7d1SEP - 7 data bits; 1 stop bit, EVEN parity control 7d1SoP - 7 data bits; 1 stop bit, ODD parity control 8d1SnP - 8 data bits; 1 stop bit, no parity control 8d2SnP - 8 data bits; 2 stop bits, no parity control 8d1SEP - 8 data bits; 1 stop bit, EVEN parity control 8d1SoP - 8 data bits; 1 stop bit, ODD parity control
Return to weighing:
See point - 11.2.2. – return to weighing.
- 30 -
bL
no
-
Backlight disabled
bL
YES
-
Backlight enabled
bL
Auto
-
Backlight in automatic mode, i.e. switched off if seconds

15. OTHER PARAMETRS

The scale enables setting parameters influencing its operation, like display backlight, “beep” sound, automatic switch off mode. These parameters are grouped in menu <P5.othr>,

15.1. Display backlight

The scale software automatically recognizes power supply mode (mains, accumulators) and adjusts display backlight mode:
bL – for power supply from mains
blbA – for power supply from batteries or accumulators

15.1.1. Display backlight for supply from mains Procedure:

Enter submenu <P5. othr> in accordance with po int 11.2 of the user manual, and:
display indication does not change for 10
- 31 -
Return to weighing:
See point - 11.2.2. – return to weighing.
Caution:
When bL=Auto, and the mass indication does not change for approximately 10 seconds, then the display backlight is automatically switched off. The backlight is automatically activated on mass indication change.

15.1.2. Display backlight for supply from batteries or accumulators

Depending on user needs (and visibility (operation of light) in the scale’s place of use) the user can adjust brightness of display backlight between 0 % and 100 %. In lower display brightness levels the operation time on batteries / accumulators in longer. For setting display brightness level, operation of display backlight is always set to Auto.
Procedure:
Enter submenu <P5. othr> in accordanc e with po int 11 .2 of the user manual, and:
Return to weighing:
See point - 11.2.2. – return to weighing.
- 32 -
Caution:
Enabled display backlight shortens scale’s operation time on batteries / accumulators.

15.2. “beep” sound – scale reaction to pressing a key Procedure:

Enter submenu <P5. othr> in accordanc e with po int 11 .2 of the user manual, and:
bEEP no - beep sound on pressing a key disabled bEEP YES - beep sound on pressing a key enabled
Return to weighing:
See point - 11.2.2. – return to weighing.

15.3. Scale automatic switch off

This function is essential to save energy while powering the scale from batteries / accumulators. If the t1 function is enabled, the scale switches off within 5 minutes, on condition there is no activity carried out on the scale (i.e. mass indication on the display does not change).
- 33 -
Function operation according to means of power supply:
Function operation
Function settings
t1 = 0 Disabled Disabled t1 = YES Enabled Enabled t1 = Auto * Disabled Enabled
Power supply
from mains
Power supply by
batteries/accumulators
* function disabled / enabled automatically according to means of scale’s power
supply.
Procedure:
Enter submenu <P5. othr> in accordance with po int 11 .2 of the user manual, and:
Return to weighing:
See point - 11.2.2. – return to weighing.
- 34 -

15.4. Charging and discharging batteries / accumulators

If during scale supplying from batteries / accumulators the software detects too low voltage value, then it is signaled by displaying
pictogram. It is a
message ordering the user to immediately replace the batteries with new ones or charge the accumulators. Blinking
pictogram denotes
accumulators’ charging process.

15.4.1. Battery / accumulator voltage level check

This function is used to checking the level of voltage in batteries / accumulators. The function operates only if:
The scale is in weighing mode,
The scale is supplied from batteries / accumulators.
Procedure:
While in weighing mode simultaneously press
and keys. The display shows message bAtt for approximately 1 second, which is followed by information on batteries / accumulators charging status in % visible for approximately 2 seconds.
On displaying battery / accumulator status, the scale automatically returns to weighing mode.

15.4.2. Operation of battery / accumulator charge / discharge indication

In case the scale is supplied by batteries / accumulators, and the software detects too low voltage level (voltage rate fall below 18 % of permissible
value), the display shows a pictogram
(bat low) which recommends
- 35 -
immediate replacing used batteries with new ones or charging the accumulators.
Low voltage signaling:
pictogram appears on the display,
If a scale operates with low battery pictogram visible on the display, after some time the software automatically switches off the scale to protect the accumulators from excessive discharge,
Accumulator charging process is signaled by blinking pictogram
(blinking interval: 2s) on the display.

15.4.3. Accumulator charging options

The function provides enabling or disabling charging option of NiMH accumulators.
a) Parameter <5.5.CHr6> is set to <no>:
Pictogram
is not displayed, charging option is disabled.
On scale start, the display shows message <bAtt>.
b) Parameter <5.5.CHr6> is set to <YES>:
Pictogram
is blinking while accumulators charging process
(blinking interval 2s), charging option is enabled.
On scale start the display shows message <nImh>,
In case the accumulators are defected or missing, the pictogram
is blinking quickly (with very short intervals ~ 0,5s).
Caution:
The indicator of a medical scale comes standard with a set of new NiMH accumulators, size AA (R6), and a power adapter.
Procedure:
Enter submenu <P5. othr> in accordance with point 11.2 of the user manual, and:
- 36 -
CHr6 YES - function enabled (pictogram appears on the
display)
CHr6 no - function disabled (pictogram does not appear on
the display)
Return to weighing:
See point - 11.2.2. – return to weighing.

15.4.4. Accumulator charging process The indicator of a medical scale comes standard with a set of new NiMH

accumulators, size AA (R6), and a power adapter. On scale’s commissioning it is obligatory to form the accumulators, which extends their life-span. The process requires full charging and complete discharging the accumulators. New accumulators have the optimum capacity after completing three charging / discharging cycles.
Forming process:
1. Plug the power adapter to scale’s indicator.
2. Charge the accumulators for approximately 12 hours – charging time specified for accumulators with capacity 2200mAh.
3. After 12 hours unplug the power adapter from the indicator.
4. Operate the scale with supply from accumulators until its automatic switches off.
- 37 -
5. After scale’s automatic switch off repeat the forming process twice by acting as specified in points 1 to 4 above. The forming process enables obtaining optimum capacity of the accumulators.

16. WORKING MODES

Medical scales, while in weighing mode have two working modes, which are enabled on pressing of
key:
Weighing with peak hold (lock of measurement result on the display (hoLd)
Determining BMI factor.

16.1. Means of selecting a working mode

On changing – selecting a working mode, the scale displays name of the working mode for approximately 1 second. The software stores data on recently used working mode, and it is displayed as the first one on next
pressing of
key.
Procedure:

16.2. Weighing with lock of measurement resul t o n th e display - hoLd

While in weighing mode press key and select working mode hoLd. On enabling the mode, the user should set its operation parameters. The display automatically previews the first of available options.
- 38 -
hoLd
Print
-
Scale operation with peak hold of measurement result
key;
hoLd
StAb
-
Scale operation with automatic locking of
key;
hoLd
OFF
-
Exit to weighing mode on pressing key.
enabled on pressing key, the lock is cancelled on taking the load off the weighing platform and pressing
measurement result, the lock is cancelled on taking the load off the weighing platform and pressing
On accepting selecting working mode by pressing software automatically moves to displaying mass indication window:
Software confirmation for enabling HoLd mode, is appearance of -OK­marker in the top section of the display.
key, the
- 39 -
Caution:
In case the mass indication is below set value of S_Lo parameter, then the measurement result will not be locked (HoLd function will not be activated).
Mode operation:
Load scale’s weighing platform. On stabilization of measurement result the display locks mass indication, as required by enabled HoLd mode.
Take the weighed load off scale’s weighing platf orm,
Before carrying out the following measurement press
key.

16.3. Determining BMI factor

The mode of determining body mass index BMI is dedicated for adults over 18 years old, therefore the software comprises the following limitations:
The mode is not available in baby scales (models WPT 6/15D and WPT 10/20D).
Body mass limitation for mode activation - min 10kg,
Height limitation for mode activation - from 1m to 2,5m.
Mode operation:
Load scale’s weighing platform,
Press function key
to select working mode bnni,
If measurement result is stable, the display indicates (for approximately 2 seconds) the mass value of weighed patient followed by blinking value of height expressed in m eter s ,
If measurement result is unstable, the display indicates dashes < - - - - - - - - - - > until stable measurement result is obtained. Then the display indicates (for approximately 2 seconds) the mass value of weighed patient followed by blinking value of height expressed in meters.
- 40 -
H
1.85 m
W
75,82 kg
BMI
22,2
2
H
W
BMI =
where:
W
-
body mass in [kg]
H -
height in [m]
BMI
-
factor’s value rounded to first decimal place.
Use indicator’s keyboard to type in the height of weighed patient, where:
- selects a digit, - selects digit value,
On typing in the height data, accept it by pressing key, and the display indicates the value of BMI factor:
The following pressing of
key causes printing the following data on
a connected printer:
Take of the load of the scale’s weighing platform,
Return to weighing mode by pressing
The BMI factor is determined using the following relation:
key.
,
Caution: In case the mass indication while entering bnni mode is below set value of S_Lo parameter, then the displays shows a message < -Lo- > and the
scale returns to weighing mode.
- 41 -

17. SC ALE ADJUST M ENT

Available in non-verified scales only
Ensuring high weighing accuracy requires periodical controlling scale’s indications and correcting the adjustment factor stored in scale’s memory – the process is known as scale adjustment. Adjustment should be carried out on scale start of operation or if dynamic change of temperature occurs. Before carrying out adjustment procedure remove load from the weighing platform.

17.1. Adjustment Procedure:

Enter submenu <P6.CAL> in accordance with po int 11 .2 of the user manual, and:
The software displays the following messages / commands:
During this time the scale adjusts its start mass, and after completing
the procedure, the mass of adjustment weight is displayed (e.g. 3.000kg).
- 42 -
Place the required weight on the weighing platform and press key. The adjustment process starts automatically, which is signaled by the displaying the following message:
The completion of the adjustment procedure is signaled by the following message:
Take off the adjustment weight from the weighing platform, after which message <donE> is displayed for approximately 1s and the scale returns to displaying name of submenu: adjustment:
Adjustment process can be aborted at optional moment by pressing
key, which is signaled by displaying the following message:
Caution:
1. Remember to carry out scale’s adjustment process when there is no load on the scale’s weighing platform!
2. If the adjustment process (span adjustment) lasts longer than 15 seconds, error message <Err8> is displayed and short beep sound is
emitted. Press
key and repeat adjustment procedure under more
stable ambient conditions!
- 43 -
Return to weighing with procedure o f saving changes:
See point - 11.2.2. – return to weighing.

17.2. Start mass adjustment

If the scale does not require the full adjusting process, or the user does not have sufficient number of adjustment weights, it is possible to adjust the start mass only.
Procedure:
Enter submenu <P6.CAL> in accordance with po int 11 .2 of the user manual, and:
The software displays the following messages / commands:
After completing start mass adjustment process, the scale returns to
displaying parameter’s name:
Start mass adjustment process can be aborted at optional moment by pressing
key, which is signaled by displaying the following
message:
- 44 -
Caution:
If the start mass adjustment lasts longer than 15 seconds, error message <Err8> is displayed and short beep sound is emitted. Press
key and
repeat adjustment procedure under more stable ambient conditions!
Return to weighing with procedure o f saving changes:
See point - 11.2.2. – return to weighing.

18. COOPER ATION WITH A PRINTER

Each time the
key is pressed causes sending a signal containing current mass value and a measuring unit. Depending on setting of STAB parameter, the printout is released with stable or unstable (instantaneous) value. Depending on setting of REPL parameter, printout is automatic or manual.
Medical scales enable cooperating with one of the following printers
KAFKA series:
a) KAFKA
Printout of weighing result and measuring unit only
b) KAF KA 1/Z
This printer features an internal real time clock. Both date and time are printed on plugging the printer to mains and connecting it to the scale.
c) KAFKA SQ S
This printer features an internal real time clock and possibility of running statistics from measurements. Statistic contents: quantity of samples, total mass of weighed samples, average value, standard deviation, variation factor, min value, max value, difference max – min.
- 45 -
- manually
-
After pressing key,
- continuously
-
After sending a controlling command,
- on a request from the computer
-
After sending a controlling command
Cable diagram:
Cable: scale – KAFKA printer

19. COOPERATION WITH A COMPUTER

Medical scales enable communicating with a PC computer featuring an RS 232 interface.
Sending weighing results to a computer can be carried out:
Caution:
In medical scales enabling continuous transmission and automatic printout by setting an appropriate parameters is not available for a user.
- 46 -
Cable diagram:
Cable: scale - computer
Medical scales enable cooperating with a computer software “SCALE EDITOR”. The software option of displaying indicator’s main weighing
window features the most important data on weighing process previewed on scale’s display. The software enables simple scale configuration, including designing printouts according to user needs, editing main software settings, editing RS 232 parameters, etc. Detailed description of software cooperation with a scale is provided in the software’s tab “Help...”.

20. COMM UNIC ATION PROTOCOL

20.1. General information

A. A character based communication protocol scale-terminal is
designed for establishing communication between a RADWAG scale and a peripheral devices via RS 232C interface.
B. It consists of commands sent from a peripheral device to the scale
and responses from the scale.
C. Responses are sent from the scale on each receipt of a command as
a reaction for a specific command.
D. Commands forming the communication protocol enable obtaining
data on scale’s status and influencing its operation, e.g.: request measurement results from the scale, zero indication, etc.
- 47 -
Command
Command description
S
SUI
CU1
Switch on continuous transmission in current measuring unit
XX_A CR LF
XX_D CR LF
XX_I CR LF
XX _ ^ CR LF
XX _ v CR LF
ES_CR LF

20.2. A set of commands recognized by the indicator

Z T TO
Zero scale Tare scale Give tare value Send stable measurement result in basic measuring unit
SI SU
Immediately send measurement result in basic measuring unit Send stable measurement result in current measuring unit Immediately send measurement result in current measuring unit
C1 C0
CU0 PC
Switch on continuous transmission in basic measur ing unit Switch off continuous transmission in basic measur ing unit
Switch off continuous transmission in current mea suring unit Send all implemented commands
Caution:
1. Each command must end with CR LF characters;
2. Sending a following command without waiting for a response for an already sent command may result in scale’s loosing (skipping) some of sent commands. The best policy for communication is not sending another command until a response for a former command has been received.

20.3. Responses format for commands sent from computer level

On receipt of a command the indicator sends a response in one of the following formats:
command understood and in progress command carried out (appears only after the command XX_A) command understood but not accessible at this moment command understood but max range is exceeded command understood but min range is exceeded Command carried out (completed)
- 48 -
XX _ E CR LF
an error occurred on command carrying out (time limit exceeded balance’s characteristic param et er)
Z_A CR LF
- command understood and in progress
-
Z_A CR LF
- command understood and in progress
-
Z_A CR LF Z_E CR LF
- command understood and in progress
-
Z_I CR LF
- command understood but not accessible at this moment
T_A CR LF
- command understood and in progress
-
T_A CR LF
- command understood and in progress
-
T_A CR LF T_E CR LF
- command understood and in progress
-
T_I CR LF
- command understood but not accessible at this moment
while waiting for stable measurement result (time limit is
XX - stands for a name of sent command _ - substitutes spaces

20.4. Command’s description

20.4.1. Zero scale

Format Z CR LF Accessible responses:
Z_D CR LF
Z_^ CR LF
command carried out
command understood but zeroing range exceeded
time limit exceeded while waiting for stable measurement result

20.4.2. Tare scale

Format: T CR LF Accessible responses:
T_D CR LF
T_v CR LF
command carried out
command understood but tarring range exceeded
time limit exceeded while waiting for stable measurement result

20.4.3. Give tare value

Format: TO CR LF Response: TO_TARA CR LF - command carried out
- 49 -
stability marker
S_A CR LF
- command understood and in progress
S_I CR LF
- command understood but not accessible at this moment
S_A CR LF MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
stability marker
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response is immediate with mass value in basic weighing unit
Frame format:
1 2 3 4 5-6 7-15 16 17 18 19 20 21
T O space
Tare - 9 characters with right justification Unit - 3 characters with left justification

20.4.4. Send stable measurement result in basic measuring unit

Format: S CR LF Accessible responses:
space tare space unit CR LF
S_E CR LF
- time limit exceeded while waiting for stable measurement result
Frame format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space
Example: S CR LF – command sent from a computer
S _ A CR LF - command understood and in progress S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response
is mass value in basic measuring unit.

20.4.5. Immediately send measurement result in basic measuring unit

Format: SI CR LF Accessible responses:
space sign mass space unit CR LF
- 50 -
stability
marker
SU_A CR LF
- command understood and in progress
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
marker
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
space sign mass space unit CR LF
Example: S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response of mass value in basic weighing unit

20.4.6. Send stable measurement result in current weighing unit

Format: SU CR LF Accessible responses:
SU_E CR LF
- time limit exceeded while waiting for stable measurement result
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U space
Example: S U CR LF – command from a computer S U _ A CR LF – command understood and in progress S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.
stability
space sign mass space unit CR LF
- 51 -
20.4.7. Immediately send measurement result in current measuring
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- mass value in current measuring unit is returned immediately
stability marker
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
stability
marker
unit
Format: SUI CR LF Accessible responses:
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
space sign mass space unit CR LF
Example: S U I CR LF – command from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out,
immediate response of mass value in current measuring unit

20.4.8. Switch on continuous transmission in basic measuring unit

Format: C1 CR LF Accessible responses:
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
space sign mass space unit CR LF
- 52 -

20.4.9. Switch off continuous transmission in basic measuring unit

C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and in progress
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
marker
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and in progress
Format: C0 CR LF Accessible responses:

20.4.10. Switch on continuous transmission in current measuring unit

Format: CU1 CR LF Accessible responses:
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
stability
space sign mass space unit CR LF

20.4.11. Switch off continuous transmission in current measuring unit

Format: CU0 CR LF Accessible responses:

20.4.12. Send all implemented commands Format: PC CR LF

Accessible responses: PC_- >_Z,T, TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command carried
out, terminal has sent all implemented commands.
- 53 -
Stability marker
[space] if stable
[v] if error of exceeding range to “-“ occurs
sign
[space] for positive values or [-] for negative values
mass
9 characters with decimal point and right
justification
unit
3 characters with left justification
command
3 characters with left justification

20.5. Manual printout

Medical scale enables generating manual printouts. The manual printout is generated after placing a weighed load on scale’s weighing platform,
stabilization of measurement result and pressing
key.
Caution:
If the scale is a verified one, then printout of unstable / instantaneous data in disabled.
Format of mass printout:
1 2 3 4 -12 13 14 15 16 17 18
stability
marker
space sign mass space unit CR LF
[?] if unstable [^] if error of exceeding range to “+” occurs
Example 1: _ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.
Example 2: ? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.
Example 3: ^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.
- 54 -
Err2
-
Value beyond the zero range
Err3
-
Value beyond the tare range
Err4
-
Adjustment mass or start mass beyond the acceptable range (±1% for weight, ±10 for start mass)
Err5
-
Mass of single part (piece) below scale’s reading unit.
Err7
-
Too short scale’s switch off time (recommended time more than 3 sec.)
Err8
-
Tarring / Zeroing operation time exceeded
null
-
Zero value from the AD converter
FULL2
-
Measurement range (Max. capacity) exceeded
LH
-
Start mass error, indication beyond acceptable range (from -5% to +15% of start mass)

20.6. Continuous transmission

Scale indicator enables printing mass value in continuous transmission in basic and in current measuring unit. Continuous transmission mode in medical scales can be enabled by sending a command via RS 232C interface (see point 20.4 of this user manual).

20.7. Printout configuration General information

If data contained on a standard printout is excessive or insufficient, and there is a need to modify it according to user needs, please use SCALE EDITOR computer software. The freeware is available for downloading from RADWAG website: http://www.radwag.com

21. ERROR MESSAGES

Caution:
1. Errors: Err2, Err3, Err4, Err8, null, LH, that appear on the display are also accompanied by a short beep sound (about 1 sec.).
2. Error FULL2 that appears on the display is also accompanied by a continuous sound until the cause of error (excessive load on the weighing platform) disappears.
- 55 -

22. SCALE TECHNICAL PARAMETERS

22.1. Chair scale

Technical data: WPT/K 250C
Max capacity 250kg Min load 2kg Readability [d] 100g Verification interval [e] 100g Tare range -250kg
Power supply
Working temperature Average operation time on accumulators 35 hours Interface RS 232
100-240VAC/12VDC
and 6xAA (NiMH)
from 0°C to +40°C

22.2. Personal scales

Technical data:
Max capacity 60/150kg 100/200kg Min load 0,4/1kg 1/2kg Readability [d] 20/50g 50/100g Verification interval [e] 20/50g 50/100g Tare range -150kg -200kg Height meter max range - 2m - 2m
Power supply
Working temperature Average operation time on
accumulators Interface RS 232
WPT
60/150 O
WPT
60/150 OW
230 V 50 Hz/11V AC, I
and 6xAA (NiMH)
from 0°C to +40°C
100/200 O
35 hours
WPT
= 600mA
max
100/200 OW
WPT
- 56 -
Average operation time on accumulators

22.3. Bed scales

Technical data: WPT/4B 500C WPT/8B 300C
Max capacity 500kg 300kg Min load 4kg 2kg Readability [d] 200g 100g Verification interval [e] 200g 100g Tare range -500kg -300kg
Power supply
Working temperature
Interface RS 232
230 V 50 Hz/11V AC, I
and 6xAA (NiMH)
from 0°C to +40°C
35 hours
= 600mA
max

22.4. Baby scales

Technical data: WPT 6/15D WPT 10/20D
Max capacity 6/15kg 10/20kg Min load 40/100g 100/200g Readability [d] 2/5g 5/10g Verification interval [e] 2/5g 5/10g Tare range -15kg -20kg
Power supply Working temperature
Average operation time on accumulators 35 hours Interface RS 232
230 V 50 Hz/11V AC, I
and 6xAA (NiMH)
from 0°C to +40°C
= 600mA
max
- 57 -

23. TROUBLE SHOOTING

Problem Cause Solution
Connect power adapter to charge the accumulator / accumulators.
Check the correctness of accumulator(s) installation (polarization).
Go to submenu “othr” and change parameter “5.4 t1” to “no”.
Unload the weighing platform. In short period of time the scale should indicate zero.
Scale does not turn on
Scale turns off automatically
After turning on message “LH” is displayed
Discharged accumulator / accumulators,
No accumulators (not installed or improperly installed)
Parameter “t1” set to “YES” (power saving mode)
Load on scale’s weighing platform during powering up.

24. ADDITIONAL EQUIPMENT

Accessories:
KAFKA printer cable - P0136,
Computer cable - P0108,
EPSON printer cable - P0151,
Power cord for car lighter 12V DC - K0047,
Thermal printer - KAFKA,
Dot matrix printer - EPSON,
Additional display in plastic housing - WD- 4/1
(available only as the factory installed option with a new scale),
RS232 / RS485 converter for PUE C/31 – KR-01,
RS232 / Ethernet converter for PUE C/31 - KR-04.
Computer software:
„Scale Editor” computer software,
"RAD-KEY" computer software,
"PW-WIN" computer software.
- 58 -
Table 201
Manufacturer declaration and information – electromagnetic emission
The DEVICE is intended to operate in the electromagnetic environment described below.
Recipient or user of the DEVICE should make sure that it operates in such an environment.
Emissivity tests
Compliance
Electromagnetic environ m e nt – information
Emission of RF
Group 1
The DEVICE produces energy in the range of RF only
proximity.
Emission in the range
Class B
The DEVICE is relevant to operate in all premises,
supplies dwelling houses.
RF – frequency from the interval o electromagnetic spectrum which is between the lower
transmission. Limit frequencies are 9 kHz and 3 000 GHz
Table 202
Manufacturer declaration and information - electromagnetic immunity
The DEVICE is intended to operate in the electromagnetic environment described below. Recipient or user of the DEVICE should make sure that it operates in such an environment.
Immunity tests
Electromagnetic
Electrostatic
+/- 6kV contact
air
+/- 6kV contact
The floor should be wooden,
concrete or made of ceramic
at least 30%.
Transient states
+/- 2 kV
for I/O lines
+/- 2 kV
for I/O lines
The quality of mains should
Surge
+/- 1 kV
associative mode
+/- 1 kV
associative mode
The quality of mains should

25. EMC INFORMATION

according to CISPR 11
as a subsequence of internal functionality. Owing to that the emission in the range of RF is minimal and it is unlikely for it to interfere with anything in the near
of RF according to CISPR 11
including living surface and premises directly connected to the low-tension system (mains) that
range of radio frequencies of long waves and the infrared range; frequencies useful for radio
Testing level IEC
discharge (ESD) according to IEC 61000-4-2
and impulses according to IEC 61000-4-5
60601
discharge
+/- 8kV through­discharge
for supply lines
+/- 1 kV
Met requirements
discharge
+/- 8kV through-air discharge
for supply lines
+/- 1 kV
environment – information
plates. If floors are covered with a synthetic material relative humidity should be
be on the on the office- or hospital-level environment.
disturbances according to IEC 61000-4-5
differential mode
+/- 2 kV
differential mode
+/- 2 kV
be on the on the office- or hospital-level environment.
- 59 -
Short supply voltage drops and
< 5% UT
through 5 seconds
< 5% UT
through 5 seconds
The quality of mains should
continuous supply during the
Magnetic field with
61000-4-8
3A/m
Magnetic field with frequency
Caution UT is an alternative current (AC) of mains before implementing the test level.
RF – frequency from the interval o electromagnetic spectrum which is between the lower
transmission. The limit frequencies are 9 kHz and 3 000 GHz
Table 204
Manufacturer declaration and information - electromagnetic immunity
The DEVICE is intended to operate in the electromagnetic environment described below. Recipient or user of the DEVICE should make sure that it operates in such an environment.
Testing
60601
Portable devices for wireless
Advisable minimal distance
changes of supply voltage in supply lines according to IEC 61000-4-11
frequency the same as the supply voltage according to IEC
range of radio frequencies of long waves and the infrared range; frequencies useful for radio
(> 95% voltage drop UT) through 0.5 period
40% UT (60% voltage drop UT) through 5 periods
70% UT (30% voltage drop UT) through 25 periods
< 5% UT (> 95% voltage drop UT)
(> 95% voltage drop UT) through 0.5 period
40% UT (60% voltage drop UT) through 5 periods
70% UT (30% voltage drop UT) through 25 periods
< 5% UT (> 95% voltage drop UT)
be on the on the office- or hospital-level environment. If a DEVICE user requires
breaks in the mains it is advisable that the DEVICE is supplied from UPS or batteries.
the same as the supply voltage should be on the level typical for office or hospital environment
Immunity tests
level IEC
Met requirements
Electromagnetic environment –
information
communication and cellular telephones should not be used within the distance from the DEVICE lower than the minimal recommended and that includes cables/harness. This distance can be calculated using the equations applicable for frequencies of transmitter operation.
- 60 -
Electromagnetic
IEC 61000-4-6
3 Vrms
3 Vrms
d=1,2√P
Electromagnetic
3 V/m
3 V/m
d=1,2√P 80 MHz to 800 MHz
Where P is a maximal transmitter output power
Caution 1: For 80 MHz and 800 MHz higher frequency range is taken.
persons.
(a) Powers sourcing from fields of specific transmitters such as base stations in cellular
greater than 3 V/m.
RF – frequency from the interval o electromagnetic spectrum which is between the lower range
transmission. Limit frequencies are 9 kHz and 3 000 GHz
CAUTION! Portable devices for RFI communication can
disturbances within the RF range
26 kHz to 80 MHz
according to
disturbances within the RF range
80 MHz to 1 GHz
d=2,3√P 800 MHz to 2,5 GHz according to IEC 61000-4-3
value in watts (W) according to information from a manufacturer.
d is a recommended, minimal distance in
meters (m).
The power of electromagnetic disturbances from stationary transmitters of RF, settled in conditions on the spot (a), should be lower than the level of compliance for every frequency range (b). Interference can appear in the proximity of devices marked with the following sign:
Caution 2: Given information are not applicable in every situation. Propagation of electromagnetic waves is influenced by absorption and reflection from surfaces, objects and
telephony, radio transmitters, amateur radios, AM and FM radio transmission and TV transmission cannot be theoretically estimated. To estimate an electromagnetic environment testing local conditions needs to be considered. If the measured field power on the spot of DEVICE operation overpasses an acceptable level of compliance, there should be tested if the DEVICE operates properly. If an inappropriate operation appears it may be necessary make some preventions such as displacement of the DEVICE. (b) For frequencies outside the range 150 kHz to 80 MHz, the field power should not be
of radio frequencies of long waves and the infrared range; frequencies useful for radio
influence ELECTRICAL MEDICAL DEVICES
- 61 -
Table 206
Recommended minimal distance between portable devices for RFI
communication and the DEVICE
The DEVICE intended to operate in the electromagnetic environment, where electromagnetic
maximal output power of communication devices.
Given maximal
Minimal distance adequate for the transmitter frequencies
[m]
150 kHz to 80 MHz
d
= 1,2√P
80 MHz to 800 MHz
d
= 1,2√P
800 MHz to 2,5 GHz
d
= 2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73 1 1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters with the maximal output power not mentioned above, the advisable minimal
persons..
RF – frequency from the interval o electromagnetic spectrum which is between the lower
transmission. Limit frequencies are 9 kHz and 3 000 GHz
CAUTION! Operating cellular telephones (in the proximity
CAUTION! Operation in the proximity of (e.g. 1 m) devices
unstable operation of the DEVICE.
CAUTION! Using accessories and cables other than
resistance of the DEVICE.
disturbances of RF are under control. Recipient or user of the DEVICE can prevent electromagnetic interference by keeping minimal distances between portable devices for wireless communication (transmitters) and the DEVICE prompted below, according to
transmitter’s
power
[W]
distance d in meters can be estimated by equation adequate to the frequency of the transmitter, where
CAUTION 1: If f= 80MHz and f=800MHz higher frequency range is taken. CAUTION 2: Given information are not applicable in every situation. Propagation of
electromagnetic waves is influenced by absorption and reflection from surfaces, objects and
range of radio frequencies of long waves and the infrared range; frequencies useful for radio
P
is a rated maximal power in watts (W).
of 2.8 m) can cause unstable operation of the DEVICE.
for microwave therapy or shortwave therapy can cause
described in present manual may result in increased emission of electrom ag netic waves and decreased wear
- 62 -
CAUTION! The DEVICE should not be used in the proximity
ack of disturbance/influence
intended to be used.
- symbol „product number”
- symbol „temperature limitations” (+ temperatures)
- symbol „ please see the user manual”
of other electric medical equipment and put one over another. If such situation appears the DEVICE operation should be observed to see the l on the operation of the DEVICE in the configuration it is

26. APPENDIX A

26.1. Pictograms on the data plate

- symbol „manufacturer” (+ name & address)
- symbol „production date” (+ production year)
- symbol „serial number”
- symbol „B type application part”
Example:
- WPT 60/150 OW personal scale

26.2. CE marking

- 63 -
MANUFACTURER
OF ELECTRONIC WEIGHING INSTRUMENTS
RADWAG WAGI ELEKTRONICZNE
26 – 600 Radom, Bracka 28
POLAND
Phone +48 48 384 88 00, fax. + 48 48 385 00 10
e-mail: export@radwag.com
www.radwag.com
Loading...