RADWAG MW-01 Series, MW-01 ver.1, MW-01 ver.2 User Manual

Page 1
User manual
Weighing module MW-01
Manual number: ITKU-29-05-08-14-A
BALANCES AND SCALES
www.radwag.com
Page 2
AUGUST 2014
2
Page 3
CONTENT
1. APPLICATION............................................................................................................................5
2. PRECAUTIONARY MEASURES................................................................................................5
3. WARRANTY CONDITIONS........................................................................................................5
4. TECHNICAL PARAMETERS .....................................................................................................6
4.1.
INPUT/OUTPUT parameters ................................................................................................7
5.
CONSTRUCTION .......................................................................................................................7
5.1. External view.........................................................................................................................8
5.2. Dimensions ...........................................................................................................................8
5.3. Interfaces ..............................................................................................................................8
5.3.1.
PROFIBUS .................................................................................................................. 9
5.4.
Diagrams of connection cables...........................................................................................10
6. INSTALLATION of “MwMANAGER” software.......................................................................12
6.1. Minimal hardware requirements..........................................................................................12
6.2. Installation procedure..........................................................................................................13
7. PC PROGRAM DESCRIPTION................................................................................................16
7.1. Weighing window ................................................................................................................ 16
7.2. Application settings .............................................................................................................17
7.2.1. Connection settings...................................................................................................17
7.2.2. Language ..................................................................................................................19
7.2.3.
Other .........................................................................................................................20
7.3.
Parameters .........................................................................................................................20
7.3.1. User Parameters .......................................................................................................21
7.3.2. Communication settings ............................................................................................22
7.3.3. Functions I/O ............................................................................................................. 24
7.4. Functions ............................................................................................................................26
7.4.1. Dosing .......................................................................................................................26
7.4.2. Checkweighing ..........................................................................................................30
7.4.3. Input/output state.......................................................................................................31
8. WEIGHING................................................................................................................................32
8.1. .......................................................................................................................32 Terms of use
8.2.
Zeroing................................................................................................................................33
8.3. Tarring.................................................................................................................................34
8.4. Weighing on dual range scales...........................................................................................34
8.5. Toggling between weight units............................................................................................35
9. SC LE PARAMETERS............................................................................................................36 A
9.1.
Autozero function ................................................................................................................ 36
9.2. Median filter ........................................................................................................................37
9.3. Filter ....................................................................................................................................37
10. CHECKWEIGHING.................................................................................................................38
10.1. limit..............................................................................................................................38 LO
10.2.
MIN/MAX limit ...................................................................................................................38
11. DOSING ..................................................................................................................................39
12. PARAMETERS IN FILE..........................................................................................................40
12.1. Saving to file......................................................................................................................41
1 .2. Uploading file data ............................................................................................................42 2
13.
OFFLINE MODE .....................................................................................................................44
3
Page 4
14. COMMUNICATION PROTOCOL............................................................................................45
14.1. General information ..........................................................................................................45
14.2. A set of commands ...........................................................................................................45
14.3. Respond message format.................................................................................................46
14.4. Command’s description ....................................................................................................46
14.4.1. Zeroing....................................................................................................................46
14.4.2. Tarring.....................................................................................................................47
14.4.3. Get tare value ......................................................................................................... 47
14.4.4. Set tare value..........................................................................................................47
14.4.5. esult in basic unit .........................................................................48 Send the stable r
14.4.6.
Send the result immediately in basic unit ...............................................................48
14.4.7. Send the stable result in current unit ......................................................................49
14.4.8. Send the result immediately in current unit.............................................................49
14.4.9. Switch on continuous transmission in basic unit.....................................................50
14.4.10. Switch off continuous transmission in basic unit...................................................50
14.4.11. Switch on continuous transmission in current unit ................................................50
14.4.12. Switch off continuous transmission in current unit ................................................51
14.4.13. Set lower threshold ...............................................................................................51
14.4.14. Set upper threshold...............................................................................................51
14.4.15. Read lower threshold ............................................................................................ 52
14.4.16. Read upper threshold ........................................................................................... 52
14.4.17. Send all implemented commands.........................................................................52
15. COMUNICATION MODULE PROFIBUS................................................................................53
15.1. General information ..........................................................................................................53
15.2. work ..........................................................53 Setting instrument’s address in a Profibus net
15.3.
Memory map .....................................................................................................................54
15.3.1. Output address ....................................................................................................... 54
15.3.2. Input address..........................................................................................................55
15.4. Description of variables.....................................................................................................55
15.4.1. riables......................................................................................................55 Output va
15.4.2.
Input variables ........................................................................................................57
16. MODBUS COMMUNICATION PROTOCOL...........................................................................60
16.1. Basic information ..............................................................................................................60
16.2. Addressing a device in ModBus network ..........................................................................61
16.3. Register tables in ModBus network ..................................................................................61
16.3.1. Table of output registers.........................................................................................62
16.3.2. Table of input registers ........................................................................................... 63
16.4. Software for controller PLC SIEMENS S7-300 ................................................................. 63
17. ERROR MESSAGES..............................................................................................................64
4
Page 5

1. APPLICATION

MW-01 weighing module is intended for conftruction of industrial load cell scales. Depending on the needs, communication with the module may be established by the following interfaces: RS232, RS485, Ethernet, Profibus and ModBus protocol. MW-01 Module may cooperate with PUE 5, PUE 7 terminals or PC computer.
Operating of the weighing module MW-01 from a PC level is carried out with
a computer software “MwManager”. A detailed description of the application
is included in the further section of the MW-01 user manual.

2. PRECAUTIONA RY ME ASUR ES

A. Please, read carefully this user manual and use the device according
to its intended use;
B. Weighed loads should be placed possibly in the central setion of the
weighing platform;
C. The weighng platform should be loaded with objects which gross
weight does not exceed the maximal capacity of the scale;
D. Do not leave heavy loads on the weoghing patform for longer period
of time;
E. In case of failure, immediately disconnect the power supply;
F. Devices that are to be decommissioned should be utilized according
to valid legal regulations.

3. WARRANTY CONDITIONS

A. RADWAG is obliged to repair or change those elements that appear
to be faulty because of production and construction reason,
B. Defining defects of unclear origin and outlining methods of their
elimination can be carried out only in participation of the user and the manufacturer representatives,
C. RADWAG does not take any responsibility connected with defecting
or loss deriving from unauthorized or inappropriate (not adequate to manuals) production or service processes,
5
Page 6
D. Warranty does not cover:
Mechanical failures caused by inappropriate exploitation of the device or failures of thermal or chemical origin or
Defects caused by atmospheric discharge, overvoltage in mains or other random event,
Maintenance activities (cleaning of the weighing module).
E. Warranty loss appears if:
A repair is carried out by an unauthorized service,
Intrusion into mechanical or electronic construction
of unauthorized people,
Removing or destroying protection stickers for the weighing module
F. The detailed warranty conditions are listed in the warranty certificate.
G. Contact with Authorized Service: +48 48 384 88 00 ext. 106 or 107.

4. TECHNICAL PARAMETERS

MODEL MW-01 ver.1 MW-01 ver.2
Interface
RS232, RS485,
Ethernet
RS232, PROFIBUS
Power Supply 8 ÷ 30VDC
Housing metal
IP rating IP 66
Working temperature -10°C ÷ 40°C
OIML class III
Number of verification intervals 10 000
Max. increase of input signal 19,5mV
Min. voltage per verification division 0,5µV
Min. load cell impedance 80
Max. load cell impedance 1200
Load cell excitation voltage 5VDC
Load cell connectivity 4 or 6 wire shielded cable
Number of weighing platforms 1
2INE/3OUT Standard
Transmission Speed RS232 RS485 Default 57600 bit/s
6
Page 7
4.1. INPUT/OUTPUT parameters
The module features 2 optoisolated inputs, 2 reed relay outputs and fast transistorized output type open collector.
OUT1 – OUT2 Parameters
Number of outputs 2 Type of outputs Reed operating contact Conductor intersection 0.14 – 0.5mm2 Maximum switchable current 200mA Maximum conducted voltage 50VDC
Output OUT3 Parameters
Number of outputs 1 Type of outputs Transistor Umax 5VDC Imax 100mA
Input Parameters
Number of inputs 2 Type of inputs Optocoupled Conductor intersection 0.14 – 0.5mm2 Control voltage range 5 -24V DC

5. CONSTRUCTION

Weighing module comprises metal housing. Signal cables are fed through cable glands. Communication with the peripheral devices is established by one of the following interfaces: RS485, RS232, Ethernet, Profibus.
The weighing module enables cooperation with indicators or PC computers. It features 2 optoisolated inputs, 2 reed relay outputs and one fast transistor output. The weighing module is supplied form 8÷30VDC.
7
Page 8
5.1. External view
Fig.1 Weighing module MW-01
5.2. Dimensions
Fig.2 Dimensions of the MW-01
5.3. Interfaces
MW-01 ver.1 has cables fed through cable glands. MW-01 ver.2 has 2 sockets (PROFIBUS IN and PROFIBUS OUT), RS232 is fed though cable gland.
8
Page 9
Fig.3 Location and marking of interfaces for MW-01 main board
5.3.1. PROFIBUS
MW-01 module, version 2, is equipped with input and output socket for PROFIBUS. The volatage, indispensable for correct operation of a terminator is 5VDC on an output socket. Available socket standard is M12 5 pin B coded (for DP PROFIBUS).
Sockets’ topology:
Pin1 – NC
PROFIBUS IN
(male)
PROFIBUS OUT
(female)
Pin2 – A Pin3 – NC Pin4 – B Pin5 – NC
Pin1 - +5V Pin2 – A Pin3 – GND Pin4 – B Pin5 – NC
9
Page 10
5.4. Diagrams of connection cables RS232 cable
RS485 cable
Fig. 4 PT0145 Cable
Fig. 5 PT0012 Cable
10
Page 11
Ethernet cable
IN/OUT cable
Fig. 6 PT0014 Cable
Fig. 7 PT0146 Cable
11
Page 12

6. INSTALLATION of “MwMANAGER” software

Caution:
In order to install the program on the computer with older version
“MwManager” softwre, uninstall the previous version first,
The installation instruction was made for Windows 7 and it is compatibile with former versions of MS Windows
Proper operation of the program reguires installing Microsoft .NET Framework ver. 2.0 or higher. It is accessible on Microsoft website:
http://www.microsoft.com/downloads/details.aspx?displaylang=pl& FamilyID=0856eacb-4362-4b0d-8edd-aab15c5e04f5
Proper operation of the program reguires installing all accessible ServicePacks for your operating system.
Due to continuous software updates there may occur some discrepancies between the instruction and the program,
RADWAG takes no responsibility for the consequences of the software operation, and any errors resulting from the incorrect software operation,
RADWAG takes no responsibility for loss or protection of any data caused by the incorrect utilization of the software or computer.
6.1. Minimal hardware requirements
Proper operation of the software requires a computer with below specified configuration:
A PC computer with installed OS MS Windows 2000/XP/2003/Vista/7,
processor 2.4 GHz or faster,
RAM 512 MB or more (recommended 1 GB),
At least 1 GB of free space on the hard drive,
Display with minimum resolution 800 x 600,
DVD drive.
12
Page 13
6.2. Installation procedure
1. On receiving an installation file of MwManager x.x.x.x.exe software,
run installation file, select language version and press OK key.
2. Press in the welcome window:
13
Page 14
3. Browse window:
If necessary, change the software insallation folder, and press
.
4. Window for selecting tasks:
On marking / unmarking available option, press key.
14
Page 15
5. Installation window:
In order to install the software, press „Install” key.
6. Software installation completion window:
To finish installation process, press „Finish” key.
15
Page 16
7. A shortcut will be created on computer’s desktop.

7. PC PROGRAM DESCRIPTION

Operation of the weighing module from PC computer level requires
application ”MwManager”. The application can operate in MS Windows environment with installed add-on .NET framework 2.0. The program
enables calibrationg the module, displaying current mass indication, tarring, zeroing, setting filters and functions for corresponding inputs and outputs.
Caution:
1. This manual complies with the software “MwManager” starting from version 1.0.2.1
2. The entered values are accepted by pressing Enter key. The changes are saved in the weighing module on pressing Save key. All temporary parameters settings that are not saved in the weighing module are highlighte in red.
7.1. Weighing window
and weighing module program Prj215 from version 2.2.
Fig.8 View of the software’s weighing window
On completing initializating procedure, the weighing window displays below pictograms:
- precise zero indication
- stable measurement result
- weighing unit
- weighing platform number
16
Page 17
If the indication is other than zero – press the zeroing key.
Function keys:
-
-
Zeroing
Taring
7.2. Application settings
Tab comprises the settings for conneting the weighing module, choosing interface language version and other software options.

7.2.1. Connection settings

In tab
, key opens a window for establishing
connection with the weighing module.
Fig.9 Connection settings’ window
17
Page 18
Description:
Means of connection Interface for connecting to the weighing module
RS 232 Connection via port RS232
TCP/IP Ethernet connection
RS 485 RS 485 connection
Offline
The offline mode is used for saving and editing all indispensable parameters of the configuration file.
RS232:
Port Baud rate Baud rate for the RS232 interface. 57600 bps by default
Parity Parity bit parameter. “None” by default
Data bits Number of data bits. 8 by default
Stop bits Number of stop bits. 1 by default
Choosing a COM port number to which the weighing module is plugged
TCP/IP:
Adres IP IP address of the device, default setting 192.168.0.2 Port Port set in the weight module, default setting 4001.
RS485:
Port Choosing a COM port number to which the weighing module is plugged Baud rate Baud rate for the RS485 interface. 57600 bps by default
Address Weighing module address in the net
Caution:
1. The default baud rate of RS232 and RS485 interfaces in the MW-01 is set to 57600 bps;
2. If case of problems with establishing a connection to the weighing module, set and check accessible baud rates or connect through Ehternet interface;
3. If the display is set to port RS232, then connecting the computer software through the same port is impossible. Display operation on this port is restored on restarting the module (unplugging and repeated plugging to mains).
18
Page 19
Keys description:
Disconnect Communication disconnected
Reconnect Switching off and on (communication restart) communication with the device
Connect Establishing connection with the weighing module

7.2.2. Language

Use tab
, and key to open a window enabling
changing the language of the software.
Fig.10 Language selection window
On selecting the language version, press “Apply” key to save the changes.
Accessible language versions:
Czech
English
French
German
Polish
Spanish
19
Page 20

7.2.3. Other

Use tab
, key to start other software options.
Fig.11 Other options window
Tick “autostart” option, so that on switching on the software automatically
connects to the weighing module, by the default of last activated means of connection.
7.3. Parameters
Tab
comprises all factory and user parameters, and also the
communication parameters of the weighig module and functions of 2 inputs and 3 outputs.
20
Page 21

7.3.1. User Parameters

Use tab
, key to open a window with user parameters
of the weoghing module. The parameters are visible and edittable for each software user.
Fig.12 User parameters window
List of user parameters:
Autozeroing - tick to enable autozeroing
Beep - Sound signal
Median filter - tick to enable median filter
Filter - tick to enable averaging filter
Current unit - Toggling between weighing units in the weighing window
21
Page 22

7.3.2. Communication settings

Use tab , key to open a window with communication
parameters of the weighing module and configuration of peripheral devices. The parameters are visible and edittable to all software users, who establish cmmunication with the weighing module.
Ethernet
Fig. 13 Communication parameters window for Ethernet
List of fields with description:
IP address - Device’s IP address, default 192.168.0.2 Subnet mask - Ethernet subnet mask, default 255.255.255.0 Default gate - Ethernet default gate, default 192.168.0.1 Port - TCP communication port, default 4001.
Timeout -
Inactivity time in which the device breaks communication, expressed in seconds, range 0 – 300 [s].
22
Page 23
RS 232/485
Fig.14 RS 232/485 comunication parameters window
List of fields with description:
Module address -
Baud rate RS232 -
Baud rate RS485 -
Weighuing module address in RS 485 network (the network requires setting different address for each of the devices), default set to 1. Available range from 1 to 254.
Baud rate setting for RS 232 communication interface. Default 57600 bps
Baud rate setting for RS 485 communication interface. Default 57600 bps
Caution:
Indicator series PUE7 and other devices may use the address for Ethernet connection (TCP/IP) and connection through RS232. In such case, set parameter „Module address” to default value 1 for MW-01 and in the settings of PUE7.
On changing the communication parameters, save the settings and restart the weighing module (by unplugging and repeated plugging to mains), so that the inserted changes are effective. Remember, that the new parameters are inserted in the connection settings window with the weighing module.
23
Page 24
Devices / Interface
List of fields with description:
Additional display port
Modbus - Selecting Modbus protocol type and operating interface,
Offset adr. Modbus
Interface setting for the additional display for the communication
-
protocol of LCD display.
Offsetting memory addressing for Modbus protocol, Default value 1 for correct operation of PLC controller.
-
Caution! In case of testing software ProfiModBusTester set the value to 0. Available values 0 - 255.

7.3.3. Functions I/O

The weighing module features two inputs and three outputs.
Use tab
, key to open a window enabling software
user to access configurating functions of the weighing module inputs and outputs.
24
Page 25
Inputs’ configuration
Accessible input functions:
No Tarring Zeroing Start dosing Stop dosing
Outputs’ configuration
Accessible output functions:
None Stable MIN stable MIN OK stable OK
Output inactive
Signalling stable weighing result over LO limit
Signalling stable weighing result over LO limit but below MIN limit
Signalling unstable weighing result over LO limit but below MIN limit
Signalling stable weighing result between MIN and MAX limits
Signalling unstable weighing result between MIN and MAX limits
Fig.15 I/O configuration window
Input inactive
Platform tarring
Platform zeroing
Initiate batching process
Stop batching process
25
Page 26
MAX stable MAX
Signalling stable weighing result over MAX limit
Signalling unstable weighing result over MAX limit
Notice:
If a function is assigned to a specific output and the same output is used for bulk or fine dosing then on batching start and continuation, the outputs will be activated compatibly to the dosing parameters. End of dosing process causes switching over the functions to outputs.
7.4. Functions
Use tab to functions of dosing (batching), checkweighing, inputs
and outputs status and simulation.

7.4.1. Dosing

Use tab process.
, key to open a window enabling setup of dosing
Fig.16 Window of dosing parameters
26
Page 27
Bargraph
The dosing window features a graphic bar visualizing mass indication within the weighing range of the weighing module
Ticking the additional option extends bar graph scaling to 120 % of the maximum dosing limit. If the fine dosing threshold is disabled, then the bar graph is scaled according to the bulk dosing criteria.
Fig.17 Bar graph scaling for bulk dosing limit
Fig.18 Bar graph scaling for bulk and fine dosing limits
Fig.19 Bar graph for low weight without scaling
27
Page 28
Fig.20 The same bar graph with enabled scaling option
Dosing parameters
Fig.21 Window for setting dosing parameters
The dosing process can consist of one or two phases depending on the needs.
List of fields with description:
Bulk dosing limit
Mass value for which the high state of selected outputs is set in the first dosing phase,
Output No. for bulk dosing
Fine dosing limit
Output No. for fine dosing
Selection of output or outputs numbers assigned to the first dosing phase,
Mass value for which the high state of selected outputs is set in the second dosing phase,
Selection of output or outputs numbers assigned to the second dosing phase,
28
Page 29
Accessible outputs for specific dosing phase:
1 2 3 1 & 2 1 & 3 2 & 3 1 & 2 & 3
-
High state on the first reed relay output
High state on the second reed relay output
High state on the third transistor output
Activating the first and second reed relay output
Activating the first reed relay output and the third transistor output
Activating the second reed relay output and the third transistor output
High state for all outputs
Dosing switched off for specific limit
Dosing status
The dosing status window displays the current status of dosing process
Description:
Dosing process status: DOSING – dosing in progress
Dosing status
DISCONTINUED – dosing terminated by pressing STOP key. STOP – dosing stoppage, COMPLETED – Dosing completed.
Simulation of inputs’ operation
Inputs simulation enables simulating the operation of a function assigned to a specific input. The first input corresponds to the upper key, and the bottom key simulates operation of a function assigned to the second input.
29
Page 30
Dosing simulation
The lower part of the window comprises keys for starting and stopping dosing process independently on the functions assigned to the inputs. They enable starting and stopping the dosing process.

7.4.2. Checkweighing

Start option
and press key to open a window
enabling checkweighing setting.
Fig.22 Checkweighing settings window
List of fields with description:
LO limit Min limit Max limit
Output signals MIN, OK, MAX only if it is over (net) LO limit
Net mass value below which the MIN limit is signalled, above OK
Net mass value for which the OK limit is signalled, above MAX
30
Page 31
Signalling function with limits:
MIN
OK
MAX
Caution:
The checkweighing signalization in the software is accessible on setting the functions to outputs.

7.4.3. Input/output state

Start option
and press key to open a window enabling
setting the input and output signalling and setting output status.
Fig.23 Window of inputs and outputs status
31
Page 32
Software inputs and outputs are numbered according the weighing module documentation.
Active input / output (high state)
Inactive input / output (low state)
The simulation of output operation is possible on pressing the output number which is activated immediately, provided there is no function assigned to the output.
The simulation of inputs is accessible in the dosing window.

8. WEIGHING

Place weighed load on the weighing platform. As the stable measurement pictogram
8.1. Terms of use
In order to ensure a long lasting operation and correct measuring of weighed loads it is advised to:
Place the loads on the weighing platform gently and shocklessly:
is displayed, read the measurement result.
Place the loads in the centre of the weighing platform (eccentric weighing errors are outlined in the norm PN-EN 45501 sections
3.5 and 3.6.2):
32
Page 33
Do not load the weighing platform with concentrated force:
Avoid side loads and particularily side shocks:
8.2. Zeroing
In order to zero the mass indication in program „MwManager” press
key in the weighing window (the top right corner) or recall the zeroing function. The display should indicate mass equal to zero and pictograms:
and should be displayed.
Zeroing is equivalent to setting a new zero point comprehended by the scale as a precise zero. Zeroing is possible only for stable display status.
Zeroing can be carried out by pressing the external push-button connected to an input configured to zeroing function.
33
Page 34
Caution:
Zeroing display status can only be carried out within the ±2% of the maximal scale capacity. If a zeroed indication is beyond the range ±2% of maximal scale indication, the display indicates error message Err2.
The procedure of defining inputs - keys is described in section 6.3.3 of this manual.
8.3. Tarring
In order to determine net weight place a container on the weighing platform
and on indication staibilization press
The display should indicate mass equal to zero and pictograms: Net i
key or recall the tarring function.
should be displayed.
While using tarring function pay attention not to exceed the maximal measuring range of a scale. When taking the load and its packaging off the weighing platform the display indicates value that is a sum of tarred weigts preceded by minus sign.
Tarring can be carried out by pressing the external push-button connected to an input configured to tarring function
Caution:
Tarring cannot be carried out if the scale display indicates zero or negative values. In such case the display indicates error message Err3.
The procedure of defining inputs - keys is described in section 6.3.3 of this manual.
8.4. Weighing on dual range scales
Switching between the 1 without operator’s interference, at the point of reaching the Max of the 1
st
range and the 2nd range takes place automatically,
st
range.
Weighing in range II is signalled by displaying pictogram in the top left
corner of the display. When unloading the weighing platform, the indication
returns to zero. Weighing in the 2
nd
range continuous until the indication
returns to AUTOZERO zone.
34
Page 35
Fig. 24 Weighing window in the 2
When in AUTOZERO zone, the indication retirns to precise zero, which is signalled by pictogram
and pictogram of the 2
from the display. Weighing returns automatically to the 1
8.5. Toggling between weight units
The weight unit can be changes in the main window of “MwManager”
software by pressing
key.
nd
range
nd
range disappears
st
range.
Fig. 25 The main window with changed weighing unit
Selection options:
If the basic weighing unit [kg], the following units are accessible:
[kg, lb, oz, ct, N, g]. For verifies scales, [lb, oz, N] are not accessible;
If the basic weighing unit is [g], the following units are accessible:
[g, kg, lb, oz, ct, N] For verifies scales, [lb, oz, N] are not accessible.
35
Page 36

9. SCALE PARAMETERS

Users can adjust the scale to ambient conditions (filtering level) or user needs (autozero operation) and also determine the minimum weight limit for operation of some basic scale functions.
The parameters are accessible in tab
> .
List of scale parameters:
Autozero
Median Filter
Filter
9.1. Autozero function
In order to ensure precise scale indications the software “AUTOZERO” function has been introduced. It is designed to automatically control and correct scale’s zero indication.
When autozero function is enabled, the subsequent results are compared in constant time intervals. If two subsequent results differ less than the declared AUTOZERO range the scale automatically sets new zero point and pictograms of stability –
and zero state – are displayed.
If AUTOZERO is enabled, then each measurement always starts from the precise zero point. There are, however, cases in which the function may disturb measuring process. It is for instance while very slow loading the weighing platform (for instsance pouring a load). In such case the zero correcting set can correct the actual indication of load placed on the scale’s weighing platform.
Procedure:
Enter the User parameters window by pressing choose parameter <Autozeroing> and set the appropriate value.
36
key,
Page 37
Accessible settings:
NO YES
- autozero disabled
- autozero enabled
9.2. Median filter
The intended use of the median filtering is eliminating shortlasting interference of impulse character (e.g. mechanical shocks).
Procedure:
Enter User parameters window by pressing
key, choose
parameter <Median Filter> and set its appropriate value.
Accessible settings:
None - median filter disabled
0.5, 1, 1.5, 2, 2.5 - median filter enabled
9.3. Filter
The intended use of the averaging filter is adjustment to existing ambient conditions.
Procedure:
Enter User parameters by pressing
key, choose parameter
<Filter> and set its appropriate value.
Accessible settings:
No, Very Fast, Fast, Average, Slow
Caution:
The higher filtering level the longer stabilization time of measurement result.
37
Page 38

10. CHECKWEIGHING

Checkweighing is a function that aims at precise weighing a sample with pre-defined low anf high weighing limits (checkweighing thresholds (LO – sample mass too low, HI – sample mass too high, OK – sample mass correct).
Such software solution is a very good means for quick mass value evaluation with no need for contiinuous monitoring the measurement result. The above mentioned weighing status (LO, OK, HI) have their graphic visualization presented on scale’s display.
The status is also indicated by optical signalization or controlled by the set of external devices.
> 0 <
Lo
MIN MAX
Fig. 26 Illustration of intervals for checkweighing function
Caution:
The checkweighing configuration is described in section 7.4.2 of this manual.
10.1. LO limit
Parameter <LO limit> determines the net weight indicated on the display
which activates opertaion of outputs for status MIN, OK, MAX.
Procedure:
Enter parameter <LO limit>, insert the value of LO limit and press
Min thres.
OK
Max thres.
HI
Enter key. Save changes using
key.
10.2. MIN/MAX limit
Parameter <MIN limit> in checkweighing determines the net weight limit
for which the checkweighing status is switched between MIN and OK.
Parameter <MAX limit > in checkweighing determines the net weight limit
for which the checkweighing status is switched between OK and MAX.
38
Page 39
Output signalization is activated over the set net value of LO LIMIT.
Procedure:
Enter parameter <MIN threshold> or <MAX threshold>, insert the limit value and press Enter key. Save changes using
key.

11. DOSING

Dosing is a function enabling precise load measurement to a pre-defined target value.
Parameter <Bulk dosing limit> determines in fast (bulk) dosing
the value of net mass for which one or a few outputs are activated.
Parameter <Fine dosing limit> determines in slow (fine) dosing the value
of net mass for which one or a few outputs are activated.
Procedure:
Enter parameter <Bulk dosing limit> or <Fine dosing limit>, insert the limit value and press Enter key. Save changes using
key.
Changes are confirmed by a message box:
If any changes in limits values are introduced and but not saved, then
it is possible to preview the currently set limits by pressing key.
The readout is confirmed by a message box:
39
Page 40
Caution:
The dosing function and its parameters are described in section 7.4.1 of this manual.

12. PARAMETERS IN FILE

“MwManager” software features saving set parameters in a file format *.sav. The function can be utilized for saving weighing module settings
in a backup copy, which is restorable in case of module defect or further utilization of the parameters while configuring larger number of weighing modules.
Fig.27 An instance of a windpw with accessible option for saving and uploading a file
40
Page 41
File format:
MW01_(factory number)_YY-MM-DD_HH-MM.sav
12.1. Saving to file Procedure:
As the weighing module parameters are set, press key to save them in a file.
Then, using system operation window browse for file saving directory,
and press
key.
Fig.28 System window “Save as”
Default saving directory: C:\Program Files\RADWAG\MwManager\Save\
Correctly saved parameters cause displaying the below message box:
41
Page 42
Caution:
The system window look depends on installed OS version, and may differ fro mthe one visible in Fig.28.
12.2. Uploading file data Procedure:
In order to upload parameters from a file, press
key,
Then, using system dialogue window, see fig. 29, select an earlier
saved file, and press
key
Fig.29 System window „Open”
42
Page 43
Next, using window for uploading groups of parameters select one or all parameters to be uploaded to the weighing module.
Fig.30 Window for selecting groups of parameters
The selected parameters are grouped according to their location
in the main software menu.
Parameters that are correctly uploaded are confirmed by the below message box:
43
Page 44

13. OFFLINE MODE

The Offline mode enables starting different software options with no need to
connect to a weighing module. This mode of connection is created to enable saving parameters without phsycial connecting to a weighing device.
Fig. 31. Window for starting Offline mode.
Procedure:
On software start, enter connection settings tab and in the connection
settings options select Offline, and press
key,
The weighing window displays a communicate as presented in Fig. 31,
Set parameters and save them into a file according to description in
section 12.1 of this manual.
44
Page 45

14. COMMUNICATION PROTOCOL

14.1. General information
A. Serial communication protocol “weighing module – terminal”
is designed for establishing communication between a MW-01 weighing module and a peripheral device through interfaces: RS232, RS485, Ethernet;
B. The protocol consists of commands sent from a peripheral
device to the weighing module, and responses sent inversely;
C. The responses are sent from the weighing module on each receipt
of a command. Each response is a reaction for a specific command;
D. Commands which form serial communication protocol enable checking
device’s status, and trigger its reaction, e.g. obtaining measurement result from the weighing module, preview mass displaying window, etc.
14.2. A set of commands
Commands
Z T OT UT S SI SU SUI C1 C0 CU1 CU0 DH UH ODH OUH PC
Description of commands
Zeroing
Tarring
Get tare value
Set tare value
Send the stable result in basic unit
Send the result immediately in basic unit
Send the stable result in current unit
Send the result immediately in current unit
Switch on continuous transmission in basic unit
Switch off continuous transmission in basic unit
Switch on continuous transmission in current unit
Switch off continuous transmission in current unit
Set lower threshold
Set upper threshold
Read lower threshold
Read upper threshold
Send all implemented commands
45
Page 46
Notice:
1. Each command have to be terminated in CR LF;
2. The best Policy for communication is not sending another command until the former answer has been received.
14.3. Respond message format
After sending a request message you can receive:
XX_A CR LF XX_D CR LF XX_I CR LF XX _ ^ CR LF XX _ v CR LF XX _ OK CR LF ES_CR LF
XX _ E CR LF
command accepted and in progress
command completed (appears only after XX_A)
command comprehended but cannot be executed
command comprehended but time overflow error appeared
command comprehended but the indication below the
Command done
Command not comprehended
error while executing command – time limit for stable result exceeded (limit time is a descriptive parameter of the scale)
XX - command name
_ - substitutes spaces
14.4. Command’s description

14.4.1. Zeroing

Syntax Z CR LF
Possible answers:
Z_A CR LF Z_D CR LF
Z_A CR LF Z_^ CR LF
Z_A CR LF Z_E CR LF
Z_I CR LF
- command accepted and in progress
- command completed
- command accepted and in progress
- command comprehended but zero range overflow appeared
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed
46
Page 47

14.4.2. Tarring

Syntax: T CR LF
Possible answers:
T_A CR LF T_D CR LF
T_A CR LF T_v CR LF
T_A CR LF T_E CR LF
T_I CR LF
- command accepted and in progress
- command completed
- command accepted and in progress
- command comprehended but tare range overflow appeared
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed

14.4.3. Get tare value

Syntax: OT CR LF Reply: OT_TARA CR LF – command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
O T space tare space unit space CR LF
Tare - 9 characters justified to the right Unit - 3 characters justified to the left
Notice:
Tare values are always send in calibration unit.

14.4.4. Set tare value

Syntax: UT_TARE CR LF, where TARE – tare value
Possible replies:
UT_OK CR LF UT_I CR LF ES CR LF
- command completed
- command correct, but not accessible at the moment
- command incorrect (e.g. incorrect tare format)
47
Page 48
Notice:
Use dots as decimal points in tare values.

14.4.5. Send the stable result in basic unit

Syntax: S CR LF
Possible answers:
S_A CR LF S_E CR LF
S_I CR LF S_A CR LF
MASS FRAME
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space stability space sign mass space unit CR LF
Example: S CR LF – computer command
S _ A CR LF - command accepted and in progress S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.

14.4.6. Send the result immediately in basic unit

Syntax: SI CR LF
Possible answers:
SI_I CR LF MASS FRAME
- command comprehended but cannot be executed at the moment
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space unit CR LF
48
Page 49
Example: S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value
in basic unit is returned immediately.

14.4.7. Send the stable result in current unit

Syntax: SU CR LF
Possible answers:
SU_A CR LF SU_E CR LF
- command accepted and in progress
- timeout while waiting for stable results
SU_I CR LF SU_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U space stability space sign mass space unit CR LF
Example: S U CR LF – computer command
S U _ A CR LF - command accepted and in progress S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass value
in current unit is returned.

14.4.8. Send the result immediately in current unit

Syntax: SUI CR LF
Possible answers:
SUI_I CR LF MASS FRAME
- command comprehended but cannot be executed
- mass value in current unit is returned immediately
49
Page 50
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF
Example: S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned

14.4.9. Switch on continuous transmission in basic unit

Syntax: C1 CR LF
Possible answers:
C1_I CR LF C1_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space unit CR LF

14.4.10. Switch off continuous transmission in basic unit

Syntax: C0 CR LF
Possible answers:
C0_I CR LF C0_A CR LF
- command comprehended but cannot be executed
- command comprehended and executed

14.4.11. Switch on continuous transmission in current unit

Syntax: CU1 CR LF
50
Page 51
Possible answers:
CU1_I CR LF CU1_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF

14.4.12. Switch off continuous transmission in current unit

Syntax: CU0 CR LF
Possible answers:
CU0_I CR LF CU0_A CR LF
- command comprehended but cannot be executed
- command comprehended and executed

14.4.13. Set lower threshold

Syntax: DH_XXXXX CR LF, where: XXXXX – mass format
Possible answers:
DH_OK CR LF ES CR LF
- command executed
- command not comprehended (wrong mass format)

14.4.14. Set upper threshold

Syntax: UH_XXXXX CR LF, where: XXXXX – mass format
Possible answers:
UH_OK CR LF ES CR LF
- command executed
- command not comprehended (wrong mass format)
51
Page 52

14.4.15. Read lower threshold

Syntax: ODH CR LF Possible answers: DH_MASA CR LF - command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
D H space mass space unit space CR LF
Mass - 9 characters justified to the right Unit - 3 characters justified to the left

14.4.16. Read upper threshold

Syntax: OUH CR LF Possible answers: UH_MASA CR LF - command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
U H space mass space unit space CR LF
Mass - 9 characters justified to the right Unit - 3 characters justified to the left

14.4.17. Send all implemented commands

Syntax: PC CR LF
Possible answers:
PC_A_”Z,T,S,SI,SIA,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT, PC" – command executed, the weighing module has sent all the implemented
commands.
52
Page 53

15. COMUNICATION MODULE PROFIBUS

15.1. General information
Communication module Profibus ensures exchanging data between a superordinate controlling unit (master) and a weighing module MW-01 (slave) in accordance with protocol Profibus DP.
The superordinate unit (master):
cyclically reads inputs signals fro mthe weighing module MW-01,
cyclically saves outputs status to the weighing module MW-01.
Functions of Profibus communication with the weighing module MW-01:
Mass readout from the weighing module
Tarring the weighing module
Zeroing the weighing module
Reading module’s status
Reading current measuring unit
Setting and reading tare value
Setting and reading LO threshold value
START / STOP of dosing process
Setting and reading the value of bulk (fast) dosing threshold
Setting and reading the value of fine (slow) dosing threshold
Setting outputs status
Reading the status of dosing process
Reading inputs status
Setting and reading Min threshold value
Setting and reading Max threshold value
15.2. Setting instrument’s address in a Profibus network
The address of the weighing module MW-01 in Profibus network has to be set in accordance with module addressing specification. (See point 7.3.2
of this user manual – „Communication settings”). Field module address
is used to set corresponding address of the device in Profibus network.
53
Page 54
15.3. Memory map

15.3.1. Output address

Address
0 1 2 3 4 5 6 7 8 9
Offset
0 1 2 3 4 5 6 7 8 9
M T J S LO ST SW MIN MAX DS DW
M M M M T T T T J J
S S LO LO LO LO - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - ST ST SW SW MIN MIN
MIN MIN MAX MAX MAX MAX DS DS DS DS
DW DW DW DW - - - - - -
- - - - - - - - - -
- Mass for a platform, 4 bytes, float
- Tare for a platform, 4 bytes, float
- Current measuring unit of a platform, 2 bytes, word
- Platform status, 2 bytes, word
- Lo threshold of a platform, 4 bytes, float
- Process status, 2 bytes, word
- Inputs status, 2 bytes, word
- MIN threshold, 4 bytes, float
- MAX threshold, 4 bytes, float
- Bulk (fast) dosing threshold, 4 bytes, float
- Fine (slow) dosing threshold, 4 bytes, float
Table. 1
54
Page 55

15.3.2. Input address

Address
Offset
0 1 2 3 4
0 1 2 3 4 5 6 7 8 9
C C CP CP - - T T T T
LO LO LO LO SW SW MIN MIN MIN MIN
MAX MAX MAX MAX DS DS DS DS DW DW
DW DW - - - - - - - -
- - - - - - - - - -
Table. 2
C CP T LO SW MIN MAX DS DW
- command, 2 bytes, word
- command with a parameter, 2 bytes, word
- Tare for a platform, 4 bytes, float
- Lo threshold of a platform, 4 bytes, float
- Outputs status, 2 bytes, word
- MIN threshold, 4 bytes, float
- MAX threshold, 4 bytes, float
- Bulk (fast) dosing threshold, 4 bytes, float
- Fine (slow) dosing threshold, 4 bytes, float
15.4. Description of variables

15.4.1. Output variables

Readout of output variables enables obtaining data on instrument’s status.
Notice:
All output values, except for mass, are displayed in adjustment/calibration unit.
Name of output variable address length [word] d ata type
mass 0 2 float
tare 4 2 float
unit 8 1 word
status 10 1 word
55
Page 56
LO 12 2 float
process status 64 1 word
input status 66 1 word
MIN 68 2 float
MAX 72 2 float
bulk (fast) dosing threshold 76 2 float
fine (slow) dosing threshold 80 2 float
Table. 3
mass – mass value in current unit is returned,
tare – tare value in calibration/adjustment unit is returned,
unit – determines current (displayed) measuring unit:
Byte no.
Unit
g gram kg kilogram ct carat lb pound oz ounce N Newton
B5 B4 B3 B2 B1
0 0 0 0 0 0 0 0 0 0 0 0 0 0
1
1
1
0 0
0 0 0
1
0
1
0
0
0 0 0
B0 Dec
0 1
1 2
0 4 0 8 0 16 0 32
status – determines scale’s status
Status
bytes
0 correct measurement (no error message from scale) 1 1 stable measurement 2 2 scale in zero indication 4 3 scale tarred 8 4 scale in second measuring range 16 5 scale in third measuring range 32 6 scale reports NULL error 64 7 scale reports LH error 128 8 scale reports FULL error 256
Task Dec
For instance:
byte no. B8 B7 B6 B5 B4 B3 B2 B1 B0 value
0 0 0 0 1 0 0 1 1
56
Page 57
The scale does not report an error, the measurement is stable in the second measuring range.
• LO – the value of LO threshold in calibration/adjustment unit is returned.
• Process status – determines the status of dosing process
Decimal value of
a variable (Dec)
0 1 2 3
process status
process disabled (inactive) 0 0
dosing start 0 1
dosing stop 1 0
end of dosing process 1 1
byte no.
B1 B0
Inputs status – determines the status of inputs in a weighing module
MW-01:
input status byte no.
In2 In1 B1 B0
OFF OFF 0 0 0
OFF ON 0 1 1
ON OFF 1 0 2
ON ON 1 1 3
Decimal value of
a variable (Dec)
MIN – set value of MIN Threshold (in calibration/adjustment unit)
is returned.
MAX – set value of MAX Threshold (in calibration/adjustment unit) is returned.
Bulk (fast) dosing threshold – set value of bulk (fast) dosing
threshold (in calibration/adjustment unit) is returned.
Bulk (slow) dosing threshold – set value if fine (slow) dosing threshold (in calibration/adjustment unit) is returned.

15.4.2. Input variables

Saving the input variables into the weighing module MW-01 enables controlling module’s operation.
57
Page 58
Notice:
All input values are set in reation to calibration/adjustment unit.
Name of input variable address length [word] data type
command 0 1 word
complex command 2 1 word
Complex command parameters
tare 6 2 float
LO 10 2 float
output status 14 1 word
MIN 16 2 float
MAX 20 2 float
bulk (fast) dosing threshold 24 2 float
fine (slow) dosing threshold 28 2 float
Table. 4
command – basic command. Setting of an appropriate command’s
byte is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
0 scale zeroing 1
1 scale tarring 2
5 dosing process start 32
6 dosing process stop 64
task Dec
For instance: 0000 0000 0010 0000 – command carried out, dosing process start
complex command - Setting of an appropriate command’s byte
is directly responsible for carrying out a task, in accordance with below table:
58
Page 59
command
bytes
0 setting tare value 1
1
2 setting outputs status 4
3
4
5 setting the value of bulk (fast) dosing threshold 32
6 setting the value of fine (slow) dosing threshold 64
task Dec
setting LO threshold value
setting MIN threshold value setting MAX threshold value
2
8
16
For instance: 0000 0000 0000 0010 – command carried out, LO threshold set
to value given in LO parameter (address 10); See Table 4.
Notice:
A complex command requires setting an appropriate address parameter from 6 to 28. See Table.4.
tare – complex command parameter – tare value (in calibration /
adjustment unit).
LO – complex command parameter – LO threshold value (in calibration
/ adjustment unit).
outputs status – complex command parameter determining outputs status of the weighing module MW-01.
Output status byte no.
Out3 Out2 Out1 B2 B1 B0
OFF OFF OFF 0 0 0 0
OFF OFF ON 0 0 1 1
OFF ON OFF 0 1 0 2
OFF ON ON 0 1 1 3
ON OFF OFF 1 0 0 4
ON OFF ON 1 0 1 5
ON ON OFF 1 1 0 6
ON ON ON 1 1 1 7
Dec
59
Page 60
MIN - complex command parameter – MIN threshold value
(in calibration / adjustment unit).
MAX - complex command parameter – MAX threshold value
(in calibration / adjustment unit).
bulk (fast) dosing threshold – complex command parameter – the value of bulk (fast) dosing threshold (in calibration / adjustment unit).
fine (slow) dosing threshold – complex command parameter – the
value of fine (slow) dosing threshold (in calibration / adjustment unit).
Notice:
A command or a command with parameter is executed once on detecting its corresponding byte setting. Should a command with the same byte be re­executed, the byte must be reseted.
For instance:
command address 1 address 0
tarring 0000 0000 0000 0010
zeroing command bytes 0000 0000 0000 0000
tarring 0000 0000 0000 0010

16. MODBUS COMMUNICATION PROTOCOL

16.1. Basic information
The weighing module MW-01 enables complete operation of programmable controllers (PLC), which communicate through Modbus protocol class 0. The weighing module in a network using modbus standard operates as slave on
an accessible interfaces, RS485 or Ethernet. According to the standard,
Modbus class 0, the device uses two functions:
Read multiple registers (fc 3), for cyclic readout of signals from the weighing module,
Write multiple registers (fc 16), for cyclic record of the weighing module statuses.
60
Page 61
The weighing module MW-01 can operate in one of three modes:
RS484 - RTU (device interface RS485, frame RTU),
TCP/IP - RTU (device interface Ethernet, frame RTU),
TCP/IP - Open ModBus (device interface Ethernet, frame Open
Modbus).
The functionalitz of Modbus protocol is based on a scheme used in Profibus protocol, as described in point 15.3 and 15.4 of this user manual.
Caution:
Modbus address offset in the module that cooperates with PLC controller is set to value 1, i.e. deafult setting. In case of a weighing module cooperating with a test computer software, set the value to 0.
16.2. Addressing a device in ModBus network
Address of the weighing module MW-01 in modbus netowrk has to be set in accordance to module addressing specification. (see point 7.3.2 of this user
manual – „Communication settings”). Use tab RS232/485, field module address to set address of a device appropriate for the Modbus network.
The default port of the Ethernet interface (TCP/IP) in Modbus network is
set to 502.
16.3. Register tables in ModBus network
Numbers on the output and input register tables are described below,
and they are compatible with the Offset of Modbus address set for value
0. In case of a value setting from 0 do 255, add the set value of Offset of Modbus address to the output or input register number.
61
Page 62

16.3.1. Table of output registers

The weighing module uses 42 addresses for function Read Holding Registers as listed in below table.
Name of output variable address
Mass 0 2 float
Tare 4 2 float
Unit 8 1 word 4 Status 10 1 word 5
Lo 12 2 float
N. A.
N. A. N. A.
N. A. N. A.
N. A. N. A.
N. A. N. A.
N. A. N. A. N. A.
N. A. N. A.
N. A.
Process status (Stop, Start) 64 1 word 32 Inputs status 66 1 word 33
Low threshold value 68 2 float
High threshold value 72 2 float
Bulk (fast) dosing threshold value 76 2 float
value
N. A. – Register inactive.
16 2 float
20 2 float
24 1 word 12 26 1 word 13
28 2 float
32 2 float
36 2 float
40 1 word 20 42 1 word 21
44 2 float
48 2 float
52 2 float
56 1 word 28 58 1 word 29
60 2 float
80 2 float
Length [words
Data
type
Read Holding
Registers
0 1 2 3
6 7 8
9 10 11
14 15 16 17 18 19
22 23 24 25 26 27
30 31
34 35 36 37 38 39 40 Fine (slow) dosing threshold 41
62
Page 63

16.3.2. Table of input registers

The weighing module uses 16 addresses for function Write Multiple Registers as listed in below table.
Name of input variable address
Command 0 1 word 0 Command with parameter 2 1 word 1
N. A.
tare 6 2 float
Lo 10 2 float
Outputs status 14 1 word 7
MIN 16 2 float
MAX 20 2 float
Bulk (fast) dosing threshold 24 2 float
Fine (slow) dosing threshold 28 2 float
Length [word]
4 1 word 2
Data type
Write Multipe
Registers
3 4 5 6
8
9 10 11 12
13 14 15
N. A. - Register inactive.
16.4. Software for controller PLC SIEMENS S7-300
Description of software Siemens S7-300. The software enables communicating of the weighing module MW-01 with the Siemiens S7-300 controller through Open Modbus TCP protocol. The software is designed for below hardware specification. Unit CUP: S7-313 2DP (6ES7 313-6CE00-0AB0 /V1.0) Communication processor: CP 343-1 (6GK7 343-1EX10-0XE0 / V2.0)
Minimum software requirements: Programming environment: STEP 7 V5.4 + SP3 Additional libraries: SIEMENS OPEN MODBUS TCP
Description:
Exchanging data with a weighing module MW01 is carried out through
data block DB1 of the following structure.
63
Page 64
Name Type Description Data acquired from the MW01 weighing module when a flag is set to
Read_Write = FALSE
Mass REAL Mass from the weighing module
Tare REAL Tare value from the weighing module
Unit WORD Measuring unit from the weighing module
Status WORD Weighing status from the weighing module
LO REAL LO threshold the weighing module
Data saved to the MW01 weighing module when a flag is is set to Read_Write = TRUE
CMD WORD Command without a parameter sent to the module
CMD_PARAM WORD Command with a parameter sent to the module
RESERVED WORD N/A
Tara1 REAL Tare value for a command with a parameter
LO1 REAL Value of LO threshold with a parameter
OutPutsState WORD Output status for a command with a parameter
MIN REAL Min value for for a command with a parameter
MAX REAL Max value for a command with a parameter
FastDosing REAL Value of bulk (fast) dosing threshold for a command with
SlowDosing REAL Value of fine (slow) dosing threshold for a command with
Read_Write BOOL Flag for switching between date readout and record
a parameter
a parameter
17. ERROR MESSAGES
Err2 Err3 Err8 NULL FULL HI
LH
- Value beyond zero range
- Value beyond tare range
- Exceeded time of tarring / zeroing operation
- Zero value from A/D converter
- Measuring range exceeded
- Display range exceeded Start mass error, indication out of range
­(from -5% to +15% of start mass)
64
Page 65
BALANCES AND SCALES
RADWAG BALANCSES AND SCALES
POLAND, 26 – 600 Radom, Bracka 28
Phone: +48 48 38 48 800, fax. + 48 48 385 00 10
export@radwag.com
www.radwag.com
65
Loading...