MAC 50/1
M AC 50
MAC 110
MAC 210
Max capacity
50 g
50 g
110 g
210 g
Readability
0,1 mg
1 mg
1 mg
1 mg
Tare range
- 50 g
- 50 g
- 110 g
- 210 g
Max sample
mass
50 g
50 g
110 g
210 g
Readability of
readout
Repeatability
reading
+/-0,1% (sample ~ 2g), +/-0,02% (sample ~10g)
Drying
range
max. 160 °C
Heating
halogen (WH)
heater in metal cover (NS)
Drying modes
4 drying modes: standard, quick, step, mild
Auto switch off
option
3 modes: automatic, time defined, manual
Working
temperature
+10 °C - +40 °C
Power supply
230V 50Hz AC / 400W
Display
LCD (backlit)
Pan size
Φ 90 mm, h = 8 mm
Drying
dimensions
120 x 120 x 20 mm
Net weight /
Gross weight
4,9 / 6,4 kg
Packaging
dimensions
470 x 380 x 336 mm
1. TECHNICAL DATA
Type
moisture
of moisture
content
temperature
module
0,0001% 0,001 %
max. 250 °C (WH)
infrared emitter
chamber
6
WARNING
Dangerous situations of high risk. These
CAUTION
Dangerous situations of low risk. These
situations can cause damage of the
instrument or its functions, loss of data as
ATTENTION
Important information related to the moisture
Dimensions:
2. GENERAL SAFETY INFORMATION
2.1. Definitions of signals and warning signs
Safety precautions are marked with special descriptions and warning signs.
They inform and warn a user of possible dangers. Ignoring the safety warnings
may cause injuries, damage of the moisture analyzer, its inappropriate
operation and errors of measurements.
2.1.1. Warning descriptions
situations can cause serious injuries or death
if the safety precautions are not taken.
well as minor or moderate body injuries.
analyser.
2.1.2. Warning signs
Electric shock risk
7
Nominal voltage for a moisture analyzer is 230 VAC. It
autions for low voltage
core power supply cable with grounding pin comes
If necessary, an extension cord can be used as long as it
s and has a protective ground
conductor. Intended disconnection of the grounding cable is
Acid / Corrosion
Danger
Flammable or explosive substances
Toxic substances
Hot surfaces
2.2. Safety information
CAUTION!
The use of moisture analyzer MA 3Y regardless of both, safety
information and user manual guidelines may cause health damage and
even death
WARNING:
means that safety usage prec
devices must be abide by while operating the device.
Threestandard with a moisture analyzer.
meets
the applicable standard
8
forbidden.
CAUTION:
Drying chamber cannot be opened during operation (drying
process). It is due to the fact that a halogen lamp and its
prevent heat from building up and to keep your analyzer
drying chamber), the moisture analyzer must be switched
Wait until all the components have cooled down. Do not
Same types of samples require taking particular safety precautions.
They can pose a danger for people and objects. It is always the user who
is liable for possible damages caused by the use of an inappropriate
sample.
Substances that release aggressive vapours (e.g. acids)
the heating process. In this case, it is recommended to
housing parts and cause corrosion.
glass shield may reach the temperature of up to 400 C.
When setting up the moisture analyzer leave enough space
to
from
overheating. Leave about 20 cm around the instrument and
about 1 m above.
Air vents that are located in the housing cannot be covered,
sealed or blocked in any other way.
Do not put any flammable substances on, under or near the
moisture analyzer.
Be particularly careful when removing the sample from the
drying chamber: the sample itself, the drying chamber,
shields
and the pan can still be extremely hot.
In case of any maintenance work (cleaning the inside of the
off.
any modification to the heating module.
CAUTION:
perform
Corrosion
during
work with
small samples. Otherwise, vapours can condense on cold
WARNING:
Fire or explosion
9
Flammable or explosive substances, substances containing
solvents or releasing flammable or explosive gases or
vapours.
For this type of samples apply the drying temperatures as
low as possible to prevent flames or explosion. During the
necessary to wear protective glasses and gloves. The
irritations (eyes, skin or respiratory system), illnesses or
Perform a risk analysis when in doubt relating to the sample
characteristics before carrying out the procedure.
analysis it is
samples
should be relatively small.
Under no circumstances can the instrument be left
Unsupervised!
WARNING:
Substances containing toxic and caustic or corrosive
components
Substances that release toxic gases or vapours can cause
even death. Dry such substances only in fume hood.
Under no circumstances should the instrument be used in an area with
any
risk of explosion. The moisture analyzer is not designed to operate in
hazardous areas.
2.3. Intended use
A moisture analyzer MA 3Y series is designed to determine relative moisture
content in small samples of various substances, determine dry mass content in
small samples and determine mass of weighed objects.
MA 3Y ensures fast and precise determination process of water content in a
tested sample, and application of touch screen display considerably simplifies
operation and improves functionality of the moisture analyzer. A moisture
analyzer MA 3Y series can be used to determine humidity content of different
materials.
At the initial stage of measurement, the device precisely determines the mass
of an object placed on instrument’s weighing pan. As the mass reading is
stabilized, the sample is quickly heated by a halogen lamp, an IR emitter or a
metal heater causing humidity evaporation from the tested sample. While
sampling, the moisture analyzer is continuously checking the loss of mass, and
on calculation, it displays current moisture content in a tested sample.
Compared to conventional methods of humidity content determination of
various substances, application of moisture analyzer MA 3Y series
10
considerably shortens measurement time and simplifies testing procedure. MA
3Y series allows for setting multiple parameters which influence the procedure
of moisture content determination in a sample, such as: temperature, time,
drying modes, etc.
2.4. Inappropriate use
CAUTION!
Do not open the drying chamber during drying process. Moisture analyzer
features a halogen lamp which is a very powerful heat source. Thus, user
should pay special attention no to touch those elements of a moisture analyzer
that get hot while drying procedure (i.e.: disposable pan, pan handle, and inner
shields of the drying chamber).
Remember that some of tested samples may become dangerous if heated
(appearance of poisoning vapours, danger of ignition or explosion).
Moisture analyzer MA 3Y series is not intended for dynamic weighing. Even if
small amounts of a sample are added to taken off the weighing pan, the mass
readout should only be taken only on stabilization of measurement result
(appearance of stability pictogram
on the display).
Do not place any magnetic materials on the weighing pan, as this can cause
damage of the measuring system of the instrument. Be sure to avoid impact
shock and overloading the moisture analyzer in excess of the prescribed
maximum measuring range (max capacity), minus any possible tare weight that
has been applied).
Never use the moisture analyzer in an environment endangered by an
explosion!
This moisture analyzer is not adjusted for operation in explosive areas. There
must not be any modification made to the moisture analyzer.
2.5. Principles of safety use
Moisture analyzer is in accord with all binding safety regulations. Nevertheless
there are exceptional circumstances that may cause danger.
It is forbidden to open an instrument housing. Inside there are no parts that
would require maintenance, repair or replacement carried out by a user. In
case of any problems it is recommended to contact RADWAG service or a
distributor. It must be always remembered to use the device in accordance with
this service manual and especially to follow instructions on installation and
configuration of a new device.
Use of a moisture analyzer MA 3Y series conversely to safety principles and
user manual guidelines may be hazardous to operator’s health and life. It is
obligatory to acknowledge the safety principles listed in the user manual:
- use a moisture analyzer only to determine humidity content in
samples and determine mass of a tested sample. Any other use of
the moisture analyzer may be dangerous both to the device and the
user,
- before commissioning the moisture analyzer, make sure that the
nominal power of the device specified on its data plate, is compatible
11
with the supply in the mains to which the moisture analyzer will be
plugged in,
- replecemant of halogen lamps can only be carried out by authorized service,
- protect moisture analyzer against contacts with liquids, it might lead
to electrocution, fire, emission of substances containing toxic or
caustic vapour, emission of explosive substances.
2.6. Warranty
Warranty does not cover the following cases:
• Not observing the regulation listed in user manual,
• Using the moisture analyzer conversely to its intended use,
• Any modifications of moisture analyzer or cases when its housing is
opened (damaged protective stickers),
•Mechanical defects and defects caused by media, liquids, water and
natural wearing off,
• Improper placing of defects of electrical network / mains,
• Overloading of measuring mechanism of the moisture analyzer.
2.7. Monitoring metrological parameters of the instrument
Metrological characteristics of the moisture analyzer requires periodical
inspection carried out by its operator. Inspection frequency is conditioned by
ambient conditions in which the moisture analyzer is used, types of performed
processes and accepted quality management system in an organization.
2.8. Data included in this user manual
Please read the user manual carefully before instrument’s plugging to mains
and startup, even if the user is experienced with this type of moisture
analyzers.
2.9. Staff competence
Moisture analyzer MA 3Y series should only be operated and maintained by
personnel who is trained and experienced in using this type of instruments.
In order to use the moisture analyzer, first read the user manual. Keep these
instructions for the future reference.
Do not make any structural modifications. Additional equipment, which can be
connected to the moisture analyzer, should be supplied by RADWAG or an
authorized distributor.
12
2.9.1. Protective clothing
The use of protective clothing is highly recommended while working with the
instrument in order to take safety precautions against potential hazards
resulting from tested samples and ingredients.
Use the following while carrying out tests:
• protective apron,
• protective glasses,
• protective gloves (while working with hazardous chemical
substances).
Before the use of the above-mentioned protective clothing, make sure that they
are designed to be used with specific samples and they are not damaged.
3. TRANSPORT I STORAGE
3.1. Delivery check
Please check the packaging immediately upon delivery and the device during
unpacking for any visible signs of external damage.
3.2. Packaging
Please retain all parts of the original packaging should the moisture analyzer
be transported in the future. Only the original packaging should be used for
dispatching the moisture analyzer. Before packing, disconnect all attached
cables and remove any loose/movable parts (weighing pan, shields, inserts).
Place moisture analyzer and its components in their original packaging, and
protect them against damage during transport.
4. UNPACKING, ASSEMBLING AND STARTUP
4.1. Assembling and place of use
•A moisture analyzer should be stored and used in locations free of
vibrations and shakes, free of air movement and dust, located at the
max altitude of 2000 above sea level,
•Moisture analyzer’s place of use should ensure good air circulation
around the instrument (approximately 20 cm fee space around the
moisture analyzer and 1 m free space over it),
•Ambient air temperature in the weighing room should not exceed the
range of:
+10 °C ÷ +40 °C,
13
•Ambient relative humidity should not exceed 80% in the temperature
up to 31°C, and decrease linearly to 50 % of relative humidity in
temperature 40°C,
•A moisture analyzer should be located on a stable wall console desk
or a stable working table which is not affected by vibrations and
distant from heat sources,
•Take special safety measures when weighing magnetic objects, as
part of the balance is a strong magnet.
4.2. Unpacking
Carefully remove the moisture analyzer from its packaging, remove the plastic
and foil transport protective elements. Gently place the moisture analyzer in its
intended place of use.
Assembly the components of drying chamber, following guidelines from below
figure:
Assembling moisture analyzer components:
- Assembly insert of the drying
chamber’s basis
- Assembly drying chamber
cover,
- Assembly drying pan handle,
- Assembly drying pan bracket,
- Assembly the disposable drying
pan.
Setting drying pan’s supporter (bracket).
14
4.3. Level setting
When assembling the pan supporter
(bracket) pay special attention to the
correct positioning of its mandrel. The
mandrel features a cut for its unique
positioning against the pan handle, thus
preventing their contact and friction.
Setting pan’s bracket:
- On assembling the pan bracket onto
the mandrel, turn the bracket slightly,
so that the cut on the mandrels are
located in their unique and correct
position,
- When turning the mandrel use most
gentle movements not to damage
moisture analyzer’s measuring
system.
Before plugging to mains, level the moisture analyzer using
adjustable feet. Turn the feet in a way that the air bubble of
the level is located centrally.
4.4. Standard delivery components
• A moisture analyzer MAC series
• Insert of drying chamber basis.
• Drying chamber shield.
• Drying pan handle.
• Drying pan bracket.
• Disposable pan.
• Power cord.
• User manual (CD).
4.5. Cleaning
Remember to unplug the moisture analyzer from mains before any cleaning
activities.
Clean the moisture analyzer using a damp cloth by gentle rubbing
contaminated places. Remember to remove the drying pan from the drying
chamber before its cleaning. If the drying pan assembly is installed while
being clened it may damage instrument’s measuring system, thus
remove it from the bracket.
15
4.6. Plugging to mains
A moisture analyzer can be plugged to mains only by means of original
power cord, which comes standard with the moisture analyzer. Rated
voltage (specified on device’s data plate) must be compatible with mains
rated voltage.
The power cord can be connected only to socket with ground contact. Plug the
power cord to the moisture analyzer. The moisture analyzer’s power plug is
located at the back of instrument’s housing.
On plugging to mains moisture analyzer’s displays software’s name and
number, after which the indication changes to 0.000 g (in case of a moisture
analyzer with measuring accuracy 1 mg) or 0.0000 g (in case of a moisture
analyzer with measuring accuracy 0,1 mg). In mass indication is other than
zero press O/T key located on the overlay.
4.7. Connecting peripheral equipment
The moisture analyzer must be unplugged from the mains before connecting or
disconnecting any peripheral equipment (printer, PC computer). Use only
peripheral equipment recommended by the manufacturer with your moisture
analyzer. These have been ideally coordinated to your moisture analyzer. On
connecting a peripheral device, plug the moisture analyzer to mains.
16
/ ending drying process
key, abandoning introduced changes / exit one
change of parameters
sending display status to
peripheral device (Print) or confirmation of selected
zeroing moisture analyzer’s
On/Off key – switch on/off display of moisture
other subassemblies are continuously supplied, and
the moisture analyzer is in standby mode.
5. KEYBOARD – FUNCTION KEYS
Start/Stop key – starting
according to selected drying mode.
Esc
level higher in menu structure.
Group of navigation keys –
value; moving in moisture analyzer’s menu.
Print/Enter key –
value or function (Enter).
ZERO/TARE key –
indication.
analyzer. On switching of the display of the device,
17
changes type of data displayed during
for selecting drying
and after drying process.
Display key –
Setup key – enter to main menu.
Test menu key – function key
modes.
6. START-UP
On plugging to mains the moisture analyzer carries out a display test (all
pictograms and digits appear on the display). Then the display shows software
name and number, and enters weighing mode.
6.1. Moisture analyzer temperature stabilization period
Before start of measuring processes, it is necessary to wait until the moisture
analyzer is thermally stabilized. It is a period of so called moisture analyzer’s
self-heating.
For moisture analyzer MAC series which before plugging to mains was stored
in room temperature, self-heating period takes approximately 30 minutes. For
moisture analyzers that were stored in much lower temperatures before
plugging to mains (e.g. during winter period) thermal stabilization should last
approximately 4 hours.
During self-heating period the indications on moisture analyzer’s display may
change. Correct operation of a moisture analyzer is possible within
temperature range specified in the technical parameters, see point 1.
It is recommended that ambient temperature changes at moisture analyzer’s
place of use are very small (slow).
18
7. USER MENU
Selecting group of parameters one by
one upwards / changing parameter value by one value
activating. On pressing the key, the display indicates the
User menu is divided into 5 basic function groups.
7.1. Moving through user menu
An operator moves through moisture analyzer’s menu using keys on overlay’s
keyboard:
Setup key. Entering main menu.
Navigating key UP –
upwards.
Navigating key DOWN – Selecting group of parameters one
by one downwards / changing parameter value by one value
downwards.
Navigating key RIGHT – Selecting group of parameters for
first parameter in a selected group.
Navigating key LEFT – Exit to previous menu level, e.g. to
main menu.