KR-04 converter has been designed to be used with scales or terminals that are not equipped
with Ethernet interface. It enables connection of scales with RS232 interface to LAN network.
Converter types:
KR-04-1: plastic housing, RS232 cable with DB9/F connector. Intended for scales equipped
with PUE C31 (plastic housing), laboratory balances and other devices featuring RS232
interface with DB9/M connector. KR-04-1 converter is equipped with Ethernet interface of RJ45
type.
KR-04-2: stainless steel housing, RS232 cable with hermetic M12 connector. Intended for
scales equipped with the following terminals: PUE C/31H, PUE 4/7. KR-04-2 converter features
Ethernet cable fed through cable gland.
KR-04-3: stainless steel housing, RS232 cable with hermetic M12 connector. Intended for
scales equipped with the following terminals: PUE C/31H, PUE 4/7. KR-04-3 converter features
hermetic Ethernet M12 connector.
Page 5
- 5 -
3.1. External view
KR-04-1 converter
1 – Ethernet RJ45 port
2 – RS232 cable with DB9/F connector
3 – power supplier
Dimensions of KR-04-1 converter (plastic housing)
Page 6
- 6 -
KR-04-2, KR-04-3 converter
Dimensions of KR-04-2, KR-04-3 converter ( stainless steel housing)
Page 7
- 7 -
3.2. Connectors
KR-04-1 interfaces
1 – Ethernet RJ45 port
2 – cable bush: power supplier
3 – cable bush: RS232 cable (cable with RJ45 connector)
KR-04-2, KR-04-3 interfaces
1 – power cord cable gland
2 – RS232 cable gland (cable with a hermetic connector)
KR-04 converter setup is carried out using Network Enabler Administrator PC software
(installed on a computer). The computer and the converter communicate via Ethernet.
Procedure:
1. Install Network Enabler Administrator PC software.
2. Establish communication between the computer and the converter via LAN, use crossed
cable terminated with RJ45 connectors on both sides. For KR-04-3 converter use
RADWAG-manufactured P0198 cable and a switch.
3. Connect the converter to the mains.
4. Run Network Enabler Administrator PC software.
5. Click „Zoom” icon, the software will search for network-connected converters.
Page 10
- 10 -
6. List of converters operating in LAN network is displayed.
7. Select the respective converter (double click using left mouse button).
8. Go to „Configuration” window, select „Network” tab.
9. Tick „Modify” box by field that is to be changed.
10. Enter all necessary data.
11. Select „Serial” tab – RS232 setup.
Page 11
- 11 -
Flow Control
NONE
FIFO
Enable
Interface
RS232
Baud rate
9600
Parity
NONE
Data bits
8
Stop Bits
1
12. Tick „Modify” box and select Port that is to be modified (double click using left mouse
button).
Default settings:
Page 12
- 12 -
13. Select „Operating Mode” tab.
By default „Operating mode” is set to „TCP Server Mode”.
By default „Local TCP Port” is set to 4001.
14. Press „OK” button to confirm changes.
15. Exit Network Enabler Administrator PC software.
Setup parameters may vary depending on individual needs.
Network Enabler Administrator PC software is to be downloaded from www.radwag.com.
Page 13
- 13 -
Page 14
-14-