HRP.R, the series of professional high resolution weighing platforms, perfectly suits
every single branch of industry requiring great weighing accuracy, precision and
repeatability of measurement in challenging industrial environment.
Automatic adjustment performed by means of an internal adjustment weight is
a warranty for a remarkably effective operation and repeatability of indications. The
automatic adjustment saves time and lets to avoid performing the process
manually, making it comfortable and easy.
High resolution weighing platform, HRP.R, features the following interfaces: 1xRS
232, 2xIN (PRINT/TARE).
Data safety and archiving is assured by software dedicated for RADWAG terminals.
Features:
•Brand new design of the electromagnetic mechanism housed in a
stainless steel tight casing,
• Overload protection,
• An in-build adjustment system with and adjustment weight,
• IP rating: IP67,
• Interfaces: RS232, and dedicated 2 in, enabling cooperation with external
buttons (TARE/PRINT),
- 9 -
Page 10
1.3 Usage precautions
Do not use the scale as a dynamic weighing instrument. Even if small quantities of
weighed material are added or removed from the weighing pan of the instrument,
the reading should be taken only after stabilisation of the measurement results. Do
not place any magnetic materials on the weighing pan, as this can cause damage
to the measuring system of the instrument.
Be sure to avoid impact shock and overloading the scale in excess of the
prescribed maximum measuring range (max capacity), minus any possible tare
weight that has been applied.
Never use the scale in an environment where explosion is possible. This scale has
not been adjusted for operation in hazardous areas.
There must not be any modification made to the scale.
1.4 Warranty
Warranty is invalid for the following:
• non- observation of the guidelines of this user manual,
• use of the scale other than specified in this manual,
• alteration to or opening of the device,
• mechanical damage and damage caused by media, water, wear and tear,
• inappropriate assembling or defects of electric installation,
• overloading of the measuring instrument.
1.5 Metrological par ameters monitoring
Metrological characteristics of the scale require periodical inspection to be carried
out by its user. Inspection frequency is dependent on ambient conditions in which
the scale is used, types of performed processes and accepted quality management
system in organization.
1.6 User manual significance
Please read the user manual carefully before assembling and start-up, even if you
are experienced with this type of weighing instruments.
1.7 Staff training
This scale should only be operated and maintained by personnel who is trained and
experienced in using this type of scales.
- 10 -
Page 11
2. TRANSPORT AND STORAGE
2.1 Delivery check
Please check the packaging immediately upon delivery, while unpacking the device
make sure that there are no visible signs of external damage.
2.2 Packaging
Please retain all parts of the original packaging should the scale be transported in
the future. Only the original packaging shall be used for shipping the scale. Before
packing, disconnect all attached cables and remove any loose/movable parts
(weighing pan, anti-draft shield, etc.). Please place scale and its components in the
original packaging, thus protecting them against damage during transport.
- 11 -
Page 12
3. UNPACKING, INSTALLING AND TESTING
3.1 Place of use and ins tallation
•The scale should be stored and used in locations free of vibrations and
shakes, and free of air movement and dust.
• Ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C.
• Ambient relative humidity should not exceed 80%.
• During operation, ambient temperature in the weighing room should not
change more than 3 °C within one hour.
•The device should be located on a stable wall console desk or a stable
working table, not affected by vibrations, distant from heat sources.
•Since part of the scale is a strong magnet it is necessary to take special
precaution when weighing magnetic objects. Should such loads be
weighed, use under-pan weighing option, with this the weighed load is
removed from an area influenced by the magnet. The hook for under-pan
weighing is installed in scale base.
•In order to avoid the influence of static electricity on the measurement
process, ground the scale housing. The grounding bolt is located at the
back of the scale housing.
3.2 Unpacking
Cut the sealing tape. Take the scale out of the packaging. Open the accessories
box and take all necessary components. Upon placing the scale in the desired
location, assembly its weighing pan and other parts following an instruction.
Be as gentle as possible so as not to damage the scale mechanism by accident.
- 12 -
Page 13
HRP 150 platform installation
1. Take the scale out of its packaging, make sure that you touch „OUTER
FRAME” only.
2. Connect the terminal and the weighing module.
- 13 -
Page 14
3. Install mandrels for the weighing pan.
4. Disassembly transport lock by undoing screws and removing fixing washers.
- 14 -
Page 15
5. Install the weighing pan.
3.3 Levelling
Before plugging the device to the mains, level it by turning its
feet. Keep turning the feet until an air bubble takes central
position of the level indicator ring. Make sure that your device is
placed firmly on all its feet.
3.4 Maintenance and cleaning
1. Be careful while cleaning the device so as not to cause any damages to
the mechanism.
2. Use a dry flannel cloth in order to clean the device components (mild
cleanser may be used if it does not contain any abrasive substances).
3.5 Connecting to the mains
The device can be connected to the mains only with a power adapter that comes
standard with it. Nominal power supply of the power adapter (specified on the data
plate of the power adapter) must be compatible with the power from the mains.
Plug the scale to the mains – the power supply socket is located at the back of the
housing.
- 15 -
Page 16
First, display unit test is carried out (all the elements and pictograms are backlit for
a short time), next, the name and the program number appear consequently, the
indication is ZERO (with scale-related reading unit). During scale start-up, internal
adjustment mechanism is tested (single location and elevation of the internal mass
adjustment).
If the indication is different from zero, press
button.
3.6 Connecting peripherals
It is inevitable to disconnect the scale from the mains before coupling it to any
peripheral devices (printer, PC computer, additional display, computer keypad,
etc.). Follow this rule also when replacing any coupled peripheral device with a new
one.
Connect only the equipment recommended by the manufacturer.
When the devices are coupled, plug the scale to the mains.
- 16 -
Page 17
4. BASIC FUNCTIONS
Unit-related functions:
• weighing in grams [g],
• weighing in milligrams [mg] (devices with d≤0.001g),
• weighing in kilograms [kg] (devices with d≥0.01g,
• setting parameter of measurement result averaging,
• value release,
• display backlight,
• beep signal.
Communication ports related functions (RS232, USB, Wi-Fi)
• setting communication port,
• setting baud rate value,
• continuous data transmission in basic measuring unit,
• continuous data transmission in current measuring unit,
• automatic operation mode.
Device can be used to weigh loads under the weighing platform – i.e. under-pan
weighing option. The software enables generating a report from adjustment
process in a form of a printout containing modifiable data (see CAL REPORT).
- 17 -
Page 18
5. KEYBOARD
Press to switch the display ON/OFF. If switched off,
F9 button of the computer keyboard.
ed for particular
F10
send current display status to a peripheral device
accept selected value of a parameter or
function (ENTER).
components other than the display are powered, the device
is in stand-by mode.
Press to enter databases and data record
database: user, product, tare.
button of the computer keyboard.
Press to enter functions of an active operating mode.
F11 button of the computer keyboard.
Press to select a particular operating mode.
F5 button of the computer keyboard.
Press to change measuring unit.
Press to
(PRINT) or to
Press to zero the indication.
Press to tare the indication.
Press to start adjustment process.
F6 button of the computer keyboard.
- 18 -
Page 19
Press to enter home menu.
Navigating arrows for operating menu or changing
F7 button of the computer keyboard.
parameter value.
- 19 -
Page 20
6. START-UP
When plugged to the mains, the scale displays program name and number, next it
proceeds to the weighing mode.
6.1 Temperature stabilization period
Prior to measurement it is necessary to wait until the scale stabilizes thermally,
For scales stored in much lower temperatures before plugging to the mains (e.g.
during winter period), thermal stabilisation period shall take 4 – 8 hours. During the
thermal stabilization, the indications on the display panel can change. For correct
operation it is recommended that ambient temperature changes at place of use
were insignificant (slow to change).
- 20 -
Page 21
7. USER MENU
group of parameters for activation.
Press to delete a character while editing text values, e.g. user
The menu comprises 6 parameter groups. Each group features its own symbol
starting with a capital letter, P.
7.1 Operating user menu
The user operates the menu pressing overlay buttons:
Press to enter the main menu.
Press to select group of parameters one by one downwards or to
change parameter value by one value down.
Press to select group of parameters one by one upwards or to
change parameter value by one value up.
Press to select a particular
On pressing the button, the display indicates the first parameter
of a selected group.
Press to exit to the previous menu level, e.g. to the main menu.
Press to abandon parameter changes.
Press to accept / confirm introduced changes.
name.
- 21 -
Page 22
Press to i
nsert a character while editing text values, e.g. user
P1
ADJUSTMENT
P1.1
INT.CALIB
| [internal adjustment]
P1.2
EXT.CALIB
| [external adjustment]
P1.3
USER CALIBRATION
| [user adjustment]
P1.4
CALIBRATION TEST
| [adjustment test]
P1.5
AUTO.CALIB.
|
NONE
[automatic adjustment]
P1.6
AUTO.CALIB.C
| [time of automatic adjustment]
P2
OPERATING MODE
P2.1
ACCESSIBILITY
| [setting accessibility of individual
modes in course of operation]
P2.2
WEIGHING
| [weighing function settings]
P2.3
PARTS COUNTING
| [parts counting function settings]
P2.4
CHECKWEIGHING
| [checkweighing function settings]
P2.5
DOSING
| [dosing function settings]
P2.6
PERCENT
WEIGHING
| [percent weighing function
settings]
P2.7
DENS. OF SOLIDS
| [settings for determinig solids
density]
P2.8
DENS. OF LIQUIDS
| [settings for determinig liquids
density]
P2.9
ANIMAL WEIGHING
| [animal weighing function settings]
P2.10
STATISTICS
| [statistics function settings]
P2.11
TOTALISING
| [totalising function settings]
P2.12
PEAK HOLD
| [peak hold function settings]
P2.13
PIPETTES CALIB.
| [pipettes calibration function
settings]
P3
COMMUNICATION
P3.1
COM 1
| [transmission parameters port
COM 1]
P3.2
COM 2
| [transmission parameters port
COM 2]
P3.3
WIFI
| [transmission parameters port
WIFI]
P4
DEVICES
P4.1
COMPUTER
|
COM 1/COM
2/USB/WIFI/NONE
[PC connection port]
CONT.TRANSM.
|
CALIBRATION
UNIT/NONE
[]turning on/off
transmission
name.
UNIT/CURRENT
- 22 -
continuous
Page 23
INTERVAL
|
0.1
[setting intervals for
continuous
E2R SYSTEM
|
YES/NO
[turning on/off
cooperation with E2R
system]
P4.2
PRINTER
|
COM1/COM2/USB/WI
PC/NONE
[printer connection port]
P4.3
BARCODE
READER
|
COM1/COM2/NONE
[barcode reader port]
P4.4
ADD.DISPLAY
|
COM1/COM2/NONE
[additional display port]
P4.5
EXT.BUTTONS
|
TARE/PRINT
P5
PRINTOUT
P5.1
CAL.REPORT
| [contents of the adjustment report]
P5.2
HEADER
| [contents of the header printout]
P5.3
GLT PRNT.
| [contents of the weighing result
prnt.]
P5.4
FOOTER
| [contents of the footer printout]
P5.5
NSD.PRN.1
| [project of non-standard printout 1]
P5.6
NSD.PRN.2
| [project of non-standard printout 2]
P5.7
NSD.PRN.3
| [project of non-standard printout 3]
P5.8
NSD.PRN.4
| [project of non-standard printout 4]
P5.9
VARIABLE 1
| [project of variable 1]
P5.10
VARIABLE 2
| [project of variable 2]
P6
MISC
P6.1
LANGUAGE
|
POLISH/ENGLISH
[menu language]
P6.2
ACCESS LEV.
|
ADMIN
[access levels for
editing menu]
P6.3
KEY SOUND
|
YES/NO
[key sound]
P6.4
BACKLIGHT
|
70
[display backlight
level]
P6.5
BACKLIT OFF
|
NONE/0.5/1/2/3/5
[backlight turn-off
time interval]
P6.6
AUTO OFF
|
NONE/1/2/3/5/10
[display turn-off time
interval]
P6.7
DATE
|
2015.05.27
[date settings]
P6.8
TIME
|
8:53:00 A
[time settings]
P6.9
DATE FORM.
|
YYYY.MM.DD
MM.DD.YYYY
[date format]
P6.10
TIME FORM.
|
12H/24H
[time format]
P6.11
GLP
AUTOTEST
| [carrying out autotest
for the device]
transmission]
FI/PENDRIVE/USB
YYYY.DD.MM
DD.MM.YYYY
- 23 -
Page 24
P7
INFO
P7.1
BALANCE ID
|
P7.2
SCALE TYPE
|
P7.3
PROG. VER.
|
P7.4
TEMP.
| [temperature]
P7.5
SETUP PRNT.
| [scale parameters
printout]
P8
UNITS
P7.1
ACCESSIBILITY
| [declaration of units to
operation]
P7.2
START UNIT
| [selection of a start
unit, unit active on
scale start-up]
P7.3
USER UNIT U1
| [parameter defining
user unit 1]
P7.4
USER UNIT U2
| [parameter defining
user unit 2]
7.2 Return to weighing
CAUTION!
Introduced modifications will be saved permanently upon
abandoning the menu (on return to weighing).
Press ESC button several Times.
be available for scale
- 24 -
Page 25
8. LOGGING IN
In order to have full access to user parameters and be enabled to edit databases,
the scale operator should log in as <ADMINISTRATOR> each time running the
scale. The software enables the entry of 10 users with various access rights.
First logging in procedure
•Go to the main window and press button or a respective function
button (to which the < LOG IN> shortcut is assigned), next select <LOG
IN> option; or press
button to enter User Database, select
<ADMIN> user.
•Press button for confirmation, dialog window for entering the
password is displayed .
• Enter the password:„1111”, and press button for confirmation.
• The program returns to the main window.
• When logged for the first time, enter users and assign them with
appropriate permissions levels (read section 15 for detailed information)
The next time you log in, select a user from the list and enter a respective
password. The program is run with access level of the selected user.
If any user is logged in, a pictogram
is displayed.
Logout procedure
• Select <NONE> option of the list of available users
• The program returns to the main window, the display shows no logged-in
user state (no logged-in user, no pictogram on the display).
Permissions levels
The software comprises three permissions levels: administrator, advanced
operator, user.
On switching on, the display is constantly active enabling carrying out measuring
processes regardless of the fact that no user is logged in.
- 25 -
Page 26
Permissions
level
Operator
Free editing of parameters of <Readout> submenu.
The operator can preview information recorded in
Advanced
Free editing of parameters of the following submenus:
<Readout>; <Working modes>; <Communication>;
Administrator
Access to all user parameters and functions, editing
Permissions level dependent access to edition of user parameters, databases and
software functions.
Available operations
Modification of settings for <Misc.> parameter group, except
for settings for <Date and Time>.
The operator can start and carry out all weighing processes.
<Databases>, (s)he can define universal variables.
operator
<Peripherals>; <Misc.>.
Access to <Date and Time> submenu denied.
The advanced operator can start and carry out all weighing
processes.
databases enabled.
- 26 -
Page 27
9. WEIGHING
YES
NO
YES
NO
NO
NO
9.1 Good weighing practic e
In order to ensure long lasting use of the device and reliable measurement of
weighed loads, follow the guidelines below:
•Start the scale with no load on a weighing pan (permissible value of load
on the weighing pan on scale start is ±10% of its maximal capacity).
• Load the weighing pan steadily avoiding shocks:
• Place weighed loads centrally on the weighing pan:
• Avoid side loading, in particular side shocks:
The scale requires adjusting before running the weighing process or in case of
drastic change of ambient conditions at a workstation (e.g. ambient temperature
- 27 -
Page 28
variation at a workstation greater than 3°C). The procedure of scale adjustment is
described further down this user manual.
•Prior to measurement, it is recommended to load the weighing pan a few
times with mass close to scale max capacity.
•Check if unloaded scale indicates “precise zero” and whether
measurement is stable
• Press UNITS button, to set a measuring unit:
• [g], [mg], [ct], and if enabled in a service mode (non-verified scales) [kg],
•Place weighed object on the weighing pan and read result only on
measurement result stabilization.
•Mass indication of a load placed on the weighing pan can be tarred
multiple number of times by pressing →T←/Insert (pay CAUTION not to
exceed maximal capacity of a scale by applying multiple tare function).
The scale shall stay plugged to the mains in between measurement series. It is
recommended to switch off the scale display by pressing ON/OFF button. On
repeated ON/OFF button pressing, the scale is ready for operation, the
measurement can be carried out again.
9.2 Zeroing
Zeroing is a function allowing to zero mass indication. In order to zero mass
indication, press button.
Mass indication of zero value shall be displayed together with precise zero
and stability
Zeroing process is an equivalent for determining new zero point, recognized by the
scale as precise zero. Zeroing is possible only for stable status of display
indication.
markers.
- 28 -
Page 29
CAUTION!
Zeroing the display indication is possible only within the ±2% range of instrument’s
maximal capacity. If the zeroed value is above the ±2% of the maximal capacity,
then the software indic ates an error message Err2.
9.3 Tarring
Tarring is a function allowing to determine net weight of a measured object. In order
to determine net weight of the object, place object’s container (packaging) on the
weighing pan, and on stabilization of measurement result press
key. The
display should indicate mass equal to zero and symbols: Net and . On taking
off the weighed load and its packaging from instrument’s weighing pan, the display
indicates sum of total tarred mass with minus sign.
The software also enables assigning tare value to a product from a database.
Then, on product selection from a database, the software automatically uploads
data on tare value for the specific product.
CAUTION!
Tarring negative values is impossible. On tarring negative values the scale
responds with an error message Err3. In such case zero scale indication and
repeat tarring procedure.
Manual tare determination
Procedure:
•For any operating mode, press quick access button F, to which the option
<ENTER TARE> has been assigned in <SHORTCUTS> settings (read <F shortcut keys> secti on for det a iled in struc tio n) .
• Wait for a respective window to open.
• Use arrow buttons to enter tare value and press button.
• The scale returns to weighing mode, the display indicates entered tare
value with minus „–" sign.
Selecting tare out of TARE DATABASE
Procedure:
•For any operating mode, press quick access button F, to which the option
<SELECT TARE> has been assigned in <SHORTCUTS> settings (read
<F shortcut keys> section for detailed ins truction), or press button
firts and click <SELECT TARE> next.
•Wait for the first packaging weight, recorded in tare database, to be
displayed.
- 29 -
Page 30
•Use arrow buttons to select the tare which is to be recalled and press
button,
•The scale returns to the weighing mode, the display indicates entered tare
value with minus „–" sign.
or
• For any operating mode, press button.
• Enter b3 <TARE> option.
• Wait for the first packaging weight, recorded in tare database, to be
displayed.
•Use arrow buttons to select the tare which is to be recalled and press
button.
•The scale returns to the weighing mode, the display indicates entered tare
value with minus „–" sign.
AUTOTARE
Function allowing for automatic tarring of the packaging during the weighing
process when packaging mass for each of the product is different. Read section 9.5
for AUTOTARE function overview.
Deleting tare
Entered tare value can be deleted by pressing
button of the scale’s overlay or
by entering tare value of 0.000g (see description above).
9.4 WEIGHING mode settings
The software allows setup of operating parameters (filters, value release and
autozero option, deleting the last digit and other settings) separately for each
working mode.
It enables customizing the instrument and utilizing its properties depending on the
user’s needs and expectations, or on specific requirements for selected working
mode (e.g. DOSING); as a result the device operation is quick and easy.
Settings:
P2.2.1 – READOUT
P2.2.1.1 – FILTER [V.FAST/ FAST / NORMAL / SLOW / V. SLOW]
P2.2.1.2 – RESULT CONF. [REL. / FAST / FAST+REL.]
P2.2.1.3 – AUTOZERO [YES / NO]
P2.2.1.4 – LAST DIGIT [ALWAYS / NEVER / WHEN STAB.]
P2.2.1.5 – AMBIENT CONDITIONS (STABLE/UNSTABLE)
P2.2.2 – AUTOTARE [YES / NO]
- 30 -
Page 31
P2.2.3 – PRINT MODE
P2.2.3.1 – MODE [EACH / WHEN STAB. / AUTO]
P2.2.3.2 – AUTO. PROG
P2.2.4 – INFORMATION [TARE / NETTO / GROSS / USER / PRODUCT / /NSTD.
INF. / NONE]
P2.2.5 – INF. NSTD. (non-standard information – any text, max 19 characters)
P2.2.6 – HOT KEYS (available options depending on the working mode)
P2.2.6 1 – F1 [NONE / ENTER TARE/ …]
P2.2.6.2 – F2 [NONE / ENTER TARE / …]
P2.2.6.3 – F3 [NONE / ENTER TARE/ …]
P2.2.6.4 – F4 [NONE / ENTER TARE / …]
9.4.1 Filter level setting
Filter shall be set depending on the ambient and environment conditions. For the
best possible conditions the filter can work in a very fast mode (V.FAST value for
parameter P.2.2.1.1 Filter); however, if the conditions are bad (shakes, drafts), the
filter should be set to slow or very slow option (SLOW or V. SLOW value for
parameter P.2.2.1.1 Filter). The effectiveness of the filter is different for the
weighing range. The filter works slower when “approaching” the weighed mass, it
works more intensively for weighed mass within the set range of the filter
(parameter for setting filter range is accessible only from the service menu – the
user does not have an access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer
(SLOW and V. SLOW).
CAUTION!
The higher filter level, the longer weighing time.
- 31 -
Page 32
9.4.2 Value release
Since ambient conditions at a workplace vary, it is necessary to determine the
value release parameter in a most preferable way enabling scale adaptation,
parameter options are: FAST.+REL., FAST or RELIABLE. Depending on selected
option, weighing time is either shorter or longer.
9.4.3 Autozero function
The software features autozero function (Auto) ensuring precise mass indication.
This function automatically controls and corrects zero indication. When Autozero is
enabled, it compares scale indications at declared time interval e.g. 1s, providing
that weighing pan is unloaded and display indication is close to zero. If results vary
less than declared AUTOZERO range e.g. one division, scale zeros automatically,
marker of stable measurement result –
displayed.
If AUTOZERO function is enabled, then each weighing process starts from precise
zero point. There are, however, some cases when this function can be a disturbing
factor for the measuring process; e.g. very slow placing of load on the weighing pan
(load pouring). Here, zero indication correction can also correct actual indication of
loaded mass.
and precise zero marker – are
- 32 -
Page 33
9.4.4 Last digit display
Function enables displaying the last digit on a display.
9.4.5 Ambient conditions
Ambient conditions parameter features two settings: STABLE and UNSTABLE.
Selecting STABLE mode makes the scale work much faster, ie. weighing takes
much less time than for UNSBABLE mode. This parameter relates to ambient and
environmental conditions in which the scale operates. If the ambient conditions are
unstable it is recommended to use UNSTABLE mode. Normally the parameter is
set to STABLE option.
9.5 AUTOTARE
AUTOTARE function is designed to enable quick determination of net weight for
loads with different tare values, being weighed one after another.
When function is active (<P2.2.2 AUTOTARE> value set to <YES>), the operating
process is as follows:
• Make sure that the weighing pan is empty and press zeroing button,
• Put product packaging on a weighing pan,
- 33 -
Page 34
•Wait until the measurement indication is stable,automatic tarring of the
packaging weight proceeds (Net marker appears in the upper part of the
display),
• Put product that is to be packed on the packaging;
• Display shows a net weight of the product;
• Take off the product together with the packaging;
• The scale cancels tare value (packaging weight saved in scale storage
during the first step of the operating process) after the gross weight value
(set in < P 2.2.3.2 AUTO THRES> parameter) has been exceeded;
•Put packaging of a next product on a weighing pan, after measurement
stabilization automatic tarring of the packaging weight proceeds (Net
marker appears in the upper part of the display);
•Put a next product that is to be packed.
For correct operation of the scale with AUTOTARE function it is necessary to adjust
the threshold value.
<P 2.2.3.2 AUTO THRES > parameter is connected with the following functions:
• automatic tare,
• automatic operation,
There is no automatic tarring until the scale operates within the gross mass value
range set in < P 2.2.3.2 AUTO THRES > parameter.
9.6 Printout mode
Printout function is designed to set the printing mode, i.e. button activity.
Printout mode settings:
•<WHEN STAB> the stable measurement result, along with the settings for
parameter <GLP PRINTOUT>, is sent to the printer port. On pressing
button at the moment when the result has not stabilized yet (no
sign on a display), the scale software waits for the stable indication
before sending it to the port.
•<EACH> every single pressing of button results in sending the
measurement result to the printer port along with the settings for
parameter <GLP PRINTOUT>. Every measurement result is sent (stable
and unstable). The <?> sign appears at the beginning of the printing frame
when the result is unstable.
- 34 -
Page 35
•<AUTO> - select this option to enable automatic printing of
measurements. If this option has been selected, remember to select
<AUTO THRES > parameter to suit your needs.
Automatic operation procedure:
•Press button to zero the scale (marker of stable measurement
and zero marker are displayed).
•Load the weighing pan, the scale sends the first stable measurement to
the printer port.
• Unload the weighing pan.
• Next measurement is possible when the indication is lower than the set
value of <AUTO THRES.> parameter (next measurement does not require
zero value).
Procedure for settings modification:
For automatic operation it is necessary to adjust the threshold value.
<P 2.2.3.2 AUTO THRES > parameter is connected with the following functions:
• automatic tare,
• automatic operation,
- 35 -
Page 36
For automatic operation function, the measurement is not send from the computer
to the printer as long as the mass measurement stays within the set value range
<P 2.2.3.2 PROG THRES>.
9.7 Temporary measuring unit
Temporary measuring unit function is designed to enable selection of a measuring
unit which is to be indicated next to mass reading during the operation. It is
available for any mode but <PARTS COUNTING> and <PERCENT WEIGHING>.
The set measuring unit is in use from the moment of its activation until its change or
switching the scale off and on.
Information function is designed to enable additional information display (in
a bottom line). Depending on the needs, a user can choose the following options,
which are shown while working in <WEIGHING> mode:
- 36 -
Page 37
<BARGRAF> option presents amount of used weighing capacity in a graphic form,
it concerns 0 – MAX range.
Display with bargraph option activated for scale with 1000 g Max capacity: 500 g
heavy load placed on a weighing pan means that 50% of max capacity has been
used (half-filled bargraph bottom line).
It is possible to enable <BARGRAF> option for the following modes: PARTS
COUNTING, CHECKWEIGHING, DOSING, PERCENT WEIGHING, WEIGHING,
ANIMAL WEIGHING, STATISTICS, TOTALIZING, PEAK HOLD.
For CHECKWEIGHING mode the bargraph is enriched with threshold markers, for
DOSING mode, target weight marker and target weight tolerance area are added.
Because of LCD display limitations, the markers do not reflect the exact points for
threshold or target weight, they approximately demonstrate settings and let one
perform selection in a rough manner. The markers function is to help user in course
of scale operation.
- 37 -
Page 38
9.9 Non-standard informa tion
Function designed to enable declaring non-standard information, which is to be
displayed in the bottom line of the display. A user can insert any text consisting of
maximum 19 characters.
It should be remembered that if the declared non-standard information is to be
visible, parameter P2.1.2 should be set to < NSTD. INF. > option.
- 38 -
Page 39
9.10 F shortcut keys
Function designed to enable assigning quick access for weighing functions which
are to be accessible by pressing F1, F2, F3 or F4 keys. As a user you can choose
between the following options for <WEIGHING> mode: <NONE / ENTER TARE /
can be freely assigned to any of the F buttons. For other modes there are more
available options (see further sections of this manual).
- 39 -
Page 40
10. UNITS
P8 UNITS parameter group is designed to enable the user to change availability of
mass units in-course of scale operation and to define two user units, thus positively
effecting comfort and speed of operation. It is possible to change unit during
weighing mode operation or during operation of other modes. PARTS COUNTING
and PERCENT WEIGHING are exceptions.
10.1 Units availability
By pressing button, the user may declare units to be available for selection
of temporary unit. Units with parameter value set to <YES> are available for
selection in specified operating modes.
Procedure:
10.2 Start unit selection
Upon selection of start unit, the scale activates with the specified unit for these
modes where change of the unit is possible.
Procedure:
- 40 -
Page 41
10.3 User unit
The user may declare two units. The user unit is a result of indication multiplied by
coefficient given for the user unit.
The units are displayed as [u1] – user unit 1, and [u2] – user unit 2.
Procedure for declaration of a coefficient for the user unit:
- 41 -
Page 42
11. MISCELLANEOUS PARAMETERS
P6
MISC
P6.1
LANGUAGE
|
POLISH/ENGLISH
[menu language]
P6.2
ACCESS LEV.
|
ADMIN
[access levels for
editing menu]
P6.3
KEY SOUND
|
YES/NO
[key sound]
P6.4
BACKLIGHT
|
70
[display backlight
level]
P6.5
BACKLIT OFF
|
NONE/0.5/1/2/3/5
[backlight turn-off
time interval]
P6.6
AUTO OFF
|
NONE/1/2/3/5/10
[display turn-off time
interval]
P6.7
DATE
|
2015.05.27
[date settings]
P6.8
TIME
|
8:53:00 A
[time settings]
P6.9
DATE FORM.
|
YYYY.MM.DD
MM.DD.YYYY
[date format]
P6.10
TIME FORM.
|
12H/24H
[time format]
P6.11
GLP
AUTOTEST
| [carrying out autotest
for the scale]
Miscellaneous parameters menu is designed to provide scale customization
providing more comfortable operation. The parameters are to be found in
parameters group P6 MISC.
Settings modification for particular parameters of this parameter group proceeds
likewise as described in the previous section.
YYYY.DD.MM
DD.MM.YYYY
Enter P6 MISC. menu.
11.1 Available menu languages
Parameter designed to enable selection of language.
Available languages: POLISH, ENGLISH, GERMAN, SPANISH, FRENCH,
TURKISH, CZECH, ITALIAN, HUNGARIAN.
11.2 Accesss levels
Parameter designed to enable selection of permissions level for scale operation.
Available access levels: ADMIN. / USER. / ADV.
Depending on the option selected, a user of a specific permissions level can
access and make changes to the scale settings.
- 42 -
Page 43
11.3 ‘Beep’ sound – reaction on pressing function button
Parameter designed to enable switching on/ switching off the ‘beep’ sound
responsible for informing a user about pressing any key of scale overlay.
NO - ‘beep’ sound switched off
YES - ‘beep’ sound switched on.
11.4 Backlight and adjusting display brightness
Parameter designed to enable setting the brightness of the backlight, or switching
off the display brightness completely.
100 - maximum brightness of the backlight
10 - minimum brightness of the backlight
NONE - display brightness switched off
11.5 Backlit turn-off time
Parameter <P6.5 BACKLIT OFF> is designed to activate display stand-by mode
when weighing process is not being performed (stable indication is a necessary
condition for activation of the stand-by mode).
NONE – backlit turn-off time not activated
0.5; 1; 2; 3; 5 – time given in minutes
If the software registers stable indication for a specified time interval, set in
parameter <P6.5 BACKLIT OFF>, than the display goes out immediately.
The display gets backlit again upon change of indication (no stability pictogram on
the display) or pressing any key on the scale keypad. Blanking works also when
scale menu is entered.
11.6 Auto switch-off
Parameter <P6.6 AUTO OFF> is designed to enable use of function responsible for
automatic display deactivation (functions as button ). Upon display
deactivation, the other subassemblies are powered and the scale turns to stand-by
mode.
NONE– auto switch-off not activated
- 43 -
Page 44
1; 2; 3; 5;10– time given in minutes
If the software registers stable indication for a specified time interval, set in
parameter <P6.6 AUTO OFF> settings, than the display is turned-off immediately
(inactive backlight function, no indication on the display, clock displayed).
To run the scale, it is necessary to press button
keypad. The scale automatically returns to weighing.
Scale cannot be turned off if any process is started or if the menu is entered.
located on the terminal
11.7 Date
Parameter designed to enable setting the current date.
Procedure:
11.8 Time
Parameter designed to enable setting the current time.
Procedure:
11.9 Date format
Parameter designed to enable altering the date format on the printout
[YYYY.MM.DD / YYYY.DD.MM / DD.MM.YYYY / MM.DD.YYYY]
Where:
YYYY – year
- 44 -
Page 45
MM – month
DD – day
11.10 Time format
Parameter designed to enable altering time format on the printout [12H / 24H].
For the value [12H], in parameter P6.6 TIME, letters <A> or <P> are displayed next
to the time value.
Where:
A - hours before noon,
P - hours after noon.
Letters AM or PM are printed out next to the time value.
11.11 GLP Autotest
AUTOTEST function is designed to aid a user in evaluating scale operation and
diagnosing the reasons for occurrence of weighing errors that exceed the maximum
permissible values for a given scale model.
By means of a simple, repeatable and fully documented way, the function enables
optimizing scale settings to maintain the best possible repeatability and weighing
time at workstation. The main purpose of the function is the aforementioned
parameters monitoring, possible to be performed at an optional moment, and
saving records from the carried out tests in the form of printed reports on the tests,
the reports are automatically generated at the end of examination.
GLP Autotest tests repeatability for putting the interval weight on a weighing pan
and repeatability of determining error of indication, wherein the error is referred to
scale maximum capacity.
Testing procedure:
• deposit internal weight twice,
• deposit internal weight ten times,
• perform scale adjustment,
• calculate the value of standard deviation,
• print a report.
- 45 -
Page 46
Test results:
*Deviation for Max.
*Value of repeatability of indication expressed as standard deviation
--------- GLP Autotest: Report ---------
Balance type PS 3000.R2
Balance ID 400010
User Admin
Software rev. v.0.4.9
Date 2013.07.16
Time 09:17:16
-----------------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Average
Value release Fast &Reliable
-----------------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature
…………..........................................
PROCEDURE:
Press
The software starts GLP autotest procedure, the procedure is carried out
automatically, no operator interference is required till the end.
A user can stop the procedure at any time by pressing
process window.
Once the procedure is completed, the value of standard deviation of all
measurements is shown in the main display, and message <RESULT> in the
bottom line, the final report is printed out automatically (see the exemplary report
above). It is possible to print out the report again by pressing
Report is not saved, therefore when abandoning this level, it is deleted from the
scale volatile memory.
Press
Press
button to run parameter <P6.9 GLP AUTOTEST>.
button of the GLP
button.
button to abandon the window, the program returns to the main menu.
button repeatedly to return to weighing.
- 46 -
Page 47
- 47 -
Page 48
12. INFORMATION ON THE DEVICE
INFO menu includes information on the device and its software. The parameters
are strictly informative.
Once the parameter <PRINT. SET> is chosen, the scale settings are sent to the
printer (all parameters). The parameters are transferred via port selected for
<PRINTERS>.
- 48 -
Page 49
13. ADJUS TMENT PRO CEDURE
Adjustment procedure is performed in order to ensure the highest weighing
accuracy, it is required to carry out the adjustment procedure once a while. For
adjustment performance, introduce a corrective factor of indications to scale
memory, remember to refer the said factor to a mass standard.
Adjustment should be carried out:
• before the beginning of weighing procedure,
• if long breaks between successive measuring series occur,
• if temperature inside the scale changes more than: 3°C.
Types of adjustment:
•Automatic internal adjustment
* triggered by scale temperature change
* triggered by elapsing time
•Manual internal adjustment
* triggered by pressing
•Adjustment with external weight
* of declared mass which cannot be modified
* of any weight, but not lower than 30% of maximum range.
13.1 Internal adjustment
Adjustment process can be initiated automatically and manually.
For manual activation press
button. The procedure is carried out
automatically by an automatic adjustment system, no operator interference is
required till the end. Respective information on successive process stages is
displayed.
Automatic adjustment process cycle:
•The software detects the necessity of carrying out an adjustment
procedure, it is signalled by a thermometer pictogram and <Cal> sign at
the top of the display.
- 49 -
Page 50
• Successive weighings can be carried out for about 2 minutes, no longer.
• After 2-minute time interval, message ‘CAL_30’ is displayed, countdown
timer counting seconds from 30..29..28 to 0 is run.
•The user has 30 seconds to make a decision on next steps:
ono action shall be taken in order to enable start of the adjustment
procedure,
oEsc button shall be pressed in order to enable completion of the
weighing procedure. With Esc button pressed, weighing
procedure can be continued, last weighing result is displayed.
Message ‘CAL_30’ reappears on the screen in about 5 minutes.
•The adjustment process can be postponed repeatedly, nevertheless it
shall be remembered that postponing the adjustment for a long time may
lead to greater errors occurring in course of weighing process. The errors
are the effect of temperature changes and as a consequence changes of
scale sensitivity.
Automatic adjustment process takes place for three different cases:
• Adjustment on plugging the scale to mains – for verified devices.
• Adjustment triggered by scale temperature change. The scale is equipped
with a very precise system for temperature monitoring. At each adjustment
process, the temperature is saved in the system. The next adjustment is
automatically initiated if temperature difference between current and last
saved value is greater than 3°C.
•Adjustment triggered by elapsing time. It is possible to declare time
intervals which are criteria for adjustment performance. Settings allow the
adjustment to be carried out every 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12
hour(s).
- 50 -
Page 51
P1
ADJUSTMENT
P1.1
INT.CALIB
| [internal adjustment]
P1.2
EXT.CALIB
| [external adjustment]
P1.3
USER
CALIBRATION
| [user adjustment]
P1.4
CALIBRATION
TEST
| [adjustment test]
P1.5
AUTO.CALIB.
|
NONE/TIME/TEMP/BOTH
[automatic
adjustment]
P1.6
AUTO.CALIB.C
| [time of automatic
adjustment]
13.2 Adjustment menu parameters
INT.CALIB – internal adjustment
Start of an internal adjustment process. Adjustment is carried out automatically, no
user assistance is required. If a weighing pan is loaded, the terminal displays
a command ordering to unload it. If weighing pan is loaded with a relatively small
weight (up to ±10% of max capacity) then adjustment process is carried out
automatically.
or
EXT.CALIB – external adjustment (by means of external adjustment weight)
Adjustment performed with an external weight, value of which is saved in factory
parameters.
USER CALIBRATION - user adjustment (by means of external adjustment weight)
Adjustment performed with an external weight of any mass within scale range,
however not lower than 30% of Max range.
- 51 -
Page 52
CALIBRATION TEST – adjustment process test
Fuction designed to enable comparison of intenal adjustment weight with its value
saved in scale memory. The process is carried out automatically, its result is
displayed on the screen (if the scale is connected via communication port with
a computer or a printer, the adjustement test result is printed out). Press ESC to go
back to the previous menu.
AUTO.CALIB. – automatic adjustment (for scale with internal adjustment)
Specyfying factor that is to trriger the automatic internal adjustment
NONE neither time nor temperature (automatic adjustment turned off),
TEMP. change of scale inside temperature,
TIME time interval set in P1.6 AUTO CALIB. C.,
BOTH time and temperature.
AUTO.CALIB.C – automatic adjustment time interval (for scale with internal
adjustment)
Specyfying time interval, determining how often automatic adjustment process shall
be initiated.
13.3 Manual adjustment
13.3.1 Internal adjustment
Press Cal button
or
- 52 -
Page 53
The scale performs internal adjustment automatically. During adjustment process
do not load the weighing pan with any load. The message <DO NOT TURN OFF CALIBRATION> is displayed in the bottom line. Once adjustment process is
completed, scale saves its result in memory and returns to weighing mode.
CAUTION!
- Press Esc button to abort adjustment process.
- If a weighing pan is loaded, message informing about an error is displayed.
Adjustment process is aborted, scale returns to the weighing mode. Adjustment
process may be repeated upon removal of the excess load.
13.3.2 External adjustment
The external adjustment shall be carried out with an external mass standard /
weigh of class F
.
1
Run an external adjustment process,
a command ordering to unload the weighing pan, <REMOVE MASS> (the weighing
pan must be empty), is displayed. When the weighing pan is unloaded, press
button. The scale determines mass of an empty pan, message <CALIBRATION> is
displayed in the bottom line. Next, message <PUT MASS> and weight value to be
placed on the weighing pan are displayed, e.g.200.000g (depending on the scale
type). Put an external adjustment weight of displayed weight value on a pan and
press button. The scale determines the weight, message <CALIBRATION>
is displayed in the bottom line. On completing adjustment process the scale returns
to submenu P1.2 EXT.CALIB.
13.3.3 User adjustment
The user adjustment shall be carried out with an external mass standard / weigh of
class F
.
1
Run the user adjustment process,
- 53 -
Page 54
First declare the mass of a weight that is to be used for adjustment. The mass must
be ≥ 30% Max capacity.
Once the mass of the weight is entered and confirmed, a message prompting the
user to remove the weight from the pan is displayed: <REMOVE MASS> (the
weighing pan must be empty). Follow the instruction, next press
scale determines weight of an unloaded pan, message <CALIBRATION> is shown
in the bottom line. Next, message <PUT WEIGHT> and mass value to be placed on
the weighing pan are displayed, e.g.200.000g (depending on the scale type). Put
an external adjustment weight of displayed weight value on a pan and press
button. The scale determines the mass, message <CALIBRATION> is displayed in
the bottom line. On completing adjustment process the scale returns to submenu
P1.2 EXT.CALIB.
button, the
13.4 Report on adjustment – printout
At the end of each adjustment process or adjustment test, an adjustment report is
generated automatically and sent to communication port COM 1. The content of the
report is declared in menu P5.1 CAL REPORT.
The instruction on how to declare settings for this option is described in section
about printouts.
The report can be printed out via a printer connected to the device or it can be sent
to the computer and saved as a file for archiving purposes.
- 54 -
Page 55
14. DETERMINING PRINTOUT CONTENT
P5.1
CAL.REPORT
P5.1.1
PROJECT
|
P5.1.2
CALIB. TYPE
|
YES
P5.1.3
USER
|
YES
P5.1.4
PROJECT
|
YES
P5.1.5
DATE
|
YES
P5.1.6
TIME
|
YES
P5.1.7
BALANCE ID
|
YES
P5.1.8
CAL.DIFFER.
|
YES
P5.1.9
DASHES
|
YES
P5.1.10
SIGNATURE
|
YES
Variable
Overview
PROJECT
Option enables naming the project (name associated with a
particular type of weighing). The name may consist of
CALIB TYPE
Option enables printing out the type of the adjustment being
carried out.
USER
Option enables printing out the name of a logged-in user.
PROJECT
Option enables printing out the name of the project (see
parameter Project).
DATE
Option enables printing out the date of the carried out
adjustment.
TIME
Option enables printing out the time of the carried out
adjustment.
BALANCE ID
Option enables printing out the scale ID number.
14.1 Report on adjustment
CAL. REPORT group of parameters is designed to enable declaring data that is to
be printed on an adjustment printout.
maximum 16 characters.
- 55 -
Page 56
CAL. DIFFER
Option enables printing out the difference between mass of
ustment weight measured during the last adjustment
DASHES
Option enables printing out dashes that separate the date of
a printout from a signature.
SIGNATURE
Option enables providing an area for the signature of a user
performing the adjustment.
P5.2
HEADER
P5.2.1
DASHES
|
YES
P5.2.2
WORKING MODES
|
YES
P5.2.3
DATE
|
YES
P5.2.4
TIME
|
YES
P5.2.5
SCALE TYPE
|
YES
P5.2.6
BALANCE ID
|
YES
P5.2.7
USER
|
YES
P5.2.8
PRODUCT
|
YES
P5.2.9
VARIABLE 1
|
YES
P5.2.10
VARIABLE 2
|
YES
For the parameters described above, one of these values must be selected:
NO - do not print
YES - print
Report example:
an adj
and the current measured mass of this weight.
14.2 Header printout
HEADER group of parameters is designed to enable declaring data that is to be
printed on header printout.
- 56 -
Page 57
P5.2.11
EMPTY LINE
|
YES
P5.2.12
CAL.REPORT
|
YES
P5.2.13
NON-STANDART
PRNT.
|
NONE
NSTD. PRN. 2
NSTD. PRN.4
Variable
Overview
DASHES
Option enables printing out separating dashes.
WORKING MODE
Option enables printing out the name of a working
mode.
DATE
Option enables printing out the date of the carried out
adjustment.
TIME
Option enables printing out the time of the carried out
adjustment.
SCALE TYPE
Option enables printing out the scale type.
BALANCE ID
Option enables printing out the scale ID number.
USER
Option enables printing out the name of a logged-in
user.
PRODUCT
Option enables printing out the name of a currently
selected product.
VARIABLE 1
Option enables printing out the value of VARIABLE 1.
VARIABLE 2
Option enables printing out the value of VARIABLE 2.
EMPTY LINE
Option enables printing out an empty separating line.
CAL.REPORT
Option enables printing out a report from the last
adjustment, according to the settings declared for the
adjustment report printout (see sec. 14.1 of this user
manual).
NSTD. PRNT.
Option enables printing out one of 100 non-standard
standard printout
described further down this user manual.
NSTD.PRN. 1
NSTD.PRN.3
printouts on the footer printout. You can choose one of
the following options: NONE / nonname. The way of entering non-standard printouts is
For the parameters described above, one of these values must be selected:
Option enables printing out the date of the carried
/ NSTD. PRN.4
For instruction on entering non-standard printouts read later sections of this user
manual.
Printout example
14.3 GLP PRINTOUT – measurement result section
GLP PRINTOUT group of parameters is designed to enable declaring data that is
to be printed on the measurement result printout.
P5.3 W YDRUK GLP, to grupa parametrów umożliwiająca zadeklarowanie danych,
które znajdą się na wydruku pomiaru.
STANDART
PRNT.
NSTD.PRN. 1
NSTD.PRN.3
- 58 -
Page 59
out adjustment.
TIME
Option enables printing out the time of the carried
out adjustment.
USER
Option enables printing out the name of a logged-in
user.
PRODUCT
Option enables printing out the name of a currently
selected product.
VARIABLE 1
Option enables printing out the value of VARIABLE
1.
VARIABLE 2
Option enables printing out the value of VARIABLE
2.
NET
Option enables printing out net weight value in a
basic unit (calibration unit).
TARE
Option enables printing out the tare value in the
current unit.
GROSS
Option enables printing out the gross mass value in
the current unit.
CURR.RES
Option enables printing out the current
measurement result (NET weight) in a current unit.
CAL.REPORT
Option enables printing out a report from the last
adjustment, according to the settings declared for
the adjustment report printout (see sec. 14.1 of this
user manual).
NSTD. PRNT.
Option enables printing out one of 100 non-standard
printouts on the footer printout. You can choose one
printouts is described further down this user manual.
P5.4
FOOTER
P5.4.1
WORKING MODES
|
YES
of the following options: NONE / non-standard
printout name. The way of entering non-standard
For instruction on entering non-standard printouts read later sections of this user
manual.
Printout example:
14.4 Footer printout
FOOTER group of parameters is designed to enable declaring data that is to be
printed on footer printout.
- 59 -
Page 60
P5.4.2
DATE
|
YES
P5.4.3
TIME
|
YES
P5.4.4
SCALE TYPE
|
YES
P5.4.5
BALANCE ID
|
YES
P5.4.6
USER
|
YES
P5.4.7
PRODUCT
|
YES
P5.4.8
VARIABLE 1
|
YES
P5.4.9
VARIABLE 2
|
YES
P5.4.10
DASHES
|
YES
P5.4.11
EMPTY LINE
|
YES
P5.4.12
CAL.REPORT
|
YES
P5.4.13
SIGNATURE
|
YES
P5.4.14
NON-STANDART
PRNT.
|
NONE
NSTD. PRN. 2
NSTD. PRN.4
Variable
Overview
WORKING MODE
Option enables printing out the name of a working
mode.
DATE
Option enables printing out the date of the carried out
adjustment.
TIME
Option enables printing out the time of the carried out
adjustment.
SCALE TYPE
Option enables printing out the scale type.
BALANCE ID
Option enables printing out the scale ID number.
USER
Option enables printing out the name of a logged-in
user.
PRODUCT
Option enables printing out the name of a currently
selected product.
CUSTOMER
Option enables printing out the name of a currently
selected Customer.
VARIABLE 1
Option enables printing out the value of VARIABLE 1.
VARIABLE 2
Option enables printing out the value of VARIABLE 2.
DASHES
Option enables printing out separating dashes.
EMPTY LINE
Option enables printing out an empty separating line.
CAL.REPORT
Option enables printing out a report from the last
adjustment, according to the settings declared for the
adjustment report printout (see sec. 14.1 of this user
NSTD.PRN. 1
NSTD.PRN.3
- 60 -
Page 61
manual).
SIGNATURE
Option enables providing an area for the signature of a
user performing the adjustment.
NSTD. PRNT.
Option enables printing out one of 100 non-standard
printouts on the footer printout. You can choose one of
standard printout
described further down this user manual.
the following options: NONE / nonname. The way of entering non-standard printouts is
For the parameters described above, one of these values must be selected:
For instruction on entering non-standard printouts read later sections of this user
manual.
Printout example:
14.5 Non-standard printouts
The software enables entering 4 non-standard printouts. Each of them can consist
of approximately 160 characters.
Non-standard printout may include:
•Variables depending on the working mode and other needs (mass, date
etc.)
•Permanent text form the user menu, however only CAPITAL letters shall
be used, without Polish characters
•Non-standard printout can have approximately 160 characters (typed as a
single line)
- 61 -
Page 62
14.5.1 Inserting texts
%V
Current net mass in the current unit
Variables with the same values for all operating modes.
%% Printout of a “%” character
%N Current net mass in the basic unit
%G Current gross mass in the basic unit
%T Current tare mass in the basic unit
%D Current date
%M Current time
%I Scale number
%R Program number
%P Project number
%U User number
%F Name of a current function – working mode
%C Date and time of the last adjustment
%K Type of the last adjustment
%S Currently weighed product
%Y Deviation for the last adjustment
%1 Variable 1
%2 Variable 2
Variables with values depending on the currently used working mode.
Variable Description Mode for which the variable is active
%W Standard mass 1 pcs PARTS COUNTING
%H High threshold
%L Low threshold
%A Target mass DOSING
%B Reference mass PERCENT WEIGHING
Non-standard characters used in designing non-standard printouts.
\\ a single „\” character
\C CRLF
\R CR
CHECKWEIGHING
- 62 -
Page 63
\N LF
\T
Tabulator
\F Form feed (for PCL printers)
%E Crop the paper for EPSON printers
Every single printout can contain max 160 characters (letters, numerals, nonstandard characters, spaces). A user can apply non-standard characters depending
on type of data that is to be printed out.
Enter printout content settings and design the printout using respective data
variables and characters for text format.
- 63 -
Page 64
Example 2:
To crop the paper after the printout has been carried out by EPSON printer (if the
printer is equipped witth an autocutter blade) the user must select (for a given
printout: HEADER, GLP PRINTOUT or FOOTER) an option of non-standard
printout 1, 2, 3 or 4 with <%E> value available, then select this printout for a given
printout settings.
In such case <SUFFIX> command should stay empty.
Paper must be cropped underneath the FOOTER.
Example settings:
• P5.4.14 STANDARD PRINTOUT | NSD. PRN. 1
• P5.5 NSD. PRN. 1 | %E
Inserting text
•by means of the scale
Press to select a character to be replaced, or to move a
cursor or an active (blinking) character to the right.
Press to select a character to be replaced, or to move a
cursor or an active (blinking) character to the left.
Press to change the character by one value down
Press to change the character by one value up
Press to delete a character
Press to insert a character
•by means of USB-type computer keyboard
Connect a computer keyboard of USB type to the scale in order to provide
easier and quicker editing of the printouts.
To insert any text, select a respective menu option and, using the
keyboard, type the text. Next press Enter button for confirmation.
CAUTION!
It is important to type variables, used for non-standard printouts, in capital
letters.
- 64 -
Page 65
14.6 Variables
Variable is defined as alphanumeric data which can be linked to printouts, products
or other information related to weighing. Every variable is characterized by its
content, the content must be given. Variables are used for entering various data
during the weighing process, e.g. serial number or batch number. The program
allows to enter two variables. Each can consist of max 32 characters.
In order to input a variable content, the user needs to enter variable settings
(parameter P5.9 – VARIABLE 1 or P5.10 – VARIABLE 2) and enter the respective
values using direction keys (arrows) on the scale keypad or a computer keyboard.
Procedure for entering texts is the same as for non-standard printouts.
- 65 -
Page 66
15. DATABASES
Scale software features three editable databases (USERS, PRODUCTS, TARES)
and two non-editable databases storing measurements (WEIGHINGS, ALIBI).
Data saved within particular databases:
USERS – 100 different users.
PRODUCTS – 1000 different products.
TARES – 100 different masses of the packaging.
WEIGHINGS – 10 000 consecutive measurements
ALIBI – 100 000 consecutive measurements
Operations on USERS, PRODUCTS and TARES databases:
Deleting the selected record
Inserting a new record
For detailed information on inserting and deleting records read later sections of this
user manual.
15.1 Users
USERS – 100 different users.
Each user record is characterized by the following data:
NAME (30 characters), CODE (6 characters),
PASSWORD (8 characters, digits only),
ACCESS (USER, ADVANCED, ADMIN),
LANGUAGE (any of the available).
The software comprises three permissions levels: administrator, advanced
operator, user.
On switching on, the display is constantly active enabling carrying out measuring
processes regardless of the fact that no user is logged in.
- 66 -
Page 67
Permissions
level
Operator
Free editing of parameters of <Readout> submenu.
The operator can preview information recorded in
Advanced
Free editing of parameters of the following submenus:
<Readout>; <Working modes>; <Communication>;
Administrator
Access to all user parameters and functions, editing
Permissions level dependent access to edition of user parameters, databases and
software functions.
Available operations
Modification of settings for <Misc.> parameter group, except
for settings for <Date and Time>.
The operator can start and carry out all weighing processes.
<Databases>, (s)he can define universal variables.
operator
<Peripherals>; <Misc.>.
Access to <Date and Time> submenu denied.
The advanced operator can start and carry out all weighing
processes.
databases enabled.
Adding a new user – procedure:
- 67 -
Page 68
With new user name entered, provide remaining data.
User code – Max 6 characters
User password – Max 8 numbers
Access level
Language
Once all of the data is entered, return to the weighing mode.
In order to select a user, follow the logging-in procedure described in section 8.
- 68 -
Page 69
Removing a user – procedure:
• Enter the user database – read the description above.
• Select the user that is to be removed from the list.
• Press
• The software shows <DELETE?> message in the bottom line.
• Confirm by pressing button.
• The software removes the selected user from the list.
• Return to the weighing mode
button.
15.2 Products
PRODUCTS – 1000 different products.
Each product record is characterized by the following data:
NAME (30 characters),
CODE (6 characters),
EAN (16 characters),
MASS (accuracy of a reading unit),
TARE (mass of the packaging relating to a particular product, accuracy of a reading
unit),
MIN (low limit for the <CHECKWEIGHING> mode, to be inserted with the accuracy
of a reading unit),
MAX (high limit for the mode <CHECKWEIGHING>, to be inserted with the
accuracy of a reading unit),
TOLERANCE (tolerance limits to [±] for <DOSING> mode, entered as a % of the
target mass).
In order to add a product, enter products database, adding new user and adding
new product procedures are likewise.
Provide product details.
- 69 -
Page 70
15.3 Tares
TARES – 100 different weights of the packaging.
Each tare record is characterized by the following data:
- 70 -
Page 71
NAME (30 characters),
TARE (packaging weight, accuracy of a reading unit).
In order to add tare, enter tares database, adding new user and adding new tare
procedures are likewise.
Provide packaging details.
15.4 Weighings
Weighings database is non-editable, i.e. the weighing-related data is recorded
automatically. A user has a possibility of viewing the recorded data and printing it
out or exporting it to a PENDRIVE (read later sections for more information on
export).
The software allows you to record and store up to 10 000 measurements carried
out on the scale. This occurs automatically, after pressing the <PRINT> button, no
additional actions or settings change is needed.
Additional data is saved along with the measurement.
• Date of the measurement
• Time of the measurement
• Measurement result (mass)
• Tare value
• Name of the product that has been weighed
• Person carrying out the measurement (logged-in user)
• Working mode for which the measurement has been carried out
- 71 -
Page 72
•Value of variable 1 and 2
The software records the measurements in a so called loop, i.e. when the
measurement 10 001 is recorded, the measurement 1 is automatically deleted from
the scale memory.
The recorded measurements cannot be deleted.
As a user you can view and print out the recorded data.
Procedure:
Each measurement is recorded with its individual number. The format is: b4.4.n,
where <n> is the consecutive number of the recorded measurement. In the bottom
line, the date and time are displayed for every single measurement.
To swap between measurements recorded in database, use the arrow buttons,
or . Pressing one of these buttons lets you to move to the next
measurement either up or down the list.
To view the remaining data relating to a respective measurement, first select the
measurement in question, next press button:
The software automatically shows measurement-related data, the data is presented
in the bottom line of the device display.
- 72 -
Page 73
To swap between measurement-related data, press
relating to the measurement can be printed out by selecting the option <PRINT>
or . The data
and pressing
Record example:
Date 21.06.2013
Time 13:05:02
User
Product
Tare 0.000 g
Gross 0.000 g
--------------Cal. report.--------------
Cal. Type Internal
User
Project 1234567890123459
Date 16.07.2013
Time 13:27:09
Balance ID 10353870
Cal. diff. -0.004 g
Data to be printed out depends on the settings of parameter P5.3 GLP PRINTOUT.
Parameters selected for the printout (<YES> option), are printed out also on
measurement result printout for WEIGHINGS database (read sec. 14.3.)
15.5 ALIBI memory
The device is equipped with “ALIBI” memory allowing you to record and store up to
100 000 measurements carried out on the scale.
With “ALIBI” memory installed, the measurements record occurs automatically, it is
run by pressing <PRINT> button, you do not need to take any additional actions or
change settings.
Additional data is saved along with the measurement.
• Date of the measurement
• Time of the measurement
• Measurement result (mass)
• Tare value
• Person carrying out the measurement (logged-in user)
• The name of the product that has been weighed
The software records the measurements in a so called loop, i.e. when the
measurement 10 001 is recorded, the measurement 1 is automatically deleted from
the scale memory.
The recorded measurements cannot be deleted.
As a user you can view and print out the recorded data.
Procedure:
Each measurement is recorded with its individual number. The format is: b4.5.n,
where <n> is the consecutive number of the recorded measurement. In the bottom
line, the date and time are displayed for every single measurement.
- 74 -
Page 75
To swap between measurements recorded in database, use the arrow buttons,
or . Pressing one of these buttons lets you to move to the next
measurement either up or down the list.
To view the remaining data relating to a respective measurement, first select the
measurement in question, next press button:
The software automatically shows measurement-related data, the data is presented
in the bottom line of the device display.
To swap between measurement-related data, press or . The data
relating to the measurement can be printed out by selecting the option <PRINT>
and pressing
button.
- 75 -
Page 76
ALIBI record example:
Date 19.06.2013
Time 6:48:41
Result 199.90 g
Tare 0.000 g
User SMITH
Product PILL
To return to the weighing mode press
button repeatedly.
- 76 -
Page 77
16. DATABASE EXPORT AND IMPORT
Option designed to allow the user to:
•Archive data relating to the carried out weighments – WEIGHMENT
database and ALIBI database.
•Copy the products databases and tares databases between devices of this
series.
This can be performed by means of an external flash memory, USB, with <FAT file system>.
Plug the memory stick into USB 1 – TYPE A port.
The device automatically detects the memory stick, the message enabling
operations relating to export or import of the database is displayed.
Parameter options:
Database EXPORT
Database IMPORT
- 77 -
Page 78
16.1 Database export
To export the database, select EXPORT option.
The following functions are available:
• Export of all the databases
• Export of users database
• Export of product database
• Export of tares database
• Export of weighings
• Export of weighings recorded in ALIBI memory
• Export of user parameters
After selecting <ALL DATABASES> option, the scale software creates files on the
memory stick. The files are of relevant names, data from individual databases is
recorded in them. The files are characterized by special extensions, saved data is
encoded in a way that the files cannot be read or seen by standard computer
programs.
Special computer software manufactured by RADW AG company is used to read
data from database files: ALIBI and WEIGHINGS. The data from files storing
information from PRODUCTS, USERS, and TARES databases is read
automatically by the software when option <IMPORT> is enabled.
- 78 -
Page 79
16.2 Database import
<IMPORT> function is designed to allow transfer of data, recorded in scale
databases, form one scale to another. This is a quick and reliable way for inserting
the data without any mistakes.
To import the database, plug a memory stick to USB port, next select IMPORT
option and choose one of the following:
The following options are available:
• Import of all databases
• Import of users databases
• Import of products databases
• Import of tares databases
• Import of user parameters
The data from ALIBI and WEIGHINGS databases cannot be imported.
16.3 Measurement data printout
The software allows the user to save data, relating to a measurement, on an
external flash memory, USB.
Procedure:
• Plug a memory stick into USB port.
• Press button to abandon <IMPORT/EXPORT> option, which is
automatically started.
•Enter parameter P4.2.1 <DEVICES/PRINTER/PORT> and set
<PENDRIVE> option.
• Return to the weighing mode
• Now every single pressing of button results with measurement data
record (data set for GLP0 PRINTOUT is recorded) in a text file which is
automatically created by the software. The file name is: printout.txt.
•Switch the scale off by pressing button, do it in order the data was
saved in the file. Now you can remove the pen drive from the port and
read the data on a computer.
- 79 -
Page 80
The data can be printed out on any printer connected to the computer.
More data can be recorded in the same file. The software adds any new data to the
file that has already been created on a memory stick. Consequently, a user can
continue saving measurements in the same file once created.
CAUTION!
USB memory stick shall feature <FAT file system>.
When software detects the memory stick, menu <P9 IMPORT/EXPORT> becomes
accessible.
All the above described operations relating to export and import Can be performed
using this menu.
- 80 -
Page 81
17. OPERATION MODES
• Weighings
• Parts Counting
• Checkweighing
• Dosing
• Percent Weighing
• Animal Weighing
• Density Determination of solids
• Density Determination of liquids
• Statistics
• Totalising
• Peak Hold
• Pipettes Calibration
In order to run a particular mode press
button, next select it from the list.
- 81 -
Page 82
Once the button
shown.
or - press to select the working mode,
- press to enter the selected working mode.
For overview of operation modes setup, read later sections of this manual.
has been pressed, the name of the first available function is
17.1 Operation modes accessibility
This group of parameters is designed to let you declare which modes are to be
available after pressing
You can switch off the modes will not be used while working with the device. To do
it set the availability parameter to <NO> option.
Procedure:
button.
- 82 -
Page 83
17.2 Parts Counting – equal weight pieces
Parts Counting mode allows you to count small objects of the same weight.
• Run Parts Counting function
•
When Parts Counting mode is run for the first time, the reference sample weight
equals 0.0000g.
Parts Counting mode when restarted operates with the most recently used
reference sample weight value, either determined or entered.
17.2.1 Parts Counting mode setup
The software allows individual setup for each operation mode. Some parameters
are identical for all operation modes. To overview identical parameters description,
read section relating to <WEIGHING> mode.
This section overviews parameters that are individual for <PARTSCOUNTING>
mode.
- 83 -
Page 84
F shortcut keys
NONE
no function assigned to the button
ENTER SAMPLE
function initiates an option of entering the reference
weight of a single piece
DETERMINE SAMPLE
function initiates an option of automatic determination of
of known quantity
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <PARTS COUNTING> mode:
weight for a single piece taken from a number of samples
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.2.2 Reference Sample Weight determination with the use of sample of
known quantity
While determining the weight of a single piece, ACAI function (Automatic Accuracy
Correction) is in use.
Means of operation of ACAI function:
•Pieces quantity (on adding) on the weighing pan has to be greater than
before.
•Pieces quantity (on adding) on scale weighing pan must be less than twice
the amount displayed before adding parts.
•Current quantity of pieces must be within the ± 0,3 tolerance of the total
value.
•Measurement result has to be stable.
Procedure:
•Put the container on the pan and tare its weight,
- 84 -
Page 85
•Press F button to which the function <ENTER SAMPLE> is assigned, wait
to see the editing window <SAMPLE QUANTITY>
•Use navigating buttons or to select the correct sample
quantity.
•For option: sample quantity (displayed value <0000>) enter demanded,
use navigating buttons.
•Confirm sample quantity, message <PLACExx PCS> is displayed
- 85 -
Page 86
•Put declared number of pieces in the container, wait until the result is
stable (the symbol is displayed), confirm the weight by pressing
button,
•The software automatically counts a single piece weight and enters
<PARTSCOUNTING> mode, number of pieces on the pan (pcs) is
displayed. In the bottom line, a single piece weight value is displayed (if
the option has been selected for the function <INFORMATION>).
CAUTION!
Remember that:
•The total weight of all pieces put on the weighing pan must not be greater
than the device maximum weighing range.
•The weight of a single piece cannot be lower than 0,1 reading unit. If this
condition is not met you shall see message: <Sample mass too low>.
•While determining pieces quantity, wait for the stability pictogram
next confirm the quantity.
•You can confirm the declared quantity by pressing
button only after
the stability pictogram is displayed. Otherwise, the scale will not accept
the measurement.
17.2.3 Entering Reference Sample Weight
Procedure:
•Press F button to which the function <ENTER SAMPLE> is assigned, wait
to see the editing window <ENTER SAMPLE>
,
•Use navigating buttons to enter known single piece weight value.
- 86 -
Page 87
• Confirm the entered weight value by pressing button,
• The software automatically counts a single piece weight and enters
<PARTSCOUNTING> mode, number of pieces on the pan (pcs) is
displayed. In the bottom line, a single piece weight value is displayed (if
the option has been selected for the function <INFORMATION>).
17.2.4 Return to weighing mode
17.3 Checkweighing
Checkweighing is a working mode that uses two thresholds (LOW and HIGH) in
order to check mass of the samples. It is generally assumed that the mass is
correct if it is contained within the threshold values.
- 87 -
Page 88
F shortcut keys
NONE
no function assigned to the button
ASSIGN
THRESHOLDS
function initiates an option of ascribing checkweighing
thresholds
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <CHECKWEIGHING> mode:
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.3.1 Declaring threshold values
Procedure:
•Press F button to which the function <ASSIGN THRESHOLDS> is
assigned, the editing window is displayed and the value of MIN low
threshold can be ascribed. The ascribed value must be given in a current
unit.
•Use navigating buttons to enter low threshold value
- 88 -
Page 89
• Confirm the entered threshold value by pressing button,
• The software automatically proceeds to the editing window of MAX high
threshold value, which is to be given in the current unit. Use the navigating
arrows to set a particular value.
• Confirm the entered threshold value by pressing button,
• The software proceeds to the <CHECKWEIGHING> mode, values of
declared thresholds are displayed in the bottom line (if such an option has
been selected for the function <INFORMATION>).
•At the top of the display the message <Min> is shown, it indicates weight
value of load placed on the weighing pan in relation to low limit weight
- 89 -
Page 90
value. Bottom line of the display, if <BARGRAPH> option is enabled, shall
<Min>: mass lower than the value of the low threshold
<Ok>: mass contained within thresholds
<Max>: mass higher than the value of the high threshold
demonstrate weight indication, in a graphic form, in relation to set
thresholds.
17.3.2 Return to weighing mode
17.4 Dosing
Dosing mode comprises process of sample weighing, wherein the said sample
weighing operation is performed until target mass is reached. The target mass is
defined along with dosing tolerance value. The tolerance value is set as
a percentage of the target mass.
Example:
Target weight = 100.000g
Tolerance = 2,5% (2,5% of 100g, which amounts to 2,5g)
- 90 -
Page 91
i.e.: the software accepts correctly dosed value within the following thresholds: from
NONE
no function assigned to the button
ENTER SAMPLE
function initiates an option of entering the reference
weight along with tolerance (one procedure)
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
97,500g to 102,500g.
F shortcut keys
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <DOSING> mode:
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.4.1 Entering target weight
Procedure:
•Press F button to which function <ENTER SAMPLE> is assigned, the
editing window is displayed. The target weight value must be given in a
current unit.
- 91 -
Page 92
• Use navigating buttons to enter the target weight
• Confirm the entered target weight by pressing button,
• The software automatically sets tolerance of target mass dosing.
• Use navigatingbuttons to set the tolerance value
• Confirm the entered tolerance value by pressing button,
• The software automatically enters the <DOSING> mode, target weight
value is displayed within the main section with a minus sign, reference
sample weight value (target weight) is displayed in the bottom line (if such
option has been selected for the function <INFORMATION>).
•At the top of the display the message <Min> is shown, it indicates weight
value of load placed on the weighing pan in relation to target weight value.
- 92 -
Page 93
Bottom line of the display, if <BARGRAPH> option is enabled, shall
<Min>: mass lower than the Target Value - Tolerance
<Ok>: mass contained within tolerance Target Value +/- Tolerance
<Max>: mass greater than the Target Value + Tolerance
demonstrate weight indication, in a graphic form, in relation to target
weight and set tolerance value.
17.4.2 Return to weighing mode
17.5 Percent Weighing
The software has an option of control of deviation (in percentage) between
measured load weight and specified reference sample weight. The sample weight
can be determined by means of a weighing process or it can be entered into
memory by the user.
F shortcut keys
- 93 -
Page 94
As a user you can assign weighing mode functions for quick access. The functions
NONE
no function assigned to the button
ENTER SAMPLE
function initiates an option of entering reference weight
for control of weight deviation
DETERMINE SAMPLE
function initiates an option of determining reference
weigh for control of weight deviation
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <PERCENT WEIGHING> mode:
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.5.1 Reference Weight determination by weighing
Procedure:
•Press F button to which function <DETERMINE SAMPLE> is assigned,
<PUT 100%> editing window is displayed.
- 94 -
Page 95
•Put reference weight (as 100%) on the pan, wait for the indication to be
stable (
symbol is displayed) and confirm the weight by pressing
button.
•The software automatically enters the measured load value as a reference
weight and enters <PERCENT WEIGHING> mode, 100.000% value is
displayed within the main section, reference sample weight value (target
weight) is displayed in the bottom line (if such option has been selected for
the function <INFORMATION>)
17.5.2 Entering reference weight
Procedure:
•Press F button to which function <ENTER SAMPLE> is assigned,
<ENTER SAMPLE> editing window is displayed.
•Using navigating buttons to enter the known reference weight.
- 95 -
Page 96
• Confirm the entered weight by pressing button.
• The software automatically enters <PERCENT WEIGHING> mode,
0.000% value is displayed within the main section, reference sample
weight value (target weight) is displayed in the bottom line (if such option
has been selected for the function <INFORMATION>)
17.5.3 Return to weighing mode
17.6 Animal Weighing
Animal Weighing mode enables to reliably determine weight of objects in motion. In
principle, these types of objects generate unstable measurement, thus it requires
using a different filtering method of measurement signal.
- 96 -
Page 97
NONE
no function assigned to the button
START
function initiates weighing of the object placed on the
manual mode (AUTOSTART parameter
set to NO value)
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
17.6.1 Additional mode settings
Animal Weighing mode features standard weighing mode parameters along with
supplementary parameters designed to improve this mode operation.
Supplementary parameters:
•AVERAGING TIME – time period for measurement records analysis,
obtained data is used to determine average measurement result.
•THRESHOLD - value expressed in mass measuring units, to start
measurement, the indication value has to exceed value set in the
threshold.
•AUTOSTART - determines start criterion for measurements: whether they
are initiated manually (on pressing a button or selecting START), or
automatically.
When the parameter is set to <YES> option, the measurement starts
automatically if indication on display exceeds the value of set threshold.
The next measurement can start on removing the weighed object from a
pan (the indication has to return below the value set in threshold), loading
a new object, and exceeding the threshold’s value by the indication.
Remember to adjust settings of the above parameters prior to tests. The set values
shall be respective to needs and expectations resulting from working environment.
F shortcut keys
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <ANIMAL WEIGHING> mode:
weighing pan -
- 97 -
Page 98
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.6.2 Running the process manually – means of operation
To manually start the process, go to the mode setup, select an <AUTOSTART>
parameter, next select <NO> value.
Procedure:
•With modified settings, return to the main window, to do it press
button repeatedly.
•Now, set the averaging time in seconds (time period during which the
software collects measurements and determines average result on their
basis).
• Do not set <THRESHOLD> parameter as it is not required for this mode.
• Select <ANIMAL WEIGHING> mode.
- 98 -
Page 99
•Put a container for the load on the weighing pan, wait for a stable
indication and carry out tarring.
•Next, enter the mode options and start the measurement procedurę, follow
the below procedure.
- 99 -
Page 100
•With completed proces the measurement result is locked and
automatically printed.
• To finish the measurement press button.
• The program automatically returns to the main window of Animal Weighing
mode.
17.6.3 Running the process automatically – means of operation
To automatically start the process, go to the mode setup, select an <AUTOSTART>
parameter, next select <YES> value. Follow procedure described in the previous
subsection. Additionally set <AVERAGING TIME> and <THRESHOLD>
parameters.
Select <ENTER TARE> parameter, do it to carry out the measurement in a
container (TARE) for this mode, for <ENTER TARE> parameter overview read
‘Weighing’ section.
Procedure:
•Enter the weight of the container, next put the container on the pan and
place the object that is to be weighed in the container, the process of
measurement begins automatically after exceeding the set threshold
weight.
•On weighing process completion, the result is locked and automatically
printed.
• Press button to finish the measurement.
• The program automatically returns to the main window of Animal Weighing
mode.
- 100 -
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.