RADWAG BALANCES 3Y User Manual

Page 1
User manual no.: IMMU-02-07-09-14-ENG
BALANCES 3Y SERIES
ultra-microbalances UYA 3Y
analytical balances XA 3Y
− analytical balances XA 3Y.A
analytical balances AS 3Y
precision balances APP 3Y
BALANCES AND SCALES
RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com www.radwag.com
Page 2
- 2 -
SEPTEMBER 2014
Page 3
- 3 -
Table of contents
1. BASIC INFORMATION ....................................................................... 9
2. UNPACKING AND INSTALLATION .................................................... 11
2.1. Ultra-microbalaces UYA 3Y .................................................................... 11
2.2. Balances XA 3Y; XA 3Y.A ....................................................................... 12
2.3. Balances AS 3Y series ........................................................................... 13
2.4. Balances PS 3Y series ........................................................................... 14
2.5. Balances APP 3Y series ......................................................................... 14
2.6. Balance cleaning .................................................................................. 15
2.7. Plugging to mains ................................................................................. 21
2.8. Connecting peripheral equipment ........................................................... 21
3. BALANCE DESIGN ........................................................................... 22
3.1. Dimensions ......................................................................................... 22
3.2. Interfaces ............................................................................................ 26
3.3. Balances with a wireless connection between the terminal and the weighing
module (an option) .............................................................. 27
Start-up procedure ............................................................................... 28 3.3.1.
Icons description .................................................................................. 28 3.3.2.
Settings .............................................................................................. 29 3.3.3.
4. STARTUP ......................................................................................... 30
5. BALANCE KEYBOARD – FUNCTION KEYS ......................................... 30
6. SOFTWARE STRUCTURE .................................................................. 31
7. SOFTWARE WEIGHING WINDOW .................................................... 32
8. LOGGING ........................................................................................ 33
9. OPERATING USER MENU ................................................................. 35
9.1. Terminal’s keyboard ............................................................................. 35
9.2. Return to weighing mode ...................................................................... 36
10. WEIGHING ...................................................................................... 37
10.1. Selection of measuring unit ................................................................... 37
10.2. Means of correct weighing procedure ...................................................... 38
10
.3. Balance leveling ................................................................................... 39
10.4. Zeroing ............................................................................................... 39
10.5. Tarring ................................................................................................ 40
10.6. Weighing on dual range balances (ref. balance PS 200/2000.3Y series) ...... 41
10.7. Use of air buoyancy compensation factor ................................................. 41
10.8. Additional parameters on weighing process ............................................. 45
10.9. Minimum sample .................................................................................. 47
10.10. Cooperation with TITRATORS ............................................................. 51
11. ADJUSTMENT .................................................................................. 52
11.1. Internal adjustment .............................................................................. 52
11.2. External adjustment ............................................................................. 52
11.3. User adjustment .................................................................................. 53
11.4. Adjustment test ................................................................................... 53
11.5. Automatic adjustment ........................................................................... 53
11.6. Automatic adjustment time ................................................................... 54
Page 4
- 4 -
11.7. Adjustment report printout .................................................................... 54
11.8. GLP project ......................................................................................... 54
11.9. Adjustment history ............................................................................... 55
12. USERS ............................................................................................. 56
13. ACCESS LEVEL ................................................................................. 57
14. PROFILES ........................................................................................ 59
14.1. Creating a profile ................................................................................. 59
14.2. Profile structure ................................................................................... 60
Settings .............................................................................................. 61 14.2.1.
Working modes .................................................................................... 61 14.2.2.
Readout .............................................................................................. 61 14.2.3.
Measuring units .................................................................................... 63 14.2.4.
15. WORKING MODES – general information ........................................ 64
15.1. Working mode selection ........................................................................ 65
15.2. Parameters related to working modes ..................................................... 66
15.3. Quick access keys ................................................................................ 66
Proximity sensors ................................................................................. 67 15.3.1.
Automatically opened door .................................................................... 67 15.3.2.
15.4. Information ......................................................................................... 68
15.5. Printouts ............................................................................................. 68
15.6. Profiles ............................................................................................... 70
16. PARTS COUNTING ........................................................................... 71
16.1. Additional settings of parts counting mode .............................................. 72
16.2. Parts counting – quick access keys ......................................................... 73
16.3. Setting standard (reference) mass by inserting determined part mass ........ 73
16.4. Setting standard (reference) mass by weighing ........................................ 73
16.5. Acquiring part mass from database......................................................... 74
16.6. Updating part mass in the database ........................................................ 74
16.7. Parts counting procedure ....................................................................... 75
.8. Checkweighing function in parts counting mode ....................................... 75
16
16.9. Dosing function in parts counting mode ................................................... 76
17. CHECKWEIGHING............................................................................ 78
17.1. Use of checkweighing thresholds ............................................................ 79
17.2. Additional settings of checkweighing mode .............................................. 79
18. DOSING .......................................................................................... 80
18.1. Use of database of products in dosing process ......................................... 81
18.2. Additional settings of dosing mode ......................................................... 82
19. PERCENT SETUP .............................................................................. 83
19.1. Comparison of sample and (reference) mass standard .............................. 84
19.2. Checkweighing, dosing functions in percent setup mode ............................ 85
19.3. Interpreting the function by use of a bargraph ......................................... 86
19.4. Additional settings of Percent Setup mode ............................................... 86
20. DENSITY ......................................................................................... 87
20.1. Set for density estimation of liquids and solid objects ............................... 88
20.2. Assembly ............................................................................................ 89
20.3. Determining density of solid object ......................................................... 90
20.4. Determining density of liquids ................................................................ 91
Page 5
- 5 -
20.5. Density of air ....................................................................................... 92
20.6. Additional settings of density mode ........................................................ 93
20.7. Printouts ............................................................................................. 94
20.8. Report on completed density determination processes .............................. 96
21. ANIMAL WEIGHING ........................................................................ 97
21.1. Setting of animal weighing mode............................................................ 98
21.2. Additional settings of animal weighing mode ............................................ 99
22. FORMULATION .............................................................................. 100
22.1. Additional settings of formulation mode .................................................101
22.2. Formulation – quick access keys ...........................................................102
22.3. Adding formulas to the Database of Formulas .........................................102
22.4. Using formulas in weighing ...................................................................104
22.5. Printouts ............................................................................................106
22.6. Report on completed formula making processes ......................................107
23. STATISTICS .................................................................................. 109
23.1. Setting of keys and workspace for Statistics mode ..................................110
23.2. Additional settings of Statistics mode .....................................................110
23.3. Parameters related to a series of measurements .....................................111
24. PIPETTE CALIBRATION ................................................................. 113
24.1. Additional settings of pipettes calibration mode .......................................117
24.2. Pipettes calibration – quick access keys .................................................118
24.3. Adding a pipette to the Database of Pipettes ...........................................118
24.4. Printouts ............................................................................................119
24.5. Activating procedure ............................................................................120
24.6. Report on completed calibration processes .............................................122
25. DIFFERENTIAL WEIGHING ............................................................ 124
25.1. Additional settings of differential weighing mode .....................................125
25.2. Differential weighing – quick access keys ...............................................125
25.3. Introducing a series to the Database of Series ........................................127
25.4. An example of process for differential weighing mode ..............................127
25.5. Copy tare ...........................................................................................134
25
.6. Using option “SAMPLE SELECTION”........................................................134
25.7. Deleting a value ..................................................................................135
25.8. Printouts ............................................................................................135
26. STATISTICAL QUALITY CONTROL – SQC ....................................... 137
26.1. Working mode activating procedure .......................................................138
26.2. Additional settings of the SQC mode ......................................................140
26.3. Carrying out a control process ..............................................................140
26.4. Report on product control .....................................................................141
27. PREPACKAGED GOODS CONTROL .................................................. 143
27.1. Working mode activating procedure .......................................................143
27.2. Control settings...................................................................................144
27.3. Local settings of PGC mode ..................................................................145
27.4. Editing product for control ....................................................................145
27.5. Control start procedure ........................................................................148
27.6. Abort control procedure .......................................................................149
Page 6
- 6 -
27.7. Aborted control restoring procedure, power loss case ..............................150
27.8. Non-destructive average tare control mode ............................................150
27.9. Non-destructive empty – full control mode .............................................156
27.10. Destructive empty-full and full-empty control mode ............................158
27.11. Control according to internal criteria ..................................................158
27.12. A report on determining average tare value ........................................160
27.13. A report on product control ..............................................................161
28. MASS CONTROL ............................................................................. 162
28.1. Working mode activating procedure .......................................................162
28.2. Mass control global settings ..................................................................162
28.3. Mass control process............................................................................162
28.4. Printouts ............................................................................................164
28.5. Report on completed mass control processes ..........................................165
29. DATABASES................................................................................... 166
29.1. Processes carried out on databases .......................................................167
29.2. Products .............................................................................................168
29.3. Weighing records ................................................................................170
29.4. Clients ...............................................................................................171
29.5. Formulation ........................................................................................171
29.6. Reports on formulation ........................................................................171
29.7. Reports on density ..............................................................................172
29.8. Controls .............................................................................................173
29.9. Average tares .....................................................................................174
29.10. Pipettes .........................................................................................175
29.11. Reports on pipette calibration processes ............................................176
29.12. Series ............................................................................................176
29.13. A report on SQC ..............................................................................177
29.14. Minimal sample weight.....................................................................178
29.15. Mass controls..................................................................................180
29.16. Ambient conditions ..........................................................................182
29.17. Packages........................................................................................182
29.18. Warehouses ...................................................................................182
29
.19. Printouts ........................................................................................183
29.20. Universal variables ..........................................................................185
29.21. Operating database .........................................................................185
Export database of weighing records to file ........................................... 185 29.21.1.
Delete database ................................................................................ 187 29.21.2.
Delete weighments and reports ........................................................... 188 29.21.3.
30. COMMUNICATION ......................................................................... 189
30.1. RS 232 port settings ............................................................................189
30.2. ETHERNET port settings .......................................................................189
30.3. WiFi settings .......................................................................................190
30.4. TCP protocol settings ...........................................................................191
31. PERIPHERAL DEVICES................................................................... 191
31.1. Computer ...........................................................................................192
31.2. Printer ...............................................................................................192
31.3. Barcode scanner .................................................................................194
Page 7
- 7 -
31.4. Transponder card reader ......................................................................195
31.5. Additional display ................................................................................196
31.6. Ambient conditions module ...................................................................197
31.7. Tablet feeder (automatic feeder) ...........................................................197
32. INPUTS / OUTPUTS ....................................................................... 198
33. OTHER PARAMETERS ..................................................................... 200
33.1. Interface language ..............................................................................200
33.2. Date and time setting ..........................................................................200
33.3. “Beep” sound ......................................................................................200
33.4. Volume ..............................................................................................201
33.5. Sleep mode ........................................................................................201
33.6. Display brightness ...............................................................................201
33.7. Touch panel calibration ........................................................................201
33.8. Level control .......................................................................................202
33.9. Sensor sensitivity ................................................................................202
33.10. Proximity sensors delay ...................................................................203
33.11. Autotest .........................................................................................203
33.12. Start logo .......................................................................................207
33.13. Export of system events...................................................................207
34. UPDATE ......................................................................................... 208
35. SYSTEM INFORMATION ................................................................. 208
36. MOVIES ......................................................................................... 209
37. COMMUNICATION PROTOCOL ....................................................... 211
37.1. A set of commands ..............................................................................211
37.2. Responses format for commands sent from computer level ......................212
37.3. Manual printout / Automatic printout .....................................................223
38. CONNECTING PERIPHERAL DEVICES ............................................. 225
39. DIAGRAMS OF CONNECTING CABLES ............................................ 225
40. ERROR MESSAGES ......................................................................... 226
41. ADDITIONAL EQUIPMENT ............................................................. 227
42. APPENDIX A – Variables for printouts........................................... 228
42.1. List of variables ...................................................................................228
42.2. Variables formatting ............................................................................237
43. APPENDIX D – Settings of CITIZEN printer ................................... 239
44. APPENDINX E - Settings of ZEBRA printers .................................. 240
45. APPENDIX F – Communication with barcode scanners .................. 240
46. APPENDIX G – Menu structure ...................................................... 241
Page 8
- 8 -
Page 9
- 9 -

1. BASIC INFORMATION

INTENDED USE
Balances 3Y series are intended to precise mass measurement of weighed loads conducted in laboratory conditions.
PRECAUTION MEASURES
Before first use of the balance, it is highly recommended to carefully
read this User Manual, and operate the balance as intended.
Do not operate the touch panel using sharp edged tools (knife,
screwdriver, etc.).
Place weighed loads in the center of balance’s weighing pan.
Load the balance’s weighing pan with loads that gross mass does not
exceed instrument’s measuring range (maximal capacity).
Do not leave heavy loads on balance’s weighing pan for a longer period
of time.
In case of defect immediately unplug the instrument from mains.
Balances to be decommissioned, should be decommissioned in
accordance with valid legal regulations.
Do not use the balance is areas endangered with explosion. Balance Y2
series is not designed to operate in EX zones.
WARRANTY CONDITIONS
A. RADWAG feels obliged to repair or exchange all elements that appear to be
faulty by production or their construction,
B. Defining defects of unclear origin defects and means of their elimination can
only be realized with assistance of manufacturer and user representatives,
C. RADWAG does not bear any responsibility for defects, losses or accidents
resulting from unauthorized or inadequate performing of production or service processes,
D. Warranty does not cover:
Mechanical defects caused by product exploitation other than intended,
defects of thermal and chemical origin, defects caused by lightning, overvoltage in the power network or other random event,
Balance defects if it is utilized contrary to its intended use,
Page 10
- 10 -
Balance defects, if service claims removing or destroying product’s
protective stickers which protect the balance’s housing against unauthorized access.
Mechanical defects or defects caused by liquids and natural wear,
Balance defects caused by inappropriate setting of a defect of electric
power network,
Defects caused by overloading balance’s mechanical measuring system,
Maintenance activities (cleaning).
E. Loss of warranty takes place if:
A repair is carried out outside RADWAG sales office or authorized service
point,
Service claims intrusion into mechanical or electronic construction by
unauthorized people,
Other version of the operating system is installed in a balance,
The balance does not bear company’s protective stickers.
F. Detailed warranty conditions are listed on a service card.
Supervision over balance’s metrological parameters
Metrological parameters of a balance need to be checked by a user in determined time intervals. Inspection frequency is conditioned by ambient conditions in which a balance is used, kind of carried out processes and adopted quality management system.
The manual’s significance
It is very important to read the user manual carefully before switching on and starting up balance operation, even if the user is experienced and has worked with the this type of balance before.
Balance user training
A balance should be utilized and supervised only by users who are trained and experienced in such type of weighing instruments.
Page 11
- 11 -

2. UNPACKING AND INSTALLATION

2.1. Ultra-microbalaces UYA 3Y

Carefully unpack the balance, remove plastic, cardboard and foil packagings together with protective elements and using gently movements place the balance onto its location. Assembly the weighing pan and other elements following the diagram.
After ultra-microbalance installing at operating place, following should be set:
1 – pan shield 2 – bottom ring 3 – balance pan 4 – glass cover
After the elements setting, additional devices should be connected and balance should be plugged to power source. Power socket is located on back side of electronic module.
Page 12
- 12 -

2.2. Balances XA 3Y; XA 3Y.A

A – balance with d=0,01mg unit
elements should be set up.
After placing the balance in its intended
B – balance with d=0,1mg unit
After balance installation at operating place, remove the transport protection (9) ­press the safety catch gently and turn it accordingly to <OPEN> instruction, next remove the safety catch but remember to retain it should the device be transported in the future. Next, assembly the following parts:
bottom shield of the drying
chamber (5)
centering ring (4)
weighing pan (2)
pan shield (3) or (1)
After the elements setting, additional devices should be connected and balance should be plugged to power source. Power socket is located on back side of balance housing.
Pipettes calibration Remove the pan and put on its place:
- pipettes calibration chamber
- inside the chamber, pan and other
Balances XA 3Y.F
place of use, assembly all of its standard components:
- bottom shield of the weighing chamber (4)
- bottom ring (3)
- standard weighing pan (1)
- pan shield (2)
After assembling all of the components and connecting peripheral devices plug the balance to power source using a dedicated power adapter. Power socket is located on back side of balance housing.
Before weighing filters remove the pan shield (2), the standard weighing pan (1) and the bottom ring (3). Next, inside the weighing chamber place a glass shield (6) and assembly a pan-stand for weighing filters (5).
on completing the above process zero/tare balance indication. The system is ready weighing filters procedures.
CAUTION! Any activities have to be carried out very carefully and gently not to damage the measuring system of the balance.
Page 13
- 13 -

2.3. Balances AS 3Y series

back of balance’s housing.
Carefully remove the balance from its packaging, remove the plastic and foil transport protective elements. Gently place the balance in its intended place of use. Install the weighing pan, and other elements according to below scheme.
Analytical balances AS 3Y series A – balance with reading unit d=0,01mg
B – balance with reading unit d=0,1mg
On installing the balance in its intended place of use, remove transport
lock (1) – press the lock gently and turn it accordingly to OPEN/CLOSE instruction, next remove the lock. Please retain the lock should the device be transported in the future. Next assembly:
2 – bottom shield of the weighing chamber 3 – centering ring 4 – weighing pan 5 – pan shield
On assembling all components, connect peripheral devices to the balance, and plug the balance to mains. Power socket is located at the
Balance with transport lock mounted on
Page 14
- 14 -

2.4. Balances PS 3Y series

Precision balances PS 3Y series
back of balance’s housing.
Carefully remove the balance from its packaging, remove the plastic and foil transport protective elements, and gently place the balance in its intended place of use. Install the weighing pan, and other elements according to below scheme.
A – balance with reading unit d=1mg
B – balance with reading unit d=10mg
On installing the balance in its intended place of use, assembly:
1 – weighing pan mandrels 2 – weighing pan 3 – pan shield /in case of d=1mg/
On assembling all components, connect peripheral devices to the balance, and plug the balance to mains. Power socket is located at the

2.5. Balances APP 3Y series

Gently take the balance out of its packaging, remove plastic, carton, foil and transport protections, and gently place the balance in its intended place of use.
After unpacking unscrew bolt 1 which locks the adjustment mechanism.
Page 15
- 15 -
Before installing the balance in its intended place
of use, remove transport protections (1) and
On removing the protections and assembling the
assembly the weighing pan (2) (for APP 10.3Y, one of the pans only).
On assembling the components, connect all peripheral devices to the balance and then plug the instrument to the mains. Power socket is located at the back of balance’s housing.
weighing pan, connect the terminal to the socket located at the back of balance’s housing, and connect all peripheral devices. Only then plug the balance to mains (mains socket is located at the back of balance’s housing).

2.6. Balance cleaning

1. Disassembly the weighing pan and portable elements, depending on the balance
type the elements differ (refer to UNPACKING AND INSTALLATION for details).
Use the most gentle movements so as not to damage the mechanism. It is recommended to use tweezers while removing the pan from microbalances.
2. If possible, use vacuum keyboard cleaner to remove the dust from the weighing chamber (especially recommended for microbalances)
3. Glass parts should be cleaned using a dry flannel cloth (mild cleanser
may be used if it does not contain any abrasive substances) – for draft shield disassembly instruction go to the next section of this manual
4. Use a dry flannel cloth in order to clean disassembled parts (mild cleanser may be used if it does not contain any abrasive substances)
CAUTION:
Cleaning balance’s weighing pan if installed may damage instrument’s measuring mechanism.
For ease of draft shield cleaning process it is allowed to remove its glass parts as it has been presented below.
Page 16
- 16 -
XA 3YA series balances
Steps:
1. Unscrew and remove top glass protection, next slide the top glass out.
2. Remove left and right side glass. Before removing the glass completely pull its protection as it has been presented in the picture. Left and right side glass are not interchangeable therefore it must be remembered which is left, which is right. Upon cleaning procedure completion left and right side glass must be mounted to a correct place.
3. Undo the screws protecting the front glass to remove it.
Page 17
- 17 -
4. Remove the weighing pan and the pan shield so as not to damage the mechanism while carrying out the cleaning process.
Thus prepared draft shield and its glass parts may be thoroughly cleaned. Carry out the cleaning process carefully using the most gentle movements. Pay attention to dirt and any small items, they could affect balance operation if they got inside the balance through the opening that serves to mount the weighing pan.
Upon completion of the cleaning process assembly the balance, take actions like before but in a reverse order. Pay a special attention to left and right side glass, assembly it on the correct side of the balance. While sliding the glass back onto its place remember to pull the protection (like before when dismantling the draft shield).
XA 3Y series balance
Steps:
1. Unscrew and remove top glass protection, next slide the top glass out.
Page 18
- 18 -
2. Replace the back glass.
3. Remove left and right side glass. Left and right side glass are not interchangeable therefore it must be remembered which is left, which is right. Upon cleaning procedure completion left and right side glass must be mounted to a correct place.
4. CAUTION: the front glass cannot be dismounted.
5. Remove the weighing pan, the pan shield and the bottom shield of the weighing chamber so as not to damage the mechanism while carrying out the cleaning process.
Page 19
- 19 -
Thus prepared draft shield and its glass parts may be thoroughly cleaned. Carry out the cleaning process carefully using the most gentle movements. Pay attention to dirt and any small items, they could affect balance operation if they got inside the balance through the opening that serves to mount the weighing pan.
Upon completion of the cleaning process assembly the balance, take actions like before but in a reverse order. Pay a special attention to left and right side glass, assembly it on the correct side of the balance. While sliding the glass back onto its place remember to pull the protection (like before when dismantling the draft shield).
AS 3Y series balance
Steps:
1. Unscrew and remove top glass protection, next slide the top glass out.
Page 20
- 20 -
2. Remove the back glass
3. Remove left and right side glass. Left and right side glass are not interchangeable therefore it must be remembered which is left, which is right. Upon cleaning procedure completion left and right side glass must be mounted to a correct place.
4. CAUTION: the front glass cannot be dismounted.
5. Remove the weighing pan, the pan shield and the bottom shield of the weighing chamber so as not to damage the mechanism while carrying out the cleaning process.
Page 21
- 21 -
Thus prepared draft shield and its glass parts may be thoroughly cleaned. Carry out the cleaning process carefully using the most gentle movements. Pay attention to dirt and any small items, they could affect balance operation if they got inside the balance through the opening that serves to mount the weighing pan.
Upon completion of the cleaning process assembly the balance, take actions like before but in a reverse order. Pay a special attention to left and right side glass, assembly it on the correct side of the balance. While sliding the glass back onto its place remember to pull the protection (like before when dismantling the draft
shield).

2.7. Plugging to mains

The balance should be plugged to mains using the original power adapter, which is balance’s standard equipment. Plug the power adapter’s plug to balance’s socket located at the back of the housing.

2.8. Connecting peripheral equipment

Use only accessories and peripheral equipment recommended by the manufacturer of your balance. The balance must be disconnected from the mains before connecting or disconnecting any peripherals (printer, PC computer, computer keyboard). On connecting the peripherals, plug the balance to mains.
Page 22
- 22 -

3. BALANCE DESIGN

3.1. Dimensions

Ulctramicrobalance UYA 3Y series
Balance XA 3Y.A series
Page 23
- 23 -
Balance XA 3Y series
Balance XA 3Y.F series
Page 24
- 24 -
Balance AS 3Y series
Balance PS 3Y series
Page 25
- 25 -
Balance APP 3Y series
Page 26
- 26 -

3.2. Interfaces

3 – USB port
Description of available interfaces
1 – Ethernet RJ45 2 – RS232 interface (COM1)
4 – IN / OUT, RS232 (COM2)
Interfaces RS232 and IN/OUT
RS232 - socket DB9/M (male), front view:
Pin2 - RxD Pin3 - TxD Pin5 - GND
IN/OUT, RS232 socket DSUB15/F (female), front view:
Pin1 - GNDWE Pin2 - OUT1 Pin3 - OUT2 Pin4 - COMM Pin5 - 6÷9VDC Pin6 - IN4 Pin7 - IN3 Pin8 - TxD2 Pin9 - 5VDC Pin10 - GNDRS Pin11 - IN2 Pin12 - IN1 Pin13 - RxD2 Pin14 - OUT4 Pin15 - OUT3
Page 27
- 27 -

3.3. Balances with a wireless connection between the terminal and the weighing module (an option)

The terminal operates on bateries, wireless connection ensures its cooperation with the weighing module. This, in contrast to a standard design, allows to eliminate cable connecting the devices.
Wireless connection solution is marked by letter <B>, eg. AS 220.3Y.B.
Features:
Maksimum range: 10 m Maxumum operation time of the terminal (without the necessity to charge
baterries): 8 h
Usage:
1. Lack of additional sources of vibration
Terminal located elsewhere than the weighing bench eliminates vibrations being a result of operation. This ensures stable working conditions for balances with the highest. No specific weighing benches are needed.
2. Comfort of weighing in fume cupboards and laminar flow cabinets
Wireless connection between a terminal and a weighing module positively influences weighment process making it more comfortable and safer.
3. Elimination of unfavourable influence factors
Due to wireless connection the weighing module may be completely separated from the operator by being placed in an anti-draft chamber. This makes weighing of contaminated and toxic substances possible without putting one’s health and life to risk.
4. Improved pipetting ergonomics
Pipettes calibration process requires precision and quickness of operation. Wireless connection redefines comfort of operation which will definitely be appreciated by the user.
Wireless solutions are equipped with 2 power suppliers. One used for supplying the weighing module, the other for powering the terminal.
Page 28
- 28 -

Start-up procedure 3.3.1.

No.
Icon
Description
1
2
On unpacking and placing the balance onto its workstation it is necessary to connect the weighing module and the terminal to mains. Next the user shall proceed like in case of a standard design (description to be found further down this manual). Main window preview for a wireless connection solution is presented below.
Main window of a wireless connection solution features two extra icons when compared to a standard design:
1. battery charge status icon
2. wireless connection icon
Operation and functions are identical like for standard design balances.

Icons description 3.3.2.

Battery charge status icon:
1
2
3
Battery charging
Battery charged, terminal connected to mains
Battery charged, terminal not connected to mains
4
5
Wireless connection icon:
Battery status (about 50%), terminal not connected to mains
Minimal battery status, necessity of connecting the terminal to mains
Page 29
- 29 -
No.
Icon
Description
1
Correct connection of wireless modules – correct cooperation
2
Connection interrupted – loss of connection
Terminal connected to mains, battery charged, no connection of the terminal with the weighing module.

Settings 3.3.3.

For ensuring long enough operation while powering the device by means of battery, the correct parameters of group OTHER must be set, <Sleep mode> <Display brightness>. For more information refer to point 33 of this manual titled Other Parameters.
Page 30
- 30 -

4. STARTUP

Key
Description
Sending measurement result to a plugged printer or computer
On plugging the balance to mains, instrument’s diode ON/LOAD
located on indicator’s housing is lit up.
Press powering key located on the upper left section of terminal’s
overlay. Within a few seconds, the OS Windows CE and RADWAG software start loading, which is signaled by flickering red diode ON/LOAD.
On completing the startup procedure, the instrument’s displays main
window of the balance software.
The balance starts up with no user logged in. In order to start operation,
a user has to log in. (logging procedure is described further in this user manual).
CAUTION: Remember to start the balance with no load on the weighing pan.

5. BALANCE KEYBOARD – FUNCTION KEYS

Switching the instrument ON / OFF
Zeroing
Tarring
Function key (entering instrument’s menu)
Working mode selection, (programmable key)
User profile selection, (programmable key)
Internal adjustment, (programmable key)
Page 31
- 31 -

6. SOFTWARE STRUCTURE

ADJUSTMENT
USERS
PROFILES
DATABASES
COMMUNICATION
PERIPHERALS
INPUTS/OUTPUTS
ACCESS LEVEL
OTHERS
UPDATE
INFORMATION ON
SYSTEM
MOVIES
The structure of balance’s main menu is divided into function groups. Each group comprises parameters grouped by their reference. Description of each menu group is provided further in this user manual.
List of groups - Parameters
Balance’s main menu is accessed by pressing SETUP function key or soft key located in the bottom tool bar of balance’s display. The menu comprises parameters referring to balance settings, functions and profiles.
Page 32
- 32 -

7. SOFTWARE WEIGHING WINDOW

The software’s main window is divided into 4 main sections:
The upper part of the touch panel displays data on active working mode,
logged operator, date, time, active connection to a computer and current level status of a balance.
Below there is a weighing window, indicating measurement result and
current measuring unit
Under which there is a grey coloured workspace containing additional data
on weighing process and activities
Caution:
Data contained in the workspace is optionally configurable. Configuring process is described in point 15.4 of this user manual;
Below the workspace there is a set of soft keys:
Caution:
Balance user can declare the presence of function (quick access) keys. Declaring process is described in point 15.3 of this user manual;
Page 33
- 33 -

8. LOGGING

Full access to balance’s parameters and editing databases requires logging to
instrument’s software as an operator with <Administrator> access level. The
logging procedure should be carried out on each switching on the balance.
Log in procedure
While in main weighing window of the application, press <Log in> text
located in the upper window bar, which opens a window with operators
database including < Admin>entry,
On entering the < Admin> entry, the software activates on-screen
keyboard for entering operator’s password,
Insert password “1111” and accept it by pressing key,
The software automatically returns to displaying the main window, and
the display’s upper bar indicates word <Admin> instead of <Log in>,
After first logging in, the administrator should add user profiles and set
their corresponding access level (the procedure of assigning access levels
is described further in this user manual, see point 12 and 13).
On future logging in, select a user from the list, and on entering password balance’s software initiates operation with access level set for a corresponding user.
Log out proc edure
While in main weighing window of the application, press name of a logged
operator located in the display’s upper bar. A window with operators database is opened,
Press <Log out> soft key (located as position no. 1 in the list of
operators)
The software returns to the main window, and the display’s upper bar
indicates word <Log in> instead of name of a logged operator.
Page 34
- 34 -
Access level
Authorization
Access level
None (guest)
No access to editing all of user parameters.
No access to menu <Databases>.
Operator
Access to editing parameters in submenu: <Profiles;
Advanced
Access to editing parameters in submenu: <Profiles;
Administrator
Access to all user parameters, functions and editing
Software of balances Y2 series provides four access levels: administrato r , advanced operator, operator and none (anonymous logging).
On switching on the balance, its display is constantly active, enabling carrying out mass measuring processes. These measurements, however, cannot be saved in balance’s database unless a user with at least minimal access level is logged in. The minimal access level enables carrying out mass measuring processes and saving them in balance’s databases using available function keys.
Below table indicates access to editing user parameters, databases and software functions depending on active access level.
it is not possible to accept a weighing record nor to start a process.
Readout> and settings in group of parameters <Others>, except for <Date and Time>. The operator can start and carry out all weighing processes, and has access to <Export weighing database to a file> and preview data in <Databases>. An operator can define universal variables.
operator
Readout>; <Working modes>; <Communication>; <Peripherals>; <Others> except for <Date and Time>. The advanced operator can start and carry out all weighing processes, and can erase old data from the <Databases>.
databases.
Page 35
- 35 -

9. OPERATING USER MENU

Reading printout template from a file format *.lb
device to terminal’s USB port)
Moving in menu of balances 3Y series is intuitive. The touch panel makes the operation of balance’s software very simple. Pressing a function key, a soft key or area on the display initiates an assigned function or process.

9.1. Terminal’s keyboard

or
Enter main menu
Scroll menu upwards
Scroll menu downwards,
Fast up and down menu scrolling,
Accept changes,
Leave a function unchanged (without saving changes),
Add items to a database,
Search for items in a database by date
Search for items in a database by name
Search for items in a database by code
Printout of an item from a database
Clear the editing field
Activate / deactivate on-screen keyboard
(function key is active on plugging a data storage
Selection of variables from a list for a printout
Moving (exiting) one level up in menu structure
Page 36
- 36 -

9.2. Return to weighing mode

Changes in instrument’s memory are saved permanently on returning to weighing mode.
Procedure:
- Press key for a few times, until the display indicates to the weighing mode.
- press soft key in the upper bar on the display. The software immediately returns to displaying the main window.
Page 37
- 37 -

10. WEIGHING

Verified
balance
Load a weighed object on balance’s weighing pan. On stabilization of weighing result, indicated by stability marker visible on the side of balance’s display, read the measurement result.
Saving / printout of the measurement result is available on pressing <PRINT> key:
In case of verified balances – only stable measurement result can be saved or printed (stability marker visible on balance’s display)
In case of non-verified balances – stable or unstable measurement result can be saved or printed (regardless of stability marker presence). If unstable measurement result is printed then it is
accompanied by question mark <?> in front of printed mass value.

10.1. Selection of measuring unit

Change of measuring unit is carried out by pressing the measuring unit icon visible next to the value of measurement result on balance’s display. Pressing the measuring unit icon opens a list with available units. On selecting one of them, the software automatically calculates the indicated mass value to the active measuring unit.
Available measuring units:
Unit denotation
gram [g] yes
milligram [mg] yes *
kilogram [kg] yes *
carat [ct] yes *
pound [lb] no
ounce [oz] no
ounce Troy [ozt] no
pennyweight [dwt] no
Taele Hong Kong [tlh] no
Taele Singapore [tls] no
Taele Taiwan [tlt] no
Momme [mom] no
Grain [gr] no
Newton [N] no
Tical [ti] no
Page 38
- 38 -
* - Accessibility of measuring units is conditioned by type of balance and its
YES
NO
YES
NO
NO
NO
verification status. In non-verified balances all measuring units including those outside the International System of Units are available.

10.2. Means of correct weighing procedure

In order to ensure long lasting use of a balance, correct and reliable
measurement of weighed loads, follow below procedures:
Start the balance with no load on the weighing pan
(permissible value of load on the weighing pan on balance start is ±10% of its maximal capacity).
Load the weighing pan steadily avoiding shocks:
 Place weighed loads centrally on the weighing pan:
 Avoid side loading, in particular side shocks:
Page 39
- 39 -

10.3. Balance leveling

Balance Y2 series features AutoLEVEL System, which continuously monitors balance’s level status during its operation. The level status is signaled in the upper right corner of balance’s display. The system continuously controls level, and on detecting level change it indicates appropriate information on the display and/or signals an alarm. Then the software opens a window for adjusting the level.
Balance leveling procedure
Press level status pictogram< > located in the upper right corner of the display.
The display indicates a control panel of leveling function. Balance operator previews level status, and balance’s top elevation.
Level the balance by turning the adjustable feet in a way indicated by
blinking pictograms on the display < > – the level point should move towards the center of the level circle.
As the level point is in the center point, its colour changes from red to green – which is a confirmation for correct level status.
CAUTION: The balance is equipped with Automatic Level Control mechanism. Description to be found in point 33.7 of this manual.

10.4. Zeroing

In order to zero mass indication press key. The mass indication on the display should equal zero, and precise zero and stability markers should appear.
Zeroing process is tantamount to determining new zero point recognized by the balance as precise zero. Zeroing is possible only in stable status of display indication.
Caution:
Zeroing the display indication is possible only within the ±2% range of instrument’s maximal capacity. If zeroed value is above the ±2% of the maximal capacity, then the software indicates an error message Err2.
Page 40
- 40 -

10.5. Tarring

In order to determine net weight of a weighed object, place object’s container (packaging) on balance’s weighing pan, and on stabilization of measurement
result press key. The display should indicate mass equal to zero and
symbols: Netand . On taking off the weighed load and its packaging from
instrument’s weighing pan, the display indicates sum of total tarred mass with minus sign.
The software also enables assigning tare value to a product from a database. Then, on product selection from a database, the software automatically uploads data on tare value for the specific product.
Caution:
Tarring negative values is impossible. On tarring negative values the balance responds with an error message. In such case zero balance’s indication and repeat tarring procedure.
Manual tare determination
Procedure:
• While in optional mode press quick access key ,
• Which opens numeric keyboard on balance’s display,
Using numeric keys insert desired tare mass and press key,
The balance returns to weighing mode, and the display indicates inserted
tare value with minus “–“ sign.
Deleting tare
The tare value indicated on balance’s display can be deleted by pressing ZERO key on balance’s front panel, or using programmable function key <Deactivate tare>
PROCEDURE 1 – on taking the tarred load off balance's weighing pan
Press ZERO key
The NET marker is deleted, and new zero point of the balance is determined
PROCEDURE 2 – when the tarred load in on balance’s weighing pan
Press ZERO key
The NET marker is deleted, and new zero point of the balance is determined
If tare value exceeds 2% of balance’s maximal capacity, the display
indicates error message –Err- (forbidden operation)
PROCEDURE 3 - when the tarred load in on balance’s weighing pan or on
taking the tarred load off balance's weighing pan
• Press programmable key <Deactivate tare>
The NET marker is deleted,
• The display indicates tare value
Press <Restore tare> key to restore the last tare value
Page 41
- 41 -
→0←
→0←

10.6. Weighing on dual range balances (ref. balance PS 200/2000.3Y series)

Transition from weighing with accuracy of the I weighing range to weighing with accuracy of theII weighing range takes place automatically on
exceeding Max1 200g (without operator’s interference). On entering the
accuracy of the II weighing range, the display indicates < II> pictogram on its
left side, the last digit is simultaneously blanked, and last but one digit is differentiated. From now on the balance continues weighing in the II weighing range.
I weighing range
II weighing range
In order to return to weighing with the accuracy of the I weighing range
follow below procedure:
• take the load off balance’s weighing pan;
as the indication returns to zero and symbols:
press
key.
and are lit up,
The balance returns to weighing with readability d1=0.001g and the marker of
II weighing range <II> is blanked.

10.7. Use of air buoyancy compensation factor

The application enables correcting mass measurement errors during:
1. Weighing loads that density considerably differs from the density of standard used to adjusting the balance. Regularly, the balance is adjusted with a steel mass standard with density ~8.0g/cm3 or a brass mass standard with density ~8.7g/cm3. If loads made of other substances are weighed, then ratio indicated on below chart should apply.
Below chart presents the value of mass correction in relation to the density of weighed load, assuming that air density is constant and equals
1.2 kg/m3.
Page 42
- 42 -
Correction value [mg]
Sample mass [g]
Density of weighed load
Error value in relation to density of weighed sample
2. Testing the changes of sample mass within a few hours’ time, if: sample mass is relatively constant (minor changes). In such case it is assumed, that considerable effect on the measurement result rests on air density changes, which is affected by changes of pressure, temperature and humidity.
In order to carry out a reliable measurement, apart from being familiar with relations applying to ambient conditions, an operator should know the measuring methods, features of tested sample, air density in the laboratory room and density of weighed sample.
FUNCTIONING
The application provides two ways to use the air buoyancy compensation procedure.
1. Inserting the specified value of air density and specified value of weighed sample density to balance’s memory.
On inserting these values the application automatically calculates the correction factor for weighed mass and indicates corrected value of weighed mass on balance’s display.
Page 43
- 43 -
In order to avoid errors, the corrected mass value is accompanied by <!>
symbol visible on balance’s display and a printout.
2. The application semi-automatically determines the value of air density. The specified density of the weighed sample is inserted to balance’s memory by an operator.
Determination of air density is carried out using a set of two mass standards. One of them is made of stainless steel, and the other of aluminum. Based on mass indications of the two standards, the software automatically calculated the air density and saves the value in balance’s memory on operator’s confirmation. Then the density of weighed sample has to be inserted to balance’s memory.
On inserting these values the application automatically calculates the correction factor for weighed mass and indicates corrected value of weighed mass on balance’s display.
As in case 1, the corrected mass value is accompanied by <!> symbol
visible on balance’s display and a printout.
Air buoyancy conmpensation mode is enabled / disabled from the level of user menu. The application is operating in weighing mode only.
STARTUP AND OPERATION OF AIR BUOYANCY COMPENSATION
PROCEDURE
• While in main window press grey coloured workspace
• Select option <Settings>
Select parameter < Air buoyancy compensation>
Accessible settings
- Air buoyancy compensation – YES/NO
- Sample density (insert the density value of weighed sample). If an operator uses products from databases, then on product selection its density value is automatically uploaded.
- Air density – for selecting the means of inserting the air density value used for calculating buoyancy correction factor.
Settings:
VALUE – selecting this option opens a window for inserting determined value of air density (e.g.: determined using other methods). The inserted value is used to calculating buoyancy compensation factor. The value is assigned automatically on carrying out the air density determination (on its completion by
pressing < > key).
Page 44
- 44 -
ONLINE – selecting this option downloads current air density value from a connected THB sensor, if such is connected to the balance, or from internal sensor installed in the balance. If a balance features both types of sensors (THB module and internal sensor), then data from the THB module supersedes the
other, and parameters Setup/Ambient Conditions/Ambient Conditions Module require setting to SAVE or SAVE & ALERTS
mode. If a balance features only one type of the sensors for temperature, humidity and pressure recording, then correct ONLINE operation requires setting of balance parameters:
Internal sensors only - Setup/Ambient Conditions/Ambient Conditions Module set to NONE,
External THB module only - Setup/Ambient Conditions/Ambient Conditions Module set to SAVE or
SAVE & ALERTS mode.
Additionally, for correct balance cooperation with the THB module set port’s transmission parameters to compatible with those of the THB module. the THB module transmission parameters are detailed on its data plate.
On returning to the weighing mode, the display indicates an additional symbol
<!>, as indicated on below drawing.From now on the value of indicated mass
is corrected including the air buoyancy compensation factor and sample density.
The measurement result can only be compensated by a correct value if an operator inserts to balance’s memory the actual value on the air density and correct density of weighed sample.
Caution:
If the <AIR BUOYANCY COMPENSATION> mode is set to <ONLINE> value, then balance display does not indicate pictogram <!> which means that parameter Setup/Ambient Conditions/Ambient Conditions Module is set to SAVE or SAVE & ALERTS mode, and there is no physical connection between a balance and a THB module, or the parameters for communicating the module with the balance are set incorrectly. In such case plug a THB module to balance’s ports COM 1 or COM 2 and set appropriate cooperation parameters, such as specified on module’s data plate.
Page 45
- 45 -

10.8. Additional parameters on weighing process

Changes in settings of functions related to weighing process can in some extend program balance reactions.
Procedure:
1. Press grey coloured workspace
2. Balance display indicates menu: Settings, Keys, Information, Printouts, Profile
3. Press one of available submenu options and select item for modification,
Menu <Settings> - additional options related to weighing mode
Menu <Keys> - defining quick access keys Menu <Information> - supplementary data on weighing process displayed in
the grey coloured workspace Menu <Printouts> - selection of printout type Menu <Profile> - selection of profile, active during balance operation
Menu <SETTINGS> - contains supplementary data on weighing process, such
as:
Tare mode
SINGLE,
mass value stored in balance memory on single pressing of TARE key, the following pressing of TARE key determines new value of tare. Selection of product or packaging with assigned tare value causes automatic deleting of previously assigned tare value;
SUM OF ACTIVE,
totalizing applied tare values for a product or a packaging (resulting from selecting a product or a packaging from corresponding database) with a possibility of increasing determined tare value by manual tare introduced manually using balance’s numeric keyboard. On following determining of tare value for a product or a packaging, the tare value introduced manually is deactivated.
SUM OF ALL,
totalizing all introduced tare values.
AUTOTARE
Means of operation: Each first stable measurement result is tarred. The NET pictogram flashes on the display. The operator can determine net mass of weighed load. On taking off the load from balance’s weighing pan, and as the indication returns to autozero zone, the software automatically deactivates Tare value.
Automatic footer printou t
Accessible options:
MODE - No – manual footer printout,
Page 46
- 46 -
Sum of measurements – the condition for printing a footer is
exceeding mass value set in parameter <Threshold>
Number of measurements – the condition for printing a footer
is carrying out a pre-defined number of measurements (a batch) set in parameter <Threshold>
THRESHOLD – setting the value of threshold determining footer
printout. For option <Sum of measurements> the value is determined in measuring unit [g], and for option <Number of measurements> the value is not measurable, and determined by number of carried out measurements.
Printout mode / printout release
Function key PRINT / printout release, (manual operation)
Never – printout deactivated First stable – the first stable measurement is printed Each stable – all stable measurements are set for printing
Each – printout of all measurements (stable and
unstable), in case of a verified balance only stale measurement results are printable (as in setting <Each stable>)
AUTOMATIC MODE
Never – printout deactivated First stable – the first stable measurement result obtained after
placing a load on balance’s weighing pan is recorded, record of the following measurement result takes place only on unloading the weighing pan, returning of the mass indication below set threshold’s value, and placing another load on balance’s weighing pan
Last stable – accepted is the last stable measurement recorded
before taking the load off the weighing pan. The record takes place on taking the load off the balance’s weighing pan, and returning of the mass indication below set threshold’s value.
THRESHOLD
mass value obligatory for operation of automatic printout. Set in grams.
Printout
Contains type of printout that is related to a working mode. Printout takes place on pressing PRINT key on balance’s overlay.
Accessible options:
STANDARD PRINTOUT
Possible declaration of printout content: HEADER, WEIGHING DATA and FOOTER. printout components marked as <YES> in the menu are printed on pressing printout activating function key.
NON-STANDARD PRINTOUT
Page 47
- 47 -
Database of printouts enables selecting one of available non­standard printouts visible in menu <PRINTOUTS>, or designing a unique printout which is automatically added to the databases.
Caution: Means of designing printouts is described in point 15.5 of this user manual.
Air buoyancy compensation
Contains parameters allowing the user to switch on the correction and enter data relating to density of the sample and to density of the air.
CAUTION: function operates only for weighing mode.
For instruction and information on air buoyancy compensation operation
see point 10.7.

10.9. Minimum sample

The weighing mode settings comprise a function of <Minimum sample>. The function is enabled on specifying in the database <Database/Minimum sample > parameters on methods for determining the minimum sample and the values of minimal sample weight specific for a method. In standard version, this database item is not filled in with data.
Activities related to determining minimum sample and adding data to the database: <Databases/Minimum samples> are accessi b le only for an authorized “RADWAG” personnel.
Should balance operator use the minimum sample mode, and data on minimal sample weight are not entered to balance menu, then please contact the local representative of “RADWAG” company. An authorized RADWAG representative determines the minimum sample dependent on packaging mass. The process is carried out with mass standards, in balance place of use and in accordance with the requirements of the accepted quality system. The obtained values are entered to balance software: <Databases/Minimum samples>. A minimum sample determining method may have a few tare values defined, along with assigned values of minimum samples, expiry dates of the carried out measurements and entered data. The settings are not to be
edited by balance user.
Using <Minimum sample> mode guarantees that the results of completed weighing processes are within set tolerance limits, and they comply with the requirements of the quality system accepted by an organization.
CAUTION: the mode operates in the weighing mode only.
Page 48
- 48 -
Available options:
METHOD
It is a means for designating an accepted quality standard. Pressing a field displays a window with a list of minimal sample weight values stored in balance’s memory, following a criterion on determining the minimum sample. Adding a new method is available only from the menu level <Databases /Minimum samples>
MODE
Lock – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user whether the weighed mass is below or above the minimum sample limit. Additionally the software disables accepting the measurement result if it is below the minimum sample limit
Warn – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user whether the weighed mass is below or above the minimum sample limit. The user can accept a measurement that is below the minimum sample limit, but on a printout such measurement is preceded by an asterisk (*).
CAUTION: Means of setting a method of minimum sample determining is described in point 27.9. of this user manual: <Databases /Minimum sample>. RADWAG authorized personnel is only authorized to add new values of minimum samples.
Weighing with application of <MINIMUM SAMPLE> mode
If during a weighing process a user wants to be informed whether a
measurement exceeds the minimum sample limit for a specific weighing
threshold, then the <Minimum sample> mode has to be enabled in the
additional setting of the weighing mode.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts and Profiles
3. Select menu <Settings> - additional options on weighing process
4. Press < Minimum sample> field
5. Below window is opened
Page 49
- 49 -
6. Press <Method> field
7. Which opens a window containing a list of minimum sample methods stored in balance database.
8. Select one of available entries.
9. The software returns to the previous menu window
10. Press <Mode> field
11. Which opens a window with parameter settings. Select one of below options:
Lock – on selecting this option and during the weighing process the
display indicates corresponding pictograms that inform a user whether the weighed mass is below or above the minimum sample limit. Additionally the software disables accepting the measurement result if it is below the minimum sample limit
Warn – on selecting this option and during the weighing process the
display indicates corresponding pictograms that inform a user whether the weighed mass is below or above the minimum sample limit. The user can accept a measurement that is below the minimum sample limit, but on a printout such measurement is preceded by an asterisk (*).
12. On setting the parameters exit the menu.
13. The mass field on the display is supplemented by an additional informative pictogram. The pictogram changes during weighing process and indicates current status of mass in relation to the pre-determined value of minimum sample.
- Mass below the set value of minimum
sample
- Mass above the set value of minimum
sample for a given tare range
Page 50
- 50 -
Meaning of the pictograms of minimum sample mode:
Mass below the set value of minimum sample.
Mass above or equal to the value of minimum sample.
Mass below the set value of minimum sample. A pictogram with a clock informs on approaching expiring of the validity period of the minimum sample (it is displayed two weeks before the determined expiry date).
Mass above the set value of minimum sample. A pictogram with a clock informs on approaching expiring of the validity period of the minimum sample (it is displayed two weeks before the determined expiry date).
The validity term of the minimum sample has expired. Changes are needed in the minimum sample settings. Only authorized RADWAG personnel can change the minimum sample settings.
Caution: If more than one reference tare value is programmed (with
corresponding minimum sample mass) then the indicated value automatically moves to a range corresponding to the weight of the tarred container. At the same time the required minimum mass is changed.
Page 51
- 51 -

10.10. Cooperation with TITRATORS

In order to ensure correct cooperation with TITRATORS it is necessary to set <Mass for titrator> option to <YES> value. The setting must be done for standard printout content. Any other variables for this printout must be off.
If the option is on, the user shall notice an icon informing about specific mass printout format accepted by TITRATORS. The icon is located in the top toolbar of the main window.
Page 52
- 52 -

11. ADJUSTMENT

Balances 3Y series feature automatic internal adjustment system which ensures correct measurement accuracy. Menu <ADJUSTMENT> contains functions controlling operation of balance adjustment process, including options:

11.1. Internal adjustment

Internal adjustment process utilizes an internal weight built in balance’s housing. <Internal adjustment> function key activates automatic adjustment process. On its completion, balance’s display indicates a message box on process completion and its status.
CAUTION:
Balance adjustment requires stable measurement conditions (free from air breeze and vibrations), adjustment process has to be carried out with empty weighing pan.

11.2. External adjustment

External adjustment is carried out using an external adjustment weight, with appropriate accuracy and mass relating to balance’s maximal capacity and readability. The process takes place semi-automatically, and the following process phases are indicated on balance’s display.
CAUTION:
External adjustment is available only in balances which are not subject to conformity assessment (verification).
Process course:
Enter submenu < Adjustment > and select option: “ External
adjustment”,
Balance’s display indicates the following message box:
If there is a load on balance’s weighing pan, unload it,
Press key. The balance determines start mass, which is indicated by
a message box: “Start mass determination” on balance’s display
On determining the start mass, the balance display’s a message box:
Page 53
- 53 -
Acting according to the command, place an ordered weight/standard on
balance’s weighing pan and press key,
On completing the adjustment procedure the balance indicates a message
box:
On confirming the message box with key, the balance returns to
weighing mode.

11.3. User adjustment

User adjustment is carried out with an optional standard, which mass ranges between 0,15 Max and Max. Adjustment procedure is compatible with the external adjustment process, but before its start the software opens a message box for entering mass of a standard used for user adjustment.
CAUTION:
User adjustment is available only in balances which are not subject to conformity assessment (verification).
In order to start user adjustment, enter submenu < Adjustment >, and select option: “ User adjustment”. Then follow the commands indicated on
balance’s display.

11.4. Adjustment test

< Adjustment test > function enables comparing the result of internal automatic adjustment with the value of internal weight saved in balance’s factory parameters. The comparison is used for determining balance’s sensitivity drift over time.

11.5. Automatic adjustment

This menu option is used for selecting a factor which determines start of automatic adjustment process. Accessible options:
None – automatic adjustment disabled
Page 54
- 54 -
Time – adjustment takes place in time intervals declared in menu
<Automatic adjustment time > (11.6)
Temperature – adjustment is triggered by temperature change only Both – adjustment activation is triggered both by temperature changes
and time interval changes
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are not subject to conformity assessment (i.e. non-verified balances).

11.6. Automatic adjustment time

< Automatic adjustment time > determines time interval in which automatic internal adjustment of a balance is activated. The time interval is declared in hours and ranges between 1 and 12 hours.
Setting time interval of automatic internal adjustment:
Select option < Automatic adjustment time >Using displayed menu select appropriate time interval (given in hours)
which is a time gap elapsing from the last carried out internal automatic adjustment until activating the following automatic internal adjustment.
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are not subject to conformity assessment (i.e. non-verified balances).

11.7. Adjustment report printout

Parameter < Report printout > determines whether or not a report on automatic internal adjustment should be automatically printed on its completion.
Setting automatic report printout on adjustment completion. Go to parameter: < Report printout > and select option <YES>.

11.8. GLP project

GLP is one of means for maintaining documentation from work in accordance with adopted quality system. Data selected for printing are printed on each release of a report on balance adjustment.
Balance operator can use in a GLP report below listed information and signs:
adjustment (adjustment mode) working mode (name of a working mode)date, time, user, balance type, balance ID,level status,nominal mass, current massdifferencetemperature,
Page 55
- 55 -
blank line, dashes,signature,non-standard printout.

11.9. Adjustment history

Contains data on all carried out adjustment processes. The record is carried out automatically. Each entry on adjustment comprises basic data on completed process. Balance menu enables displaying the list of completed adjustment processes, and each report is printable.
Printing a report on adjustment process. Enter submenu < Adjustment >, and: <Adjustment history>, then select for printing adjustment entry from a
list. On displaying details of a record, press print soft key < > on display’s upper bar.
Hint:
If memory for records on completed adjustments is full, then the oldest record is automatically erased.
If internal procedures of an organization require maintaining complete documentation from all carried out adjustment processes, then the list with records on adjustment should be periodically printed and filed.
Searching for adjustment record
Balance enables searching for a specific record from completed adjustment processes:
– Press search icon and insert date of adjustment process.
Exporting data on completed adjustment process
Connect a data storage device to balance’s USB port. Press<Data export> key located in the upper right corner of balance’s display. The process is fully automatic, and on its completion, a file with extension *.tdb is saved on a data storage device connected to the USB port. The file is editable using Excel spreadsheet or a text editor.
Page 56
- 56 -

12. USERS

Menu “Users” contains list of balance operators, who are authorized to operate the instrument. The following data can be defined for each balance user:
 Name  Code  Password  Access level  Language  Profile  Card no.
Adding new user can only be carried out by balance’s Administrator.
Procedure of adding a new user: In menu Users press <Add> soft key,
Which opens a message box on the display:<Create new record>, accept by pressing corresponding key
Define all necessary fields for a new created balance user
Caution:
Search for a user in database of users by code or name.
Edit data on a user:
− Press field with name of a user
− The display indicates data on a specific user
− Select and change necessary data
Deleting a user can only be carried out by balance’s Administrator.
To delete a user:
− Press and hold user name
− A menu is opened referring to a user record
− Select option <Delete>
Page 57
- 57 -

13. ACCESS LEVEL

Access level in a balance determines scope of activities that a user can carry out. This menu can only be modified by balance’s Administrator.
Anonymous user
Balance Administrator can grant access level to a balance user who is not logged in (i.e. Anonymous user).
Procedure:
Enter group of parameters < Access level>, select option < Anonymous user>, and set appropriate access level for the anonymous user. Available access levels for an anonymous user: Guest, User, Advanced User, Administrator.
Caution: Setting <Guest>access lev el causes that logged user has no permission to change any settings on a balance.
Date and time
Balance default settings enable a user logged as the Administrator changing
date and time settings. However, the software also enables changing required access level to modify option < Date and time>.
Procedure:
Enter group of parameters < Access level>, select option < Date and time>, and set desired access level required to modify the settings. Available access levels for changing date and time settings are: Guest, User, Advanced User, Administrator.
Caution:
Setting <Guest> provides free access to date and time settings (no need to log in).
Printouts
Balance default settings enable a user logged as the Administrator editing
default printout templates. However, the software also enables changing required access level to modify option < Printouts>.
Procedure:
Enter group of parameters < Access level>, select option < Printouts>,
and select one of available options: Guest, User, Advanced User, Administrator.
Caution: Setting <Guest>access lev el provides free access to printouts settings (no need to log in).
Page 58
- 58 -
Databases
Balance Administrator is also authorized to set access level required to modifying each of the databases.
Procedure:
Enter group of parameters < Access level>, select option < Databases>, and set desired access level: Guest, User, Advanced User, Administrator for each of the databases.
Caution:
Setting <Guest>access lev el causes that access to editing each of the databases is free.
Page 59
- 59 -

14. PROFILES

A Profile is a data pack determining:
o functioning of working modes, e.g. parts counting, percent setup, etc., o what data is displayed during working mode operation, o which function keys are active, o which measuring units are accessible o which criteria are mandatory for balance’s speed of operation and
measurement stability,
Balance software enables creating numerous profiles, which in practice provides:
o each balance user to create their own and individual operating
environment
o balance operation can be easily programmed by activating function keys
and information on process which are (improving ergonomics of operation)

14.1. Creating a profile

A default profile in a balance is named <Home>. Balance Administrator can create new profiles by:
− Copying an already existing profile and its modification,
− Creating a new profile.
Copying an existing profile Procedure:
Enter balance’s main menu by pressing Setup key,
Enter submenu < Profiles>,
Press and hold an entry with profile name that should be copied,
A menu is opened referring to available options:
o Edit o Delete o Copy o Cancel
Select option <Copy>
A new profile is created named <Copy name>, and all setting are identic
with the copied profile,
After copying change necessary data in a profile: (name, etc.)
Creating a new profile
Procedure:
Enter balance’s main menu by pressing Setup key,
Enter submenu < Profiles>,
Press key, which opens a message box: <Create new record?>,
Accept the message box by pressing key. The software automatically
adds a new record and enters its editing mode.
Caution:
Adding a profile is only enabled after logging in as an Administrator.
Page 60
- 60 -
Settings
This menu enables setting an individual profile’s name (a
default on profile selecting).
Working
Contains the following submenu:
Printouts
Readout
Contains the following submenu:
Last digit
Units
Menu enables declaring the start unit, the supplementary
acceleration force in balance’s operation place.
Deleting a profile Procedure:
Enter balance’s main menu by pressing Setup key,
Enter submenu < Profiles>,
Press and hold an entry with profile name that should be deleted,
A menu is opened with a list. Select option <Delete> from the list,
A message box is displayed which requires confirming profile’s deleting:
<Confirm to delete>,
Accept the message box by pressing key, the profile is deleted.
Caution:
Deleting a profile is only enabled after logging in as an Administrator
.

14.2. Profile structure

Each profile contains the following entries:
sequence of alphanumeric characters), and declaring a default working mode (the selected mode is activated as
modes
 Additional setting of a working mode  Function keys  Displayed information
 Filter  Value release  Autozero  Autozero: Dosing
unit, 2 custom units and entering the value of gravitational
Page 61
- 61 -

Settings 14.2.1.

Name
Default working mode FILTER
On entering this option, the display opens a message box with keyboard. Insert name of a profile and accept it by
pressing key. The name is assigned to the profile.
On entering this option the user can select a specific working mode, which is set as default in a profile. In case option <None> is selected, then on selecting the profile, the balance activates last used working mode.

Working modes 14.2.2.

On entering this option, the display opens a window containing all accessible working modes. The user can introduce their settings to each of the working modes which are activated on selecting a specific profile.
In each of the working modes, the user can change the following parameters:
Settings
contain specific parameters relating to a working mode and universal settings, such as: result control, tare mode, automatic footer printout, printout mode, printout.
Functions of quick access keys
declaring quick access keys, which are visible in the bottom display bar
Information
declaring information which is visible in the grey coloured workspace
Printout
declaring type of printout of defining a non-standard printout

Readout 14.2.3.

The function enables the user to adjust balance operation to ambient conditions (filter settings) or individual user needs. Menu <Readout> contains the following elements:
Each measurement signal before being displayed is electronically processed for the purpose of obtaining corrected parameters specific for stable measurement result, i.e. ready to read.
Page 62
- 62 -
Balance user can influence to some extend the range of signal processing
by selecting an appropriate FILTER value. Available options:
recommended range of filter is between average/normal ÷ very slow.
applied value release
If the function is enabled, the following measurement results are
o very fast o fast o average / normal o slow o very slow
While setting the filtering level the user should consider the actual operating conditions of a balance. In case of very good operating conditions set the filter to average/normal or fast, in case of rough conditions set the filter to slow or very slow.
Caution:
in case of precision balances 3Y series the recommended range of filter
is between very fast
÷
average/normal;
in case of analytical balances and microbalances 3Y series the
Value release
This parameter determines displaying the stability marker for a
measurement result.
There are 3 available settings of value release parameter
o fast o fast + reliable /recommended/ o reliable
Caution: The speed of stability marker occurrence depends on applied filter value and
Autozero function
The function is to automatically monitor and correct zero indication of a balance.
compared to each other in constant time intervals. If the results differ less than declared AUTOZERO range, e.g. 1 division, the balance is automatically zeroed, and the markers of stable indication and precise zero are displayed.
Active AUTOZERO means, that each measurement starts from the precise zero point. There are, however, cases where the function may disturb the weighing process. For instance during very slow load placing or pouring onto the balance’s weighing pan. In such case, the correcting system of zero indication may also correct the actual indication of a load placed on the weighing pan.
Accessible settings: NO - autozero function disabled YES - autozero function enabled
Page 63
- 63 -
Autozero function: Dosing
The function sets autozero operation mode to default one set for
dosing mode.
Last digit
Ambient conditions
much less time than in case of setting to value unstable.
Accessible values:
NO - autozero operation is automatically disabled on entering
dosing mode
YES - autozero operation is automatically enabled on entering
dosing mode
The function determines visibility of the last decimal place indicated on measurement result. The functions provides three settings:
- ALWAYS: al the digits visible
- NEVER: last digit invisible
- WHEN STABLE: last digit visible only for a stable
measurement
This parameter refers to ambient conditions of the balance. By default the
parameter is set to value Stable but it is advisable to set the parameter to value Unstable for unstable ambient conditions. Stable setting results in faster operation, i.e. weighing takes
Unstable
Stable

Measuring units 14.2.4.

Balance user can declare in a profile: start unit, supplementary unit and two custom measuring units.
A custom measuring unit features:
o A multiplier o A name (3 characters)
If a custom unit is designed, then its name is added to the list of accessible measuring units.
This menu additionally enables inserting the value of gravitational acceleration force for balance’s place of use. It is obligatory should a balance be used to determine mass in [N].
Page 64
- 64 -
Percent setup
standard (reference).
Formulation
specifying its ingredients and their mass.

15. WORKING MODES – general information

A balance 3Y series in its standard version features the following working modes:
Weighing
Means of operation: weight of a load is determined through an indirect measurement. A balance measures gravitational force which attracts the load. An obtained result is processed to a digital format and displayed in a form of measurement result.
Parts counting
Means of operation: based on a determined mass of a single part it is possible to count another parts, assuming that mass of a single part is determined with sufficient accuracy, and that the following parts are equal in mass.
Checkweighing
Means of operation: control of sample mass with applied thresholds. A user should specify the value of low threshold <LO> and high threshold <HI>.
Dosing
Means of operation: a user should specify sample’s target mass to be obtained by pouring.
Means of operation: control of percent ratio of a sample in relation to a standard (reference). Obtained data provides percent ratio on how test sample differs from the accepted
Density
Means of operation: based on Archimedes principle, a balance determines density of solids and liquids. The mode requires an optional density determination kit.
Animal weighing
Means of operation: mass measurement takes place with application of filters dampening animal moves on a weighing pan, thus enabling obtaining a correct measurement result.
Means of operation: by adding a sequence of ingredients, a user can prepare an optional mixture or formulation. Before mixing balance software requires designing a formulation by
Statistics
Means of operation: carried out measurements are used to calculate statistical data, such as Min, Max, deviation, etc.
Pipette calibration
Means of operation: calibration of pipettes according to procedures listed in ISO 8655 or according to user
Page 65
- 65 -
requirements.
Differential weighing
Statistical Quality Control Control of Prepacked Goods
(mode not available in balance’s standard version)
(mode not available in balance’s standard version)
Means of operation: analysis of mass sample change over
time.
Working mode is intended to carrying out different types of product packing processes and aimed at monitoring and / or controlling the packing process. It enables detecting excess or lack of product quantity in a packaging.
Working mode intended for CPG (Control of Prepacked Goods) processes in accordance with the regulation on prepacked goods.
Mass control
Working mode intended to carrying out quick statistical control of samples in accordance with the requirements on a quality system and/or internal standards.
The settings of separate working modes include special functions specific to a mode. They enable adapting mode’s operation to user’s individual needs. The special settings are activated on selecting a corresponding profile. A detailed description of special functions is provided while presenting each of working modes.

15.1. Working mode selecti on

Changing working mode:
o Press name of active working mode, displayed in the left corner of the
upper bar.
o List of available working modes is displayed, o Select name of a desired working mode, and press it.
Page 66
- 66 -

15.2. Parameters related to w o rking modes

Each working mode has programmable parameters determining its functioning. Procedure of determining these parameters:
1. Press grey coloured workspace area
2. Below menu is displayed:
o <Settings> - additional options related to a mode o <Keys> - defining quick access keys o <Information> - selecting information displayed in the workspace o <Printouts> - selecting type and content of a printout o <Profile> - selecting a profile, which is active during balance’s
operation
3. Press corresponding menu item and select area for modification,
Description of basic parameters accessible in <Settings> tab is provided in point 10.8 of this user manual “Additional parameters on weighing process”. Other parameters specific to each of working modes are presented in working modes description.

15.3. Quick access keys

Balance user can define up to 7 quick access keys, which are displayed in the bottom bar.
On assigning a function to a specific key, a corresponding soft key appears, and it is located in the bottom navigation bar of the main window.
It is a so called quick access key, dedicated for most often used functions
and processes.
Page 67
- 67 -

Proximity sensors 15.3.1.

This menu section comprises settings on automatic opening and closing
Settings for a balance MYA 3Y series cover:
o Left door / the option enables changing the position of the door on the left
Settings for a balance XA 3Y.A series cover:
causes opening the door of the weighing chamber.
The balance comprises two proximity sensors enabling wave control of the balance therefore the user does not have to either press any key on the overlay or touch the screen. The program recognizes 4 types of motions performed near the sensor:
1. Placing hand near the left sensor <Left proximity sensor>
2. Placing hand near the right sensor <Right proximity sensor>
3. Motion to the left <Proximity sensor: left motion>
4. Motion to the right <Proximity sensor: right motion>
CAUTION: using option LEFT MOTION or RIGHT MOTION, it is necessary to
remember that <Proximity sensors delay> parameter must be set to value 500
ms!
To each of the motions, one of the available options may be assigned: [None; Profile; Adjustment; Zeroing; Tarring; Tare setting; Tare switch off; Reset tare; Packaging; Print; Header printing; Footer printing; Unit; Universal Variable 1; Universal Variable 2; Universal Variable 3; Universal Variable 4; Universal Variable 5; Confirm; Abort; User; Left door; Right door; Open/Close door; Parameters; Product, Warehouse; Client]. The selected option is recognized by the program and performed when activated by a particular motion. In order to ensure correct operation the user must remember to set the right sensors sensitivity and the right value for proximity sensors delay (for more information refer to point 33of this manual).

Automatically opened door 15.3.2.

door of a balance featuring such mechanism, i.e. balances MYA 3Y and XA 3Y.A series. The settings may refer to the quick access keys as well as the infrared proximity sensors located on the overlay of balance’s terminal.
side of the weighing chamber
o Right door / the option enables changing the position of the door on the
left side of the weighing chamber
o Open / Close door / the option enables changing the position of the door,
e.g. if the weighing chamber is opened, then activating the option causes closing it. If the weighing chamber is closed, then activating the option causes opening the door located on the right side of the weighing chamber,
o Open / the option enables opening the door of the weighing chamber o Close / the option enables closing the door of the weighing chamber o Open / Close door / the option enables changing the position of the door,
e.g. if the weighing chamber is opened, then activating the option causes closing it. If the weighing chamber is closed, then activating the option
Page 68
- 68 -

15.4. Information

Information on weighing process is displayed in the grey coloured workspace.
- NO, information disabled
Standard printouts
selected, then the variable is disabled, i.e. it is not printed.
create it.
Non-standard printout
It contains maximally 6 pieces of information. If larger amount of data is selected, then only the first 6 are displayed. Each parameter features two options:
- YES, information enabled (displayed in the workspace)

15.5. Printouts

Menu printouts comprises two separate sections. The first one is standard printouts, the other one is non-standard printouts.
comprises three internal sections featuring different variables. For each variable set option YES to include it in a standard printout. If option NO is
PROCEDURE:
1. Press field with name of a template to be edited(Header – Weighing – Footer) and select variables to be printed
2. If a printout is a non-standard one,
o HEADER
− Dashes
− Working mode
− Date
− Time
− Balance type
− Balance ID
− User
− Level status
− Client
− Warehouse
− Product
− Packaging
− Universal variable 1 … 5
− Empty line
− GLP report
Page 69
- 69 -
o WEIGHING
Non-standard printout
Non-standard printout
BASIC PRINCIPLES OF USING PRINTOUTS
described in point 15.3 of this user manual.
− N (number of measurements)
− Date
− Time
− Level status
− Client
− Warehouse
− Product
− Packaging
− Universal variable 1 … 5
− Net
− Tare
− Gross
− Current measurement result
− Supplementary weighing unit
− Mass
o FOOTER
− Working mode
− Date
− Time
− Balance type
− Balance ID
− User
− Level status
− Client
− Warehouse
− Product
− Packaging
− Universal variable 1 … 5
− Dashes
− Empty line
− GLP report
− Signature
1. Press PRINT key on balance’s overlay to print variables that are contained in section WEIGHING of the standard printout, and if they are attributed = YES (see above list of variables in printouts).
2. Variables attributed as YES present in the HEADER and FOOTER are
printed ONLY on pressing Print Header and Print Footer soft
keys. The soft keys have to be added to the bottom bar of the display, as quick access keys.
Adding quick access keys on the bottom bar of the display is
Page 70
- 70 -
Print data from header
Print data from footer
Caution: Measuring units for printing mass indication:
Net – main measuring unit (adjustment unit)
Tare – main measuring unit (adjustment unit)
Gross – main measuring unit (adjustment unit)
Current result – currently displayed measuring unit
Supplementary unit – supplementary measuring unit
Mass – main measuring unit (adjustment unit)
Non-standard printout
Non-standard printout
a printout may contain: TEXTS and VARIABLES(which are acquired from
the software on printout).
Each non-standard printout is a separate project, featuring specific name, by which it is identified, and saved in the database of printouts.
PROCEDURE:
1. Press option <Non-standard printout>
2. Press <Add> key
3. A window is opened with data such as Name/Code/Project
4. Set name and code of a printout
5. Press <Project> key
6. A keyboard is opened on the display for editing the printout
7. Use keyboard to design printout template, the printout may contain
texts and variables.
Caution:
User can add a printout by importing configured texts from data storage devices
connected to balance’s USB port.
Printout’s name is ONLY a NAME, and it is not part of its content. Means of designing a non-standard printout template is described in point 24.11
<Printouts> of this user manual

15.6. Profiles

Profiles are described in point 14, Profiles, of this user manual.
Page 71
- 71 -

16. PARTS COUNTING

Working mode< Parts counting> enables determining quantity of small
parts (objects) with equal part mass. Counting is based on determined mass of a single part, which is:
o determined from a reference quantity of parts o acquired from database of products o inserted manually as a numeric value
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Parts counting> mode, the software automatically returns to the
main window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Gross o Standard mass o Low threshold o High threshold o Target value
On selecting the parts counting mode, the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Give mass of 1 part – editing field for inserting mass of a single part
6. Set mass of 1 part – setting mass of a single part from optional number of parts, e.g.: from 10 pcs, 20 pcs, 75 pcs, etc.
Page 72
- 72 -

16.1. Additional settings of parts counting mode

The additional settings enable adjusting the working mode to user’s needs and requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and Profile
3. Press <Settings> key
4. The display indicates functions related to weighing and parts counting modes.
Parts counting mode features the following optional functions:
ACAI, Automatic Accuracy Correction:
o YES, mass of a single part is updated during counting process o NO, mass of a single part is not updated
Means of operation of ACAI function:
1. Number of parts (on adding) on balance’s weighing pan has to be
greater than has been previously.
2. Number of parts (on adding) on balance’s weighing pan must be less than twice the amount of which was visible on the display before adding parts.
3. Current quantity of parts must be within the ± 0,3 tolerance of the total value.
4. Measurement result has to be stabilized.
Minimal reference mass: 1unit, 2units, 5units, 10units, it is the
minimal mass value of a single part. Unless the condition is met, counting process cannot start.
Result control:
o YES, print and save only those measurements which are included
within the Lo and Hi thresholds.
o NO, all measurements are printed and saved.
Other functions of menu <Settings>:
− Tare mode
− Automatic footer printout
Page 73
- 73 -
− Printout mode / Value release
− Printouts
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.

16.2. Parts counting – quick access keys

Each of working modes features a set of default quick access keys, which are automatically displayed on mode activation. The set of keys can be modified by selecting other quick access keys to the bottom bar of the display. Such process requires appropriate operator’s access level.

16.3. Setting standard (reference) mass by inserting determined part mass

Procedure:
Press < Give mass of 1 part> key, which opens an editing window
<Reference mass> with an on-screen keyboard,
Insert value of a single part and accept it by pressing key, the balance
returns to working mode < Parts counting> with automatic accuracy correction function enabled.
Caution:
In case the single part mass is determined as lower than 0,1 of balance’s reading unit, the balance display’s a message: <Value too low
>.

16.4. Setting standard (reference) mass by weighing

Procedure:
Place a container on balance’s weighing pan and tare its mass,
Press < Set mass of 1 part>, which opens an editing window
<Reference quantity> with an on-screen keyboard,
Insert desired value (number of parts) and accept by pressing key,
which displays a command: <Load: xx parts> (where xx – denotes set
value of parts),
Load requested number of parts on the weighing pan. On stabilization of
measurement result ( pictogram visible on the display) accept the mass by pressing key,
The software automatically recalculates mass of a single part, moves to
< parts counting> mode and displays number of parts loaded on the
weighing pan with unit (pcs).
Caution:
Remember:
Total mass of all parts loaded on the weighing pan must not exceed the maximal
capacity (weighing range) of the balance;
Page 74
- 74 -
Total mass of all parts loaded on the weighing pan must not be lower than value
declared in parameter “Minimal reference mass”. Unless this condition is met, the balance displays a message: <Too low sample mass>;
Mass of a single part must not be lower than 0,1 of balance’s reading unit.
unless this condition is met, the balance displays a message: <Too low part mass >.

16.5. Acquiring part mass from database

A product record in the database of products features set of information which identifies it. One of them is mass, which is used during part counting process.
Procedure:
When in working mode < Parts counting> press < Products database> key, and then select desired product from displayed list.
Insert standard (reference) mass to balance’s memory
Procedure for adding standard (reference) mass of a single part to the database of products:
a) Press < Setup > key followed by pressing < Database > key b) In the Database of Products press < Products > c) Press name of a product and edit data in field no. [5] Mass d) Return to <Parts counting> mode.
If there are no data in the database:
a) Press < Setup > key followed by pressing < Database > key
b) In the Database of Products press < Add > key c) Accept the process of adding a new record in the database d) Fill in field referring to a product, including field no. [5] Mass e) Return to <Parts counting> mode.

16.6. Updating part mass in the database

Determined mass of a single part can be assigned to a product in the database of products. This option is applied while using ACAI (Automatic Accuracy Correction) function in order to determine part mass with high accuracy.
Procedure:
a) Determine mass of a single part
b) Press < Database> key c) Press and hold finger on name of a product, which mass is updated, d) Contextual menu is displayed,
Page 75
- 75 -
e) Select option <Assign standard>, reference mass is saved in a product
record under entry <Mass>.

16.7. Parts counting procedure

The first step in parts counting mode is obtaining data on mass of a single part. Select one of available options:
Give mass value of a single part (see point 16.3) and place parts on balance’s weighing pan, balance displays totalized parts.
Set mass of a single part from a given quantity of parts (see point
16.4.), the balance additionally shows pictogram of the ACAI function < > (if enabled). Place parts on balance’s weighing pan, the balance displays totalized parts.
Acquire mass of a single part from database of products (see point
16.5.) by selecting a desired product record. Place parts on balance’s weighing pan, the balance displays totalized parts.
Caution:
All additional elements (i.e. packaging) have to be tarred before starting the parts counting process.

16.8. Checkweighing function in parts counting mode

Parts counting process can be aided by checkweighing function, i.e. control whether indication is within set thresholds.
Checkweighing requires setting values of two thresholds, described as:
− LOW threshold [Min= … parts]
− HIGH threshold [Max= … parts]
and enabling bargraph (set its attribute to YES), which displays below relation:
CURRRENT NO. OF PARTS / CHECKWEIGHING THRESHOLDS.
Defining the values of Low and High thresholds is carried out in the Database
of Products while editing Product entry, or using a quick access key < Defining thresholds>.
Page 76
- 76 -
Procedure
Press grey coloured workspace and press option <Information>
Set Bargraph option to YES, return to parts counting mode
Press grey coloured workspace and press option <Keys>
Assign <Checkweighing thresholds> option to one of the quick access keys
displayed in the bottom bar
Return to parts counting mode
Press < Checkweighing thresholds> key and insert values for LOW and
HIGH thresholds, return to parts counting mode
Under the measurement result there is bargraph displayed. Colour of the
bargraph corresponds to the status of parts counting process.
o Yellow bargraph’s colour: current number of parts below set value of
Low threshold
o Green bargraph’s colour: current number of parts within set value of
thresholds
o Red bargraph’s colour: current number of parts above set value of High
threshold

16.9. Dosing function in parts counting mode

Parts counting process can be aided by dosing function, i.e. control whether indication does not exceed set target value.
Dosing requires setting a value, which is the target, for instance 100 parts, and percent tolerance from the target value. The target value is displayed as a bracket on a bargraph.
Defining the value of target mass is carried out using a quick access key
< Target Value>.
Procedure
Press grey coloured workspace and press option <Information>
Set Bargraph option to YES, return to parts counting mode
Press grey coloured workspace and press option <Keys>
Assign <Target Value> option to one of the quick access keys displayed in
the bottom bar
Return to parts counting mode
Page 77
- 77 -
Press < Target Value > key and insert number of parts recognized as
the target value,
If tolerance applies, set its value (range between 0 – 100%)
Under the measurement result there is bargraph displayed, containing:
o Current number of parts on the weighing pan (marked with colours –
see point 16.7.)
o The value of Target Mass (black marker)
Caution: The checkweighing and dosing functions can operate simultaneously in the parts counting mode. In such case the dosing tolerance is controlled by Lo and Hi thresholds of the checkweighing function.
Page 78
- 78 -

17. CHECKWEIGHING

Working mode< Checkweighing > enables controlling sample mass
using two threshold (Low and High). Usually it is accepted, that mass indication is correct if it is included within the thresholds’ values.
Working mode activating procedure
While in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
Select < Checkweighing> mode, the software automatically returns to
the main window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Supplementary unit o Low threshold o High threshold o Difference – denoting the “distance” of current measurement result from
the center of Lo-Hi section
On selecting the checkweighing mode the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Set tare – field for setting numeric value of tare
6. Checkweighing thresholds – for declaring the values of Low and High threshold
Page 79
- 79 -

17.1. Use of checkweighing thresholds

Use of checkweighing thresholds can be carried out by:
Selection of Product < > for which Low and High thresholds have been
already set
Inserting the numeric value of thresholds < >, in this case the
thresholds are not referred to any product from database
PROCEDURE 1 – Selection of product from Database of Products
Press < Database of Products> quick access key
Using list of products select one to be weighed
Automatically, in the grey workspace the values of thresholds are
displayed
Under the measurement result there is bargraph displayed. Colour of the
bargraph corresponds to the current status of mass:
o Yellow bargraph’s colour: mass value below Low threshold o Green bargraph’s colour: mass value within set value of thresholds o Red bargraph’s colour: mass above set value of High threshold
PROCEDURE 2 – manual inserting checkweighing thresholds
Press < Checkweighing thresholds> quick access key
Press Low Threshold key and determine its value
Accept set value by pressing < > key
Press High Threshold key and determine its value
Accept set value by pressing < > key
Caution:
The value of High threshold has to be bigger than the value of Low threshold.

17.2. Additional settings of checkweighing mode

The additional settings enable adjusting the working mode to user’s needs and requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and Profile,
3. Press <Settings> key,
4. The display indicates functions related to checkweighing:
− Result control
− Tare mode
− Automatic footer printout
− Printout mode / Value release
− Printout
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.
Page 80
- 80 -

18. DOSING

Working mode< Dosing> enables carrying out the sampling process, until
obtaining a pre-defined target mass.
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Dosing > mode, the software automatically returns to the main
window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Tare o Gross o Target value o Tolerance – percent [%] value related to the target value o Product
On selecting the dosing mode, the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Set tare – field for setting numeric value of tare
6. Target value – for declaring the Target value of dosing process
Page 81
- 81 -

18.1. Use of database of products in dosing process

When weighing, it is possible to use the value of target mass assigned to a product in the database of products or temporarily determine custom target values of mass. In the database of products, target mass of a product is its mass field.
PROCEDURE 1 – Selection of product from Database of Products
Press < Database of Products> quick access key
Using list of products select one to be weighed
Automatically, in the grey workspace the target value and the tolerance
value are displayed
The display indicates the negative value of the target value
Under the measurement result there is bargraph displayed. Colour of the
bargraph corresponds to the current status of mass :
o Yellow bargraph’s colour: mass value below the Target Value –
Tolerance
o Green bargraph’s colour: mass value within the tolerance field: Target
Value +/- Tolerance
o Red bargraph’s colour: mass above the Target Value + Tolerance
PROCEDURE 2 – manual inserting the value of target mass
Press < Target value > quick access key
In an opened window give the target value and tolerance
Return to weighing
Page 82
- 82 -
Caution:
If the Target Value is acquired from the Database of Products, then fields Target Value and Tolerance contain data referring to selected product. The data can be edited and modified.

18.2. Additional settings of dosing mode

The additional settings enable adjusting the working mode to use’s needs and requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and Profile,
3. Press <Settings> key,
4. The display indicates functions related to dosing process: Result control, Tare mode, Automatic footer printout, Printout mode / Value release, Printouts
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.
Page 83
- 83 -

19. PERCENT SETUP

Working mode< Percent setup> enables comparing weighed load to a
reference mass (standard). The process is expressed in percent [%].
Additionally, the percent setup process can be aided by DOSING and CHECKWEIGHING processes. The supplementary modes and a bargraph are not enabled automatically.
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Percent setup> mode, the software automatically returns to the
main window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Supplementary unit o Gross o Reference mass (standard) o Low threshold – percent [%] value of the Reference Mass (Standard) o High threshold – percent [%] value of the Reference Mass (Standard)
On selecting the percent setup mode the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Percent setup: set reference mass (standard)
6. Percent setup: set as 100 %
Page 84
- 84 -

19.1. Comparison of sample and (reference) mas s standard

Comparison of samples with a mass standard can be carried out through:
Giving mass of a standard, using < Give reference mass> soft key;
Accepting current mass loaded on the weighing pan as a reference
mass, using < Set as 100%> soft key;
Selecting a product from database of products, where parameter mass
of a product entry is defined, using < Database of Products> soft key.
PROCEDURE 1 – manual declaration of reference mass
Press < Give reference mass > soft key
In an opened window type in value of reference mass and accept it by
pressing < > key
All weighed products are compared with the reference mass, and the
display indicates difference between weighed objects in [%].
PROCEDURE 2 – accepting currently loaded mass as reference (standard)
Place a sample on balance’s weighing pan
On stabilization of measurement result press < Set as 100%> soft
key
The display indicates 100.000%, mass is accepted as reference, and
automatically saved in (reference) mass standard field.
Unload the sample from balance’s weighing pan
All samples weighed afterwards are compared with the reference mass,
and the display indicates the difference, expressed in [%], for each weighed sample in relation to the reference mass.
PROCEDURE 3 – selecting a product from Database of Products
Press < Database of Products> soft key, select a product entry to be
weighed from a list of displayed products
On selecting, the grey coloured workspace automatically updates its
content with data on Reference Mass
Mass of the weighed product is automatically inserted in field: reference
mass, which is activated by pressing < > soft key
The display indicates 0.00 % (if the weighing pan is unloaded)
All samples weighed afterwards are compared with the reference mass,
and the display indicates the difference, expressed in [%], for each weighed sample in relation to the reference mass.
Page 85
- 85 -

19.2. Checkweighing, dosing functions in percent setup mode

Percent setup mode can be aided by checkweighing and dosing functions. Access to the supplementary functions is given by determining corresponding soft keys in the bottom bar of the display.
Values of the supplementary functions have to be determined as percentage.
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and Profile,
3. Press <Settings> key,
4. The display indicates a list of soft keys, functions and proximity sensors,
5. Press one of available options and assign a corresponding function.
CHECKWEIGHING
Uses two checkweighing thresholds defined in [%] for controlling sample mass.
PROCEDURE:
1. Press Checkweighing Thresholds < > soft key,
2. Press Low Threshold key and insert its value in [%],
3. Accept by pressing < > key,
4. Press High Threshold key and insert its value in [%],
5. Accept by pressing < > key.
Caution:
The value of High Threshold has to be set higher than the value of Low Threshold.
DOSING
Uses the Target Value expressed in [%], which has to be reached during weighing (pouring) a sample. The target value also features a [+/-] tolerance, determining the area within which the measurement is recognized as correct.
PROCEDURE:
1. Press < Target Value > key,
2. Insert the target value expressed in [%],
Page 86
- 86 -
3. If tolerance applies, set its value,
4. Accept by pressing < > key,
5. Press High Threshold key and insert its value in [%],
6. Accept by pressing < > key

19.3. Interpreting the function by use of a bargraph

The dosing and checkweighing functions are aided by graphic indication, in a form of a bargraph. Below see an example of simultaneous operation of the two functions.
a) checkweighing thresholds< > are set as Low Threshold = 90%,
High Threshold = 110%
b) Target value = 105% ; tolerance = 5% < >
c) Reference mass = 19.986 g < >

19.4. Additional settings of Percent Setup mode

The additional settings enable adjusting the working mode to user’s needs and requirements. Functions related to checkweighing: result control, tare mode, automatic footer printout, printout mode / value release, printouts.
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.
Page 87
- 87 -

20. DENSITY

Working mode< Density > features three separate modules. The first
one is used to determine density of solids, the second is for the density of liquids, and the third one determines density of air. The third module is available in balances AS 3Y, XA 3Y and MYA 3Y series. Carrying out the density procedures requires installing an optional density kit (additional equipment) appropriate to a model of used balance.
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Density> mode, the software automatically returns to the
main window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Procedure Solids o Weighing 1 o Weighing 2 o Standard liquid Water o Temperature 22 o Density of standard liquid 0.9978 g/cm
On selecting the density mode the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Density of solid
6. Density of liquid
o
C
3
Page 88
- 88 -

20.1. Set for density estimation of liquids and solid objects

1
Beaker basis
8
Hook
Weighing pan with stand
Top weighing pan of density determining kit for solids
3
Plunger
10
Flexible connector
Bottom weighing pan of density determining kit for solids
5
Thermometer handle
12
Additional weight
Additional stand for set of pans or a plunger
Additional set of pans for
density is lower than density of water
For balances XA i AS series
The set comprises:
2
4 Beaker 11
6 Thermometer 13
Plunger flexible
7
connector
For balances PS series
9
14
determining density of solids, which
Page 89
- 89 -
The set comprises:
1
Weighing pan with stand
7
Thermometer
Bottom weighing pan of
solids
3
Flexible connector
9
Plunger
Top weighing pan of density determining kit for solids
Additional stand for set of pans or a plunger
Additional set of pans for
density of water
XA Balances
AS Balances
PS Balances
2
density determining kit for
8 Thermometer handle
4
10 Beaker basis
5 Beaker 11
6 Hook 12
determining density of solids, which density is lower than
CAUTION:
− Particular elements shall be stored in a respective box.
Weighing pans and plunger shall not be put down onto the weighing bench,
it may result in damage of particular parts. If either weighing pan or plunger is not to be used, it should be placed back on a hook.

20.2. Assembly

CAUTION:
If upon assembly the –NULL- message is shown than the set shall be loaded with weights (12), the device shall be zeroed or tared. Thus prepared device may be used for density estimation.
Page 90
- 90 -

20.3. Determining density of solid object

o
BA
A
ρρ
=
Before starting the procedure, it is necessary to determine parameters of the process, such as:
Type of liquid
o Distilled water o Ethanol o Other liquid with determined density
Liquid temperature (required if used liquid is distilled water or ethanol)
Liquid density It is set automatically if liquid type Water or Ethanol is used and on inserting liquid temperature. If used liquid is <Other> then it has to be inserted manually.
The density of solids is calculated according to a formula:
ρ - density of a sample A - sample’s mass measured in the air B - sample mass measured in liquid ρ o - liquid density
PROCEDURE
1. Assembly the density determination kit
2. Should the density value be saved in a product record of the database of products, set parameter <Assign density to product> in <Settings> and activate product using quick access key <Product>,
3. Press < Density of solids > soft key
4. A window with menu is opened, set appropriate values for fields: Standard liquid, Temperature, Density of standard liquid,
5. Press < START > key
6. The balance is ready to start density determining process of solids.
7. Load a sample on the TOP pan of the density kit, and on stabilization of
measurement result press < > key,
8. Load the sample on the BOTTOM pan of the density kit which is immersed
in liquid, and on stabilization of measurement result press < > key,
9. The display indicates result of density determining process,
10. Press < > key to finish the procedure.
Caution:
Pressing < > key starts another density determining process with the same settings.
Page 91
- 91 -

20.4. Determining density of liquids

d
V
BA
+
=
ρ
It is determining mass of a plunger if weighed in the air and in tested liquid. The density of tested liquid is calculated according to a formula:
ρ - density of liquid A - plunger mass measured in the air B - plunger mass measured in tested liquid V - plunger’s volume d - air density ( max 0,001 g/cm3 )
Before starting the procedure, assembly the density determination kit, and insert to balance’s memory volume of the plunger.
Press < Density of liquid > key
The display opens a menu, press < Plunger’s volume> key and insert
the numeric value of volume specified on plunger’s hanger,
The balance is ready to start density determining process of liquid.
PROCEDURE
1. Assembly the density determination kit
2. Should the density value be saved in a product record of the database of products, set parameter <Assign density to product> in <Settings> and activate product using quick access key <Product>,
3. Press < START > key,
4. Follow displayed commands,
5. Carry out measurement in the air, and on stabilization of measurement
result press < > key,
6. Carry out measurement in tested liquid, and on stabilization of
measurement result press < > key,
7. The display indicates result of determined liquid density,
8. Press < > key to finish the procedure.
Caution:
Pressing < > key starts another density determining process with the same settings
Page 92
- 92 -

20.5. Density of air

Density of air < > is a piece of information required to calculate correction of measurement result related to air BUOYANCY compensation. Another data required in this process is the density of weighed sample. The function of air density determining is enabled only in balances with reading unit lower than d=1mg. Determining density of air requires using a dedicated set of mass standards (additional equipment), specific for a balance model.
PROCEDURE:
1. Press < Air density > quick access key,
2. The display indicates options related to this mode
3. Press field with mass of steel standard and insert its value as provided in its calibration certificate,
4. Press field with mass of aluminium standard and insert its value as
provided in its calibration certificate,
5. Press field with density of steel mass standard and insert the density
value,
6. Repeat process for the aluminium mass standard,
7. Press START field – is ready to start density determining process of the
air.
8. Load steel mass standard on balance’s weighing pan and on stabilization
of measurement result accept it by pressing < > key,
Page 93
- 93 -
9. Load aluminium mass standard on balance’s weighing pan and on
stabilization of measurement result accept it by pressing < > key,
10. The balance automatically calculates the density of air, and indicates its
value on the display,
11. Press < > key to finish the procedure.
The value of determined air density is automatically assigned to field <Air density> in menu <Settings/ Air buoyancy compensation> in working mode: weighing.

20.6. Additional settings of density mode

The additional settings enable adjusting the working mode to user’s needs and requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and Profile,
3. Press <Settings> key,
4. The display indicates functions related to density determining process, such as:
o Request sample number
o Measuring unit: the balance enables choosing between the
o Assign density to a product: YES/NO – if option <YES> is
Available settings:
NO – data on sample number is not required; applicable in case of
a single measurement
YES – before each measurement a window requesting inserting
sample number is automatically displayed; applicable in case of a series of measurements
following units: [g/cm3], [kg/m3], [g/l]. Selected unit is activated for all density determining options and also in final printout.
selected, the software automatically assigns the value of determined density of solid or liquid to field <Density> in a
Page 94
- 94 -
product record of the database of products. To use this function,
Balance ID
before starting the density determining process, using database of product select a product for which the density procedure is carried out (if density has already been determined for this product, it is replaced with new value),
o Tare mode, o Printout mode / Value release, o Printout
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.

20.7. Printouts

Option Printouts enables setting the contents of printouts in a standard printout and also determining non-standard printouts.
Standard printout
comprises four internal blocks, each featuring different variables. Each variable can be attributed with parameter YES – i.e. present on a
printout, or NO – i.e. not printed.
Settings for header printout, weighing printout, and footer printout are specified in point. 15.5; below description refers to setting a printout template from density mode. A user can design content of a report for each of carried out procedures. On pressing <Density printout template> option a window is opened enabling setting the content of each report.
Content of each of reports:
o SOLID
Working mode
Procedure
Sample no.
User
Balance type
Page 95
- 95 -
Date
Non-standard printout
Mass of aluminium standard
Time
Standard liquid
Temperature
Density of standard liquid
Weighing 1
Weighing 2
Density
Volume
Product
Warehouse
Client
Empty line
Dashes
Signature
Non-standard printout
o LIQUID
Working mode
Procedure
Sample no.
User
Balance type
Balance ID
Date
Time
Plunger volume
Temperature
Weighing 1
Weighing 2
Density
Product
Warehouse
Client
Empty line
Dashes
Signature
o AIR
Working mode
Procedure
Sample no.
User
Balance type
Balance ID
Date
Time
Mass of steel standard
Page 96
- 96 -
Density of steel standard
Density of aluminium standard
Non-standard printout
Weighing 1
Weighing 2
Density
Product
Warehouse
Client
Empty line
Dashes
Signature

20.8. Report on completed density determination processes

On completing any type of density determining processes, i.e. solid, liquid
or air, a report is generated. The report is saved in the database < Density reports > . The files are named by their date and time (hour) of
density determining process.
An example of report on density determination of solid.
--------------- Density ----------------
---------------- Solid ---------------­ User Admin Balance ID 400015 Date 2011.10.07 Time 10:08:09 Standard liquid Water Temperature 22 °C Density of standard liquid 0.9978 g/cm3 Weighing 1 10.526 g Weighing 2 2.586 g Density 1.322776 g/cm3
------------------------------------------------- Signature
Page 97
- 97 -
Use of data contained in the header and footer.
Should data from header and/or footer be used, follow below procedure:
o Press < Print header> soft key before starting density determination
procedure,
o Carry out density determination procedure
A report is printed automatically on completing the second
measurement (Weighing 2),
When density determination result is displayed, it is possible to print
it for numerous times by multiple pressing of < PRINT > key,
o Press < Print footer > soft key after completing the density
determination procedure.
Caution:
Remember to select data to be printer in the header, footer and report. For details see section “PRINTOUTS”.

21. ANIMAL WEIGHING

Working mode< Animal weighing> enables reliable determining mass of
objects in motion. In principle, this type of objects generate unstable measurement, thus it requires using a different filtering method of measurement signal.
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Animal weighing> mode, the software automatically returns to
the main window and displays pictogram in the upper bar,
The grey coloured workspace contains the following data:
o Averaging time 5 o Threshold 10
Page 98
- 98 -
On selecting the animal weighing mode, the display contains the following quick access keys in the bottom bar:
1. Setup – access to balance’s menu
2. Print header – print of data declared in the header
3. Print footer – print of data declared in the footer
4. Database of products – selection of products from corresponding database
5. Set tare
6. Animal weighing

21.1. Setting of animal weighing mode

Depending on required mass analysis of weighed object, the balance enables setting internal criteria of the function.
PROCEDURE:
1. press < Animal weighing > key,
2. the display indicates option referring to this working mode, such as:
Averaging time
Amount of time during which measurement records of weighed sample are analysed. Obtained data is used to determine measurement result.
Automatic operation
Determines start criterion for measurements: whether they are initiated manually /on pressing a key/ or automatically. A measurement starts automatically if indication on balance’s display exceed the value of set threshold. The following measurement can start on unloading the weighed object from balance’s pan (the indication has to return below the value set in threshold), loading a new object on balance’s pan and exceeding the threshold’s value by the indication.
Page 99
- 99 -
Threshold
It is a value expressed in mass measuring units. in order to start measurement, the indication has to exceed value set in the threshold.
3. set mode’s operating parameters and return to weighing,
4. load weighed object on balance’s weighing pan and press < > key,
5. on completing the measurement, the display indicates and “locks” the
measurement result,
6. the following measurement is carried out on pressing < > key:
- in case of manual operation mode, press < > key,
- in case of automatic operation mode, unload weighed object from the weighing pan and load the following object to be weighed.

21.2. Additional settings of animal weighing mode

The additional settings enable adjusting the working mode to user’s needs and requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. Press menu <Settings>,
3. The display indicates functions related to checkweighing, such as:
Result control
Tare mode
Automatic footer printout
Printout mode / Value release
Printout
Means of using the above functions are given in point 10.8. “Additional parameters on weighing process”.
Page 100
- 100 -

22. FORMULATION

Means of verifying ingredient’s
mass TOLERANCE
Means of verifying ingredient’s
mass THRESHOLDS
Working mode< Formulation> enables preparing mixtures out of several
ingredients. The process is conducted automatically. Balance user, while preparing mixtures, can:
use database of formulas, where formula records are stored. In such
case, balance’s software assists in weighing ingredients of a mixture by displaying corresponding commands or information in the grey coloured workspace;
prepare a mixture without using database of formulas. Then, a user
controls weighing of an ingredient, weighing sequence and quantity.
Should database of formulas be used, first a formulation has to be created in the database, and then it is recalled (activated) for use. Creating a formulation is carried out only from the level of Database of Formulas, and it is described later in this section.
Working mode activating procedure
while in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of available working modes,
select < Formulation> mode, the software automatically returns to the
main window and displays pictogram in the upper bar.
The grey coloured workspace contains the following data:
o Formulation o Ingredient o Ingredient no. o Ingredient mass o Target value (mass) o Sum
Caution: In the grey coloured workspace a bargraph can be enabled (displayed). It is a graphic
interpretation of correctly weighed ingredient’s mass with +/- tolerance, acc or di ng to s et tolerance thresholds.
Loading...