RADWAG AS 60/220.R2, XA 82/220.R2, PS 210.R2, PS 360.R, PS 600.R2 User Manual

...
Page 1
Number of instruction: IMMU-03-07-05-14-ENG
LCD RADWAG BALANCES
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS
RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone +048 48 38 48 800, fax +48 48 385 00 10 Export department +0-48 48 366 80 06
www.radwag.com
Page 2
- 2 -
MAY 2014
Page 3
- 3 -
TABLE OF CONTENTS
1. TECHNICAL DATA .................................................................................7
1.1. Connection ports....................................................................................... 11
1.2. PS R2.H series ......................................................................................... 11
2. BASIC INFORMATION ........................................................................ 14
2.1. Intended use ............................................................................................. 14
2.2. Inappropriate use ...................................................................................... 14
2.3. Warranty ................................................................................................... 14
2.4. Monitoring the metrological parameters of the instrument ........................ 15
2.5. Data included in this user manual ............................................................. 15
2.6. Staff training ............................................................................................. 15
3. TRANSPORT AND STORAGE ........................................................... 15
3.1. Delivery check .......................................................................................... 15
3.2. Packaging ................................................................................................. 15
4. UNPACKING, ASSEMBLING AND START-UP .................................. 16
4.1. Assembling and place of use .................................................................... 16
4.2. Unpacking ................................................................................................. 17
4.3. Levelling ................................................................................................... 22
4.4. Cleaning ................................................................................................... 22
4.5. Plugging to mains ..................................................................................... 27
4.6. Connecting additional equipment.............................................................. 27
5. BASIC FUNCTIONS ............................................................................ 28
6. KEYBOARD ......................................................................................... 29
7. START-UP ........................................................................................... 30
7.1. Connecting peripheral devices ................................................................. 30
7.2. Balance temperature stabilisation period .................................................. 30
8. USER MENU ....................................................................................... 31
8.1. Moving through user menu ....................................................................... 31
9. LOGGING IN ....................................................................................... 34
10. WEIGHING .......................................................................................... 36
10.1. Required weighing procedures ................................................................. 36
10.2. Zeroing ..................................................................................................... 38
10.3. Taring ....................................................................................................... 38
10.4. Settings for mode <WEIGHING> .............................................................. 40
10.4.1. Setting level of filtering ........................................................................... 41
10.4.2. Value release ......................................................................................... 41
10.4.3. Autozero function ................................................................................... 42
10.4.4 Displaying the last digit .......................................................................... 42
10.4.5 Balance ambient conditions ................................................................... 43
10.5 AUTOTARE .............................................................................................. 43
10.6 Printing mode ........................................................................................... 44
10.7 Temporary measuring unit ........................................................................ 46
10.8 Start-up measuring unit ............................................................................ 46
10.9 Availability of measuring units .................................................................. 47
10.10 User unit ................................................................................................... 48
10.11 Information ................................................................................................ 49
10.12 Non-standard information ......................................................................... 50
Page 4
- 4 -
10.13 F button shortcuts ..................................................................................... 51
10.14 Dual range balance (NPS 200/2000.R2) .................................................. 52
11 OTHER PARAMETERS ....................................................................... 53
11.1. Available menu languages ........................................................................ 53
11.4 Access levels ............................................................................................ 53
11.5 ‘Beep’ sound – reaction on pressing function button ................................ 53
11.6 Backlight and adjusting display brightness ............................................... 54
11.7 Date .......................................................................................................... 54
11.8 Time .......................................................................................................... 54
11.9 Date format ............................................................................................... 54
11.10 Time format ............................................................................................... 55
11.11 GLP autotest ............................................................................................. 55
12 INFORMATION ABOUT THE BALANCE ............................................. 57
13 BALANCE ADJUSTMENT ................................................................... 58
13.1. Internal adjustment .................................................................................... 59
13.2. Menu adjustment settings ......................................................................... 60
13.3. Manual adjustment ................................................................................... 63
13.3.1. Internal adjustments ................................................................... 63
13.3.2. External adjustment .................................................................... 64
13.3.3. User adjustment ......................................................................... 64
13.4. Adjustment report printout......................................................................... 65
14. DETERMINING CONTENT OF A PRINOUT ....................................... 66
14.1. Adjustment report ..................................................................................... 66
14.2. Header printout ......................................................................................... 68
14.3. Measurement result printout – GLP PRINTOUT ....................................... 70
14.4. Footer printout .......................................................................................... 72
14.5. Non-standard printouts ............................................................................. 74
14.5.1. Inserting texts ............................................................................. 74
14.6. Variables ................................................................................................... 77
15. DATABASE .......................................................................................... 78
15.1. Users ........................................................................................................ 78
15.2. Products.................................................................................................... 81
15.3. Tares ........................................................................................................ 82
15.4. Weighments .............................................................................................. 83
15.5. ALIBI ......................................................................................................... 86
16. EXPORT AND IMPORT OF DATABASE ............................................. 88
16.1. Export database ........................................................................................ 89
16.2. Import database ........................................................................................ 90
16.3. Measurement data print-out ...................................................................... 90
17. BALANCE WORKING MODES ............................................................ 92
17.1. Setup for availability of working modes ..................................................... 93
17.2. Counting parts of the same mass ............................................................. 94
17.2.1. Setup for mode COUNTING PIECES ......................................... 95
17.2.2. Setting the reference mass by determining from the sample of
known quantity............................................................................ 96
17.2.3. Setting the reference mass by entering its mass ........................ 98
17.2.4. Returning to the weighing mode ................................................. 99
17.3. Checkweighing ....................................................................................... 100
Page 5
- 5 -
17.3.1. Declaration of mass thresholds ................................................ 101
17.4. Dosing .................................................................................................... 103
17.4.1. Setting a target weight by entering values ................................ 104
17.4.2. Returning to the weighing mode ............................................... 106
17.5. Deviations % against the mass of the standard ...................................... 106
17.5.1. Setup for reference mass determined by weighing .................. 107
17.5.2. Setup for reference mass by typing its mass ............................ 108
17.5.3. Returning to the weighing mode ............................................... 109
17.6. Animal weighing...................................................................................... 109
17.6.1. Additional setting for animal weighing mode ............................ 109
17.6.2. Method of operation to manually start the process ................... 111
17.6.3. Method of operation to automatically start the process ............ 113
17.7. Density of solids...................................................................................... 114
17.7.1. Density determination of solids ................................................. 117
17.8. Density of liquid ...................................................................................... 121
17.8.1. Density of liquids determination ................................................ 121
17.9. Statistics ................................................................................................. 124
17.8.2. Means of operation ................................................................... 125
17.8.3. Deleting statistics ..................................................................... 128
17.8.4. Returning to the weighing mode ............................................... 129
17.10. Totalising .............................................................................................. 129
17.8.5. Means of operation ................................................................... 130
17.8.6. Returning to the weighing mode ............................................... 133
17.11. Peak hold ............................................................................................. 133
17.8.7. Means of operation ................................................................... 134
17.8.8. Returning to the weighing mode ............................................... 135
17.12. Pipettes calibration ............................................................................... 136
17.8.9. Additional settings of pipettes calibration mode ........................ 136
17.8.10. Means of operation ................................................................... 137
17.8.11. Returning to the weighing mode ............................................... 141
18. COMMUNICATION ............................................................................ 142
18.1. RS 232 ports settings ............................................................................. 142
18.2. WIFI port settings.................................................................................... 143
19. PERIPHERALS .................................................................................. 145
19.1. Computer ................................................................................................ 145
19.1.1. Port for computer connection ................................................... 145
19.1.2. Continuous transmission .......................................................... 146
19.1.3. Cooperation with E2R .............................................................. 146
19.2. Printer ..................................................................................................... 147
19.3. Bar code reader ...................................................................................... 149
19.4. Additional display .................................................................................... 149
19.5. External buttons ...................................................................................... 150
20. WORKING WITH EXTERNAL DEVICES SUCH AS PRINTER OR
COMPUTER ................................................................................ 152
20.1. Cable pinout diagram .............................................................................. 153
20.2. Format of data setting ............................................................................. 154
20.3. Format of data sent on pressing of PRINT button................................... 154
20.3.1. Format of data sent as response for commands generated from a
computer .................................................................................. 154
Page 6
- 6 -
21. COMMUNICATION PROTOCOL ....................................................... 156
21.1. List of commands .................................................................................... 156
21.2. Responses format for commands sent from computer level ................... 157
21.3. Commands description ........................................................................... 157
22. ERROR MESSAGES ......................................................................... 162
23. UNDER HOOK WEIGHING ............................................................... 163
24. ADDITIONAL EQUIPMENT ............................................................... 164
24.1. Anti-vibration table .................................................................................. 164
24.2. Rack for under hook weighing ................................................................ 164
24.3. Density determination kit (for solids and liquids) ..................................... 164
24.4. Additional display .................................................................................... 164
Page 7
- 7 -
1. TECHNICAL DATA
Balance XA series:
XA 52.R2 XA 82/220.R2
Max capacity 52 g 82/220 g
Min capacity 1 mg 1 mg
Readability 0,01 mg 0,01/0,1 mg
Tare range
-52 g -220 g
Repeatability
0,02 mg 0,02 /0.08 mg
Linearity
± 0,06 mg ± 0,06/0,2 mg
Eccentric load deviation
0,06mg 0,2mg
Pan size
Ø85mm
Dimensions:
Page 8
- 8 -
Balance AS series:
AS
60/220.R2
AS
160.R2
AS
220.R2
AS
310.R2
Max capacity 60/220g 160g 220g 310g
Min capacity 1mg 10 mg 10 mg 10 mg
Readability 0.01/0.1mg 0.1mg 0.1mg 0.1mg
Tare range -220g -160 g -220 g -310 g
Pan size Φ 85 mm
Sensitivity drift 1 ppm/°C in temperature +10 ° - +40 °C
Working temperature
+10 ° - +40 °C
Power supply 12 ÷ 16 V DC / 2.1 A
Dimensions:
Page 9
- 9 -
Balance PS series with weig hing pan 128 x 128 mm:
Dimensions:
Type balance
PS
200/2000.R2
PS
210.R2
PS
360.R2
PS
600.R2
PS
750.R2
PS
1000.R2
Max capacity
200/2000g 210g 360g 600g 750g 1000g
Tare range
-2000g -210g -360g -600g -750g -1000g
Min capacity
20mg 20mg 20mg 20mg 20mg 20mg
Readability
1/10mg 1mg 1mg 1mg 1mg 1mg
Working temperature
+10°C - +40°C
Power supply
12 ÷ 16 V DC / 2,1 A
Sensitivity drift
2 ppm/°C in temperature +10°C -+ 40°C
Pan size
128x128 mm
Page 10
- 10 -
Balance PS series with weig hing pan 195 x 195 mm:
Type balance
PS
1200.R2
PS
2100.R2
PS
3500.R2
PS
4500.R2
PS
6000.R2
Max capacity 1200g 2100g 3500g 4500g 6000g
Tare range -1200g -2100g -3500g -4500g -6000g
Min capacity 500mg 500mg 500mg 500mg 500mg
Readability 10mg 10mg 10mg 10mg 10mg
Working temperature
+10°C - +40°C
Power supply 12 ÷ 16 V DC / 2,1 A
Sensitivity drift 2 ppm/°C in temperature +10°C - +40°C
Pan size 195x195mm
Dimensions:
Page 11
- 11 -
Balance APP series:
APP
10.R2
APP
35.R2
Max capacity 10 kg 35 kg
Tare range -10 kg -35 kg
Min capacity 2 g 2 g
Readability 0.01 g 0.1 g
Working temperature
+10 °C - +40 °C
Power supply 12 ÷ 16 V DC / 2,1 A
1.1. Connection ports
1
- power port
2
- COM 2 port (e.g. additional display)
3
- COM 1 port (e.g. printer)
4
- USB 2 port (e.g. computer)
5
- USB 1 port (e.g. computer keyboard)
1.2. PS R2.H series
PS R2.H series redefine the level of sta ndard for precision balances. Not only do they share all the features of R series balances, but ca n a lso wo rk in adverse operating conditions (condensed dust, drops of water falling down at different angles typical for IP 54).
PS.R.H balances are offered with round pans of two possible sizes: ø115mm and ø170mm. Balances with pans of a smaller size feat ure draft shield as well.
Page 12
- 12 -
Additional asset of PS R2.H balances are their interfaces build-in a hermetic closed housing which is separated from the balance. The interfaces include 2×RS 232, USB type A, USB type B, and WiFi optionally.
Balance housing is made of plastic, whereas its pan of stainless steel.
PS R2.H series metrological parameters are identical like those of standard design balances.
Balance with pan Φ 115 mm
Balance with pan Φ 170 mm
Page 13
- 13 -
Balance assembly
Upon unpacking the balance remove any protective elements and assembly the
balance:
-
assembly the metal pan
(2) on the balance base, mounted onto the balance
- assemlby draft shield (1), only for balances with pan Φ 115 mm
- plug the power adapter interface into the port of
interfaces housing, next connect the power adapter to mains. The balance will
activate automatically.
CAUTION:
While clean ing it is advis able to fol low below pre cautions. Ad ditionally it is a must to disconnect the device from mains and unplug all the peripherals (printers, computer etc.) rememberin g at the same time to protect ports by means of stoppers. Only thus prepared device can be cleaned. Upon completion of the cleaning process the device may be connected to mains for futher operation.
Page 14
- 14 -
2. BASIC INFORMATION
2.1. Intended use
The purchased balanced serves to determine the weighing value of loads in laboratory environment. It is intended for application as a non-automatic weighing instrument only, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan. Weighing result should be read only after stable reading has been obtained.
2.2. Inappropriate use
Do not use the balance as a dynamic weighing instrument. Even if small quantities of weighed material are added or removed from the weighing pan of the instrument, the reading should be taken only after stabilisation of the measurement results. Do not place any magnetic materials on the weighing pan, as this can cause damage to the measuring system of the instrument. Be sure to avoid impact shock and overloading the balance in excess of the prescribed maximum measuring range (max capacity), minus any possible tare weight that has been applied. Never use the balance in an environment where explosion is possible. This balance has not been adjusted for operation in explosive areas. There must not be any modification made to the balance.
2.3. Warranty
Warranty is invalid for the following:
non- observation of the guidelines of this user manual,
use of the balance other than specified in this manual,
alteration to or opening of the device,
mechanical damage and damage caused by media, water, wear and
tear,
inappropriate assembling or defects of electric installation,
overloading of the measuring instrument.
Page 15
- 15 -
2.4. Monitoring the metrological parameters of the instrument
Metrological characteristics of the balance require periodical inspection to be carried out by its user. Inspection frequency is dependent on ambient conditions in which the balance is used, types of performed processes and accepted quality management system in organisation.
2.5. Data included in this user manual
Please read the user manual carefully before assembling and start-up, even if the user is experienced with this type of weighing instruments.
2.6. Staff training
This balance should only be operated and maintained by personnel who is trained and experienced in using this type of balances.
3. TRANSPORT AND STORAGE
3.1. Delivery check
Please check the packaging immediately upon delivery and the device during unpacking for any visible signs of external damage.
3.2. Packaging
Please retain all parts of the original packaging should the balance be transported in the future. Only the original packaging should be used for shipping the balance. Before packing, disconnect all attached cables and remove any loose/movable parts (weighing pan, anti-draft shield, etc.). Please place balance and its components in their original packaging, and protect them against damage during transport.
Page 16
- 16 -
4. UNPACKING, ASSEMBLING AND START-UP
4.1. Assembling and place of use
The balance should be stored and used in locations free of vibrations and shakes, free of air movement and dust,
ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C,
ambient relative humidity should not exceed 80%,
during balance operation, ambient temperature in the weighing room should
not change more than 3°C within one hour,
the balance should be located on a stable wall console desk or a stable working table which is not affected by vibrations and distant from heat sources,
take special precaution when weighing magnetic objects, as part of the balance is a strong magnet. Should such loads be weighed, use under hook weighing option, which removes the weighed load from area influenced by the balance’s magnet. For assembling the hook for under hook weighing
option see the bottom section of balance’s housing,
in order to avoid the influence of static electricity on the measurement process, ground the balance’s housing. The grounding bolt is located at the
back of the balance’s housing.
Page 17
- 17 -
4.2. Unpacking
Cut the sealing tape. Take the balance out of the packaging. Open the box with accessories and take the items necessary for correct balance operation. Upon placing the balance in the desired location assembly its weighing pan and other parts in accordance with the following instruction.
Steps:
- open side doors of the weighing
chamber,
- inside the weighing chamber place
lower shield of the weighing
chamber (5),
- place centering ring (4) upon the
lower shield,
- assembly the weighing pan (1)
- assembly anti-draft shield (1),
- close side doors of the weighing
chamber,
- plug the balance to mains, pin of
power adaptor should be connected
to socket located in the back of
balance housing
AS R.2 balance
Balance with a mounted transport lock.
Page 18
- 18 -
AS balance with d=0.01/0.1 mg AS balance with d=0.1 mg
Page 19
- 19 -
open side doors of the weighing chamber,
remove transport lock (1) – press the lock gently and turn it accordingly to
OPEN/CLOSE instruction, next remove the lock; please retain the lock should the
device be transported in the future,
inside the weighing chamber place lower shield of the weighing chamber (2),
put centering ring [framing facing downwards] (3),
in the centre of centering ring assembly the weighing pan (4),
put anti-draft shield in the chamber (5),
close side doors of the weighing chamber,
plug the balance to mains, pin of power adaptor should be connected to socket in
the back of balance housing.
Balance PS
Remove the adhesive tape. Carefully remove the balance from its packaging. Remove all the elements necessary for correct operation from the accessory box. Carefully remove the plastic and foil transport protective elements and gently place the balance in its intended place of use.
Assemble the weighing pan, and other elements according to below schema:
Assembling a balance with reading interval 10mg:
- remove the adhesive tape
protecting the mass spring on one
of the rubber bumpers (1),
- assemble the weighing pan (2) on
the rubber bumpers (1),
- after assembling the balance
components, check whether the weighing pan is firmly placed on the
rubber bumpers.
Assembling a balance with reading interval 1mg:
- remove the adhesive tape protecting the
mass spring on one of the rub
ber bumpers
(1),
- assemble the weighing pan (2) on the
Page 20
- 20 -
rubber bumpers (1),
- assemble the glass anti-draft shield (3)
- after assembling the balance components,
check whether the weighing pan is firmly
placed on the rubber bumpers.
Balance APP
Remove the adhesive tape. Carefully remove the balance from its packaging. Remove all the elements necessary for correct operation from the accessory box. APP 10.R1 and APP 10.R2 balances are equipped with 2 weighing pans which can be used in turns. It must be remembered that balance activation is only possible with one weighing pan assembled at a time. Attempt of activating the balance with two weighing pans assembled simultaneously will result in display of an “-LH-“ error. In such case one of the pans needs to removed.
Carefully take the balance out of the packaging, remove transport protective elements and gently place the balance onto its workplace.
Remove the weighing pan and remove transport protective elements as shown in the picture above (if these elements are attached).
After the unpacking of the balance unscrew the bolt (1) which blocks the
calibration mechanism.
Page 21
- 21 -
Assemble the weighing pan (for APP 10.R, one of the pans only).
Make sure the weighing pan is placed properly on the shock absorbers:
the weighing pan cannot touch the housing and should be stable.
Place the balance at the workplace.
Levell the balance (see the instruction below).
Using the wrench (3), after adjusting the balance level (see below); tighten
the bolt (2) on the foot (1) to the balance’s footing until the point of resistance. Repeat the activity for each foot of the balance.
After removing the protective elements of the weighing pan and after unblocking the internal calibration mechanism (only R2 series balances), plug the indicator into the port which is on the back of the housing. Next, plug in all the additional devices. Only then can the balance be connected to mains (the power port is located at the back of the balance’s housing).
Page 22
- 22 -
4.3. Levelling
Before plugging to mains, level the balance by turning the adjustable feet in a way that the air bubble of the level is in its central ring. The balance should be placed at the workplace firmly on all of its feet.
4.4. Cleaning
Balance should be cleaned with damp cloth by gentle wiping smudges of dirt.
Cleaning of the weighing pan when assembled may cause damage of the measuring system of the balance. Hence, it is crucial to remove it from the
balance before cleaning.
In order to clean the glass walls of the draft shield of the XA R2 and AS R2 series balance, it is acceptable to remove the walls of the draft shield following the below instruction.
Steps for XA.R2 series balance:
1. Twist off and remove the top glass wall safety frame, and slide the top glass wall out of the slideway
Page 23
- 23 -
2. Remove the back glass wall.
Page 24
- 24 -
3. Remove the side glass walls. The side glass walls are not replaceable, so it is important to remember which one is left and right, and after cleaning install them back properly in the draft shield.
4. CAUTION: do not remove the front glass wall.
5. Remove the weighing pan, the weighing pan safety rings and the bottom metal plates, so as not to damage the balance mechanism durin the cleaning.
Thus prepared draft shield and glass walls can be properly cleaned. All the operations should be done carefully. Pay special CAUTION to the place where the weighing pan was installed: dirt and other small elements might enter the balance construction through this openiing, which might negatively influence the balance parametres.
Upon completion of the cleaning proces reassembly the balance. To do it repeat steps 1-5 in a reverse order. Be especially careful while assembling
side glass walls, they must take the right place. While sliding the glass
back onto its place remember to pull the protection (like before when disma ntling the draft shield).
Page 25
- 25 -
Steps for AS.R2 series balance:
1. Twist off and remove the top glass wall safety frame, and slide the top glass wall out of the slideway.
2. Remove the back glass wall.
3. Remove the side glass walls. The side glass walls are not replaceable, so it is important to remember which one is left and right, and after cleaning install them back properly in the draft shield.
Page 26
- 26 -
4. CAUTION: do not remove the front glass wall.
5. Remove the weighing pan, the weighing pan safety rings and the bottom metal plates, so as not to damage the balance mechanism durin the cleaning.
Thus prepared draft shield and glass walls can be properly cleaned. All the operations should be done carefully. Pay special CAUTION to the place where the weighing pan was installed: dirt and other small elements might enter the balance construction through this openiing, which might negatively influence the balance parametres.
Page 27
- 27 -
4.5. Plugging to mains
Balance can be connected to mains only with a power adapter that comes standard with the balance. Nominal power supply of the power adapter (specified on the data plate of the power adapter) should be compatible to the power from mains.
Plug the balance to mains – the plugging socket is located at the back of balance housing. The test of the display unit takes place after connecting to the power (all the elements and pictograms are backlit for a short time). Next, the name and the program number appears consequently, the indication will reach ZERO level (with reading unit depending on the b alance). During the process of starting the balance, the test of the internal mass adjustment mechanism occurs (single location and elevation of the internal mass adjustment).
If the indication is different from zero, please press
button.
CAUTION: If the balance is “verified”, automatic adjustment/calibration occurs right after it has been switched on.
4.6. Connecting additional equipment
It is inevitable to plug the computer out of the main before connecting additional equipment or changing it (printer, PC computer, additional display, computer keypad, etc.). Connect only the equipment recommended by the balance manufacturer. When the devices are connected, plug the balance to the main.
Page 28
- 28 -
5. BASIC FUNCTIONS
functions related to measuring units
- weighing in grams [g],
- weighing in milligrams [mg] (in balances with d0.001g),
- weighing in kilograms [kg] (in balances with d0.01g,
- weighing in carats [ct],
- weighing in units from beyond SI (available only in non-verified
instruments):
[lb], [oz], [ozt], [dwt], [tlh], [tls], [tlt], [tlc], [mom], [gr], [ti], [N], [baht], [tola]
functions related to working modes
- counting pieces,
- checkweighing (control +/-),
- dosing,
- percent setup,
- density determination of solids,
- density determination of liquids,
- animal weighing,
- statistics
- totalising
- peak hold
functions related to setting a balance to ambient conditions at a workstation
- enabling and disabling autozero system,
- setting parameter of measurement result averaging,
- value release,
- display backlight,
- beep signal,
functions related to use of communication ports (RS232, USB, WIFI)
- setting communication port,
- setting bound rate,
- continuous data transmission in basic measuring unit,
- continuous data transmission in current measuring unit,
- automatic operation mode.
Balance can be used to weigh loads under the weighing platform – i.e. under hook weighing option. Balance software enables generating a report from adjustment process in a form of a printout containing modifiable data (see CAL REPORT)
Page 29
- 29 -
6. KEYBOARD
ON/OFF button enables switching on and off balance’s
display. If switched off balance components other than the
display are powered, and balance is in stand-by mode.
F9 button of the computer keyboar d.
Di
rect access to data stored in a databese: user, product,
tare. F10 button of the computer keyboard.
Function button, which enables quick entering the settings
of an active working mode.
F11
button of the computer keyboard
MODE button for selecting balance’s working mode. F5 button of the computer keyboard
UNITS button changes measuring units.
PRINT/ENTER button-
sends current display status to a peripheral device (PRINT) or accepts selected value of a parameter or function (ENTER).
ZERO button– zeroing of balance’s indication.
TARE –taring of balance’s indication.
Adjustment –
function button of immediate initiating the
adjustment / calibration process.
F6
button of the computer key board.
Function button for entering the main menu of a balance. F7 button of the computer keyboar d
Page 30
- 30 -
Navigating arrows for moving in balance’s menu or
changing parameter value.
7. START-UP
When plugged to mains, the balance displays name and number of software, and afterwards proceeds to weighing mode.
7.1. Connecting peripheral devices
The balance must be unplugged from the mains before connecting or disconnecting peripheral devices (printer or computer). Use only peripheral devices recommended by the manufacturer with your balance. These have been matched to your balance. After connecting a peripheral device, plug the balance to mains.
7.2. Balance temperature stabilisation period
Before start of measuring processes, it is necessary to wait until the balance is thermally stabilised. For balances that were stored in much lower temperatures before plugging to mains (e.g. during winter period) thermal stabilisation should last approximately 4 hours for NPS balances, and 8 hours for NAS and NAPP balances. During the thermal stabilization, the indications on the display panel can change. Correct operation of a balance is possible within temperature range specified in technical parameters, see point 1. It is recommended that ambient temperature changes at balance’s place of use are very small (slow to change).
Page 31
- 31 -
8. USER MENU
Balance menu is divided into 6 basic function groups. Each group has individual
name starting with capital letter P.
8.1. Moving through user menu
A user moves through balance menu using buttons on overlay’s keyboard:
Setup button. Entering balance’s main menu.
Selecting group of parameter one by one downwards or
changing parameter value by one value downwards.
Selecting group of parameters one by one upwards or changing parameter value by one value upwards.
Selecting group of paramete
rs for activating. On pressing the
button, the display indicates the first parameter in a selected
group
Exit to previous menu level, e.g. to main menu.
Abandon parameter changes.
Accept / confirm introduced changes.
Deleting a character while editing the text values of the menu, e.g. user name.
Inserting a character while editing the text values of the menu, e.g. user name
Page 32
- 32 -
Name of function groups and their content:
P1 ADJUSTMENT
P1.1 INT. CALIB. | [internal adjustment] P1.2 EXT. CALIB. | [external adjustment] P1.3 USER CALIBRATION | [user adjustment]
P1.4 CALIBRATION TEST | [adjustment test] P1.5 AUTO. CALIB. |NONE [automatic adjustment] P1.6 AUTO. CALIB. C. |1 [time of automatic adjustment]
P2 OPERATION MODE
P2.1 ACCESSIBILITY | [settings for the accessibility of individual modes while working with the balance] P2.2 WEIGHING | [setting for the function weighing] P2.3 COUNTING PCS | [settings for the function counting pieces]
P2.4 CHECKWEIGHING | [settings for the function checkweighing] P2.5 DOSING | [settings for the function dosing] P2.6 DEVIATIONS | [settings for the function deviations % against the mass of the standard]
P2.7 DENS. OF SOLIDS | [settings for determining density of solids] P2.8 DENS OF LIQUIDS | [settings for determining density of liquids] P2.9 ANIMAL WEIGHING | [settings for the function animal weighing] P2.10 STATISTICS | [settings for the function statistics]
P2.11 TOTALISING | [settings for the function totalising] P2.12 PEAK HOLD |[settings for the function peak hold]
P3 COMMUNICATION
P3.1 COM 1 | [transmission parameters port COM 1] P3.2 COM 2 | [transmission parameters port COM 2] P3.3 WIFI | [transmission parameters port WIFI]
P4 DEVICES
P4.1 COMPUTER; PORT |COM 1/COM 2/USB/WIFI/NONE [PC connection port]
CONT. TRANSM. | CALIBRATION UNIT./ CURRENT UNIT/ NONE
[turning on/off continuous transmission]
E2R SYSTEM | YES/ NO [turning on/off cooperation with E2R system]
P4.2 PRINTER
|COM 1/COM 2/USB/WIFI/PENDRIVE/USB PC/NONE
[printer connection port]
P4.3 BARCODE READER |COM 1/COM 2/NONE [barcode reader port]
Page 33
- 33 -
P4.4 ADD. DISPLAY |COM 1/COM 2/NONE [additional display port] P4.5 EXT. BUTTONS |TARE / PRINT
P5 PRINTOUT
P5.1 CAL. REPORT | [contents of the adjustment report] P5.2 HEADER | [contents of the header print-out] P5.3 GLP PRNT. | [contents of the weighment result prnt.] P5.4 FOOTER | [contents of the footer print-out] P5.5 NSD.PRN.1 | [project of non-standard print-out 1] P5.6 NSD.PRN. 2 | [project of non-standard print-out 2] P5.7 NSD.PRN. 3 | [project of non-standard print-out 3] P5.8 NSD.PRN. 4 | [project of non-standard print-out 4] P5.9 VARIABLE 1 | [project of variable 1] P5.10 VARIABLE 2 | [project of variable 2]
P6 OTHER
P6.1 LANGUAGE |POLISH / ENGLISH [menu language] P6.2 ACCESS LEV. |ADMIN [access levels for editing menu] P6.3 KEY SOUND |YES/NO [key sound]
P6.4 BACKLIGHT |70 [display backlight level] P6.5 DATE |2013.05.27 [data settings] P6.6 TIME |8:53:00 A [time settings] P6.7 DATE FORM.|YYYY.MM.DD/YYYY.DD.MM/DD.MM.YYYY/MM.DD.YYYY [date format] P6.8 TIME FORM. |12H/24H [time format] P6.9 GLP AUTOTEST | [carrying out autotest for the balance]
P7 INFO
P7.1 BALANCE ID | P7.2 SCALETYPE. | P7.3 PROG. VER. |
P7.4 TEMP. | P7.5 SETUP PRNT.
|
CAUTION
Changes introduced in balance memory will be saved, when abandoning the menu (on return to weighing). Press ESC button several times.
Page 34
- 34 -
9. LOGGING IN
In order to have full access to the user’s parameters and editing of database, the
person operating the balance, should log in as <ADMINISTRATOR> each time
they turn it on. The software enables the entry of 10 users with varying degrees of authority.
The first login procedure
While in the main window select <LOG IN>,available by pressing ,
or by pressing one of the function buttons to which the shortcut <LOG
IN> is assigned, or after pressing the button
, enter the users’
database and select the user <ADMIN>
After selecting the < ADMIN> button and confirming the choice by
pressing
, the program asks you to enter the user’s password,
You must enter „1111” and confirm by pressing ,
The program returns to the main window,
After the first login, you should first enter the users and assign the
appropriate access levels to them (procedures are described later in this manual, see section 14).
The next time you log in, select a user from the list and enter the password. After that the program starts working with the permissions for the selected user.
If any user is logged in, the pictogram
is displayed.
Logout procedure
In order to log out, select <NONE> from the list of available users
The program returns to the main window, and the display shows no
logged-in user. (no logged-in user, no pictogram on the display).
Page 35
- 35 -
Access levels
The balance software features three access levels: ADMINISTRATOR, ADVANCED, USER.
When you turn the balance on, the display is active all the time, which enables carrying out mass measuring processes.
The table below shows the access to the edition of user parameters, databases and software functions depending on the access level.
Access Access levels
USER
Access to parameters in the submenu: <Reading> and
setting in parameter group <Others> except for <Date and Time>. All the weighing processes can be started and carried out at this time. The user has the access to the information preview in <Databases>, and can also define universal variables.
ADV
Access to editing of parameters in the submenu:
<Reading>; <Working modes>; <Communication>; <Devices>; <Others> except for <Date and Time>. All the weighing processes can be started and carried out at this time.
ADMIN
Access to all the user parameters, functions and
editing databases.
Page 36
- 36 -
10. WEIGHING
10.1. Required weighing procedures
In order to ensure long lasting use of a balance and reliable
measurement of weighed loads, follow below procedures:
Start the balance with no load on the weighing pan
(permissible value of load on the weighing pan on balance start is ±10% of its maximal capacity).
Load balance’s weighing pan steadily with no shocks:
YES
NO
Place weighed loads centrally on the weighing pan:
YES
NO
Avoid side loading, in particular side shocks:
NO
NO
Page 37
- 37 -
Before the start of weighing process or in case of drastic change of ambient conditions at a workstation (e.g. ambient temperature change at a workstation more than 3°C) the balance requires adjusting. The procedure of balance adjustment is described further in this user manual.
Before the start of weighing procedure, it is recommended to load the balance’s weighing pan a few times with mass close to balance max
capacity,
Check if unloaded balance indicates “precise zero”
and whether
measurement is stable
, if not, press /Delete button,
Press UNITS button, to set a measuring unit:
[g], [mg], [ct], and also if enabled in factory menu (non-verified balances) [kg], [lb], [oz], [ozt], [dwt], [tlh], [tls], [tlt], [tlc] [mom], [gr], [ti], [N], [baht], [tola]
Place weighed object on balance’s weighing pan and read result only on stabilisation of measurement result,
Mass indication of a load placed on balance’s weighing pan can be tarred
for multiple Times by pressing T
/Insert (pay CAUTION not to exceed
maximal capacity of a balance by applying multiple tare function).
During times between carrying out the following measurement series do not unplug the balance from mains. It is recommended to switch off the balance’s display by
pressing ON/OFF button. On repeated pressing of the ON/OFF button the balance
is ready for operation and carrying out the following measurements.
Page 38
- 38 -
10.2. Zeroing
In order to zero mass indication press button.
The mass indication on the display should equal zero, and precise zero and stability marker
should appear.
Zeroing process is tantamount to determining new zero point recognised by the balance as precise zero. Zeroing is possible only in stable status of display indication.
Caution:
Zeroing the display indication is possible only within the ±2% range of
instrument’s maximal capacity If the zeroed value is above the ±2% of the maximal capacity, then th e softw are ind ic ates an error mes sage Err2.
10.3. Taring
In order to determine the net weight of a weighed object, place object’s
container (packaging) on balance’s weighing pan, and on stabilisation of
measurement result press
button. The display should indicate mass
equal to zero and symbols: Net and
.
On taking off the weighed load and its packaging from instrument’s weighing pan, the display indicates sum of total tarred mass with minus sign.
The software also enables assigning tare value to a product from a database. Then, on product selection from a database, the software automatically uploads data on tare value for the specific product.
Caution:
Taring negative values is impossible. On tarring negative values the balance responds with an error message Err3. In such case zero balanc e’s indication and repeat tarring procedure.
Page 39
- 39 -
Manual tare determination
Procedure:
While in optional mode press quick access button F, to which the option <ENTER TARE> is assigned in <SHORTCUTS> settings (description of the setting further in this user manual, poi nt <F button shortcuts>),
which opens a window,
use the arrow buttons to enter tare value and press
button,
the balance returns to weighing mode, and the display indicates
entered tare value with minus „–" sign.
Selection of tare from TARE DATABASE
Procedure:
while in any working mode, press F – quick access button, to which
option <SELECT TARE> is assigned in <SHORTCUTS>
(description of the setting further in this user manual, point <F button
shortcuts), or choose option <SELECT TARE> after pressing
button,
which displays the first mass value of the mass packages that was saved in tare database,
use the arrow buttons to select the tare which is to be recalled and
press
button,
the balance returns to the weighing mode, and the display indicates
entered tare value with minus „–" sign.
Or
While in any working mode press
button,
Enter b3 option <TARE>
Which displays the first mass value of the mass packages that was
saved in tare database,
Use the arrow buttons to select the tare which is to be recalled and
press
button,
The balance returns to the weighing mode, and the display indicates
entered tare value with minus „–" sign.
Page 40
- 40 -
AUTOTARE Function allowing for automatic taring of the packaging during the weighing process, when packaging mass for each of the product is different. Description of this function is to be found further down this manual.
Deleting tare
Entered tare value can be deleted by pressing
button on the balance’s overlay or by entering tare value of 0.000g (see description above).
10.4. Settings for mode <WEIGHING>
The software allows the set-up of operating parameters (filters, value release and autozero option, deleting the last digit and other settings) separately for each working mode. It enables customising the instrument and utilising properties depending on the user’s needs and expectations or specific requirements for selected working mode (e.g. DOSING); as a result the work is quick and easy.
Available settings
P2.2.1 – READOUT P2.2.1.1 – FILTER [V.FAST/ FAST / NORMAL / SLOW / V. SLOW] P2.2.1.2 – RESULT CONF. [REL. / FAST / FAST+REL.] P2.2.1.3 – AUTOZERO [YES / NO] P2.2.1.4 – LAST DIGIT [ALWAYS / NEVER / WHEN STAB.] P2.2.1.5 – AMBIENT CONDITIONS (STABLE/UNSTABLE) P2.2.2 – AUTOTARE [YES / NO] P2.2.3 – PRINT MODE P2.2.3.1 – MODE [EACH / WHEN STAB. / AUTO] P2.2.3.2 –AUTO. PROG P2.2.4 – UNITS P2.2.4.1 – AVAILABILITY P2.2.4.2 –START UNIT [G]
P2.2.4.3 – USER UNIT1 P2.2.4.4 – USER UNIT2 P2.2.5 – INFORMATION [TARE / NETTO / GROSS / USER / PRODUCT /
/NSTD. INF. / NONE] P2.2.6 – INF. NSTD. (non-standard information – any text, max 19 characters) P2.2.7 – SHORTCUTS (available options depending on the working mode) P2.2.7 1 – F1 [NONE / ENTER TARE/ …] P2.2.7.2 – F2 [NONE / ENTER TARE / …] P2.2.7.3 – F3 [NONE / ENTER TARE/ …] P2.2.7.4 – F4 [NONE / ENTER TARE / …]
Page 41
- 41 -
10.4.1. Setting level of filtering
Depending on the balance working conditions you should set the filter. In ideal conditions you can set the filter to be very fast (parameter value P.2.2.1.1 Filter to be V.FAST); however, if the working conditions are bad (shakes, drafts), the filter should be set to slow or very slow (parameter value P.2.2.1.1 to SLOW or V. SLOW). The effectiveness of the filter is different for the weighing range. The filter works slower when “approaching” the weighed mass, it works more intensively when the weighed mass is placed within the set range of the filter (parameter filter range is accessible only from the service menu – the user does not have an access to it). Depending on the adjusted filter, the weighing time will be shorter (V.FAST and FAST) or longer (SLOW and V. SLOW)
CAUTION
The higher level of filtering, the longer the weighing time.
10.4.2. Value release
Due to various ambient conditions at workplaces, and in order to set balance to present conditions, it is necessary to determine the value release parameter
as: FAST.+REL., FAST or RELIABLE. Depending on selected option,
weighing time is either shorter or longer.
Page 42
- 42 -
10.4.3.
Autozero function
In order to ensure balance’s precise mass indication, autozero (Auto) software
parameter has been introduced. The application of this function is automatic control and correction of zero indication. When the function is enabled, it compares balance indications at declared time interval e.g. 1s, on condition that weighing pan is unloaded and display indication is close to zero. If results vary less than declared AUTOZERO range e.g. one division, balance will zero automatically, marker of stable measurement result–
and precise zero
marker –
will be displayed.
If AUTOZERO function is enabled, then each weighing process starts from precise zero point. There are, however, some case when this function can be a disturbing factor in the measuring process; for instance very slow placing of load on the weighing pan (e.g. load pouring) in such case zero indication correction can also correct actual indication of loaded mass.
10.4.4 Displaying the last digit
Function enables displaying the last digit on the balance display.
Page 43
- 43 -
10.4.5 Balance ambient conditions
This parametre has two settings: STABLE and UNSTABLE. Setting on STABLE mode makes the balance work much faster than on UNSBABLE mode. This parametre relates to the ambient conditions in which the balance operates. If the ambient conditions are unstable it is recommended to use UNSTABLE mode. Normally the parametre is set on: STABLE.
10.5 AUTOTARE
In case of loads with different tare values being weighed one after another automatic tare function is usuful for quick determination of the weighed loads net mass. When function is active (<P2.2.2 AUTOTARE> value set to <YES>), the operating process is as follows:
Make sure that the weighing pan is empty and press button responsible for zeroing,
Put product packaging on weighing pan,
After measurement stabilization automatic taring of the packaging
mass will proceed (Net marker will appear in the upper part of the
display),
Put product that is to be packed on the packaging;
Display will show a net mass of the product;
Page 44
- 44 -
Take off the product together with the packaging;
The balance will cancel tara value (the packaging mass saved in
balance storage during the first step of the operating process) after
the gross mass value (set in < P 2.2.3.2 AUTO THRES>
parameter) has been exceeded;
Put packaging of a next product on weighing pan, after measurement stabilization automatic taring of the packaging mass
will proceed (Net marker will appear in the upper part of the
display);
Put a next product that is to be packed.
For a correct operation of the balance with AUTOTARE function, the threshold value must be adjusted.
<P 2.2.3.2 AUTO THRES > parameter is connected with the following functions:
- automatic tare,
- automatic operation,
There will be no automatic taring until the balance operates within the gross mass
value range set in < P 2.2.3.2 AUTO THRES > parameter.
10.6 Printing mode
This function enables setting the printing mode, i.e. button activity
.
A user has the possibility of selecting the following settings:
<WHEN STAB> the stable measurement result, along with the settings for parameter < GLP PRINTOUT> are sent to the printer port. Pressing
the button
, when the result is not stable (no sign on
display), the balance software will send the measurement result to the
port after reaching the stable condition of the measurement.
<ANYONE> every single pressing of button results in sending the measurement result to the printer port along with the settings for parameter <GLP PRINTOUT>. Every measurement result is sent (stable and unstable). The <?> sign appears at the beginning of the printing frame when the result is unstable.
This function applies only to non-verified balances.
<AUTO> - select this option to enable automatic printing of measurements. If this option has been selected, remember to select
<AUTO THRES > parameter to suit your needs.
Page 45
- 45 -
Automatic operation proceeds as follows:
Press button to zero the balance (marker of stable
measurement
and zero marker are shown on display)
Place the load, the balance sends the first stable measurement to
the printer port,
Remove the load from the balance, Next measurement is possible when the reading on the display is
lower than the set value of < AUTO THRES. > parameter (next measurement does not require zero value).
This procedure applies to changing settings:
For automatic working mode adjust the threshold value.
< AUTO THRES. > parameter is connected with the following functions:
- automatic tare,
- automatic operation,
For automatic operation function the measurement will not be send from the computer to the printer as long as the mass measurement stays withing the set
value range < P 2.2.3.2 PROG THRES>.
Page 46
- 46 -
10.7 Temporary measuring unit
This function is only available in <WEIGHING> mode. In other modes measurements are only possible in basic unit [g]. Function enables selecting a measuring unit which will be indicated next to mass reading during the weighing process. Set measuring unit will be in use from the moment of its activation until its change or switching off and on the balance.
Each press of the
button causes change of the measuring unit.
Available settings:
For verified balances, a user can select from the following: [g], [mg] or [kg], [ct]
For non-verified balances, a user can select from the following measuring units: [g], [mg], [kg], [ct], [lb], [oz], [ozt], [dwt], [tlh], [tls], [tlt], [tlc], [mom], [gr], [ti], [N]
10.8 Start-up measuring unit
This function is only available in <WEIGHING> mode. In other modes measurements are only possible in basic unit [g]. After selecting the start-up unit, when switched on, the balance will work in <WEIGHING> mode with this very unit. The units which can be selected are determined by the balance’s status, i.e. whether the balance is verified or non-verified.
Procedure:
Page 47
- 47 -
10.9 Availability of measuring units
This function is only available in <WEIGHING> mode. In other modes measurements are only possible in basic unit [g]. The user can declare the weighing units which will be available when selecting
temporary unit by pressing
button. For these units the parameter
availability is set to <YES>. These parameters can be selected by the user.
Procedure:
Page 48
- 48 -
10.10 User unit
A user can declare two weighing units. User unit is the result of balance indication multiplied by the factor that was introduced for user unit.
User units are marked as [u1] – user 1 unit and [u2] – user 2 unit.
The way of declaring factor for an individual user is shown below
User units are available only for non-verified balances.
Page 49
- 49 -
10.11 Information
Function enables displaying additional information in the bottom line. Depending on the needs, a user can choose the following options, which are
shown while working in mode <WEIGHING>:
Page 50
- 50 -
10.12 Non-standard information
Function enables declaring non-standard information, which is to be displayed in the bottom line of the display. A user can insert any text consisting of maximum 19 characters.
It should be remembered though, that the declared non-standard information is
visible. Therefore, option P2.1.2 should be set to < NSTD. INF. >.
Page 51
- 51 -
10.13 F button shortcuts
Function enables assigning quick access option for weighing functions which are accessible by choosing F1, F2, F3 or F4. A user can choose the following
options in <WEIGHING> mode, which can be assigned to any of the F
buttons< NONE / ENTER TARE / PRINT HEADER / PRINT FOOTER / VARIABLE 1 / VARIABLE 2>. There are different options available in other modes (see further on in this manual).
Page 52
- 52 -
10.14 Dual range balance (NPS 200/200 0.R2)
Balance NPS 200/2000.R2 series is a dual range weighing instrument. The accuracy of the I weighing range is d
1
=0.001g, and of the II weighing range is
d
2
=0.01g.
Transfer from weighing with the accuracy of the I weighing range to weighing with the accuracy of the II weighing range takes place automatically on exceeding
Max
1
200g (with no user activity needed). On switching to weighing with the
accuracy of the II weighing range, the display signals by indicating
symbol
on its left side, and an additional marker of last but one digit of the weighing result.
From now on the balance weighs mass with the accuracy of the II weighing range.
To return to weighing in the accuracy of the I weighing range:
Take the weighed load off the weighing pan
As the indication returns to zero and pictograms
→0←
and are lit,
press
button.
The balance returns to weighing with the accuracy of the I weighing range, and II
weighing range
pictogram and marker of the last but one digit are blanked.
Page 53
- 53 -
11 OTHER PARAMETERS
A user can set up parameters which have an influence on working with the balance.
These parameters are included in group P6 OTHER.
Changing settings for particular parameters is the same as described in the previous paragraph of this user manual.
P6 OTHER
P6.1 LANGUAGE |POLISH / ENGLISH [menu languages] P6.2 ACCESS LEV. |ADMIN [access levels for editing menu]
P6.3 KEY SOUND. |YES/NO [key sound] P6.4 BACKLIGHT |70 [display backlight level]
P6.5 DATE |2013.05.27 [date settings] P6.6 TIME |8:53:00 A [time settings] P6.7 DATE FORM. |YYYY.MM.DD / YYYY.DD.MM / DD.MM.YYYY / MM.DD.YYYY [date format] P6.8 TIME FORM. |12H / 24H [time format] P6.9 GLP AUTOTEST | [carrying out autotest for the balance] P6.10 SETUP PRNT. | [balance parameters print-out]
Enter group P6 OTHER menu.
11.1. Available menu languages
Parameter which enables selecting the language of the balance menu descriptions. Available languages: POLISH, ENGLISH, GERMAN, SPANISH, FRENCH, TURKISH, CZECH, ITALIAN, HUNGARIAN
11.4 Access levels
Enables choosing access levels when entering the balance menu. Available access levels: ADMIN. / USER. / ADV. Depending on the option selected, a user of a specific access level can access and make changes to the settings in the balance.
11.5 ‘Beep’ sound – reaction on pressing function button
Enables switching on/ switching off the ‘beep’ sound, which informs a user about pressing any button on balance’s overlay.
NO - ‘beep’ sound switched off YES - ‘beep’ sound switched on.
Page 54
- 54 -
11.6 Backlight and adjusting display brightness
Enables setting the brightness of the backlight or switching off the display brightness completely.
100 - maximum brightness of the backlight
10 - minimum brightness of the backlight
NONE - display brightness switched off
11.7 Date
Enables setting the current date.
As follows:
11.8 Time
Enables setting the current time.
As follows:
11.9 Date format
Enables altering the date format on the printout [YYYY.MM.DD / YYYY.DD.MM / DD.MM.YYYY / MM.DD.YYYY]
Where: YYYY – year MM – month DD - day
Page 55
- 55 -
11.10 Time format
Enables choosing the time format on the printout [12H / 24H].
For the set value [12H], in parameter P6.6 TIME letters <A> or <P> will be
displayed next to the time value.
Where:
A stands for hours before noon, P stands for hours after noon. Letters AM or PM are printed out next to the time value.
11.11 GLP autotest
AUTOTEST function is designed to aid a user in assessing balance’s operation and diagnosing the reasons for occurrence of errors in weighing which exceed the maximum permissible values for a given balance model. By means of a simple, repeatable and fully documented way the function enables optimizing balance’s settings to maintain the best possible repeatability and weighing time at workstation. The main purpose of the function is the possibility of monitoring the above-mentioned parameters at optional moment and saving records from the carried out tests in the form of printed reports of the tests that are automatically generated at the end of examination.
The test controls repeatability of placing the interval weight and determining error of indication with reference to balance’s maximum capacity.
Testing procedure:
two times loading of the internal weight, followed by 10 times loading of the internal weight
balance internal adjustment
calculating the value of standard deviation
report printout
Test results:
*Deviation for Max. *Value of repeatability of indication expressed as standard deviation
Page 56
- 56 -
Example of a report:
.....................................................
--------- Autotest GLP: Report --------- Balance type PS 3000.R2 Balance ID 400010 User Admin Software rev. v.0.4.9 Date 2013.07.16 Time 09:17:16
---------------------------------------- Number of measurements 10 Reading unit 0.001/0.01 g Internal weight mass 1402.094 g Filter Normal Value release Fast &Reliable
---------------------------------------- Deviation for Max. -0.118 g Repeatability 0.0088 g Signature
........................................
PROCEDURE:
Parameter <P6.9 GLP AUTOTEST > is initiated by pressing
button.
The balance software starts the autotest GLP procedure and it is carried out automatically from now on to the end.
A user can stop the procedure at any time by pressing
button in the window. Once the procedure is finished, the value of standard deviation of all measurements is shown in the main display, and message <RESULT> in the bottom line, and the final report is printed out automatically (as shown above). It is
possible to print out the report again by pressing
button. Report is not saved, therefore when abandoning this level, it is deleted from the balance volatile memory.
To abandon the window press
button and the program returns to the main
menu. To return to weighing press button repeatedly.
Page 57
- 57 -
12 INFORMATION ABOUT THE BALANCE
This menu includes information about the balance and its installed software. The parameters are strictly informative.
Once the parameter <PRINT. SET>> is chosen, the balance settings are sent to the printer (all parameters). The parameters are transferred via port that was chosen for <PRINTERS>.
Page 58
- 58 -
13 BALANCE ADJUSTMENT
In order to ensure the highest weighing accuracy, it is recommended to periodically introduce to balance memory a corrective factor of indications in relation to a mass standard – i.e. balance adjustment.
Adjustment should be carried out:
- Before the beginning of weighing procedure,
- If long breaks between following measuring series occur,
- If temperature inside the balance changes more than: 3°C.
Types of adjustment:
- Internal automatic adjustment * triggered by temperature change * triggered by elapsing time
- Manual internal adjustment
* initiated by pressing
- Adjustment with external weight * with declared mass which cannot be modified * of any weight, but not lower than 30% of maximum range.
CAUTION
In case of verified balances (with internal automatic adjustment system) only the automatic internal adjustment and manual internal adjustment system are available for a user. Remember to carry out the adjustment process when there is no load on the pan! In case there is load on the weighing pan, the display will indicate a command <RANGE EXCEEDED> . In such a case the load should be removed from the weighing pan and the process of the adjustment should be finished. Adjustment process can be aborted if necessary by pressing Esc at any time during the process.
Page 59
- 59 -
13.1. Internal adjustment
Adjustment process can be initiated automatically and manually. Manual means of activating adjustment procedure is achieved by pressing
. System of automatic adjustment will carry out the process fully
automatically and will inform the user on successive process stages.
Cycle of automatic adjustment process:
- Balance software detects the necessity of carrying out adjustment and
signals it by displaying a thermometer pictogram and <Cal> at the top of the display.
- During this time interval of about 2 minutes within which weighing
procedure can be completed
- As the 2 minute time elapses, balance display indicates message CAL_30
and starts counting down from 30..29..28 to 0 (indicated value is the counter),
- Balance user has 30 seconds to make a decision
In order to start adjustment, do not take any actions
In order to complete weighing procedure, press Esc. When pressed,
balance returns to weighing procedure and displays last weighing
result. In about 5 minutes balance indicates CAL_30 message again.
- The adjustment process can be postponed for multiple times, but it needs
to be pointed out that postponing of adjustment for a long time may lead to larger errors of weighing process. The errors are the effect of temperature changes and as a consequence changes of balance sensitivity.
Page 60
- 60 -
Automatic adjustment process takes place in three different areas:
Adjustment on plugging the balance to mains – verified balances.
Adjustment triggered by temperature change inside the balance. The
balance is equipped with very precise system for monitoring temperature. At each adjustment process, the temperature is saved in the system. The next adjustment is automatically initiated if temperature changes more than 3°C from the last saved temperature.
Adjustment triggered by elapsing time. It is possible to declare time intervals which are criteria for balance adjustment. Accessible settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours since last carried out adjustment procedure,
CAUTION
Declaring time intervals for balance adjustment is available only in non-verified balances. In case of verified balances the time interval between the following adjustm ents is set for 3 hour s.
13.2. Menu adjustment settings
P1 CALIBRATION
P1.1 INT.CALIB. | [internal adjustment] P1.2 EXT.CALIB. | [external adjustment]
P1.3 USER CALIB. | [user adjustment] P1.4 CALIB. TEST | [adjustment test] P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH [automatic
adjustment]
P1.6 AUTO CAL. C. | [time of automatic adjustment]
Page 61
- 61 -
P1.1 INT. CALIB. – Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically, and no user assistance is required. If weighing pan of the balance is loaded, balance will display a command to unload the weighing pan. If weighing pan is loaded with a relatively small mass (up to 4% of balance max capacity) then adjustment process will be carried out automatically, but measuring results may differ by the amount that was present on balance weighing pan.
or
P1.2 EXT. CALIB. – External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.
P1.3 USER CALIB. – User adjustment (with an external weight)
Adjustment with external weight at any mass in the balance range, however, not lower than 30% of max range.
CAUTION
Function unavailable in verified balances.
Page 62
- 62 -
P1.4 CALIB. TEST – Adjustment test
This fuction enables comparing the result of internal automatic adjustment with the value of internal weight saved in balance’s factory parameters. The process is conducted automatically and its result is shown on the display (if the balance is connected via RS 232 port with a computer or
printer, the adjustement result will be printed out). Press ESC to go back to
the previous menu.
P1.5 AUTO. CALIB. T. – Temperature of automatic adjustment (balances with internal adjustment)
Determination of factor which signals start of automatic internal adjustment
NONE - no factor will cause start of adjustement
(adjustment automatically turned off)
TEMP. - adjustment triggered by change of temperature TIME - adjustment triggered by time interval set in P1.6 AUTO CALIB.
C.
BOTH - adjustment triggered by time and temperature.
CAUTION
Function unavailable in non-verified balances.
P1.6 AUTO CALIB. C. - Time of automatic adjustment (balances with internal adjustment system)
Determination of time interval, after which automatic adjustment process is
initiated.
CAUTION
Function unavailable in verified balances.
Page 63
- 63 -
13.3. Manual adjustment
13.3.1. Internal adjustments
Press Cal button
or
The balance performs internal adjustment automatically. During adjustment process do not load the weighing pan with any load. The
message is displayed <DO NOT TURN OFF CALIBRATION> in the
bottom line. Once adjustment process is completed, balance saves its result in memory and returns to weighing mode.
CAUTION
- Press Esc button to abort adjustment process.
-If, during adjustment process, weighing pan of a balance is loaded, the display will show an error message. Adjustment will be stopped. As weighing pan is unloaded, the adjustment process will continue and be completed.
Page 64
- 64 -
13.3.2. External adjustment
The external adjustment in balances NPS series should be carried out with external mass standard / weigh class F
1
Start of external adjustment process,
Balance will display a command ordering unloading of weighing pan
<REMOVE MASS> (weighing pan must be empty). When weighing pan is
unloaded, press
button.
The balance determines mass of empty pan and message <CALIBRATION> is displayed in the bottom line. It is followed by message message <PLACE MASS> in the bottom bar, and in the main window the mass value to be placed on the weighing pan e.g. 200.000g (depending on the type of
balance). Place an external adjustment weight with mass determined on
balance’s display and press
button. The balance determines the mass and message <CALIBRATION> is displayed in the bottom line. On completing adjustment process the balance returns to submenu P1.2
INT.CALIB.
If a balance is verified, it is not possible for a user to carry out external adjustment process.
13.3.3. User adjustment
The external adjustment in balances NPS series should be carried out with external mass standard / weigh class F
1
Start of external adjustment process,
Page 65
- 65 -
The first procedure of the process is to enter the mass of the weight that is to be used for adjustment. The mass must be 30% Max capacity.
Once the mass of the weight is entered and confirmed, the message is
displayed which asks you to remove the weight from the pan <REMOVE
MASS> (weighing pan must be empty). After that, press button, the
balance determines the weight of unloaded pan and message
<CALIBRATION> is shown in the bottom line . Next, message <PLACE MASS>is displayed in the bottom bar, and in the main window the value of the mass to be placed e.g. 200.000g (depending
on balance type). Place the weight of the indicated mass and press
button. The balance determines the mass and message <CALIBRATION> is
displayed in the bottom line. After the adjustment process, the balance
returns to submenu P1.2 EXT.CALIB.
If a balance is verified, it is not possible for a user to carry out external adjustment process.
13.4. Adjustment report printout
At the end of each adjustment process or adjustment test, an adjustment report is generated automatically and sent to communication port COM 1. The content of the report is declared in menu P5.1 CAL REPORT.
The description of how to declare the settings for this option is included further in this manual, in the section on printouts.
The report can be printed out via the printer connected to the balance or it can
be sent to the computer and saved as a file for archiving.
Page 66
- 66 -
14. DETERMINING CONTENT OF A PRINOUT
14.1. Adjustment report
P5.1 CAL. REPORT, is a group of parameters which enable declaring data that is
on an adjustment printout.
P5.1 CAL. REPORT
P5.1.1 PROJECT | P5.1.2 CALIB. TYPE | YES P5.1.3 USER | YES
P5.1.4 PROJECT | YES P5.1.5 DATE | YES P5.1.6 TIME | YES P5.1.7 BALANCE ID | YES P5.1.8 CAL.DIFFER. | YES
P5.1.9 DASHES | YES P5.1.10 SIGNATURE | YES
P5.1.1 PROJECT
Option enables entering the name of the project (e.g. associated with a specific type of weighment). The name can be up to 16 characters. For instance, a project name entered into the balance in capital letters, looks like the figure below:
P5.1.2 CALIB. TYPE.
Enables printing out the type of the adjustment being carried out.
P5.1.3 USER
Enables printing out the name of logged-in user.
Page 67
- 67 -
P5.1.4 PROJECT
Enables printing out the name of the project. (see parameter P5.0.1)
P5.1.5 DATE
Enables printing out the date of the carried out adjustment.
P5.1.6 TIME
Enables printing out the time of the carried out adjustment.
P5.1.7 BALANCE ID
Enables printing out the balance ID number.
P5.1.8 CAL. DIFFER.
Enables printing out the difference between the mass of adjustment weight measured during the last adjustment and the current measured mass of this weight.
P5.1.9 DASHES
Enables printing out the dashes that separate the date of printout from the signature
P5.1.10 SIGNATURE
Enables printing out the signature of a user performing the adjustment.
Choose one of the options for the above-mentioned parameters.
NO - do not print on the report YES - print on the report
Example of a report:
Page 68
- 68 -
14.2. Header printout
P5.2 HEADER, is a group of parameters which enable declaring data that is on the
header printout.
P5.3 HEADER
P5.2.1 DASHES | YES P5.2.2 WORKING MODES | YES P5.2.3 DATE | YES
P5.2.4 TIME | YES P5.2.5 SCALE TYPE | YES P5.2.6 BALANCE ID | YES P5.2.7 USER | YES P5.2.8 PRODUCT | YES
P5.2.9 VARIABLE 1 | YES P5.2.10 VARIABLE 2 | YES P5.2.11 EMPTY LINE | YES P5.2.12 CAL.REPORT | YES
P5.2.13 NON-STANDART PRNT. | NONE / NSTD.PRN. 1 /
NSTD. PRN. 2 / NSTD.PRN.3
/ NSTD. PRN.4
P5.2.1 DASHES
Enables printing out a line of separating dashes.
P5.2.2 WORKING MODES
Enables printing out the name of balance’s working mode.
P5.2.3 DATE
Enables printing out the date in the header.
P5.2.4 TIME
Enables printing out the time in the header.
P5.2.5 SCALE TYPE
Enables printing out the scale type.
Page 69
- 69 -
P5.2.6 BALANCE ID
Enables printing out balance ID number.
P5.2.7 USER
Enables printing out the name of a logged-in user.
P5.2.8 PRODUCT
Enables printing out the name of a selected product.
P5.2.9 VARIABLE 1
Enables printing out the value of VARIABLE 1.
P5.2.10 VARIABLE 2
Enables printing out the value of VARIABLE 2.
P5.22.11 EMPTY LINE.
Enables printing out an empty separating line.
P5.2.12 CAL. REPORT
Enables printing out a report from the last adjustment, according to the settings declared for the adjustment report printout (see sec. 12.1 in this manual).
P5.2.13 NSTD. PRINT.
Enables printing out one of the four non-standard printouts on the header printout. You can choose one from the following options: NONE / NSTD. PRN 1 / NSTD.PRN.2/NSTD.PRN. 3 / NSTD. PRN.4 The way of entering non-standard printouts is described further in this manual.
For the parameters described above i.e. P5.2.1 to P5.2.12 you should choose
from:
NO - do not print in the header YES - print in the header
For the parameter P5.2.13, you should choose one of the following options:
NONE / NSTD.PRN. 1 / NSTD. PRN 2 / NSTD.PRN 3 / NSTD. PRN.4
Page 70
- 70 -
Example of a header printout:
14.3. Measurement result printout – GLP PRINTOUT
P5.3 GLP PRINTOUT, is a group of parameters which enable declaring data that is
on the measurement result printout.
P5.3 GLP PRNT.
P5.3.1 DATE | YES P5.3.2 TIME | YES
P5.3.3 USER | YES P5.3.4 PRODUCT | YES P5.3.5 VARIABLE 1 | YES P5.3.6 VARIABLE 2 | YES P5.3.7 TARE | YES
P5.3.8 GROSS | YES P5.3.9 CURR.RES. | YES P5.3.10 CAL. REPORT | YES P5.3.11 NSTD.PRNT. | NONE / NSTD.PRN. 1 /NSTD.PRN. 2 /
NSTD. PRN.3 / NSTD.PRN. 4
P5.3.1 DATE
Enables printing out the date of measurement.
P5.3.2 TIME
Enables printing out the time of measurement.
P5.3.3 USER
Enables printing out the name of a logged-in user.
Page 71
- 71 -
P5.3.4 PRODUCT
Enables printing out the name of the product that was weighed.
P5.3.5 VARIABLE 1
Enables printing out the value of VARIABLE 1.
P5.3.6 VARIABLE 2
Enables printing out the value of VARIABLE 2.
P5.3.7 TARE
Enables printing out the tare value.
P5.3.8 GROSS
Enables printing out the gross mass value.
P5.3.9 CURR. RES.
Enables printing out the current measurement result (NET mass) in the current unit.
P5.3.10 CAL. REPORT
Enables printing out a report from the last adjustment, according to the settings declared for the adjustment report printout (see sec. 12.1 in this manual).
P5.3.11 NSTD. PRNT.
Enables printing out one of the four non-standard printouts on the header printout. You can choose one from the following options: NONE / NSTD. PRN 1 / NSTD.PRN.2/NSTD.PRN. 3 / NSTD. PRN.4 The way of entering non-standard printouts is described further in this manual.
Example of a printout:
Page 72
- 72 -
14.4. Footer printout
P5.4 FOOTER, is a group of parameters which enable declaring the data that is on
the footer printout.
P5.4 FOOTER
P5.4.1 WORKING MODES | YES P5.4.2 DATE | YES P5.4.3 TIME | YES
P5.4.4 SCALE TYPE | YES P5.4.5 BALANCE ID | YES P5.4.6 USER | YES P5.4.7 PRODUCT | YES P5.4.8 VARIABLE 1 | YES
P5.4.9 VARIABLE 2 | YES P5.4.10 DASHES | YES P5.4.11 EMPTY LINE. | YES P5.4.12 CAL.REPORT | YES
P5.4.13 SIGNATURE | YES P5.4.14 NSTD. PRNT. | NONE / NSTD.PRN. 1 /
NSTD.PRN.2/NSTD.PRN.3 /
NSTD. PRN.4
P5.4.1 WORKING MODES
Enables printing out of the name of balance’s working mode.
P5.4.2 DATE
Enables printing out of the date in the footer.
P5.4.3 TIME
Enables printing out the time in the footer.
P5.4.4 SCALE TYPE
Enables printing out the scale type.
P5.4.5 BALANCE ID
Enables printing out balance ID number.
Page 73
- 73 -
P5.4.6 USER
Enables printing out the name of a logged-in user.
P5.4.7 PRODUCT
Enables printing out the name of a selected product.
P5.4.8 VARIABLE 1
Enables printing out the value of VARIABLE 1
P5.4.9 VARAIBLE 2
Enables printing out the value of VARIABLE 2
P5.4.10 DASHES
Enables printing out a line of separating dashes
P5.4.11 EMPTY LINE
Enables printing out an empty separating line
P5.4.12 CAL. REPORT
Enables printing out a report from the last adjustment, according to the settings declared for the adjustment report printout (see sec. 12.1 in this manual).
P5.4.13 SIGNATURE
Enables printing out the signature of the user performing the weighments.
P5.4.14 NSTD. PRNT.
Enables printing out one of the four non-standard printouts on the header printout. You can choose one from the following options: NONE / NSTD. PRN 1 / NSTD.PRN.2/NSTD.PRN. 3 / NSTD. PRN.4 The way of entering non-standard printouts is described further in this manual.
For the parameters described above P5.4.1 to P5.4.13 you should choose from:
NO - do not print in the footer
YES - print in the footer
For the parameter P5.4.14, you should choose one of the following options:
NONE / NSTD.PRN. 1 / NSTD. PRN 2 / NSTD.PRN 3 / NSTD. PRN.4
Page 74
- 74 -
Example of a footer printout:
14.5. Non-standard printouts
The balance’s software enables entering 4 non-standard printouts. Each of them can have approximately 160 characters.
Non-standard printout can include:
Variables depending on the working mode and other user’s needs (mass,
date etc.)
Permanent text in the user menu, it should be remembered, however, to
use CAPITAL letters and without Polish characters
Non-standard printout can have approximately 160 characters (typed as a
single line)
14.5.1. Inserting texts
Variables in all modes and with the same values
%% Print-out of a “%” character
%V Current NET mass in the current unit
%N Net mass in the current unit
%G Gross mass in the current unit
%T Tare mass in the current unit
%D
Current date
%M Current time
%I Balance number
%R Program number
%P Project number
%U User number
%F Name of current function – working mode
%C Date and time of the last adjustment
Page 75
- 75 -
%K Type of the last adjustment
%S
Currently weighed product
%Y Weighing deviation to the last adjustment
%1 Variable 1
%2 Variable 2
Variables depending on the currently used working mode
Variable Description Mode in which the variable is active
%W
Standard mass 1 pcs
COUNTING PIECES
%H High threshold
CHECKWEIGHING
%L Low threshold
%A Target mass DOSING
%B Reference mass DEVIATIONS
Non-standard characters used in designing non-standard print-outs
\\ a single „\” character
\C CRLF
\R CR
\N LF
\T Tabulator
\F Form feed (for PCL printers)
%E Crop the paper for EPSON printers
Every single print-out can have max 160 characters (letters, numerals, non­standard characters, spaces). A user can apply non-standard characters in order to the variables depending on the needs.
Example 1:
“RADWAG” DATE: <current measurement date> TIME: <current measurement time> PRODUCT MASS: <current mass indicator>
*****SIGNATURE:.........
<current working mode>
Page 76
- 76 -
You should enter the printout of the content setting and design the printout using appropriate variables and text format characters.
Example 2:
To crop the paper after the printout had been carried out by EPSON printer (if the printer is equipped witht an autocutter blade) the user must select (for a given printout: HEADER, GLP PRINTOUTor FOOTER) an option of non-standard printout 1,2,3 or 4 with <%E> value available and select this printout for a given printout settings.
In such case <SUFFIX> command should stay empty.
Paper must be cropped underneath the FOOTER. Example settings:
P5.4.14 STANDARD PRINTOUT | NSD. PRN. 1
P5.5 NSD. PRN. 1 | %E
The way of inserting texts
By the means of balance
Choice of the character to be changed. Moving the cursor or an active character (blinking) to the right.
Choice of the character to be changed. Moving the cursor or an active character (blinking) to the left.
Change of the character by one value down
Change of the character by one value up
Deleting a character
Page 77
- 77 -
Inserting a character
By the means of computer keyboard of the USB type
A computer keyboard of the USB type can be connected, which enables easier and quicker editing of the print-outs. In order to insert text, you should enter the appropriate option in the menu and with help of the keyboard insert the text and confirm by pressing Enter key.
CAUTION: It is important to type variables used for non-standard printouts in capital letters.
14.6. Variables
Variables are alphanumeric inrormation which can be linked to the printouts, products or other information related to weighing. For every variable enter its contents. Variables can be used for entering e.g. serial number or batch number during the products weighing. The program allows to enter two variables. Each can contain max 32 characters. In order to enter a variable, you need to enter the variable settings (parametre P5.9 – VARIABLE 1 or P5.10 – VARIABLE 2) and enter its contentents using direction keys (arrows) on the balance keypad or a computer keyboard. Entering texts procedure is the same as for non-standard printouts.
Page 78
- 78 -
15. DATABASE
The balance software has 3 databases that can be edited (USERS, PRODUCTS, TARES) as well as 2 databases (WEIGHMENTS AND ALIBI), in which all the measurements carried out on the balance are saved.
Data range that can be saved in particular databases:
USERS – 100 different users. PRODUCTS – 1000 different products. TARES – 100 different masses of the packages. WEIGHMENTS – 10 000 consecutive measurements ALIBI – 100 000 consecutive measurements
Operations on databases: USERS, PRODUCTS and TARES:
Deleting the selected item in the database
Inserting a new item in the database
The way of inserting and deleting an item is described below.
15.1. Users
USERS – 100 different users.
It is possible to introduce the following date for each user:
NAME (30 characters), CODE (6 characters), PASSWORD (8 characters, just numbers), ACCESS (USER, ADVANCED, ADMIN), LANGUAGE (any from the available).
Access levels
The balance software has three access levels: USER, ADVANCED, ADMINISTRATOR.
Once the balance is switched on, the display is active for the whole time, which enables carrying out mass measurements even without a logged-in user.
The table below shows the access to the edition of user parameters, databases and software functions depending on the access level.
Page 79
- 79 -
Access Access levels
USER
Access to parameters in the submenu: <Reading> and
setting in parameter group <Others> except for <Date and Time>. All the weighing processes can be started and carried out at this time. The user has the access to the information preview in <Databases>, and can also define universal variables.
ADV
Access to editing of parameters in the submenu:
<Reading>; <Working modes>; <Communication>; <Devices>; <Others> except for <Date and Time>. All the weighing processes can be started and carried out at this time.
ADMIN
Access to all the user parameters, functions and
editing databases.
In order to add a user, follow the scheme shown below, add a user and assign a name to them.
Page 80
- 80 -
Once the user’s name is added, follow the procedures shown below:
User’s code – Max 6 characters
User’s password – Max 8 numbers
Select the access level
Select the language
Once all of the data is inserted, you can return to the weighing mode. In order to select a user, follow the logging-in procedure described in section 9 in this manual.
In order to remove a user should follow this procedure:
Enter the user database – as in the description above
Page 81
- 81 -
Select the user that is to be removed from the list
Press
button
The software shows <DELETE?> in the bottom line
Confirm by pressing
button
When confirmed, the software removes the selected user from the list
Return to the weighing mode
15.2. Products
PRODUCTS – 1000 different products. The following data can be inserted for each
product: NAME (30 characters), CODE (6 characters), EAN (16 characters), MASS (with the accuracy of balance reading unit), TARE (mass of the packaging relating to a particular product with the accuracy of balance reading unit), MIN (low threshold for the mode WEIGHING>, should be inserted with the accuracy of balance reading unit), MAX (high threshold for the mode <CHECKWEIGHING>, should be inserted with the accuracy of balance reading unit), TOLERANCE (tolerance thresholds to [±] in mode <DOSING> entered as a % of the target mass).
In order to add a product, you should enter in the products database and add the name of the product (follow the procedure as in the section above).
Next, insert the details about the product.
Page 82
- 82 -
15.3. Tares
TARES – 100 different masses of the packaging. The following data can be
inserted for each packaging: NAME (30 characters), TARE (mass of the packaging, enter the value with the accuracy of balance reading unit).
In order to add tare – mass of the packaging, enter the tare database and insert a name for the tare (follow the procedure as in the section above).
Page 83
- 83 -
Next, insert the details about the packaging.
15.4. Weighments
Weighments database is non-editable i.e. the data relating to the weighments is
saved automatically. A user has a possibility of viewing this data and printing it out or exporting to PENDRIVE; if such a need occurs (procedure is described further on in this manual).
The balance software allows you to save and store up to 1000 measurements carried out on the balance. This occurs automatically, after a single press of the button <PRINT>, without a need of additional actions or settings change. The additional data is saved along with the measurement.
Date of the measurement
Date of the measurement
Measurement result (mass)
Tare value
Name of the product that has been weighed
Person carrying out the measurement (logged-in user)
Working mode in which the measurement has been carried out
Value of variable 1 and 2
The software saves the measurements in a so called loop, i.e. when the measurement 1001 is saved, the measurement 1 is automatically deleted from the balance’s memory.
Page 84
- 84 -
The measurements saved in the balance’s memory cannot be deleted.
It is possible for a user to view and print out the data saved in the memory.
Procedure:
Each measurement is saved with its individual number. The format is: b4.4.n, where <n> is the consecutive number of the saved measurement. In the bottom line the date and time are displayed for every single measurement.
To move through the saved measurements database, use the arrow
buttons
or . To move to the next measurement either up or down the
list, press one of the buttons.
To view the remaining data relating to the measurement, press
after selecting
the measurement in question:
The software will automatically show the data relating to the measurement in the bottom line of the balance display.
To move through the data relating to the measurement, press
or
The data relating to the measurement can be printed out by selecting the option
<PRINT> and pressing
button.
Page 85
- 85 -
Example of a printout Date 21.06.2013 Time 13:05:02 User Product
Tare 0.000 g Gross 0.000 g
0.000 g
--------------Cal. report.--------------- Cal. type Internal User Project 1234567890123459 Date 16.07.2013 Time 13:27:09 Balance ID 10353870 Cal. diff. -0.004 g
---------------------------------------- Signature
........................................
The data that is to be printed out, depends on the settings in parameter P5.3 GLP PRINTOUT. Depending on what data is set in this parameter for the print-out (by selecting <YES> the data is printed out also in WEIGHMENTS database
(see sec. 13.3.)
Page 86
- 86 -
15.5. ALIBI
The balance is equipped with “ALIBI”, a type of memory that allows you to save and store up to 100 000 measurements carried out on the balance. If the “ALIBI” memory is installed in the balance, the saving of the measurements occurs automatically, by pressing <PRINT> button, without a need of additional actions or settings change. The additional data is saved along with the measurement.
Date of the measurement
Time of the measurement
Measurement result (mass)
Tare value
Person carrying out the measurement (logged-in user)
The name of the product that has been weighed
The software saves the measurements in a so called loop, i.e. when the measurement 1001 is saved, the measurement 1 is automatically deleted from the balance’s memory.
The measurements saved in the balance’s memory cannot be deleted.
It is possible for a user to view and print out the data saved in the “ALIBI” memory.
Procedure:
Each measurement is saved with its individual number. The format is: b4.5.n, where <n> is the consecutive number of the saved measurement. In the bottom line the date and time are displayed for every single measurement.
To move through the saved measurements database, use the arrow
or
. To move to the next measurement either up or down the list, press one of
the buttons.
Page 87
- 87 -
To view the remaining data relating to the measurement, press after selecting the measurement in question:
The software will automatically show the data relating to the measurement in the bottom line of the balance display.
To move through the data relating to the measurement, press
or
The data relating to the measurement can be printed out by selecting the option
<PRINT> and pressing
button.
Example of a print-out Date 19.06.2013 Time 6:48:41 Result 199.90 g Tare 0.000 g User SMITH Product PILL
To return to the weighing mode press
button repeatedly.
Page 88
- 88 -
16. EXPORT AND IMPORT OF DATABASE
This option allows you to:
Archive the data relating to the weighments carried out– WEIGHMENT database and ALIBI database
Copy the products databases and tares databases between the balances of this series
This can be accomplished by using external memory PENDRIVE. Plug a PENDRIVE in USB 1 – TYPE A port if you want to use this option.
The balance automatically sees the external memory drive and the message enabling operations relating to the export or import of the database is displayed.
The following options are available when entering this parameter:
EXPORT database
IMPORT database
Page 89
- 89 -
16.1. Export database
To export the database, enter the option EXPORT.
The following functions are available:
Export of all the databases
Export of users databases
Export of product databases
Export of tares databases
Export of weighments
Export of weighments saved in ALIBI memory
Export of users parameters
After selecting the option <ALL DATABASES>, the balance software creates files on the PENDRIVE. The files are of relevant names and are saved in individual databases. The files have special extensions and the data saved in them is encoded in a way that the files cannot be read or seen in standard computer programs.
Special computer softwares manufactured by RADWAG company are used to read the data from the database files: ALIBI and WEIGHMENTS. The data from the files, in which the data from the PRODUCTS, USERS, and TARES databases is saved, is read automatically by the balance software in the option <IMPORT>.
Page 90
- 90 -
16.2. Import database
The function <IMPORT> allows you, amongst others, to transfer the data saved in the balance database. In this balance, the data to be entered in the new balance is entered. This is a quick and reliable way to enter the data without any mistakes.
To import the database, plug a PENDRIVE in USB port and next select IMPORT and choose one of the following options
The following options are available:
Import of all databases
Import of users databases
Import of products databases
Import of tares databases
Import users parameters
The data from the ALIBI and WEIGHMENTS databases cannot be imported.
16.3. Measurement data print-out
The balance software allows you to save the data relating to the measurement on the external memory - PENDRIVE. The operation can be performed as follows:
Plug a pendrive in USB port
Abandon <IMPORT/EXPORT> option, which is automatically started by
pressing
button
In the parameter P4.2.1 <DEVICES/PRINTER/PORT> set the option <PENDRIVE>
Return to the weighing mode
From this moment on, every single pressing of the button
leads to the saving the measurement data (compatible with the settings for GLP0 PRINTOUT) in a text file, which is automatically created by the balance
software. The file name is: printout.txt.
Turn the balance off by pressing
button so that the data is saved in the file. Only then can you remove the pendrive from the port and read the data on a computer.
Page 91
- 91 -
The data can be printed out on any printer connected to the computer.
More data can be saved in the same file. The balance software will add the date in the file that has already been created on the pendrive. Consequently, a user can continue with saving the measurements in the same file once created.
CAUTION: When the balance software sees the PENDRIVE, the menu <P8 IMPORT/EXPORT> is accessible. All the operations relating to export and impor t that ar e des cr ibed abo ve can be performed using this menu.
Page 92
- 92 -
17. BALANCE WORKING MODES
- Weighments
- Counting pieces
- Checkweighing
- Dosing
- Deviations % against the mass of the standard
- Animal weighing
- Density determination of solids
- Density determination of liquids
- Statistics
- Totalising
- Peak hold
In order to initiate a particular mode press
button, and follow by
selecting the mode from the list.
Page 93
- 93 -
Once the button has been pressed, the name of the first available
function is shown.
or - select the working mode
- enter the selected working mode
The setup for the functions is described further in this manual.
17.1. Setup for availability of working modes
In this group of parameters, a user declares functions which are to be available
for the user after pressing
button.
It is possible for a user to switch off the functions that are not used during working with the balance. This can be done by setting the availability
parameter to <NO>.
Follow the scheme shown below:
Page 94
- 94 -
17.2. Counting parts of the same mass
The balance, in the standard version, is equipped with the option of counting small objects of the same mass.
Start Counting Pieces function
When the function is initiated for the first time, the standard mass equals
0.0000g. If the sample mass is determined and has been used in the mode
<COUNTING PIECES> then the software accepts the mass of recently used
standard as the standard mass.
Page 95
- 95 -
17.2.1. Setup for mode COUNTING PIECES
The software allows entering settings for each working mode. Some the settings are identical in all working modes. They are described in section relating to the mode <WEIGHING>.
This section covers only the settings relating to the mode <COUNTING
PIECES>.
F button shortcuts
The function enables the option of quick access for weighing functions, which are available when buttons F1, F2, F3 or F4 are pressed.
In the mode <COUNTING PIECES> the following options are available, which a user can assign to each of the F buttons:
NONE – no function assigned to the button ENTER SAMPLE – function initiates the option of entering the reference
mass of a single piece
DETERMINE SAMPLE – function initiates the option of automatic
determination of the mass of a single piece from a number of samples of known quantity
SELECT PRODUCT – function initiates the option of selecting a product
from product database
LOG IN – function initiates the option of selecting and logging-in of a
balance’s user
ENTER TARE – function initiates the option of inserting packaging mass SELECT TARE – function initiates the option of selecting packaging
mass from product database
PRINT HEADER – function initiates the print-out of a designed header PRINT FOOTER – function initiates the print-out of a designed footer VARIABLE 1 – function initiates the selection and editing of variable 1 VARIABLE 2 – function initiates the selection and editing of variable 2
The way of declaring the function is described in the section relating to the weighing mode settings <F button shortcuts>.
Page 96
- 96 -
17.2.2. Setting the reference mass by determining from the sample of known quantity
While determining the mass of a single piece the function ACAI (Automatic Accuracy Correction) is in use.
Means of operatio n of ACAI function:
- Number of pieces ( on adding) on balance’s weighing pan has to be
greater than has been previously
- Number of pieces (on adding) on balance’s weighing pan must be less
than twice the amount of which was visible on the display before adding parts
- Current quantity of parts must be within the ± 0,3 tolerance of the total
value,
- Measurement result has to be stabilised.
Procedure:
Place the container on the pan and tare its mass,
Press one of the F buttons, to which the function <ENTER SAMPLE>
is assigned, and the editing window <SAMPLE QUANTITY> is
displayed
Use the navigating buttons or to select the correct
sample quantity.
For the option: any quantity (displayed value <0000> enter a numer
using navigating buttons).
Page 97
- 97 -
Zatwierdzić wybraną liczność próbki, zostanie wyświetlony komunikat
<PLACExx PCS>.
Place the declared number of pieces in the container and when the
result is stable (the symbol
is displayed) and confirm their mass by
pressing button,
The balance software automatically counts a single sample mass and
enters the mode <COUNTING PIECES> displaying the number of pieces which are on the pan (pcs). In the bottom line, a single sample
mass value is shown (if the option for the function
<INFORMATION>has been selected).
CAUTION:
It should be remembered that:
Page 98
- 98 -
The total mass of all the pieces put on the weighing pan must not be greater
than the balance’s maximum weighing range.
The mass of a single piece cannot be lower than the 0,1 reading unit of the
balance. If this condition is not met, the balance displays the message: <Too low sample mass>.
While determining the number of piece s, wait for the stability pictogram ,
in order to confirm the quantity.
Only after the stability pictogram is displayed can you confirm the declared
quantity by pressing
button. Otherwise, the balance will not accept the
measurement.
17.2.3. Setting the reference mass by entering its mass
Procedure:
Press one of the F buttons, to which the function <ENTER SAMPLE>
is assigned, and the editing window <ENTER SAMPLE> is displayed
Use the arrow buttons to insert the known a single sample mass value.
Confirm the entered mass by pressing button,
Page 99
- 99 -
The balance software automatically counts a single sample mass and
enters the mode <COUNTING PIECES> displaying the number of pieces which are on the pan (pcs). In the bottom line, a single sample
mass value is shown (if the option for the function
<INFORMATION>has been selected).
17.2.4. Returning to the weighing mode
Page 100
- 100 -
17.3. Checkweighing
Checkweighing is a working mode that uses two thresholds (LOW and HIGH) in order to check the mass of the samples. It is generally assumed that the mass is correct if it is contained within the threshold values.
F button shortcuts
The function enables the option of quick access for weighing functions, which are available when buttons F1, F2, F3 or F4 are pressed.
In the mode <CHECKWEIGHING> the following options are available, which a user can assign to each of the F buttons:
NONE – no function assigned to the button ASSIGN THRESHOLDS – function initiates the option of ascribing
thresholds for the checkweighing
SELECT PRODUCT – function initiates the option of selecting a product
from product database
LOG IN – function initiates the option of selecting and logging-in of a
balance’s user
ENTER TARE – function initiates the option of inserting packaging mass SELECT TARE – function initiates the option of selecting packaging
mass from product database
PRINT HEADER – function initiates the print-out of a designed header PRINT FOOTER – function initiates the print-out of a designed footer VARIABLE 1 – function initiates the selection and editing of variable 1 VARIABLE 2 – function initiates the selection and editing of variable 2
The way of declaring the function is described in the section relating to the weighing mode settings <F button shortcuts>.