RadPro TLD Cube User Manual

RadPro
Radiation protection
Environmental research
Medical dosimetry
Personnel Dosimetry
Experimental physics
Checking irradiated food
...Radiation Protection for the Radiation Professionals...
Manual TLD Reader
International GmbH
TLD Cube
i
Versatile and user friendly
TLD Cube is a small, lightweight and portable TLD Reader for measuring, analyzing and evaluating of thermoluminescent materials. Modern design and software combine easy operation and excellent performance in this newly developed reader. The ceramic heater with defined nitrogen cooling in the rotating drawer accepts all standard TLD shapes.
TLD Cube is available in two versions as TLD Cube 400 with heating capability up to 400°C and as TLD Cube 600 which provides heating capabilities up to 600°C.
The software includes calibration and QA features, data storage and export options as xml file or Excel file to meet all typical requirements.
Applications
All information in this brochure is subject to technical changes without notice.
Technical Specifications
Heater
State-of-the-art ceramic contact heater with thermocouple
Measuring Chamber
Sealed chamber with nitrogen circulation pump and cooling function
Base Plate
Stainless steel, exchangeable
Base Plate types
Chip 3.2 mm x 3.2 mm, Rod 1 mm x 6 mm, Disc Ø 4.5 mm, Multi-Purpose Ø 10 mm
Powder Dish
Inner Ø 8 mm, outer Ø 10 mm
Capacity
1 Sample per load
Linear heating ramps
Standard 0.1 K/s - 20 K/s (up to 40K/s possible)
Heating program
Multi-step program, adjustable temperatures +/- 1 K
Max. heater temperature
400C (TLD Cube 400) 600°C (TLD Cube 600)
Readout time
Freely adjustable
Accuracy
1 % S.D. (for multiple readout)
Stability
Better than 1 % during 8 h operation
Test light
Stability < 0.5 %, colour blue
Detector
Thermoelectric temperature stabilization
Data output
LAN to PC
Measurement range
7 decades
Signal processing
24 bit, full digital evaluation system
Neutral Density Filter
Exchangeable
Power supply voltage
100 – 240 V, 50 - 60 Hz
Power consumption
200 VA
Shielding and cooling gas
Nitrogen, approx. 300 ml / measurement
Software
State-of-the-art Windows software with full calibration and export functions
Dimension approx.
15 cm x 17 cm x 29 cm (W x D x H)
Weight approx.
5.5 kg
Carrying Case approx.
48 cm x 25 cm x 25 cm, 11 kg (incl. reader)
Accessories
N2 gas pressure reducer, metric
Programmable TLD Oven
Vacuum Tweezers
incl. Vacuum Pump
Storage dish for 120 TLD
elements
RadPro
International GmbH
...Radiation Protection for the Radiation Professionals...
Burger Straße 28
42929 Wermelskirchen
Germany
Phone: +49 2196 889803
Email: sales@radproint.de
Web: www.radpro-int.com
Loading...