Radiodetection T631 TDR Operating Manual

Page 1
T631 TDR
High Specification Coaxial Cable Fault Locator
Operating Manual
Issue 6
Refer to Preface and Safety Instructions before operating
Radiodetection Ltd.
Bristol
BS14 0AF
United Kingdom
Tel: +44 (0) 117 976 7776
Fax: +44 (0) 117 976 7775
E-mail: sales.uk@radiodetection.spx.com
00901209-6
Page 2
© Radiodetection Ltd 2007. All rights reserved. This document is copyright and may not be copied or reproduced or transmitted or modified or used, in whole or in part, in any
way whatsoever, without the specific prior wr
itten consent of Radiodetection Ltd.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 2
Page 3
CONTENTS
Page
1. Preface & Safety 5
2. Introduction 7
3. Front Panel - Controls & Indicators 11
4. Power Sources 16
5. Display 19
6. Operating Modes 20
7. Operating Instructions 21
7.1 General 21
7.2 Switch On 21
7.3 Programming from the Help Menu 22
7.4 Setting Dielectric (PVF) 23
7.5 Connecting to cable under test 27
7.6 Locating a cable fault 27
7.7 Use of left hand cursor 28
7.7.1 Eliminating the length of the test lead 29
7.7.2 Measuring between two features 29
T631 OPERATINGMANUAL 3
Radiodetection Ltd. 00901209-6
Page 4
CONTENTS
Page
7.7.3 Resetting the LH cursor 30
7.8 Auto Fault Find 30
7.9 Return Loss Measurements 32
7.10 Filter 38
8. Use of the Memory 39
8.1 Storing a trace 39
8.2 Recalling a trace 39
8.3 Comparing a live and stored trace 39
8.4 Difference between a live and stored trace 40
9. RS232 Printing and PC Interface 41
10. Help Menu 43
11. Error Messages 44
12. Specification 45
13. Product Safety Data 53 Appendix – Bicotest Product Range 57
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 4
Page 5
1. PREFACE & SAFETY
1.1 HEALTH AND SAFETY AT WORK ACT 1974 SECTION 6.1 (C)
Thisproductistestedandsuppliedinaccordancewithpublishedspecificationsand,whenused in normal or prescribed applications and within the parameters specified, it will not cause danger or hazard to health or safety.
All usage of the product must be in accordance with the Operating Manual for this equipment and any workonthe electricalcomponentshoused within the machine must be undertaken by qualified personnel.
T631 OPERATINGMANUAL 5
Radiodetection Ltd. 00901209-6
Page 6
1.2 SAFETY PRECAUTIONS
Forconnectiontolivecablesupto600 VRMSusethe Bicotest T631F Blocking Filter.
Only fit 2 Ah ‘C’size NiCad cells. Do not charge the batteries when the ambient temperature is below 0°C.
AC adaptor and its carry pouch are not waterproof.
AC adaptor is for indoor use only.
The main unit is safe and conforms to IEC 1010. The AC adaptor conforms to the manufacturers safety standards.
1.3 BATTERIES
The T631 is provided with a set of 2 Ah rechargeable NiCad cells. These are supplied in the discharged state and must be charged for 24 hours before use. Charge the batteries as describedin section 4.Note that full rated capacity maynotbeachievedforthefirstthreecycles of use.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 6
Page 7
2. INTRODUCTION
The Model T631 is a pulse reflection (TDR) Cable Test Set, for locating cable faults and evaluating changes in impedance caused by connectors, taps, terminations etc. Pulses transmitted into a cable are reflected by cable imperfections. The transmitted pulse and the reflectedpulse(s) are shown on thedisplay.Thetimetaken by the pulse totraveltothe faultand return is ameasure of the distance to the fault.Distance to fault is displayedon the screen after the cursor is positioned to coincide with the start of the fault pulse. The type of fault can be determined by analysis of the displayed waveform.
Althoughdesignedprincipallyfor50,75and93 coaxial cables it is equally effectiveonothercable types as a fault locator.
Impedance mismatches can be measured in terms of Return Loss (dBRL) with the location displayed in feet, metres or time.
Principle features include a 15 location memory for trace storage with ability to transfer to a printer or PC, and dual cursors for point to point measurements.
T631 OPERATINGMANUAL 7
Radiodetection Ltd. 00901209-6
Page 8
2.1 SAMPLE TRACES
1.
Open circuit/high impedance series faults
Note: Positive (upward) reflection
2.
Short circuit/low impedance shunt faults Note: Negative (downward) reflection
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 8
Transmit Pulse Reflected Pulse
Transmit Pulse Reflected Pulse
Page 9
The T631 will display:
a) A “live” trace. b) A stored trace recalled from the memory. c) A live and stored trace together for comparison. d) The difference between a live and stored trace.
2.2 POWER SOURCE
The T631 is powered by 8 rechargeable cells in a compartmentaccessiblefrom the rear of the T631, or from an external DC power source via the DC jack.
When operated from batteries, the T631 will automatically switch off before the batteries are completely discharged.
T631 OPERATINGMANUAL 9
Radiodetection Ltd. 00901209-6
Page 10
2.3 MAIN FEATURES
Figure 1 shows the front panel of T631.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 10
Page 11
3. FRONT PANEL
3.1 CONTROLS
Switches the unit ON. To conserve battery life the unit will switch off 5 mins after the last key operation. If automatically switched off all measurement parameters will be restored when switched on again.
Switches the unit OFF. This will reset the instrument to the 6 m range, (or equivalentinfeetortime),GainA1,2 nspulseandLINEmode.PVF,unitsand impedance will remain as they were before switch off.
Rotary control to adjust the contrast of the screen display.It is recommended that it is set to the mid position on switch on to obtain a display.
CURSORS – two independent cursors
Adjusts the position of the appropriate cursor. Cursors move left or right as indicated.Ifhelddownthecursorinitiallymovesslowlyandwillthenaccelerate. The cursor being moved is displayed as a solid line, the passive cursor is a broken line.
T631 OPERATINGMANUAL 11
Radiodetection Ltd. 00901209-6
Page 12
PULSE WIDTHS
Enablesthevarious pulse widths availableoneachrange to be selected.The pulse widths are either automatically selected (AUTO) or set to a specific value (MANUAL) via the HELP menu.
HORIZONTAL EXPANSION (ZOOM)
Enablesthetraceaboutthesolidcursortobeexpandedbyfactorsof2,4and8, dependingonrangeselected(seeSection7.7andSection12,Specificationfor further details).
OPERATION
Selects the operating mode (see Section 6 for further details).
DIELECTRIC
Used to set the appropriate velocity factor for the cable dielectric. The value willincrementordecrement.Ifhelddownthevaluechangesslowlyatfirstand then quickens.Also act as numbers 6 and 11.
RANGE
Selects the displayed range.The range will increase or decrease and if held down will step through the available ranges. Also act as numbers 7 and 12.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 12
Page 13
SHIFT
Adjusts the vertical position of the live trace. If held down the trace shifts
– slowly at first and then quickens.Also act as numbers 8 and 13.
AMPLITUDE
Adjusts the vertical amplitude of the trace. If held down the unit will step through the available gain stages.Also act as numbers 9 and 14.
MEMORY
Enablesalivetracetobestored into any of 15 memorylocations.Also acts as number 10.
Enables any of 15 stored traces to be recalled.Also acts as number 15.
BACKLIGHT
Switches the display backlight ON and OFF. The backlight will switch off automatically after 5 minutes.Also acts as number 5.
T631 OPERATINGMANUAL 13
Radiodetection Ltd. 00901209-6
Page 14
INTERFACE
Calls up a menu from which the displayed trace and data may be transferred toaprinter,or memorised waveformsmaybetransferredtoa PC,or reloaded from a PC.Also acts as number 4.
AUTO FAULT FIND
Causes the unit to search for the next significant reflection, startingfrom the right hand cursor position (see Section 7.8). Also acts as number 3.
HELP MENU
Accesses a “Help” menu for setting machine parameters (see Section 10). Also acts as number 2.
RETURN LOSS
Enables the return loss between the transmitted pulse and a selected reflected pulse to be determined. Also acts as number 1.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 14
Page 15
3.2 CONNECTORS Input/Output
BNC socket to connect to the cable under test.
9 way D connector to interface with a printer or PC.
DC jack to supply 12 – 20 V external DC. Note centre is positive.
3.3 CHARGE INDICATOR Battery Charging
Charge Indicator illuminates when the internal battery is being charged.
T631 OPERATINGMANUAL 15
Radiodetection Ltd. 00901209-6
Page 16
4. POWER SOURCES
4.1 INTERNAL BATTERY
4.1.1 RECHARGING
Thecellsmaybe recharged by connecting the ACadaptortothe DC power jack and leaving the equipment switched off. The charge indicator will illuminate while the batteries are being charged.Full charge is achieved in 14 hours. A fully charged battery will give approximately 8 hours operation if the backlight is not used. (see Section 12 ‘Power Source’ for details of charging voltages).
Note:Whenfittingnewcells the initial charge period should be 24 hoursand14hoursthereafter.
4.1.2 MAXIMISING BATTERY LIFE
Do not recharge until the Battery Low warning is displayed. Do not charge for more than 24 hours. Stay within the RECOMMENDED temperature limits shown in the specification (Section 12).
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 16
Page 17
4.1.3 REPLACEMENT
Figure 2 shows the rear view of the unit. Cells are accessed by removing the back
cover which is secured by 2 fasteners. See specification (Section 12) for cell type. Itis recommended that cellsare replaced as
a set of 8.
T631 OPERATINGMANUAL 17
Radiodetection Ltd. 00901209-6
Figure 2 – Battery Access
Page 18
4.2 EXTERNAL DC
Theunitwillrunfromanexternal12 – 20 VDCpowersourceviatheDCjack.This maybeeither the AC adaptor or any other DC supply that meets the specification.Centre is positive.
Notethat the batter ies arecharged when external 15 – 20VDCpoweris applied and the unit is switched off – DO NOT FIT NON-RECHARGEABLE BATTERIES.
4.3 MEMORY BACK UP BATTERY
The memory back up battery is a lithium manganese cell which is part of the main PCB assembly.
The life expectancy is typically four years and replacement is recommended every two years during a routine service to avoid possible loss of stored traces.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 18
Page 19
5. DISPLAY
T631 OPERATINGMANUAL 19
Radiodetection Ltd. 00901209-6
Figure 3 – T631 Display
Page 20
6. OPERATING MODES
The unit has 4 display modes.
1. Line (L) thetraceobtainedfromthecableconnectedtotheBNCsocketisdisplayed.
2. Memory (M) the trace from a selected memory is displayed.
3. Dual (L & M) the trace obtained from the cable connected to the BNC socket and a
selected trace recalled from the memory are displayed simultaneously.
4. DIFF (L – M) the trace displayedisthedifferencebetweenthetraceof the cable connected
to the BNC socket and a selected trace from the memory.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 20
Page 21
7. OPERATING INSTRUCTIONS
7.1 GENERAL
For correct operation the cable under test should be taken out of service with all sources of supply removed.If this is not possibleand the cable is energised or likely to become energised then the blocking filter accessory should be used.
RF signals present on telecommunications cables may corrupt the display.
7.2 SWITCH-ON
Powertheunitfrompreviouslycharged internal batteries or fromtheACadaptor,orexternal DC source.
Press the on button and the display should appear. Rotate the contrast control if necessary to obtain a well defined trace. On receipt from the factory the unit will powerup in the LIVE cable mode, on the 6 m range, with
thedistancereadoutinmetres,apropagationv elocityfactor (PVF)of0.667andimpedance75 Ω. The left hand cursor,(dotted), will be at the startof the transmitted pulse (0 m). The right hand
cursor, (solid), will be approximately one third of the way across the display.
T631 OPERATINGMANUAL 21
Radiodetection Ltd. 00901209-6
Page 22
Thedistanceunits,impedanceandthePropagation VelocityFactorMode (i.e.PVF,V,or
V 2
)
can be re-programmed via the HELP menu and once programmed will be set automatically on switch on thereafter.
Should the unit switch off automatically, the last established machine settings are restored when the unit is switched on.
7.3 PROGRAMMING FROM THE HELP MENU
7.3.1 DISTANCEUNITS
Press HELP key to obtain the Menu. Selects UNITS by pressing the key indicated. Select Feet, Metres or Time by pressing the appropriate key.
7.3.2 DIELECTRIC PARAMETERS
Press HELP key to obtain the Menu. Select DIELECTRIC by pressing the key indicated. Select PVF, V, or
V 2
by pressing the appropriate key.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 22
Page 23
7.3.3 IMPEDANCE
Press HELP key to obtain the Menu. Select IMPEDANCE by pressing the key indicated.
Select 50, 75 or 93 by pressing the appropriate key.
7.3.4 PULSE SELECTION
A fixed pulse width may be selected by pressing HELP,PULSE WIDTH and then choosing the required pulse.This pulse width will then apply to all ranges and the pulse width control will not function until auto is restored, or the unit is switched off.
Auto pulse selection allows the instrument to select a pulse appropriate to the selected range. This mode is selected by pressing HELP and then selecting PULSE WIDTH and AUTO.
7.4 SETTING DIELECTRIC (PVF)
7.4.1 FROM CABLE TYPE
Theunitcanbedirectlyprogrammedfromthelist of cable types containedintheHELPMenu. To access the Menu, press HELP, press the key appropriate to CABLE TYPE. From the list of
options select the appropriatecategory and then the particularcabletype.The available cable types are shown in the following table.
T631 OPERATINGMANUAL 23
Radiodetection Ltd. 00901209-6
Page 24
On returning to normal operation the display will confirm the selected PVF value.
Note:The PVF values are nominal values and variations may be encountered. The DIEL keys are then used to set the specific value.
The PVF (DIEL) can also be set using the DIEL controls to the value appropriateto the cable under test.
OPTION MAIN TYPE SUB TYPE DIELECTRIC SETTING
M/µs
(PVF) V V/2
1 TELEPHONE POLY PE
JELLY FILLED PAPER (0.83 uF) PAPER (0.72 uF)
0.667
0.64
0.72
0.88
200 192 216 264
100
96 108 132
2 POWER PILC 25 kV
XLPE MIC
0.54
0.52
0.41
162 156 123
81
78
62
3 CATV QR PARA III
PARA I T, TR TX, TX10 RG6, RG11, RG59
0.88
0.82
0.87
0.89
0.82
264 246 261 267 246
132 123 131 134 123
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 24
Page 25
OPTION MAIN TYPE SUB TYPE DIELECTRIC SETTING
M/µs
PVF V V/2
4 IBM TYPE 1
TYPE 2 TYPE 3 TYPE 6 TYPE 9
0.78
0.78
0.62
0.78
0.69
234 234 186 234 207
117 117
93 117 104
5 DATA RG58
RG58U THICK ETHERNET THIN ETHERNET RG11, RG6
0.78
0.66
0.78
0.67
0.82
234 198 234 201 246
117
99 117 101 123
For conversion 1 metre = 3.28 feet.
T631 OPERATINGMANUAL 25
Radiodetection Ltd. 00901209-6
Page 26
IfthevalueofthePVForthecable type is not knownthenthePVFcanbedeterminedasfollows.
7.4.2 DIELECTRIC CONSTANT(PERMITTIVITY) KNOWN
PVF =
1
ε
Example: For polyethylene ε = 2.25
PVF ==
1
225
0667..
7.4.3 SHORT LENGTH OF SAME TYPE OF CABLE
1. Measure physical length of the sample.
2. Obtain a reading for the apparent length of the cable using any PVF value.
3. Adjust the PVF value until the distance reading indicated is the same as the measured
length.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 26
Page 27
7.5 CONNECTING TO CABLE UNDER TEST
Connect the cable under test directly to the BNC socket or via the test lead supplied. UnlesstheLHcursorismovedtocoincidewiththeendofthetestlead,thedistanceread out will
include the length of the test lead which must be subtracted from the reading obtained.
7.6 LOCATING A CABLE FAULT
Set the DIEL appropriate to the cable type (see Section 7.4).Select a RANGE to coverthe full cable length.Use “AUTO”Pulse Width selection unless a specific pulse width is preferred.
Adjust the AMP controls until the fault reflection is observed. If the fault reflection is less than halfway across the screen select a shorter range and a
narrower pulse width for optimum results. If there is no fault reflection there are two possibilities.
1. Thecablelength is longer thanits recorded length. In this case increase the rangeuntil a reflection is seen.
2. The fault is very close to the position at which the instrument is connected. In this case reduce the range until a reflection is seen. In both cases, use may be made of the auto fault find facility (see Section 7.8).
T631 OPERATINGMANUAL 27
Radiodetection Ltd. 00901209-6
Page 28
AdjusttheRight Hand cursor controls to bringthe RH cursor to the pointatwhich the startofthe reflected pulse just leaves the horizontal as shown.
The distance to the fault is as indicated on the LCD. For closer examination, press ZOOM to expand the trace and reposition the cursor more
precisely . Remember to subtract the length of the test lead if used.
7.7 USE OF THE LEFT HAND CURSOR
The left hand cursor can be moved to:
1. Eliminate the length of the test lead.
2. Measure between any two features on the trace.
3. Whenmovedthe LH cursorbecomes the livecursor and changes to a solid line, with the
RH cursor becoming a broken line.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 28
Page 29
7.7.1 ELIMINATINGTHE LENGTH OF THE TEST LEAD
Connect the test lead to the unit. Select 3 m or 6 m range. Adjust the controls so that the reflection from the end of the test lead is obtained. Using the LH cursor controls move the LH cursor to the start of the reflection. Connect the test lead to the cable under test and locate the fault as described in Section 7.6.
NB: The distance to fault is now that from the end of the test lead.
7.7.2 MEASURING BETWEEN TWO FEATURES
Using the LH cursor controls set the LH cursor to the start of the feature on the LH side. Using the RH cursor controls set the RH cursor to the start of the feature on the RH side. The RH cursor is then live and becomes a solid line with the LH cursor a broken line. Thedistance read out is the distance between the
two cursors as shown. IntheZOOMmode,ifacursorisnotonthescreen
it cannot be moved.Therefore if a measurement
T631 OPERATINGMANUAL 29
Radiodetection Ltd. 00901209-6
Page 30
between two features is to be made in ZOOM mode, position cursors at the relevant positions, then ZOOM and make any fine adjustments as necessary. The readout is always the distance between the cursors, even if one is not on the screen. Since expansion takes place around the live(solid) cursor, the ZOOM facility can be activated by either the left orright cursor by moving the relevant cursor and thereby making it solid.
7.7.3 RESETTING THE LH CURSOR
The LH cursor may be reset to `0' for normal measurements by switching the unit off and on. Note that the instrument is calibrated such that cable lengths are measured from the co-ax
socket.Dueto the internalwiring of the instrument, this may result in the left hand cursor being set a short distance into the displayed transmit pulse, especially on the short pulses. This is correct and ensures accurate cable length measurements.
7.8 FIND FAULT – ENSURE RANGE SETTING IS FULL AND NOT EXPANDED (ZOOM)
The “FIND FAULT” function is a means whereby the unit will automatically scan a cable and locate the nearest significant feature and configure the parameters of the unit to optimally display the feature.
Pressing the key a second or subsequent time locates the next significant feature along the cable.
Itis essential that the PVF is set to the valueappropriate to thecable type in orderthat distance measurements are accurate (see Section 7.4).
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 30
Page 31
Having obtained the display the parameters may then be altered if required. This feature is intended for use on coaxial cableswhere the reflection from a fault may be quite
small, so the sensitivity used is quite high. On a power cable or twisted pair telephone cable, mismatches due to joints, or change of cable type may give rise to many reflections which the instrument will detect even though they are not faults.
7.8.1 OPERATION
The FIND FAULT function starts at the location of the RH cursor and searches to the right. Therefore for a full search the RH cursor should first be placed just to the rightof the transmit pulse.
Pressing the FIND FAULT key will start the search. This may take several seconds. While searching, “AFF” is displayed in place of the usual mode display.
Searching is carried out on all ranges from the one appropriate to the starting cursor position (6 m range if the cursor starts on zero) to 12 Km. Gain and Pulse Widths appropriate to the range are used.
When a fault is found,the unit will beep twice and place the RH cursor just before the fault.Ifno fault is found the unit will beep 4 times, and return to the original conditions.
Pressing FIND FAULTagain will cause the next fault to be found.Note that “FIND FAULT” may change the machine settings.
T631 OPERATINGMANUAL 31
Radiodetection Ltd. 00901209-6
Page 32
7.9 RETURN LOSS (dBRL) MEASUREMENTS
The T631 has two principles of operation.
1. Simple fault location.
2. Qualitative measurement and location.
In the Simple Fault Location mode the pulse width is automatically selected with each range to onebest suited to provide a strong reflection from the fault, in orderto easily identify and locate it.Nosignificant conclusion can be drawn from the amplitude of the reflected pulse although its polarity is indicative of the type of fault.
i.e. Upwards – positive – high series impedance
Downwards – negative – low shunt impedance
In order to gain a measure of the significance of the reflection it is necessary to compare the amplitude of the reflected pulse with that of the transmitted pulse.
The amplitude of the reflection resulting from an impedance change, relative to that of the transmittedpulseis a measure of theimpedancemismatch at that point andthe ratio is referred to as the reflection coefficient.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 32
Page 33
Reflection coefficient
ρ=
Ar At
where
Ar = amplitude of reflected pulse At = amplitude of transmitted pulse
The amplitude of the reflected pulse cannot normally exceed that of the transmitted pulse and therefore the reflection co-efficient cannot exceed unity. It is usual therefore to express the reflection in percentage termsor millirhos (mρ) where:
1000 mρ = 1 rho = 100 % reflection ImpedancemismatchescanalsobeexpressedlogarithmicallyintermsofReturnLosswhere:
Return Loss = 20 log
10
At Ar
dB
or Return Loss = 20 log
10
1
ρ
dB
T631 OPERATINGMANUAL 33
Radiodetection Ltd. 00901209-6
At
Ar
Page 34
A conversion chart is given below:
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 34
Reflection Coefficient/Return Loss Conversion Chart
Page 35
The foregoing assumes an ideal “lossless" cable.In practice, measurements obtained include the effect of pulse attenuation due to cable loss and requires correction in order to obtain true return loss.
i.e.True Return Loss = Measured Return Loss – Pulse Attenuation Sincethe amplitude of the reflection varies with the pulsewidth it is essentialwhen determining
Return Loss to use the SAME PULSE WIDTH for all measurements. This is facilitated by the HELP Menu. Press HELP and then the appropriate key for
PULSE WIDTH. From the list of options select the pulse appropriate to the measurement. On returning to the trace the selected pulse width will be confirmed on the display,and the PULSE key will be inoperative.
TheoutputimpedanceoftheT631mustbesettomatch the impedance of the cableundertest.
7.9.1 MEASUREMENT OF RETURN LOSS (dBRL)
Through the HELP Menu select the IMPEDANCE to match the cable under test. Similarly select the PULSE WIDTH appropriate to the measurement. To minimise any possible mismatches the cable under test should be connected directly to the
unit using matched connectors.
T631 OPERATINGMANUAL 35
Radiodetection Ltd. 00901209-6
Page 36
Select the appropriate RANGE & PVF. Set the AMP to A1 (0 db) on the LCD. Ensure that the LH cursor is set to the start of the transmitted pulse or a particular pulse if an
interfeature measurement is required. Increase the AMP control until the desired reflection is obtained. Move the RH cursor to coincide with the start of the reflected pulse. Check the distance reading to confirm the location - note the distance. Press dBRL and the Return Loss in dB will be calculated and displayedon the LCD together
with the corresponding reflection co-efficient. The result is “Measured Return Loss” which requires correction for Pulse Attenuation. Press RS232 key to print results.Note: this option is not displayed on screen but is available.
Ensure printer lead is plugged into the instrument and printer and that the printer is ready to print (see section 9).
Press key 8 to return to normal operation.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 36
Page 37
Notes a) Note that a positive dB value (or mrho value <1000) indicates a return loss (i.e. the
reflectionissmallerthanthetransmitpulse).Undercertainconditionsitispossibleforthe reflectionto have a greatervoltageamplitude than the transmit pulse.In this case the dB value will be shown as negative and the mrho figure will be greater than 1000.
b)
If the return loss is outside the range of the instrumentthenXXX dB and XXX mρ will be displayed.
c) The position of the cursor prior to pressing dBRL is important as the signal level at that
point is used as the zero reference.Be sure to place the cursor on the flat section of the trace just before the pulse.
d) Whenmeasuring very smallreflections switching the filter incan improve the performance
of the return loss function.
7.9.2 MEASUREMENT OF PULSE ATTENUATION
This should be determined using a sample length of the same type of cable as that under test. Connect the cable directly to the T631 with the end shorted or open. Measure the Return Loss and the length. Since the cable ends provides a 100% (0dB)
reflection, the loss measured is entirely due to the pulse attenuation.
T631 OPERATINGMANUAL 37
Radiodetection Ltd. 00901209-6
Page 38
Divide the loss obtained by the cable length and obtain the pulse attenuation loss in dB per metre or per foot.
Multiply by the distance reading for the cable length noted in the Return Loss measurement to obtain the appropriate pulse attenuation.
Obtain the TRUE RETURN LOSS from: True Return Loss = Measured Return Loss – Pulse Attenuation.
Note: Pulse Attenuation (Cable Loss) is frequency dependent and will be higher for narrow pulse widths than wider pulse widths. The pulse width used for measuring pulse attenuation MUST be the same as that used to measure the Return Loss.
AlibraryofPulseAttenuation(CableLoss)fordifferentcabletypesfordifferentpulsewidthscan be compiled using the procedure detailed in “Measurement of Pulse Attenuation”.
7.10 FILTER
A low pass filter is available from the Help menu. It is used to reduce noise, especially on high gain settings.It should not be used with the 2 ns and 10 ns pulses.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 38
Page 39
8. USE OF THE MEMORY
8.1 STORING A TRACE
Obtain the desired LIVE trace using the method previously described in Section 7. Press the SAVE key and then select the desired Memorylocation by pressing the appropriate
key. The live trace is now stored in the selected memory location, together with the associated
machine settings. Any trace present in the memory will be overwritten and lost.
8.2 RECALLING A STORED TRACE
Press the RECALL key and then select the desired memory location. The stored trace will then be displayed on the LCD and memory location confirmed. The machine parameters are set in accordance with the stored trace, and cannot be altered.
8.3 COMPARING A LIVE AND STORED TRACE
Press the MODE key and select L & M. The displayed memory waveform will be the last one that was accessed by either SAVE or RECALL.
T631 OPERATINGMANUAL 39
Radiodetection Ltd. 00901209-6
Page 40
Recall the appropriate stored trace from memory as described earlier. The machine parameters are set in accordance with the stored trace. If not already connected to thecable under test, connect to the cable and observe that boththe
live trace and the recalled trace are overlaid. TheSHIFT control may be used toseparate the traces for easier identification and comparison.
The live trace moves relative to the stored trace.
8.4 DIFFERENCE BETWEEN A LIVE AND STORED TRACE
Press the MODE key and then select L – M. Recalltheappropriatestoredtracefrommemoryasdescribedearlier.The machine parameters
are set in accordance with the stored trace. If not already connected to the cable under test, connect to the cable and observe the trace
which is the difference between the live and recalled trace. ToviewtheLIVEtraceandrecalledtraceseparatelyuse the MODE keyandthenselectLIVE(L)
OR MEMORY (M) as required.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 40
Page 41
8.5 HELP MENU
When in the MEMORY mode, if the HELP menu is accessed, those keys that affect set parameters are disabled.
9. RS232 – PRINTING AND PC INTERFACE
Screen information and Memory contents can be transferred to a printer or PC via the RS232 interface.
9.1 PRINTING
AsuitableprinteristheSeikoDPU411 – 21Bavailablein220 V50 Hzor120 V60 Hzversions. Otherprintersmayrequireadifferentinterfacecableandprinterset-upanddetailsoftheRS232 interface is given in the specification Section 12.
T631 OPERATINGMANUAL 41
Radiodetection Ltd. 00901209-6
Page 42
9.2 PRINTER SET-UP SEIKO DPU 411-21B
The DIP switch on the printer must be set as shown below.
Connecttheinterfacecablesuppliedwith the printerbetween the printer and the RS232 port. Obtain the required display on the screen, either a live trace, a recalled memory trace, or an
L & M or L – M trace. Depress the RS232 key and select menu option 1 “PRINTER”. The display will then display
“PRINTING” as the data is transferred to the printer.
9.3 PC TRANSFER
A PC software kit, X600 TRACEability™ is available. This enables a cable database to be established.Data from all memories can be transmitted to, and received from, a PC.See the manual supplied with the software for full instructions.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 42
Page 43
10. HELP MENU MACHINE SET UP
1. Impedance 4. Dielectric 6. Pulse Width 93
1 PVF 1 2 ns 1
75
2 V 2 10 ns 2
50
3 V/2 3 30 ns 3 Help Menu 7 Help Menu 7 100 ns 4 Operation 8 Operation 8 300 ns 5
1200 ns 6
2. Filter 5. Cable Type Auto 7 Filter ON/OFF Telephone 1 Operation 8
Power 2
3. Units CATV 4 7. Help Text Feet 1 IBM 4 Explanation 1 Metres 2 Data 5 2 Time 3 Help Menu 7 3 Help Menu 7 Operation 8 4 Operation 8 6
7 8
8. Operation
T631 OPERATINGMANUAL 43
Radiodetection Ltd. 00901209-6
Page 44
11. ERROR MESSAGES
WhenswitchedontheT631performs a selftest.Ifitfailsthe followingmessageisdisplayed:
INTERNAL MALFUNCTION ERROR XX
PLEASE REFER TO MANUAL
Please contact your supplier and report the error number.
.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 44
Page 45
12. SPECIFICATION
All specifications in this section assume a PVF of 0.667. 1 metre is equivalent to 3.28 feet or 10 nanoseconds.
RANGES AND RESOLUTION:
Maximum Displayed
Range Nominal
Horizontal
Expansion (Zoom)
Resolution at
Max Zoom
3 m x1 13 mm
6 m x1 26 mm 12 m x1, x2 26 mm 25 m x1, x2, x4 26 mm 50 m x1, x2, x4, x8 26 mm
100 m x1, x2, x4, x8 52 mm 200 m x1, x2, x4, x8 0.1 m 400 m x1, x2, x4, x8 0.2 m
800 m x1, x2, x4, x8 0.4 m 1600 m x1, x2, x4, x8 0.8 m 3200 m x1, x2, x4, x8 1.6 m 6400 m x1, x2, x4, x8 3.2 m
12800 m x1, x2, x4, x8 6.4 m
T631 OPERATINGMANUAL 45
Radiodetection Ltd. 00901209-6
Accuracy Typically <0.5%
(48 m range and above)
Dielectric PVF 0.300 - 0.999
V 90 - 300 m/µs V/2 45 - 150 m/µs
Cursors
2
Pulse Characteristics:
Waveform
Amplitude
Pulse Widths
Source Impedance
Connector Protection
Mode Selection
Filter
Sine Squared Nominally 2.5 V into selected
impedance 2, 10, 30, 100, 300, 1200 ns half
height nominal 50, 75 or 93 BNC Female 250 V RMS, 0-60 Hz Auto or Manual 16 Mhz Low Pass Filter
Page 46
VERTICAL DEFLECTION:
SENSITIVITY
Gain Setting Signal for
Full Scale Deflection
A1 A2 A3 A4 A5 A6 A7 A8 A9 Aa Ab
16 V
8 V 4 V
2 V 960 mV 480 mV 240 mV 120 mV
60 mV 30 mV 15 mV
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 46
Page 47
T631 OPERATINGMANUAL 47
Radiodetection Ltd. 00901209-6
AUTO MODE PULSE SELECTION
Nominal Range (@PVF = 0.667)
Default Pulse
(ns)
Available Pulses
(ns)
3 m
6 m 12 m 24 m 48 m 96 m
192 m 384 m 800 m
1600 m
& above
2
2 10 10 30
100 100 300 300
1200
2, 10 2, 10 2, 10, 30 2, 10, 30 2, 10, 30, 100 2, 10, 30, 100 2, 10, 30, 100, 300 2, 10, 30, 100, 300, 1200 2, 10, 30, 100, 300, 1200 2, 10, 30, 100, 300, 1200
Page 48
Return Loss Measurement Range: 0 dB to > 30 dB
Display Modes Line (L)
Memory (M) Comparison (L & M) Difference (L-M)
Memories 15 - Stores trace displayed in Mode L.
Interface RS232 to printer or PC.
Connector 9 pin D Male. Configuration 4800 baud, no parity, 8 bits, 1 stop bit.
PIN FUNCTION DIRECTION
2 Received Data (RXD) Input 3 Transmit Data (TXD) Output 5 Ground 6 Data Set Ready (DSR) Input 7 Request To Send (RTS) Output
Display 240 x 128 pixel LCD
240 x 100 pixel waveform area
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 48
Page 49
Backlight LED with auto switch off (5 mins)
Keyboard Sealed membrane
Power Sources Battery - internal rechargeable NiCad provides 8 hours operation (excluding backlight).
Cells - 8 x R14
External DC:
Operating 12 to 20 V 0.25 A Charging 15 to 20 V 0.25 A Connector 2.1 x 5.5 x 9.5 mm plug centre positive (reverse polarity protection).
Dimensions 300 x 183 x 75 mm
Weight 2.5 kg (including batteries & carry bag)
Accessories Carry bag - soft weatherproof bag with shoulder strap and accessory pouch.
AC Adaptors - 120 V or 230 V 50/60 Hz (see separate specification). Test Lead - 3 metre 75 BNC/BNC + BNC/Crocodile clip + BNC/F Adaptor Operating Manual
Safety IEC 1010-1
T631 OPERATINGMANUAL 49
Radiodetection Ltd. 00901209-6
Page 50
AC Adaptors UK
230 V ±10% 50 Hz 90 mA Output 15 V DC 400 mA
North America 120 V ±10% 60 Hz 180 mA Output 15 V DC 400 mA
Europe 230 V ±10% 50 Hz 90 mA Output 15 V DC 400 mA
Environmental & Safety (Main Unit)
Safety Maximum permissible voltage
on input connector for safe operation
BS EN 61010-1 : 1993 and IEC 1010-1 : 1990 and Amendment 1 : 1992 30 V RMS
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 50
Page 51
Temperature Operating temperature:
Including batteries: 0°C to +50°C Excluding batteries: -5°C to +50°C
Storage temperature: Including batteries: -20°C to +50°C Excluding batteries: -20°C to +65°C
Recommended temperature limits to maximise battery life (for Varta RSH2K cells)
Damp Heat, Steady State
Damp Heat, Cyclic
Charging: +10°C to +35°C Discharging: -5°C to +45°C Storage: 0°C to +45°C
BS 2011, part 2.1 Ca : 1977 (IEC 68-2-3 : 1969) 40°C, 93% RH, 4 days
BS 2011, part 2.1 Db : 1981 (IEC 68-2-30 : 1980) 25°C, 95% RH, 12 hr 55°C, 93% RH, 12 hr 6 cycles
Low Air Pressure BS 2011 part 2.1 M : 1984 (IEC 68-2-13 : 1983)
Non operational: 150 mb 16 hours Operational: 533 mb 30 minutes
T631 OPERATINGMANUAL 51
Radiodetection Ltd. 00901209-6
Page 52
Random Vibration BS 2011 part 2.1 Fdb : 1973
5 to 150 Hz, 0.005g
2
/Hz 2 hours in each of 3 planes (in soft carry case)
Shock
Bump
BS EN 60068-2-27 : 1993 part 2, test Ea (IEC 68-2-27 : 1987) 50g, 11ms (in soft carry case)
BS EN 60068-2-29 : 1993 part 2, test Eb (IEC 68-2-29 : 1987) 40g, 6ms, 1000 bumps in each of 3 axes (in soft carry case)
Free Fall
Water and dust Protection
BS EN 60068-2-32 : 1993 part 2, test Ed (IEC 68-2-32 : 1975) 1m (in soft carry case)
BS EN 60529 (IEC 529 : 1989) To IP54
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 52
Page 53
13. PRODUCT SAFETY DATA
The T631 is a Pulse Echo TestSet that provides visual indication of cable faults. The T631 is tested and supplied in accordance with our published specifications,and when used in normal or prescribed applications within the parameters specified for electrical and mechanical performance, will not present any danger or hazard to health or safety, provided normal engineering and safety practices are observed.
Doubt relating to any aspect of usage of this instrument must be referred to Bicotest Limited.
13.1 QUALITY
Bicotest Limited is a BS EN ISO 9001 registered company.
13.2 POWER SUPPLY
a) The instrument is fitted with eight rechargeable NiCad batteries type R14 with capacity of
2 Ah. Replacement batteries must be of similar type, preferably Varta type RSH2K. A quickblowfusethatis not accessible to theoperatorisfittedtoprotectthebatterycircuit.
b) The instrument can also be powered with an external 12 to 20 V DC source 0.25 A as
outlined in specification (section 12). A quick blow fuse that is not accessible to the operator is fitted to offer protection.
c) The instrument can also be poweredwith an optional AC Adaptor (forfurther details see
specification, section 12).
T631 OPERATINGMANUAL 53
Radiodetection Ltd. 00901209-6
Page 54
13.3 ROUTINE SERVICING
ItisrecommendedthattheinstrumentisreturnedtoBicotestLimitedeverytwoyearsforservice and calibration checks. The instrument is fitted with eight rechargeable NiCad batteries type R14 with capacity of 2 Ah. Replacement batteries must be of a similar type, preferably Varta type RSH2K.
13.4 COMPOSITION/TOXIC HAZARDS
Under normal conditions of use, storage and handling, the T631 presents no toxic hazards, however, in certain circumstances the following could apply:
a)
Incineration
Theinstrument houses NiCad batteries andthesemustnot be incinerated.Additionallysome
of the electronic components included in the assembly are constructed with resins and other chemicals which produce toxic fumes during incineration. It is required that the instrument is submitted to the correct authority for disposal in accordance with local by-laws.
b)
Acidic or Caustic Compounds
Some of the electronic components included in the assembly, particularly the electrolytic capacitors contain acidic compounds. In the event of any damaged items coming into contact with the skin, the affected area should be washed with clean, cold water. In the event of eye contamination, thoroughly irrigate with recognised eyewash and seek urgent medical assistance.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 54
Page 55
c)
Physical Damage
Some of the components used in the assembly may contain very small quantities of toxic materials.Thereexistsaremotepossibilitythatphysicallydamagedcomponentsmaypresenta toxic hazard. As a general precaution avoid unnecessary contact with damaged electronic components and arrange fordisposal in accordance with local legislation that may currentlybe in force.
13.5 TRANSPORT AND HANDLING
The instrument is supplied in a soft carry bag which offers adequate protection under normal working conditions. For transportation over long distances the instrument should be suitably packed in a box filled with shock absorbing material such as bubblepack or corrugated cardboard.
13.6 STORAGE
The instrument should be stored in a dry, clean environment. The NiCad batteries will self discharge within three months and it will be necessarytofully charge the batteries which takes 14 hours using the recommended AC adaptor outlined in specification (section 14).
No hazard is anticipated during storage.
13.7 DISPOSAL
When disposing of electrical and electronic equipment or packaging materials, exercise precautions that are required by local legislation.If in doubt, contact the local authority.
T631 OPERATINGMANUAL 55
Radiodetection Ltd. 00901209-6
Page 56
13.8 SAFE USE
The T631 is designed to be used by suitably trained personnel following the procedures and instructions described in this operating manual. Additionally, the following points should be noted:
a)
Personal Protection/Protective Clothing
NotnecessaryforoperatingtheT631,providingthatnormalsafeworkingpracticeisobserved.
b)
Working Environment
No special precautions are needed for operating the T631. Appropriate precautions must be observedforpotentiallyhazardous,working environments such as construction site installations, electricity substations, explosive atmospheres, etc.
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 56
Page 57
T631 OPERATINGMANUAL 57
Radiodetection Ltd. 00901209-6
Other products available from Bicotest:
VIXXON™ S3000 Intermittent LV Fault Location System
The unique system solution to intermittent LV Faults
BOXXER™ S4000 Power Cable Fault Location System
The system solution to intermittent, flashing and high resistance HV faults
AXXIS™ S5000 Sheath Fault Location System
The cost-effective, system solution to cable sheath faults
P109 and P122 DC and ACTest Sets
HV test sets for proof testing on cable and equipment
VICTORT120A
For safe, hand-held fault sectioning on medium voltage networks
M225 Cable and Pipe locator
For location and route tracing of cable and pipes
LEXXI™T810
A universal hand-heldTDRforuseonalltypesof metallic cable,includingliveACwhenusedwiththeblockingfilterlead
T600 SeriesTDRs
World class, user-friendly cable fault locators which feature interface with PCs via X600TRACEability software
T631F,T620FS,T620FD Blocking Filters
Single and dual input LV mains blocking filters for use with T600 series TDRs
T272 High Resistance Cable Fault Locator
For location of high resistance insulation faults of up to 200 megohms
Please contact our sales Office for further information on the above systems.
Page 58
Radiodetection Ltd. 00901209-6
T631 OPERATING MANUAL 58
The Bicotest Comprehensive
HV and LV Fault Location Process
TDR
<300 Ohms
>300 Ohms
Intermittent
Fault
Fault Triggers
DAU
Establish
Cable Length
Prelocate
Fault
Prelocate
Fault
Pinpoint Fault
Pinpoint Fault
Locate Fault
Locate Fault
Sheath
Fault
Resistance HV
Faults
Intermittent,
Flashing or High
TDR
VIXXON S3000 SP
TM
VIXXON S3000 SP
TM
BOXXER
S4000
TM
BOXXER
S4000
TM
AXXIS
S5000
TM
AXXIS
S5000
TM
LV
Fault Diagnosis
HV Fault Diagnosis
BOXXER S4000
TM
LV & HV Fault Diagnosis AXXIS S5000
TM
T272
High Resistance
Fault Locator
LINXX
B110
TM
FIX
FAULT
Locate Fault
Loading...