Racal Dana 1990 Service and user manual

Page 1

The Electronics Group

Page 2

PUBLICATION NUMBER TH 9099 ISSUE 3.5.89

UNIVERSAL COUNTER

RACAL-DANA

RACAL-DANA INSTRUMENTS INC.

4 Goodyear Street, P.O. Box C-19541, Irvine, Ca 92713, USA. Telephone: (714) 859-8999, TWX: 910-595-1136, TLX: 678-341

RACAL-DANA INSTRUMENTS LTD.

Hardley Industrial Estate, Hythe, Southampton, England SO4 6ZH Telephone: Southampton (0703) 843265 Telex: 47600 Cables: Technico, Hythe, Southampton

RACAL-DANA INSTRUMENTS S.A.

18 Avenue Dutartre, 78150 Le Chesnay, France Telephone: (3) 955 88 88. TLX: 697 215

RACAL-DANA NSTRUMENTS GmbH

Frankenforsterstrasse 21, 5060 Bergisch Gladbach 1, Federal Republic of Germany. Telephone: 02204 60081. TLX: 8878428

RACAL-DANA INSTRUMENTS ITALIA sri

Via Mecenate 84/A, 20138 Milano Ml, Italy. Telephone: (02) 5062767/5052686/503444. TLX: 315697

PUBLICATION DATE: DECEMBER 1987

Copyright © 1987 by Racal-Dana Instruments Ltd. Printed in England. All rights reserved. This book or parts thereof may not be reproduced in any form without written permission of the publishers.

Page 3

LETHAL VOLTAGE WARNING

VOLTAGES WITHIN THIS EQUIPMENT ARE SUFFICIENTLY HIGH TO ENDANGER LIFE.

COVERS MUST NOT BE REMOVED EXCEPT BY PERSONS QUALIFIED AND AUTHORISED TO DO SO AND THESE PERSONS SHOULD ALWAYS TAKE EXTREME CARE ONCE THE COVERS HAVE BEEN REMOVED.

Page 4

RESUSCITATION

TREATMENT OF THE NON-BREATHING CASUALTY

or switch

SHOUT FOR HELP. TURN OFF WATER, GAS OR SWITCH OFF ELECTRICITY IF POSSIBLE

Oo this immediately. If not possible

don't waste time searching for a tap

REMOVE FROM DANGER 72 WATER, GAS, ELECTRICITY FUMES, ETC.

Safeguard yourself when removing casualty from hazard If casualty still in contact with electricity, and the supply cannot be isolated, stand on dry non-conducting material (rubber mat, wood, linoleum). Use rubber aloves, dry clothing, length of dry rope or

wood to pull or push casualty away from the hazard.

ORSTRUCTION TO BREATHING

If casualty is not breathing start ventilation at once.

SEND FOR DOCTOR AND AMBULANCE
DOCTOR AMBULANCE HOSPITAL Nearest First Aid Post
TELEPHONE TELEPHONE TELEPHONE

Crown Copyright; reprinted by Racal Group Services Ltd. with the permission of Her Majesty's Stationery Office

Page 5

Page 6

HANDBOOK AMENDMENTS

Amendments to this handbook (if any), which are on coloured paper for ease of identification, will be found at the rear of the book. The action called for by the amendments should be carried out by hand as soon as possible.

Page 7

TABLE OF CONTENTS

Paragraph Title Page
SECTION 1 TECHNICAL SPECIFICATION 1-1
SECTION 2
1
2
3
4
6
8
9
10
11
12
13
15
16
17
18
19
20
21
22
23
24
25
26
26
26
26
27
28
29
30
34
GENERAL DESCRIPTIONIntroductionMeasurement FunctionsFrequency A FunctionPeriod A FunctionTime Interval FunctionTotal A by B FunctionPhase A rel B FunctionTotal A FunctionSignal Input ChannelsLow-Pass FilterNull FunctionDelay FunctionDelay FunctionDol Average FunctionSpecial FunctionsError IndicationDisplay FormatHold FeatureResolution and Gate TimeStandby ModeInitializationOptions AvailableFrequency Standards (04X Options)Reference Frequency Multiplier (Option 10) 2-1
2-1
2-1
2-1
2-2
2-2
2-2
2-2
2-2
2-2
SECTION 3
1
3
5
6
9
11
13
14
14
15
16
PREPARATION FOR USE Unpacking Power Supply AC Line Voltage Setting Line Fuse Power Cord Functional Check Frequency Standard Trigger Level Output Preparation For Use With the GPIB Introduction Connection to the GPIB Address Setting and Display 3-1
3-1
3-1
3-2
3-2
3-2
3-3
3-3
3-3
3-3
3-3
3-4
3-4

Contents 1

1990 FD 477

Page 8
18
21
25
28
29
31
33
34
35
35
35
37
40
43
46
48
49
51
GPIB Check
3-6
3-7
3-8
3-9
3-10
3-10
3-10
3-10
3-10
3-11
3-12
3-13
3-14
3-15
3-16
3-17
SECTION 4 OPERATING INSTRUCTIONS 4-1
4-1
4-1
2 Front Panel Items 4-1
3 Rear Panel Items 4-7
4 Frequency Measurement 4-10
5 Period Measurement 4-11
6 Time Interval Measurement 4-12
7 Totalize Measurement 4-13
7 Total A by B (Electrical 4-13
8 Total A (Manual) 4-15
9 Phase Measurement 4-16
10 Ratio Measurement 4-17
11 Time Measurement 4-18
13 Setting and Measuring the Trigger Level 4-19
14 Instrument Checks 4-20
15 Display Resolution 4-20
20 Setting the Display Resolution 4-21
21 4-21
23 4-22
27 4-23
27 Use of the Delay 4-23
29 Chapting the Delay 4-24
30 Unanging the Delay 4-24
32
22
Avenage of 100 Deadings 4-24
38 Special Functions 4-24
1_25
38 4-25
40 4-25
4 1 Frror Codes 25
4-25
42 Clearing the Error Codes 4-26
Page 9
43
43
44
46
47
49
50
Using the Battery Pack Option
SECTION 5
1
2
3
8
9
10
11
13
14
16
20
22
24
27
28
29
30
32
OPERATION VIA THE GPIB5-1Introduction5-1GPIB Operating Modes5-1Talk Only Mode5-1Addressed Mode5-2Data Output Format5-2Deferred Commands and Immediate Commands5-2Service Request5-4Status Byte5-4Explanation of Response to Interface Messages5-6Local Lockout5-6Device Clear and Selected Device Clear5-8Go to Local5-8Untalk and Unlisten5-8Input Command Codes5-8Special Function Register5-14
SECTION 6
1
3
6
6
11
27
27
32
39
39
42
46
46
46
50
54
54
56
63
63
63
69
83
83
87
PRINCIPLES OF OPERATION6-1Introduction6-1Functional Systems6-1The Channel A and Channel B System6-3Functional Description6-4The Measurement System6-5Functional Description6-7The Display System6-7Functional Description6-8Circuit Description6-8The Keyboard System6-9Functional Description6-10Circuit Description6-11The Microprocessor System6-12Functional Description6-12Circuit Description6-13The Standby and IRQ System6-14Functional Description6-15The Power Supply System6-17Functional Description6-15The Power Supply System6-17Functional Description6-17Functional Description6-17

1990

FD 511

Page 10
93
93
96
104
104
108
119
119
123
125
128
133
135
142
142
142
163
The Frequency Standard System6-18Functional Description6-19Circuit Description6-19The Reference Frequency Multiplier (Option 10)6-20Functional Description6-20Circuit Description6-21The GPIB Interface (Option 55)6-23Introduction6-23Address Setting and Recognition6-23Writing to the Bus6-23Writing to the Bus6-24Serial Poll6-25Data Transfer Between Microprocessors6-25The Battery Pack (Option 07)6-26Functional Description6-26Technical Description6-30
SECTION 7 MAINTENANCE 7-1
1 Introduction
2 Test Equipment Required 7-1
4 Dismantling and Reassembly 7-1
4 Introduction 7-1
5 Instrument Covers 7-3
7 Front Panel 7-3
9 Rear Panel
11 Display Board
13 Special Functions for Diagnostic Purposes
14 Special Function 21
15 Special Function 70
16 Special Function 71
17 Fault Finding 7-5
10 Sotting Un After Dengin 7-23
10 Introducion 7-23
21 Channel A Input Suctom 7-23
25 Channel R Input System 7-25
20 Enguancy Doublan 19,1238 (Ontions OAA and OAR) 7-27
29 Poference Erequency Multiplier 19-1164 (Option 10) 7-27
33 Reference frequency multiplier 19-1104 (option 10) 7-27
Rattony Back 11-1625 (Option 07) 7-28
35
45
Internal Enguency Standard, Routine Calibration 7-31
45 Standard Oscillator and TYCO Option 7-31
43 Overed Occillator Options OAA and OAB
47 Overall Specification Check 7-33
40 Introduction 7-33
40
51
Channel A Sonsitivity DVD
51 Channel R Sensitivity DVD 7-36
57 Attenuator DVD 7-22
50 Filton DVD 7-20
61 Time Interval DVD 7 A0
63 Thispan Loval DVD 7 11
65 ΔC/DC Coupling PVP 7_42
05 AU/DU UUUPINNY EVE

Contents 4

Page 11
67 External Standard Ir
(Peference Option 0)
nput Sensitivity PVF ) 7 4 2
69 10 MHz Standard Outp but Level PVP ••• •• /-43
(Reference Option 02 2 only) • ••• 7-44
11 Internal Frequency S standard PVP •. • /-45
SECTION 8 PARTS LIST AND CIRCUIT DI IAGRAMS 8-1
Parts List : BNC Mounting Board 19-3003 ••• 8-1
Parts List : Display Assembly 19-3002 . 8-2
Parts List : Motherboard Assembly 19-300 04 8-4
Parts List : GPIB Assembly 19-1146 •••••••• 8-14
Parts List : Reference Frequency Multip lier Assembly 19-116 8-16
Parts List : Reference Frequency Doubler Assembly 19-1238 8-19
Parts List : Oscillator Assembly 19-1208 3 • • • ••• 8-21
Parts List : Battery Pack Assembly 11-16 525 • • • ••• 8-22
Internal Lavout: Fig. 1
Display Assembly 19-3002: Component Lavout Fig. 2
Circuit Diagram Fia. 3
Motherboard Assembly 19-3004: Component Layout Fig. 4
Circuit Diagram Fig. 5
Circuit Diagram • • Fig. 6
Circuit Diagram Fig. 7 -
GPIB Assembly 19-1146: Component Layout • • Fig. 8
1 Circuit Diagram Fig. 9
Reference Frequency Multiplier 19-1164: Component Layout • • Fig. 10
Circuit Diagram • ••• Fig. 11
Reference Doubler 19-1238: Component Layout ••• •• • ••• Fig. 12
0 Circuit Diagram ••• •• • ••• Fig. 13
Dettony Roand Acomply 19-1208: ••• •• • ••• ⊢1g. 14
Dattery Dudru Assimply 19-1203: ••• •• rig. 15
SWILLI DUdru Assembly 19-1242:
Rattony Dack Assembly 11 1625:
Component Layout ••• •• • ••• rig. 17
Interconnections: CITCUIL DIAYRAM • • • ••• riy. ⊥/
10
riy. TQ
Page 12
Table Title Page
3.1
3.2
GPIB Connector Pin Assignment Address Switch Settings 3-4
3-5
4.1
4.2
Resolution and Gate Time
Error Codes
4-21
4-25
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
5.10
5.11
5.12
5.13
5.14
5.15
5.16
5.17
5.18
Output Message Format
Function Letters
Status Byte Format
Response to Bus Messages
Permitted Terminators
Instrument Preset Code
Measurement Function Codes
Input Control Codes
Measurement Control Codes
Store and Recall Codes
Store and Recall Codes
Numerical Input Format
Numerical Input Ranges
Resolution Selection
Special Function Codes
Special Functions
Output Message Format
Service Request Codes
Alphabetic List of Command Codes
5-3
5-4
5-5
5-7
5-9
5-9
5-10
5-10
5-10
5-11
5-12
5-13
5-14
5-14
5-14
5-15
5-15
5-16
6.1 Control Signals 6-8
7.1
7-2
7-3
7-4
Test Equipment Required Special Functions Special Function Standard Accuracy Special Frequency Standard Accuracy Special Function Standard Option (Standard Oscillator and TXCO Option) 7-2
7-4
7-29
7-32
7-5 Internal Frequency Standard Accuracy 7-33
/ -0
7 7
Channel A Sensitivity 1-35
/ - / Channel A Sensitivity 1-30
/-ð 1-51
/-9 /-3/
/-10 Lime Interval PVP Readings /-41
/-11 Irigger Level Output Voltages /-41
/-12 Internal Frequency Standard Accuracy (Standard Oscillator and TXCO Option) 7-45
7-13 Internal Frequency Standard Accuracy 7-46

Contents 6

Page 13

LIST OF ILLUSTRATIONS

Fig.
3.1
3.2
TitlePagePCB to Motherboard ConnectionConnector Alignment3-15
4.1 Use of Stop Circuit Delay 4-23
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
6.12
6.13
Functional Block Diagram6-2The Channel A and Channel B System6-3The Measurement System6-6Basic Recipromatic Counting Technique6-6The Display System6-9The Keyboard System6-11The Microprocessor System6-12The Standby and IRQ System6-14The Power Supply System6-179423 and 9444 Oscillators6-18The Reference Frequency Multiplier6-20Pulse Generator Waveforms6-21The Battery Pack6-27
7.1
7.2
7.3
7.4
7.5
7.6
7.7
7.8
7.9
7.10
7.10
7.11
7.12
7.13
7.14
Fault Finding Flowchart Part 17-7Fault Finding Flowchart Part 27-9Fault Finding Flowchart Part 37-11Fault Finding Flowchart Part 47-13Fault Finding Flowchart Part 57-15Fault Finding Flowchart - GPIB Part 17-17Fault Finding Flowchart - GPIB Part 27-19Fault Finding Flowchart - GPIB Part 37-21Location of R33, R72, and PL157-23Connection for Channel A Input System Adjustment7-24Location of R36, R73, and PL157-25Connections for Channel B Input System Adjustment7-26Connections for Internal Frequency Standard Adjustment7-31(Standard Oscillator and TXCO Option)7-31
7.15
7.16
Location of Standard Oscillator OSC1
7.17
7.18
7.19
7.20
7.21
7.22
7.23
7.24
7.25
7.26
7.27
7.28
Connections for Channel A Sensitivity PVP
r • L J (Ovened Oscillator Options) 7-4

1990 FD 511

Contents 7

Page 14

TECHNICAL SPECIFICATION

Freg.)

Input Characteristics Frequency Range DC to 120MHz DC coupled 10Hz to 120MHz AC coupled Input B DC to 100MHz DC coupled 10Hz to 100MHz AC coupled Sensitivitv Sine Wave 25mVrms DC to 100MHz 50mVrms to 120MHz Pulse 75mV p-p. 5ns min. width Dynamic Range 75mV to 5V p-p to 50MHz (x 1 attenuation) 75mV to 2.5V p-p to 100MHz 150mV to 2.5V p-p to 120MHz Signal Operating Range x 1 attenuation x 10 attenuation ±25V Input Impedance (nominal) Separate Mode (x1/x10) 1 Mohm //≪45pF Common Mode (x1) 500kohm //≪55pF Common Mode (x10) 1 Mohm //≪55pF Maximum Input (without damage) 260V(DC + ACrms), DC to 2kHz v1 attenuation Decreasing to 5Vrms, at 100kHz and above x10 attenuation 260V(DC + ACrms), DC to 20kHz Decreasing to 50Vrms at 100 kHz and Coupling Low Pass Filter 50kHz nominal (Input A selectable) Trigger Slope Attenuator v1 or v10 nominal Trigger Level Range Manual x1 attenuation ±2.8V typical x10 attenuation +28V typical

Trigger Level Output

Range Accuracv* x1 (sep/common) x10 (separate mode)

±10mV ± 5% of V output ±100mV ± 7% of V output

±2.8V typical

Impedance

10kohm nominal

Measurement Modes
Frequency A

Range Digits Displayed LSD Displayed (Hz)

DC to 120MHz 3 to 8 digits plus overflow E x 10-D (D = No. of digits, E = Freq. rounded up to next decade)*

Resolution *(Hz) Accuracy *(Hz) Time Interval Range Single Averaged

Input Separate

Trigger Slopes

Resolution* (sec) Accuracy* (sec)

Input A START Input B STOP +ve or -ve Selectable START and

Input A START and STOP

100ns to 8 x 105 sec

10ns to 8 x 10⁵ sec

100ns min. (10ns with averaging)

±LSD ± (Trig. Error* x Freq.)/Gate

± Resolution ± (Timebase Error x

±LSD ± 5ns rms ± Trig Error*

± Resolution ± (Timebase Error x TI) ± Trigger Level Timing Error*

Time Delav

Available on Time Interval and Totalise

Range 200µs to 800ms nominal
Step Size 1ms nominal. (Extra step to 200µs)
Accuracy ±0.1% setting ±50µs

Derind A

Range 8.3ns to 1.7 x 10 3 sec
Digits Displayed 3 to 8 digits plus overflow
LSD Displayed (sec) P x 10 -D (D = No. of digits, P = Period rounded up to next decade)*
Resolution* (sec) ±LSD ± (Trig. Error* x Period)/Gate
Time
Accuracy* (sec) ±Resolution ± (Timebase Error x
Period)
Ratio A/B

Specified for higher frequency applied to Input A

Range

LSD Displayed (for 4-8 digits selected)

\left(\frac{10}{\text{Freq. B x Gate Time}}\right) , rounded to nearest decade*

DC to 100MHz on both inputs

Resolution*

±LSD ± (Trig. Error B*/Gate Time) x Ratio

Accuracy*

+ Resolution

*see definitions **A differential delay between amplifiers

Page 15
Totalise A by B and Total A
Gate Time
Input
Start/Stop
Range
Maximum Rate
Minimum Pulse Width
Accuracy
Input A
Electrical (Input B) or Manual
10 12 - 1 (8 most significant digits
displayed)
10 8 events/sec.
5ns min. at trigger points
±1 count
(Frequency, Period
and Ratio modes)
Automatically determined by
resolution selected.
(Range 1msec-10sec)
Resolution
Selected (Digits) Gate Time*(s)
8 10
7 1
6 0.1
5 0.01
4,3 0.001
Phase (A rel. to B) Display 8 – digit, high brightness, 14mm LED
Range 0.1° to 360° exponent digit
LSD Displayed Normal (averaged)
1° (0.1°) to 100kHz
10° (1°) to 1MHz
100° (10°) to 5MHz
Power Requirements
Voltage (AC)
90–110V
103-127V
193-237V externally selectable
Resolution* (degrees) ±LSD ± (TI Resolution/Period A) x 207-253V
360° Frequency 45-440Hz
Accuracy (degrees) Rating 35VA Max
Time Operating
Temperature Range
0° to +50°C
(0° to +40°C with battery pack)
Start/Stop
Bange
Manual
40ms to 8 x 10 5 sec
Storage Temperature
Range
-40°C to +70°C (-40°C to +60°C with battery pack)
Resolution ± 40ms Environmental Designed to meet MIL-T-28800,
Accuracy ± Resolution ± (Timebase Error x DEF-STD 66/31 and IEC 68
Time) Safety Designed to meet the requirements of
IEC348 and follow the guidelines of
UL1244
100 Average RFI/EMC MIL-STD-461B
Function Displays average value of 100 measurements Weight Net 3.6kg (8lb) excl. battery
Averaging Time 2.5 sec. + (100 x single measurement
time)
Shipping 5.5kg (12lb) excl. battery
8.75kg (19.3lb) incl. battery
Null Dimensions 331 x 218 x 101mm
(13.03 x 8.58 x 3.98 ins)
Available on all measurements except Phase and Check Shipping Dimensions 430 x 360 x 280mm
Function Displays (Result – Null) (16.9 x 14.2 x 11.0 ins)
Entry Range ±1 x 10 -9 to ±1 x 10 10 to 8 significant figures

Single Cycle (Hold) Enables a single measurement to be initiated and held

General

Internal Timebase

rystal Controlled
Frequency 10MHz
Aging 2 x 10 -6 in the first year
Temperature Stability ±1 x 10 -5 over the range 0 to +50°C
Adjustment Internal
Options
Option 02

-Frequency Standard Input/Output

Frequency Standard Output
Frequency 10MHz
Amplitude TTL levels giving approx. 1Vp-p into 50 ohms
Impedance 90 ohms nominal
Reverse Input ±15V max

*see definitions

Page 16
External Standard Input
Frequency 10MHz (see also Option 10 for other frequencies)
Single Amplitude 300mVrms min.
(Sine Wave) 10Vrms max.
Impedance 1 kohm nominal at 1V p-p
Option 04T
Temperature Compension sated Crystal Oscillator
Frequency 10MHz
Aging Rate 3 x 10 -7 /month
1 x 10 -6 in the first year
Temperature Stability ± 1 x 10 -6 over the range
0 to 40°C (operable to +50°C)
Adjustment Via rear panel
Option 04A
Ovened Oscillator
Frequency 10MHz
Aging Rate 3 x 10 -9 /day averaged over 10 days after 3 months continuous operation
Temperature Stability ± 3 x 10 -9 /°C averaged over range 0° to +45°C (operable to +50°C)
Warm Up Typically ±1 x 10 -7 within 6 minutes
Adjustment Via rear panel
Option 04B
High Stability Ovened Oscillator
Frequency 10MHz
Aging Rate 5 x 10 -10 /day averaged over 10 days after 3 months continuous operation
Temperature Stability ±6 x 10 -10 /°C averaged over range 0° to 50°C
Warm Up ±1 x 10 -7 within 20 minutes
Adjustment Via rear panel

Option 07

Rechargeable Battery Pack and External DC Operation
Battery Type Sealed lead-acid cells
Battery Life Typically 5.75 hours at +25°C
(24 hrs on standby)
Battery Condition Display indicates battery low
External DC 11-16V via socket on rear panel
(-ve ground, not isolated)
Option 10
Reference Frequency Multiplier
Input Frequency 1,2,5 or 10MHz (±1 x 10⁻⁵)
Input Amplitude
and Impedance As for external standard input

1990 FD 477

Opti

GPIB

Cont

Outc

IEEE Subs

Hand

Read

Defi

I SD

(Leas

on 55
Interface Designed to comply with IEEE-STD-
488 (1978) and to conform with the
guidelines of IEEE-STD-728 (1982)
rol Capability All functions and controls
programmable except power on/off,
trigger levels and standby/charge
ut Engineering format (11 digits and exponent)
-STD-488
sets
SH1, AH1, T5, TE0, L4, LE0, SR1, RL1,
PP0, DC1, DT1, C0, E2
lshake Time 250µs to 1ms/character dependent or message content.
l Rate Typically 22/sec dependent upon
measurement function
nitions
t Significant Digit) In Frequency and Period modes
display automatically upranges at 1.1 >
decade and downranges at 1.05 x
Accuracy and Resolution Trigger Error =

(seconds)

Expressed as an rms value.

\left(\frac{(e_i^2 + e_n^2)}{s^2}\right) (rms) where e. = input amplifier rms noise (typically 200µVrms in

120MHz bandwidth) en = input signal rms noise in 120MHz bandwidth S = Slew rate at trigger point V/sec Suffix 1 denotes START edge Suffix 2 denotes STOP edge In Frequency A and Period A, triggering is always on positive aoina edae

Trigger Level Timing Error
Trigger Level Timing Error (Seconds) = 0.035
typically = 0.018 \left(\frac{1}{S1} - \frac{1}{S2}\right)

S1 = Slew rate on START edge V/sec. S2 = Slew rate on STOP edge V/sec.

Trigger level output accuracy is referenced to the centre of the hysteresis band.

Gate Time

The nominal gate time indicated is set by the resolution selected in Frequency, Period. Ratio and Check modes. It is the value which is used in the calculation of LSD and Resolution. The true gate time will be extended from this value by up to one period of the input signal(s) on Frequency A, Period A and Ratio A/B.

Supplied Accessories

Power Cord Spare Fuse Operator's Manual

Page 17

Ordering Information

1990 120MHz Universal Counter
Options and Accessories
02 Frequency Standard Input/Output 11-9000
04T** ТСХО 11-1713
*04A** Oven Oscillator 11-1710
*04B** High Stability Oven Oscillator 11-1711
07† Battery Pack 11-1625
*10 Reference Frequency Multiplier 11-1645
55† GPIB Interface 11-9001
60 Handles 11-1730
60A Rack Mounting Kit (Fixed, Single) 11-1648
60B Rack Mounting Kit (Fixed, Double) 11-1649
61 Hard Carrying Case 15-0773
61M Protectomuff Case 15-0736
65 Chassis Slides (incl. Rack Mounts) 11-1716
Telescopic Antenna 23-9020
High Impedance Probe (1MΩ) 23-9104

* Option 02 must also be purchased with these options

** Only one frequency standard may be fitted at any one time. The standard reference will be supplied unless option 04T, 04A or 04B is specified

+ The battery pack and GPIB options cannot both be fitted

1990 FD 477

Page 18
INTRODUCTION

1 The Racal-Dana universal counter Model 1990 is a microprocessor controlled instrument using recipromatic measurement techniques. The instrument offers a comprehensive range of functions and easy to use controls.

MEASUREMENT FUNCTIONS
Frequency A Function

2 The Frequency A function is used to measure the frequency of the signal applied to the channel A input. A resolution of seven digits is available with a one-second gate time.

Period A Function

3 The Period A function is used to measure the period of the waveform applied to the A channel input. A number of periods, depending upon the resolution (and therefore the gate time) selected, are measured, and the average value is displayed.

Time Interval Function

  • 4 The Time Interval function is used to make single-shot measurements of the time interval between:
    • (1) An event occurring at the channel A input and a later event at the channel B input (using separate input channels).
    • (2) Two events occurring at the channel A input (using a common input channel).
  • 5 The arming of the stop circuit can be delayed for a time set by the operator. This prevents the measurement interval being stopped prematurely by spurious pulses, such as those caused by contact bounce.
Page 19
Total A by B Function

  • 6 The Total A by B function permits events occurring at the channel A input to be totalized. The counting interval is controlled by electrical start and stop signals applied to the channel B input, where alternate edges start and stop the measurement.
  • 7 Delayed arming of the stop circuit is available in the Total A by B mode to prevent spurious triggering.
Phase A rel B Function

8 The Phase A rel B function is used to measure the phase difference between the waveform applied to the A channel input and that applied to the channel B input. The phase difference is displayed in degrees, and indicates the phase lead at the channel A input.

Ratio A/B Function

9 The Ratio A/B function is used to measure the ratio of the frequency applied to the channel A input to that applied to the channel B input.

Total A Function

10 The Total A function totalizes events occurring at the Channel A input. The counting interval is controlled by successive operations of the front panel HOLD (start/stop) key.

Time Function

11 The time function is used to measure the time interval between successive operations of the front panel HOLD (start/stop) key, ie. stop watch.

CHECK FUNCTION

12 With the Check function selected a number of functional tests of the instrument's circuits can be made without the use of additional test equipment. Although these tests do not check the instrument's performance to its published specification, they can be used to verify that the equipment is operating correctly following receipt or transportation to a new location. A suitable functional check procedure is given in Section 3.

SIGNAL INPUT CHANNELS

13 Signal input channels A and B are fully independent, but provision is made for connection of the signal at the channel A input into both channels. When this is done, the channel B input socket is isolated from channel B.

1990 FD 477

Page 20

  • 14 Each channel is provided with independent controls to permit the selection of:
    • (1) AC or DC input coupling.
    • (2) X1 or X10 input attenuation.
    • (3) Positive- or negative-slope trigger.
    • (4) Manually-set input trigger level.

The trigger levels are set by the front panel potentiometers which also have a set zero position.

The trigger voltage levels in use are also available at pins mounted on the front panel of the instrument. The input trigger voltage range is typically ±2.8 V. The voltage should be multiplied by 10 when the attenuator is selected.

LOW-PASS FILTER

15 An internal low-pass filter can be introduced to reduce the bandwidth of channel A to 50 kHz (nominal).

NULL FUNCTION

16 With the NULL function active the instrument displays the difference between the measured value and the value held in the internal NULL store.

DELAY FUNCTION

17 In Time Interval and Total A by B, the stop circuit can be delayed. With the delay function active, the stop circuit is prevented from being triggered prematurely by spurious signals.

100 AVERAGE FUNCTION

18 Enabling this function increases the resolution by taking 100 measurements and displaying the averaged value. The signal must be repetitive and asynchronous with the counter standard.

1990 FD 477

Page 21
SPECIAL FUNCTIONS

19 A number of special functions are available to the operator. These provide test procedures and operating facilities additional to those available by operation of the front panel controls. Details are given in Section 4 of this manual.

ERROR INDICATION

20 Certain errors in the operation of the instrument will result in the generation of error codes, which will be displayed. Details are given in Section 4 of this manual.

DISPLAY FORMAT

21 The display uses an engineering format, with an eight digit mantissa and one exponent digit. Overflow of the most significant digits can be used to increase the display resolution.

HOLD FEATURE

22 The hold feature allows readings to be held indefinitely. A new measurement cycle is initiated using the RESET key.

RESOLUTION AND GATE TIME

23 In the Total A by B, Total A and Time modes, the counting interval (gate time) is controlled by the time interval between the start and stop signals at the channel B input, or between successive operations of the HOLD key. In the Frequency A, Period A and Ratio A/B modes, the gate time is determined by the display resolution selected. In Phase mode, the gate time is fixed and the display resolution is determined by the input signal frequency. Details of the relationship between gate time and resolution for each measurement mode are given in Section 4 of this manual.

Page 22
STANDBY MODE

24 When the instrument is switched to standby, the internal frequency standard continues to operate but the measuring circuits are switched off. If the battery pack option is fitted and an external power supply is connected, the battery is charged at the full rate.

INITIALIZATION

25 When the instrument is first switched on, or when it is initialized via the GPIB, it is set to the following conditions:

Measurement Function FREQ A
Display Resolution 6 digits
Channel A and B Inputs AC coupling
Positive-slope trigger
LF filter disabled
Common input disabled
Delay
Delay Store
Disabled
1 ms
Null Function
Null Store
Disabled
O
100 Avg Disabled
Hold Disabled
Special Functions Off
OPTIONS AVAILABLE
Frequency Standards (04X Options)

26 A wide range of internal frequency standard options is available. The technical specifications are given in Section 1 of this manual. The frequency standard can be changed, if required, by the customer: instructions are given in Section 3.

Reference Frequency Multiplier (Option 10)

27 The reference frequency multiplier is an internally-mounted, phaselocked multiplier, which permits the use of external frequency standard signals at 1 MHz, 2 MHz, 5 MHz or 10 MHz. The multiplier can be fitted by the customer: instructions are given in Section 3.

1990 FD 477

Page 23

Reference Input and Output (Option 02)

28 The instrument may be operated using an external frequency standard. The instrument will operate from the external standard, in preference to the internal standard, whenever the signal at the EXT STD INPUT socket is of sufficient amplitude. It will revert to operation from the internal standard automatically if the external standard is removed. The 10 MHz STANDARD OUTPUT connector provides a signal derived from whichever standard is in use at the time.

GPIB Interface (Option 55)

29 An internally mounted IEEE-488-GPIB interface is available. This permits remote control of all the instrument's functions except the power ON/OFF, standby switching and trigger levels. The interface can be fitted by the customer: instructions are given in Section 3. The GPIB interface cannot be fitted to an instrument already fitted with the battery pack option. An adapter, Racal-Dana part number 23-3254, to convert the connector to the IEC 625-1 standard is available as an accessory.

Battery Pack (Option 07)

  • 30 Fitting the internal battery permits the instrument to be used in locations where no suitable AC supply is available. The option also allows operation from an external DC supply with the internal/external switch set to EXTERNAL 11-16V DC position.
  • 31 The battery is trickle-charged whenever the instrument is operated from an AC supply. Charging at the full rate is carried out when the instrument is switched to the standby mode. A full charge requires approximately 14 hours.
  • 32 The instrument will operate continuously for approximately 53/4 hours from a fully-charged battery. It will switch off automatically when the battery approaches the discharged condition. The STBY/CHRG indicator starts to flash approximately 15 minutes before this occurs. The battery life can be extended by use of the Battery-Save facility.
  • 33 The battery pack can be fitted by the customer. Instructions are given in Section 3. The battery pack cannot be fitted to an instrument already fitted with the GPIB interface option.

1990 FD 477

Page 24
Rack Mounting Kits

  • 34 The following kits, permitting the instrument to be mounted in a standard 19-inch rack are available:
    • Single instrument, fixed-mount kit (Option 60A). (Racal-Dana part number 11-1648). The mounted instrument occupies half the rack width and is two rack units (3.5 inches) in height. The instrument is mounted offset in the rack and may be at either side.
    • Double instrument, fixed-mount kit (Option 60B). (Racal-Dana part number 11-1649). The panel of the mounting kit occupies the full rack width and is two rack units (3.5 inches) in height. Two instruments can be mounted side-by-side.
  • 35 All the kits can be fitted by the customer. Instructions are given in Section 3.
Page 25
UNPACKING

  • 1 Unpack the instrument carefully to avoid unnecessary damage to the factory packaging.
  • 2 If it becomes necessary to return the instrument to Racal-Dana Instruments for calibration or repair, the original packaging should be used. If this is not possible, a strong shipping container should be used. Ensure that sufficient internal packing is used to prevent movement of the instrument within the container during transit.
POWER SUPPLY
AC Line Voltage Setting

  • 3 Before use, check that the AC voltage selector is set correctly for the local AC supply. The voltage range already set can be seen through a window in the selector board retaining clamp to the left of the AC power plug.
  • 4 If it is necessary to change the setting, proceed as follows:
    • (1) Undo the selector board retaining clamp on the rear panel.
    • (2) Withdraw the board.
    • (3) Replace the board with the required voltage setting positioned so that it will show through the window in the retaining clamp.
    • (4) Replace the retaining clamp.
Page 26
Line Fuse

Check that the rating of the line fuse is suitable for the AC voltage range in use. The fuse should be of the 1/4 in x 11/4 in, glass cartridge, surge-resisting type. The required rating is:

90 V to 127 V: 500 mAT (Racal-Dana part number 23-0052). 193 V to 253 V: 250 mAT (Racal-Dana part number 23-0056).

Power Cord

  • 6 The 1990 is a Safety Class 1 instrument, which is designed to meet international safety standards. A protective ground terminal, which forms part of the power-input connector on the rear panel, is provided. Each instrument is supplied with a 3-core power cord. Only the power cord supplied should be used to make electrical connection to the power-input connector.
  • AC power for the instrument must be taken from a power outlet incorporating a protective ground connector. When the green/yellow conductor of the power cord is joined to this connector, the exposed metalwork of the instrument is grounded. The continuity of the protective ground connection must not be broken by the use of 2-core extension cords or 3-prong to 2-prong adapters.
  • 8 Connection of the power cord to the power outlet must be made in accordance with the standard color code.
European America
Line
Neutral
Brown
Blue
ı Black
White
Ground (Earth) Green/Yellow Green
FUNCTIONAL CHECK

  • 9 The check given in paragraph 10 tests the operation of most of the instrument's circuits to establish whether the instrument is functioning correctly. The procedure should be followed when the instrument is first taken into use, and after transportation to a new location. It does not check that the instrument is operating to the published specification. Detailed specification tests are given in Section 7 of the maintenance manual.
  • 10 (1) Connect the instrument to a suitable AC supply.
    • (2) Switch the instrument on. Check that the instrument typenumber appears in the display for approximately two seconds, followed by a number which indicates the software version and issue number.

1990 FD 477

Page 27

  • (3) Press the FUNCTION & key until the CHECK indicator lights. Check that the display shows 10.00000 E 6 Hz and that the GATE indicator is flashing.
  • (4) Verify that the RESOLUTION indicator is lit. Press the RESOLUTION + key three times, ensuring that the resolution of the display is decreased by one digit each time.
  • (5) Press the RESOLUTION 1 key four times to increase the resolution to seven digits, and check that the GATE indicator flashes about once a second.
  • (6) Press the RECALL key. Check that all LEDs, with the exception of REM, ADDR, SRQ, GATE, TRIG A, TRIG B and STBY/CHRG flash on and off every two seconds. If the GPIB option is installed, the REM, ADDR and SRQ indicators should be lit.
  • (7) Press the RESET/CONTINUE key to return to the check functions.
FREQUENCY STANDARD (Input/Output) Option 02)

  • 11 If it is intended to use an external frequency standard, the output of the frequency standard should be connected to the EXT STD INPUT connector on the rear panel of the instrument. The connection should be made using coaxial cable. Switch on the frequency standard and the instrument: check that the EXT STD indicator on the front panel of the instrument lights.
  • 12 A 10 MHz signal, derived from the frequency standard in use, is available at the 10 MHz STD OUT connector on the rear panel of the instrument. If this signal is used, the connection should be made using coaxial cable.
TRIGGER LEVEL OUTPUT

13 The trigger levels in use on channels A and B are available via pins on the instrument front panel. If required, connection to the pins should be made using a clip-on probe or small crocodile clip.

PREPARATION FOR USE WITH THE GPIB

Introduction

14 The instrument must be prepared for use in accordance with the instructions given in Paragraphs 3 to 8 before the instructions given in this section are implemented.

1990 FD 477

Page 28

Connection to the GPIB

15

Connection to the GPIB is made via a standard IEEE-488 connector, mounted on the rear panel. The pin assignment is given in Table 3.1. An adapter, Racal-Dana part number 23-3254, to convert the connector to the IEC 625-1 standard is available as an optional accessory.

TABLE 3.1

Pin Signal Line Pin Signal Line
1 DIO 1 13 DIO 5
2 DIO 2 14 DIO 6
3 DIO 3 15 DIO 7
4 DIO 4 16 DIO 8
5 EOI 17 REN
6 DAV 18 Gnd (6)
7 NRFD 19 Gnd (7)
8 NDAC 20 Gnd (8)
9 IFC 21 Gnd (9)
10 SRQ 22 Gnd (10)
11 ATN 23 Gnd (11)
12 SHIELD 24 Gnd (5 and 17)
GPIB Connector Pin Assignment
Address Setting and Display

16 The interface address is set using five switches, A1 to A5, which are mounted on the rear panel. The permitted address settings, in binary, decimal and ASCII character form, are given in Table 3.2. The GPIB address set can be displayed, in decimal form, by pressing

RECALL LOCAL

If the address is changed, this key sequence must be repeated to display the new address. The instrument is returned to the measurement mode by pressing

CONTINUE

17 For addressed operation, the TALK ONLY switch must be in the logic '0' position (down). When this switch is in the logic '1' position, the interface is switched to the talk-only mode. The settings of switches A1 to A5 are then irrelevant.

1990 FD 477

Page 29

TABLE 3.2

Address Switch Settings

and the state of the
А5
00000001111111000000000111111111 A4 Sh
Set
00001111000011110000011111000001111 A3
001100110011001100110011001 A2 :H
IGS
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
A1
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
9
30
DECIMA
AI
S!" #$%&- ()*+,/Ø123456789:;√=∕ ASCII
LISTEN
ADDRESS
DDRESS C
@ABCDEFGHHJKLMNOPQRSTUVWXYZU/// ASCII
TALK
ADDRESS
ODES

1990 FD 477

Page 30
GPIB CHECK

  • 18 The procedure which follows checks the ability of the instrument to accept, process and send GPIB messages. The correct functioning of the instrument under local control should be verified before the procedure is attempted.
  • 19 The recommended test equipment is the Hewlett-Packard HP-85 GPIB controller, with the I/O ROM in the drawer. It is assumed that the select code of the controller I/O port is 7, and that the address of the instrument is 15 (to change the address, see Paragraph 16). If any other controller or select code/address combination is used, the GPIB commands given in the following paragraphs will require modification. The controller should be connected to the GPIB interface of the instrument via a GPIB cable. No connection should be made to the channel A or B inputs.
  • 20 Successful completion of the GPIB check proves that the instrument's GPIB interface is operating correctly. The procedure does not check that all the device-dependent commands can be executed. However, if the GPIB interface works correctly and the instrument operates correctly under local control, there is a high probability that it will respond to all device-dependent commands.

Remote and Local Message Check

  • 21 Switch the instrument on. Check that the REM, ADDR and SRQ indicators flash on and off once. If the indicators do not flash, or if they flash continuously, there is a fault on the GPIB board.
  • 22 Test as follows:
Action HP-85 Code Your Controller
Send the REN message true,
together with the
instrument's listen address
REMOTE 715

Check that the REM indicator lights.

23 Test as follows:

Action HP-85 Code Your Controller
Send the device-dependent command CK OUTPUT 715; "CK"

Check that the ADDR indicator lights and that the Check mode is selected.

Page 31
Action HP-85 Code Your Controller
Send the instrument's
listen address followed by
the GTL message
LOCAL 715

Check that the REM indicator is off. The ADDR indicator will also be off if the controller used sends the unlisten message (UNL) true automatically. This is the case when using the HP-85.

Local Lockout and Clear Lockout Check

25 Test as follows:

Action HP-85 Code Your Controller
Send the REN message true,
together with the
instrument's listen address
REMOTE 715
Send the LLO message LOCAL LOCKOUT 7

Check that the REM indicator lights. Operate the LOCAL key on the front panel and verify that the REM indicator remains lit.

26 Test as follows:

Action HP-85 Code Your Controller
Send the REN message false LOCAL 7

Check that the REM indicator is off.

27 Test as follows:

Action HP-85 Code Your Controller
Send the REN message true,
together with the
instrument's listen address
REMOTE 715

Check that the REM indicator lights. Press the LOCAL key and verify that the REM indicator turns off.

1990 FD 477

Page 32

Data Output Check

Test as follows:

Action HP-85 Code Your Controller
Set the instrument to the
check mode by sending the
listen address, followed by
the device-dependent command
CK
OUTPUT 715; "CK"
Prepare a store to receive
a 21-byte data string
DIM Z$ 21
Send the instrument's talk
address. Store the 21-byte
data string in the
prepared store
ENTER 715; Z$
Display the contents of the store DISP Z$

Check that the display reads CK+000010.00000E+06 with the cursor moved to the next line, indicating that carriage return (CR) and line feed (LF) have been accepted.

SRQ and Status Byte Check

29 Test as follows:

Action HP-85 Code Your Controller
Send the REN message true REMOTE 7
Set the instrument to send
the SRQ message when an
error is detected, and force
the generation of error code
05 by sending the device-
dependent commmand XXX
OUTPUT 715;"IPXXX"
Store the status of the GPIB
interface of the controller,
in binary form, as variable T
STATUS 7, 2; T
Display the status of the SRQ
line
DISP"SRQ=";BIT(T,5)

Check that the HP-85 displays SRQ=1, the SRQ status bit is at logic '1' or the SRQ line is < 0.8 V). Check that the SRQ indicator on the instrument is lit.

1990 FD 477

Page 33

30 Test as follows:

Action HP-85 Code Your Controller
Conduct a serial poll and
store the status byte as
variable R
R = SPOLL (715)
Display variable R DISP "R="; R

Check that the SRQ indicator is turned off when the serial poll is made. The value of R should be 101 (in binary form, R should be 000000001100101). If using an HP-85 controller, check that the ADDR indicator is turned off.

Device Clear and Selected Device Clear Check

31 Test as follows:

Action HP-85 Code Your Controller
Set the instrument to the
Total A by B mode by
sending the listen address,
followed by the device-
dependent command TA
OUTPUT 715;"TA"
Send the DCL message true CLEAR 7

Check that the function indicated on the instrument front panel changes to FREQ A.

32 Test as follows:

Action HP-85 Code Your Controller
Reset the instrument to the
Total A by B mode by
sending the listen address,
followed by the device-
dependent command TA
OUTPUT 715;"TA"
Send the SDC message true CLEAR 715

Check that the function indicated on the instrument front panel changes to FREQ A.

Page 34

IFC Check

33 Test as follows:

Action HP-85 Code Your Controller
Send the ATN message false
Send the IFC message true
RESUME 7
ABORTIO 7

Check that the ADDR indicator is turned off.

TALK ONLY Selector Test

  • 34 (1) Set the TALK ONLY switch in the instrument rear panel to '1'. Check that the REMOTE indicator is turned off and the ADDR indicator lights.
    • (2) Set the TALK ONLY switch to '0'. Check that the ADDR indicator turned is off.

OPTION FITTING INSTRUCTIONS

Single-Instrument Fixed Rack Mounting Kit 11-1648 (Option 60A)

35 The kit comprises:

Item Qty Racal-Dana Part Number
Short mounting bracket 1 16-0643
Long mounting bracket 1 16-0644
Screw, M4 x 16 4 24-7733
Crinkle washer M4 4 24-2802
Spacer, plain M4x5 4 24-4112
Screw, M6 x 16 4 24-7995
Cup washer, M6 4 24-2809
Caged nut, M6 4 24-2240

36 Assemble the kit to the instrument as follows:

  • (1) Disconnect the AC power cord at the rear panel.
  • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
  • (3) Remove the bottom cover by sliding it towards the rear of the instrument.
  • (4) Remove the instrument's feet from the bottom cover.
  • (5) Replace the bottom cover. Replace and secure the bezel.

1990 FD 477

Page 35

  • (6) Remove the four blind grommets from the sides of the instrument. This will reveal two threaded holes in each side frame.
  • (7) At one side of the instrument, secure a mounting bracket to the side frame, using two spacers, M4 screws and crinkle washers. Position the spacers between the mounting bracket and the side frame.
  • (8) Repeat step (7) at the other side of the instrument.
  • (9) Fit the cup washers to the M6 screws. Offer the instrument up to the rack in the required position, and secure the brackets to the rack using the M6 screws and nuts.
Double-Instrument Fixed Rack Mounting Kit 11-1649 (Option 60B)

37 The kit comprises:

Item Qty Racal-Dana Part Number
Short mounting bracket 2 16-0643
Screw, M4 x 16 4 24-7733
Crinkle washer, M4 4 24-2802
Spacer, plain, M4 x 5 4 24-4112
Spacer, female 2 14-1583
Spacer, male 2 14-1584
Mating plate 1 13-2000
Rivet, plastic 4 24-3211
Screw, M6 x 16 4 24-7995
Cup washer, M6 × 4 24-2809
Caged nut, M6 4 24-2240

38 Prepare both instruments as follows:

  • (1) Disconnect the AC power cord at the rear panel.
  • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
  • (3) Remove the bottom cover by sliding it towards the rear of the instrument.
  • (4) Remove the instrument's feet from the bottom cover.
  • (5) Replace the bottom cover. Replace and secure the bezel.
  • (6) Remove the four blind grommets from the sides of the instrument. This will reveal two threaded holes in each side frame.
  • (7) Remove two buffers from the bezel at the side which is to be at the centre of the rack.

1990 FD 477

Page 36

  • 39 Assemble the kit to the instruments as follows:
    • (1) At the sides which are to be at the centre of the rack, secure the female spacers to one instrument and the male spacers to the other. The spacers screw into the threaded holes in the side frames.
    • (2) At the other side of each instrument, secure a mounting bracket to the side frame, using two plain spacers, M4 screws and crinkle washers. Position the spacers between the mounting bracket and the side frame.
    • (3) Fit the male spacers on one instrument into the female spacers on the other.
    • (4) Position the mating plate to bridge the gap between the bezels. Secure it by pushing the plastic rivets through the plate into the buffer holes.
    • (5) Fit the cup washers to the M6 screws. Offer the two instruments up to the rack in the required position, and secure the brackets to the rack using the M6 screws and nuts.
TCXO Frequency Standard, 11-1713 (Option 04T)

40 The kit comprises:

Item Qty Racal-Dana Part Number
Plate assembly 1 11-1610
Oscillator PCB 1 1 9- 1208
Crinkle washer M3 3 24-2801
Screw, M3 x 6 3 24-7721

Installation

  • 41 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument.
    • (4) If an ovened frequency standard is fitted, remove this. If the basic frequency standard is in use, remove link LK1 and the frequency controls blanking plate from the rear panel.
    • (5) Secure the PCB to the plate assembly, using an M3 screw and washer from the kit. The screw should be passed through the mounting hole in the board and screwed into the threaded spacer of the plate assembly. The component side of the board should be towards the plate assembly.

1 99 0 FD 477

Page 37

Fig 3.1 PCB to Motherboard Connection

  • (6) Connect the PCB to the motherboard at PL14, ensuring that the socket fits over the portion of PL14 indicated in Fig 3.1.
  • (7) Secure the plate assembly to the rear panel, using two M3 screws and washers. The screws pass through the holes adjacent to the FREQ STD ADJUST aperture and screw into the plate assembly.
  • (8) Replace the top cover. Replace and secure the bezel.

Removal

  • 42 (1) Remove the two screws adjacent to the FREQ STD ADJUST aperture in the rear panel.
    • (2) Pull the PCB and plate assembly upwards until the board is disconnected from the motherboard.

Ovened Frequency Standards 11-1710 and 11-1711 (Options 04A and 04B) The kit comprises:

Item Qty Racal-Dana Part Number
Oscillator assembly 1 9444 for 11-1710
9423 for 11-1711
Crinkle washer, M3
Screw, M3 x 6
2
2
24-2801
24-7721

Installation

  • 44 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument.
    • (4) If an ovened frequency standard is fitted, remove this as in para. 41. If the basic frequency standard is in use, remove link LK1 and the frequency controls blanking plate from the rear panel.

1990 FD 477

43

Page 38

  • (5) Connect the frequency standard into the motherboard at PL14, ensuring that the socket fits over the portion of PL14 indicated (see illustration in para. 41).
  • (6) Secure the oscillator assembly to the rear panel of the instrument, using the M3 screws and washers. The screws pass through the holes adjacent to the FREQ STD ADJUST aperture and screw into the oscillator assembly.
  • (7) Replace the top cover. Replace and secure the bezel.
    • NOTE: When either of the ovened frequency standard options is fitted, the ref. input and output option should also be fitted.

Removal

45

  • (1) Remove the two screws adjacent to the FREQ STD ADJUST aperture in the rear panel.
    • (2) Lift the oscillator assembly out of the chassis and disconnect the flying lead from the motherboard at PL14.

Reference Frequency Multiplier Option 11-1645 (Option 10)

46 The kit comprises:

Item Qty Racal-Dana Part Number
Frequency multiplier 1 19-1164
Screw, M3 x 6 2 24-2801

  • 47 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument.
    • (4) Remove the frequency standard, if it is an ovened type.
    • (5) Remove the shorting link from between pins 8 and 9 on PL16.

NOTE: This link should be stored in a safe place. It must be replaced if Option 10 is removed from the instrument.

  • (6) Connect the frequency multiplier PCB to the motherboard at PL16 and PL17, with the threaded spacers towards the righthand side frame.
  • (7) Secure the PCB to the side frame, using the M3 screws and washers.

1990 FD 477

Page 39

  • (8) Replace and secure the frequency standard if it was removed in (5).
  • (9) Replace the top cover. Replace and secure the bezel.

NOTE: Where this option is fitted, the ref. input and output option must also be fitted.

Reference Input/Output Option 11-9000 (Option 02)

  • 48 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws securing the board to the rear panel, and remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument. If the temperature compensated crystal oscillator (TCXO) is fitted, remove the screws holding this to the rear panel.
    • (4) Remove the four screws securing the rear panel to the side members and ease the rear panel back from the chassis. Take care not to strain the connecting wires.
    • (5) Remove the two plastic plugs covering the EXT STD INPUT and 10 MHz STD OUTPUT connector holes.
    • (6) Remove the nuts from the BNC connectors. Pass the connectors through the rear panel and secure the BNCs with the nuts on the outside.
    • (7) Replace the rear panel, taking care to align the connectors PL20/SK20. Note that the connectors PL19/SK19 are staggered when fitted correctly, see Fig 3.2:

Fig 3.2 Connector Alignment

  • (8) Replace the screws holding the rear panel.
  • (9) Replace the top cover. Replace and secure the bezel.
Page 40

GPIB Option 11-9001 (Option 55)

49 The kit comprises:

Item Qty Racal-Dana Part Number
GPIB board assembly 1 19-1146
Bracket 2 11-1728
Speednut 2 24-0146
Shakeproof washer, M3 2 24-2813
Screw, M3 x 6 2 24-7721
Washer, M4 2 24-2802
Screw, M3 x 6 2 24-7721
Screw, M4 x 8 4 24-7730
-

NOTE:

This option cannot be fitted to an instrument already fitted with the battery pack option.

  • 50 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument.
    • (4) Remove the blanking plate from the rear panel by pushing out the plastic rivets from the inside of the instrument.
    • (5) Slide a speednut onto each of the two brackets. Ensure that the flat non-threaded face of the speednut is uppermost.
    • (6) Secure one bracket to each sideframe of the instrument using the M4 screws and crinkle washers supplied. Ignore the two holes near the rear of the instrument in the RH sideframe.
    • (7) Hold the GPIB board, component side down, with the GPIB connector towards the rear panel. Connect the ribbon-cable to the motherboard at SK4.
    • (8) Tilt the GPIB board, and lower it into the instrument, easing the GPIB connector into the shaped cut-out in the rear panel of the instrument.
    • (9) Line-up the holes in the GPIB board with the speednuts (move the speednuts slightly if necessary). Secure the board with the two self-tapping screws.
    • (10) Secure the bracket which carries the GPIB connector to the rear panel, using the two M3 screws and shakeproof washers.
Page 41

NOTE:

The screws and washers provide the ground connection between the GPIB connector and the instrument chassis. Tighten the screws firmly to ensure that a good connection is obtained.

(11) Replace the top cover. Replace and secure the bezel.

Battery Pack Option 11-1625 (Option 07)

51

The kit comprises:

Item Qty Racal-Dana Part Number
PCB assembly 1 11-1722
Mounting bracket 1 11-1599
Battery pack 1 11-1723
Cover plate 1 13-2040
Crinkle washers, M3 2 24-2801
Screws, M3 2 24-7721
Crinkle washers, M4 6 24-2802
Plain washers, M4 2 24-2705
Screws, M4 6 24-7730
Spare fuse, 3AT 1 23-0069
Plastic rivet 1 24-0252

NOTE:

This option cannot be fitted to an instrument already fitted with the GPIB interface option without first removing that option.

  • 52 (1) Disconnect the AC power cord at the rear panel.
    • (2) Remove the two screws which secure the bezel to the rear panel: remove the bezel.
    • (3) Remove the top cover by sliding it towards the rear of the instrument.
    • (4) Remove the blanking plate from the rear panel by pushing out the plastic rivets from the inside of the instrument.
    • (5) If a PCB-mounted frequency standard is fitted, remove the two screws adjacent to the FREQ STD ADJUST aperture.
    • (6) Remove the four screws which secure the rear panel to the side frames.
    • (7) Ease the rear panel away from the instrument until it disconnects from the motherboard at PL19 and PL20.
    • (8) Hold the PCB assembly with the switches towards the rear of the instrument and the PCB connector pointing downwards.
Page 42

  • (9) Lower the assembly into the chassis and connect the PCB to the motherboard at PL21, taking care that it mates correctly.
  • (10) Replace and secure the rear panel.
  • (11) If a PCB-mounted frequency standard is fitted, secure it to the rear panel with the screws removed in (5).
  • (12) Position the cover plate over the switches protruding through the rear panel. Secure the cover plate and the rear panel to the PCB assembly, using the M3 screws and washers.
  • (13) Secure the mounting bracket to the right-hand side frame, using two M4 screws and washers. The horizontal flange should be towards the top of the instrument.
  • (14) Position the battery pack within the chassis, with the supporting lugs resting on the mounting bracket. Secure the battery pack to the left-hand side frame, using two M4 screws and washers.
  • (15) Secure the supporting lugs to the mounting bracket, using M4 screws and washers.
  • (16) Connect the flying lead on the battery pack to the connector on the PCB assembly.
  • (17) Replace the top cover. Replace and secure the bezel.
Page 43

INTRODUCTION

1 The instrument should be prepared for use in accordance with the instructions given in Section 3. If the instrument is being used for the first time, or at a new location, pay particular attention to the setting of the AC voltage selector.

DESCRIPTION OF CONTROLS, INDICATORS AND CONNECTORS

Front Panel Items

Reference Item Description
1 Display
  • A 7-segment, LED, digital display, used to display:
  • (1) The result of a measurement.
  • (2) A number awaiting entry into an internal store.
  • (3) A number recalled from an internal store.
  • (4) Error indications.

1990 FD 477

Page 44
Reference Item Description
The display format is in engineering
format, with an 8-digit mantissa and
a 1-digit exponent. The exponent is
normally a multiple of three.
The exponent digit is blanked, and
should be assumed to be zero, for:
(1) Display of phase mode
measurement results.
(2) Totalize measurement results
having less than nine digits.
0/F Indicator Lights when the measurement result
has overflowed the eighth digit of
the display.
REM Indicator Lights when the instrument is operating under remote control.
ADDR Indicator Lights when the instrument is acting as a listener or as a talker.
SRQ Indicator Lights when the instrument generates a service request.
EXT STD Indicator Lights when the instrument is operating from an external frequency standard.
GATE Indicator Lights while a measurement cycle is in progress.
Display Units
Indicators
The Hz indicator lights for a
frequency display. The S indicator
lights for a time display. Neither
indicator lights for a display of
phase angle, ratio, total, or a
number.
2 DELAY VALUE
Control Indicator
Lights when a delay value is being
displayed. The displayed delay
value can be stepped up or down
using the f and f keys.
Page 45

Reference Item Description
3 RESOLUTION Control
Indicator
Lights to show that the resolution
of the display, and, therefore, the
measurement period (gate time) can
be changed by means of the f or f
control keys.
4 Step-Up ∮ and
Step-Down ↓ Keys
Used to step the display resolution
or the displayed delay value up or
down.
5 Function Selector The functions can be selected in
turn using the FUNCTION
keys.
The function selection 'wraps round'
at both ends.
6 HOLD (Start/Stop)
Key
Successive operations select or de-
select the Hold (single-shot
measurement) mode. The indicator
lights when the instrument is in the
Hold mode. Readings are triggered
using the RESET key.
When the instrument is in the Total
A or Time modes, successive
operations of the key start and stop
the measurement cycle.
Page 46
Reference Item Description
7 RESET/CONTINUE
(LOCAL) Key
This key has three functions.
RESET
Clears the display and triggers a
new measurement cycle when the
instrument is in the measurement
mode.
CONTINUE
Returns the instrument to the
measurement mode and triggers a
measurement cycle, following the
display of a number recalled from
store.
LOCAL
Returns the instrument to local
control from remote GPIB control
provided local lockout is not set.
8 STBY/CHRG Key Successive operations switch the
instrument into and out of the
standby state. The indicator lights
when the instrument is in the
standby state.
If the battery pack option is
installed the indicator flashes when
the battery approaches the
discharged state. The battery is
charged at the full rate when the
instrument is in standby and
external power is applied.
9 POWER Switch Controls the AC or DC power to the instrument.
10 RECALL Key Used in conjunction with NULL, DELAY and RESET keys.
RECALL NULL displays the value in the Null store.
RECALL DELAY displays the delay
value and lights the DELAY VALUE
indicator. The delay value can be
changed using the Delay value 4 and
4 keys.
RECALL RESET displays the GPIB address when this option is fitted.
Page 47

Reference Item Description
NULL Key and
indicator
Key enables and disables the NULL
function. At the time that the NULL
function is enabled, the currently-
displayed value is stored in the
Null register.
The indicator lights when NULL is selected.
DELAY Key and
indicator
Key enables and disables the DELAY
function. The delay used is that
currently in the delay value store.
The indicator lights when DELAY is selected.
11 100 AVG Key and indicator Key enables and disables the 100 AVG function.
The indicator lights when 100 AVG is selected.
12 Measurement Channel
Controls
The A and B channels have almost identical controls.
TRIG LEVEL
Controls
Used to set the trigger level, which
can be continuously adjusted over
the range -2.8 V to +2.8 V min. A
switched 0 V position is also
available, giving maximum
sensitivity for frequency
measurements.
Page 48
Reference Item Description
TRIG LEVEL
Outputs
The trigger levels in use on the A
and B channels are available at two
terminals. The voltage range is
typically ± 2.8 V , regardless of
whether or not the X10 attenuator is
selected.
AC/DC Key Used to select AC or DC coupling of
the input signal. The indicator
lights when DC coupling is selected.
Trigger Slope Key Used to select the positive-going,
or negative-going, , edge of the
input waveform for triggering.
The indicator lights when the
positive-going edge is selected.
X10/X1 Key Used to select attenuation of the
input signal. With X10 selected the
input is attenuated by a factor of
10. The indicator lights when X10
is selected.
13 TRIG Indicators Channels A and B are provided with trigger indicators.
(1) , Indicator permanently lit.
Trigger level too low or signal
input held in high state.
(2) Indicator flashing.
Channel being triggered.
(3) Indicator permanently off.
Trigger level too high or signal
input held in low state.
14 FILTER Key Successive operations enable and
disable the channel A input filter.
The indicator lights when the filter
is enabled.
15 COM A Key Used to connect the channel A input
to channels A and B in parallel
(common configuration).
The indicator lights when the common
configuration is selected.
Input Connectors All inputs are BNC connectors.
Page 49

Rear Panel Items

Reference Item Description AC Power Input Plug A standard connector for the AC power supply. An RFI filter is incorporated. 2 Line Voltage Voltage selection is changed by Selector externally repositioning a printed circuit board. The voltage selected can be seen through a window in its retaining clamp. 3 Line Fuse A 1/4 in x 11/4 in, anti-surge, glass cartridge fuse. The required fuse ratings for different line voltage ranges are shown on the panel and in Section 3 of this manual. 4 10 MHz STD OUTPUT A BNC connector, providing a 10MHz option 02 only output signal locked to the frequency standard in use.

3

Page 50
Reference Item Description
5 EXT STD INPUT
option O2 only
A BNC connector for connecting an
external frequency standard. The
instrument will operate from the
external frequency standard whenever
a signal of suitable frequency and
amplitude is applied.
The frequency required is 10 MHz
unless the reference frequency
multiplier option is fitted. With
this option, frequencies of 1 MHz,
2 MHz, 5 MHz and 10 MHz are
acceptable.
6 FREQ. STD.
ADJUST
This aperture provides access to allow adjustment of the optional internal frequency standards.
7 GPIB Option
GPIB Address
Switches
Switches A1 to A5 define the listen
and talk addresses for GPIB
operation in the addressed mode.
The talk-only switch must be in the
'Ø' position.
With the talk-only switch in the '1'
position the instrument is set to
the talk-only condition. The
positions of switches A1 to A5 are
then irrelevant.
GPIB Connector An IEEE-488-1978 standard connector
used to connect the instrument to
the GPIB.
An adapter, Racal-Dana part number
23-3254, to convert the connector to
the IEC 625-1 standard is available
as an accessory.
Page 51

Reference Item Description
8 Battery-Pack Option
DC Power Input Plug Permits the instrument power to be derived from an external DC supply.
Battery NORMAL/SAVE
Switch
Used to select the Battery-Save facility.
INTERNAL/EXTERNAL
DC Supply Switch
Used to select operation from the
internal battery or an external DC
supply
DC Supply Fuse A 1 / 4 in x 1 1 / 4 in glass cartridge fuse
of the anti-surge type. The
required rating is 3 AT.
Page 52

FREQUENCY MEASUREMENT

4

  • (1) Switch the power on.
    • (2) Select the FREQ A measurement mode, using the function selector ① •
    • (3) Set the AC/DC coupling ② and attenuator ③ as required.
  • (4) Connect the signal to be measured to the channel A input (4). CAUTION: SIGNAL LEVEL ENSURE THAT THE INPUT SIGNAL DOES NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.
  • (5) Set the manual trigger level to the required value (5). Check that the channel A TRIG indicator (6) flashes.

NOTE: Put the trigger control (5) to 0 V for maximum sensitivity. Any other position sets a trigger level between ±2.8 V.

  • (6) Select the required display resolution (7) .
  • (7) If a frequency below 50 kHz is to be measured in the presence of noise, enable the filter (8) .
  • (8) If operation in the hold mode is required, select HOLD (9) and press the RESET key (10).
  • (9) Check that the GATE indicator (1) flashes on during the measurement period.

1990 FD 477

Page 53

PERIOD MEASUREMENT

5

  • (1) Switch the power on.
    • (2) Select the PERIOD A measurement mode, using the function selector (1).
    • (3) Set the AC/DC coupling ② and attenuator ③ for channel A, as required.
    • (4) Connect the signal to be measured to the channel A input (4) .

CAUTION: SIGNAL LEVEL ENSURE THAT THE INPUT SIGNAL LEVEL DOES NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

(5) Set the manual trigger level to the required value (5) . Check that the channel A TRIG indicator (6) flashes.

NOTE: Put the trigger control (5) to 0 V for maximum sensitivity. Any other position sets a trigger level between ±2.8 V.

  • (6) Select the required display resolution (7) .
  • (7) If hold mode operation is required, select HOLD (8) and press the RESET key (9).
  • (8) Check that the GATE indicator (1) flashes on during the measurement period.

1990 FD 477

Page 54

TIME INTERVAL MEASUREMENT

(1) Switch the power on.

6

  • (2) Select the T.I. A → B measurement mode, using the function selector (1).
  • (3) Set the AC/DC coupling (2), attenuator (3), and slope (4), as required. If the start and stop signals are from the same source, select COM A (5).
  • (4) Connect the start signal to the channel A input 6. If a separate stop-signal source is used, connect the stop signal to the channel B input 7 and set the associated input controls.

CAUTION: SIGNAL LEVEL ENSURE THAT THE INPUT SIGNALS DO NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

(5) Set the manual trigger levels to the required values (8) (9). Check that the TRIG indicators (10) (11) flash.

NOTE: If required, monitor the trigger output terminals ① and set the trigger to a specific voltage level.

  • (6) Select the required display resolution (12) .
  • (7) If a delay to the stop circuit is required, set the required delay value and enable the delay (13).
  • (8) If hold mode operation is required, select HOLD (14) and press the RESET key (15) .
  • (9) Check that the GATE indicator (16) flashes on during the measurment period.

1990 FD 477

Page 55

TOTALIZE MEASUREMENT

Total A by B (Electrical)

A procedure for Total A (Manual) totalize is given in para. 8.

  • (1) Switch the power on.
  • (2) Select the TOTAL A by B measurement mode using the function selector (1) .
  • (3) Set the AC/DC coupling (2), attenuator (3) and slope (4) as required for both channels.

NOTE: The channel A slope switch selects the slope of the events which are counted. The measurement period starts on the slope of the B channel signal selected by the channel B slope switch, and stops on the opposite slope.

(4) Connect the signal to be totalized to the channel A input (5) and the control signal to the channel B (6) input.

CAUTION: SIGNAL LEVELS ENSURE THAT THE SIGNAL LEVELS DO NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

1990 FD 477

Page 56

  • (5) Set the manual trigger levels to the required values (7) (8) Check that the TRIG indicators (9) (10) flash.
  • (6) If a delay to the stop circuit is to be used, set the required delay value and enable the delay (11).
  • (7) If hold mode operation is required, select HOLD (12) and RESET (13) .
  • (8) Check that the GATE indicator (14) flashes on during the measurement period.
Page 57

Total A (Manual)

  • (1) Switch the power on.
  • (2) Select the TOTAL A measurement mode, using the function selector (1). The HOLD indicator (2) will light.
  • (3) Set the AC/DC coupling (3), attenuator (4) and slope (5) of channel A as required.
  • (4) Connect the signal to be totalized to the channel A input (6).

CAUTION: SIGNAL LEVEL ENSURE THAT THE INPUT SIGNAL DOES NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

  • (5) Set the manual trigger level (7) to the required value. Check that the TRIG indicator (8) flashes.
  • (6) Start and stop a measurement using the HOLD key (9). The HOLD indicator (2) will be turned off and the GATE indicator (10) will light during the measurement period. The displayed result is cumulative over successive measurement cycles. If required, clear the display after a measurement cycle by pressing the RESET key (11).

1990 FD 477

8

Page 58

PHASE MEASUREMENT

g

  • (1) Switch the power on.
    • (2) Select the PHASE A rel B measurement mode, using the function selector (1).
    • (3) Set the AC/DC coupling (2), attenuator (3) and slope (4) as required.
    • (4) Connect the signals to be compared to the channel A (5) and B (6) inputs (the larger and cleaner signal to channel A for maximum accuracy).

CAUTION SIGNAL LEVELS ENSURE THAT THE INPUT SIGNALS DO NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

  • (5) Set the manual trigger levels to the required values (7) (8). Check that the TRIG indicators (9) (10) flash.
  • (6) If hold mode operation is required, select HOLD (1) and press the RESET key (12).
  • (7) Check that the GATE indicator 13 flashes on during the measurement cycle.

NOTE: The phase measurement is always positive, and is the angle by which the signal applied to channel A leads that applied to channel B.

19 90 FD 477

Page 59

RATIO MEASUREMENT

  • 10 (1) Switch the power on.
    • (2) Select the RATIO A/B measurement mode, using the function selector (1).
    • (3) Set the AC/DC coupling (2) and attenuator (3) as required.
    • (4) Connect one of the signals to channel B (4) and the other to channel A (5). The lower frequency signal should be connected to channel B.
CAUTION: SIGNAL LEVEL ENSURE THAT THE INPUT SIGNALS DO NOT EXCEED THE DAMAGE LEVELS SPECIFIED IN SECTION 1 OF THIS MANUAL.

  • (5) Set the manual trigger levels to the required values (6) (7). Check that the TRIG indicators (8) (9) flash.
  • (6) Select the required display resolution (10) .
  • (7) If hold mode operation is required, select HOLD (11) and press the RESET key (12).
  • (8) Check that the GATE indicator 13 flashes on during the measurement period.
Page 60

TIME MEASUREMENT

  • 11 (1) Switch the power on.
    • (2) Select the TIME measurement mode using the function selector
    • (3) Start and stop the time measurement using the HOLD (start/stop) key (2).
  • 12 The HOLD indicator is turned off and the GATE indicator lit during the measurement period. The displayed result is cumulative over successive measurement cycles. If required, the display can be cleared, after a measurement cycle, by pressing RESET (3).
Page 61

SETTING AND MEASURING THE TRIGGER LEVEL

  • 13 (1) The trigger level is set using the rotary controls for channel A (1) and channel B (2) . Each control can be set independently for positive or negative levels typically between +2.8 V and -2.8 V with centre 0 V. Rotating the control clockwise gives a positive trigger level and anticlockwise gives a negative trigger level.
    • (2) For maximum sensitivity on small signals a switched 0 V position is available by turning the appropriate control 1 or 2 fully anti-clockwise until the control clicks.
    • (3) The trigger voltage set can be measured at the output terminals (3) (4), using a DVM, oscilloscope or other high input impedance instrument.
Page 62

INSTRUMENT CHECKS

  • 14 (1) Switch power on.
    • (2) Select the CHECK mode, using the function selector (1). Check that 10 MHz is displayed. Resolution can be changed using the RESOLUTION key (2).
    • (3) To check the front panel LEDs press RECALL (3). All LEDs, with the exception of REM, ADDR, SRQ, GATE, TRIG A, TRIG B and STBY/CHRG, will flash on and off every two seconds. If the GPIB option is fitted, the REM, ADDR and SRQ will flash in time with the other LEDs.
DISPLAY RESOLUTION

  • For all measurement functions other than TOTAL A by B, TOTAL A and PHASE A rel B, the resolution refers to the number of zeros displayed when no signal is applied at the input. The resolution can be set to display 3 to 8 digits. A 10% overrange of the display is permitted without a change of range. Because of this, an additional digit with a value of 1 may appear at the more significant end of the display when measurements are made. If the 10% display overrange is exceeded, with eight digits selected, the overflow LED will light. An overflow can also occur when NULL or 100 AVG is enabled. The overflow digit can be displayed by decreasing the resolution.
  • 16 With some measurement functions, the number of digits appearing may be less than the selected resolution to ensure that they are rounded to meaningful values.
  • 17 When RATIO measurements are made, no more than eight digits are displayed, regardless of the resolution selected.

1990 FD 477

Page 63

  • 18 For the TOTAL A by B and TOTAL A measurement functions the display shows the true total of events counted from 1 to 99999999. For higher totals the exponent is used.
  • 19 For the PHASE A rel B measurement, up to four digits may be displayed for frequencies up to 100 kHz and up to three digits for higher frequencies. Leading zeros are suppressed. For frequencies above 100 kHz the resolution of the display is reduced, and placeholding (half size) zeros are displayed as the least-significant digits.
Setting the Display Resolution

20 Whenever the resolution control indicator is lit, the resolution can be changed using the step-up and step-down keys.

GATE TIME

21 For the frequency, period and ratio measurement functions, the gate time is related to the resolution selected, as shown in Table 4.1.

TABLE 4.1
Resolution and Gate Time
Resolution Gate Time
8 10 s
7 1 s
6 100 ms (see NOTE 2)
5 10 ms
4 1 ms
3 1 ms (see NOTE 3)

NOTE 1:

The gate times shown are nominal. Due to the use of the recipromatic counting technique the gate time may be extended by up to one period of the input signal on FREQ A, PERIOD A and RATIO A/B.

NOTE 2: A resolution of 6 is selected when the instrument is first switched on.

NOTE 3: With a resolution of 3 selected, measurements are averaged.

1990 FD 477

Page 64

22 For the PHASE A rel B measurement function the gate time depends upon the signal frequency. The gate time is approximately 50 ms for frequencies above 200 Hz, but will be increased at lower frequencies.

NULL FEATURE

23 The null feature allows a displayed value to be entered into the internal NULL store. When the null feature is enabled (NULL indicator lit) the display indicates

(measured value minus the value held in the NULL store).

  • 24 The null feature is avilable with all functions except phase and check.
  • 25 (1) Use the procedures given in paragraphs 4 to 12 to set up the instrument to display the measurement required. If nulling from a value already in the NULL store, press

The value in the NULL store will be displayed.

(2) To enable NULL, press

NULL .

The NULL indicator will light. The displayed value will be entered into the NULL store. When a new measurement is made the display indicates the difference between the measured value and the value in the NULL store.

(3) To disable the null facility, press

NULL .

The NULL indicator will go out and the display will indicate the measured value. The value in the NULL store is unchanged.

26 The value held in the NULL store can be displayed at any time by pressing

RECALL NULL .

To return the instrument to the status existing before the NULL store contents were displayed press

CONTINUE .

1990 FD 477

Page 65
STOP CIRCUIT DELAY (HOLD OFF)

Use of the Delay

  • 27 The stop circuit can be delayed when the T.I. A → B or the TOTAL A by B measurement function is selected. The required delay is entered into an internal store by the operator. The delay function can then be enabled and disabled as required. The delay is set to 10 ms when the instrument is first switched on.
  • 28 The delay can be used to prevent the stop circuit being triggered prematurely by spurious signals, such as those resulting from contact bounce. The principle is shown in Fig 4.1.

Fig 4.1 Use of Stop Circuit Delay

Page 66

Displaying the Delay

29 The value of delay held in the store can be displayed by pressing

Changing the Delay

  • 30 The delay value stored is changed using the DELAY VALUE | or | keys. Pressing RESET/CONTINUE or DELAY stores the new delay value and returns the instrument to measurement mode. If DELAY has been pressed, the newly entered delay value is also enabled.
  • 31 The permitted range of delay is from 200 µs to 800 ms. The value in the delay store is retained when the instrument is switched to standby.
Enabling and Disabling the Delay

32 The stop delay is enabled and disabled by means of the DELAY key. The DELAY indicator lights when the delay is enabled.

AVERAGE OF 100 READINGS

  • 33 This feature allows one extra digit of resolution to be obtained by taking 100 readings into store and displaying the software average value. The average of 100 readings can be used with the Time Interval, Phase Measurement, Frequency A, Period A and Ratio A/B.
  • 34 To enable the averaging facility press the 100 AVG key. The 100 AVG indicator will light.
  • 35 With the appropriate signals connected the instrument will take 100 measurements, during which time the gate indicator will flash 100 times. The average value is then displayed. For reliable operation the signal must be constant during the measurement period and asynchronous (not related) to the counters frequency standard.
  • 36 A new sequence can be started at any time by pressing

RESET / CONTINUE

37 To disable the averaging feature press 100 AVG . The 100 AVG indicator will go out.

Page 67
SPECIAL FUNCTIONS
Frequency B

38 This function can be activated in FREQUENCY A mode only, by pressing:

DELAY

and holding the key down for three seconds. The DELAY indicator will then light and the counter will measure frequency B to the resolution selected.

39 To disable Frequency B press:

DELAY or RESET / CONTINUE .

The DELAY indicator will go out. Note that normal delay mode is not available in Frequency.

LED Check

40 When in CHECK mode this function is activated by pressing:

RECALL .

The LED Check causes the display and single LEDs to be alternately on and then off at a 3-second rate. All LEDs are checked, including the GPIB LEDs (if GPIB is fitted), except the STBY/CHRG, GATE and TRIG LEDs. To revert to the CHECK function press

RESET / CONTINUE .

ERROR CODES

41 The instrument is able to detect a number of error states, which are indicated on the display. The meanings of the error codes are shown in Table 4.2

TABLE 4.2

Error Codes
Display Error
Er 01 Phase measurement attempted on signals of different frequencies.
Er 02 Measurement result too large or too small for the display.
Er 03 Overflow of internal counters.
Er 04 Number entry error (GPIB only).
Er 05 GPIB programming error (GPIB only).
Er 06 Phase inputs greater than 5 MHz.
Er 50 Basic check function error.

1990 FD 477

Page 68

Clearing the Error Codes

  • 42 Error code Er 01 is cleared by:
    • (1) Making a phase measurement on signals of equal frequency.
    • (2) Selecting another measurement function.

Error codes Er 02 and Er 03 are cleared by:

  • (1) Obtaining a measurement result that is within range.
  • (2) Selecting another measurement function

Error code Er 06 is cleared by:

  • (1) Reducing the input frequency to less than 5 MHz.
  • (2) Selecting another measurement function.

USING THE BATTERY PACK OPTION

Power Supply Changeover

43 When the battery pack option is installed, the instrument can be powered from the internal battery, an external DC supply of 11V to 16V, or an external AC supply. If the instrument is operating from either the DC supply or the battery, it will automatically change to operation from the AC supply when this is connected. The battery will not take over from either the AC or the DC supply if the supply fails. An external DC supply will not take over from the AC supply if the AC supply fails.

Battery-Low Indication

  • 44 When the instrument is operating from the internal battery, or from an external DC supply, the STBY/CHRG indicator will start to flash as the supply voltage approaches the minimum permissible level. This occurs regardless of whether the instrument is in the standby mode or not. When operating from the battery, the instrument can be used in the measurement mode for approximately 15 minutes after the indicator commences flashing.
  • 45 When the voltage of the battery or the external DC supply reaches the minimum permissible level, the instrument shuts down completely.
Operating Instructions

46 Instructions for preparing the instrument to make measurements are given in the following paragraphs. No other change in the operating procedure is required.

1990 FD 477

Page 69
Operation From the Battery

  • (1) Set the internal/external switch on the rear panel to INTERNAL BATTERIES.
    • (2) Set the BATTERY SAVE/NORMAL switch to NORMAL.
    • (3) Switch the instrument on.
    • (4) Check that the instrument goes through the normal switch-on sequence. If the STBY indicator is flashing, or if there is no display, charge the battery.
  • 48 If the battery-save facility is to be used, set the BATTERY SAVE/NORMAL switch to BATTERY SAVE. The instrument will remain in the measurement mode for approximately one minute and will then switch to standby. It can be returned to the measurement mode for a further period of one minute by pressing the STBY/CHRG key.
Operation From an External DC Supply

  • 49 (1) Ensure that the instrument is switched off.
    • (2) Connect the DC supply to the DC power-input plug on the rear panel. The mating connector is a 2.1 mm coaxial socket.

CAUTION: SUPPLY POLARITY THE POSITIVE SIDE OF THE SUPPLY MUST BE CONNECTED TO THE CENTER CONDUCTOR.

  • (3) Set the internal/external switch on the rear panel to EXTERNAL 11-16V.
  • (4) Switch the instrument on. Check that the instrument goes through the normal switch-on sequence.
Battery Charging

50 The battery is trickle-charged whenever the instrument is operated from an AC supply and the internal/external switch is set to INTERNAL BATTERIES. To charge the battery at the full rate, connect the instrument to an external AC or DC supply, switch on and select the standby mode.

1990 FD 477

.,

Page 70
INTRODUCTION

1 The instrument must be prepared for use in accordance with the instructions given in Section 3. If the instrument is being used for the first time, or at a new location, pay particular attention to the setting of the AC line voltage selector.

GPIB OPERATING MODES

2 The instrument can be operated via the GPIB in either the addressed mode or the talk-only mode.

TALK-ONLY MODE

  • 3 The talk-only mode may be used in systems which do not include a controller. Such a system permits remote reading of the instrument's measurement data, but the instrument is operated by means of the front-panel controls as described in Section 4.
  • 4 The rate at which measurements are made is determined by the instrument. The output buffer is updated at the end of each measurement cycle, overwriting the previous measurement data if this has not been transferred to the listener.
  • 5 The transfer of data from the instrument to the listener is triggered by the listener. The instrument's output buffer is cleared when the data transfer is complete. Problems arising from the differences between the measurement rate and data transfer trigger rate are resolved according to the following protocol:
    • (1) If data transfer is in progress at the end of a measurement cycle, the updating of the output buffer is delayed. The data transferred will relate to the previous measurement cycle.
    • (2) If the data transfer trigger occurs during a measurement cycle and the output buffer is empty, data transfer will be delayed until the buffer is updated. The data transferred will then relate to the latest measurement cycle.
    • (3) If a measurement cycle is completed before the results of the previous cycle have been transferred to the listener, the buffer will be updated. The data for the previous cycle will be overwritten and lost.
Page 71

  • The rate at which measurements are made can be controlled in the following ways:
    • (1) The gate time of the instrument (duration of the measurement cycle) can be controlled by choosing an appropriate display resolution.
    • (2) The instrument can be operated in the hold mode. Single measurement cycles can be triggered, when required, by means of the RESET key.
  • 7 The format of the data output is described in Table 5.1.

ADDRESSED MODE

8 In addressed-mode operation, all the instrument's functions, except the power ON/OFF, standby switching and trigger levels, can be controlled by means of device-dependent commands, sent via the bus, when the instrument is addressed to listen. The measurements made, and data regarding the instrument's status, can be read via the bus when the instrument is addressed to talk. If the instrument is addressed to talk when the output buffer is empty, no data transfer can take place and bus activity will cease. Data transfer will commence when the output buffer is updated at the end of the next measurement cycle.

DATA OUTPUT FORMAT

9 The same output message format is used for the transmission of measured values and numbers recalled from the instrument's internal stores. The message consists of a string of 21 ASCII characters for each value transmitted. These are to be interpreted as shown in Table 5.1. The units should be assumed to be Hz, seconds, degrees or a ratio, depending upon the commands previously given to the instrument.

DEFERRED COMMANDS AND IMMEDIATE COMMANDS

10 Some commands (known as Deferred Commands) are accepted until a terminating character or message is received, see Table 5.5. The whole string will then be obeyed. Other commands (known as Immediate Commands) are obeyed as soon as they are received. These are indicated, in Table 5.18, by an asterisk.

EXAMPLE: OUTPUT 716; FA ANS SRS5 S81 CR LF

Because SRS is an immediate command, Frequency A, A Channel Negative Slope, and 5 Digit Resolution will all be set following receipt of SRS5.

199 0 FD 477

Page 72

Output Message Format

Byte No Interpretation Permitted ASCII Characters
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Function letter
Function letter
Sign of measurement
Most significant digit
Digit
Digit
Digit
Digit
Digit
Digit
Digit
Digit
Digit
Least significant digit
Exponent indicator
Sign of exponent
More significant digit
Less significant digit
Carriage return
Line Feed
See Table 5.2
+ or -
0 to 9
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
E + or -
0 to 9
CR
LF

NOTE 1:

Bytes 4 to 15 will always include 11 digits and a decimal point. Zeros will be added, where necessary, in the more significant digit positions.

NOTE 2:

The exponent indicated by bytes 18 and 19 will always be a multiple of three.

Page 73

Function Letters

Function Function Letters
Frequency A FA
Period A PA
Time interval TI
Total A by B TA
Phase PH
Ratio A/B RA
Check CK
Total A TM
Time TC
Recalled Data Function Letters
Unit type UT
Resolution RS
Null store NL
Delay time DT
Special function SF
Master software issue number MS
GPIB software issue number GS

NOTE:

Spaces are substituted for the function letters when special function 81 is active.

SERVICE REQUEST

  • 11 The instrument can be set, by means of device-dependent commands, to generate the service request message (SRO) when:
    • (1) (2) (3) (4) A measurement cycle is completed
    • A change of frequency standard occurs
    • An error state is detected
    • Any combination of (1), (2) and (3),
  • 12 The generation of the SRQ may also be inhibited. The necessary commands are given in Table 5.17. Option (3) of Paragraph 10 is selected when the instrument is first switched on.
STATUS BYTE

13 The format of the status byte, generated in response to a serial poll, is given in Table 5.3.

1990 FD 477

Page 74
Status Byte Format
DIO Line Function
1 LSB
2 Number of error detected (binary)
3 MSB (See NOTE 1)
4 '1' = frequency standard changed
5 '1' = reading ready (See NOTE 2)
6 '1' = error detected
7 '1' = service requested
8 '1' = gate open

NOTE 1:

The error code numbers which can occur are:

  • 1 Phase measurement attempted on waveforms of differing frequency.
  • 2 Result out of range of the display
  • 3 Overflow of internal counters
  • 4 Error in numerical entry
  • 5 Syntax error in GPIB command
  • 6 Phase measurement attempted on signal above 10 MHz.

No measurement data string is available if error code 1, 2 or 3 is generated.

NOTE 2: Regardless of the SRQ mode in use, the SRQ message that a reading is ready is not generated following a data-recall operation.

NOTE 3:

The errors are cleared as follows:

  • Error 1: Correct the difference in input frequencies or change the measurement mode in use.
  • Error 2: The error is cleared when an in-range measurement is completed.
  • Error 3: The error is cleared when an in-range measurement is completed.
  • Error 4: The error is cleared when a valid numerical entry is made.
  • Error 5: The command string will be correctly executed up to the point at which the error occurs. The remainder of the string will be hand-shaken, but not executed. The error is cleared when the next valid command is received.
  • Error 6: Correct the input frequency or change the measurement mode in use.

1990 FD 477

Page 75

EXPLANATION OF RESPONSE TO INTERFACE MESSAGES

  • 14 The instrument will respond to all valid device-dependent commands which are received after it has been addressed to listen. Devicedependent commands are recognized as such because they are transmitted with the attention (ATN) message false.
  • 15 The instrument also responds to a number of multi-line interface messages. These are recognized because they are transmitted with the ATN message true. Refer to Table 5.4, which gives the instrument's response to different bus messages. The following paragraphs detail the instrument's response to these messages. Any multi-line message not specifically mentioned is hand-shaken, but is otherwise ignored.
Address Messages

  • 16 The instrument responds to address messages defined by the setting of the address switches, A1 to A5, on the rear panel.
  • 17 On receipt of its listen address, the instrument becomes a listener. If it has previously been addressed to talk it ceases to act as a talker. If in the local control state when the address is received, the instrument goes to the remote control state provided that the REN message is true.
  • 18 On receipt of its talk address, the instrument becomes a talker. If it has previously been addressed to listen it ceases to act as a listener. If in the local control state when the address is received, it will remain under local control.
  • 19 If the instrument has been addressed to talk, and then receives the talk address of another device, it ceases to act as a talker.
Local Lockout

  • 20 The instrument will respond to the local lockout (LLO) message regardless of its addressed state. The return-to-local function of the LOCAL key on the front panel is disabled (the RESET/CONTINUE function remains enabled when in local control).
  • 21 Local lockout is cleared by sending the remote enable (REN) message false. This returns all devices on the bus to the local control state.
Device Clear and Selected Device Clear

  • 22 The instrument only responds to the device clear (DCL) message and the selected device clear (SDC) message when it is in the remote control state. It will only respond to the SDC message if it is a listener, but will respond to the DCL message regardless of its addressed state.
  • 23 The instrument responds to either message by reverting to the functions and settings of the power-up state. No change is made to the condition of the GPIB interface.

1990 FD 477

Page 76

Response to Bus Messages

Message Addressed State Instrument Response
Address Any For listen address:
Becomes a listener and goes to
the remote control state. If
previously addressed to talk,
ceases to act as a talker.
For talk address:
Becomes a talker. If previously
addressed to listen, ceases to
be a listener.
For talk address of another
device:
If previously addressed to talk,
ceases to be a talker.
Local Lockout
(LLO)
Any LOCAL key disabled.
(Cleared by sending the REN
message false).
Device Clear
(DCL)
Any, but must
be in remote
control.
Reverts to nower-up state
Selected Device Clear
(SDC)
Listen and in remote control
Serial Poll Enable
(SPE)
Any Enters the serial poll mode state
(SPMS). If addressed to talk
while in this state, sends the
status byte.
Serial Poll Disable
(SPD)
Any Enters the serial poll idle state
(SPIS). If addressed to talk
while in this state, sends data
in the output message format.
Group Execute Trigger
(GET)
Listen, and no
measurement
cycle in
progress
Takes a measurement.
Go to Local
(GTL)
Listen Reverts to local control.
Untalk (UNT)
Unlisten (UNL)
Talk
Listen
Ceases to be a talker.
Ceases to be a listener.
The ADDR indicator is turned off.
Page 77

Serial Poll Enable and Serial Poll Disable

  • 24 The instrument responds to both the serial poll enable (SPE) message and the serial poll disable (SPD) message regardless of its addressed state.
  • 25 The instrument responds to the SPE message by entering the serial poll mode state (SPMS). If the instrument is addressed to talk while in this state, it will put its status byte onto the bus instead of its normal data output string.
  • 26 The instrument responds to the SPD message by leaving the SPMS and entering the serial poll idle state (SPIS). If the instrument is addressed to talk while in this state, it will put its data output string onto the bus provided data is available in the output buffer.
Group Execute Trigger

27 The instrument responds to the group execute trigger (GET) message provided that it is a listener and no measurement cycle is in progress. Except for the inability to retrigger during a measurement cycle, the response to the GET message is the same as to the device-dependent command T2.

Go to Local

28 The instrument responds to the go to local (GTL) message provided that it is a listener. The instrument reverts to the local control state, but remains addressed to listen. It will return to remote control on receipt of the first byte of a device-dependent command.

Untalk and Unlisten

29 If addressed to talk, the instrument will go to the talker idle state (TIDS) on receipt of the untalk message. If addressed to listen, it will go to the listener idle state (LIDS) on receipt of the unlisten message. The ADDR indicator will be turned off.

INPUT COMMAND CODES

30 When the instrument is addressed to listen it can be controlled by means of device-dependent commands given in the following tables:

Table 5.6 Instrument Preset Code Table 5.7 Measurement Function Codes Table 5.8 Input Control Codes Table 5.9 Measurement Control Codes Table 5.10 Store and Recall Codes Table 5.11 Numerical Input Format Table 5.12 Numerical Input Ranges

Table 5.13 Resolution Selection Table 5.14 Special Function Codes Table 5.15 Special Functions Table 5.16 Output Message Format

Table 5.17 Service Request Codes

1990 FD 477

Page 78

31 If more than one command is to be sent, no delimiters are required. If necessary, commas, spaces and semicolons may be included in command strings as an aid to clarity without affecting the operation of the instrument. Each command string must be followed by an endof-string terminating group. The permitted terminating groups are shown in Table 5.5.

TABLE 5.5

Permitted Terminators

1 2 3 4 5 б
LF LF
EOI true
CR
EOI true
CR LF CR LF
EOI true
Last Character
EOI true
TABLE 5.6
Instrument Preset Code
Function Code
Set instrument functions and settings to the power-up state IP

TABLE 5.7

Measurement Function Codes
Function Code
Frequency A FA
Period A PA
Time interval TI
Total A by B TA
Phase of A relative to B PH
Ratio A/B RA
Check CK
Total A TM
Time TC
Page 79
Function Code
A Channel B Channel
AC coupling selected
DC coupling selected
Positive slope trigger selected
Negative slope trigger selected
X10 attenuator disabled
X10 attenuator enabled
A channel filtering enabled
A channel filtering disabled
A and B channels separate
A and B channels common
AAC
ADC
APS
ANS
AAD
AAE
AFE
AFD
BAC
BDC
BPS
BNS
BAD
BAE
BCS
BCC

Input Control Codes

TABLE 5.9
Measurement Control Codes
Function Code
Select continuous measurement mode TØ (see NOTE 1)
Select one-shot measurement mode T1 (see NOTE 2)
Take one measurement or start total A or time measurement T2 (see NOTE 3)
Stop total A or time measurement T3 (see NOTE 3)
Read present value without stopping totalize measurement RF (see NOTE 4)
Null disabled ND
Null enabled NE
Delay disabled DD
Delay enabled DE
100 Average disabled NA
100 Average enabled AE
Reset (Stop measurement cycle and clear output buffer) RE

NOTE 1:

When making continuous measurements the output buffer is updated at the end of each gate period. If the buffer is being read via the GPIB when the gate period ends, updating is delayed until reading is complete.

NOTE 2: When one-shot measurements are being made, the output buffer is cleared each time command T2 is received. The measurement made must, therefore, be read before a further measurement cycle is triggered.

1990 FD 477

Page 80
NOTE 3:

When making totalize measurements, commands T2 and T3 are used with TM. In this mode the readings made in successive totalize periods are cumulative. The RE command is used to reset the count to zero when required.

NOTE 4: The RF command (reading on the fly) must be sent each time a reading is required. The reading is obtained when the instrument is made a talker.

TABLE 5.10

Store and Recall Codes
Function Code
Recall unit type RUT
Store display resolution number SRS
Recall display resolution number RRS
Store null value SN (see NOTE 1)
Recall null value RN (see NOTE 1)
Store delay value SDT
Recall delay value RDT
Recall special function register RSF
Recall master software issue number RMS
Recall GPIB software issue number RGS

NOTE 1:

Numbers to be stored should follow the store command. The format to be used for numerical entry is given in Table 5.11. The limiting values for numerical entries are given in Table 5.12.

NOTE 2:

The instrument returns to the measurement mode automatically at the completion of a store or recall operation.

NOTE 3: No SRQ message is generated for recalled data.

1990 FD 477

Page 81

Byte Interpretation Permitted ASCII Characters
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Sign of mantissa
Most significant digit
Digit
Digit
Digit
Digit
Digit
Digit
Digit
Least significant digit
Exponent indicator
Sign of exponent
More significant digit
Less significant digit
+ or -
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or .
0 to 9 or

NOTE 1: Spaces, nulls or zeros occurring immediately before byte 1 will be ignored.

NOTE 2: Byte 1 may be omitted. A positive mantissa will then be assumed.

NOTE 3:

Bytes 2 to 11 may contain up to nine digits and a decimal point. If more than nine digits are entered without a decimal point, excess digits will be truncated. The excess digits will, however, increase the power of ten stored.

If fewer than nine digits are required the unused bytes may be omitted.

NOTE 4: Spaces or nulls entered between bytes 11 and 12 will be ignored.

NOTE 5: The exponent group, bytes 12 to 15, may be omitted.

NOTE 6: Byte 13 may be omitted or transmitted as a space. In either case a positive exponent will be assumed.

1990 FD 477

Page 82

NOTE 7: Byte 15 may be omitted for a single-digit exponent.

NOTE 8: Units are assumed to be seconds for delay time.

TABLE 5.12

Function Command
Code
Numerical Limits
Low High
Resolution SRS 3 8
Null store SN 1 x 10-9 1 x 10 10
-1 x 10 10 -1 x 10 -9
Delay time SDT 200 x 10-6 0.8
Numerical Input Ranges

NOTE 1:

Delay time entries will be rounded up before storage, as follows:

1 ms to 800 ms in multiples of 1 ms, or a fixed delay of 200 µs.

NOTE 2:

Resolution entries will be rounded down to the next integer. The related gate times are shown in Table 5.13.

NOTE 3:

The exponential numbers format is shown in Table 5.11.

Page 83

Number of digits in Freq.,
Period, Time and Check.
Gate Time Resolution number
8 10 s 8
7 1 s 7
6 100 ms 6
5 10 ms 5
4 1 ms 4
3 1 ms 3

TABLE 5.14

Special Function Codes
Function Code
Enter special function nn in special function register Snn

NOTE 1: The list of special functions is given in Table 5.15.

NOTE 2:

A special function number entered in the register while the special functions are enabled will be enabled immediately.

TABLE 5.15

Special Functions

Function Number Number Function Description
20 Normal operation
21 Frequency B
70 Basic 10 MHz check
71 LED check
80 Leading letters in output string
81 No leading letters in output string
Special Function Register

32 The special functions are stored in a register organised by decades. Only decades 2, 7 and 8 are used in this instrument. The first digit of a special function number indicates its position in the register and the second digit is stored in that location. If recalled, the special function register will be transmitted as a string of 21 ASCII characters. These are interpreted in Table 5.16.

1990 FD 477

Page 84
Byte Output Characters Notes
1
2
'S' or space
'F' or space
Spaces transmitted if
SF 81 is programmed
3
4,5
6
7-10
11
'+'
0
0 or 1
0
0 or 1
0 = SF20 selected
1 = SF21 selected
0 = SF70 selected
1 = SF71 selected
12
13-15
16
17
18, 19
20
21
0 or 1
0
'E'
'+'
0
CR
LF
0 = SF80 selected
1 = SF81 selected

Output Message Format

TABLE 5.17

Service Request Codes

Function ode
Inhibit generation of SRQQSRQ generated when error is detectedQSRQ generated for measurement readyQSRQ generated for measurement ready or error detectedQSRQ generated for frequency standard changeoverQSRQ generated for frequency standard changeover or error detectedQSRQ generated for measurement ready or frequency standardQSRQ generated for measurement ready or frequency standardQSRQ generated for measurement ready or frequency standardQSRQ generated for measurement ready, error detected or frequencyQSRQ generated for measurement ready, error detected or frequencyQSRQ generated for measurement ready, error detected or frequencyQSRQ generated for measurement ready, error detected or frequencyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement readyQSRQ generated for measurement rea
Q1
Q2
Q3
Q4
Q5
Q6
Q7

NOTE: SRQ is not generated by data recalled from store.

1990 FD 477

Page 85

Alphabetic List of Command Codes

Code Code
AAC
AAD
AAE
ADC
AFD
AFE
AFS
BAC
BAC
BAC
BAC
BAC
BAC
BAC
BAC
BAC
BAC
A channel, AC coupling
A channel X10 attenuator disabled
A channel X10 attenuator enabled
A channel, DC coupling
100 Average enabled
A channel filtering disabled
A channel filtering enabled
A channel, -ve slope
A channel, +ve slope
B channel, AC coupling
B channel X10 attenuator disabled
B channel X10 attenuator enabled
A and B channels common
A and B channel separate
B channel, DC coupling
B channel, -ve slope
B channel, +ve slope
Check
Delay disabled
Delay enabled
Frequency A
Instrument preset *
100 Average function disabled
ND
NE
PA
PH
Qn
RDT
RE
RGS
RMS
RN
RRS
RUT
SN
SN
SN
SN
SN
STA
TC
TI
Tn
Null disabled *
Null enabled *
Period A
Phase A relative to B
SRQ mode *
Ratio A/B
Recall delay time
Reset measurement
Read total so far
Recall GPIB software issue
Recall master software issue number *
Recall null value *
Recall null value *
Recall special function *
Recall special function *
Recall special function *
Store delay time
Store null value *
Store resolution *
Total A by B
Time
Time interval
Measurement mode or Start/Stop reading

NOTE:

n represents a single digit. * = Immediate Command, see Deferred Commands and Immediate Commands on Page 5-2.

Page 86
INTRODUCTION

  • 1 This section describes the principles of operation of the instrument, with respect to a number of block diagrams in the text, and describes the significant features of the circuits used with respect to the circuit diagrams given in Section 8. The block diagrams are annotated with the main circuit references to simplify cross referencing between the block diagram and circuit diagram.
  • In the circuit descriptions the integrated circuits are referred to by the circuit reference given on the appropriate circuit diagram. Note that a separate series of numbers, starting at IC1, is allocated to each assembly. Where an integrated circuit package contains more than one circuit, suffix letters are used to distinguish between them. Where it is required to identify a particular pin of an integrated circuit, the circuit reference, with suffix letter if appropriate, is followed by an oblique stroke and the required pin number.
FUNCTIONAL SYSTEMS

  • 3 The instrument contains eight functional systems. These are:
    • (1) The channel A and channel B system.
    • (2) The measurement system.
    • (3) The display system.
    • (4) The keyboard system.
    • (5) The microprocessor system.
    • (6) The standby and IRQ system.
    • (7) The power supply system.
    • (8) The frequency standard system.

1990 FD 511A

Page 87

  • The functional relationship between the systems is illustrated in Fig 6.1. The measurement system is internally configured by the microprocessor system according to the instructions entered via the keyboard or GPIB system. The signal to be measured and the signal from the frequency standard are fed to the measurement system. The measured result is passed to the microprocessor system. If mathematical manipulation of the result is required, this is performed by the microprocessor before the final output is passed to the display or GPIB system.
  • 5 The standby and IRQ system handles instructions to switch to standby, received from the keyboard system or the battery pack option, and interrupt requests made by other systems.

1990 FD 511A

4

Page 88
THE CHANNEL A AND CHANNEL B SYSTEM
Functional Description

  • 6 The channel A and channel B system processes the signals applied at the channel A and channel B inputs to produce differential pairs of signals which are fed to the measurement system. A block diagram is given in Fig 6.2.
  • 7 Each channel includes relay-controlled circuits which allow selection of AC/DC coupling and X1/X10 attenuation. The common A configuration (channel B signal disconnected and channel A signal connected to both amplifiers in parallel) can be selected.
  • 8 The channel amplifiers feature separate high frequency and low frequency paths. The crossover frequency is nominally 5 kHz. Signal filtering can be introduced, in channel A only, by disconnecting the high frequency amplifier path and increasing the bandwidth of the low frequency path to 50 kHz nominal. The signals from the high and low frequency paths are combined, and drive a Schmitt trigger output stage.
  • 9 The trigger levels for the two channels are derived from manual controls on the front panel.
  • 10 Control signals for the system relays are supplied from the microprocessor system.

Fig 6.2 The Channel A and Channel B System

1990 FD 511A

Page 89
Circuit Description

  • 11 The circuit diagram is shown in Fig 5 in Section 8.
  • 12 When energised, RLA gives DC coupling of the input signal. With RLA de-energised the signal is AC coupled via C1. R3 limits the current surge which occurs if DC coupling is selected while C1 is in the charged state.
  • 13 The X1/X1O attenuator is formed by R9, R1O, and RLC. With RLC deenergised, the attenuator has a series element R9 and a shunt element formed by R1O and the amplifier impedance of 1MΩ in parallel. The attenuation is 20 dB (nominal). With RLC energised R9 is short circuited, giving 0 dB attenuation.
  • 14 The signal is then fed to a combined limiter and fixed attenuator. R16, R21 and R22 form the attenuator; with D1, D2, D3 and D4 limiting the signal to about ±3.5 V. The attenuator insertion loss is about 1.5 dB.
  • 15 The high-frequency channel buffer is a compound follower (Q1 and Q3) with a gain from Q1 gate to Q3 emitter of approximately 0.94.
  • 16 The low-frequency channel buffer IC21 and Q5, receives its input from the potential divider R21, R22. The gain from the divider input to Q5 emitter is approximately 0.94. Any offset in the system can be nulled by adjusting R33.
  • 17 When RLG is de-energised (channel A filter not selected) the signals from the two buffers are combined at the base of Q7 by the network C15 and R39. These components act as a low-pass filter to the output of the low frequency buffer, and as a high-pass filter to the output of the high frequency buffer. The crossover frequency is 5 kHz.
  • 18 The signal at Q7 emitter is fed to the Schmitt trigger, IC2Oa, via the diode bridge formed by D9a, D9b, D1Oa and D1Ob. This protects the input of IC2Oa by limiting the signal swing to approximately ±2.5 V.
  • 19 The differential output of IC2Oa forms the input to the measuring system. The hysteresis of IC2Oa, and therefore the channel sensitivity, can be set by adjusting R72.
  • 20 The trigger level from the display board is attenuated by R58 and R54, giving a loss of 1.5 dB to compensate for the limiter loss in the signal path. It is then fed into the low-frequency buffer via R49. Feedback, taken from the emitter of Q7 to IC21/2 via R47 and R48, makes IC21/2 a virtual earth point, and the gain from the R49/R54 junction to the emitter of Q7 is -0.94. A 1 V DC level at the channel A input and a 1 V trigger level therefore combine to give 0 V at Q7 emitter. Thus the selected trigger point on the input signal is always brought to 0 V at Q7 emitter.

1990 FD 511A

Page 90

  • 21 The trigger levels are derived from the supply rails on the display board. The circuit diagram is shown in Fig. 3 in Section 8. For channel A, variable resistor R3 is connected across the +5 V, -5.2 supplies, and the wiper taps off part of this voltage. The tapped voltage passes via S20 (normally closed) and is attenuated via R2 and the load resistances on the motherboard. R7 provides a buffered output to the front panel.
  • 22 With the trigger control set to the switched 0 V position. S20 opens and the trigger level is grounded via the load resistors R58 and R54 on the motherboard. This gives an accurately set 0 V trigger level with maximum signal sensitivity.
  • 23 The channel B trigger level control operates in a similar way.
  • 24 When the channel A low-pass filter is selected, RLG is energised. This open circuits the high frequency channel, and connects C16 across the low frequency channel. The low frequency channel bandwidth is then nominally 50 kHz.
  • 25 The circuit of channel B is similar to that of channel A, but is not provided with a low-pass filter. Energizing RLE connects the signal applied at the channel A input to both channel amplifiers, while deenergising RLF isolates the channel B input. RLE and RLF always operate simultaneously and in the opposite sense to each other.
  • 26 The relays are controlled by the microprocessor system. The voltage levels on the control lines are latched in IC7, shown in Fig 6 in Section 8.
THE MEASUREMENT SYSTEM
Functional Description

  • 27 The measurement circuits of the instrument are provided by two custom-built integrated circuits. These are the Multiple Counter and Control (MCC) circuits, MCC1 and MCC2. A block diagram is shown in Fig 6.3.
  • 28 The circuits within MCC1 and MCC2 are configured by the microprocessor according to the measurement function in use. The recipromatic counting technique is used. With this technique the measured signal, not the counter clock pulses, controls the start and stop of the measurement period (gate time) as shown in Fig 6.4. The gate time therefore extends over an integral number of cycles of the measured waveform. The gate time is measured by counting the clock pulses which occur while the gate is open.
  • 29 For all measurement functions the signals to be measured are fed directly to MCC2.

1990 FD 511A

Page 91

Fig 6.3 The Measurement System

Fig 6.4 Basic Recipromatic Counting Technique

1990 FD 511A

Page 92

  • 30 At the end of each measurement period MCC1 generates an interrupt request for the microprocessor system. The registers within MCC1 are addressed using the address bus and the MCC SELECT line. The measured value is transferred to the microprocessor system via the multiplexed bus.
  • 31 The internal and external frequency standard inputs are both fed to MCC2. The system will operate from the external standard provided that the input is of sufficient amplitude. A 10 MHz output, derived from the frequency standard in use, is made available at a socket on the rear panel when the Frequency Standard Input/Output option (Option 02) is fitted.

Circuit Description

32 The circuit diagram is shown in Fig 6 in Section 8.

Measured Signal Input

33 For all measurement functions, the differential outputs from channel A and channel B are applied to the measuring circuit at IC18/15, 16, 17 and 18.

Reference Frequency

  • 34 The internal reference signal is applied to IC18/2 and the external reference signal, if present, to IC18/3. A buffered version of the external reference is present at IC18/24, and is applied to the detector D13/C47/R79. The detector output is fed to IC11/6, and is read periodically by the microprocessor. If the level is above the TTL logic '1' threshold, the microprocessor sets IC18/38 to logic '0' and the measurement system switches to use the external reference.
  • 35 A 10 MHz signal, derived from the frequency standard in use, is present at IC18/37, and is fed via PL19 pin 1 to the 10 MHz STD OUTPUT socket (if fitted) on the rear panel.
  • 36 A 10 MHz reference signal, derived from the internal frequency standard is present at IC18/36. This signal is applied to IC4/24.

Microprocessor Clock and Timer

37 A 5 MHz clock signal for the microprocessor (and the GPIB microprocessor if fitted) is taken from IC4/2. A 39.0625 kHz clock signal for the microprocessor timer is taken from IC4/4.

Control Signals

38 The logic levels on lines QO to Q4, between IC4 and IC18 are shown in Table 6.1. These levels are stable if no signals are applied to any of the channel inputs.

1990 FD 511A

Page 93

TABLE 6.1

Control Signals

Measurement Control Line
Function QO Q1 Q2 Q3 Q4
FREQ A
FREQ B
PERIOD A
T.I. A-B
TOTAL A by B
RATIO A/B
CHECK
TOTAL A
TIME
1
1
0
1
1
1
1
0
0
0
0
1
0
0
1
0
1
0
1
1
1
1
1
1
1
1
GATE
1
1
0
0
0
1
1
0
0

  • NOTE (1) The FREQ B function is obtained with special function 21 active. (FREQ A with delay).
    • (2) In CHECK, Q3 will be '1' for GATE indicator on, and will be '0' for GATE indicator off.
THE DISPLAY SYSTEM
Functional Description

  • 39 A block diagram of the system is given in Fig 6.5. The GPIB indicators, the GATE indicator, the channel A and channel B TRIGGER indicators and the STANDBY indicator are held on or off by control signals from other systems. The remainder of the display is multiplexed under the control of the display drivers.
  • 40 To update the display, the microprocessor selects the appropriate display driver, using the MODE 1 and MODE 2 control lines. A string of nine 8-bit words (a control word and eight data words) is then put onto the bus. Each word is entered into a memory within the display driver under the control of the STROBE signal.
  • 41 The display driver puts the data words onto its output bus in turn. For each data word, the appropriate numeric indicator or group of LEDs is enabled by a signal on its control line.

1990 FD 511A

Page 94

LED NUMERIC DATA TO ALL LEDS INDICATO FROM MICROPROCESSO FRDM GPIB SYSTEM FROM MEASUREMENT FROM CHANNEL A AND PL SYSTEM PART OF MOTHERBOARD PART OF MOTHERBOARD

Fig 6.5 The Display System

Circuit Description

42 The circuit diagram is shown in Fig 3 in Section 8. The GPIB indicators, LP1, LP2 and LP3, are driven via SK1 from the GPIB system. The GATE indicator, LP5, is driven from the measurement system via a driver stage, shown in Fig 9, and SK2 pin 13. The TRIG indicators, LP6 and LP7, are driven from the channel A and B system via driver stages, shown in Fig 9, and SK2 pins 7 and 3. The STANDBY indicator, LP4, is driven via SK1 pin 8 from the standby and interrupt system. The remaining LED indicators and the numeric indicator DI5 are controlled by the display driver, IC2. Numeric indicators DI1 to DI4 are controlled by IC3.

1990 FD 511A

Page 95

  • 43 Display data are stored in memory within IC2 and IC3. To change the data, the microprocessor puts a control word on the port B bus. The microprocessor writes this word into the display drivers by means of a negative pulse applied to the DISPLAY STROBE line at SK1 pin 4. The control word determines the operating mode of the display drivers.
  • 44 The microprocessor then selects the display driver required by setting a logic 'O' on the appropriate MODE line, at SK1 pin 3 or 6. Eight words containing display data are written into the selected display driver via the port B bus, controlled by eight negativegoing pulses on the DISPLAY STROBE line.
  • 45 The output of each display driver is multiplexed, under the control of an internal clock. Eight-bit display data (for seven segments + decimal point or eight LED indicators) are put onto the device output bus (pins 1 to 4 and 24 to 27). A positive pulse is then applied to the enablement line of the device or group of indicators which is to display the data. The enablement line waveforms consist of 500µs positive-going pulses at approximately 250 pps.
THE KEYBOARD SYSTEM
Functional Description

  • 46 A block diagram of the system is given in Fig 6.6. The encoding of the keyboard data is performed within the system without microprocessor action. An interrupt request (IRQ) is made to the microprocessor when encoding is complete. Data transfer is initiated by the KEYBOARD ENABLE signal from the microprocessor.
  • 47 The 18 keys are organised as a 16-key matrix and two extra keys. When a key is pressed, its position is encoded into a 5-bit word. One bit, carried on the KEYBOARD EXTEND line, indicates whether the key is in the matrix or is one of the extra keys. The remaining bits indicate the position of the key.
  • 48 When a key is pressed, the encoder examines both the matrix and the extra keys, and generates a 4-bit code representing the key position.
  • 49 If the key pressed is one of the extra two, the keyboard extend line is pulled low. If the key is in the main matrix the KEYBOARD EXTEND is isolated from the key line by D1, and remains at logic 1.

199 0 FD 511A

Page 96

Fig. 6.6 The Keyboard System

Circuit Description

  • 50 The circuit diagram is given in Fig 3 in Section 8. The keys are organised as a 16-key matrix and two extra keys, having common row lines connected to the encoder at IC1/7, 8, 10 and 11. The matrix has column lines, connected to IC1/1, 2, 3 and 4.
  • 51 The encoder normally holds the row lines at logic 'O'. When a key is pressed the corresponding column line is pulled to logic 'O'. The encoder then scans the keyboard and stores a 4-bit code, corresponding to the row and column of the key, in an internal register.
  • 52 The KEYBOARD EXTEND line indicates whether or not the key is in the matrix. The inputs to IC1 are normally held at logic '1', so that SK2 pin 11 is at logic '1'. If one of the extra keys (S6 or S11) is pressed, SK2 pin 11 will go to logic '0'. The column lines of the matrix are isolated from the inputs of IC1 by D1, so that the logic level at SK2 pin 11 is not changed when a key in the matrix is pressed.

1990 FD 511A

Page 97

53 When the key-position code has been stored, the encoder sets the KEYBOARD DATA READY line, at SK2 pin 4, to logic '1' giving a microprocessor interrupt. The microprocessor sets IC1/13 to logic '0', using the KEYBOARD ENABLE line, and the encoder puts the 4-bit code onto the bus. The microprocessor reads the code and the state of the KEYBOARD EXTEND line to find which key has been pressed.

THE MICROPROCESSOR SYSTEM

Functional Description

  • 54 A block diagram of the system is given in Fig 6.7. The microprocessor used has a 5-bit bus for the high-order address bits and an 8-bit multiplexed bus which is used for the low-order address bits and for data. The low-order address bits are strobed into the address latch, which holds them on an 8-bit address bus, to free the multiplexed bus for data.
  • 55 Two latches, fed from port B of the microprocessor, are used to maintain voltage levels on the instrument control lines. A third latch is used to read the status of the instrument flags via port B. The latches and registers for the connection of the multiplexed bus to the measurement system are in the measurement system, and are controlled by the MCC SELECT signal. The display data latches are in the display system, and are controlled by strobe and chip select signals obtained from port A.

1990 FD 511A

Page 98
Circuit Description

  • 56 The circuit diagram is given in Fig 6 in Section 8. The microprocessor clock and timer signals are generated in the measurement system, and are fed to IC3/39 and IC3/37. A RESET signal is generated in the standby and IRQ system when the instrument is switched on or off, and is fed to IC3/1.
  • 57 The microprocessor bus for the high-order address bits is designated A8 to A12. The multiplexed bus, used for the low-order address bits and for data is designated B0 to B7. The microprocessor also has two input/output ports PAO to PA7 and PBO to PB7.

Multiplexed Bus Operation

  • 58 The microprocessor puts IC3/6 (ADDRESS STROBE) to logic '1' and IC3/4 (DATA STROBE) to logic '0'. This enables the address latch, IC2 (IC2/11 at logic '1') disables the ROM, IC5 (IC5/20 at logic '1') and disables the address decoder, IC14 (IC14/6 at logic '0').
  • 59 The address is put onto lines BO to B7 and A8 to A12. When the lines have settled the ADDRESS STROBE line is taken to logic 'O'. The low-order bits of the address are latched into IC2, and are held on address lines AO to A7. Lines BO to B7 are now free for use as a data bus.

Address Decoding

  • 60 The levels on address lines A6 to A8 are decoded in IC14 to provide the following outputs:
    • (1) MCC SEL, the chip-select signal for IC4.
    • (2) GPIB SEL, the chip-select signal for the GPIB address decoder.
    • (3) Y6, the chip select signal for output latch IC8.
    • (4) Y7, the chip select signal for output latch IC7.
  • 61 These outputs are only available when IC14 is enabled by a logic '1' at IC14/6 and a logic '0' at IC14/4 and 5. The level at IC14/6 is set by the DATA STROBE output at IC3/4, which is at logic '1' when the multiplexed bus is available for data transfer. All outputs from IC14 are decoded from addresses with lines A9 to A12 at logic '0', when IC14/4 and 5 are held at logic '0' by the output from IC9a, b and d.
Input and Output Latches

62 The logic levels required on the instrument control lines and on the PAGE line (most significant bit of RAM address) are set into the output latches, IC7 and IC8, from data port B of the microprocessor. The latch strobe signals are decoded in IC14. Data may be read by the microprocessor from the input latch, IC11. The latch strobe signal is provided via data port A of the microprocessor.

1990 FD 511A

Page 99
THE STANDBY AND IRQ SYSTEM
Functional Description

  • 63 The system generates reset signals for the microprocessor and GPIB interface, and the standby switching signal for the power supply system. It also combines the IRQ signals from the GPIB interface, the measurement system and the keyboard system for connection to the microprocessor. A block diagram is given in Fig 6.8.
  • 64 Reset signals for the microprocessor and the GPIB interface are generated whenever power is applied to or removed from the instrument's power supply system.
  • 65 On switching to standby, the standby signal from the keyboard system sets the standby IRQ latch. The latch outputs provide the standby IRQ and a standby flag for the microprocessor system. The standby IRQ output also clocks the standby ON/OFF latch to the set state. This provides signals to switch the power supply to standby, light the STANDBY indicator and disable IC13b, so inhibiting the other IRQs. At the end of the microprocessor interrupt routine the standby IRQ latch is reset, removing the standby IRQ. The state of the standby ON/OFF latch is not changed.

1990 FD 511A

Page 100

  • 66 While the instrument is in standby, the input to IC12b is held low and the IRQ input to the microprocessor is held high via D14. This inhibits all IRQs. The output from IC12b also holds the GPIB interface reset via Q9c.
  • 67 On return from standby, the standby IRQ latch is again set by the standby signal from the keyboard system. The standby ON/OFF latch is clocked to the reset state, the power supply is returned to normal operation and IC13b is enabled. The input to IC12b rises as C61 charges, removing the reset signal from the GPIB interface and enabling the microprocessor IRQ input. The microprocessor is now able to accept the IRQ from IC13a. At the end of the restart sequence the standby IRQ latch is reset.
  • 68 When the encoder in the keyboard system has data ready to be read by the microprocessor, the keyboard IRQ latch is clocked via the KEYBOARD DATA READY line. The latch outputs provide the keyboard IRQ and a keyboard IRQ flag. Once the keyboard has been identified as the source of the interrupt, the latch is reset by the microprocessor.
Circuit Description

69 The circuit diagram is shown in Fig 6 in Section 8.

Reset Circuit

  • 70 The RESET signal is generated in the circuit containing Q10, Q9a, d and e, and C60. When the instrument is switched on, the input to IC12f is held low until C60 charges through R96, Q9a and R90. The output at IC12f/12 goes to logic '1' when power is applied, but drops to logic '0' after approximately 300 ms. This output is inverted by IC12e to provide the microprocessor reset and by Q9c to provide the GPIB reset.
  • 71 If there is a reduction in the +5 V STANDBY supply, due to the instrument being switched off or to power failure, the potential across R91 falls. The potential at Q10 emitter is maintained by the charge in C60, so Q10 conducts. The current in R97 makes the base of Q9d positive, so the transistor conducts and holds the base of Q10 low until C60 is completely discharged. This ensures that a good reset action is obtained, even if the power is quickly restored.
Standby Operation

  • 72 On switching to standby, PL1 pin 14 is taken to 0 V by the STANDBY key. Debouncing is provided by R89 and C57. The leading edge of the signal is sharpened in IC12c, C58, R93 and IC12d, and sets the standby IRQ latch, IC13c and d.
  • 73 The negative-going output from IC13c/10 is passed via IC13a, IC12a and R101 to IC3/2, to provide a microprocessor interrupt. The positive-going output from IC13d/11 forms the standby IRQ flag (read by the microprocessor via IC11 during the interrupt routine) and clocks the standby latch, IC10b, to the set state.

1990 FD 511A

Loading...