QUOTICARE
Thanks for choosing the Digital Blood Pressure Monitor.
Please read this instruction manual before using your unit.
Please keep this instruction manual safe for future use.
Thank you for purchasing Fully Automatic Upper Arm Style Digital
Blood Pressure Monitor.
The unit uses the oscillometric method of blood pressure
measurement. It means the unit detects the movement of your
blood through your brachial artery, and converts your blood
pressure into a digital reading. The unit is simple to use because a
stethoscope is not needed while using an oscillometric monitor.
The unit stores automatically 60 sets of measurement values. You
can read the stored data conveniently by pressing the memory
button.
The unit comes with the following components:
• Main Unit
• Arm Cuff
• Instruction Manual printed in English
Symbols Use
SYMBOLS
USEDININ
THIS
INSTR
UCTION
MANUALAL
WARNING
CAUTION
WARNING
CAUTION
Symbols Use
SYMBOLS
USED
THIS
INST
UCTION
Indicates a potentially hazardous situation
avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury to the user or
patient or damage to the equipment or other property.
Transport package shall be kept away from rain.
AUTHORISED REPRESENTATIVE IN THE
EUROPEAN COMMUNITY
Refer to instruction manual/booklet
MANUFACTURER
SERIAL NUMBER
Type BF applied part
The marking of electrical and electronics devices
according to Directive 2002/96/EC.
The device,accessories and the packaging have to be disposed
of waste correctly at the end of the usage.
Please follow Local Ordinances or Regulations for disposal.
MAN
which, if not
To assure the correct use of the product basic safety measures
should always be followed including the precautions listed below:
Read all information in the instruction manual and any other
literature included in the box before using the unit.
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results
may be dangerous.Follow the instructions of your healthcare
provider.
Operate the unit only as intended. Don't use for any other purpose.
The unit is intended to use in measuring blood pressure and pulse
rate for adult only, not recommended for neonatal baby at home or
medical centre.
Do not use a cellular phone near the unit. It may result in
operational failure.
Please avoid using in high radiant area in order to make your
measuring data correctly.
Do not disassemble or attempt to repair the unit or components.
Do not use the equipment where flammable gas (such as
anesthetic gas, oxygen or hydrogen) or flammable liquid (such as
alcohol) are present.
Do not use a mobile phone or other devices that emit
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit.
Indicating that too frequent measurements can cause injury to the
PATIENT due to blood flow interference;
Remove the batteries if the unit will not be used for Three months
or More.
Regarding the application of the CUFF over a wound, as this can
cause further injury;
Regarding the effect of blood flow interference and resulting harmful
injury to the PATIENT caused by continuous CUFF pressure due to
connection tubing kinking;
2
3
Warning and safety notices
Warning and safety notices
Know Your Unit
Regarding the application of the CUFF and its pressurization on
any limb where intravascular access or therapy, or an arteriovenous (A-V) shunt, is present because of temporary interference
to blood flow and could result in injury to the PATIENT;
Regarding the application of the CUFF and its pressurization on
the arm on the side of a mastectomy;
Regarding the information that pressurization of the CUFF can
temporarily cause loss of function of simultaneously used
monitoring ME EQUIPMENT on the same limb;
Regarding the need to check (for example, by observation of the
limb concerned) that operation of the AUTOMATED
SPHYGMOMANOMETER does not result in prolonged
impairment of the circulation of the blood of the PATIENT.
When the arm is oppressed by air pressure,please loosen
CUFF or remove batteries.
Not touch the patient and battery output simultaneously when
measurement.
Warning: not allow to use the luer connetors.If Luer lock connectors
are used in the construction of tubing, there is a possibility that they
might be inadvertently connected to intravascular fluid systems,
allowing air to be pumped into a blood vessel.
4
Body
Air Socket
Display
Battery Indicator
Inflation/Deflation Symbol or
WHO Indicator
User Symbol
Memory & Times
Irregular Heartbeat Symbol
Date/Time
Display
Memory Button
Setting Button
START/STOP Button
Battery Compartment
Systolic Blood Pressure
Diastolic Blood Pressure
Average Value Symbol
Unit Indicator
Pulse
Heartbeat Symbol
5
Know Your Unit
Know Your Unit
Arm Cuff
Fit for 22cm~36cm
range of upper arm
perimeter
If your arm cuff is broken or not functional, please use a new cuff.
A new arm cuff does not include an air plug. Please continue to use the
old air plug on the new arm cuff.
Air Plug
(Insert to Air Socket)
Hose
Metal
Power
3 AA Alkaline Batteries
Instruction Manual
Please keep the instruction manual
well after using.
Warning:
Do not use rechargeable batteries.
6
Press the “ ” button.
All symbols appear,then the cuff
starts to inflate automatically.
7
Remove the battery cover
Adjusting Settings
1. To set year
+
+
-
+
-
-
+
-
+
+
-
-
from the battery
compartment.
Adjusting Settings
1. To set year
1.1 All information will appear
when the batteries are
installed at the first time.After
2 seconds the year flashes.
Finally, press the “SET” button
and continue with step 2.
Install 3 “AA” size batteries
so the + (positive) and - (negative)
polarities match the polarities
of the battery compartment
as indicated.
Replace the battery cover.
Do not use rechargeable batteries.
Use four identical 1.5V alkaline batteries type AA/LR06.
Replace both of the batteries if the symbol of low battery
appears on the display.
Never leave any low battery in the battery compartment
since they may leak and cause damage to the unit.
Disposal of used batteries should be carried out in accordance
with the national regulations for the disposal of batteries.
8
1.2 The mark of year appears and flashes after press the “ SET”
Setting button for 3
4 seconds
, while the unit is off
. Finally,
~
press the “ SET” button again and continue with step 2.
Press the
Long press the button and the value quickly changes.
“MEM”
button to increase one year.
The year can be set between 2014 and 2050, and it will return
to 2014 when the display of year up to 2050.
9
You can stop setting anytime when you press the “ ”
button to save the current setting and turn off.
Adjusting Settings
Adjusting Settings
4. To set unit
Adjusting Settings
3. To set the 12 hour or the 24 hour
2. To set date and time
2. To set date and time
Press the “MEM” button to change the month and press the
button and continue with next step. Then set the day, the
“SET”
hour and the minute by using the same way. Continue with step 3.
Adjusting Settings
3. To set the 12 hour or the 24 hour
Then press the “MEM” button again to change the current the 12 hour
or the 24 hour.
Press the “SET”
The 12 hour or
the 24 hour Setting
4. To set unit
button
again and
Continue with step 4
12 hour format24 hour format
.
Then press the “MEM” button again to change the current unit.
Press the “ ” button to save the current setting and turn off.
Unit Setting
KPa&mmHg
“mmHg”display“kPa”display
You can stop setting anytime when you press the “ ”
button to save the current setting and turn off.
10
11
Adjusting Settings
Adjusting Settings
Before Taking A Measurement
5. To set user
Before Taking A Measurement
5. To set user
The mark of a user will appear when pressing the “SET” button,
then press the “SET” button again to change the current user.
Finally, press the “ ” button to confirm user and turn off.
1.Avoid eating,drinking alcohol, smoking, exercising and bathing for
30 minutes before taking a measurement and 5 minutes should
elapse before the first reading is taken.
2.Stress raises blood pressure.Avoid taking measurements during
stressful times. Relax as much as possible and no talking during the
measurement procedure.
3.The cuff can be applied to your left or right Arm.Always measure
your blood pressure on the same Arm.
4.Measurements should be taken in a quiet place.
5.Position the unit at heart level throughout the measurement.
6.Keep a record of your blood pressure and pulse readings for your
physician.A single measurement does not provide an accurate
indication of your true blood pressure.You need to take and record
several readings over a period of time. Try to measure your blood
pressure at the same time each day for consistency.
7.Wait 30~60 seconds between measurements.The wait time allows
the arteries to return to the condition prior to taking the blood
pressure measurement.You may need to increase the wait time
depending on your individual physiological characteristics.
12
13
Proper Use of the Arm Cuff
8
7
6
5
Proper Use of the Arm Cuff
4
3
2
Make sure the air plug is
1
inserted properly in the
main unit.
Remove all clothes from
your upper arm allowing
the cuff to fit directly on
the skin.
Sit in a chair with your
feet flat on the floor.
Place your arm on a
table so as to level the
cuff to your heart.
Proper Use of the Arm Cuff
Proper Use of the Arm Cuff
Put your arm through the
loop, then pull it up to the
position of your upper arm.
The hose runs down the
inside of your arm. The
bottom of the cuff should be
approximately 1~2cm above
your elbow.
Wrap the cuff tightly around
your upper arm by using the
Velcro strip. Do not leave
any spacing between the
arm and the cuff.
1 ~ 2cm
Pass the end of the
cuff through the metal,
and keep the hose
outward.
Relax your arm, keep the
palm upward and the
fingers curving naturally,
then turn on the unit and
start to measure.
14
15
Take a Measurement
NOTE:
Take a Measurement
Take a Measurement
Take a Measurement
1
Press the “ ” button.
All display symbols appear on the screen. The cuff starts to
inflate automatically.
2
Inflation stop automatically and the measurement is
All symbols display
Deflating automaticall y
started.
The heartbeat symbol flashes once a pulse is detected.
Start to in fla te
3
The blood pressure and pulse rate are displayed when
Inflating
Maximum inflation
the measurement completed.
The cuff is deflated
automatically, and all of the
measurement results are
stored in the memory with
date and time.
IHB
The “ ” symbol will be
displayed if
is detected.
irregular heartbeat
"kPa" display
16
Start to in fla te
"IHB" display"mmHg" display
4
Press the “ ” button to turn off the unit.
The unit will automatically turn off after 2 minutes if you have
forgotten to turn it off.
NOTE:
To stop the inflation or measurement, press the “ ” button in
anytime,the unit will stop inflating and then release and turn off.
17
To delete the memory value
Use the Memory Function
Use the Memory Function
Use the Memory Function
TO
read the measurement value
To read the average value
The unit stores the blood pressure and pulse rate in the memory after
completing a measurement every time. It can automatically store 90 sets of
measurement values . The oldest record is deleted to save the latest
measurement value when more than 90 sets.
The unit also calculates an average reading based on the values of the latest
3 times measurement taken.
To read the average value
Press the “MEM” Memory button. The average value symbol will
display.
Turn to next
Use the Memory Function
To delete the memory value
In the memory reading mode,
.........................................
press
the “
SET” Memory
button for 4-6
seconds the unit will display “ ”.
Press the “
SET
” Memory button again and LCD will display “ ”.
If you are sure all the data of corresponding user can be deleted, press
the “ ” button, all the corresponding user’s data will be deleted
and the unit will turn off.
the average value of t
3 times measurement taken
read the measurement value
Press the “MEM” Memory button to view the latest measurement
value with the measuring date
and time
he latest
.............................................
.
Press the “SET”Setting button to view the oldest recorded value
with the measuring date and time.
Long press the button and the value quickly changes.
Forward
Backward
Alternating display
The 1st latest
measurement value
The 2nd latest
measurement value
The 60th latest
measurement value
1819
Note: if you have set “ ”, but do not press a button for 2 minutes,
the product will shut off automatically, but the corresponding
user’s blood pressure data will not be deleted. The delete
operation must be confirmed by user.
About Blood Pressure
IHB Introduction
IHB Introduction
Irregular Heartbeat Symbol( )
When the unit detects an irregular rhythm two or more times during
the measurement, the irregular heartbeat Symbol will appear on the
display with the measurement values.
An irregular heartbeat
rhythm is defined as a
rhythm that is more than
25% slower or 25% faster
pressure
from the average rhythm
detectedwhile the monitor
is measuring thesystolic
blood pressure and the
diastolic blood pressure.
pressure
If the Irregular Heartbeat Symbol( ) displays with your
measurement results, we recommend you consult your physician.
Follow the directions of your doctor.
Normal Hear tbeat
Pulse
Blood
Irregular Heartbeat
Short Long
Pulse
Blood
About Blood Pressure
Blood Circulation
The blood circulation is responsible for supplying the body with oxygen.
Blood pressure is the pressure exerted on the arteries.
The systolic blood pressure value (higher pressure or top value) represents
the blood pressure produced by contraction of the heart muscle.
The diastolic blood pressure value (lower pressure or lower value) represents
the blood pressure produced by relaxation of the heart muscle.
Systolic Blood PressureDiastolic Blood Pressure
Contraction of
the heart muscle
Extrude blood
Pressure increased
to the blood vessel
Higher pressure
Relaxation of the
heart muscle
Blood return heart
Pressure decreased
to the blood vessel
Lower pressure
20
21
Exceptional Situations
Exceptional Situations
Reference Material: Journal of Hypertension 1999, Vol 17 No.2
After each measurement is completed, the LCD display will show your
position automatically on the six segments of the bar indicator which
corresponds to World Health Organization (WHO) Blood Pressure Indicator.
Reference Material: Journal of Hypertension 1999, Vol 17 No.2
Diastolic Blood Pressur (mmHg)
110
100
90
Normal systolic value
85
Normal blood pressure
80
Optimal
blood pressure
(target value)
*Note!
When a person’s systolic and diastolic pressures fall into different categories,
the higher category should apply.
Severe hypertension
Moderate hypertension
Mild hypertension
120 130 140 160 180
Systolic Blood Pressure (mmHg)
22
Error Indicators
The following symbol will appear on the display when measuring abnormally.
Symbol
CauseCorrection
The course of
inflating appears
error.
When measurement
fails.
When the batteries
power are too low.
Wrap the cuff correctly and tightly.
Ensure the connection is close
between the air plug and the unit.
Inflate over again after ensuring.
Do not move your arm and body,
and keep quiet.
Measure over again according to
correct way.
Replace all of the worn batteries
with new ones.
Eliminate Fault
ProblemCauses and Solutions
No power.
No display appears on the
display screen.
Measurement values
appear too high or too low.
Replace all the worn batteries with 4 new
ones.
Check the battery installation for proper
placement of the batteries polarities.
Blood pressure varies constantly. Many
factors including stress, time of day, how
you wrap the cuff, may affect your blood
pressure. Review the sections "Proper Use
of the Arm Cuff" and "Take a Measurement".
23
Troubleshooting and Maintenance
Troubleshooting and Maintenance
Maintenance
To protect your unit from damage, please observe the following:
Do not disassemble the unit.
Do not fold the cuff or tubing tightly.
Do not wash the cuff or immerse it in water.
Do not use volatile liquids to clean the main unit.
Do not use petrol, thinners or similar solvents to clean the cuff.
Do not subject the unit to strong shocks or vibrations (for example,
dropping the unit on the floor).
Do not subject the main unit and the cuff to extreme temperatures,
humidity, moisture or direct sunlight.
Do not carry out repairs of any kind yourself. If a defect occurs,
Please contact the local distributor.
If needed, please use ultraviolet radiation for disinfection.
※
gasoline
thinner
naphtha
The unit should be cleaned with a soft, dry cloth.
Use a soft, moistened cloth and soap to clean the arm cuff.
Substitution of a cuff different from that supplied might result in
measurement error.
We will provide circuit diagrams, component part lists, descriptions,
calibration instructions, or other information to assist service
personnel to repair parts.
The service personnel needs to be the personnel who is
trained by the manufacture and has got the relevant
qualification certificate.
The accuracy of this blood pressure monitor has been carefully
tested and is designed for a long service life. We recommend that the
performance be checked every 2 years and after maintenance and
repair, by utilizing the manometer mode and verifying the accuracy
of the manometer at least at 50 mmHg (6,7 kPa) and 200 mmHg
(26,7 kPa). Please contact the ocal distributor for details.
24
BP101U
180
2×90 Sets of Measurement Values
3 AA Alkaline Batteries
Within 2 Minutes
2 Years
Cuff, Instruction Manual
105
(50 ~ 104 F
10~40℃
15 ~ 90% RH
80~105 kPa
(-4 ~ 140 F
-20~60℃
Note: Subject to technical modification without prior notice.
10 ~ 95% RH
25
Battery Indicator
)
(Noncondensing
)
(Noncondensing
)
)
EMC Table
Guidance and manufacture’s declaration – electromagnetic emission
Emission test
Compliance
Electromagnetic environment – guidance
Guidance and manufacture’s declaration
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
IEC 60601 test
Compliance
level
level
Electromagnetic environment - guidance
Guidance and manufacture’s declaration – electromagnetic immunity
Statement
Statement
The unit is intended to be used to measure blood pressure
(systolic and diastolic) and heart rate from the upper arm by using
the oscillometric method.
The unit is intended for usingin only adult population,not applied
to the other populations such as neonatal baby.
It can’t be used while the wirst (arm) has bleeding or wound to
avoid the blood flowing from the wound in pressurzing.
The device,part and batteries have to be disposed of waste correctly at
the end of the usage.
Please follow Local Ordinances or Regulations for disposal.
Applied part: CUFF.
Protection Class: Internally powered equipment.
Applied Part Type: Type BF.
Moisture Protection: IPX0,continue operation.
altitude<2000m;
overvoltage:II;
pollution degree:2
The risk of patient and user can be lowered to acceptable level.
The unit might not meet its performance specification if stored or used
outside the following specified temperature , humidity and altitude
ranges.
The unit satisfies the requirements of IEC60601-1 Medical electrical
equipment,
IEC 60601-1-2:2007: Electromagnetic compatibility –Requirements
and tests
IEC80601-2-30:Non-invasive sphygmomanometers - Part 1:
Requirements and tes t methods f o r non-automated measurement type
EN 1060-3:Non-invasive sphygmomanometers - Part 3:
Supplementary requirements for electro-mechanical blood pressure
measuring systems
EMC Table
Guidance and manufacture’s declaration
Guidance and manufacture’s declaration – electromagnetic emission
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer of the user of the Digital Blood Pressure Monitor should
assure that it is used in such an environment.
Emission test
RF emissions
RF emission
IEC 61000-3-2
fluctuations/
flicker emissions
IEC 61000-3-3
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of Digital Blood Pressure Monitor should assure
that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz/60Hz)
magnetic field IEC
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of Digital Blood Pressure Monitor should assure
that it is used in such an environment.
Compliance
CISPR 11
CISPR 11
Harmonic
emissions
Voltage
61000-4-8
Group 1
Class B
Not applicable
Not applicable
Guidance and manufacture’s declaration – electromagnetic immunity
IEC 60601 test
level
±6 kV contact
±8 kV air
3V/m3V/m
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration – electromagnetic immunity
Electromagnetic environment – guidance
The Digital Blood Pressure Monitor use RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The Digital Blood Pressure Monitor is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Compliance
±6 kV contact
Electromagnetic environment - guidance
level
Floors should be wood, concrete or
ceramic tile. If floor are covered with
±8 kV air
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
2627
EMC Table
Recommended separation distances between
portable and mobile RF communications equipment and the Digital Blood Pressure Monitor.
output power
of transmitter
Separation distance
(m)
according to frequency of transmitter
Compliance
level
EMC Table
EMC Table
Immunity test
IEC 60601 test
level
Electromagnetic environment - guidance
Recommended separation distance
IEC 60601 test
Immunity test
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Digital Blood Pressure
Monitor is used exceeds the applicable RF compliance level above, the Digital Blood
Pressure Monitor should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
Digital Blood Pressure Monitor.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
3V/m3V/m
80 MHz to 2.5
level
GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the Digital Blood Pressure Monitor,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
b
EMC Table
Recommended separation distances between
portable and mobile RF communications equipment and the Digital Blood Pressure Monitor.
Pressure Monitor is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Digital Blood Pressure
Monitor can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Digital
Blood Pressure Monitor as recommended below, according to the maximum output power
of the communications equipment.
output power
of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE l At 80 MHz and 800 MHz. the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations .Electromagnentic propagation is
affected by absorption and reflection from structures, objects and people.
80 MHz to 800 MHz
Separation distance
according to frequency of transmitter
0.1170.233
0.3690.738
1.1672.333
3.6897.379
11.66723.333
(m)
800 MHz to 2.5GHz
2829
Warranty Information
Warranty Information
The unit, excluding the unit cuff, is guaranteed to be free of defects in
workmanship and materials for a period of two years from the date
listed on the purchase record, when used in accordance with the
instructions provided with the unit.The unit cuff is warranted to be free
from defects in materials and workmanship appearing within one
year from the date listed on the purchase record when the unit is
used in accordance with the instructions provided with the unit.
This guarantee does not cover damage incurred by use of the unit
not in accordance with the instructions, accidental damage, or being
tampered with or serviced by unauthorized service agents.
For repair under this warranty.Our authorized service agent must be
advised of the fault within the period of the warranty.This warranty
covers parts and labor only under normal operations.A transportation
fee or freight fee that may be incurred will be the owner’s
responsibility.Any defect resulting from natural causes,e.g.,flood,
hurricane etc., is not within this guarantee.
Unit subjected to misuse, abuse, and neglect of these manual
contents, non-instructional purposes;unauthorized repair or
modifications will be excluded from this warranty. This guarantee
specifically excludes expendables and consumables, for example
batteries.All warranty claims must be directed to the distributor
responsible for the sale of the device. The content of this warranty is
subject to change without further notice.
WARNING: No modification of this equipment is allowed.