Quantum Odyssey, Q-Vision Service manual

ODYSSEY, Q-VISION
HF Series™
X-ray Generators
Service Manual
Manual Part No. DC30-011 Revision W
EU Authorized Representative: Medizintechnik Berlin GmbH Altentreptower, Strasse 59 12683 Berlin - Germany Phone: +49-302-82 4726 Fax: +49-302-82 6382 E-mail: Medizintechnik.Berlin.GmbH@t-online.de
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC. (QMI)
This Service Manual is incomplete without Operator’s Manual part number DC30-010.
Copyright© 2010 QMI
Quantum Medical Imaging, LLC
2002-B Orville Drive North Ronkonkoma, NY 11779-7661 USA Phone: (631) 567-5800 Fax: (631) 567-5074 E-mail: info@qmiteam.com www.quantummedical.net
Made in U.S.A.

Revision History

REVISION DATE TYPE OF MODIFICATION
E 1/12/02 Incorporated ECO 0647, 0651, 0687 F 2/25/02 Incorporated ECO 0714, 0718, 0738, 0742 G 5/15/02 Incorporated ECO 0642, 0748, 0753, 0770, 0787, 0803, 0809,
0821, 0827, 0829, 0830, 0847, 0859, 0881, 0888 H 9/26/02 Incorporated ECO 0787 J 11/11/02 Incorporated ECO 0936, 1003, 1009, 1010, 1017, 1028, 1030 K 3/31/03 Incorporated ECO 1093, 1102, 1103, 1107, 1115, 1124, 1146 L 7/11/03 Incorporated ECO 1255, 1256 M 7/29/03 Incorporated ECO 1269, 1270, 1272 N 12/19/03 Incorporated ECO 1144, 1216, 1347, 1359, 13 62, 1380, 1383,
1400 P 9/10/04 Incorporated ECO 1346, 1438, 1485, 1518 Q 4/22/05 Incorporated ECO 1565, 1570, 1600, 1606, 1607 R 12/20/05 Incorporated ECO 1660, 1663, 1679 S 5/22/07 Assigned new part number to Non-QMI HSS ver sio n man ua l;
Incorporated ECO 1697, 1715, 1757, 1794, 1890 T 5/8 /2 00 8 Incorporate ECO 1895, 19 20 , 19 32 , 19 7 7, 198 5 U 11/21/2008 Incorporate ECO 2045, 2050, 2068, 2093, 2105 V 9/1/2009 Incorporate ECO 1980, 2160, 2166, 2169 W 1/15/2010 Incorporate ECO 2220, 2228, 2276
Page
Number
i thru xii W 4-1 thru 4-20 W B-1 thru B-56 W
1-1 thru 1-22 W 5-1 thru 5-96 W C-1 thru C-7 W
Rev
Page
Number
Rev
Page
Number
Rev
2-1 thru 2-56 W 6-1 thru 6-67 W D-1 thru D-66 W 3-1 thru 3-72 W A-1 thru A-16 W
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging i
Revision History
THIS PAGE INTENTIONALLY LEFT BLANK
Revision W HF Series X-ray Generators - Service Manual
ii Quantum Medical Imaging

Table of Contents

CHAPTER 1 - SPECIFICATIONS
SYSTEM CONFIGURATIONS ................................................................. 1-3
PHYSICAL SPECIFICATIONS ................................................................. 1-5
GENERATOR CABINET SPECIFICATIONS ......................................... 1-5
OPERATOR CONTROL PANEL SPECIFICATIONS ............................... 1-5
ODYSSEY HF Control Panel: ............................... ...................... 1-5
Q-VISION HF Control Panel: .................................................... 1-5
SYSTEM SPECIFICATIONS .............................................................. 1-5
POWER REQUIREMENTS .................................................... .................. 1-6
SINGLE-PHASE X-RAY GENERATORS .. ............................................. 1-7
25 kW Single Phase (Model QG-25) .... ....................................... 1-7
25 kW Single Phase (Model QG-25-5)* ...................................... 1-7
32 kW Single Phase (Model QG-32) .... ....................................... 1-7
32 kW Single Phase (Model QG-32-5)* ...................................... 1-7
40 kW Single Phase (Model QG-40) .... ....................................... 1-8
40 kW Single Phase (Model QG-40-5)* ...................................... 1-8
32 kW Single Phase, Stored Energy ........................................... 1-8
40 kW Single Phase, Stored Energy ........................................... 1-8
50 kW Single Phase, Stored Energy ........................................... 1-8
THREE-PHASE X-RAY GENERATORS ................................................ 1-9
32 kW Three Phase (Model QG-32-2) ........................................ 1-9
32 kW Three Phase (Model QG-32-3) ........................................ 1-9
40 kW Three Phase (Model QG-40-2) ........................................ 1-9
40 kW Three Phase (Model QG-40-3) ........................................ 1-9
50 kW Three Phase (Model QG-50-2) ........................................ 1-9
50 kW Three Phase (Model QG-50-3) ........................................ 1-9
65 kW Three Phase (Model QG-65) ......................... .................1-10
80 kW Three Phase (Model QG-80) ......................... .................1-10
HIGH-SPEED STARTER POWER REQUIREMENTS ............................1-10
GENERATOR AUXILIARY POWER SUPPLIES ........................... ...............1-15
PERFORMANCE SPECIFICATIONS ........................................................1-15
SYSTEM OPTIONS ..............................................................................1-19
SYSTEM OPERATING ENVIRONMENT ....................................... ............1-19
NON-OPERATING ENVIRONMENT .......... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .... .. .. ..1-20
CHAPTER 2 - ASSEMBLY & INSTALLATION
OVERVIEW .................. .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. .. . 2-3
PRE-INSTALLATION GUIDELINES ................................................... 2-3
Verify Equipment Location ........................................................ 2-3
Electrical Requirements ............................................................ 2-5
UNPACKING .................................................................................. 2-5
INSTALLATION .............................. ............... ............... ................. ....... 2-6
REQUIRED TOOLS AND MATERIALS ............................. .................. 2-6
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
iii
Table of Contents
CHAPTER 2 - ASSEMBLY & INSTALLATION (CONT’D)
GENERATOR CABINET SETUP ......................................................... 2-6
Transformer Line Tap Adjustment (Single Phase Non-SE Systems) 2-8
Transformer Line Tap Adjustment (Single Phase SE Systems) ....2-10
AC Input (Mains) Connection ...................................................2-12
High Voltage (H.V.) Tank ......................... ...............................2-15
High Voltage Cables ................. .. .. .. .. .. .. .... .. ... .. .. .. .. .... .. .. ... .. .. ...2-18
INTERFACE BOARD SETTINGS & CONNECTIONS ............................2-21
Door Interlock Switch Connection ............................................2-22
"In Use" Light Switch Connection .............................................2-22
Collimator Connections ............................................................2-23
Tube Stand/Peripheral Equipment Connections .........................2-23
Internal +28 VDC Power Supply Option (QG-28VPS) .................2-23
Tube Rotor Boost Voltage Verification (Systems without
High (Dual) Speed Starter Option) ..........................................2-24
RADIOGRAPHIC SYSTEM EQUIPMENT CONNECTIONS ....................2-25
Tube Rotor Cable (Systems without High (Dual) Speed
Starter Option) ......................................................................2-25
Tube Rotor Cable (Systems with AID External High
Speed Starter QG-HSS Option Only) ........................................2-27
Tube Rotor Cable (Systems with Internal High Speed
Starter Q-HSS Option Only) ................................ ....................2-30
Table/Wall Bucky Connections .................................................2-34
Table/Wall Ion Chamber Connections (AEC Option Only) ...........2-34
Interfacing to Canon Digital Radiographic Receptor
(When Applicable) .................................................................2-35
Interfacing to CMT Digital Radiographic Receptor
(When Applicable) .................................................................2-36
Interfacing to Siemens Ion Chambers (When Applicable) ...........2-36
Interfacing to Pausch Tomography System (When Applicable) ...2-39
OPERATOR CONTROL PANEL SETUP ..............................................2-41
ODYSSEY Operator Control Panel ............................................2-41
OCP Pedestal Mounting (QG-PDL Option) .................................2-41
OCP Wall Mounting (QG-WM Option) ................................ .......2-43
Q-VISION Operator Control Panel ............................................2-44
Integrated Workstation Operator Control Panel .........................2-46
Optional Remote Expose Switch Connection .............................2-47
INITIAL POWER UP ......................................................................2-47
TRANSFORMER TAP VALIDATION ..................................................2-48
Q-VISION OPERATOR CONTROL PANEL - NETWORK
SETTINGS VERIFICATION ............................................................2-49
ROTOR DRIVE BOARD A10 (AY40-013T) PERFORMANCE
VERIFICATION .............. ..............................................................2-55
CHAPTER 3 - CALIBRATION
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. 3-3
REQUIRED TOOLS AND MATERIALS ............................................... 3-3
Revision W HF Series X-ray Generators - Service Manual
iv Quantum Medical Imaging
Table of Contents
CHAPTER 3 - CALIBRATION (CONT’D)
SYSTEM CONFIGURATION/CALIBRATION ..............................................3-3
ACCESSING SERVICE MODE ...........................................................3-4
SYSTEM CONFIGURATION ........................................ ......................3-6
Printer On/Off Setting ........................ .. ..... .. .. .... .. .. .... ... .. .... .. .... .3-7
Generator Configuration/Derating Settings .................................3-8
AEC Setting .............................................................................3-9
Rotor Drive Setting ................ .... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .. .... .. . 3-10
Edit Facility/Welcome Facility Utility ......................................... 3-11
Tube Type Setting ...... .... .. .. ... .. .. .. .... .. .. .. ... .. .. .. .... .. .. .. ... .. .. .... .. . 3-12
Edit Digital Settings ................................................................3-13
Ion Chamber/Receptor Configuration .......................................3-14
Tomo Setting ............................ ............................................. 3-18
SYSTEM SETTINGS ......................................................................3-21
Set Date Screen ....... .. .... .... ... .. .... .. .. .... .. ... .... .. .. .... .. ..... .. .. .... .. . 3-22
Set Time Screen ................. ... .. .. .. .. .. .... .. ... .. .. .. .. .... .. .. ... .. .. .. .. ... 3-23
Film/Screen Defaults Screen ...................................................3-23
APR Format Setting ................................................................3-25
Language Setting ................................................................... 3-26
Back Up MAS Setting ................................ .............................. 3-27
Generator Timeout Setting ..................................................... . 3-28
Date/Time Format Setting ......................................................3-28
SYSTEM CALIBRATION ................................................................. 3-29
GENERAL CALIBRATION PROCEDURE NOTES ................................3-32
Film Speed Setting on non-AEC Systems .................................. 3-33
A/D Calibration ......................................................................3-34
kV Calibration Function ...........................................................3-37
Filament Offset Calibration ...................................................... 3-39
MA Calibration ..... ..................................................................3-43
AEC CALIBRATION .......................................................................3-46
AEC Calibration Procedures ..................................................... 3-48
NON-AEC REPRODUCIBILITY VERIFICATION .................................3-66
AEC REPRODUCIBILITY VERIFICATION (ONLY PERFORM
ON UNITS EQUIPPED WITH QG-AEC OPTION) .............................. 3-67
NON-AEC LINEARITY VERIFICATION .............................................3-67
AEC LINEARITY VERIFICATION (ONLY PERFORM ON
UNITS EQUIPPED WITH QG-AEC OPTION) ...................................3-68
MAXIMUM DEVIATION: KVP, TIME, MA, MAS ................................. 3-68
CHAPTER 4 - THEORY OF OPERATION
OVERVIEW .......................................................................................... 4-3
OPERATOR CONTROL PANEL CIRCUIT ............................................4-3
Microcontroller U3 Operation .....................................................4-3
Flash Memory U2 .....................................................................4-3
EEPROM U17 Operation ............................................................4-4
LCD Controller U12 Operation ........ .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .. .4-4
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
v
Table of Contents
CHAPTER 4 - THEORY OF OPERATION (CONT’D)
Switching Power Supply Circuit ................................................. 4-5
Front Panel Key Activation Detection Circuit ............................... 4-5
Jumpers Installation ............................................................. .... 4-5
TechVision Compatibility .......................................................... 4-5
GENERATOR LOGIC CIRCUIT ......................................................... 4-5
Generator/OCP Communication ................................................ 4-6
Master and Slave Microcontroller Operation ............................... 4-6
FILAMENT CONTROL REGULATOR CIRCUIT .................................... 4-7
KVP & FILAMENT CONTROL CIRCUIT .. ............................................ 4-7
LINE MONITOR CIRCUIT (NON-SE SINGLE-PHASE AND
THREE-PHASE SYSTEMS) .............................................................. 4-8
Input Line Monitor Circuit ............................................ ............. 4-8
Capacitor Discharge Circuit ....................................................... 4-9
Auto-Shutoff Circuit ................................................................. 4-9
Three-Phase Line Monitor Circuit .............................................4-10
LINE MONITOR CIRCUIT (STORED ENERGY SYSTEMS) ...................4-10
Battery Monitor Circuit ............................................................4-10
Storage Capacitor Discharge Circuit .........................................4-10
Generator Auto-Shutoff Circuit .................................................4-11
Battery Charger Circuit ............................................................4-11
POWER SUPPLY CIRCUIT ..............................................................4-12
+48 VDC Power Supply ...........................................................4-12
+15 VDC Power Supply ...........................................................4-12
+24 VDC Power Supply ...........................................................4-12
ROTOR BOARD CIRCUIT (NON-HIGH SPEED STARTER SYSTEMS) ...4-13 ROTOR BOARD CIRCUIT (QUANTUM INTERNAL HIGH
SPEED STARTER SYSTEMS) .........................................................4-13
HSS Control Board A20A1 (AY40-035S) ....................................4-14
HSS Drive Board A20A2 (AY40-036S) .......................................4-15
AEC BOARD CIRCUIT ....................................................................4-16
Standard Type AEC .................................................................4-16
Universal Type AEC .................................................................4-17
+300 VDC Power Supply .........................................................4-18
AEC Control Circuitry ...............................................................4-18
Tomography Circuitry ..............................................................4-20
CHAPTER 5 - SERVICE INSTRUCTIONS
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. 5-3
DISPOSAL OF BATTERIES AND ACCUMULATORS
(DIRECTIVE 2006/66/EC) .............................................................. 5-3
SERVICE MAINTENANCE ................................................................ 5-4
Visual Inspection ..................................................................... 5-4
TROUBLESHOOTING ............... ........... ............ ........... .......... ........... 5-5
Warning Messages ................................................................... 5-5
Error Codes ............................................................................. 5-5
Optional +28 VDC Power Supply Troubleshooting .....................5-48
Revision W HF Series X-ray Generators - Service Manual
vi Quantum Medical Imaging
Table of Contents
CHAPTER 5 - SERVICE INSTRUCTIONS (CONT’D)
SERVICE REPORTS .................................................................. .....5-49
Accessing Service Reports - ODYSSEY X-ray Generators ............5-49
Accessing Service Reports - Q-VISION Operator Control Panel ...5-56
Accessing Service Reports - Integrated X-ray Generators ..........5-63
REMOVAL/REPLACEMENT PROCEDURES ........................................5-72
Generator Cabinet Cover Removal ...........................................5-72
Generator Cabinet Cover Replacement .....................................5-72
High Voltage (H.V.) A17 Tank Removal .................................... 5-73
High Voltage (H.V.) Tank A17 Installation ................................ 5-74
Power Modules A18 and A19 (P/N AY20-047) Removal ............5-75
Power Modules A18 and A19 (P/N AY20-047) Replacement .......5-77
Logic Board A1 (P/N AY40-006S) Removal ...............................5-78
Logic Board A1 (P/N AY40-006S) Replacement .........................5-78
KVP Control Board A2 (P/N AY40-003S) Removal .....................5-80
KVP Control Board A2 (P/N AY40-003S) Replacement ...............5-80
Filament Control Board A5 (P/N AY40-007T) Removal ............... 5-82
Filament Control Board A5 (P/N AY40-007T) Replacement ........5-82
Power Supply Board A6 (P/N AY40-005T) Removal ...................5-83
Power Supply Board A6 (P/N AY40-005T) Replacement .............5-83
Line Monitor Board A8 Removal .............................................. 5-84
Line Monitor Board A8 Replacement ...................... ..................5-84
Stored Energy Systems Line Monitor/Charger Board A8
(AY40-028T) Removal ........................ .. ..... .. .... .. .. .... ... .. .... .. .. . 5-85
Stored Energy Systems Line Monitor/Charger Board
(AY40-028T) A8 Replacement ................................................5-86
Rotor Drive Board A10 (P/N AY40-013T) Removal .................... 5-87
Rotor Drive Board A10 (P/N AY40-013T) Replacement .............. 5-87
Standard AEC Control Board A11 (P/N AY40-031S) Removal .....5-88
Standard AEC Control Board A11 (P/N AY40-031S) Replacement 5-88
Universal AEC Control Board A11 (P/N AY40-027S) Removal .....5-89
Universal AEC Control Board A11 (P/N AY40-027S) Replacement 5-89 OCP Microprocessor Board A16 (P/N AY40-004S1) Removal
(Not Applicable on Q-VISION or Integrated Generators) ..........5-90
OCP Microprocessor Board A16 (P/N AY40-004S1) Replacement
(Not Applicable on Q-VISION or Integrated Generators) ..........5-90
Q-HSS Equipped Systems Only: High-Speed Starter Control
Board A20A1 (P/N AY40-035S) Removal .................................5-91
Q-HSS Equipped Systems Only: High-Speed Starter Control
Board A20A1 (P/N AY40-035S) Replacement ...........................5-91
Q-HSS Equipped Systems Only: High-Speed Starter Driver
Board A20A2 (P/N AY40-036S) Removal .................................5-92
Q-HSS Equipped Systems Only: High-Speed Starter Driver
Board A20A2 (P/N AY40-036S) Replacement ...........................5-92
Stored Energy System - Battery Tray Removal .........................5-93
Stored Energy System - Battery Tray Replacement ...................5-95
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
vii
Table of Contents
CHAPTER 5 - SERVICE INSTRUCTIONS (CONT’D)
REPLACEMENT PARTS AND ORDERING INFORMATION ...................5-96
ORDERING INFORMATION .......... ... .... .. .. .... .. .. ..... .. .... .. .. .... .. ... .... .. .5-96
CHAPTER 6 - DIAGRAMS
Single-Phase (Non-STORED ENERGY) X-ray Generators, 25, 32 and
40 kW Parts Location Diagram (Sheet 1 of 2) ....................................... 6-3
Single-Phase (Non-STORED ENERGY) X-ray Generators, 25, 32
and 40 kW Parts Location Diagram (Sheet 2 of 2) ................................. 6-4
Single-Phase (STORED ENERGY) X-ray Generators, 20, 32 40 and
50 kW Parts Location Diagram (Sheet 1 of 2) ....................................... 6- 5
Single-Phase (STORED ENERGY) X-ray Generators, 20, 32, 40 and
50 kW Parts Location Diagram (Sheet 2 of 2) ....................................... 6-6
Three-Phase X-ray Generators, 32 and 40 kW Parts
Location Diagram (Sheet 1 of 2) .......................................................... 6-7
Three-Phase X-ray Generators, 32 and 40 kW Parts Location
Diagram (Sheet 2 of 2) ....................................................................... 6-8
Three-Phase X-ray Generators, 50, 65 and 80 kW Parts Location
Diagram (Sheet 1 of 2) ....................................................................... 6-9
Three-Phase X-ray Generators, 50, 65 and 80 kW Parts Location
Diagram (Sheet 2 of 2) ......................................................................6-10
Three-Phase X-ray Generators, 50, 65 and 80 kW (with QMI-HSS)
Parts Location Diagram (Sheet 1 of 2) .................................................6-11
Three-Phase X-ray Generators, 50, 65 and 80 kW (with QMI-HSS)
Parts Location Diagram (Sheet 2 of 2) .................................................6-12
Generator System Cabling Diagram (Internal High-Speed Starter) ..........6-13
Generator System Cabling Diagram (External High-Speed Starter Systems) 6-14
System Cabling Diagram (with Q-Connect Option) ................................6-15
HF Generator to Canon/TechVision System Interconnection Diagram
(Non-Widescreen) .............................................................................6-16
HF Generator to Canon/TechVision System Interconnection
Diagram (Widescreen) ....... ................................................................6-17
17 x 17 Midwest Bucky Interconnection Diagram ..................................6-18
17 x 17 Progeny and L-F Bucky Interconnection Diagram ..................... .6-19
14 x 36 Midwest Bucky Interconnection Diagram ..................................6-20
Single-Phase 208 - 260 VAC Input (Non-STORED ENERGY) X-ray
Generators, 25, 32 and 40 kW Interconnection Diagram .......................6-21
Single-Phase 380 - 480 VAC Input (Non-STORED ENERGY) X-ray
Generators, 25, 32 and 40 kW Interconnection Diagram .......................6-22
Single-Phase (STORED ENERGY) X-ray Generator, 20 kW
Interconnection Diagram ................................................................. ...6-23
Single-Phase (STORED ENERGY) X-ray Generators, 32, 40 and
50 kW Interconnection Diagram .........................................................6-24
Three-Phase (380-480) Input (32, 40 and 50 kW without High
Speed Starter) X-ray Generators Interconnection Diagram ....................6-25
Revision W HF Series X-ray Generators - Service Manual
viii Quantum Medical Imaging
Table of Contents
CHAPTER 6 - DIAGRAMS (CONT’D)
Three-Phase (208-240) Input (32, 40 and 50 kW without High
Speed Starter) X-ray Generators Interconnection Diagram ................... 6-26
Three-Phase (50, 65 and 80 kW with AID High Speed Starter) X-ray
Generators Interconnection Diagram ..................................................6-27
Three-Phase (32, 40 and 50 kW without High Speed Starter, with
Universal AEC Board) X-ray Generators Interconnection Diagram .........6-28
Three-Phase (65 and 80 kW with AID High Speed Starter, Universal
AEC Board) X-ray Generators Interconnection Diagram ........................6-29
Three-Phase (208-240) Input (32, 40 and 50 kW without High Speed Starter, with Universal AEC Board) X-ray Generators Interconnection Diagram AEC6-30 Three-Phase (208-240) Input (50 kW with QMI High Speed Starter) X-ray Generator
Interconnection Diagram .......... .. .... ... .... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .... .. .. . 6-31
Three-Phase (208-240) Input (50 kW with QMI High Speed Starter and
Tomography Interface) X-ray Generator Interconnection Diagram ........ 6-32
Three-Phase (65 and 80 kW with QMI High Speed Starter) X-ray
Generator Interconnection Diagram ....................................................6-33
Three-Phase (65 and 80 kW with QMI High Speed Starter and Tomography
Interface) X-ray Generator Interconnection Diagram ...........................6-34
Operator Control Panel - ODYSSEY/QUEST Wiring Diagram ................... 6-35
Q-Connect Option System Interconnection Diagram .............................6-36
Operator Control Panel Circuit Block Diagram (Non-Q-VISION OCP) ...... . 6-37
Generator Logic Circuit Block Diagram ................................................. 6-38
Filament Regulator Control Circuit Block Diagram ................................. 6-39
KVP & Filament Control Circuit Block Diagram ..................................... . 6-40
Single-Phase (Non-STORED ENERGY) and Three-Phase Line Monitor
Circuit Block Diagram ........................................................................6-41
Single-Phase STORED ENERGY Line Monitor Circuit Block Diagram ........6-42
Interface Board A9 (AY40-023T/AY40-023T1) Schematic Diagram ......... 6-43
Interface Board, Digital Receptor A9 (AY40-034T) Interface Diagram .....6-44
Interface Board, Digital Receptor A9 (AY40-061T) Interface Diagram .....6-45
Power Supply Board A6 (AY40-005T) Schematic Diagram 46
Rotor Circuit Block Diagram ................................................................6-47
QMI High-Speed Starter Circuit Block Diagram .....................................6-48
Logic Board A1 (AY40-006S) Test Point/Jumper Location Diagram .........6-49
KVP Control Board A2 (AY40-003S) Test Point/Jumper Location Diagram 6-50 Filament Control Board A5 (AY40-007T) - Test Point/Jumper Location
Diagram ............................... .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. . 6-51
Power Supply Board A6 (AY40-005T) - Test Point/Jumper Location
Diagram ............................... .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. . 6-52
Line Monitor Board A8 (AY40-022T1), Single-Phase Systems - Test
Point/Jumper Location Diagram ......................................................... 6-53
Line Monitor/Battery Charger Board A8 (AY40-028T), Stored-Energy
Systems - Test Point/Jumper Location Diagram ...................................6-54
Line Monitor Board A8 (AY40-022T2/T3), Three-Phase Systems - Test
Point/Jumper Location Diagram ......................................................... 6-55
Line Monitor Board A8 (AY40-053S2/AY40-053S3), Three-Phase QMI
High-Speed Starter Systems - Test Point/Jumper Location Diagram ...... 6-56
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
ix
Table of Contents
CHAPTER 6 - DIAGRAMS (CONT’D)
Interface Board (No HSS Non-Digital) A9 (AY40-023T) - Terminal
Block/Jumper Location Diagram ..........................................................6-57
Digital Interface Board (No HSS with Digital Receptor) A9 (AY40-034T) -
Terminal Block/Jumper Location Diagram ............................................6-58
Interface Board (HSS Non-Digital) A9 (AY40-061T) Terminal Block/Jumper
Location Diagram ............................................................................. .6-59
Digital Interface Board (HSS with Digital Receptor) A9 (AY40-062T)
Terminal Block/Jumper Location Diagram ............................................6-60
Rotor Drive Board A10 (AY40-013T) - Terminal Block Location Diagram .6-61 Standard AEC Board A11 (AY40-031S) - Test Point/Jumper Location
Diagram ...........................................................................................6-62
Universal AEC Board A11 (AY40-027S) - Test Point/Jumper Location
Diagram ...........................................................................................6-63
OCP Control Board A16A1 (AY40-004S1) - Test Point/Jumper
Location Diagram ............................................................................. .6-64
OCP Interface Board A1A1 (AY40-073S) - Test Point/Jumper Location
Diagram ...........................................................................................6-65
QMI High-Speed Starter Control Board A20A1 (AY40-035S) - Test
Point/Potentiometer Location Diagram ................. ...............................6-66
QMI High-Speed Starter Driver Board A20A2 (AY40-036S) - Test
Point/Fuse Location Diagram ..............................................................6-67
APPENDIX A - AEC CALIBRATION WITH DIGITAL RECEPTOR OR CR FILM PLATE
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. A-3
SPECIAL CANON DR SYSTEM CONFIGURATION INSTRUCTIONS ....... A-3
AEC CALIBRATION WITH DIGITAL RECEPTOR OR CR FILM PLATE
INSTALLED .................................................................................. A-4
AEC Calibration Instructions - One Ion Chamber Per Panel .........A-8
AEC Calibration Instructions - Single Panel, Dual Ion Chambers A-13
APPENDIX B - CALIBRATION (AGFA DIREX)
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. B-3
REQUIRED TOOLS AND MATERIALS ............................................... B-4
SYSTEM CONFIGURATION/CALIBRATION ............................................. B-4
ACCESSING SERVICE MODE ...... .. ... .... .. .. .... .. .. ..... .. .. .... .. .... .. ... .... .. .. B-4
FACTORY RESET ........................................................................... B-8
Factory Reset .......................................................................... B-8
Backup Configuration Utility ...................................................... B-9
GENERATOR CONFIGURATION ..................................................... B-10
Generator Settings ................................................ ................. B-13
Tube Settings ........................................................................ B-15
Ion Chamber Configuration .............. .... .. .... ... .. .... .. .. .... .. ... .... .. B-17
Film/Screen Default Settings ................................................... B-21
Revision W HF Series X-ray Generators - Service Manual
x Quantum Medical Imaging
Table of Contents
APPENDIX B - CALIBRATION (AGFA DIREX) (CONT’D)
Copy Film Screen Utility ..........................................................B-23
GENERATOR CALIBRATION ..........................................................B-25
General Calibration Procedure Notes ........................................B-26
A/D Calibration ................ .. ... .. .. .. .. .... .. .. ... .. .. .. .... .. .. .. ... .. .. .. .... . B-27
kV Calibration ........................................................................ B-30
Filament Offset Calibration ......................................................B-32
MA Calibration ......................................................................B-37
AEC CALIBRATION ........... .. .... ..... .. .... .. .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .B-40
Ion Chamber Balancing (Three-field Chambers) .......................B-43
AEC MAS Balance Verification .......................... .... .. .. ..... .. .. .... .. .B-45
Film Screen Selection .............................................................B-47
The AEC Calibration Menu ......................................................B-50
Optical Density Calibration ......................................................B-51
AEC Optical Density Calibration .......... .... ... .. .... .. .. .... .. ..... .. .. .... . B-52
AEC KV Calibration Procedure .................................................B-55
APPENDIX C - LOG4VIEW ERROR LOGGER APPENDIX D - (QG-DIG-CXDI CANON/QUANTUM
INTEGRATION) CALIBRATION
OVERVIEW .................. .............................................. ......................... D-3
REQUIRED TOOLS AND MATERIALS ............................................... D-4
ACCESSING GENERATOR SERVICE TOOLS ...................................... D-4
Accessing Generator Service Tools - Q-VISION Operator
Control Panel ......................................................................... D-5
Accessing Generator Service Tools - Workstation-Based
Operator Control Panel ........................................................... D-7
SYSTEM CONFIGURATION/CALIBRATION ............................................D-10
GENERATOR CONFIGURATION .....................................................D -12
Generator Settings Tab .............. .............................................D-12
Generator Calibrations Tab .....................................................D-13
Tube Settings Tab ..................................................................D-13
Physical Receptors Tab ...................... .. .. ..... .. .. .... .. .. ..... .. .. .... .. .D-15
Receptors Tab .......................................................................D-16
Ion Chambers Tab ...................... .. .... .. ..... .. .. .... .. .. .... ... .. .... .. .. .D-18
Licensing Tab ........................................................................D-20
GENERATOR CALIBRATION ..........................................................D-21
A/D Tab .................... .. .. .... ... .. .. .. .. .. .. .... ... .. .. .. .. .. .... .. ... .. .. .. .. .. .D-24
kVp Calibration Tab ............... .. .... .. .. .... .. ... .... .. .... .. .. ..... .. .. .... .. .D-28
Filament Offset Calibration ......................................................D-31
MA Calibration Tab .................................................................D-41
AEC CALIBRATION ........... .. .... ..... .. .... .. .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .D-45
Ion Chamber Balancing (Three-field Chambers) .......................D-47
AEC MAS Balance Verification .......................... .... .. .. ..... .. .. .... .. .D-50
The AEC Calibration Tab ............................... ..........................D-52
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
xi
Table of Contents
APPENDIX D - (QG-DIG-CXDI CANON/QUANTUM INTEGRATION) CALIBRATION (CONT’D)
Base OD Calibration Procedure .......................... ..................... D-56
AEC kVp Calibration Procedure ...............................................D-58
Copy Receptor Calibration Tab ................................................ D-62
Revision W HF Series X-ray Generators - Service Manual
xii Quantum Medical Imaging
Chapter
1
SPECIFICATIONS
1-1
1-2
Chapter 1 Specifications

SYSTEM CONFIGURATIONS

The tables below show model numbers of components in each type of HF Series X-ray Generator system (hereinafter referred to as the "HF Series X-ray Generators").
Note: The X-Ray Control Models shown in the following tables are not used if the HF generator system is integrated with computed radiographic (CR) or digital radiographic (DR) workstation computer (e.g., Agfa NX, Canon CXDI, etc.).
Single-Phase Model Generators
TM
SYSTEM
TYPE
ODYSSEY HF 32 kW
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 40 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
SYSTEM
TYPE
CATALOG
NO.
QG-3200
QG-4000
QGV-32 QG-32 QG-32G QGV-32-5 QG-32-5 QG-32G-5 QGV-40 QG-40 QG-40G QGV-40-5 QG-40-5 QG-40G-5
Three-Phase Model Generators
CATALOG
NO.
SYSTEM
MODEL
QG-32 QG-32G QG-3200
QG-32-5 QG-32G-5 QG-3200
QG-40 QG-40G QG-4000
QG-40-5 QG-40G-5 QG-4000
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL*
QGV-32
QGV-40
X-RAY
CONTROL
MODEL
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 50 kW
QG-3200
QG-4000
QG-5000
QG-32-2 QG-32G-2
QG-3200
QG-32-3 QG-32G-3 QG-40-2 QG-40G-2
QG-4000
QG-40-3 QG-40G-3 QG-50-2 QG-50G-2
QG-5000
QG-50 QG-50G
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-3
Chapter 1 Specifications
Three-Phase Model Generators
SYSTEM
TYPE
ODYSSEY HF 65 kW
ODYSSEY HF 80 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
Q-VISION HF 50 kW
Q-VISION HF 65 kW
Q-VISION HF 80 kW
CATALOG
NO.
QG-6500 QG-65 QG-65G QG-6500
QG-8000 QG-80 QG-80G QG-8000
QGV-32-2 QG-32-2 QG-32G-2 QGV-32-3 QG-32-3 QG-32G-3 QGV-40-2 QG-40-2 QG-40G-2 QGV-40-3 QG-40-3 QG-40G-3 QGV-50-2 QG-50-2 QG-50G-2 QGV-50 QG-50 QG-50G
QGV-65 QG-65 QG-65G QGV-65
QGV-80 QG-80 QG-80G QGV-80
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL
QGV-32
QGV-40
QGV-50
STORED ENERGY (SE) Model Generators
SYSTEM
TYPE
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 50 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
Q-VISION HF 50 kW
CATALOG
NO.
QG-3200-SE QG-32-SE QG-32G-SE QG-3200
QG-4000-SE QG-40-SE QG-40G-SE QG-4000
QG-5000-SE QG-50-SE QG-50G-SE QG-5000
QGV-32-SE QG-32-SE QG-32G-SE QGV-32-SE
QGV-40-SE QG-40-SE QG-40G-SE QGV-40-SE
QGV-50-SE QG-50-SE QG-50G-SE QGV-50-SE
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL
Revision W HF Series X-ray Generators - Service Manual
1-4 Quantum Medical Imaging
Chapter 1 Specifications

PHYSICAL SPECIFICATIONS

The following are physical specifications of the HF Series X-ray Generators:

GENERATOR CABINET SPECIFICATIONS

Width: 24.0 inches (609.6 mm) Depth: 20.0 inches (508.0 mm) Height: 34.7 inches (881.4 mm) Weights:
Single-Phase (non-SE units): 298 lbs. (135.2 kg) 32, 40 and 50 kW Stored
Energy (SE) units: 431 lbs. (195.5 kg) 50, 65, 80 kW Three-Phase
units): 380 lbs. (712.4 kg) w/o internal HSS
388 lbs. (176.0 kg) w/internal HSS

OPERATOR CONTROL PANEL SPECIFICATIONS

Note: X-Ray Control Panels below are utilized with HF gener­ator systems integrated with computed radiographic (CR) or digital radiographic (DR) workstation computers (e.g., Agfa NX, Canon CXDI, etc.).
The Q-VISION all-in-one operator control is not powered from the generator cab­inet; it requires a separate 100-240 VAC, 3A power input.
ODYSSEY HF Control Panel:
Width: 17.9 inches (454.7 mm) Depth: 12.25 inches (311.2 mm) Height (without pedestal): 3.3 inches (83.8 mm) Height (with pedestal): 36.5 inches (927.4 mm) Weight (without pedestal): 5 lbs. (2.3 kg) Weight (with pedestal): 25 lbs. (11.3 kg)
Q-VISION HF Control Panel:
Width: 18.7 inches (476.0 mm) Depth (without stand): 1.9 inches (49.0 mm) Depth (with stand): 7.125 inches (181.0 mm) Height: 14.4 inches (365.0 mm) Weight: 13 lbs. (5.9 kg)

SYSTEM SPECIFICATIONS

System Compatibility: Compatible with all Quantum Medical
Tube Compatibility: Toshiba Models E7239X, E7242X, E7252X,
Imaging tube stands, overhead tube cranes, radiographic tables and wall stands
E7254X, E7255X, E7813X, E7869X Varian Models RAD-8, RAD-13, RAD-14, RAD-21, RAD-44, RAD-56, RAD-60, RAD-68, RAD-74, RAD-92, A192, A272, A292
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-5
Chapter 1 Specifications
Collimators: Huestis Manual, Model 150MC
Note: The HF Series radiographic systems are compatible with any
automatic collimating system that provides an isolated nor­mally open switch that closes when the Automatic Collima­tion System presents an “EXPOSURE READY” mode to the control; they are also compatible with any beam limiting device that complies with 21 CFR Subchapter J intended for radiographic equipment.

POWER REQUIREMENTS

Dunlee Models PX 1302, PX 1429, PX 1436, PX 1312, DU 309, DA 1092, MX100 replacement tube for GE Radiographic Systems
(For tube types not listed, consult factory.) Huestis Automatic, Model 150PBL
Huestis Selectable, Model 150A, Progeny Linear MC150 (Manual) Progeny Linear II and Linear IV (Automatic)
In order to assure the proper operation of our equipment and subsequently, the produc­tion of high-qua lity radiographs, it is es se ntial that all pow er r e q u irements, as sp ec if ied in the tables that follow, are suitably met by the facility in which the equipment is installed. This includes usage of the proper size conductor between distribution transformer and dis­connect switch, connection to a supply transformer of sufficient capacity (specified in terms of KVA), insuring the AC input voltage tolerance meets specification (±5% or ±10% depending on model), and voltage regulation of 5% at full load is maintained. All wiring and grounding at the installation site must be in accordance with applicable National Elec­tric Code (NEC) and/or local electrical codes.
THe generator requires a 3-phase "Y" connection, measured line-to-line, balanced to earth or neutral. The "Y" connection can be 4-wire (L1, L2, L3, Neutral) and GND or 3-wire (L1, L2, L3) and GND. The 4-wire connection may be required by local electri­cal code and in connection where a roughly balanced 3-phase source exists (e.g., when using a delta-to-Y transformer).
Note that failure to comply with the above requirements not only may result in improper generator operation and/or performance, but it will also void the warranty, as the warranty is exclusive of the following conditions:
"Failure of customer to prepare the site or provide power requirements or
operating environmental conditions in compliance with any applicable instructions or recommendations of QMI.
"Failure of customer to provide the proper incoming power required to support
the equipment in accordance with the recommendation of QMI.
"Improper or extraordinary use of the Product, improper maintenance of the
Product, or failure to comply with any applicable instructions or recommendations of Quantum Medical Imaging, LLC.
Revision W HF Series X-ray Generators - Service Manual
1-6 Quantum Medical Imaging
Chapter 1 Specifications
NOTE: All single-phase x-ray generator input ratings assume the use of a
single-phase distribution transformer. When utilizing two legs of a 3­phase transformer, review the National Electrical Code and any appli­cable local electrical code(s) to ensure the correct rating distribution transformer (in terms of KVA) is used to assure line stability. This configuration would require an increase in the overall power (KVA) of the distribution transformer, depending on whether the input line transformer is Wye or Delta type. Under no circumstances should a Delta type transformer be used for input power to a 3-phase genera­tor.

SINGLE-PHASE X-RAY GENERATORS

Non-Stored Energy Single-phase x-ray genera tors require three-wire line inputs in either of the following two configurations:
one hot wire, one neutral wire, and one earth ground wire
two hot wires and one earth ground. Refer to Table 1-1 for power requirements of single-phase (non-stored energy)
HF Series Generators.
25 kW Single Phase (Model QG-25)
Line Voltage: 208-260 VAC (±5%), 100 Amps, 50/60 Hz Momentary Current: 200 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
25 kW Single Phase (Model QG-25-5)*
Line Voltage: 380 - 480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 110 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32)
Line Voltage: 208-260 VAC (±5%), 120 Amps, 50/60 Hz Momentary Current: 240 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32-5)*
Line Voltage: 380 - 480 VAC (±10%), 75 Amps, 50/60 Hz Momentary Current: 140 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-7
Chapter 1 Specifications
40 kW Single Phase (Model QG-40)
Line Voltage: 208-260 VAC (±5%), 150 Amps, 50/60 Hz Momentary Current: 300 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Single Phase (Model QG-40-5)*
Line Voltage: 380-480 VAC (±10%), 100 Amps, 50/60 Hz Momentary Current: 170 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
40 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
50 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
Note: Models QG-25-5, QG-32-5, and QG-40-5 are not UL classified.
Revision W HF Series X-ray Generators - Service Manual
1-8 Quantum Medical Imaging
Chapter 1 Specifications

THREE-PHASE X-RAY GENERATORS

Three-phase x-ray generators require three "hot" wires and one earth ground wire. Refer to Tables 1-2 and 1-3 for power requirements of three-phase HF Series Generators.
32 kW Three Phase (Model QG-32-2)
Line Voltage: 208-240 VAC (±5%), 70 Amps, 50/60 Hz Momentary Current: 140 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Three Phase (Model QG-32-3)
Line Voltage: 380-480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 117 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Three Phase (Model QG-40-2)
Line Voltage: 208-240 VAC (±5%), 85 Amps, 50/60 Hz Momentary Current: 170 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Three Phase (Model QG-40-3)
Line Voltage: 380-480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 125 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
50 kW Three Phase (Model QG-50-2)
Line Voltage: 208-240 VAC (±5%), 110 Amps, 50/60 Hz Momentary Current: 220 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
50 kW Three Phase (Model QG-50-3)
Line Voltage: 380-480 VAC (±10%), 80 Amps, 50/60 Hz Momentary Current: 144 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-9
Chapter 1 Specifications
65 kW Three Phase (Model QG-65)
Line Voltage: 380-480 VAC (±10%), 90 Amps, 50/60 Hz Momentary Current: 190 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
80 kW Three Phase (Model QG-80)
Line Voltage: 380-480 VAC (±10%), 100 Amps, 50/60 Hz Momentary Current: 200 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation

HIGH-SPEED STARTER POWER REQUIREMENTS

A High-Speed Starter (HSS) unit is optional on Models QG-40 and QG-50, and is standard equipment on Models QG-65 and QG-80.
The external HSS (QG-HSS) configuration requires a separate 200 - 240 VAC, 20A line input. Note that a separate 200- 240 VAC line input is erators equipped with Internal High Speed Starter (Q-HSS) configuration.
not required
on gen-
Revision W HF Series X-ray Generators - Service Manual
1-10 Quantum Medical Imaging
Chapter 1 Specifications
QG-25
QG-32
QG-40
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAG E (VAC)
DISCONNECT
TO GENE R ATO R
15 FT. (5 M) M A X .
*SERVICE
RATING
*RECOMMENDED
DISTR IBU T IO N
TRANSFORMER
CAPACITY
MAX. INPUT LINE
RESISTANCE
(AT FULL-LOAD
RMS)
NOTE: All installations must meet the requirements of local electrical codes. The above values are recommended as typical (unless otherwise specified).
*
1.
±5%)
Line Regulation: 5% at full load.
7. Ground wire size is 4 AWG.
8. Specifications are subject to change.
9. Specifications for single-phase “stored en ergy” Models QG-20-SE, QG-32-SE, QG-40-SE and QG-50-SE are as follows: 105-130 VAC (±10%), 10 Amps, 50/60 Hz 210-250 VAC (±10%), 10 Amps, 50/60 Hz
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. Input Voltage: 208-257 ( VAC 50/60 Hz (configured at time of installation).
6.
QG-250
QG-320, QG-3200, QGV-32
QG-400, QG-4000, QGV-40
GENERATOR
CATALOG NO.
208
25
32
40
240
208
240
208
240
#4 AWG
#4 AWG
#2 AWG
#2 AWG
#1 AWG
#1 AWG
100 AMPS
80 AMPS
120 AMPS
100 AMPS
150 AMPS
120 AMPS
30 KVA
30 KVA
40 KVA
40 KVA
50 KVA
50 KVA
0.06 OHMS
0.08 OHMS
0.04 OHMS
0.05 OHMS
0.035 OHMS
0.04 OHMS
Table 1-1. HF Series Generators -
Electrical Wiring Requirement s (208 - 260 VAC Single-Phase Input Units)
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-11
Chapter 1 Specifications
*NOTE: All installations must meet the requirements of local electrical codes.
The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 380-480 (±10%) VAC 50/60 Hz (configured at time of installation)
. Ground wire size is 4 AWG.
7. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAGE (VAC)
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*SERVICE
RATING
*RECOMMENDED
DISTRIBUTION
TRANSFORMER
CAP ACITY
MAX. INPUT LINE
RESISTANCE
(A T F U L L-LOAD
RMS)
GENERATOR
CATALOG NO .
25
32
40
QG-25-5
QG-32-5
QG-40-5
QG-250-5
QG-320-5, QG-3200-5, QGV-32-5
QG-400-5, QG-4000-5, QGV-40-5
380
#6 AWG 60 AMPS
32 KVA
0.028 OHMS
415 #6 AWG 55 AMPS 32 KVA 0.028 OHMS
480 #6 AWG 40 AMPS 32 KVA 0.032 OHMS
380 #6 AWG 75 AMPS 40 KVA 0.024 OHMS
415 #6 AWG 70 AMPS 40 KVA 0.024 OHMS
480
#6 AWG 60 AMPS
40 KVA
0.028 OHMS
380 #4 A WG 100 AMPS 50 KVA 0.015 OHMS
415
#4 AWG 90 AMPS
50 KVA
0.020 OHMS
480
#6 AWG 70 AMPS
50 KVA
0.024 OHMS
Table 1-2. HF Series Generators -
Electrical Wiring Requirements (380 - 480 VAC Single-Phase Input Units)
Revision W HF Series X-ray Generators - Service Manual
1-12 Quantum Medical Imaging
Chapter 1 Specifications
QG-32-3,
QG-40-3
QG-50
QG-65
QG-80
32
(QG-32
-3)
40
(QG-40-3)
50
65
80
G E
N E R A
T O
R
M O
D E L
N O .
MAX. POWER
O UTPUT (kW)
N O M
I
N A L
L I
N E
V O L
T A G
E (
V A C )
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*
S E R V I
C E
R A T I
N G
*
R E C O M
M E
N D E
D
D I
S T R I
B U T I
O N
T R A N
S F O
R M
E R
C A P
A C I
T Y
MAX. INPUT LINE
RESISTANCE
(AT FULL-LO AD RMS)
380#6 AWG60 AMP
S
50 KV
A
0.23 OHMS380#4 AWG80 AMPS65 KVA0.17 OHMS380#4 AWG90 AMPS80 KVA0.12 OHMS380#4 AWG100 AMPS100 KVA0.10 OHMS440#6 AWG60 AMPS50 KVA0.33 OHMS440#6 AWG80 AMPS65 KVA0.20 OHMS440#4 AWG90 AMPS80 KVA0.17 OHMS440#4 AW
G
S
100 KVA
0.14 OHMS
480#6 AWG50 AMPS50 KVA0.4 OHMS480#6 AWG75 AMPS65 KVA0.24 OHMS480#4 AWG90 AMPS80 KVA0.19 OHMS480
#4 AWG100 AMPS100 KVA0.16 OHM
S
QG-320, QG-3200, QGV
-32-3, QG-400, QG-4000, QG
V-40-3
QG-500, Q
G-5000,
Q
GV-50
QG-6500, QGV-65
QG-8000, QGV-80
G E
N E
R A T O R
C A T
A L
O G
N O .
*NOTE: All installations must meet the requirements of local electrical codes.
The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 380-480 (±10%) VAC 50/60 Hz (configured at time of installation)
. Voltage is three-phase Y-configuration, measured line-to-line, balanced to earth
ground. (Note: Additional Neutral wire may be used if required by local electrical code.)
7. Ground wire size is 4 AWG.
8. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
Table 1-3. HF Series Generators - Electrical Wiring Requirements
(380-480 VAC Three-Phase Input Units)
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-13
Chapter 1 Specifications
QG-32-2
QG-40-2
QG-50-2
208
32
40
50
240
208
240
208
240
#6 AWG
#6 AWG
#4 AWG
#6 AWG
#4 AWG
#4 AWG
70 AMPS
60 AMPS
90 AMPS
70 AMPS
110 AMPS
90 AMPS
40 KVA
40 KVA
50 KVA
50 KVA
65 KVA
65 KVA
0.026 OHMS
0.023 OHMS
0.017 OHMS
0.024 OHMS
0.014 OHMS
0.016 OHMS
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAGE (VAC)
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*SERVICE
RATING
*RECOMMENDED
DISTRIBUTION
TRANSFORMER
CAPACITY
MAX. INPUT LINE
RESISTANCE
(AT FULL-LOAD
RMS)
QG-320, QG-3200, QGV-32-2
QG-400, QG-4000, QGV-40-2
QG-500, QG-5000, QGV-50-2
GENERATOR
CATALOG NO.
*NOTE: All installations must meet the requirements of local electrical codes. The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 208-257 (±5%) VAC 50/60 Hz (configured at time of installation)
. Voltage is three-phase Y-configuration, measured line-to-line, balanced to earth
ground. (Note: Additional Neutral wire may be used if required by local electrical code.)
Line Regulation: 5% at full load.
8. Ground wire size is 4 AWG.
9. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
7.
Table 1-4. HF Series Generators -
Electrical Wiring Requirements (208-240 VAC Three-Phase Input Units)
Revision W HF Series X-ray Generators - Service Manual
1-14 Quantum Medical Imaging
Chapter 1 Specifications

GENERATOR AUXILIARY POWER SUPPLIES

The HF Series X-ray Generator supplies the following supply voltages for powering X-ray room equipment:
24 VDC, 1.5A
24 VAC, 150W (collimator lamp not to exceed 150W)
115 VAC, 3A
NOTE: The above voltage outputs are not to be used as power

PERFORMANCE SPECIFICATIONS

Table 1-5 provides performance specifications of single-phase (non-stored energy) HF Series Generators.
Table 1-5. HF Series Single-Phase Generator Performance Specifications
sources for collimator lamps or for inductive loads (e.g., motorized tables, solenoids, etc.)
Cat. No.
N/A
QG-3200,
QGV-32
Models
N/A
QG-32,
QG-32-5
Maximum kW N/A 32 40 mA Stations;
Small Focus (S) Large Focus (L)
kVp Range (kVp) N/A 40-125 40-125 kVp increments (kVp/step) N/A 1.0 1.0 150 kVp option N/A Yes Yes Time Range (sec.)* N/A 0.001 - 6.3 0.001 - 6.3 Minimum Exposure Time
(sec.) mAs Range† High-Speed Starter
Ripple Voltage (output) N/A 5% 5%
* Time measured at 75% of the peak kVp waveform
NOTE: AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS.
mAs is tube dependent
N/A
N/A 0.001 0.001 N/A 0.025-600 0.025-600
N/A No No
25S, 75S, 150S
100L, 200L, 250L,
320L, 400L, 500L
Cat. No.
QG-4000,
QGV-40
Models
QG-40,
QG-40-5
25S, 75S, 150S
100L, 200L, 250L, 320L,
400L, 500L
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-15
Chapter 1 Specifications
Refer to Table 1-6 for performance specifications of single-phase stored energy (SE) HF Series Generators.
Table 1-6. HF Series Single-Phase STORED ENERGY Generator Performance Specifications
Cat. No.
N/A
QG-3200-SE,
QGV-32-SE
Model QG-32-SE
N/A
(STORED ENERGY)
Maximum kW N/A 32 40 50 mA Stations;
Small Focus (S) Large Focus (L) N/A
kVp Range (kVp) N/A 40-125 40-125 40-125 kVp increments
(kVp/step) 150 kVp optional? N/A No No No Time Range
(sec.)* Minimum Expo-
sure Time (sec.)* mAs Range High-Speed Starter
Ripple Voltage (output)
N/A 1.0 1.0 1.0
N/A 0.001 - 6.3 0.001 - 6.3 0.001 - 6.3
N/A 0.001 0.001 0.001 N/A 0.025-400 0.025-500 0.025-500
N/A No No No N/A 5% 5% 5%
25S, 50S, 75S, 150S
100L, 160L, 200L,
250L, 320L, 400L
Cat. No.
QG-4000-SE,
QGV-40-SE
Model QG-40-SE
(STORED ENERGY)
25S, 50S, 75S, 150S
100L, 200L, 250L,
320L, 400L, 500L
Cat. No.
QG-5000-SE,
QGV-50-SE
Model QG-50-SE
(STORED ENERGY)
25S, 50S, 75S,
150S 100L, 200L, 250L, 320L, 400L, 500L,
600L
* Maximum exposure time is a function of kV and mAs settings and the age of the
batteries; Time measured at 75% of the peak kVp waveform NOTE: AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS.
mAs is tube dependent; the generator may not reach maximum mAs due t o tube type. In AEC mode, the mAs is limited to 600.
Revision W HF Series X-ray Generators - Service Manual
1-16 Quantum Medical Imaging
Chapter 1 Specifications
Refer to Table 1-7 for performance specifications of three-phase HF Series Gener­ators.
Table 1-7. HF Series Three-Phase Generator Performance Specifications
Cat. No.
QG-3200-2,
QG-3200-3
QGV-32-2,
QGV-32-3
Model
Model QG-32-2,
QG-32-3
Maximum kW 32 40 50 65 80 mA Stations; Small Focus (S) Large Focus (L)
kVp Range (kVp) 40-125
kVp increments (kVp/step) 150 kVp optional? Time Range (sec.)* Minimum Expo­sure Time (sec.) mAs Range† 0.025-600 0.025-600 0.025-800 0.025-800 0.025-800 High-Speed
Starter Ripple Voltage (output)
25S, 75S, 150S
100L, 200L, 250L, 320L,
400L, 500L
1.0 1.0 1.0 1.0 1.0
Yes Yes Yes Standard Standard
0.001 - 6.3 0.001 - 6.3 0.001 - 6.3 0.001 - 6.3 0.001 - 6.3
0.001 0.001 0.001 0.001 0.001
No Option Option Yes Yes
5% 5% 5% 5% 5%
Cat. No.
QG-4000-2,
QG-4000-3
QGV-40-2,
QGV-40-3
Model
Model QG-40-2,
QG-40-3
25S, 75S, 150S
100L, 200L, 250L, 320L,
400L, 500L
40-125
(40-150 with
QG-150 option)
Cat. No.
QG-5000-2,
QG-5000-3
QGV-50,
QGV-50-2
Model
QG-50-2,
QG-50
25S, 75S, 150S
100L, 200L, 320L, 400L,
500L, 650L
40-125
(40-150 with
QG-150 option)
Cat. No.
QG-6500,
QGV-65
Model
Model QG-65
25S, 75S, 150S
100L, 200L, 320L, 400L,
500L, 650L, 800
40-150 40-150
25S, 75S, 150S
Cat. No.
QG-8000,
QGV-80
Model
Model QG-80
100L, 200L, 320L, 400L, 500L, 650L,
800
* Time measured at 75% of the p eak kVp waveform
NOTE: AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS.
mAs is tube dependent; the generator may not reach maximum mAs due to tube type. In AEC mode, the mAs is limited to 600.
High-Speed Starter Duty Cycle: do no t to exceed two activations within any one minute period
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-17
Chapter 1 Specifications
Accuracy Specifications
kVp Accuracy: kVp = ± ( 4.0% kV )
mA Accuracy: mA = ± ( 5.0% mA + 1 mA )
Time Accuracy:
mAs Accuracy:
mAs = ± ( 5.0% mAs + 0.5 mAs )
time = ± ( 2.0% time + 1 msec. ) time = ± ( 1.0% time + 1 msec. ) time = ± ( 0.5% time + 1 msec. )
Table 1-8. Accuracy Specifications
Revision W HF Series X-ray Generators - Service Manual
1-18 Quantum Medical Imaging
Chapter 1 Specifications

SYSTEM OPTIONS

The following lists the available options for HF Series X-Ray Generators. Note that some options, such as QG-150 and Q-HSS, may only be used on certain generator models (refer to Tables 1-5 through 1-7 for compatibility):
R80-HS: Exposure Hand Switch and R80-FS Foot Switch
QG-PDL: Operator Control Panel Pedestal Mount (used with ODYSSEY control panel)
QG-WM: Operator Control Panel Wall Mount (used with ODYSSEY control panel)
QGV-WM: Operator Control Panel Wall Mount (used with Q-VISION control panel)
QG-150: 150 kV High Voltage Transformer (standard equipment on Models QG-65 and QG-80)
Q-HSS: Internal High (Dual) Speed Starter Unit (standard equipment on Models QG-65 and QG-80
Q-Connect: Enables ethernet connection to a PC computer using Quantum’s "QBus" network integration software; required on systems utilizing a touch­screen operator control panel
QG-28VPS: Internally mounted Auxiliary +28 VDC/8A Power Supply (only available with three-phase input generators)
QG-TOMO: Provides interface for operation with Siemens and Pausch tomographic systems (not available with Q-VISION HF Series OCP’s)
TechVision: Supplementary x-ray control panel, mounts on tube support device. Refer to the TechVision Remote Control Panel Service and Operator’s Manuals for complete installation and operation instructions.
NOTE: The use of ACCESSORY equipment not complying with the
equivalent safety requireme nts of this equipment may lead to a reduced level of safety of the resulting system. Consider­ation relating to the choice shall include:
-use of the accessory in the PATIENT VICINITY
-evidence that the safety certification of the ACCESSORY has been performed in accordance with the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized national standard.

SYSTEM OPERATING ENVIRONMENT

Ambient Temperature: +10°C to +40°C Relative Humidity: 30 to 75%, non-condensing Atmospheric Pressure: 700 hPa to 1060 hPa Typical Heat Output of
Generator Cabinet: 3600 BTU per hour Typical Heat Output of
operator control panel (OCP): 130 BTU per hour
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-19
Chapter 1 Specifications
NOTE: The above calculations are typical, based upon average usage.
Air circulation around the electronic cabinet must be provided. This includes the sides, back, front and top of cabinet.

NON-OPERATING ENVIRONMENT

Ambient Temperature: -18°C to +70°C Relative Humidity: 20 to 95%, non-condensing Atmospheric Pressure: 500 hPa to 1060 hPa
Figure 1-1. HF Series X-ray Generator Cabinet Dimensions
Revision W HF Series X-ray Generators - Service Manual
1-20 Quantum Medical Imaging
Chapter 1 Specifications
Figure 1-2. Odyssey HF Series X-ray Generator Operator Control
Panel Dimensions
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-21
Chapter 1 Specifications
18.7
(476.0)
14.4
(365.0)
7.125
(181.0)
Figure 1-3. Q-VISION HF Series X-Ray Generator Operator Control
Panel Dimensions
Revision W HF Series X-ray Generators - Service Manual
1-22 Quantum Medical Imaging
Chapter
2
ASSEMBLY &
INSTALLATION
2-1
2-2
Chapter 2 Assembly & Installation

OVERVIEW

This chapter provides procedures for installing the HF Series x-ray generator, from pre-installation guidelines to final system inspection prior to operational readiness. These procedures should be reviewed carefully before beginning the actual installa­tion. Preparing the X-r ay generator f or operation require s completion of the following tasks:
Installing the X-ray generator cabinet
Installing the Operator Control Panel (OCP)
Making the Electrical Cable Connections
•Power-up
Calibrating the generator
Inspecting the system
Verifying programmed techniques
WARNING! Equipment installation and servicing procedures should be performed by properly trained and qualified service personnel only.

PRE-INSTALLATION GUIDELINES

While site planning is the customer's responsibility, the following are recom­mended guidelines.
Verify Equipment Location
In radiology site preparation, it is essential to verify the location and space allocated for the equipment is sufficient. See Figures 1-1 through 1-3 for HF Series x-ray generator dimensions. The room plan should show the following for the equipment:
All working and parked positions
Overall dimensions
Total equipment weights and power outputs
Structural mounting and attachment methods
Ensure sufficient air circulation space is provided between the rear side of Generator Cabinet and wall (4.0" min.). The floor where the generator is installed should be flat and approximately level (minor shimming is accept­able) and capable of supporting a load of 500 lbs. (226.8 kg).
If the generator is equipped with the "Q-Connect option" (ethernet adapter for connection to Q-VISION or for integration with a Digital Radiographic Sys­tem), the associated computer monitor functions as the generator operator control panel (OCP). If the workstation is qualified to standard UL 60950/IEC
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-3
Chapter 2 Assembly & Installation
2.5 m
1.5 m
1.5 m
1.5 m
60950 for Information Technology Equipment, the workstation must be placed outside the radius of the patient environment, shown in Figure 2-1. Only devices conforming to UL 60601-1/IEC 60601-1 may be placed within this environment. Consult with the user of the system when determining the location for the workstation computer; consider all areas the patient may access, including the room entrance and/or exit, when establishing the patient environment.
Figure 2-1. Patient Environment
Revision W HF Series X-ray Generators - Service Manual
2-4 Quantum Medical Imaging
Chapter 2 Assembly & Installation
NOTE
Examine all car­tons an d cr a t es carefully at time of delivery. If damage is appar­ent, have del i v­ery driver write a "Damaged Ship­ment Note" on copies of freight bill, sign it, and file appropriate carrier claim. Should you dis­cove r co nc e aled damage, immedi­ately notify the transporting age nt and ask for an "Inspection of Damage". Carrier will not accept concealed dam­age claim if filed after 15 days from date of receipt of mer­chandise.
Electrical Requirements
In addition to providing for the physical requirements of the equipment, it is also vital to prepare for the power requirements. Refer to Chapter 1, SPECIFICATIONS for model-dependent power specifications. This information should be used to prepare system wiring and cabling.

UNPACKING

The HF Series x-ray generator is shipped in a single crate. To unpack, pro­ceed as follows:
1. Remove packing list and use it as a guide to open the remaining cartons. Do not dispose of packing material until packing list is matched with actual parts received. Should there be a shortage or damage, notify Quantum Medical Imaging, LLC.’s customer service department immedi­ately . The manufacturer is relieved of any responsibility for damage dur­ing shipment after unit is picked up by the carrier.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-5
WARNING! The X-ray generator cabinet can weigh as much as 430 lbs. (195 kg), depend­ing on configuration. Use lifting straps and hoist or other suitable lifting device to trans­port generator. The H.V. Tank weighs approximately 100 lbs. and requires two per­sons to lift.
2. Remove the Generator Cabinet from its crate (lifting straps and hoist or other suitable lifting device is recommended).
3. Using a hand truck from the rear side of the cabinet, lift the generator and place it as close to its final installation location as possible.
WARNING! When unpacking the generator; keep upright at all times. Verify there is no
damage caused during shipping, with no apparent oil leakage from H.V. Tank; check that the tank lid and oil plug (in tank lid) are secure.
4. Locate the OCP and Pedestal (optional) or Wall Mount Bracket (optional) and place on the side (not applicable on Q-VISION equipped systems or with inte­grated systems (e.g., Agfa, Canon, etc.).
Chapter 2 Assembly & Installation

INSTALLATION

The following paragraphs provide information necessary for installing the Generator Cabinet and Operator Control Panel (OCP), including all required external equipment cable connections with the X-ray generator.
CAUTION! This equipment contains electrostatic sensitive devices. Observe proper grounding pre­cautions before handling components or printed circuit boards.

REQUIRED TOOLS AND MATERIALS

The following lists the tools and materials required to perform the procedures included in this chapter:
Hoist with lifting straps (or other suitable lifting device)
Tape measure
Standard hand tools
Drill bit, suitable size for installing floor mounting hardware
Bubble level
Various wiring connectors and cable ties

GENERATOR CABINET SETUP

1. Loosen six (6) phillips head screws (3 each on left and right sides of cabinet) and remove three (3) phillips head screws (in lower front side) securing gen­erator cabinet cover to generator chassis (see Figure 2-1a). Grasp cover on each side, lift up slightly and remove.
Figure 2-1a. Generator Cabinet Cover Screw Locations
Revision W HF Series X-ray Generators - Service Manual
2-6 Quantum Medical Imaging
Chapter 2 Assembly & Installation
WARNING! Approximately 200 VDC is present on each battery tray. This voltage is live even when generator cabinet power switch is set to OFF. Use extreme caution when servicing battery tray(s).
2. On stored energy (SE) units, open the two (2) modular fuse holders located on the front of each battery tr ay. Note that there are two pairs of fuses on 32 kW and 40 kW units, one pair for each battery tray.
3. If mounting generator to floor, mark locations of the four mounting holes on floor in accordance with dimensions given in Figure 2-2. Drill four holes using proper diameter drill bit (depending on mounting hardware used).
Figure 2-2. Generator Cabinet Mounting Hole Locations
4. Position the generator chassis in its mounting location and place a level on flat horizontal surface of chassis. If chassis is not level, add shim(s) below generator as required until level.
5. After all cable connections that require access to the rear of the generator are complete, secure cabinet to floor using 3/8" diameter mounting hardware suitable for floor construction type.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-7
Chapter 2 Assembly & Installation
NOTE
STORED ENERGY (SE) models can operate between 105 and 130 VAC and non-SE, single-phase models between 200 and 270 VAC. Refer to Chapter 1, SPEC­IFICATIONS.
Transformer Line Tap Adjustment (Single Phase Non-SE Systems)
WARNING! Extreme care MUST be exercised to ensure the safety of service or other personnel in the area when working with live AC mains.
Proceed as follows:
1. Using a calibrated DVM, measure the AC input (mains) voltage (i.e., line­to-line or line-to-neutral, when applicable).
2. Verify that power to x-ray generator is disconnected.
3. Adjust transformer taps by relocating the leads on the top side of trans­former terminal block TB2 to match the input line voltage as closely as possible. Refer to Figure 2-3 for 208 - 260 VAC single-phase input line tap configurations, and Figure 2-4 for 380 - 480 VAC single-phase input line tap configurations.
Revision W HF Series X-ray Generators - Service Manual
2-8 Quantum Medical Imaging
Figure 2-3. Single-Phase 208 - 260 VAC Single-Phase Input
AC Transformer Line Tap Connections
Chapter 2 Assembly & Installation
Figure 2-4. Single-Phase 380 - 480 VAC Single-Phase Input
AC Transformer Line Tap Connections
4. For single-phase non-SE models, it is necessary to insure that the line resistance meets the maximum allowed as stated in Chapter 1, SPECIFI­CATIONS . If line resistance exceeds the specification, the gener ator must be derated. Refer to Chapter 3, CALIBRATION for derating instructions.
5. Verify Line Monitor Board A8 "line input jumper" is configured correctly based on the installation site’s measured line input voltage as follows:
a. For 208 - 260 VAC input models (QG-25, QG-32 and QG-40), refer to
Line Monitor Board A8 (AY40-022T1) Test Point/Jumper Location Dia­gram in Chapter 6, DIAGRAMS and, if necessary, set "line input" jumpers to match the line input supply voltage (i.e., 208 - 260 VAC, three "240" jumpers are installed). Also, for 208 - 260 V AC line input configurations, fuses A8F1 and A8F3 should be 1/2A.
b. For 380 - 480 VAC input models (QG-25-5, QG-32-5 and QG-40-5),
refer to Line Monitor Board A8 (AY40-022T2) Test Point/Jumper Loca­tion Diagram in Chapter 6, DIAGRAMS and, if necessary, set "line input" jumpers to match the line input supply voltage (i.e., for 342­403 VAC input, install jumper"380"; for 404-459 VAC input, install jumper "430"; for 460-528 VAC input, install jumper "480"). Also, for 380 - 480 VAC line input configurations, fuses A8F1 and A8F3 should be 1/4A.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-9
Chapter 2 Assembly & Installation
Transformer Line Tap Adjustment (Single Phase SE Systems)
WARNING! Extreme care MUST be exercised to ensure the safety of service or other personnel in the area when working with live AC mains.
1. Using a calibrated DVM, measure the AC input (mains) voltage (i.e., line­to-line or line-to-neutral, when applicable).
2. Verify that power to x-ray generator is disconnected.
3. Adjust transformer taps by relocating the leads on the left side of trans­former terminal block TB1 to match the input line voltage as closely as possible. Refer to Figure 2-5 for STORED ENERGY (SE) input line tap configurations.
6. Refer to Line Monitor/Battery Charger Board A8 (AY40-028T) Test Point/ Jumper Location Diagram in Chapter 6, DIAGRAMS a nd verify "line input" jumpers are configured to match the line input supply voltage (115 or 230 VAC). Also, for 115 VAC line input configurations, fuses A8F5 and A8F7 should be 2A; for 230 VAC line input configurations, these fuses should be 1A.
Figure 2-5. STORED ENERGY Input AC Transformer Line Tap Connections
Revision W HF Series X-ray Generators - Service Manual
2-10 Quantum Medical Imaging
Chapter 2 Assembly & Installation
198 VAC
208 VAC
220 VAC
240 VAC
198 VAC
208 VAC
220 VAC
240 VAC
198 VAC
208 VAC
220 VAC
240 VAC
PHASE A PHASE B
LA40-056 REV. C
234523452345
380 VAC
415 VAC
440 VAC
480 VAC
380 VAC
415 VAC
440 VAC
480 VAC
380 VAC
415 VAC
440 VAC
480 VAC
PHASE A
PHASE B
LA40-033 REV. F
56785
6785678
380-480 VAC LINE INPUT CONFIGURATION
208-240 VAC LINE INPUT CONFIGURATION
TA P ADJUSTMENT
LEADS
WARNING! Extreme care MUST be exercised to ensure the safety of service or other personnel in the area when working with live AC mains.
Proceed as follows:
1. Using a calibrated DVM, measure the AC input (mains) voltage (i.e., line­to-line).
2. Verify that power to x-ray generator is disconnected.
3. Adjust transformer taps by relocating the leads on the top side of trans­former terminal block TB2 to match the input line voltage as closely as possible. Refer to Figure 2-6 for input line tap configurations.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-11
Figure 2-6. Three-Phase Input AC Transformer Line Tap Connections
4. For three-phase models, it is necessary to insure that the line resistance meets maximum allowed as stated in Chapter 1, SPECIFICATIONS. If line resistance exceeds the specification, the generator must be derated. Refer to Chapter 3, CALIBRATION for derating instructions.
Chapter 2 Assembly & Installation
5. On Line Monitor Board A8 (part number AY40-053S2 used on 380-480 VAC line input configuration, part number AY40-053S3 used on 208-240 VAC line input configuration), verify 380/430/480 jumper is configured correctly based on incoming line voltage measurement as follows (see Line Monitor Board A8 Chapter 6, DIAGRAMS):
AC INPUT VOLTAGE (MEASURED) JUMPER POSITION
342-403 VAC 380 404-459 VAC 430 460-528 VAC 480 198-252 VAC 240 (3 jumpers)
AC Input (Mains) Connection
WARNING! Verify incoming power supply (mains disconnect) is turned OFF and all mains cables are de-energized before connecting to the generator.
All electrical connections must be made in compliance with local and/or NEC requirements. The generator is factory configured to accept line input speci­fied when the order was placed. However, if it is necessary to change the generator line input configuration, refer to the previous procedure entitled "Transformer Line Tap Adjustment".
Non-STORED ENERGY (SE) version generators require a mains disconnect box (with switch) before the x-ray generator (refer to your local code). This switch should be within reach of the operator. A separate earth ground, #6 AWG insulated wire, must be installed. This wire must be the only wire con­nected to the generator ground terminal.
Revision W HF Series X-ray Generators - Service Manual
2-12 Quantum Medical Imaging
Chapter 2 Assembly & Installation
GENERATOR CHASSIS
LINE INPUT
CABLE CLAMP
(TYPICAL)
TERMINAL
BLOCK TB1
GROUND
LUG
For non-SE generators, connect AC input cable as follows:
1. Mount cable clamp (included with system) to the hole on the lower right or left side of generator cabinet chassis (see Figure 2-7).
Figure 2-7. Line Input Cable Clamp Installation
2. Route the AC input cable through cable clamp to Terminal Block TB1 and tighten cable clamp.
3. On single-phase systems: a. Connect ground wire to ground lug mounted on rear wall of genera-
tor chassis, to the left of terminal block TB1 (see Figure 2-8).
Figure 2-8. Single-Phase Systems - Ground Lug and Terminal
Block TB1 Locations
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-13
Chapter 2 Assembly & Installation
GROUND
LUG
TERMINAL
BLOCK TB1
b. Connect mains line and neutral wires to TB1 term i n al s L1 an d L2 ,
respectively. Do not turn mains power on at this time.
4. On three-phase systems: a. Connect Phase A, B and C wires to TB1 terminals L1 through L3,
respectively. Do not turn mains power on at this time. (See Figure 2-9).
Figure 2-9. Ground Lug and Terminal Block TB1 Locations
b. If available, connect Neutral wire to TB1 terminal L4 (N). (Neutral
connection may be required; check local electrical code.)
5. On STORED ENERGY (SE) type generators, the battery tray fuse(s) are shipped in the non-inserted position to minimize battery discharge and high voltage hazards. With fuse in holder, close the fuse holder(s). The SE system can be plugged into a standard switched 115/230 VAC grounded receptacle. The switch controlling the receptacle should be within reach of the operator control console. For 115 V AC installations, a 25-foot power cord with a hospital-grade plug is factory pre-wired to the generator. For 230 VAC installations, a 25-foot "HAR" power cord is fac­tory pre-wired to the generator.
If power cord is replaced, ensure that impedance between the protective contact in the mains plug and any accessible metal part which is protec­tively earthed does not exceed 0.1 Ohms.
Revision W HF Series X-ray Generators - Service Manual
2-14 Quantum Medical Imaging
Chapter 2 Assembly & Installation
HV TANK (TOP VIEW)
PRESSURE RELIEF SCREW: AT TIME OF INSTALLATION, TURN SCREW COUNTER­CLOCKWISE TWO (2) TURNS AND LEAVE IN THIS POSITION.
High Voltage (H.V.) Tank
The HF Series X-ray Generator is shipped with the H.V. Tank installed at the factory . Locate the Pressure R elief Screw on the top side of the H.V. T ank near the front right corner (see Figure 2-10). To permit venting, loosen screw two (2) turns (counterclockwise) and leave in this position.
Figure 2-10. H.V. Tank (Top View)
The following instructions are provided should it be necessary to install the H.V. Tank. When installing the H.V. Tank, keep it upright at all times. See Figure 2-11 and proceed as follows:
WARNING! After completing the following proce­dure, double-check the H.V. Tank wiring connec­tions to make sure they are correctly connected. Incorrect wiring can cause damage to the genera­tor. Ensure that ground wires are connected.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-15
Chapter 2 Assembly & Installation
CAUTION! The H.V. Tank weighs approximately 92 lbs. (41.7 kg). Lift with proper assistance.
1. Loosen four (4) screws securing the clear plastic high-voltage acrylic pro­tection panel from the generator chassis and remove panel.
2. Position H.V. Tank on Tank Mounting Brackets so mounting holes in tank align with those in mounting brackets. Secure tank to brack ets using four (4) 1/4-20 x 3/4" hex head screws, flat washers and lock washers.
3. Connect H.V. Tank cable terminal PC1 to Power Module A19 IGBT lower right terminal. Tighten screw securely.
4. Connect H.V. Tank cable terminal PC2 to right terminal lug on Capacitor C7. Tighten nut securely.
5. Connect H.V. Tank cable terminal PA1 to left terminal lug on Capacitor C4. Tighten nut securely.
6. Connect H.V. Tank cable terminal PA2 to Power Module A18 lower left IGBT terminal. Tighten screw securely.
7. Locate ground wires for H.V. Tank (yellow/green) and securely fasten to H.V. Tank ground stud (GND) using 8/32 kep nut (provided on stud).
8. Insert plug A17P1 (16-pin plug on Cable Assembly AY51-017) into con­nector A17J1 on the H.V. Tank and tighten plug securely.
9. Re-install the clear plastic high-voltage protection panel on the generator chassis and re-tighten four (4) mounting screws.
Revision W HF Series X-ray Generators - Service Manual
2-16 Quantum Medical Imaging
Chapter 2 Assembly & Installation
Figure 2-11. H.V. Tank Connections
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-17
Chapter 2 Assembly & Installation
BEFORE INSERTING CABLE, LOOSEN SET SCREW
COAT WITH INSULATING COMPOUND BEFORE INSERTING
1
TIGHTEN RING NUT ONTO RECEPT ACLE
2
FULLY TIGHTEN SET SCREW
3
High Voltage Cables
It is important that these instructions be followed for the proper installation of the High Voltage Cable terminals into the tube housing and H.V. Tank receptacles. Use proper handling to ensure cable terminals and receptacles are clean and free of moisture. Connect High Voltage Cables, respecting proper anode and cathode polarity, as follows:
Figure 2-12. High Voltage Cable Connection to H.V. Tank
Revision W HF Series X-ray Generators - Service Manual
2-18 Quantum Medical Imaging
Chapter 2 Assembly & Installation
ROTOR
CABLE
ALL OTHER
CABLES
GENERATOR CABINET (VIEWED FROM REAR)
HIGH VOLTAGE CABLES
CAUTION! When routing the High Voltage Cables into the generator cabinet through the access openings, the anode and cathode cable must each be routed separate from all other system cables, except the Tube Rotor Cable, as shown in Figure 2-13. Doing so will reduce the possibility of elec­tromagnetic noise coupling and subsequent deg­radation of system performance.
Figure 2-13. Routing High Voltage Cables into the Generator Cabinet
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-19
Chapter 2 Assembly & Installation
1. If not already done so, assemble onto the High Voltage Cable end a cable retaining nut, grounding washer and strain relief (see Figure 2-14). Ensure the 10-32 x 1/4" locking screw is turned out so as not to interfere with engagement.
Figure 2-14. High Voltage Cable Assembly (Typical)
2. The High Voltage Cable terminals and tube housing and H.V. Tank recep­tacles must be clean and dry. Be sure compression gaskets are in place on the cable terminal.
3. To insure proper locking and grounding of the cable sheath, do not omit the compression gasket on the cable terminal, as it is necessary for proper contact between cable nut and flare.
4. Insert the cable terminal into the receptacle socket (watch the nipple on the terminal to ensure correct positioning of the contact pins). Coat the
entire surface of the cable terminal insulator with insulating compound (supplied with cables) using clean, dry plastic spoon or wooden stick. Do not use fingers. Fully cover the entire surface
with a heavy coat (approximately 1/16" to 1/8"). Build a pyramid at the tip of the terminal with the height of the compound even with the contact ends.
5. With terminal key aligned with receptacle key way, insert the cable termi­nal into receptacle until all excess compound has oozed out. Make sure the contact pins are engaged in the holes in the socket insulator. Wipe off any excess compound.
6. Tighten the cable nut while holding the cable firmly in place. Re-tighten after calibration as some loosening may take place after the housing has been heated and cooled. With cable nut properly tightened, secure with 10-32 x 1/4" locking screw.
Revision W HF Series X-ray Generators - Service Manual
2-20 Quantum Medical Imaging
Chapter 2 Assembly & Installation
24VAC
24 VAC
CHASSISGND.24VD
C
24V RTN
SPARE
15V RTN
15 VD
C
CHASSIS
GND
.
COL RDY
SLO-BLO
250 VAC
8 AMP
220VAC
110VAC
DOOR INTERLOCK TERMINAL BLOCK TB6
TUBESTAND
TERMINAL
BLOCK TB2
COLLIMATOR
CONNECTIONS
TERMINAL BLOCK TB1
EXPOSURE
HOLD
CONNECTIONS
(AUTO COLL.)
X-RAY ON LIGHT
TERMINAL BLOC K TB5
JP1 INSTALLED ON SYSTEMS WITH MANUAL COLLIMATOR
JP2 INSTALLED ON SYSTEMS WITH NON-BUCKY TABLE RECEPTOR
JP3 INSTALLED
ON SYSTEMS WITH
NON-BUCKY WALL
RECEPTOR
TABLE BUCKY POWER
TERMINAL BLOCK TB3
WALL STAND BUCKY POWER
TERMINAL BLOCK TB4
MIDWEST BUCKY SWITCHES
(SEE T ABLE /WALL BUCKY
CONNECTION INSTRUCTIONS)
TUBE ROTOR BOOST VOLTAGE JUMPER (REV. R AND HIGHER BOARDS ONLY)

INTERFACE BOARD SETTINGS & CONNECTIONS

The generator’s Inte rface Board A9 pr ovides the means for ele ctrically connecting external system components, such as collimators, tube stands, buckies, etc., and verifying the correct Rotor Boost Voltage setting (non-high speed starter systems only). Figure 2-15 shows connection locations on a typical non-digital (film­based) system Interface Board; Figure 2-16 shows connection locations on a typ­ical digital imaging system Interface Board.
Figure 2-15. Non-Digital System Interface Board A9 - Connection Locations
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-21
Chapter 2 Assembly & Installation
1
24VAC
24VAC
CH GND
24V
24V RTN
SPARE
COLL RDY
15V RTN
15V
GND
D2
D4
D7
D8
D12
D13
D14
K3
R11
R12
R16
R17
R19
TB5
D20
D19
F18
F19
F13
DOOR INTERLOCK TERMINAL BLOCK TB6
TUBESTAND
TERMINAL
BLOCK TB2
COLLIMATOR
CONNECTIONS
TERMINAL BLOCK TB1
EXPOSURE
HOLD
CONNECTIONS
(AUTO COLL.)
X-RAY ON LIG HT
TERMINAL BLOCK TB5
TABLE BUCKY POWER
TERMINAL BLOCK TB3
TUBE ROTOR BOOST VOLTAGE JUMPER
(REV. J AND HIGHER
BOARDS ONLY)
WALL STAND BUCKY POWER
TERMINAL BLOCK TB7
Figure 2-16. Digital System Interface Board A9 - Connection Locations
Door Interlock Switch Connection
To wire a door interlock circuit, remove jumper and connect door interlock wires (supplied by installer) to terminal block TB6 pins 1 and 2 on Interface Board A9 [(P/N AY40-023T/or AY40-061T on internal high-speed starter sys­tems) or (AY40-034T/or A Y40-062T on internal high-speed starter systems)]. See Figures 2-15 or 2-16 for terminal board TB6 location. When the door is closed, the circuit will be continuous (contacts closed).
"In Use" Light Switch Connection
To wire an "X-ray On" or "In-Use light", connect wires from light circuit (sup­plied by installer) to terminal board TB5 pins 1 and 2 on Interface Board A9 [(P/N AY40-023T/or AY40-061T on internal high-speed starter systems) or (AY40-034T/or AY40-062T on internal high-speed starter systems)] (On AY40-062T boards, pin 3 is not used). See Figures 2-15 or 2-16 for terminal block TB5 location. When the X-ray generator is in either PREP or EXPOSE states, the circuit will be continuous (contacts closed).
Revision W HF Series X-ray Generators - Service Manual
2-22 Quantum Medical Imaging
Chapter 2 Assembly & Installation
P2-1 (+)
P2-2 (-)
OUTPUT
VOLTAGE
ADJ. POT.
Collimator Connections
Terminal block TB1 (see Figures 2-15 or 2-16 for terminal block TB1 location) provides 24 VAC (TB1-1 and TB1-2) and chassis ground (TB1-3) outputs for connection to a collimator. On systems equipped with a manual collimator, verify jumper A9JP1 is installed; remove A9JP1 if system is equipped with an automatic collimator.
Also available on terminal block TB2 is the COL RDY terminal (at TB2-4) for connection to automatic collimator with "exposure hold" interlock (use CHAS­SIS GND terminal TB2-7 for return). Refer to collimator manual for specific connection details.
Tube Stand/Peripheral Equipment Connections
Interface Board A9 Terminal block TB2 (see Figures 2-15 or 2-16 for terminal board TB2 location) provides +28 VDC (TB2-1) and +28 VDC RTN (TB2-2) terminals for connection of a tube stand power input cable, when applicable. Maximum load current is 1.5A at these terminals.
Internal +28 VDC Power Supply Option (QG-28VPS)
If generator is equipped with +28 VDC Internal Power Supply Option (QG-28VPS), see Figure 2-16a and connect the tubestand/wallstand’s power supply input leads as follows:
1. Positive lead to terminal P2-1 (+) on power supply
2. Negative lead to terminal P2-2 (-) on power supply
Figure 2-16a. Internal +28 VDC Power Supply Option (QG-28VPS) Location
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-23
Chapter 2 Assembly & Installation
The output of the +28 VDC Power Supply is adjusted depending on the type of Quantum product it will supply power to, as indicated below:
Table 2-1a: Internal +28 VDC Power Supply Option (QG-28VPS) Adjustment
Quantum Product Type Adjust Power Supply Output To:
Model RS-580 Overhead Tube Crane +30 VDC Model QV-800 Universal System +28 VDC Model QW-420-T, QW-420-T-D Vertical
Tilting Wall Stand
When using the +28 VDC Power Supply Option to provide power for a Model RS-580 Overhead Tube Crane, the +30 VDC output is intended for operating the tube crane’s transverse and longitudinal electromagnetic locks (brakes). Refer to the RS-580 Service Manual for connection instructions.
CAUTION! The Internal +28 VDC Power Supply Option (QG-28VPS) is only available with 3-phase generators, and is intended to supply power to a single, external device.
Tube Rotor Boost Voltage Verification (Systems without High (Dual) Speed Starter Option)
Ensure Interface Board A9 [P/N A Y40-023T (non-digital systems, revision R or later boards only)/AY40-023T1 (non-digital stored energy systems, r evision K or later boards only)/AY40-034T (digital systems, revision J or later boards only)] Tube Rotor Boost V oltage Jumper is configured correctly depending on the type of tube the generator is being interfaced with according to table below:
Table 2-1: Tube Rotor Boost Vol tage Jumper Setting
+28 VDC
TUBE TYPE
Toshiba E7252X, E7813X, E7242X, E7239X
All others (220 VAC stator) Pins 2-3 220 VAC Stator
JUMPER
POSITION
Pins 1-2 100 VAC Stator
ROTOR BOOST VOLTAGE
SETTING
Revision W HF Series X-ray Generators - Service Manual
2-24 Quantum Medical Imaging
Chapter 2 Assembly & Installation

RADIOGRAPHIC SYSTEM EQUIPMENT CONNECTIONS

The following paragraphs provide instructions for cable connections between the HF Series x-ray generator and the x-ray tube, table, and wall bucky, as necessary. Refer to Generator Interconnection Diagrams in Chapter 6, DIAGRAMS for addi­tional guidance.
Ensure the generator is configured with the correct value rotor phase capacitor C1 (on the Generator door) for tube installed in the system (refer to Table 2-2).
Table 2-2: Standard Single-Speed Rotor Phase Capacitor Configuration
Tube Type Rotor Phase Capacitors
E7239X, E7242X 25 uF E7254X, E7255X, ALL VARIAN,
ALL DUNLEE (including MX100 Replace­ment for GE Radiographic System)
E7813X, E7252X 45 uF
Tube Rotor Cable (Systems without High (Dual) Speed Starter Option)
Route the X-ray Tube Rotor Cable AY51-014 through the middle cable access opening located in the top rear wall of the generator chassis. Connect one end of the cable to the appropriate terminals on Terminal Block TB1 on Rotor Drive Board A10 (P/N AY40-013T) as shown in Figure 2-17. Connect the opposite end to the X-ray tube’s terminal strip (refer to appropriate intercon­nection diagram in Chapter 6, DIAGRAMS for tube connections).
30 uF
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-25
Chapter 2 Assembly & Installation
TERMINAL BLOCK TB1
POTENTIOMETER R20 TUBE CURRENT ADJ.
POTENTIOMETER R24 TUBE CURRENT ADJ.
1MAIN 2 PHASE 3 COMMON 4 NOT USED 5 NOT USED 6 GROUND
7TEMP
TERMINAL
SIGNAL
TEST POINT 7
ROTOR RUN
TEST POINT 10
+15 VDC
TEST POINT 4
PHASE
TEST POINT 1 COMMON
TEST POINT 2 MAIN
TEST POINT 3
AC RETURN
TEST POINT 13
60 VAC
TEST POINT 12
200 VAC
U3
TP3
R26
R15
R12
MOV4
MOV3
J3
R37
PD2
TB1
K2
D18
C14
R35
TP1
TP2
+
C16
Q10
U6
PD1
Q9
R46
J5
TP4
R20
R25
R24
D4
D5
+
C7 D7 D12
R36
R27
R19
Q6
J4
D8
TP15
U5 7 8
D15
C15
R41
Q8
TP14
R30
R29
R11
Q7
D17
R22
R45
TP7
TP10
D10
D1
TP8
U4
D3
Q5
R6 R9
D14
R3
Q3
C4
D6
D9 D11
C1
TP6
10W
R2
10W
R14
PD3
MOV2
C8
TP13
J1
TP12
Q1
MOV1
C6
Q4
C2
PD4
TP11
R4
C3
U1
Q2
U2
R10
D2
R8
R7
R5
TP9
TP5
R33
R32
R17 R16
7
1
1
3
B
CE
B
B
EEC
1
2
ECEBCBC
9
10
1
1
4
T2
T1 G
T1
T2
G
B
CE
9
Revision W HF Series X-ray Generators - Service Manual
2-26 Quantum Medical Imaging
Figure 2-17. Rotor Board AY40-013T Tube Rotor Cable Connections
Chapter 2 Assembly & Installation
Tube Rotor Cable (Systems with AID External High Speed Starter QG-HSS Option Only)
Refer to Figure 2-18, the AID High Speed Starter installation manual and/or the appropriate interconnection diagram in Chapter 6, DIAGRAMS for high speed starter connections.
tems equipped with the externally-connected AID High Speed Starter option.
1. Connect one end of X-ray Tube Rotor Cable AY51-032T to the appropriate connector/terminals on AID High Speed Starter LC102 terminal block TS2 (see Figure 2-18).
2. Connect the opposite end to the X-ray tube’s terminal strip.
3. Route the X-ray Tube Rotor Cable AY51-033 (Toshiba tubes) through the middle cable access opening located in the top rear wall of the generator chassis.
4. Connect one end of the High Speed Starter Cable AY51-033 to the appro­priate connector/terminals on AID High Speed Starter terminal strips TB1 and TB2 and connect to Cable Assembly AY51-032T (temperature sense connector).
Note: This procedure applies only to sys-
5. Connect the opposite end of the High Speed Starter Cable AY51-033 to connector J8 on Logic Board A1, AY40-006S.
6. Connect 220 VAC power input to High Speed Starter.
7 . The AID High Speed Starter was factory pre-configured for the x-ra y tube
supplied with the generator. Verify boost and brake settings are correct according tube manufacturer’s data sheets (ref er to Table 2-3). Verify the high speed starter is configured with the correct capacitors for the tube used with the system: Toshiba E7252X and E7813X should be configured for 45 uF. All other tube types require 30 uF.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-27
Chapter 2 Assembly & Installation
Table 2-3: AID High Speed Starter DIP Switch Settings
TUBE TYPE
TOSHIBA E7813X, E7252X
TOSHIBA E7254X, E7255X
VARIAN (ALL 3 INCH DIA.)
VARIAN (ALL 4 INCH DIA.)
START
VOLTAGE
(180 HZ)
220 VAC (#6 tap*)
490 VAC (use 370 VAC tap #10*)
320 VAC (use 370 VAC tap #10*)
450 VAC (use 370 VAC tap #10*)
RUN VOLTAGE (180 HZ)
60 VAC (75 VAC #3 tap*)
165 VAC (75 VAC #3 tap*)
100 VAC (75 VAC #3 tap*)
100 VAC (75 VAC #3 tap*)
HSS SWITCH SW1 (BOOST
TIME)
1.2 seconds: SW1-1 = OFF SW1-2 = ON SW1-3 = ON SW1-4 = OFF SW1-5 = OFF
2.0 seconds: SW1-1 = OFF SW1-2 = ON SW1-3 = OFF SW1-4 = ON SW1-5 = OFF
1.0 seconds SW1-1 = ON SW1-2 = OFF SW1-3 = ON SW1-4 = OFF SW1-5 = OFF
2.2 seconds SW1-1 = ON SW1-2 = ON SW1-3 = OFF SW1-4 = ON SW1-5 = OFF
HSS SWITCH
SW2 (BRAKE
TIME)
2.0 seconds: SW2-1 = OFF SW2-2 = ON SW2-3 = OFF SW2-4 = ON SW2-5 = OFF
2.0 seconds: SW2-1 = OFF SW2-2 = ON SW2-3 = OFF SW2-4 = ON SW2-5 = OFF
2.0 seconds: SW2-1 = OFF SW2-2 = ON SW2-3 = OFF SW2-4 = ON SW2-5 = OFF
2.0 seconds: SW2-1 = OFF SW2-2 = ON SW2-3 = OFF SW2-4 = ON SW2-5 = OFF
Revision W HF Series X-ray Generators - Service Manual
2-28 Quantum Medical Imaging
Chapter 2 Assembly & Installation
Figure 2-18. AID High Speed Starter Connections
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-29
Chapter 2 Assembly & Installation
Tube Rotor Cable (Systems with Internal High Speed Starter Q­HSS Option Only)
The Q-HSS High Speed Starter is factory pre-configured for the x-ray tube supplied with the generator and, with the exception of the Rotor Cable, does not require any field connections or adjustments. Tables 2-3a and 2-3b are provided in the event tube replacement is necessary or if the HSS Power Driver Board AY40-036S needs replacement. The table indicates the proper connections of the various transformer voltage inputs (380, 220, 120, 80, 60 and 0 VAC) to the HSS Driver Board at Termin al Block TB1 for each of the compatible tubes used with t he generator.
Also, ensure the generator is configured with the correct value rotor capaci­tors for the installed tube as indicated in Table 2-3a.
Table 2-3a: Quantum High-Speed Starter Rotor Phase Capacitor Configuration
Tube Type Rotor Phase Capacitors
E7813X, E7252X, E7869X Install one (1) 40 uF cap (low speed), two (2) 12 uF in series
(high speed)
E7254X, all Varian, all Dunlee (except MX100 GE replacement, see below)
Dunlee MX100 Replace­ment for GE Radiographic System
If not already connected, connect transformer input wires to HSS Driver Board AY40-036S terminal block TB1 plug for corr ect tube type in ac cordance with table below:
Table 2-3b: HSS Driver Board AY40-036S Boost and Run Voltage Requirements: Transformer
TUBE MODEL
TOSHIBA E7255, E7254
Install one (1) 25 uF cap (low speed), two (2) 12 uF in series (high speed)
Install one (1) 25 uF cap (low speed), two (2) 15 uF in series (high speed)
Input Wires to Terminal Block Plug TB1
TB1-1 TB1-3 TB1-5 TB1-6 TB1-7
HIGH SPEED
BOOST
VOLTAGE WIRE #5
(380 VAC)
LOW SPEED
BOOST
VOLTAGE WIRE #4
(220 VAC)
HIGH SPEED
RUN
VOLTAGE WIRE #3
(120 VAC)
LOW SPEED
RUN
VOLTAGE
WIRE #2A
(80 VAC)
COMMON
WIRE #1
(0 VAC)
TOSHIBA E7252, E7813
WIRE #4
(220 VAC)
WIRE #3
(120 VAC)
WIRE #2A
(80 VAC)
WIRE #2
(60 VAC)
WIRE #1
(0 VAC)
Revision W HF Series X-ray Generators - Service Manual
2-30 Quantum Medical Imaging
Chapter 2 Assembly & Installation
Table 2-3b: HSS Driver Board AY40-036S Boost and Run Voltage Requirements: Transformer
Input Wires to Terminal Block Plug TB1
TB1-1 TB1-3 TB1-5 TB1-6 TB1-7
TUBE MODEL
VARIAN RAD 92, RAD 60, RAD 56, RAD 44, A292, A272; DUNLEE DA 1092, DU309, PX 1302, PX 1312, PX 1429, PX 1436
VARIAN RAD 21, RAD 14, RAD 13, A192; TOSHIBA E7869X DUNLEE MX100 Replacement for GE
HIGH SPEED
BOOST
VOLTAGE WIRE #5
(380 VAC)
WIRE #5
(380 VAC)
LOW SPEED
BOOST
VOLTAGE
WIRE #4
(220 VAC)
WIRE #4
(220 VAC)
HIGH SPEED
RUN
VOLTAGE WIRE #3
(120 VAC)
WIRE #2A
(80 VAC)
LOW SPEED
RUN
VOLTAGE WIRE #2
(60 VAC)
WIRE #2 (60 VAC)
COMMON
WIRE #1
(0 VAC)
WIRE #1
(0 VAC)
NOTE 1: Stow unused WIRE #2 (60 VAC) or WIRE #2A (80 VAC) in unused terminal block
TB1-2 location.
NOTE 2: Terminal block location TB1-4 is not used.
Non-Overhead Tube Crane Rotor Cable Wiring Instructions
Refer to Figure 2-18a and the steps below to connect the Rotor Cable to the HSS Driver Board x-ray tube on a system Crane.
1. Route the X-ray Tube Rotor Cable from the tube through the middle cable access opening located in the top rear wall of the generator chassis.
2. Connect the MAIN (BLK) conductor to terminal 1 on HSS Driver Board A20A2 Terminal Block TB2.
3. Connect the PHASE (RED) conductor to terminal 3 on HSS Driver Board A20A2 Terminal Block TB2.
4. C onnect the COMMON (WHT) conductor to terminal 6 on HSS Driver Board A20A2 Terminal Block TB2.
5. Connect the TEMP SENSE (ORN) conductor to terminal 1 on HSS Driver Board A20A2 Terminal Block TB4.
6. Connect the TEMP SENSE (GRN) conductor to terminal 2 on HSS Driver Board A20A2 Terminal Block TB4.
not using
an Overhead Tube
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-31
Chapter 2 Assembly & Installation
L4A
L5A
L5B
L4B
1
B
B
B
E
E
E
DANGER
HIGH VOLTAGE
HIGH VOLTAGE
DANGER
C17
C21
C27
C28
C29
C30
C33
C37
C42
C47
C49
C51
J1
J3
2
0
R5
R11
R13
R15
R16
R17
R18
R20
R26
R42
R43
R47
R50
R54
R55
R58
R60
TB1
TP1
TP2
TP6
TP8
TP10
U9
R2
R67
R68
R72
R71
R73
R28
D6
D11
D12
D14
D4
D31
D22
R63
R64
R66
C7
C6
C2
L4
L5
B
L
K
RED
W
H
T
O
R
N
G
R
N
R
O
T
O
R
C
A
B
L
E
A
2
0
A
2
T
B
2
A
2
0
A
2
T
B
4
H
S
S
D
R
I
V
E
R
B
O
A
R
D
A
2
0
A
2
(
A
Y
4
0
-
0
3
6
S
)
R
O
T
O
R
C
A
B
L
E
C
O
N
N
E
C
T
I
O
N
S
:
M
A
I
N
(
B
L
K
)
T
O
A
2
0
A
2
T
B
2
-
1
P
H
A
S
E
(
R
E
D
)
T
O
A
2
0
A
2
T
B
2
-
3
C
O
M
M
O
N
(
W
H
T
)
T
O
A
2
0
A
2
T
B
2
-
6
T
E
M
P
S
E
N
S
E
(
O
R
N
)
T
O
A
2
0
A
2
TB4
-
1
T
E
M
P
S
E
N
S
E
(
G
R
N
)
T
O
A
2
0
A
2
TB4
-
2
7. Connect the opposite end to the X-ray tube’s terminal strip.
Figure 2-18a. Non-Overhead Tube Crane System Rotor Cable Connections
Revision W HF Series X-ray Generators - Service Manual
2-32 Quantum Medical Imaging
Overhead Rotor Cable Wiring Instructions
Refer to Figure 2-18b and the steps below to connect the Rotor Cable to the x-ray tube on a system using an Overhead Tube Cranes:
1. Route the X-ray Tube Rotor Cable from the tube through the middle cable
2. Connect the MAIN (BLK 0) conductor to terminal 1 on HSS Driver Board
3. Connect the PHASE (BLK 1) conductor to terminal 3 on HSS Driver Board
4. Connect the COMMON (BLK 2) conductor to terminal 6 on HSS Driver
access opening located in the top rear wall of the generator chassis.
A20A2 Te rminal Block TB2.
A20A2 Te rminal Block TB2.
Board A20A2 Terminal Block TB2.
Chapter 2 Assembly & Installation
L4A
L5A
L5B
L4B
1
B
B
B
E
E
E
DANGER
HIGH VOLTAGE
HIGH VOLTAGE
DANGER
C17
C21
C27
C28
C29
C30
C33
C37
C42
C47
C49
C51
J1
J3
2
0
R5
R11
R13
R15
R16
R17
R18
R20
R26
R42
R43
R47
R50
R54
R55
R58
R60
TB1
TP1
TP2
TP6
TP8
TP10
U9
R2
R67
R68
R72
R71
R73
R28
D6
D11
D12
D14
D4
D31
D22
R63
R64
R66
C7
C6
C2
L4
L5
B
L
K
0
BLK 1
B
L
K
2
B
L
K
W
H
T
R
O
T
O
R
C
A
B
L
E
A
2
0
A
2
T
B
2
A
2
0
A
2
T
B
4
H
S
S
D
R
I
V
E
R
B
O
A
R
D
A
2
0
A
2
(
A
Y
4
0
-
0
3
6
S
)
R
O
T
O
R
/
T
E
M
P
CABL
E CON
NECTI
ONS
:
M
A
I
N
(
B
L
K
0
)
T
O A20
A
2TB2-
1
P
H
A
S
E
(
B
L
K
1) TO A20A2
TB2
-
3
C
O
M
M
O
N
(
B
L
K
2) T
O
A2
0
A2TB2
-
6
T
E
M
P
S
E
N
S
E
(
BLK) TO A
20A2T
B4-
1
T
E
M
P
S
E
N
S
E
(
WHT)
TO A2
0A2TB4-
2
T
U
B
E
T
E
MP
CAB
L
E
L
X
1
-
D
160
X
6
.
1
s
w
;3w
s
5. Connect the TEMP SENSE (BLK) conductor to terminal 1 on HSS Driver Board A20A2 Terminal Block TB4.
6. Connect the TEMP SENSE (WHT) conductor to terminal 2 on HSS Driver Board A20A2 Terminal Block TB4.
7. Connect the opposite end to the X-ray tube’s terminal strip.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-33
Figure 2-18b. Overhead Tube Crane System Rotor Cable Connections
Chapter 2 Assembly & Installation
Table/Wall Bucky Connections
Wall stand and/or table buckies are interconnected to the HF Series x-ray generator via Interface Board A9, P/N AY40-023T terminal blocks A9TB4 (Wall Stand) and A9TB3 (Table) or Digital Interface Board AY40-034T termi­nal blocks A9TB7 (Wall Stand) and A9TB3 (Table) using 5-conductor, 18AWG cable (P/N EL75-025), which is provided with the table and/or wall stand.
IMPORTANT! When interfacing to a 14" x 17" Midwest table or wall bucky, switches A9SW1 or A9SW2, respectively, must be set to the "M" posi­tion. For 14" x 36" Midwest, Progeny, or LF type buckies, switches A9SW1 or A9SW2 must not be set to the "M" position.
Refer to bucky manufacturer’s installation manual (provided with system) for specific bucky wiring information and to Figures 2-15, 2-16, and in Chapter 6, DIAGRAMS, Figures 6-6 through 6-9. Bucky cable entrance/exit is provided through the access opening in the top rear panel of the generator chassis. DO NOT route cables through an opening containing a High Voltage Cable (see Figure 2-13). When connecting a table bucky, remove jumper JP2 on Interface Board A9. When connecting a wall bucky, remove jumper JP3 on Interface Board A9.
Table/Wall Ion Chamber Connections (AEC Option Only)
Systems equipped with standard type AEC (AEC Board AY40-031S): If using a wall ion chamber, connect wall ion chamber cable to AEC Board
A11, P/N AY40-031S connector J2. If using a table ion chamber, connect table ion chamber cable to AEC Board A11 connector J4. Table 2-4 shows AID type ion chamber connector pin assignments. Route the ion chamber cable(s) as necessary to exit through the access opening in top rear side of generator chassis, and secure as required. DO NOT route through an open­ing containing a High Voltage Cable (see Figure 2-13).
Table 2-4: AID ICX-153 or ICX-1153 Ion Chamber 9-Pin Sub-D Connector Pin-Outs
9-PIN SUB-D
CONNECTOR PIN
1NONE 2 FIELD 2 SELECT 3 FIELD 1 SELECT 4RESET 5OUTPUT 6 FIELD 3 SELECT 7 NEGATIVE SUPPLY 8 POSITIVE SUPPLY 9GROUND
FUNCTION
Revision W HF Series X-ray Generators - Service Manual
2-34 Quantum Medical Imaging
Chapter 2 Assembly & Installation
(Note: Perform ion chamber balance procedure in accordance with manufac­turer’s instructions provided with chamber.)
Typically, when the HF Series x-ray generator is shipped with an AID ICX ion chamber, the chamber jumpers are pre-set in the following configuration:
Table 2-5: AID ICX-153 or ICX-1153 Ion Chamber Jumper Configuration
JUMPER POSITION FUNCTION
61135A W1 1-2 & 3-4 LOW ACTIVE RESET 61135A W2, W3, W4 1-2 & 3-4 LOW ACTIVE FIELD SELECT 61135A W5, W6 1-2 & 3-4 A-B-C = 2-1-3 61135A W7 2-3 POSITIVE OUTPUT SIGNAL
CAUTION! The ion chamber must be electrically isolated from the bucky to ensure proper AEC sys­tem performance.
For systems equipped with universal type AEC (AEC Board AY40-027S): Up to four ion chambers can be used. Connect ion chambers as follows:
Wall#1 ion chamber to A11J8
Wall#2 ion chamber to A11J12
Table ion chamber to A11J10
Auxiliary ion chamber to A11J13
Note that to connect a Siemens ion chamber AEC cable to universal AEC Board A11 requires an interface cable assembly AY51-061 for each chamber. This cable assembly adapts the ion chamber connector to 9-pin D-Sub type used on the AEC Board, and provides a separate 300 VDC wire f or connection to A11TB1.
Route the ion chamber cable(s) as necessary to exit through the access open­ing in top rear side of generator chassis, and secure as required. DO NOT route through an opening containing a High Voltage Cable (see Figure 2-13).
Interfacing to Canon Digital Radiographic Receptor (When Applicable)
The digital radiographic receptor is interconnected to the HF Series x-ray gen­erator via Interface Board A9, P/N AY40-034T terminal blocks A9TB7 (Wall Stand) and A9TB3 (Table), respectively, using 4-conductor, 18AWG cable. Refer to the digital receptor manufacturer’s installation manual for more detailed wiring instructions.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-35
Chapter 2 Assembly & Installation
Digital receptor cable entrance/exit is provided through the access opening in the top rear panel of the generator chassis. DO NOT route signal cables through an opening containing a High Volt age Cable (see Figure 2-13). When connecting to a table digital receptor, remove jumper JP2 on Interface Board A9. When connecting to wall digital receptor, remove jumper JP3 on Inter­face Bo ard A9.
Interfacing to CMT Digital Radiographic Receptor (When Applicable)
Refer to CMT digital receptor installation manual for detailed wiring instruc­tions.
Interfacing to Siemens Ion Chambers (When Applicable)
Note: When interfacing the generator to a Siemens table/wall ion chamber and a Quantum table/wall ion chamber (AID type), only the Siemens ion chambers require the use of cable assembly AY51-061. The Quantum table/wall ion chamber can be connected directly to AEC Board A11 (AY40-027S) connector J10 (TABLE) or J8 (WALL).
Interfacing of the HF Series x-ray generator to Siemens table and/or wall ion chambers, such as in Siemens Tomography System, is facilitated using the "universal" type AEC Board A11 (AY40-027S) located on the front side of the generator cabinet PCB door. The VERTIX wall stand and MULTIX table ion chambers are connected as shown in Figure 2-19. Note that cable assemblies AY51-061 shown in the illustration should already be connected to both the MULTIX table AEC cable EK14 and the VERTIX wall stand AEC cable. Any additional ion chambers used with the system are connected as described in "Table/Wall Ion Chamber Connections (AEC Option Only)" procedure included in this chapter.
Figure 2-19. AEC Ion Chamber Connections (Siemens Tomography System)
Revision W HF Series X-ray Generators - Service Manual
2-36 Quantum Medical Imaging
Chapter 2 Assembly & Installation
SECURE ONE ION
CHAMBER CABLE
CONNECTOR HERE
SECURE OTHER
ION CHAMBER
CABLE CONNECTOR
HERE
Secure the ion chamber cables to the PCB Door using cable ties as shown in Figure 2-20.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-37
Figure 2-20. Siemens Ion Chamber Cable Management
Interfacing to Siemens Tomography System (When Applicable)
The Siemens tomo interface cable (supplied with the Siemens system) has attached to it at one end three plugs identified as "KK1B G1,G2 X10", "KK1B G1,G2 X11" and "KK1B G1,G2 X12". Plugs "X10" (TABLE), "X11" (W ALL) and "X12" (TOMO SWEEP TIME) will need to be cut off and the wire ends stripped to allow connection to the universal AEC Board A11 (AY40-027S) pluggable
Chapter 2 Assembly & Installation
terminal blocks A11TB10, A11TB12 and A11TB11, respectively. Refer to Fig­ure 2-21 for Siemens tomo interface cable wire connections. Secure the tomo interface cable securely using cable ties.
Figure 2-21. Siemens Tomo/Ion Chamber Interface Cable Interconnection Diagram
Check that jumper JP1 is installed on Universal AEC Board A11 (AY40-027S). If not, install a jumper.
Revision W HF Series X-ray Generators - Service Manual
2-38 Quantum Medical Imaging
Chapter 2 Assembly & Installation
Interfacing to Pausch Tomography System (When Applicable)
To interface the HF Series generator to Pausch tomo system, two lengths of 10-conductor shielded cable is required (supplied with the HF Series genera­tor). No wire terminations are required for interconnecting the cables. Ref er to Figure 2-22 for Pausch tomo/generator interface wiring. After interfacing, secure the tomo interface cable securely using cable ties as required.
Check that jumper JP1 is not installed on Universal AEC Board A11 (AY40-027S). If installed, remove this jumper.
IMPORTANT! Be sure to re-program the tomo sweep times on the Pausch tomography system to correspond with available time settings selectable at the HF Series generator Operator Control Panel.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-39
Chapter 2 Assembly & Installation
Figure 2-22. Pausch Tomography System Interconnection Diagram
Revision W HF Series X-ray Generators - Service Manual
2-40 Quantum Medical Imaging
Chapter 2 Assembly & Installation

OPERATOR CONTROL PANEL SETUP

The HF Series X-ray Generator is available with several different types of operator control panel (OCP) configurations. These include:
ODYSSEY Operator Control Panel
Q-VISION Touch-Screen Operator Control Panel
Integrated Workstation Control Panel (e.g., Canon DR Workstation, Agfa NX Workstation, etc.)
The paragraphs that follow describe the installation and setup procedures for each type control panel.
ODYSSEY Operator Control Panel
The ODYSSEY Operator Control Panel (OCP) is provided with a 50-ft. inter­connect cable. Extended length OCP cables are available; part number QG­CBL65 is 65-ft. length, and part number QG-CBL85 is 85-ft. length. The OCP can be placed directly on a shelf, table top, or on top of the Generator Cabi­net, resting on its rubber-tipped feet. It may also be mounted on a pedestal (requires QG-PDL option) or on a wall-mounted bracket (requires QG-WM option). The following are installation instructions for each option.
OCP Pedestal Mounting (QG-PDL Option)
1. On systems equipped with Pedestal Mount option QG-PDL, mount the OCP on the pedestal as follows: Turn the OCP over and rest it on a surface covered with soft material.
2. Remove four (4) OCP front panel screws.
3. Turn over the OCP. Lift up OCP front panel to access OCP mounting holes (if necessary, insert a small flat blade screwdriver under the edge of the panel, in the upper left side, and lift up panel).
4. P osition the OCP on the pedestal aligning the pedestal mounting holes in the bottom of OCP with the mount ing holes in the top flange of the ped­estal.
5. Secure the OCP to the pedestal using four (4) 1/4-20 x 3/4" pan head phillips screws, flat washers and nuts.
6. Re-install four (4) OCP front panel screws.
CAUTION! The OCP serial cable shield (yellow/ green wire) provides a protective earth ground from the OCP to the generator cabinet. Verify both ends of the OCP serial cable are properly secured to grounded mounting screws as described in the following step.
7. Connect OCP serial cable (P/N AY51-018) between connector A16J1 on the OCP and connector J5 on CPU Board A1 (P/N AY40-006T).
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-41
Chapter 2 Assembly & Installation
8. Connect the yellow/green ground wires at each end of OCP serial cable to the nearest mounting screw on OCP Control Board A16A1 (AY40-004S) and Logic Board A1 (AY40-006S).
9. Ensure the connector screws and PCB mounting screws are fully tight­ened.
CAUTION! The OCP serial cable shield is electri­cally connected to the connector shells at either end of the cable. It is recommended the OCP serial cable connectors not be removed so as not to compromise integrity of cable shield connec­tions. If necessary to remove connector, verify that after the connectors are re-installed, there is less than 0.1 ohm resistance between connector shell at one end of cable and the connector shell at opposite end of the cable.
10. If mounting OCP to floor, either transfer drill four mounting holes in floor using the OCP base plate as a guide or mark the floor in accordance with dimensions given in Figure 2-23, and then drill holes where marked. Drill holes using a #W drill bit (0.390" diameter). After OCP cable connection is complete, secure OCP to floor using 3/8" ha rdware suitable for floor type.
Figure 2-23. OCP Mounting Hole Locations
Revision W HF Series X-ray Generators - Service Manual
2-42 Quantum Medical Imaging
Chapter 2 Assembly & Installation
MOUNTING CONFIGURATION A
(SIDE VIEW)
MOUNTING CONFIGURATION B
(SIDE VIEW)
1.00
11.01
10.25
(REAR VIEW)
6.50
O.203 THRU 6X
6.50
1.00
.44 2X
(TOP VIEW)
15.00
(DIMENSIONS SHOWN
IN INCHES)
OCP Wall Mounting (QG-WM Option)
Figure 2-24. QG-WM Wall Mount Option Installation Configurations
On systems equipped with Wall Mounting option QG-WM, mount the OCP to the wall as follows:
1. Secure the Wall Mount to the wall using hardware suitable for wall con­struction type (each screw capable of supporting 15 lbs. pull load).
2. Turn the OCP over and rest it on a surface covered with soft material.
3. Remove four (4) OCP front panel screws.
4. Turn over the OCP. Lift up OCP front panel (if necessary, insert a small flat blade screwdriver under the edge of the panel, in the upper le ft side , and lift up panel) to access OCP mounting holes.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-43
Chapter 2 Assembly & Installation
5. Position the OCP on the Wall Mount bracket aligning the mounting holes in the bottom of OCP with the mounting holes in the wall bracket.
6. Secure the OCP to the Wall Mount Bracket using four (4) 1/4-20 x 3/4" pan head phillips screws, flat washers and nuts.
7. Re-install four (4) OCP front panel screws.
8. Connect one end of the OCP serial cable (P/N AY51-018) to connect or J3 on the OCP. Route the OCP serial cable through the access opening in the Wall Mount Bracket and connect to connector J5 on CPU Board A1 (P/N AY40-006T). Ensure the connector screws are fully tightened. DO NOT route serial cable through an opening containing a High Voltage Cable (see Figure 2-13).
Q-VISION Operator Control Panel
The Q-VISION Operator Control Panel (OCP) is an all-in-one PC, with the computer integrated with the touch-screen monitor in a single case. The Q­VISION OCP is provided with a Power/Expose Control Box (AY20-27 1), which contains a two-stage, hand-held PREP/EXPOSE switch and a built-in genera­tor ON/STANDBY rocker switch for remotely powering on/off the Generator. Cables for interconnection of the Q-VISION OCP and Power/Expose Control Box to the generator are available in either 50-ft. or a 100-ft. interconnect lengths, as specified at the time of order. In order to interface Q-VISION with the generator, the generator must be equipped with the "Q-Connect" option, which provides an OCP Interface Board (AY40-073S) that enables Ethernet connectivity to the generator’s Logic Board AY40-006S.
The Q-VISION OCP monitor is intended for use as a "touch-screen" input device; however, a keyboard and mouse are provided for use by field service technicians if and when it becomes necessary to perform operations external to the Q-VISION program environment (such as when performing operating system functions).
The Q-VISION OCP can be placed directly on a flat surface such as a table top or desk, or on can be wall-mounted using Wall Mount option QGV-WM. Refer to the Q-Connect Option System Interconnection Diagram in Chapter 6, DIA­GRAMS and proceed as follows:
1. Connect AC power adapter plug to power jack on the All-in-One PC (Q-VISION) in accordance with manufacturer’s instructions (provided with PC).
2. If optional keyboard and/or mouse are provided, connect these in accor­dance with manufacturer’s instructions (provided with PC).
3. Connect Ethernet cable from All-in-One PC (Q-VISION) to an Ethernet Switch provided with Q-VISION OCP.
4. Connect Ethernet cable EL75-058-XXX between Ethernet Switch and Gen­erator OCP Interface Board (A Y40-073S) connector J1 (see Figure 2-24a).
5. Install the Ethernet Switch Power Supply connector, and plug transformer into 110 VAC, 60Hz electric outlet (220 VAC, 50Hz power source applica-
Revision W HF Series X-ray Generators - Service Manual
2-44 Quantum Medical Imaging
Chapter 2 Assembly & Installation
GENERATOR LOGIC BOARD (AY40-006S)
ETHERNET CABLE CONNECTION
OCP INTERFACE BOARD (AY40-073S)
tions are compatible with the supplied dual v oltage power adaptor). See Figure 2-24b.
Figure 2-24a. OCP Interface Board - AY40-073S (Ethernet Connection)
Figure 2-24b. Ethernet Switch Box (Grounded)
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-45
Chapter 2 Assembly & Installation
6-PIN MODULAR PLUG
6. The Power/Expose Control Box (AY20-271) must be installed between the Generator and the "Non-patient Area" in which the Q-VISION OCP and the Operator will be positioned. The Power/Expose Control Box should be securely mounted in close proximity to the Operator’ s position for easy access to the ON/ STANDBY Switch, and in clear view of the Q-VISION OCP. A 30-inch expansion cable for the 2-stage Expose Switch is hard­wired to the Power/Expose Control Box.
Figure 2-24c. Power/Expose Control Box
7. A prefabricated 50-ft. or 100 ft. cable with 6-pin modular plugs (AY51­243-XXX), is supplied for connect ing the Power/Expose Cont rol Box to the generator OCP Interface Board (A Y40-073S ). See Figur e 2-24c. DO NOT route cable through an opening in the generator containing a High Volt­age Cable (see Figure 2-13).
8. On systems provided with a Wall Mount Kit (option QGV-WM), install the wall mount components (VESA Mount Adapter, VESA Mount Plate, and Wall Mount Bracket) in accordance with instructions provided with the Wall Mount Kit.
9. Ensure wall attachment hardware is suitable for wall construction type (each screw capable of supporting 25 lbs. pull load).
Integrated Workstation Operator Control Panel
If the generator is to be integrated with an external digital imaging worksta­tion computer for operation as a generator control panel, the "Q-Connect" option must be installed in the generator. This option provides an Ethernet adapter to enable connection of the generator to the workstation personal computer (PC). T ypically, this adapter is factory installed, and is ready for use when the unit is shipped. If the Q-Connect Ethernet adapter is needed,
Revision W HF Series X-ray Generators - Service Manual
2-46 Quantum Medical Imaging
Chapter 2 Assembly & Installation
DANGER HIGH VOLTAGE
RESET
11
1
10
20
1
1
1
1
1
1
RS-232
1
4
1
TP10
R23
D19
JP11 C43
R18
C41
JP12
U10
C27
U19
R35
R34
TP9
JP7
R13
R12
JP6
JP16
R11
R16
TP1
D16
PD6PD7
R37
D2
D4 D3
TB1
P
R
E
P
E
X
P
O
S
E
C
O
M
M
G
N
D
upgrade kits are available (contact Quantum Technical Support for more information).
Refer to the Integration System Service Manual provided, depending on the the type system integration, for system interconnection and softwar e installa­tion instructions. A general Q-Connect Option System Interconnection Dia­gram is provided in Chapter 6, DIAGRAMS in this manual for reference.
Optional Remote Expose Switch Connection This option is not applicable on generators equipped with Q-Con-
nect option. Optional remote expose switch (R80-HS) is wired into the OCP
Board A16A1 (P/N AY40-004S) at TB1 as shown in Figure 2-25, below. If desired, removing jumpers JP6 and JP7 on OCP Control Board A16A1 (AY40­004S1) disables the operation of the PREP and EXPOSE buttons on the oper­ator control panel (see figure below).
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-47
Figure 2-25. Operator Control Panel (OCP) Control Board Terminal Block TB1
Connections for Remote Expose Switch

INITIAL POWER UP

With the generator completely assembled and when electrical checks are com­plete, the system is ready for initial power up. Proceed as follows:
Chapter 2 Assembly & Installation
1. Set ON/OFF circuit breaker CB1 on the generator cabinet to ON position.
2. Set the power on/standby switch on the Operator Control Panel or Power/ Expose Control Box on Q-VISION equipped generators to the on ( ) posi-
tion.
3. The system begins the start up sequence (the "Welcome Screen" is displa yed and "Testing. Standby." is flashing). After a few seconds, the normal opera­tion screen will display.
IMPORTANT! If the system is a STORED ENERGY (SE) type, and it has been longer than one month since being shipped from factory, allow the sys­tem approximately four hours to re-charge the batteries before taking any exposures. The sys­tem must be turned on with auto-shutoff disabled to re-charge. Refer to "Generator Time-out Set­ting" paragraph in Chapter 3, CALIBRATION for detailed instructions.
4. If necessary, the operator control panel LCD display intensity can be increased or decreased by adjusting the rotary potentiometer on the rear side of the operator console.

TRANSFORMER TAP VALIDATION

1. On single-phase non-SE versions, using a DMM, measure the DC voltage across the (+) and (-) terminals of the Anode DC capacitor bank (i.e., C13­C16). The voltage should be 400 (±20) VDC. If not, re-check line voltage and adjust transformer taps as required. (Refer to "Transformer Line Adjust­ment" procedure in this chapter for detailed instructions.)
2. On SE versions, verify that 220VAC (±10%) exists between A8P1-4 and A8J1-7. If these voltages are out of range, test the incoming AC line and con­firm the correct taps have been selected at TB1. Monitor the charger output voltages at A8TP2 (positive supply) and A8TP3 (negative supply) while allow­ing the chargers to recharge the battery tray(s) until each test point rises to ±205 (±1.0) VDC. Note: Charging to these levels may take up to 2 hours. If these levels cannot be achieved, check the integrity of the batteries in the affected side of each tray.
3. On three-phase models (see Figures 6-49 and 6-50), correct transformer tap connections will yield 220 VAC measured between phases A, B and C as follows:
a. Phase A located on Line Monitor Board A8 Fuse F4. b. Phase B located on Line Monitor Board A8 Fuse F5. c. Phase C located on Line Monitor Board A8 Fuse F6. d. In addition, using a DMM, measure 265 VAC between terminal block
TB1 (neutral) and each Phase A, B and C.
Revision W HF Series X-ray Generators - Service Manual
2-48 Quantum Medical Imaging
Chapter 2 Assembly & Installation
DC voltage measurement should be made across the (+) and (-) terminals of the anode DC capacitor bank (i.e., capacitors C13 - C16). The measured voltage should be +400 (±20) VDC. If not, re-check line voltage and adjust transformer taps as required to achieve as close to +400 VDC as possible.

Q-VISION OPERATOR CONTROL PANEL - NETWORK SETTINGS VERIFICATION

NOTE! Both the generator cabinet and Q-VISION network IP addresses are set at the factory to assure initial communication between the two components. If the generator will be conncetred to the internet (such as when utilizing the remote servicing feature), the factory default IP addresses must be changed in accordance with facility Inoformation Technology (IT) administra­tor’s requirements.
If system is equipped with a Q-VISION operator control panel, verify Generator Network Settings as follows:
1. Set ON/OFF circuit breaker CB1 on the generator cabinet to ON position.
2. Set the power on/standby switch on the Power/Expose Control Box to the on () position.
3. The system begins the start up sequence. After a few seconds, the Q- VISION control panel screen will display.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-49
Chapter 2 Assembly & Installation
4. Touch the Help button located on the lower right corner of the Q­VISION screen (shown on right). The Generator Help Services Screen (Figure 2-25a) appears:
Figure 2-25a. Generator Help Services Screen
5. Touch the Login button at the lower left corner of the screen. A password entry keypad displays.
Figure 2-25b. Enter Password Screen
6. In the Enter Password screen, enter the default Service Technician password (58623) using the keypad, then select Enter. The word "Login" is replaced with "FieldTech" on the Generator Help Services screen to indicate successful login.
Revision W HF Series X-ray Generators - Service Manual
2-50 Quantum Medical Imaging
Chapter 2 Assembly & Installation
7. On the Generator Help Services screen, select Service option. The General Service Options Menu appears:
Figure 2-25c. General Service Options Menu
8. Select Network Settings. The following screen appears:
Figure 2-25d. Network Settings Screen
9. The Network Settings Screen displays the network configuration settings of the personal computer housed within the all-in-one Q-VISION operator con­trol panel. Verify the fields are set as follows:
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-51
Chapter 2 Assembly & Installation
ONLY PRESS THESE BUTTONS WHEN DIRECTED TO DO SO BY TECHNICAL SUPPORT
•DHCP = Off
IP Address = 10.0.0.71
Subnet mask = 255.255.255.1
Default gateway = 10.0.0.1
Preferred DNS server = 10.0.0.3
10. If necessary, each of these fields can be modified by selecting the field, and then entering the correct information using the data input Touch Pad that appears. Note that the touch pad display includes numerical input buttons 0 through 9, including a decimal point, a plus sign (+), a minus sign (-), a clear (Clr) button, a delete (Del), a Cancel button and an Enter button; however, erroneous input via the numbers buttons will cause the displayed address to turn red, indicating that an error has been made either in acceptable value or entry format. Should an input error occur, it can be corrected by touching the Del button, followed by touching the correct key. The Clr button clears the Address window, and allows immediate re-entry of Address values.
11. Touch the Apply button to save new settings.
12. Return to the Generator Help Services Screen by touching the Back to Ser- vice button at lower left corner of screen.
13. On the Generator Help Services screen, select Generator Service. The Generator Service Options Menu appears:
Figure 2-25e. General Service Options Menu
Revision W HF Series X-ray Generators - Service Manual
2-52 Quantum Medical Imaging
Chapter 2 Assembly & Installation
CURRENT IP ADDRESS OF GENERATOR Q-NODE
14. Select Service Configuration. The following screen appears:
Figure 2-25f. Service Configuration Screen
15. The Service configuration Screen displays the network configuration settings of the generator cabinet, specifically of the Q-Node board (located on the Logic Board A1), which is a component of the Q-Connect generator option. The current Generator IP Address appears in the Generator IP Address field as shown in Figure 2-25f (10.0.0.175 is assigned at the factory as the default address).
16. Touch the Identify button. An audible tone will be emitted by the Power/ Expose Control Box if the Q-VISION OCP and the generator cabinet IP are communicating properly via the ethernet network connection.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-53
Chapter 2 Assembly & Installation
17. If the generator IP address needs to be changed, press the Load button to causes the lower portion of the "Service Configuration" screen to populate with data relevant to the Generator with the IP Address shown in the top por­tion of the screen, as shown below:
Figure 2-25g. Service Configuration Screen
18. T o change the IP Address, SubnetMask, or DefaultGateway field, touch the ... button to the right of the field to open a data input Touch Pad, which oper­ates in the same manner as described previously on the Network Settings Screen.
19. After all changes are made, touch Save button. This saves the new IP Address settings for the Generator.
Note: The Generator Q-Node and Q-VISION Subnet Addresses must match in order for the Service Configuration program to become operational. Once communication between the Q­VISION OCP Computer and the Generator has been established, the technician may then set the Generator IP Address in accordance with instructions provided by the cognizant IT professional for the facility at which the System is installed.
Revision W HF Series X-ray Generators - Service Manual
2-54 Quantum Medical Imaging
Chapter 2 Assembly & Installation
2.5 - 3.0 ms
14.0 to
15.0V(peak)
A10TP15
Waveform
14.0 to
15.0V(peak)
A10TP14
Waveform

ROTOR DRIVE BOARD A10 (AY40-013T) PERFORMANCE VERIFICATION

Note: This procedure is not required on systems equipped with high speed starter option.
Refer to Rotor Drive Board A10 (AY40-013T) Terminal Block Location Diagram in Chapter 6, DIAGRAMS and verify Rotor Drive Board adjustments as follows:
1. Set A10R20 to approximately mid range.
2. Connect one channel of oscilloscope to A10TP15 and second channel to A10TP14.
3. Press and hold the PREP key for approximately 3 seconds. Verify that during the run cycle, the amplitude of the phase and main pulses are +14 VDC to +15 VDC and waveforms are as shown in Figure 2-26 (i.e., the peaks of waveforms at TP15 and TP14 should be between 2.5 and 3.0 milliseconds apart).
4. Press and hold PREP key for approximately 3 seconds.
5. Verify the peak v oltage at A10TP14 after PREP has been pressed for at least 2
6. Set A10R24 to approximately midrange.
7. Press and hold PREP key for approximately 3 seconds.
Figure 2-26. Rotor Board Phase and Main Pulses
seconds (rotor in boost mode) is A10R20 to achieve the correct voltage.
greater than
14 V
. If necessary , adjust
PEAK
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 2-55
Chapter 2 Assembly & Installation
8. Verify the peak voltage at A10TP15 after PREP key has been pressed for at least 2 seconds (rotor in boost mode) is
adjust A10R24 to achieve the correct voltage.
9. Connect a DMM between A10TB1-1 and A10TB1-3.
10. Press the PREP key and verify the voltages as follows:
• For 110 VAC stator tubes (Toshiba E7242X, E7813X, E7252X, E7239X)
96 - 140 VAC during boost, 54 - 66 VAC during run
• For 220 VAC stator tubes (all others tubes not listed above)
183 - 227 VAC during boost, 54 - 66 VAC during run
11. Install cabinet cover.
12. Proceed with system calibration procedure described in Chapter 3, CALIBRATION.
greater than
14 V
. If necessary ,
PEAK
Revision W HF Series X-ray Generators - Service Manual
2-56 Quantum Medical Imaging
Chapter
3
CALIBRATION
3-1
3-2
Chapter 3 Calibration

OVERVIEW

This chapter provides instructions for configuring and calibrating the HF Series X-ray generator following installation. For additional information, NEMA Standards Publica­tion/No. XR8-1979(R1994), Test Methods for Diagnostic X-ray Machines for Use Dur­ing Initial Installation is recommended as a reference.
IMPORTANT NOTE: If the generator is equipped with the Q-Connect option" (Ethernet Adapter) for Agfa DiRex Integration (i.e., the Agfa NX workstation computer functions as the operator control panel), proceed to Appendix B, CALIBRATION (Agfa DiRex) for generator configuration and calibration instructions. If the generator is equipped with the "Q-Connect option" (Ethernet Adapter) for use with a Q-VISION operator control panel (OCP), or it is integrated with a workstation computer other than an Agfa DiRex Integration (e.g., the Canon DR Integration System, Agfa NX-DR Series Integration), refer to Appendix D, CALIBRATION (Insight - Canon) for generator configuration and calibration instructions.

REQUIRED TOOLS AND MATERIALS

The following lists the tools and materials required to perform the procedures contained in this chapter:
Digital Voltmeter: Fluke Model 87 (or equivalent)
kV and mAs meter: Radcal Dynalyzer III (or equivalent)
Radiation meter (0-1000 mR)
Oscilloscope, 50 Mhz storage: Tek 2210 (or equivalent)
5 CM, 10 CM, 15 CM, 25 and 30 CM acrylic phantoms (required for calibrating systems equipped with QG-AEC option).

SYSTEM CONFIGURATION/CALIBRATION

The HF Series x-ray generator has been factory configured and calibrated. It is nec­essary, however, to perform the following procedures to validate system compliance. Perform the procedures in the sequence presented, unless otherwise instructed.
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging, LLC 3-3
Chapter 3 Calibration
PRESS TO
ACCESS
PASSWORD
MENU
FIELD SELECT KEYS

ACCESSING SERVICE MODE

Calibration and configuration can only be performed by accessing the X-ray gen­erator’s Service Mode. To gain access to the Service Mode, password entry is required. This is to prevent unintended and/or unauthorized changes to system settings. To access the Service Menu, proceed as follows:
1. Set the ON/OFF circuit breaker CB1 on the generator cabinet to ON position.
2. Set the power on/standby switch on the Operator Control Panel (OCP) to the standby ( ) position, then to the on ( ) position.
3. The "Welcome Screen" appears with "Testing. Stand by." displayed in the lower portion of screen (see Figure 3-1.).
Figure 3-1. Welcome Screen
4. Within five seconds of start up, press the upper left field select key as shown in Figure 3-1). The Password menu displays:
Revision W HF Series X-ray Generators - Service Manual
Figure 3-2. Password Menu
3-4 Quantum Medical Imaging, LLC
Chapter 3 Calibration
5. In the Password menu, enter the default Service Technician password (58623) using the field select keys next to the corresponding numbers. (This password enables access to Service Mode for modification of system configuration, calibration and settings and for access to the APR Edit utility.)
6. Upon validation of the password, the Service Mode Menu displays:
Figure 3-3. Service Menu
The Service Menu offers a selection of service-oriented system utilities, each of which is described briefly below:
Configuration utility: Provides access to system configuration settings,
such as AEC type, rotor drive settings, tube type, etc.
Calibration utility: Provides access to system calibration functions, such
as filament, mA, kVp calibration and, if equipped with QG-AEC option, AEC and film/screen calibrations.
Reports utility: Provides access to Service Reports menu, which
provides Generator, Tube, Error Warnings, and Diagnostic reports (refer to Chapter 5, SERVICE MAINTENANCE for more information).
Self-Test utility: Currently not accessible.
Last Calibration Date: Currently not accessible.
Settings utility: Provides access to system settings, such as the
language, date, time, date format (international or U.S.) and backup time/mAs settings.
APR Edit utility: Used to edit existing APR screens or use to create new
screens (refer to Chapter 4, APR EDITOR in the operator’s manual for more information).
Pressing the key next to the Exit field returns the system to the Start Up screen (see Figure 3-1).
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging, LLC 3-5
Chapter 3 Calibration

SYSTEM CONFIGURATION

Press the key next to the Configuration field to access system configuration functions. The Service Configuration Menu displays:
Figure 3-4. Service Configuration Menu
The Service Configuration Menu provides access to the system's configuration settings. For proper operation of the generator, the settings must accurately reflect the actual configuration of both the generator and the components to which it is interfaced. Typically, these settings are made at the factory in accor­dance with sales order specifications. However, it is recommended that you review the current configuration settings for accuracy. The following briefly describes the various Service Configuration Menu selections:
Printer: Enables data output to an external label printer when one is interfaced to the generator
Generator Model setting: Accesses the generator maximum power output derating setting (from 80% to 100%). Access to other functions is restricted to authorized QMI factory personnel only
AEC setting: displays AEC option setting (i.e., AEC "Standard", AEC "Universal", AEC OFF)
Rotor Drive setting: Accesses the rotor drive settings - 60 Hz (no high speed starter), 50 Hz (no high speed starter), Quantum internal high­speed starter (QMI HSS), or AID external high-speed starter (AID HSS)
Edit Welcome/Facility: Accesses the facility name text editing and, on ODYSSEY systems, the dealer name text editing utilities, which provide functions necessary to edit the information appearing on the "Welcome Screen" displayed during start-up
Main: returns the system to the Service Menu (see Figure 3-3)
Tube Type setting: Accesses the Tube Type Setting screen, for selection of x-ray tube model used with system
Edit Digital: Accesses the digital receptor settings including digital receptor type (i.e., Canon, Konica, CMT, Generic, or NONE), Universal system type (i.e., Quantum QV-800, Pausch COSMOS-2, or NONE), and PREP HOLD setting (on or off).
Revision W HF Series X-ray Generators - Service Manual
3-6 Quantum Medical Imaging, LLC
Chapter 3 Calibration
Ion Chamber Configuration setting: Accesses the Ion Chamber Configuration screen, which is used to configure each ion chamber interfaced with the system (two separate ion chambers, of either slope or level type, can be interfaced to the generator)
Tomo setting - Accesses the Tomo configuration screens for setting up the system for interfacing with the Siemens or Pausch tomography systems (available settings are Siemens, Pausch, or OFF (none))
To change a configuration setting, press the key next to the configuration selec­tion to be changed. After all configuration settings are set appropriately, press the key next to the Main field. The configuration settings are saved in non-vola­tile read-only memory (NVROM) and the system returns to the main Service Menu (see Figure 3-3).
The following paragraphs provide specific instructions for setting each of the available configuration selections.
Printer On/Off Setting
When configured to do so, the HF Series X-ray generator will print post-exposure factors (kVp, mA, time, mAs, etc.) to an externally interfaced device such as a label printer. The label printer must comply with the follow­ing serial communication protocol:
1 stop bit, no parity
between the printer and the RS-232 A16J2 9-pin connector on the rear panel of the Operator Control Panel.
The Service Configuration Screen displays the current printer setting. When set to On, the generator will provide serial output data to an externally con­nected printer. When set to Off, serial data output for a printer is disabled.
Note: The printer setting has no affect on upload or download functionality using Q-Ware software.
. To connect a printer, attach an RS-232 serial cable
9600 baud rate, no handshake, 1 start bit,
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging, LLC 3-7
Chapter 3 Calibration
PRESS THIS KEY TO SET THE POWER OUTPUT DERATING PERCENTAGE
Generator Configuration/Derating Settings
The Generator Configuration Screen, shown below, displays the generator model and the power output derating setting (presented as a percentage of total power output (kW) capacity).
Figure 3-5. Generator Configuration Screen
The power output derating function is used to limit the power output of the system to a percentage of the maximum power capacity. For example, on a 50 kW system, setting the "KW%" to 90% will limit the maximum power out­put to 45 kW (50kW x .9 = 45kW). To set the power output derating, pro­ceed as follows:
1. Press the key next to the KW% field to choose a derating percentage (choices are 80%, 85%, 90%, 95% or 100% [i.e., no derating]).
2. Press the key located next to the Save field. The derating percentage setting is saved in non-volatile read-only memory (NVROM) and the sys­tem returns to the Service Configuration Menu (see Figure 3-4).
Revision W HF Series X-ray Generators - Service Manual
3-8 Quantum Medical Imaging, LLC
Loading...