Quantum Odyssey, Q-Vision Service manual

ODYSSEY, Q-VISION
HF Series™
X-ray Generators
Service Manual
Manual Part No. DC30-011 Revision W
EU Authorized Representative: Medizintechnik Berlin GmbH Altentreptower, Strasse 59 12683 Berlin - Germany Phone: +49-302-82 4726 Fax: +49-302-82 6382 E-mail: Medizintechnik.Berlin.GmbH@t-online.de
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC. (QMI)
This Service Manual is incomplete without Operator’s Manual part number DC30-010.
Copyright© 2010 QMI
Quantum Medical Imaging, LLC
2002-B Orville Drive North Ronkonkoma, NY 11779-7661 USA Phone: (631) 567-5800 Fax: (631) 567-5074 E-mail: info@qmiteam.com www.quantummedical.net
Made in U.S.A.

Revision History

REVISION DATE TYPE OF MODIFICATION
E 1/12/02 Incorporated ECO 0647, 0651, 0687 F 2/25/02 Incorporated ECO 0714, 0718, 0738, 0742 G 5/15/02 Incorporated ECO 0642, 0748, 0753, 0770, 0787, 0803, 0809,
0821, 0827, 0829, 0830, 0847, 0859, 0881, 0888 H 9/26/02 Incorporated ECO 0787 J 11/11/02 Incorporated ECO 0936, 1003, 1009, 1010, 1017, 1028, 1030 K 3/31/03 Incorporated ECO 1093, 1102, 1103, 1107, 1115, 1124, 1146 L 7/11/03 Incorporated ECO 1255, 1256 M 7/29/03 Incorporated ECO 1269, 1270, 1272 N 12/19/03 Incorporated ECO 1144, 1216, 1347, 1359, 13 62, 1380, 1383,
1400 P 9/10/04 Incorporated ECO 1346, 1438, 1485, 1518 Q 4/22/05 Incorporated ECO 1565, 1570, 1600, 1606, 1607 R 12/20/05 Incorporated ECO 1660, 1663, 1679 S 5/22/07 Assigned new part number to Non-QMI HSS ver sio n man ua l;
Incorporated ECO 1697, 1715, 1757, 1794, 1890 T 5/8 /2 00 8 Incorporate ECO 1895, 19 20 , 19 32 , 19 7 7, 198 5 U 11/21/2008 Incorporate ECO 2045, 2050, 2068, 2093, 2105 V 9/1/2009 Incorporate ECO 1980, 2160, 2166, 2169 W 1/15/2010 Incorporate ECO 2220, 2228, 2276
Page
Number
i thru xii W 4-1 thru 4-20 W B-1 thru B-56 W
1-1 thru 1-22 W 5-1 thru 5-96 W C-1 thru C-7 W
Rev
Page
Number
Rev
Page
Number
Rev
2-1 thru 2-56 W 6-1 thru 6-67 W D-1 thru D-66 W 3-1 thru 3-72 W A-1 thru A-16 W
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging i
Revision History
THIS PAGE INTENTIONALLY LEFT BLANK
Revision W HF Series X-ray Generators - Service Manual
ii Quantum Medical Imaging

Table of Contents

CHAPTER 1 - SPECIFICATIONS
SYSTEM CONFIGURATIONS ................................................................. 1-3
PHYSICAL SPECIFICATIONS ................................................................. 1-5
GENERATOR CABINET SPECIFICATIONS ......................................... 1-5
OPERATOR CONTROL PANEL SPECIFICATIONS ............................... 1-5
ODYSSEY HF Control Panel: ............................... ...................... 1-5
Q-VISION HF Control Panel: .................................................... 1-5
SYSTEM SPECIFICATIONS .............................................................. 1-5
POWER REQUIREMENTS .................................................... .................. 1-6
SINGLE-PHASE X-RAY GENERATORS .. ............................................. 1-7
25 kW Single Phase (Model QG-25) .... ....................................... 1-7
25 kW Single Phase (Model QG-25-5)* ...................................... 1-7
32 kW Single Phase (Model QG-32) .... ....................................... 1-7
32 kW Single Phase (Model QG-32-5)* ...................................... 1-7
40 kW Single Phase (Model QG-40) .... ....................................... 1-8
40 kW Single Phase (Model QG-40-5)* ...................................... 1-8
32 kW Single Phase, Stored Energy ........................................... 1-8
40 kW Single Phase, Stored Energy ........................................... 1-8
50 kW Single Phase, Stored Energy ........................................... 1-8
THREE-PHASE X-RAY GENERATORS ................................................ 1-9
32 kW Three Phase (Model QG-32-2) ........................................ 1-9
32 kW Three Phase (Model QG-32-3) ........................................ 1-9
40 kW Three Phase (Model QG-40-2) ........................................ 1-9
40 kW Three Phase (Model QG-40-3) ........................................ 1-9
50 kW Three Phase (Model QG-50-2) ........................................ 1-9
50 kW Three Phase (Model QG-50-3) ........................................ 1-9
65 kW Three Phase (Model QG-65) ......................... .................1-10
80 kW Three Phase (Model QG-80) ......................... .................1-10
HIGH-SPEED STARTER POWER REQUIREMENTS ............................1-10
GENERATOR AUXILIARY POWER SUPPLIES ........................... ...............1-15
PERFORMANCE SPECIFICATIONS ........................................................1-15
SYSTEM OPTIONS ..............................................................................1-19
SYSTEM OPERATING ENVIRONMENT ....................................... ............1-19
NON-OPERATING ENVIRONMENT .......... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .... .. .. ..1-20
CHAPTER 2 - ASSEMBLY & INSTALLATION
OVERVIEW .................. .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. .. . 2-3
PRE-INSTALLATION GUIDELINES ................................................... 2-3
Verify Equipment Location ........................................................ 2-3
Electrical Requirements ............................................................ 2-5
UNPACKING .................................................................................. 2-5
INSTALLATION .............................. ............... ............... ................. ....... 2-6
REQUIRED TOOLS AND MATERIALS ............................. .................. 2-6
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
iii
Table of Contents
CHAPTER 2 - ASSEMBLY & INSTALLATION (CONT’D)
GENERATOR CABINET SETUP ......................................................... 2-6
Transformer Line Tap Adjustment (Single Phase Non-SE Systems) 2-8
Transformer Line Tap Adjustment (Single Phase SE Systems) ....2-10
AC Input (Mains) Connection ...................................................2-12
High Voltage (H.V.) Tank ......................... ...............................2-15
High Voltage Cables ................. .. .. .. .. .. .. .... .. ... .. .. .. .. .... .. .. ... .. .. ...2-18
INTERFACE BOARD SETTINGS & CONNECTIONS ............................2-21
Door Interlock Switch Connection ............................................2-22
"In Use" Light Switch Connection .............................................2-22
Collimator Connections ............................................................2-23
Tube Stand/Peripheral Equipment Connections .........................2-23
Internal +28 VDC Power Supply Option (QG-28VPS) .................2-23
Tube Rotor Boost Voltage Verification (Systems without
High (Dual) Speed Starter Option) ..........................................2-24
RADIOGRAPHIC SYSTEM EQUIPMENT CONNECTIONS ....................2-25
Tube Rotor Cable (Systems without High (Dual) Speed
Starter Option) ......................................................................2-25
Tube Rotor Cable (Systems with AID External High
Speed Starter QG-HSS Option Only) ........................................2-27
Tube Rotor Cable (Systems with Internal High Speed
Starter Q-HSS Option Only) ................................ ....................2-30
Table/Wall Bucky Connections .................................................2-34
Table/Wall Ion Chamber Connections (AEC Option Only) ...........2-34
Interfacing to Canon Digital Radiographic Receptor
(When Applicable) .................................................................2-35
Interfacing to CMT Digital Radiographic Receptor
(When Applicable) .................................................................2-36
Interfacing to Siemens Ion Chambers (When Applicable) ...........2-36
Interfacing to Pausch Tomography System (When Applicable) ...2-39
OPERATOR CONTROL PANEL SETUP ..............................................2-41
ODYSSEY Operator Control Panel ............................................2-41
OCP Pedestal Mounting (QG-PDL Option) .................................2-41
OCP Wall Mounting (QG-WM Option) ................................ .......2-43
Q-VISION Operator Control Panel ............................................2-44
Integrated Workstation Operator Control Panel .........................2-46
Optional Remote Expose Switch Connection .............................2-47
INITIAL POWER UP ......................................................................2-47
TRANSFORMER TAP VALIDATION ..................................................2-48
Q-VISION OPERATOR CONTROL PANEL - NETWORK
SETTINGS VERIFICATION ............................................................2-49
ROTOR DRIVE BOARD A10 (AY40-013T) PERFORMANCE
VERIFICATION .............. ..............................................................2-55
CHAPTER 3 - CALIBRATION
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. 3-3
REQUIRED TOOLS AND MATERIALS ............................................... 3-3
Revision W HF Series X-ray Generators - Service Manual
iv Quantum Medical Imaging
Table of Contents
CHAPTER 3 - CALIBRATION (CONT’D)
SYSTEM CONFIGURATION/CALIBRATION ..............................................3-3
ACCESSING SERVICE MODE ...........................................................3-4
SYSTEM CONFIGURATION ........................................ ......................3-6
Printer On/Off Setting ........................ .. ..... .. .. .... .. .. .... ... .. .... .. .... .3-7
Generator Configuration/Derating Settings .................................3-8
AEC Setting .............................................................................3-9
Rotor Drive Setting ................ .... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .. .... .. . 3-10
Edit Facility/Welcome Facility Utility ......................................... 3-11
Tube Type Setting ...... .... .. .. ... .. .. .. .... .. .. .. ... .. .. .. .... .. .. .. ... .. .. .... .. . 3-12
Edit Digital Settings ................................................................3-13
Ion Chamber/Receptor Configuration .......................................3-14
Tomo Setting ............................ ............................................. 3-18
SYSTEM SETTINGS ......................................................................3-21
Set Date Screen ....... .. .... .... ... .. .... .. .. .... .. ... .... .. .. .... .. ..... .. .. .... .. . 3-22
Set Time Screen ................. ... .. .. .. .. .. .... .. ... .. .. .. .. .... .. .. ... .. .. .. .. ... 3-23
Film/Screen Defaults Screen ...................................................3-23
APR Format Setting ................................................................3-25
Language Setting ................................................................... 3-26
Back Up MAS Setting ................................ .............................. 3-27
Generator Timeout Setting ..................................................... . 3-28
Date/Time Format Setting ......................................................3-28
SYSTEM CALIBRATION ................................................................. 3-29
GENERAL CALIBRATION PROCEDURE NOTES ................................3-32
Film Speed Setting on non-AEC Systems .................................. 3-33
A/D Calibration ......................................................................3-34
kV Calibration Function ...........................................................3-37
Filament Offset Calibration ...................................................... 3-39
MA Calibration ..... ..................................................................3-43
AEC CALIBRATION .......................................................................3-46
AEC Calibration Procedures ..................................................... 3-48
NON-AEC REPRODUCIBILITY VERIFICATION .................................3-66
AEC REPRODUCIBILITY VERIFICATION (ONLY PERFORM
ON UNITS EQUIPPED WITH QG-AEC OPTION) .............................. 3-67
NON-AEC LINEARITY VERIFICATION .............................................3-67
AEC LINEARITY VERIFICATION (ONLY PERFORM ON
UNITS EQUIPPED WITH QG-AEC OPTION) ...................................3-68
MAXIMUM DEVIATION: KVP, TIME, MA, MAS ................................. 3-68
CHAPTER 4 - THEORY OF OPERATION
OVERVIEW .......................................................................................... 4-3
OPERATOR CONTROL PANEL CIRCUIT ............................................4-3
Microcontroller U3 Operation .....................................................4-3
Flash Memory U2 .....................................................................4-3
EEPROM U17 Operation ............................................................4-4
LCD Controller U12 Operation ........ .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .. .4-4
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
v
Table of Contents
CHAPTER 4 - THEORY OF OPERATION (CONT’D)
Switching Power Supply Circuit ................................................. 4-5
Front Panel Key Activation Detection Circuit ............................... 4-5
Jumpers Installation ............................................................. .... 4-5
TechVision Compatibility .......................................................... 4-5
GENERATOR LOGIC CIRCUIT ......................................................... 4-5
Generator/OCP Communication ................................................ 4-6
Master and Slave Microcontroller Operation ............................... 4-6
FILAMENT CONTROL REGULATOR CIRCUIT .................................... 4-7
KVP & FILAMENT CONTROL CIRCUIT .. ............................................ 4-7
LINE MONITOR CIRCUIT (NON-SE SINGLE-PHASE AND
THREE-PHASE SYSTEMS) .............................................................. 4-8
Input Line Monitor Circuit ............................................ ............. 4-8
Capacitor Discharge Circuit ....................................................... 4-9
Auto-Shutoff Circuit ................................................................. 4-9
Three-Phase Line Monitor Circuit .............................................4-10
LINE MONITOR CIRCUIT (STORED ENERGY SYSTEMS) ...................4-10
Battery Monitor Circuit ............................................................4-10
Storage Capacitor Discharge Circuit .........................................4-10
Generator Auto-Shutoff Circuit .................................................4-11
Battery Charger Circuit ............................................................4-11
POWER SUPPLY CIRCUIT ..............................................................4-12
+48 VDC Power Supply ...........................................................4-12
+15 VDC Power Supply ...........................................................4-12
+24 VDC Power Supply ...........................................................4-12
ROTOR BOARD CIRCUIT (NON-HIGH SPEED STARTER SYSTEMS) ...4-13 ROTOR BOARD CIRCUIT (QUANTUM INTERNAL HIGH
SPEED STARTER SYSTEMS) .........................................................4-13
HSS Control Board A20A1 (AY40-035S) ....................................4-14
HSS Drive Board A20A2 (AY40-036S) .......................................4-15
AEC BOARD CIRCUIT ....................................................................4-16
Standard Type AEC .................................................................4-16
Universal Type AEC .................................................................4-17
+300 VDC Power Supply .........................................................4-18
AEC Control Circuitry ...............................................................4-18
Tomography Circuitry ..............................................................4-20
CHAPTER 5 - SERVICE INSTRUCTIONS
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. 5-3
DISPOSAL OF BATTERIES AND ACCUMULATORS
(DIRECTIVE 2006/66/EC) .............................................................. 5-3
SERVICE MAINTENANCE ................................................................ 5-4
Visual Inspection ..................................................................... 5-4
TROUBLESHOOTING ............... ........... ............ ........... .......... ........... 5-5
Warning Messages ................................................................... 5-5
Error Codes ............................................................................. 5-5
Optional +28 VDC Power Supply Troubleshooting .....................5-48
Revision W HF Series X-ray Generators - Service Manual
vi Quantum Medical Imaging
Table of Contents
CHAPTER 5 - SERVICE INSTRUCTIONS (CONT’D)
SERVICE REPORTS .................................................................. .....5-49
Accessing Service Reports - ODYSSEY X-ray Generators ............5-49
Accessing Service Reports - Q-VISION Operator Control Panel ...5-56
Accessing Service Reports - Integrated X-ray Generators ..........5-63
REMOVAL/REPLACEMENT PROCEDURES ........................................5-72
Generator Cabinet Cover Removal ...........................................5-72
Generator Cabinet Cover Replacement .....................................5-72
High Voltage (H.V.) A17 Tank Removal .................................... 5-73
High Voltage (H.V.) Tank A17 Installation ................................ 5-74
Power Modules A18 and A19 (P/N AY20-047) Removal ............5-75
Power Modules A18 and A19 (P/N AY20-047) Replacement .......5-77
Logic Board A1 (P/N AY40-006S) Removal ...............................5-78
Logic Board A1 (P/N AY40-006S) Replacement .........................5-78
KVP Control Board A2 (P/N AY40-003S) Removal .....................5-80
KVP Control Board A2 (P/N AY40-003S) Replacement ...............5-80
Filament Control Board A5 (P/N AY40-007T) Removal ............... 5-82
Filament Control Board A5 (P/N AY40-007T) Replacement ........5-82
Power Supply Board A6 (P/N AY40-005T) Removal ...................5-83
Power Supply Board A6 (P/N AY40-005T) Replacement .............5-83
Line Monitor Board A8 Removal .............................................. 5-84
Line Monitor Board A8 Replacement ...................... ..................5-84
Stored Energy Systems Line Monitor/Charger Board A8
(AY40-028T) Removal ........................ .. ..... .. .... .. .. .... ... .. .... .. .. . 5-85
Stored Energy Systems Line Monitor/Charger Board
(AY40-028T) A8 Replacement ................................................5-86
Rotor Drive Board A10 (P/N AY40-013T) Removal .................... 5-87
Rotor Drive Board A10 (P/N AY40-013T) Replacement .............. 5-87
Standard AEC Control Board A11 (P/N AY40-031S) Removal .....5-88
Standard AEC Control Board A11 (P/N AY40-031S) Replacement 5-88
Universal AEC Control Board A11 (P/N AY40-027S) Removal .....5-89
Universal AEC Control Board A11 (P/N AY40-027S) Replacement 5-89 OCP Microprocessor Board A16 (P/N AY40-004S1) Removal
(Not Applicable on Q-VISION or Integrated Generators) ..........5-90
OCP Microprocessor Board A16 (P/N AY40-004S1) Replacement
(Not Applicable on Q-VISION or Integrated Generators) ..........5-90
Q-HSS Equipped Systems Only: High-Speed Starter Control
Board A20A1 (P/N AY40-035S) Removal .................................5-91
Q-HSS Equipped Systems Only: High-Speed Starter Control
Board A20A1 (P/N AY40-035S) Replacement ...........................5-91
Q-HSS Equipped Systems Only: High-Speed Starter Driver
Board A20A2 (P/N AY40-036S) Removal .................................5-92
Q-HSS Equipped Systems Only: High-Speed Starter Driver
Board A20A2 (P/N AY40-036S) Replacement ...........................5-92
Stored Energy System - Battery Tray Removal .........................5-93
Stored Energy System - Battery Tray Replacement ...................5-95
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
vii
Table of Contents
CHAPTER 5 - SERVICE INSTRUCTIONS (CONT’D)
REPLACEMENT PARTS AND ORDERING INFORMATION ...................5-96
ORDERING INFORMATION .......... ... .... .. .. .... .. .. ..... .. .... .. .. .... .. ... .... .. .5-96
CHAPTER 6 - DIAGRAMS
Single-Phase (Non-STORED ENERGY) X-ray Generators, 25, 32 and
40 kW Parts Location Diagram (Sheet 1 of 2) ....................................... 6-3
Single-Phase (Non-STORED ENERGY) X-ray Generators, 25, 32
and 40 kW Parts Location Diagram (Sheet 2 of 2) ................................. 6-4
Single-Phase (STORED ENERGY) X-ray Generators, 20, 32 40 and
50 kW Parts Location Diagram (Sheet 1 of 2) ....................................... 6- 5
Single-Phase (STORED ENERGY) X-ray Generators, 20, 32, 40 and
50 kW Parts Location Diagram (Sheet 2 of 2) ....................................... 6-6
Three-Phase X-ray Generators, 32 and 40 kW Parts
Location Diagram (Sheet 1 of 2) .......................................................... 6-7
Three-Phase X-ray Generators, 32 and 40 kW Parts Location
Diagram (Sheet 2 of 2) ....................................................................... 6-8
Three-Phase X-ray Generators, 50, 65 and 80 kW Parts Location
Diagram (Sheet 1 of 2) ....................................................................... 6-9
Three-Phase X-ray Generators, 50, 65 and 80 kW Parts Location
Diagram (Sheet 2 of 2) ......................................................................6-10
Three-Phase X-ray Generators, 50, 65 and 80 kW (with QMI-HSS)
Parts Location Diagram (Sheet 1 of 2) .................................................6-11
Three-Phase X-ray Generators, 50, 65 and 80 kW (with QMI-HSS)
Parts Location Diagram (Sheet 2 of 2) .................................................6-12
Generator System Cabling Diagram (Internal High-Speed Starter) ..........6-13
Generator System Cabling Diagram (External High-Speed Starter Systems) 6-14
System Cabling Diagram (with Q-Connect Option) ................................6-15
HF Generator to Canon/TechVision System Interconnection Diagram
(Non-Widescreen) .............................................................................6-16
HF Generator to Canon/TechVision System Interconnection
Diagram (Widescreen) ....... ................................................................6-17
17 x 17 Midwest Bucky Interconnection Diagram ..................................6-18
17 x 17 Progeny and L-F Bucky Interconnection Diagram ..................... .6-19
14 x 36 Midwest Bucky Interconnection Diagram ..................................6-20
Single-Phase 208 - 260 VAC Input (Non-STORED ENERGY) X-ray
Generators, 25, 32 and 40 kW Interconnection Diagram .......................6-21
Single-Phase 380 - 480 VAC Input (Non-STORED ENERGY) X-ray
Generators, 25, 32 and 40 kW Interconnection Diagram .......................6-22
Single-Phase (STORED ENERGY) X-ray Generator, 20 kW
Interconnection Diagram ................................................................. ...6-23
Single-Phase (STORED ENERGY) X-ray Generators, 32, 40 and
50 kW Interconnection Diagram .........................................................6-24
Three-Phase (380-480) Input (32, 40 and 50 kW without High
Speed Starter) X-ray Generators Interconnection Diagram ....................6-25
Revision W HF Series X-ray Generators - Service Manual
viii Quantum Medical Imaging
Table of Contents
CHAPTER 6 - DIAGRAMS (CONT’D)
Three-Phase (208-240) Input (32, 40 and 50 kW without High
Speed Starter) X-ray Generators Interconnection Diagram ................... 6-26
Three-Phase (50, 65 and 80 kW with AID High Speed Starter) X-ray
Generators Interconnection Diagram ..................................................6-27
Three-Phase (32, 40 and 50 kW without High Speed Starter, with
Universal AEC Board) X-ray Generators Interconnection Diagram .........6-28
Three-Phase (65 and 80 kW with AID High Speed Starter, Universal
AEC Board) X-ray Generators Interconnection Diagram ........................6-29
Three-Phase (208-240) Input (32, 40 and 50 kW without High Speed Starter, with Universal AEC Board) X-ray Generators Interconnection Diagram AEC6-30 Three-Phase (208-240) Input (50 kW with QMI High Speed Starter) X-ray Generator
Interconnection Diagram .......... .. .... ... .... .. .. .... .. .. ..... .. .. .... .. .. ..... .. .... .. .. . 6-31
Three-Phase (208-240) Input (50 kW with QMI High Speed Starter and
Tomography Interface) X-ray Generator Interconnection Diagram ........ 6-32
Three-Phase (65 and 80 kW with QMI High Speed Starter) X-ray
Generator Interconnection Diagram ....................................................6-33
Three-Phase (65 and 80 kW with QMI High Speed Starter and Tomography
Interface) X-ray Generator Interconnection Diagram ...........................6-34
Operator Control Panel - ODYSSEY/QUEST Wiring Diagram ................... 6-35
Q-Connect Option System Interconnection Diagram .............................6-36
Operator Control Panel Circuit Block Diagram (Non-Q-VISION OCP) ...... . 6-37
Generator Logic Circuit Block Diagram ................................................. 6-38
Filament Regulator Control Circuit Block Diagram ................................. 6-39
KVP & Filament Control Circuit Block Diagram ..................................... . 6-40
Single-Phase (Non-STORED ENERGY) and Three-Phase Line Monitor
Circuit Block Diagram ........................................................................6-41
Single-Phase STORED ENERGY Line Monitor Circuit Block Diagram ........6-42
Interface Board A9 (AY40-023T/AY40-023T1) Schematic Diagram ......... 6-43
Interface Board, Digital Receptor A9 (AY40-034T) Interface Diagram .....6-44
Interface Board, Digital Receptor A9 (AY40-061T) Interface Diagram .....6-45
Power Supply Board A6 (AY40-005T) Schematic Diagram 46
Rotor Circuit Block Diagram ................................................................6-47
QMI High-Speed Starter Circuit Block Diagram .....................................6-48
Logic Board A1 (AY40-006S) Test Point/Jumper Location Diagram .........6-49
KVP Control Board A2 (AY40-003S) Test Point/Jumper Location Diagram 6-50 Filament Control Board A5 (AY40-007T) - Test Point/Jumper Location
Diagram ............................... .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. . 6-51
Power Supply Board A6 (AY40-005T) - Test Point/Jumper Location
Diagram ............................... .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. .. .. .. .. ... .... .. .. .. . 6-52
Line Monitor Board A8 (AY40-022T1), Single-Phase Systems - Test
Point/Jumper Location Diagram ......................................................... 6-53
Line Monitor/Battery Charger Board A8 (AY40-028T), Stored-Energy
Systems - Test Point/Jumper Location Diagram ...................................6-54
Line Monitor Board A8 (AY40-022T2/T3), Three-Phase Systems - Test
Point/Jumper Location Diagram ......................................................... 6-55
Line Monitor Board A8 (AY40-053S2/AY40-053S3), Three-Phase QMI
High-Speed Starter Systems - Test Point/Jumper Location Diagram ...... 6-56
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
ix
Table of Contents
CHAPTER 6 - DIAGRAMS (CONT’D)
Interface Board (No HSS Non-Digital) A9 (AY40-023T) - Terminal
Block/Jumper Location Diagram ..........................................................6-57
Digital Interface Board (No HSS with Digital Receptor) A9 (AY40-034T) -
Terminal Block/Jumper Location Diagram ............................................6-58
Interface Board (HSS Non-Digital) A9 (AY40-061T) Terminal Block/Jumper
Location Diagram ............................................................................. .6-59
Digital Interface Board (HSS with Digital Receptor) A9 (AY40-062T)
Terminal Block/Jumper Location Diagram ............................................6-60
Rotor Drive Board A10 (AY40-013T) - Terminal Block Location Diagram .6-61 Standard AEC Board A11 (AY40-031S) - Test Point/Jumper Location
Diagram ...........................................................................................6-62
Universal AEC Board A11 (AY40-027S) - Test Point/Jumper Location
Diagram ...........................................................................................6-63
OCP Control Board A16A1 (AY40-004S1) - Test Point/Jumper
Location Diagram ............................................................................. .6-64
OCP Interface Board A1A1 (AY40-073S) - Test Point/Jumper Location
Diagram ...........................................................................................6-65
QMI High-Speed Starter Control Board A20A1 (AY40-035S) - Test
Point/Potentiometer Location Diagram ................. ...............................6-66
QMI High-Speed Starter Driver Board A20A2 (AY40-036S) - Test
Point/Fuse Location Diagram ..............................................................6-67
APPENDIX A - AEC CALIBRATION WITH DIGITAL RECEPTOR OR CR FILM PLATE
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. A-3
SPECIAL CANON DR SYSTEM CONFIGURATION INSTRUCTIONS ....... A-3
AEC CALIBRATION WITH DIGITAL RECEPTOR OR CR FILM PLATE
INSTALLED .................................................................................. A-4
AEC Calibration Instructions - One Ion Chamber Per Panel .........A-8
AEC Calibration Instructions - Single Panel, Dual Ion Chambers A-13
APPENDIX B - CALIBRATION (AGFA DIREX)
OVERVIEW .................. .. ... .. .. .. .... .. .. .. ... .. .. .... .. .. .. .. .. ... .. .... .. .. .. .. .. ... .... .. .. B-3
REQUIRED TOOLS AND MATERIALS ............................................... B-4
SYSTEM CONFIGURATION/CALIBRATION ............................................. B-4
ACCESSING SERVICE MODE ...... .. ... .... .. .. .... .. .. ..... .. .. .... .. .... .. ... .... .. .. B-4
FACTORY RESET ........................................................................... B-8
Factory Reset .......................................................................... B-8
Backup Configuration Utility ...................................................... B-9
GENERATOR CONFIGURATION ..................................................... B-10
Generator Settings ................................................ ................. B-13
Tube Settings ........................................................................ B-15
Ion Chamber Configuration .............. .... .. .... ... .. .... .. .. .... .. ... .... .. B-17
Film/Screen Default Settings ................................................... B-21
Revision W HF Series X-ray Generators - Service Manual
x Quantum Medical Imaging
Table of Contents
APPENDIX B - CALIBRATION (AGFA DIREX) (CONT’D)
Copy Film Screen Utility ..........................................................B-23
GENERATOR CALIBRATION ..........................................................B-25
General Calibration Procedure Notes ........................................B-26
A/D Calibration ................ .. ... .. .. .. .. .... .. .. ... .. .. .. .... .. .. .. ... .. .. .. .... . B-27
kV Calibration ........................................................................ B-30
Filament Offset Calibration ......................................................B-32
MA Calibration ......................................................................B-37
AEC CALIBRATION ........... .. .... ..... .. .... .. .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .B-40
Ion Chamber Balancing (Three-field Chambers) .......................B-43
AEC MAS Balance Verification .......................... .... .. .. ..... .. .. .... .. .B-45
Film Screen Selection .............................................................B-47
The AEC Calibration Menu ......................................................B-50
Optical Density Calibration ......................................................B-51
AEC Optical Density Calibration .......... .... ... .. .... .. .. .... .. ..... .. .. .... . B-52
AEC KV Calibration Procedure .................................................B-55
APPENDIX C - LOG4VIEW ERROR LOGGER APPENDIX D - (QG-DIG-CXDI CANON/QUANTUM
INTEGRATION) CALIBRATION
OVERVIEW .................. .............................................. ......................... D-3
REQUIRED TOOLS AND MATERIALS ............................................... D-4
ACCESSING GENERATOR SERVICE TOOLS ...................................... D-4
Accessing Generator Service Tools - Q-VISION Operator
Control Panel ......................................................................... D-5
Accessing Generator Service Tools - Workstation-Based
Operator Control Panel ........................................................... D-7
SYSTEM CONFIGURATION/CALIBRATION ............................................D-10
GENERATOR CONFIGURATION .....................................................D -12
Generator Settings Tab .............. .............................................D-12
Generator Calibrations Tab .....................................................D-13
Tube Settings Tab ..................................................................D-13
Physical Receptors Tab ...................... .. .. ..... .. .. .... .. .. ..... .. .. .... .. .D-15
Receptors Tab .......................................................................D-16
Ion Chambers Tab ...................... .. .... .. ..... .. .. .... .. .. .... ... .. .... .. .. .D-18
Licensing Tab ........................................................................D-20
GENERATOR CALIBRATION ..........................................................D-21
A/D Tab .................... .. .. .... ... .. .. .. .. .. .. .... ... .. .. .. .. .. .... .. ... .. .. .. .. .. .D-24
kVp Calibration Tab ............... .. .... .. .. .... .. ... .... .. .... .. .. ..... .. .. .... .. .D-28
Filament Offset Calibration ......................................................D-31
MA Calibration Tab .................................................................D-41
AEC CALIBRATION ........... .. .... ..... .. .... .. .. .... .. ... .... .. .. .... .. .. ..... .. .... .. .D-45
Ion Chamber Balancing (Three-field Chambers) .......................D-47
AEC MAS Balance Verification .......................... .... .. .. ..... .. .. .... .. .D-50
The AEC Calibration Tab ............................... ..........................D-52
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging
xi
Table of Contents
APPENDIX D - (QG-DIG-CXDI CANON/QUANTUM INTEGRATION) CALIBRATION (CONT’D)
Base OD Calibration Procedure .......................... ..................... D-56
AEC kVp Calibration Procedure ...............................................D-58
Copy Receptor Calibration Tab ................................................ D-62
Revision W HF Series X-ray Generators - Service Manual
xii Quantum Medical Imaging
Chapter
1
SPECIFICATIONS
1-1
1-2
Chapter 1 Specifications

SYSTEM CONFIGURATIONS

The tables below show model numbers of components in each type of HF Series X-ray Generator system (hereinafter referred to as the "HF Series X-ray Generators").
Note: The X-Ray Control Models shown in the following tables are not used if the HF generator system is integrated with computed radiographic (CR) or digital radiographic (DR) workstation computer (e.g., Agfa NX, Canon CXDI, etc.).
Single-Phase Model Generators
TM
SYSTEM
TYPE
ODYSSEY HF 32 kW
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 40 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
SYSTEM
TYPE
CATALOG
NO.
QG-3200
QG-4000
QGV-32 QG-32 QG-32G QGV-32-5 QG-32-5 QG-32G-5 QGV-40 QG-40 QG-40G QGV-40-5 QG-40-5 QG-40G-5
Three-Phase Model Generators
CATALOG
NO.
SYSTEM
MODEL
QG-32 QG-32G QG-3200
QG-32-5 QG-32G-5 QG-3200
QG-40 QG-40G QG-4000
QG-40-5 QG-40G-5 QG-4000
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL*
QGV-32
QGV-40
X-RAY
CONTROL
MODEL
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 50 kW
QG-3200
QG-4000
QG-5000
QG-32-2 QG-32G-2
QG-3200
QG-32-3 QG-32G-3 QG-40-2 QG-40G-2
QG-4000
QG-40-3 QG-40G-3 QG-50-2 QG-50G-2
QG-5000
QG-50 QG-50G
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-3
Chapter 1 Specifications
Three-Phase Model Generators
SYSTEM
TYPE
ODYSSEY HF 65 kW
ODYSSEY HF 80 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
Q-VISION HF 50 kW
Q-VISION HF 65 kW
Q-VISION HF 80 kW
CATALOG
NO.
QG-6500 QG-65 QG-65G QG-6500
QG-8000 QG-80 QG-80G QG-8000
QGV-32-2 QG-32-2 QG-32G-2 QGV-32-3 QG-32-3 QG-32G-3 QGV-40-2 QG-40-2 QG-40G-2 QGV-40-3 QG-40-3 QG-40G-3 QGV-50-2 QG-50-2 QG-50G-2 QGV-50 QG-50 QG-50G
QGV-65 QG-65 QG-65G QGV-65
QGV-80 QG-80 QG-80G QGV-80
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL
QGV-32
QGV-40
QGV-50
STORED ENERGY (SE) Model Generators
SYSTEM
TYPE
ODYSSEY HF 32 kW
ODYSSEY HF 40 kW
ODYSSEY HF 50 kW
Q-VISION HF 32 kW
Q-VISION HF 40 kW
Q-VISION HF 50 kW
CATALOG
NO.
QG-3200-SE QG-32-SE QG-32G-SE QG-3200
QG-4000-SE QG-40-SE QG-40G-SE QG-4000
QG-5000-SE QG-50-SE QG-50G-SE QG-5000
QGV-32-SE QG-32-SE QG-32G-SE QGV-32-SE
QGV-40-SE QG-40-SE QG-40G-SE QGV-40-SE
QGV-50-SE QG-50-SE QG-50G-SE QGV-50-SE
SYSTEM
MODEL
GENERATOR
CABINET
MODEL
X-RAY
CONTROL
MODEL
Revision W HF Series X-ray Generators - Service Manual
1-4 Quantum Medical Imaging
Chapter 1 Specifications

PHYSICAL SPECIFICATIONS

The following are physical specifications of the HF Series X-ray Generators:

GENERATOR CABINET SPECIFICATIONS

Width: 24.0 inches (609.6 mm) Depth: 20.0 inches (508.0 mm) Height: 34.7 inches (881.4 mm) Weights:
Single-Phase (non-SE units): 298 lbs. (135.2 kg) 32, 40 and 50 kW Stored
Energy (SE) units: 431 lbs. (195.5 kg) 50, 65, 80 kW Three-Phase
units): 380 lbs. (712.4 kg) w/o internal HSS
388 lbs. (176.0 kg) w/internal HSS

OPERATOR CONTROL PANEL SPECIFICATIONS

Note: X-Ray Control Panels below are utilized with HF gener­ator systems integrated with computed radiographic (CR) or digital radiographic (DR) workstation computers (e.g., Agfa NX, Canon CXDI, etc.).
The Q-VISION all-in-one operator control is not powered from the generator cab­inet; it requires a separate 100-240 VAC, 3A power input.
ODYSSEY HF Control Panel:
Width: 17.9 inches (454.7 mm) Depth: 12.25 inches (311.2 mm) Height (without pedestal): 3.3 inches (83.8 mm) Height (with pedestal): 36.5 inches (927.4 mm) Weight (without pedestal): 5 lbs. (2.3 kg) Weight (with pedestal): 25 lbs. (11.3 kg)
Q-VISION HF Control Panel:
Width: 18.7 inches (476.0 mm) Depth (without stand): 1.9 inches (49.0 mm) Depth (with stand): 7.125 inches (181.0 mm) Height: 14.4 inches (365.0 mm) Weight: 13 lbs. (5.9 kg)

SYSTEM SPECIFICATIONS

System Compatibility: Compatible with all Quantum Medical
Tube Compatibility: Toshiba Models E7239X, E7242X, E7252X,
Imaging tube stands, overhead tube cranes, radiographic tables and wall stands
E7254X, E7255X, E7813X, E7869X Varian Models RAD-8, RAD-13, RAD-14, RAD-21, RAD-44, RAD-56, RAD-60, RAD-68, RAD-74, RAD-92, A192, A272, A292
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-5
Chapter 1 Specifications
Collimators: Huestis Manual, Model 150MC
Note: The HF Series radiographic systems are compatible with any
automatic collimating system that provides an isolated nor­mally open switch that closes when the Automatic Collima­tion System presents an “EXPOSURE READY” mode to the control; they are also compatible with any beam limiting device that complies with 21 CFR Subchapter J intended for radiographic equipment.

POWER REQUIREMENTS

Dunlee Models PX 1302, PX 1429, PX 1436, PX 1312, DU 309, DA 1092, MX100 replacement tube for GE Radiographic Systems
(For tube types not listed, consult factory.) Huestis Automatic, Model 150PBL
Huestis Selectable, Model 150A, Progeny Linear MC150 (Manual) Progeny Linear II and Linear IV (Automatic)
In order to assure the proper operation of our equipment and subsequently, the produc­tion of high-qua lity radiographs, it is es se ntial that all pow er r e q u irements, as sp ec if ied in the tables that follow, are suitably met by the facility in which the equipment is installed. This includes usage of the proper size conductor between distribution transformer and dis­connect switch, connection to a supply transformer of sufficient capacity (specified in terms of KVA), insuring the AC input voltage tolerance meets specification (±5% or ±10% depending on model), and voltage regulation of 5% at full load is maintained. All wiring and grounding at the installation site must be in accordance with applicable National Elec­tric Code (NEC) and/or local electrical codes.
THe generator requires a 3-phase "Y" connection, measured line-to-line, balanced to earth or neutral. The "Y" connection can be 4-wire (L1, L2, L3, Neutral) and GND or 3-wire (L1, L2, L3) and GND. The 4-wire connection may be required by local electri­cal code and in connection where a roughly balanced 3-phase source exists (e.g., when using a delta-to-Y transformer).
Note that failure to comply with the above requirements not only may result in improper generator operation and/or performance, but it will also void the warranty, as the warranty is exclusive of the following conditions:
"Failure of customer to prepare the site or provide power requirements or
operating environmental conditions in compliance with any applicable instructions or recommendations of QMI.
"Failure of customer to provide the proper incoming power required to support
the equipment in accordance with the recommendation of QMI.
"Improper or extraordinary use of the Product, improper maintenance of the
Product, or failure to comply with any applicable instructions or recommendations of Quantum Medical Imaging, LLC.
Revision W HF Series X-ray Generators - Service Manual
1-6 Quantum Medical Imaging
Chapter 1 Specifications
NOTE: All single-phase x-ray generator input ratings assume the use of a
single-phase distribution transformer. When utilizing two legs of a 3­phase transformer, review the National Electrical Code and any appli­cable local electrical code(s) to ensure the correct rating distribution transformer (in terms of KVA) is used to assure line stability. This configuration would require an increase in the overall power (KVA) of the distribution transformer, depending on whether the input line transformer is Wye or Delta type. Under no circumstances should a Delta type transformer be used for input power to a 3-phase genera­tor.

SINGLE-PHASE X-RAY GENERATORS

Non-Stored Energy Single-phase x-ray genera tors require three-wire line inputs in either of the following two configurations:
one hot wire, one neutral wire, and one earth ground wire
two hot wires and one earth ground. Refer to Table 1-1 for power requirements of single-phase (non-stored energy)
HF Series Generators.
25 kW Single Phase (Model QG-25)
Line Voltage: 208-260 VAC (±5%), 100 Amps, 50/60 Hz Momentary Current: 200 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
25 kW Single Phase (Model QG-25-5)*
Line Voltage: 380 - 480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 110 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32)
Line Voltage: 208-260 VAC (±5%), 120 Amps, 50/60 Hz Momentary Current: 240 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32-5)*
Line Voltage: 380 - 480 VAC (±10%), 75 Amps, 50/60 Hz Momentary Current: 140 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-7
Chapter 1 Specifications
40 kW Single Phase (Model QG-40)
Line Voltage: 208-260 VAC (±5%), 150 Amps, 50/60 Hz Momentary Current: 300 Amps (@208VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Single Phase (Model QG-40-5)*
Line Voltage: 380-480 VAC (±10%), 100 Amps, 50/60 Hz Momentary Current: 170 Amps @ 415 VAC Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
40 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
50 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz Momentary Current: 13 Amps @ 115 VAC Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
210-250 VAC (±10%), 10 Amps, 50/60 Hz 7 Amps @ 230 VAC
Note: Models QG-25-5, QG-32-5, and QG-40-5 are not UL classified.
Revision W HF Series X-ray Generators - Service Manual
1-8 Quantum Medical Imaging
Chapter 1 Specifications

THREE-PHASE X-RAY GENERATORS

Three-phase x-ray generators require three "hot" wires and one earth ground wire. Refer to Tables 1-2 and 1-3 for power requirements of three-phase HF Series Generators.
32 kW Three Phase (Model QG-32-2)
Line Voltage: 208-240 VAC (±5%), 70 Amps, 50/60 Hz Momentary Current: 140 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
32 kW Three Phase (Model QG-32-3)
Line Voltage: 380-480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 117 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Three Phase (Model QG-40-2)
Line Voltage: 208-240 VAC (±5%), 85 Amps, 50/60 Hz Momentary Current: 170 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
40 kW Three Phase (Model QG-40-3)
Line Voltage: 380-480 VAC (±10%), 60 Amps, 50/60 Hz Momentary Current: 125 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
50 kW Three Phase (Model QG-50-2)
Line Voltage: 208-240 VAC (±5%), 110 Amps, 50/60 Hz Momentary Current: 220 Amps (@220 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
50 kW Three Phase (Model QG-50-3)
Line Voltage: 380-480 VAC (±10%), 80 Amps, 50/60 Hz Momentary Current: 144 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-9
Chapter 1 Specifications
65 kW Three Phase (Model QG-65)
Line Voltage: 380-480 VAC (±10%), 90 Amps, 50/60 Hz Momentary Current: 190 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation
80 kW Three Phase (Model QG-80)
Line Voltage: 380-480 VAC (±10%), 100 Amps, 50/60 Hz Momentary Current: 200 Amps (@440 VAC) Long Term Current: less than 1.0 Amp Mode of Operation: Intermittent operation

HIGH-SPEED STARTER POWER REQUIREMENTS

A High-Speed Starter (HSS) unit is optional on Models QG-40 and QG-50, and is standard equipment on Models QG-65 and QG-80.
The external HSS (QG-HSS) configuration requires a separate 200 - 240 VAC, 20A line input. Note that a separate 200- 240 VAC line input is erators equipped with Internal High Speed Starter (Q-HSS) configuration.
not required
on gen-
Revision W HF Series X-ray Generators - Service Manual
1-10 Quantum Medical Imaging
Chapter 1 Specifications
QG-25
QG-32
QG-40
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAG E (VAC)
DISCONNECT
TO GENE R ATO R
15 FT. (5 M) M A X .
*SERVICE
RATING
*RECOMMENDED
DISTR IBU T IO N
TRANSFORMER
CAPACITY
MAX. INPUT LINE
RESISTANCE
(AT FULL-LOAD
RMS)
NOTE: All installations must meet the requirements of local electrical codes. The above values are recommended as typical (unless otherwise specified).
*
1.
±5%)
Line Regulation: 5% at full load.
7. Ground wire size is 4 AWG.
8. Specifications are subject to change.
9. Specifications for single-phase “stored en ergy” Models QG-20-SE, QG-32-SE, QG-40-SE and QG-50-SE are as follows: 105-130 VAC (±10%), 10 Amps, 50/60 Hz 210-250 VAC (±10%), 10 Amps, 50/60 Hz
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. Input Voltage: 208-257 ( VAC 50/60 Hz (configured at time of installation).
6.
QG-250
QG-320, QG-3200, QGV-32
QG-400, QG-4000, QGV-40
GENERATOR
CATALOG NO.
208
25
32
40
240
208
240
208
240
#4 AWG
#4 AWG
#2 AWG
#2 AWG
#1 AWG
#1 AWG
100 AMPS
80 AMPS
120 AMPS
100 AMPS
150 AMPS
120 AMPS
30 KVA
30 KVA
40 KVA
40 KVA
50 KVA
50 KVA
0.06 OHMS
0.08 OHMS
0.04 OHMS
0.05 OHMS
0.035 OHMS
0.04 OHMS
Table 1-1. HF Series Generators -
Electrical Wiring Requirement s (208 - 260 VAC Single-Phase Input Units)
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-11
Chapter 1 Specifications
*NOTE: All installations must meet the requirements of local electrical codes.
The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 380-480 (±10%) VAC 50/60 Hz (configured at time of installation)
. Ground wire size is 4 AWG.
7. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAGE (VAC)
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*SERVICE
RATING
*RECOMMENDED
DISTRIBUTION
TRANSFORMER
CAP ACITY
MAX. INPUT LINE
RESISTANCE
(A T F U L L-LOAD
RMS)
GENERATOR
CATALOG NO .
25
32
40
QG-25-5
QG-32-5
QG-40-5
QG-250-5
QG-320-5, QG-3200-5, QGV-32-5
QG-400-5, QG-4000-5, QGV-40-5
380
#6 AWG 60 AMPS
32 KVA
0.028 OHMS
415 #6 AWG 55 AMPS 32 KVA 0.028 OHMS
480 #6 AWG 40 AMPS 32 KVA 0.032 OHMS
380 #6 AWG 75 AMPS 40 KVA 0.024 OHMS
415 #6 AWG 70 AMPS 40 KVA 0.024 OHMS
480
#6 AWG 60 AMPS
40 KVA
0.028 OHMS
380 #4 A WG 100 AMPS 50 KVA 0.015 OHMS
415
#4 AWG 90 AMPS
50 KVA
0.020 OHMS
480
#6 AWG 70 AMPS
50 KVA
0.024 OHMS
Table 1-2. HF Series Generators -
Electrical Wiring Requirements (380 - 480 VAC Single-Phase Input Units)
Revision W HF Series X-ray Generators - Service Manual
1-12 Quantum Medical Imaging
Chapter 1 Specifications
QG-32-3,
QG-40-3
QG-50
QG-65
QG-80
32
(QG-32
-3)
40
(QG-40-3)
50
65
80
G E
N E R A
T O
R
M O
D E L
N O .
MAX. POWER
O UTPUT (kW)
N O M
I
N A L
L I
N E
V O L
T A G
E (
V A C )
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*
S E R V I
C E
R A T I
N G
*
R E C O M
M E
N D E
D
D I
S T R I
B U T I
O N
T R A N
S F O
R M
E R
C A P
A C I
T Y
MAX. INPUT LINE
RESISTANCE
(AT FULL-LO AD RMS)
380#6 AWG60 AMP
S
50 KV
A
0.23 OHMS380#4 AWG80 AMPS65 KVA0.17 OHMS380#4 AWG90 AMPS80 KVA0.12 OHMS380#4 AWG100 AMPS100 KVA0.10 OHMS440#6 AWG60 AMPS50 KVA0.33 OHMS440#6 AWG80 AMPS65 KVA0.20 OHMS440#4 AWG90 AMPS80 KVA0.17 OHMS440#4 AW
G
S
100 KVA
0.14 OHMS
480#6 AWG50 AMPS50 KVA0.4 OHMS480#6 AWG75 AMPS65 KVA0.24 OHMS480#4 AWG90 AMPS80 KVA0.19 OHMS480
#4 AWG100 AMPS100 KVA0.16 OHM
S
QG-320, QG-3200, QGV
-32-3, QG-400, QG-4000, QG
V-40-3
QG-500, Q
G-5000,
Q
GV-50
QG-6500, QGV-65
QG-8000, QGV-80
G E
N E
R A T O R
C A T
A L
O G
N O .
*NOTE: All installations must meet the requirements of local electrical codes.
The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 380-480 (±10%) VAC 50/60 Hz (configured at time of installation)
. Voltage is three-phase Y-configuration, measured line-to-line, balanced to earth
ground. (Note: Additional Neutral wire may be used if required by local electrical code.)
7. Ground wire size is 4 AWG.
8. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
Table 1-3. HF Series Generators - Electrical Wiring Requirements
(380-480 VAC Three-Phase Input Units)
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-13
Chapter 1 Specifications
QG-32-2
QG-40-2
QG-50-2
208
32
40
50
240
208
240
208
240
#6 AWG
#6 AWG
#4 AWG
#6 AWG
#4 AWG
#4 AWG
70 AMPS
60 AMPS
90 AMPS
70 AMPS
110 AMPS
90 AMPS
40 KVA
40 KVA
50 KVA
50 KVA
65 KVA
65 KVA
0.026 OHMS
0.023 OHMS
0.017 OHMS
0.024 OHMS
0.014 OHMS
0.016 OHMS
GENERATOR
MODEL NO.
MAX. POWER
OUTPUT (kW)
NOMINAL LINE
VOLTAGE (VAC)
DISCONNECT
TO GENERATOR
15 FT. (5 M) MAX.
*SERVICE
RATING
*RECOMMENDED
DISTRIBUTION
TRANSFORMER
CAPACITY
MAX. INPUT LINE
RESISTANCE
(AT FULL-LOAD
RMS)
QG-320, QG-3200, QGV-32-2
QG-400, QG-4000, QGV-40-2
QG-500, QG-5000, QGV-50-2
GENERATOR
CATALOG NO.
*NOTE: All installations must meet the requirements of local electrical codes. The above values are recommended as typical (unless otherwise specified).
*
1.
Input Voltage: 208-257 (±5%) VAC 50/60 Hz (configured at time of installation)
. Voltage is three-phase Y-configuration, measured line-to-line, balanced to earth
ground. (Note: Additional Neutral wire may be used if required by local electrical code.)
Line Regulation: 5% at full load.
8. Ground wire size is 4 AWG.
9. Specifications are subject to change.
All wire must be copper.
2. Actual wire size depends on power quality and distance between mains power and the generator.
3. All wiring and grounding must be in accordance with National Electric Code (NFPA #70) or local electrical codes.
4. The disconnect switch must be located within reach of the operator.
5. . 6
7.
Table 1-4. HF Series Generators -
Electrical Wiring Requirements (208-240 VAC Three-Phase Input Units)
Revision W HF Series X-ray Generators - Service Manual
1-14 Quantum Medical Imaging
Chapter 1 Specifications

GENERATOR AUXILIARY POWER SUPPLIES

The HF Series X-ray Generator supplies the following supply voltages for powering X-ray room equipment:
24 VDC, 1.5A
24 VAC, 150W (collimator lamp not to exceed 150W)
115 VAC, 3A
NOTE: The above voltage outputs are not to be used as power

PERFORMANCE SPECIFICATIONS

Table 1-5 provides performance specifications of single-phase (non-stored energy) HF Series Generators.
Table 1-5. HF Series Single-Phase Generator Performance Specifications
sources for collimator lamps or for inductive loads (e.g., motorized tables, solenoids, etc.)
Cat. No.
N/A
QG-3200,
QGV-32
Models
N/A
QG-32,
QG-32-5
Maximum kW N/A 32 40 mA Stations;
Small Focus (S) Large Focus (L)
kVp Range (kVp) N/A 40-125 40-125 kVp increments (kVp/step) N/A 1.0 1.0 150 kVp option N/A Yes Yes Time Range (sec.)* N/A 0.001 - 6.3 0.001 - 6.3 Minimum Exposure Time
(sec.) mAs Range† High-Speed Starter
Ripple Voltage (output) N/A 5% 5%
* Time measured at 75% of the peak kVp waveform
NOTE: AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS.
mAs is tube dependent
N/A
N/A 0.001 0.001 N/A 0.025-600 0.025-600
N/A No No
25S, 75S, 150S
100L, 200L, 250L,
320L, 400L, 500L
Cat. No.
QG-4000,
QGV-40
Models
QG-40,
QG-40-5
25S, 75S, 150S
100L, 200L, 250L, 320L,
400L, 500L
HF Series X-ray Generators - Service Manual Revision W Quantum Medical Imaging 1-15
Chapter 1 Specifications
Refer to Table 1-6 for performance specifications of single-phase stored energy (SE) HF Series Generators.
Table 1-6. HF Series Single-Phase STORED ENERGY Generator Performance Specifications
Cat. No.
N/A
QG-3200-SE,
QGV-32-SE
Model QG-32-SE
N/A
(STORED ENERGY)
Maximum kW N/A 32 40 50 mA Stations;
Small Focus (S) Large Focus (L) N/A
kVp Range (kVp) N/A 40-125 40-125 40-125 kVp increments
(kVp/step) 150 kVp optional? N/A No No No Time Range
(sec.)* Minimum Expo-
sure Time (sec.)* mAs Range High-Speed Starter
Ripple Voltage (output)
N/A 1.0 1.0 1.0
N/A 0.001 - 6.3 0.001 - 6.3 0.001 - 6.3
N/A 0.001 0.001 0.001 N/A 0.025-400 0.025-500 0.025-500
N/A No No No N/A 5% 5% 5%
25S, 50S, 75S, 150S
100L, 160L, 200L,
250L, 320L, 400L
Cat. No.
QG-4000-SE,
QGV-40-SE
Model QG-40-SE
(STORED ENERGY)
25S, 50S, 75S, 150S
100L, 200L, 250L,
320L, 400L, 500L
Cat. No.
QG-5000-SE,
QGV-50-SE
Model QG-50-SE
(STORED ENERGY)
25S, 50S, 75S,
150S 100L, 200L, 250L, 320L, 400L, 500L,
600L
* Maximum exposure time is a function of kV and mAs settings and the age of the
batteries; Time measured at 75% of the peak kVp waveform NOTE: AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS.
mAs is tube dependent; the generator may not reach maximum mAs due t o tube type. In AEC mode, the mAs is limited to 600.
Revision W HF Series X-ray Generators - Service Manual
1-16 Quantum Medical Imaging
Loading...
+ 463 hidden pages