1. Attach the coolant lines to the back of the
laser head and back of the power supply:
blue to blue and red to red.
2. Attach the double-connector side of the
electrical cable to the laser head. These
connectors have red dots indicating proper
orientation. Note
: It is easier to attach the
bottom connector first. Ensure that the
connectors are fully seated. Note
: Yo u
should hear a “click”.
3. Attach the other end of the D-Connector to the back of the power supply
and tighten the thumb screws completely to avoid connector contact
damage. Note
: Never operate the laser system without the connectors of
the electrical cable fully inserted and thumbscrews tightened.
∗ If you received a DI Cartridge that is not attached to the coolant lines,
instructions to attach the DI Cartridge are on the opposite side of this
Quick Start Guide.
4. Attach the remote box to the remote box interface on the front of
the power supply. Thread the silver collar on the connector until it
is flush with the front of the power supply.
5. Connect the power cord to the back of the power supply.
6. Remove the white plastic shipping cover from the front of the laser head.
Do not discard.
uick Start Guide For Your ULTRA Laser System
WARNING: THIS IS A CLASS IV LASER. READ AND FOLLOW YOUR USER’S MANUAL BEFORE OPERATING TO AVOID DAMAGE OR PERSONAL INJURY.
ULTRA Laser Head / Standard ICE
450
Power Supply
1
UNPACK your system. Save your packing material.
Locate the following items:
Laser Head
ICE
450
Power Supply
Fill Bottle/Vent Tube
Coolant Lines* and Electrical Cable
Remote Box
Accessory Kit (including keys)
Power Cord
Coolant (Distilled Water)
WARNING: OPERATING YOUR LASER WITHOUT COOLANT WILL CAUSE CATASTROPHIC FAILURE.
2
CABLE CONNECTION
Electrical Cables
Coolant Lines
Vent Tube Connector
(Connect to the ICE450 Vent Port)
Fill Bottle Connector
(Connect the vent tube here)
3
1. Connect the vent tube to the fitting on the front of the reservoir
bottle.
2. Connect the vent tube to the reservoir vent port on the front of the
power supply.
3. Connect the fill bottle connector to the reservoir fill port on the front
of the power supply.
4. Add coolant to the fill bottle until water rises into the vent tube.
5. Insert the key into the power key switch and
turn it to the ON (“I”) position.
6. Continue to add coolant until the coolant lines
are full and the coolant level is visible in the
vent tube (above the minimum coolant level
indicator). Note:
The pink illuminated section
indicates the actual coolant level.
7. Disconnect the fill and vent tubes.
8. Secure laser head to the table using three 6-32
UNC screws.
FILL
Reservoir Vent Port
Reservoir Fill Port
Minimum Coolant
Level Indicator
Fill Bottle Hose
(OVER)
Fill Bottle Connector
(Connect to the ICE450 Fill Port)
Power Key Switch
Remote Box Interface
Vent Tube Connector
(Connect to the Fill Bottle Connector)
DOC00071 Rev C

Quick Start Guide For Your ULTRA Laser System
ULTRA Laser Head / Standard ICE
450
Power Supply
4
CONGRATULATIONS!
Your system is ready for full operation.
CONSULT CHAPTER 2 OF THE USER’S MANUAL
BEFORE CONTINUING FOR APPROPRIATE EYE
PROTECTION AND SAFETY PRECAUTIONS.
1. Ensure that the red emergency stop button on top of
the remote box is pulled out.
2. See System Summary to verify the desired
settings on the remote box. See the User’s
Manual for more details of operation.
3. Press the “Flashlamp Start Button” on the
remote box.
4. Open the manual shutter located on the side of
the laser head.
5. Lasing will begin when the “Q-Switch Start
Button” is pressed on the remote
box.
Note:
For systems with harmonics, it may take several
minutes for desired energy output to stabilize.
DOC00071 Rev C
START UP
Shutter OPEN Shutter CLOSED
FOR TECHNICAL SUPPORT
PLEASE CALL
CUSTOMER SERVICE:
QUANTEL USA: 1-800-914-8216
QUANTEL: (33)1-6929-1700
Emergency Stop Button
Your laser is pre-configured to operate at
full energy and Pulse Repetition
Frequency (PRF).
The only routine maintenance required on
the ULTRA system is deionization filter
replacement and flashlamp replacement.
Cycle the coolant at least 30 minutes
every month and replace the flashlamp as
required. See the User's Manual for
details.
If you need to ship the laser for any
reason, see the User's Manual for details
on draining the system for transport to
prevent freezing.
Flashlamp Start Button
Q-Switch Start Button
*DI Cartridge (not attached to the coolant lines):
If your DI Cartridge is not attached to the coolant lines,
please follow the steps below:
1. Remove the gray shipping fittings on the DI cartridge
by using two fingers to hold down the base of the fitting
while pulling straight out on the top of the fitting (See
Item 1). Repeat on the opposite side.
2. Remove the two gray caps from the fittings on the
coolant line (Item 2).
3. Attach the two sides of the DI cartridge to the coolant
lines at the fittings on both ends (Items 3- 4). Make
sure that the “Flow” arrow on the DI Cartridge is
pointing in the correct direction (away from the power
supply). Return to the “Cable Connection” section of
this Quick Start Guide.
1
2
3
4