Qiagen Hybrid Capture Rotary Shaker 1 User Manual

Sample to Insight__
September 2018
Hybrid Capture® System Rotary Shaker 1 User Manual
-2240E (230 V)
2
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
Trademarks: QIAGEN®, Sample to Insight®, digene®, HC2®, Hybrid Capture® (QIAGEN Group). Registered names, trademarks, etc., used in this document, even when not specifically marked as such, are not to be considered unprotected by law. © 2018 QIAGEN, all rights reserved.
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
3
Contents
1 Introduction ..................................................................................................... 5
1.1 General information ................................................................................ 5
1.1.1 Technical assistance ........................................................................ 5
1.1.2 Version management ....................................................................... 5
1.2 Intended use .......................................................................................... 5
2 Safety Information ............................................................................................ 6
2.1 Proper use ............................................................................................. 7
2.2 Electrical safety ...................................................................................... 9
2.3 Biological safety ..................................................................................... 9
2.4 Waste disposal .................................................................................... 11
2.5 Symbols ............................................................................................... 12
3 Installation ..................................................................................................... 14
3.1 Unpacking ........................................................................................... 14
3.2 Start up ............................................................................................... 14
4 Functional Description ..................................................................................... 16
4.1 Operational modes ............................................................................... 17
5 General Operation ......................................................................................... 19
5.1 Loading Microplates ............................................................................. 19
5.2 Continuous operation ............................................................................ 20
5.3 Timer-terminated operation .................................................................... 20
5.4 Timer function – Accumulated time .......................................................... 21
5.5 Timer function – Remaining time ............................................................. 21
4
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
5.6 Beeper Preference ................................................................................. 22
6 Maintenance .................................................................................................. 23
6.1 Cleaning and decontamination .............................................................. 23
6.2 Regular maintenance ............................................................................. 24
6.3 Detaching and replacing the shaker platform ........................................... 25
6.4 Replacing the fuse ................................................................................. 26
6.5 Shaking speed verification ..................................................................... 27
6.6 Servicing ............................................................................................. 28
7 Troubleshooting .............................................................................................. 29
8 Technical Data ............................................................................................... 33
8.1 Operating conditions ............................................................................ 33
8.2 Transport conditions .............................................................................. 35
8.3 Storage conditions ................................................................................ 35
Appendix A – Waste Electrical and Electronic Equipment (WEEE) .................................. 36
Appendix B – Warranty ............................................................................................ 37
Appendix C — FCC Declaration ................................................................................ 38
Ordering Information ................................................................................................ 40
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
5
1
Introduction
The Hybrid Capture System (HCS) Rotary Shaker 1 is an instrument designed to shake microplates and consists of a shaker base and a shaker platform.
Read this user manual prior to operating the HCS Rotary Shaker 1.
1.1 General information
1.1.1 Technical assistance
For technical assistance and more information, please see our Technical Support Center at www.qiagen.com/TechSupportCenter or contact QIAGEN Technical Services or a local distributor.
1.1.2 Version management
This document is Hybrid Capture System Rotary Shaker 1 User Manual; see the front cover of this user manual for document number and revision.
1.2 Intended use
The HCS Rotary Shaker 1 is intended for use in conjunction with digene Hybrid Capture 2 (HC2®) DNA tests. The HCS Rotary Shaker 1 is designed for shaking microplates.
6
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
2 Safety Information
This manual contains information about warnings and cautions that must be followed by the user to ensure safe operation of the HCS Rotary Shaker 1 and to maintain the instrument in a safe condition.
WARNING The term WARNING is used to inform you about situations that could result
in personal injury to you or other persons. Details about these circumstances are provided to avoid personal injury to you or other persons.
CAUTION
The term CAUTION is used to inform you about situations that could result in damage to the instrument or other equipment. Details about these circumstances are provided to avoid damage to the instrument or other equipment.
Before using the instrument, it is essential to read this manual carefully and to pay particular attention to any details it contains concerning hazards that may arise from the use of the instrument.
The details given in this manual are intended to supplement, not supersede, the normal safety requirements prevailing in the user’s country.
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
7
2.1 Proper use
WARNING/
CAUTION
WARNING/
CAUTION
WARNING/
CAUTION
Risk of personal injury and material damage
Improper use of the HCS Rotary Shaker 1 may cause personal injuries to the user or damage to the instrument. The HCS Rotary Shaker 1 must only be operated by qualified personnel who have been appropriately trained.
Risk of personal injury and lnstrument damage
Prior to operation, secure the HCS Rotary Shaker 1 to a level, smooth, stable surface by firmly pressing down on the four (4) corners of the unit, creating a strong suction to the work surface (DO NOT place on a bench mat). Failure to do so will cause excess vibration and could cause personal injury, Instrument damage, and/or property damage.
Risk of lnstrument damage Spills should be removed promptly. DO NOT use a cleaning agent or solvent on the front panel which is abrasive or harmful to plastics, nor one which is flammable. Always ensure the power is disconnected from the unit prior to any cleaning.
WARNING Risk of personal injury
Always wear shatterproof eye protection.
8
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
WARNING Risk of material damage
WARNING/
CAUTION
WARNING/
CAUTION
Only a QIAGEN Field Service Specialist can service or repair the instrument. The only exceptions are the maintenance activities listed
in the
“Maintenance” section, page 23, of this user manual.
Risk of personal injury and material damage
Do not immerse the HCS Rotary Shaker 1 in water or pour liquids over the instrument as electrical shock may occur.
Risk of improper operation Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these can interfere with the proper operation.
Take the following precautions when operating or working near the HCS Rotary Shaker 1:
Always make certain that the shaker platform and any hardware are secure before
operating the HCS Rotary Shaker 1.
Load the HCS Rotary Shaker 1 symmetrically. Avoid unbalanced loads. When shaking a
single microplate, a second empty microplate must be placed diagonally to balance the load. Likewise, when shaking 3 microplates, an empty microplate must be placed in the fourth position to balance the load.
Do not use solvents and flammables on or near the HCS Rotary Shaker 1.
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
9
Use the instrument in a dry, clean environment.
Wipe down the base and platform of the HCS Rotary Shaker 1 after each use with a
soft, dry cloth.
Clean up any spills immediately.Do not allow dust to accumulate on the unit.If needed, the shaker platform may be removed and cleaned with a cloth dampened in a
dilute detergent solution. See “Maintenance,” page 23, for additional instructions.
Avoid cold starts: Unit is not designed to start after being in a cold room environment.
Bring unit into cold room from a room temperature environment, operate and remove unit from cold room as soon as operation is complete.
2.2 Electrical safety
Only operate the HCS Rotary Shaker 1 with the power cord provided with the instrument. For satisfactory and safe operation of the HCS Rotary Shaker 1, it is essential that the line power cord is connected to true electrical earth (ground).
2.3 Biological safety
WARNING Hazardous substances
T
he products used with this instrument may contain hazardous substances. When working with chemicals, always wear a suitable lab coat, disposable gloves and protective goggles. For more information, please consult the appropriate safety data sheets (SDSs). These are available online in PDF format at www.qiagen.com/safety where you can find, view and print the SDS for each QIAGEN kit and kit component. For further information see the instructions for use that comes with the kit.
10
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
WARNING Risk of exposure to hazardous material
Shake hazardous samples only in appropriate containment vessels.
WARNING/
CAUTION
Risk of personal injury and material damage
Consider any laboratory equipment used for research or clinical analysis a potential biohazard that requires decontamination before reuse.
WARNING Risk of personal injury
Sodium hypochlorite solution is caustic; wear rubber gloves and eye
protection when handling it.
To dispose of the HCS Rotary Shaker 1, follow all national, state and local health and safety regulation and laws for disposing of laboratory waste. For disposal of Waste Electrical and Electronic Equipment (WEEE compliance), see “Appendix A – Waste Electrical and Electronic Equipment (WEEE),” page 36.
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
11
2.4 Waste disposal
Waste may contain certain hazardous chemicals or contagious/biohazardous materials and must be collected and disposed of properly in accordance with all national, state and local health and safety regulations and laws.
12
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
2.5 Symbols
The following symbols may be found on the instrument, in this user manual or on labels associated with the instrument.
Symbol
Location
Description
On the instrument General warning sign
Type plate on the instrument
CE mark for Europe
Type plate on the i
nstrument
In vitro diagnostic medical device
Type plate on the i
nstrument
RoHS mark for China (the restriction of the use of certain hazardous substances in electrical and electronic equipment)
Type plate on the i
nstrument
Waste Electrical and Electronic Equipment (WEEE)
SN:
Type plate on the instrument
Serial number
Type plate on the i
nstrument
Manufacturer
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
13
Symbol
Location
Description
Type plate on the Instrument
RCM mark for Australia
Type plate on the Instrument
The instrument complies with applicable standards for electrical safety of laboratory equipment
UDI Label o
n the
instrument
Global trade item number
Instrument box label F
ragile, handle with care
Instrument box label
Consult instructions for use
Front cover of this user manual
Authorized representative in the European Community
Front cover of this user ma
nual
Catalog number
14
Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
3 Installation
3.1 Unpacking
Before using the HCS Rotary Shaker 1 for the first time, examine the exterior carton and the equipment itself for damage. In the event of shipping damage, contact your local QIAGEN representative or QIAGEN Technical Services.
Carefully unpack the instrument and verify the contents of the package, which should contain the following components of the instrument:
HCS Rotary Shaker 11 power cord
If any of these items are missing, contact your local QIAGEN representative or QIAGEN Technical Services. Save the original packaging until the instrument is operated successfully.
3.2 Start up
Make sure to select the correct voltage rating by checking the type plate on the side of the instrument. Record the serial number, located on the type plate, for future reference. Turn the unit upside down, gently rest it on the shaking platform, and inspect the suction cup feet. Remove any dust or debris from the suction cup feet with 70% Isopropyl alcohol and a lint-free cloth. Likewise, prepare the surface where the HCS Rotary Shaker 1 will be placed by wiping it with 70% Isopropyl alcohol and a lint-free cloth.
Place the HCS Rotary Shaker 1 on a level, smooth, stable surface near a grounded electrical outlet. Allow at least 8 cm (3 inches) clearance on all sides of the instrument for proper ventilation. Make sure the shaker platform will not touch other objects while in operation.