pyroscience Piccolo2 User Manual

1
Piccolo2
FIBER-OPTIC OXYGEN METER
USER MANUAL
2
Document Version 2.04 Refers to Pyro Oxygen Logger Software version >3.2
The Piccolo2 is manufactured by
Pyro Science GmbH
Hubertusstr. 35 52064 Aachen Germany
Phone +49 (0)241 5183 2210 Fax +49 (0)241 5183 2299 Email info@pyro-science.com Internet www.pyro-science.com
Registered: Aachen HRB 17329, Germany
3
TABLE OF CONTENT
1 OVERVIEW .............................................................................. 5
2 SAFETY GUIDELINES .............................................................. 6
3 INTRODUCTION TO THE PICCOLO2 ......................................... 8
4 SOFTWARE INSTALLATION .................................................... 9
5 OXYGEN SENSOR TYPES....................................................... 10
5.1 ROBUST PROBE ....................................................................... 10
5.2 DIPPING PROBE ....................................................................... 11
5.3 SENSOR SPOTS ........................................................................ 11
5.4 FLOW-THROUGH CELLS ............................................................ 13
5.5 RESPIRATION VIALS .................................................................. 14
5.6 NANOPROBES ......................................................................... 16
5.7 CONNECTING THE SENSORS AND OPTICAL FIBERS ......................... 18
5.8 CLEANING AND MAINTENANCE OF THE SENSORS........................... 20
6 THE SOFTWARE "PYRO OXYGEN LOGGER" ........................... 22
6.1 MAIN WINDOW ........................................................................ 22
6.2 SETTINGS................................................................................ 30
6.2.1 Basic Settings .................................................................... 31
6.2.2 Advanced Settings ............................................................. 32
6.2.3 Conditions in the Sample ................................................... 33
6.2.4 Options ............................................................................. 34
6.3 RAW DATA WINDOW ................................................................ 36
7 CALIBRATION OF OXYGEN SENSORS .................................... 38
7.1 CALIBRATION .......................................................................... 38
7.1.1 Calibration Mode: Factory .................................................. 39
7.1.2 Calibration Mode: 1-Point .................................................. 40
7.1.3 Calibration Mode: 2-Point .................................................. 42
7.1.4 Calibration Mode: Custom .................................................. 44
7.2 ADVANCED ADJUSTMENTS ......................................................... 45
8 CALIBRATION STANDARDS ................................................... 47
8.1 THE AIR CALIBRATION STANDARD .............................................. 47
4
8.1.1 Ambient Air ....................................................................... 49
8.1.2 Water-Vapor Saturated Air ................................................ 49
8.1.3 Air Saturated Water ........................................................... 50
8.2 THE 0% STANDARD .................................................................. 51
8.2.1 Water Mixed with a Strong Reductant ................................ 51
8.2.2 Water Flushed with Nitrogen Gas ....................................... 51
8.2.3 Nitrogen Gas ..................................................................... 52
9 CALIBRATION OF CONTACTLESS SENSORS .......................... 53
9.1 CALIBRATION PROCEDURE ......................................................... 53
9.2 MANUAL BACKGROUND COMPENSATION .................................... 54
10 APPENDIX ............................................................................. 56
10.1 SPECIFICATIONS OF THE PICCOLO2 .............................................. 56
10.2 TROUBLESHOOTING ................................................................. 57
10.3 MEASURING PRINCIPLE ............................................................. 58
10.4 OPERATING SEVERAL PICCOLO2 IN PARALLEL ................................ 60
10.5 DEFINITION OF OXYGEN UNITS ................................................... 61
10.6 TABLE OF OXYGEN SOLUBILITY .................................................. 63
10.7 EXPLANATION OF THE SENSOR CODE .......................................... 65
5
1 Overview
The ultra-compact Piccolo2 is a fiber-optic oxygen meter integrated in a small USB stick housing. Despite its small size, it features the proven REDFLASH technology from Pyro Science (see Appendix 10.3 for more details). The Piccolo2 can be used in combination with a variety of optical oxygen sensors, like robust probes, dipping probes or contactless sensors (sensor spots, respiration vials, flow-through cells, nanoprobes). The optimized optics of the Piccolo2 enable contactless oxygen measurements up to a window thickness of 20 mm. The sensors can be used in water samples (dissolved oxygen, DO), as well as in the gas phase (gaseous oxygen, O2). Simply plug the Piccolo2 into a USB port of your Windows PC or Windows tablet, connect the oxygen sensor of your choice and start measuring with the comfortable logging software "Pyro Oxygen Logger".
More information concerning our products can be found at
www.pyro-science.com
or contact us under info@pyro-science.com
Your Pyro Science Team
6
2 Safety Guidelines
The Piccolo2 is a laboratory instrument to be used at constant experimental temperatures with optical oxygen sensors (optodes) from Pyro Science. In order to guarantee an optimal performance of the Piccolo2, please follow these operation instructions and safety guidelines.
Please note that opening the housing will void the warranty. There are no serviceable parts inside the device.
The Piccolo2 and the sensors should be used in the laboratory by qualified personnel only, following the operation instructions and safety guidelines of this manual. They should be kept and stored out of reach of children in a secure place under dry and clean conditions at room temperature, avoiding moisture, dust, corrosive conditions and heating of the instrument.
The Piccolo2 and the sensors are not intended for medical, military or other safety-relevant areas. They must not be used for applications in humans; not for in vivo examination on humans, not for human-diagnostic or therapeutic purposes. The sensors must not be brought in direct contact with foods intended for consumption by humans.
Please follow the appropriate laws and guidelines for safety like EEC directives for protective labor legislation, national protective labor legislation, safety regulations for accident prevention and safety data-sheets from manufacturers of chemicals used during the measurements.
When used in the field, extreme environmental conditions (like high humidity, dust, and exposure to (salt) water or intense solar radiation) can cause damage to the Piccolo2.
7
Before using the Piccolo2 and the sensors, carefully read the instructions and user manuals.
In case of problems or damage, disconnect the instrument and mark it to prevent any further use! Consult Pyro Science for advice! There are no serviceable parts inside the device. Please note that opening the housing will void the warranty!
The Piccolo2 is not watertight, is sensitive to corrosive conditions and to changes in temperature causing condensation. Avoid temperatures above 50°C (122°F) or below 0°C (32°F). Avoid any elevated humidity causing condensing conditions.
Handle the sensors with care, especially after removal of the protective cap! Prevent mechanical stress to the fragile sensing tip! Avoid strong bending of the fiber cable!
Calibration and application of the sensors are on the user’s authority, as well as data acquisition, treatment and publication!
The sensors and the oxygen meter Piccolo2 are not intended for medical, diagnostic, therapeutic, or military purposes or any other safety-critical applications. The sensors must not be used for applications in humans and must not be brought in direct contact with foods intended for consumption by humans.
The sensors should be used in the laboratory by qualified personnel only, following the user instructions and the safety guidelines of the manual, as well as the appropriate laws and guidelines for safety in the laboratory!
Keep the sensors and the oxygen meter Piccolo2 out of reach of children!
8
3 Introduction to the Piccolo2
The Piccolo2 is a miniature USB-driven fiber-optic oxygen meter for usage in the laboratory. It is compatible to a variety of oxygen sensors from Pyro Science, like robust probes, dipping probes and contactless sensors, including sensor spots, respiration vials, flow­through cells and nanoprobes.
These sensors are available in versions for the full range (0­50% O2, max. range 0-100% O2), and selected sensor types are additionally available for the trace range (0-10% O2). The Piccolo2 utilizes a measuring principle based on red light excitation and lifetime detection in the near infrared using the proven REDFLASH indicators (REDFLASH technology, see Appendix 10.3 for more details).
The Piccolo2 can be connected directly to a Windows PC with the USB plug, providing energy supply and data exchange with the PC. If a greater distance between the PC and the Piccolo2 is needed, the delivered USB extension cable can be used.
Comfortable calibration and logging functions are provided by the logging software Pyro Oxygen Logger, available as free download.
IMPORTANT: The Piccolo2 is intended for oxygen measurements at constant and defined temperatures! Do not submerse the housing of the Piccolo2 into liquid samples! This destroys the device. Protect the Piccolo2 against water spray!
9
4 Software Installation
IMPORTANT: Do not connect the Piccolo2 to your PC before the
Pyro Oxygen Logger software has been installed. The software will
install the appropriate USB-drivers automatically.
System requirements:
PC or tablet with Windows XP / Vista / 7 / 8 / 10 (but not
Windows RT) and min. 200 MB free disk space
Installation steps:
download the installer package for the newest version of the Pyro
Oxygen Logger software from the Pyro Science homepage:
www.pyro-science.com/downloads.html
unzip, then start the installer and follow the instructions
connect the Piccolo2 to a free USB port of the computer
After successful installation, a new program group "Pyro Oxygen Logger" is added to the start menu, and a shortcut named "Oxygen Logger" can be found on the desktop.
10
5 Oxygen Sensor Types
For the Piccolo2 meter, Pyro Science offers a range of optical oxygen probes and contactless oxygen sensors, comprising sensor spots, flow-through cells, respiration vials and nanoprobes. For an overview of all available oxygen sensor types, please visit the Pyro
Science website.
5.1 Robust Probe
Characteristics: robust probe for Piccolo2 (item no. OPROB3) with
stainless steel tubing 3 mm in diameter and
30 mm in length
oxygen sensor with optical isolation located
in the 3 mm disc at the tip of the tubing (sensor end, see chapter 5.7)
fiber cable 3 mm in diameter and 1 m in
length
Note: The fiber cable needs to be handled with care, as strong bending can lead to breakage and damage of the robust probe.
Applications:
standard laboratory applications in gases and liquids for aqueous liquids: permanent stirring is recommended, as
diffusion limitation of oxygen towards the 3 mm tip has a measurable effect on the response time in liquid samples (this diffusion limitation is negligible in gas samples)
11
5.2 Dipping Probe
Characteristics: dipping probe for Piccolo2 (item no. OPDIP20)
with
stainless steel tubing 3 mm in diameter and
200 mm in length
oxygen sensor with optical isolation located
in the 3 mm disc at the tip of the tubing (sensor end, see chapter 5.7)
Applications:
standard laboratory applications in gases and liquids full probe length can be dipped into the sample in aqueous liquids: permanent stirring is recommended, as
diffusion limitation of oxygen towards the 3 mm tip has a measurable effect on the response time in liquid samples (this diffusion limitation is negligible in gas samples)
Important: No part of the Piccolo2 device itself must be submersed into liquid samples!
5.3 Sensor Spots
Measuring principle: oxygen sensor spots (item no. OXSP5)
oxygen measurements with contactless read-out
through transparent container walls
12
sensor spot with REDFLASH indicator needs to
be glued to the inner container wall
excitation and detection of oxygen indicator
luminescence emission through the transparent container wall with an optical fiber fixed at the outer container wall
Characteristics:
PET foil coated with REDFLASH indicator and optical isolation standard diameter of 5 mm to be glued with the green back side on transparent, clean and dry
inner container walls (e.g. glass or acrylic glass)
black sensing surface has to be covered completely by the gaseous
or aqueous sample
can be autoclaved few cycles at 121°C for 15 min (requires
calibration afterwards)
Accessories for readout: 0-10 mm thick transparent windows
basic spot adapter SPADBAS: 12 mm in diameter, needs to be
fixed tightly or glued to the outer container wall at the spot position
optical fiber PICFIB2: black fiber cable, 3 mm in
diameter, 1 m in length, connecting the basic spot adapter SPADBAS with the Piccolo2 meter (see also chapter 5.7) or, alternatively,
optical fiber rod PICROD2: stainless steel tubing, 3 mm diameter,
40 mm length, connecting the basic spot adapter SPADBAS with the Piccolo2 meter (see also chapter 5.7)
transparent silicone glue SPGLUE: based on acetic acid, for gluing
sensor spots and basic spot adapters to the container wall
The PICFIB2 and PICROD2 are also compatible with the flow­through cells and respiration vials from Pyro Science.
13
Note: A change of position of the spot adapter after calibration of the sensor spot might require a new calibration.
Accessories for readout: 10-20 mm thick transparent windows
optical fiber rod PICROD3: ca. 50 mm in length, needs a
customized fixation on the outer container wall (not compatible with the spot adapter SPADBAS)
The optical fiber rod PICROD3 is not compatible with the basic spot adapter SPADBAS. The user has to build a custom fixation at the wall of the sample container.
Applications:
multi-sampling measurements measurements of oxygen at greater scales measurements in closed sample containers,
e.g. in respiration chambers, bioreactors, cell biological approaches and industrial process water-monitoring
CAUTION: Oxygen measurements in air-tight containers or setups require special precautions, like constant conditions during measurements. More details on request.
Options:
trace range version (item no. TROXSP5): for measurements close
to 0% O2 (range: 0-10% O2)
5.4 Flow-Through Cells
Characteristics: flow-through cells (item no. OXFTC, OXFTC2)
flow-through cells with integrated oxygen sensor available as small version OXFTC and as large version OXFTC2
14
equipped with Luer lock connectors on both tubing ends included Luer lock adapters allow connection of gas-
tight tubings (e.g. stainless steel, ISO-VERSINIC®) with about 1.0-2.0 mm inner diameter to the OXFTC and about 3.0-4.0 mm to the OXFTC2
Accessories for readout:
optical fiber PICFIB2: black fiber cable, 3 mm in diameter, 1 m in
length, connecting the flow-through cell with the Piccolo2 meter (see also chapter 5.7)
Applications:
contactless measurement of oxygen in a
gaseous or liquid sample pumped through the cell
CAUTION: For liquids, a flow rate of ca. 10-100 mL/min for the
OXFTC and of 20-500 mL/min for the OXFTC2 is recommended!
Lower flow rates are in principle possible; however the user has to ensure sufficient gas tightness of the connecting tubing.
Options:
trace range version (item no. TROXFTC): for measurements close
to 0% O2 (range: 0-10% O2)
5.5 Respiration Vials
Characteristics: respiration vials (item no. OXVIAL4, OXVIAL20)
ready assembled respiration vials with integrated stripe of the
oxygen sensitive REDFLASH indicator with optical isolation
available with a (uncalibrated) volume of ca. 4 mL (OXVIAL4)
or ca. 20 mL (OXVIAL20)
15
elongated shape of the integrated sensor allows
oxygen measurements at different heights within the vial
Note: A change of position of the adapter ring after sensor calibration might require new calibration.
Accessories for readout:
adapter rings ADVIAL4 / ADVIAL20: for easy fixation of the optical
fiber PICFIB2 to OXVIAL4 / OXVIAL20
optical fiber PICFIB2: black fiber cable, 3 mm in diameter, 1 m in
length, can be fixed in the adapter ring with the clamping screw, connecting the respiration vial with the Piccolo2 meter (see also chapter 5.7)
Applications:
for oxygen measurements in e.g. stirred aqueous liquids or
gaseous sample
small-scale respirometry and metabolic rate measurements of e.g.
cell cultures, eggs, larvae, small crustaceans, small fish, water-, plant-, algal-samples etc.
CAUTION: Oxygen measurements in air-tight vials require special precautions, like e.g. constant temperature. More details on request.
16
5.6 Nanoprobes
Measuring principle: oxygen nanoprobes (item no. OXNANO)
nanoparticles coated with oxygen-
sensitive REDFLASH indicator
dispersible in aqueous solutions, like e.g.
water, culture media
contactless readout from outside through
transparent windows
Characteristics:
oxygen nanoprobes OXNANO in 5 x 10 mg units dispersible in aqueous solutions ultra-fast response times simple batch calibration measurements at different positions or in
different vials containing the same dispersion
Accessories for readout: vials, microwell plates
basic spot adapter SPADBAS: 12 mm in diameter, needs to be
fixed tightly or glued to the outer container wall
optical fiber PICFIB2: black fiber cable, 3 mm in diameter, 1 m in
length, connecting the basic spot adapter with the Piccolo2 meter (see also chapter 5.7) or, alternatively,
optical fiber rod PICROD2: stainless steel tubing, 3 mm in
diameter, 40 mm in length, connecting the basic spot adapter with the Piccolo2 meter (see also chapter 5.7)
Accessories for readout: microfluidic applications
optical fiber with integrated focusing lens PICFIB2-LNS: black fiber
cable, 1 m in length, stainless steel tip 3 mm in diameter, connecting the basic spot adapter or custom fixation with the
Piccolo2 meter (see also chapter 5.7) or, alternatively,
17
optical fiber rod with integrated focusing lens PICROD2-LNS:
stainless steel tubing, 3 mm in diameter, 40 mm in length, connecting the basic spot adapter or custom fixation with the
Piccolo2 meter (see also chapter 5.7)
Applications:
real-time monitoring high through-put screening in e.g. microwell
plates
contactless measurements of fast processes,
like enzymatic reactions
in microfluidic devices (microfluidic chip,
lab-on-a-chip device)
more details on request
CAUTION: Oxygen measurements in closed vials require special precautions, like constant conditions during measurements. More details on request.
18
5.7 Connecting the Sensors and Optical Fibers
The robust probe OPROB3, the dipping probe OPDIP20, and the optical fibers PICFIB2 / PICFIB2-LNS or the optical fiber rods
PICROD2 / PICROD3 / PICROD2-LNS must be connected to the
sensor port of the Piccolo2 with their instrument end. At the instrument end IE, the core of the optical fiber protrudes ca. 1 mm from the fiber jacket. For OPROB3, OPDIP20 and PICFIB2 the instrument end IE is marked with "Instrument" on the attached label.
First, remove the protective caps from both ends of the optical probe / fiber / rod. Then slightly unscrew and loosen the nut at the sensor port of the Piccolo2. The nut must not be removed from the sensor port. Typically, unscrewing the nut with a single half turn (180°) is sufficient. Insert now the instrument end IE carefully into the sensor port of the Piccolo2, the insertion length is about 24 mm (see pictures with insertion of a) a dipping probe and b) an optical fiber rod).
19
Insert the optical probe / fiber / rod carefully as deep as possible (ca. 24 mm) into the sensor port of the Piccolo2!
Fix the optical probe / fiber / rod by screwing down the nut tightly onto the sensor port of the Piccolo2.
a)
b)
Then, insert the adapter end (AE) of the optical fibers PICFIB2 /
PICFIB2-LNS or optical fiber rods PICROD2 / PICROD2-LNS into
the basic spot adapter, adapter ring or flow-through cell, and fix it with the clamping screw.
For OPROB3 and OPDIP20, insert the sensor end (SE) into the gas or (stirred) aqueous liquid sample.
20
5.8 Cleaning and Maintenance of the Sensors
Oxygen sensors for the Piccolo2 can be sterilized with ethylene oxide (EtO) and can be cleaned with ethanol, peroxide (3% H2O2) or soap solution. They can be applied in gas phases, in aqueous solutions, as well as in ethanol, methanol or isopropanol (robust / dipping probe: only short-term application in diluted ethanol, methanol or isopropanol). Other organic solvents and gaseous chlorine (Cl2) interfere with the oxygen sensor reading. No cross­sensitivity is found for pH 1-14, CO2, CH4, H2S and any ionic species.
A signal drift of the sensor can indicate photo-bleaching of the oxygen-sensitive REDFLASH indicator, depending on the ambient light intensity, as well as the intensity of the excitation light and the sample interval. This can necessitate new calibration of the sensor and eventually also readjustment of the Sensor Settings (LED intensity; see also chapter 6.2.2). In case of sensor spots, re­positioning of the spot fiber on the sensor spot and subsequent new calibration might be necessary. If the signal intensity is getting too low (typically after several millions of data points), as indicated by the horizontal indicator bar in the main window of the Pyro
Oxygen Logger software and by the respective warning (see
chapter 6), the sensor needs to be replaced.
After finalization of the measurements, the sensor tip of the robust and dipping probe should be rinsed carefully with demineralized water. Especially after application in seawater, it is recommended to clean the sensor thoroughly with demineralized water. After drying, put on carefully the protective caps onto both ends of the probe.
For sensor spots and respiration vials, wet cotton swabs can be used for carefully wiping the sensing surface. Rinse the sensing
21
surface sufficiently with demineralized water afterwards to remove small particles and let it dry before storage.
After drying, put the respiration vials into the corresponding black bag and store them in a secure, dry and dark place at room temperature!
Store sensors in a dry, dark and secure place at room temperature.
22
6 The Software "Pyro Oxygen Logger"
This chapter describes all functions of the Pyro Oxygen Logger software excluding sensor calibration, which is described in detail in chapter 7 and 9.
6.1 Main Window
After start of the Pyro Oxygen Logger software the following main window is shown:
As default, uncalibrated oxygen sensor readings (raw value) are shown, which give only a qualitative information of the actual oxygen level.
Note that placing the cursor on many elements in the window will show a short description ("tool tip"). By clicking on the right mouse button and selecting "Description and Tip" a more detailed description might be available additionally.
23
The sensor readings are displayed in a numeric display (D) and in a chart recorder (C) in the chosen oxygen unit (UD). The color and appearance of each graph can be changed by clicking on the color­control (CC), opening a pop-up menu. With Common Plots, Color,
Line Style, Line Width, Interpolation, and Point Style the chart appearance can be changed. Clicking on the small rectangular button (to the left of CC) allows hiding / showing the respective graph.
The visible time frame of the chart recorder (C) can be changed by moving the bar along the scroll bar (SCB). Switching off the Autoscroll button will allow inspection of older data which are not visible anymore in the time frame, e.g. during long-time measurements.
The description of the sensor, as defined in the Settings by the Sensor Code, is shown in the description display (DD). The Signal
24
Intensity (SI) of the oxygen sensor is shown as a horizontal
indicator bar just underneath the numeric display (D).
A reasonable oxygen sensor shows signal intensities well above 50 (typically 50-500)1. If the signal intensity drops below 50, the indicator bar (SI) turns gradually from grey to red, indicating that the sensor gets degraded soon.
At signal intensities <10, the warning Low signal will appear in the warning display (WD). At a signal intensity <5, the
display changes to NaN (Not a Number), indicating that the signal is too low and the sensor needs to be replaced (or moved to another position on the sensor spot).
Besides the warning Not calibrated for an uncalibrated oxygen sensor, the warning display (WD) can show the following warnings:
Low signal The sensor is either not connected or needs to be replaced by a new one. In case of contactless sensors it might indicate that the distance between the optical fiber and the sensor spot is too large. (For advanced users: increase the LED intensity and / or the amplification in the Advanced Settings).
Signal too high There might be too much ambient light on the sensor tip or sensor spot. Avoid direct sun light exposure or strong direct illumination with a lamp. (For advanced users: decrease the LED intensity and / or the amplification in the Advanced Settings)
Bad reference This indicates internal problems of the Piccolo2. Please contact Pyro Science for support.
1
Note: Exceptions are trace oxygen sensors. During the air calibration at 21% O2, these sensors naturally show a very low signal intensity (as low as 10). But the signal intensity will strongly increase when a trace oxygen sensor is applied within its specified range of 0-10% O2.
25
Refer to the Troubleshooting in chapter 10.2.
The actual Compensation Temperature (i.e. the temperature used for calculating the oxygen values) is shown in the temperature display (TD) in units of degree Celsius (°C).
If a sensor is not yet calibrated, the warning Not Calibrated is shown on the right-hand side of the Calibrate button. As long as the sensors are not calibrated, the data are shown in units of "raw value" reflecting the measured oxygen levels only qualitatively. In order to switch to quantitative oxygen units, the sensor settings need to be adjusted in the Settings and calibration has to be performed by clicking on the button Calibrate. The calibration procedure is explained in detail in chapter 7.
The adjustment of the Settings using the button Settings is described in detail in chapter 6.2.
The button Save Setup can be used to save the current settings and calibration data of the Piccolo2. They can be reloaded anytime by pressing the button Load Setup. This allows e.g. to switch between different laboratory setups with a single Piccolo2. This function might be useful also if different computers are used for calibration and for actual measurements. You might calibrate the sensors with the first computer, save the configuration with Save Setup. By transferring this file and also the oxygen meter Piccolo2 to a second computer, you can load this configuration again with Load Setup, giving you calibrated sensors ready for measurement.
Note, that for this procedure identical software versions of the
Pyro Oxygen Logger must be installed on both computers.
26
Save Setup and Load Setup might be useful also if e.g. a Piccolo2 should be used repeatedly with several oxygen sensors. Initially, each sensor needs to be calibrated only once and the configuration of each sensor is saved with Save Setup. If later on a measurement should be performed with a specific sensor, it is only necessary to load the configuration for this sensor with Load Setup (but it must be checked if the calibration is still valid).
All current settings and calibration data are automatically saved when closing the Pyro Oxygen Logger software and are automatically loaded again at the next startup. The Last Setup loaded is shown underneath the Settings button.
The button Flash Logo causes a short flashing (ca. 1-2 sec) of the status LED of the Piccolo2. Several Piccolo2 meters can be connected to the PC in parallel and multiple measurements can be performed by opening the Pyro Oxygen Logger software a number of times corresponding to the number of connected Piccolo2 meters. The different windows operate completely independent of each other and are assigned to exactly one Piccolo2. This allows measurements in different setups at the same time. The flashing of the status LEDs can help to assign a specific logger window to the corresponding Piccolo2 (more details in chapter 10.4).
Clicking on Raw Data opens a pop-up window Oxygen Sensor Raw Data which is described in chapter 6.3.
Clicking on the button Bar Graph will open the Bar Graph Window. Here, different parameters can be chosen to be displayed in the bar graph by clicking next to "Select data type", e.g. the Signal Intensity (mV) of the connected oxygen sensor.
27
A Measurement is started by clicking on the measurement start button (MSB). The arrow in the button turns from dark green to light green indicating that a measurement is in progress. Clicking on it again will stop the measurement.
The mode of Measurement can be chosen as single data point acquisition, continuous sampling (default setting) or as continuous sampling limited to a defined time interval. The duration of this time interval can be adjusted in the duration display (DUD) shown as hour (HH): minutes (MM): seconds (SS).
The Sample Interval (s) for continuous sampling can be defined in the field designed with set. Setting the sample interval to 0.25 will give the maximal possible scan rate. The exact maximal rate depends on the settings. The actual sample interval is shown in the display actual and is displayed in red if the actual ≠ the set sample interval.
Acquired data can be smoothed by a Data Smoothing (range 110, default: 3, a value of 1 means no data smoothing). For continuous or duration measurements with a sample interval <10 s, data smoothing is done by a simple running average (e.g. with Data Smoothing=5 always the last 5 sampled data points are averaged). However, for single data point measurements and for continuous or duration measurements with sample intervals >10 s, the data smoothing is done by averaging repetitive measurements (e.g. with Data Smoothing=3 for each data point 3 oxygen measurements are performed as fast as possible sequentially, and the average of these 3 measurements is displayed as the new data point).
28
IMPORTANT: By default the displayed data are not automatically saved to a data file.
To activate data logging, click on the red start button (SB) of Log to File. Select a file name in the appearing file dialog. The saved data files are simple text-files with the file extension ".txt", which can be imported easily into common spreadsheet programs.
Thereafter, the indicators Comment and File Path are shown in the main window additionally. In the field Comment, the measure­ments can be commented and this comment is then saved together with the next data point into the data file.
During data logging, the data file can be displayed and opened by clicking on the button Show File. The data logging is indicated by the light green arrow in the grey Log to File button and can be stopped by clicking this button again.
NOTE: During data logging, the buttons Settings and Calibrate are not active and cannot be used before Log to File is stopped.
29
The display of the data in the charts can be changed by different chart tools arranged underneath the chart recorder.
The button with the magnifying glass offers different zoom options. After clicking the button with the hand, the user can click on the chart and move the whole area while keeping the mouse button pressed.
The button Clear Graph offers the option to clear the graph and to zero the time scale. Note, that this will not affect the saved data in the data file.
The unit of the x-axis can be changed with the selector Time Scale. The time scale can be displayed in Seconds (s), Minutes (min), Hours (h), Relative Time (HH:MM:SS), Absolute Time (HH:MM:SS), Absolute Time and Date, and numbered Data Points.
The scales of the y- and x-axis can be adjusted by clicking on Adjust Scales, opening a pop-up window:
The upper (Maximum) and lower limits (Minimum) and the Increment of the Y Scale (Oxygen) and of the X Scale (Time) can be changed by clicking on the respective selectors or by double-clicking directly onto the field and entering the values manually (changing these parameters will automatically deactivate the autoscaling). Autoscaling for all axes can be activated with the Autoscale button. The arrow in the button turns from dark green to light green indicating that Autoscale is activated. By default only the y-axis Oxygen is in Autoscale mode.
30
6.2 Settings
To open the dialog window Piccolo2 Settings click on the Settings button in the Main Window:
The Settings can be adjusted only when data logging is not active.
In the Settings the user has to enter the Sensor Code of the connected oxygen sensor and to define (1) the Sensor Settings, (2) the Conditions in the Sample under investigation and (3) the oxygen Units for the measurements.
In the Channel 1 tab, the oxygen sensor can be activated / deactivated by clicking on the button Activate. Activation is indicated by a change from dark to light green of the arrow in the button. A text describing the connected sensor type appears on the right-hand side of this button after the Sensor Code (see label on sensor) has been entered. This description will be shown in the description display (DD) of the main window and in the data file as well.
31
Please take care that the sensor code of the sensor connected to the Piccolo2 is entered into the field Sensor Code in the Channel 1 tab of the window Piccolo2 Settings. It includes information for optimal sensor settings and for calibration data needed for the (rough) factory calibration and for the 1-point calibration. The first letter of the sensor code defines the sensor type. A detailed explanation of the sensor code is given in chapter 10.7.
The oxygen units can be selected by the selector Units. The selectable units include raw value (default), % air saturation, % O2, mL/L, µmol/L, mg/L (ppm), hPa (mbar), mmHg (Torr), dphi and µg/L (ppb). For measurements in a Gas Phase only the units raw value, % O2, hPa (mbar), mmHg (Torr) and dphi can be selected, whereas for measurements of dissolved oxygen in a Water sample (DO) all units except % O2 can be selected. For details please refer to chapter 10.5.
NOTE: The chart in the main window is automatically cleared after the Settings have been modified. Readjustments in the Settings might require also new calibration of the sensor.
If changes of the Settings require new sensor calibration, a warning Not Calibrated appears on the right-hand side of the Calibrate button in the main window.
6.2.1 Basic Settings
The Sensor Settings can be adjusted in a Basic or an Advanced mode. The first-time user is advised to work with the Basic Settings. Please ensure that the correct sensor code (see label on the connected sensor) has been entered in the field Sensor Code.
The Measuring Mode can be adjusted gradually between low drift (1) and low noise (128) of the sensor signal by moving the arrow with the cursor along the scale,
32
thereby changing the oxygen measuring time. For standard applications, an intermediate mode (8 or 16) is advisable.
The Fiber Length (m) of the connected robust probe (OPROB3) or of the connected optical fiber (PICFIB2) for contactless sensors must be entered, which is typically 1 m (refer to chapter 9.2 for more details). For the dipping probe OPDIP20 and for the optical fiber rods PICROD2 and PICROD3, please enter a Fiber Length of
0 m (zero m).
NOTE: Ensure that the correct sensor code has been entered. If the
sensor is not yet calibrated, a warning Not Calibrated is displayed in the main window.
6.2.2 Advanced Settings
If Advanced Sensor Settings are chosen, more complex setting controls will be shown. Please ensure that the correct sensor code (see label on sensor) has been entered in the field Sensor Code.
The Advanced Measuring Parameters comprise the LED Intensity for excitation of the REDFLASH
indicator (in %) and Amplification of the sensor
signal. As a rule of thumb, the LED Intensity should be 10-30% for robust / dipping probes, but can be increased up to 100% for contactless sensors (sensor spots, flow-through cells, respiration vials). The Amplification should be typically chosen as 80x, 200x or 400x. Note that varying the LED Intensity and the Amplification has direct influence on the signal intensity and therefore on the signal-to-noise-ratio.
The Oxygen Measuring Time defines the integration time for acquisition of a single data point. Shorter measuring times provide low long-term drift, whereby longer measuring times assure less noise. The maximal possible value is 128 ms.
33
NOTE: If using the Advanced Sensor Settings, it is recommended to perform a 2-Point calibration of the oxygen sensor. Later readjustments in the Advanced Settings might require also new calibration of the sensor.
The Fiber Length (m) of the connected robust probe (OPROB3) or of the connected optical fiber (PICFIB2) for contactless sensors must be entered, which is typically 1 m (refer to chapter
9.2 for more details). For the dipping probe
OPDIP20 and for the optical fiber rods PICROD2
and PICROD3, please enter a Fiber Length of 0 m.
Alternatively, it is possible to select Manual Background Compensation, which is described in detail in chapter 9.2. Finally, the background compensation can be completely deactivated by selecting No Background Compensation.
NOTE: Generally, it is advised to select Fiber Length (m) and to enter the fiber length there. The alternative options Manual Background Compensation or No Background Compensation are only intended for advanced users (see also chapter 9.2).
6.2.3 Conditions in the Sample
The next step is determination of the Conditions in the Sample, which can be Water for dissolved
oxygen (DO) or a Gas Phase.
The temperature of the environmental sample has to be measured with a thermometer, kept constant and entered manually as Fixed Temperature (°C). This "Compensation Temperature" will be displayed in the main window (TD).
34
IMPORTANT: The measurements must be performed at constant temperature and constant environmental conditions!
It is important to keep the temperature constant during calibration and measurements, as
the luminescence of the REDFLASH indicator is temperature
dependent and
the conversion of some oxygen units is dependent on
temperature.
The atmospheric pressure (e.g. read from a separate barometer) must be entered manually as Fixed Pressure (mbar). Common conditions refer to 1013 mbar (default).
If the actual atmospheric pressure cannot be determined on site, it is also possible to enter the Elevation (m) above sea level. For this click on Elevation and enter the actual elevation in meters. This procedure will calculate only the average atmospheric pressure for this elevation; therefore, this option is less precise than measuring the actual atmospheric pressure.
The Salinity (g/L) of the environmental sample needs to be measured and entered, e.g. in case of saline water. For measurements in gas samples this value has no relevance (and is not active). For measurements in liquid samples, the salinity is only relevant if a concentration unit was selected (e.g. mg/L or µmol/L).
6.2.4 Options
In the panel Options, it is possible to designate a specific name to the connected Piccolo2 by typing a name in the field Device Name e.g. "water sample #12". This device name is then displayed in the top line of the main window. This option is useful especially if several Piccolo2 devices are operated in parallel in order to distinguish the opened logger windows.
35
The maximum number of data points kept in the graphs can be changed by the selector Max. Data Points in Graphs (default:
10800). A change of the number of data points will clear the graphs, and high values (>>10000) might decrease the maximum sample rate (increase the actual sample interval).
Advanced Options (only for advanced users):
The USB communication speed can be adjusted e.g. for improving the maximum sampling rate (default: 57600).
Activation of the button Enable High-Speed Sampling will enable adjustment of a Sample Interval <0.25 s in the main window (and disable Max. Data Points in Graphs).
The Data Export is relevant only for OEM modules or third-party software.
36
6.3 Raw Data Window
The Raw Data Window is mostly intended for trouble shooting and advanced users. During standard measurements it is not needed in general. After clicking on the Raw Data button in the main window the Oxygen Sensor Raw Data window will open.
NOTE: While the Oxygen Sensor Raw Data window is opened, all raw values are saved also in the data file in additional columns behind the standard data columns.
The panel of the oxygen channel (Chan 1) shows the phase shift as "delta phi" (dphi, in °). dphi is the actual measured raw value which is used for internal calculation of oxygen concentration Oxygen (µM), Oxygen partial pressure (hPa), Oxygen in % air saturation (% air sat) and Oxygen (% O2) (see also chapter
10.3).
The Signal Intensity (in mV) gives a measure of quality of the oxygen measurement and is also displayed in the horizontal bar indicator in the main window.
Ambient Light (in mV) gives a measure of the ambient light entering the sensor from outside. At too high ambient light levels, the detector of the Piccolo2 might get saturated, indicated by the warning Signal too high in this window and in the warning display of the main window.
On the left side, the Status and different warnings related to the signal and reference intensity (too low, too high) can be indicated.
37
On the right side of the channel tab, a graph can be activated, showing the dphi (°) and Signal Intensity (mV) in the graph (default setting). Plotting of additional parameters can be activated by clicking on the small rectangular button next to the color control of the respective parameter.
38
7 Calibration of Oxygen Sensors
This chapter describes the possible calibration modes for oxygen sensors connected to the Piccolo2 meter using the logger software
Pyro Oxygen Logger.
For sensor calibration, the temperature in the calibration standard, the atmospheric pressure and the relative humidity of the ambient air are important parameters ensuring high precision of calibration. It is on the user´s authority to measure and adjust these parameters manually.
This chapter covers only the necessary steps during the calibration procedure. For details regarding preparation of calibration standards refer to chapter 8.
Please note that during oxygen sensor calibration, the Sample Interval is automatically set to 0.5 s and the Data Smoothing to 5, ensuring fast determination of a precise mean value during sensor calibration. After completion of the calibration, the program returns to the former settings automatically.
7.1 Calibration
To calibrate a sensor, click on the button Calibrate in the main window. Note that during data logging this button cannot be used until Log to File is stopped.
A dialog window Oxygen Sensor Calibration will open in which the calibration mode can be selected:
39
Factory calibration (for a quick, rough calibration): taking the 0% and the air calibration values from the sensor code; advised only
for rough measurements or testing purposes.
1-Point: taking the 0% value from the sensor code and the air
calibration value from a manual calibration for precise measurements around 21% O2.
2-Point: taking the 0% and the air calibration value from a manual calibration for precise measurements over the full range (0-21% O2 or 0-100% dissolved O2 (DO)).
NOTE: It is recommended to determine the air calibration value in air saturated water if measurements will be performed in water samples (aqueous liquids).
For advanced users and applications, a Custom Mode can be selected, allowing the user to combine freely the possible calibration types for the air and the 0% calibration.
7.1.1 Calibration Mode: Factory
NOTE: Factory Calibration is intended only for rough measurements and testing purposes. It is only possible if the correct Sensor Code has been entered in the Settings (see chapter
6.2).
40
If the calibration mode Factory Calibration is chosen, ensure that the correct sensor code has been entered in the Settings (as displayed in 2. Adjust Calibration Conditions of the Oxygen Sensor Calibration window). If the sensor code displayed is not correct, click on Finish, open Settings, enter the correct Sensor Code and repeat Factory Calibration.
After clicking on Finish factory calibration is completed, and the program returns to the main window.
7.1.2 Calibration Mode: 1-Point
NOTE: The calibration mode 1-Point is only possible if the correct
Sensor Code has been entered in the Settings (see chapter 6.2).
The calibration mode 1-point is selected by clicking on the button 1-Point. This mode uses a manual calibration in an air calibration standard for adjusting the air calibration value. The 0% calibration value is taken from the Sensor Code. The preparation of appropriate air calibration standards is described in chapter 8.1.
Depending on the air calibration standard used, select Water (DO) for dissolved oxygen or Gas Phase.
The air calibration standard (see also chapter 8.1) can be:
ambient air of known humidity, water-vapor saturated air or air saturated water (100% air saturation).
The oxygen level in the calibration standard (unit: %O2) can be freely chosen in Oxygen (%O2). If the air calibration standard is based on ambient air or air saturated water, then this value should be kept at 20.95% O2 (default), representing the standard oxygen volume fraction in ambient air. However, other values can be adjusted if custom calibration gases are used, of e.g. 5% O2, which might be useful when using trace oxygen sensors.
41
The temperature of the calibration standard needs to be measured, kept constant and entered manually in Fixed Temperature.
In addition, the actual atmospheric pressure in the calibration standard can be entered manually in Pressure (mbar). Normal conditions refer to 1013 mbar (default setting).
If the actual atmospheric pressure cannot be determined on site, alternatively it is possible to enter the actual Elevation (m) in meters above sea level. For this click on Elevation and enter the actual elevation. Please note, that this option takes only the elevation-dependent pressure change into account, but not the variations due to actual weather conditions. Therefore, determining the actual atmospheric pressure with a barometer gives more precise results.
The relative Humidity (%RH) of the gas phase (e.g. air) needs to be adjusted. If a calibration standard with water-vapor saturated air is used (see chapter 8.1.2), it must be adjusted to 100% RH. Otherwise, the humidity must be determined with a hygrometer and must be entered.
Now place the oxygen sensor into the air calibration standard.
NOTE: Ensure constant calibration conditions and a constant temperature in the calibration standard!
42
Wait for steady-state until the sensor reading is stable by observing the graph. Also ensure stable temperature during calibration.
Note that the button Set Air will be highlighted as soon as the oxygen reading is within the expected range for the connected sensor type (the latter does not apply for custom air calibration with ≠20.95% O2).
If the oxygen readings have reached their steady-state, click on Set Air, and the current oxygen sensor reading is taken for the air calibration. If the oxygen reading seems to be out of the expected range, a warning will be shown offering the option to repeat calibration (it is not recommended to continue without checking the calibration standard and the sensor!).
A completed air calibration is indicated by the green indicator Calibrated at Air.
For Factory 0% Calibration, no further steps are necessary (refer to chapter 7.1.3). Ensure that the correct sensor code has been entered in the Settings.
7.1.3 Calibration Mode: 2-Point
The calibration mode 2-point is selected by clicking on the button 2-Point. In this mode both the air calibration value and the 0% calibration value are determined in specially prepared calibration standards. Preparation of appropriate 0% and air calibration standards is explained in chapter 8.
The air calibration standard and the calibration conditions need to be defined and entered as described for the 1-Point calibration (see chapter 7.1.2).
Now place the oxygen sensor into the air calibration standard.
43
NOTE: Ensure constant calibration conditions and a constant temperature in the calibration standard!
Wait for steady-state until the sensor reading is stable by observing the graph. Ensure also stable temperature conditions in the standard during calibration.
Note that the button Set Air will be highlighted as soon as the oxygen readings are within the expected range for the connected sensor type (the latter does not apply for custom air calibration with ≠20.95% O2).
If the oxygen sensor readings have reached their steady-state, click on Set Air, and the current oxygen sensor reading is taken for air calibration. If the oxygen reading seems to be out of the expected range, a warning will be shown offering the option to repeat calibration (it is not recommended to continue without checking the calibration standard and the sensor!).
A completed air calibration is indicated by the green indicator Calibrated at Air.
Subsequently, insert the oxygen sensor into the 0% calibration standard. Wait for steady-state until the sensor reading is stable by observing the graph. Ensure stable temperature conditions during calibration.
44
Note that the button Set 0% will be highlighted as soon as the oxygen readings are within the expected range for the connected sensor type.
If the oxygen sensor readings have reached their steady-state, click on Set 0%, and the current oxygen sensor reading is taken for 0% calibration. If the oxygen reading is out of the expected range, a warning will be shown offering the option to repeat calibration (it is not recommended to continue without checking the calibration standard and the sensor!).
A completed 0% calibration is indicated by the green indicator Calibrated at 0%.
Finally, click on Finish for returning to the main window.
7.1.4 Calibration Mode: Custom
The custom calibration mode is selected by clicking on the button Custom Mode. This mode allows to freely combine all possible calibration types for the air calibration and the 0% calibration. The air calibration type can be selected by clicking on the "Air Calibration Selector". The 0% calibration type can be selected by clicking on the "0% Calibration Selector".
45
The following air calibration types can be selected:
Factory Air Calibration (refer to chapter 7.1.1) Air Calibration (refer to chapter 7.1.2)
For the 0% calibration the following types can be selected:
Factory 0% Calibration (refer to chapter 7.1.1) 0% Calibration (refer to chapter 7.1.3)
7.2 Advanced adjustments
This section is for advanced users only!
For very advanced applications, it is possible to manipulate all internal calibration parameters of the Piccolo2 manually. This option is accessible by selecting Custom Mode and subsequently clicking on Advanced: Adjust Manually, which opens a separate window showing all internal calibration parameters.
Here, the Upper Calibration Point (default: Partial Volume of Oxygen 20.95% O2) can be defined. The calibration conditions (Temperature, Air Pressure, Humidity) need to be measured and entered, as well as the Temperature (°C) at 0% for the 0% Calibration Point.
46
Further, the phase shift "dphi" (dphi, see chapter 10.3) for the 0% calibration standard (dphi 0% in °) and for the air calibration standard (dphi 100% in °) can be adjusted manually.
The Background Amplitude (in mV) and the Background dphi (phase shift in °) can be adjusted (refer to chapter 9.2 for more details). These two values are relevant only if background compensation for measurements with contactless sensors has been activated in the Advanced Settings.
The parameters f, m, F, kt, mt, Tofs, and tt are needed for internal calculation of the oxygen concentration data. These parameters are specific constants for the REDFLASH indicators, and are automatically adjusted for the selected Sensor Type in the Settings. Unless otherwise communicated by Pyro Science, it is
strongly advised to leave these parameters at their default values.
47
8 Calibration Standards
8.1 The Air Calibration Standard
The Air Calibration standard can be
ambient air water-vapor saturated air air saturated water (100% air saturation)
Always use a proper lab stand for mounting the oxygen sensor!
All air calibration standards described in the following rely on the virtually constant oxygen content in the earth’s atmosphere of about 20.95% O2 in dry air. Slight deviations might be given in closed rooms occupied by many people (or e.g. candles, combustion engines) consuming the oxygen. So if in doubt, ensure good ventilation of the room with fresh air e.g. by opening a window for some minutes.
Furthermore, the relative humidity of the air causes deviations from the ideal value of 20.95% O2. Simply speaking, the water vapor in humid air replaces a fraction of the oxygen, resulting in a diminished oxygen level of e.g. 20.7% O2. For temperatures around and below 20°C, this effect fortunately causes a maximum deviation of about 0.5% O2 only. However, for higher temperatures at 30°C or even 40-50°C, the humidity of the air has a significant influence on the actual oxygen level. For example, ambient air at body temperature (37°C) with 100% relative humidity contains only 19.6% O2 compared to dry air with 20.95% O2.
During calibration of oxygen sensors, there are two possibilities to take the humidity into account:
48
(1) The relative humidity and the temperature of the ambient air must
be determined during calibration and entered into the software. Then, the Pyro Oxygen Logger software will automatically calculate the real oxygen level under these conditions.
(2) The calibration standard is prepared in a closed vessel either filled
with water or partly filled with e.g. wet cotton wool or a wet sponge. This ensures a constant humidity of 100% RH and there is no need to measure the humidity.
Option (2) is utilized for the calibration standards "Water-Vapor Saturated Air" (see section 8.1.2) and "Air Saturated Water" (see section 8.1.3).
Another parameter even more important for the air calibration standard is the atmospheric pressure. The principle parameter measured by the oxygen sensors is not the partial volume (i.e. "% O2"), but the partial oxygen pressure (i.e. "mbar") (see also chapter
10.5). So, an oxygen level of e.g. 20.7% O2 (determined as described above by a given humidity and temperature) is converted internally by the Pyro Oxygen Logger software into a partial pressure of oxygen essentially by multiplying the relative oxygen level with the atmospheric pressure of e.g. 990 mbar (see chapter 10.5):
0.207 x 990 mbar = 205 mbar
giving a partial oxygen pressure of e.g. 205 mbar. This is the essential calibration value used internally by the Pyro Oxygen
Logger software. The atmospheric pressure can be influenced 1) by
weather changes (e.g. varying between ca. 990 mbar and 1030 mbar at sea level) and 2) by the elevation above sea level (e.g. at 1000 m elevation the typical atmospheric pressure is about 900 mbar compared to 1013 mbar at sea level).
49
Thus, in summary, there are three important parameters to be known for the air calibration standard:
Temperature (°C) Relative Humidity (% RH) Atmospheric Pressure (mbar)
8.1.1 Ambient Air
If ambient air is used as the air calibration standard, the following parameters must be independently determined in parallel
- air temperature (thermometer),
- air humidity (hygrometer),
- atmospheric pressure (barometer),
and entered into 2. Adjust Calibration Conditions of the Oxygen Sensor Calibration window. Ensure that these three parameters are kept constant during the calibration procedure.
Then, the dry oxygen sensor connected to the Piccolo2 is exposed to the ambient air. Otherwise, follow the calibration procedures given in chapter 7.
For precise calibrations in ambient air, it is important that the measuring tip / surface of the oxygen sensor is completely dry. Wet sensor tips will cause undefined humidity levels around the sensor tips. And even worse, the evaporation of water drops would cool down the sensor tips causing undefined temperatures.
8.1.2 Water-Vapor Saturated Air
Enclose wet cotton wool into a flask (e.g. DURAN flask) with a lid prepared with one hole for the oxygen sensor connected to the
Piccolo2 and another hole for a temperature sensor. Typically,
about 1/3 to 1/2 of the flask volume is filled with wet cotton wool, while the other volume fraction is left free for inserting the tip of
50
the oxygen sensor and a temperature sensor. Otherwise, follow the calibration procedures given in chapter 7.
8.1.3 Air Saturated Water
Fill an appropriate amount of water into a flask (e.g. Duran flask) with a lid prepared with holes for inserting the oxygen sensor and a temperature sensor. For about 10 minutes, stream air through the water with an air stone connected to an air pump (available as commercial equipment for fish aquaria). Alternatively, if no air pump is available, fill water into the flask leaving >50% air in the headspace, close it with a lid and shake the flask strongly for about 1 minute. Open the lid shortly for ventilating the headspace with fresh air. Close it again and shake the flask for 1 more minute. Insert the oxygen sensor and temperature sensor into the flask. Ensure that the tips of the sensors are immersed in the water. Otherwise, follow the calibration procedures given in chapter 7.
Please consider that streaming air through water may cause cooling of the water. Ensure correct temperature determination!
51
8.2 The 0% Standard
The 0% calibration standard can be
water mixed with a strong reductant water flushed with nitrogen gas (N2) nitrogen gas (N2)
8.2.1 Water Mixed with a Strong Reductant
Fill an appropriate amount of water into a glass flask (e.g. Duran flask) with a lid prepared with holes for inserting the oxygen sensor and a temperature sensor. Add a strong reductant, like sodium dithionite (Na2S2O4) or sodium sulfite (Na2SO3) at a concentration of 30 g/L, creating oxygen-free water by chemical reaction. It is not recommended to use saline water (e.g. seawater) for this, because the high salinity of the water might prevent a proper dissolution of the reductant. Stir the solution until the salt is completely dissolved and let the solution stand for about 15 minutes. Insert the oxygen sensor and temperature sensor into the flask, and ensure that the sensor tips are completely immersed into the water. Otherwise, follow the calibration procedures given in chapter 7.
Do not store the sensors in this solution and rinse them sufficiently with demineralized water after calibration.
8.2.2 Water Flushed with Nitrogen Gas
Fill water into a glass flask (e.g. Duran flask) with a lid prepared with holes for inserting the oxygen sensor and a temperature sensor. Close it and stream nitrogen gas through the water for about 10 minutes. You might speed up this process by first boiling the water (and thereby removing all dissolved gases) and then stream the nitrogen gas through it during cooling. Insert the oxygen and temperature sensor into the flask, let it equilibrate and perform calibration as described in chapter 7.
52
Please consider that streaming N2 gas through water may cause cooling of the water. Ensure correct temperature determination of the 0% calibration standard!
8.2.3 Nitrogen Gas
Flush 100% nitrogen gas through a glass flask (e.g. Duran flask) with a lid prepared with holes for inserting the oxygen sensor and a temperature sensor. Ensure that all air has been replaced by the nitrogen gas before performing calibration. Insert the oxygen and temperature sensor into the flask, let it equilibrate and perform calibration as described in chapter 7.
Ensure that no ambient air enters the flask again during the calibration process. Convectional gas transport is a very fast process! It is therefore advised to keep flushing the flask with nitrogen gas during the complete calibration process!
Please consider that nitrogen gas from gas bottles might be significantly cooled down by the decompression process. Ensure correct temperature determination of the calibration standard!
53
9 Calibration of Contactless Sensors
For preparing a setup with contactless oxygen sensors, please refer to the chapters 5.3-5.7. More details on oxygen nanoprobes in microfluidic applications are available on request.
9.1 Calibration Procedure
In general, the calibration procedure for contactless sensors (e.g. sensor spots, flow-through cells, respiration vials) is the same as for the robust and dipping probe described in chapters 7 and 8. However, if a 1-point or a 2-point calibration should be performed, the calibration standards have to be filled directly into the vessel containing the sensor spot, into the tubing of the flow-through cell or into the respiration vial.
If "Ambient Air" is used for the air calibration standard (see chapter
8.1.1), a good air circulation of the ambient air into the dry setup is important. For precision applications without the possibility to ensure a dry setup for the calibration procedure, the alternative air calibration standards "Water-Vapor Saturated Air" (see also chapter 8.1.2) or "Air Saturated Water" (see also chapter 8.1.3) should be preferred. In the first case, some part of the inner volume of the setup can be filled with e.g. wet cotton wool ensuring 100% RH around the oxygen sensor position. In the latter case, the inner volume of the setup is simply filled with air saturated water prepared as described in chapter 8.1.3. Ensure that the oxygen sensor is completely covered with air saturated water!
54
9.2 Manual Background Compensation
The calibration of contactless sensors (e.g. sensor spots, flow­through cells, respiration vials, nanoprobes) includes a compensation of potential background fluorescence from the fiber-optic cable connecting the Piccolo2 with the contactless oxygen sensor. Based on the Fiber Length (m) entered in the Settings (see chapters 6.2.1 and 6.2.2), a background signal for compensation is estimated automatically by the Pyro Oxygen
Logger software. So the user usually does not notice the
background compensation at all. For standard applications this should be the preferred procedure.
But for precision applications and especially for low signal intensities (e.g. <50 mV), a manual background compensation can be performed by the user alternatively. For this, Manual
Background Compensation must be selected in the Advanced Settings (see chapter 6.2.2). After opening the calibration window
by clicking on Calibrate, a separate Background Compensation window will open automatically:
Here, the background fluorescence of the connected optical fiber can be compensated. For this it is important that
55
the instrument end of the Optical Fiber is connected to the
Piccolo2 (see 5.7) and
the adapter end of the Optical Fiber is not attached to the
sensor (i.e. disconnect the adapter end from the spot adapter, adapter ring or from the flow-through cell)
Then, wait for steady-state and press the button Take Actual Values.
Alternatively, the button Keep Last Values can be used if the sensor spots are calibrated again with the same optical fiber, which was background compensated before. Then, the last values for background compensation are kept.
It is also possible to manually enter values for the Background and
dphi (°) into the field Manual and subsequently, clicking on Take Manual Values. If you manually enter zero for Background, no
background compensation is performed.
After background compensation is finished, the window closes and the program proceeds with the main oxygen sensor calibration window (see chapter 7). It is important that, for the subsequent calibration process, the Optical Fiber is attached to the sensor spot position again, e.g. by connecting the adapter end again to the spot adapter, adapter ring or flow-through cell.
Please ensure that during background compensation the Optical
Fiber is not connected to the contactless sensor.
Please ensure that during the subsequent calibration process the
Optical Fiber is attached again to the contactless sensor.
Keep in mind that the position of the spot adapter or adapter ring should not be changed after calibration of the sensor spot; otherwise it has to be calibrated again.
56
10 Appendix
10.1 Specifications of the Piccolo2
Dimensions
15.5 x 15.5 x 54 mm
Weight
ca. 20 g
Interface
USB 2.0
Power Supply
5VDC from USB-port, ca. 3-5mA average current consumption with max. 40mA peaks (of about 10­200ms duration) during an oxygen measurement
Supported operating systems
Windows 2000, XP, VISTA, 7,8, 10
(but not Windows RT)
Operating temperature
0 to 50ºC
Max. relative humidity
Non-condensing conditions
Oxygen channels
1
Oxygen measuring principle
lifetime detection of REDFLASH indicator luminescence
Excitation wavelength
620 nm (orange-red)
Detection wavelength
760 nm (NIR)
Max. sample rate
4 samples per second
Max. sample rate with enabled “high
speed sampling” in Settings->Options
ca. 10-20 samples per second
*Please note, that the oxygen sensors have different temperature ranges (typ. 0­50°C specified, -20°C to 70°C not specified).
57
10.2 Troubleshooting
How to respond to the warnings shown in the Pyro Oxygen Logger software:
Signal Too High
Too much ambient light exposed to the sensor, or amplification is too high, or LED intensity is too high:
darken the surrounding and / or decrease the Amplification in the Advanced settings
and / or decrease the LED Intensity in the Advanced settings
Low Signal
Sensor signal is too low:
check whether the sensor cable is connected increase the Amplification in the Advanced settings and / or increase the LED Intensity in the Advanced settings replace the sensor, the tip might be broken / bleached
Bad Reference
Internal problem of the electronics contact Pyro Science
58
10.3 Measuring Principle
The REDFLASH technology is based on the unique oxygen­sensitive REDFLASH indicator showing excellent brightness. The measuring principle is based on the quenching of the REDFLASH
indicator luminescence caused by collision between oxygen
molecules and the REDFLASH indicator immobilized on the sensor tip or surface. The REDFLASH indicators are excitable with red light (more precisely: orange-red at a wavelength of 610-630 nm) and show an oxygen-dependent luminescence in the near infrared (NIR, 760-790 nm).
The REDFLASH technology impresses by its fast response times, high precision, high reliability, low power consumption, and low cross-sensitivity. The red light excitation significantly reduces
59
interferences caused by autofluorescence and reduces stress in biological systems. The REDFLASH indicators show much higher luminescence brightness than competing products working with blue light excitation. Therefore, the duration of the red flash for a single oxygen measurement could be decreased from typically 100 ms to now typically 10 ms, significantly decreasing the light dose exposed to the measuring setup. Further, due to the excellent luminescence brightness of the REDFLASH indicator, the actual sensor matrix can be prepared much thinner now, leading to fast response times of the Pyro Science oxygen sensors.
The measuring principle is based on a sinusoidally modulated red excitation light. This results in a phase-shifted sinusoidally modulated emission in the NIR. The Piccolo2 measures this phase shift (termed "dphi" in the software). The phase shift is then converted into oxygen units based on the Stern-Vollmer-Theory.
60
10.4 Operating several Piccolo2 in parallel
The fiber-optic oxygen meter Piccolo2 offers one oxygen channel. In order to realize extendable multichannel systems with higher channel numbers, several Piccolo2 can be operated at a single PC as described in the following:
Connect each Piccolo2 to a free USB port of your PC. If the PC does not provide a sufficient number of USB ports, you can use an external USB-hub.
The Pyro Oxygen Logger software now has to be started separately for each connected
Piccolo2. So, if you want to operate e.g. 4 different Piccolo2, you
have to start the Pyro Oxygen Logger software 4 times, which will open 4 Pyro Oxygen Logger windows on your desktop. The different windows operate completely independent from each other, and each of them will be associated to exactly one Piccolo2. In order to check which window is associated to a specific Piccolo2, simply press the Flash Logo button in the main window of the Pyro
Oxygen Logger software, which causes a short flashing (1-2 sec.) of
the red status LED at the associated Piccolo2.
When closing the Pyro Oxygen Logger software, all settings and all current calibration data are saved in a "setup file", which is automatically loaded at the next startup. This setup file is saved specifically for each Piccolo2 serial number, i.e. each Piccolo2 in the setup described above keeps its own settings and calibration data.
61
10.5 Definition of Oxygen Units
phase shift dphi
The phase shift dphi is the fundamental unit measured by the optoelectronics in the Piccolo2 (see chapter 10.3). Please note, that dphi is not at all linearly dependent on the oxygen units, and increasing oxygen levels correspond to decreasing dphi values, and vice versa! As a rule of thumb, anoxic conditions will give about dphi=53, whereby ambient air will give about dphi=20.
raw value raw value
Definition: raw value = %O2 (uncalibrated)
The unit raw value is the default unit for uncalibrated sensors and shows qualitative oxygen sensor readings only.
partial pressure pO2 hPa = mbar
Used in: gas and water phase
For a calibrated sensor, the partial oxygen pressure pO2 in units of hPa (equivalent to mbar) is the fundamental oxygen unit measured by the Piccolo2.
partial pressure pO2 Torr
Definition: pO2[Torr] = pO2[hPa] x 759.96 / 1013.25 Used in: gas or water phase
volume percent pV %O2
Definition: pv = pO2[hPa] / p
atm
x 100%
Used in: gas
with p
atm
: actual barometric pressure
62
% air saturation A % air sat
Definition: A[% a.s.] = 100% x pO2 / p100O2 Used in: water phase
with p100O2 = 0.2095 ( p
atm
– p
H2O
(T) )
p
H2O
(T) = 6.112mbar x exp ( 17.62 T[°C] / (243.12 + T[°C])) pO2: actual partial pressure p
atm
: actual barometric pressure T: actual temperature p
H2O
(T): saturated water vapor pressure at temperature T
Dissolved O2 concentration C µmol/L
Definition: C [µmol/L] = A[% a.s.] / 100% x C
100
(T,P,S)
Used in: water phase
with C
100
(T,P,S): interpolation formula for dissolved oxygen concentration in units of µmol/L at temperature T, atmospheric pressure P and Salinity S (see chapter 10.6).
Dissolved O2 concentration C mg/L = ppm
Definition: C [mg/L] = C [µmol/L] x 32 / 1000 Used in: water phase
Dissolved O2 concentration C mL/L
Definition: C [mL/L] = C [µmol/L] x 0.02241 Used in: water phase
63
10.6 Table of Oxygen Solubility
The following table shows the equilibrium oxygen concentration C
100
(T, P=1013mbar, S) in units of µmol/L at standard atmospheric pressure of 1013 mbar as a function of water temperature in units of °C and salinity in units of PSU ("practical salinity unit" ≈ g/L). In order to correct these values for the actual atmospheric pressure p
atm
, the following formula has to be applied:
C
100
(T,P,S) = C
100
(T,P=1013mbar,S) x p
atm
/ 1013mbar
References: Garcia, HE and Gordon, LI (1992)
Oxygen solubility in seawater: Better fitting equations.
Limnol. Oceanogr. 37: 1307-1312
Millero, FJ and Poisson, A (1981)
International one-atmosphere equation of state of seawater.
Deep Sea Res. 28A: 625-629
64
Sal
(PSU)
Temp
0
(°C)
5
10 15 20 25 30 35 40
0
456.6
398.9
352.6
314.9
283.9
257.9
235.9
217.0
200.4
2
450.4
393.6
348.1
311.1
280.6
255.0
233.3
214.7
198.3
4
444.2
388.5
343.7
307.3
277.3
252.1
230.8
212.4
196.3
6
438.1
383.3
339.4
303.6
274.0
249.3
228.3
210.2
194.3
8
432.1
378.3
335.1
299.9
270.8
246.5
225.8
207.9
192.3
10
426.1
373.3
330.8
296.2
267.6
243.7
223.3
205.7
190.3
12
420.3
368.4
326.7
292.6
264.5
240.9
220.9
203.6
188.4
14
414.5
363.5
322.5
289.1
261.4
238.2
218.5
201.4
186.5
16
408.8
358.7
318.4
285.5
258.3
235.5
216.1
199.3
184.6
18
403.2
354.0
314.4
282.1
255.3
232.8
213.7
197.2
182.7
20
397.7
349.3
310.4
278.6
252.3
230.2
211.4
195.1
180.8
22
392.2
344.7
306.5
275.2
249.3
227.6
209.1
193.0
179.0
24
386.8
340.2
302.6
271.9
246.4
225.0
206.8
191.0
177.1
26
381.5
335.7
298.7
268.5
243.5
222.5
204.5
189.0
175.3
28
376.2
331.2
294.9
265.3
240.6
219.9
202.3
187.0
173.5
30
371.0
326.9
291.2
262.0
237.8
217.4
200.1
185.0
171.7
32
365.9
322.5
287.5
258.8
235.0
215.0
197.9
183.0
170.0
34
360.9
318.3
283.9
255.7
232.2
212.5
195.7
181.1
168.2
36
355.9
314.1
280.3
252.5
229.5
210.1
193.6
179.2
166.5
38
351.0
309.9
276.7
249.5
226.8
207.7
191.4
177.3
164.8
40
346.2
305.8
273.2
246.4
224.1
205.4
189.3
175.4
163.1
65
10.7 Explanation of the Sensor Code
The oxygen sensors are delivered with an attached sensor code which has to be entered in the Settings (refer to chapter 6.2). The following figure gives a short explanation about the information given in the sensor code.
Example Code: XB7-532-205
Sensor Type
LED Intensity
Amplification
Factory Calibration C0
Factory Calibration C100
Sensor Type
X Robust / Dipping Oxygen Probe (full range) T Oxygen Sensor Spot / FTC (trace range) S Oxygen Sensor Spot / FTC (full range) P Oxygen Nanoprobes
LED Intensity
A 10% E 40% B 15% F 60% C 20% G 80% D 30% H 100%
Amplification
4 40x 6 200x 5 80x 7 400x
66
C0 (Factory Calibration at 0% O2)
dphi0 = C0 / 10
C100 (Factory Calibration at 100% O2)
dphi100 = C100 / 10
The values of factory calibration are valid for the following calibration conditions:
Partial Volume of Oxygen (% O2) 20.95 Temperature at both calibration points (°C) 20.0 Air Pressure (mbar) 1013 Humidity (% RH) 0
Loading...