Pyromation Dual J-T-E-K User Manual

Page 1
THERMOCOUPLE THERMOMETER
Dual J-T-E-K
MODEL NO. 28-02800
5211 Industrial Road
Fort Wa yne, Indiana U. S.A. 46825
(260) 484-2580
(260) 482-6805 (Fax)
A1299-0689
Edition 06
Page 2
CERTIFICA TE OF CONFORMANCE
This thermometer was calibrated using equipment traceable to the National Institute of Standards and Technology (NIST).
This instrument conforms to NIST monograph 175 revised to ITS-90.
The accuracy of the thermometer at the time of calibration was within specifications stated in the operating manual.
Model No.:__________________________ Serial Number:______________________ Date placed in service:________________ T o purchase an NIST certificate of traceability
with test data and test date for meter and probe, please contact:
PYROMA TION
5211 Industrial Road
Fort Wa yne, Indiana U. S.A. 46825
260-484-2580
Page 3
INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements in Celsius, or Fahrenheit, using a wide range of thermocouple types. The temperature range for each type ther­mocouple is listed in the following chart.
TYPE °C °F
J –200 to 1000 –328 to 1832 K –250 to 1372 –418 to 2501 T –250 to 400 –418 to 752 E –250 to 1000 –418 to 1832
1
Page 4
The instrument is designed for easy operation and includes the following f eatures:
• Operator selection of temperature scale
• Resolution of 0.1° from 150°C to 999.9°C (238°F to 999.9°F)
• LCD with three four-digit displays
• T wo (2-blade f emale) ANSI mini-connector inputs
• Hold feature for temporarily retaining a reading
• T wo-point field calibration capability
• Low battery warning
• Stores up to 25 readings
• Scrolls through all stored readings
• Displays MIN and MAX readings
• Scrolls between T1, T2, and T1-T2 readings
• Built-in tilt stand for easy hands-free operation
2
Page 5
SAFETY PRECAUTIONS
WARNING
LOW LEVEL SIGNALS SUPPLIED BY STAN­DARD THERMOCOUPLES. UNDER NO CIR­CUMSTANCES SHOULD THE INPUT VOL TAGE EXCEED THE SPECIFIED 50V RMS.
CAUTION
MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS.
CAUTION
STRUMENT . DEAD BATTERIES CAN LEAK AND CAUSE DAMA GE T O UNIT.
DANGER
MA Y ALSO BE PRESENT A T THE BA TTER Y TER­MINALS. ALW A YS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
WARNING
PHERE, BA TTERIES MUST ONL Y BE CHANGED IN AN AREA KNOWN T O BE NON-HAZARDOUS.
A VERTISSEMENT
D’ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WARNING
PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
THIS INSTRUMENT IS DESIGNED TO ACCEPT
DO NOT USE OR STORE THIS INSTRUMENT IN
WEAK BA TTERIES SHOULD NOT BE LEFT IN THE IN-
VOLTAGES PRESENT AT THE THERMOCOUPLES
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L’INFLAMMATION
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
3
Page 6
SPECIFICATIONS
THERMOCOUPLE PROBES
Typ e Temperature Range
T ype J: –200°C to 1000°C (–328°F to 1832°F ) T ype K: –250°C to 1372°C (–418°F to 2501°F ) Type T: –250°C to 400°C (–418°F to 752°F ) T ype E: –250°C to 1000°C (–418°F to 1832°F )
Accuracy > –150°C: ±0.1% of reading, ±0.4°C
(±0.7°F)
Accuracy < –150°C: ±0.25% of reading, ±1°C
(±2°F)
Linearization: Conforms to NIST monograph 175
revised to ITS-90.
Input Protection: 50V rms Conversion Rate: Reading 0.6 seconds per
update.
Connector: T wo-b lade female ANSI mini-connector
inputs.
Battery
Size: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6
Life: 750 hours continuous, typical
Display: LCD with 0.4 in high characters main
readout and 0.2 in high characters secondary displays, 4 digits each display plus v arious an­nunciators.
4
Page 7
Temperature/Humidity Range Operating:
Stated Accuracy: 18°C to 28°C Useful Range: 0°C to 40°C Storage: 40°C to 65°C Humidity: 10% to 90%
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 grams (8 ounces) Ingress protection: Meets IEC-529 IP-54 for dust
and water resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically safe operation in hydrogen atmospheres per Class I, Division 1, Groups A, B, C, and D hazardous (classified) locations for UL per UL913 and CSA per C22.2 No. 0-M1982 and No. 157-M1987. Maximum surface temperature: 135°C (T4); UL file No. E182612 (1997).
Compliance (For CE Mark):
EN61326-1/A1: 1998 (EU EMC Directive)
(64°F to 82°F) (32°F to 104°F) (40°F to 149°F) (non-condensing)
5
Page 8
BATTERY INSTALLATION AND
REPLACEMENT
CAUTION
INSTRUMENT. DEAD BATTERIES CAN LEAK AND CAUSE D AMAGE TO UNIT.
DANGER
MA Y ALSO BE PRESENT A T THE BA TTER Y TER­MINALS. ALW A YS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
WARNING
PHERE, BA TTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN T O BE NON-HAZARDOUS.
AVERTISSEMENT
D’ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WEAK BA TTERIES SHOULD NOT BE LEFT IN THE
VOLTAGES PRESENT AT THE THERMOCOUPLES
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L’INFLAMMATION
6
Page 9
If battery indicator turns on, battery life is approxi­mately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
A T THIS POINT BA TTER Y MUST BE CHANGED . IF BATTERY VOLTAGE GOES TOO LOW, DIS­PLA Y WILL GO BLANK.
See SPECIFICA TIONS f or battery type.
1. Before changing battery , turn instrument off and disconnect thermocouple(s).
2. Loosen screw and lift battery cover off back of case.
3. Remove the two AA batteries.
4. Insert two new batteries observing polarity.
5. Install cover and tighten screw.
7
Page 10
CONNECTING A THERMOCOUPLE
Note: Be sure your instrument setting (J, T, E or K)
matches the thermocouple you are using. Use the correct thermocouple type for your instru-
ment. Using an incorrect thermocouple type will result in erroneous readings. Thermocouples are color coded by type using the North American ANSI color code as follows:
TYPE COLOR
J Black K Yellow T Blue E Purple
Thermocouple connectors have one wide blade and one narrow blade.
wards.
Connect thermocouple(s) to receptacle at
top of instrument.
Do not force connector in back-
If only one thermocouple is being used, it can be connected to either T1 or T2. The thermometer will automatically determine which connector is being used. When two thermocouples are used, connect one to the T1 connector and one to the T2 connector.
Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be reversed. The red wire is negative for thermocouple wires manufac­tured in North America.
8
Page 11
91011
Page 12
Page 13
Page 14
QUICK SETUP
Note: Review wa rnings on page 6.
1. Install batteries.
2. Connect thermocouple(s).
3. Press the ON/OFF key. The thermometer per­forms a self-test and all display digits and indi­cators, as shown below, should remain on for approximately one second.
4. Use J-T -E-K key to select the correct thermo­couple and the °C/°F key to select the desired scale.
If a thermocouple has not yet been connected to one or more inputs on the instrument, you will see this display:
This message also appears if a thermocouple is broken. No measurements can be made while this warning is displayed.
12
Page 15
COMPLETE SETUP PROCEDURE
The setup procedure is used to select the tempera­ture scale, resolution, and thermocouple type.
NOTE
Selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced.
SELECTING TEMPERA TURE SCALE
Select °C or °F by pressing the °C/ °F key. Each time the key is pressed the temperature scale will switch. Switching between °C and °F can be done at any time during operation.
Each time you turn the instrument on, it will power up with the same settings that were set when the unit was last turned off.
SELECTING RESOLUTION
Select 0.1° or 1° resolution by pressing the
0.1°/1° key . This key toggles between the two reso­lutions. When measuring temperature above 999.9° or below 150°C the thermometer automatically autoranges to 1° resolution.
SELECTING THERMOCOUPLE TYPE
Select the thermocouple type, either J, T, E, K, by pressing the J-T-E-K key. Each push of the key will step to the next type thermocouple. The se­lected thermocouple type is indicated by the an­nunciators in the lower right corner of the display .
Continue with the OPERATING PROCEDURES section or turn the instrument OFF by pressing the ON/OFF key again. The displa y will blank.
13
Page 16
OPERATING PROCEDURES
The unit will always power up with the upper dis­play showing T1 unless T1 is not connected. If only T2 is connected at power up, the upper dis­play shows T2 and the lower left displa y indicates T1 is open. If thermocouples are not connected to T1 or T2 at power up, the upper display indicates an open T1 and the lower right displa y indicates an open T2.
If you are in single probe mode and the probe is moved to the other connector or to restore the dual probe mode, turn power OFF then back ON.
For optimum operation, allow about one minute f or ambient temperature stabilization. If the unit has been stored at an extreme ambient condition, more time may be needed.
BASIC TEMPERA TURE MEASUREMENTS
Check that the thermometer is turned on, the probe(s) connected, the desired resolution 0.1° or 1° is selected, and the desired scale °C or °F, is selected.
Single Probe Measurements
The thermometer will automatically determine if T1 or T2 connection is being used.
Initially the upper display will indicate the measured temperature and the annunciator will indicate which probe is making the measurement. One of the lower displays will indicate open and the annunciator will indicate which probe is open. Pressing the T1/T2 key will cause the thermometer to go into the single probe differential mode described later.
T emperatures > 999.9° hav e a resolution of 1° and therefore no decimal will be present.
14
Page 17
Dual Probe Measurements
The thermometer will automatically determine if one or two thermocouples are connected. When two thermocouples are connected, the upper main dis­play will initially show T1 and the lower right dis­play will show T2. Pressing the T1/T2 k ey once will cause the upper display to show T2 and the lower left display to show T1. Pressing the T1/T2 key again will cause the upper display to show T1-T2, the lower left displa y to show T1, and the lower right display to show T2.
MAXIMUM READINGS
The maximum reading function displays the maxi­mum reading since power up or since the last time the clear function was used. The maxim um read­ing function is ideal for monitoring unattended op­erations while continually displaying e very tempera­ture change that occurs. The maximum and mini­mum values are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost. Factory Calibration will be maintained.
15
Page 18
Single Probe Measurements
The displayed information depends on whether you are using a single probe in the dual probe mode or have changed to the single probe differential mode by pressing the T1/T2 key.
If you are using a single probe and in the dual probe mode, momentarily press the MAX s key. The MAX annunciator turns on. If the probe is connected to T1 the maximum reading will be shown in the lower left display. If the probe is connected to T2 the maximum reading will be shown in the lower right display. If a higher maximum occurs while in the MAX mode, the higher reading will be displayed. To turn off the maximum reading, press the MAX s key again.
If you are in the single probe differential mode, momentarily press the MAX s key. The MAX and MIN annunciators turn on and both the maximum and minimum readings appear in the lower displays. The minimum will be on the left display and the maximum on the right display . If a higher maximum or lower minimum occurs while in the MAX mode, the higher reading will be displayed. T o turn off the maximum reading, press the MAX s key again.
16
Page 19
Dual Probe Measurements
Momentarily press the MAX s key. The MAX an­nunciator turns on, and both maximum T1 and T2 readings appear in the lower display . The T1 read­ing is on the left display and the T2 reading is on the right display . Press MAX s key again to cancel.
Clearing a Maximum Reading
Press the CLEAR key then the MAX s key. The maximum memory will be cleared.
MINIMUM READINGS
The minimum reading function displays the mini­mum reading since power up or since the last time the clear function was used. While continually dis­playing ev ery temperature change that occurs, the maximum and minimum values are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost.
17
Page 20
Single Probe Measurements
The displayed information depends on whether you are using a single probe in the dual probe mode or have changed to the single probe differential mode by pressing the T1/T2 key.
If you are using a single probe and in the dual probe mode, momentarily press the MIN t k ey . The MIN annunciator turns on. If the probe is connected to T1 the minimum reading will be shown in the lower left display . If the probe is connected to T2 the mini­mum reading will be shown in the lower right dis­play. If a lower minimum occurs while in the MIN mode, the lower reading will be displayed. To turn off the minimum reading, press the MIN t key again.
If you are in the single probe differential mode, momentarily press the MIN t key. The MAX and MIN annunciator turns on and both the maximum and minimum readings appear in the lower displays. The minimum will be on the left display and the maximum on the right display . If a lower minim um or higher maximum occurs while in the MIN mode, the new readings will be displayed. To turn off the minimum reading, press the MIN t key again.
Dual Probe Measurements
Momentarily press the MIN t key. The MIN an­nunciator turns on, and both minimum T1 and T2 readings appear in the lower display . The T1 read­ing is on the left display and the T2 reading is on the right display. Press the MIN t key again to cancel.
Clearing a Minimum Reading
Press the CLEAR key then press the MIN t key . The minimum memory will be cleared.
18
Page 21
DIFFERENTIAL READINGS (T1-T2)
The differential function is used to compare two different measurements and display the diff erence between them. Diff erential measurements can be made using either one probe or two probes.
Single Probe Measurements
For single probe measurements, one measurement is stored as the reference measurement, then each following measurement is compared to the refer­ence. The reference measurement is viewed on the lower left display , the diff erential value is shown in the lower right display and the main display shows the present T1 or T2 reading.
1. Place probe at first measurement point.
2. Press the T1/T2 key to store this ref erence tem­perature. The ∆T and M= annunciators will be on and the reference temperature will be dis­played on the lower left display. The T indi­cates that the differential mode is selected and the M= indicates that a reference temperature has been stored. Any time after this that the T1/ T2 key is pressed, the M= value will be updated to the present value shown in the main display.
3. Place the probe at the second measurement point. The main display will immediately show the new measured temperature and the lower right display will immediately show the difference between the reference temperature and the present temperature point.
19
Page 22
Dual Probe Measurement
1. Connect the two probes to the two points of measurement. The lower left displa y will indicate the T1 temperature, the lower right display will indicate the T2 temperature.
2. The main display can be scrolled between dis­playing T1, T2 or the differential temperature T1­T2 by repeatedly pressing the T1/T2 key . T o dis­play the differential temperature press the T1/T2 key until the T1-T2 annunciator to the right of the main display is on.
HOLD FUNCTION
Press the HOLD key to retain the reading on the display. Press HOLD key again for normal operation.
Maximum/Minimum Hold Readings
Press the MAX s key, then press the HOLD
key . T o turn off the HOLD function and return to
normal operation, press the HOLD key again. T o clear the maximum readings, press the CLEAR
key, then the MAX s key.
Minimum/Hold
Press the MIN t ke y , then press the HOLD key. T o turn off the HOLD function and return to normal operation press the HOLD key again.
T o clear the minimum readings, press the CLEAR key, then the MIN t key.
STORED READINGS
The store function allows you to store up to 25 read­ings. Both T1 and T2 readings are stored. To store readings proceed as follows:
20
Page 23
1. Momentarily press the STORE ke y . Both T1 and T2 are stored. The upper main displa y will mo­mentarily show the storage location number and the STO annunciator will be on.
After three seconds the storage number will be re­placed with the temperature reading but the STO annunciator remains on to indicate a temperature reading has been stored. The STORE ke y may be pressed as fast as once per second.
2. Repeat step 1 for all the points to be recorded up to a maximum of 25. Each time the ST ORE key is pressed the new reading will be stored and the upper main display will show the stor­age location number for about 3 seconds. After 25 stored readings, the next time the STORE key is pressed the main display will indicate FULL.
RECALL READINGS
This function allows the stored readings to be re­called. RECALL shows each stored reading indi­vidually . T o recall readings proceed as f ollows:
1. Momentar ily press the RECALL key, the first stored reading will be displayed on the lower dis­play and the store number “1” will be displa y ed on the upper display. The STO, and RCL an­nunciators will be on.
2. To step through the readings, press the MIN t key or the MAX s key. Each press of the key will advance to the next reading in se­quence.
3. To return to normal operation press any other key except ON/OFF, MAX s, MIN t or RE-
CALL.
Clearing Stored Readings
Press the CLEAR key, then the STORE key. The stored readings in memory will be cleared.
21
Page 24
CALIBRATION
The calibration function allows both single point and dual point calibration of the thermometer. Single point calibrates the offset only . Dual point calibrates the offset first then calibrates the slope. The ther­mometer can be calibrated at any temperature. When two probes are used, a match calibration matches the T1 and T2 offsets.
It is not necessary to perform a field calibration to obtain specified accuracies. Use the calibration feature to improve thermometer/probe accuracy or to compensate for thermocouple probe calibration drift.
The thermometer has a memory retention capabil­ity to hold calibration values even while power is off or the battery is removed.
When you restart, there is no need to recalibrate. Each of the four thermocouple types may be indi­vidually calibrated and stored.
NOTE
The temperature at which water boils (100°C/ 212°F) is at sea level and standard atmospheric pressure using distilled water. Changes in altitude and barometric pressure will cause the boiling tem­perature to change.
As a general rule, the boiling temperature of water will decrease 1°C (1.8°F) for ev ery 1000 feet above sea level. For e xample, at an elev ation of 5300 feet (1600 meters), water will boil at approximately
94.7°C (202.5°F).
Other liquids may also be used as calibration stan­dards. Consult a chemical handbook f or their freez­ing (melting) point and boiling point properties.
22
Page 25
When calibrating at freezing (0°C or 32°F) it is rec­ommended to use crushed ice made of distilled water in a dewar flask. Add crushed ice to top of flask. Top off flask with distilled water. Continue to add crushed ice to maintain tightly packed crushed ice/water in flask.
CALIBRA TION PROCEDURES
Calibration Procedure (One Probe Detected)
This calibration function provides for both an offset and slope field calibration of either T1 or T2. If only one probe is detected, the procedure applies to that probe, and the main display will indicate T1 or T2 as applicable. For proper calibration the f ollowing conditions must be observed:
• The slope point must be a higher temperature than
the offset point.
• The difference must be at least 20°C.
• Use two points based on the expected high and
low temperatures. Temperatures measured out­side of these limits may no longer meet accuracy specifications.
Proceed as follows to calibrate the probes.
1. Place the probe at the lower known reference temperature.
2. Offset Calibration: Momentarily press the CAL key to enter the calibration mode, the CAL an­nunciator will flash. The temperature is display ed on the main display and Lo is displayed on the lower left display. “Lo” signifies that this is the offset point.
23
Page 26
3. Allow the reading to stabilize. If the displayed temperature is higher or lower than the reference temperature, use the MAX s key to increase the displayed reading or the MIN t key to de­crease the displayed reading until the ref erence temperature is displayed. The MIN t or MAX s key must be pressed at least once. The CAL annunciator should be blinking during this procedure.
4. Press the HOLD key to lock the offset cali­bration in and advance to the slope calibration. T o return to normal operation press any key e x­cept CAL or HOLD .
5. Slope Calibration: Place the probe at the higher reference temperature.
6. Allow the reading to stabilize. If the displayed temperature is higher or lower than the reference temperature, use the MAX s key to increase the displayed reading or the MIN t key to de­crease the displayed reading until the ref erence temperature is displayed. The MIN t or MAX s key must be pressed at least once. The CAL annunciator should be blinking during this pro­cedure.
7. Press the HOLD key to lock the calibration in.
24
Page 27
Calibration Procedure (Tw o Probes Detected)
This calibration function provides for both an offset and slope field calibration of either or both T1 and T2. F or proper calibration the following conditions must be observed:
• The slope point must be a higher temperature than the offset point.
• The difference must be at least 20°C.
• Use two points based on the expected high and low temperatures. Temperatures measured out­side of these limits may no longer meet accuracy specifications.
Proceed as follows to calibrate the probes.
1. Place the probe at the lower known reference
temperature.
2. Offset Calibration: Momentarily press the CAL
key to enter the calibration mode, the CAL an­nunciator will flash. The temperature is display ed on the main display and Lo is displayed on the lower left display. “Lo” signifies that this is the offset point.
3. Allow the reading to stabilize. If the displayed
temperature is higher or lower than the reference temperature, use the MAX s key to increase the displayed reading or the MIN t key to de­crease the displayed reading until the ref erence temperature is displayed. The MIN t or MAX s key must be pressed at least once. The CAL annunciator should be blinking during this procedure.
4. Press the HOLD key to lock the T1 calibr a-
tion in and advance to T2 offset calibration, or press the CAL key to skip T1 calibration. In either case you will now be in T2 offset calibr ation.
25
Page 28
5. Allow the reading to stabilize. If the display ed temperature is higher or lower than the refer­ence temperature, use the MAX s key to in­crease the displayed reading or the MIN t key to decrease the displa yed reading until the reference temperature is displayed.
The MIN t or MAX s key must be pressed at least once. The CAL annunciator should
remain blinking during this procedure.
6. Press the HOLD key to lock the T2 offset calibration in, or press the CAL key to skip T2 offset calibration and proceed with T1 slope.
If T1 or T2 offset calibration is skipped, then the slope calibration for that input will also be skipped.
7. Slope Calibration: Place the probe(s) at the higher reference temperature.
8. The temperature is displayed on the main dis­play and "Hi" is displayed on the lo wer left dis­play .
9. Allow the reading to stabilize. If the display ed temperature is higher or lower than the refer­ence temperature, use the MAX s key to increase the displayed reading or the MIN t ke y to decrease the displayed reading until the reference temperature is displayed.
The MIN t or MAX s key must be pressed at least once. The CAL annunciator should
remain blinking during this procedure.
10. Press the HOLD key to lock the T1 slope calibration in and advance to T2 slope calibra­tion, or press the CAL key to skip T1 slope calibration. In either case y ou will now be in T2 slope calibration.
26
Page 29
11. Press the HOLD key to lock the T2 slope calibration in, or press the CAL key to skip T2 slope calibration. In either case you will now be in T1 = T2 offset calibration. To return to normal operation press any key except CAL key or HOLD key.
If either T1 or T2 offset or slope were modified, CAL 1 or 2 will be lit.
12. The MATCH annunciator should be blinking. When both readings (T1 and T2) are stable, press the HOLD key to lock the T1 = T2 offset calibration in. The MATCH annunciator will remain lit.
13. This completes the two-point calibration.
14. To return to normal operation press any key except ON/OFF key or the CAL key.
Matching T1 and T2 Procedure
This procedure requires both probes to be at the same temperature.
1. Place both probes at the same temperature and
let stabilize.
2. Momentarily press the CAL key until the MA TCH
annunciator is blinking. When both readings (T1 and T2) are stable, press the HOLD key to lock in the calibration.
Clearing an Individual Cal Point: Press CAL key until the desired point is displayed, then press
CLEAR key . Clearing all Cal Points: Press CLEAR key, then
CAL key. Unit reverts to factory calibration with no
offset or slope compensation.
27
Page 30
FIELD CALIBRATION LOCK OUT AND RE-ENABLE
The calibration lockout feature , prev ents any field calibration changes. The loc kout remains in eff ect until a lockout re-enable has been performed. Use the following procedures to lockout or re-enab le the field calibration operation.
Lockout Procedure
1. Turn the thermometer off.
2. Simultaneously press and hold the CAL and the
CLEAR keys down and momentarily press the ON/OFF key. Continue to hold the CAL and CLEAR keys for at least 5 seconds.
Re-Enable Procedure
1. Turn the thermometer off.
2. Simultaneously press and hold the HOLD and
the CAL k eys down and momentarily press the
ON/OFF key . Continue to hold the HOLD and CAL ke ys until the display blanks .
28
Page 31
MAINTENANCE AND
TROUBLESHOOTING
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and chang­ing of the batteries.
WARNING
PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents. Use mild detergents, never immerse nor use e xcessive fluid.
BA TTERIES
If there is no display when the thermometer is turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary, ref er to the BA TTER Y INSTALLATION AND REPLACEMENT section for replacement procedure.
All stored readings are retained until cleared, even if the batteries are removed for long periods.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS-
29
Page 32
SERVICE
WARNING
INTRINSIC SAFETY .
A VERTISSEMENT
PROMETTRE LA SECURITE INTRINSEQUE.
There are no internal adjustments or user replace­able parts.
If “Err” f ollowed b y the numbers 1 through 9 is dis­played (see example belo w) return unit for service. Note that “Err” alone ma y be displa yed during im­proper field calibration.
Note: Serial number label is located inside battery compartment.
SUBSTITUTION OF COM­PONENTS MAY IMPAIR
LA SUBSTITUTION DE COMPOSANTS PEUT COM-
30
Page 33
TROUBLESHOOTING
The following chart lists the most probable faults. There are no internal adjustments or user replace­able parts. If this does not solve the prob lem refer service to your dealer.
F AULT ACTION
No display when Check condition of turned on. batteries.
Check that batteries are inserted properly.
Display shows Out of range indication.
- - - ­Display shows Open thermocouple
OPEn connection. Display shows If displayed at any time
Err other than during field
calibration, return instrument for service.
If Err 1 through Err 9 remains on the display , return instrument for service.
31
A-1925-34
Rev. G
Printed in U.S.A.
071000
Page 34
Page 35
WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifi­cations. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Re­pair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instruc­tions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. F or your protection, pack the product carefully and in­sure it against possible damage or loss. Any dam­ages resulting from improper packaging are your responsibility .
TECHNICAL ASSISTANCE
If you have any questions about the use of this prod­uct, contact the Manufacturer or authorized seller.
Trademarks bearing the ® symbol in this publication are reg­istered in the U.S. and in other countries.
A-3477-60
Rev. F
Printed in U.S.A.
061000
Loading...