Pyromation 28-02700-T User Manual

OPERATING MANUAL
Thermocouple Thermometer Models
28-02700-J (Type J) 28-02700-K (Type K) 28-02700-T (Type T)
5211 Industrial Road
Fort Wa yne, Indiana U .S.A. 46825
(260) 484-2580
(260) 482-6805 (Fax)
Edition 05
CERTIFICATE OF
CONFORMANCE
This thermometer was calibrated using equipment traceable to the National Institute of Standards and Technology (NIST).
This instrument conforms to NIST monograph 175 revised to ITS-90.
The accuracy of the thermometer at the time of calibration was within specifications stated in the operating manual.
Model No.:_________________________ Serial Number:______________________ Date placed in service:________________ T o purchase an NIST certificate of traceability
with test data and test date for meter and probe, please contact your dealer or:
PYROMA TION
5211 Industrial Road
Fort Wa yne, Indiana U. S.A. 46825
260-484-2580
INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements. The instru­ment is designed for easy operation and includes the following features:
• Operator selection of Celsius or Fahrenheit scale
• Resolution of 0.1° from –99.9 to 299.9°
• Four-digit LCD
• Hold feature for temporarily retaining a reading
• Field calibration capability
• Low battery warning
• Built-in tilt stand for easy hands-free operation
• T wo blade f emale ANSI mini-connector input
• Operates with a wide selection of probes
1
SAFETY PRECAUTIONS
WARNING
LOW LEVEL SIGNALS SUPPLIED BY STAN­DARD THERMOCOUPLES. UNDER NO CIR­CUMSTANCES SHOULD THE INPUT V OLT A GE EXCEED THE SPECIFIED 50V RMS.
CAUTION
MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS.
CAUTION
STRUMENT . DEAD BATTERIES CAN LEAK AND CAUSE D AMAGE T O UNIT.
DANGER
ALSO BE PRESENT AT THE BATTERY TERMI­NALS. ALWAYS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
WARNING
PHERE, BA TTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN TO BE NON-HAZARDOUS.
A VERTISSEMENT
D'ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WARNING
PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
THIS INSTRUMENT IS DESIGNED TO ACCEPT
DO NOT USE OR STORE THIS INSTRUMENT IN
WEAK BA TTERIES SHOULD NOT BE LEFT IN THE IN-
VOLTAGES PRESENT AT THE THERMOCOUPLE MAY
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L'INFLAMMATION
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
2
SPECIFICATIONS
THERMOCOUPLE PROBES
Typ e Temperature Range
T ype J: –200°C to 1000°C (–328°F to 1832°F ) T ype K: –250°C to 1372°C (–418°F to 2501°F ) T ype T: –250°C to 400°C (–418°F to 752°F )
Out of range display: - - - ­Resolution: 0.1°/1° autoranging, 0.1° from
–99.9° to 299.9°, 1° outside this range.
Accuracy
> –99.9°: +(0.2% of reading +0.5°C)
< –99.9°: +(0.25% of reading, +1°C)
Display: 4-digit LCD with 0.5 in high numerals. Display update rate: 0.6 sec per update. Input: One thermocouple with ANSI connector. Input Protection: 50V rms Battery: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6.
Battery Life: 750 hours continuous, typical. Low battery indication
Battery Symbol on when 8 to 10 hours of battery life remains. Battery symbol blinking: replace battery
Operating Conditions
Stated accuracy: 18°C to 28°C (64°F to 82°F) Useful range: 0°C to 40°C (32°F to 104°F) Storage: –40°C to 65°C (–40°F to 149°F) Humidity (non-condensing): 10% to 90%
+(0.2% of reading +0.9°F)
+(0.25% of reading +2°F)
3
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 g rams (8 ounces) Ingress protection: Meets IEC-529 IP-54 for dust
and water-resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically safe operation in hydrogen atmospheres per Class I, Division 1, Groups A, B, C and D hazardous (classified) locations for UL per UL913 and CSA per C22.2 No. 0-M1982 and No. 157-M1987. Maximum surface temperature: 135°C (T4); UL file No. E182612 (1997).
Compliance: (For CE Mark)
EN61326-1/A1: 1998 (EU EMC Directive)
4
BATTERY INSTALLATION AND
REPLACEMENT
CAUTION
INSTRUMENT. DEAD BATTERIES CAN LEAK AND CAUSE D AMAGE TO UNIT .
DANGER
ALSO BE PRESENT AT THE BATTERY TERMI­NALS. ALWAYS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
WARNING
PHERE, BA TTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN TO BE NON-HAZARDOUS.
AVERTISSEMENT
D'ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WEAK BA TTERIES SHOULD NOT BE LEFT IN THE
VOLTAGES PRESENT AT THE THERMOCOUPLE MAY
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L'INFLAMMATION
5
If battery indicator turns on, battery life is appro xi­mately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
AT THIS POINT, BA TTER Y MUST BE CHANGED. IF BATTERY VOLTAGE GOES TOO LOW, DIS­PLA Y WILL GO BLANK.
See SPECIFICA TIONS for battery type.
1. Before changing battery , turn instrument off and disconnect thermocouple.
2. Loosen screw and lift battery cover off back of case.
3. Remove the two AA batteries.
4. Insert two new batteries observing polarity.
5. Install cover and tighten screw.
6
CONNECTING A THERMOCOUPLE
Use the correct thermocouple type for your instru­ment. Using an incorrect thermocouple type will result in erroneous readings. Thermocouples are color coded by type using the North American ANSI Color Code as follows:
TYPE COLOR
J Black K Yellow T Blue
Thermocouple connectors have one wide blade and one narrow blade.
wards.
Connect thermocouple to receptacle at top
of instrument as shown in the following illustration.
Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be reversed. The red wire is negative for thermocouple wires manufac­tured in North America.
Do not force connector in back-
7
8910
OPERA TION
Note: Be sure the thermocouple you are using
matches the instrument setting indicated on the unit.
1. Press the ON/OFF key. The thermometer per­forms a self-test and all display digits and indi­cators, as shown below, should remain on for approximately two seconds.
2. If a thermocouple is not connected or if the ther­mocouple is defective, the display will indicate OPEn.
3. Select °C or °F by pressing the °C/°F key . Tem­perature scale can be switched at any time. The thermometer remembers the last selected scale the next time you turn the power on.
4. For optimum instrument accuracy, allow one minute for ambient temperature stabilization. If the unit has been stored at an extreme ambient condition, more time will be needed.
Thermocouples are sensitive at the tip or measur­ing junction. When taking temperature measure­ments, allow time for the reading to stabilize . Mul­tiplying the time constant of the probe by 5 will give you the approximate time required.
Note that the readings will automatically change to 1° resolution below –99.9° or above 299.9°.
If desired, press the HOLD key to retain the read­ing on the display . Press HOLD key again for nor­mal operation.
If the display shows "- - - -" the temperature read­ing is out of range of the instrument.
FIELD CALIBRATION
The thermometer is factory calibrated and does not require calibration before use. The CAL function allows single point calibration of the thermometer, at 0°C (32°F) to compensate for thermocouple off­set error. It is not necessary to perform a field cali­bration to obtain specified meter accuracy . Use the field calibration feature to improve thermometer/ probe accuracy or to compensate for thermocouple drift.
1. Follow steps 1 through 4 in OPERATION sec­tion.
2. Pack sensing end of probe in a container tightly packed with crushed ice and filled with distilled water. Allow temperature to stabilize.
3. Press and hold the CAL key for 5 seconds to enter the calibration mode, the CAL annunciator on the display starts blinking. Release the CAL key .
4. If the measured temperature is from −10°C to 10°C (14 to 50°F), when the temperature read­ing is stable, press the CAL key .
The CAL indicator will stop blinking and the read­ing will be set to 0°C (32°F). The CAL indicator will remain turned on, indicating a field calibra­tion is active.
If “Err” is displayed, either the displa yed reading is outside the above limits or the batteries are weak.
CANCELING A FIELD CALIBRA TION
1. Turn the thermometer off.
2. Hold the CAL key down while pressing the ON/OFF key.
The field calibration is canceled and the thermom­eter reverts to the default factory calibration. The CAL annunciator is now turned off.
11
FIELD CALIBRATION LOCK OUT AND RE-ENABLE
The calibration lockout feature, pre v ents any field calibration changes. The loc kout remains in effect until a lockout re-enable has been performed. Use the following procedures to lockout or re-enab le the field calibration operation.
Lockout Procedure
1. Turn the thermometer off.
2. Simultaneously press and hold the CAL and the
°C/°F keys down and momentarily press the ON/ OFF key. Continue to hold the CAL and °C/°F
keys for at least 5 seconds.
Re-Enable Procedure
1. Turn the thermometer off.
2. Simultaneously press and hold the HOLD and the CAL keys do wn and momentarily press the
ON/OFF key. Continue to hold the HOLD and CAL keys until the display b lanks.
12
MAINTENANCE
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and chang­ing of the batteries.
CLEANING
WARNING
SPHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents. Use mild detergents, never immerse nor use e xcessive fluid.
BA TTERIES
If there is no display when the thermometer is turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary, ref er to the BA TTER Y INSTALLA TION AND REPLACEMENT section for replacement procedure.
TO PREVENT IGNITION OF A HAZARDOUS ATMO-
13
SERVICE
WARNING
SUBSTITUTION OF COM­PONENTS MAY IMPAIR
INTRINSIC SAFETY .
A VERTISSEMENT
LA SUBSTITUTION DE COMPOSANTS PEUT COM-
PROMETTRE LA SECURITE INTRINSEQUE.
There are no internal adjustments or user replaceable parts.
If “Err” followed by the numbers 1 through 9 is displayed (see example below) retur n unit for service. Note that “Err” alone ma y be displa yed during improper field calibration.
Note: Serial number label is located inside battery compartment.
14
A-1925-32
Rev. E
Printed in U.S.A.
050798
15
WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifi­cations. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Re­pair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the W arranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instruc­tions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. F or your protection, pack the product carefully and in­sure it against possible damage or loss. Any dam­ages resulting from improper packaging are your responsibility .
TECHNICAL ASSISTANCE
If you have any questions about the use of this prod­uct, contact the Manufacturer or authorized seller.
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
A-3477-57
Rev. E
Printed in U.S.A.
051000
Loading...