Pulsar Measurement Ultra Twin User guide

Page 1
Ultra Twin
Instruction Manual
Page 2
2
Page 3
PULSAR MEASUREMENT
3
ULTRA TWIN (FIFTH EDITION REV 1)
May 2021 Part Number M-192-0-005-1P
COPYRIGHT
© Pulsar Measurement, 2005 -21. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any form without the written permission of Pulsar Process Measurement Limited.
WARRANTY AND LIABILITY
Pulsar Measurement guarantee for a period of 2 years from the date of delivery that it will either exchange or repair any part of this product returned to Pulsar Process Measurement Limited if it is found to be defective in material or workmanship, subject to the defect not being due to unfair wear and tear, misuse, modification or alteration, accident, misapplication or negligence.
DISCLAIMER
Pulsar Measurement neither gives nor implies any process guarantee for this product and shall have no liability in respect of any loss, injury or damage whatsoever arising out of the application or use of any product or circuit described herein.
Every effort has been made to ensure accuracy of this documentation, but Pulsar Measurement cannot be held liable for any errors.
Pulsar Measurement operates a policy of constant development and improvement and reserves the right to amend technical details, as necessary.
The Ultra Twin shown on the cover of this manual is used for illustrative purposes only and may not be representative of the actual Ultra Twin supplied.
CONTACT For technical support, please contact: Europe: supporteurope@pulsarmeasurement.com
Outside Europe: supportnorthamerica@pulsarmeasurement.com If you have any comments or suggestions about this product, please contact: Europe: europe@pulsarmeasurement.com
Outside Europe: northamerica@pulsarmeasurement.com Pulsar Measurement website: www.pulsarmeasurement.com
United States
11451 Belcher Road South Largo, FL 33773 888-473-9546
Canada
16456 Sixsmith Drive Long Sault, Ont. K0C 1P0 855-300-9151
United Kingdom
Cardinal Building, Enigma Commercial Centre
Sandy’s Road, Malvern
WR14 1JJ 00 44 (0)1684 891371
Page 4
4
CONTENTS
Chapter 1: Start Here… ........................................................................................................ 10
About this Manual ............................................................................................................ 10
About the Ultra Twin ....................................................................................................... 11
Functional Description .................................................................................................... 11
Product Specification ....................................................................................................... 13
EU Certificate of Conformity ......................................................................................... 15
Chapter 2 Ultra Twin Installation ..................................................................................... 16
Unpacking ............................................................................................................................ 16
Power Supply Requirements......................................................................................... 16
Location ................................................................................................................................ 17
Dimensions .......................................................................................................................... 18
Terminal connection details .......................................................................................... 22
Power ..................................................................................................................................... 23
Transducer ........................................................................................................................... 23
ATEX ....................................................................................................................................... 23
FM ........................................................................................................................................... 24
Voltage selector and fuse location............................................................................. 26
Preparation for Operation ............................................................................................. 27
Maintenance ....................................................................................................................... 28
Chapter 3 How To Use Your Ultra Twin ........................................................................ 29
Operating the Controls ................................................................................................... 29
Run Mode ............................................................................................................................. 31
Program Mode ................................................................................................................... 31
Points of Measurement (P1 and P2) .......................................................................... 31
How to Access Program Mode .................................................................................... 32
Hot Keys ................................................................................................................................ 33
Menu Keys ........................................................................................................................... 34
Numeric Keys ...................................................................................................................... 34
Using the Menu System ................................................................................................. 35
Directly Editing Parameters ........................................................................................... 36
Test Mode ............................................................................................................................ 37
Page 5
PULSAR MEASUREMENT
5
Using the Serial Interface ............................................................................................... 38
Parameter Defaults ........................................................................................................... 40
Chapter 4 Quick Setup Guide ........................................................................................... 41
Enter Program Mode ....................................................................................................... 41
Choose Quick Setup ......................................................................................................... 41
Level or Volume ................................................................................................................. 42
Example 1 Level Monitoring with Alarms ................................................................ 48
Example 2 Level Monitoring and Control (up or down) .................................... 50
Example 3 Volume Application .................................................................................... 53
Example 4 Differential Control ..................................................................................... 56
Pump ...................................................................................................................................... 60
Quick Setup ......................................................................................................................... 61
Example 2 Sump Control (pump down) ................................................................... 66
Example 3 Reservoir Control (pump Up) ................................................................. 68
Flow ........................................................................................................................................ 70
Exponential Devices ......................................................................................................... 77
Calculations ......................................................................................................................... 80
BS3680 Flumes (P700 = 2) ............................................................................................. 84
Rectangular Flume Calculations .................................................................................. 85
U-Throated Flume Calculations ................................................................................... 86
Example 2 BS3680 U-Throated Flume ...................................................................... 87
BS3680 Weirs (P700 = 3) ................................................................................................ 89
Rectangular Weir Calculations ..................................................................................... 89
V-Notch Weir Calculations ............................................................................................ 90
BS3680 Rectangular Broad Crested Weir ................................................................ 93
Rectangular Broad Crested Weir Calculations ....................................................... 93
Special Devices ................................................................................................................... 95
Palmer Bowlus and H-Flume Calculations .............................................................. 96
V-Notch Angle Weir (Non-BS 3680) Calculations ................................................ 96
Universal Calculations (P700=6) .................................................................................. 97
Page 6
ULTRA TWIN INSTRUCTION MANUAL
6
Chapter 5 Parameter Guide ............................................................................................... 98
Menu System ...................................................................................................................... 98
Top Level Menu ................................................................................................................. 98
Application Menu .............................................................................................................. 99
Relays Menu ...................................................................................................................... 100
Pump “Advanced” Menu .............................................................................................. 101
Digital Inputs Menu........................................................................................................ 102
Data Logs Menu .............................................................................................................. 103
Volume (When P100 = 6)............................................................................................. 104
OCM Menu (When P100 = 4 or 5) ........................................................................... 105
Display Menu .................................................................................................................... 106
mA Output 1 Menu ........................................................................................................ 107
mA Output 2 Menu ........................................................................................................ 107
Compensation Menu ..................................................................................................... 108
Stability Menu .................................................................................................................. 108
Echo Processing Menu .................................................................................................. 109
System Menu .................................................................................................................... 110
Device Comm Menu ...................................................................................................... 111
Test Menu .......................................................................................................................... 112
Application Parameters................................................................................................. 113
Operation ........................................................................................................................... 113
Operation ........................................................................................................................... 113
Dimensions ........................................................................................................................ 115
Relay Parameters ............................................................................................................. 117
Alarms .................................................................................................................................. 119
Pumps .................................................................................................................................. 126
Control ................................................................................................................................. 130
Miscellaneous ................................................................................................................... 134
Common Relay parameters......................................................................................... 136
Pump “Advanced” Parameters ................................................................................... 137
Pump Run On ................................................................................................................... 137
Page 7
PULSAR MEASUREMENT
7
Starting ................................................................................................................................ 137
Stopping ............................................................................................................................. 138
Pump Exercising .............................................................................................................. 138
Wall Cling ........................................................................................................................... 139
Digital Inputs ..................................................................................................................... 140
Digital Input Parameters .............................................................................................. 146
Digital Inputs ..................................................................................................................... 148
Digital Input ....................................................................................................................... 150
Data Log parameters ..................................................................................................... 151
Totaliser Audits ................................................................................................................ 151
Temperature ...................................................................................................................... 151
Pump Logs ......................................................................................................................... 152
Volume ................................................................................................................................ 153
Conversion ......................................................................................................................... 154
Breakpoints ........................................................................................................................ 159
Tables ................................................................................................................................... 160
OCM ..................................................................................................................................... 161
PMD Setup ......................................................................................................................... 161
Dimensions ........................................................................................................................ 164
Calculations ....................................................................................................................... 168
Breakpoints ........................................................................................................................ 168
Tables ................................................................................................................................... 169
Average Flow .................................................................................................................... 169
Display Parameters ......................................................................................................... 170
Options ................................................................................................................................ 170
Failsafe ................................................................................................................................. 172
Auxiliary ............................................................................................................................... 173
Totaliser ............................................................................................................................... 176
Bargraph ............................................................................................................................. 178
mA Output 1 Parameters ............................................................................................. 179
Page 8
ULTRA TWIN INSTRUCTION MANUAL
8
Range ................................................................................................................................... 179
Operation ........................................................................................................................... 180
Setpoint ............................................................................................................................... 180
mA1 Limits ......................................................................................................................... 181
mA1 Trim ............................................................................................................................ 181
mA1 Failsafe ...................................................................................................................... 182
mA1 Allocation ................................................................................................................. 182
mA Output2 Parameters .............................................................................................. 183
Range ................................................................................................................................... 183
Operation ........................................................................................................................... 184
Setpoint ............................................................................................................................... 184
mA2 Limits ......................................................................................................................... 185
mA2 Trim ............................................................................................................................ 185
mA2 Failsafe ...................................................................................................................... 186
mA2 Allocation ................................................................................................................. 186
Compensation Parameters .......................................................................................... 188
Offset.................................................................................................................................... 188
Temperature ...................................................................................................................... 188
Velocity ................................................................................................................................ 189
Stability Parameters ....................................................................................................... 190
Damping ............................................................................................................................. 190
Indicator .............................................................................................................................. 190
Rate ....................................................................................................................................... 190
Filters .................................................................................................................................... 191
Echo Processing Parameters ....................................................................................... 192
Transducer 1 Status ........................................................................................................ 192
Transducer 2 Status ........................................................................................................ 193
DATEM Parameters......................................................................................................... 194
System Parameters ......................................................................................................... 195
Backup ................................................................................................................................. 195
System Information ........................................................................................................ 196
Page 9
PULSAR MEASUREMENT
9
Date & Time ...................................................................................................................... 197
LED Colour ......................................................................................................................... 197
Watchdog ........................................................................................................................... 198
Daylight Saving Time ..................................................................................................... 199
Device Comm. .................................................................................................................. 203
RS232 Set Up .................................................................................................................... 203
RS 485 Set Up ................................................................................................................... 203
Remote Alarm ................................................................................................................... 204
Test Parameters ............................................................................................................... 206
Simulation .......................................................................................................................... 206
Hardware ............................................................................................................................ 207
Chapter 6 Troubleshooting .............................................................................................. 209
Chapter 7 Disposal .............................................................................................................. 210
Notes ........................................................................................................................................ 211
Page 10
ULTRA TWIN INSTRUCTION MANUAL
10
CHAPTER 1: START HERE…
Congratulations on your purchase of a Pulsar Ultra Twin. This quality system has been developed over many years and represents the latest in high technology ultrasonic level measurement and control.
It has been designed to give you years of trouble-free performance, and a few minutes spent reading this operating manual will ensure that your installation is as simple as possible.
About this Manual
It is important that this manual is referred to for correct installation and operation. There are various parts of the manual that offer additional help
or information as shown.
Tips
TIP : Look f or th i s icon through out yo u r Puls a r Mea s u rement manual to fin d help f u l inf o r matio n and a n swers to freq uently a sked q uestio n s .
Additional Information
References
See Also
References to other parts of the manual.
Additional Information
At various parts of the manual, you will find sections like this that explain specific things in more detail.
Page 11
PULSAR MEASUREMENT
11
About the Ultra Twin
Ultra Twin has two independent points of measurement, the wall mount model provides a dedicated display to each point of measurement, whilst the fascia model, whilst in RUN, will show detail of one point of measurement in the main display line, with the second point being displayed on the auxiliary display line. In both models, the display will provide information relevant to the point of measurement selected whilst in RUN and PROGRAM mode
Ultra Twin combines premium specification with high performance in a most versatile system which is quickly configurable offering a choice of applications in any combination, between the two points of measurement of three specific applications i.e., level or volume measurement, pump control or flow measurement.
Functional Description
Ultra Twin sends a transmit pulse to the transducer(s), which emits an ultrasonic pulse perpendicular to the transducer face, and the returned echo is sent back to the Ultra Twin. The time taken to receive the echo is measured and the distance from the transducer face to the surface being monitored is calculated.
Ultra Twin can measure from zero to 40m from the transducer to the surface being monitored, dependent on the application chosen and transducer used.
Six user-definable relays can be programmed to activate alarms, pump starters, or other control equipment, and can be allocated to either point of measurement. Also provided are four user definable digital inputs on the wall mount model and seven on the fascia mount model, which can be allocated to either point of measurement.
Page 12
ULTRA TWIN INSTRUCTION MANUAL
12
There is an isolated 4-20 mA output for each point of measurement that can be connected to a recorder or PLC, to monitor level space, distance, volume, OCM head or flow (dependant on the application chosen), independently from that shown on the display. There is an RS232 port, so that the Ultra Twin can be operated remotely by a PC or other equipment.
Ultra Twin can be programmed either by the built-in keypad (standard), or by PC via the RS 232 Serial Interface (optional).
All parameters are stored in non-volatile memory, so are retained in the event of power interruption. A second backup copy of all parameters can also be retained in the Ultra Twin memory, in case an alternative set of parameters needs to be stored.
The system utilises the unique DATEM software (Digital Adaptive Tracking of Echo Movement). This is a proven digital mapping technique developed especially for the Pulsar Ultra range, which gives the system unequalled
ability when identifying the “true target level” in the face of competing
echoes from pipes, pumps or other obstructions. Coupled with the powerful, long-range abilities of the ‘all new’ dB transducer range, the Ultra Twin lives up to its reputation as the most reliable ultrasonic level measurement system available.
The Pulsar Ultra Twin ultrasonic level controller has been designed to provide maintenance-free fit and forget performance.
The Ultra Twin can show level, space, distance, on the display. The relays can be programmed to activate alarms, pump starters, or other control equipment. In addition, the digital inputs can be used to modify pump and control regimes to optimise performance. There is an isolated 4-20 mA output that can be connected to a chart recorder or PLC, to monitor level, space or distance, independently from that shown on the display. There is an RS232 port, so that the Ultra Twin can be operated remotely by a PC or other equipment.
The Ultra Twin is programmed by the built-in keypad or by PC via the RS 232 Serial Interface (optional). All the parameters are stored in non-volatile memory, so are retained in the event of power interruption. A second backup copy of all parameters can also be retained in the Ultra Twin, in case a previous set of parameters needs to be restored.
Page 13
PULSAR MEASUREMENT
13
Product Specification
PHYSICAL
Wall Mount Outside dimensions
235 x 184 x 120 mm
Weight
Nominal 1 kg
Enclosure material/description
Polycarbonate, flame resistant to UL94-5V
Cable entry detail
10 cable entry knock outs, 1 x M16, 5 x M20 underside, 4 x 18mm (PG11) at rear
Fascia Outside dimensions
200 x 112 x 108 mm
Weight
Nominal 1.3kg
Enclosure material/description
Stainless steel back, Polycarbonate UL94-V0 front and bezel
Transducer cable extensions
2-core screened
Maximum separation
1000m (500m for mmWAVE)
ENVIRONMENTAL
IP Rating (Wall)
IP65
IP Rating (Fascia)
IP64
Max. & min. temperature (electronics)
-20 ºC to +50 ºC
Flammable atmosphere approval
Safe area: compatible with approved dB transducers (see transducer spec' sheet)
CE Approval
See EU Declaration of Conformity
PERFORMANCE
Accuracy
0.25% of the measured range or 6 mm (whichever is greater). mmWAVE ± 2mm.
Resolution
0.1% of the measured range or 2 mm (whichever is greater)
Max. Range
Dependant on transducer (maximum 40m dB40)
Min. Range
Dependent upon application and transducer (minimum zero dB Mach3)
Rate Response
Fully adjustable
ECHO PROCESSING
Description
DATEM (Digital Adaptive Tracking of Echo Movement)
OUTPUTS
Analogue Outputs
Isolated (floating) output (to 150V) of 4-20 mA or 0-20 mA into 500 (user programmable and adjustable)
0.1% resolution
Page 14
ULTRA TWIN INSTRUCTION MANUAL
14
Digital output
Full Duplex RS232
Volt free contacts, number, and rating
6 form "C" (SPDT) rated at 5A at 240V AC
DIGITAL INPUTS
Wall Mount x 4 Fascia Mount x 7
Min. Input Voltage 4.5VDC. Max. Input Voltage 30VDC (Max Current 3mA). 24VDC Input Supply maximum total current 24mA.
DISPLAY
Wall Mount x 2 Fascia Mount x 1
6 digits plus 12-character text, plus bar graph with direction indicators, remote communicator identifier, and program/run/test mode indicators.
PROGRAMMING
On-board programming
By integral keypad
PC programming
Via RS232
Programming security
Via passcode (user selectable and adjustable)
Programmed data integrity
Via non-volatile RAM, plus backup
SUPPLY
Power Supply
115V AC + 5% / -10% 50/60 Hz, 230V AC + 5% / -10% 50/60 Hz, DC 18 - 36V 10W maximum power (typically 6W)
Fuses
100mA at 230V AC (fitted as standard to wall units) 200mA at 115V AC (fitted as standard to fascia units)
Pulsar Measurement operates a policy of constant development and improvement and reserve the right to amend technical details, as necessary.
Page 15
PULSAR MEASUREMENT
15
EU Certificate of Conformity
Page 16
ULTRA TWIN INSTRUCTION MANUAL
16
CHAPTER 2 ULTRA TWIN INSTALLATION
Unpacking
Power Supply Requirements
The Ultra Twin can operate from AC supply or from a DC battery. The AC is 115V +5%/-10% 50/60Hz or 230V +5%/-10% 50/60Hz, depending on the position of the selector switch. The DC is 18-36V. In all cases the Ultra Twin will typically consume 6W of power, with a maximum of 10W.
Important Information
All shipping cartons should be opened carefully. When using a box cutter, do not plunge the blade deeply into the box, as it could potentially cut or scratch equipment components. Carefully remove equipment from each carton, checking it against the packing list before discarding any packing material. If there is any shortage or obvious shipping damage to the equipment, report it immediately to Pulsar Process Measurement Limited.
Page 17
PULSAR MEASUREMENT
17
Location
The Ultra Twin must be mounted in a non-hazardous (safe) area, and the transducer fitted in the hazardous area.
When choosing a location to mount the enclosure, bear in mind the following:
Ensure that the Ultra Twin is installed in a “Safe”, non-hazardous
area.
For a clear view of the LCD display, it is recommended that it is
mounted at eye level.
The mounting surface is to be vibration free.
The ambient temperature is between -20°C and 50°C.
There should be no high voltage cables or inverters nearby,
Important Information
All electronic products are susceptible to electrostatic shock, so follow proper grounding procedures during installation.
Page 18
ULTRA TWIN INSTRUCTION MANUAL
18
Dimensions
Wall Mount
The dimensions of the wall fixing holes are as shown below:
The Ultra Twin should be mounted by drilling three holes suitable for size 8 screws (length to suit your application), and fixing the top screw in place. Hang the unit on this and fix the two remaining screws by removing the terminals access cover to access the pre-drilled holes.
30mm (1.18")
142mm
(5.59")
198mm (7.80inches)
Page 19
PULSAR MEASUREMENT
19
The full dimensions of the Wall enclosure are as shown below:
24.5mm
24.5mm
48mm
(0.96")
(1.89")
(0.96")
212mm (8.35 inches)
235mm (9.25 inches)
153mm (6.02 inches)
184mm (7.24 inches)
120mm (4.72 inches)
27mm
28mm
28mm
28mm
30mm
(1.06")
(1.10")
(1.10")
(1.10")
(1.18")
Page 20
ULTRA TWIN INSTRUCTION MANUAL
20
Cable Entry
There are 6 cable gland knockouts on the base of the wall mount Ultra Twin (5 x M20, 1 x M16) and 4 on the rear (4 x 18mm dia (PG11)). Select which ones you wish to take out, and remove them by using a circular cutter, such as a tank cutter. Take care not to damage the circuit board inside whilst undertaking this. Do not use a hammer, as this may cause damage to the enclosure.
It is recommended that you use suitable cable glands to ensure that the ingress rating is maintained and that they be tightened to the manufacturers recommended settings.
Fascia Mount
The dimensions of the wall fixing holes are as shown below:
Page 21
PULSAR MEASUREMENT
21
The full dimensions of the fascia mount enclosure are as shown below:
162 mm (6.37”)
20mm
(0.78”)
72mm (2.83”)
15mm (0.62”)
16mm (0.63”)
200 mm (7.87“)
112 mm
(4.41”)
Page 22
ULTRA TWIN INSTRUCTION MANUAL
22
Terminal connection details
Wall Mount
The terminal strip is as detailed as below. There is also a wiring diagram inside the terminals access cover.
Fascia Mount
The terminal details are as illustrated below:
Page 23
PULSAR MEASUREMENT
23
Power
The Ultra Twin can operate from mains AC and automatically from DC or battery backup in the event of power failure or can be operated permanently from DC or batteries.
Transducer
The transducer should be installed, and connected, in accordance with the installation instructions contained in the Transducer User Guide. Wire the transducer to the Ultra Twin’s transducer terminals, as follows:
Transducer 1
TERMINAL CONNECTION DETAILS
Unit Type
Red:
Power
White: Signal
Black:
0Volts
Green: Screen
Wall Mount
45
46
47
47
Fascia Mount
32
31
30
30
Transducer 2
TERMINAL CONNECTION DETAILS
Unit Type
Red:
Power
White:
Signal
Black:
0Volts
Green: Screen
Wall Mount
48
49
50
50
Fascia Mount
35
34
33
33
When using 2-core screened extension cable, the Black and Green wires of the transducer should be connected to the screen of the extension cable, which in turn should be connected to the appropriate 0 volts terminal of the Ultra Twin.
ATEX
For EEx m (Zone 1) applications a transducer certified to Sira 02ATEX5104X is used, and must be supplied via a 4000A breaking fuse,
which is fitted as standard to the Ultra Twin level controller. For EEx ia (Zone 0) a transducer certified to Sira 02ATEX2103X is used,
which must be connected to the Ultra Twin via an external Zener barrier.
Page 24
ULTRA TWIN INSTRUCTION MANUAL
24
FM
For EEx m (Zone 1) applications a transducer certified to FM Class I Div 1 Group A, B, C & D, ClassII Div 1 Group E, F & G, Class III is used, and must
be supplied via a 1500A breaking fuse, which is fitted as standard to the Ultra Twin level controller.
Restrictions do not use in the presence of these groups of Chemicals, Aliphatic Hydrocarbons, Ketones or Esters
For EEx ia (I.S.) a transducer certified to FM Class I Div 1 Group A, B, C & D, ClassII Div 1 Group E, F & G is used, which must be connected to the Ultra Twin via an external Zener barrier.
See transducer label for certification details.
Relay Outputs
The five relays can be programmed for a variety of alarms, pump control, or other process functions. The relay contacts are all rated at 5A at 240V AC. All connections should be such that the short circuit capacity of the circuits to which they are connected, is limited by fuses rated so that they do not exceed the relay rating.
Current Output
This is an isolated (floating) mA output (to 150 V), of 4 - 20mA or 0 - 20mA, and the load should not exceed 500 .
Current Input (Optional)
This feature is available as an option only. Please consult Pulsar for further details. The current input is an isolated (floating) mA input (to 150 V), 4 ­20mA or 0 -20mA.
Important Information
When using the Ultra Twin to measure the differential level between the two points of measurement then transducer one must be located on the upstream side of the application.
Page 25
PULSAR MEASUREMENT
25
Temperature Input (Optional)
The external temperature sensor allows more localised compensation of the measured distance due to changes in temperature.
There are two models, Type A and Type B as follows:
TYPE
RANGE
A
-25° to 50°C
B
-25° to 125°C
The temperature sensor should be placed close to the point of measurement.
The Temperature sensor connections are as follows:
DESCRIPTION
TEMPERATURE
SENSOR
ULTRA TWIN
TERMINAL
Power Supply
Terminal 1
Terminal 27
Return
Terminal 2
Terminal 28
Temp Source (P1-852, P2-852), should be set to option 4 or 5 depending on the sensor range, set 4 for type A and 5 for type B (see above), the range is specified on the label of the sensor.
92mm
182mm
100mm
M20 ENTRY
IP67 ABS HOUSING
316 STAINLESS STEEL
6mm
Page 26
ULTRA TWIN INSTRUCTION MANUAL
26
RS2323 Serial Interface
If required, you can connect to the serial interface to operate your Ultra Twin remotely.
Voltage selector and fuse location
Wall Mount
The voltage selector switch and mains fuse is located inside the terminal compartment, to the left of the mains terminals, as illustrated below:
Fascia Mount
The voltage selector switch and mains fuse, is located under the removable cover at the bottom of the unit, as illustrated below.
230
115
AC POWER FUSE
Page 27
PULSAR MEASUREMENT
27
DON’T FORGET
Make sure you move the voltage selector switch to the correct position for your power supply.
Preparation for Operation
Before switching on, check the following:
The Ultra Twin is mounted correctly and is in a ‘safe’ area. The power supply is correctly installed. The voltage selector switch is in the correct position. The relays are connected correctly.
Important Information
If the equipment is installed or used in a manner not specified in this manual, then the protection provided by the equipment may be impaired.
Important Information
Before applying AC power (mains), make sure you have correctly selected the voltage selector switch, as detailed in the preceding pages.
Please note that all Fascia units are set to 115 volts AC with a 200mA fuse fitted, and all Wall units are supplied set to 230 volts AC for safety reasons, and a 100mA fuse fitted as standard.
Never operate the Ultra Twin with terminal access exposed. An external switch or circuit breaker should be installed near to the Ultra
Twin to allow the supply to be removed during installation and maintenance. In addition, the relay contacts should also have a means of isolating them from the Ultra Twin.
Interconnecting cables must be adequately insulated in accordance with local regulations. Strip back 30 mm of the outer insulation of the cable. Strip 5 mm of insulation from the end of each conductor. Twist all exposed strands of the conductor together. Insert the stripped conductor into the terminal block as far as it will go and tighten the terminal block screw. Ensure that all strands are firmly clamped in the terminal block and that there is no excess bare conductor showing, and no stray strands.
Page 28
ULTRA TWIN INSTRUCTION MANUAL
28
Maintenance
There are no user serviceable parts inside Ultra Twin, except the mains fuse. If you experience any problems with the unit, then please contact Pulsar Measurement for advice.
To clean the equipment, wipe with a damp cloth. Do not use any solvents on the enclosure.
Important Information
The unique DATEM software comes into operation as soon as power is applied and is designed to monitor a moving level or target with the transducer in a fixed position.
If, after any period of use, it should become necessary to move the transducer, for any reason, from its original operating position, switch off the Ultra Twin, before proceeding, in order to prevent any undesirable updates to the DATEM trace. If after moving the transducer the reading is not as expected, please refer to Chapter 6 Troubleshooting.
Important Information
Please note that the on-board Lithium battery, mounted to the processor PCB, is not user serviceable.
Page 29
PULSAR MEASUREMENT
29
CHAPTER 3 HOW TO USE YOUR ULTRA TWIN
Operating the Controls
Display
On the wall mount model, there are two identical displays, by default, the top display will provide information on the current mode of operation, and status of the remote communication for point 1 (transducer 1), while the bottom display provides the same information for point 2 (transducer 2).
In the case of the fascia model, while in the RUN mode, the single display will provide information for point 1 (transducer 1) on the main display line and the auxiliary display provides the same information for point 2 (transducer 2).
When in Program Mode the display of the fascia mount model can be “toggled” by pressing to change from point 1 to point 2 to access
parameters on each point. While in the Run Mode the displays will show, the current level reading and
its units of measure, along with status messages with regards to the Transducer, Echo reception and Fail-Safe Mode. Additionally, they can be programmed independently to provide status messages on alarms, pumps etc.
When in the Program mode the display is used to read information on the Menu System, Point of Measurement, Parameter Number and parameter details and values, which can be entered.
During Test Mode, the display is used to monitor the simulated level. A bar graph is also provided which will provide a visual reading of the level, in percentage of span.
Page 30
ULTRA TWIN INSTRUCTION MANUAL
30
1. Displays the current mode of operation.
2. Main 6-digit display:
Run Mode; current measurement displayed, dependent on mode and measurement units chosen, and value of hotkey selected. Program Mode; displays parameter number and values entered for parameters. Test Mode; displays simulated level.
3. Auxiliary Display, scrolling twelve-digit display.
Run Mode; displays measurement units, status messages on signal and transducer, details of Hot Key function selected. It can be also programmed to provide notification messages on alarms and pumps etc. For full details please refer to Display Parameters in the relevant parameter listing. Program Mode; displays menu and sub menu headings, parameter details and options.
4. Communicator status, this displays the status of remote PC
connection and velocity sensor operation.
5. Bar graph display, this gives visual indication of measurement in %
of span.
6. Level indicators:
Run Mode; indicates in which direction the level is moving. Program Mode: indicates at which level of the menu system you are
at.
1 2 3
4
0%
5
6
Run Mode Program Mode Test Mode 100%
REMOTE COMMUNICATOR OFF
XXXXXXXXXXXX
000.000
Page 31
PULSAR MEASUREMENT
31
There are two main operating modes for your Ultra Twin, Run Mode and Program Mode. There is also a Test Mode, used for checking the set-up. All modes are now described.
Run Mode
This mode is used once the Ultra Twin has been set up in program mode. It is also the default mode that the unit reverts to when it resumes operation after a power failure.
When the Ultra Twin is switched on for the first time, it will display, in metres, the distance from the transducer face to the target. All relays by default are switched off.
After programming is complete, any relays that are set will operate when the level reaches the relevant setpoint, and the LED’s will change colour (unless specifically switched off).
Program Mode
This mode is used to set up the Ultra Twin or change information already set. You must use either the built-in keypad (standard) or, alternatively the unit can be set up with a PC via the RS232 Serial Interface.
Entering a value for each of the parameters that are relevant to your application provides all the programming information.
Points of Measurement (P1 and P2)
All menu options are the same for Point 1 and Point 2. Measurement points can be “toggled” by pressing to change from point 1 to point 2
to access parameters on each point.
On the Fascia unit the LED will light up green on Relay 1 for Point 1, and Relay 2 for Point 2.
Page 32
ULTRA TWIN INSTRUCTION MANUAL
32
How to Access Program Mode
To enter program mode on the Ultra Twin, you simply enter the passcode, via the keypad, followed by the ENTER key. The default passcode is 1997, so you would press the following:
Once you have entered the program mode the Ultra Twin will automatically access point 1 menu system, and the top display will show “Program Mode”
in the Mode Status Line and “Quick Setup” in the Auxiliary Display Line, in
the case of the wall mount model, the bottom display, point 2, will be blank.
To change from one point to the other point’s menu system press the hot key, whilst in any Main Menu heading, e.g. Quick Setup, Application etc. and you will toggle between the two points and their relevant menu systems
Important Information
There is a time-out period of 15 minutes when in program mode. After which time the run mode will resumed if you do not press any key.
ENTER
1
9 9 7
Page 33
PULSAR MEASUREMENT
33
Hot Keys
There are five hot keys on the keypad, which can be used to quickly access common parameters for viewing only, while in Run Mode. Pressing the hot key once will display the first parameter, then repeated pressing will display the others, then the Ultra Twin reverts to Run Mode. In program mode, they have different functions, the functions are shown below.
HOT KEY
RUN MODE
PROGRAM MODE
When application is Flow, view non­resettable totaliser(s). View and reset the resettable totaliser(s). When application is Pump, view information on total pump running hours, and individual pump running hours.
Not used with Ultra Twin
Displays echo confidence, echo strength, height above loss limit (HALL), average noise, peak noise, and temperature
Not used with Ultra Twin
Total number of pump starts, and individual pump starts.
Reset parameter to default setting
Instantaneous mA output
Not used with Ultra Twin
Dependent on application. Displays distance, level, space, volume or rate of change of level.
*Toggle between Point 1 & 2 Main Menu System. When programming relays toggle relay setpoints between Ultra Twin’s units of measure and % of span.
Not used with Ultra Twin
Takes you to the last parameter edited when you first enter program mode.
Shows details of function type, firmware revision and serial number
Enter decimal point.
*When using a Fascia mount unit and this hotkey is pressed, the relay light will alternate from Relay 1 to Relay 2 indicating the change between Point 1 & 2.
mA
n
Page 34
ULTRA TWIN INSTRUCTION MANUAL
34
Menu Keys
The menu keys have the following functions:
HOT KEY
FUNCTION
1) Arrow keys for moving left and right around the menu
system.
2) Used in test mode to simulate the level moving up and
down.
1) Used to confirm each action (e.g. select a menu option)
2) Used to confirm questions asked by the Ultra Twin, such as
before restring factory defaults.
Used to navigate up a level in the menu system, and back to run mode.
Used to cancel a value entered in error
Numeric Keys
These keys are used for entering numerical information during programming.
CANCEL
ENTER
1
2
3 4 5
6
7
8 9 0
.
+
-
Page 35
PULSAR MEASUREMENT
35
There are two means of editing parameters, directly or using the menu system. Each is now described.
Using the Menu System
The menu system has been designed to make the changing of parameters very simple. There are two levels of menu: Main Menu and Sub Menu.
Main Menu
The main or top menu is common to both points of measurement and when you first access the program mode your Ultra Twin will display the menu system for point 1. To change form one point to point 2 menu system, press the hot key, whilst in any Main Menu heading, e.g. Quick Setup, Application etc. and you will toggle between the two points and their relevant menu systems.
Sub Menu
Any sub-menu and the parameters contained in it relating to Point 1 (Transducer 1) is pre-fixed P1, sub-menus and parameters relating to Point 2 (Transducer 2) are pre-fixed P2. Menus and parameters which are common to both Points (both transducers) are pre-fixed P* e.g., P*104 Measurement Units.
On the display, there is a line of text that displays the menu system. Pressing the arrow keys scrolls the display between the top-level menu items, (as shown below, starting at Quick Setup).
As you press the cursor keys to scroll left and right between these, you can press ENTER at any time, to select the desired menu heading, and take you to the sub-menu.
Quick Setup
Application
Relays
Digital Inputs
Data Logs
mA 2
Output
Display
OCM
P100 = 4 or 5
Compensation
System
Test
Device
Comm
Echo
Process
Volume
P100 = 8 or 9
mA 1
Output
Pump
Advance
Stability
Page 36
ULTRA TWIN INSTRUCTION MANUAL
36
Each of these options, along with their sub-menus are described later in this manual. When you move down into the sub-menu, you can scroll round using the arrow keys press ENTER to go to the required section of parameters. Once you have reached the relevant section, scroll through the parameters, and enter the necessary information. To enter the information, use the numeric keys and then press ENTER, you will then see the message
Saved!” If you press CANCEL, then the change you made will not be saved, and the message “Unchanged!!” will be displayed.
When you have finished, press CANCEL to go back to the previous level. When you have reached the top level, then the Ultra Twin will ask for confirmation before allowing you to go back into run mode. This is done by pressing ENTER at the display prompt.
Directly Editing Parameters
If you already know the number of the parameter, that you wish to look at or edit, simply type the number in at any time while you are in the menu system. Thus, if you are in either the menu or sub-menu level by pressing a numeric key, you can enter the parameter number directly and jump straight there. You cannot type a parameter number whilst at parameter level, only at one of the two menu levels.
When you are at a parameter, the text line rotates automatically displaying the parameter name, number, the applicable units, and the maximum and minimum figure you can enter. The top line shows the value you are setting.
Once you have accessed a parameter, you can either just look at it, or change it.
Important Information
You can tell which part of the menu system you are in, as the up/down level indicators, (arrows) next to the bar graph will indicate as follows:
Top level menu: Down arrow on, to indicate you can move down.
Sub-menu: Up and Down arrows on, to indicate you can move up to
the top level, and down to parameter level.
Parameter Level: Up arrow on, to indicate you can move up to sub-
menu level.
Parameter Editing: No arrows on.
Page 37
PULSAR MEASUREMENT
37
Once a parameter has been changed, press ENTER and you will see the message Saved!”. If you press ‘CANCEL, then the change you made will not
be saved, and the message “Unchanged!!” will be displayed.
You can jump straight to the last parameter you edited, by pressing ‘+/- when you first enter program mode.
Test Mode
Test mode is used to simulate the application and confirm that all parameters and relay setpoints have been entered as expected. During simulation, there is a choice of whether the relays will change state (hard simulation) or not (soft simulation), but the LED’s will always change colour as programmed, and the mA output will change in accordance to the chosen mode of operation. If you wish to test the logic of the system that the relays
are connected to then select hard simulation, but if you do not wish to change the relay state, then select a soft simulation.
There are two simulation modes, automatic and manual. Automatic simulation will move the level up and down between empty level or the pre­determined Start Level (P*983) and Pump/Control relay switch points, if you wish to change the direction of the level movement e.g., to go beyond relay setpoints, this can be done by using the arrow keys. In manual simulation, using the arrow keys will allow you to move the level up and down as required.
To enter simulation, first go to program mode. Using the menu system, select menu item ‘Test’, then sub-menu item ‘P1 or P2 Simulation’. Simply change the value of the parameter P*980 to one of the following:
1= Manual soft simulation 2= Automatic soft simulation 3= Manual hard simulation 4= Automatic hard simulation
To return to program mode, press CANCEL and test mode will end. When in manual simulation, by default test mode will move the level by
0.1m steps. Altering the increment (P*981) will change this value.
Page 38
ULTRA TWIN INSTRUCTION MANUAL
38
In automatic mode, the rate at which the level moves up and down is set by the increment (P981 in metres, and the rate (P*982) in minutes, which can be changed to make the level move up and down faster. E.g. if increment (P*981) is set for 0.1m and rate (P*982) is set to 1 min then the level will increase or decrease at a rate of 0.1m/min. To make the simulated level move slower, decrease the value in increment (P*981) or increase the value in rate (P*982). To make the simulated level move faster, increase the value in increment (P*981) or decrease the value in rate (P*982).
Using the Serial Interface
The RS232 serial interface is used to communicate between the Ultra Twin and a PC using the optional Ultra PC and other associated Pulsar software packages, to obtain information such as data logging and view echo traces upload, download and save parameter files. In addition, it can also be used to control or obtain information using a standard PC or other computer base equipment. To do so, the settings for control are as follows: baud rate 19,200, 8 data bits, no parity, 1 stop bit.
The device should be connected as shown in Chapter 2 Ultra
Twin Installation.
To use the device remotely, you need to log on to start, and log off when finished. When logged on, the Ultra Twin will show ‘Remote ON’ on the display, and “Communicator OFF” when logged off.
All commands should be followed by a carriage return. When logged on, the unit will respond either OK (or a value) if the command is accepted, or NO if it is not.
To log on, send the command /ACCESS:pppp where pppp is the passcode (P922). To log off, send the command /ACCESS:OFF To read a parameter value, send the command /Pxxx where xxx is the parameter you wish to read, and the Ultra Twin will
respond with the parameter value. To set a parameter, send the command /Pxxx:yy where xxx is the parameter number, and yy is the value you wish to
set it to.
Page 39
PULSAR MEASUREMENT
39
Other commands you can use are:
/DISTANCE (shows current distance)
/LEVEL (shows current level)
/SPACE (shows current space)
/RATE (shows current rate)
/VOLUME (shows current volume)
/TEMP (shows current temperature)
/CURRENTOUT1 (show the mA output 1 value)
/CURRENTOUT2 (show the mA output 2 value)
/BACKUP1 (take backup of parameters to area 1)
/BACKUP2 (take backup of parameters to area 2)
/RESTORE1 (restore parameters from area 1)
/RESTORE2 (restore parameters from area 2)
Please consult Pulsar Measurement or contact your local Pulsar representative for further details and a full list of available commands.
Page 40
ULTRA TWIN INSTRUCTION MANUAL
40
Parameter Defaults
When you first switch the Ultra Twin on, it will be reading the distance from the face of the transducer to the surface. It will be indicating in metres, as shown on the display. All relays are set OFF.
The date (P*931) and time (P*932) in the Ultra Twin were set at the factory, but may need checking, and amending if, for example the application is in a time zone other than GMT, see relevant parameter listing for full details.
In some applications, it is easier to empty the vessel, take a reading from the Ultra Twin for distance and then setup the empty level to this figure.
Once you are satisfied with the installation, and Ultra Twin is reading what you would expect in terms of distance from the face of the transducer to the material level, then you can proceed with programming, for the intended application. It is sensible to program all the required parameters at the same time. The system will be then set-up.
Note: The span is automatically calculated from the empty level, so the empty level should be entered first.
Factory Defaults
When first installing the Ultra Twin, or subsequently moving or using the unit on a new application, before proceeding to program the unit for its intended application it is recommended that you ensure that all parameters are at their default values by completing a Factory Default P930, as described in the relevant unit type parameter guide.
Page 41
PULSAR MEASUREMENT
41
CHAPTER 4 QUICK SETUP GUIDE
This quick set-up guide shows you how to get up and running in a few minutes in just four easy steps after installing your Ultra Twin.
Enter Program Mode
First you need to go from run mode into program mode. Assuming the passcode is the default 1997, then you should enter the following on the keypad:
Choose Quick Setup
Now you need to go into the quick setup. You will see on the menu the words ‘Quick Setup’, which is the first item on the menu system. By default, the Ultra Twin will always access point 1 menu system, to change to point 2 menu, press the hot key. Try pressing the two arrow keys to see some more menu options, but return to Quick Setup, and press:
1 = Level or Volume (Level/Vol) 2 = Pump Control (Pump) 3 = Open Channel Flow measurement (OCM Flow)
APPLICATION
UNIT TYPE
1 = Level/Vol
When selected the Ultra Twin will be configured for Level/Volume measurement.
2 = Pump
When selected the Ultra Twin will be configured for pump control.
3 = Flow
When selected the Ultra Twin will be configured for OCM measurement.
ENTER
1
9 9 7
ENTER
This takes you to the common
applications parameters, and you will see
some options appearing on the display.
Page 42
ULTRA TWIN INSTRUCTION MANUAL
42
Level or Volume
If you want to set-up a level or volume application, as described in the following examples, then choose 1 for Level/Vol. You will then be given a choice of 1 = Level or 2 = Volume.
Choose Your Application
If you want to set-up a basic level monitoring application, as described in the following example 1, then choose 1 for Level/Vol, and then 1 for level.
If you want to set-up a level monitoring application with control relays, as described in the following example 2, then choose 1 for Level/Vol. Followed by 1 for level and choose either control down (press 1) or control up (press 2).
If you want to set-up a volume application, as described in the following example 3, then choose 1 for Level/Vol. Followed by 2 for volume, you then need to decide if any control function is required and choose the appropriate option no control (press 0), control down (press 1) or control up (press 2).
Once you have chosen your application you will be asked a series of questions which are answered by choosing the appropriate option as detailed in the flow chart below. Once all the questions have been answered you will be prompted to provide further information, as detailed in the tables below, to complete the programming of the unit.
Page 43
PULSAR MEASUREMENT
43
Level/Volume Quick Setup Menu
Application
1 = Level/Vol
For Each Alarm
1 = Set to Relay 1
2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5
6 = Set to Relay 6
0 = No Control
1 = Control Down
2 = Control Up
No. of Control Relays
1 = 1 Control Relay 2 = 2 Control Relay 3 = 3 Control Relay 4 = 4 Control Relay 5 = 5 Control Relay
6 = 6 control Relay
For each Cntl. Relay
1 = Set to Relay1
2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5 6 = Set to Relay 6
How Many Alarm
0 = No Alarms 1 = One Alarm
2 = Two Alarms
3 = Three Alarms
4 = Four Alarms
5 = Five Alarms
6 = Six Alarms
List will be truncated
according to the
number of Control
relays selected
For Each Alarm
1 = High Alarm
2 = Low Alarm
3 = Hi Hi Alarm
4 = Lo Lo Alarm
5 = Loss of Echo
1 = Level
2 = Volume
Page 44
ULTRA TWIN INSTRUCTION MANUAL
44
Wait….
If you have selected a Volume Application, you will now be prompted to enter details required for the calculation of volume.
PARAMETER
DEFAULT
DESCRIPTION
P101 Transducer
2 = dB6
Type of transducer being used
P102 Material
1 = Liquid
Material in the vessel, either liquid or solid. If the solid lays flat, then it can be entered as liquid.
P104 Measurement Units
1 = metres
Select units to be used for programming measurement information.
P105 Empty Level
6.00m
Distance from the face of the transducer to the material at the bottom of the vessel.
P106 Span
5.70m
Distance from the empty level (0% full) to span (100% full).
PARAMETER
DEFAULT
DESCRIPTION
P600 Vessel Shape
0=Cyl. Flat Base
Shape of vessel being monitored.
P601-P603 Vessel Dimensions
Dependent on vessel shape selected.
Enter Vessel dimensions as required P605 Volume
units
3 = Cubic m
Selects volume units required.
P607 Max Volume
Read Only
Displays the calculated Volume in P605 units.
Page 45
PULSAR MEASUREMENT
45
For More Options Hit Enter
PARAMETER
DEFAULT
DESCRIPTION
P213 / P214 Relay 1 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P223 / P224 Relay 2 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P233 / P234 Relay 3 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P243 / P244 Relay 4 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P253 / P254 Relay 5 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P263 / P264 Relay 6 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P830 mA Out Range
2 = 4-20mA
Determines the mA output range. 0 = Off, 1 = 0 to 20mA 2 = 4 to 20mA, 3 = 20 to 0mA, 4 = 20 to 4mA.
P870 Fill Damping
10.00 m/min
Rate of maximum fill rate (set above the actual fill rate of the vessel).
P871 Empty Damping
10.00 m/min
Rate of maximum empty rate (set above the actual empty rate of the vessel).
Page 46
ULTRA TWIN INSTRUCTION MANUAL
46
The default values used for determining the relay setpoints, when setting Alarm and Control relays, via the Quick Setup menu are entered as a % of span and are as follows.
APPLICATION
NO. OF
CTL
RELAYS
CTL RELAY
NUMBER
ON
SETPOINT
OFF
SETPOINT
Control Down
One
Control 1
80%
20%
Control Down
Two
Control 1 Control 2
80% 70%
20% 20%
Control Down
Three
Control 1 Control 2 Control 3
80% 70% 60%
20% 20% 20%
Control Down
Four
Control 1 Control 2 Control 3 Control 4
80% 70% 60% 50%
20% 20% 20% 20%
Control Down
Five
Control 1 Control 2 Control 3 Control 4 Control 5
80% 70% 60% 50% 40%
20% 20% 20% 20% 20%
Control Down
Six
Control 1 Control 2 Control 3 Control 4 Control 5 Control 6
80% 70% 60% 50% 40% 30%
20% 20% 20% 20% 20% 20%
Page 47
PULSAR MEASUREMENT
47
APPLICATION
NO. OF
CTL
RELAYS
CTL RELAY
NUMBER
ON
SETPOINT
OFF
SETPOINT
Control Up
One
Control 1
20%
80%
Control Up
Two
Control 1 Control 2
20% 30%
80% 80%
Control Up
Three
Control 1 Control 2 Control 3
20% 30% 40%
80% 80% 80%
Control Up
Four
Control 1 Control 2 Control 3 Control 4
20% 30% 40% 50%
80% 80% 80% 80%
Control Up
Five
Control 1 Control 2 Control 3 Control 4 Control 5
20% 30% 40% 50% 60%
80% 80% 80% 80% 80%
Control Up
Six
Control 1 Control 2 Control 3 Control 4 Control 5 Control 6
20% 30% 40% 50% 60% 70%
80% 80% 80% 80% 80% 80%
RELAY
FUNCTION
RELAY ID
ON
SETPOINT
OFF
SETPOINT
Alarm
HiHi
90%
85%
Alarm
High
85%
80%
Alarm
Low
10%
15%
Alarm
LoLo
5%
10%
Page 48
ULTRA TWIN INSTRUCTION MANUAL
48
Example 1 Level Monitoring with Alarms
A vessel, containing liquid that has a variation in level that is to be monitored, with a high-level alarm set on Relay 1 and low-level alarm set on Relay 2.
This application is to be assigned to Point 1 (Transducer 1)
In this example, when the level rises to 2.38 m, relay 1 will come on until the level drops to 2.24 m when it will turn off. If the level drops to 0.28 m, then relay 2 will come on until it rises 0.42 m when it will turn off.
The display will show the level in the tank.
The mA output will be representative of level where 4mA = empty level (0%) and 20mA = 2.8m (100%).
Empty Distance (P105), 3.5m
100%. Span (P106), 2.8m
85%. High alarm on (P213), 2.38m
80%. High alarm off (P214), 2.24m
0%, Empty level
10%. Low alarm on (P223), 0.28m
15%, Low alarm off (P224), 0.42m
Page 49
PULSAR MEASUREMENT
49
To program the Ultra Twin for Example 1 Level Monitoring with Alarms by using the Quick Setup menu proceed as follows. If required to access Program Mode, key in the passcode 1997 and press ENTER,
At the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and ENTER.
QUESTION
OPTION
Level/Volume
1 = Level App.
No. of alarms
2 = 2 Alarms
Type alarm 1
1 = High
Alarm no.1
1 = Set to relay 1
Type alarm 2
2 = Low
Alarm no.2
2 = Set relay 2
Xducer (P101)
2 = dB6
Material (P102)
1 = Liquid
Measurement units (P104)
1 = Metres
Empty Level (P105)
3.5m
Span (P106)
2.8m
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to the Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is displayed, and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either by the main menu system or directly via parameter number and changed as necessary.
Page 50
ULTRA TWIN INSTRUCTION MANUAL
50
Example 2 Level Monitoring and Control (up or down)
A vessel, containing a liquid that has a variation in level that is to be monitored, and when the level reaches a specific point, the vessel is pumped down, with the fluid being transferred to another process. The pump will be assigned to Relay 1 a High Alarm to Relay 2 and Low Alarm to Relay 3.
This application is to be assigned to Point 2 (Transducer 2)
In this example, there is a control relay (relay 1), which will come on if the level rises to 2.24m, and go off when the level drops to 0.56m (control down). If the level rises to 2.4m, then the high-level alarm (relay 2) will come on until the level drops to 2.24m. If the level falls to 0.28m, then the low­level alarm (relay 3) will come on until the level rises to 0.42m.
Alternatively, if it is a control up application, then the on and off points for the control relay are reversed, so the control device comes on when the level is at 0.56m and goes off when it rises to 2.24m.
Empty Distance (P105), 3.5m
100%. Span (P106), 2.8m
85%. High alarm on (P213), 2.38m
80%. High alarm off (P214), 2.24m
0%, Empty level
10%. Low alarm on (P233), 0.28m
15%, Low alarm off (P234), 0.42m
80%. Control (down) on (P213), 2.24m
20%. Control (down) off (P214), 0.56m
Page 51
PULSAR MEASUREMENT
51
The display will show the level in the tank and the mA output will be representative of level where 4mA = empty level (0%) and 20mA = 2.8m (100%).
To program the unit for Example 2 Level Monitoring and Control by using the Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER.
At the Quick Setup menu press the hot key and toggle to Point 2 display and press ENTER and then as prompted, by the questions, select the relevant option and ENTER.
Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and press ENTER.
QUESTION
OPTION
Level/Volume
1 = Level App.
Control
1 = Control down
No. of controls
1 = 1 Relay
Control no.1
1 = Set to relay 1
No. of alarms
2 = 2 Alarms
Type alarm 1
1 = High
Alarm no.1
1 = Set to relay 2
Type alarm 2
2 = Low
Alarm no.2
The unit knows that only Relay 3 is available and so will automatically set Alarm 2 to Relay 3.
Xducer (P101)
2 = dB6
Material (P102)
1 = Liquid
Measurement units (P104)
1 = Metres
Empty Level (P105)
3.5m
Span (P106)
2.8m
Page 52
ULTRA TWIN INSTRUCTION MANUAL
52
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to the Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is displayed and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either by the main menu system or directly via parameter number and changed, as necessary.
Page 53
PULSAR MEASUREMENT
53
Example 3 Volume Application
A cylindrical tank with a diameter of 2m and a flat base that is typically used to temporarily hold liquid, and you wish to know the volume of liquid. You also require a high and low alarm and when the level reaches a specific point, the vessel is pumped down, with the fluid being transferred to another process. The application is to be assigned to Point 1 (Transducer 1)
In this example, there is a control down relay (relay 1), which will come on if the level rises to 2.24m, and go off when the level drops to 0.7m. (control down). If the level rises to 2.38m, then the high-level alarm (relay 2) will come on until the level drops to 2.24 m. If the level falls to 0.28m, then the low-level alarm (relay 3) will come on until the level rises to 0.42m.
The display will show the volume of fluid in the tank and the mA output will be representative of Volume where 4mA = empty (0%) and 20mA = Max Volume (100%).
Empty Distance (P105), 3.5m
100%. Span (P106), 2.8m
85%. High alarm on (P213), 2.38m
80%. High alarm off (P214), 2.24m
0%, Empty level
10%. Low alarm on (P233), 0.28m
15%, Low alarm off (P234), 0.42m
80%. Control (down) on (P213), 2.24m
20%. Control (down) off (P214), 0.7m
Page 54
ULTRA TWIN INSTRUCTION MANUAL
54
To program the Ultra Twin for Example 3 Volume Application with Control by using the Quick Setup menu proceed as follows. If required access the Program Mode and key in the passcode 1997 and press ENTER.
Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and ENTER.
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to the Run Mode.
QUESTION
OPTION
Level/Volume
2 = Volume App.
Control
1 = Control down
No. of controls
1 = 1 Relay
Control no.1
1 = Set to relay 1
No. of alarms
2 = 2 Alarms
Type alarm 1
1 = High
Alarm no.1
1 = Set to relay 2
Type alarm 2
2 = Low
Alarm no.2
The unit knows that only Relay 3 is available and so will automatically set Alarm 2 to Relay 3.
Xducer (P101)
2 = dB6
Material (P102)
1 = Liquid
Measurement units (P104)
1 = Metres
Empty Level (P105)
3.5m
Span (P106)
2.8m
Vessel shape (P600)
0 = Cylindrical flat base
Vessel dimensions
Enter vessel dimensions as requested (depends on vessel shape chosen)
Volume units
Select as required
Max Volume (Read only)
Displays the Max volume as calculated by the Ultra Twin
Page 55
PULSAR MEASUREMENT
55
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is displayed and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either by the main menu system or directly via parameter number and changed, as necessary.
Page 56
ULTRA TWIN INSTRUCTION MANUAL
56
Example 4 Differential Control
In this example the Ultra Twin is being used to control a rake on a screen, which is filtering out solids in the inlet flow to a wastewater treatment plant.
A high alarm has been assigned to Point 1 (Transducer 1), on the
upstream side and a low alarm, to Point 2 (Transducer 2) on the downstream side. The Diff. Control, to operate the rake is on relay 1, high
alarm, on Transducer 1 (upstream), is on relay 2 and, low alarm, on Transducer 2 (downstream) is on relay 3.
Important Notice
In this example, the transducers are mounted at the same height. If the transducers are mounted at different heights, ensure that the empty levels are correct such that there is no differential present when the level is zero on both sides.
0%. Empty Level
Empty Distance (P105), 3.5m
100%. Span (P106), 2.8m
85%. Pt 1, high alarm on (P223), 2.38m
80%. Pt 1, high alarm off (P224), 2.24m
10%. Pt 2, low alarm on (P233), 0.28m
15%. Pt 2, low alarm off (P234), 0.42m
1%. Differential control on (P214), 0.03m
5%. Differential control off (P223), 0.14m
{
Point 1
(Up Stream)
Point 2
(Down Stream)
Page 57
PULSAR MEASUREMENT
57
This will operate as follows, when the level rises on the upstream side and/or the level on the downstream side falls, resulting in a differential of 0.14m, (anywhere within the working span), indicating that the screen is blocked, relay 1 will come on and operate the rake. Once the level on the inflow has decreased and the differential level falls to 0.03m relay 1 will switch off the rake.
Should the level on the upstream side rise, for any reason, to a level of
2.38m, relay 2 will operate to give a high alarm, once the level has fallen back to 2.24m the alarm will go off. A falling level in the downstream side, for any reason, will operate relay 3 at 0.28m giving an alarm for low level, once the level has risen again to a value of 0.42 m relay 3 will reset.
To program the Ultra Twin for Example 4 Differential Control by using the Quick Setup menu proceed as follows. Access the Program Mode and key in the passcode 1997 and press ENTER. At the Quick Setup menu press
ENTER and as prompted, by the questions, select the relevant option and ENTER.
QUESTION
OPTION
Level/Pump /Flow
1 = Level/Vol
Level or Volume
1 = Level
Control
1 = Control Down
No. of Control relays
1 = 1 Control
Control No. 1
1 = Set to Relay 1
No. of alarms
1 = 1 Alarm
Type Alarm 1
1 = High Alarm
Alarm No. 1
Set to Relay 2
Xducer (P1-101)
2 = dB6
Material (P1-102)
1 = Liquid
Measurement units (P1-104)
1 = Metres
Empty Level (P1-105)
3.5m
Span (P1-106)
2.8m
Press CANCEL to come out of the Quick setup menu for point 1 and press the hotkey to switch to point 2.
At the Quick Setup menu for point 2 press ENTER and as prompted by the questions, select the relevant option, and ENTER.
Page 58
ULTRA TWIN INSTRUCTION MANUAL
58
QUESTION
OPTION
Level/Pump /Flow
1 = Level/Vol
Level or Volume
1 = Level
Control
0 = No Control
No. of alarms
1 = 1 Alarm
Type Alarm 1
2 = Low Alarm
Alarm No. 1
Set to Relay 3
Xducer (P1-101)
2 = dB6
Material (P1-102)
1 = Liquid
Measurement units (P1-104)
1 = Metres
Empty Level (P1-105)
3.5m
Span (P1-106)
2.8m
When prompted “For more options hit ENTER”, press ENTER. Use the left and right arrow keys and the ENTER key to access the following parameters and change their values to those shown below.
Press ENTER to save the new values.
PARAMETER
VALUE
P*213, R1 Set 1
0.14m
P*214, R1 Set 2
0.03m
Press CANCEL and when Quick Setup is displayed scroll across to the Relays menu. Press ENTER and press ENTER again when * Relay 1 is shown on the screen. Scroll across to P*216 and set the following. This will set up the differential control relay.
PARAMETER
VALUE
P*216, R1 Allocation
5 = Diff. -2
After pressing ENTER to save the parameter, press CANCEL until Relays is displayed on the screen.
Page 59
PULSAR MEASUREMENT
59
On the wall mount model, to display the Differential on the main display line of the upper LCD, point 1 Level on the auxiliary display of the upper LCD and Point 2 Level on the main display on the lower LCD, change the following parameters. Press the hotkey to switch back to point 1. The upper LCD should now show Relays. Press the Right arrow key until Display is shown on the screen. Press ENTER and press ENTER again when P1 Options is displayed on the screen. Use the left and right arrow keys and ENTER key to change the following parameter and press ENTER to save the new value.
PARAMETER
VALUE
P1-805, Disp. Source
5 = Diff. -2
After pressing ENTER to save the parameter, press CANCEL to display P1 Options on the screen. Press the Right arrow key until P1 Auxiliary is displayed and press ENTER.
Use the left and right arrow keys and ENTER key to change the following parameter and press ENTER to save the new value.
PARAMETER
VALUE
P1-816, Aux Source
1 = Point 1
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press ENTER, and the Ultra Twin will return to the Run Mode.
Page 60
ULTRA TWIN INSTRUCTION MANUAL
60
Pump
If you want to set-up a pump application, as described in the following examples, then choose 2 for pump. You will then be given a choice of 1 = Level App. 2 = Pump Down or 3 = Pump Up.
Choose Your Application
If you want to set-up a pump down (sump control) application, as described in the following example 1 then choose 2 for pump followed by 2 for pump down.
If you want to set-up a pump up (reservoir control) application, as described in the following example 2 then choose 2 for pump followed by 3 for pump up.
Once you have chosen your application you will be asked a series of questions which are answered by choosing the appropriate option as detailed in the flow chart below. Once all the questions have been answered you will be prompted to provide further information, as detailed in the tables below, to complete the programming of the unit.
Page 61
PULSAR MEASUREMENT
61
Quick Setup
Quick Setup
For Each Alarm
1 = High Alarm 2 = Low alarm 3 = Hi Hi Alarm 4 = Lo Lo alarm 5 = Loss of Echo
1 = Level
2 = Pump Down
3 = Pump Up
How Many Pumps
1 = One Pump
2 = Two Pumps
3 = Three Pumps 4 = Four Pumps 5 = Five Pumps 6 = Six Pumps
For each Pump Relay
1 = Set to Relay 1 2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5 6 = Set to Relay 6
How Many Alarms
0 = No Alarms 1 = One Alarm 2 = Two Alarms 3 = Three Alarms 4 = Four Alarms 5 = Five Alarms 6 = Six Alarms List will be truncated according to the number of Control relays selected.
Pump Duty
1 = Fixed Duty Assis 2 = Fixed Duty Backup 3 = Alt Duty Assist 4 = Alt Duty Backup 5 = Duty Backup & Ass 6 = Ser Ratio Duty Ass 7 = Ser Ratio Duty B’up 8 = FOFO Alt Duty Ass 9 = Ser Ratio Stby 10 = 2 Pump Set
For each Alarm
1 = Set to Relay 1 2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5 6 = Set to Relay 6
Wait…
Page 62
ULTRA TWIN INSTRUCTION MANUAL
62
PARAMETER
DEFAULT
DESCRIPTION
P101 Transducer
2 = dB6
Type of transducer being used.
P102 Material
1 = Liquid
Material in the vessel, either liquid or solid. If the solid lays flat, then it can be programmed as liquid.
P104 Measurement Units
1 = metres
Select units to be used for programming measurement information.
P105 Empty Level
6m
Distance from the face of the transducer to the material at the bottom of the vessel.
P106 Span
5.7m
Distance from the empty level (0% full) to span (100% full).
For More Options Hit Enter
PARAMETER
DEFAULT
DESCRIPTION
P213 / P214 Relay 1 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P223 / P224 Relay 2 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P233 / P234 Relay 3 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P243 / P244 Relay 4 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P253 / P254 Relay 5 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
P263 / P264 Relay 6 ON/OFF setpoints
Factory pre-set as a % to appropriate level according to the span already entered. See tables below.
Either Alarm or Level control. Depends on application.
Page 63
PULSAR MEASUREMENT
63
PARAMETER
DEFAULT
DESCRIPTION
P830 mA Out Range
2 = 4 to 20mA
Determines the mA output range. 0 = Off, 1 = 0 to 20mA, 2 = 4 to 20mA, 3 = 20 to 0mA, 4 = 20 to 4mA.
P870 Fill Damping
10m/min
Rate of maximum fill rate (set above the actual fill rate of the vessel).
P871 Empty Damping
10m/min
Rate of maximum empty rate (set above the actual rate of the vessel).
The default values used for determining the relay setpoints, when setting Alarm and Pump relays, via the Quick Setup menu are entered as a % of span and are as follows.
APPLICATION
NO. OF PUMPS
PUMP
NUMBER
ON
SETPOINT
OFF
SETPOINT
Pump Down
One
Pump 1
80%
20%
Pump Down
Two
Pump 1 Pump 2
80% 70%
20% 20%
Pump Down
Three
Pump 1 Pump 2 Pump 3
80% 70% 60%
20% 20% 20%
Pump Down
Four
Pump 1 Pump 2 Pump 3 Pump 4
80% 70% 60% 50%
20% 20% 20% 20%
Pump Down
Five
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5
80% 70% 60% 50% 40%
20% 20% 20% 20% 20%
Pump Down
Six
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5 Pump 6
80% 70% 60% 50% 40% 30%
20% 20% 20% 20% 20% 20%
Page 64
ULTRA TWIN INSTRUCTION MANUAL
64
APPLICATION
NO. OF PUMPS
PUMP
NUMBER
ON
SETPOINT
OFF
SETPOINT
Pump Down
One
Pump 1
50%
20%
Pump Down
Two
Pump 1 Pump 2
50% 70%
20% 20%
Pump Down
Three
Pump 1 Pump 2 Pump 3
50% 60% 70%
20% 20% 20%
Pump Down
Four
Pump 1 Pump 2 Pump 3 Pump 4
40% 50% 60% 70%
20% 20% 20% 20%
Pump Down
Five
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5
40% 50% 60% 70% 75%
20% 20% 20% 20% 20%
Pump Down
Six
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5 Pump 6
40% 50% 60% 70% 75% 80%
20% 20% 20% 20% 20% 20%
Page 65
PULSAR MEASUREMENT
65
APPLICATION
NO. OF PUMPS
PUMP
NUMBER
ON
SETPOINT
OFF
SETPOINT
Pump Up
One
Pump 1
50%
20%
Pump Up
Two
Pump 1 Pump 2
50% 30%
20% 20%
Pump Up
Three
Pump 1 Pump 2 Pump 3
50% 40% 30%
20% 20% 20%
Pump Up
Four
Pump 1 Pump 2 Pump 3 Pump 4
60% 50% 40% 30%
20% 20% 20% 20%
Pump Up
Five
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5
60% 50% 40% 30% 75%
20% 20% 20% 20% 20%
Pump Up
Six
Pump 1 Pump 2 Pump 3 Pump 4 Pump 5 Pump 6
40% 50% 60% 70% 75% 80%
20% 20% 20% 20% 20% 20%
RELAY
FUNCTION
RELAY ID
ON
SETPOINT
OFF
SETPOINT
Alarm
HiHi
90%
85%
Alarm
High
85%
80%
Alarm
Low
10%
15%
Alarm
LoLo
5%
10%
Important Notice
When using the Quick Setup Menu relays will be allocated to the point of measurement you are currently setting up and the availability of relays will depend on the number of relays used when setting up the previous point of measurement via the Quick Setup Menu for that point.
Page 66
ULTRA TWIN INSTRUCTION MANUAL
66
Example 2 Sump Control (pump down)
A sump is typically used to temporarily hold water or effluent, and when the level reaches a specific point, the sump is pumped down, with the fluid being transferred to another process.
In this example, there are two pumps, which will be set to alternate duty assist, so they come on alternately. Pump 1 is to be set to relay 1, pump 2 to relay 2, and the high-level alarm to relay 3.
This will operate as follows. During normal operation, pump 1 will come on at 0.84m and pump down to 0.56m. The setpoints are then shifted to pump 2, which will come on first next time. During peak periods, when pump 1 cannot cope, pump 1 will come on at 0.84m, pump 2 will come on at 1.4m, and pump down to 0.56m. The setpoints are then shifted to pump 2, which will come on first next time.
If neither pump can cope, and the level rises to 2.38m, then the alarm relay (relay 3) will come on and go off when the level falls to 2.24m. This will indicate insufficient capacity of the pumps. The display will show the level in the sump and the mA output will be representative of level where 4mA = empty level (0%) and 20mA = 2.8m (100%).
Empty Distance (P105), 3.5m
100%. Span (P106), 2.8m
85%. High alarm on (P233), 2.38m
80%. High alarm off (P234), 2.24m
0%, Empty level
20%. Pump 1+2 off (P214, 224), 0.56m
30%. Pump 1 on (P213), 0.56m
50%. Pump 2 on (P223), 1.4m
Page 67
PULSAR MEASUREMENT
67
To program the Ultra Twin for Example 2 Sump control (pump down) by using the Quick Setup menu proceed as follows. If required to access Program Mode, key in the passcode 1997 and press ENTER.
At the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and ENTER.
QUESTION
OPTION
Application
2 = Pump
Level/Pump Up/Down
2 = Pump Down
No. of Pumps
2 = 2 Pumps
Pump Duty
3 =Alt Duty Assist
Pump No. 1
1 = Set to relay 1
Pump No. 2
2 = Set to relay 2
No. of Alarms
1 = 1 Alarm
Type Alarm 1
1 = High
Alarm No. 1
5 = Set to relay 5
Xducer (P1-101)
2 = dB6
Measurement units (P-1104)
1 = Metres
Empty Level (P1-105)
3.5m
Span (P1-106)
2.8m
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to the Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is displayed, and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either by the main menu system or directly via parameter number and changed as necessary.
Page 68
ULTRA TWIN INSTRUCTION MANUAL
68
Example 3 Reservoir Control (pump Up)
A sump is typically used to temporarily hold water or effluent, and when the level reaches a specific point, the sump is pumped down, with the fluid being transferred to another process.
In this example, there are two pumps, which will be set to alternate duty assist, so they come on alternately. Pump 1 is to be set to relay 2, pump 2 to relay 3, and the low-level alarm to relay 1. This will operate as follows:
During normal operation, pump 1 will come on at 1.96m and pump up to
2.24m. The setpoints are then shifted to pump 2, which will come on first next time. During peak periods, when pump 1 cannot cope, pump 1 will come on at 1.96m and pump 2 will come on at 1.4m and pump up to 2.24m. The setpoints are then shifted to pump 2, which will come on first next time. If both pumps cannot cope, and the level falls to 0.28m, then the alarm relay (relay 3) will come on and go off when the level rises to 0.42m. This will indicate insufficient capacity of the pumps.
The display will show the level in the sump and the mA output will be representative of level where 4mA = empty level (0%) and 20mA = 2.8m (100%).
50%. Pump 2 on (P233), 1.4m
0% - Empty level
80%. Pump 1&2 off (P224, P234), 2.24m 10% - Low alarm on (P213), 0.28m
Empty distance (P105), 3.5m
70%. Pump 1 on (P223), 1.96m
100%. Span (P106), 2.8m
15%. Low alarm off (214), 0.42m
Page 69
PULSAR MEASUREMENT
69
To program the Ultra Twin for Example 3 Reservoir Control (pump up) by using the Quick Setup menu proceed as follows. If required access the Program Mode and key in the passcode 1997 and press ENTER
Using the ‘right arrow key, go to Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and ENTER.
QUESTION
OPTION
Application
2 Pump
Level/Pump Up/Down
2 = Pump Up
No. of Pumps
2 = 2 Pumps
Pump Duty
3 =Alt Duty Assist
Pump No. 1
1 = Set to relay 1
Pump No. 2
2 = Set to relay 2
No. of alarms
1 = 1 Alarm
Type Alarm 1
2 = Low
Alarm No. 1
3 = Set to Relay 3
Xducer (P101)
2 = dB6
Measurement units (P104)
1 = Metres
Empty Level (P105)
3.5m
Span (P106)
2.8m
Programming is now complete, and the unit can be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to the Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is displayed, and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either by the main menu system or directly via parameter number and changed, as necessary.
Page 70
ULTRA TWIN INSTRUCTION MANUAL
70
Flow
If you want to set-up a flow application, as described in the following examples, then choose 3 for flow. You will then be given a choice of Primary Measuring Devices to choose from.
Choose Your Application
There are five categories of Primary Measuring Device, which are all described in this chapter. They are exponential, BS3860 flumes, BS3860 weirs, special and universal.
Calculations for flow can be performed using absolute or ratiometric calculations. The answer will be the same, the choice of calculation method being limited to the amount of information available, with regards to the primary measuring device.
For ratiometric calculation it is normally sufficient to know the maximum flow at maximum head for the device in question. All types of primary measuring devices can be set up with a choice of alarms.
If you want to set-up a basic exponential device, as described in the following example 1, then choose 3 for Flow, followed by 1 for exponent. You then need to select the primary measuring device for your application from the following available options: suppressed rectangular weir,
Cipolletti (trapezoidal) weir, Venturi flume, Parshall flume, Leopold Lagco flume, V notch weir or other, for any other type of exponential
device. To set-up an application for a BS3680 flume, as described in the following
example 2, then choose 3 for Flow followed by 2 for 3680 Flume. You then need to select the primary measuring device for your application from the following available options: rectangular flume with or without hump, U- throated flume with or without hump.
To set-up an application for a BS3680 weir, as described in the following example 3, then choose 3 for Flow followed by 3 for 3680 Weir. You then need to select the primary measuring device for your application from the following available options: rectangular weir, V notch full 90o (90degrees),
V notch half 90o (53 degree 8 minutes), V notch quarter 90o (28 degree 4 minutes) or a Broad Crested Weir.
Page 71
PULSAR MEASUREMENT
71
To set-up an application for a device contained in special, choose 3 for Flow followed by 5 for Special. You then need to select the primary measuring
device for your application from the following available options: Palmer Bowlus flume, H-flume or a V notch, other than BS3680.
For devices, which do not match any of the above devices the application can be setup using a universal flow calculation, to select this option choose 3 for Flow followed by 6 for universal. You then need to select the primary measuring device for your application from the following available options: linear flow or curved flow.
Once you have chosen your application you will be asked a series of questions which are answered by choosing the appropriate option as detailed in the flow chart below. Once all the questions have been answered you will be prompted to provide further information, as detailed in the tables below, to complete the programming of the unit.
Page 72
ULTRA TWIN INSTRUCTION MANUAL
72
Quick Setup Menu
PMD Type
0 = Off (No Calculation) 1 = Exponential 2 = BS3680 Flumes 3 = BS3680 Weirs
4 = Not Available
5 = Special 6 = Universal
For Each Alarm
1 = Set to Relay 1 2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5
6 = Set to relay 6
Exponential
1 = Supp. Rect. 2 = Trapezoid 3 = Venturi 4 = Parshall 5 = Leopold L. 6 = V-Notch 7 = Other
BS3680 Flumes
1 = Rectangular 2 = Rect. + Hump 3 = U-Throat 4 = U-Throat + Hump
BS3680 Weirs
1 = Rectangular 2 = V-Notch 90o (Full 90o) 3 = V-Notch 53
o
8’ (Half 90
o
)
4 = V-Notch 28
o
4’ (Qtr 90
o
)
5 = Broad Crested
Special
1 = Palmer Bowlus 2 = H-Flume 3 = V-Notch Angle
Calculation
1 = Absolute 2 = Ratiometric
How Many Alarms
0 = No Alarms
1 = One Alarm
2 = Two Alarms 3 = Three Alarms 4 = Four Alarms 5 = Five Alarms 6 = Six Alarms
For Each Alarm
1 = High Alarm 2 = Low Alarm 3 = Hi Hi Alarm 4 = Lo Lo Alarm 5 = Loss of Echo
Application
3 = Flow
Universal
1 = Linear flow 2 = Curved flow
Page 73
PULSAR MEASUREMENT
73
Wait ….
PARAMETER
DEFAULT
DESCRIPTION
P101 Transducer
1 = dB Mach3
Type of transducer being used
P706 Volume units
1 = Litres
Units of flow as on display and used for calculations. 1 = Litres 2 = Cubic metres 3 = Cubic Feet 4 = UK gallons 5 = US gallons 6 = Mil. USG
P707 Time Units
1 = per second
Units of time that volume units will be displayed and calculated in
P104 Measurement units
1 = Metres
Used to enter dimensions and displayed where appropriate: 1 = Metres 2 = Centimetres 3 = Millimetres 4 = Feet 5 = Inches
P105 Empty Level
2.425m
Distance from the end of the transducer horn (dB Mach3) or face of the transducer to the material at the bottom of the measuring element
P703 Minimum head
0.00m
Distance from empty point (P105) to zero flow.
P704 Max head
2.425m
Distance from zero flow to max flow. It should be noted that any change to P704 updates P106 Span and vice versa.
P824 Totaliser Allocation
1 = Point 1
Enables the Totaliser to a specific point of flow measurement or a combination of flow when both points set to measure Flow. For full list of options see P824 in Chapter 5 Parameter Guide.
P815 Aux Mode
2 = Level
Enables the Auxiliary display line to display additional information whilst in RUN mode. For full list of options see P815 in Chapter 5 Parameter Guide.
Page 74
ULTRA TWIN INSTRUCTION MANUAL
74
PARAMETER
DEFAULT
DESCRIPTION
P816 Aux Source
0 = Off
Determines which point or combination of points, that the Auxiliary display line will relate to. For full list of options see P816 in Chapter 5 Parameter Guide.
P823 Totaliser Multiplier
4 = *1
Sets the factor by which the calculated volume will be divided or multiplied by before being displayed: 1 = /1000 2 = /100 3 = /10 4 = *1 5 = *10 6 = *100 7 = *1,000 8 = *10,000 9 = *100,000 10 = *1,000,000
The remaining parameters required to finalise the setup of your application will follow on immediately from the above. These parameters relate to details required to carry out the calculation for flow and will be dependent on the Primary Measuring Device chosen and the method of calculation chosen, please enter values for the parameters concerned as requested.
PARAMETER
DEFAULT
DESCRIPTION
P705 Max flow
0.00
When requested enter the known maximum flow rate, in units of volume (P706) and Time (P707) which occurs at maximum head (P704)
P710 Dimension A
0
When requested, enter, in measurement units (P104) the required dimension.
P711 Dimension B
0
When requested, enter, in measurement units (P104) the required dimension.
P712 Dimension C
0
When requested, enter, in measurement units (P104) the required dimension.
P713 Dimension D
0
When requested, enter, in measurement units (P104) the required dimension.
P717 Exponent
Dependent on chosen PMD
Where available the unit will automatically enter the default exponent value for the PMD chosen, but this can be changed if required. When P700 = 7 (Other), enter the exponent value as defined by the manufacturer of the PMD.
P718 K Factor
Enter the ‘K’ factor for the PMD. Obtained from the manufacturer’s specifications
Page 75
PULSAR MEASUREMENT
75
For More Options Hit Enter
PARAMETER
DEFAULT
DESCRIPTION
P213 / P214 Relay 1 ON/OFF setpoints
Depends on application
Set required Alarm Setpoints.
P223 / P224 Relay 2 ON/OFF setpoints
Depends on application
Set required Alarm Setpoints.
P233 / P234 Relay 3 ON/OFF setpoints
Depends on application
Set required Alarm Setpoints.
P243 / P244 Relay 4 ON/OFF setpoints
Depends on application
Set required Alarm Setpoints.
P253 / P254 Relay 5 ON/OFF setpoints
Depends on application
Set required Alarm Setpoints.
P263 / P264 Relay 5 ON/OFF
setpoints
Depends on application
Set required Alarm Setpoints.
P708 Flow Decimal
2
Sets the number of decimal points required in the flow rate display.
P709 Flow cut off
5%
Enter, as a percentage maximum flow. The minimum flow rate to be added to the totaliser.
P830 mA Out range
2 = 4 -20mA
What the mA output uses for the range. 0= Off, 1= 0 to 20 mA, 2= 4 to 20 mA, 3= 20 to 0 mA, 4= 20 to 4 mA.
P870 Fill Damping
10m/min
Rate of maximum fill rate (set above the actual fill rate of the vessel)
P871 Empty Damping
10m/min
Rate of maximum empty rate (set above the actual empty rate of the vessel)
Page 76
ULTRA TWIN INSTRUCTION MANUAL
76
The default values used for determining the relay setpoints, when setting Alarm relays, via the Quick Setup menu are entered as a % of span and are as follows.
RELAY
FUNCTION
RELAY ID
ON
SETPOINT
OFF
SETPOINT
Alarm
HiHi
90%
85%
Alarm
High
85%
80%
Alarm
Low
10%
15%
Alarm
LoLo
5%
10%
Important Notice
When using the Quick Setup Menu relays will be allocated to the point of measurement you are currently setting up and the availability of relays will depend on the number of relays used when setting up the previous point of measurement via the Quick Setup Menu for that point.
Page 77
PULSAR MEASUREMENT
77
Exponential Devices
If the primary measuring device is a simple exponential device, then an exponent value is required. The Ultra win will automatically enter the exponent value for the device chosen as detailed in the table below:
EXPONENT TYPE
PMD SHAPE EXAMPLE
EXPONENT (P717)
Suppressed Rectangular Weir (Without end contractions)
1.50, automatically set by the unit.
Cipolletti (Trapezoidal) Weir
1.50, automatically set by the unit.
Venturi Flume
1.50, automatically set by the unit.
Parshall Flume
Automatically calculated according to the throat size.
Leopold Lagco Flume
1.55
V-Notch Weir
2.50
Other
As per manufacturer
Value to be set as required.
Contracted Rectangular Weir (With end contractions)
1.50
Page 78
ULTRA TWIN INSTRUCTION MANUAL
78
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For Suppressed/Contracted Rectangular, Trapezoidal and V-notch, weirs, the head is measured upstream at a minimum distance of 3 times maximum head from the weir plate to ensure the surface of the liquid is not affected by turbulence or drawdown. (See DRWG. 1)
In the case of a Venturi flume the point of measurement should be 150 mm upstream from the beginning of the converging section and for a Parshall flume 2/3 the length of the converging section upstream of the throat section. (See DRWG 2 and 3).
3 x Hmax minimum
FLOW (q)
DRWG. 1
DRWG. 2
DRWG. 3
FLOW (q)
FLOW (q)
VENTURI FLUME
PARSHALL FLUME
Page 79
PULSAR MEASUREMENT
79
For a Leopold Lagco flume the head is measured at a point upstream of the beginning of the converging section as detailed in the table below. (See DRWG 4).
FLUME SIZE
POINT OF MEASUREMENT
mm
inches
mm
inches
100 - 305
4 – 12 25
1.0
380
15 32
1.3
455
18 38
1.5
530
21 44
1.8
610
24 51
2.1
760
30 64
2.5
915
36 76
3.0
1065
42 89
3.5
1220
48 102
4.0
1370
54 114
4.5
1520
60 127
5.0
1675
66 140
5.5
1830
72 152
6.0
When any Other device is chosen please consult the manufacturer of the device for details of where the point of measurement should be located but ensure that it is chosen such that the surface of the liquid is not affected by turbulence or drawdown.
FLOW (q)
THROAT
DIVERGING
CONVERGING
TRANSDUCER MOUNTED MIN. BLANKING DISTANCE (P107) ABOVE MAX. HEAD
POINT OF MEASUREMENT
DRWG. 4
Page 80
ULTRA TWIN INSTRUCTION MANUAL
80
Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula (s) as follows:
EXPONENT TYPE
FORMULA
EXPONENT
K FACTOR
Suppressed Rectangular Weir (Without end contractions)
Q = KLh* Where: Q=Flow K=K Factor L=Crest length of weir h=head *=exponent
1.50 Automatically selected by the Ultra Twin
Automatically calculated, dependent on measurement, flow and time units chosen.
Cipolletti (Trapezoidal Weir)
Q = KLh* Where: Q=Flow K=K Factor L=Crest length of weir h=head *=exponent
1.50 Automatically selected by the Ultra Twin
Automatically calculated, dependent on measurement, flow and time units chosen
Venturi Flume
Q=Kh* Where: Q=Flow K=K Factor h=head *=exponent
1.50 Automatically selected by the Ultra Twin
Enter value of K Factor (P718) as required
Parshall Flume
Q=Kh* Where: Q=Flow K=K Factor h=head *=exponent
Automatically calculated, dependent on throat size (P719)
Automatically calculated, dependent on measurement, flow and time units chosen
Page 81
PULSAR MEASUREMENT
81
EXPONENT TYPE
FORMULA
EXPONENT
K FACTOR
Leopold Lagco Flume
Q=KD
0.0953hx
Where: Q =Flow K=K factor D=pipe
diameter h=head *=exponent
1.55 Automatically selected by the Ultra Twin
Automatically calculated, dependent on measurement, flow and time units chosen
V-Notch Weir
Q=Kh
x
Where: Q =Flow K=K factor h=head *=exponent
2.50 Automatically selected by the Ultra Twin
Automatically calculated, dependent on measurement flow and time units chosen.
Other
Q=Kh*
Enter value as required
Enter value as required
Contracted Rectangular Weir (With end contractions)
Q=K(L-0.2*h)h* Where: Q=Flow K=K Factor L=Crest length of weir H=head
x
=exponent
1.50 Automatically selected by the Ultra Twin
Automatically calculated, dependent on measurement flow and time units chosen.
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q = q
cal
(h/h
cal
)x
Where:
q = flowrate
q
cal
= flowrate at maximum head (705)
h = head
h
cal
= maximum head (P704)
x = exponent (determined as in absolute calculation above)
Page 82
ULTRA TWIN INSTRUCTION MANUAL
82
Example 1 ‘V’ Notch Weir
In this example, it is required to calculate the flow through a Simple Exponential Device, which on this occasion is a V-Notch Weir. Ratiometric calculation will be used, to use the customers declared maximum flow, there is no requirement for alarms and the flow rate is to be displayed in litres/second. The totaliser is to record the flow in cubic metres but is not to be displayed during RUN.
This application is to be assigned to point 1 (Transducer 1).
The distance from the end of the transducer horn (dB Mach 3) to zero flow (P1-105) is 1 metre and max head (P71-04) is 0.4 metres, maximum flow (P1-705) is known to be 96.5 litres/second.
Page 83
PULSAR MEASUREMENT
83
To program the Ultra Twin for Example 1 V-Notch Weir by using the Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER.
Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and press ENTER.
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to Run Mode.
QUESTION
OPTION
Application
3 = Flow
PMD Type
1 = Exponent
Exponent
6 = V-Notch
Calculation
2 = Ratiometric
No. of alarms
0 = No alarms
Xducer
1 = dB Mach3
Volume units (P1-706)
1 = Litres
Time units (P1-707)
1 = Per second
Measurement units (P*104)
1 = metres
Empty Level (P1-105)
1.00 metres
Minimum head (P1-703)
0.00 metres
Maximum head (P1-704)
0.40 metres
Totaliser Alloc. (P1-824)
1 = On
Aux. Mode (P1-815)
2 = Level
Aux Source (P1-816)
0 = Off
Totaliser multiplier
7 = 1,000
Max flow
96.5
Page 84
ULTRA TWIN INSTRUCTION MANUAL
84
BS3680 Flumes (P700 = 2)
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For a Rectangular and U-throated flume, the head is measured at 3 to 4
times the maximum head upstream from the beginning of the converging section, to ensure the surface of the liquid is not affected by turbulence.
(See DRWG 6)
3-4 x hmax
DRWG. 6
THROAT LENGTH
DIM C (P712)
DIM A (P710)
APPROACH WIDTH
DIM B (P711)
THROAT WIDTH/DIA.
FLOW (q)
BS3680 FLUME
Page 85
PULSAR MEASUREMENT
85
Rectangular Flume Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5gn0.5
CsCvCdbh
1.5
Where:
q = flowrate gn = gravitational acceleration (nominal value = 980.66 cm/s2)
Cs = shape coefficient (value = 1) Cv = velocity coefficient calculated by Ultra Twin (P721) Cd = discharge coefficient calculated by Ultra Twin (P722) b = throat width P711
h = head
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal(Cv/Cvcal
)(Cd/C
dcal
)(h/h
cal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by Ultra Twin (P721)
C
vcal
= velocity coefficient at maximum head
Cd = discharge coefficient calculated by Ultra Twin (P722) C
dcal
= discharge coefficient at maximum head
h = head
h
cal
= maximum head P704
Page 86
ULTRA TWIN INSTRUCTION MANUAL
86
U-Throated Flume Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5
g
n
0.5
CuCvCdbh
1.5
Where:
q = flowrate gn = gravitational acceleration, (nominal value = 980.66 cm/s2) h = head
Cu= shape coefficient calculated by Ultra Twin (P724) Cv = velocity coefficient calculated by Ultra Twin (P721) Cd = discharge coefficient calculated by Ultra Twin (P722) b = throat width P711
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula:
q= q
cal(Cv/Cvcal
)(Cd/C
dcal
)(Cu/C
ucal
)(h/h
cal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by Ultra Twin (P721)
Cv
cal
= velocity coefficient at maximum head
Cd = discharge coefficient calculated by Ultra Twin (P722) Cd
cal
= discharge coefficient at maximum head
Cu = shape coefficient P724 Cu
cal
= shape coefficient at maximum head
h = head h
cal
= maximum head P704
Page 87
PULSAR MEASUREMENT
87
Example 2 BS3680 U-Throated Flume
In this example, it is required to calculate to BS3680 the flow through a U­Throated Flume without any hump. Absolute calculation will be used, and there is a requirement for an alarm to indicate a low flow condition which will be set to relay 1. The flow rate is to be displayed in cubic meters/hour and the totaliser is also to record the flow in cubic metres, the resettable totaliser is to be displayed during RUN.
The application is to be assigned to point 2 (Transducer 2).
The distance from the end of the transducer horn (dB Mach3) or face of the transducer to zero flow (P2-105) is 1 metre and max head (P2-704) is 0.4 metres, maximum flow(P2-705) will be calculated by the Ultra Twin as
725.17 cubic metres/hour.
The dimensions of the flume are as follows:
Approach Channel diameter (Dim “A”) P2-710 = 0.7m Throat diameter (Dim “B”) P2-711 = 0.5m Throat length (Dim “C”) P2-712 = 1m
Page 88
ULTRA TWIN INSTRUCTION MANUAL
88
To program the Ultra Twin for Example 2 BS3680 U-Throated Flume by using the Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER. Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, select the relevant option and press ENTER.
QUESTION
OPTION
Application
3 = Flow
PMD Type
2 = 3680 Flume
3680 Flumes
3 = U-Throat
Calculation
1 = Absolute
No. of alarms
1 = 1 Alarm
Type alarm 1
2 = Low
Alarm No.1
1 = Set to relay 1
Transducer (P2-101)
1 = dB Mach3
Volume units (P2-706)
2 = Cubic metres
Time Units (P2-707)
4 = Per hour
Measurement units (P*104)
1 = Metres
Empty level (P2-105)
1.00m
Minimum head (P2-703)
0.00m
Maximum head (P2-704)
0.40m
Totaliser Alloc. (P2-824)
1 = On
Aux. Mode (P2-815)
1 = Yes
Aux Source (P2-816)
1 = Point 1
Totaliser multiplier (P2-823)
7 = 1,000
Approach dia. (Dim A) (P2-710)
0.70m
Throat dia. (Dim B) (P2-711)
0.50m
Throat length (Dim C) (P2-712)
1.00m
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively, access the setpoints by the main menu system or directly accessing the parameters.
Page 89
PULSAR MEASUREMENT
89
BS3680 Weirs (P700 = 3)
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For a Rectangular and V-notch weir, the head is measured at a point 4 to 5 times the maximum head upstream from the weir plate, to ensure the surface of the liquid is not affected by turbulence or drawdown. (See DRWG
7)
Rectangular Weir Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce 2/3(2gn)
0.5
beh
e
1.5
Where: q = flowrate Ce = discharge coefficient calculated by Ultra Twin (P723)
gn = gravitational acceleration (nominal value = 980.66 cm/s2) be =effective approach width where b is approach width (Dim “A”) P710 he = effective head
4-5 x hmax minimum
DRWG. 7
BS3680 WEIR
Page 90
ULTRA TWIN INSTRUCTION MANUAL
90
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
calCe/Cecal(he/hecal
)
1.5
Where: q = flowrate
q
cal
= flowrate at maximum head P705
Ce = discharge coefficient calculated by Ultra Twin (P723)
Ce
cal
= discharge coefficient at maximum head he = effective head he
cal
= effective head at maximum head
V-Notch Weir Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce8/15tan(theta/2) (2gn)
0.5h2.5
Where:
q = flowrate Ce = discharge coefficient calculated by the Ultra Twin (P723) theta = V-notch angle gn = gravitational acceleration (nominal value = 980.66 cm/s2) h = head
The Ultra Twin pre-sets the angle (theta) on selection of the chosen device this angle is 90° for a BS 3680 full 90°V notch weir, 53° 8 minutes in the case of the BS3680 half 90°V notch weir and 28° 4 minutes in the case of the BS3680 quarter 90°V notch.
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q = q
calCe
(h)/Ce(h
cal
)(h/h
cal
)
2.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Ce(h) = discharge coefficient for head Ce(h
cal
) = discharge coefficient for maximum head
h = head
h
cal
= maximum head P704
Page 91
PULSAR MEASUREMENT
91
Example 3 BS3680 Rectangular Weir
In this example, it is required to calculate to the flow through a BS3680 Rectangular weir. Absolute calculation will be used, and there is a requirement for an alarm to indicate a high flow condition to be set to relay
3. The flow rate is required to be displayed in litres/minute and the totaliser is to record the flow in cubic metres, the resettable totaliser is to be displayed during RUN.
This application is to be assigned to point 1 (Transducer 1).
The distance from the end of the transducer horn to zero flow (P105) is 1 metre and max head (P704) is 0.4 metres, maximum flow (P705).
Approach width (Dim “A”) P710 = 0.5 m Crest width (Dim “B”) P711 = 0.3 m Crest Height (Dim “C”) P712 = 0.3 m
Page 92
ULTRA TWIN INSTRUCTION MANUAL
92
To program the Ultra Twin for Example 3 BS3680 Weir by using the Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER.
Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and press ENTER.
QUESTION
OPTION
Application
3 = Flow
PMD Type
2 = 3680 Weir
3680 Flumes
1 = Rectangular
Calculation
1 = Absolute
No. of alarms
1 = 1 Alarm
Type alarm 1
1 = High
Alarm No.1
3 = Set to relay 3
Transducer (P1-101)
1 = dB Mach3
Volume units (P1-706)
1 = Litres
Time Units (P2-707)
1 = Per Second
Measurement units (P*104)
1 = Metres
Empty level (P1-105)
1.00m
Minimum head (P1-703)
0.00m
Maximum head (P1-704)
0.40m
Totaliser Alloc. (P1-824)
1 = Point 1
Aux. Mode (P1-815)
7 = Totaliser
Aux Source (P1-816)
1 = Point 1
Totaliser multiplier (P1-823)
7 = 1,000
Approach dia. (Dim A) (P1-710)
0.70m
Throat dia. (Dim B) (P1-711)
0.50m
Throat length (Dim C) (P1-712)
1.00m
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively, access the setpoints by the main menu system or directly accessing the parameters.
Page 93
PULSAR MEASUREMENT
93
BS3680 Rectangular Broad Crested Weir
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107.
The head is measured at a point 3 to 4 times the maximum head upstream from the weir crest, to ensure the surface of the liquid is not affected by turbulence or drawdown.
Rectangular Broad Crested Weir Calculations
Calculations Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5
Ceb(gh3)
0.5
Where:
q = flowrate Ce = discharge coefficient calculated by Ultra Twin P723 b = approach width P710 g = gravitational acceleration (nominal value = 980.66 cm/s2) h = head
Page 94
ULTRA TWIN INSTRUCTION MANUAL
94
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
calCe/Cecal(he/hecal
)
1.5
Where: q = flowrate
q
cal
= flowrate at maximum head P705
Ce = discharge coefficient calculated by Ultra Twin P723
Ce
cal
= discharge coefficient at maximum head he = effective head he
cal
= effective head at maximum head
Page 95
PULSAR MEASUREMENT
95
Special Devices
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107.
In the case of a Palmer Bowlus flume the point of head measurement should be half the value of Dim “A” P710 upstream of the device.
For a H-Flume the head measurement is taken at a point downstream from the flume entrance as detailed in the table below:
FLUME SIZE-DIM A
(P710)
POINT OF MEASUREMENT
cm
feet
cm
inches
15.25
0.5
4.7
1.88
23.00
0.75
6.7
2.69
30.05
1.0
9.1
3.63
45.70
1.5
13.5
5.378
61.00
2.0
17.9
7.19
76.20
2.5
22.5
9.00
91.45
3.0
27.2
10.88
137.15
4.5
40.5
16.19
V-notch angle weirs, the head is measured upstream of the weir plate at a minimum distance of 3 times maximum head to ensure the surface of the liquid is not affected by turbulence or drawdown. See Exponential devices, above, for further details.
Page 96
ULTRA TWIN INSTRUCTION MANUAL
96
Palmer Bowlus and H-Flume Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = f(h)
Where:
q = flowrate f = is an 8th degree polynomial solution for h (head)
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal
f(h)/f(h
cal
)
Where:
q = flowrate q
cal
= flowrate at maximum head P705 f(h) = a polynomial solution for h (head) f(h
cal
) = a polynomial solution for h
cal
(maximum head)
V-Notch Angle Weir (Non-BS 3680) Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce 8/15 tan (theta/2)(2gn)
0.5
(h = kh)
2.5
Where: q = flowrate
Ce = discharge coefficient calculated by Ultra Twin (P723) theta = V-notch angle gn = gravitational acceleration h = head kh = compensated head
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal
(h+kh/h
cal
+kh)
2.5
Where: q = flowrate
q
cal
= flowrate at maximum head P705
h = head kh = compensated head
Page 97
PULSAR MEASUREMENT
97
Universal Calculations (P700=6)
Point of Measurement
The transducer must be above the maximum head P704 by at least the near blanking distance P107. For all Universal calculation applications, the point at which the head is measured should be chosen such that the surface of the liquid is not affected by turbulence.
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = q(h) Where: q = flowrate
q(h) = flowrate for head
The desired number of Breakpoints, (P730 - P793) are to be entered in pairs in values of head and corresponding flow. (Minimum of 2 pairs of Breakpoints is required).
Page 98
ULTRA TWIN INSTRUCTION MANUAL
98
CHAPTER 5 PARAMETER GUIDE
This chapter describes all the parameters in your Ultra Twin 140, as they appear in the menu system.
Menu System
Shown below is a set of charts to show you how all the various functions and features can be found using the menu system.
For further details and a full description of all parameters refer to the Parameter Listings and Descriptions section of this chapter.
Top Level Menu
Quick Setup
Application
Relays
Pump
Advanced
Digital Inputs
Data Logs
Volume
P100 = 6
mA 2
Output
mA 1
Output
Display
Compensation
Stability
System
Test
Device Comm
Echo
Processing
OCM
P100 = 4 or 5
Page 99
PULSAR MEASUREMENT
99
Application Menu
P1 & P2
System Units
P1 P2
Operation
P1 & P2
Distances
P1*04
Measurement
Units
P1-105 P2-105
Empty Distance
P1-100 P2-100
Mode
P1-101 P2-101
Transducer
P1-102 P2-102
Material
P1-106 P2-106
Span
P1-107 P2-107
Near Blanking
P1-108 P2-108
Far Blanking
Important Notice
In the following parameters, * denotes the point of measurement being used (* = parameter selectable in Point 1 and/or in Point 2). All menu options are the same for Point 1 and Point 2. Measurement points can be “toggled” by pressing to change from point 1 to point 2 to access parameters on each point.
Page 100
ULTRA TWIN INSTRUCTION MANUAL
100
Relays Menu
Relay 1
P*210
Type
P*211
Function
P*212
Alarm ID
or
Pump
Group
P*214
Set 2
P*213
Set 1
P*215
Set 3
As
Required
P*216
Allocat.
P*217
Closures
P*218
Fail Safe
P*219
R1
Max.Rate
If
P210=2
Relay 2
Relay 3
Relay 4
Relay 5
Relay 6
P*220
Type
P*230
Type
P*240
Type
P*250
Type
P*260
Type
P*221
Function
P*231
Function
P*241
Function
P*251
Function
P*261
Function
P*222
Alarm ID
or
Pump
Group
*P232
Alarm ID
or
Pump
Group
P*242
Alarm ID
or
Pump
Group
P*252
Alarm ID
or
Pump
Group
P*262
Alarm ID
or
Pump
Group
P*223
Set 1
*P233
Set 1
P*243
Set 1
P*253
Set 1
P*263
Set 1
P*224
Set 2
P*234
Set 2
P*244
Set 2
P*254
Set 2
P*264
Set 2
P*225
Set 3
As
Required
P*235
Set 3
As
Required
P*245
Set 3
As
Required
P*255
Set 3
As
Required
P*265
Set 3
As
Required
P*226
Allocat.
P*236
Allocat.
P*246
Allocat.
P*256
Allocat.
P*266
Allocat.
P*227
Closures
P*237
Closures
P*247
Closures
P*257
Closures
P*267
Closures
P*228
Fail Safe
P*238
Fail Safe
P*248
Fail Safe
P*258
Fail Safe
P*268
Fail Safe
P*249
R4
Max.Rate
If
P240=2
P*259
R5
Max.Rate
If
P250=2
P*269
R6
Max.Rate
If
P260=2
P*229
R2
Max.Rate
If
P220=2
P*239
R3
Max.Rate
If
P230=2
Loading...