Pulsar Measurement FlowCERT Lite Product Manual

Page 1
FlowCERT Lite
Instruction Manual
Page 2
2
Page 3
PULSAR MEASUREMENT
3
FLOWCERT LITE (FOURTH EDITION REV 1)
February 2021 Part Number M-150-L-004-1P
COPYRIGHT
© Pulsar Measurement, 2016 -21. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any form without the written permission of Pulsar Process Measurement Limited.
WARRANTY AND LIABILITY
Pulsar Measurement guarantee for a period of 2 years from the date of delivery that it will either exchange or repair any part of this product returned to Pulsar Process Measurement Limited if it is found to be defective in material or workmanship, subject to the defect not being due to unfair wear and tear, misuse, modification or alteration, accident, misapplication, or negligence.
DISCLAIMER
Pulsar Measurement neither gives nor implies any process guarantee for this product and shall have no liability in respect of any loss, injury or damage whatsoever arising out of the application or use of any product or circuit described herein.
Every effort has been made to ensure accuracy of this documentation, but Pulsar Measurement cannot be held liable for any errors.
Pulsar Measurement operates a policy of constant development and improvement and reserves the right to amend technical details, as necessary.
The FlowCERT Lite shown on the cover of this manual is used for illustrative purposes only and may not be representative of the actual FlowCERT Lite supplied.
CONTACT For technical support, please contact: Europe: supporteurope@pulsarmeasurement.com
Outside Europe: supportnorthamerica@pulsarmeasurement.com If you have any comments or suggestions about this product, please contact: Europe: europe@pulsarmeasurement.com
Outside Europe: northamerica@pulsarmeasurement.com Pulsar Measurement website: www.pulsarmeasurement.com
United States
11451 Belcher Road South Largo, FL 33773 888-473-9546
Canada
16456 Sixsmith Drive Long Sault, Ont. K0C 1P0 855-300-9151
United Kingdom
Cardinal Building, Enigma Commercial Centre
Sandy’s Road, Malvern
WR14 1JJ 00 44 (0)1684 891371
Page 4
4
CONTENTS
Chapter 1: Start Here… ........................................................................................................... 9
About this Manual ............................................................................................................... 9
About the FlowCERT ........................................................................................................ 10
Functional Description .................................................................................................... 11
Product Specification ....................................................................................................... 12
EU Certificate of Conformity ......................................................................................... 14
Chapter 2 FlowCERT Lite Installation ............................................................................. 15
Unpacking ............................................................................................................................ 15
Power Supply Requirements......................................................................................... 15
Safety Symbols ................................................................................................................... 16
Location ................................................................................................................................ 17
Dimensions .......................................................................................................................... 18
Wall Mount .......................................................................................................................... 18
Cable Entry ........................................................................................................................... 20
Terminal connection details .......................................................................................... 20
Terminal connections ...................................................................................................... 20
Transducer ........................................................................................................................... 21
ATEX ....................................................................................................................................... 21
FM ........................................................................................................................................... 21
Internal Temperature Compensation ........................................................................ 24
Mounting Bracket ............................................................................................................. 25
Voltage Selector and Fuse Location .......................................................................... 27
Preparation for Operation ............................................................................................. 28
Maintenance ....................................................................................................................... 28
Chapter 3 How To Use Your FlowCERT Lite ................................................................ 29
Operating the Controls ................................................................................................... 29
Run Mode ............................................................................................................................. 30
Program Mode ................................................................................................................... 31
How to Access Program Mode .................................................................................... 31
Hot Keys ................................................................................................................................ 32
Menu Keys ........................................................................................................................... 33
Page 5
PULSAR MEASUREMENT
5
Numeric Keys ...................................................................................................................... 33
Using the Menu System ................................................................................................. 34
Directly Editing Parameters ......................................................................................... 35
Test Mode ............................................................................................................................ 36
Using the Serial Interface ............................................................................................... 37
Parameter Defaults ........................................................................................................... 39
Chapter 4 Quick Setup Guide ........................................................................................... 40
Enter Program Mode ....................................................................................................... 40
Choose Quick Setup ......................................................................................................... 40
Choose Your Application ............................................................................................... 41
Quick Setup ......................................................................................................................... 43
Exponential Devices (P700 = 1) ................................................................................... 47
Point of Measurement .................................................................................................... 48
Calculations ......................................................................................................................... 50
Example 1 V-Notch Weir ................................................................................................ 52
BS3680 Flumes (P700 = 2) ............................................................................................. 54
Calculations ......................................................................................................................... 55
Rectangular Flume ............................................................................................................ 55
U-Throated Flume ............................................................................................................. 56
Example 2 BS3680 U-Throated Flume ...................................................................... 57
BS3680 Weirs (P700 = 3) ................................................................................................ 59
Calculations ......................................................................................................................... 59
BS3680 Rectangular Weir .............................................................................................. 59
BS3680 V-Notch Weir...................................................................................................... 60
Example 3 BS3680 Rectangular Weir ........................................................................ 61
BS3680 Rectangular Broad Crested Weir ................................................................ 63
Calculations ......................................................................................................................... 63
Special Devices (P700 =5) .............................................................................................. 65
Calculations ......................................................................................................................... 66
Page 6
FLOWCERT INSTRUCTION MANUAL
6
Universal Calculations (P700=6) .................................................................................. 67
Calculations ......................................................................................................................... 67
Chapter 5 Parameter Guide ............................................................................................... 68
Menu System ...................................................................................................................... 68
Top Level Menu ................................................................................................................. 68
Application Menu .............................................................................................................. 69
Relays Menu ........................................................................................................................ 70
Data Logs Menu ................................................................................................................ 71
OCM Menu .......................................................................................................................... 72
Display Menu ...................................................................................................................... 73
mA Output Menu .............................................................................................................. 73
Compensation .................................................................................................................... 74
Stability Menu .................................................................................................................... 74
Echo Processing Menu .................................................................................................... 75
System Menu ...................................................................................................................... 76
Device Comm Menu ........................................................................................................ 77
Test Menu ............................................................................................................................ 78
Chapter 6 Parameter Listing and Description ............................................................ 79
Application Parameters................................................................................................... 79
Operation ............................................................................................................................. 79
Dimensions .......................................................................................................................... 80
Relay Parameters ............................................................................................................... 82
Alarms .................................................................................................................................... 83
Pumps .................................................................................................................................... 87
Control ................................................................................................................................... 91
Miscellaneous ..................................................................................................................... 93
Common Relay parameters........................................................................................... 95
Data Log parameters ....................................................................................................... 96
Totaliser Audits .................................................................................................................. 96
Temperature ........................................................................................................................ 96
Page 7
PULSAR MEASUREMENT
7
OCM Parameters ............................................................................................................... 98
PMD Setup ........................................................................................................................... 98
Dimensions ........................................................................................................................ 102
Calculations ....................................................................................................................... 108
Breakpoints ........................................................................................................................ 108
Tables ................................................................................................................................... 109
Average Flow .................................................................................................................... 109
Display Parameters ......................................................................................................... 110
Options ................................................................................................................................ 110
Failsafe ................................................................................................................................. 111
Auxiliary ............................................................................................................................... 112
Totaliser ............................................................................................................................... 114
Bargraph ............................................................................................................................. 116
mA Output Parameters ................................................................................................. 116
Range ................................................................................................................................... 116
Operation ........................................................................................................................... 117
Setpoint ............................................................................................................................... 117
Limits .................................................................................................................................... 118
Trim ....................................................................................................................................... 118
Failsafe ................................................................................................................................. 119
Allocation ........................................................................................................................... 119
Compensation Parameters .......................................................................................... 119
Offset.................................................................................................................................... 119
Temperature ...................................................................................................................... 120
Velocity ................................................................................................................................ 120
Stability Parameters ....................................................................................................... 121
Damping ............................................................................................................................. 121
Indicator .............................................................................................................................. 122
Rate ....................................................................................................................................... 122
Page 8
FLOWCERT INSTRUCTION MANUAL
8
Filters .................................................................................................................................... 123
Echo Processing Parameters ....................................................................................... 124
Transducer Status ............................................................................................................ 124
System Parameters ......................................................................................................... 125
Passcode ............................................................................................................................. 125
Backup ................................................................................................................................. 125
System Information ........................................................................................................ 126
Date & Time ...................................................................................................................... 127
LED Colour ......................................................................................................................... 127
Watchdog ........................................................................................................................... 128
Daylight Saving Time ..................................................................................................... 129
Device Comm. .................................................................................................................. 133
RS232 Set Up .................................................................................................................... 133
RS 485 Set Up ................................................................................................................... 133
Remote Alarm ................................................................................................................... 134
Test Parameters ............................................................................................................... 136
Simulation .......................................................................................................................... 136
Hardware ............................................................................................................................ 137
Chapter 7 Troubleshooting .............................................................................................. 140
Chapter 8 Disposal .............................................................................................................. 141
Notes ........................................................................................................................................ 142
Page 9
PULSAR MEASUREMENT
9

CHAPTER 1: START HERE…

Congratulations on your purchase of a Pulsar FlowCERT Lite. This quality system has been developed over many years and represents the latest in high technology ultrasonic level measurement and control.
It has been designed to give you years of trouble-free performance, and a few minutes spent reading this operating manual will ensure that your installation is as simple as possible.

About this Manual

It is important that this manual is referred to for correct installation and operation. There are various parts of the manual that offer additional help
or information as shown.
Tips
TIP: Look f o r this icon t h r oughout y o u r Pulsar Me a s u r e ment manual to find help f u l informatio n a n d answers t o f r equentl y a s ked questio n s .
Additional Information
References
See Also
References to other parts of the manual.
Additional Information
At various parts of the manual, you will find sections like this that explain specific things in more detail.
Page 10
FLOWCERT INSTRUCTION MANUAL
10

About the FlowCERT

The FlowCERT lite open channel flow meter is a state-of-the-art ultrasonic flow metering and control system which provides comprehensive flow monitoring with data logging (optional) and control functions for a complete range of flumes, weirs and channels.
Easy calibration via the “Quick Set Up” menu and maintenance free “fit and forget” performance means that you can install the FlowCERT LITE rapidly
and with confidence. FlowCERT LITE is equipped with five user definable relays enabling initiation
of control or indication devices such as pumps, samplers and remote totalisers, along with an isolated mA output to provide remote indication while local indication of flow and total flow is provided by a back lit LCD display and bar graph.
Page 11
PULSAR MEASUREMENT
11

Functional Description

FlowCERT Lite sends a transmit pulse to the transducer, which emits an ultrasonic pulse perpendicular to the transducer face, and the returned echo is sent back to the FlowCERT Lite. The time taken to receive the echo is measured and the distance from the transducer face to the surface being monitored is calculated.
FlowCERT Lite can measure from zero to 16m (52.49 feet) from the face of the transducer to the surface being monitored, dependent on the application chosen and transducer used.
The relays can be programmed to activate alarms, pump starters, or other control equipment. There is an isolated 4-20 mA output that can be connected to a recorder or PLC, to monitor, level space, distance, OCM head, or OCM flow, independently from that shown on the display. Additionally, there is an RS232 port, so that the FlowCERT Lite can be operated remotely by a PC or other equipment.
FlowCERT Lite can be programmed either by the built-in keypad (standard), or by PC via the RS 232 Serial Interface (optional). All parameters are stored in non-volatile memory, so are retained in the event of power interruption. A second backup copy of all parameters can also be retained in the FlowCERT Lite memory, in case an alternative set of parameters needs to be stored.
Five user definable control relays with individual setpoints and intelligent performance logging software features ensure maximum control versatility.
The system utilises the unique DATEM software (Digital Adaptive Tracking of Echo Movement). This is a proven digital mapping technique developed especially for the Pulsar Ultra range, which gives the system unequalled
ability when identifying the “true target level” in the face of competing
echoes from pipes, pumps, or other obstructions. Coupled with the powerful ‘all new’ dB transducer range, the FlowCERT Lite lives up to its reputation as the most reliable ultrasonic level measurement system available.
The Pulsar FlowCERT Lite ultrasonic level controller has been designed to provide maintenance-free fit and forget performance.
Page 12
FLOWCERT INSTRUCTION MANUAL
12

Product Specification

PHYSICAL
Wall Mount Outside Dimensions
235 x 184 x 120 mm (9.25 x 7.24 x 4.72 inch)
Weight
Nominal 1kg (2.2lbs)
Enclosure Material/Description
Polycarbonate, flame resistant to UL94-5V
Cable Entry Detail
10 cable entry knock outs, 1 x M16, 5 x M20 underside, 4 x 18mm dia (PG11) at rear
Transducer Cable Extensions
-core screened
Maximum Separation
1000m (3,280 ft), 500m (1,640 ft) for mmWave
ENVIRONMENTAL
IP Rating (Wall)
IP65 (NEMA 4X), IP20 when open
Max. & min. temperature (electronics)
-20 ºC to +50 ºC (-4˚F to +120˚F)
Flammable atmosphere approval
Safe area: compatible with approved dB transducers (see transducer spec' sheet)
CE Approval
See EU Declaration of Conformity
PERFORMANCE
Accuracy
0.1% of the measured range (dB3 and external temperature compensation or internal compensation using twin sunshields)
Resolution
0.5mm (dB3 and external temperature compensation)
Max. Range
Dependent on application and transducer (maximum 16m (52.49ft) dBR16)
Min. Range
Dependent upon application and transducer (minimum zero dB Mach3)
Rate Response
Fully adjustable
ECHO PROCESSING
Description
DATEM (Digital Adaptive Tracking of Echo Movement)
Page 13
PULSAR MEASUREMENT
13
OUTPUTS
Analogue output
Isolated (floating) output (to 150V) of 4-20 mA or 0-20 mA into 500Ω (user programmable and adjustable) 1µA resolution
Digital output
Full Duplex RS232
Volt free contacts, number, and rating
5 form "C" (SPDT) rated at 5A at 115V/240V AC
Display
6 digits plus 12-character text, plus bar graph with direction indicators, remote communicator identifier, and program/run/test mode indicators
PROGRAMMING
On-board programming
By integral keypad
PC programming
Via RS232
Programming security
Via passcode (user selectable and adjustable)
Programmed data integrity
Via non-volatile RAM, plus backup
SUPPLY
Power Supply
115V AC +5% / 10% 50/60 Hz, 230V AC +5% / 10% 50/60 Hz. DC 22-28V 10W maximum power (typically 6W)
Fuses
100 mA at 210-230 VAC (fitted as standard to wall unit) 200 mA at 105-115 VAC
REMOTE COMMUNICATOR
Enclosure protection
IP68
Max and Min Temperature
-20 ºC to +70 ºC (-4˚F to +158˚F)
Pulsar Measurement operates a policy of constant development and improvement and reserve the right to amend technical details, as necessary.
Page 14
FLOWCERT INSTRUCTION MANUAL
14

EU Certificate of Conformity

Page 15
PULSAR MEASUREMENT
15

CHAPTER 2 FLOWCERT LITE INSTALLATION

Unpacking

Power Supply Requirements

FlowCERT lite can operate from AC supply or from a DC battery. The AC is either 115 +5%/-10% 50/60Hz or 230V +5%/-10% 50/60Hz, depending on the position of the selector switch. The DC is 22-28V. In all cases the FlowCERT lite will typically consume 6W of power, with a maximum of 10W. If FlowCERT lite has both an AC and DC supply available then the AC supply source will be automatically sensed and used, should the AC supply be removed for any reason then the DC supply will take over.
The AC and DC wiring should be completed using either 1.5–2.5mm2 (16 – 14AWG) stranded or 1.5–4mm2 (16 – 14AWG) solid wire, with all terminals being tightened to (0.5Nm) (4.5in. lbs.). An external supply isolator/circuit breaker (AC or DC) must be fitted and labelled to identify the instrument to which it refers.
Important Information
All shipping cartons should be opened carefully. When using a box cutter, do not plunge the blade deeply into the box, as it could potentially cut or scratch equipment components. Carefully remove equipment from each carton, checking it against the packing list before discarding any packing material. If there is any shortage or obvious shipping damage to the equipment, report it immediately to Pulsar Process Measurement Limited.
Page 16
FLOWCERT INSTRUCTION MANUAL
16

Safety Symbols

Detailed below are descriptions and meanings of safety/warning symbols that are used on the FlowCERT Lite and in this manual.
SYMBOL
DESCRIPTION
DIRECT CURRENT (DC)
ALETRNATING CURRENT (AC)
PROTECTIVE CONDUCTOR TERMINAL
CAUTION (Refer to accompanying Documents)
!
Page 17
PULSAR MEASUREMENT
17

Location

The FlowCERT Lite must be mounted in a non-hazardous (safe) area, and the transducer fitted in the hazardous area
When choosing a location to mount the enclosure, bear in mind the following:
Ensure that the FlowCERT Lite is installed in a “Safe”, non-hazardous
area.
For a clear view of the LCD display, it is recommended that it is
mounted at eye level.
The mounting surface is to be vibration free.
The ambient temperature is between -20°C and 50°C.
Page 18
FLOWCERT INSTRUCTION MANUAL
18
There should be no high voltage cables or inverters nearby,

Dimensions

Wall Mount

The dimensions of the wall fixing holes are as shown below:
The FlowCERT Lite should be mounted by drilling three holes suitable for size 8 screws (length to suit the application) and fixing the screw in place. Hang the unit on this and fix the two remaining screws by removing the terminal access cover to access the pre-drilled holes.
198mm (7.80”)
30mm (1.18”)
142mm
(5.59”)
FIXING HOLES
Important Information
All electronic products are susceptible to electrostatic shock, so follow proper grounding procedures during installation.
Page 19
PULSAR MEASUREMENT
19
The full dimensions of the enclosure are as shown below:
235mm (9.25”)
184mm (7.24”)
212mm (8.35”)
0.96”
(24.5mm)
1.89”
(48mm)
0.96”
(24.5mm)
1.06”
(27mm)
1.10”
(28mm)
1.10”
(28mm)
1.10”
(28mm)
1.18”
(30mm)
142mm (5.59”)
4.72” (120mm)
Page 20
FLOWCERT INSTRUCTION MANUAL
20

Cable Entry

There are 6 cable gland knockouts on the base of the FlowCERT Lite (5 x 20mm (0.79"), 1 x 16mm (0.63")) and 4 on the rear (4 x 18mm dia (0.73")). Select which ones you wish to take out, and remove them by using a circular cutter, such as a tank cutter. Take care not to damage the circuit board inside whilst undertaking this. Do not use a hammer, as this may cause damage to the enclosure. It is recommended that you use suitable cable glands to ensure that the ingress rating is maintained.

Terminal connection details

There is a wiring diagram inside the terminal access cover, and is shown below:

Terminal connections

Power

The FlowCERT can operate from mains AC and automatically from DC or battery backup in the event of power failure or can be operated permanently from DC or batteries.
Page 21
PULSAR MEASUREMENT
21

Transducer

The transducer should be installed, and connected, in accordance with the installation instructions contained in the Transducer User Guide. The entire range of standard dB/dBR transducers is certified for use in hazardous areas and different models are available for use in Zone 1 or Zone 0. DBR16 is also available for use up to Zone 0. Wire the transducer to the FlowCERT Lite transducer terminals as follows:

Transducer 1

TERMINAL CONNECTION DETAILS
Unit Type
Green: Screen
Red:
Power
White:
Signal 3 Wall Mount
37
38
39
40
When using 2-core screened extension cable, the Black and Green wires of the transducer should be connected to the screen of the extension cable
which in turn should be connected to the appropriate 0 volts’ terminal of the
FlowCERT Lite.

ATEX

For Zone 1 applications a transducer certified to Sira 02ATEX5104X is used, and must be supplied via a 4000A breaking fuse, which is fitted as standard to the FlowCERT Lite.
For Zone 0 a transducer certified to Sira 02ATEX2103X is used, which must be connected to the FlowCERT controller via an external Zener barrier.
FM
For EEx m (Zone 1) applications a transducer certified to FM Class I Div 1 Group A, B, C & D, ClassII Div 1 Group E, F & G, Class III is used, and must
be supplied via a 1500A breaking fuse, which is fitted as standard to the FlowCERT level controller. Restrictions do not use in the presence of these groups of Chemicals, Aliphatic Hydrocarbons, Ketones or Esters.
For EEx ia (I.S.) a transducer certified to FM Class I Div 1 Group A, B, C & D, ClassII Div 1 Group E, F & G is used, which must be connected to the FlowCERT via an external Zener barrier. See transducer label for certification details.
Page 22
FLOWCERT INSTRUCTION MANUAL
22

Relay Outputs

The five relays can be programmed for a variety of alarms, pump control, or other process functions. The relay contacts are all rated at 5A at 240V AC. All connections should be such that the short circuit capacity of the circuits to which they are connected, is limited by fuses rated so that they do not exceed the relay rating.

Current Output

This is an isolated (floating) mA output (to 150 V), of 4 - 20mA or 0 - 20mA, and the load should not exceed 500 .

Temperature Input

The FlowCERT Lite can be used with an external temperature sensor and a dB3 transducer, allowing more localised compensation of the measured distance due to changes in temperature.

There are 2 types of external models, Type A and Type B and their ranges are as follows:
TYPE
RANGE
A
-25° to 50°C
B
-25° to 125°C
Important Information
The temperature sensor should be placed close to the point of measurement and out of the effects of direct sunlight. This allows for the temperature sensor to measure the true air temperature, thereby giving the most accurate results.
Page 23
PULSAR MEASUREMENT
23
The Temperature sensor connections are as follows:
DESCRIPTION
TEMPERATURE SENSOR
FLOWCERT LITE
Power Supply
Terminal 1
Terminal 29
Return
Terminal 2
Terminal 28
Temp Source (P852), should be set to option 4 or 5 depending on the sensor range, set 4 for type A and 5 for type B (see above), the range is specified on the label of the sensor.
92mm
182mm
100mm
M20 ENTRY
IP67 ABS HOUSING
316 STAINLESS STEEL
6mm
Page 24
FLOWCERT INSTRUCTION MANUAL
24

Internal Temperature Compensation

The FlowCERT Lite can also be used with a dB3 transducer fitted with twin Pulsar sunshields, allowing a more localised compensation of the measured distance due to changes in temperature
Parts supplied in the sun shield kit should be:
QTY
DESCRIPTION
2
Sunshield
1
Washer 1 (60mm O.D. x 33mm I.D.)
1
M20 Thread adapter
1
Washer 2 (34mm O.D. x 21mm I.D.)
1
Mounting bracket kit
1
M20 Nut
The upper sun shield is held in place by the mounting bracket and the M20 nut. The bracket can be fixed using the supplied ‘U’ bolts.
The assembly should be mounted perpendicular to the water surface at the correct position as specified by the structure designer/manufacturer.
Important Information
The temperature sensor should be placed close to the point of measurement and out of the effects of direct sunlight. This allows for the temperature sensor to measure the true air temperature, thereby giving the most accurate results.
Page 25
PULSAR MEASUREMENT
25

Mounting Bracket

10 mm
22 mm
25 mm
75 mm
75 mm
30 mm
30 mm
90 mm
30 mm
100 mm
14 mm
Page 26
FLOWCERT INSTRUCTION MANUAL
26
Temp Source (P852), should be set to option 2, so the temperature from the transducer is always used.
The unit is connected as follows:
CABLE COLOUR
DESCRIPTION
TERMINAL NO.
Green
Screen
37
Red
Power
38
White
Signal
39
Black
0 Volts
40

RS232 Serial Interface

If required, you can connect to the serial interface, to operate your FlowCERT Lite remotely.
Page 27
PULSAR MEASUREMENT
27

Voltage Selector and Fuse Location

The voltage selector switch, and mains fuse is located inside the terminal compartment. This is left of the mains terminals, as illustrated below:
DON’T FORGET
Make sure you move the voltage selector switch to the correct position for your power supply
Important Information
Before applying AC power (mains), make sure you have correctly selected the voltage selector switch, as detailed in the preceding pages.
Please note that all units are supplied set to 230 volts AC for safety reasons, and a 100mA fuse fitted as standard.
Never operate the FlowCERT Lite with terminal access exposed.
An external switch or circuit breaker should be installed near to the FlowCERT Lite to allow the supply to be removed during installation and maintenance. In addition, the relay contacts should also have a means of isolating them from the FlowCERT Lite.
Interconnecting cables must be adequately insulated in accordance with local regulations. Strip back 30 mm of the outer insulation of the cable. Strip 5 mm of insulation from the end of each conductor. Twist all exposed strands of the conductor together. Insert the stripped conductor into the terminal block as far as it will go and tighten the terminal block screw. Ensure that all strands are firmly clamped in the terminal block and that there is no excess bare conductor showing, and no stray strands.
Important Information
If the equipment is installed or used in a manner not specified in this manual, then the protection provided by the equipment may be impaired.
Page 28
FLOWCERT INSTRUCTION MANUAL
28

Preparation for Operation

Before switching on, check the following:
FlowCERT Lite is mounted correctly and is in a ‘safe’ area. The power supply is correctly installed. The voltage selector switch is in the correct position. The relays are connected correctly.

Maintenance

There are no user serviceable parts inside FlowCERT Lite, except the mains fuse. If you experience any problems with the unit, then please contact Pulsar Measurement for advice.
To clean the equipment, wipe with a damp cloth. Do not use any solvents on the enclosure.
Important Information
The unique DATEM software comes into operation as soon as power is applied and is designed to monitor a moving level or target with the transducer in a fixed position.
If, after any period of use, it should become necessary to move the transducer, for any reason, from its original operating position, switch off the FlowCERT Lite, before proceeding, in order to prevent any undesirable updates to the DATEM trace. If after moving the transducer the reading is not as expected, please refer to Chapter 8 Troubleshooting.
Page 29
PULSAR MEASUREMENT
29

CHAPTER 3 HOW TO USE YOUR FLOWCERT LITE

Operating the Controls

Display

The display provides information on the current mode of operation, and status of the remote communication. Whilst in the Run Mode it will display the current level reading and its units of measure, along with status messages with regards to the Transducer, Echo reception and Fail Safe Mode. Additionally, it can be programmed to provide status messages on alarms, pumps etc. When in the Program mode the display is used to read information on the Menu System, Parameter Number and parameter details and values, which can be entered. During Test Mode, the display is used to monitor the simulated level. A bar graph is also provided which will provide a visual reading of the level, in percentage of span.
1. Displays the current mode of operation
2. Main 6-digit display:
Run Mode; current measurement displayed, dependent on mode and measurement units chosen, and value of hotkey selected. Program Mode; displays parameter number and values entered for parameters. Test Mode; displays simulated level.
1 2 3
4
0%
5
6
Run Mode Program Mode Test Mode 100%
REMOTE COMMUNICATOR OFF
XXXXXXXXXXXX
000.000
Page 30
FLOWCERT INSTRUCTION MANUAL
30
3. Auxiliary Display, scrolling twelve-digit display.
Run Mode; displays measurement units, status messages on signal and transducer, details of Hot Key function selected. It can be also programmed to provide notification messages on alarms and pumps etc. For full details please refer to Display Parameters in the relevant parameter listing. Program Mode; displays menu and sub menu headings, parameter details and options.
4. Communicator status, this displays the status of remote PC
connection and velocity sensor operation.
5. Bar graph display, this gives visual indication of measurement in %
of span.
6. Level indicators:
Run Mode; indicates in which direction the level is moving. Program Mode; indicates at which level of the menu system you are
at.
There are two main operating modes for your FlowCERT Lite, Run Mode and Program Mode. There is also a Test Mode, used for checking the set-up. All modes are now described.

Run Mode

This mode is used once the FlowCERT Lite has been set up in program mode. It is also the default mode that the unit reverts to when it resumes operation after a power failure.
When the FlowCERT Lite is switched on for the first time, it will display, in metres, the distance from the transducer face to the target. All relays by default are switched off.
After programming is complete, any relays that are set will operate when the
level reaches the relevant setpoint, and the LED’s will change colour (unless
specifically switched off).
Page 31
PULSAR MEASUREMENT
31

Program Mode

This mode is used to set up the FlowCERT Lite or change information already set. You must use either the built-in keypad (standard) or, alternatively the unit can be set up with a PC via the RS232 Serial Interface.
Entering a value for each of the parameters that are relevant to your application provides all the programming information.

How to Access Program Mode

To enter program mode on the FlowCERT Lite, you simply enter the passcode, via the keypad, followed by the ENTER key. The default passcode is 1997, so you would press the following:
Important Information
There is a time-out period of 15 minutes when in program mode. After which time the run mode will resumed if you do not press any key.
ENTER
1
9 9 7
Page 32
FLOWCERT INSTRUCTION MANUAL
32

Hot Keys

There are five hot keys on the keypad, which can be used to quickly access common parameters for viewing only, while in Run Mode. Pressing the hot key once will display the first parameter, then repeated pressing will display the others, then the FlowCERT reverts to Run Mode. In program mode, they have different functions, the functions are shown below.
HOT KEY
RUN MODE
PROGRAM MODE
When application is Flow, you can view the non-resettable totalizer. View and reset the resettable totalizer.
Not used with FlowCERT Lite
Displays echo confidence, echo strength, height above loss limit (HALL), average noise, peak noise and temperature
Not used with FlowCERT Lite Not used with FlowCERT Lite
Reset parameter to default setting
Instantaneous mA output
Not used with FlowCERT Lite
Dependent on application. Displays distance, level, space, flow or rate of change of level, velocity (if velocity sensor attached)
Toggle relay setpoints between units of measure and % of span.
Not used with FlowCERT lite
Takes you to the last parameter edited when you first enter program mode.
Shows details of function type, firmware revision and serial number
Enter decimal point.
mA
n
Page 33
PULSAR MEASUREMENT
33

Menu Keys

The menu keys have the following functions:
HOT KEY
FUNCTION
1) Arrow keys for moving left and right around the menu
system.
2) Used in test mode to simulate the level moving up and
down.
1) Used to confirm each action (e.g. select a menu option)
Used to navigate up a level in the menu system, and back to run mode.
Used to cancel a value entered in error

Numeric Keys

These keys are used for entering numerical information during programming.
CANCEL
ENTER
1
2
3 4 5
6
7
8 9 0
.
+
-
Page 34
FLOWCERT INSTRUCTION MANUAL
34
There are two means of editing parameters, directly or using the menu system. Each is now described.

Using the Menu System

The menu system has been designed to make the changing of parameters very simple. There are two levels of menu: Main Menu and Sub Menu.
On the display, there is a line of text that displays the menu system. Pressing the arrow keys scrolls the display between the top-level menu items, (as the example shown below, starting at Quick Setup).
As you press the cursor keys to scroll left and right between these, you can press ENTER at any time, to select the desired menu heading, and take you to the sub-menu.
Each of these options, along with their sub-menus, are described later in this manual. When you move down into the sub-menu, you can scroll round using the arrow keys, press ENTER to go to the required section of parameters.
Once you have reached the relevant section, scroll through the parameters, and enter the necessary information. To enter the information, use the
numeric keys and then press ENTER, you will then see the message “Saved!”
If you press CANCEL, then the change you made will not be saved, and the message “Unchanged!!” will be displayed.
When you have finished, press CANCEL to go back to the previous level. When you have reached the top level, then the FlowCERT Lite will ask for confirmation before allowing you to go back into run mode. This is done by pressing ENTER at the display prompt.
Quick Setup
Application
Relays
Data Logs
OCM
Display
mA Output
Compensation
Stability
Echo Process
System
Device Comm
Test
Page 35
PULSAR MEASUREMENT
35

Directly Editing Parameters

If you already know the number of the parameter, that you wish to look at or edit, simply type the number in at any time while you are in the menu system. Thus, if you are in either the menu or sub-menu level by pressing a numeric key, you can enter the parameter number directly and jump straight there. You cannot type a parameter number whilst at parameter level, only at one of the two menu levels.
When you are at a parameter, the text line rotates automatically displaying the parameter name, number, the applicable units, and the maximum and minimum figure you can enter. The top line shows the value you are setting.
Once you have accessed a parameter, you can either just look at it, or change it.
Once a parameter has been changed, press ENTER and you will see the
message “Saved!”. If you press ‘CANCEL, then the change you made will not be saved, and the message “Unchanged!!” will be displayed.
You can jump straight to the last parameter you edited, by pressing ‘+/- when you first enter program mode.
Important Information
You can tell which part of the menu system you are in, as the up/down level indicators, (arrows) next to the bar graph will indicate as follows:
Top level menu: Down arrow on, to indicate you can move down.
Sub-menu: Up and Down arrows on, to indicate you can move up to the top
level, and down to parameter level.
Parameter Level: Up arrow on, to indicate you can move up to sub-menu
level.
Parameter Editing: No arrows on.
Page 36
FLOWCERT INSTRUCTION MANUAL
36

Test Mode

Test mode is used to simulate the application and confirm that all parameters and relay setpoints have been entered as expected. During simulation, there is a choice of whether the relays will change state (hard
simulation) or not (soft simulation), but the LED’s will always change colour
as programmed, and the mA output will change in accordance to the chosen mode of operation. If you wish to test the logic of the system that the relays
are connected to then select hard simulation, but if you do not wish to change the relay state, then select a soft simulation.
There are two simulation modes, automatic and manual. Automatic simulation will move the level up and down between empty level or the pre­determined Start Level (P983) and Pump/Control relay switch points, if you wish to change the direction of the level movement e.g. to go beyond relay setpoints, this can be done by using the arrow keys. In manual simulation, using the arrow keys will allow you to move the level up and down as required.
To enter simulation, first go to program mode. Using the menu system, select menu item ‘Test’, then sub-menu item ‘Simulation’. Simply change the value of the parameter P980 to one of the following:
1= Manual soft simulation 2= Automatic soft simulation 3= Manual hard simulation 4= Automatic hard simulation
To return to program mode, press CANCEL and test mode will end. When in manual simulation, by default test mode will move the level by
0.1m steps. Altering the increment (P981) will change this value. In automatic mode, the rate at which the level moves up and down is set by
the increment (P981 in metres, and the rate (P982) in minutes, which can be changed to make the level move up and down faster. E.g. if increment (P981) is set for 0.1m and rate (P982) is set to 1 min then the level will increase or decrease at a rate of 0.1m/min. To make the simulated level move slower, decrease the value in increment (P981) or increase the value in rate (P982). To make the simulated level move faster, increase the value in increment (P981) or decrease the value in rate (P982).
Page 37
PULSAR MEASUREMENT
37

Using the Serial Interface

The RS232 serial interface is used to communicate between the FlowCERT Lite and a PC using the optional Ultra PC and other associated Pulsar software packages, to obtain information such as data logging and view echo traces upload, download and save parameter files. In addition, it can also be used to control or obtain information using a standard PC or other computer base equipment. To do so, the settings for control are as follows: baud rate 19,200, 8 data bits, no parity, 1 stop bit.
The device should be connected as shown in Chapter 2 FlowCERT Lite
Installation.
To use the device remotely, you need to log on to start, and log off when finished. When logged on, FlowCERT Lite will show ‘Remote ON’ on the display, and “Communicator OFF” when logged off.
All commands should be followed by a carriage return. When logged on, the unit will respond either OK (or a value) if the command
is accepted, or NO if it is not. To log on, send the command /ACCESS:pppp where pppp is the passcode (P922). To log off, send the command /ACCESS:OFF To read a parameter value, send the command /Pxxx where xxx is the parameter you wish to read, and the FlowCERT will
respond with the parameter value. To set a parameter, send the command /Pxxx:yy where xxx is the parameter number, and yy is the value you wish to
set it to.
Page 38
FLOWCERT INSTRUCTION MANUAL
38
Other commands you can use are:
/LEVEL (shows current level)
/SPACE (shows current space)
/HEAD (shows current OCM head)
/FLOW (shows current OCM flow)
/TEMPERATURE (shows current temperature)
/CURRENTOUT1 (show the mA output 1 value)
/CURRENTOUT2 (show the mA output 2 value)
/CURRENTIN (show the mA input value)
/BACKUP1 (take backup of parameters to area 1)
/BACKUP2 (take backup of parameters to area 2)
/RESTORE1 (restore parameters from area 1)
/RESTORE2 (restore parameters from area 2)
Please consult Pulsar Measurement or contact your local Pulsar representative for further details and a full list of available commands.
Page 39
PULSAR MEASUREMENT
39

Parameter Defaults

When you first switch the FlowCERT Lite on, it will be reading the distance from the face of the transducer to the surface. It will be indicating in metres, as shown on the display. All relays are set OFF.
The date (P931) and time (P932) in the FlowCERT Lite were set at the factory, but may need checking, and amending if, for example the application is in a time zone other than GMT, see relevant parameter listing for full details.
In some applications, it is easier to empty the vessel, take a reading from the FlowCERT Lite for distance and then setup the empty level to this figure.
Once you are satisfied with the installation, and FlowCERT Lite is reading what you would expect in terms of distance from the face of the transducer to the material level, then you can proceed with programming, for the intended application. It is sensible to program all of the required parameters at the same time. The system will be then set-up.
Note: The span is automatically calculated from the empty level, so the empty level should be entered first.
Factory Defaults
When first installing the FlowCERT, or subsequently moving or using the unit on a new application, before proceeding to program the unit for its intended application it is recommended that you ensure that all parameters are at their default values by completing a Factory Default P930, as described in the relevant unit type parameter guide.
Page 40
FLOWCERT INSTRUCTION MANUAL
40

CHAPTER 4 QUICK SETUP GUIDE

This quick set-up guide shows you how to get up and running within a few minutes of installing your FlowCERT Lite.

Enter Program Mode

First you need to go from run mode into program mode. Assuming the passcode is the default 1997, then you should enter the following on the keypad:

Choose Quick Setup

Now you need to go into the quick setup. You will see on the menu the
words ‘Quick Setup’, which is the first item on the menu system. Try pressing
the two arrow keys to see some more menu options, but return to Quick Setup, and press
ENTER
1
9 9 7
ENTER
ENTER
This takes you to the ‘Quick Setup Menu’.
This takes you to the common applications menu, and several options will appear on the display.
Important Notice
If you have already setup a common application, then there will be a number shown other than 0, and you will see messages showing what the current setup is. If you want to reset this and start again, press 0 (which will reset all the quick setup parameters), otherwise pressing ENTER will allow you to edit the parameters that have been set.
Page 41
PULSAR MEASUREMENT
41

Choose Your Application

There are six categories of Primary Measuring Device, which are all described in this chapter. They are Exponential, BS3860 flumes, BS3860 weirs, Area/Velocity, Special and Universal.
Calculations for flow can be performed using absolute or ratiometric calculations. The answer will be the same, the choice of calculation method being limited to the amount of information available, with regards to the primary measuring device.
For ratiometric calculation it is normally sufficient to know the maximum flow at maximum head for the device in question. All types of primary measuring devices can be set up with a choice of alarms.
If you want to set-up a basic Exponential device, as described in the following example 1, then choose 1. You then need to select the PMD for your application from the following available options: Suppressed
Rectangular Weir, Cipolletti (trapezoidal) weir, Venturi flume, Parshall flume, Leopold Lagco flume, V notch weir, Contracted Rectangular weir
or other, for any other type of exponential device. To set-up an application for a BS3680 flume, as described in the following
example 2, then choose 2. You then need to select the PMD for your application from the following available options: Rectangular flume with or without hump, U-throated flume with or without hump.
To set-up an application for a BS3680 weir, as described in the following example 3, then choose 3. You then need to select the PMD device for your application from the following available options: Rectangular weir, V notch
full 90o (90degrees), V notch half 90o (53 degree 8 minutes), V notch quarter 90o (28 degree 4 minutes) or a Broad Crested weir.
To set-up an application for Area Velocity, then choose 4. You then need to select the PMD for your application from the following available options: U-
channel (circular bottom with straight sides), Rectangular channel, Trapezoidal channel, Round pipe or a Fixed pipe.
To set-up an application for a device contained in Special, choose 5. You then need to select the PMD for your application from the following available options: Palmer Bowlus flume, H-flume or a V notch, other than BS3680.
Page 42
FLOWCERT INSTRUCTION MANUAL
42
For devices which do not match any of the above devices the application can be setup using a Universal flow calculation, to select this option choose 6. You then need to select the PMD for your application from the following available options: Linear flow or Curved flow.
Once you have chosen your application you will be asked a series of questions which are answered by choosing the appropriate option as detailed in the flow chart below. Once all of the questions have been answered you will be prompted to provide further information, as detailed in the tables below, in order to complete the programming of the unit.
Page 43
PULSAR MEASUREMENT
43

Quick Setup

Universal
1 = Linear Flow 2 = Curved Flow
BS3680 Weirs
1 = Rectangular 2 = V-Notch 90° (Full 90°) 3 = U-Throat 4 = U-Throat with Hump
Exponent
1 = Rectangular 2 = Rect. with Hump 3 = U-Throat 4 = U-Throat with Hump 5 = Leopold 6 = V-Notch 7 = Other 8 = Contracted Rect.
Quick Setup
PMD Type
0 = Off (No calculation) 1 = Exponential 2 = BS3680 Flumes 3 = BS3680 Weirs 4 = Not Available 5 = Special 6 = Universal
BS3680 Flumes
1 = Rectangular 2 = Rect. with Hump 3 = U- Throat 4 = U-Throat with Hump
Special
1 = Palmer Bowlus 2 = H–Flume 3 = V-Notch Angle
Calculation
1 = Absolute
2 = Ratiometric
How Many Alarms
0 = No Alarms 1 = 1 Alarm 2 = 2 Alarms 3 = 3 Alarms 4 = Four Alarms 5 = Five Alarms
For Each Alarm
1 = High Alarm 2 = Low Alarm 3 = Hi Hi Alarm 4 = Lo Lo Alarm 5 = Loss Of Echo
For Each Alarm
1 = Set to Relay 1 2 = Set to Relay 2 3 = Set to Relay 3 4 = Set to Relay 4 5 = Set to Relay 5
Page 44
FLOWCERT INSTRUCTION MANUAL
44
Wait….
PARAMETER
DEFAULT
DESCRIPTION
P101 Transducer
1 = dB3
Select transducer to be used.
P706 Volume Units
1 = Litres
Units of flow on display and used for calculations. 1=litres 2=Cubic metres 3=Cubic feet 4=UK gallons 5=US gallons 6=Mil. USG
P707 Time Units
1 = Second
Units of time that flow units will be displayed and calculated in.
1=units/sec 2=units/min 3=units/hour 4=units/day
P104 Measurement Units
1 = Metres
Units used to enter dimensions and displayed where appropriate.
1=Metres 2=Centimetres 3=Millimetres 4=Feet 5=Inches
P105 Empty Level
2.70m
Distance from the face of the transducer (lower transducer on DUET) or end of the horn (dB Mach3) or lens face of a mmWAVE to the material at the bottom of the measuring element.
P703 Minimum Head
0.00m
Distance from empty point (P105) to zero flow.
P704 Maximum Head
2.575m
Distance from zero flow to max flow. It should be noted that any change from P704 updates P106 Span and vice versa
P824 Totaliser Enable
1 = On
Enables the flow totaliser, P820, options are 0=Off, 1=On. Note this totaliser can be viewed during run via the totaliser “Hot Key” It can be reset but only via P820, whilst in “program mode”.
P816 Totaliser (R)
0 = No
Enables the totaliser to be displayed in the
auxiliary display whilst in “run mode”. Note this totaliser is “resettable”, whilst in “run mode”, via
the totaliser hot key.
Page 45
PULSAR MEASUREMENT
45
PARAMETER
DEFAULT
DESCRIPTION
P823 Totaliser
Multiplier
4 = *1
Sets the factor by which the calculated volume will be divided or multiplied by before it is displayed.
1 = /1000 3 = /10 5 = *10 7 = *1000 9 = *100,000
2 = /100 4 = *1 6 = *100 8 = *10,000 10 = *1,000,000
The remaining parameters required to finalise the setup of your application will follow on immediately from the above. These parameters relate to details required to carry out the calculation for flow and will be dependent on the PMD chosen and the method of calculation chosen, please enter values for the parameters concerned as requested.
PARAMETER
DEFAULT
DESCRIPTION
P705 Max Flow
0.00
If Ratiometric is selected: When requested enter the known maximum flowrate, in units of volume (P706) and Time (P707) which occurs at maximum head (P704). If Absolute is selected: The maximum flowrate will be shown here as calculated by the unit.
P710 Dimension A
0
When requested enter, in measurement units, P104, the required dimension.
P711 Dimension B
0
When requested enter, in measurement units, P104, the required dimension.
P712 Dimension C
0
When requested enter, in measurement units, P104, the required dimension.
P717 Exponent
Dependent on chosen PMD
Where available the FlowCERT will automatically enter the default exponent value for the PMD chosen, but this can be changed if required. When P700=7 (Other), enter the exponent value as defined by the manufacturer of the PMD.
P718 K Factor
Enter the ‘K’ factor for the PMD, obtained from the manufacture’s specification
P719 Throat Width
Enter the Throat Width for the Parshall Flume, obtained from the manufacturer’s ‘specification
Page 46
FLOWCERT INSTRUCTION MANUAL
46
For More Options Hit Enter….
PARAMETER
DEFAULT
DESCRIPTION
P213 / P214 Relay 1 ON/OFF setpoints
depends on
application
Set required Alarm Setpoints.
P223 / P224 Relay 2 ON/OFF setpoints
depends on
application
Set required Alarm Setpoints.
P233 / P234 Relay 3 ON/OFF setpoints
depends on
application
Set required Alarm Setpoints.
P243 / P244 Relay 4 ON/OFF setpoints
depends on
application
Set required Alarm Setpoints.
P253 / P254 Relay 5 ON/OFF setpoints
depends on
application
Set required Alarm Setpoints.
P708 Flow Decimal
2
Set the number of decimal points required in the flow rate display
P709 Flow Cut Off
5.00%
Enter as a percentage of maximum flow, the minimum flow rate to be added to the totaliser.
P830 mA Out Range
2= 4 to 20
mA
What the mA output uses for the range. 0= Off, 1= 0 to 20 mA, 2= 4 to 20 mA, 3= 20 to 0 mA, 4= 20 to 4 mA.
P870 Fill Damping
10m/min
Rate of maximum fill rate (set above the actual fill rate of the vessel).
P871 Empty Damping
10m/min
Rate of maximum empty rate (set above the actual empty rate of the vessel).
Default values used for determining the relay setpoints, when setting Alarm relays, via the Quick Setup menu are entered as a % of span and are as follows:
PARAMETER
ALARM ID
ON SETPOINT
OFF SETPOINT
Alarm
Hi Hi
90%
85%
Alarm
High
85%
80%
Alarm
Low
10%
15%
Alarm
Lo Lo
5%
10%
Page 47
PULSAR MEASUREMENT
47

Exponential Devices (P700 = 1)

If the primary measuring device is a simple exponential device, then an exponent value is required. The FlowCERT will automatically enter the exponent value for the device chosen as detailed in the table below.
EXPONENT TYPE
PMD SHAPE EXAMPLE
EXPONENT (P717)
Suppressed Rectangular Weir (Without end contractions)
1.50, automatically set by the unit.
Cipolletti (Trapezoidal) Weir
1.50, automatically set by the unit.
Venturi Flume
1.50, automatically set by the unit.
Parshall Flume
Automatically calculated according to the throat size.
Leopold Lagco Flume
1.55
V-Notch Weir
2.50
Other
As per manufacturer
Value to be set as required.
Contracted Rectangular Weir (With end contractions)
1.50
Page 48
FLOWCERT INSTRUCTION MANUAL
48

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For Suppressed/Contracted Rectangular, Trapezoidal and V-notch, weirs, the head is measured upstream at a minimum distance of 3 times maximum head from the weir plate to ensure the surface of the liquid is not affected by turbulence or drawdown. (See DRWG. 1)
In the case of a Venturi flume the point of measurement should be 150 mm upstream from the beginning of the converging section and for a Parshall flume 2/3 the length of the converging section upstream of the throat section. (See DRWG 2 and 3).
3 x Hmax minimum
FLOW (q)
DRWG. 1
DRWG. 2
DRWG. 3
FLOW (q)
FLOW (q)
2/3 L
L
VENTURI FLUME
PARSHALL FLUME
Page 49
PULSAR MEASUREMENT
49
For a Leopold Lagco flume the head is measured at a point upstream of the beginning of the converging section as detailed in the table below. (See
DRWG 4).
FLUME SIZE
POINT OF MEASUREMENT
mm
inches
mm
inches
100 - 305
4 – 12 25
1.0
380
15 32
1.3
455
18 38
1.5
530
21 44
1.8
610
24 51
2.1
760
30 64
2.5
915
36 76
3.0
1065
42 89
3.5
1220
48 102
4.0
1370
54 114
4.5
1520
60 127
5.0
1675
66 140
5.5
1830
72 152
6.0
When any Other device is chosen please consult the manufacturer of the device for details of where the point of measurement should be located but
FLOW (q)
THROAT
DIVERGING
CONVERGING
TRANSDUCER MOUNTED MIN. BLANKING DISTANCE (P107) ABOVE MAX. HEAD
POINT OF MEASUREMENT
DRWG. 4
Page 50
FLOWCERT INSTRUCTION MANUAL
50
ensure that it is chosen such that the surface of the liquid is not effected by turbulence or drawdown.

Calculations

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula (s) as follows:
EXPONENT TYPE
FORMULA
EXPONENT
K FACTOR
Suppressed Rectangular Weir (Without end contractions)
Q = KLh* Where: Q=Flow K=K Factor L=Crest length of weir h=head *=exponent
1.50 Automatically selected by the FlowCERT
Automatically calculated, dependent on measurement, flow and time units chosen.
Cipolletti (Trapezoidal Weir)
Q = KLh* Where: Q=Flow K=K Factor L=Crest length of weir h=head *=exponent
1.50 Automatically selected by the FlowCERT
Automatically calculated, dependent on measurement, flow and time units chosen
Venturi Flume
Q=Kh* Where: Q=Flow K=K Factor h=head *=exponent
1.50 Automatically selected by the FlowCERT
Enter value of K Factor (P718) as required
Parshall Flume
Q=Kh* Where: Q=Flow K=K Factor h=head *=exponent
Automatically calculated, dependent on throat size (P719)
Automatically calculated, dependent on measurement, flow and time units chosen
Page 51
PULSAR MEASUREMENT
51
EXPONENT TYPE
FORMULA
EXPONENT
K FACTOR
Leopold Lagco Flume
Q=KD
0.0953hx
Where: Q =Flow K=K factor D=pipe diameter h=head *=exponent
1.55 Automatically selected by the FlowCERT
Automatically calculated, dependent on measurement, flow and time units chosen
V-Notch Weir
Q=Kh
x
Where: Q =Flow K=K factor h=head *=exponent
2.50 Automatically selected by the FlowCERT
Automatically calculated, dependent on measurement flow and time units chosen.
Other
Q=Kh*
Enter value as required
Enter value as required
Contracted Rectangular Weir (With end contractions)
Q=K(L-0.2*h)h*
1.50 Automatically selected by the FlowCERT
Automatically calculated, dependent on measurement flow and time units chosen.

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q = q
cal
(h/h
cal
)x
Where:
q = flowrate
q
cal
= flowrate at maximum head (705)
h = head
h
cal
= maximum head (P704)
x = exponent (determined as in absolute calculation above)
Page 52
FLOWCERT INSTRUCTION MANUAL
52

Example 1 V-Notch Weir

In this example it is required to calculate the flow through a Simple Exponential Device, which on this occasion is a V-Notch Weir. The K factor for the weir is unknown so ratiometric calculation will be used, there is no requirement for alarms and the flow rate is to be displayed in litres/second. The totaliser is to record the flow in cubic metres but is not to be displayed during RUN.
The distance from the face of the lower transducer (DUET) to zero flow (P105) is 1 metre and max head (P704) is 0.4 metres, maximum flow(P705) is known to be 96.5 litres/second.
Transducer mounted Min. blanking distance (P107) above max. head
h
Minimum 3 x hmax
TYPICAL SIMPLE EXPONENTIAL DEVICE
DRWG. 5
Page 53
PULSAR MEASUREMENT
53
To program the FlowCERT Lite for Example 1 V-Notch Weir by using the
Quick Setup menu proceed as follows. If required access the Program Mode , key in the passcode 1997 and press ENTER. Using the ‘right’ arrow
key, go to the Quick Setup menu press ENTER and as prompted, by the questions, select the relevant option and press ENTER.
QUESTION
OPTION
PMD Type
1 = Exponent
Exponent
6 = V-Notch
Calculation
2 = Ratiometric
No. of alarms
0 = No alarms
Transducer
1 = dB3
Volume units
1 = Litres
Time units
1 = Per second
Measurement units
1 = Metres
Empty Level
1.00m
Minimum head
0.00m
Maximum head
0.40m
Totaliser enabled
1 = On
Totaliser (R)
1 = Yes
Totaliser multiplier
4 = *1
Maximum flow
96.5
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the FlowCERT Lite will return to run mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively, access the setpoints by the main menu system or directly accessing the parameters.
Page 54
FLOWCERT INSTRUCTION MANUAL
54

BS3680 Flumes (P700 = 2)

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For a Rectangular and U-throated flume, the head is measured at 3 to 4
times the maximum head upstream from the beginning of the converging section, to ensure the surface of the liquid is not affected by turbulence.
(See DRWG 6)
3-4 x hmax
DRWG. 6
THROAT LENGTH
DIM C (P712)
DIM A (P710)
APPROACH WIDTH
DIM B (P711)
THROAT WIDTH/DIA.
FLOW (q)
BS3680 FLUME
Page 55
PULSAR MEASUREMENT
55

Calculations

Rectangular Flume

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5gn0.5
CsCvCdbh
1.5
Where:
q = flowrate gn = gravitational acceleration (nominal value = 980.66 cm/s2)
Cs = shape coefficient (value = 1) Cv = velocity coefficient calculated by FlowCERT (P721) Cd = discharge coefficient calculated by FlowCERT (P722) b = throat width P711
h = head

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal(Cv/Cvcal
)(Cd/C
dcal
)(h/h
cal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by FlowCERT (P721)
C
vcal
= velocity coefficient at maximum head
Cd = discharge coefficient calculated by FlowCERT (P722) C
dcal
= discharge coefficient at maximum head
h = head
h
cal
= maximum head P704
Page 56
FLOWCERT INSTRUCTION MANUAL
56

U-Throated Flume

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5
g
n
0.5
CuCvCdbh
1.5
Where:
q = flowrate gn = gravitational acceleration (nominal value = 980.66 cm/s2) h = head
Cu= shape coefficient calculated by FlowCERT (P724) Cv = velocity coefficient calculated by FlowCERT (P721) Cd = discharge coefficient calculated by FlowCERT (P722) b = throat width P711

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula:
q= q
cal(Cv/Cvcal
)(Cd/C
dcal
)(Cu/C
ucal
)(h/h
cal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by FlowCERT (P721)
Cv
cal
= velocity coefficient at maximum head
Cd = discharge coefficient calculated by FlowCERT (P722) Cd
cal
= discharge coefficient at maximum head
Cu = shape coefficient P724 Cu
cal
= shape coefficient at maximum head
h = head h
cal
= maximum head P704
Page 57
PULSAR MEASUREMENT
57

Example 2 BS3680 U-Throated Flume

In this example, it is required to calculate to BS3680 the flow through a U­Throated Flume without any hump. Absolute calculation will be used, and there is a requirement for an alarm to indicate a low flow condition which will be set to relay 1. The flow rate is to be displayed in cubic meters/hour and the totaliser is also to record the flow in cubic metres, the resettable totaliser is to be displayed during RUN.
The distance from the face of the lower transducer (DUET) to zero flow (P105) is 35 inches and max head (P704) is 15 inches, maximum flow(P705) will be calculated by the FlowCERT as 0.1784 Mil USG/day
The dimensions of the flume are as follows:
Approach Channel diameter (Dim “A”) P710 = 28 inches Throat diameter (Dim “B”) P711 = 20 inches Throat length (Dim “C”) P712 = 40 inches
Page 58
FLOWCERT INSTRUCTION MANUAL
58
To program the FlowCERT Lite for Example 2 BS3680 U-Throated Flume by using the Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER. Using the ‘right’ arrow key, go to the Quick Setup menu press ENTER and as prompted, select the relevant option and press ENTER.
QUESTION
OPTION
PMD Type
2 = 3680 Flume
3680 Flumes
3 = U-Throat
Calculation
1 = Absolute
No. of alarms
1 = 1 Alarm
Type alarm 1
2 = Low
Alarm No.1
1 = Set to relay 1
Transducer
1 = dB3
Volume units
2 = Cubic metres
Time Units
4 = Per hour
Measurement units
1 = Metres
Empty level
1.00m
Minimum head
0.00m
Maximum head
0.40m
Totaliser enabled
1 = On
Totaliser (R)
1 = Yes
Totaliser multiplier
7 = 1000
Approach dia. (Dim A)
0.70m
Throat dia. (Dim B)
0.50m
Throat length (Dim C)
1.00m
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the FlowCERT will return to Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively, access the setpoints by the main menu system or directly accessing the parameters.
Page 59
PULSAR MEASUREMENT
59

BS3680 Weirs (P700 = 3)

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107.
For a Rectangular and V-notch weir, the head is measured at a point 4 to 5 times the maximum head upstream from the weir plate, to ensure the surface of the liquid isn’t affected by turbulence or drawdown. (See DRWG 7)

Calculations

BS3680 Rectangular Weir

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce 2/3(2gn)
0.5
beh
e
1.5
Where: q = flowrate Ce = discharge coefficient calculated by FlowCERT (P723)
gn = gravitational acceleration (nominal value = 980.66 cm/s2) be =effective approach width where b is approach width (Dim “A”) P710
he = effective head
4-5 x hmax minimum
DRWG. 7
BS3680 WEIR
Page 60
FLOWCERT INSTRUCTION MANUAL
60

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
calCe/Cecal(he/hecal
)
1.5
Where: q = flowrate
q
cal
= flowrate at maximum head P705
Ce = discharge coefficient calculated by FlowCERT (P723)
Ce
cal
= discharge coefficient at maximum head he = effective head he
cal
= effective head at maximum head

BS3680 V-Notch Weir

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce8/15tan(theta/2)(2gn)
0.5h2.5
Where:
q = flowrate Ce = discharge coefficient calculated by FlowCERT (P723) theta = V-notch angle gn = gravitational acceleration (nominal value = 980.66 cm/s2) h = head
The FlowCERT pre-sets the angle (theta) on selection of the chosen device this angle is 90° for a BS 3680 full 90°V notch weir, 53° 8 minutes in the case of the BS3680 half 90°V notch weir and 28° 4 minutes in the case of the BS3680 quarter 90°V notch.

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q = q
calCe
(h)/Ce(h
cal
)(h/h
cal
)
2.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Ce(h) = discharge coefficient for head Ce(h
cal
) = discharge coefficient for maximum head
h = head
h
cal
= maximum head P704
Page 61
PULSAR MEASUREMENT
61

Example 3 BS3680 Rectangular Weir

In this example it is required to calculate to the flow through a BS3680 Rectangular weir. Absolute calculation will be used, and there is a requirement for an alarm to indicate a high flow condition to be set to relay
3. The flow rate is required to be displayed in litres/minute and the totaliser is to record the flow in cubic metres, the resettable totaliser is to be displayed during RUN. The transducer to be used is a dB3.
The distance from the face of the transducer to zero flow (P105) is 1 metre and max head (P704) is 0.4 metres, maximum flow (P705).
Approach width (Dim “A”) P710 = 0.5 m Crest width (Dim “B”) P711 = 0.3 m Crest Height (Dim “C”) P712 = 0.3 m
Page 62
FLOWCERT INSTRUCTION MANUAL
62
To program the FlowCERT Lite for Example 3 BS3680 Weir by using the
Quick Setup menu proceed as follows. If required access the Program Mode, key in the passcode 1997 and press ENTER. Using the ‘right’ arrow
key, go to the Quick Setup menu press ENTER and as prompted, select the relevant option and press ENTER.
QUESTION
OPTION
PMD Type
3 = 3680 Weir
3680 Flumes
1 = Rectangular
Calculation
1 = Absolute
No. of alarms
1 = 1 Alarm
Type alarm 1
1 = High
Alarm No.1
3 = Set to relay 3
Transducer
1 = dB3
Volume units
1 = Litres
Time Units
2 = Per minute
Measurement units
1 = Metres
Empty level
1.00m
Minimum head
0.00m
Maximum head
0.40m
Totaliser enabled
1 = On
Totaliser (R)
1 = Yes
Totaliser multiplier
7 = 1000
Approach dia. (Dim A)
0.50m
Crest Width. (Dim B)
0.30m
Crest Height (Dim C)
0.30m
Programming is now complete, and the unit can now be returned to the run mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the FlowCERT will return to Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively, access the setpoints by the main menu system or directly accessing the parameters.
Page 63
PULSAR MEASUREMENT
63

BS3680 Rectangular Broad Crested Weir

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107.
The head is measured at a point 3 to 4 times the maximum head upstream from the weir crest, to ensure the surface of the liquid is not affected by turbulence or drawdown.

Calculations

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = (2/3)
1.5
Ceb(gh3)
0.5
Where:
q = flowrate Ce = discharge coefficient calculated by FlowCERT (P723) b = approach width P710
g = gravitational acceleration (nominal value = 980.66 cm/s2) h = head
Page 64
FLOWCERT INSTRUCTION MANUAL
64

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
calCe/Cecal(he/hecal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Ce = discharge coefficient calculated by FlowCERT (P723)
Ce
cal
= discharge coefficient at maximum head he = effective head he
cal
= effective head at maximum head
Page 65
PULSAR MEASUREMENT
65

Special Devices (P700 =5)

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107. In the case of a Palmer Bowlus flume the point of head measurement should be half the value of Dim “A” P710 upstream of the device. For an H-Flume the head measurement is taken at a point downstream from the flume entrance as detailed in the table below:
FLUME SIZE-DIM A (P710)
POINT OF MEASUREMENT
cm
feet
cm
inches
15.25
0.5
4.7
1.88
23.00
0.75
6.7
2.69
30.05
1.0
9.1
3.63
45.70
1.5
13.5
5.378
61.00
2.0
17.9
7.19
76.20
2.5
22.5
9.00
91.45
3.0
27.2
10.88
137.15
4.5
40.5
16.19
Transducer mounted Min. blanking distance (P107) above max. head
Point of Measurement
See table below
FLOW (q)
Page 66
FLOWCERT INSTRUCTION MANUAL
66
V-notch angle weirs, the head is measured upstream of the weir plate at a minimum distance of 3 times maximum head to ensure the surface of the liquid is not affected by turbulence or drawdown. See Exponential devices, above, for further details.

Calculations

Absolute

Palmer Bowlus Flume and H-Flume - If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = f(h)
Where:
q = flowrate f = is an 8th degree polynomial solution for h (head)

Ratiometric

Palmer Bowlus Flume and H-Flume - If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal
f(h)/f(h
cal
)
Where:
q = flowrate q
cal
= flowrate at maximum head P705 f(h) = a polynomial solution for h (head) f(h
cal
) = a polynomial solution for h
cal
(maximum head)
V-Notch Angle Weir (Non-BS 3680)

Absolute

If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = Ce 8/15 tan (theta/2)(2gn)
0.5
(h = kh)
2.5
Where:
q = flowrate Ce = discharge coefficient calculated by FlowCERT (P723) theta = V-notch angle gn = gravitational acceleration h = head kh = compensated head
Page 67
PULSAR MEASUREMENT
67

Ratiometric

If the flow calculation is to be ratiometric P702 = 2 the flow will be calculated using the formula: q= q
cal
(h+kh/h
cal
+kh)
2.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
h = head kh = compensated head

Universal Calculations (P700=6)

Point of Measurement

The transducer must be above the maximum head P704 by at least the near blanking distance P107. For all Universal calculation applications, the point at which the head is measured should be chosen such that the surface of the liquid is not affected by turbulence.

Calculations

Absolute

Universal Head Vs Flow - If the flow calculation is to be absolute P702 = 1 the flow will be calculated using the formula: q = q(h) Where:
q = flowrate q(h) = flowrate for head
The desired number of Breakpoints, (P730 - P793) are to be entered in pairs in values of head and corresponding flow. (Minimum of 2 pairs of Breakpoints is required).
Page 68
FLOWCERT INSTRUCTION MANUAL
68

CHAPTER 5 PARAMETER GUIDE

This chapter describes all the parameters in your FlowCERT, as they appear in the menu system.

Menu System

Shown below is a set of charts to show you how all the various functions and features can be found using the menu system.
For further details and a full description of all parameters refer to the Parameter Listings and Descriptions section of this chapter.

Top Level Menu

Quick Setup
Application
Relays
Data Logs
OCM
Display
mA Output
Compensation
Stability
Echo Process
System
Device Comm
Test
Page 69
PULSAR MEASUREMENT
69

Application Menu

Operation
Distances
P100
Mode
P101
Transducer
P104
Measurement
Units
P102
Material
P105
Empty Level
P106 Span
P107
Near Blanking
P108
Far Blanking
Page 70
FLOWCERT INSTRUCTION MANUAL
70

Relays Menu

Relay 1
P210 Type
P211
Function
P212
Alarm ID
or
Pump
Group
P214
Set 2
P213 Set 1
P216
Allocat.
P217
Closures
P218
Fail Safe
P219
R1
Max.Rate
If
P210=2
Relay 2
Relay 3
Relay 4
Relay 5
P220 Type
P230 Type
P240 Type
P250 Type
P221
Function
P231
Function
P241
Function
P251
Function
P222
Alarm ID
or
Pump
Group
P232
Alarm ID
or
Pump
Group
P242
Alarm ID
or
Pump
Group
P252
Alarm ID
or
Pump
Group
P223 Set 1
P233 Set 1
P243 Set 1
P253 Set 1
P224
Set 2
P234
Set 2
P244
Set 2
P254
Set 2
P226
Allocat.
P236
Allocat.
P246
Allocat.
P256
Allocat.
P227
Closures
P237
Closures
P247
Closures
P257
Closures
P228
Fail Safe
P238
Fail Safe
P248
Fail Safe
P258
Fail Safe
P249
R4
Max.Rate
If
P240=2
P259
R5
Max.Rate
If
P250=2
P229
R2
Max.Rate
If
P220=2
P239
R3
Max.Rate
If
P230=2
P215 Set 3
As
required
P225 Set 3
As
required
P235 Set 3
As
required
P245 Set 3
As
required
P255 Set 3
As
required
Page 71
PULSAR MEASUREMENT
71

Data Logs Menu

Totaliser
Audit
Temperature
P460
Total Date 1
P461
Totaliser 1
P462, 464,
466, 468, 470,
472, 474, 476
Total Dates
2 to 9
P463, 465,
467, 469, 471,
473, 475, 477
Totalisers
2 to 9
P478
Total Date 10
P580
Min. Temp
P479
Totaliser 10
P581
Min. Temp.
Date
P582
Min. Temp.
Time
P583
Max. Temp.
P584
Max. Temp.
Date
P585
Max. Temp.
Time
P586
Current
Temperature
Page 72
FLOWCERT INSTRUCTION MANUAL
72

OCM Menu

PMD
Setup
Dimensions
Calcs
Break
Points
Tables
Average
Flow
P700 PMD
Type
P701
Primary
Measuring
Device
P702
Calculation
P703
Minimum
Head
P704
Maximum
Head
P705
Maximum
Flow
P706
Volume
Units
P707
Time Units
P708
Flow
Decimal
P709
Flow Cutoff
P710
Dimension
“A”
P711
Dimension
“B”
P712
Dimension
“C”
P713
Dimension
“D”
P714
Roughness
Coefficient
P715
Water
Temperature
P717
Exponent
Value
P719
Parshall
Throat Width
P718
K Factor
P720 Area
P721
Calc. 1
P722
Calc. 2
P723
Calc. 3
P724
Calc. 4
P725
Calc. 5
P792 Head
Bkpt 32
P731
Flow
Bkpt. 1
P732–P791
Head
&
Flow
Bkpts.
2 to 31
P730 Head
Bkpt. 1
P793 Flow
Bkpt 32
P796
Reset
Bkpts.
P797
Number
of
Bkpts.
Set
P863
Average
Flow
P864
Average
Time
Page 73
PULSAR MEASUREMENT
73

Display Menu

mA Output Menu

Range
Operation
Setpoint
Limits
Trim
Fail Safe
Allocation
P830
mA Out
Range
P831
mA Out
Mode
P834
Low
Value
P836
Low
Limit
P835
High
Value
P838
Low
mA
Trim
P840
Fail
Mode
P841
Allocation
P837
High
Limit
P839 High
mA
Trim
Options
Fail Safe
Auxiliary
P802
Display
Offset
P801
Decimal
Places
P804
Display
Conversion
P808
Fail Mode
P809
Fail Time
P810 Units
P811
Alarms
P812
Pumps
P813
Control
P814 Misc.
P800
Display
Units
Bargraph
Totaliser
P829
Bargraph
P820
Totaliser
P821
Totaliser
(R)
P822
Totaliser
Decimal
P823
Totaliser
Multiplier
P824
Totaliser
Enable
P816
Totaliser
(R)
Page 74
FLOWCERT INSTRUCTION MANUAL
74

Compensation

Stability Menu

Damping
Indicator
Rate
Filters
P870
Fill
Damping
P871
Empty
Damping
P872
Fill
Indicator
P873
Empty
Indicator
P874
Rate Update
P875
Rate Time
P876
Rate Distance
P877
Rate Value
P878
Lower Cutoff
P880
Gate Mode
P881
Fixed Distance
P882
Process Filter
P884
Peak Percent
P860
Sound
Velocity
Offset
Temperature
Velocity
P851
Measurement
Offset
P852
Temperature
Source
P854
Fixed
Temperature
Page 75
PULSAR MEASUREMENT
75

Echo Processing Menu

Xdr. 1 Status
P900
Xdr. 1 Status
P901
Echo
Confidence
P902
Echo Strength
P903
Average Noise
P904
Peak Noise
P905
Sensitivity
P906
Side Clearance
Page 76
FLOWCERT INSTRUCTION MANUAL
76

System Menu

Passcode
Backup
System
Info
Date
&
Time
LED
Colour
Daylight
Saving
P921
Enable Code
P925
Parameter
Backup
P926
Software
Revision
P931
Date
P935
Off
Colour
P970
DST
Enable
P922
Passcode
P927
Hardware
Revision
P928
Serial
Number
P929
Site
Ident.
P930
Factory
Default
P932 Time
P933
Date
Format
P936
Alarm
Colour
P937
Pump
Colour
P938
Control
Colour
P939
Misc.
Colour
P971
DST
Difference
P972
DST
Start Time
P973
Start Day
P974
Start Week
P975
Start
Month
P976
DST
End Time
P977
End Day
P978
End Week
P979
End Month
Page 77
PULSAR MEASUREMENT
77

Device Comm Menu

RS485 Setup
(Optional)
If Comms.
Type is
Modbus
P130
Device
Mode
P131
Protocol
P132
Device
Addr.
P133
Device
Baud
RS232
Setup
P61
Comms Baud
Remote
Alarm
P144
Call Type
P145
Tel. No. 1
P146
Tel. No. 2
P147
Tel. No. 3
P134
Parity
P135
Stop Bit
P136
Data
Format
P137
Delay in
milli sec.
If Comms.
Type is
Profibus
P132
Device
Addr.
P143
Addr. Lock
Page 78
FLOWCERT INSTRUCTION MANUAL
78

Test Menu

Simulation
P980
Simulate
P981
Increment
Hardware
P990
Self-Test
P991
Hard Test
P992
mA Out Test
P993
mA In Test
P994
Xdr. Test
P982
Rate
P983
Start Level
P984
Inc. Change
Page 79
PULSAR MEASUREMENT
79

CHAPTER 6 PARAMETER LISTING AND DESCRIPTION

Application Parameters

Operation

P100 Mode of Operation

This parameter sets the mode of operation, when in run mode, and can be set to one of the following:
OPTION
DESCRIPTION
1 = Distance (Default)
Display shows the distance from the transducer face to the surface of the material measured.
2 = Level
Shows how full a vessel is.
3 = Space
Shows how empty a vessel is.
4 = OCM Head
Display shows how high the head is.
5 = OCM Flow
Display shows the instantaneous flow.

P101 Xducer (Transducer)

This parameter should be set to the transducer being used with the unit, and can be set to one of the following:
OPTION
DESCRIPTION
1 = dB3 (Default)
Transducer is a dB3. Range = 0 to 3.00 metres
2 = dB6
Transducer is a dB6. Range = 0.3 to 6.00 metres
3 = dB10
Transducer is a dB10. Range = 0.3 to 10 metres
4 = dB15
Transducer is a dB15. Range = 0.5 to 15.00 metres
7 = dBS6
Transducer is a dBS6. Range = 0.2 to 6.00 metres
8 = dBMach3
Transducer is a dBMach3. Range 0.3 to 1.50 metres
*11 = dBR16
Transducer is a mmWAVE radar. Range = 0.077 to 16 metres
*12 = dBR8
Transducer is a mmWAVE radar. Range = 0.077 to 8 metres
Important Notice
*Please consult your local Pulsar distributor for the versions of firmware that the mmWAVE radars are available in.
Page 80
FLOWCERT INSTRUCTION MANUAL
80

P102 Material

This parameter should be set to the type of material being monitored.
OPTION
DESCRIPTION
1 = Liquid (Default)
Used for liquids and flat solid materials.
2 = Solid
Used for sold material that is heaped or at an angle.
3 = Closed Tank
Use for applications within a closed tank or where a secondary echo response may become focused to create a larger echo than the first.

Dimensions

P104 Measurement Units

This parameter sets the units you want to use for programming and display.
OPTION
DESCRIPTION
1 = metres (Default)
All units of measurement are Metres
2 = cm
All units of measurement are Centimetres
3 = mm
All units of measurement are Millimetres
4 = feet
All units of measurement are Feet
5 = inches
All units of measurement are Inches

P105 Empty Level

This parameter is to be set to the maximum distance from the face of the transducer to the empty point, in P104 Measurement Units. Note this value affects span as well, (see the following important information notices), so should be set before span.
Important Notice
When changing the Empty Distance (P105) you can also recalculate the values for the Span so that it equals the empty distance (P105) minus Near Blanking (P107) and the Relay Setpoints, so that they remain at the same percentage values of the empty distance as they were before you changed the empty distance (P105). You’ll be asked the question “Recalculate Span?” Yes (enter 1), Span is then recalculated. Any other answer will leave the span at its original value. You’ll then be asked if you want to “Recalculate Setpoints?”, if you choose yes (enter 1), all Relay Setpoints will be recalculated as a % of the new empty distance. Any other answer will leave the setpoints at their original values.
Page 81
PULSAR MEASUREMENT
81

P106 Span

This parameter should be set to the maximum distance from the Empty Level (P105) to the maximum material level. It is automatically set to be
equal to the Empty Level (P105) less the Near Blanking distance (P107) when you set the empty level.

P107 Span

This parameter is the distance from the face of the transducer that is not measurable and is pre-set to the minimum value dependant on the Transducer (P101) selected. It should not be set to less than this figure, but can be increased, typically to ignore close in obstructions.
TRANSDUCER
NEAR BLANKING DISTANCE
P101 = dBMach3
Default blanking distance = 0.00 metres
P101 = dB3
Default blanking distance = 0.125 metres
P101 = dB6
Default blanking distance = 0.30 metres
P101 = dB10
Default blanking distance = 0.30 metres
P101 = dB15
Default blanking distance = 0.50 metres
P101 = dBS6
Default blanking distance = 0.20 metres
P101 = dBR16
Default blanking distance = *0.077 metres
P101 = dBR8
Default blanking distance = *0.077 metres
*The signal emanates from the curved face of the radar, but for the purposes of measurement it is taken from the drip shield.

P108 Far Blanking Distance

This is the distance (as a percentage of the empty level P105) beyond the empty point that the unit will be able to measure, and by default is pre-set to 20% of the empty level.
If the surface being monitored can be extended beyond the empty level (P105) then the far blanking distance can be increased to a max. of 100% of the empty level, provided it does not exceed the max range of the transducer being used. This parameter is always entered as a % of the empty level.
Important Notice
When using the dB Mach 3 the empty distance is measured from the end of the horn to the empty point in P104 Measurement Units.
Page 82
FLOWCERT INSTRUCTION MANUAL
82

Relay Parameters

All relay related parameters are prefixed with a 2**. The second digit of the three-figure parameter number denotes the relay
number as follows:
21* parameters for Relay 1 22* parameters for Relay 2 23* parameters for Relay 3 24* parameters for Relay 4 25* parameters for Relay 5
The third digit selects specific parameter for the setting of the relays, which can be selected individually and results in the following parameter numbers for each relay:
Relay 1 210 to 218 Relay 2 220 to 228 Relay 3 230 to 238 Relay 4 240 to 238 Relay 5 250 to 238

P210, 220, 230, 240, 250 - Relay Type

This parameter defines what type each relay should be, see the table below for available options:
OPTION
DESCRIPTION
0 = Not in use (Default)
Relay is not in use or programmed and the LED will always be off.
1 = Alarm
Relay is programmed as an alarm relay, which will de- energise ON, and energise OFF. This will ensure an alarm is raised if the power fails to the unit.
2 = Pump
Relay is programmed as a pump relay, which will energise ON, and de-energise OFF.
3 = Control
Relay is programmed as a control relay, which will energise ON, and de-energise OFF.
4 = Miscellaneous
Relay is programmed as a miscellaneous relay, which will energise ON, and de-energise OFF.
Page 83
PULSAR MEASUREMENT
83

Alarms

P210, 220, 230, 240, 250 = 1 (Alarm) The second parameter for each relay determines the function of the alarm.

P211, 221, 231, 241, 251 – Relay Function

OPTION
DESCRIPTION
0 = Off (Default)
Relay will not operate.
1 = Level
Alarm is based on the level in the vessel, and the type of level alarm (P212, 222, 232, 242, 252) and two setpoints must be set (P2b3, 223, 233, 243, 253 & 214, 224, 234, 244,
254). Setpoints are entered in display units or % of span as referenced to empty level.
2 = Rate of Change
Alarm is based on the rate of change of level in the vessel, and the type of rate of change alarm (212, 222, 232, 242,
252), and two setpoints must be set (213, 223, 233, 243, 253 & 214, 224, 234, 244, 254). Setpoints are entered in Display Units per minute or % of span per minute and a negative value should be entered for a Rate Alarm on a de-creasing level, and a positive value for an increasing level.
3 = Temperature
Alarm is based on the temperature, and the type of temperature alarm (P212, 222, 232, 242, 252) and two setpoints must be set (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254). The temperature used depends on the temperature source selected (P852). Setpoints are entered in
o
C.
4 = Loss of Echo
Alarm is raised if the Failsafe Timer (P809) expires. No setpoints are required.
5 = Loss of Clock
Alarm is raised if the expires. No setpoints are required.
13 = Loss of Comms
Alarm is raised if there is a comms failure. No setpoints are required. Available in firmware version 7.5.0 and greater
Important Notice
The Loss of Echo, Loss of Clock and Loss of Comms will also be shown on the display as “Lost Echo”, Lost Clock” and Lost Comm” respectively.
Page 84
FLOWCERT INSTRUCTION MANUAL
84
The third parameter for each relay determines the Alarm ID for the relay you wish to set.

P212, 222, 232, 242, 252 – Relay Alarm ID

When P211, 221, 231, 241, 251 =4 (Loss of Echo) or 5 (Loss of Clock) This parameter has no function and will not be displayed.
P211, 221, 231, 241, 251 = 1 (Level), 2 = Rate of Change or 3 (Temp) This parameter defines which alarm type, or identification, the relay should
respond to, as follows:
OPTION
DESCRIPTION
SETPOINTS
1 = General (Default)
Relay goes “ON” when the
value reaches the ON
setpoint and goes “OFF”
when the value reaches the OFF setpoint.
P213, 223, 233 is ON Setpoint. P214, 224, 234 is OFF Setpoint
2 = High
Relay goes “ON” when the
value rises to the ON
setpoint and goes “OFF”
when the value lowers to the OFF setpoint.
ON> OFF Relay Setpoints P213, 223, 233 and P214, 224, 234 Setpoints, can be set in any order as the unit
‘knows’ that you are
setting a high-level alarm.
3 = HiHi
Same as 2 = High, but different identifier
4 = Low
Relay goes “ON” when the
value lowers to the ON
setpoint and goes “OFF”
when the value rises to the OFF setpoint.
ON<OFF Relay Setpoints P213, 223, 233 and P214, 224, 234. Setpoints, can be set in any order as the
unit ‘knows’ that you are
setting a low-level alarm.
5 = LoLo
Same as 4 = Low, but different identifier
Page 85
PULSAR MEASUREMENT
85
OPTION
DESCRIPTION
SETPOINTS
6 = In bounds
Relay goes “ON” if value is
inside the zone between the two setpoints.
Relay Setpoints, P213, 223, 233 and P214, 224, 234 can be set in any order as the unit ‘knows’ that you are setting an inbounds alarm.
7 = Out of bounds
Relay goes “ON” if value is
outside the zone between the two setpoints.
Relay Setpoints P213, 223, 233 and P214, 224, 234 can be set in any
order as the unit ‘knows’
that you are setting an out of bounds alarm.
The fourth parameter and the fifth parameter for each relay set the Alarm
ON” and “OFF” points. For a high alarm, the “ON” is set higher than OFF”. For low alarm then “ON” is set lower than OFF”. See the
appropriate alarm ID, table (P212, 222, 232) for further information.
When P211, 221, 231, 241, 251 = 4 (Loss of Echo) or 5 (loss of Clock)
This parameter has no function and will not be displayed.
When P211, 221, 231, 241, 251 = 1 (Level), 2 (Rate of Change), 3 (Temp), 11 (Flow) or 12 (Velocity).

P213, 223, 233, 243, 253 – Relay Setpoint 1

Determines the ON’ or ‘OFF’ point for the alarm according to the ID selected.

P214, 224, 234, 244, 254 – Relay Setpoint 2

Determines the ‘ON’ or ‘OFF’ point for the alarm according to the ID selected.
To set figures in % press the hotkey to show and enter % figure, relative to empty level.
Page 86
FLOWCERT INSTRUCTION MANUAL
86
Important Notice
Setpoints are entered in values according to the function selected. Level - entered in Display Units or % of span as referenced to Empty Level.
Rate of Change - entered in Display Units per minute or % of span per minute.
For an alarm on an increasing level enter setpoints as a positive value, for an alarm on a decreasing level enter setpoints as a negative value. Temperature - entered in ºC.
See the appropriate alarm function, table (P211, 221, 231, 241 and 251) for further information.
Page 87
PULSAR MEASUREMENT
87

Pumps

P210, 220, 230, 240, 250 = 2 (Pump)

When a relay is being used for a pump function, the second parameter determines the pump duty that will be used to determine the operating cycle.

P211, 221, 231, 241, 251 – Relay Function

This parameter defines which pump duty the relay should respond to as follows:
PUMP DUTY
DESCRIPTION
0 = Off (Default)
Relay is always de-energised
1 = Fixed duty assist
All pumps are used to assist each other (run at the same time) and each pump has its own setpoints. (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254).
2 = Fixed duty backup
If a pump fails to meet the demand (due to malfunction, intake blockage etc.), then it is stopped, and another pump shall take over. Each pump has its own setpoints. (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254).
3 = Alternate duty assist
All pumps are used to assist each other (run at the same time). Each pump has its own setpoints, (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254) but each time all pumps have stopped, the setpoints are sequentially rotated between the pumps to ensure equal pump use.
4 = Alternate duty backup
If a pump fails to meet the demand (due to malfunction, intake blockage etc.), then it is stopped, and another pump shall take over. Each pump has its own setpoints, (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254) but each time all pumps have stopped, then the setpoints are sequentially rotated between the pumps to ensure equal pump use.
Page 88
FLOWCERT INSTRUCTION MANUAL
88
PUMP DUTY
DESCRIPTION
5 = Duty backup and assist
First pump comes on, if it cannot cope, it goes off and next pump comes on (duty backup). This continues until the last pump comes on and if it cannot cope the first pump comes back on to assist the last pump (duty assist) if the level continues to rise all other pumps will come on (assist) in turn until the level decreases to the pump off points. Each pump has its own setpoints, (P213, 223, 233, 243, 253 & P214, 224, 234, 243, 253).
6 = Service ratio duty assist
All pumps are used to assist each other (run at the same time) and each pump has its own setpoints (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254). And a service ratio setting. The third setpoint (P215, 225, 235, 245, 255) is used to set the service ratio. Each time a pump is required to start then the pump with the least running hours (with respect to the service ratio) is started (i.e. the setpoints are re-assigned accordingly). For example, if two pumps A and B have the service ratio set to 2 and 1 respectively, then pump A will operate for twice as many hours as pump B.
7 = Service ratio duty backup
If a pump fails to meet the demand (due to malfunction, intake blockage and so on), then it is stopped, and another pump shall take over. Each time a pump is required to start then the pump with the least running hours (with respect to the service ratio) is started (i.e. the setpoints are re-assigned accordingly). Each pump has its own setpoints (P213, 223, 233, 243, 253 & P214, 224, 234, 244, 254). The third setpoint (P215, 225, 235, 245, 255) is used to set the service ratio. For example, if two pumps A and B have the service ratio set to 2 and 1 respectively, then pump A will operate for twice as many hours as pump B.
Page 89
PULSAR MEASUREMENT
89
PUMP DUTY
DESCRIPTION
8 = First On First Off Alternate duty assist
The first pump switched on is the first pump to be switched off, etc. regardless of the set points, so the setpoints are dynamically changed to enable this.
9 = Service Ratio Standby
When a service ratio duty is being used, on all other pumps in use, the standby pump can be started on a ratio basis only, when it will assume the setpoints of the next pump to start. The third setpoint (P215, 225, 235, 245, 255) is used to set the service ratio.
10 = Two Pump Sets
There are four pumps. Two rotate their start-up sequence with each other. If the two pumps cannot keep up, the level rise to the setpoints of the other two pumps which take over and rotate their sequence with each other.
The third parameter for each relay determines the pump group. You can have two groups of pumps, and all similar duties within that group will operate together.

P212, 222, 232, 242, 252 – Relay Pump Group

By default, all pump groups are set to 1, but if you want to have another group, then set this parameter to 2, for each pump relay that should operate together as part of a second group.
Important Notice
The pumps are started and stopped at the “ON” and “OFF” setpoints. To pump down (reduce level) then set “ON” higher than “OFF”. To pump up (increase level) then set “ON” lower than “OFF”.
Page 90
FLOWCERT INSTRUCTION MANUAL
90
The fourth parameter and the fifth parameter for each relay set the pump ON” and “OFF” points, which are entered in Measurement units P104. For
pump down the “ON” is set higher than OFF”. For pump up then “ON” is
set lower than OFF”. See the appropriate pump duty, function table (P212, 222, 232, 242, 252) for further information.

P213, 223, 233, 243, 253 – Relay Setpoint 1

This parameter determines the ‘ON’ point of the pump.

P214, 224, 234, 244, 254 – Relay Setpoint 2

This parameter determines the ‘OFF’ point of the pump.
The sixth parameter will determine the service ratio that will be used to switch the pump, when pump duty selected is a Service Ratio duty.

P215, 225, 235, 245, 255 – Relay Setpoint 3

This parameter determines the Service Ratio in values of %. See the appropriate pump duty function, table (P211, 221, 231, 241, 251), for further information.

P219, 229, 239, 249, 259 – Relay Max Rate

This parameter will allow a pump to be switched at a pre-determined Rate of change of Level, irrespective of the “ON” level setpoint P213, 223, 233, 243, 253. Once a pump relay has been switched “ON” by the pre-determined Rate of Change, it will remain energised until the level reaches the “OFF
level setpoint P214, 224, 234, 244, 254. Max. Rate is entered in Measurement Units (P104) per minute and can be
entered as either positive (increasing level) or negative (decreasing level) values.
Page 91
PULSAR MEASUREMENT
91

Control

P210, 220, 230, 240, 250 = 3 (Control)
When a relay is being set up as a control relay, the second parameter that will be displayed in the menu determines its function.

P211, 221, 231, 241, 251 – Relay Function

This function allows the relay to be assigned to specific control functions (other than pumps and alarms) several of these functions work in relation to time.
This can be used to activate devices based on elapsed time or running cycles, such as a timed rake control to keep a ram lubricated if idle for long periods, or flush valve operation.
PUMP DUTY
DESCRIPTION
0 = Off (Default)
Relay is always de-energised
1 = Time
Relay will energise ON” after the Cycle time that is set in Relay Setpoint 2 (P214, 224, 234, 244, 254). And turns “OFF”, de-energises, after the On-Time Period that is set in Relay Setpoint 1 (P213, 223, 233, 243, 253)
2 = Step Time
Step Time Control allows relays to be used to control a device, such as a motorised valve or gate, in order to maintain the level within two predetermined points. Relays will energise ON” when Step Time condition is in effect and de-energises OFF” when Step Time goes off. One relay will be required to control an increase in
level, (‘open’ the device) and a second relay is required
to control a decrease in level, (‘close’ the device). Alarm ID (P212, 222, 232, 242, 252) is used to assign the relay
to control either the open or close condition. Step Time Control relay requires three setpoints. The first set point (P213, 223, 233, 243, 253) determines the level, at which the relay is to be activated, (N.B. level setpoint for open relay, increase the level, must be lower than the setpoint for the close relay, decrease the level). The relay will energise ON” after the Limit time that is set in Relay Setpoint 3 (P215, 225, 235, 245, 255). And turns “OFF”, de-energises, after the Drive Period that is set in Relay Setpoint 2 (P214, 224, 234, 244, 254).
Page 92
FLOWCERT INSTRUCTION MANUAL
92
The third parameter for each relay determines the assignment or condition of the relay, where required.

P212, 222, 232, 242, 252 – Relay Alarm ID/Pump Group

P211, 221, 231, 241, 251 = 2 (Step Time)
If the relay is selected for Step Time, then this parameter is used to assign the relay to the 0 = Open condition (increase level) or 1 = Close condition (decrease level).
The fourth parameter, fifth parameter and sixth parameter are set to determine the switch points, “ON” and “OFF” for the relay and where required the order of start. See control function, table (P211, 221, 231, 241, 251) for further information.

P213, 223, 233, 243, 253 – Relay Setpoint 1

P211, 221, 231, 241, 251 = 1 (Time)
This parameter determines the “Time Period” that the relay will remain “ON”. Relay Setpoints are entered in Minutes. See the appropriate relay function
tables (P211, 221, 231, 241, 251) for further information.
P211, 2210, 231, 241, 251 = 2 (Step Time)
Relay Setpoint 1 is entered in values of Measurement Units (P104) See the appropriate relay function tables (P211, 221, 231, 241, 251) for
further information.

P214, 224, 234, 244, 254 – Relay Setpoint 2

P211, 221, 231, 241, 251 = 1 (Time)
This parameter determines the “Cycle Time” for the operation of the relay. See the appropriate relay function tables (P211, 221, 231, 241, 251) for
further information.
P211, 2210, 231, 241, 251 = 2 (Step Time)
Relay Setpoints are entered in Seconds to set Drive Period, the time that the relay will remain ON’.
See the appropriate relay Function tables (P211, 221, 231, 241, 251) for further information.
Page 93
PULSAR MEASUREMENT
93

P215, 225, 235, 245, 255 – Relay Setpoint 3

P211, 221, 231, 241, 251 = 2 (Step Time)
This parameter is used to determine the Limit Time between each Drive Period. Relay Setpoints are entered in Minutes, during which time the relay will remain OFF.
See the appropriate relay Function tables (P211, 221, 231, 241, 251) for further information.

Miscellaneous

P210, 220, 230, 240, 250 = 4 (Miscellaneous)

When a relay is set to be a miscellaneous relay, the second parameter determines its function.

P211, 221, 231, 241, 251 – Relay Function

This function allows the relay to work in relation to a clock or a specific event and will be set to activate in relation to Real Time.
PUMP DUTY
DESCRIPTION
0 = Off (Default)
Relay is always de-energised
1 = Clock
Relay will energise ON at a specified time each day as set in Relay Setpoint 1 (P213, 223, 233, 243, 253). And turns OFF, de-energises, after the specified-On Time period as set in Relay Setpoint 2 (P214, 224, 234, 244, 254).
2 = Totaliser
Relay will energise ON momentarily each time the specified volume has passed as set in Relay setpoint 1 (P213, 223, 233, 243, 253), this parameter sets the multiplication factor which will be applied to the on board totaliser (P820) to determine the switch point of the relay. E.g. if the totaliser is set to totalise in cubic metres and the relay is required to provide a closure every 10,000 litres Relay setpoint 1 would be set to 10. Relay setpoint 2 (P214, 224, 234, 244, 254) is used to select the time the relay will remain closed in seconds.
Important Notice
The pumps are started and stopped at the “ON” and “OFF” setpoints. To pump down (reduce level) then set “ON” higher than “OFF”. To pump up (increase level) then set “ON” lower than “OFF”.
Page 94
FLOWCERT INSTRUCTION MANUAL
94
The third parameter has no function when miscellaneous relay is chosen and will not be displayed.
The fourth parameter, and fifth parameter, are set to determine the switch points, “ON” and “OFF” for the relay. See miscellaneous function table (P211, 221, 231, 241, 251) for further information.
P210, 220, 230, 240, 250 = 1 (Clock)

P213, 223, 233, 243, 253– Relay Setpoint 1

Relay Setpoints are entered in Hours & Minutes (HH:MM) to set Time at which relay will energise. Default = 00:00 (HH:MM)

P214, 224, 234, 244, 254– Relay Setpoint 2

Relay Setpoints are entered in seconds to set the time period that the relay
will remain ‘ON’. Default = 0.00 secs.
P210, 220, 230, 240, 250 = 2 (Totaliser)

P213, 223, 233, 243, 253– Relay Setpoint 1

Relay Setpoints are entered as a factor by which the on board totaliser (P820) should be multiplied by to provide a relay closure. Default = 0.00

P214, 224, 234, 244, 254– Relay Setpoint 2

Relay Setpoints are entered in seconds to set the time period that the relay will remain ‘ON’. Default = 0.00 secs.

P219, 229, 239, 249, 259 – Relay Max Rate

This parameter will allow a pump to be switched at a pre-determined Rate of change of Level, irrespective of the “ON” level setpoint P213, 223, 233, 243, 253. Once a pump relay has been switched “ON” by the pre-determined Rate of Change, it will remain energised until the level reaches the “OFF
level setpoint P214, 224, 234, 244, 254. Max. Rate is entered in Measurement Units (P104) per minute and can be
entered as either positive (increasing level) or negative (decreasing level) values.
Page 95
PULSAR MEASUREMENT
95

Common Relay parameters

P216, 226, 236, 246, 256 – Relay Allocation

This parameter determines which input the relay will act on. You can set it to the transducer (default), or the optional auxiliary (current) input, or an average of the two. In most cases, this will not need to be changed from the default.
OPTION
DESCRIPTION
1 = Xducer 1 (Default)
Relay acts on Xducer 1 calculated values

P217, 227, 237, 247, 257 – Relay Closures

The FlowCERT will record how many times each relay is closed, this parameter displays the number of times the relay has activated since the relay has been in use. It can be reset with any value.

P218, 228, 238, 248, 258 – Relay Fail Safe

The FlowCERT has a general fail-safe parameter P808. However, this can be overridden so that each individual relay has its own independent fail-safe mode.
This parameter determines what the relay will do in the event of the Failsafe
Time (P809) expiring.
OPTION
DESCRIPTION
0 = Default
Relay assumes system default mode set in P808
1 = Hold
Relay remains in its current state
2 = De-energise
Relay will De-energise
3 = Energise
Relay will energise
Page 96
FLOWCERT INSTRUCTION MANUAL
96

Data Log parameters

The data log parameters contain the following information:

Totaliser Audits

P460 to P479 Total Audits

Parameters P460-P479 show the date and total flow for the last ten days, the first on the list are the most recent and last ones are the oldest. When all ten total audits are full the oldest is pushed out and all totals increment through to allow the new days total to be registered in the first day’s total audit parameter allocation.

P480 Clear Logs

This parameter enables all of the Total Audits (P460 – P479) to be cleared to factory default values.

Temperature

The following parameters give information on temperature conditions seen by the Temperature source (P852) in ºC. All of these parameters are read only and cannot be changed, though if P852 is changed they will be reset.

P580 Minimum Temperature

This parameter displays the minimum temperature recorded.

P581 Minimum Temperature Date

This parameter displays the date when the minimum temperature was recorded.

P582 Minimum Temperature Time

This parameter displays the time when the minimum temperature was recorded.

P583 Maximum Temperature

This parameter displays the maximum temperature recorded.
Important Notice
The pumps are started and stopped at the “ON” and “OFF” setpoints. To pump down (reduce level) then set “ON” higher than “OFF”. To pump up (increase level) then set “ON” lower than “OFF”.
Page 97
PULSAR MEASUREMENT
97

P584 Maximum Temperature Date

This parameter displays the date when the maximum temperature was recorded.

P585 Maximum Temperature Time

This parameter displays the time when the maximum temperature was recorded.

P585 Current Temperature

This parameter displays the current temperature.
Page 98
FLOWCERT INSTRUCTION MANUAL
98

OCM Parameters

PMD Setup

P700 Primary Measuring Device Type

This parameter is used to select the type of Primary Measuring Device and enable additional parameters required to calculate the flow of the particular Primary Measuring Device chosen (P701). Options are as follows:
0 = Off (Default)
1 = Exponent 2 = BS3680 Flume 3 = BS3680 Weir 4 = Area/Velocity 5 = Special 6 = Universal

P701 Primary Measuring Device

Enter the Primary Measuring Device used:
If P700 = 1 Exponent
1 = Suppressed Rectangular Weir 2 = Trapezoidal (Cipolletti) Weir 3 = Venturi Flume 4 = Parshall Flume 5 = Leopold Lagco Flume 6 = V-Notch Weir 7 = Others
8 = Rectangular Weir with End contractions
If P700 = 2 BS3680 Flume
1 = Rectangular 2 = Rectangular with hump 3 = U-throated 4 = U-throated with hump
If P700 = 3 BS3680 Weir
Page 99
PULSAR MEASUREMENT
99
1 = Rectangular 2 = V-Notch Full 90 degree (full 90°) 3 = V-Notch 53 degree 8’ (half 90°) 4 = V-Notch 28 degree 4’ (quarter 90°) 5 = Broad Crested (Rectangular) Weir
If P700 = 5 Special
1 = Palmer Bowlus Flume 2 = H-Flume 3 = V-Notch angle (other than BS3680)
If P700 = 6 Universal
Where the Primary Measuring device does not match any of the devices contained in the above categories then a universal volume calculation can be performed. A head Vs flow chart is used, to enter a number of Breakpoints for head and flowrate (P730-793), which is either provided by the manufacturer or created based on the dimensions of the device.
1 = Universal Linear calculation 2 = Universal Curved flow calculation

P702 Calculation

Select the required calculation method, both will give the same answer, but the difference is the information required to complete the calculation. For ratiometric it is normally sufficient to know the maximum flow at the maximum head. Choose between:
1 = Absolute 2 = Ratiometric (Default)

P703 Minimum Head

This parameter is used to enter the distance, above empty, that represents zero head and flow. This feature is used in Primary Measuring Devices
where the zero reference is at a higher level than the channel bottom, at the point of measure. Enter distance in Measurement Units P104.
Page 100
FLOWCERT INSTRUCTION MANUAL
100

P704 Maximum Head

Enter the head value that represents maximum flow, enter in Measurement Units P104.
Note any change to the value of this parameter will be reflected in P106 (Span) and vice versa.

P705 Maximum Flow

When P702 = 2 Ratiometric enter the flow rate value that occurs at maximum head (P704), enter in volume units (P706) per time units
(P707). When P702 = 1 Absolute, and all relevant flow parameters have been
entered, the maximum flow that occurs at maximum head P704 will be calculated, after the unit is returned to RUN mode, and displayed in this parameter in volume units (P706) per time units (P707).
Loading...