Pulmonetic Systems LTV User manual

®
LTV
Series Ventilator
Operator’s Manual
Document Revision History
Document Revision Level Revision Date
Rev A, ECO 390 January 1999 Rev B, ECO 450 February 1999 Rev C, ECO 478 June 1999 Rev D, ECO 685 October 1999 Rev E, ECO 774 March 2000 Rev F, ECO 868 September 2000
Rev G, ECO 956 December 2000
Rev H, ECO 1127 February 2001
Rev J, ECO 1299 May 2001
Rev K, ECO 1496 October 2001
Rev L, ECO 1722 April 2002
Rev M, ECO 1795 June 2002
Rev N, ECO 1978 August 2002
Contact Information
Pulmonetic Systems, Inc.
17400 Medina Rd., Suite 100
Minneapolis, Minnesota 55447-1341 Phone: (763) 398-8300 Customer Care Center: (800) 754-1914 Fax: (763) 398-8400 E-mail: info@pulmonetic.com Website: http://www.pulmonetic.com
LTV 1000, LTV 950, LTV 900, LTV 800 and LTV Pulmonetic Systems, Inc.
Copyright © 1997, 1998, 1999, 2000, 2001, 2002 Pulmonetic Systems, Inc., Minneapolis, Minnesota.
®
are trademarks belonging to
Operator’s Manual LTV® Series Ventilator Page i p/n 10664, Rev. N
Warranty
Pulmonetic Systems warrants that the LTV® Series Ventilator is free from defects in material and workmanship for a period of one (1) year from the date of shipment, or 8,800 hours as measured on the usage meter, whichever comes first, with the following limitations:
1) Patient circuit components, including hoses, exhalation manifold, and other
associated parts are warranted for sixty (60) days from date of shipment.
Pulmonetic Systems will, at its option, either repair, replace, or issue credit for products that prove to be defective during the warranty period.
For warranty service or repair, the product must be returned to Pulmonetic Systems or a service facility designated by Pulmonetic Systems, shipping prepaid by the Buyer.
LIMITATION OF WARRANTY
Ordinary maintenance, as specified in the LTV
®
Series Ventilator Operator’s and Service
Manuals, is not covered under the forgoing warranty. The forgoing warranty does not apply to defects resulting from:
1) Improper or inadequate maintenance of the unit;
2) Improper use or misuse of the unit;
3) Unauthorized modifications or repairs to the unit;
4) Use of the unit with unauthorized accessories, e.g. external battery or AC adapter.
5) Operation of the unit outside the specified environment.
NO IMPLIED WARRANTIES
This warranty is exclusive. There are no other warranties expressed or implied.
LIMITATION OF LIABILITY
Pulmonetic Systems shall not be liable for loss of profits, loss of use, consequential damages, or any other claim based on breach of warranty. Pulmonetic Systems, Inc.’s liability for damages of any kind shall be limited to the purchase price of the defective unit.
Page ii LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Notices
The LTV® Series Ventilator complies with limitations as specified in IEC 601-1-2 for Medical Products. It does however, use and radiate radio frequency energy.
The function of this machine may be adversely affected by the operation of other nearby equipment, such as high frequency surgical diathermy equipment, short-wave therapy equipment, defibrillators or MRI equipment.
CAUTION: Federal law restricts this device to sale by or on the order of a physician.
FM 45841
BS EN ISO 9001/EN 46001
BS ISO 13485
European Regulatory Requirements per 93/42/EEC Medical Device Directives
Pulmonetic Systems, Inc.’s European Representative for vigilance reporting within the European Community is:
BREAS
®
MEDICAL
Företagsvägen 1
435 33 Mölnlycke Sweden
Phone:+46 31 86 88 00
Fax:+46 31 86 88 10
E-mail: breas@breas.se
Website: http://www.breas.com
Any product malfunctioning issues that fall under Medical Device Directives Essential Requirements should be directed to BREAS MEDICAL.
Operator’s Manual LTV® Series Ventilator Page iii p/n 10664, Rev. N
Notice To Operators
Unsafe Operation - Operating the LTV® Series Ventilator without a complete and thorough understanding of its attributes is unsafe and may cause harm to the patient. It is important that this manual be read and understood in its entirety before operating the ventilator.
Warnings and Cautions Section - Read the section on Warnings and Cautions carefully before operating the LTV® Series Ventilators.
Use and Maintenance - Any questions regarding installing, operating, or maintaining the LTV® Series Ventilators, should be directed to a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
Avis important
Fonctionnement dangereux - L'opération d'un ventilateur de la série LTV® sans une excellente compréhension de ses attributs est dangereuse et risque de blesser le patient. Il est très important de lire et de comprendre entièrement ce manuel avant de faire fonctionner le ventilateur.
Section Avertissements et Attention - Lire attentivement la section Avertissements et Attention avant de procéder à l'opération des ventilateurs de la série LTV®.
Utilisation et entretien - En cas de questions concernant l'installation, l'opération ou l'entretien des ventilateurs de la série LTV®, veuillez vous adresser à un technicien de service certifié de Pulmonetic Systems ou directement à Pulmonetic Systems, Inc.
Page iv LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Table of Contents
Chapter 1 - Introduction .............................................................. 1-1
Operator’s Safety Information...................................................................................................1-2
Warnings................................................................................................................................... 1-3
Cautions.................................................................................................................................. 1-11
Symbols .................................................................................................................................. 1-17
Chapter 2 - Ventilator Overview.................................................. 2-1
Intended Use ............................................................................................................................ 2-1
Power/Supplies Required .........................................................................................................2-2
Information/Assistance .............................................................................................................2-3
Chapter 3 - Breath Types ............................................................ 3-1
Breath Types ............................................................................................................................ 3-1
Volume Control Breaths ........................................................................................................ 3-2
Pressure Control Breaths...................................................................................................... 3-3
Pressure Support Breaths..................................................................................................... 3-5
Spontaneous Breaths............................................................................................................ 3-6
Chapter 4 - Ventilation Modes..................................................... 4-1
Control Mode ............................................................................................................................ 4-1
Assist/Control Mode.................................................................................................................. 4-1
SIMV Mode ............................................................................................................................... 4-2
CPAP Mode .............................................................................................................................. 4-4
Apnea Backup .......................................................................................................................... 4-5
NPPV ........................................................................................................................................ 4-6
Volume / Pressure Ventilation ..................................................................................................4-7
Bias Flow .................................................................................................................................. 4-7
Chapter 5 - Using the Controls and Indicators ........................... 5-1
Ventilator Controls .................................................................................................................... 5-1
Setting a Control....................................................................................................................... 5-2
Variable Controls ...................................................................................................................... 5-2
Buttons...................................................................................................................................... 5-3
Set Value Knob......................................................................................................................... 5-3
Extended Features ................................................................................................................... 5-3
Mechanical Controls ................................................................................................................. 5-4
Bright, Dim and Blank Control Displays.................................................................................... 5-4
Flashing Controls...................................................................................................................... 5-5
Dashes...................................................................................................................................... 5-5
Control Limiting......................................................................................................................... 5-5
Control Locking......................................................................................................................... 5-6
Control Retention...................................................................................................................... 5-6
Operator’s Manual LTV® Series Ventilator Page v p/n 10664, Rev. N
Chapter 6 - Controls .................................................................... 6-1
Assist/Control / SIMV/CPAP Modes......................................................................................... 6-1
Breath Rate............................................................................................................................... 6-2
Control Lock.............................................................................................................................. 6-3
High Pressure Limit .................................................................................................................. 6-4
Inspiratory / Expiratory Hold Control Button ............................................................................. 6-5
Inspiratory Hold ..................................................................................................................... 6-6
Expiratory Hold...................................................................................................................... 6-8
Inspiratory Time ...................................................................................................................... 6-10
Low Minute Volume ................................................................................................................ 6-11
Low Pressure.......................................................................................................................... 6-12
Low Pressure O2 Source (Option) .......................................................................................... 6-13
Manual Breath ........................................................................................................................6-17
O2 % (Option).......................................................................................................................... 6-18
On / Standby........................................................................................................................... 6-20
PEEP Valve ............................................................................................................................ 6-21
Pressure Control (Option).......................................................................................................6-22
Pressure Support.................................................................................................................... 6-24
Select...................................................................................................................................... 6-26
Sensitivity................................................................................................................................ 6-27
Set Value Knob....................................................................................................................... 6-28
Silence / Reset........................................................................................................................ 6-29
Tidal Volume........................................................................................................................... 6-30
Volume / Pressure Mode (Option).......................................................................................... 6-31
Chapter 7 - Displays and Indicators ............................................ 7-1
Displays .................................................................................................................................... 7-1
Airway Pressure........................................................................................................................ 7-1
Display Window ........................................................................................................................ 7-1
Indicators .................................................................................................................................. 7-1
Battery Level............................................................................................................................. 7-2
Charge Status........................................................................................................................... 7-4
External Power ......................................................................................................................... 7-5
NPPV ........................................................................................................................................ 7-6
Patient Effort ............................................................................................................................. 7-6
Vent Inop .................................................................................................................................. 7-6
Chapter 8 - Monitored Data ......................................................... 8-1
Automatic or Manual Data Display Scrolling ............................................................................ 8-2
PIP xxx cmH2O ......................................................................................................................... 8-3
MAP xx cmH2O......................................................................................................................... 8-3
PEEP xx cmH2O ....................................................................................................................... 8-3
f xxx bpm .................................................................................................................................. 8-3
Vte xxx ml ................................................................................................................................. 8-3
Page vi LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
VE xx.x L................................................................................................................................... 8-3
I:E xx:xx ...................................................................................................................................8-4
xxx Lpm ......................................................................................................................8-4
Chapter 9 - Ventilator Alarms...................................................... 9-1
APNEA, APNEA xx bpm........................................................................................................... 9-2
BAT EMPTY ............................................................................................................................. 9-3
BAT LOW.................................................................................................................................. 9-5
DEFAULTS ............................................................................................................................... 9-6
DEFAULTS SET ....................................................................................................................... 9-8
DISC/SENSE ............................................................................................................................ 9-9
HIGH O2 PRES ...................................................................................................................... 9-10
HIGH PRES ............................................................................................................................ 9-11
HW FAULT ............................................................................................................................. 9-12
INOP ....................................................................................................................................... 9-13
LMV OFF ................................................................................................................................ 9-14
LMV LPPS OFF ...................................................................................................................... 9-15
LOCKED ................................................................................................................................. 9-16
LOW MIN VOL........................................................................................................................ 9-17
LOW O2 PRES ....................................................................................................................... 9-18
LOW PRES............................................................................................................................. 9-19
LPPS OFF .............................................................................................................................. 9-20
NO CAL DATA, NO CAL Monitor Display .............................................................................. 9-21
POWER LOST........................................................................................................................ 9-22
POWER LOW ......................................................................................................................... 9-23
REMOVE PTNT...................................................................................................................... 9-24
RESET.................................................................................................................................... 9-25
WARMUP xx........................................................................................................................... 9-26
XDCR Fault............................................................................................................................. 9-27
Chapter 10 - Extended Features ............................................... 10-1
Navigating the Extended Features Menus .............................................................................10-2
Alarm Operations.................................................................................................................... 10-3
Alarm Volume...................................................................................................................... 10-3
Apnea Interval ..................................................................................................................... 10-3
High Pressure Alarm Delay................................................................................................. 10-4
Low Peak Pressure Alarm................................................................................................... 10-4
Exit....................................................................................................................................... 10-4
Vent Operations...................................................................................................................... 10-5
Operator’s Manual LTV® Series Ventilator Page vii p/n 10664, Rev. N
Variable Rise Time.............................................................................................................. 10-6
Variable Flow Termination................................................................................................... 10-7
Variable Time Termination .................................................................................................. 10-8
Pressure Control Flow Termination..................................................................................... 10-9
Leak Compensation .......................................................................................................... 10-10
NPPV Mode....................................................................................................................... 10-11
Control Unlock................................................................................................................... 10-12
Language Selection........................................................................................................... 10-12
Software Version............................................................................................................... 10-13
Usage Meter...................................................................................................................... 10-13
Communications Setting ................................................................................................... 10-13
Set Date............................................................................................................................. 10-14
Set Time ............................................................................................................................ 10-15
Date Format ...................................................................................................................... 10-15
PIP LED............................................................................................................................. 10-16
Model Number / Serial Number......................................................................................... 10-16
Valve Home Position......................................................................................................... 10-17
Set Defaults....................................................................................................................... 10-17
Exit..................................................................................................................................... 10-17
Transducer Autozero......................................................................................................... 10-18
Airway Pressure Transducer Autozero ............................................................................. 10-18
Bi-directional Flow Transducer Differential Autozero........................................................ 10-19
Exhalation Flow Transducer Differential Autozero - Narrow ............................................. 10-20
Exhalation Flow Transducer Differential Autozero - Wide ................................................ 10-21
Real Time Transducers..................................................................................................... 10-22
Chapter 11 - Ventilator Checkout Tests ................................... 11-1
Alarm Test .............................................................................................................................. 11-4
Display Test ............................................................................................................................ 11-5
Control Test ............................................................................................................................ 11-7
Leak Test ................................................................................................................................ 11-9
Vent Inop Alarm Test............................................................................................................ 11-11
Set Defaults .......................................................................................................................... 11-13
Exit........................................................................................................................................ 11-14
Chapter 12 - Operating Procedure ............................................ 12-1
Procedure for Turning the Ventilator On................................................................................. 12-1
Before Connecting the Ventilator to a Patient ........................................................................ 12-2
Procedure for Control Mode Set Up ....................................................................................... 12-4
Procedure for Assist / Control Mode Set Up........................................................................... 12-5
Procedure for SIMV Mode Set Up .......................................................................................... 12-6
Procedure for CPAP Mode Set Up ......................................................................................... 12-7
Procedure for NPPV Mode Set Up ......................................................................................... 12-7
Procedure for NPPV Mode Set Up ......................................................................................... 12-8
Page viii LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Procedure for Turning the Ventilator Off................................................................................. 12-9
LTV® Ventilator Settings Checklist........................................................................................ 12-10
Chapter 13 - Cleaning, Disinfecting and Sterilizing .................. 13-1
Cleaning the Ventilator ...........................................................................................................13-1
Cleaning the Fan Filter ........................................................................................................... 13-2
Cleaning the Inlet Filter........................................................................................................... 13-3
Cleaning the Exhalation Valve and Reusable Patient Circuit................................................. 13-4
Chapter 14 - Power and Battery Operation............................... 14-1
Using the AC Adapter............................................................................................................. 14-2
Using An External Battery....................................................................................................... 14-3
Using The Automobile Cigarette Lighter Adapter................................................................... 14-6
Replacing the Adapter Fuse................................................................................................ 14-9
Using the LTV/LTM Power Splitter Cable Assembly ............................................................ 14-10
Caring For The Internal Battery ............................................................................................ 14-11
Battery Disposal.................................................................................................................... 14-11
Chapter 15 - Troubleshooting.................................................... 15-1
Displays and Buttons.............................................................................................................. 15-2
Ventilator Performance........................................................................................................... 15-5
Power and Battery Operation ............................................................................................... 15-14
Alarms................................................................................................................................... 15-16
Checkout Test Failures......................................................................................................... 15-20
Test Lung Operations ........................................................................................................... 15-22
Appendix A - Ventilator Specifications ....................................... A-1
Appendix B - Set Up / Maintenance............................................. B-1
Recommended Maintenance Schedule....................................................................................B-1
Service Assistance ...................................................................................................................B-2
Appendix C - Installation and Checkout .....................................C-1
Installation and Setup ...............................................................................................................C-1
Operator’s Manual LTV® Series Ventilator Page ix p/n 10664, Rev. N
Unpacking the Ventilator – Instructions.................................................................................C-1
Protective Boots ....................................................................................................................C-2
Protective Boot Removal.......................................................................................................C-3
Protective Boot Installation....................................................................................................C-6
LTM/LTV Mounting Bracket...................................................................................................C-9
LTM/LTV Mounting Bracket Installation ..............................................................................C-10
LTM/LTV Mounting Bracket Removal .................................................................................C-13
Patient Breathing Circuit – Connection Instructions............................................................C-16
Oxygen Lines – Connection Instructions.............................................................................C-18
Patient Assist Call System – Connection Instructions ........................................................C-19
Communications Port..........................................................................................................C-20
LTM™ Graphics Monitor .....................................................................................................C-20
Using the Remote Alarm Cable...........................................................................................C-21
Checking the Ventilator for Proper Operation .....................................................................C-22
Ventilator Proper Operation Worksheet ..............................................................................C-24
Appendix D - Principles of Operation .......................................... D-1
Overview...................................................................................................................................D-1
Appendix E - Event Trace ............................................................ E-1
Event Codes .............................................................................................................................E-3
Event Codes by Code # ........................................................................................................E-3
Event Codes by Event Name ................................................................................................E-5
Appendix F - Glossary.................................................................. F-1
Appendix G - Index ...................................................................... G-1
Page x LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Chapter 1 - INTRODUCTION
This Operator’s Manual contains detailed information and instructions which when adhered to, ensure the safe and effective set up, use and simple maintenance of the
®
LTV
Series Ventilators.
It is designed for use by Respiratory Therapists or other qualified, and trained personnel under the direction of a physician and in accordance with applicable state laws and regulations. It contains the following:
Ventilator Overview
Installation and Checkout
Using the Controls and Indicators
Monitored Data
Ventilator Alarms
Extended Features
Ventilator Checkout tests
Operating Procedure
Troubleshooting
Cleaning, Disinfecting and Sterilizing
Set Up / Maintenance
Power and Battery Operation
Service tests, calibration, and major maintenance operations are described in the LTV Series Ventilator Service Manual (P/N 10665).
®
Operator’s Manual LTV® Series Ventilator Page 1-1 p/n 10664, Rev. N
Operator’s Safety Information
All Operators are to read and understand the following information about Warning, Caution and Note statements before operating the LTV
®
Series Ventilator.
, WARNING !
Warning” statements contain information about circumstances or practices that may cause serious and undesirable results, or expose the patient or operator to danger.
, MISE EN GARDE
Les « Avertissements » contiennent des informations relatives à des situations ou des pratiques pouvant entraîner des résultats graves et indésirables, ou pouvant exposer le patient ou l'opérateur à des dangers.
, Caution !
Caution” statements contain information about circumstances or practices that may result in equipment damage.
, Avertissement
Les énoncés « Attention » contiennent des informations à propos des circonstances ou des pratiques pouvant causer des dommages à l'équipement.
ª Note
Note” statements contain additional information to assist in the proper operation of the LTV® Series Ventilators.
ª Remarque
Les énoncés « Remarque » contiennent des informations supplémentaires pour aider à l'opération adéquate des ventilateurs de la série LTV.
Page 1-2 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Warnings
, WARNING !
Untrained Personnel – Only properly trained personnel should operate the ventilator. The LTV Series Ventilator is a restricted medical device designed for use by Respiratory Therapists or other properly trained and qualified personnel under the direction of a physician and in accordance with applicable state laws and regulations.
Leak Testing the Patient Breathing Circuit – The patient circuit must be leak tested in the VENT CHECK mode before connection to the patient. In addition, the Ventilator Checkout mode should be used to check for correct operation of the ventilator alarm, displays and controls. Harm to the patient or ineffective ventilation may result from failure to leak test the patient breathing circuit before connection to a patient. When using a heated humidifier, include it in the circuit when performing leak testing.
Adjustable and Critical Alarms – For safety purposes, all adjustable alarms and all critical alarms must be checked to insure proper operation.
Alarms Function Verification - All alarms must be verified as functioning properly on a daily basis. If any alarm malfunctions, immediately contact a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
Patient Monitoring - Patients who are dependent on a ventilator should be constantly monitored by qualified personnel. Such personnel should be prepared to address equipment malfunctions and circumstances where equipment becomes inoperative. An alternative method of ventilation should be available for all patients dependent on the ventilator, and qualified personnel should be fully familiar with emergency ventilation procedures.
Alternative Ventilation - It is recommended that an alternative means of ventilating the patient be available at all times and that all ventilator operators be fully familiar with emergency ventilation procedures.
Fire or Explosion - Operation of the LTV Series Ventilators in the presence of flammable gases could cause a fire or explosion. Under no circumstances is the ventilator to be operated when explosive gases are present. The presence of nitrous oxide or flammable anesthetics presents a danger to the patient and operator.
Patient Breathing Circuit Disconnection - Inadvertent disconnection of the patient from the patient breathing circuit can be dangerous.
Critical Alarms - Failure to set the critical alarms such as the Low Minute Volume alarm and the Low Pressure alarm may cause non-detection (no alarm) for a disconnection of the lower sense line or the exhalation valve drive line.
Exhalation Valve Diaphragm – Patient ventilation may be ineffective or dangerous if the exhalation valve diaphragm is damaged or worn out. The exhalation valve diaphragm must be inspected on a daily basis and replaced whenever necessary.
BAT EMPTY Alarm - A BAT EMPTY alarm indicates the internal battery is almost depleted. Connect the ventilator to an external power source immediately.
INOP Alarm - If an INOP alarm occurs during operation, ventilate the patient using an alternative method, disconnect the ventilator, and immediately contact a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
Operator’s Manual LTV® Series Ventilator Page 1-3 p/n 10664, Rev. N
, WARNING !
NO CAL Condition - Operation of the LTV Series Ventilator under a NO CAL condition may result in inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient from the ventilator, provide an alternative method of ventilation and immediately contact a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
XDCR FAULT Alarm - Continued operation of the LTV Series Ventilator with an activated XDCR FAULT alarm may result in inaccurate flow and volume measurements. Should this condition occur, disconnect the patient from the ventilator, provide an alternative method of ventilation and immediately contact a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
Personal Injury and Electric Shock - Operation of the LTV Series Ventilator if any of its panels have been removed may result in electrical shock to the patient or operator. All servicing must be performed by a certified Pulmonetic Systems service technician.
NPPV Mode – NPPV1 is not a life support mode and is not suitable for patients that require life support ventilation. NPPV Mode should only be used for supplemental ventilation of non-life support patients.
NPPV Mode - When operating in NPPV1 mode, many of the standard alarms are disabled. This may result in reduced ventilation accuracy should a problem occur. Carefully read Chapter 4 - Ventilation Modes, NPPV, before selecting this mode of operation.
Audible Alarms - Failure to immediately identify and correct audible alarm situations may result in serious patient injury.
Equipment Malfunction or Failure - The LTV Series Ventilator has alarms to notify operators of certain conditions and to cease operating upon detecting possible danger. In the event of equipment failure, all ventilator operators should have an alternative method of ventilation available and be fully familiar with emergency ventilation procedures.
Improperly Functioning Ventilator - Operation of a ventilator that does not appear to be working properly may be hazardous. If the ventilator is damaged, fails Ventilator Checkout tests or malfunctions in any way, discontinue its use and immediately contact a certified Pulmonetic Systems service technician or Pulmonetic Systems, Inc.
Ventilator Checkout Tests – Be aware that gas is not delivered to the patient during these tests. Disconnect the patient from the ventilator and ventilate the patient using an alternative method before running the Ventilator Checkout tests.
Ventilator Checkout and Maintenance Modes - The LTV Series Ventilator does not deliver gas during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT MTNCE) and should not be used to ventilate a patient during these tests.
1
NPPV, Non-Invasive Positive Pressure Ventilation
Page 1-4 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
, WARNING !
Inspired Oxygen (FIO2) Concentration – If the patient has a variable respiratory rate, his/her minute ventilation will fluctuate. If exact concentrations of inspired oxygen (FIO2) are required to be delivered to the patient, it is recommended that an accurate oxygen analyzer2 with alarms be used.
Unauthorized Parts or Accessories – Serious harm to the patient may result from the use of unauthorized parts or accessories. Only items expressly approved by Pulmonetic Systems may be used in conjunction with the LTV Series Ventilators.
Unapproved Adapters – Only Pulmonetic Systems Accessories should be used to connect the ventilator to Patient Assist Call Systems. These accessories incorporate safety features to reduce the risk of shock. Do not attempt to modify these accessories in any way.
Patient Assist Call Connector – Do not apply more than 25V rms or 32VDC to the Patient Assist Call connector.
Ventilator Service and Repair - All servicing or repair of the LTV Series Ventilator must be performed only by a service technician certified by Pulmonetic Systems.
Disabled Oxygen Inlet Pressure Alarms - When the oxygen blending option is not installed, the Oxygen Inlet Pressure Alarms are disabled.
Patient Circuits – Pulmonetic Systems Patient Circuits, Exhalation Valve Assemblies and Water Traps are shipped clean, not sterile.
Ultra Violet Light Sensitivity – The material used in the tubing of the “Re-usable” Patient Circuits is not UV stable. Avoid exposure of the tubing to UV light.
PEEP Valve Rotation – Attempting to adjust the PEEP valve counterclockwise past zero (0) may damage the PEEP valve assembly or cause circuit leaks.
Accessories Mounting Screws - Refer to the information contained in Pulmonetic Systems Replacement Screws Kit, P/N 11149, to determine the appropriate accessories mounting screws or accessories replacement screws location, type and length to use when removing or exchanging external accessories on an LTV Series Ventilator.
Mounting Screw Use – Internal damage to the ventilator may result if the wrong length mounting screws are used when installing or removing external accessories.
Specific Boot Replacement Screw Location - One leg of the upper protective boot has an additional screw hole (furthest from the end of the leg);
On earlier version ventilators (screw was located in the upper hole in the leg of the boot) the use of a 3/16” mounting screw is required.
On current version ventilators (screw was located in the lower hole in the leg of the boot) the use of a 1/4” mounting screw is required.
2
In compliance with ANSI Z-79.10, 1997 standards
Operator’s Manual LTV® Series Ventilator Page 1-5 p/n 10664, Rev. N
, WARNING !
Specific Boot Installation Screw Location - One leg of the upper protective boot has an additional screw hole (furthest from the end of the leg);
On earlier version ventilators, the screw hole will align with the upper hole in the boot and requires the use of the 1/4” mounting screw.
On current version ventilators, the screw hole will align with the lower hole in the boot and requires the use of the 5/16” mounting screw.
Specific Bracket Installation Screw Location - One leg of the LTM/LTV mounting bracket has a circular notch just above the elongated screw slot;
On earlier version ventilators, the screw hole behind this slot will align in the upper half of the slot (nearest the circular notch) and requires the use of a 5/16” mounting screw.
On current version ventilators, the screw hole behind this slot will align in the lower half of the slot (furthest from the circular notch) and requires the use of a 3/8” mounting screw.
Specific Bracket Installation Screw Location - One leg of the LTM/LTV mounting bracket has a circular notch just above the elongated screw slot;
On earlier version ventilators, the screw hole behind this slot will align in the upper half of the slot (nearest the circular notch) and requires the use of a 5/16” mounting screw.
On current version ventilators, the screw hole behind this slot will align in the lower half of the slot (furthest from the circular notch) and requires the use of a 3/8” mounting screw.
Specific Bracket Replacement Screw Location – One leg of the LTM/LTV mounting bracket has a circular notch just above the elongated screw slot;
On earlier version ventilators (screw was positioned in the upper half of the mounting bracket leg screw slot) the use of a 3/16” mounting screw is required.
On current version ventilators (screw was positioned in the lower half of the mounting bracket leg screw slot) the use of a 1/4” mounting screw is required.
Patient Circuit Accessories - The use of accessories such as Speaking Valves, Heat­Moisture Exchangers and Filters create additional patient circuit resistance and in the event of a disconnection, may impede the generation of a Low Pressure Alarm. Ensure that the Low Pressure Alarm settings accommodate these types of accessories when used in combination with patient circuits.
Low Minute Volume Control Settings - The Low Minute Volume control should be set to its highest clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.
Page 1-6 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
,MISE EN GARDE
Personnel non qualifié - Seul le personnel qualifié doit opérer le ventilateur. Le ventilateur de la série LTV est un dispositif médical restreint conçu pour être utilisé par les inhalothérapeutes ou autres personnes qualifiées, et par le personnel qualifié sous la supervision d'un médecin et en conformité avec les lois et règlements applicables.
Contrôle de l’étanchéité du circuit respiratoire du patient – L'étanchéité du circuit respiratoire du patient (vérification de ventilation) doit être vérifiée en mode VENT CHECK avant le raccordement au patient. En outre, on doit utiliser le mode Ventilator Checkout (vérification du ventilateur) afin de s’assurer du fonctionnement adéquat de l’alarme, des affichages et des commandes du ventilateur. Le défaut de vérifier l’étanchéité du circuit respiratoire du patient avant le raccordement à un patient peut être nocif pour le patient ou provoquer une ventilation inefficace. Lorsqu’un humidificateur chauffant est employé, il convient de l’inclure dans le circuit en procédant à la vérification de l’étanchéité.
Alarmes ajustables et critiques - Afin d'assurer l'opération sécuritaire des ventilateurs de la série LTV, toutes les alarmes ajustables doivent être réglées avant l'opération. De plus, toutes les alarmes critiques (par exemple, alarme de basse pression), doivent être inspectées avant de laisser le patient seul.
Vérification du fonctionnement des alarmes - Toutes les alarmes sonores et visuelles doivent être vérifiées quotidiennement. Si une des alarmes fonctionne de façon inadéquate, contactez votre technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Surveillance du patient – Un personnel qualifié doit constamment surveiller les patients qui sont reliés à un ventilateur. Le personnel doit être en mesure de s’occuper des défectuosités de fonctionnement de l’équipement ainsi que des circonstances où ce dernier devient inopérant. Une forme de ventilation alternative doit être disponible à tous les patients reliés au ventilateur et le personnel qualifié devrait être pleinement familier avec les procédures de ventilation d’urgence.
Ventilation alternative - Il est recommandé qu'un moyen alternatif de ventilation soit disponible en tout temps, et que tous les opérateurs de ventilateur soient pleinement familiers avec les procédures de ventilation d'urgence.
Feu ou explosion - L'opération des ventilateurs de la série LTV en présence de gaz inflammables peut causer un feu ou une explosion. Le ventilateur ne doit être opéré sous aucune circonstance en présence de gaz. La présence d'oxyde nitreux ou d'anesthésiques inflammables représente un danger pour le patient et l'opérateur.
Débranchement du circuit respiratoire du patient - Le débranchement accidentel du circuit respiratoire du patient peut s'avérer dangereux.
Alarmes critiques – Le défaut de définir les alarmes critiques telles que l’alarme basse ventilation-minute et l’alarme basse pression peut causer une non-détection (absence d’alarme) pour un débranchement du tube de détection inférieur ou du tube d’entraînement de la soupape d’expiration.
Diaphragme de la soupape d'expiration - Une ventilation inefficace ou dangereuse pour le patient peut résulter si le diaphragme de la soupape est endommagé ou usé. Le diaphragme de la soupape d'expiration doit être vérifié quotidiennement, et remplacé au besoin.
Operator’s Manual LTV® Series Ventilator Page 1-7 p/n 10664, Rev. N
,MISE EN GARDE
Alarme BAT EMPTY - Une alarme BAT EMPTY indique que la pile interne est pratiquement à plat. Branchez immédiatement le ventilateur à une source d'alimentation externe.
Alarme INOP - Si une alarme INOP survient au cours de l'opération, ventilez le patient à l'aide de la méthode alternative, retirez immédiatement le ventilateur du service, et contactez immédiatement votre technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Condition NO CAL - L'opération continue du ventilateur de la série LTV sous condition NO CAL peut résulter en mesures de pression et de volume erronées. Si cette condition
se présente, le ventilateur doit être retiré du service, et vous devez immédiatement contacter votre technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Alarme XDCR FAULT - L'opération continue du ventilateur de la série LTV avec une alarme XDCR FAULT activée peut résulter en mesures de débit et de volume erronées. Si cette condition se présente, le ventilateur doit être retiré du service, et vous devez immédiatement contacter votre technicien de service certifié de Pulmonetic Systems ou Pulmonetic Systems Inc.
Blessures personnelles et chocs électriques - L'opération d’un ventilateur de la série LTV alors que ses panneaux sont enlevés, peut causer un choc électrique au patient ou à l'opérateur. Tout entretien doit être effectué par un technicien de service certifié de Pulmonetic Systems.
Mode NPPV – Le mode NPPV n’est pas un mode de maintien des fonctions vitales continu et il n’est pas approprié pour les patients qui ont besoin d’une ventilation continue pour le maintien des fonctions vitales. Le mode NPPV ne doit être utilisé que comme ventilation supplémentaire pour les patients qui ne nécessitent pas de maintien des fonctions vitales.
Mode NPPV – Lorsque l’appareil fonctionne en mode NPPV, bon nombre des alarmes standards sont désactivées. Par conséquent, si un problème survient, la précision de la ventilation pourrait diminuer. Assurez-vous de lire attentivement le chapitre 4 – Types de respiration et modes de ventilation, mode NPPV avant de choisir ce mode de fonctionnement.
Alarmes sonores - L'échec à identifier et à corriger dans l'immédiat les situations d'alarmes sonores peut causer des blessures au patient.
Mauvais fonctionnement ou panne de l'équipement - Des dispositifs électromécaniques peuvent mal fonctionner ou subir une panne. Le ventilateur de la série LTV a été conçu avec des alarmes, pour détecter et aviser les opérateurs de certaines conditions, et pour cesser d'opérer en cas de conditions d'opération dangereuses. En cas de panne de l'équipement, tous les opérateurs du ventilateur devraient avoir une forme de ventilation alternative à leur disponibilité, et être pleinement familiers avec les procédures de ventilation d'urgence.
Ventilateurs fonctionnant de façon inadéquate - L'opération d'un ventilateur dont le fonctionnement semble inadéquat peut représenter un danger. Si le ventilateur est endommagé, s'il échoue les tests de vérification du ventilateur ou s'il fonctionne de façon inadéquate, suspendez l'utilisation de ce ventilateur et contactez immédiatement votre technicien de service certifié de Pulmonetic Systems Inc.
Page 1-8 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
,MISE EN GARDE
Tests de vérification du ventilateur – Noter que le gaz n’est pas transmis au patient au cours de ces tests. Débrancher le patient du ventilateur et ventiler le patient à l’aide d’une forme de ventilation alternative avant de procéder aux tests de vérification du ventilateur.
Modes Vérification et Entretien du ventilateur - Le ventilateur de la série LTV ne transmet pas le mélange de gaz en mode Vérification du ventilateur (VENT CHECK) ou en mode Entretien du ventilateur (VENT MTNCE), il ne devrait donc pas être utilisé pour ventiler un patient durant l'exécution de ces tests.
Concentration d’oxygène inspiré (FIO2) – Si la fréquence respiratoire du patient est variable, sa ventilation-minute va fluctuer. Lorsqu’une concentration exacte d’oxygène inspiré (FIO2) est nécessaire pour une transmission au patient, il est recommandé d’utiliser un analyseur de niveau d’oxygène2 précis, comportant des alarmes.
Pièces, accessoires et options non autorisées - Des dommages à l'équipement ou des blessures au patient peuvent survenir suite à l'utilisation de pièces, accessoires et options non autorisées. Seuls les éléments expressément approuvés par Pulmonetic Systems doivent être utilisés en conjonction avec les ventilateurs de la série LTV.
Accessoires non approuvés – L’utilisation d’accessoires qui ne sont pas expressément approuvés par Pulmonetic Systems pourrait entraîner des conditions dangereuses. Seuls les accessoires de Pulmonetic Systems devraient être utilisés pour brancher les ventilateurs aux systèmes d’aide aux patients. Ces accessoires comportent des caractéristiques de sécurité pour réduire les risques de choc. N’essayez pas de modifier ces accessoires d’aucune façon.
Connecteur d’appel d’aide aux patients – Ne mettez pas plus de 25 V efficace ou 32 V c.c. au connecteur d’appel d’aide aux patients.
Entretien et réparation du ventilateur - Tout entretien ou réparation du ventilateur de la série LTV ne doit être effectué que par un technicien de service certifié de Pulmonetic Systems.
Alarmes de pression d'entrée de l'oxygène désactivées - Lorsque l'option de mélange d'oxygène n'est pas activée, les alarmes de pression d'entrée de l'oxygène sont désactivées.
Circuits du patient – Les circuits du patient du Pulmonetic Systems, les valves expiratoires et les collecteurs d’eau sont expédiés propres, mais pas stériles.
Sensibilité à la lumière ultraviolette – Les matériaux utilisés pour la tubulure des circuits du patient ne sont pas stables sous rayons UV. Éviter d’exposer la tubulure à la lumière UV.
Rotation de la valve de pression expiratoire positive – Si vous essayez d’ajuster la valve de pression expiratoire positive en sens inverse des aiguilles d’une montre passé zéro (0), vous pourriez endommager la valve de pression expiratoire positive ou causer une fuite dans le circuit.
Vis de montage des accessoires – Voir les renseignements fournis dans la trousse de vis de remplacement de Pulmonetic Systems, numéro de pièce 11149, pour déterminer l’emplacement, le type et la longueur des vis de montage d’accessoires ou des vis de remplacement pour accessoires à utiliser lors de la dépose ou de l’échange d’accessoires externes sur un ventilateur de la série LTV.
Utilisation des vis de montage – Vous pourriez causer des dommages internes au ventilateur si des vis de montage de mauvaise longueur sont utilisées lors de l’installation ou de la dépose des accessoires externes.
Operator’s Manual LTV® Series Ventilator Page 1-9 p/n 10664, Rev. N
,MISE EN GARDE
Emplacement des vis de remplacement d’un gaine spécifique – Une patte de la gaine protectrice supérieure possède un trou de vis supplémentaire (le plus éloigné de l’extrémité de la patte);
Sur les anciennes versions des ventilateurs (la vis se trouvait dans le trou supérieur de la patte de la gaine), vous devez utiliser une vis de montage de 3/16”.
Sur la version actuelle des ventilateurs (la vis se trouve dans le trou inférieur de la patte de la gaine), vous devez utiliser une vis de montage de 1/4”.
Emplacement des vis d’installation d’un gaine spécifique – Une patte de la gaine protectrice supérieure possède un trou de vis supplémentaire (le plus éloigné de l’extrémité de la patte);
Sur les anciennes versions des ventilateurs, le trou de la vis s’alignera au trou supérieur de la gaine et vous devez utiliser une vis de montage de 1/4".
Sur la version actuelle des ventilateurs, le trou de la vis s’alignera au trou inférieur de la gaine et vous devez utiliser une vis de montage de 5/16”.
Emplacement de la vis d’installation d’un support spécifique – L’une des pattes du support de montage du LTM/LTV possède une encoche circulaire juste au-dessus de la fente allongée pour la vis;
Sur les anciennes versions des ventilateurs, le trou de la vis à côté de cette fente s’aligne à la moitié supérieure de la fente (la plus proche de l’encoche circulaire) et vous devez utiliser une vis de montage de 5/16”.
Sur la version actuelle des ventilateurs, le trou de la vis à côté de cette fente s’aligne à la moitié inférieure de la fente (la plus éloignée de l’encoche circulaire) et vous devez utiliser une vis de montage de 3/8”.
Emplacement de la vis de remplacement d’un support spécifique – L’une des pattes du support de montage du LTM/LTV possède une encoche circulaire juste au-dessus de la fente allongée pour la vis;
Sur les anciennes versions des ventilateurs (la vis se trouvait dans la moitié supérieure de la fente pour la vis sur la patte du support de montage), vous devez utiliser une vis de montage de 3/16”.
Sur la version des ventilateurs (la vis se trouve dans la moitié inférieure de la fente pour la vis sur la patte du support de montage), vous devez utiliser une vis de montage de 1/4”.
Accessoires du circuit du patient - L’utilisation d’accessoires tels que les membranes vocales, les échangeurs thermohydriques et les filtres, produit une résistance additionnelle dans le circuit de patient et en cas de débranchement, elle risque d’empêcher la génération de l’alarme de basse pression. S’assurer que les paramètres de l’alarme de basse pression s’adaptent à ces types d’accessoires lorsqu’ils sont utilisés avec les circuits du patient.
Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques, effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d., sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio­respiratoire) s’avère pertinente.
Page 1-10 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Cautions
, Caution !
Ventilator Sterilization – To avoid irreparable damage to the LTV Series Ventilator, do not attempt to sterilize it.
Cleaning Agents – To avoid damaging the ventilator’s plastic components and front panel, do not use cleaning agents containing ammonium chloride, other chloride compounds, more than 2% glutaraldehyde, phenols, or abrasive cleaners.
Ventilator Immersion - Do not immerse the ventilator in liquids. Reusable Patient Circuit Components - To avoid degradation of the reusable patient
circuit components, do not exceed the following constraints:
50 cleaning cycles or 1 year (whichever comes first)
Steam Autoclave:
Pressure: 20 PSIG
Temperature: 275°F (135°C)
Time: 6 minutes
Liquid Sterilizing Agent:
The use of liquid agents containing more than 2% glutaraldehyde.
Pasteurization:
A 30-minute warm water detergent and a 30-minute 165°F (74°C) hot water cycle.
Drying in a sterile drier for more than 1 hour or 140°F (59°C).
Gas (ETO):
Temperature: 131°F (55°C)
Differential Pressure Ports - A low pressure air nozzle with flow less than 10 liters per minute should be used for cleaning the differential pressure ports.
Exhalation Valve Cleaning - Do not pour or spray liquid cleaners into the exhalation valve.
Patient Wye Installation – After cleaning, install the patient wye in the patient circuit so the proximal sense lines are oriented up while operating.
Care of the Exhalation Valve - The exhalation valve is a delicate assembly and may be damaged if;
Care is not exercised when handling or cleaning it.
Cleaning instruments or foreign bodies are inserted into it.
High-pressure gas nozzles are used to dry it.
Front Panel Cleaning – Do not pour or spray liquid cleaners onto the front panel. Care of Bacterial Filters – If bacterial filters are used in conjunction with the LTV Series
Ventilator, comply with all procedures as specified by the filter manufacturer. Wet or Damp Filters - Do not install a wet or damp filter into the LTV Series Ventilators.
This could damage the ventilator.
Operator’s Manual LTV® Series Ventilator Page 1-11 p/n 10664, Rev. N
, Caution !
Proximal Sense Lines - Do not remove the proximal sense lines from the patient wye. Automobile Cigarette Lighter and Power Outlets – Automobile cigarette lighter and
power outlets are normally wired for a positive center contact and ground sleeve contact. Connecting the ventilator to an improperly wired outlet will cause the adapter fuse to blow and may damage the adapter or the ventilator.
Automobile Cigarette Lighter Outlet Power Rating - Running a ventilator from an improperly rated automobile cigarette lighter outlet (less than 20 amperes) may cause a fuse in the automobile to blow, causing the ventilator and possibly other accessories in the automobile to stop operating.
Automobile Cigarette Lighter Adapter - Do not operate the ventilator from the Automobile Cigarette Lighter Adapter while starting the vehicle or when jump starting the automobile battery. Doing so may cause damage to the ventilator.
Automobile Cigarette Lighter Adapter Tip - Use care when disconnecting the Automobile Cigarette Lighter Adapter after use, its tip may be hot.
Automobile Cigarette Lighter Outlet – Depending on the condition of the automobile battery, whether the automobile is turned off, being started or running, automobile cigarette lighter outlets can provide varying levels of voltage (in some, the outlet only operates when the vehicle is running). Verify which power source the ventilator is using by checking the EXTERNAL POWER LED on the ventilator.
Remote Alarm - Always verify that the remote alarm properly reports the LTV Series Ventilator alarms before use.
Remote Alarm - Always follow the remote alarm manufacturer’s usage and maintenance requirements to guarantee proper function of the device.
External Battery Pack - The External Battery Pack should only be connected to the LTV Series Ventilators using the Pulmonetic Systems External Battery Cable (PN
10802). This cable is pre-wired and properly terminated to ensure safe connection of
the External Battery Pack to the ventilator. Electrical Grounding – In the event of a loss of electrical protective ground, touching
the ventilator could result in electrical shock. To ensure grounding and avoid this danger, use only the unmodified power cord originally supplied with the LTV Series Ventilators, maintained in good condition and connected to a properly wired and grounded electrical power outlet.
Electrostatic Shock – The use of electrically conductive hoses and tubing is not recommended. The use of such materials may result in damage to the ventilator from electrostatic discharge.
External DC Power Source or External Battery - When connecting the LTV Series Ventilators to an external DC power source or external battery, use only the approved method and connectors specified in Chapter 14 - Power and Battery Operation.
Page 1-12 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
, Caution !
AC Power Source - When connecting the ventilator to an AC power source, use only the approved LTV AC Power Adapter.
AC Power Earth Ground Validity – If the validity of the AC power earth ground connection is in doubt, use the internal battery, an external battery, or an external DC power source to operate the LTV Series Ventilator.
Fuse Fire Hazard – Replacement of existing fuses with fuses with different voltage or electrical current ratings may cause a fire.
Storage Temperature - Storing the LTV Series Ventilator at temperatures above 60°C (140°F) for long periods can damage the internal battery and cause expected battery duration to degrade.
Patient Assist Call Connector – Do not apply more than 25V rms or 32VDC to the Patient Assist Call connector.
Ventilator Checkout Tests - LTV Series Ventilator Checkout tests must be performed before connecting the patient to the ventilator. Rerun the tests monthly and whenever a question about the ventilator’s operation arises.
Release Button - To avoid damaging the ventilator or the power connector, push the release button on the connector before removing it from the ventilator power port or the power port pigtail connector.
Operator’s Manual LTV® Series Ventilator Page 1-13 p/n 10664, Rev. N
, Avertissement
Stérilisation du ventilateur - Afin d'éviter des dommages irréparables au ventilateur de la série LTV, ne tentez pas de stériliser ce dernier.
Produits de nettoyage - Afin d'éviter d'endommager les composants plastiques et le panneau frontal du ventilateur, n'utilisez pas des produits de nettoyage contenant : chlorure d'ammonium, composés de chlorure, plus de 2% de glutaraldéhyde, ou phénol.
Immersion du ventilateur - Ne pas immerger le ventilateur dans des liquides, incluant les produits stérilisants.
Composants réutilisables du circuit du patient – Pour éviter la dégradation des composants réutilisables du circuit du patient, ne dépassez pas les limites suivantes:
50 cycles de nettoyage ou 1 an (le premier des deux prévalant)
Autoclave à vapeur:
Pression : 20 lb/po2
Température : 275°F (135°C)
Durée : 6 minutes
Agent de stérilisation liquide:
L’utilisation d’agents liquide contenant plus de 2% de glutaraldéhyde.
Pasteurisation:
Un cycle avec détergent à l’eau tiède pendant 30 minutes et à l’eau chaude à 165°F (74°C) pendant 30 minutes.
Séchage dans un séchoir stérile pendant plus de 1 heure ou à 140°F (59°C). Gaz (ETO):
Température : 131°F (55°C)
Ports de pression différentielle - Une source de gaz à débit faible (moins de 10 ppm) doit être utilisée pour le nettoyage des fluides et de débris des ports de pression différentielle.
Nettoyage de la soupape d'expiration - Ne pas asperger une solution nettoyante dans la soupape d'expiration.
Installation de la soupape d'expiration - Après le nettoyage, installez la soupape d'expiration dans le circuit du patient de sorte que les lignes de détection soient alignées vers le haut pendant l’opération.
Entretien de la soupape d'expiration - La soupape d'expiration est une pièce fragile et peut être endommagée si :
Des précautions ne sont pas prises lors de sa manipulation ou de son nettoyage.
Des instruments de nettoyage ou des corps étrangers sont insérés dans celle-ci.
Des pistolets de gaz à haute-pression sont utilisés pour l'assécher.
Nettoyage du panneau frontal - Ne pas asperger des solutions nettoyantes ou les laisser s'écouler sur le panneau frontal.
Entretien des filtres bactériens - Les filtres bactériens ne devraient pas être immergés dans un liquide. Un autoclave à vapeur devrait être utilisé pour le nettoyage des filtres bactériens.
Page 1-14 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
, Avertissement
Filtres mouillés ou humides - Ne pas installer des filtres mouillés ou humides dans les ventilateurs de la série LTV. Cela pourrait endommager le ventilateur.
Conduites de détection – N’enlevez pas les conduites de détection qui se trouvent sur les divisions en Y du circuit du patient.
Allume-cigare et prises de courant – L’allume-cigare et les prises de courant sont habituellement câblés de façon à obtenir un contact central positif et un contact du manchon à la terre. Le branchement du ventilateur dans une prise qui n’est pas câblée adéquatement aura pour effet de faire sauter le fusible de l’adaptateur et pourrait endommager l’adaptateur ou le ventilateur.
Puissance nominale des prises d’allume-cigare – Le branchement d’un ventilateur à une prise d’allume-cigare qui ne possède pas la tension suffisante (moins de 20 ampères) peut faire griller un fusible de l’automobile, causant ainsi l’arrêt du ventilateur et éventuellement, celui d’autres accessoires de l’automobile.
Adaptateur pour allume-cigare – Ne faites pas fonctionner le ventilateur à l’aide de l’adaptateur pour allume-cigare lorsque vous démarrez le véhicule ou lorsque vous faites une connexion provisoire de la batterie d'un véhicule. Vous pourriez ainsi endommager le ventilateur.
Embout adaptateur pour allume-cigarette d’automobile - Après l'utilisation, débrancher l’adaptateur pour allume-cigarette d’automobile avec précaution car son embout peut être chaud.
Prise d’allume-cigare d’automobile – Selon la condition de la batterie de l’automobile, si le moteur est coupé, démarré ou est en marche, les prises d’allume-cigare d’une automobile peut générer des niveaux de tension variés (sur certains modèles, la prise ne fonctionne que si le moteur est en marche). Vérifier la source d’alimentation utilisée par le ventilateur indiquée par la DEL EXTERNAL POWER du ventilateur.
Alarme à distance – Assurez-vous toujours que l’alarme à distance indique de façon adéquate les alarmes du ventilateur LTV avant d’utiliser le ventilateur.
Alarme à distance – Suivez toujours les exigences d’utilisation et d’entretien du fabricant de l’alarme à distance afin d’assurer le fonctionnement adéquat de l’appareil.
Bloc-piles externe – Le bloc-piles externe ne doit être branché qu’aux ventilateurs de la série LTV à l’aide du câble pour piles externes de Pulmonetic Systems (No pièce 10802). Ce câble est précâblé et ses terminaisons assurent une connexion sécuritaire entre le bloc-piles externe et le ventilateur.
Mise électrique à la terre - En cas de perte de la mise électrique à la terre de protection, toutes les pièces conductrices peuvent transmettre un choc électrique. Pour éviter un choc électrique, n'utilisez que le cordon d'alimentation d'origine non modifié fourni avec les ventilateurs de la série LTV, maintenus en bonne condition, et branchés à une prise adéquatement câblée et mise à la terre.
Choc électrostatique – L'utilisation de tuyaux et de tubes conductibles n'est pas recommandée. L'utilisation de ces matériaux risque de causer une décharge électrostatique qui endommagerait le ventilateur.
Operator’s Manual LTV® Series Ventilator Page 1-15 p/n 10664, Rev. N
, Avertissement
Source de courant continu ou pile externe - Lorsque vous branchez les ventilateurs de la série LTV sur une source de courant continu ou sur une pile externe, utilisez seulement les méthodes et les connecteurs approuvés spécifiés au chapitre 14 ­Alimentation et opération avec pile.
Source d'alimentation c.a. - Lorsque vous branchez le ventilateur sur une source d'alimentation c.a., utilisez l'adaptateur c.a. LTV approuvé.
Validité de la mise à la terre de l'alimentation c.a. - Si vous doutez de la validité de la mise à la terre de l'alimentation c.a., utilisez la pile interne, une pile externe ou une source externe de courant continu, pour opérer le ventilateur de la série LTV.
Danger d'incendie des fusibles - Le remplacement des fusibles existants par des fusibles de type, d'ampérage et de courant électrique différent peut causer un incendie.
Température d'entreposage - L'entreposage du ventilateur de la série LTV à des températures supérieures à 60° C (140° F) durant des périodes prolongées peut endommager la pile interne et causer l'usure prématurée de la pile.
Connecteur d’appel d’aide aux patients - Ne mettez pas plus de 25 V efficace ou 32 V c.c. au connecteur d’appel d’aide aux patients.
Tests de vérification du ventilateur - Les tests de vérification du ventilateur doivent être effectués avant de relier le patient au ventilateur. Effectuez les tests sur une base mensuelle, et lors de doutes relativement à l'opération adéquate du ventilateur.
Bouton de déclenchement – Pour éviter d'endommager le ventilateur ou le connecteur d'alimentation, appuyer sur le bouton de déclenchement situé sur le connecteur avant de le retirer du port d'alimentation du ventilateur ou du raccord de queue de cochon du port d'alimentation.
Page 1-16 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N
Symbols
Symbol Compliance
3
Title Application
ISO 3864
(Prev. IEC 348)
Symbol No.
IEC 417
Symbol No.
417-IEC-5016
IEC 417
Symbol No.
417-IEC-5035
IEC 417
Symbol No.
417-IEC-5019
IEC 417
Symbol No.
417-IEC-5333
IEC 417
Symbol No.
417-IEC-5031
B.3.1
Caution (refer to accompanying documents)
Used to direct the user to the instruction manual where it is necessary to follow certain specified instructions where safety is involved.
Fuse To indicate the fuse boxes, for example,
and their location.
Output To identify an output terminal when it is
necessary to distinguish between inputs and outputs.
Protective earth (ground) To identify any terminal which is intended
for connection to an external protective conductor for protection against electric shock in case of a fault or the terminal of a protective earth (ground) electrode.
Type BF equipment. To mark a type BF equipment complying
with IEC
Publication 601.
Direct Current To indicate on the rating plate that the
equipment is suitable for direct current only; to identify relevant terminals.
IEC 417
Symbol No.
417-IEC-5032
IEC 417
Symbol No.
417-IEC-5172
Alternating current To indicate on the rating plate that the
equipment is suitable for alternating current only; to identify relevant terminals.
Class II equipment To identify equipment meeting safety
requirements specified for Class II equipment.
3
Reference IEC Medical Electrical Equipment, 2nd. Edition 1988
Operator’s Manual LTV® Series Ventilator Page 1-17 p/n 10664, Rev. N
Page 1-18 LTV® Series Ventilator Operator’s Manual p/n 10664, Rev. N