PTW Freiburg PTW-UNITEST 47001, PTW-UNITEST 47002, PTW-UNITEST 47003 User Manual

User Manual
PTW-UNITEST
(Electrical Check Device)
Type 47001, Type 47002 and Type 47003
D235.131.0/4 2005-12 Sa

General Information

PTW-UNITEST
(Electrical Check Device)
General Information
The product bears the CE-mark "CE-0124" in
accordance with the Council Directive 93/42/EEC about Medical Devices and fulfills the essential requirements of Annex 1 of this directive. The product is a class IIb device (MDD).
The user manual is an integral part of the
product. It should always be kept near the product. Observance of the manual is a pre­requisite for proper product performance and correct operation.
Operator safety, specified measuring accuracy
and interference-free operation can be guar­anteed only if original products and parts are used. Furthermore only the accessories listed in this manual are approved by PTW-Freiburg and may be used in conjunction with the prod­uct, or else accessories whose use has been expressly permitted by PTW-Freiburg. Safe operation and proper product performance are not guaranteed if accessories or consumables from other manufacturers are used.
This manual is in conformity with the product
specifications and all applicable safety stan­dards valid at printing date. All rights are re­served for devices, circuits, techniques, soft­ware and names referred to in the manual.
Electromagnetic immunity: The product fully
complies with the electromagnetic immunity requirements of standard IEC 60601-1-2 `Electromagnetic Compatibility - Medical Elec­trical Equipment`.
No part of the technical documentation may be
reproduced without written permission from PTW-Freiburg.
PTW-Freiburg works in strict accordance with
a quality management system which is con­tinuously updated according to national and international standards.
PTW-Freiburg cannot be held liable for dam-
ages resulting from the use of accessories, consumables from other manufacturers or when the user ignores the instructions and in­formation given in this manual.
The warranty period is 1 (one) year and begins
on the day of delivery. It is unaffected by repairs covered by the war­ranty regulations.
PTW-Freiburg considers itself responsible for
safety, reliability and performance of the prod­uct only, if assembly, extension, readjustment, modification or repair is carried out by PTW-Freiburg or by persons authorized by PTW-Freiburg, and if the product is used in compliance with the technical documentation.
PTW-FREIBURG
Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de
PTW-UNITEST
(Electrical Check Device)
Contents
Contents
General Information 2
1 Intended Use 4
2 For Your Safety 4
3 Description of the Instrument 7
4 Description of Function 7
5 Measurements with the UNITEST 8
5.1 Hints for the use of the UNITEST 8
5.2 Preparing the measurements 8
5.3 Measurements 8
6 Evaluation of the measuring results 11
6.1 Consistency check of the response in the different measuring ranges 11
6.2 Electrical constancy checks 11
7 Cleaning, Maintenance, Disposal 12
7.1 Cleaning 12
7.2 Preventive Maintenance 12
7.3 Disposal of the Product 12
8 Technical Specifications 13
Literature 14
D235.131.0/4 3
PTW-UNITEST
(Electrical Check Device)

Intended Use

1 Intended Use
UNITEST is an electrical check device for the sur­veillance of the PTW dosemeters listed in chap­ter 5. No mains connection or battery is needed to operate UNITEST; it is simply connected to the dosemeter instead of an ionization chamber. Using UNITEST the following checks can be done:
Check of the consistency of the response in all
doserate and dose measuring ranges.
Electrical constancy check of the different ampli-
fier ranges and of the internal high voltage sup­ply of the dosemeter.
It must be observed that the connection systems of UNITEST and dosemeter are compatible. Use of adapter cables is not permitted.
When the UNITEST is not in use, it should be stored in a location protected against humidity. The connector should be protected with a dust protec­tion cap.
2 For Your Safety
The following safety statements are divided into
DANGER, WARNING, CAUTION and NOTE.
Definitions
DANGER
Indicates an imminent hazard. If not avoided, the hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the hazard could result in minor injury or product / property damage.
NOTE
Provides useful information to assure that you get the most from your equipment.
PTW-UNITEST
(Electrical Check Device)
For Your Safety
Safety Information
DANGER
Explosion Hazard – The product is not suitable for operation in areas of risk where an explo- sion hazard may occur. Explosion hazards may be caused by the use of combustible an­aesthetics, skin-cleansing agents and dis­infectants.
Furthermore the product is not suitable for ap­plication in oxygen-enriched atmospheres. The atmosphere is considered to be oxygen­enriched when more than 25 % of oxygen or ni­trous oxide is added to the ambient air.
WARNING
Shock Hazard – Strictly observe the following warnings. Failure to do so may endanger the lives of the patient, the user and other persons involved.
Before using the product, the operator must
ascertain that it is in correct working order and operating condition.
Before putting the device into operation,
visually inspect all connection cables for signs of damage. Damaged cables and con­nectors must be replaced immediately.
Devices on which moisture condensation
has developed as a result of temperature changes must not be switched on unless completely dry.
Liquids must not enter the device. If liquids
have entered the device, it must be thor­oughly inspected before being used again.
Exclusion of operation in the patient envi­ronment:
Neither the product nor any peripheral device may be operated within the patient environment (see Figure 1).
Exclusion of operation as device with pa­tient contact:
The device is not for use in direct contact with the patient. Neither UNITEST nor any periph­eral device may have contact to the patient.
Never touch the patient and open connectors of the device at the same time.
WARNING
Suffocation Hazard – Dispose of the packag­ing material, observing the applicable waste­control regulations. Keep the packaging mate­rial out of children's reach.
Patient Hazard – The product is a medical electrical device and must only be handled by persons who are trained in the use of such equipment and are capable of applying it prop­erly. The operator must be trained in the use of the device.
Risk of Poisoning – Chemicals required for application or maintenance of the device, for in­stance, must under all circumstances be stored, prepared, and kept at hand in their specific containers. Failure to observe this instruction may result in severe consequences for the pa­tient.
CAUTION
Equipment damage - Always observe the am­bient conditions as indicated in the `Technical Specifications`.
NOTE
Please observe the user manuals of all con­nected devices!
Set up the device so that the operator has a clear, unobstructed view of the control panel.
D235.131.0/4 5