PTW UNIDOS webline Service Manual

Page 1
Service Manual
webline
UNIDOS
D776.133.0/1
2006-06 Se/Sa
Page 2
webline
Contents
Contents
1 For Your Safety 3
2 Spare Parts List 3
3 Regularly Maintenance 4
3.1 Checkups (Function tests) 4
3.2 Constancy Test 5
3.3 Maintenance 5
3.4 Further Tests 6
3.5 System Information 6
4 Troubleshooting 7
5 Functional units 7
5.1 Front panel unit 7
5.2 Measuring Module 8
5.3 Main Board 9
6 Service Software 10
2 D776.133.0/1
Page 3
For your safety
webline

1 For Your Safety

CAUTION
This document is a supplement to the UNIDOS conjunction with that manual.
The safety statements appearing in the indi­cated user manual are applicable and must be observed.
WARNING
In addition to these safety statements, please note that devices connected to the interfaces described in this manual must fulfill the re­quirements of the IEC standard 60950.
webline
user manual and is valid only in

2 Spare Parts List

The following spare parts are available for the UNIDOS
Description PTW ID No. or
Desiccator cartridge L137004 O-ring for desiccator car-
tridge Cover for desiccator car-
tridge Fuse set for mains opera-
tion Set of rechargeable batter-
ies
webline
:
Order Designation
L612037
L603024
2 x L251013
6 x L389026
PTW-FREIBURG
Physikalisch Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de
D776.133.0/1 3
Page 4

3 Regularly Maintenance

3.1 Checkups (Function tests)

The following checkups have to be executed regu­larly according to the recommendation or the na­tionally valid laws / rules.
Regularly Maintenance
webline
Test
Interval Recommendation Comment
Visual inspection of the device and the cables
daily and after breaks (over several hours)
Current offset + zeroing daily and after
breaks (over several hours)
Detector test daily and after
breaks (over several hours), after changing the detector
Check desiccator
weekly recommended
Safety inspections
annually acc. law refer to national laws / rules
Inspections of measurement
2 years acc. law refer to national laws / rules
technique
To execute the tests correctly, please observe the instructions given in the corresponding sections of the user manual of the UNIDOS
webline
.
strongly recom­mended
strongly recom­mended
repeat after each changing of the detector
recommended HV check at detector
Safety inspections must only be handled by per­sons who are trained in the use of such equipment and are capable of applying it properly. The opera­tor must be trained in the use of the device.
The UNIDOS
webline
must be inspected at regular intervals according to national regulations at PTW­Freiburg or a qualified calibration laboratory.
Safety inspections and inspections of measurement technique must also be performed after repairs or after each influence which may have changed the behavior of the device.
National valid laws or rules may regulate shorter intervals.
4 D776.133.0/1
Page 5
Regularly Maintenance

3.2 Constancy Test

The following tests should be executed regularly. They make it possible to recognize slow alterations of the characteristics of device or detector.
webline
Test Detector test
Interval
Accessories
weekly - HV check at detector; only range Low
Comment
Device test
weekly - checks device only; only range Low
Long term stability weekly UNITEST check HV and all 3 measuring ranges;
no complete function test
Long term stability monthly check device no complete function test
To execute the tests correctly, please observe the instructions given in the corresponding sections of the user manual of the UNIDOS
webline
.
The test results have to be recorded from the first start of operation in order to be able to make state­ments about a change. Addionally for detector tests: the record has to be separate for each detec­tor.
The checkup of the device can be performed most extensive with an UNITEST (refer to the user man­ual of the UNITEST).
A complete function test can take place however only through a inspection of measurement tech­nique.

3.3 Maintenance

Description Cleaning + care Rechargeable battery
- handling – Changing fuses
To execute the tests correctly, please observe the instructions given in the corresponding sections of the user manual of the UNIDOS
Interval Comment if required max. 3 month discharge and cha rge NiMH batteries
if required
webline
.
D776.133.0/1 5
Page 6

3.4 Further Tests

Ethernet-interface The Ethernet interface can be checked using the VNC Viewer.
TRIP Output The function of the TRIP OUTPUT can be checked as follows: set up the threshold and apply a signal which exceeds the threshold (e.g. using an UNITEST).
To execute the tests correctly, please observe the instructions given in the corresponding sections of the user manual of the UNIDOS
webline
.

3.5 System Information

Information about the version of firmware and hardware and the status of supply voltages of the device can be read out via service e-mail (refer to corresponding sections of user manual UNI-
webline
DOS
The device monitors the supply voltages continu­ously (Note: These information can be read out only via service e-mail.)
Example: answer e-mail on a service e-mail at unidos.webline9@ptw.de:
).
Regularly Maintenance
webline
UNIDOS webline SERVICE 2006-02-01 12:04:09
System information
Device name: UNIDOS webline Type number: T10021 Serial number: 5 Hardware revision: A02 Firmware: 1.12-1 Calibration date: Next calibration:
CPUBoard SerNr.: 0 CPUBoard Rev.: A01
Measuring Unit
Firmware: 1.00-9 Serial number: 1 Calibration: 1 Last Zeroing: 2006-02-01 12:03:12
Power Supply
Firmware: 1.01 Serial number: 0 Calibration: 1
Main: 12.017V, OK 12V: 12.004V, OK 5V: 4.991V, OK
3.3V: 3.294V, OK Bat: 11.133V Battery State: 0 Battery Level: 0
System-Internals
StandbyState: 0 System Flags: 0 Debug Flags: 0 Storetime: 2006-02-01 11:04 MB_BootCounter: 100 MB_ErrorBits: 0 MB_ControlWord: 0 MB_AccessErrCnt0: 0 MB_AccessErrCnt1: 0 MB_AccessErrCnt2: 0 MB_AccessErrCnt3: 0 MU_Ready: 100 MU_ErrorBits: 0 MV ErrorCounter: 0 MV ErrorCntTotal: 0 AllCorrectionsOff: 0 StabilizationTimeLeft: 0 SpurTrapCounter: 0 MbComMultipleTelegrams: 0 PSinfo: 4 DB Error:: 0 INET_State: 0 UDP_State: 0 TCP Error: 0 E-Task Error: 0 E-Task Time: 0 E-Task Data: 0
6 D776.133.0/1
Page 7
Troubleshooting
webline

4 Troubleshooting

The instruction manual of UNIDOS the procedures to apply, if problems occur during the operation of the device or error messages are shown on the display. The particular sections are:
10.1 Error Messages on the Display
10.2 Checking the Measured Value
10.3 Problems Operating the UNIDOS
webline
describes
webline

5 Functional units

Opening of the device and working at the device must be performed only by PTW-Freiburg or by persons authorized by PTW-Freiburg.
WARNUNG
Shock Hazard – Dangerous voltages become touchable with the opening of the device. For the connecting systems BNT and TNC: the en­tire enclosure of the measuring module is on the potential of the high voltage. In order to switch off the device safely, it must be turned off and separated from the power line. Moreover, all connection cables must be disconnected in order to prevent contact to external voltages. At last the rechargeable batteries must be re­moved.

5.1 Front panel unit

CCFL Inverter for
backlight
TFT Display
Power
Figure 1: Front panel unit
The front panel unit is a compact unit which con­tains and connects the subunits shown above.
Plastic foil keyboard and navigation knob
Main CPU board
I/O
The main CPU executes the firmware of the device with all aspects like graphic user interface, display control, presentation of measured values, commu­nication with interfaces, etc. The main CPU stores all device settings, detector data and also the re­sults of system tests.
D776.133.0/1 7
Page 8
In case of exchanging the front panel unit all these data have to be saved first and later on entered into the new front panel unit by means of the service software.
The main CPU contains also the real-time clock and the associated back-up battery.

5.2 Measuring Module

Functional units
webline
Input
ADC
Amplifier
HV measurement
Desiccator
HV (to detector)
External HV
Internal
HV
HV source selection HV polarity
µController
EEPROM
Figure 2: Measuring module
The measuring module is designed as an inde­pendent unit and can be exchanged with another one of the same connecting system.
Serial I/O
Control
DC DC
Power
Electrical Isolation
It stores all relevant amplifier data like calibration factors and zeroing results as well as calibration factors for measuring high voltage. With older edi­tions of the measuring module, only the internal HV can be measured and monitored.
Since the results of the system tests are stored in the main CPU, these data are however no more comparable after a replacement of the measuring module.
8 D776.133.0/1
Page 9

5.3 Main Board

webline
Functional units
Supply voltages
Power supply
units
Voltages
measurement
Relay
AC / DC
Charge control
Wide range
Electrical Isolation
Figure 3: Main board
The main board contains the individual subunits for generation and monitoring of the supply voltages as well as for charging the rechargeable batteries. It fulfills connection functions beside that. Its name originates from its physical size and implies in no way, that it houses the main CPU.
Mains
Connector &
Filter
Rear panel
Rechargeable
batteries
Host I/O (SERIAL)
Ethernet
TRIP Output
The calibration factors for measuring of the supply voltage are stored in the main board.
The results of the voltage monitoring can be read out via service e-mail. It is not necessary to execute measurements inside the device.
D776.133.0/1 9
Page 10

6 Service Software

The service software has to fulfill three tasks:
Backup of the device settings as well as the
detector data base to a connected PC
Update of the device firmware
Restore of the device settings as well as the
detector data base
If a firmware update from PTW-Freiburg is released for the UNIDOS ware version on the newest version by means of this software.
The software uses exclusively the RS232 interface for communications.
webline
, the user can update the firm-
Service Software
webline
A firmware update over the Ethernet interface is not possible.
The software will probably be available in June
2006.
10 D776.133.0/1