PTW DIAMENTOR E2 User Manual

Page 1
User Manual
DIAMENTOR E2
T11033
from SN 02000 and higher
D651.131.00/05 en 2012-11 Gü/Hn/Kau
Page 2
General Information
General Information
The product bears the CE-mark "CE-0124" in
accordance with the Council Directive 93/42/EEC about Medical Devices and fulfills the essential requirements of Annex I of this di­rective. The product is a class IIb device (MDD).
Furthermore, the DIAMENTOR E2 helps fulfill
the requirements of the EC patient directive 97/43/EURATOM regarding the acquisition and documentation of the radiation exposure of pa­tients during X-ray examinations.
Product standard
The product fully complies with the require­ments of the following standards: IEC 60580
Electrical safety
The product fully complies with the require­ments of the following standards: IEC 60601-1 ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1
Electromagnetic immunity
The product fully complies with the require­ments of the following standards: IEC 60601-1-2 IEC/CISPR 11 class B
The user manual is an integral part of the prod-
uct. It should always be kept near the product. Observance of the manual is a prerequisite for proper product performance and correct opera­tion.
Operator safety, specified measuring accuracy
and interference-free operation can be guaran­teed only if original products and parts are used. Furthermore only the accessories listed in this manual are approved by PTW-Freiburg and may be used in conjunction with the product, or else accessories whose use has been ex­pressly permitted by PTW-Freiburg. Safe op­eration and proper product performance are not guaranteed if accessories or consumables from other manufacturers are used.
PTW-Freiburg cannot be held liable for dam-
ages resulting from the use of accessories, consumables from other manufacturers or when the user ignores the instructions and informa­tion given in this manual.
The warranty period is 1 (one) year and begins
on the day of delivery. It is unaffected by repairs covered by the war­ranty regulations.
PTW-Freiburg considers itself responsible for
safety, reliability and performance of the prod­uct only, if assembly, extension, readjustment, modification or repair is carried out by PTW-Freiburg or by persons authorized by PTW-Freiburg, and if the product is used in compliance with the technical documentation.
In case of any questions concerning service,
support or warranty please contact your sup­plier.
This manual is in conformity with the product
specifications and all applicable safety stan­dards valid at printing date. All rights are re­served for devices, circuits, techniques, soft­ware and names referred to in the manual.
PTW-Freiburg reserves the right to make modi-
fications. Please contact PTW or your local representa­tive for the most current information concerning the products of interest.
No part of the technical documentation may be
reproduced without written permission from PTW-Freiburg.
PTW-Freiburg is registered manufacturer ac-
cording to the ElektroG (Elektro- und Elek­tronikgeräte-Gesetz). Elektro-Altgeräte-Register (EAR) Registration number DE15599992.
2
D651.131.00/05 en
Page 3
General Information
PTW-Freiburg works in strict accordance with a
quality management system which is continu­ously updated according to national and inter­national standards.
PTW-FREIBURG
Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de
D651.131.00/05 en 3
Page 4
Table of Contens
Table of Contents
1 INTENDED USE 5
2 FOR YOUR SAFETY 5
3 OPERATING CONTROLS AND INDICATORS 11
4 FIRST TIME OPERATION 12
4.1 Setting up the DIAMENTOR E2 12
4.2 Connecting the Supply Voltage 12
4.3 Connecting the Chamber and Printer and Putting into Operation 15
4.4 Mounting the ionization chamber 15
4.5 Basic Device Settings (Setup Menu) 16
5 PERFORMING A MEASUREMENT 17
5.1 Constancy Check 17
5.2 Measurement 17
5.3 Air Density Correction 18
5.4 Print out 18
6 ERROR MESSAGES AND TROUBLESHOOTING 19
7 CALIBRATION 20
8 PERIPHERALS 22
8.1 Printer 22
8.2 PC 22
9 CLEANING, PREVENTIVE MAINTENANCE AND DISPOSAL
9.1 Cleaning 23
9.2 Preventive Maintenance 23
9.3 Disposal of the Product 24
10 TECHNICAL SPECIFICATIONS
23
25
11 ACCESSORIES AND SPARE PARTS 27
12 SPECIAL ADJUSTMENTS 27
13 ELECTROMAGNETIC COMPATIBILITY (EMC) ACCORDING TO IEC 60601-1-2 28
APPENDIX: CHINA ELECTRONIC INDUSTRY STANDARD COMPLIANCE 32
LITERATURE 34
INDEX 35
4
D651.131.00/05 en
Page 5
Intended Use

1 Intended Use

The DIAMENTOR E2 is a device for measurement of the dose area product and the dose area product rate in diagnostic radiology. It fulfills the require­ments of IEC 60580.
The dose area product is measured with an ioniza­tion chamber mounted directly to the light beam diaphragm housing.
The DIAMENTOR E2 is a single channel micro­processor-controlled measuring system.
The DIAMENTOR E2 is equipped with a serial con­nector (socket SERIAL) to connect a label printer or a PC.
This manual only describes the DIAMENTOR E2.
A detailed description of the RS232 interface is
supplied on request (document number D651.131.1).

2 For Your Safety

The following safety statements are divided into
DANGER, WARNING, CAUTION and NOTE.
Definitions
DANGER
indicates an imminent hazard. If not avoided, the hazard will result in death or serious injury.
WARNING
indicates a hazard. If not avoided, the hazard could result in death or serious injury.
CAUTION
indicates a potential hazard. If not avoided, this hazard could result in minor personal injury or product/property damage.
NOTE
Provides useful information to assure that you get the most from your equipment.
D651.131.00/05 en 5
Page 6
For Your Safety
Safety Information
DANGER
Explosion Hazard – The product is not de-
signed for use in areas of medical locations where an explosion hazard may occur. An ex­plosion hazard may result from the use of flammable anesthetics, skin cleansing agents and disinfectants. Furthermore, the product is suitable for use in an oxygen-enriched atmos­phere only with certain restrictions. The atmos­phere is oxygen-enriched when the room air contains more than 25 % oxygen or nitrous oxide.
WARNING
Shock Hazard – Strictly observe the following
warnings. Failure to do so may endanger the lives of the patient, the user and other persons involved.
The product is a system device of
IEC 60601-1 protection class II in
combination with functional earth. It may be
put into operation only when connected to a properly installed power outlet with earthing contact.
The installation of the product may only be
performed by trained electro-technical staff.
Before using the product, the user must as-
certain that it is in correct working order and operating condition.
Inspect the connection cables for damage
before use. Replace damaged cables and connectors immediately.
When disconnecting the device from the
power line, first remove the connector from the wall outlet. Then disconnect the cable from the device.
Liquids must not enter neither the display
unit nor the adapter. If liquids have entered the product, it must be thoroughly inspected before being used again.
Extension cords must not be used.
Operation in the patient environment:
With all peripheral devices connected, the product represents a medical system. The product may be operated in the patient envi­ronment only if
no peripheral devices are connected to the
serial interface or
the peripheral devices connected meet the
requirements IEC 60601-1/ ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1
the peripheral devices connected meet the
requirements IEC 60950-1 / UL 60950-1 and are operated outside the patient environ­ment.
Exclusion of operation as device with pa­tient contact:
The device is not for use in direct contact
with the patient. Neither the device nor any peripheral device may have contact to the patient.
Never touch the patient and open connec-
tors of the device at the same time.
Exclusion of operation as a controlling in­strument:
The device is only for use to measure the
dose within radiation field. The product must not be used to control radiodiagnostic equipment or radiotherapy units.
Devices on which moisture condensation
has developed as a result of temperature changes must not be switched on unless completely dry.
6
D651.131.00/05 en
Page 7
For Your Safety
WARNING
Shock Hazard – Strictly observe the following
warnings. Failure to do so may endanger the lives of the patient, the user and other persons involved.
Use of peripheral devices:
Peripheral devices (PC, printer) may only be
connected if they meet the requirements of IEC 60950-1 / UL 60950-1 or of standard IEC 60601-1 ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1. These devices may only be operated in the patient environment, if it is guaranteed that they meet the requirements of IEC 60601-1/ ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1.
Devices may be connected to other devices
or to parts of systems only if it has been as­certained that this connection does not im­pair the safety of the patient, the operator or the environment.
If the device specifications do not contain in-
formation as to connecting the device to other equipment, you must consult the manufacturer of the other equipment or an expert about the effects of the connection on the patient, the operator or the environment.
Always observe standards IEC 60601-1/
ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1.
WARNING
Suffocation Hazard – Dispose of the packag-
ing material, observing the applicable waste­control regulations. Keep the packaging mate-
rial out of children's reach. Patient Hazard – The product is a medical
electrical device and must only be handled by persons who are trained in the use of such equipment and are capable of applying it prop­erly. The operator must be trained in the use of the device.
Risk of Poisoning – Chemicals required for
application or maintenance of the device, for in­stance, must under all circumstances be stored, prepared, and kept at hand in their specific con­tainers. Failure to observe this instruction may result in severe consequences for the patient.
WARNING Equipment failure - Electromagnetic fields are
capable of interfering with the proper perform­ance of the device.
The DIAMENTOR E2 should not be used in
the immediate vicinity of or placed on top of or below other equipment. However, if the application requires an arrangement of de­vices as described above, the DIAMENTOR E2 should be watched to en­sure its proper functioning in the specific ar­rangement.
The customer or user of the
DIAMENTOR E2 should assure that it is used in an electromagnetic environment as described in IEC 60601-1-2.
D651.131.00/05 en 7
Page 8
For Your Safety
WARNING
Bodily Injury! Equipment Damage!
Modifications, additions or maintenance of the equipment may be conducted only by suitably qualified and trained persons.
Spare parts must be approved by the manufac­turer.
After any modifications of the equipment, ap­propriate inspections and tests have to be car­ried out.
Failure to comply may lead to fatal or serious personal injury and / or damage to the equip­ment, as well as void the warranty.
WARNING The system has no means of isolation from the
supply voltage. When the mains plug is con­nected to the power line or the cable is clamped into the DC supply, the system is powered, even if the DIAMENTOR E2 is turned off.
CAUTION
Equipment damage - Handle the product with
care to avoid equipment damage.
Verify that the power ratings indicated on the device are the same as those of your power supply system.
Always use the product in compliance with the user manual. Otherwise, the intended protection can be reduced.
Use the product only in conjunction with the de­vices approved for this purpose by PTW and/or the devices listed in the section "Intended Use".
The ambient conditions specified in the "Tech­nical Specifications" section must be ensured at all times.
CAUTION
Equipment damage - If you move the X-ray
unit do not bump into doorframes or walls to avoid damage to the DIAMENTOR E2.
8
D651.131.00/05 en
Page 9
For Your Safety
NOTE
Please observe the user manuals of all con­nected system components!
NOTE
Set up the device so that the operator has a clear, unobstructed view of the control panel.
NOTE
Do not position the device so that it is difficult to operate the disconnecting device.
Symbols on Product and Nameplate
Symbol Description
Observe the user manual.
Please refer to user manual.
The product bears the CE-mark.
The product is certified by VDE (German Association for Electrical, Electronic and Information Tech­nologies)..
The product is certified by UL (Un­derwriters Laboratories) for both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.
Figure 1: Definition of patient environment
The device is only suitable for use with direct current.
Manufacturer and date of manu­facture
Reference number
Serial number
Separate collection for electrical and electronic equipment! (refer also to section Disposal of
the Product)
Labeling according to “Administra­tion on the Control of the Pollution caused by Electronic Information Products (ACPEIP)“ (China RoHS) (refer also to Appendix)
The connection of the DIAMENTOR E2 to the ground wire does not have any protection func­tion here, this serves as function earth.
D651.131.00/05 en 9
Page 10
For Your Safety
Information for Installation
Inside the patient Outside the patient Not medical used area environment environment
Power
supply unit
T99052
Power
limiter
T26060
or
Printer / PC
DIAMENTOR E2
PTW
T11033
IEC 60950-1
or
Figure 2: Note for installation in the patient environment
Printer / PC
IEC 60950-1
10
D651.131.00/05 en
Page 11
Operating Controls and Indicators

3 Operating Controls and Indicators

Figure 3: DIAMENTOR E2
1 Socket SERIAL for connection of label printer or PC (RS232 interface; RJ12 Western Plug) 2 Folding stand 3 Connection ionization chamber (connecting system A ) 4 Connection adapter 5 On/Off switch Device switched on
Device switched off
6 Display (8 digit seven-segment display)
7 TEST ( 8 PRINT (
menu mode. If the print mode is set to ”Prn off” the print function is inactive.
9 RESET (
mode
D651.131.00/05 en 11
) button to conduct a device test and for cursor control in the menu mode
) button to print out the measured value on a connected printer and for cursor control in the
) button to reset the reading to zero and to confirm settings and quit a menu in the menu
Page 12

4 First Time Operation

First Time Operation

4.1 Setting up the DIAMENTOR E2

The DIAMENTOR E2 can be set up by means of the folding stand on the back of the device.
Alternatively, the device can be fixed to the wall or a plain surface by means of the velcro type ribbon supplied with the DIAMENTOR E2. To do so, fasten the self-adhesive back of the ribbon to the back of the DIAMENTOR E2 and the opposite ribbon to the wall or surface.
4.2 Connecting the Supply Volt­age
WARNING
The DIAMENTOR E2 must only be connected
to one of the two ways described below.
Use only the listed original cable from PTW.
Observe the installation notes of the supply voltage.
WARNING The system has no means of isolation from the
supply voltage. When the mains plug is con­nected to the power line or the cable is clamped into the DC supply, the system is powered, even if the DIAMENTOR E2 is turned off.
The DIAMENTOR E2 can be connected with two different adapters for supply voltage (refer to fig­ure 2).
1. Connection with PTW power supply T99052
2. Connection with power limiter T26060 to an external supply voltage according to IEC 60601­1/ ANSI/AAMI ES 60601-1 / CAN/CSA-C22.2.No.60601-1
To connect the DIAMENTOR E2 follow the descrip­tion of the used adapter.
1
PTW
Mains
(110... 230)VAC
2
External Supply Voltage
Double / Reinforced insulation
IEC 60601-1 2 MOPP
24 VDC+/-10%
Power Supply T99052
PTW
T26060
Figure 4: DIAMENTOR E2 connecting options
PTW T25024-2
PTW T25024-2
PTW
DIAMENTOR E2
PTW
DIAMENTOR E2
PTW
Chamber
PTW
Chamber
12
D651.131.00/05 en
Page 13
First Time Operation

4.2.1 Connection with PTW power supply T99052

WARNING
The device is designed to meet the require­ments of protection class II. It may be operated only with PTW power supply T99052. Extension cords must not be used.
The DIAMENTOR E2 is connected via external power supply T99052 to the mains (refer to 3, item 1).
CAUTION FOR USA Installation in USA - In case of a 240 VAC-
line voltage the DIAMENTOR E2 must only be connected to center tapped circuits.
The DIAMENTOR E2 is connected to the power supply via supply cable T25024-2.
Figure
Place the power supply so that mechanical
damages and penetration of liquids are avoided. Do not place the power supply on the ground! Avoid injury hazard due to cable loops!
Connect the power supply unit to the adapter
socket (refer to necting the power supply unit to the power sup­ply line, verify that the ratings of your local power line are those indicated on the power supply.
Figure 3, item 4). Before con-

4.2.2 Connection with power limiter T26060 to external supply voltage

CAUTION
Make sure that the release of the installation manufacturer is on your hand before you con­nect the DIAMENTOR E2 via power limiter T26060 to the installation.
WARNING
The power source has to guarantee safety by means of 2 MOPPs (Means Of Patient Protec­tion) according to IEC 60601-1/ ANSI/AAMI ES 60601-1/ CAN/CSA-C22.2.No.60601-1. The DIAMENTOR E2 may be operated only with power limiter T26060. Extension cords must not be used.
The connection of the power limiter T26060 re­quires technical know-how and may only be performed by trained electrotechnical staff ac­cording to IEC 60601-1/ANSI/AAMI ES 60601-1 /CAN/CSA-C22.2.No.60601-1.
The DIAMENTOR E2 is connected to the power limiter T26060 via supply cable T25024-2 and to a low voltage supply (24 VDC) (refer to item 2).
Figure 3,
Place the power limiter so that mechanical
damages and penetration of liquids are avoided. Do not place the power limiter on the ground! Avoid injury hazard due to cable loops!
Connect the power limiter unit via supply cable
T25024-2 to the adapter socket (refer to 3, item 4). Before connecting the power limiter to the DC supply voltage, verify that the ratings of the DC supply voltage are those indicated on the power limiter.
D651.131.00/05 en 13
Figure
Page 14
First Time Operation
Installation of the power limiter T26060 (refer to
Figure 5)
The DIAMENTOR E2 may be operated only with power limiter T26060 in compliance with the follow­ing conditions.
The earth and enclosure leakage currents of the
whole installation have to be measured when putting into operation. The earth and enclosure leakage current may not exceed the maximum values of 300 µA (safety class I) or 150 µA (safety class II) depending on the safety class.
The user replaceable mains (line) fuse must be
an approved type acceptable to the authorities where the equipment is sold.
The complete system has to be evaluated ac-
cording to the requirements of IEC 60601-1 ANSI/AAMI ES 60601-1/ CAN/CSA-C22.2.No.60601-1 by the end user.
To achieve the requirements for electro-
magnetic compatibility it is recommended to connect the DC power supply output to GND and earth (refer to tion has no protective function (function earth).
Maximum Earth and Enclosure Leakage Current Measurement according to IEC 62353
Class I equipment: The maximum leakage current is 300 µA.
Class II equipment: The maximum leakage current is 150 µA.
Figure 5). This earth connec-
Power Supply
Double / Reinforced insulation IEC 60601-1
2 MOPP 24 VDC ±10%
Power GND
24 VDC
white
brown
GND
Earth Connection
Figure 5: Installation of the power limiter T26060
PTW
T26060
14
D651.131.00/05 en
Page 15
First Time Operation

4.3 Connecting the Chamber and Printer and Putting into Operation

Always turn off the device before connecting or
removing peripheral devices.
Connect the chamber to 3 (refer to Figure 3).
Use the original cables from PTW (available in various lengths up to 40 meters (130 ft)).
Socket SERIAL (Figure 3, item 1): Connect the
printer cable or RS232 interface cable for PC.
Turn on the DIAMENTOR E2 with the On/Off
switch (refer to version will be displayed for about 2 seconds, followed by a display test. After completion of
the test the DIAMENTOR E2 is operational.
Figure 3, item 5). The firmware

4.4 Mounting the ionization chamber

General information
The DIAMENTOR E2 can be operated with every PTW-DIAMENTOR chamber featuring connecting system A (
The chamber calibration factor must be entered for the connected chamber (section ing the chamber, do not forget to enter the new chamber calibration factor (section
The DIAMENTOR E2 connection cables are avail­able in various lengths (max. 40 meters (130 ft.)). Use only the original cables from PTW.
Figure 3, item 3).
4.5). When replac-
4.5.)
D651.131.00/05 en 15
Page 16
First Time Operation

4.5 Basic Device Settings (Setup Menu)

General information
The setup menu will be activated by pressing the TEST and RESET buttons simultaneously for about 2 seconds In the menu mode the buttons are used for cursor control, to change and confirm settings and to quit a menu.
All adjustments performed in the setup menu are saved to the non-volatile memory.
Selecting a measuring quantity
When the setup menu is called up, the current set­tings for the measuring quantity are shown. The settings available for the measuring quantity are
dose area product doSE
dose area product rate doSEr
During irradiation display
of the detected dose area product rate; after irradiation display of the total dose
area product Auto
Entering the chamber calibration factor
Check the setting for the chamber calibration factor. The original chamber factor K
is indicated on the
G
chamber housing.
Each time you connect another chamber, you should enter the new chamber calibration factor as described below.
To change the chamber calibration factor confirm the setting for the measuring quantity with RESET (
). The menu item for setting the chamber calibra-
tion factor CHAF will be called up automatically.
The current setting for the chamber calibration fac­tor will be displayed. Change the current digit with PRINT () and move to the next digit with TEST ().Chamber calibration factors can be set in the range (0.01 ... 9.99).
Confirm the selected setting with RESET (). After confirmation, the measuring mode will be entered automatically. The print mode can be switched on / off with TEST button (display Prn on / Prn off).
PC can be connected only with disabled print mode.
If the print mode is set to ”Prn off” the print function of PRINT button is inactive. To obtain a printout of the measured values, first activate print mode.
Use the TEST () button to toggle between the available settings. Confirm the selected setting with RESET (). After confirmation, the menu item for setting the chamber calibration factor will be called up automatically.
NOTE
Ionization chambers calibrated by PTW­Freiburg are suitable for over couch tubes and under couch tubes. To compensate for the cali­bration error inherent in the system, the cham­ber calibration factor may be multiplied by a cor-
rection factor. The correction factor for over
couch tubes is 1.10, for under couch tubes,
0.95 (refer to section 7).
We recommend to have the system calibrated annually or every 2 years by specially trained staff on site or in our calibration laboratory of PTW-Freiburg.
16
D651.131.00/05 en
Page 17
Performing a Measurement

5 Performing a Measurement

5.1 Constancy Check

CAUTION
Faulty measurement - Check the device per-
formance every day, before putting the DIAMENTOR E2 into operation.
Before using the DIAMENTOR E2, check that its measurement range matches the examina­tion technique that will be applied.
Reset the DIAMENTOR E2 before each exami­nation.
The DIAMENTOR E2 will conduct an internal con­stancy check as soon as the TEST button is pressed for 2 seconds. The return value of the hardware test is the proportional relation between measured value and an internal test value. After about 5 seconds the DIAMENTOR E2 returns auto­matically to the measuring mode.
After the device has been turned on, the firmware version has been displayed for about 2 seconds and the display test has been completed success­fully measured values are displayed according to the selected mode as soon as radiation is detected.
„Auto“ measuring mode (Setup: Auto)
During irradiation display of the detected dose area product rate in µGym²/s; after irradiation display of the total dose area product in µGym².
„Dose rate“ measuring mode (Setup: doSEr)
Display of the dose area rate in µGym²/s. This mode is indicated by a square at the left margin of the display.
„Dose“ measuring mode (Setup: doSE)
Display of the total dose area product in µGym². In this mode the display will be refreshed every sec­ond.
If the proportional relation between measured value and test value is not within the range of 95 ... 105 the device needs to be calibrated before it can be used again.

5.2 Measurement

NOTE
Allow for a pause of at least 10 seconds be­tween the activation of the device and the first measurement (section 10).
Press the RESET button before starting the first measurement to bring the DIAMENTOR E2 into a defined initial state.
CAUTION
Faulty measurement - Accessories placed be-
tween the chamber and the patient, such as a pediatric filter or beam applicator, will reduce the dose area product relevant for the patient. In this situation the DIAMENTOR E2 will display measuring values that are too high. According to national regulations in Germany, however, it is sufficient to record the measuring value and the accessories in use for reconstruction of the radiation exposure level. If these accessories are used for all measurements, the DIAMENTOR E2 should be recalibrated with the accessories installed.
Equipment damage - The DIAMENTOR E2
should be operated only when an ionization chamber is connected or when the chamber connection at the device is covered.
D651.131.00/05 en 17
Page 18
Performing a Measurement

5.3 Air Density Correction

To get the highest possible measuring accuracy the indicated / determined value of the DIAMENTOR E2 must be multiplied with a correc­tion factor to the air density correction.
The air density correction factor is calculated by the following formula:
Tp
K
= 1 013 hPa
p
0
p = current air pressure T
0
T = current temperature
According to device, the correction can perhaps be executed over a correction menu or an used soft­ware.
0
=
D
= 293 K (= 20°C)
Tp
0

5.4 Print out

The print mode can be switched on / off (display
Prn on / Prn off) over setup menu (refer to sec-
tion
4.5).
Set print mode to “Prn on
Connect the printer to the DIAMENTOR E2
(refer to
Press the PRINT button to obtain a printout of
the measured value.
To start a new measurement press the RESET
button to reset the display to zero.
Figure 3, item 1).
18
D651.131.00/05 en
Page 19
Error Messages and Troubleshooting

6 Error Messages and Troubleshooting

Message Cause Remedy
OFL-Err (Overflow
Error)
Error 51 communication error Turn the device off and on again. When
Error 54 Memory malfunction Turn the device off and on again. When
HU-Err Chamber voltage missing, device is un-
The dose area product has exceeded
20 000 000 µGym²
or
the dose area product rate has exceeded
15 000 µGym²/s
able to measure
Press the RESET button.
the message appears again the device
must be repaired.
the message appears again the device
must be repaired.
Disconnect the chamber cable from the
DIAMENTOR E2. When the message
continues appearing the device is defec-
tive and must be repaired.
When the message clears the chamber
cable is defective and must be replaced.
D651.131.00/05 en 19
Page 20

7 Calibration

Calibration
Electrical Calibration
The DIAMENTOR E2 is electrically calibrated in the factory. You can verify the calibration as described in section
5.2. If the electrical calibration values are outside the admissible range, the device should be returned to PTW-Freiburg for repair, as the electri­cal calibration cannot be corrected on site.
Radiological Calibration
The chamber calibration can be verified on site (see below) or in the PTW calibration laboratory. We recommend to have the system calibrated annually or every 2 years in our calibration laboratory of PTW-Freiburg.
Some Basic Facts On Chamber Calibration (PTW-Freiburg Calibration Laboratory)
For verification of the chamber calibration fac-
tor only the ionization chamber (without the measuring system) needs to be returned to the factory.
The ionization chambers calibrated in the PTW
calibration laboratories can be used with over­couch and undercouch tubes.
In order to take into account the different ab-
sorption conditions of overcouch and under­couch tubes, an additional absorber equivalent to 0.35 mm AI is placed between the X-ray tube and the ionization chamber during the factory calibration. Therefore the measure­ment error of overcouch tube installations where there is no absorber between chamber and patient is up to - 9 %. In conjunction with undercouch tubes the absorption of the patient couch situated between the patient and the chamber is equivalent to 0.5 to 0.7 mm AI, which results in a measurement error of up to + 5 %.
Enter the new chamber calibration factors as
described in section chamber calibration factor K
4.5. For this, multiply the by a correction
G
factor of 1.10 for over-couch tubes and of
0.95 for under-couch tubes.
Verifying the Chamber Calibration in Conjunc­tion with the X-Ray System
Specially trained staff will be able to perform the chamber calibration on site. The required accesso­ries/tools are:
1 diagnostic dosemeter (e.g., PTW-DIADOS,
PTW-DALi, PTW-NOMEX or PTW-UNIDOS)
1 X-ray film, 13 cm x 18 cm (for radiography)
1 measurement phantom (for fluoroscopy)
1 adapter for DIADOS detector (part no.
T20002), (for fluoroscopy)
Check the electrical calibration as described in sec­tion
5.1.
Radiography
Make adjustments for a field size of approx.
10 cm x 10 cm and mark the area on the pa­tient couch.
Place the X-ray film on the patient couch and
make a radiograph (70 kV and 40 mAs).
Develop the film (only briefly to obtain sharp
edges) and measure the exact size of the field (noting down the result A).
Place the detector of the diagnostic dosemeter
in the center of the area marked on the patient couch. Make a radiograph (70 kV and 40 mAs).
The chamber calibration factors are noted
down on the calibration certificate.
20
D651.131.00/05 en
Page 21
Calibration
Note down the reading M
Reference
shown on the
diagnostic dosemeter.
Note down the dose area product reading
shown on the DIAMENTOR E2 (DAP
Measured
).
Calculate the dose area product
DAP
Calc
= M
Reference
x A where A is the field size.
Repeat the measurement 3 times, noting down
the results.
Fluoroscopy (undercouch tube)
Place the measurement phantom without de-
tector on the patient couch (phantom at the bot­tom).
Start fluoroscopy and determine the field size A
in the plane of the phantom by means of the phantom depicted on the screen.
Fluoroscopy (overcouch tube)
Place the measurement stand without detector
on the patient couch (phantom at the top).
Start the fluoroscopy and determine the field
size in the plane of the phantom by means of the phantom depicted on the screen.
Turn the measurement stand and put the detec-
tor and its holder in place (phantom at the bot­tom, detector at the top).
Start another fluoroscopy for approx.
30 seconds.
Note down the reading M
Reference
shown on the
diagnostic dosemeter.
Note down the dose area product reading
shown on the DIAMENTOR E2 (DAP
Measured
).
Turn the measurement stand and put the detec-
tor and its holder in place (phantom at the top, detector at the bottom).
Start another fluoroscopy for approx.
30 seconds.
Note down the reading M
Reference
shown on the
diagnostic dosemeter.
Note down the dose area product reading
shown on the DIAMENTOR E2 (DAP
Measured
).
Calculate the dose area product
DAP
Calc
= M
Reference
x A where A is the field size.
Repeat the measurement 3 times, noting down
the results.
Calculate the dose area product
DAP
Calc
= M
Reference
x A where A is the field size.
Repeat the measurement 3 times, noting down
the results.
Calculating the New Chamber Calibration Factor
Calculate the mean value of (dose area prod-
uct) DAP
Calculate the mean value of (dose area prod­uct) DAP
.
Calc
Measured
.
Calculate the new chamber calibration factor
K
, using the mean values and the old cham-
new
ber calibration factor:
DAP
K
new
=
DAP
Calc
Measured
K
old
Saving the New Chamber Calibration Factor
Enter the new chamber calibration factor as de­scribed in section
4.5.
D651.131.00/05 en 21
Page 22

8 Peripherals

Peripherals

8.1 Printer

WARNING
Please read the printer manual before putting the printer into operation!
The DIAMENTOR E2 may be operated in the patient environment only if the connected printer meets the requirements of IEC 60950-1 / UL 60950-1 and if the printer is operated out­side the patient environment.
The connected printer meets the requirements of IEC 60950-1/ UL 60950-1 only if it is oper­ated with the supplied original power supply.
To obtain a printout of the measured value connect the printer to the DIAMENTOR E2 ( item 1).
Switch on the printer first or simultaneously with the
DIAMENTOR E2. Set the print mode to “Prn on
(see section
The printout contains the device designation „PTW DIAMENTOR E2“ and the measured value:
4.5).
PTW DIAMENTOR E2
Figure 3,

8.2 PC

WARNING
Please read the printer manual before putting the printer into operation!
Operate the DIAMENTOR E2 in the patient en­vironment only if the connected PC fulfills the requirements of the IEC 60950-1 / UL 60950-1 and is installed outside the patient environment.
The PC connected to the device must comply with the standards IEC 60601-1/ ANSI/AAMI ES 60601-1/ CAN/CSA-C22.2.No.60601-1 if it is to be oper­ated in the patient environment.
WARNING
Before connecting a PC via the RS232 inter-
face, set the print mode to “Prn off” (see sec-
tion 4.5).
Use only the original RS232 interface cable T25030-3.
It is also possible to connect a PC to the RS232 interface of the DIAMENTOR E2 (socket SERIAL).
22
12.34 uGym2
When connecting a PC to the DIAMENTOR E2, please pay close attention to the safety information in section
A detailed description of the interface is supplied on request: document number D651.131.1.
2.
D651.131.00/05 en
Page 23
Cleaning, Preventive Maintenance and Disposal

9 Cleaning, Preventive Maintenance and Disposal

9.1 Cleaning

Ionization chambers
The outer shells of the ionization chambers are made of macrolon (polycarbonate). The electrode surfaces are made conductive by vacuum metalliza­tion. In this process a transparent metal coating is applied to the electrode. In order to keep the coat­ing transparent, ionization chambers must not be treated or come into contact with alcohol or sol­vents.
Cleaning agents and solvents contained in disinfec­tants (alcohols) have proved to be particularly harmful.
Wipe the chambers clean with a moist cloth.
Please use a common disinfection agent for
medical instruments that is listed in your national hygiene institute (e.g., VAH in Germany). If your country does not provide such a list we recom­mend using either an aldehyde based solution or a solution with a quaternary ammonium com­pound.
Surface
Connection cables
disconnect the cables from the devices, pulling
on the plug, not on the cable
for cleaning, rub the cables down with a cloth
moistened with soap water or disinfectant
never immerse the cable in liquid!

9.2 Preventive Maintenance

WARNING
Hazards to Persons! Equipment Damage!
Only specially trained staff is allowed to perform the following service tasks.
Further service tasks may only be carried out by PTW-Freiburg or by persons authorized by PTW-Freiburg.
Check before each use
Before each use, visually inspect the device
and the cables for signs of mechanical dam­age
CAUTION
Equipment damage - Do not use disinfectants
on a phenol base or peroxide compounds to disinfect the surface.
First of all, disconnect the devices from the
power line, before attempting to clean or disin­fect them.
Rub the devices down with a moist cloth; no
liquid must enter the devices. Due to the sur­face geometry, a thorough disinfection is not possible. The device cannot be sterilized.
D651.131.00/05 en 23
Also, conduct a constancy check as described
in section
If damages or malfunctions are identified, the
device must be repaired before it is used again.
Repair
Repairs may only be carried out by PTW-Freiburg or by persons / companies authorized by PTW-Freiburg.
5.1.
Page 24
Cleaning, Preventive Maintenance and Disposal
Safety inspections
Only regularly inspected product s are fail-safe. To preserve the functional security and operational safety, the safety inspections are to be executed at regular intervals according to national regulations.
These inspections must only be performed by inde­pendent persons with adequate training and experi­ence.
It is recommended to execute the safety inspections annually or every 2 years.
Technical Inspection of the Measuring Systemm
The product must be inspected at regular intervals according to national regulations. It is recom­mended to perform the technical inspections of the measuring system annually or every 2 years at PTW-Freiburg or a qualified calibration laboratory.

9.3 Disposal of the Product

The typical lifetime of the DIAMENTOR E2 is about 10 years.
At the end of the product life ears the DIAMENTOR E2 components must be disposed of in compliance with the applicable waste control regulations. The different materials must be sepa­rated and recycled as appropriate.
DIAMENTOR E2 does not contain hazardous mate­rials.
The electronic components must be recycled ac­cording to local regulations.
The cost for an eventual return at the end of the product life time is to be borne by the customer.
Inspections must also be performed after repairs or after each influence which may have changed the behavior of the product.
Calibration
The exact calibration factor and supplementary data are included in the calibration certificate.
We recommend to have the DIAMENTOR E2 cali­brated annually or every 2 years at PTW-Freiburg or by specially trained staff on site.
24
D651.131.00/05 en
Page 25
Technical Specifications

10 Technical Specifications

Only values for which tolerances or limits are specified are guaranteed. Values for which no tolerances are specified are for information only.
Device designation and use DIAMENTOR E2 T11033
dose area product meter according to IEC 60580 for measurements in diagnostic radiology
Manufacturer PTW-Freiburg
Display LED, 8 digit seven-segment display; red
Number of measuring channels 1
Measuring principle detection of irradiation by means of ionization chambers, analog
and digital signal processing, digital readout on LED
Operating mode continuous operation
Measuring quantities Dose area product, dose area product rate
Linearity deviation < 2.5 %
Offset current (zero drift) < 0.2 pA at 25°C
Settling time max.2 s
Operational readiness 10 s after power up
Nominal useful range dose area prod­uct (DAP) rate
Dose area product (DAP)
Unit: µGym² Measuring range: (0.1 ... 20 000 000) µGym² Accuracy: ± 1 digit Digital resolution: 0.01 µGym²
Dose area product (DAP) rate
Unit: µGym²/s Measuring range: (0.1 ... 15 000) µGym²/s Accuracy: ± 1 digit Digital resolution: 0.01 µGym²/s
(0.01 .... 15 000) µGycm²/s
D651.131.00/05 en 25
Page 26
Technical Specifications
Chamber voltage + 300 VDC ± 10 V
Current limitation of the chamber volt-
< 0.1 mA
age
Power 24 VDC ± 10 %
Power input 3 W max.
Protection class II according to IEC 60601
Electromagnetic compatibility
Suitable for use in an electromagnetic environment according to IEC 60601-1-2
Dimensions (W x H x D) 83 mm x 36 mm x 151 mm
Weight 250 g
Adapter
Power supply T99052
Input (100 ... 240) VDC / 0,33-0,18 A / (47 ... 63) Hz Output 24 VDC / 0,62 A
Power limiter T26060
Input 24 VDC ± 10 % Output 24 VDC ± 10 % / 9 VA
Environmental conditions Transport and storage
Ambient temperature: (-20 ... +60) °C, (-4 … +140) °F Rel. humidity: (10 ... 85) %, no condensation (max. 20 g/m³ absolute humidity) Atmospheric pressure: (600 ...1 200) hPa
Operation
Ambient temperature: (+10 ... +40) °C, (+50 ... +104) °F Rel. humidity: (10 ... 80) %, no condensation (max. 20 g/m³ absolute humidity) Atmospheric pressure: (700 ... 1 060) hPa
Max. operation altitude 3 000 m a.s.l
26
D651.131.00/05 en
Page 27
Accessories and Spare Parts

11 Accessories and Spare Parts

Order No.
T11033 DIAMENTOR E2, DAP meter
T25024-2 Supply cable 24 VDC
T99052 Power supply according to
IEC 60601-1
T26060 Power limiter
T25030-3 RS232 interface cable
L104078 Wall mount for DIAMENTOR E2
on request DIAMENTOR chamber
This list of accessories and spare parts is valid at the time of document accessories and spare parts is available on request.
creation. The current list of

12 Special Adjustments

General information
CAUTION
The adjustments described below must only be carried out by specially trained service techni­cians.
Call up the Service menu by pressing the TEST and PRINT buttons simultaneously. In the Service menu mode the buttons are used for cursor control, to change and confirm settings and to quit a menu.
Electrical calibration
When the Service menu is called up the current setting for the electrical calibration factor CAL will be shown. The factor is adjustable in the range (0.01 ... 9.99).
Change the current digit with PRINT () and move to the next digit with TEST (). Confirm the se­lected setting with RESET (). After confirmation, the menu for setting the test value will be entered automatically. The electrical calibration factor will be saved to the non-volatile memory.
Setting the test value
The menu for setting the test value will be entered automatically when the setting for the electrical calibration factor CAL in the Service menu (see above) has been confirmed with RESET().
The display shows dashes (---) and after a short time the 4-digit test value. With RESET () this value will be confirmed and the measuring mode will be entered automatically. The test value will be saved to the non-volatile memory.
D651.131.00/05 en 27
Page 28
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2 / Special Adjustments
13 Electromagnetic Compati­bility (EMC) according to
WARNING Equipment failure!
IEC 60601-1-2
The DIAMENTOR E2 should not be used in immediate proximity to other devices, or while stacked with other devices. Should this be nec-
WARNING Equipment failure!
The use of cables longer than the specified ones can lead to a deterioration of the electro­magnetic compatibility of the DIAMENTOR E2.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
Radio-Frequency Emis­sions
IEC/CISPR 11
Group 1
The DIAMENTOR E2 uses RF energy only for its inter­nal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
essary, the DIAMENTOR E2 should be ob­served in order to ensure operation according to
prescribed conditions and regulations.
Radio-Frequency Emis­sions
IEC/CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Class B
complies
complies
The DIAMENTOR E2 is suitable for use in all estab­lishments, including domestic establishments and those directly connected to the low-voltage power supply net­work that supplies buildings used for domestic pur­poses.
1
1
The DIAMENTOR E2 fulfils the requirements of IEC/CISPR 11 Class B, anyhow according to its intended use it is not for
operation in domestic establishments.
28
D651.131.00/05 en
Page 29
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below. The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage dips, short inter­ruptions and voltage variations on power supply input lines
IEC 61000-4-11 (
tested using power supply
T99042
)
IEC 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input / output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U (> 95 % dip in U
T
) for
T
0.5 cycle
40 % U (60 % dip in U
T
) for
T
5 cycles
70 % U (30 % dip in U
T
) for
T
25 cycles
< 5 % U (> 95 % dip in U
T
) for 5 s
T
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input / output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U (> 95 % dip in U
T
) for
T
0.5 cycle
40 % U (60 % dip in U
T
) for
T
5 cycles
70 % U (30 % dip in U
T
) for
T
25 cycles
< 5 % U (> 95 % dip in U
T
) for 5 s
T
Electromagnetic Environ­ment - Guidance
Floors should be wood, con­crete or ceramic tile. If floors are covered with synthetic material, the relative humid­ity should be at least 30 %.
Mains power quality should be that of a typical commer­cial or hospital environment.
Mains power quality should be that of a typical commer­cial or hospital environment.
Mains power quality should be that of a typical commer­cial or hospital environment.
If the user of the DIAMENTOR E2 requires continued operation during power mains interruptions, it is recommended that the DIAMENTOR E2 be pow­ered from an uninterruptible power supply or a battery
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical com­mercial or hospital environ­ment.
Note: UT is the a.c. mains voltage prior to application of the test level.
D651.131.00/05 en 29
Page 30
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2 / Special Adjustments
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 Test Level
3 V
rms
150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
Compliance Level
V
= 10 V
1
rms
E
= 3 V/m
1
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment are used no closer to any part of the DIAMENTOR E2, including cables, than the recom­mended separation distance calculated from the equation appropriate for the frequency of the trans­mitter.
Recommended Separation Distance
2
d =
5.3
P
=
V
1
1.2 P
3
5.3
d =
E
=
1.2 P
P
1
80 MHz to 800 MHz
7
=
d =
E
1
2.3 P 800 MHz to 2.5 GHz
P
where P is the power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter­mined by an electromagnetic site survey, than the compliance level in each frequency range.
a)
are less
b)
Interference may occur in the vicinity of equipment marked with the following symbol:
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the
measured field strength exceeds the RF compliance level above, observe the DIAMENTOR E2 to verify normal opera-
tion in each use location. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the DIAMENTOR E2.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re-
flection from structures, objects and people.
2
different test level from product standard IEC 60580 (IEC 61000-4-6) was used
3
calculation with test level of IEC 60601-1-2
30
D651.131.00/05 en
Page 31
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
Recommended separation distances between portable and mobile RF communications equipment and
the DIAMENTOR E2
The DIAMENTOR E2 is intended for use in an electromagnetic environment in which radiated RF distur­bances are controlled. The customer or the user of the DIAMENTOR E2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the DIAMENTOR E2 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum out-
put power of transmit-
Separation distance according to frequency of transmitter
(m)
ter (W)
0.01
0.1
1
10
100
150 kHz to 80 MHz
5.3
d =
P
V
= 1.2 P
1
0.12 0.12 0.23
0.38 0.38 0.73
1.2 1.2 2.3
3.8 3.8 7.3
12 12 23
80 MHz to 800 MHz
5.3
d =
E
P
1
= 1.2 P
800 MHz to 2.5 GHz
7
d =
E
1
= 2.3 P
P
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp­tion and reflection from structures, objects and people.
D651.131.00/05 en 31
Page 32
Appendix: China Electronic Industry Standard Compliance / Special Adjustments

Appendix: China Electronic Industry Standard Compliance

附录:中华人民共和国电子产业标准
This Supplement concerns China Electronic Industry Standard Compliance
此附录涉及中华人民共和国电子产业标准的相关规定
The following product pollution control information is provided according to SJ/T11364-2006 M arking for Control of Pollution caused by Electronic Information Products.
根据SJ/T11364-2006标准规定的《电子信息产品污染控制标识要求》特提供以下有关产品污染控制方面的信息。
1. Explanation of Pollution Control Label
污染控制标识说明
This symbol indicates the product contains hazardous materials in excess of the limits estab-
lished by the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Cer- tain Hazardous Substances in Electronic Information Products. The number in the symbol is the
Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information products will not result in any severe environmental pollution, any bodily injury or damage to any assets. The unit of the period is “Year”.
In order to maintain the declared EFUP, the product shall be operated normally according to the instructions and environmental conditions as defined in the product manual, and periodic mainte­nance schedules specified in Product Maintenance Procedures shall be followed strictly.
Consumables or certain parts may have their own label with an EFUP value less than the prod­uct. Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in accordance with the Product Maintenance Procedures.
This product must not be disposed of as unsorted municipal waste, and must be collected sepa­rately and handled properly after decommissioning.
该标识表明本产品含有的有毒有害物质超出SJ/T11363­2006标准规定的《电子信息产品中有毒有害物质的限量要求》。标识中的数字为本产品的环保使
用期限EFUP,表明在正常使用的条件下,电子信息产品内含的有毒有害物质或元素不会发生 外泄或突变,用户使用该电子信息产品不会对环境造成任何严重污染或对人身、财产造成任何严重 损害的期限。单位为年。
32
为保证所声明的环保使用期限,应按产品手册中所规定的指示和环境条件进行正常使用,并严格遵 守产品维护程序中规定的定期维护和保养日程。
产品中的耗件或某些零部件可能有单独的标识,其环保使用期限有可能产品本身的环保使用 期限。应按产品维护程序期更换这些耗件或零部件,以保证所声明的整个产品的环保使用期限。
本产品在使用寿命结束后不可作为普通生活垃圾处理,必须另行收集并作妥善处理。
D651.131.00/05 en
Page 33
Appendix: China Electronic Industry Standard Compliance
2. Name and Concentration of Hazardous Substances
有毒有害物质的名称及含量
Table 1 - Hazardous substances’ name and concentration.
1 - 有毒有害物质的名称及含量
Component Name
部件名称
Display 显示设备
Others 其它
(Pb)
(铅)
(Hg)
(汞)
X X X X X X
X O X X X X
Hazardous substances’ name
有毒有害物质的名称
(Cd)
(镉)
(Cr(VI))
(六价铬)
(PBB)
(多溴联苯)
(PBDE)
(多溴二苯醚)
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part
is below the limit requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials
used for this part is above the limit requirement in SJ/T11363-2006
Data listed in the table represents best information available at the time of publication.
This table shows where these substances may be found in the supply chain of PTW electronic information
products, as of the date of sale of the enclosed product. Note that some of the component types listed above may or may not be a part of the enclosed product.
O: 表示该有毒有害物质在该部件所有均质材料中的含量均在SJ/T11363-2006 标准规定的限量要求以下。
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出SJ/T11363-2006 标准规定的限量要求。
此表所列数据为直至发布时所能获得的最全面信息。
此表表明自本产品销售之日起在PTW电子信息产品的供应链上何处可能找到以上所述有毒有害物质。请注
意,以上所列的部件类型中的一些可能不属于本产品。
D651.131.00/05 en 33
Page 34

Literature

[1] [Council Directive 93/42/EEC concerning medical devices
(Medical Device Directive - MDD)
[2] Council Directive 97/43/EURATOM on health protection of
individuals against the dangers of ionizing radiation in rela-
tion to medical exposure
[3] IEC 60601-1-2
Medical electrical equipment
Part 1-2: General requirements for basic safety and essential
performance -
Collateral Standard: Electromagnetic compatibility - Re-
quirements and tests
[4] IEC/CISPR 11
Industrial, scientific and medical (ISM) radio-frequency
equipment - Radio disturbance characteristics - Limits and
methods of measurement.
[5] IEC 60950-1
Information technology equipment - Safety
Part 1: General requirements
[6] IEC 60601-1 / ANSI/AAMI ES 60601-1/
CAN/CSA-C22.2.No.60601-1
Medical electrical equipment –
Part 1: General requirements for basic safety and essential
performance
[11] IEC 61000-4-2
Electromagnetic compatibility (EMC) -
Part 4-2: Testing and measurement techniques -
Electrostatic discharge immunity test
[12] IEC 61000-4-3
Electromagnetic compatibility (EMC) -
Part 4-3: Testing and measurement techniques -
Radiated, radio-frequency, electromagnetic field immunity
test
[13] IEC 61000-4-4
Electromagnetic compatibility (EMC) -
Part 4-4: Testing and measurement techniques - Electrical
fast transient/burst immunity test
[14] IEC 61000-4-5
Electromagnetic compatibility (EMC) -
Part 4-5: Testing and measurement techniques -
Surge immunity test
[15] IEC 61000-4-6
Electromagnetic compatibility (EMC) -
Part 4-6: Testing and measurement techniques -
Immunity to conducted disturbances, induced by radio-
frequency fields
[16] IEC 61000-4-8
Electromagnetic compatibility (EMC) -
Part 4-8: Testing and measurement techniques -
Power frequency magnetic field immunity test
[7] IEC 60580
Medical electrical equipment
Dose Area Product Meter
[8] IEC 62353
Medical electrical equipment –
Recurrent test and test after repair of medical electrical
equipment
[9] IEC 61000-3-2
Electromagnetic compatibility (EMC) -
Part 3-2: Limits -
Limits for harmonic current emissions (equipment input cur-
rent 16 A per phase)
[10] IEC 61000-3-3
Electromagnetic compatibility (EMC) -
Part 3-3: Limits -
Limitation of voltage changes, voltage fluctuations and flicker
in public low-voltage supply systems, for equipment with
rated current 16 A per phase and not subject to conditional
connection
[17] IEC 61000-4-11
Electromagnetic compatibility (EMC) -
Part 4-11: Testing and measurement techniques -
Voltage dips, short interruptions and voltage variations im-
munity tests
[18] IEC 60522
Determination of the permanent filtration of X-ray tube as-
semblies
34
D651.131.00/05 en
Page 35

Index

A
Accessories and Spare Parts ........................................ 27
Air Density Correction ................................................... 18
Appendix ....................................................................... 32
C
Calibration ..................................................................... 24
Calibration factor, ionization chamber ........................... 16
Calibration, electrical..................................................... 27
Cleaning........................................................................ 23
Connecting.............................................................. 12, 15
Constancy Check .......................................................... 17
D
Disposal ........................................................................ 24
E
Electromagnetic compatibility.................................. 26, 28
Error Messages............................................................. 19
F
First Time Operation ..................................................... 12
G
General Information ........................................................ 2
I
Intended Use................................................................... 5
Ionization chamber, mounting ....................................... 15
L
Literature .......................................................................34
M
Maintenance .................................................................. 23
Measurement.................................................................17
O
Operating Controls and Indicators ................................. 11
P
PC .................................................................................22
Peripherals ....................................................................22
Print out .........................................................................18
Printer............................................................................22
R
Repair ............................................................................23
RS232 interface.............................................................22
S
Safety Information ...........................................................5
Safety inspections .........................................................24
Setting up ......................................................................12
Setup menu, basic settings............................................16
Special Adjustments ......................................................27
T
Technical Inspection of the Measuring System .............24
Technical Specifications ................................................25
Troubleshooting.............................................................19
D651.131.00/05 en 35
Loading...