Page 2
Instruction Manual DIAMENTOR E [D151.131.0/3]
Security Information
- 2 - Security Information
The instruction manual is an integral part of the device. It should always be kept near the device. Observance of the manual is a prerequisite for proper device performance and correct operation.
Operator safety, specified measuring accuracy and interference-free operation can be guaranteed only if original devices and parts are used. Furthermore only the accessories listed in this manual are approved by PTW-Freiburg and may be used in conjunction with the device, or else accessories whose use has been expressly permitted by PTW-Freiburg. Safe operation and proper device performance are not guaranteed if accessories or consumables from other manufacturers are used.
PTW-Freiburg cannot be held liable for damages resulting from the use of accessories or consumables from other manufacturers.
The warranty period is 1 (one) year and begins on the day of delivery. It is unaffected by repairs covered by the warranty regulations.
Devices on which moisture has developed as a result of temperature changes may not be switched on unless completely dry.'
PTW- Freiburg considers itself responsible for safety, reliability and performance of the device only, if assembly, extension, readjustment, modifi-cation or repair is carried out by PTW­Freiburg or by persons authorized by PTW­Freiburg,and if the device is used in compliance with the technical documentation.
This technical documentation is in conformity with the device specifications and all applicable safety standards valid at printing date. All rights are reserved for devices, circuits, techniques and names appearing in the manual.
No part of the technical documentation may be reproduced without written permission from PTW-Freiburg.
CE-Marking
The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about Medical Devices and fulfills the essential requirements of Annex 1 of this directive. The product is a class IIb device (MDD).
Electromagnetic immunity
The product fully complies with the electromagnetic immunity requirements of standard EN 60601-1-2 `Electromagnetic Compatibility - Medical Electrical Equipment`.
Safety Instructions
Dispose the packaging material according the applicable waste disposal regulations and keep it out of children's reach.
The product is a medical device. Only person with sufficient knowledge and skill may use the device.
Before using the device, the operator must ascertain that it is in correct working order and operating condition.
The operator must be trained in the use of the de- vice.
Set up the device so that the operator has a clear and unobstructed view of the control panel.
The product is not suitable for operation in areas of risk where an explosion hazard may occur. Explosion hazards may be caused by the use of combustible anesthetics, skin-cleansing agents and disinfectants. Furthermore the device is of restricted suitability for application in oxygen-enriched atmospheres. The atmosphere is considered to be oxygenenriched when more than 25% of oxygen or nitrous oxide is added to the ambient air.
Set up the device in a location which affords sufficient ventilation. The air gaps of the device must not be obstructed. Always observe the ambient conditions as indicated in the `Technical Specifications`.
Page 3
Instruction Manual DIAMENTOR E [D151.131.0/3]
Operation in the patient environment
The product represents a medical system. The product may be operated in the patient environment.
Exclusion of operation as device with patient contact
Neither the device nor any peripheral device may have contact to the patient.
Exclusion of operation as controlling instrument
The device is only for use to measure the dose within radiation field. The product must not be used to control radiodiagnostic equipments.
Safety Instruction to avoid electrical hazard
The product is a device of IEC 601 protection class II.
Before connecting the device to the power line, make sure that line voltage and frequency are equivalent to the ratings indicated on the nameplate of the device.
Before putting the device into operation, visually inspect all connection cables for signs of damage. Damaged cables and connectors must be replaced immediately.
- 3 - Security Information
Never touch the patient and open connectors of the device at the same time.
Please observe the manual !
PTW-Freiburg Physikalisch Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 Freiburg GERMANY Phone: ++49-761-49055-0 Fax: ++49-761-49055-70
E-Mail: ptw@ptw.de
This symbol stands for:
Page 4
Instruction Manual DIAMENTOR E [D151.131.0/3]
1 Introduction .............................................................................................................................5
2 Connecting the DIAMENTOR E to the Mains .....................................................................6
3 Description of the DIAMENTOR E.......................................................................................7
4 Description and Mounting of the Ionization Chamber........................................................8
5 Connection and Care of the Ionization Chamber.................................................................9
- 4 - Contents
Contents
Page:
6 Adjusting the DIAMENTOR E ............................................................................................ 10
6.1 Adjustment during the Installation ....................................................................................... 10
6.2 Checking the Calibration after Installation........................................................................... 10
7 Checking the Stability ........................................................................................................... 12
8 Further Information..............................................................................................................13
8.1 Influence of Air Temperature and Humidity ........................................................................13
8.2 Checking the Leakage...........................................................................................................13
8.3 Influence of Field Size and Dose Rate..................................................................................13
8.4 Wall mounting ......................................................................................................................14
9 Time Schedule for regular Checks.......................................................................................15
10 Technical Data .....................................................................................................................16
Page 5
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 5 - Introduction
1 Introduction
It is a well recognized goal to keep the patient's load during diagnostic X-ray examinations as low as possible. Besides the possibilities to adjust the voltage, the current and the filtration (i.e. the doserate), adjustable quantities during the examinations are the beam area and the irradiation time. The task of the diagnostic dosemeter DIAMENTOR E is to measure the product of dose rate, exposure time and area in cGycm². This product is a measure for the patient's load.
The German regulation ("Röntgenverordnung") suggests to measure and to record the dose area product. The DIAMENTOR E simplifies these recordings considerably.
The instrument consists of ionization chamber, electrometer with built-in display, connection cable and separate power supply.
The area of the ionization chamber must always be bigger than the cross section of the beam. The ionization chamber automatically determines the product of the different field sizes and the dose, i.e. the dose area product. The patient's load is recorded during fluoroscopy as well as during radiography. Nonstop alternating operation between fluoroscopy and radiography is possible.
The device is calibrated in such a way that one step of the counter is equivalent to 1 cGycm².
The DIAMENTOR E complies with the rules laid down in the German regulation DIN 6819.
The working principle of the DIAMENTOR E is based on the fact that X-rays inside the special DIAMENTOR ionization chamber produce a well defined amount of charge during the time of exposure. This amount of charge is proportional to the product of dose and beam area and is shown on a LC display with 4 1/2 digits.
Page 6
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 6 - Connecting the DIAMENTOR E to the Mains
2 Connecting the DIAMENTOR E to the Mains
The DIAMENTOR E gets its power via an external power supply with + 15 V DC (see picture 1). For a mains voltage of 115 V AC it is recommended that the PTW power supply T99021 or T99008 is used. For a mains voltage of 230 V AC, it is recommended that the PTW power supply T99009 or T99008 is used.
If the power outlet is different from the mains connector of the power supply, please use an appropriate adapter.
The DIAMENTOR E with the accompanying power supply is a class II instrument. The unit is intended for continuous operation.
Picture 1: Power Cable
Picture 2: DIAMENTOR E, general view
Page 7
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 7 - Description of the DIAMENTOR E
3 Description of the DIAMENTOR E
Picture 3: DIAMENTOR E, connectors
1. Socket for chamber cable
2. Socket for power cable
3. Test button
4. On/Off switch
5. Calibration potentiometer
The calibration potentiometer allows to adjust the response of the DIAMENTOR E in accordance with the ionization chamber in use. See also chapter 6.2.
If the test button (picture 3, 3) is pressed, a charged capacitor is connected to the amplifier in­put. This results in a well defined number of counts. The status of the counter after this proce­dure allows to estimate the response of the amplifier and makes an adaptation of the used chamber possible. A detailed description of this subject is given in chapters 6 and 7 of this manual.
If the high voltage is not sufficient or missing at all, the DIAMENTOR E shows a colon between the second and third and between the third and fourth digit of the display. This also happens immediately after switching on the DIAMENTOR E until the high voltage has reached its final value.
Page 8
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 8 - Description and Mounting of the Ionization Chamber
4 Description and Mounting of the Ionization Chamber
The standard ionization chamber type B (no 57523) has the outer dimensions 17 cm x 18.5 cm x 1.7 cm for a maximum diaphragm opening of 14.5 cm x 14.5 cm.
On special request the chamber type A (no 5754) can be supplied. This chamber has the outer dimensions 21 cm x 18.5 cm x 1.7 cm for a maximum diaphragm opening of 18 cm x 18 cm.
Guide rails of the appropriate size are fitted to the ionization chamber to enable the user to slide the chamber into the guides found a most light beam diaphragms. Both types of chambers are 70 % transparent to visible light and do not impair the use of the light beam diaphragm. The ionization chambers present an additional beam filtration of approximately 0.2 mm Al equivalent according to DIN 6863 part 2. Further information about the ionization chambers can be extracted from the data sheets.
The ionization chamber is connected to the DIAMENTOR E via cable type K3S. This cable has an inner conductor which is connected to the collecting electrode, a first inner shield connected to the guard potential, a second little shield connected to the polarizing voltage of + 500 V and a third outer shield connected to earth. The plugs and sockets of the cable are also connected to the outer shield (earth).
Page 9
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 9 - Connection and Care of the Ionization Chamber
5 Connection and Care of the Ionization Chamber
The connection between the ionization chamber and the DIAMENTOR E is performed by the cable T26004. This is a special cable with cable core and three concentric shieldings. The metal parts of plug and socket are connected to the outer shielding which has mass potential. When installing the cable, it must be sure that plug and socket of the cable do not touch metallic parts of the X-ray machine, to avoid current loops.
Do not clean the surfaces of the ionization chambers with alcohol or other solvent materials. Use only a moistened cloth or a tissue with one of the following disinfectants:
1. 2% Lysoform
2. 2% Korsolin
3. 1% watery solution of Sokrena
4. 5 % Sagrotan 5 5 % Gigasept
Page 10
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 10 - Adjusting the DIAMENTOR E
6 Adjusting the DIAMENTOR E
6.1 Adjustment during the Installation
The DIAMENTOR E is calibrated in the factory. During this calibration the ionization chamber gets a chamber factor KG. The DIAMENTOR E itself gets a characteristic factor QT. The product of these two factors gives the check reading nc:
nc = QT K
This check reading nc must be adjusted by using the calibration potentiometer (see picture 3,
5).
The procedure should be performed as follows:
Connect the chamber to the DIAMENTOR E. Press the test button for approximately 1 s, then release it and read the value on the display. If this displayed value differs from the calculated nc, the calibration potentiometer must be adjusted. Repeat this procedure until the correct value of nc is displayed. The RESET button must be pressed before each test.
G.
(1)
6.2 Checking the Calibration after Installation
During the factory calibration the chamber factor kG is measured with an absorber of 0.35 mm Al equivalent between the ionization chamber and the X-ray focus. This absorber simulates a mean value between undercouch and overcouch X-ray tube installations.
If the DIAMENTOR E is used on an overcouch tube (with no additional absorption between chamber and patient), the factory calibration causes a reading that is approximately 9 % too low. When the ionization chamber is used on an undercouch X-ray tube (where there is approxi-mately 0.5 mm - 0.7 mm Al equivalent between chamber and patient), the reading of the DIAMENTOR E is approximately 3 % too high.
A precise check of the calibration after installation is only possible by comparison with a reference dosemeter. This check should be done annually.
Page 11
Instruction Manual DIAMENTOR E [D151.131.0/3]
Calibration procedure:
a) The couch should be positioned horizontally. b) The diaphragm should be adjusted to give a field size of 10 cm x 10 cm (this can be
measured accurately by exposing a film in this plane).
c) The distance between focus and the measurement plane should be noted (this distance
should be around 100 cm). d) A reference dosemeter should be placed in the reference plane. e) A comparison of the dose area product as indicated by the DIAMENTOR E with that
calculated from the reference dosemeter (dose times area) should then be made at a
range of set kVp values across the normal operating range of the X-ray unit. The com-
parison at a range of field sizes (with fixed kVp and mAs) may also be performed. f) The mean values of these comparisons will enable the user to derive a suitable
calibration factor or to adjust the sensitivity of the meter, so that the indicated and
measured values match. This calibration should be performed at the typically used
doserates.
- 11 - Adjusting the DIAMENTOR E
Page 12
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 12 - Checking the Stability
7 Checking the Stability
The stability of the calibration can be checked with the test button (picture 3, 3) as follows:
Press the test button for approximately 1 second and then release it. By doing so, an amount of charge is fed into the amplifier. This amount of charge is constant within ± 1 %. If the sensitivity of the DIAMENTOR E has not changed, the check reading nc must also be constant.
If it is not (i.e. the check reading deviates by more than ± 5 % from that check reading calcu­lated with the values in the test certificate), a calibration according to chapter 6 should be per­formed.
Page 13
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 13 - Further Information
8 Further Information
8.1 Influence of Air Temperature and Humidity
As long as the DIAMENTOR E is used in the range + 10 °C to + 40 °C, the response changes by less than ± 5 %.
If the DIAMENTOR E is used in environments with high humidity (for example 30 °C, 75 % relative humidity), a film of moisture might cover parts of the instrument or chamber and can cause an increase of the leakage current (see chapter 8.2).
If such a leakage current is observed, the DIAMENTOR E should be placed for about 24 hours in dry warm air with a maximum temperature of 50 °C. It is advisable to carry out such a drying procedure after each transport over longer distances.
8.2 Checking the Leakage
The reading of the DIAMENTOR E may be affected by leakage current. If a not irradiated ionization chamber causes a reading on the display, leakage current most probably is the reason.
A check for leakage is carried out (preferably over night) by leaving the instrument switched on and having the chamber connected but not irradiated. Not more that one digit per hour should appear (see also chapter 9).
8.3 Influence of Field Size and Dose Rate
The smallest current is delivered to the DIAMENTOR E when the chamber is used with the smallest field size and the lowest exposure rate.
The greatest current is delivered to the DIAMENTOR E when the chamber is irradiated with the largest field size and the highest exposure rate. In the range from 100 cGycm²/h to 30,000 cGycm²/s, the response changes by less than ± 5 %.
Page 14
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 14 - Further Information
8.4 Wall mounting
If required, the DIAMENTOR E can be attached to the wall in an easy way. Just screw the delivered screws into the wall and make sure that they have a perpendicular distance of 140 mm. Please make sure that the maximum distance between the wall and the head of the screw is 6 mm. Now you can slide the DIAMENTOR E over the screws through the holes in the rear of the instrument. The DIAMENTOR E is finally fixed by now sliding it a little bit downwards.
Page 15
Instruction Manual DIAMENTOR E [D151.131.0/3]
- 15 - Time Schedule for regular Checks
9 Time Schedule for regular Checks
The proper function of the DIAMENTOR E can be checked with the help of the test button (picture 3, 3) and by a calibration according to chapter 6 of this instruction manual.
It is recommended to perform the check with the test button regularly (especially if there are hints concerning changed or irregular measuring results). It is also recommended to perform the calibration according to chapter 7 every second year.
If leakage occurs that can not be fixed by the drying procedure described in chapter 8.1, a check of the DIAMENTOR E, chamber cable and chamber should be performed: First check the leakage with the DIAMENTOR E alone, i.e. without a cable or a chamber con­nected. If then the results are normal, the problem most probably is either in the cable or in the chamber. If the results are not normal, then most probably there is a problem with the electron­ics of the DIAMENTOR E. In that case the DIAMENTOR E must be send back to PTW­Freiburg or to an authorized technician.
W a r n i n g !
The DIAMENTOR E should only be repaired by PTW-Freiburg or by authorized
service personnel.
Page 16
Instruction Manual
±
- 16 - Technical Data DIAMENTOR E [D151.131.0/3]
10 Technical Data
Device designation and use: DIAMENTOR E type 11002 dose area product meter
according to IEC 580 for measurements in diagnostic radiology
Manufacturer: PTW-Freiburg
Mode of operation: Continuous operation
Electrical Resolution: (0.4 ... 1.4) nC/digit
Radiological Resolution: 1 cGycm²
Maximum counting rate: 30 kHz
Range of dose area product rate: max. 30.000 cGycm²/s
min. 0.03 cGycm²/s
Display: LCD with 4 1/2 digits
Warm-up time: 10 min
Response time: 1 s
Overall uncertainly of the measuring
±
25 %
system according to DIN 6819 Accuracy of the display:
3 % ± 1 digit
Mains: 115 V / 230 V AC + 10 % / - 15 %
±
50 ... 60 Hz
3 Hz, with separate power supply
Power supply 15V DC; 300 mA
External power supply - PTW T 99021
PTW T 99009 (recommended for use with 230 AC) Euro Plug
-
PTW T99008 (recommended for use from 85 V ... 264 V AC
-
(recommended for use with 115 AC) US Plug
wide input range)
Classification according to IEC601-1 Class II
Normal Conditions Temperature: Humidity:
10 0C ... 40 0C 10 % ... 75 % rel. humidity, max. 20 g/m³
Dimensions: 13.7 cm x 19.0 cm x 4.4 cm (LxWxD)
Weights: electrometer 516 g
power supply T99021 and T99009 525 g power supply T99008 370 g
Chamber voltage 475 V
Chamber connector: Type V
Page 17
Instruction Manual
- 17 - Technical Data DIAMENTOR E [D151.131.0/3]
Chamber: PTW type B, type no. 57523
Filtration of the chamber: equivalent to 0.2 mm Al
Transparency of the chamber: > 70 %