in accordance with the Council Directive
93/42/EEC about Medical Devices and fulfills
the essential requirements of Annex I of this
directive. The product is a class IIb device
(MDD).
– The product was granted the type approval for
legal calibration in accordance with § 40 of
the Austrian Measurement and Verification Act
(Maß- und Eichgesetz, MEG) by the Austrian
Federal Office of Metrology and Surveying
(Bundesamt für Eich- und Vermessungswesen, BEV) in Vienna. For detailed information,
refer to section "Technical Specifications".
– Product standards
The product fully complies with the requirements of the following standards:
– IEC 61674
– Electrical Safety
The product fully complies with the requirements of the following standards:
– IEC 61010-1
– Electromagnetic compatibility
The product fully complies with the requirements of the following standards:
– IEC 61326-1
– Operator safety, specified measuring accu-
racy, and interference-free operation can
only be guaranteed if original products and
parts are used. Furthermore, only the accessories listed in this manual are approved by
PTW-Freiburg and only they or other accessories whose use has been expressly permitted by PTW-Freiburg may be used in conjunction with the product. Safe operation and
proper product performance are not guaranteed if accessories or consumables from other
manufacturers are used.
– PTW-Freiburg cannot be held liable for dam-
age resulting from the use of accessories
or consumables from other manufacturers or
when the user ignores the instructions and
information given in this manual.
– The warranty period is 1 (one) year and begins
on the day of delivery. It is unaffected by
repairs covered by the warranty regulations.
– PTW-Freiburg only considers itself responsi-
ble for the safety, reliability, and performance
of the product if the assembly, extension, readjustment, modification, or repair is carried out
by PTW-Freiburg or by persons authorized by
PTW-Freiburg, and if the product is used in
compliance with the technical documentation.
– The user manual is an integral part of the prod-
uct. It should always be kept near the product.
Observance of the manual is a prerequisite for
proper product performance and correct operation.
2D594.131.00/03 en
– In case of any questions concerning the ser-
vice, support, or warranty, please contact your
supplier.
– This manual is in conformity with the product
specifications and all applicable safety standards that are valid as at the printing date. All
rights are reserved for devices, circuits, techniques, software, and names as referred to in
the manual.
Page 3
CONNY II T11007
General Information
– PTW-Freiburg reserves the right to make mod-
ifications.
Please contact PTW or your local representative for the most current information concerning the products of interest.
– No part of the technical documentation may
be reproduced without written permission from
PTW-Freiburg.
– PTW-Freiburg is a registered manufacturer
according to the ElektroG (Elektro- und Elektronikgeräte-Gesetz).
Elektro-Altgeräte-Register (EAR). Registration
number DE15599992
Figure 1:Definition of the patient environment...............................................................................................12
Figure 2:CONNY II, front panel..................................................................................................................... 15
Figure 3:Detector of the CONNY II............................................................................................................... 16
Figure 4:Preferred direction of radiation........................................................................................................ 17
Figure 5:Attaching the detector to the NORMI MAM analog test object....................................................... 18
Figure 6:Battery compartment on the back of the device............................................................................. 28
D594.131.00/03 en5
Page 6
Intended Use
The CONNY II diagnostic dosemeter is a dose
and dose rate meter for constancy tests of X-ray
installations in diagnostic radiology.
CONNY II is suitable for constancy test of medical
X-ray installations for radigraphy, fluoroscopy and
mammogrphy.
Note:In the United States, federal law restricts
this device to sale by or on the order of
a physician.
CONNY II T11007
Intended Use
6D594.131.00/03 en
Page 7
Safety Information
CONNY II T11007
Safety Information
This is the safety alert symbol. It is used
to alert the user to potential hazards.
Obey all safety messages that follow this
symbol to avoid any possible bodily injury
or equipment damage.
All safety messages consist of the following components:
– Safety alert symbol and signal word
– Type of danger
– Source of danger
– Consequence
– Measures to prevent hazards.
Signal Words
DANGER
Indicates an imminent hazard. If not avoided,
the hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result in death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the
hazard can result in minor or moderate injury.
NOTICE
Indicates a potential hazard. If not avoided, the
hazard can result in product damage.
HINT
Provides application tips or other useful information to ensure that you get the most from
your equipment.
D594.131.00/03 en7
Page 8
CONNY II T11007
Safety Information
Safety Information
Strictly observe the following safety information concerning the signal words DANGER and
WARNING. Failure to do so endangers the lives
of the patient, user, and other persons involved.
DANGER
Operation in areas where an explosion may
occur or in oxygen-enriched atmospheres.
Explosion Hazard!
Do not operate the product in areas of risk
where an explosion may occur. Explosion
hazards may be caused, for example, by the
use of combustible anesthetics, skin-cleansing
agents, and disinfectants.
Do not operate the product in oxygen-enriched atmospheres. The atmosphere is
considered to be oxygen-enriched when more
than 25 % of oxygen or nitrous oxide is added
to the ambient air.
DANGER
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
– Switch on products on which moisture con-
densation has developed as a result of temperature changes only after they are completely dry.
– Liquids must not enter the product. If liquids
have entered the product, dry the product
or wait for it to dry completely. The product must be thoroughly inspected by service
personnel before being used again.
– Do not use extension cords.
DANGER
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
– Before using the product, verify that it is in
correct working order and operating condition.
– Before putting the product into operation,
inspect all connection cables, connectors,
housing, and covers for signs of damage.
Replace damaged cables and connectors
immediately.
8D594.131.00/03 en
Page 9
CONNY II T11007
Safety Information
DANGER
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Exclusion of operation in the patient environment:
Do not operate the product nor any peripheral devices in the patient environment (see
Figure 1).
Exclusion of operation with patient contact:
The product is not intended for use in direct
contact with the patient. Ensure that neither the
product nor any peripheral devices come into
contact with the patient.
DANGER
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Use of peripheral devices:
Connect only peripheral devices (PC, printer)
that meet the requirements of the standard
IEC 60950-1 (UL 60950-1) or the standard
IEC 61010-1.
Connect devices to other devices or to parts
of systems only if it has been ascertained that
this connection does not impair the safety of the
patient, the operator, or the environment.
If the device specifications do not contain information regarding the connection of the device
to other equipment, consult, for example, the
manufacturers of the other equipment or other
informed experts to verify that the connection
does not impair the safety of the patient, operator, or environment. Always observe the standard IEC 61010-1.
D594.131.00/03 en9
Page 10
CONNY II T11007
Safety Information
WARNING
Improper handling.
Hazards to Persons!
The product is a medical device and must
only be handled by persons who are trained
in the use of such equipment and are capable
of applying it properly. The operator must be
trained in the use and operation of the product.
WARNING
Magnetic and electrical fields are capable of
interfering with the proper performance of the
product.
Product Failure!
– When operating the product, make sure that
all external devices operated in its vicinity comply with the relevant EMC requirements. X-ray equipment, MRI devices, radio
systems, cell phones, etc. are possible
sources of disturbance as they may emit
higher levels of electromagnetic radiation.
Keep the product away from these devices
and verify its performance before use.
– Do not use cables longer than specified.
This might impair the electromagnetic compatibility characteristics of the product.
WARNING
Radiation is a source of risk, particularly when
the product is operated inappropriately.
Risks to Persons!
Exclusion of operation as a controlling
instrument:
Use the product as a measuring device only. Do
not use measuring results to control an equipment that produces radiation, e.g., a radiodiagnostic equipment or a radiotherapy unit.
WARNING
Application for uses other than those stated in
the user manual.
Bodily Injury! Equipment Damage!
Always apply the product in compliance with its
intended use.
Failure to do so may lead to severe or even fatal
injuries and/or equipment damage and may
void the warranty.
WARNING
Hazards originating from other system components.
Hazards to Persons! Equipment Damage!
– Do not use the product in the immedi-
ate vicinity or placed on top of or below
other equipment. However, if the application requires an arrangement of devices
as described above, the product should be
monitored in order to ensure its proper functioning in the specific arrangement.
– Make sure that the product is used in an
electromagnetic environment as described
in IEC 61326-1.
10D594.131.00/03 en
Observe the safety information provided in the
user manuals of the system components.
Page 11
CONNY II T11007
Safety Information
CAUTION
Use of the product without observing the user
manual.
Bodily Injury! Equipment Damage!
Always use the product in compliance with the
user manual. Otherwise, the provided protection may be impaired.
Use the product only in conjunction with
the products approved by PTW for this purpose and/or the products listed in the section
"Intended Use".
Use the product within the rated range of use
as indicted in the section "Technical Specifications".
Handle the product with care.
NOTICE
To avoid damage to the cables and connectors,
observe the following points:
– Do not kink the cables. Observe the stated
minimum bend radius of the cables. Never
exert pressure on the cables. Do not step
on the cables. Do not roll trolleys over the
cables. Do not pull or twist the cables.
– Disconnect the cables from the devices dur-
ing transport and when the devices are not
in use. Do not let the cables hang down
unrestrained.
– The connectors must always be clean. Do
not allow connectors to lie on the floor.
Always protect connectors when pulling
them through cable conduits.
D594.131.00/03 en11
Page 12
CONNY II T11007
OE YY
i XXX
Safety Information
HINT
Observe the user manuals of all connected
devices!
HINT
Set up the device so that the operator has a
clear, unobstructed view of the indicators and
controls and can operate the controls easily.
Symbols on the Nameplate
SymbolDescription
Refer to the user manual!
The product bears the CE-mark.
The product was granted the
type approval for legal calibration under Austrian law. For
detailed information, refer to section "Technical Specifications".
Manufacturer and date of manufacture
Reference number
Serial number
Separate collection for electrical
and electronic equipment!
(refer also to the section "Disposal of the Product")
Figure 1:Definition of the patient environment
Labeling according to
"Administration on the Control of the Pollution caused by
Electronic Information Products
(ACPEIP)"
(China RoHS)
(refer also to the "Appendix")
For a description of additional symbols, if
applicable, please refer to the equipment
description.
12D594.131.00/03 en
Page 13
Operating Manual - Device Description
Operating Manual
1Device Description
CONNY II T11007
1.1Application
The diagnostic dosemeter CONNY II is suitable
for measurements free in air or at the radiation
entrance side of the test object.
The CONNY II can be used with medical
X-ray equipment for radiography, fluoroscopy,
and mammography (anode-filter combinations:
W/2.5 mm Al at 70 kV and 100 kV as well as with
Mo/30 µm Mo at 30 kV).
The CONNY II can be used for image quality
assurance in diagnostic X-ray departments for
constancy tests according to DIN 6868-3, -4, -7,
and -13. For this purpose, a test object complying with the requirements of the standards shall
be placed in the path of the X-rays, and the dose
occurring at the radiation entrance side of the test
object during radiography shall be measured. At
X-ray systems used for fluoroscopy, the dose rate
at the radiation entrance side of the test object
shall be determined by measuring the dose for a
defined period of time.
When operated within the rated range of use, the
CONNY II fulfills the requirements for diagnostic
dosemeters set forth in IEC 61674.
The CONNY II consists of a battery operated
electrometer and an undetachable semiconductor
detector.
It can be used in combination with test objects of
the NORMI series.
D594.131.00/03 en13
Page 14
CONNY II T11007
Operating Manual - Device Description
1.2Features
The CONNY II consists of a battery operated electrometer with a connected semiconductor detector.
During a measurement, the measuring quantity
chosen before starting the measurement is displayed. After termination of the measurement the
measured dose, the mean dose rate or irradiation
time can be displayed.
Features:
– Battery operated measuring instrument with
voltage monitoring (LOW BAT message)
– Display of the firmware version for approx. 2 s
after switching on the device
– Automatic zeroing after switching on the
device
– After termination of the measurement: display
of the measured dose, mean dose rate or irradiation time by toggling with the MODE key
– Display of a mean dose rate value for irradia-
tion times of more than 8 s
– Measures Gy, Gy/s and s
– Over range message (OL)
– Auto-Shutoff after 5 min without irradiation or
key operation
– Auto start and Auto reset functions
– Toggling between the 30 kV and 70/100 kV
voltage range possible
– Two measuring ranges (LOW and HIGH);
manual selection of a measuring range
– Display of the measuring quantity selected
with MODE during the measurement
14D594.131.00/03 en
Page 15
CONNY II T11007
Operating Manual - Device Description
1.3Operating Controls and Indicators
Figure 2:CONNY II, front panel
Operating controls and indicatorsSymbol
1
Display of the measuring value and unit
2
Display of selected kV range, measuring mode and measuring range
3
Key for selection of the kV range
4
Key for selection of the measuring range
5
Key for selection of the measuring mode
6
Key to turn on or to switch off the device
D594.131.00/03 en15
---
---
kV
RANGE
MODE
ON / OFF
Page 16
Figure 3:Detector of the CONNY II
CONNY II T11007
Operating Manual - Putting into Operation
2Putting into Operation
The CONNY II is battery operated and features a
voltage monitoring function. If the battery voltage
decreases below a lower limiting value the device
switches off automatically. The battery lifetime for
continuous operation is approximately 18 hours.
A battery symbol appears behind the measuring
value approximately 3 hours before the auto shutoff.
HINT
When rechargeable batteries are used the
LOW BAT message is displayed earlier.
Switching on
Turn the CONNY II on with the ON/OFF switch.
•
➔ During the display of the firmware version,
all measuring ranges are zeroed automatically.
➔ After zeroing, the CONNY II is ready for
operation.
➔ Default values:
Radiation quality: 70/100 kV
Measuring range: HIGH
Measuring mode: DOSE
Start a measurement by irradiating the detector
(refer to 3 "Measuring ").
The CONNY II features an auto shutoff function
which switches off the device after 5 min without
irradiation or key operation.
16D594.131.00/03 en
Page 17
3Measuring
CONNY II T11007
Operating Manual - Measuring
3.1General Information
Default settings
After power-up, the following default values are
preset in the CONNY II:
– Radiation quality: 70/100 kV
– Measuring range: HIGH
– Measuring mode: DOSE
Before starting a measurement, the defaults can
be adapted to the measuring conditions with the
kV, RANGE, and MODE keys.
– Fluoroscopy:
measuring range LOW (5 µGy/s ... 3 mGy/s)
– Radiograpy:
measuring range HIGH (0.5 ... 300) mGy/s
Detector
The detector of the CONNY II can be attached to
the test object by means of the supplied reclosable fastener. The detector surface must be irradiated perpendicularly (Figure 4), and the entire
detector must be located in the radiation beam.
Figure 4:Preferred direction of radiation
Measuring quantities
The CONNY II can be used to measure the following measuring quantities:
– air kerma rate (dose rate mode)
– air kerma (dose mode)
– time (irradiation or exposure time)
Measuring ranges, refer to 5 "Technical Specifi-
cations"
When fastening the detector to the test object,
take care not to cover measuring areas or test
structures with the detector cable.
D594.131.00/03 en17
Page 18
Figure 5:Attaching the detector to the NORMI
MAM analog test object
CONNY II T11007
Operating Manual - Measuring
3.2Performing a
Measurement
Place the CONNY II on the patient couch
•
outside the radiation beam and fold out the
bracket at the back of the device if necessary.
Before the measurement, attach the detector
•
of the CONNY II to the test object. The detector must be irradiated perpendicularly, and it
must be completely located in the radiation
beam; the reference point is at the center of
the entrance window.
Use the ON/OFF switch to turn the CONNY II
•
on; at the same time, all measuring ranges will
be zeroed automatically.
Adapt the preset parameters to the measur-
•
ing conditions with the kV, RANGE, and MODE
keys.
– Measuring range for fluoroscopy:
LOW (dose rates in the range from
5 µGy/s to 3 mGy/s)
– Measuring range for radiography:
HIGH (dose rates in the range from
0.5 to 300 mGy/s)
Start a measurement by exposing the detector
•
to radiation. The CONNY II features an autostart function, which means that a measurement will start automatically when the detector
detects radiation. The zero adjustment of the
last switch-on will be taken into account.
18D594.131.00/03 en
Page 19
CONNY II T11007
Operating Manual - Measuring
Measurement readout
While a measurement is in progress, the keyboard
of the device is locked and the parameter set with
the MODE key is displayed. For the duration of the
measurement, the letter M is visible in the upper
left of the display.
HINT
About dose rate measurements:
If the selected measuring quantity is dose rate,
the mean dose rate will be displayed first after
the radiation is switched off. The first and the
last 3 seconds of the measuring interval are not
included in the calculation.
The minimum measuring time is 8 seconds
(effective measuring time of 2 s + waiting period
of 2 x 3 s that is not taken into account).
The value is averaged over a maximum of
20 seconds (t
the minimum measuring time is reached, the
flashing label "Time" will be displayed in addition to the dose rate. If a measurement is terminated during this period, the readout is reset to
0.00 mGy/s. If the measurement takes longer,
the message "Time" will be displayed permanently. After the end of the measurement, the
mean dose rate will be displayed with the label
"mean".
The measured value should always be displayed
with a minimum resolution of 1 % (100 digit). If
this resolution is not available in the HIGH measuring range, switch to the LOW measuring range.
If measurements in the LOW measuring range
exceed the maximum dose rate, a condition indicated by the message OL, switch to the HIGH
measuring range.
–23 s to t
end
– 3 s). Before
end
Interrupting a measurement
If a measurement is interrupted for less than
3 seconds, it will continue after the interruption. If
the interruption lasts longer than 3 seconds, the
measurement will be canceled and the current
measured value will be displayed. If a measuring
signal becomes available again afterwards, the
readout will be reset to zero and a new measurement will be started (auto-reset). The values of the
previous measurement are no longer available in
this case!
Terminating a measurement and reading the
measured values
After the measurement, the measured parameters dose, dose rate, and time can be displayed by
pressing the MODE key. The last parameter set
with the MODE key will be displayed during the
next measurement if the device remains switched
on.
The CONNY II features an auto-shutoff function,
which means that the device switches off if no
radiation is detected and no key is pressed for a
period of 5 minutes.
HINT
All measured values will be lost when the
CONNY II is switched off. Therefore, be sure to
read off the measured values within 5 minutes
after the measurement!
D594.131.00/03 en19
Page 20
4Error Messages
CONNY II T11007
Operating Manual - Error Messages
Message
OLinstead of the measuring valuedose measurement:
L-Errinstead of the measuring value
T-Errinstead of the measuring value
Position on the displayRemarks
over range of the (dose) range (9.999 Gy),
max. dose rate in the measuring range exceeded
dose rate measurement:
during over range of the dose rate range
time measurement:
over range of the (time) range (999.9 s); measurement of
dose and dose rate still possible
irradiation time measurement not possible because signal
(only irradiation time measurement)
(only irradiation time measurement)
is too low, an estimated value for the irradiation time is dis-
played
irradiation time measurement not possible (no rectangular
pulse)
E1upper left partzeroing not possible (e.g. irradiation during switching on)
E2upper left partcalibration factor exceeds the limit
E3upper left partcheck sum error;
send the CONNY II back to PTW-Freiburg for repair
20D594.131.00/03 en
Page 21
CONNY II T11007
OE 06
i 110
Technical Manual
Technical Manual
5Technical Specifications
Only values with specified tolerances or limits are guaranteed. Values without tolerances are for informational purposes only.
Device designationDiagnostic dosemeter CONNY II T11007
ManufacturerPTW-Freiburg
Intended useDose and dose rate measurement for constancy tests of X-ray
installations for radiography
Type approval certificate in
Austria
Mode of operationcontinuous operation
DisplayLCD 4 digits
Detectorsemiconductor detector with entrance window 16 mm x 9 mm, ref-
asDosemeter for constancy checks in diagnostic
X-ray applications
atAustrian Federal Office of Metrology and Sur-
veying (Bundesamt für Eich- und Vermessungswesen, BEV)
according to§ 40 des Maß- und Eichgesetzes (MEG),
BGBl. Nr. 152/1950, in der geltenden Fassung
approval
sign
issued toPTW-Freiburg
erence point in 5.7 mm depth under the center of the window
Nominal useful energy range in
mode
70/100 kV
30 kV
Energy dependence of the display within the nominal useful
range
D594.131.00/03 en21
70/100 kV ± 10 % (W anode 2.5 mm Al)
30 kV ± 10 % (Mo anode 30 µm Mo)
± 5 % with variation of tube voltage by ± 10 %
Page 22
CONNY II T11007
Technical Manual
Device designationDiagnostic dosemeter CONNY II T11007
Measuring quantitiesair kerma in Gy,
air kerma rate in Gy/s,
irradiation time [s]
Measuring ranges and resolution
Range LOW
70/100 kV
Range LOW
30 kV
Range HIGH
70/100 kV
Range HIGH
30 kV
Dose measurement:
resolution
lower limit
upper limit
20 nGy
2 µGy
9.999 Gy
50 nGy
5 µGy
1)
9.999 Gy
1)
2 µGy
200 µGy
9.999 Gy
5 µGy
500 µGy
1)
9.999 Gy
Dose rate measurement:
resolution
lower limit
upper limit (typ.)
50 nGy/s
5 µGy/s
3 mGy/s
100 nGy/s
10 µGy/s
5 mGy/s
5 µGy/s
500 µGy/s
300 mGy/s
10 µGy/s
1 mGy/s
500 mGy/s
Time measurement
resolution
lower limit
upper limit
10 ms
1 s
999.9 s
10 ms
1 s
1)
999.9 s
1)
10 ms
1 s
999.9 s
10 ms
1 s
1)
999.9 s
1)
Leakage≤ 2 Digit in all dose rate ranges (for at least 1 min acc. to IEC
61674).
The zero point is measured and compensated each time the
device is switched on. The auto-reset/auto-start function ensures
that leakage will not occur after a dose measurement.
1)
Repeatability≤ 1 %
Linearity of the dose measure-
± 2 %
ment
Long-term stabilitymax. 2 % p. A.
Nominal useful ranges
Temperature
Humidity
Atmospheric pressure
Maximum operating altitude
Group 1The product uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
Class BThe product is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
24D594.131.00/03 en
Page 25
CONNY II T11007
Service Manual - Cleaning
Service Manual
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
Ensure that the service tasks described below
are only performed by specialized staff.
In addition, ensure that all other service tasks
are only performed by PTW-Freiburg or by a
person authorized by PTW-Freiburg.
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
Before carrying out any service tasks on the
CONNY II, de-energize the device.
Do not open the housing of the CONNY II.
6Cleaning
General Information on Cleaning
DANGER
Liquids have entered the product.
Shock Hazard! Erroneous Measurements!
Have the product inspected by service personnel before putting it into service again.
NOTICE
To avoid surface damage, do not use disinfectants on a phenol base or peroxide compounds
to disinfect the surface.
The product cannot be sterilized or completely
disinfected.
Clean the product simply by wiping it down with
a dry or slightly moistened cloth. Use water or a
mild soap solution.
Alternatively, you can use a common disinfection
agent for medical instruments which is specified
in the list of disinfectants of your national hygiene
institute (e.g., VAH in Germany). If your country
does not provide such a list, we recommend either
a solution on an aldehyde base or a solution with
a quaternary ammonium compound.
D594.131.00/03 en25
Page 26
CONNY II T11007
Service Manual - Preventive Maintenance
Surfaces and Connection Cable
Clean the surfaces and the connection cable as
follows:
Only wipe the product down. Liquids must
•
not be allowed to enter the product. Do not
immerse the product in water or other liquids.
Do not apply spray to the product.
7Preventive Maintenance
Below, you will find a list of the maintenance measures and inspections necessary for the product.
Check before each use
Before each use, visually inspect the product
•
and cables for signs of mechanical damage.
If damage or malfunctions are identified, the
•
product must be repaired before it is used
again.
Repair
Repairs may only be carried out by PTW-Freiburg
or by persons/companies authorized by PTWFreiburg.
26D594.131.00/03 en
Page 27
CONNY II T11007
Service Manual - Preventive Maintenance
Safety Inspections
Only regularly inspected products are fail-safe. To
preserve the functional security and operational
safety, a safety inspection is to be executed at
regular intervals according to the national regulations.
These inspections must only be performed by
independent persons with adequate training and
experience.
It is recommended to execute the safety inspections every 2 years.
Technical Inspection of the Measuring
System
The product must be inspected at regular intervals according to the national regulations. It is recommended to perform technical inspections of the
measuring system every 5 years at PTW-Freiburg
or a qualified calibration laboratory.
Calibration
The exact calibration factor and supplementary
data are included in the calibration certificate.
It is recommended to have the product calibrated
every 2 years at PTW-Freiburg or by specially
trained staff on site.
Inspections must also be performed after repairs
or after each influence that may have changed the
behavior of the product.
D594.131.00/03 en27
Page 28
8Inserting Batteries
DANGER
Improper handling of batteries.
Explosion Hazard!
CONNY II T11007
Service Manual - Inserting Batteries
Remove the lid from the battery compartment
•
on the underside of the device.
Remove the four batteries.
•
Do not throw the batteries into a fire.
WARNING
Improper handling of batteries.
Bodily Injury!
If electrolyte leaks from the battery, avoid contact with skin or clothes. Immediately clean skin
or clothes with water if they have come in contact with electrolyte. In case of eye contact,
immediately rinse eyes with water and consult
doctor.
Do not open batteries.
WARNING
Improper handling of batteries.
Risk of Fire or Skin Burns!
Never short-circuit batteries.
Insert new batteries into the device. Observe
•
the correct battery polarity Figure 6. Use only
AA size batteries (refer to section 5 "Technical
Specifications").
Reinstall the lid of the battery compartment.
•
The batteries must be disposed of in compli-
•
ance with the applicable regulations.
NOTICE
Observe the following points to avoid equipment damage:
– Do not open batteries.
– Never expose batteries to water.
– Store and use the batteries at the ambient
conditions (temperature, relative humidity)
specified for the device.
– Remove discharged batteries as soon as
possible because leaking electrolyte may
damage the device.
Proceed as follows to replace the batteries:
Switch off the device.
•
28D594.131.00/03 en
Figure 6:Battery compartment on the back of
the device
The CONNY II can also be used with AA size
rechargeable batteries. The CONNY II does not
provide a charging function.
The safety information set forth above also
applies to the use of rechargeable batteries.
Page 29
CONNY II T11007
Service Manual - Disposal of the Product
9Disposal of the Product
The typical lifetime of the product is approx.
10 years.
At the end of the product lifetime, the components
of the product must be disposed of in compliance
with the applicable waste control regulations. The
different materials must be separated and recycled as appropriate.
The electronic components must be recycled
according to the local regulations.
The product does not contain hazardous materials.
The rechargeable batteries must be disposed of
in compliance with the applicable regulations.
D594.131.00/03 en29
Page 30
Literature
CONNY II T11007
Literature
[1] Council Directive 93/42/EEC concerning medical devices
(Medical Device Directive - MDD)
[2] Council Directive 97/43/EURATOM on health protection
of individuals against the dangers of ionizing radiation in
relation to medical exposure
[3] IEC 61674
Medical electrical equipment -
Dosimeters with ionization chambers and/or semi-con-
ductor detectors as used in X-ray diagnostic imaging
[4] DIN 6868-3
Image quality assurance in diagnostic X-ray departments
-
Part 3: Tests for constancy in direct radiography
[5] DIN 6868-4
Image quality assurance in diagnostic X-ray departments
-
Part 4: Constancy testing of medical X-ray equipment for
fluoroscopy
[10] IEC/CISPR 11
Industrial, scientific and medical equipment -
Radio-frequency disturbance characteristics - Limits and
methods of measurement
[11] ISTA 2A
ISTA 2 Series
Partial Simulation Performance Test Procedure -
Part 2A: Packaged-Products 150 lb (68 kg) or less
[6] DIN 6868-7
Image quality assurance in diagnostic X-ray departments
-
Part 7: Constancy testing of mammographic Xray equip-
ment
[7] DIN 6868-13
Image quality assurance in diagnostic X-ray departments
-
Part 13: RöV constancy test of projection radiography
systems with digital image receptors
[8] IEC 61010-1
Safety requirements for electrical equipment for mea-
surement, control and laboratory use -
Part 1: General requirement
[9] IEC 61326-1
Electrical equipment for measurement, control and labo-