PTW 47001, 47002, 47003 User Manual

Page 1
User Manual
PTW-UNITEST
(Electrical Check Device)
Type 47001, Type 47002 and Type 47003
D235.131.0/4 2005-12 Sa
Page 2

General Information

PTW-UNITEST
(Electrical Check Device)
General Information
The product bears the CE-mark "CE-0124" in
accordance with the Council Directive 93/42/EEC about Medical Devices and fulfills the essential requirements of Annex 1 of this directive. The product is a class IIb device (MDD).
The user manual is an integral part of the
product. It should always be kept near the product. Observance of the manual is a pre­requisite for proper product performance and correct operation.
Operator safety, specified measuring accuracy
and interference-free operation can be guar­anteed only if original products and parts are used. Furthermore only the accessories listed in this manual are approved by PTW-Freiburg and may be used in conjunction with the prod­uct, or else accessories whose use has been expressly permitted by PTW-Freiburg. Safe operation and proper product performance are not guaranteed if accessories or consumables from other manufacturers are used.
This manual is in conformity with the product
specifications and all applicable safety stan­dards valid at printing date. All rights are re­served for devices, circuits, techniques, soft­ware and names referred to in the manual.
Electromagnetic immunity: The product fully
complies with the electromagnetic immunity requirements of standard IEC 60601-1-2 `Electromagnetic Compatibility - Medical Elec­trical Equipment`.
No part of the technical documentation may be
reproduced without written permission from PTW-Freiburg.
PTW-Freiburg works in strict accordance with
a quality management system which is con­tinuously updated according to national and international standards.
PTW-Freiburg cannot be held liable for dam-
ages resulting from the use of accessories, consumables from other manufacturers or when the user ignores the instructions and in­formation given in this manual.
The warranty period is 1 (one) year and begins
on the day of delivery. It is unaffected by repairs covered by the war­ranty regulations.
PTW-Freiburg considers itself responsible for
safety, reliability and performance of the prod­uct only, if assembly, extension, readjustment, modification or repair is carried out by PTW-Freiburg or by persons authorized by PTW-Freiburg, and if the product is used in compliance with the technical documentation.
PTW-FREIBURG
Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de
Page 3
PTW-UNITEST
(Electrical Check Device)
Contents
Contents
General Information 2
1 Intended Use 4
2 For Your Safety 4
3 Description of the Instrument 7
4 Description of Function 7
5 Measurements with the UNITEST 8
5.1 Hints for the use of the UNITEST 8
5.2 Preparing the measurements 8
5.3 Measurements 8
6 Evaluation of the measuring results 11
6.1 Consistency check of the response in the different measuring ranges 11
6.2 Electrical constancy checks 11
7 Cleaning, Maintenance, Disposal 12
7.1 Cleaning 12
7.2 Preventive Maintenance 12
7.3 Disposal of the Product 12
8 Technical Specifications 13
Literature 14
D235.131.0/4 3
Page 4
PTW-UNITEST
(Electrical Check Device)

Intended Use

1 Intended Use
UNITEST is an electrical check device for the sur­veillance of the PTW dosemeters listed in chap­ter 5. No mains connection or battery is needed to operate UNITEST; it is simply connected to the dosemeter instead of an ionization chamber. Using UNITEST the following checks can be done:
Check of the consistency of the response in all
doserate and dose measuring ranges.
Electrical constancy check of the different ampli-
fier ranges and of the internal high voltage sup­ply of the dosemeter.
It must be observed that the connection systems of UNITEST and dosemeter are compatible. Use of adapter cables is not permitted.
When the UNITEST is not in use, it should be stored in a location protected against humidity. The connector should be protected with a dust protec­tion cap.
2 For Your Safety
The following safety statements are divided into
DANGER, WARNING, CAUTION and NOTE.
Definitions
DANGER
Indicates an imminent hazard. If not avoided, the hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the hazard could result in minor injury or product / property damage.
NOTE
Provides useful information to assure that you get the most from your equipment.
Page 5
PTW-UNITEST
(Electrical Check Device)
For Your Safety
Safety Information
DANGER
Explosion Hazard – The product is not suitable for operation in areas of risk where an explo- sion hazard may occur. Explosion hazards may be caused by the use of combustible an­aesthetics, skin-cleansing agents and dis­infectants.
Furthermore the product is not suitable for ap­plication in oxygen-enriched atmospheres. The atmosphere is considered to be oxygen­enriched when more than 25 % of oxygen or ni­trous oxide is added to the ambient air.
WARNING
Shock Hazard – Strictly observe the following warnings. Failure to do so may endanger the lives of the patient, the user and other persons involved.
Before using the product, the operator must
ascertain that it is in correct working order and operating condition.
Before putting the device into operation,
visually inspect all connection cables for signs of damage. Damaged cables and con­nectors must be replaced immediately.
Devices on which moisture condensation
has developed as a result of temperature changes must not be switched on unless completely dry.
Liquids must not enter the device. If liquids
have entered the device, it must be thor­oughly inspected before being used again.
Exclusion of operation in the patient envi­ronment:
Neither the product nor any peripheral device may be operated within the patient environment (see Figure 1).
Exclusion of operation as device with pa­tient contact:
The device is not for use in direct contact with the patient. Neither UNITEST nor any periph­eral device may have contact to the patient.
Never touch the patient and open connectors of the device at the same time.
WARNING
Suffocation Hazard – Dispose of the packag­ing material, observing the applicable waste­control regulations. Keep the packaging mate­rial out of children's reach.
Patient Hazard – The product is a medical electrical device and must only be handled by persons who are trained in the use of such equipment and are capable of applying it prop­erly. The operator must be trained in the use of the device.
Risk of Poisoning – Chemicals required for application or maintenance of the device, for in­stance, must under all circumstances be stored, prepared, and kept at hand in their specific containers. Failure to observe this instruction may result in severe consequences for the pa­tient.
CAUTION
Equipment damage - Always observe the am­bient conditions as indicated in the `Technical Specifications`.
NOTE
Please observe the user manuals of all con­nected devices!
Set up the device so that the operator has a clear, unobstructed view of the control panel.
D235.131.0/4 5
Page 6
NOTE
PTW-UNITEST
(Electrical Check Device)
For Your Safety
This symbol means: Refer to user manual.
This symbol means: Separate collection for electrical and electronic equipment! (refer also to section Disposal of the Product)
This symbol means: The product bears the CE-mark.
Figure 1: Definition of patient environment
Page 7
PTW-UNITEST
(Electrical Check Device)

Description of the Instrument

3 Description of the Instru-
ment
The UNITEST has 2 control elements (see Fig­ure 2):
Range switch
left selector switch to adapt to the measuring ranges of the dosemeter (see Table 1)
Start switch
right selector switch for the injection of a charge into the dosemeter.
4 Description of Function
The electrical check device UNITEST is working with passive components. UNITEST does not need a mains connection or a battery; it is using the high voltage supplied by the dosemeter.
By putting the right selector switch of UNITEST into START position, a capacitor on UNITEST is charged and the charge is injected into the meas­uring input of the dosemeter.
In accordance with the measuring range to be tested, the test current must be adjusted using the left selector switch of the UNITEST. This is actuat­ing a precision voltage divider in the UNITEST.
Figure 2: UNITEST front panel with the "RANGE" switch and the "START" switch
D235.131.0/4 7
Page 8
PTW-UNITEST
(Electrical Check Device)

Measurements with the UNITEST

5 Measurements with the UNITEST
5.1 Hints for the use of UNITEST
Always connect the UNITEST directly to the
dosemeter. Do not use extension or adapter ca­bles.
Do not move the UNITEST or the connection
cable during measurements.
Only use the UNITEST range selector switch
when the right UNITEST selector switch is in RESET position.
Check values are independent of the doseme-
ter's polarity setting. Dosemeter polarity should only be changed in the least sensitive measuring range and with the UNITEST disconnected.
5.2 Preparing the measurements
1. Switch on the dosemeter and wait for the dura­tion of the warm-up time.
5.3 Measurements
1. Adjust the UNITEST range according to Table 1 (e.g. 0.001 when using it with a UNIDOS).
2. Start at the dosemeter a charge measurement using the STA sition the right UNITEST selector switch to START. Do not move UNITEST and connection cable.
3. After the test time given in Table 1, read the measuring value and note it down.
4. Set the right UNITEST selector switch to RESET position. The dosemeter should be in the next higher measuring range and in RESET state. When using UNITEST with UNIDOS in the measuring mode "Charge" the RESET state of UNIDOS must be selected first before the UNITEST selector switch is set to RESET posi­tion.
or the MEAS button, then po-
2. Set the dosemeter to electrical units.
3. For dosemeters with variable high voltage adjust a chamber voltage of 400 V; polarity is of no im­portance. For measurements with the TANDEM adjust a chamber voltage of 250 V.
4. Set the dosemeter to charge measurement and measuring range Low. (For measurements with the TANDEM, also see the hints below.)
5. Set the right UNITEST selector switch to RESET position.
6. Connect the UNITEST to the dosemeter. Do not use extension cables.
7. Wait for about one minute.
8. Perform a zero adjustment.
NOTE
If a measurement must be repeated in the same position, wait for about 20 s after switching back to RESET position before starting a new meas­urement.
5. Repeat the steps 1 to 4 for all settings in Ta­ble 1.
CAUTION
During switching between the UNITEST ranges the right UNITEST selector switch must be in the RESET position.
Page 9
PTW-UNITEST
(Electrical Check Device)
Measurements with the UNITEST
Dosemeter setting UNITEST range
Standard range of
Test time T dosemeter display values
UNIDOS
webline
Charge Low 0,001 (84 - 92) pC 10 s
Medium 0,1 (8,4 - 9,2) nC 10 s
High 1 (84 - 92) nC 20 s
UNIDOS
Charge Low 0,001 (84 - 92) pC 10 s
High 0,1 (8,4 - 9,2) nC 20 s
Integrated Current Low 0,001 (84 - 92) pC 10 s
Medium 0,1 (8,4 - 9,2) nC 10 s
High 1 (84 - 92) nC 20 s
UNIDOS E
Charge Low 0,001 (84 - 92) pC 10 s
Medium 0,1 (8,4 - 9,2) nC 10 s
High 1 (84 - 92) nC 20 s
MULTIDOS (Channel 1 and 2)
Charge Low 0,01 (840 - 920) pC 10 s
High 1 (84 - 92) nC 20 s
TANDEM
Charge Low Current
Charge Medium Current
Charge High Current
0,01
0,1
1 (52 - 58) nC
(520 - 580) pC (52 - 58) pA
(5,2 bis 5,8) nC (520 - 580) pA
(2,6 - 2,9) nA
10 s
10 s
20 s
Table 1: The necessary settings of dosemeter and UNITEST and typical dosemeter display values after end of
the test time T
D235.131.0/4 9
Page 10
NOTE
for measurements with the TANDEM:
1. If the TANDEM is used in combination with TANSOFT the measuring procedure is the same as described in 5.2 and 5.3 because the graphical user interface of TANSOFT is similar to the user interface of a dosemeter. TANSOFT allows charge measurements by integrating the current measurement values of the TANDEM. Therefore the charge val­ues in the 3rd column of Table 1 should be considered as set values.
2. If the TANDEM is used in combination with other software which does not allow to inte­grate current values, the measuring time of the TANDEM has to be set to the time given in the 4th column of Table 1 (TIME telegram) first. Subsequently, a measurement with a fixed measuring time has to be started (MT telegram). Immediately afterwards the UNITEST selector switch on the right has to be set to START position. After the measur­ing time has elapsed the TANDEM gener­ates a telegram which contains a current measuring value. This value should match the appropriate current set value in the 3 column of Table 1. By multiplying the current value by the measuring time value from the 4th column of Table 1 a charge value can be calculated.
PTW-UNITEST
(Electrical Check Device)
Measurements with the UNITEST
rd
Page 11
PTW-UNITEST
(Electrical Check Device)

Evaluation of the measuring results

6 Evaluation of the measuring results
6.1 Consistency check of the re­sponse in the different meas­uring ranges
The dosemeter display values noted down must
be divided by the corresponding UNITEST ranges according to Table 1, for example:
84 pC / 0.001 = 84 nC
8.4 nC / 0.1 = 84 nC
Ideally, the standardized charge displays thus determined are identical.
Compare the standardized charge displays.
They must not differ by more than 1 %. Typical deviations are 0.2 % to 0.3 %.
6.2 Electrical constancy checks
The measuring results according to Table 1 can also be used for an electrical constancy check of the amplifiers and the internal high voltage supply of the dosemeter.
For a successful constancy check the following conditions must be fulfilled.
The dosemeter display must be within the stan-
dard range according to Table 1.
The dosemeter display must not differ more than
1 % from earlier measurements.
If these conditions are not fulfilled, this can be caused by the following reasons:
dosemeter is defective
dosemeter high voltage has changed
UNITEST is defective
If it is intended to use the UNITEST (and, where appropriate, together with a radioactive check de­vice) for a documented long-time constancy check of a dosemeter, the user should determine refer­ence values immediately after receiving the instru­ment. These values are not documented by the manufacturer since they vary with each individual dosemeter.
D235.131.0/4 11
Page 12
PTW-UNITEST
(Electrical Check Device)

Cleaning, Maintenance, Disposal

7 Cleaning, Maintenance, Disposal
7.1 Cleaning
Surface
Rub the device down with a moist cloth; no liquid must enter the device. Do not spray the device or the connecting plug.
WARNING
Equipment damage - Do not use disinfectants on a phenol base or peroxide compounds to disinfect the surface.
NOTE
Due to its surface geometry, the UNITEST can­not be thoroughly disinfected. Furthermore, the UNITEST cannot be sterilized.
Clean the cables as follows:
disconnect the cables from the device, pulling on
the plug, not on the cable
7.2 Preventive Maintenance
Before each use, visually inspect the device and the cables for signs of mechanical damage.
If damages or malfunctions are identified, the de­vice must be repaired before it is used again.
The UNITEST does not require any other, regular maintenance measures.
7.3 Disposal of the Product
At the end of the product life the UNITEST compo­nents must be disposed of in compliance with the applicable waste control regulations.
The different materials must be separated and re­cycled as appropriate.
The electronic components must be recycled ac­cording to local regulations.
The cost for an eventual return at the end of the product life time is to be borne by the customer.
for cleaning, rub the cables down with a cloth
moistened with soap water
never immerse the cables in liquid!
Page 13
PTW-UNITEST
(Electrical Check Device)

Technical Specifications

8 Technical Specifications
Only values for which tolerances or limits are specified are guaranteed. Values for which no tolerances are specified are for information only.
Product Electrical check device PTW-UNITEST
type 47001 (M connection system), type 47002 (TNC connection system) and type 47003 (BNT connection system)
Manufacturer PTW-Freiburg
Usage
Charge capacity 220 pF ± 2 % Voltage divider (1 : 10 : 100 : 1000) ± 0,1 %
Outer dimensions 55 mm x 180 mm x 60 mm (H x L x W)
Weight approx. 500 g
Checking of PTW dosemeters UNIDOS MULTIDOS, TANDEM
webline
, UNIDOS, UNIDOS E,
Environmental conditions for operation
Ambient temperature
Rel. humidity
Atmospheric pressure
Environmental conditions for transport and storage
Ambient temperature
Rel. humidity
Atmospheric pressure
(10 ... 40) °C
(10 ... 75) %, no condensation (max. 20 g/m³ absolute humidity)
(600 ...1200) hPa
(-20 ... +60) °C
(10... 85) %, no condensation (max. 20 g/m³ absolute humidity)
(600 ...1200) hPa
D235.131.0/4 13
Page 14
(Electrical Check Device)

Literature

[1] Council Directive 93/42/EEC concerning medical devices
(Medical Device Directive - MDD)
[2] Council Directive 97/43/EURATOM on health protection of
individuals against the dangers of ionizing radiation in rela-
tion to medical exposure
[3] IEC 60601-1-2
Medical electrical equipment
Part 1: General requirements for safety;
nd
Collateral Standard: Electromagnetic compatibility - Re-
2
quirements and tests
[4] IEC 60601-1
Medical electrical equipment
Part 1: General requirements for safety
PTW-UNITEST
Literature
[5] IEC 60601-1-1
Medical electrical equipment
Part 1: General requirements for safety;
st
Collateral Standard: Safety requirements for medical
1
electrical systems
Loading...