PTW 30010, 30011, 30012, 30013 User Manual

User Manual
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Contents
Operating Manual
Technical Manual
D596.131.00/05 en 2016-01 ext/Hn

General Information

Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
General Information
– The product bears the CE-mark "CE 0124"
in accordance with the Council Directive 93/42/EEC about Medical Devices and fulfills the essential requirements of Annex I of this directive. The product is a class IIb device (MDD).
Product standards
The product fully complies with the require­ments of the following standards: – IEC 60731
Electrical Safety
The product fully complies with the require­ments of the following standards: – IEC 61010-1
Electromagnetic compatibility
The product fully complies with the require­ments of the following standards: – IEC 60601-1-2
– PTW-Freiburg cannot be held liable for dam-
age resulting from the use of accessories or consumables from other manufacturers or when the user ignores the instructions and information given in this manual.
– The warranty period is 1 (one) year and begins
on the day of delivery. It is unaffected by repairs covered by the warranty regulations.
– PTW-Freiburg only considers itself responsi-
ble for the safety, reliability, and performance of the product if the assembly, extension, read­justment, modification, or repair is carried out by PTW-Freiburg or by persons authorized by PTW-Freiburg, and if the product is used in compliance with the technical documentation.
– In case of any questions concerning the ser-
vice, support, or warranty, please contact your supplier.
– The user manual is an integral part of the prod-
uct. It should always be kept near the product. Observance of the manual is a prerequisite for proper product performance and correct oper­ation.
– Operator safety, specified measuring accu-
racy, and interference-free operation can only be guaranteed if original products and parts are used. Furthermore, only the acces­sories listed in this manual are approved by PTW-Freiburg and only they or other acces­sories whose use has been expressly permit­ted by PTW-Freiburg may be used in con­junction with the product. Safe operation and proper product performance are not guaran­teed if accessories or consumables from other manufacturers are used.
– This manual is in conformity with the product
specifications and all applicable safety stan­dards that are valid as at the printing date. All rights are reserved for devices, circuits, tech­niques, software, and names as referred to in the manual.
– PTW-Freiburg reserves the right to make mod-
ifications. Please contact PTW or your local representa­tive for the most current information concern­ing the products of interest.
– No part of the technical documentation may
be reproduced without written permission from PTW-Freiburg.
2 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
General Information
– PTW-Freiburg is a registered manufacturer
according to the ElektroG (Elektro- und Elek­tronikgeräte-Gesetz). Elektro-Altgeräte-Register (EAR). Registration number DE15599992
– PTW-Freiburg works in strict accordance with
a quality management system that is contin­uously updated according to the national and international standards.
PTW-Freiburg
Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG Germany Tel.: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de
D596.131.00/05 en 3
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013

Table of Contents

Table of Contents
General Information........................................................................................................2
Table of Contents........................................................................................................... 4
List of Figures.................................................................................................................5
Intended Use................................................................................................................... 6
Safety Information.......................................................................................................... 7
Operating Manual...........................................................13
1 Application................................................................................................................. 13
2 Measuring Ranges.................................................................................................... 18
Technical Manual...........................................................21
3 Technical Specifications.......................................................................................... 21
4 Electromagnetic Compatibility according to IEC 60601-1-2..................................29
5 Accessories and Spare Parts.................................................................................. 35
Service Manual...............................................................37
6 Cleaning......................................................................................................................37
7 Preventive Maintenance........................................................................................... 39
8 Disposal of the Product........................................................................................... 40
Literature........................................................................................................................41
Index...............................................................................................................................42
4 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013

List of Figures

List of Figures
Figure 1: Farmer chamber types 30010, 30011, 30012................................................................................... 6
Figure 2: Farmer chamber type 30013............................................................................................................. 6
Figure 3: Definition of the patient environment...............................................................................................12
Figure 4: Drawing of the Farmer chamber type 30010 (dimensions in mm)..................................................24
Figure 5: Drawing of the Farmer chamber type 30013 (dimensions in mm)..................................................24
Figure 6: Drawing of the Farmer chamber type 30011 (dimensions in mm).................................................. 28
Figure 7: Drawing of the Farmer chamber type 30012 (dimensions in mm)..................................................28
D596.131.00/05 en 5

Intended Use

Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Intended Use
Farmer chambers types 30010, 30011, 30012, and 30013 are standard chambers for absolute and relative dosimetry in combination with preci­sion electrometers according to IEC 60731.
With their outside dimensions they are compatible with the Farmer chambers from other manufactur­ers. The chambers differ in the material used for the chamber walls and for the central electrode.
Furthermore, the Farmer chamber type 30013 is waterproof. The chambers are used for measure­ments in the radiation beam in radiotherapy or for reference purposes in secondary standard labo­ratories.
The Farmer chambers types 30010, 30011, 30012, 30013 will be referred to as detectors in the following chapters. A differentiation between the Farmer chambers types 30010, 30011, 30012, and 30013 is only made when necessary due to technical differences.
Figure 1: Farmer chamber types 30010,
30011, 30012
Figure 2: Farmer chamber type 30013
6 D596.131.00/05 en
Type 30010, 30011, 30012, 30013

Safety Information

Farmer Chamber
Ionization Chamber
Safety Information
This is the safety alert symbol. It is used to alert the user to potential hazards. Obey all safety messages that follow this symbol to avoid any possible bodily injury or equipment damage.
All safety messages consist of the following com­ponents: – Safety alert symbol and signal word – Type of danger – Source of danger – Consequence – Measures to prevent hazards.
Signal Words
DANGER
Indicates an imminent hazard. If not avoided, the hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard can result in death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the hazard can result in minor or moderate injury.
NOTICE
Indicates a potential hazard. If not avoided, the hazard can result in product damage.
HINT
Provides application tips or other useful infor­mation to ensure that you get the most from your equipment.
D596.131.00/05 en 7
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
Safety Information
Strictly observe the following safety informa­tion concerning the signal words DANGER and WARNING. Failure to do so endangers the lives of the patient, user, and other persons involved.
DANGER
Operation in areas where an explosion may occur or in oxygen-enriched atmospheres.
Explosion Hazard!
Do not operate the product in areas of risk where an explosion may occur. Explosion hazards may be caused, for example, by the use of combustible anesthetics, skin-cleansing agents, and disinfectants.
Do not operate the product in oxygen- enriched atmospheres. The atmosphere is considered to be oxygen-enriched when more than 25 % of oxygen or nitrous oxide is added to the ambient air.
DANGER
Electricity is a source of risk, particularly when the product is not in perfect operating condition or when it is operated inappropriately.
Shock Hazard!
– Switch on products on which moisture con-
densation has developed as a result of tem­perature changes only after they are com­pletely dry.
– Liquids must not enter the product. If liquids
have entered the product, dry the product or wait for it to dry completely. The prod­uct must be thoroughly inspected by service personnel before being used again.
DANGER
Electricity is a source of risk, particularly when the product is not in perfect operating condition or when it is operated inappropriately.
Shock Hazard!
– Before using the product, verify that it is in
correct working order and operating condi­tion.
– Before putting the product into operation,
inspect all connection cables, connectors, housing, and covers for signs of damage. Replace damaged cables and connectors immediately.
8 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
DANGER
Electricity is a source of risk, particularly when the product is operated inappropriately.
Shock Hazard!
Exclusion of operation in the patient envi­ronment:
Do not operate the product nor any periph­eral devices in the patient environment (see
Figure 3).
Exclusion of operation with patient contact:
The product is not intended for use in direct contact with the patient. Ensure that neither the product nor any peripheral devices come into contact with the patient.
DANGER
Electricity is a source of risk, particularly when the product is operated inappropriately.
Shock Hazard!
Use of peripheral devices:
Connect only peripheral devices (PC, printer) that meet the requirements of the standard IEC 60950-1 (UL 60950-1) or the standard IEC 61010-1.
Connect devices to other devices or to parts of systems only if it has been ascertained that this connection does not impair the safety of the patient, the operator, or the environment.
If the device specifications do not contain infor­mation regarding the connection of the device to other equipment, consult, for example, the manufacturers of the other equipment or other informed experts to verify that the connection does not impair the safety of the patient, oper­ator, or environment. Always observe the stan­dard IEC 61010-1.
D596.131.00/05 en 9
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
WARNING
Improper handling.
Hazards to Persons!
The product is a medical device and must only be handled by persons who are trained in the use of such equipment and are capable of applying it properly. The operator must be trained in the use and operation of the product.
WARNING
Magnetic and electrical fields are capable of interfering with the proper performance of the product.
Product Failure!
– When operating the product, make sure that
all external devices operated in its vicin­ity comply with the relevant EMC require­ments. X-ray equipment, MRI devices, radio systems, cell phones, etc. are possible sources of disturbance as they may emit higher levels of electromagnetic radiation. Keep the product away from these devices and verify its performance before use.
– Do not use cables longer than specified.
This might impair the electromagnetic com­patibility characteristics of the product.
WARNING
Radiation is a source of risk, particularly when the product is operated inappropriately.
Risks to Persons!
Exclusion of operation as a controlling instrument:
Use the product as a measuring device only. Do not use measuring results to control an equip­ment that produces radiation, e.g., a radiodiag­nostic equipment or a radiotherapy unit.
WARNING
Application for uses other than those stated in the user manual.
Bodily Injury! Equipment Damage!
Always apply the product in compliance with its intended use.
Failure to do so may lead to severe or even fatal injuries and/or equipment damage and may void the warranty.
WARNING
Hazards originating from other system compo­nents.
Hazards to Persons! Equipment Damage!
– Do not use the product in the immedi-
ate vicinity or placed on top of or below other equipment. However, if the applica­tion requires an arrangement of devices as described above, the product should be monitored in order to ensure its proper func­tioning in the specific arrangement.
– Make sure that the product is used in an
electromagnetic environment as described in IEC 60601-1-2.
10 D596.131.00/05 en
Observe the safety information provided in the user manuals of the system components.
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
CAUTION
Use of the product without observing the user manual.
Bodily Injury! Equipment Damage!
Always use the product in compliance with the user manual. Otherwise, the provided protec­tion may be impaired.
Use the product only in conjunction with the products approved by PTW for this pur­pose and/or the products listed in the section "Intended Use".
Use the product within the rated range of use as indicted in the section "Technical Specifica­tions".
Handle the product with care.
NOTICE
To avoid damage to the cables and connectors, observe the following points:
– Do not kink the cables. Observe the stated
minimum bend radius of the cables. Never exert pressure on the cables. Do not step on the cables. Do not roll trolleys over the cables. Do not pull or twist the cables.
– Disconnect the cables from the devices dur-
ing transport and when the devices are not in use. Do not let the cables hang down unrestrained.
– The connectors must always be clean. Do
not allow connectors to lie on the floor. Always protect connectors when pulling them through cable conduits.
D596.131.00/05 en 11
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
HINT
Observe the user manuals of all connected devices!
Symbols on the Nameplate
Symbol Description
Refer to the user manual!
The product bears the CE-mark.
Manufacturer and date of manu­facture
Reference number
Serial number
Separate collection for electrical and electronic equipment! (refer also to the section "Dis­posal of the Product")
For a description of additional symbols, if applicable, please refer to the equipment description.
Figure 3: Definition of the patient environment
12 D596.131.00/05 en

1 Application

Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Application

Operating Manual

1.1 Quick Start (Workflow)

CAUTION
Electricity is a source of risk, particularly when the product is operated inappropriately.
Shock Hazard! Equipment Damage!
Do not open the detector.
CAUTION
Improper handling.
Erroneous Measurements! Malfunction!
Before switching on the connected devices, check that all plugs are properly connected.
NOTICE
To avoid equipment damage, observe the information about electrometer connection and setup and about measurements provided in the user manual of the electrometers used.
HINT
PTW recommends transferring all components of the measuring system from the storage room to the treatment room well in advance of the measurement to allow the temperature in the devices to adapt to the ambient temperature. If the difference in temperature is high, the mea­surement quality may be impaired (particularly when measurements are performed in phan­toms).
D596.131.00/05 en 13
Connect the electrometer and switch it on.
Perform all necessary adjustments at the elec-
trometer. Verify that the high voltage is cor­rectly adjusted.
Connect the detector.
Allow for a stabilization time of 15 minutes
before starting the measurement.
For absolute dosimetry: Determine the correc-
tion factor for the air density correction. If pos­sible, set the correction factor at the electrom­eter or note it down for the correction of the measured value.
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Application

1.2 General Application Information

The detectors are intended to be used with PTW precision electrometers according to IEC 60731. The detectors are used for dose and dose rate measurement for the measuring quantities 'absorbed dose to water' and 'air kerma'. They are applied in absolute and relative dosimetry or for calibration measurements in air or in solid state phantoms.
Only the Farmer chamber, type 30013, is water­proof. These detectors are intended to be used in a water phantom.
Pre-irradiate the detector.
Check the leakage current. The leakage cur-
rent should be low and stable.
Perform a zero adjustment.
Perform the measurement.
The black line on one side of the chamber stem should face the radiation source during use.
NOTICE
To avoid damage to the detector, observe the following points:
The wall around the sensitive volume is very delicate and can easily be damaged.
Do not touch the wall with your hands or with tools.
During storage, protect the chamber wall with the build-up cap.
Handle the detectors with extreme care when inserting them into adapters (for phantoms or check devices) or removing them.
14 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Application
Operate the detectors only in the rated ranges of use indicated in section 3 "Technical Specifi-
cations". Other operating conditions may lead to
incorrect results.
HINT
Using the detectors with other electrometers or within other rated ranges of use is the sole responsibility of the user.
WARNING
High voltage is a source of risk.
Bodily Injury!
The combination of different connection sys­tems to connect the detectors requires special precaution. Connect high voltage only via an active current limitation (I
When the detectors are properly connected, high voltage is not applied to any of the exposed com­ponents.
< 0.5 mA).
max
Air Density Correction
The detectors are vented and require air density correction. The correction factor kTP for air density correction is calculated according to the following formula.
T temperature in sensitive volume [°C] P atmospheric pressure at measuring
site [hPa]
T
calibration temperature 20 °C or 22 °C
0
P
atmospheric pressure for calibration
0
1013.25 hPa
Check the aeration for proper functioning by comparison with other chambers or by using a radioactive check device.
Connection cables with an overall length of up to 100 m can be used between the detectors and the measuring device.
For measurements with
137
Cs, 60Co or higher ener­gies in air, always use the detectors with their build-up caps. Performing the measurements with these radiation qualities without build-up caps will lead to incorrect results.
D596.131.00/05 en 15
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Application
Application in Water
The detectors are not waterproof.
NOTICE
Do not use the detectors in water. Using them in water is not intended and may cause dam­age to the detectors and to the electrometer and may lead to dangerous short-circuits.
The Farmer chamber, type 30013, is waterproof.
NOTICE
To avoid damage to the detector, observe the following points:
The detectors are allowed to remain immersed in water without interruption for a maximum of 12 hours.
After use, remove the detectors from the water and dry them thoroughly.

1.3 Special Application Considerations

NOTICE
Use only accessories, such as positioner, adapter, chamber plate, or check device, approved by PTW for the different applica­tions; refer to section 5 "Accessories and Spare
Parts".
Application with the Radioactive Check Device
Proceed as follows:
Unscrew the build-up cap from the detector.
Introduce the detector holder T48002.3.003
into the chamber well of the check device all the way.
When attaching the build-up cap, make sure that there is no water between the detector and the cap.
Guard Ring
The guard ring of the detector is designed up to the sensitive volume. The potential of the guard ring is equivalent to the potential of the central electrode. When the detector is properly con­nected, high voltage is not applied to any of the exposed components.
Carefully introduce the detector into the holder
all the way. Line up the marks on the chamber stem and on the holder.
When the detectors are not in use: Screw the
build-up cap onto the detector housing.
16 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Application
Application in the Slab Phantom
Special chamber plates are available for mea­surements with the detector. Mount the detector as follows:
Unscrew the build-up cap from the detector.
Carefully insert the detector into the chamber
plate. The mark on the chamber stem should face the radiation source during use.
When the detectors are not in use: Screw the
build-up cap onto the detector housing.
D596.131.00/05 en 17
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Measuring Ranges

2 Measuring Ranges

When used in combination with the PTW electrometers, the typical measuring ranges of the detectors are as given below.
The resolution of the digital display is at least 0.5% of the indicated measuring ranges.
The upper measuring range limit of the dose rate in consideration of the recombination losses is given in section 3 "Technical Specifications".
Measuring ranges for use with a TANDEM
Measuring ranges
Dose rate
LOW MEDIUM HIGH
15 mGy/min - 3 Gy/min
150 mGy/min - 30 Gy/min
1.5 Gy/min - 300 Gy/min
Measuring ranges for use with a UNIDOS
Measuring ranges
Dose (charge)
LOW HIGH
Dose (integrated value)
LOW MEDIUM HIGH
Dose rate
LOW MEDIUM HIGH
100 µGy/min - 600 Gy/min
5 mGy/min - 30 kGy/min
500 m/min - 3.3 MGy/min
600 µGy/min - 600 mGy/min
30 mGy/min - 30 Gy/min
100 µGy - 11 mGy
5 mGy - 1.1 Gy
3 Gy/min - 3 kGy/min
18 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Measuring Ranges
Measuring ranges for use with a UNIDOS E
Measuring ranges
Dose
LOW MEDIUM
100 µGy - 325 Gy
10 mGy - 33 kGy
HIGH
Dose rate
LOW MEDIUM HIGH
600 µGy/min - 300 mGy/min
60 mGy/min - 30 Gy/min
6 Gy/min - 3 kGy/min
1 Gy - 3.3 MGy
Measuring ranges for use with a UNIDOS
Measuring ranges
Dose
LOW
100 µGy - 45 kGy MEDIUM HIGH
500 mGy - 450 MGy
Dose rate
LOW MEDIUM HIGH
600 µGy/min - 750 mGy/min
30 mGy/min - 75 Gy/min
3 Gy/min - 7.5 kGy/min
webline
5 mGy - 4.5 MGy
D596.131.00/05 en 19
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Operating Manual - Measuring Ranges
20 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications

Technical Manual

3 Technical Specifications

Only values with specified tolerances or limits are guaranteed. Values without tolerances are for informa­tional purposes only.

3.1 Farmer Chambers, Types 30010 and 30013

Device designation Farmer chambers
ionization chambers, types 30010, 30013
Manufacturer PTW-Freiburg
Application standard chambers for absolute and relative dosimetry in combination
with precision electrometers according to IEC 60731
Particular design characteris­tics
Measuring quantities air kerma and air kerma rate
Radiation quality photons, electrons, protons
Nominal sensitive volume 0.6 cm³
Effective point of measure­ment (reference point)
– shell chamber (thimble chamber) – vented – type 30010: not waterproof – type 30013: waterproof – guard ring
absorbed dose to water and absorbed dose rate to water exposure and exposure rate equivalent dose and equivalent dose rate
on detector axis, 13 mm behind tip
Reference direction of inci­dence
D596.131.00/05 en 21
radial
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Device designation Farmer chambers
ionization chambers, types 30010, 30013
Dimensions refer to Figure 4 and Figure 5
Response 20 nC/Gy
Chamber voltage nominal +400 V
maximal ±500 V
Leakage current ≤ ±4 fA
Cable leakage ≤ 1 pC/(Gy∙cm)
Cable length T30010: 1 m
T30013: 1.3 m min. bending radius 20 mm (static) min. bending radius 50 mm (for up to a maximum of 50 repetitions)
Maximum dose rate during continuous radiation
99.0 % saturation 10 Gy/s
99.5% saturation 5 Gy/s
(at nominal voltage)
Maximum dose per radiation pulse (at nominal voltage)
Ion collection time (at nomi-
99.0 % saturation
0.91 mGy
140 µs
99.5 % saturation
0.46 mGy
nal voltage)
Polarity effect ≤ ±0.5 % at 60Co
Directional response ≤ ±0.5 % for rotation around the detector axis
≤ ±0.5 % for tilting by ±5° perpendicular to the detector axis
Energy response ≤ ±2 % (70 ... 280) kV, reference 200 kV
≤ ±4 % (200 kV ... 60Co) reference 60Co
Transient periods for
temperature equilibrium pressure equilibrium
(2 ... 3) min/K ≤ 10 s
Pre-irradiation none
Stabilization time The response after 15 minutes and 2 hours differs by less than ±0.5 %
from the value after 1 hour.
Post irradiation leakage signal drops to < 0.5 % within 5 s after irradiation
Stability long-term stability: ≤ ±0.5 % per year
dose stability: ≤ ±1 % per 10 kGy 60Co
22 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Device designation Farmer chambers
ionization chambers, types 30010, 30013
Wall material 0.335 mm PMMA + 0.09 mm graphite
Area density total 56.5 mg/cm²
Electrode material Al Ø = 1.1 mm
Build-up cap 4.55 mm PMMA
Rated range of use
Chamber voltage Energy
±(100 ... 400) V photons (30 kV ... 50 MV) electrons (10 ... 45) MeV protons (50 ... 270) MeV
Field size Temperature Relative humidity Atmospheric pressure Maximum operating alti-
5 cm x 5 cm ... 40 cm x 40 cm (square field size) (10 ... 40) °C / (50 … 104) °F (10 ... 80) %, no condensation (max. 20 g/m³ absolute humidity) (700 ... 1060) hPa 3000 m above sea level
tude
Environmental conditions for transport and storage
Temperature Relative humidity Atmospheric pressure
(-20 ... +60) °C / (4 … 140) °F (10 ... 85) %, no condensation (max. 20 g/m³ absolute humidity) (600 ... 1200) hPa
Maximum dwell in water type 30013: 12h
Connecting systems PTW-M, BNT, TNC
L (only for type 30013)
D596.131.00/05 en 23
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Figure 4: Drawing of the Farmer chamber type 30010 (dimensions in mm)
Figure 5: Drawing of the Farmer chamber type 30013 (dimensions in mm)
24 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications

3.2 Farmer Chambers, Types 30011 and 30012

Device designation Farmer chambers
ionization chambers, types 30011, 30012
Manufacturer PTW-Freiburg
Application standard chambers for the absolute and relative dosimetry in combi-
nation with precision electrometers according to IEC 60731
Particular design characteris­tics
Measuring quantities air kerma and air kerma rate
Radiation quality photons, electrons, protons
Nominal sensitive volume 0.6 cm³
Effective point of measure­ment (reference point)
Reference direction of inci­dence
– shell chamber (thimble chamber) – vented – not waterproof – guard ring
absorbed dose to water and absorbed dose rate to water exposure and exposure rate equivalent dose and equivalent dose rate
on detector axis, 13 mm behind tip
radial
D596.131.00/05 en 25
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Device designation Farmer chambers
ionization chambers, types 30011, 30012
Dimensions refer to Figure 6 and Figure 7
Response 20 nC/Gy
Chamber voltage nominal +400 V
maximal ±500 V
Leakage current ≤ ±4 fA
Cable leakage ≤ 1 pC/(Gy∙cm)
Cable length 1 m
min. bending radius 20 mm (static) min. bending radius 50 mm (for up to a maximum of 50 repetitions)
Maximum dose rate during continuous radiation
99.0 % saturation 10 Gy/s
99.5 % saturation 5 Gy/s
(at nominal voltage)
Maximum dose per radiation pulse (at nominal voltage)
Ion collection time (at nomi-
99.0 % saturation
0.91 mGy
140 µs
99.5 % saturation
0.46 mGy
nal voltage)
Polarity effect ≤ ±0.5 % at 60Co
Directional response ≤ ±0.5 % for rotation around the detector axis
≤ ±0,5 % for tilting by ±5° perpendicular to the detector axis
Energy response T30011: ≤ ±12 % (280 kV ... 60Co) reference 60Co
T30012: ≤ ±2 % (70 ... 280) kV, reference 200 kV;
≤ ±4 % (200 kV ... 60Co) reference 60Co
Transient periods for
temperature equilibrium pressure equilibrium
(2 ... 3) min/K ≤ 10 s
Pre-irradiation none
Stabilization time The response after 15 minutes and 2 hours differs by less than ±0.5 %
from the value after 1 hour.
Post irradiation leakage signal drops to < 0.5 % within 5 s after irradiation
Stability long-term stability: ≤ ±0.5 % per year
dose stability: ≤ ±1 % per 10 kGy 60Co
26 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Device designation Farmer chambers
ionization chambers, types 30011, 30012
Wall material 0.425 mm graphite
Area density total 79 mg/cm²
Electrode material T30011: graphite Ø = 1.0 mm
T30012: Al Ø = 1.1 mm
Build-up cap 4.55 mm PMMA
Rated range of use
Chamber voltage Energy
±(100 ... 400) V photons T30011:
T30012: electrons (10 ... 45) MeV protons (50 ... 270) MeV
Field size Temperature Relative humidity Atmospheric pressure Maximum operating alti-
5 cm x 5 cm ... 40 cm x 40 cm (square field size) (10 ... 40) °C / (50 … 104) °F (10 ... 80) %, no condensation (max. 20 g/m³ absolute humidity) (700 ... 1060) hPa 3000 m above sea level
tude
Environmental conditions for transport and storage
Temperature Relative humidity Atmospheric pressure
(-20 ... +60) °C / (4 … 140) °F (10 ... 85) %, no condensation (max. 20 g/m³ absolute humidity) (600 ... 1200) hPa
Connecting systems PTW-M, BNT, TNC
(140 kV ... 50 MV) (60 kV ... 50 MV)
D596.131.00/05 en 27
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Technical Specifications
Figure 6: Drawing of the Farmer chamber type 30011 (dimensions in mm)
Figure 7: Drawing of the Farmer chamber type 30012 (dimensions in mm)
28 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual

4 Electromagnetic Compatibility according to IEC 60601-1-2

The following statements apply to the Farmer chambers types 30010, 30011, 30012, and
30013. The Farmer chambers will be referred to as prod­uct in these statements.
WARNING
Magnetic and electrical fields are capable of interfering with the proper performance of the product.
Product Failure!
WARNING
Magnetic and electrical fields are capable of interfering with the proper performance of the product.
Product Failure!
Do not use cables longer than specified. This
Do not use the product in the immediate vicin­ity or placed on top of or below other equip­ment. However, if the application requires an arrangement of devices as described above, the product should be monitored in order to ensure its proper functioning in the specific arrangement.
might impair the electromagnetic compatibility characteristics of the product.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions IEC/CISPR 11
Group 1 The product uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby elec­tronic equipment.
RF emissions IEC/CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker
Class B
not applicable
not applicable
The product is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply net­work that supplies buildings used for domestic pur-
1)
poses.
emissions IEC 61000-3-3
1)
The product fulfills the requirements of IEC/CISPR 11 Class B. However, it is not intended for use in domestic establishments.
D596.131.00/05 en 29
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.
Immunity
tests
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surges IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U
T
(> 95 % dip in UT) for ½ cycle
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) for 25 cycles
Compliance
level
Electromagnetic environment –
guidance
not applicable Floors should be wood, concrete or
ceramic tile. If floors are covered with synthetic material, the relative humid­ity should be at least 30 %.
not applicable
Mains power quality should be that of a typical commercial or hospital envi­ronment.
± 1 kV for input/output lines
not applicable Mains power quality should be that of
a typical commercial or hospital envi­ronment.
not applicable Mains power quality should be that of
a typical commercial or hospital envi­ronment. If the user of the product requires continued operation during power mains interruptions, it is rec­ommended that the product be pow­ered from an uninterruptible power supply or a battery.
< 5 % U
T
(> 95 % dip in UT) for 5 s
30 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity
tests
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
IEC 60601
test level
Compliance
level
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commer­cial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic environment –
guidance
D596.131.00/05 en 31
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.
Immunity
tests
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601
test level
3 V
rms
150 kHz to 80 MHz
3 V/m 80 MHz to
2.5 GHz
Compliance
level
U1 = 3 V
E1 = 3 V/m
Electromagnetic environment –
guidance
Portable and mobile RF communications equip­ment should be used no closer to any part of the product, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = (3.5 / U1) √P = 1.2 √P
d = (3.5 / E1) √P = 1.2 √P 80 MHz to 800 MHz
d = (7 / E1) √P = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmit­ter manufacturer and d is the recommended sepa- ration distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a), should be less than the compliance level in each frequency range.
b)
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
32 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the product is used exceeds the applicable RF compliance level above, the product should be observed to verify normal operation. If abnormal performance is observed, addi­tional measures may be necessary, such as re-orienting or relocating the product.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [U1] V/m = 3 V/m.
D596.131.00/05 en 33
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
Recommended separation distances between portable and
mobile RF communications equipment and the product
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the product can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmit­ters) and the product as recommended below, according to the maximum output power of the communi­cations equipment.
Rated maximum
output power
Separation distance according to frequency of transmitter
m
of transmitter
W
150 kHz to 80 MHz
d = (3.5 / U1) √P = 1.2 √P
80 MHz to 800 MHz
d = (3.5 / E1) √P = 1.2 √P
800 MHz to 2.5 GHz
d = (7 / E1) √P = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manu-
facturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
34 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Accessories and Spare Parts

5 Accessories and Spare Parts

Radioactive check device 90Sr T48012
Holder for radioactive check device T48002.3.003
TRUFIX precision mounting system for positioning of detectors in TBA sys­tems
TRUFIX precision mounting system, holder for Farmer chamber T30013 T4316/U381
PMMA chamber plate 30 cm x 30 cm T2967/29
PMMA chamber plate 40 cm x 40 cm T40005.3.021
RW3 chamber plate 30 cm x 30 cm T29672/U19
RW3 chamber plate 40 cm x 40 cm T40006.1.015
L981150
D596.131.00/05 en 35
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual - Accessories and Spare Parts
36 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Cleaning

Service Manual

WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
Ensure that the service tasks described below are only performed by specialized staff.
In addition, ensure that all other service tasks are only performed by PTW-Freiburg or by a person authorized by PTW-Freiburg.

6 Cleaning

General Information on Cleaning
DANGER
Liquids have entered the product.
Shock Hazard! Erroneous Measurements!
Have the product inspected by service person­nel before putting it into service again.
NOTICE
To avoid surface damage, do not use disinfec­tants on a phenol base or peroxide compounds to disinfect the surface.
The product cannot be sterilized or completely disinfected.
The outer shell of the detectors is made of plastic. To avoid impairing the equipment functions, the detectors must not be treated with or come in con­tact with grease, oil, alcohol, or substances con­taining solvents.
Cleaning agents and solvents contained in disin­fectants (alcohols) have proved to be particularly harmful.
D596.131.00/05 en 37
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Cleaning
Clean the product simply by wiping it down with a dry or slightly moistened cloth. Use water or a mild soap solution.
Alternatively, you can use a common disinfection agent for medical instruments which is specified in the list of disinfectants of your national hygiene institute (e.g., VAH in Germany). If your country does not provide such a list, we recommend either a solution on an aldehyde base or a solution with a quaternary ammonium compound.
Detectors and Connection Cables
Disconnect the detectors from the measuring
device (always pull on the connector, not on the cable!).
Only wipe the detectors and the connection
cables down. Do not immerse the detectors or connection cables in liquid. Liquids must not be allowed to penetrate. Do not spray anything onto the detectors and connectors. All con­tacts must remain clean and dry.
Connectors
NOTICE
To avoid equipment damage, do not immerse the connectors in liquids. Liquids must not be allowed to enter the connectors.
The connectors must always be clean. Use a soft brush to remove dust (do not blow onto the con­nector!).
Non-use
When not in use, the product should be stored in its case for protection.
The connector of the detector should be protected against humidity and dust with the supplied cap.
The chamber volume must be protected against mechanical damage by means of the build-up cap.
38 D596.131.00/05 en
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Preventive Maintenance

7 Preventive Maintenance

Below, you will find a list of the maintenance mea­sures and inspections necessary for the product.
Check before each use
Before each use, visually inspect the product
and cables for signs of mechanical damage.
Conduct a functional control. For this, you can
use a radioactive check device of PTW (refer to section 5 "Accessories and Spare Parts"). For further information on the functional control please refer to the user manual of the radioac­tive check device.
If damage or malfunctions are identified, the
product must be repaired before it is used again.
Check the leakage current after having con-
nected the detector. The detectors are designed for operation in a limited temperature and humidity range (refer to section 3 "Technical Specifications"). If a detector has been subjected to higher humid­ity this can result in increased leakage current. Such a detector can normally be restored to normal functionality by heating to 40 °C for at least 8 hours at low humidity.
Safety Inspections
Only regularly inspected products are fail-safe. To preserve the functional security and operational safety, a safety inspection is to be executed at regular intervals according to the national regula­tions.
These inspections must only be performed by independent persons with adequate training and experience.
It is recommended to execute the safety inspec­tions every 2 years.
Technical Inspection of the Measuring System
The product must be inspected at regular inter­vals according to the national regulations. It is rec­ommended to perform technical inspections of the measuring system every 2 years at PTW-Freiburg or a qualified calibration laboratory.
Inspections must also be performed after repairs or after each influence that may have changed the behavior of the product.
Repair
Repairs may only be carried out by PTW-Freiburg or by persons/companies authorized by PTW­Freiburg.
D596.131.00/05 en 39
Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Disposal of the Product
Calibration
The exact calibration factor and supplementary data are included in the calibration certificate.
For use at high energies (above 1.3 MeV cor­rection factors according to the relevant national or international dosimetry protocols must be observed (e.g., IAEA TRS 277 or TRS 398).
The product should be recalibrated at a dose rate typical for the application. Calibrations should be done at 60Co.
It is recommended to have the product calibrated every 2 years at PTW-Freiburg or by specially trained staff on site (when using a radioactive check device and performing regular check mea­surements: 6 years).

8 Disposal of the Product

The typical lifetime of the product is approx. 10 years, if used normally.
At the end of the product lifetime, the components of the product must be disposed of in compliance with the applicable waste control regulations. The different materials must be separated and recy­cled as appropriate.
The electronic components must be recycled according to the local regulations.
The product does not contain hazardous materi­als.
40 D596.131.00/05 en

Literature

Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Literature
[1] Council Directive 93/42/EEC concerning medical devices
(Medical Device Directive - MDD)
[2] Council Directive 97/43/EURATOM on health protection
of individuals against the dangers of ionizing radiation in relation to medical exposure
[3] IEC 60731
Medical electrical equipment ­Dosimeters with ionization chambers as used in radio­therapy
[4] IEC 61010-1
Safety requirements for electrical equipment for mea­surement, control and laboratory use ­Part 1: General requirement
[5] IEC 60601-1-2
Medical electrical equipment ­Part 1-2: General requirements for basic safety and essential performance ­Collateral Standard: Electromagnetic compatibility ­Requirements and tests
[6] IEC/CISPR 11
Industrial, scientific and medical equipment ­Radio-frequency disturbance characteristics - Limits and methods of measurement
[7] IEC 61000-3-2
Electromagnetic compatibility (EMC) ­Part 3-2: Limits ­Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)
[8] IEC 61000-3-3
Electromagnetic compatibility (EMC) ­Part 3-3: Limits ­Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equip­ment with rated current ≤ 16 A per phase and not subject to conditional connection
[9] IEC 61000-4-2
Electromagnetic compatibility (EMC) ­Part 4-2: Testing and measurement techniques ­Electrostatic discharge immunity test
[10] IEC 61000-4-3
Electromagnetic compatibility (EMC) ­Part 4-3: Testing and measurement techniques ­Radiated, radio-frequency, electromagnetic field immu­nity test
[11] IEC 61000-4-4
Electromagnetic compatibility (EMC) ­Part 4-4: Testing and measurement techniques ­Electrical fast transient/burst immunity test
[12] IEC 61000-4-5
Electromagnetic compatibility (EMC) ­Part 4-5: Testing and measurement techniques ­Surge immunity test
[13] IEC 61000-4-6
Electromagnetic compatibility (EMC) ­Part 4-6: Testing and measurement techniques ­Immunity to conducted disturbances, induced by radio­frequency fields
[14] IEC 61000-4-8
Electromagnetic compatibility (EMC) ­Part 4-8: Testing and measurement techniques ­Power frequency magnetic field immunity test
[15] IEC 61000-4-11
Electromagnetic compatibility (EMC) ­Part 4-11: Testing and measurement techniques ­Voltage dips, short interruptions and voltage variations immunity tests
D596.131.00/05 en 41

Index

Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Index
A
Accessories............................................................. 35
Air density correction............................................... 15
Application............................................................... 13
Application considerations, special.......................... 16
Application information, general.............................. 14
Application in water................................................. 16
B
Build-up caps........................................................... 15
C
Calibration................................................................ 40
Chamber wall........................................................... 14
Cleaning................................................................... 37
D
Dimensions Farmer chamber type 30010................ 23
Dimensions Farmer chamber type 30011................ 27
Dimensions Farmer chamber type 30012................ 28
Dimensions Farmer chamber type 30013................ 24
Disposal................................................................... 40
Q
Quick start............................................................... 13
R
Radioactive check device........................................ 16
Repair...................................................................... 39
S
Safety information...................................................... 7
Safety inspections................................................... 39
Slab phantom.......................................................... 17
Spare parts.............................................................. 35
Symbols................................................................... 12
T
Technical inspections of the measuring system....... 39
Technical specifications............................... 21, 21, 25
W
Workflow.................................................................. 13
E
Electromagnetic compatibility.................................. 29
G
General information................................................... 2
Guard....................................................................... 16
Guard ring................................................................ 16
H
High voltage............................................................. 15
I
Intended use.............................................................. 6
L
List of figures............................................................. 5
Literature.................................................................. 41
M
Maintenance............................................................ 39
Measuring ranges.................................................... 18
42 D596.131.00/05 en
Loading...