Protek A434L Operating Manual

INTRODUCTION ................................................................................ 6
FIRMWARE VERSIONS ....................................................................... 6
SAFETY INSTRUCTIONS ..................................................................... 7
CERTIFICATE .................................................................................. 11
1 GENERAL OVERVIEW ............................................................... 12
1.1 Description ....................................................................... 13
1.2 Specifications ................................................................... 13
1.2.1 Basic Specifications .................................................... 13
1.2.2 Key Measurements ..................................................... 15
1.2.3 Key Features .............................................................. 15
1.3 The Layout of A434L ......................................................... 16
1.4 Ordering Information ........................................................ 17
1.4.1 Standard Accessories .................................................. 17
1.4.2 Options ..................................................................... 19
2 PREPARATION FOR USE ........................................................... 21
2.1 General Information .......................................................... 21
2.2 Power Adaptor ................................................................. 21
2.3 Battery ............................................................................. 22
2.3.1 Caution ..................................................................... 23
2.3.2 Installing a Battery...................................................... 24
2.3.3 Charging a Battery ..................................................... 25
2.3.4 Battery Low ............................................................... 25
2.3.5 Battery Power Off ...................................................... 26
2
2.4 Front Panel ....................................................................... 27
2.4.1 Key Layout ................................................................ 27
2.4.2 ① Function Hard ........................................................ 28
2.4.3 ② Soft Keys ............................................................... 28
2.4.4 ③ Power Switch Key .................................................. 28
2.4.5 ④ Multi Keys ............................................................. 29
2.4.6 ⑤ ESC, Back, Arrow Keys ............................................ 33
2.4.7 ⑥ Enter Keys ............................................................. 34
2.4.8 ⑦ Knob Keys ............................................................. 35
2.4.9 Display ...................................................................... 35
2.5 Upper Panel ..................................................................... 36
2.5.1 DC Power Connector.................................................. 37
2.5.2 RF OUT/Reflection ...................................................... 37
2.5.3 Ethernet Port ............................................................. 37
2.5.4 USB Port ................................................................... 38
2.6 FAN ................................................................................. 38
2.6.1 FAN Operation .......................................................... 38
2.6.2 FAN alarm ................................................................. 38
3 GETTING STARTED .................................................................. 40
3.1 Power Up ......................................................................... 41
3.1.1 Initialization ............................................................... 41
3.1.2 System Information .................................................... 42
3.2 Layout of the Display Screen .............................................. 43
3.3 Function Hard Keys Structure ............................................. 45
3.3.1 System Screen Menu .................................................. 46
3.3.2 Common Menu ......................................................... 47
3.3.3 VSWR Menu .............................................................. 52
3.3.4 DTF Menu ................................................................. 54
3.3.5 Cable Loss Menu........................................................ 55
3.3.6 Power Meter Menu .................................................... 57
3.4 System Key ....................................................................... 58
3
3.5 Multi Keys ........................................................................ 60
3.5.1 Numeric Data Entry .................................................... 61
3.5.2 Specific Function Keys ................................................ 61
4 MEASUREMENT DATA ANALYSIS.............................................. 70
4.1 VSWR .............................................................................. 71
4.1.1 VSWR Display Description ........................................... 71
4.1.2 Setting Frequency ...................................................... 74
4.1.3 Calibration ................................................................ 75
4.1.4 VSWR Measurement .................................................. 77
4.1.4.1 Setting Trace Point .............................................. 79
4.1.4.2 Scale Adjustment ................................................ 79
4.1.4.3 Using Markers .................................................... 79
4.1.4.4 Limit Line ........................................................... 80
4.2 DTF ................................................................................. 81
4.2.1 DTF Display Description .............................................. 81
4.2.2 DTF Setup ................................................................. 83
4.2.3 Calibration ................................................................ 85
4.2.4 DTF Measurement...................................................... 87
4.2.4.1 Scale Adjustment ................................................ 88
4.2.4.2 Using Markers .................................................... 89
4.2.4.3 Windowing ........................................................ 89
4.3 Cable Loss ........................................................................ 91
4.3.1 Cable Loss Display Description ..................................... 91
4.3.2 Setting Frequency ...................................................... 92
4.3.3 Calibration ................................................................ 93
4.3.4 Cable Loss Measurement ............................................ 95
4.3.4.1 Scale Adjustment ................................................ 96
4.3.4.2 Using Markers .................................................... 96
4.4 Power Meter .................................................................... 97
4.4.1 Connecting Sensors.................................................... 97
4.4.2 Setting Power Meter .................................................. 99
4.4.3 Power Measurement ................................................ 100
4
4.5 Save & Load ................................................................... 101
4.5.1 Save ....................................................................... 101
4.5.1.1 Save Trace ........................................................ 102
4.5.1.2 Save Screen ...................................................... 103
4.5.1.3 Save State ........................................................ 104
4.5.2 Load ....................................................................... 105
4.5.2.1 Load Trace ....................................................... 105
4.5.2.2 Load Screen ...................................................... 107
4.5.2.3 Load State ........................................................ 107
4.5.2.4 Memory Type ................................................... 107
5 SOFTWARE UPGRADE............................................................ 108
5.1 Upgrade Method ............................................................ 109
6 WARRANTY INFORMATION .................................................... 113
Appendix 1 Band List ................................................................. 117
Appendix 2 — Cable List ................................................................ 125
5
RF Master operation and maintenance should be performed by qualified engineers with initial experience in operating of microwave circuits and PC.
The following abbreviations are used in this Manual:
PC – Personal Computer
DUT – Device Under Test
IF – Intermediate Frequency
CW – Continuous Wave
SWR – Standing Wave Ratio
The first edition of this Operating Manual applies directly to instruments that have the firmware version 1.xx.
6
Carefully read through the following safety instructions before putting the Analyzer into operation. Observe all the precautions and warnings provided in this Manual for all the phases of operation, service, and repair of the Analyzer.
The Analyzer must be used only by skilled and specialized staff or thoroughly trained personnel with the required skills and knowledge of safety precautions.
A434L complies with INSTALLATION CATEGORY II as well as POLLUTION DEGREE 2 in IEC61010–1.
A434L is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II, III, or IV.
A434L is for INDOOR USE only.
A434L is tested in stand-alone condition or in combination with the accessories supplied by GS INSTRUMENTS CO., LTD. against the requirement of the standards described in the Declaration of Conformity. If it is used as a system component, compliance of related regulations and safety requirements are to be confirmed by the builder of the system.
N
Never operate the Analyzer in the environment containing inflammable gasses or fumes.
Operators must not remove the cover or part of the housing. The Analyzer must not be repaired by the operator. Component replacement or internal adjustment must be performed by qualified maintenance personnel only.
Never operate the Analyzer if the power cable is damaged.
Never connect the test terminals to mains.
7
Electrostatic discharge can damage your Analyzer when connected or disconnected from the DUT. Static charge can build up on your body and damage the sensitive circuits of internal components of both the Analyzer and the DUT. To avoid damage from electric discharge, observe the following:
- Always use a desktop anti-static mat under the DUT.
- Always wear a grounding wrist strap connected to the desktop
anti-static mat via daisy-chained 1 MΩ resistor.
- Connect clamp on the body of the Analyzer to the body of
the DUT before you start operation.
Observe all the general safety precautions related to operation of equipment powered by mains.
The protection provided by the equipment may be impaired if the equipment is used in a manner not specified by the manufacturer.The definitions of safety symbols used on the instrument or in the Manual are listed below.
Storage
Before first use store your equipment in the factory package at environment temperature from 0 to +50 ºС and relative humidity up to 95% (at 25ºС).
After you have removed the factory package store the equipment at
environment temperature from +10 to +35 ºС and relative humidity up to 80% (at 25ºС).
Ensure to keep the storage facilities free from dust, fumes of acids and alkalies, aggressive gases, and other chemicals, which can cause corrosion.
8
Moderate Ventilation
Environment
Ambient temperature 2ºС to 50ºС
Maximum 80% RH between 2ºС and 31ºС, decreasing to 50% RH at 50ºС
Maximum altitude 2,000m
MAINS supply voltage fluctuations up to 10% of the nominal voltage
Transient over-voltages according to installation category II
Pollution degree 2
IPX0
Refers to the Manual if the instrument is marked with this symbol.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
Maintain moderate ventilation condition to prevent overheating of the product.
Warning
Never perform cleaning of the instrument if the power cable is connected
to the power outlet.Never clean the internal components of the instrument.
N
Do not position the equipment so that it is difficult to operate the power cord.
9
A chassis terminal; a connection to the
instrument’s chassis, which includes all exposed
metal structure.
CAUTION
This sign denotes a hazard. It calls attention to a procedure, practice, or condition that, if not correctly performed or adhered to, could result in injury or death to personnel.
CAUTION
This sign denotes a hazard. It calls attention to a procedure, practice, or condition that, if not correctly performed or adhered to, could result in damage to or destruction of part or the entire instrument.
Note
This sign denotes important information. It calls attention to a procedure, practice, or condition that is essential for the user to understand.
10
CERTIFICATE
11
1
GENERAL OVERVIEW
OPERATING MANUAL
Item
Sub Item
Specification
General
Max Input Power
+25dBm Damage level
Frequency Range
5MHz to 4GHz
Frequency Accuracy
< ±3ppm
Frequency Resolution
10kHz
Impedance
50Ω
Scan Speed
< 1msec /data point
Display
Single & Dual mode
Test port
N Female
1.
1.1 Description
A large number of cell site problems are caused by the antenna system, cable, or connectors. It is important to have the right instrument available when either servicing or certifying cell sites for operation.
The Protek A434L RF master is a lightweight portable diagnostic tool for an accurate detection of operational problems. The Protek A434L has all the measurement functions necessary to accurately verify the antenna system from VSWR to power measurements. In addition, the Protek A434L makes distance-to-fault measurements to accurately pinpoint the fault's location.
The Protek A434L user interface is with a front keypad and a TFT color, 7­inch display providing ease of use and control. The application specific software allows the user to easily compare and analyze measurements and generate comprehensive reports. The Protek A434L is equipped with a rechargeable and infield installable lithium-ion battery providing over four hours of operation.
1.2 Specifications
1.2.1 Basic Specifications
13
Test curve storage
Internal : Minimum 512MB External : Limited by size of USB (32G)
Screen storage
Setup storage
VSWR
Number of data points
126, 251, 501, 1001, 2001
Return loss Range
0 to -60dB
VSWR Range
1 to 65
Cable Loss
Cable loss range
0 to -30dB, 0.01dB Resolution
Interference Immunity
On-Frequency
+10dBm
On-Channel
+20dBm
DTF
Return Loss Display Range
0 to 60dB Distance Range
0 to 1250m (4125ft)
VSWR Display Range
0 to 65
Miscellaneous
Dimension
260X193X67mm
Weight
<2.45Kg include battery
A434L Voltage and Current
12Vd.c., 3.5A
Adaptor AC Input
100 to 250Vd.c., 1.5A
DC Output
12Vd.c., 5A
Battery
Li-lon (4hr operating time after full charging), 12Vd.c., 7600mAh
Environmental
Operating Temperature
0°C ~ +50°C
Storage Temperature
-40°C ~ +80°C (-40°F ~ +176°F)
Humidity
95%R.H. NO Condensation
Degree of protection
IPX0
Power Meter(Option)
Frequency Range
20MHz to 3.8GHz
Sensor Type
Average
Peak Power Sensor
-40dBm to +10dBm
Accuracy
±7%
Test Port
Precision N Female
14
Table 1.1 Basic Specifications
1.2.2 Key Measurements
High resolution VSWR Measurements Distance to Fault (DTF) Measurements Cable Loss Measurements RMS Power Measurements (optional)
1.2.3 Key Features
Rechargeable and infield replaceable lithium-ion battery Built-in world-wide signal standards and frequency
channels
7 Inch TFT color display viewable in daylight Dual display to view multiple measurements
simultaneously to reduce test time
Easy front keypad operation Superior immunity to RF interference Up to 2001 data points to locate long range problems Built-in cable menus containing >90 cables’
characteristics
User friendly menu structure Saves up to Memory (user setups, traces, screens) Alphanumeric labeling of saved data Automatic Time/Date stamp of saved data USB Port (USB 2.0) Remote firmware upgrade capability
15
Fast one-touch selection of menu item or positioning
Smart Battery management can be check Battery
Rechargeable and field replaceable Lithium Ion battery
Backlight keypad for easier use in dark environments Sun-Light Readable LCD for readability
1.3 The Layout of A434L
marker
capability
with more than 4 hours operation time
16
Figure 1.1 A434L Front Panel
Figure 1.2 A434L Rear Panel
Figure 1.3 A434L Upper Panel
1.4 Ordering Information
1.4.1 Standard Accessories
The standard accessories supplied with the A434L RF Master are as follows:
17
Description
Picture
Specification
Soft Carrying Case
AC-DC Adapter
Input: 100 ~ 240Va.c.,
1.5A
Output: 12Vd.c., 5A
Neck Strap
Hand Strap Li-ion Battery
Output: 8.4Vd.c. MAX,
7800mAh
User’s Manual
Hard Copy
18
Description
Picture
Specification
Power Sensor
(Part No: S134)
Frequency Range
: 20MHz to 3.8GHz
Sensor Type : Average
Peak Power Sensor
: -40dBm to +10dBm
Accuracy : ±7%
Test Port : Precision N Female
Battery Charger
(Part No: C122R5)
Input: 100 ~ 240Va.c.,
1.5A
Output: 12Vd.c., 5A
1.4.2 Options
The options supplied with the A434L RF Master are as follows:
19
2
PREPARATION FOR USE
OPERATING MANUAL
2.
2.1 General Information
Unpack the Analyzer and other accessories. Check the contents of the package against the list specified in section 1.4.1. Connect your A434L to the AD-DC Adaptor(12Vd.c., 5A) or Li-ion Battery(8.4Vd.c. MAX, 7800mAh) power source by means of the Power Cable supplied with the instrument.
Warm-up the Analyzer for 20 minutes after turning the power on.
Assemble the test setup using cables, connectors, fixtures, etc, which allow DUT connection to the Analyzer.
Perform calibration of the Analyzer. Calibration procedure is described in section 4.1.3.
2.2 Power Adaptor
The equipment uses following power supply.
Product: Switching Power Supply Adaptor
Model no. FSP060-DBAE1
AC INPUT: 100-240Va.c.~, 1.5A, 50-60Hz
DC OUTPUT: 12.0Vd.c. 5.0A MAX
Manufactured by: Zhonghan Electronics (Shenzhen)
Trade mark: FSP GROUPINCUC
21
2.3 Battery
A434L use CE, UL Certificated PBP-7800 Battery.
Figure 2.1 A434L Battery
CAUTION
RISK OF EXPLOSIOM IF BATTERY IS REPLACED BY AN INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO THE MANUFACTURER’S INSTRUCTIONS.
22
If the push button on the battery is pressed, the remained
battery volume will be displayed in the battery icon next to the
push button.
Each level of the battery indicator takes 20% of the battery
portion. If the battery level is three, it means the battery is charged 40% - 60%.
As illustrated below, please insert the battery pack while the
printed label is on the upside.
Non rechargeable Li-Ion battery was in analyzer. This non rechargeable Li-Ion battery is CR02032.(made by Hitachi
Maxwell Ltd)
2.3.1
Caution
Avoid placing the battery pack near heating sources of on the
place near windows.
Do not store the battery pack in the high humidity. If the battery is unused for a long time, separate it from the unit. Keep this battery pack away from children If electrolyte from the battery pack is leaking or the battery pack
smells strangely, keep the battery pack away from fire.
In case that electrolyte from the battery pack is touched by any
part of the human body, wash it immediately and go to hospital to prevent potential damage.
Use the authorized charger only. Do not abandon the battery pack in the car in the summer. Avoid any shock to the battery pack. Do not replace non rechargeable Li-Ion battery except for
certified person.
This battery has about ten-month warranty.
23
1
Open the battery cover.
2
Insert the battery (Observe correct battery polarity orientation when installing).
3
Close the battery cover.
2.3.2
Installing a Battery
Caution: Full charge the battery before first using the analyzer.
Figure 2.2 A434L Right Side View
24
1
Insert the battery in the analyzer.
2
Plug in the AC- DC adapter and switch on the external power.
3
The charge indicator icon on the screen, indicating that the battery is charging and is fully charged.
2.3.3
Charging a Battery
You may charge the battery both in the analyzer and in the external battery charger.
During charging and discharging, the battery voltage, current, and temperature are monitored. If any of the monitored conditions exceed their safety limits, the battery will terminate any further charging or discharging until the error condition is corrected.
2.3.4
Battery Low
If A434L runs out of battery, the following message will appear on the display.
25
Figure 2.3 Battery 10% or Battery 5% Screen Display
2.3.5
Battery Power Off
If the battery power is lower than 1%, the following message will appear on the display, and the Power will be Off.
26
Figure 2.4 Battery Power Off Screen Display
2.4 Front Panel
The front view of A434L is represented in figure 1.1. The front panel is equipped with the following parts:
2.4.1
Key Layout
Figure 2.5 A434L Key Layout
27
«VSWR» Activates the VSWR measurement mode.
«DTF» Activates the DTF measurement mode.
«Cable Loss» Activates the Cable Loss
measurement mode.
«Power Meter» Activates the RF Power measurement mode. An optional power sensor must be connected to the instrument before using this function
«System» Opens the system screen menu.
«CAL» Opens the calibration screen menu.
2.4.2 ① Function Hard keys
The six hard keys are located at the left of the display. The function of each key is indicated on each key. Depending on the selection of Hard Keys, the functions shown in the screen menu may vary.
2.4.3 ② Soft Keys
The six keys at the right of the display are the Soft Keys.
28
Each Soft Key has a function which varies depending on the operation mode. The function of each key is displayed at the right side of the display as a screen menu. As the LCD is touch-screen, the function can be selected by touching the screen menu.
«Power Switch» toggles between On and Off states of the Analyzer.
Note
Holding the key for more than 3 seconds on an operating Analyzer will turn off the power supply
CAUTION
Do not disconnect the power cable from the mains when the Analyzer is operating. This can damage the Analyzer software.
2.4.4 ③ Power Switch Key
2.4.5 ④ Multi Keys
Enter values printed on the keys when a value input pop-up window is prompted on the screen. Open a menu linked with the functions in blue or directly perform the specified function.
29
«Numeric & specific function» key
Multi keys operate as follows when a user is prompted to input values by pop-up window.
Numeric keys from 0 to 9 are used
to input user setting numbers.
The plus/minus (+/-) key is used to
assign positive or negative numbers in the user settings.
The period key (.) is used to input
numbers with decimal points.
Multi keys are used in VSWR, DTF, and Cable Loss measurement mode. Unless otherwise specified, multi keys do not support Power Meter mode.
«0»…«9», «.» keys
...
«0»…«9», numeric keys and decimal point used for entry of decimal numbers at the cursor location.
«±» key
«±» key is used to change the sign of a
decimal number.
«Auto Scale» key
«Auto Scale» The instrument can
automatically set the scale to the minimum and maximum values of a measurement on the Y-axis of the graph for optimum display of the traces. Every time AUTO SCALE key is pressed, the top and bottom scales are set to the minimum and maximum values with margin on the Y-axis of the screen display.
30
«AMP» key
«AMP» AMP (amplitude) defines a
manual setting for the scale on the Y­axis of the graph. It can be selected in VSWR, DTF, and Cable Loss measurement mode.
«Freq/Dist» key
«Freq/Dist» key causes different
screen menu to be displayed depending on a measurement mode. In VSWR or Cable Loss measurement modes it opens a frequency screen menu and in DTF measurement mode it opens a distance screen menu.
«Trace Point» key
«Trace Point» Trace point is used to
select the number of data points to take during a VSWR and Cable Loss measurement mode. There are four data points available: 126, 251, 501, 1001 and 2001. The default number of trace points is 1001.
«Marker» key
«Marker» A marker is used to get
the data on the specific point of a trace. A total of six markers can be displayed on the screen and each marker can be assigned independently. Placing a marker on
the trace displays the marker’s Y
coordinates next to the marker’s position. If the marker table is turned on, both X and Y coordinates of all activated markers are displayed.
31
«Trace» key
«Trace» key enables for selection of
active channel and trace. Active channel is a channel, which accepts the channel parameter setting commands. Active trace is similarly a trace, which accepts the trace parameter setting commands. Before sending commands to change a channel or a trace settings, activate the channel or trace,
«Peak» key
«Peak» This key is used to find the
peak value of a trace. Pressing this key leads to the activation of Marker CH1 and places the marker to the peak point of the trace.
«Load» key
«Load» key enables for recalling of
the Trace, Screen, and State.
«Save» key
«Save» key enables for saving of the
Trace, Screen, and State.
«Light» key
«Light» This key is used to adjust the
brightness of the LCD display. Adjustment can be made from 0 to 100% and the default setting is 50%.
32
«Scale» key
«Scale» This key is used to select a
unit of Y axis of the graph in VSWR and DTF measurement modes. Either VSWR or Return Loss can be selected. In Cable Loss measurement mode, Return Loss only can be selected.
«Hold» key
«Hold» This key is used to pause a
sweep in all measurement modes.
«ESC» key
«ESC» key is used for the following:
If there is an active data entry
field, this key cancels all
the data entered in the
field and restores the value
of the field as it was before
any new numbers or
characters were entered.
If there is no active data entry
field, this key brings the
user back to an upper level
of the soft key menu.
2.4.6 ⑤ ESC, Back, Arrow Keys
33
«Back» key
«Back»
Delete a character or number.
«Arrow» key
«Arrow» key is used for the
following:
Moves a marker position or
highlighted selection bar
one point at a time.
«ENTER» key
«ENTER» key is used for the
following:
If there is an active data
entry field, this key
completes the entry
process and assigns the
new value to the field.
If there is no active data
entry field, this key
enables the function of
a highlighted soft key.
2.4.7 ⑥ Enter Keys
34
«Knob» key
«Knob» The rotary knob is used for
the following:
Move an active marker or a
highlighted selection bar in
the list.
The Analyzer is equipped with 7” color LCD. The display can come
complete with a touch screen (TS option). The touch screen LCD allows
manipulating the Analyzer by touching the LCD screen directly with a
finger.
CAUTION
Do not press the touch screen with a
pen, screwdriver or any other sharp-
pointed object. This can damage the touch screen.
2.4.8 ⑦ Knob Keys
2.4.9 Display
35
The display consists of the control elements and area for measurement result
indication.
The main control element of the display is a vertical soft key menu bar in the right-
hand part of the screen.
The soft key menu bar consists of program soft key panels, which appear one
instead of the other on the screen. Each panel represents one of the submenus of
the soft key menu. All the panels are integrated into multilevel menu system and allow access to all the functions of the Analyzer.
Note
The top line of the screen represents
the menu bar, which enables you direct
access to the submenus of the soft key
menu. This menu is an auxiliary one and can be hidden.
2.5 Upper Panel
All Input/Output ports of the Protek A434L are located on the upside of the instrument as shown in Figure 2.6
36
Figure 2.6 A434L Upper Panel
Provides input for the DC power source via an AC-DC adapter.
In case of emergency, to avoid danger of electric shock or the like, pull the power cable out of the power outlet or the DC power connector of the instrument.
Output(Accepts) an internal(external) output(input) with a frequency range from 5MHz to 4GHz.
50ohm N-type RF Connector, maximum allowable input is +25dBm.
Ethernet port allows the user to connect the Analyzer to a LAN (Local Area Network). This connection enables the user to control the instrument using an external PC. Used for upgrading the instrument’s firmware.
10BASE-T & 100BASE-TX support
CAT5e UTP Cable uses
2.5.1 DC Power Connector
2.5.2 RF OUT/Reflection
2.5.3 Ethernet Port
37
USB ports allow connection of various external USB compatible devices, such as flash memory stick and Power Sensor. Used for either saving measurement data or an optional external power sensor.
5Vd.c. , 500mA
Type A
2.5.4 USB Port
2.6 FAN
2.6.1 FAN Operation
FAN operates if the internal temperature of the system rises.
The FAN is ON when the temperature is higher than 70.0
The FAN is OFF when the temperature is lower than 65.0
2.6.2 FAN Alarm
The message of “Fan check required!!!” will be indicated in the bottom left corner of the displaye if the FAN does not properly operates.
38
Figure 2.7 FAN Alarm Screen Display
39
3
GETTING STARTED
OPERATING MANUAL
3.
3.1 Power Up
3.1.1
Initialization
The initialization screen (Figure 3.1) appears when the instrument is started
with the indication “System Initialization”. After a successful initialization,
data loading and self-test, the VSWR measurement screen appears.
41
Figure 3.1 Boot Logo / Initialization Screen
3.1.2
System Information
Before using the instrument, verify the firmware’s version and status of the instrument.
Firmware version: For the instrument’s best performance, make
sure the latest version of firmware has been installed. Information about the latest firmware can be located at www.gsi-protek.com
Verify the system temperature is within the instrument operating
range. Depending on the storage condition, the temperature of the instrument at power up may be out of normal operating range in winter or summer season. Measurements over the operating temperature range may be out of resolution.
42
Figure 3.2 System Screen
3.2 Layout of the Display Screen
Figure 3.3 Display Screen
43
Indicates the instrument is using an external power supply. The internal battery is charged when an external power supply is connected.
Indicates the instrument is using the internal battery and shows the remaining battery capacity.
Battery Off Icon: This icon appears when a battery is not installed in the instrument.
Battery Gauge: Indicates the status of the internal battery
Measurement Mode: Indicates a current measurement mode. The
selected mode is displayed in dark blue.
Date and Time: Indicates the system clock information.
Screen Menu set: Indicates the set of screen menu selections.
Screen Menu: Indicates the selectable screen menu. Selection of
the menu can be made by pressing the soft key or touching the screen menu directly.
Trace Information
Calibration On/Off Status
Calibration Information: Calibrated frequency band and timing
Trace Points
Y Scale Unit
Trace Average (applicable to VSWR and Cable Loss
measurement mode only) before using the instrument, verify the firmware’s version and status of the instrument.
44
Span
Band Name
Cable Name (applicable to DTF measurement mode only)
Max D: Maximum measurement distance limited by user
frequency setting
VT (Relative Propagation Velocity), CL (Cable Loss) – applicable
to DTF measurement mode only
Message window: Displays the result of performed functions or
error messages.
3.3 Function Hard Keys Structure
Function hard keys on the front panel of the instrument are used to select measurement modes or perform specified functions. Refer to the following sections for the key structure to be used in each measurement modes.
Figure 3.4 Function Hard Key
45
Function Key Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
SYSTEM
Upgrade
Confirm
No
Beep
OFF ON
Sweep Mode
Conti
Sing
Instrument
Setting >>
Language >>
English
Return
Date/Time >>
Date Format
YMD MDY
DMY
Set Date
Set Time
Return
LAN >>
Apply
IP Addr
10.10.10.73
Net Mask
255.255.255.0
Gateway
10.10.10.1
Return
Alignment
3.3.1
System Screen Menu
System screen menu is used to change the basic settings of the instrument and consists of sub-menus as shown in Table 3.1.
46
Touch Screen
>>
Return
Sleep Time
Enter
Back Space
Cancel
Return
Preset
Yes
No
LED >>
10 Sec
OFF
10 Sec
20 Sec
Return
Function Key
Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
CAL
Open
Start
Back
Short
Start
Back
Load
Start
Back
VSWR
AMP
Max(Top)
Table 3.1 Menu Structure in system mode
3.3.2
Common Menu
The menu structure is common to all the measurement modes except to the Power Meter mode, shown in Table 3.2. Refer to section 3.3.6 for screen menus applicable to the Power Meter mode.
47
DTF
Cable Loss
2.00
Min(Bottom)
1.00
Limit
OFF ON
Limit Level
3.00
Dual Mode
Freq/
Dist
Start Freq
5.00 MHz
Stop Freq
4000.00 MHz
Center Freq
2002.50 MHz
Span Freq
3995.00MHz
Band List >>
Standard Band
>>
Select
Add to Custom
Return
Custom Band
>>
Select
Delete
Delete All
Return
Return
Trace Point
126
251
501
1001
2001
Marker
Marker
1 2 3 4 5 6
48
Marker View
OFF ON
Marker Type
Normal Delta
Marker
Clear All
Marker Edit
Marker Band >>
Marker Band 1
>>
Band View
OFF ON
Band Slect
Start Stop
Band Edit
Return
Marker Band 2
>>
Band View
OFF ON
Band Slect
Start Stop
Band Edit
Return
Marker Band 3
>>
Band View
OFF ON
Band Slect
Start Stop
Band Edit
Return
Clear All
Return
Trace
Trace
T1 T2 T3
T4
Trace Capture
49
Trace View
OFF ON
Clear Write
Trace Clear
Trace
Clear All
Peak
Peak Right
Peak Left
Max Search
Min Search
Load
Load Trace
(*.tra)
Select
File Manager >>
Rename
Delete
Delete All
Copy to USB
Copy All
to USB
Return
Destination >>
Trace 1
Trace 1
Trace 2
Trace 3
Trace 4
Return
Return
Load Screen
(*.jpg)
Select
File Manager >>
Rename
Delete
Delete All
Copy to USB
Copy All
to USB
50
Return
Return
Load Setup
(*.sta)
Select
File Manager >>
Rename
Delete
Delete All
Copy to USB
Copy All
to USB
Return
Return
Load From
Internal USB
Save
Save Trace
(*.tra)
Trace 1
Trace 2
Trace 3
Trace 4
Return
Save Screen
(*.png)
Save State
(*.sta)
Save to
Internal
USB
Light
Auto Scale
Scale
Hold
3.2 Menu Structure Common to all measurement mode
51
Function
Key
Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
VSWR
Freq/
Dist
Start Freq
(453.0 MHz)
Stop Freq
(464.0 MHz)
Center Freq
(458.5 MHz)
Span
(11.0 MHz)
Band List >>
(C450 P Up)
Startdard Band
Custom Band
Return
AMP
Max (Top)
0.00 dB
Min (Bottom)
40.00 dB
Limit
OFF ON
Limit Level
10.00 dB
Marker
Marker
1 2 3 4 5
6
Marker View
3.3.3
VSWR Menu
The menu structure of the VSWR hard function key and multi keys in VSWR measurement mode is shown in Table 3.3.
52
OFF ON
Marker Type
Normal
Delta
Marker Clear
All
Marker Edit
Marker Band >>
Marker Band 1
>>
Band View
OFF ON
Band Select
Start Stop
Band Edit
Return
Marker Band 2
>>
Band View
OFF ON
Band Select
Start Stop
Band Edit
Return
Marker Band 3
>>
Band View
OFF ON
Band Select
Start Stop
Band Edit
Return
Clear All
Return
3.3 Menu Structure Used for VSWR Measurement mode
53
Function Key
Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
DTF
Freq/
Dist
Start Freq
(453.0 MHz)
Stop Freq
(464.0 MHz)
Distance
(25.0 Meter)
Band List >
(C450 P Up)
Startdard Band
Custom Band
Return
Cable List >
(FSJ 250)
Startdard Band
Custom Band
Return
DTF Setting >
Apply
Velocity
Cable Loss
Unit
Meter
Feet
Windowing
>>
(Rectangular)
Rectangular
Blackman
Hamming
Hann
Return
3.3.4
DTF Menu
The menu structure of the DTF hard function key and multi keys in DTF measurement mode is shown in Table 3.4.
54
Return
AMP
Max (Top)
4
Min (Bottom)
1
Limit
OFF ON
Limit Level
1.01
Marker
Marker
1 2 3 4 5
6
Marker View
OFF ON
Marker Type
Normal
Delta
Marker Clear
All
Marker Edit
Function Key
Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
The menu structure of the Cable Loss hard function key and multi keys in Cable Loss measurement mode are shown in Table 3.5
3.4 Menu Structure Used for DTF Measurement Mode
3.3.5
Cable Loss Menu
55
Calbe Loss
Freq/
Dist
Start Freq
(453.0 MHz)
Stop Freq
(464.0 MHz)
Center Freq
(458.5 MHz)
Span
(11.0 MHz)
Band List >
(C450 P Up)
Startdard Band
Custom Band
Return
AMP
Max (Top)
0.00 dB
Min (Bottom)
40.00 dB
Limit
OFF ON
Limit Level
10.00 dB
Marker
Marker
1 2 3 4 5
6
Marker View
OFF ON
Marker Type
Normal
Delta
Marker Clear
All
Marker Edit
Marker Band >>
Band View
56
Marker Band 1
>>
OFF ON
Band Select
Start Stop
Band Edit
Return
Marker Band 2
>>
Band View
OFF ON
Band Select
Start Stop
Band Edit
Return
Marker Band 3
>>
Band View
OFF ON
Band Select
Start Stop
Band Edit
Return
Clear All
Return
Function Key
Screen Menu
1st Layer
2nd Layer
3rd Layer
4th Layer
Power Meter
Initialize/Preset
The menu structure of the Power Meter hard function key and multi keys in Power Meter is shown in Table 3.6.
Table 3.5 Menu Structure Used for Cable Loss Measurement Mode
3.3.6
Power Meter Menu
57
Frequency
Display Setup
>>
Unit
dBm Watts
External Offset
0.00 dB
Return
Mode >>
(Average)
AVG
Return
Table 3.6 Menu Structure Used in Power Meter Measurement Mode
3.4 System Key
The System key enables users to verify the system information or change the instrument settings.
Selecting the System key opens the following information:
58
Software Version
Device Version
Brightness of the display
Keypad beep On/Off setting
Sleep Mode setting: time to sleep mode
Battery remaining charge capacity
Keypad Backlight On time
Selected language
System temperature
Model / Serial Number
IP Address Info
Figure 3.5 System Status Screen
The System key opens the following screen menu:
Upgrade: Upgrades the firmware of the instrument. For detailed
upgrade procedure, refer to the section “5.Software Upgrade”.
Beep: Activates or deactivates the beep sound when keys are
pressed.
Sweep Mode: Sets the sweep mode either single or continuous in
VSWR, DTF, or Cable Loss measurements. If Sweep Mode is set Single, the message “Hold On” is displayed at the completion of a current sweep and the sweep stops. Every time the Hold key is pressed, the sweep is done once.
59
Instrument Setting
Language: Changes the language used to display the menu,
messages and information on the screen display. Currently only available in English.
Time/Date: Sets the time of the system clock.
LAN
Touch Screen
Sleep Time: Sets the time to enter into power saving mode.
Power saving mode is automatically activated when no key entry occurs during the Sleep Time.
Preset : Set Default Configuration Data
LED: Keypad Backlight On time. If no Keypad input is LED Off after
a set time.
3.5 Multi Keys
Twelve multi keys are located under the LCD display. The Multi keys serve multiple functions depending on the operation mode. The dual purpose keys are indicated in black color and the specific functions in blue color. The key is used to enter a numeric data when a user is prompted to input values. In all the other cases, the keys are used to perform the specific function.
The function and operating procedure for each multi keys are described in the following sections
60
Figure 3.6 Multi Keys Layout
3.5.1
Numeric Data Entry
Multi keys operate as follows when a user is prompted to input values by pop-up window.
Numeric keys from 0 to 9 are used to input user setting numbers. The plus/minus (+/-) key is used to assign positive or negative numbers in the user settings. The period key (.) is used to input numbers with decimal points.
3.5.2
Specific Function Keys
Multi keys are used in VSWR, DTF, and Cable Loss measurement mode. Unless otherwise specified, multi keys do not support Power Meter mode.
1. AMP
AMP (amplitude) defines a manual setting for the scale on the Y-axis of the graph. It can be selected in VSWR, DTF, and Cable Loss measurement mode.
61
Depending on the choice of an amplitude unit, allowable input value is limited to the following:
In case Y scale is VSWR:
Max(Top): Setting value can be from 1.01 to 65.00 and it
cannot be equal to or smaller than the Min(Bottom) setting value. Adjustment can be made by steps of 0.01.
Min(Bottom): Setting value can be from 1.0 to 64.99 and it
cannot be equal to or larger than the Max(Top) setting value. Adjustment can be made by steps of 0.01.
In case Y Scale is Return Loss:
Max(Top): Setting value can be from 0.0 to 59.99dB and it
cannot be equal to or smaller than the Min(Bottom) setting value. Adjustment can be made by steps of 0.01.
Min(Bottom): Setting value can be from 0.01 to 60.00dB and
it cannot be equal to or larger than the Max(Top) setting value. Adjustment can be made by steps of 0.01.
Limit line: Turns On and Off the display of the limit line on the
screen. If a measurement exceeds the limit line, the trace above the
limit line is displayed in red color.
An audible beep sound is generated.
Limit level: Sets the position of a limit line. Depending on the Y-
scale of a graph, the input value is set automatically as none for VSWR or dB for Return Loss. The value of limit level is displayed on the limit line.
The instrument takes the negative value of the Return Loss internally. Users don’t need to add minus (-) sign before typing in a value. Y-axis of a graph also doesn’t show minus sign.
62
2. Freq/Dist
Freq/Dist key causes different screen menu to be displayed depending on a measurement mode. In VSWR or Cable Loss measurement modes it opens a frequency screen menu and in DTF measurement mode it opens a distance screen menu.
VSWR/ Cable Loss Measurement Mode
Center: Sets a center frequency of the measurement to be
made.
Start: Sets a start frequency of the measurement to be made. Stop: Sets a stop frequency of the measurement to be made. Span: Sets a user-defined frequency span. Band: Opens standard or custom frequency band registered
in the instrument.
Standard: Opens the world-wide standard Band List
registered in the instrument.
- Top of List: Moves to the first page of the Band
List.
- Page Up: Moves to the previous page.
- Page Down: Moves to the next page.
- Bottom of List: Moves to the last page of the
Band List.
- Add to Custom: Copies a frequency band
registered in the Standard Band List to the Custom Band.
Custom: Opens the customized Band List registered in
the instrument.
- Top of List: Moves to the first page of the Band
List.
- Page Up: Moves to the previous page.
- Page Down: Moves to the next page.
- Bottom of List: Moves to the last page of the
Band List.
63
- Delete Custom: Deletes a frequency band from
DTF Measurement Mode
Start Frequency: Sets a starting frequency to measure DTF Stop Frequency: Sets a stop frequency to measure DTF Distance: Sets a distance to measure. The maximum
measurable distance is 1250m (4125ft).
Cable List: Opens a list of coaxial cables registered in the
instrument. About forty kinds of cables are registered in the standard cable list. The user can register additional cables to the instrument by using the application software program provided with the instrument.
Standard: Opens the Cable List registered in the
instrument.
- Top of List: Moves to the first page of the Cable
- Page Up: Moves to the previous page.
- Page Down: Moves to the next page.
- Bottom of List: Moves to the last page of the
- Add to Custom: Copies a cable registered in the
Custom: Opens a Custom Cable List registered in the
instrument.
- Top of List: Moves to the first page of the Cable
- Page Up: Moves to the previous page.
- Page Down: Moves to the next page.
- Bottom of List: Moves to the last page of the
- Delete Custom: Deletes a cable registered in the
Window: Applies video filtering to the display of the trace. If
the video filter is activated by turning the Window On, traces
the Custom Band.
List.
Cable List.
Standard Cable List to the Custom Cable List.
List.
Cable List.
Custom Cable List.
64
are reduced by smoothing out the sharp transitions, thereby enabling users easy to discriminate noises and peaks.
Setup
Velocity: Opens the Relative Propagation Velocity
parameter for data entry. Enter the propagation velocity for the type of transmission line being tested.
Cable Loss: Opens the Cable Loss parameter for data
entry. Enter the loss per meter for the type of transmission line being tested.
Unit: Selects the unit of X-axis scale to display the
measurement results in Meter or Feet.
3. Trace Point
Trace point is used to select the number of data points to take during a VSWR and Cable Loss measurement mode. There are four data points available: 126, 251, 501 and 1001. The default number of trace points is
1001.
4. Marker
A marker is used to get the data on the specific point of a trace. A total of six markers can be displayed on the screen and each marker can be assigned independently. Placing a marker on the trace displays the marker’s Y coordinates next to the marker’s position. If the marker table is turned on, both X and Y coordinates of all activated markers are displayed.
Marker
Select: Selects an Active Marker whose position is moved by
the dial knob or Up and Down arrow keys. The assigned number of an active marker is displayed in red color on the
Select screen menu and the marker’s number is also displayed
next to the marker on the trace.
Marker View: Hides or displays the selected marker on the
screen. In the same measurement mode markers appear at
65
the previous positions when the Marker View is turned off and on. If a measurement mode has been changed, markers are not restored to previous positions but move to the left end of the trace.
All Off: Turns all markers off the screen. Markers are
redisplayed on the previous position if markers are turned back on. If a measurement mode is changed, current settings are not restored.
Draw Type: Selects the maximum number of markers to be
used simultaneously. Available options are either 4CH or 6CH. If 4CH mode is selected, markers number 5 and number 6 cannot be activated. Settings changed in Draw Type are reflected in Marker CH and Marker table.
Marker Table: Displays a table on the left side of the screen
when Marker Table is activated. All X and Y coordinates of the activated markers are displayed on the table. Depending on the selection in Draw Type, the number of markers on the table is either 4 or 6.
Marker Edit: Sets the marker position manually. A pop-up
window appears for users to set the frequency and the marker position is moved to the setting frequency
Moving Markers
Knob – Turning the knob clockwise moves a marker to the
right and counter clockwise moves it to the left. The knob is used to move the marker position fast.
Up/Down Arrow Keys – Pressing the up arrow key ()
moves a marker’s position one point to the right and pressing the down arrow key (↓) moves a marker’s position one point to the left. Up/Down Arrow keys are used to move a marker’s position precisely.
As the instrument is equipped with a touch panel operation, a marker can be placed to the desired position by simply
66
touching the screen. Quickly moves an activated marker to the desired position and makes a fine adjustment using a dial knob or up/down arrow keys.
5. Trace
Captures a trace for comparison with other traces or saves traces.
Figure 3.7 Trace Screen Display
Select: Selects an active trace. Every time Select screen menu is
pressed, the active trace changes. Channel numbers are assigned to each captured traces or loaded traces. Refer to Save & Load” for procedures to load traces.
Capture: Captures a current trace on the screen and assigns a
Trace CH. Refer to “Save & Load” for procedures to save traces.
View: Hides or displays the Trace CH on the screen. Press Select
screen menu key to choose the Trace CH. Traces with View set OFF are hidden from the screen. The information about the hidden traces is also cleared from the information window shown on the upper right of the screen. Setting View On restores hidden traces and information on the window.
67
Info: Hides or displays the trace information window on the upper
right corner of the screen. This is used to view traces overlaid with the trace information window.
Clear CH: Deletes an active trace channel from the screen. The
cleared channel is not restored. It is used to select and delete a trace channel one by one when multiple channels are displayed on the screen. Verify the channels to delete with View ON/OFF function in advance settings as cleared channels cannot be restored.
Clear All: Deletes all channels from the instrument and initialize
the trace settings.
6. Peak
This key is used to find the peak value of a trace. Pressing this key leads to the activation of Marker CH1 and places the marker to the peak point of the trace.
Peak Right: Moves a marker to the nearest peak on the right. Peak Left: Moves a marker to the nearest peak on the left. Max Search: Moves a marker to the highest point of the trace. Min Search: Moves a marker to the lowest point of the trace.
7. Load
Load key enables for recalling of the Trace, Screen, and State
8. Save
Save key enables for saving of the Trace, Screen, and State.
9. Light
This key is used to adjust the brightness of the LCD display. Adjustment can be made from 0 to 100% and the default setting is 50%.
68
10. Scale
This key is used to select a unit of Y axis of the graph in VSWR and DTF measurement modes. Either VSWR or Return Loss can be selected. In Cable Loss measurement mode, Return Loss only can be selected.
11. Auto Scale
The instrument can automatically set the scale to the minimum and maximum values of a measurement on the Y-axis of the graph for optimum display of the traces. Every time AUTO SCALE key is pressed, the top and bottom scales are set to the minimum and maximum values with margin on the Y-axis of the screen display.
12. Hold
This key is used to pause a sweep in all measurement modes. The
Hold state is activated by pressing the Hold key, and it is maintained even if users change the measurement mode. The sweep resumes when the Hold key is pressed again.
When Sweep Mode is set Single in the System screen menu, a Hold
message is displayed on the screen in red color and the measurement stops at the completion of a single sweep. Pressing the Hold key it triggers another single sweep.
69
4
MEASUREMENT DATA ANALYSIS
OPERATING MANUAL
4.
4.1 VSWR
To get maximum power into a load it is required that the load impedance match the generator impedance. Any impedance difference, or mismatching would not produce maximum power transfer. A mismatch at the antenna system produces a reflective 'traveling wave' which goes in the opposite direction from the incident wave. As the two traveling waves cross each other in opposite direction, it produces an interference pattern called a "standing wave". VSWR is the ratio between the power that is sent forward to the antenna and the amount of the power that is reflected back toward the transmitter.
Some of the consequences of having a high VSWR condition in cellular services are: dropped calls, poor reception, and an overall unacceptable performance in the cell (or section of cell) covered by the base station antenna. Therefore, the VSWR of the antenna system including the feed line is one of the most critical factors in the service and maintenance of the RF transmitter systems.
4.1.1
VSWR Display Description
The screen shown in Figure 4.1 is displayed when VSWR measurement mode is selected. Various kinds of information related to the VSWR measurement are shown on the screen.
71
Figure 4.1 VSWR Measurement Screen Display
Calibration Info: Displays a calibration state on the measurement frequency
band that a user has selected. When the instrument is first turned on, the
state is “Cal Off”. The symbol “Cal On” is displayed along with the execution
time and frequency band after the calibration is successfully completed.
Trace Point: Sets the number of data points to take during a measurement
mode. Selection is made from one of five choices.
126, 251, 5001, 1001, 2001
Selecting 501 data points provides twice as many measurement points as
251, but it takes approximately twice as long for the trace to sweep and
display.
Selecting the trace points larger than needed for the measurement will result in a longer sweep time. This may hinder the user from observing the fast changing traces.
72
It is recommended to select high resolution trace points only when measuring the wide frequency band or a precise measurement is required.
The previously executed calibration is effective even after the trace points change.
Trace Average: Indicates the average value of a single sweep over the user
setting frequency band.
Freq. Span: It is a user-defined frequency band. Changing the frequency
band doesn’t affect the sweep time, but affects the calibration. Recalibration
is required if the frequency setting is changed.
Freq Band Info: The band name is displayed if the band is selected from the
band list stored in the instrument. If the user sets the start, stop, center or
span frequency manually, the band name will show “Custom”
Limit Line: Sets the upper limit value of the trace. The portion of the trace
that exceeds the limit line is displayed in red color. The captured trace by
using the Trace function does not display the exceeded portion of the trace
in different color.
Y scale unit: It is the measurement unit of the Y axis displayed for the trace.
Return Loss (dB)
VSWR
73
Action
Note
Setting Center Freq and Span.
1. Press the Freq/Dist key.
Multi key.
2. Select the
Center Freq
screen menu.
The current setting is displayed
on the Window.
3. Enter a center frequency value.
4. Press the Enter key.
5. Select the
Span
screen menu.
6. Enter a span value.
7. Press the Enter key.
Setting Start/ Stop Frequency.
1. Press the Freq/Dist key.
The current setting is cleared
when a new value is entered.
Press the ESC key to delete one
by one the numbers displayed on the pop-up window.
Pressing the ESC key repeatedly
will cancel the input mode.
The frequency input unit is in
MHz and the minimum input steps is 0.01MHz
2. Select the
Start
screen menu.
3. Enter a start frequency value.
4. Press the Enter key.
5. Select the
Stop
screen menu.
6. Enter a stop frequency value.
7. Press the Enter key.
Selection from the band list stored in the instrument.
1. Press the Freq/Dist key.
Press the Up() / Down() arrow
keys or rotate the dial knob to select a band from the list.
Select the
Page Up/Page Down
screen menu for searching bands not shown in the screen.
2. Select the
Band
screen menu.
3. Select the band from the list and press the Enter key.
4.1.2
Setting Frequency
Frequencies can be set manually or selected from a band list stored in the instrument. It is desirable to set the frequency to a value that covers the normal range of the measurement with enough margin.
74
Table 4.1 Frequency Setting Procedure
Changing the frequency settings will automatically turn calibration off with the symbol “CAL OFF” displayed on the screen. Always set the frequency before calibrating the instrument. Changing the trace points during the
measurement doesn’t affect the calibration.
4.1.3
Calibration
The instrument must be calibrated to get a reliable measurement result. For best results, set the frequency and calibrate the instrument immediately before taking a measurement.
Calibration accessories (optional)
Calibration Kit which contains one 50ohm load, one Open
standard and one Short standard
Test cable: Use a phase stable cable for reliable and
consistent measurement results
To minimize the measurement error, connect the port extension cable to the RF In port on the instrument and then connect the Cal Kit to the end of the extension cable.
75
Figure 4.2 VSWR Calibration Connection
Bending or moving the phase unstable cable while making a measurement may cause errors in the measurement. The test cable used for port extension must be phase stable in the measurement frequencies.
At the successful completion of each calibration step, the message is displayed with a beep sound.
Figure 4.2 illustrates the connection method when a port extension cable is used for calibration.
76
Action
Note
Performs Calibration after the frequency setting and test cable connections.
1. Press the CAL key.
Hard function key.
2. Connect an Open standard.
Connect CAL Kit “Open” connector to the RF out port.
3. Select the
Open/Start
screen menu.
When the
Open/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Open calibration
is finished.
4. Connect a Short standard.
Connect CAL Kit “Short” connector to the RF out port.
5. Select the
Short/Start
screen menu.
When the
Short/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Short calibration
is finished.
6. Connect the Load standard.
Connect CAL Kit “Load” connector to the RF out port.
7. Select the
Load/Start
screen
menu.
When the
Load/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Load calibration is
finished.
Calibration state is changed to “Cal On after the Open-Short-Load calibration.
To compensate errors caused by a port extension cable or adapters, perform Open-Short-Load calibration including the port extension cable.
Table 4.2 Calibration Procedure
4.1.4
VSWR Measurement
The instrument is ready to take VSWR measurement after completing the Open-Short-Load calibration using a port extension cable.
77
After calibration, do not change the connection of the port extension cable or the frequency setting. It can cause produce an error in the measurement. When the frequency setting is changed, a warning alarm will sound and the calibration state is changed to “Cal Off”. After changing the frequency setting, recalibrate the instrument using the Open­Short-Load standard.
The end of the port extension cable must be connected to the device (antenna or feed line) for VSWR measurement as shown in Figure 4.3. The result of the VSWR measurement is displayed on the screen in real time.
Figure 4.3 Connection for VSWR Measurement
78
The maximum allowable input level of the instrument is +25dBm. Do not connect the RF In port of the instrument directly to the output port of the system. An over power input degrades the performance of the instrument and may cause a malfunction of the instrument.
Do not connect the instrument to the antenna when there is a risk of lightning. Electric shock may cause a malfunction or damage the instrument.
4.1.4.1 Setting Trace Point
Adjust a trace point to change the resolution of the VSWR measurement. Changing the Trace Point doesn’t affect the calibration state.
4.1.4.2 Scale Adjustment
Press Auto Scale key to optimize the Y scale and display the entire trace. Press AMP key to set the maximum and minimum values on the Y scale
manually.
Press Scale key to select the display unit of the Y scale. Scale adjustment doesn’t affect the calibration state.
4.1.4.3 Using Markers
Depending on the draw type setting, four or six markers can be used simultaneously. The Y axis value on a current trace is displayed next to each marker on the screen, but the frequency information is not provided. Activate the marker table to see both, the level and frequency information when moving a marker to the specific frequency using the Marker Edit function.
79
Figure 4.4 Using Markers in VSWR Measurement Mode
4.1.4.4 Limit Line
By setting a limit line, users can easily check if a measurement exceeds the specified limit. It appears as a horizontal line at the value set. An alarm sounds when a trace exceeds the limit line and the exceeded portion is displayed in red color.
80
Figure 4.5 Limit Line Application
4.2 DTF
While VSWR is an indicator to express the efficiency of the antenna’s energy
transmission, DTF is a measurement to identify the fault locations in the antenna line system. Most of the antenna line system consists of the various types of coaxial cables, connectors and devices such as dividers and surge arrestors. Since VSWR is a measurement to verify the impedance discontinuity of the total feed line system, it is necessary to perform DTF measurement to identify the exact component that is contributing to the performance of the line system. The DTF measurement makes it easy to identify the fault location by displaying the relative distance of the signal reflections or discontinuities from various points of the line system
4.2.1
DTF Display Description
The screen shown in Figure 4.6 is displayed when DTF measurement mode is selected. The distance from the instrument is shown on the X-axis, while
81
the relative magnitude of the discontinuity is shown on the Y-axis. The information related to the DTF measurement is shown on the screen.
Figure 4.6 DTF Measurement Screen
Limit Line & Limit Level Info: Sets the upper limit value of a trace.
The portion of the trace that exceeds the limit line is displayed in red color. However, the captured trace by using TRACE function
doesn’t discriminate the color even if the trace exceeds the limit
line.
Calibration Info: Displays a calibration state on the measurement
frequency band that a user has selected. When the instrument is first turned on, the state is “Cal Off”. The symbol “Cal On” is displayed along with the execution time and frequency band after the calibration is successfully completed.
Maximum Distance: Displays the maximum measurable distance
within the user setting frequency band. Setting the frequency band narrow will increase the measurable distance while setting the frequency band wide will decrease the distance.
Y Scale Unit: Is the measurement unit of the Y-axis that the trace
is displayed.
Return Loss (dB) VSWR
Freq. Span: Is the user-defined frequency band. Changing the
82
frequency band doesn’t affect the sweep time, but affects the calibration. Recalibration is required if the frequency setting is changed.
Freq. Band Info: The band name is displayed if the band is selected
from the band list stored in the instrument. If the user sets the start, stop, center or span frequency manually, the band name will show “Custom”.
Cable Info: The name of user selected cable is displayed on the
screen. A cable name is displayed if the cable is selected from a Cable List stored in the instrument. If the user sets the Velocity and Cable Loss manually, the band name will show “Custom”. The following information is also displayed. Rel. Propa. Vel.: The relative propagation velocity for the
cable type selected by the user selected from the Cable List or manually set by selecting the Velocity key.
Nominal Atten.: The loss per unit distance for the cable type
that the user selected from the Cable List or set manually by selecting the Cable Loss key
By using the application program supplied with the
instrument, users can store user’s cable to the instrument. For
details, refer to the Application Program.
4.2.2
DTF Setup
Set the conditions for DTF measurements. The user setting parameters for DTF measurements are the following:
Frequency Setting: Sets the start and stop frequency to make a
measurement. If a specific frequency band has been set in VSWR measurement mode, it can be applied to DTF measurement. To change the maximum measurement distance or increase the
83
Action
Note
Frequency Setting.
Press the Freq/Dist key.
Select the
Start Freq
screen menu. {Enter start frequency value} Press the Enter key.
Additional calibration is
not necessary if a Freq Band has been set and a calibration has been
measurement resolution, is necessary to change the frequency setting.
Distance Setting: The maximum measurable distance is displayed
on the left side of the screen depending on the frequency setting. Any value within the maximum measurable distance can be set. Optimum resolution is achieved when the user setting distance is the same as the maximum measurable distance.
Cable Setting: Selects a cable type of the feed line. By using this
key, users can select the cable stored in the instrument without setting the detailed parameters of the cable.
Setup: Used to change the setting of the cable parameters or
change the distance unit. It consists of the following sub menus: Velocity: Sets the relative propagation delay of a cable. It
affects the calculation of the distance in the DTF measurement.
Cable Loss: Sets the loss per distance unit of a cable. It affects
the peak level of the discontinuity in the DTF measurement.
After calibration, do not change the connection of the port extension
cable or the frequency setting. It can cause a
frequency setting, recalibrate
A detailed procedure for DTF setup is as follows:
84
measurement error. When the frequency setting is
changed, a warning alarm will sound and the calibration
state is changed to “Cal Off”. After changing the
the instrument using the Open-Short-Load standard.
Select the
Stop Freq
screen menu. {Enter stop frequency value} Press the Enter key.
performed for the band in VSWR measurement, and the same Freq. band is used in the DTF measurement.
Distance Setting.
Select the
Distance
screen menu.
{Enter measuring distance} Press the Enter key.
The ending point can
only be set in distance setting.
The maximum
measurable distance is 1,250m (4,125ft).
Cable Setting.
Select the
Cable List
screen menu.
[Standard]/[Custom]
{Select a cable by using Knob or arrow key}
Press the Enter key.
Setup.
Setting Relative Propagation Velocity.
Select the
Velocity
screen menu.
{Enter user setting value} Press the Enter key.
Setting Cable Loss.
Select the
Cable Loss
screen menu.
{Enter user setting value} Press the Enter key.
Setting the X axis Unit.
[Meter]/[Feet]
4.2.3
Calibration
The instrument must be calibrated to get the DTF measurement results compatible with VSWR
Table 4.3 DTF Setup Procedure
85
Calibration accessories (optional)
Calibration Kit which contains one 50ohm load, one Open
standard and one Short standard
Test cable: Use a phase stable cable for reliable and consistent
measurement results
To minimize measurement errors, connect the port extension cable
to the RF In port on the instrument and then connect the Cal Kit to
the end of the extension cable.
The Figure 4.7 shows the connection diagram for calibration using a test cable. To compensate measurement errors due to the test cable or adapters, perform the Open-Short-Load (O-S-L) calibration including the test cable. For detailed calibration procedure, refer to Table 4.2 Calibration Procedure
86
Figure 4.7 DTF Calibration Connection Diagram
The maximum allowable input level of the instrument is +25dBm. Do not connect the RF In port directly to the system output port. Exposure to the overpowered input may degrade the performance of the instrument and may cause damage in the long run.
Do not connect the instrument to the antenna when there is a risk of lightning. Electric shock may cause the malfunction or breakdown of the instrument.
Bending or moving the phase unstable cable while making a measurement may cause errors in the measurement. The test cable used for port extension must be phase stable in the
measurement frequencies.
At the successful completion of each calibration step, a message is displayed with a beep sound.
4.2.4
DTF Measurement
If a port extension cable is used to interconnect the instrument with the line, measurement error can happen due to the sum of the port extension cable length and the distance to the fault. By performing the O-S-L calibration at the end of the port extension cable, the extension cable length will be compensated and the fault location can be more accurately measurable.
87
If O-S-L calibration has been done at the end of the port extension cable for DTF measurement, the length of the port extension cable is compensated automatically and is not included in the distance to the point of discontinuity.
Figure 4.8 Connection Diagram for DTF Measurement
4.2.4.1 Scale Adjustment
Adjust a trace point to change the resolution of the VSWR measurement. Changing the Trace Point doesn’t affect the calibration state.
Press the Auto Scale key to optimize the Y scale and display an
entire trace.
Press the AMP key to set the maximum and minimum values on
88
the Y scale manually.
Press the Scale key to select a display unit of the Y scale.
4.2.4.2 Using Markers
Depending on the display settings, four or six markers can be used simultaneously. The distance of a current marker position is displayed next to each marker on the screen. Activate the marker table to see both the magnitude and distance information. Use Marker Edit to move a marker to the specific distance.
Adjust a trace point to change the resolution of the VSWR measurement. Changing the Trace Point doesn’t affect the calibration state.
Figure 4.9 Using Markers in DTF Measurement Mode
4.2.4.3 Windowing
Windowing is applied when an accurate fault location cannot be verified due to the side overshoots of the trace. If the video filter is activated by turning the Window On, side overshoots of the trace are reduced by
89
smoothing out the sharp transitions thereby enabling users easy to discriminate noises and peaks. Figure 4.11 is the result of applying Blackman window to the trace on Figure 4.10. Noises around peaks are reduced and distance to the fault location is clearly verified.
90
Figure 4.10 Measurement Display Rectangular Windowing
Figure 4.11 Measurement Display Blackman Windowing
4.3 Cable Loss
The cable loss measurement feature checks the signal attenuation level of the cable system. The frequency band to measure the characteristics of a cable must be calibrated before performing the cable loss measurement.
4.3.1
Cable Loss Display Description
The screen shown in Figure 4.12 is displayed when the Cable Loss measurement mode is selected. Various kinds of information related to the Cable Loss measurement are shown on the screen.
Figure 4.12 Cable Loss Measurement Screen Display
Calibration Info: Displays the calibration state of the measurement
frequency band that a user has selected. When the instrument is first turned on, the status is “Cal Off”. The symbol “Cal On” is displayed along with the calibration time and frequency band after successful completion of calibration.
Trace Point: Sets the data points or resolution to measure the trace.
91
Action
Note
Setting Center Freq and Span.
1. Press the Freq/Dist key.
Multi key.
2. Select the
Center Freq
screen menu.
Current setting is displayed as
default on the screen.
3. Enter a center frequency value.
4. Press the Enter key.
5. Select the
Span
screen menu.
6. Enter a span value.
7. Press the Enter key.
Setting Start/ Stop Frequency.
1. Press the Freq/Dist key.
Current setting is cleared when a
new value is entered.
Press the ESC key to delete a
number displayed on the pop-up
2. Select the
Start Freq
screen menu.
3. Enter a start frequency value.
4. Press the Enter key.
Selection is made from one of five choices.
126, 251, 501, 1001, 2001
Trace Average: Indicates the average value of a single sweep over
the user setting frequency band.
Freq. Span: User-defined frequency band. Changing the frequency
band doesn’t affect the sweep time, but affects the calibration. Recalibration is required if the frequency setting is changed.
Freq. Band Info: The band name is displayed if the band is selected
from the band list registered in the instrument. If the user sets the start, stop, center or span frequency manually, the band name will indicate “Custom
Y scale unit: The unit of the y axis in the Cable Loss measurement
mode is dB.
4.3.2
Setting Frequency
The user must set the frequency band to make a Cable Loss measurement. Frequency can be set manually or chosen from a band list registered in the instrument.
92
5. Select the
Stop Freq
screen menu.
window one by one.
Pressing the ESC repeatedly will
cancel the input mode.
Input unit is MHz and minimum
input step is 0.01MHz.
6. Enter a stop frequency value.
7. Press the Enter key.
Selection from the band list stored in the instrument.
1. Press the Freq/Dist key.
Select the
Page Up/Page Down
screen menu for searching the band that doesn’t show up on the current page.
2. Select the
Band List
screen menu.
3. Select the band from the list and press the Enter key.
Table 4.4 Frequency Setting Procedure
Changing the frequency settings will automatically turn calibration off with the symbol “Cal Off” to be displayed on the screen along with an alarm sound. In this case,
recalibrate the instrument with the Open-Short-Load Calibration kit.
4.3.3
Calibration
The instrument must be calibrated to get reliable Cable Loss measurement results. For best results, set the frequency and calibrate the instrument immediately before taking measurements.
Calibration accessories (optional)
Calibration Kit which contains one 50ohm load, one Open
standard and one Short standard
To minimize the measurement error in Cable Loss measurement, don’t
use unnecessary extension cables or adapters while performing calibration. Figure 4.13 illustrates the recommended calibration method for Cable Loss measurement.
93
Action
Note
Performs Calibration after the frequency setting.
1. Press the CAL key.
Hard function key.
2. Connect an Open standard.
Connect CAL Kit “Open” connector to the RF out port.
3. Select the
Open/Start
screen
menu.
When the
Open/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Open
calibration is finished.
4. Connect a Short standard.
Connect CAL Kit “Short” connector to the RF out port.
5. Select the
Short/Start
screen
menu.
When the
Short/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Short
calibration is finished.
Figure 4.13 Port Calibration for One Port Cable Loss Measurement
94
6. Connect the Load standard.
Connect CAL Kit “Load” connector to the RF out port.
7. Select the
Load/Start
screen
menu.
When the
Load/Start
screen menu is selected, a
progress bar is displayed to show the progress.
When a progress bar is disappeared, Load
calibration is finished.
Calibration state is changed to “Cal On after the Open-Short-Load calibration.
Table 4.5 Calibration Procedure
4.3.4
Cable Loss Measurement
The instrument is ready to perform Cable Loss measurement after completion of Open-Short-Load calibration.
Figure 4.14 One Port Cable Loss Measurement Connection Diagram
The instrument must be calibrated to get reliable Cable Loss measurement results. For best results, set the frequency and calibrate the instrument immediately before taking measurements.
95
Action
Note
Make a measurement after completion of O-S-L calibration.
1. Connect the cable to measure its loss to the RF out port of the instrument.
2. Connect the Short standard of the Cal Kit to the end of the cable to be tested
Cable Loss measurement result is displayed on the screen.
Table 4.6 Cable Loss Measurement Procedure
4.3.4.1 Scale Adjustment
Press the Auto Scale key to optimize the Y scale and display an
entire trace.
Press the AMP key to set the maximum and minimum values on
the Y scale manually.
Press the Scale key to select a display unit of the Y scale.
4.3.4.2 Using Markers
Depending on the draw type setting, four or six markers can be used simultaneously. The level of a current trace is displayed next to each marker on the screen, but the frequency information is not provided. Activate a marker table to see both level and frequency information when moving a marker to the specific frequency using the Marker Edit.
96
Figure 4.15 Using Markers in Cable Loss Measurement Mode
4.4 Power Meter
The Power Meter is to measure the transmission power of the system. This function can be used only with optional external power sensors. Two kinds of power sensors, the average power sensor and peak power sensor, are available depending on the type of transmission power signals to be measured.
4.4.1 Connecting Sensors
Selecting the Power Meter function after power up will display Power Meter measurement screen with the message that “Please, connect external power sensor and press Initialize menu key” is connected.
97
Figure 4.16 Power Measurement Screen
Connect a power sensor to the USB interface port of the instrument using the provided cable as Figure 4.17. Do not connect a power sensor directly to the LPA or HPA.
Figure 4.17 Power Sensor Connection Diagram
After connecting a power sensor, select the the instrument to recognize the sensor. After successful recognition of the
Initialize
screen menu key for
98
power sensor, the sensor type is displayed on the screen as in Figure 4.18. Since no power source is connected to the sensor yet, an alarm message “-
- - -” is displayed on the screen
Figure 4.18 Power Sensor Initial Screen
No sensor type is displayed on the screen if the instrument fails to recognize a sensor type during the initialization process.
4.4.2 Setting Power Meter
All the keys used to set the power measurement are displayed as screen menu keys on the screen.
Initialize
calibration data from the sensor.
Frequency
the frequency setting affects the calibration data, be sure to set
the accurate center frequency for reliable measurement results.
: Initializes the power sensor and downloads the
: Sets the frequency of signals to measure. Note: As
99
Do not connect the power sensor directly to the output of the HPA. The power sensor will be damaged if output power greater than +10dBm is supplied directly.
Ext. Offset
or an amplifier is used before the power sensor. Note: As the
default setting is Loss, enter the level of Loss in positive value
when an attenuator is used and enter the level of gain in
negative value when an amplified is used.
Unit
units are selected as Watts, depending on the measured power
level, the units are altered as W(Watt), mW(milli Watt), or
uW(micro Watt) automatically.
Preset
: Enters the value of Gain or Loss when an attenuator
: Selects the measurement unit. Note: In the case where the
: Clears all user settings and returns to initial settings.
4.4.3 Power Measurement
After connection and initialization of a power sensor, connect the power sensor to the output port of the device
Figure 4.19 HPA (High Power Amplifier) Output Power Measurement
100
Loading...