ProteinSimple Maurice CE-SDS Application Manual

Maurice® CE-SDS Application Guide
Maurice CE-SDS Application Guide
Introduction
The CE-SDS Application Kit is a one-stop shop for all your size application development needs! It includes the CE-SDS cartridges, all sample preparation reagents and consumables that’ll let you tackle any monoclonal antibody. This Application Guide will help you every step of the way.
If you add the IgG Standard, which is produced with a known quantity of non-glycosylated heavy chain, you can test resolution and quantitation suitability to make sure your Maurice is ready to go.
CE-SDS Size Application Kit contents (P/N PS-MAK02-S)
INCLUDES QTY/KIT
Maurice CE-SDS Cartridges 2
Maurice CE-SDS Cartridge Cleaning Vials 2
Maurice clear screw caps for sample and reagent vials 100
Maurice glass reagent vials, 2 mL 100
Maurice CE-SDS orange pressure caps 12
Maurice 96-well plates 10
Maurice CE-SDS Running Buer – Top 10 vials
Maurice CE-SDS Wash Solution 2 x 20 mL
Storage conditions
Store the 25X Internal Standard (P/N 046-144) at 2-8 °C.
Everything else should be stored at room temperature.
Separation and Running buer bottles should be tightly
closed to prevent evaporation.
Running Buer
Top vials must be stored tightly closed
in the original container with two humidifying pouches.
Ordering info
This kit can be reordered by:
Phone: 1-888-607-9692, option 1
Fax: 1-408-520-4831
Email: orders@proteinsimple.com
Other things you’ll need
Maurice CE-SDS IgG Standard, PN046-039
Maurice CE-SDS Molecular Weight Markers, PN046-432
(optional)
Maurice sample vials with integrated inserts, 0.2mL,
PN046-083 (optional)
β-mercaptoethanol (>98% = 14.2 M) for reducing
conditions
Iodoacetamide (250 mM) for alkylation at non-reducing
conditions
Deionized (DI) water
Maurice CE-SDS 25X Internal Standard 2 vials
Pipettes and tips
Microcentrifuge and tubes
Ice and ice bucket
Vortexer
Water bath or thermocycler
Centrifuge with plate adapter or vial adapter (12mm,
2mL vials)
2
Maurice CE-SDS Application Guide
Guidelines for great results
Make sure you read the entire Application Guide before
getting started.
Use fresh Conditioning Solutions, Separation Matrix, and
Wash Solution for each batch.
When the Top Running Buer vial is still in the cartridge
insert, the cartridge MUST be kept in an upright position at all times.
The 25X Internal Standard, IgG Standard, and CE-SDS
Molecular Weight Markers are lyophilized. Always reseal the foil bag containing the unopened tubes with desiccant to prevent moisture absorption.
Aliquot the reconstituted 25X Internal Standard solution
into appropriately sized vials and store at -80 °C for long-term storage. For short-term storage (<1 week), the solution can be stored at 2–8 °C.
If you observe any precipitation in the 25X Internal
Standard solution, then leave the solution at room temperature and stir gently until the precipitates have dissolved completely.
Whenever you handle the cartridge or remove it from
its packaging, make sure the cartridge inlet doesn’t come in contact with any surfaces. A damaged inlet may compromise the cartridge and cause a failed injection.
Always perform the cartridge cleanup before storing,
and always store the cartridge in its original packaging at room temperature.
If you see any Separation Matrix sticking to the cartridge
inlet, soak the inlet for 5 min in DI water. Then wipe the inlet using a lint-free laboratory wipe that’s been moistened with DI water.
Each cartridge is guaranteed for up to 100 injections and
maxium of 25 batches. The injection limit of the cartridge is 200. Its RFID will keep track of how many injections have been performed and how many are left for you.
Don’t reuse reagents, vials or the clear screw caps.
Always keep the orange pressure caps paired with their respective reagents.
Orange pressure caps should be used for the CE-SDS
application only.
If you want to reuse the orange pressure caps, wash
them thoroughly with DI water, soak them overnight in DI water, rinse caps thoroughly with DI water again and air dry them before reusing.
Application overview
A successfully dened and optimized CE-SDS method gives you:
A highly reproducible peak prole
Reproducible relative migration times for each peak
NGHC/HC baseline peak resolution (≥1.5) for the reduced
IgG sample
3
Maurice CE-SDS Application Guide
Step 1:
Step 2:
Step3:
Run ProteinSimple reduced IgG Standard
Pass
Run sample
Yes
Current ~25 µA?
Yes
All peaks detected?
Yes
Fail
No
No
Call ProteinSimple
Tech Support
• Make sure separation reagents are in correct vial positions
• Always use fresh reagents
• Optimize salt and protein concentration in the sample
• Verify denaturing conditions
• Optimize injection and separation parameters (time and voltage)
Resolution sucient?
Yes
Method development
is complete
FIGURE 1. Method development strategy.
No
• Check conditioning reagents
• Perform a cartridge cleanup
• Optimize injection parameters (time and voltage)
4