WARNINGWARNING
You are cautioned that changes or modifications not espressly
approved in this document could void yout authority to operate this
equipment.
To reduce the risk of fire or electric shock,do not expose this
apparatus to rain or moisture.
VISIBLE LASER RADIATIONVISIBLE LASER RADIATION
AVOD EYE OR SKIN EXPOSUREAVOD EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED RADIATIONTO DIRECT OR SCATTERED RADIATION
AVOD EYE OR SKIN EXPOSUREAVOD EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED RADIATIONTO DIRECT OR SCATTERED RADIATION
SEMICONDUCTOR LASERSEMICONDUCTOR LASER
Wavelength:800-1700nmWavelength:800-1700nm
CLASS I LASER PRODUCTCLASS I LASER PRODUCT
To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
NOTENOTE
As the laser is harmful to the eyes, do not attempt to disassemble the cabinet.
Precautions for Use
Use batteries
At the same time, can not use different style or different capacitance batteries. And only charge the rechargeable batteries.
Avoiding condensation problems
As much as possible, avoid sudden temperature changes. Do not attempt to use the drive immediately after moving it
from a cold to a warm location, to raising the room temperature suddenly, as condensation may form with in the drive. If the
temperature changes suddenly while using the drive,
Stop using it and take out batteries for at least an hour.
Storage
When long time no use,must take out the batteries to avoid destroying the device.
1
Contents
1. Warning and Note
2. Standard
3. Product presentation
3.1 Description
3.2 On/off
3.2 Measurement Interface
3.3 Real-time and average measurement
3.4 Event list
3.5 Real-time measurement tips
3.6 Cursor selection and waveform zoom in/out
3.5 Menu
3.5.1 Menu-File menu
3.5.2 Menu-Measure settings
3.5.3 Menu-System settings
3.5.4 Menu-Module settings
3.5.5 Menu- About
3.6 TF card and USB communication20
3.7 User upgrading
3.8 Charging
3.9 Detail parameters
1
3
4
4
5
6
7
8
9
10
11
12
16
17
18
19
21
22
23
Standard
Host(with TF card)
Lithium battery
Power adapter
Carrying bag
Power cable
3
USB cable
18
19
16
1
2
Description
23
20
3
21
22
4
5
12
1314
15
6
7
8
9
10
11
17
4
OTDR optic fiber connector
1
VFL optic fiber connector
2
3
LCD
4
Zoom control button
5
Full screen button
6
Menu button
7
Power button
Confirming button
8
9
Cancel button
10
Real-time measurement button
Average measurement button
11
Up button
12
13
Left button
Right button
14
15
Down button
File operation button
16
17
Cursor select button
Power adapter socket
18
Charge indicator
19
20
Anti-dust cover
21
USB interface
TF(micro SD) card slot
22
Support frame
23
MT-7601MT-7601
Optical Time Domain Reflectometer
MENU
File
SCAN
Marker
On/off
Button is used to turn on/off the device. Keep pressing it for 2 seconds
to turn on device. Short press it again to turn off the device.
Attention: formatting SD card operation will delete all files and cannot recover, please be cautious.
Are you sure to format the SD card?
OK
12
Cancel
Module
About
2014-06-15
2014-06-17
2014-06-18
2014-06-22
Menu-File menu
Measure Set.
File
20140615
20140617
20140618
201406192014-06-19
20140622
System Set.
Measure Set.
File
20140615
20140617
20140618
201406192014-06-19
20140622
Module
2014-06-15
2014-06-17
2014-06-18
System Set.
2014-06-22
About
Module
About
2014-06-15
2014-06-17
2014-06-18
2014-06-22
13
Select "open" option, there are folders. The folder
name is the date of the file saved, which is generated
automatically by system. Files measured in the same
day will be stored in the same folder. Folder can only
be delete, but not rename. And must delete all files
before delete the folder.
Press Up or Down button to select a folder and
press button to open delete tip. Then press
button to finish deleting, or press
MENU
ESC
or button to cancel.
Menu-File menu
Measure Set.
File
.
..
133210_0001.sor
133210_0003.sor
133210_0004.sor
File Info: 1550nm 370m 10ns
System Set.
Measure Set.
File
.
..
133210_0001.sor
133210_0003.sor
133210_0004.sor
File Info 1550nm 370m 10ns
Module
System Set.
About
2014-06-18
Module
2014-06-18133210_0002.sor
2014-06-182014-06-18
Delete...
Rename...
About
2014-06-18
2014-06-18133210_0002.sor
2014-06-18
2014-06-18
14
Select one folder to display all files inside. Press Up
or Down button to select one file and press
button to display it in main interface. The file
information at the bottom is the file's main
measurement setting.
Press button can delete or rename the file.
Filename can make up of 23 digits, alphabets and
13
special symbols at most. The last four numbers_xxxx is
generated by "Automatic naming"function. Shut down
this function will not generate.
ENTER
Menu-File menu
Save
File Name:
140950
CapsbackSave
abcde
hijklmn
opqrs
vwxyz
File Name:
140950
abcde
hijklmn
opqrs
vwxyz
f
CapsbackSave
t
.
g
u
Rename
7
4
12
f
_
t
.
89
5
g
0
u
6
3
#
7
4
12
_
89
5
0
15
When finish measuring an optic fiber, press
File
button to enter storage interface.
Default filename is hour/minute/second of first time
storage. Press Up/Down/Left/Right button to select
alphabets, digits and symbols on soft keyboard. You
can input 23 alphabets at most. If "Automatic
naming" is enabled, the filename will automatically
6
generated with four digits. Without shutting down
3
the device, the following stored file's name will
#
automatically plus 1. The file format is .SOR.
On file recalling interface, press Right button can
Modify filename, as the same operations above.
Menu-Measure settings
Measure Set.
File
Wavelength
Scan Mode
Range
Pulse Width
Average Time
Splice Loss
Reflection Threshold
End Threshold
Refractive Rate
Scatter Coefficient
Splice loss---treat as an event when the loss is higher than setting value.
Reflection threshold---treat as an event when the reflection is higher than setting value.
End threshold---treat as the end of optic fiber when the loss is higher than setting value.
Refractive rate---represent the average refractive index of entire optic fiber.
Scatter coefficient--- the intrinsic value of Rayleigh Scattering.
System Set.
1550
Manual
40
1000
120
0.05
65.0
2.0
1.46832
52.1
Module
About
nm
Km
ns
s
dB
dB
dB
dB
Measure settings menu is used to set relative measurement data, which
the judgment of events list is based on. Wrong setting might leads to wrong or
missing events.
Wavelength---wavelength of laser.
Scan mode---manual and auto mode. Under auto mode, it will match the
distance, range and pulse width.
Range---match with the length of measured optic fiber, usually over one level.
Pulse width---set the pulse width of output laser. Usually, small pulse width can
measure close event, large pulse width can measure remote distance, but
enlarge event's blind area.
Average time--- can set between 5 second and 180 second as average
measurement time.
16
Menu-System settings
Measure Set.
File
Auto Power Off
Auto Name
Optical Dectector
RT AnalyseOff
Brightness
Date
Time
Length Unit
Factory Settings
Press key ENTER to edit settings.
System Set.
Module
10min
On
On
60
2013
2157
:
Metric
No
1102
About
System settings menu is used to set the device's basic functions.
Auto Power Off--- set time of automatic off or cancel this function.
Auto Name---name automatically when file is saving, can cancel this
function.
Optical Detector --- detect whether there is signal in optic fiber or
not before measuring.
RT Analyse --- set whether analyses events after real-time measuring
or not.
Brightness--- adjust brightness of LCD.
Date--- set year, month, day.
Time ---set hour, minute.
Length unit --- default is metric units.
Factory settings --- is used in resetting to default , and has no effect
on the set time and stored data.
17
Menu-Module settings
Measure Set.
File
VFL
OPM
Wavelength
Press key ENTER to enable the VFL.
Test Equ
System Set.
Close
Off
Power
1310
nm
ipment Depot - 800.517.8431 - 99 Washington Street - Melrose, MA 02176
Module
--.----.--
--.----.--
About
Module settings menu is used to set the Visual Fault Laser, Optical
Power Meter and wavelength.
VFL--- cyclically control the red laser Open -- Blink -- Close.
OPM( Optional) --- used as an ordinary optical power meter, detect
dBm
mw
range is from -60dBm to +3dBm, press Enter button to set the OPM
function On or Off, display dBm and mw values. There are six