ProMinent DULCOTEST DT1B Operating Manual

Page 1
Operating manual
A0943
DULCOTEST® DT1B Photometer
EN
Please carefully read these operating instructions before use. · Do not discard.
The operator shall be liable for any damage caused by installation or operating errors.
The latest version of the operating instructions are available on our homepage.
Page 2
Supplemental directives
General non-discriminatory approach In order to make it easier to read, this
document uses the male form in grammatical structures but with an implied neutral sense. It is aimed equally at both men and women. We kindly ask female readers for their understanding in this simplification of the text.
Supplementary information
Please read the supplementary information in its entirety.
Information
This provides important information relating to the correct operation of the unit or is intended to make your work easier.
Warning information
Warning information include detailed descriptions of the hazardous situation.
The following symbols are used to highlight instructions, links, lists, results and other elements in this document:
More symbols
Symbol Description
Action, step by step.
Outcome of an action.
Links to elements or sections of these instructions or other applicable documents.
n
List without set order.
2
Page 3
Symbol Description
Supplemental directives
[Button]
„Display /
GUI“
CODE
Display element (e.g. indicators).
Operating element (e.g. button, switch).
Screen elements (e.g. buttons, assignment of function keys).
Presentation of software elements and/or texts.
3
Page 4

Table of contents

Table of contents
1 General Information................................................................................... 5
1.1 Scope of delivery............................................................................... 6
1.2 Instructions for use............................................................................ 8
1.3 Adapter for 16 mm cuvettes............................................................. 11
1.4 Operations carried out on the device............................................... 12
2 Commissioning DT 1B............................................................................. 13
2.1 Commissioning................................................................................ 13
3 Operating Menu....................................................................................... 17
3.1 Operating Menu Options.................................................................. 17
3.2 Operating instructions...................................................................... 18
3.3 Error messages............................................................................... 18
4 Analysis methods..................................................................................... 20
4.1 Procedure instructions when using liquid reagents......................... 20
4.2 Quantitative determination using liquid reagents............................. 22
4.3 Procedure instructions when using tablets...................................... 29
4.4 Quantitative determination using tablets.......................................... 31
5 Calibration ............................................................................................... 44
6 Technical Data......................................................................................... 47
7 Consumables and Spare Parts................................................................ 49
8 Standards Complied with and Declaration of Conformity........................ 50
9 Index........................................................................................................ 51
4
Page 5

1 General Information

WARNING!
Danger from hazardous sub‐ stances!
Possible consequence: Fatal or very serious injuries.
Please ensure when handling hazardous substances that you have read the latest safety data sheets provided by the manu‐ facture of the hazardous sub‐ stance. The actions required are described in the safety data sheet. Check the safety data sheet regularly and replace, if necessary, as the hazard potential of a substance can be re-evaluated at any time based on new findings.
The system operator is respon‐ sible for ensuring that these safety data sheets are available and that they are kept up to date, as well as for producing an associated hazard assess‐ ment for the workstations affected.
General Information
5
Page 6
General Information

1.1 Scope of delivery

The following components are included as standard:
Description Quan‐
tity
n Case, blue with blue fasteners
with ProMinent sticker, Photometer Dulcotest DT1B, code
1039315
and hazard symbol
Foam insert for the case 1
Cover foam for the case 1
n Photometer DT1B
Chlorine, bromine, chlorine dioxide, ozone, pH, CyA – Battery compartment lid with O-ring
Screwdriver with clip, red 1
Countersunk screws 4
Batteries, 1.5 V alkali-manganese, type AA 4
Round cuvettes, 10 ml, d = 24 mm, h = 48 mm 3
Lid for round cuvette, 24 mm, grey 3
Sealing rings for cuvettes, grey 3
Round cuvettes, d = 16 mm, h = 90 mm 3
Lid for round cuvette, 16 mm, white 3
Adapter attachment, grey for 16 mm cuvettes 1
1
1
Syringe, 10 ml 1
Plastic mixing rod, 13 cm long 1
Cleaning brush 1
Operating instructions for Dulcotest DT1B 1
DPD -1 tablets 100
6
Page 7
General Information
Description Quan‐
tity
DPD -3 tablets 100
PHENOLRED PHOTOMETER tablets 100
CyA TEST tablets 100
CHLORINE HR (KI) tablets 100
ACIDIFYING GP tablets 100
Glycine tablets 20
7
Page 8
General Information

1.2 Instructions for use

Request the safety data sheets.
Reagents are intended for chemical analysis and access to them by unauthorised persons must not be permitted.
You must dispose of reagent solutions properly.
Observe the application possibilities, analysis instruction and matrix effects of the methods.
1. You must thoroughly clean the cuvettes, lids and stirring rod after each analysis to avoid carry-over errors.
Even slight reagent residue can lead to incorrect measurement results. For cleaning use the supplied brush.
If the fully reacted water sample is left for any length of time it may pro‐ duce stubborn coloured deposits, which you can remove using dilute (= 4 %) hydrochloric acid.
2. The outer walls of the cuvettes must be clean and dry before the analysis is carried out. Fingerprints or water droplets on the light-entry surfaces of the cuvette will result in faulty measurements. The cuvette should there‐ fore be wiped clean with a soft paper tissue (paper handkerchief) before carrying out the measurement.
3. The zero correction and the analysis must both be carried out using the same cuvette, as the cuvettes may have slightly varying tolerances.
8
Page 9
A0937
General Information
Fig. 1: Cuvette positioning (Ø 24 mm):
I. Seal II. Marking with the white triangle
4. For the zero correction and test, the cuvette must be positioned in the sample chamber in such a way that the marking with the white triangle (II.) points towards the housing marking.
5. You must carry out the zero correction and test with the cuvette lid closed. You must provide the cuvette lid with a sealing ring (I.) to prevent the entrance of light into the sample chamber.
6. Formation of bubbles on the inside walls of the cuvette will result in faulty measurements. Should this happen, seal the cuvette with the cuvette lid and tip it back and forth to dissolve/remove the bubbles before carrying out the test.
9
Page 10
A0938
General Information
7. It is important to prevent water getting into the sample chamber because this leads to incorrect measurement results.
8. Dirt in the transparent sample chamber leads to incorrect measurement results. The light entry surfaces should be checked at regular intervals and cleaned as necessary.
Standard spectacle cleaning cloths and cotton buds are suitable for cleaning.
9. Pronounced differences in temperature between the photometer and the surroundings can lead to incorrect measurements, e.g. due to the forma‐ tion of condensation in the sample chamber and on the cuvette.
10. Protect the device from direct sunlight.
Fig. 2: Correct filling of the cuvette: Left = correct / right = incorrect
11. Fill the cuvette as shown in Fig. 2.
12. Insert the reagent tablets directly from the blister pack into the water sample, without touching them with your fingers.
13. After use immediately close the dropping bottles containing the liquid reagents by closing them with the screw caps of the same colour.
14. You must observe the sequence for the addition of reagents without fail.
10
Page 11

1.3 Adapter for 16 mm cuvettes

A0941
General Information
Fig. 3: Place the adapter for 16 mm cuvettes on the sample chamber
Insert the adapter for the 16 mm cuvette as shown in Fig. 3.
This adapter is required for all analysis methods which must be carried out using a 16 mm cuvette.
11
Page 12
A0939
General Information

1.4 Operations carried out on the device

Battery replacement:
To ensure complete leak-tightness of the photometer, you must insert the gasket (2) and screw on the battery compartment lid (1).
If the battery is removed for more than 1 minute from the device, then when the device is switched on again with new inserted batteries, the date/time program appears automatically.
Fig. 4: Battery replacement
1. Battery compartment cover
2. Seal
3. Screw
4. Batteries
5. Rear of the device
12
Page 13

2 Commissioning DT 1B

2.1 Commissioning

Switching on and zero correction
Scroll memory (SM)
With multi-parameter devices the sequence of the various methods is specified. After switching on the device, the method which was last selected before the device was switched off is automatically displayed. This permits rapid access to a favoured method.
1. Switch the device on using the
[ON/OFF]
ð
2. Select the
[MODE]
ð
Zero correction
3. Fill the cuvette up to the 10-ml marking with the water sample, see Fig. 2
4. Close the cuvette using the cuv‐ ette lid
5. Position the cuvette in the sample chamber
ð
6. Press the key
key
The display outputs: The last selected
key
The display outputs:
[METHOD]
Observe the correct posi‐ tioning of the cuvette.
[METHOD]
[METHOD]
.
using the
[ZERO/TEST]
.
Commissioning DT 1B
[METHOD]
ð
approximately 8 seconds.
[0.0.0]
play.
7. Remove the cuvette from the sample chamber
Zero correction is ended.
ð
Analysis
8. Fill the cuvette up to the 10-ml marking with the water sample, see Fig. 2
9. Add reagents to the water sample
The characteristic colouring
ð
develops.
10. Close the cuvette using the cuv‐ ette lid
11. Position the cuvette in the sample chamber
12. Press the so doing adhere to any possibly required reaction time, see
flashes for
appears in the dis‐
[ZERO/TEST]
key, in
Ä
Switching the countdown func‐ tion on“ on page 14
[METHOD]
ð
approximately 3 seconds.
The result is output to the display.
flashes for
13
Page 14
Commissioning DT 1B
Switching the countdown function on
The result is automat‐ ically saved.
Repeating the analysis
Press the key again
The process runs as
ð
described here
[ZERO/TEST]
on and zero correc‐ tion“ on page 13
New zero correction
Press the key 2 seconds
The process runs as
ð
described here
[ZERO/TEST]
on and zero correc‐ tion“ on page 13
Ä „Switching
.
for
Ä „Switching
.
NOTICE!
If reaction times are not observed, the result may be incorrect measurement results.
The countdown can only be activated directly prior to a measurement.
The countdown time equals 2 minutes and is non-adjustable.
The currently running count‐ down can be ended by pressing the key [ZERO/TEST]. The measurement then takes place immediately.
For methods with a reaction time, you can optionally switch on a countdown function:
1. Press the key the key depressed
[!]
and then keep
2. Press the key
3. Release the
The countdown starts After
ð
the countdown has elapsed (2 minutes), measurement takes place automatically
[ZERO/TEST]
[!]
key
14
Page 15
Displaying stored data
The device has a ring buffer for 16 data records.
Fig. 5: Displaying stored data
1. Data record (n01 ... n16)
2. Year
3. Month/day
4. Time
5. Measured variable (e.g. chlorine, dependent upon the device ver‐ sion)
6. Value in mg/l
Commissioning DT 1B
1. With the device switched on press the
[!]
key for more than 4 seconds then
release the key again
The display immediately switches to the memory menu.
ð
2. Press the
3. Press the
4. Press the
15
[MODE]
[ZERO/TEST]
[!]
key to scroll through the 16 data records
key to scroll through the values of a data record
key to return to the
[METHOD]
display
Page 16
Commissioning DT 1B
Display background lighting
Press the
ð
[!]
key
The display's background lighting switches on or off.
During measuring the background lighting switches off automati‐ cally.
16
Page 17

3 Operating Menu

3.1 Operating Menu Options

Operating menu selection
1. The device is switched off.
Press and hold down the
[MODE]
2. Use the
on the device
ð
3. Release the
4. The [!] key allows you to select
the following operating menu items:
n
n Set date and time n User adjustment
ð
5. Use the
date and time“
item (arrow at top right and bottom left in the display)
key
[ON/OFF]
3 decimal points appear in the display.
[diS]
= read out of stored
data
The selected operating menu item is indicated by an arrow in the display.
[!]
key to select the
key to switch
[MODE]
key
operating menu
„Set
Operating Menu
Setting date and time (24-hour format)
Increase the value to be set by pressing the [MODE] key.
Lower the value to be set by pressing the [ZERO/TEST] key.
Pressing the [!] key takes you to the next value to be set.
1. Press the
ð
2. Enter the year
3. Enter the month
4. Enter the day
5. Enter the hour
6. Enter the minutes
Enter the minutes in 10 minute increments
Press the
Enter minutes in 1 minute incre‐ ments
7. After setting the minutes, press the
ð
[MODE]
The parameter to be set appears for 2 seconds.
key
[YYYY]
[MM]
[dd]
[hh]
[mm]
.
[!]
key
[!]
key
[IS SET]
play and the device auto‐ matically returns to meas‐ uring mode.
appears in the dis‐
17
Page 18
A1856
Operating Menu

3.2 Operating instructions

Display Meaning
Hi Measuring range exceeded or turbidity too
high.
Lo Measuring range undershot.
Change the batteries immediately, continuing work impossible.
btLo Battery voltage for background lighting too low,
although measurement is possible.
With a method adjusted by the user, when the result is shown in the display, an arrow appears in the
[Cal]
position.

3.3 Error messages

Display Meaning
E 27/ E 28/
Light absorption too great. Cause e.g. dirty optics.
E 29
E 10 / E 11 Calibration factor outside the permissible range.
E 20 / E 21 Detector receives too much light.
E 23 / E24 /
Detector receives too much light.
E 25
E 22 The battery power was too low during the measurement.
Replace the battery.
E 70 Cl: Factory calibration not OK / deleted
E 71 Cl: User calibration not OK / deleted
E 72 Cl HR factory calibration not in order / deleted
E 73 Cl HR user calibration not OK / deleted
18
Page 19
Display Meaning
E 80 pH: Factory calibration not OK / deleted
E 81 pH: User calibration not OK / deleted
E 82 CyA: Factory calibration not OK / deleted
E 83 CyA: User calibration not OK / deleted
Operating Menu
19
Page 20
Analysis methods

4 Analysis methods

4.1 Procedure instructions when using liquid reagents

When calculating non-directly determinable parameters from individual measured values error propagation based on the possible tolerances of the indi‐ vidual methods must be consid‐ ered.
1. Cleaning the cuvettes Many
domestic cleaning agents con‐ tain reducing substances. Use of these substances leads to false low readings when meas‐ uring chlorine/bromine/chlorine dioxide/ozone. To exclude this measurement error, the cuv‐ ettes must not contain any sub‐ stances that have a chlorine demand. Therefore the cuvettes must be stored in a sodium hypochlorite solution (0.1 g/l) for one hour and then thoroughly rinsed with deionised water.
2. For the relevant specification of
free chlorine and total chlorine use a separate set of cuvettes for each determination (see EN ISO 7393-2, section. 5.3).
Label the cuvettes set on the top and bottom so that inadvertent inter‐ changing of the cuvettes is not possible.
3. When preparing water samples you must avoid the outgassing of chlorine/bromine/chlorine dioxide/ozone, e.g. by pipetting and shaking. This applies partic‐ ularly for the dissolved gases chlorine dioxide and ozone, especially at temperatures > 30 °C. You must carry out the analysis immediately after taking the water sample.
4. The DPD colour development takes place at a pH-value of 6.2 to 6.5. Therefore the reagents contain a buffer for pH value adjustment. You must ensure strongly alkaline or acidic water samples come within a pH range between 6 and 7 (using
0.5 mol/l sulphuric acid solution or 1 mol/l sodium hydroxide sol‐ ution).
5. Concentrations above the measuring range with ...
n 4 mg/l chlorine when using
liquid reagents
n 9 mg/l bromine when using
liquid reagents
20
Page 21
Analysis methods
n 7.6 mg/l chlorine dioxide
when using liquid reagents
n 2.7 mg/l ozone when using
liquid reagents
may lead to results within
ð
the measuring range down to 0 mg/l. In this case, you must dilute the water sample with water free from oxidizing agent and repeat the measurement (plausi‐ bility test).
6. Any turbidity that occurs during the colour reaction leads to too­high results. You can prevent this error by pre-diluting the sample with oxidizing agent-free water. You must consider the diluting ratio (e.g. 1:2) when cal‐ culating the measurement result.
7. After use immediately close the dropping bottles containing the liquid reagents using the respective screw caps of the matching colour. Store the reagent set at + 6 °C to + 10 °C.
8. The DPD method used responds to many oxidising media, hence you must ensure that the selected oxidising agent is present on its own. Mixtures, e.g. from chlorine and chlorine dioxide only yield total values. These total values must then be differentiated using additional steps. To differentiate between chlorine and chlorine dioxide, see
[Chlorine with liquid reagent
method, section d.]
. To differen‐ tiate between chlorine and ozone, see
[Chlorine with liquid reagent method, section e.]
.
9. In water containing bromide and iodide (primarily seawater), the free and, as the case may be, bound halogens formed by chlorination are stated as chlorine.
A steady increase in the meas‐ ured value of a water sample therefore indicates that along‐ side the selected oxidising agent (e.g. chlorine) a further oxidising agent (e.g. bromide or iodide) is present. This addi‐ tional oxidising agent (e.g. bro‐ mide or iodide) may due to cer‐ tain circumstances (many times higher concentration, equilibria, higher temperature) bleed through into the measurement. By working quickly and reading off the measurement immedi‐ ately, the resulting error can be minimised.
These interference effects
ð
are also known to occur with these systems {combined chlorine ⇛ free chlorine} and {chloride ⇛ chlorine dioxide}.
21
Page 22
Analysis methods

4.2 Quantitative determination using liquid reagents

Chlorine with liquid reagents 0.01 ...
4.0 mg/l
Use the
[MODE]
key to select
[Cl]
.
a) Free chlorine
1. Take your cuvette set 1
2. In a 24 mm cuvette add a 10 ml
water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
3. Remove the cuvette from the
sample chamber and empty the cuvette completely
Hold the dropping bottle upright
4.
and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
Fill the cuvette up to the 10 ml marking with the water sample
5. Close the cuvette using the cuv‐
ette lid
6. Mix the contents of the cuvette
by tipping it back and forth
7. Position the cuvette in the
sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
8. Press the key
[ZERO/TEST]
[METHOD]
ð
flashes for
approximately 3 seconds.
The result in mg/l of free chlorine appears in the dis‐ play.
b) Total chlorine
1. Take your cuvette set 2
2. In a 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
3. Remove the cuvette from the sample chamber and empty the cuvette completely
Hold the dropping bottle upright
4. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
n 3 drops ➨ DPD 3 solution
Fill the cuvette up to the 10 ml marking with the water sample
5. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth
6. Place the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
22
Page 23
Analysis methods
7. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
8. The METHOD symbol flashes for approximately 3 seconds
The result in mg/l of total
ð
chlorine appears in the dis‐ play.
c) combined chlorine
[!]
key
and
. To do
Combined chlorine = total chlorine minus free chlorine
Calculate the combined chlorine
d) Chlorine together with chlorine dioxide
1. Fill a cuvette with a 10 ml water sample
2. Insert a directly from the foil into the water sample and crush the
[GLYCINE]
rod
3. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
[GLYCINE]
tablet with a stirring
tablet
4. In a second 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
5. Remove the cuvette from the sample chamber for the zero correction and empty the cuv‐ ette out
Hold the dropping bottle upright
6. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
7. Pour the contents of the first cuvette ( into the prepared cuvette
8. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth
9. Place the cuvette in the sample chamber
ð
10. Press the key
ð
11. Remove the cuvette from the sample chamber
12. Thoroughly clean the cuvette and the cuvette lid
[GLYCINE]
Observe the correct posi‐ tioning of the cuvette.
solution)
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The display value chlorine dioxide reappears in the display.
flashes for
[G]
=
23
Page 24
Analysis methods
13. Hold the dropping bottle upright and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
Fill the cuvette up to the 10 ml marking with the water sample
14. Close the cuvette using the cuv‐ ette lid
15. Mix the contents of the cuvette by tipping it back and forth
16. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette
17. Press the key
[METHOD]
ð
approximately 3 seconds.
The display value of free chlorine plus chlorine dioxide appears in the dis‐ play.
18. Remove the cuvette from the sample chamber
19. Hold the dropping bottle upright and by slow pressing, add equal sized drops into the cuvette:
[ZERO/TEST]
flashes for
[A]
= sum
22. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette
23. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
24.
[METHOD]
mately 3 seconds.
The display value
ð
of free chlorine plus com‐ bined chlorine plus chlorine dioxide appears in the dis‐ play.
Calculation:
25.
n Chlorine dioxide (mg/l) =
display value
n Free chlorine (mg/l) = dis‐
play value value
n Combined chlorine (mg/l) =
display value play value
[!]
and
key
flashes for approxi‐
[A]
[G]
[A]
. To do
[C]
= sum
[G]
x 1.9
minus display
[C]
minus dis‐
n 3 drops ➨ DPD 3 solution
20. Close the cuvette using the cuv‐ ette lid
21. Mix the contents of the cuvette by tipping it back and forth
24
Page 25
Analysis methods
d) Chlorine together with ozone
1. In a 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette completely
Hold the dropping bottle upright
3. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
n 3 drops ➨ DPD 3 solution
Fill the cuvette up to the 10 ml marking with the water sample
4. Close the cuvette using the cuv‐ ette lid
5. Mix the contents of the cuvette by tipping it back and forth
6. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
7. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
[!]
key
and
. To do
tion time to elapse
8. The METHOD symbol flashes for approximately 3 seconds
The result of Total chlorine
ð
+ ozone = display value
[A]
appears in mg/l in the dis‐ play.
9. Remove the first cuvette from the sample chamber
10. Thoroughly clean the cuvette and the cuvette lid
11. Fill a second cuvette with a 10 ml water sample
12. Insert a
[GLYCINE]
tablet directly from the foil into the water sample and crush the
[GLYCINE]
tablet with a stirring
rod
13. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
14. Hold the dropping bottle upright and by slow pressing, add equal sized drops into the first cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
n 3 drops ➨ DPD 3 solution
15. Pour the contents of the second cuvette (
[GLYCINE]
solution)
into the prepared cuvette
16. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth
25
Page 26
Analysis methods
17. Place the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
18. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
19. The METHOD symbol flashes for approximately 3 seconds
The result in mg/l of Total
ð
chlorine = display value appears in the display.
Calculation:
20.
n Ozone (mg/l) = (Display
value value
[!]
and
key
[A]
minus display
[G]
) x 0.67
. To do
[G]
Measuring tolerances when determining the chlorine con‐ tent:
0 ... 1 mg/l: ± 0.05 mg/l
> 1 ... 2 mg/l: ± 0.10 mg/l
> 2 ... 3 mg/l: ± 0.20 mg/l
> 3 ... 4 mg/l: ± 0.30 mg/l
Bromine with liquid reagents 0.02 ... 9 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[Br]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette completely
Hold the dropping bottle upright
3. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
n 3 drops ➨ DPD 3 solution
Fill the cuvette up to the 10 ml marking with the water sample
4. Use the cuvette lid to seal the cuvette
5. Mix the contents of the cuvette by tipping it back and forth
6. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
7. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press
Wait for the 2 minutes reac‐
ð
tion time to elapse
[!]
and
. To do
[ZERO/TEST]
26
Page 27
8.
[METHOD]
mately 3 seconds.
The result in mg/l of bromine
ð
appears in the display.
9.
Measuring tolerances when determining the bromine content:
flashes for approxi‐
0 ... 2.3 mg/l: ± 0.12 mg/l
> 2.3 ... 4.5 mg/l: ± 0.25 mg/l
> 4.5 ... 6.8 mg/l: ± 0.45 mg/l
> 6.8 ... 9 mg/l: ± 0.68 mg/l
Analysis methods
Chlorine dioxide with liquid reagents
0.02 ... 7.6 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[CdO]
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette completely
Hold the dropping bottle upright
3. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
.
Explanatory notes, see
ð
Ä Further information
on page 20
27
.
Fill the cuvette up to the 10 ml marking with the water sample
4. Close the cuvette using the cuv‐ ette lid
5. Mix the contents of the cuvette by tipping it back and forth
6. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
7. Press the key
[METHOD]
ð
approximately 3 seconds.
The result in mg/l of chlorine dioxide appears in the dis‐ play.
[ZERO/TEST]
flashes for
Page 28
Analysis methods
8.
Measuring tolerances when determining the chlorine dioxide content:
0 ... 1.9 mg/l: ± 0.1 mg/l
> 1.9 ... 3.8 mg/l: ± 0.2 mg/l
> 3.8 ... 5.7 mg/l: ± 0.4 mg/l
> 5.7 ... 7.6 mg/l: ± 0.6 mg/l
Ozone with liquid reagents 0.01 ... 2.7 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[O3]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette completely
Hold the dropping bottle upright
3. and by slow pressing, add equal sized drops into the cuvette:
n 6 drops ➨ DPD 1 buffer sol‐
ution
n 2 drops ➨ DPD 1 reagent
solution
n 3 drops ➨ DPD 3 solution
Fill the cuvette up to the 10 ml marking with the water sample
4. Close the cuvette using the cuv‐ ette lid
5. Mix the contents of the cuvette by tipping it back and forth
6. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
7. Press the key
[METHOD]
ð
approximately 3 seconds.
The result in mg/l of ozone appears in the display.
[ZERO/TEST]
flashes for
28
Page 29
Analysis methods
8.
Measuring tolerances when determining the ozone content:
0 ... 0.67 mg/l: ± 0.03 mg/l
> 0.67 ... 1.3 mg/l: ± 0.07 mg/l
> 1.3 ... 2.0 mg/l: ± 0.13 mg/l
> 2.0 ... 2.7 mg/l: ± 0.20 mg/l
Explanatory notes, see
ð
Ä Further information
on page 20
.

4.3 Procedure instructions when using tablets

When calculating non-directly determinable parameters from individual measured values error propagation based on the possible tolerances of the indi‐ vidual methods must be consid‐ ered.
1. Cleaning the cuvettes Many
domestic cleaning agents con‐ tain reducing substances. Use of these substances leads to false low readings when meas‐ uring chlorine/bromine/chlorine dioxide/ozone. To exclude this measurement error, the cuv‐ ettes must not contain any sub‐ stances that have a chlorine demand. Therefore the cuvettes must be stored in a sodium hypochlorite solution (0.1 g/l) for one hour and then thoroughly rinsed with deionised water.
2. For the relevant specification of
free chlorine and total chlorine use a separate set of cuvettes for each determination (see EN ISO 7393-2, section. 5.3).
29
Page 30
Analysis methods
Label the cuvettes set on the top and bottom so that inadvertent inter‐ changing of the cuvettes is not possible.
3. When preparing water samples you must avoid the outgassing of chlorine/bromine/chlorine dioxide/ozone, e.g. by pipetting and shaking. This applies partic‐ ularly for the dissolved gases chlorine dioxide and ozone, especially at temperatures > 30 °C. You must carry out the analysis immediately after taking the sample.
4. The DPD colour development takes place at a pH-value of 6.2 to 6.5. Therefore the reagents contain a buffer for pH value adjustment. You must ensure strongly alkaline or acidic water samples come within a pH range between 6 and 7 (using
0.5 mol/l sulphuric acid solution or 1 mol/l sodium hydroxide sol‐ ution).
5. Concentrations above
n 10 mg/l chlorine when using
tablets
n 22 mg/l bromine when using
tablets
n 19 mg/l chlorine dioxide
when using tablets
n 6 mg/l ozone when using
tablets
may lead to results within
ð
the measuring range down to 0 mg/l. In this case, you must dilute the water sample with water free from oxidizing agent and repeat the measurement (plausi‐ bility test).
6. Turbidity (causes incorrect measurements): For water sam‐ ples with High Calcium content* and/or high conductivity when using the
[DPD no. 1 tablet]
result can be the rendering turbid of the water sample and consequently incorrect meas‐ urements. In this case, you must you must use the alterna‐ tive reagent tablet
[DPD no. 1 High Calcium]
turbidity only appears after the addition of the
[DPD No. 3]
tablet, you can prevent this by use of the
[DPD no. 1 High Calcium]
the
[DPD no. 3 High Calcium]
tablet. The
[DPD No. 1 High Calcium]
should only be used in conjunc‐ tion with the
[DPD No. 3 High Calcium]
*exact values cannot be
ð
specified, as the formation of turbidity depends on the type and composition of the water sample.
7. The DPD method used responds to many oxidising media, hence you must ensure that the selected oxidising agent is present on its own. Mixtures,
the
. If the
and
tablet
.
30
Page 31
Analysis methods
e.g. of chlorine and chlorine dioxide only yield total values. These total values must then be differentiated using additional steps. To differentiate between chlorine and chlorine dioxide, see
[Chlorine with tablet method, section D "Chlorine together with chlorine dioxide"]
entiate between chlorine and ozone, see
To differ‐
[Chlorine with tablet method]
8. In water containing bromide and iodide, the free and, as the case may be, bound halogens formed by chlorination are stated as chlorine. A steady increase in the measured value of a water sample therefore indicates that alongside the selected oxidising agent a further oxidising agent is present. This oxidising agent may due to certain circum‐ stances (many times higher concentration, equilibria, higher temperature) bleed through into the measurement. By working quickly and reading off the measurement immediately, the resulting error can be mini‐ mised.
These interference effects
ð
are also known to occur with these systems {combined chlorine ⇛ free chlorine} and {chloride ⇛ chlorine dioxide}.

4.4 Quantitative determination using tablets

Fully dissolve tablets
Make sure that the tablets have fully dissolved before starting measurements. Otherwise incorrect measurements may result.
31
Page 32
Analysis methods
Chlorine with tablets 0.01 ... 6.0 mg/l
Use the
a) Free chlorine
1. Take your cuvette set 1
2. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[Cl]
.
Ä „Switching on and zero cor‐ rection“ on page 13
3. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops
4. Insert a directly from the foil into the water sample and crush the
[DPD No. 1]
rod
5. Fill the cuvette up to the 10 ml marking with the water sample
6. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
7. Position the cuvette in the sample chamber
ð
8. Press the key
ð
[DPD No. 1]
tablet with a stirring
Observe the correct posi‐ tioning of the cuvette.
tablet
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
flashes for
The result in mg/l of free chlorine appears in the dis‐ play.
b) Total chlorine
1. Take your cuvette set 2
2. In a 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
3. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops
4. Insert a directly from the foil into the water sample and crush the
[DPD No. 1]
rod
5. Insert a directly from the foil into the water sample and crush the
[DPD No. 3]
rod
6. Fill the cuvette up to the 10 ml marking with the water sample
7. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablets have dissolved
8. Place the cuvette in the sample chamber
ð
[DPD No. 1]
tablet with a stirring
[DPD No. 3]
tablet with a stirring
Observe the correct posi‐ tioning of the cuvette.
tablet
tablet
32
Page 33
Analysis methods
9. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
10. The METHOD symbol flashes for approximately 3 seconds
The result in mg/l of total
ð
chlorine appears in the dis‐ play.
c) combined chlorine
[!]
key
and
. To do
Combined chlorine = total chlorine minus free chlorine
Calculate the combined chlorine
d) Chlorine together with chlorine dioxide
1. Fill a cuvette with a 10 ml water sample
2. Insert a directly from the foil into the water sample and crush the
[GLYCINE]
rod
3. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
[GLYCINE]
tablet with a stirring
tablet
4. In a second 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
5. Remove the cuvette from the sample chamber for the zero correction and empty the cuv‐ ette out
6. Insert a directly from the foil into the water sample and crush the
[DPD No. 1]
rod
7. Pour the contents of the first cuvette ( into the prepared cuvette
8. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
9. Place the cuvette in the sample chamber
ð
10. Press the key
ð
11. Remove the cuvette from the sample chamber
12. Thoroughly clean the cuvette and the cuvette lid
[DPD No. 1]
tablet with a stirring
[GLYCINE]
Observe the correct posi‐ tioning of the cuvette.
tablet
solution)
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The display value (chlorine dioxide) reappears in the display.
flashes for
[G]
33
Page 34
Analysis methods
13. Fill a cuvette with a few drops of a water sample
14. Insert a directly from the foil into the water sample and crush the
[DPD No. 1]
rod
15. Fill the cuvette up to the 10 ml marking with the water sample
16. Close the cuvette using the cuv‐ ette lid
17. Mix the contents of the cuvette by tipping it back and forth, until the tablet has dissolved
18. Position the cuvette in the sample chamber
ð
19. Press the key
ð
20. Remove the cuvette from the sample chamber
21. Insert a directly from the foil into the same water sample and crush the stirring rod
22. Close the cuvette using the cuv‐ ette lid
[DPD No. 1]
tablet with a stirring
Observe the correct posi‐ tioning of the cuvette.
tablet
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The display value of free chlorine plus chlorine dioxide) appears in the dis‐ play.
[DPD No. 3]
flashes for
[DPD No. 3]
tablet with a
[A]
tablet
(sum
23. Mix the contents of the cuvette by tipping it back and forth, until the tablet has dissolved
24. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
25. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
26. The METHOD symbol flashes for approximately 3 seconds
The display value
ð
of free chlorine plus com‐ bined chlorine plus chlorine dioxide) appears in the dis‐ play.
Calculation:
27.
n Chlorine dioxide (mg/l) =
display value
n Free chlorine (mg/l) = dis‐
play value value
n Combined chlorine (mg/l) =
display value play value
[G]
[!]
key
[A]
[A]
and
. To do
[C]
(sum
[G]
x 1.9
minus display
[C]
minus dis‐
34
Page 35
Analysis methods
d) Chlorine together with ozone
1. In a 24 mm cuvette add a 10 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops
3. Insert a a from the foil into the water sample and crush the tablets using a stirring rod
4. Fill the cuvette up to the 10 ml marking with the water sample
5. Close the cuvette using the cuv‐ ette lid
6. Mix the contents of the cuvette by tipping it back and forth, until the tablets have dissolved
7. Position the cuvette in the sample chamber
ð
8. Switch the countdown on, see
[DPD No. 1]
[DPD No. 3]
Observe the correct posi‐ tioning of the cuvette.
tablet and
tablet directly
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
9. The METHOD symbol flashes for approximately 3 seconds
[!]
key
and
. To do
The result of Total chlorine
ð
+ ozone = display value appears in mg/l in the dis‐ play.
10. Remove the first cuvette from the sample chamber
11. Thoroughly clean the cuvette and the cuvette lid
12. Fill a second cuvette with a 10 ml water sample
13. Insert a directly from the foil into the water sample and crush the
[GLYCINE]
rod
14. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
15. Insert a a from the foil into the water sample and crush the tablets using a stirring rod
16. Pour the contents of the second cuvette ( into the prepared cuvette
17. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth
18. Place the cuvette in the sample chamber
ð
[GLYCINE]
tablet with a stirring
[DPD No. 1]
[DPD No. 3]
[GLYCINE]
Observe the correct posi‐ tioning of the cuvette.
tablet
tablet and
tablet directly
solution)
[A]
35
Page 36
Analysis methods
19. Switch the countdown on, see
Ä „ Switching the countdown function on“ on page 14
this, press the
[ZERO/TEST]
Wait for the 2 minutes reac‐
ð
tion time to elapse
20. The METHOD symbol flashes for approximately 3 seconds
The result in mg/l of Total
ð
chlorine = display value appears in the display.
Calculation:
21.
n Ozone (mg/l) = (Display
value value
[!]
and
key
[A]
minus display
[G]
) x 0.67
. To do
[G]
Measuring tolerances when determining the chlorine con‐ tent:
0 ... 1 mg/l: ± 0.05 mg/l
> 1 ... 2 mg/l: ± 0.10 mg/l
> 2 ... 3 mg/l: ± 0.20 mg/l
> 3 ... 4 mg/l: ± 0.30 mg/l
> 4 ... 6 mg/l: ± 0.40 mg/l
Chlorine HR with tablet 5 ... 200 mg/l
For the analysis method
with tablet 5 ... 200 mg/l“
contained in the scope of supply is required, see
Ä Chapter 1.3 „Adapter
for 16 mm cuvettes“ on page 11
1. Insert the adapter for the 16 mm round cuvette in the device, see
„Chlor HR
the adapter
.
Ä Chapter 1.3 „Adapter for 16 mm cuvettes“ on page 11
2. In a 16 mm cuvette add an 8 ml water sample and carry out a zero correction, see
Ä „Switching on and zero cor‐ rection“ on page 13
3. Insert a tablet and an tablet directly from the foil into the water sample and crush the tablets using a stirring rod
4. Close the cuvette with the cuv‐ ette lid and mix the contents by tipping back and forth until the tablet has dissolved
5. Position the cuvette in the sample chamber
ð
6. Press the key
ð
[CHLORINE HR (KI)]
[ACIDIFYING-GP]
Observe the correct posi‐ tioning of the cuvette.
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The result in mg/l of chlorine appears in the display.
flashes for
36
Page 37
Measuring tolerances when determining chlorine HR:
± 5 mg/l
Oxidising agents
lead to multiple results
All the oxidising agents contained in the water sample react like chlorine which leads to mul‐ tiple findings.
Ensure, by using suit‐ able methods and measures that for your particular process, such mul‐ tiple finding cannot occur or that such multiple findings can be allowed for in another manner.
Analysis methods
Bromine with tablet 0.02 ... 13 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[Br]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops
3. Enter a
[DPD No. 3]
the foil into the water sample and crush the tablet with a stir‐ ring rod
4. Fill the cuvette up to the 10 ml marking with the water sample
5. Use the cuvette lid to seal the cuvette
6. Mix the contents of the cuvette by tipping it back and forth, until the tablet has dissolved
7. Position the cuvette in the sample chamber
ð
8. Switch the countdown on, see
[DPD No. 1]
tablet directly from
Observe the correct posi‐ tioning of the cuvette.
tablet and a
Ä „ Switching the countdown function on“ on page 14
this, press
9. Press
[METHOD]
ð
approximately 3 seconds.
[!]
[ZERO/TEST]
and
flashes for
. To do
[ZERO/TEST]
.
37
Page 38
Analysis methods
10.
The result in mg/l of bromine appears in the display.
Measuring tolerances when determining the bromine content:
0 ... 2.3 mg/l: ± 0.12 mg/l
> 2.3 ... 4.5 mg/l: ± 0.25 mg/l
> 4.5 ... 6.8 mg/l: ± 0.45 mg/l
> 6.8 ... 9 mg/l: ± 0.68 mg/l
> 9 ... 13 mg/l: ± 0.90 mg/l
Explanatory notes, see
ð
Ä Further information
on page 29
.
Chlorine dioxide with tablet 0.02 ... 11 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[CdO]
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops of the water sample
3. Insert a directly from the foil into the water sample and crush the tablet with a stirring rod
4. Fill the cuvette up to the 10 ml marking with the water sample
5. Close the cuvette using the cuv‐ ette lid
6. Mix the contents of the cuvette by tipping it back and forth, until the tablet has dissolved
7. Position the cuvette in the sample chamber
ð
8. Press the key
ð
[DPD No. 1]
Observe the correct posi‐ tioning of the cuvette.
tablet
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The result in mg/l of chlorine dioxide appears in the dis‐ play.
flashes for
.
38
Page 39
Analysis methods
9.
Measuring tolerances when determining the chlorine dioxide content:
0 ... 1.9 mg/l: ± 0.1 mg/l
> 1.9 ... 3.8 mg/l: ± 0.2 mg/l
> 3.8 ... 5.7 mg/l: ± 0.4 mg/l
> 5.7 ... 7.6 mg/l: ± 0.6 mg/l
> 7.6 ... 11 mg/l: ± 0.8 mg/l
Ozone with tablet 0.01 ... 4 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[O3]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Remove the cuvette from the sample chamber and empty the cuvette down to the last few drops
3. Insert a a from the foil into the water sample and crush the tablets using a stirring rod
4. Fill the cuvette up to the 10 ml marking with the water sample
5. Close the cuvette using the cuv‐ ette lid
6. Mix the contents of the cuvette by tipping it back and forth, until the tablets have dissolved
7. Position the cuvette in the sample chamber
ð
8. Press the key
ð
[DPD No. 1]
[DPD No. 3]
Observe the correct posi‐ tioning of the cuvette.
tablet and
tablet directly
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The result in mg/l of ozone appears in the display.
flashes for
39
Page 40
Analysis methods
9.
Measuring tolerances when determining the ozone content:
0 ... 0.67 mg/l: ± 0.03 mg/l
> 0.67 ... 1.3 mg/l: ± 0.07 mg/l
> 1.3 ... 2.0 mg/l: ± 0.13 mg/l
> 2.0 ... 2.7 mg/l: ± 0.20 mg/l
> 2.7 ... 4.0 mg/l: ± 0.27 mg/l
Explanatory notes, see
ð
Ä Further information
on page 29
.
pH value with tablet 6.5 ... 8.4
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[PH]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Insert a
[PHENOL RED PHOTO‐ METER]
foil into the water sample and crush the tablet with a stirring rod
3. Fill the cuvette up to the 10 ml marking with the water sample
4. Close the cuvette using the cuv‐ ette lid
5. Mix the contents of the cuvette by tipping it back and forth, until the tablets have dissolved
6. Position the cuvette in the sample chamber
ð
7. Press the key
ð
tablet directly from the
Observe the correct posi‐ tioning of the cuvette.
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The result is output to the display as a pH value.
flashes for
40
Page 41
Analysis methods
8.
Measuring tolerances when determining pH:
up to ± 0.3 pH units
ð
Remarks:
For the photo‐ metric pH value determination only use [PHENOL RED] tablets with black foil printing, which are labelled with the term [PHOTOMETER].
Water samples with low carbo‐ nate hardness* may result in incorrect pH values.
*KS4.3 < 0.7 mmol/l ≜ Total alkalinity < 35 mg/l CaCO3.
pH values under
6.5 and greater than 8.4 may lead to results within the measuring range. A plausi‐ bility test (pH meter) is recom‐ mended.
The accuracy of pH values based upon colorimetric determination depends on var‐
41
Page 42
Analysis methods
ious boundary conditions (buffer capacity of the water sample, salt content, etc.).
The salt content of the water sample effects the result of the pho‐ tometric pH measurement (salt error), hence this method is not suitable for checking the func‐ tioning of a poten‐ tiometric pH measurement (DIN 19643-2 ff, paragraph 4.2.4. functional check).
Cyanuric acid with CyA test tablet 1 ... 80 mg/l
Use the
1. In a 24 mm cuvette add a 10 ml
[MODE]
water sample and carry out a zero correction, see
key to select
[CyA]
.
Ä „Switching on and zero cor‐ rection“ on page 13
2. Insert a directly from the foil into the water sample and crush the tablet with a stirring rod
3. Fill the cuvette up to the 10 ml marking with the water sample
4. Close the cuvette using the cuv‐ ette lid
5. Mix the contents of the cuvette by tipping it back and forth, until the tablets have dissolved
6. Position the cuvette in the sample chamber
ð
7. Press the key
ð
[CyA TEST]
Observe the correct posi‐ tioning of the cuvette.
tablet
[ZERO/TEST]
[METHOD]
approximately 3 seconds.
The result in mg/l of cya‐ nuric acid appears in the display.
flashes for
42
Page 43
8.
Measuring tolerances when determining the cyanuric acid content:
0 ... 25 mg/l: ± 5 mg/l
25 ... 50 mg/l: ± 8 mg/l
50 ... 80 mg/l: ± 10 mg/l
ð
Remarks:
Cyanuric acid causes a very finely distributed turbidity with a milky appearance in the water sample. Individual particles should not be traced back to the pres‐ ence of cyanuric acid, rather are due to impurities in the water sample.
Dissolve the tablet completely (tilt from side to side for approx. 1 minute). Undis‐ solved particles may result in mul‐ tiple findings.
Analysis methods
43
Page 44
Calibration

5 Calibration

1. The device is switched off. Press and hold down the
[MODE]
2. Use the on the device
ð
3. Release the
4. The [!] key allows you to select the following operating menu items:
n
n Set date and time n User calibration
ð
5. Use the
[CAL]
(arrow top right in the display)
key
[ON/OFF]
3 decimal points appear in the display.
[diS]
= read out of stored
data
The selected menu item is indicated by an arrow in the display.
[!]
key to select the
operating menu item
key to switch
[MODE]
key
Chlorine (Cl) adjustment range
User adjustment
NOTICE!
Separate adjustment of the bro‐ mine, chlorine dioxide or ozone measuring ranges is impos‐ sible. The device relies on the adjustment of the chlorine measuring range (Cl).
User adjustment (display in Adjust‐ ment mode ) =
Factory calibration (display in Adjust‐ ment mode ) =
1. Confirm the selection with
[cCAL]
[CCAL]
[MODE]
The display toggles between
ð
[CAL / METHOD]
2. Select the method to be adjusted using the
3. Fill the cuvette up to the 10-ml marking with the standard solu‐ tion, without adding reagents
.
[MODE]
key
4. Use the cuvette lid to seal the cuvette
5. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette
6. Press
[ZERO/TEST]
[METHOD]
ð
approximately 8 seconds.
flashes for
44
Page 45
Calibration
Confirmation of the zero cor‐ rection with
7. Remove the cuvette from the sample chamber and empty the cuvette completely
8. Thoroughly clean the cuvette and the cuvette lid
9. Fill the cuvette up to the 10-ml marking with a standard solution of known concentration and introduce the reagents as described under
[0.0.0]
[CAL]
alternates
.
Ä „Chlorine
with liquid reagents 0.01 ... 4.0 mg/l“ on page 22
or
Ä „a) Free
chlorine“ on page 32
10. Use the cuvette lid to seal the cuvette
11. Position the cuvette in the sample chamber
Observe the correct posi‐
ð
tioning of the cuvette.
12. Press
13. If the result matches the value
[ZERO/TEST]
[METHOD]
ð
approximately 3 seconds.
Confirmation of the result alternates with
of the standard used, (within the tolerance to be considered) you can exit Adjustment mode by pressing
flashes for
[ON/OFF]
[CAL]
.
Changing the displayed value:
Pressing the [MODE] key 1 x increases the displayed result by 1 digit.
Pressing the [ZERO/TEST] key 1 x reduces the displayed result by 1 digit.
1. Repeatedly press the keys until the displayed result matches the value of the standard used
2. Press
[ON/OFF]
The new correction factor is
ð
calculated and saved in the user adjustment level.
Confirmation of the adjust‐ ment appears in the display for 3 seconds.
45
Page 46
Calibration
Return to the factory calibration
Returning from user adjustment to factory calibration is only possible simultaneously for all methods.
With a method adjusted by the user, when the result is shown in the display, an arrow appears in the [Cal] position.
Proceed as follows to return the device to the factory calibration, pro‐ ceed as follows:
1. Device is switched off. Simulta‐ neously press
[MODE]
and
[ZERO/TEST]
2. Switch the device on using the
[ON/OFF]
ð
3. The display toggles between:
[SEL]
ð
4. or
5. The display toggles between:
[SEL]
key
After approximately 1 second, release and
[ZERO/TEST]
and
[CAL]
The device is in the as-sup‐ plied condition ( for Select).
and
[cAL]
[MODE]
.
[SEL]
stands
The device operates with an
ð
adjustment performed by the user. If the user adjust‐ ment is to be retained, switch off the device using
[ON/OFF]
the
6. Pressing simultaneously activate factory calibration for all methods
7. The display toggles between
[SEL]
8. Switch the device off using the
[ON/OFF]
[MODE]
and
key
key.
lets you
[CAL]
46
Page 47
Technical Data

6 Technical Data

Device two wavelengths, automatic wavelength selection,
colorimeter with direct measured value display
Optics LEDs, interference filter (IF) and photo sensor at the
transparent sample chamber
Wavelength specifications of the interference filter:
n 530 nm Δλ = 5 nm n 560 nm Δλ = 5 nm
Wavelength precision ± 1 nm
Photometric precision* 3% FS (Full Scale) (T = 20°C ... 25°C)
Photometric resolution 0.01 A (Absorption units)
Power supply 4 batteries (Mignon AA/LR 6)
Operating time approx. 53 h operating time or 15,000 measure‐
ments in continuous operation with background lighting switched off
Auto-OFF Automatic device switch-off 10 minutes after the last
key was pressed
Display Backlit LCD (at the press of a key)
Memory internal ring buffer for 16 data records
Time Real time clock and date
Calibration (manufac‐ turer) / Adjustment (user)
Factory calibration and user adjustment.
Return to factory calibration is possible.
Dimensions 190 x 110 x 55 mm (L x W x H)
Weight Basic device approx. 455 g (with batteries)
*measured with standard solutions
47
Page 48
Technical Data
Ambient conditions Temperature: 5 ... 40 °C
Relative humidity: 30 ... 90% (non-condensing)
Water-tight similar to IP 68 (1 hour at 0.1 m); buoyant device
*measured with standard solutions
The specified device precision is only obtained when the original reagent systems are used.
48
Page 49
Consumables and Spare Parts

7 Consumables and Spare Parts

Consumables
Material Part number
DPD-1 tablets (100 no.) 1061892
DPD-3 tablets (100 no.) 1061893
Phenol red tablets R 175 (100 no.) 305532
Cyanuric acid tablets R 263 (100 no.) 1039744
CHLORINE HR (KI) tablets (100 no.) 1075056
ACIDIFYING GP tablets (100 no.) 1075057
Glycine tablets (20 no.) 1061944
Spare parts
Material Part number
3 no. round cuvettes (d = 24 mm, h = 48 mm) with lid
1007566
(replacement cuvettes) for:
n DPD identification n Phenol red identification n Cyanuric acid identification
5 no. round cuvettes (d = 16 mm, h = 90 mm) with lid
1024072
(replacement cuvettes) for:
n Chlorine HR identification with a tablet
49
Page 50
Standards Complied with and Declaration of Conformity

8 Standards Complied with and Declaration of Conformity

Declaration of Conformity
You will find the EC Declaration of Conformity to download on our home‐ page.
Standards complied with
EC EMC Directive (2004/108/EC)
EN 61326 - 1
50
Page 51

9 Index

Index
1, 2, 3 ...
16 mm cuvette adapter ......... 11
A
Action, step by step .............2
B
Battery compartment cover ...... 12
Battery replacement ........... 12
Bromine with liquid reagents
0.02 ... 9 mg/l ................ 26
Bromine with tablet 0.02 ... 13
mg/l ........................37
C
Chlorine dioxide with liquid
reagents 0.02 ... 7.6 mg/l ....... 27
Chlorine dioxide with tablet
0.02 ... 11 mg/l ............... 38
Chlorine HR with tablet 5 ... 200
mg/l ........................36
Chlorine together with chlorine
dioxide (liquid reagents) ........ 23
Chlorine together with chlorine
dioxide (tablet) ............... 33
Chlorine together with ozone
(liquid reagents) .............. 25
Chlorine together with ozone
(tablet) ..................... 35
Combine chlorine (tablet) ....... 33
Combined chlorine (liquid
reagents) ................... 23
Consumables ................ 49
Countdown .................. 14
Cyanuric acid with CyA test tablet
1 ... 80 mg/l ..................42
D
Danger from hazardous sub‐
stances! ..................... 5
Date and time (24-hour format) ...17
Declaration of Conformity ....... 50
Degree of protection IP 68 ...... 47
Display background lighting ..... 16
Displaying stored data ......... 15
E
EN ISO 7393-2, section 5.3 ...20, 29
Error messages .............. 18
F
Free chlorine (liquid reagents) ... 22
Free chlorine (tablet) ...........32
G
General non-discriminatory
approach .....................2
I
IP 68 ....................... 47
51
Page 52
Index
L
Links to elements or sections of these instructions or other appli‐
cable documents ...............2
M
Material safety data sheet ........5
More symbols ................. 2
N
Non-discriminatory approach ..... 2
O
Ozone with liquid reagents 0.01 ...
2.7 mg/l .....................28
Ozone with tablet 0.01 ... 4 mg/l .. 39
P
pH value with tablet 6.5 ... 8.4 ....40
Photometer leak-tightness ...... 12
Q
Question: For which measured variable can I perform an adjust‐
ment? ...................... 44
Question: How can I read stored
data? .......................15
Question: How can I return to the
factory calibration? ............ 46
Question: How do I adjust the
date and time? ............... 17
Question: How do I carry out a
zero correction? .............. 13
Question: How do I carry out the
zero correction? ............... 8
Question: How do I change the
battery? .....................12
Question: How do I enter the values determined by the adjust‐
ment into the device? .......... 45
Question: How do I perform a
user adjustment? ............. 44
Question: How do I switch the background lighting of the display
off and on? .................. 16
Question: How do I switch the
countdown on? ............... 14
Question: What does the standard
scope of supply include? .........6
Question: What must it clean and
how? ........................8
Question: What sort of consuma‐
bles are available? ............ 49
Question: What spare parts are
available? ................... 49
Question: What type of error mes‐
sages are there? ..............18
Question: Where can I find the
Declaration of Conformity? ...... 50
Question: Which standards are
complied with? ............... 50
R
Repeating the analysis ......... 14
Return to the factory calibration .. 46
52
Page 53
Index
S
Safety data sheets ............. 8
Spare parts .................. 49
Standard scope of supply ........ 6
Standards complied with ........50
Switching on and zero correction . 13
T
Technical Data ............... 47
Total chlorine (liquid reagents) ... 22
Total chlorine (tablet) .......... 32
Z
Zero correction ............. 8, 13
53
Page 54
54 55
Page 55
Page 56
ProMinent GmbH
Im Schuhmachergewann 5 - 11
69123 Heidelberg, Germany
Telephone: +49 6221 842-0
Fax: +49 6221 842-419
Email: info@prominent.com
Internet: www.prominent.com
985629, 3, en_GB
© 2017
Loading...