ProMinent DULCOTEST DT1B Operating Manual

Operating manual
A0943
DULCOTEST® DT1B Photometer
EN
Please carefully read these operating instructions before use. · Do not discard.
The operator shall be liable for any damage caused by installation or operating errors.
The latest version of the operating instructions are available on our homepage.
Supplemental directives
General non-discriminatory approach In order to make it easier to read, this
document uses the male form in grammatical structures but with an implied neutral sense. It is aimed equally at both men and women. We kindly ask female readers for their understanding in this simplification of the text.
Supplementary information
Please read the supplementary information in its entirety.
Information
This provides important information relating to the correct operation of the unit or is intended to make your work easier.
Warning information
Warning information include detailed descriptions of the hazardous situation.
The following symbols are used to highlight instructions, links, lists, results and other elements in this document:
More symbols
Symbol Description
Action, step by step.
Outcome of an action.
Links to elements or sections of these instructions or other applicable documents.
n
List without set order.
2
Symbol Description
Supplemental directives
[Button]
„Display /
GUI“
CODE
Display element (e.g. indicators).
Operating element (e.g. button, switch).
Screen elements (e.g. buttons, assignment of function keys).
Presentation of software elements and/or texts.
3

Table of contents

Table of contents
1 General Information................................................................................... 5
1.1 Scope of delivery............................................................................... 6
1.2 Instructions for use............................................................................ 8
1.3 Adapter for 16 mm cuvettes............................................................. 11
1.4 Operations carried out on the device............................................... 12
2 Commissioning DT 1B............................................................................. 13
2.1 Commissioning................................................................................ 13
3 Operating Menu....................................................................................... 17
3.1 Operating Menu Options.................................................................. 17
3.2 Operating instructions...................................................................... 18
3.3 Error messages............................................................................... 18
4 Analysis methods..................................................................................... 20
4.1 Procedure instructions when using liquid reagents......................... 20
4.2 Quantitative determination using liquid reagents............................. 22
4.3 Procedure instructions when using tablets...................................... 29
4.4 Quantitative determination using tablets.......................................... 31
5 Calibration ............................................................................................... 44
6 Technical Data......................................................................................... 47
7 Consumables and Spare Parts................................................................ 49
8 Standards Complied with and Declaration of Conformity........................ 50
9 Index........................................................................................................ 51
4

1 General Information

WARNING!
Danger from hazardous sub‐ stances!
Possible consequence: Fatal or very serious injuries.
Please ensure when handling hazardous substances that you have read the latest safety data sheets provided by the manu‐ facture of the hazardous sub‐ stance. The actions required are described in the safety data sheet. Check the safety data sheet regularly and replace, if necessary, as the hazard potential of a substance can be re-evaluated at any time based on new findings.
The system operator is respon‐ sible for ensuring that these safety data sheets are available and that they are kept up to date, as well as for producing an associated hazard assess‐ ment for the workstations affected.
General Information
5
General Information

1.1 Scope of delivery

The following components are included as standard:
Description Quan‐
tity
n Case, blue with blue fasteners
with ProMinent sticker, Photometer Dulcotest DT1B, code
1039315
and hazard symbol
Foam insert for the case 1
Cover foam for the case 1
n Photometer DT1B
Chlorine, bromine, chlorine dioxide, ozone, pH, CyA – Battery compartment lid with O-ring
Screwdriver with clip, red 1
Countersunk screws 4
Batteries, 1.5 V alkali-manganese, type AA 4
Round cuvettes, 10 ml, d = 24 mm, h = 48 mm 3
Lid for round cuvette, 24 mm, grey 3
Sealing rings for cuvettes, grey 3
Round cuvettes, d = 16 mm, h = 90 mm 3
Lid for round cuvette, 16 mm, white 3
Adapter attachment, grey for 16 mm cuvettes 1
1
1
Syringe, 10 ml 1
Plastic mixing rod, 13 cm long 1
Cleaning brush 1
Operating instructions for Dulcotest DT1B 1
DPD -1 tablets 100
6
General Information
Description Quan‐
tity
DPD -3 tablets 100
PHENOLRED PHOTOMETER tablets 100
CyA TEST tablets 100
CHLORINE HR (KI) tablets 100
ACIDIFYING GP tablets 100
Glycine tablets 20
7
General Information

1.2 Instructions for use

Request the safety data sheets.
Reagents are intended for chemical analysis and access to them by unauthorised persons must not be permitted.
You must dispose of reagent solutions properly.
Observe the application possibilities, analysis instruction and matrix effects of the methods.
1. You must thoroughly clean the cuvettes, lids and stirring rod after each analysis to avoid carry-over errors.
Even slight reagent residue can lead to incorrect measurement results. For cleaning use the supplied brush.
If the fully reacted water sample is left for any length of time it may pro‐ duce stubborn coloured deposits, which you can remove using dilute (= 4 %) hydrochloric acid.
2. The outer walls of the cuvettes must be clean and dry before the analysis is carried out. Fingerprints or water droplets on the light-entry surfaces of the cuvette will result in faulty measurements. The cuvette should there‐ fore be wiped clean with a soft paper tissue (paper handkerchief) before carrying out the measurement.
3. The zero correction and the analysis must both be carried out using the same cuvette, as the cuvettes may have slightly varying tolerances.
8
A0937
General Information
Fig. 1: Cuvette positioning (Ø 24 mm):
I. Seal II. Marking with the white triangle
4. For the zero correction and test, the cuvette must be positioned in the sample chamber in such a way that the marking with the white triangle (II.) points towards the housing marking.
5. You must carry out the zero correction and test with the cuvette lid closed. You must provide the cuvette lid with a sealing ring (I.) to prevent the entrance of light into the sample chamber.
6. Formation of bubbles on the inside walls of the cuvette will result in faulty measurements. Should this happen, seal the cuvette with the cuvette lid and tip it back and forth to dissolve/remove the bubbles before carrying out the test.
9
A0938
General Information
7. It is important to prevent water getting into the sample chamber because this leads to incorrect measurement results.
8. Dirt in the transparent sample chamber leads to incorrect measurement results. The light entry surfaces should be checked at regular intervals and cleaned as necessary.
Standard spectacle cleaning cloths and cotton buds are suitable for cleaning.
9. Pronounced differences in temperature between the photometer and the surroundings can lead to incorrect measurements, e.g. due to the forma‐ tion of condensation in the sample chamber and on the cuvette.
10. Protect the device from direct sunlight.
Fig. 2: Correct filling of the cuvette: Left = correct / right = incorrect
11. Fill the cuvette as shown in Fig. 2.
12. Insert the reagent tablets directly from the blister pack into the water sample, without touching them with your fingers.
13. After use immediately close the dropping bottles containing the liquid reagents by closing them with the screw caps of the same colour.
14. You must observe the sequence for the addition of reagents without fail.
10

1.3 Adapter for 16 mm cuvettes

A0941
General Information
Fig. 3: Place the adapter for 16 mm cuvettes on the sample chamber
Insert the adapter for the 16 mm cuvette as shown in Fig. 3.
This adapter is required for all analysis methods which must be carried out using a 16 mm cuvette.
11
A0939
General Information

1.4 Operations carried out on the device

Battery replacement:
To ensure complete leak-tightness of the photometer, you must insert the gasket (2) and screw on the battery compartment lid (1).
If the battery is removed for more than 1 minute from the device, then when the device is switched on again with new inserted batteries, the date/time program appears automatically.
Fig. 4: Battery replacement
1. Battery compartment cover
2. Seal
3. Screw
4. Batteries
5. Rear of the device
12

2 Commissioning DT 1B

2.1 Commissioning

Switching on and zero correction
Scroll memory (SM)
With multi-parameter devices the sequence of the various methods is specified. After switching on the device, the method which was last selected before the device was switched off is automatically displayed. This permits rapid access to a favoured method.
1. Switch the device on using the
[ON/OFF]
ð
2. Select the
[MODE]
ð
Zero correction
3. Fill the cuvette up to the 10-ml marking with the water sample, see Fig. 2
4. Close the cuvette using the cuv‐ ette lid
5. Position the cuvette in the sample chamber
ð
6. Press the key
key
The display outputs: The last selected
key
The display outputs:
[METHOD]
Observe the correct posi‐ tioning of the cuvette.
[METHOD]
[METHOD]
.
using the
[ZERO/TEST]
.
Commissioning DT 1B
[METHOD]
ð
approximately 8 seconds.
[0.0.0]
play.
7. Remove the cuvette from the sample chamber
Zero correction is ended.
ð
Analysis
8. Fill the cuvette up to the 10-ml marking with the water sample, see Fig. 2
9. Add reagents to the water sample
The characteristic colouring
ð
develops.
10. Close the cuvette using the cuv‐ ette lid
11. Position the cuvette in the sample chamber
12. Press the so doing adhere to any possibly required reaction time, see
flashes for
appears in the dis‐
[ZERO/TEST]
key, in
Ä
Switching the countdown func‐ tion on“ on page 14
[METHOD]
ð
approximately 3 seconds.
The result is output to the display.
flashes for
13
Commissioning DT 1B
Switching the countdown function on
The result is automat‐ ically saved.
Repeating the analysis
Press the key again
The process runs as
ð
described here
[ZERO/TEST]
on and zero correc‐ tion“ on page 13
New zero correction
Press the key 2 seconds
The process runs as
ð
described here
[ZERO/TEST]
on and zero correc‐ tion“ on page 13
Ä „Switching
.
for
Ä „Switching
.
NOTICE!
If reaction times are not observed, the result may be incorrect measurement results.
The countdown can only be activated directly prior to a measurement.
The countdown time equals 2 minutes and is non-adjustable.
The currently running count‐ down can be ended by pressing the key [ZERO/TEST]. The measurement then takes place immediately.
For methods with a reaction time, you can optionally switch on a countdown function:
1. Press the key the key depressed
[!]
and then keep
2. Press the key
3. Release the
The countdown starts After
ð
the countdown has elapsed (2 minutes), measurement takes place automatically
[ZERO/TEST]
[!]
key
14
Displaying stored data
The device has a ring buffer for 16 data records.
Fig. 5: Displaying stored data
1. Data record (n01 ... n16)
2. Year
3. Month/day
4. Time
5. Measured variable (e.g. chlorine, dependent upon the device ver‐ sion)
6. Value in mg/l
Commissioning DT 1B
1. With the device switched on press the
[!]
key for more than 4 seconds then
release the key again
The display immediately switches to the memory menu.
ð
2. Press the
3. Press the
4. Press the
15
[MODE]
[ZERO/TEST]
[!]
key to scroll through the 16 data records
key to scroll through the values of a data record
key to return to the
[METHOD]
display
Commissioning DT 1B
Display background lighting
Press the
ð
[!]
key
The display's background lighting switches on or off.
During measuring the background lighting switches off automati‐ cally.
16

3 Operating Menu

3.1 Operating Menu Options

Operating menu selection
1. The device is switched off.
Press and hold down the
[MODE]
2. Use the
on the device
ð
3. Release the
4. The [!] key allows you to select
the following operating menu items:
n
n Set date and time n User adjustment
ð
5. Use the
date and time“
item (arrow at top right and bottom left in the display)
key
[ON/OFF]
3 decimal points appear in the display.
[diS]
= read out of stored
data
The selected operating menu item is indicated by an arrow in the display.
[!]
key to select the
key to switch
[MODE]
key
operating menu
„Set
Operating Menu
Setting date and time (24-hour format)
Increase the value to be set by pressing the [MODE] key.
Lower the value to be set by pressing the [ZERO/TEST] key.
Pressing the [!] key takes you to the next value to be set.
1. Press the
ð
2. Enter the year
3. Enter the month
4. Enter the day
5. Enter the hour
6. Enter the minutes
Enter the minutes in 10 minute increments
Press the
Enter minutes in 1 minute incre‐ ments
7. After setting the minutes, press the
ð
[MODE]
The parameter to be set appears for 2 seconds.
key
[YYYY]
[MM]
[dd]
[hh]
[mm]
.
[!]
key
[!]
key
[IS SET]
play and the device auto‐ matically returns to meas‐ uring mode.
appears in the dis‐
17
Loading...
+ 39 hidden pages