D1C ADULCOMETER® Controller Series D1C / Version A
Type of mounting
DControl panel installation 96 x 96 mm
WWall mounting
Operating voltage
0230 V 50/60 Hz
1115 V 50/60 Hz
2200 V 50/60 Hz (only with control panel installation)
3100 V 50/60 Hz (only with control panel installation)
424 V AC/DC
0None
2Temperature via terminal (Pt100)
4Manual temperature input
Feed forward control
0None
4Flow as 0/4-20 mA standard signal and parameter record switching
5Parameter record switching
Control input
0None
1Pause
Signal output
0None
1Standard signal 0/4-20 mA measured value
2Standard signal 0/4-20 mA control variable
3Standard signal 0/4-20 mA correction variable
42 standard signal 0/4-20 mA outputs, free programmable
Power control
GAlarm and 2 limit value relays
MAlarm and 2 solenoid valve relays
RAlarm relay and servomotor with feedback
These operating instructions describe the technical data and function of the series DULCOMETER® D1C
controller, provide detailed safety information and are divided into clear steps.
IMPORTANT
• Please observe the parts of these operating instructions applicable to your particular
version! This is indicated in the Section “Device Identification / Identity Code”!
• Correct measuring and metering is only possible in the case of impeccable operation of
the sensor. The sensor has to be calibrated / checked regularly!
NOTE
A form “Documentation of controller settings type D1C” is available under
www.prominent.com/documentation_D1C for the purpose of documenting the controller
settings.
3
Page 4
3Device Overview / Controls
Conductivity
"Change"
Button
"Branch back"
Button
"Down"
Button
DULCOMETER
STOP
START
Display field
Measured variable
inductive conductivity
Graphic
display
®
"Start/Stop"
Button
"Enter"
Button
®
"Up"
Button
STOP
START
CHANGE Button
To change over within a menu level
and to change from one variable to
another within a menu point.
START/STOP Button
Start/stop of control and metering
function.
ENTER Button
To accept, confirm or save a displayed value or status. For alarm
acknowledgement.
D1C2-cond.-002-D
UP Button
To increase a displayed numerical
value and to change variables (flashing
display).
BRANCH BACK Button
Back to permanent display or to start
of relevant setting menu.
DOWN Button
To decrease a displayed numerical
value and to change variables (flashing
display).
4
Page 5
4Functional Description
NOTE
Please refer to the description of the complete operating menu in Section 8 for a detailed
description of the individual characteristics of the DULCOMETER® D1C controller!
4.1Operating Menu
®
The DULCOMETER
preset and can be changed in the complete operating menu.
The controller is delivered with a restricted operating menu so that the D1C controller can be used
effectively in many applications from the very onset. If adaptations prove to be necessary, all relevant
parameters can then be accessed by switching over to the complete operating menu (see “General
settings”).
4.2Access Code
Access to the setting menu can be prevented by setting up an access code. The D1C controller is supplied
with the access code 5000 which permits free access to the setting menu. The calibration menu remains
freely accessible even if access to the setting menu is blocked by the code.
4.3Control
The D1C can operate as a proportional controller or as a PID controller - depending on the device version
(see identity code) and the setting.
The controlled variable is recalculated once a second. Control procedures which require rapid correction of
setpoint deviations (less than approx. 30 seconds) cannot be processed with this controller. The cycle
times must be taken into consideration when activating solenoid valves (pulse length) in the same way as
their running times when activating servomotors (3-point).
Via the control input pause, the control function (selection of controlled variable) can be switched off. The
calculation of the controlled variable starts again after cessation of "pause".
D1C controller permits settings to be made in two different menus. All values are
4.4Feed Forward Control
The D1C controller can process a signal of a feed forward control. Depending on the device version (see
identity code) and the setting, this signal can be obtained in any form of a 0–20 mA or 4–20 mA signal or as
a digital contact signal with the maximum frequencies 10 Hz or 500 Hz.
This signal can be used, for example, for flow-proportional metering (multiplicative effect) or feed forwarddependent basic load metering (additive effect). The result of control variable calculation from the
proportional or PID control is multiplied by or added to the feed forward signal. A multiplicative feed forward
variable at the level of the set rated value carries over the calculated control variable unchanged into the
controlled variable:
Controlled variable = Feed forward variable/rated value x calculated control variable
During start-up, the zero point has to be checked. The multiplicative feed forward control is not designed
for switching off permanently the actuating variable (signal ≈ 0).
An additive feed forward variable at the level of the rated value results in maximum controlled variable:
Controlled variable (max. 100 %) = Feed forward variable/rated value x max. controlled variable
+ calculated control variable
4.5Error Messages
Error messages and information are indicated on the bottom line in the permanent display 1. Errors to be
acknowledged (acknowledgement switches off the alarm relay) are indicated by the "
". Errors/notes
which still apply after acknowledgement are indicated alternately. During correction variable processing
(temperature for correction of pH-value), the value is indicated in the same line as the error/note. Faults
which are rectified of their own accord due to changed operating situations are removed from the
permanent display without the need for acknowledgement.
5
Page 6
5Display Symbols
The display of the DULCOMETER
®
D1C controller uses the following symbols:
DescriptionCommentSymbol
Limit value transgressionSymbol
Relay 1 upperleft
Relay 1 lowerleft
Relay 2 upperright
Relay 2 lowerright
Metering pump 1 (Increase conductivity)Symbol
Control offleft
Control onleft
Metering pump 2 (Reduce conductivity)Symbol
Control offright
Control onright
Symbol
Symbol
Symbol
Symbol
Symbol
Solenoid valve 1 (Increase conductivity)Symbol
Control offleft
Symbol
Control onleft
Solenoid valve 2 (Reduce conductivity)Symbol
Control offright
Symbol
Control onright
Servomotor
Control, open relay
Control, close relay
Without control
Thickness of bar
Position feedbackincreases from left to right
during opening.
Stop button pressed
Manual metering
Fault
6
Page 7
6Operation diagram
Permanent
display 1
The various menus are selected with
the CHANGE button
The menu is started with
the ENTER button
Permanent
display 2
Various
Setting menus
Access code
NOTE
Access to the setting menus can be barred with the access code!
The number and scope of setting menus depends on the device version!
BRANCH BACK to permanent display
or to relevant setting menu
Access code, correct
Parameter
setting
D1C2-cond.-007-GB
If the access code is selected correctly in a setting menu, the following setting menus are also
accessible!
If within a period of 10 minutes no button is pushed, the unit automatically branches back from
the calibrating menu or a setting menu to the permanent display 1!
BRANCH BACK without
saving setting
BRANCH BACK to
start of setting
Text 1
Selection 1
Text 2
Selection 2
CHANGE from selection to selection
Change numbers or
settings of selection
Variables flash
Text 1
Selection 1
Text 2
Selection 2
ENTER and save setting,
continue to next menu
D1C2-cond.-008-GB
7
Page 8
7Restricted Operating Menu / General Layout
The restricted operating menu permits simple operation of the most important parameters. The following
overview shows the settings which can be selected:
µS/cm
100
mea val.
fd. fwd.
reg. val.
w:
meas. parameter
setting ?
calibration cc ?
parameter record
setting ?
100 µS/cm
70 %
59 %
1000 µS/cm
PAR1
Only with parameter record 2
PAR1
Permanent display 1
Permanent display 2
only with control
(w = setpoint)
mr = meas. range
cc = cell constant
calibration cc
probe in buffer!
accept value?
1538 µS/cm 8.54
calibration cc
cc: 8.50 /cm
parameter record
Grenzwert2oben
5.00 ppmGrenzwert1unten
1.00 ppm
PS1
Positive values of setting variable:
Negative values of setting variable:
2000 µS/cm
mr :
cc :
1106 µS/cm 25,0 °C
calibration cc
cc: 8.54 /cm
cond: 100 µS/cm
8.54/cm
mr 2 : 2000 µS/cm
cc 2 : 8.54 /cm
1106 µS/cm 25,0 °C
please wait!
Increase conductivity
Reduce conductivity
Number and scope of
setting menus depends
on the device.
Access to setting menus
can be blocked with
access code.
limit PS1
setting ?
Only with control
control
settings ?
Setting in
complete
operating
menu
general setting
information
limit2upper
Grenzwert2oben
1500 µS/cm
5.00 ppm
limit1lower
Grenzwert1unten
1000 µS/cm
1.00 ppm
control
regulated value:
positive cond
negative cond
control
normal
current regulat.
value:
30 %
or
control
with dead zone
current regulat.
value:
30 %
or
control
manual
current regulat.
value:
ident-code: D1CA
DxL6xxxxxxxxx
software version
D1C-L0 FW-2.03
30 %
➞
➞
For normal control
setpoint
1000 µS/cm
Control with dead zone
setpoint 2 upper
1500 µS/cm
setpoint 1 lower
1000 µS/cm
Proportional control
ctrl. parameter
xp =
10 %
PID control
ctrl. parameter
xp =
10 %
0 s
Ti =
Td =
0 s
Manual control
manual dosing
15 %
regulated range
input HW:
rev. y
input SW:
xx.xx.xx
alarm relais
active
access c.:
operating menu:
english
=
=
restricted
5000
D1C2-cond.-010-GB
8
Page 9
Restricted Operating Menu / Description
µS/cm
Permanent display 1
100
mea val.
fd. fwd.
reg. val.
Conductivity calibration
Measurement parameters
meas. parameter
setting ?
100 µS/cm
70 %
59 %
1000 µS/cmw:
Permanent display 2
only with control
(w = setpoint)
Access to all setting menus can be blocked with an access code !
mr = meas. range
cc = cell constant
Positive values of setting variable: Increase conductivity
Negative values of setting variable: Reduce conductivity
Cell constant ccdepending on0.0001/cm0.0060/cm0.1499/cmcc can be adjusted for
sensor type0.001/cm0.150/cm1.499/cmall mr over the
used0.01/cm1.50/cm12.00/cmcomplete area
The measured value can be calibrated by changing the cell constant to the actual conductivity value
(arrow keys).
Cell constant calibration (cc)
calibration cc ?
PAR1
calibration cc
probe in buffer!
accept value?
1538 µS/cm 8.54
calibration cc
cc: 8.50 /cm
calibration cc
cc: 8.54 /cm
cond: 100 µS/cm
please wait!
D1C2-cond.-040-GB
9
Page 10
Restricted Operating Menu / Description
To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration solution (sample 1)
with a known conductivity and select the second menu option. The DULCOMETER® D1C displays the
conductivity it has calculated with the updated parameters. Press Enter when the value is constant.
Select the next menu option and enter the conductivity of the calibration solution (arrow keys). Press Enter.
The D1C displays the recalculated cell constant and saves the value.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops.
The limit value monitor and the fault diagnosis system are reset. The standard signal of the “measured value“
output is frozen.
This setting menu can be used to temporarily enable additional parameters of the second parameter record
PAR2 to be edited (in addition to the parameters from the setting menu “measurement parameter setting?“).
I.e. the “limit setting?“ and “mA output 1 (2) setting?“ setting menus are enabled to allow editing of PAR2
(designation PS2) until you exit the series of setting menus from “parameter record“ to “general setting“.
It is possible to switch between the parameter records for operation using the “feed forward control“ contact
input, e.g. during a flushing process (contact open = PAR1, contact closed = PAR2). When switching between
the parameter records, the current parameter record in the setting menu “measurement parameter
setting?”, “limit setting?“ and “mA output 1 (2) setting“ is replaced by the other parameter record
(designation PS1 or PS2).
There is no access to the calibration menus in the case of PAR2 (designation PS2).
NOTE
• The designation PAR1 indicates that the parameter record 1 is actively measuring (contact
input “feed forward control“ e.g. open).
The designation PS1 indicates that the parameter record 1 can be edited.
Exception: The parameter record 1 cannot be edited while it is actively measuring
(designation PAR1). If your controller switches the D1C to parameter set 1 while it is being
edited, the D1C returns to the permanent display.
• PAR2, no control, no standard signal outputs.
10
Page 11
Restricted Operating Menu / Description
Limit values
Access to all setting menus can be blocked with an access code !
Grenzwert2oben
limit PS1
setting ?
Type of limitLimit 1:lowerlowerLimit transgression when
transgressionLimit 2:upperupperexceeding or dropping
Setpoint100 µS/cm0,1 µS/cm-10 µS/cm210 µS/cmMeasuring range 200 µS/cm
1000 µS/cm1 µS/cm-100 µS/cm2100 µS/cmMeasuring range 2000 µS/cm
10 mS/cm0,01 mS/cm-1 mS/cm21 mS/cmMeasuring range 20 mS/cm
100 mS/cm0,1 mS/cm-10 mS/cm210 mS/cmMeasuring range 200 mS/cm
1000 mS/cm1 mS/cm-100 mS/cm2100 mS/cmMeasuring range 2000 mS/cm
Control with dead zone
2 setpoints necessary
Setpoint 2 ≥ Setpoint 1
Adjustment of measuring
range on page 9/14
Control parameter xp10 %1 %1 %500 %xp referred to measuring range
Control parameter Tioff1 s1 s9999 sFunction off = 0 s
Control parameter Tdoff1 s1 s2500 sFunction off = 0 s
Manual metering0 %1 %-100 %+100 %
Abbreviations for control variables:
xp: 100 %/Kp (inverse proportional coefficient)
: Integration time of I-controller (s)
T
i
Td: Differential time of D-controller (s)
General settings
Access to all setting menus can be blocked with an access code !
general setting
information
ident-code: D1CA
DxL6xxxxxxxxx
software version
D1C-L0 FW-2.03
All parameters of the controller can be set in the complete operating menu (see previous page). The
following overview shows the settings which can be selected:
µS/cm
Permanent display 1
100
mea. val.:
fd. fwd.:
reg. val.:
w:1000 µS/cm
meas. parameter
setting ?
cali. zero point ?
calibration cc ?
calibration α
cc :
α :
install. factor
setting ?
100 µS/cm
70 %
-59 %
PAR1
PAR1
8,54/cm
1,90 %/°C
PAR1
Permanent display 2
only with control
(w = setpoint)
Number and scope of setting menus
depends on the device.
Access to setting menus can
be blocked with access code.
limits PS1
setting ?
servomotor
setting ?
Only with
servomotor
type of sensor
setting ?
correction value
setting ?
pumps
setting ?
relays
setting ?
parameter record
setting ?
PAR1
Only with correction
variable
Only with pumps
Only with limit value
relay, solenoid valve
relay or servomotor
control
setting ?
feed forward ctrl.
setting ?
mA output 1 PS1
setting ?
mA output 2 PS1
setting ?
general setting
information
Only with control
Only with feed
forward control
Only with standard
signal output
Only with 2 standard signal outputs
D1C2-cond.-017-GB
13
Page 14
Complete Operating Menu / Description
µS/cm
100
mea. val.
fd. fwd.
reg. val.
Measuring parameter
100 µS/cm
70 %
59 %
1000 µS/cmw:
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
Permanent display 1
Permanent display 2
only with control
(w = setpoint)
mr: meas. range
cc: cell constant
α: temperature-
coefficient
Positive values of setting variable: Increase conductivity
Negative values of setting variable: Reduce conductivity
D1C2-cond.-018-GB
mr :
cc :
α :
15 mS/cm 23,8 °C
20 mS/cm
8,54/cm
1,90%/°C
mr (2) : 200 µS/cm
cc (2) : 1,120 /cm
α : 1,90%/°C
1106 µS/cm 25,0 °C
Measuring ranges mr0....2000 µS/cm0....200 µS/cmAt mr 2: 0 = off
0....2000 µS/cm(parameter record
0....20 mS/cmswitching)
0....200 mS/cmSetpoint- and limit
0....2000 mS/cmvalues are switched
to the associated
initial values
Cell constant ccdepending on0.0001/cm0.0060/cm0.1499/cmcc can be adjusted
the sensor type0.001/cm0.150/cm1.499/cmfor all mr over the
used0.01/cm1.50/cm12.00/cmcomplete area
Temperature coefficient α1.90 %/°C0.01 %/°C0 %/°C10 %/°C
The measured value can be compensated for by changing the cell constant to the actual conductivity value
(arrow keys!). The precondition is a known temperature coefficient and constant temperature.
Metering and control stops if the measurement range is changed.
Setpoints, limit values and standard signal outputs are switched to the associated starting values. Check
the settings in all menus.
The parameter sets are described on page 21.
14
Page 15
Complete Operating Menu / Description
Calibration of zero point (zp)
cali. zero point ?
PAR1
cali. zero point
probe in air!
accept value?
12 µS/cm
please wait!
D1C2-cond.-039-GB
To determine the precise zero point (zp) of the sensor, hold the sensor in the air and select the second
menu option. The DULCOMETER® D1C displays the conductivity without zero point correction. Press
Enter when the value is constant.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the “measured value“ output is
frozen.
Cell constant calibration (cc)
calibration cc ?
PAR1
calibration cc
probe in buffer!
accept value?
1538 µS/cm 8.54
calibration cc
cc: 8.50 /cm
calibration cc
cc: 8.54 /cm
LF: 100 µS/cm
please wait!
D1C2-cond.-043-GB
To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration solution
(sample 1) with a known conductivity and select the second menu option. The DULCOMETER® D1C
displays the conductivity which it has calculated with the current parameters. Press Enter when the value
is constant.
Select the next menu option and enter the conductivity of the calibration solution (arrow keys). Press Enter.
The D1C displays the recalculated cell constant and saves the value.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the “measured value“ output is
frozen.
ConductivityMeasured value0.1 µS/cm0 µS/cm200 µS/cmMeasuring range 200 µS/cm
Calibration1 µS/cm0 µS/cm2000 µS/cmMeasuring range 2000 µS/cm
solution (LF)0.01 mS/cm0 µS/cm20 mS/cmMeasuring range 20 mS/cm
0.1 mS/cm0 µS/cm200 mS/cmMeasuring range 200 mS/cm
1 mS/cm0 µS/cm2000 mS/cmMeasuring range 2000 µS/cm
15
Page 16
Complete Operating Menu / Description
Calibration of temperature coefficient α
The temperature coefficient α is determined again by means of a two point calibration. During the
calibration, the metering is reduced to the set basic load and control stops. The limit value monitor and the
fault diagnosis system are reset. The standard signal of the measured value or correction value output is
reduced to 0/4 mA.
IMPORTANT
Only values determined at 25 °C may be entered as conductivity values. The calibration
must be carried out with the same solution at both temperatures.
Access to all setting menus can be blocked with an access code !
calibration α
cc :
α :
8,54/cm
1,90 %/°C
PAR1
calibration at
temperature 1
probe in buffer !
153823,8 °C
calibration at
temperature 2
probe in buffer !
150048,5 °C
calibration α
cc :
α :
8,50/cm
2,52 %/°C
calibration at
temperature 1
accept value ?
153823,8 °C
calibration at
temperature 2
accept value ?
150048,5 °C
Enter the installation factor of the installed sensor here (arrow keys!). Determine the installation factor using
the data from the sensor documentation and the installation geometry.
identity codeSolenoid valve (SV1, SV2)relays remain active
Limit value (Limit 1/2)even with a fault.
Actuator 1,2
Servomotor
off
Cycle10 s1 s10 s9999 sfor solenoid valve
min. time1 s1 s1 scycle/2for solenoid valve
NOTE
The limit value relays can also be defined such that they react like an actuator. If e.g. a limit value
relay picked up, it will drop at a closed pause contact or for a subsequent delay time t
(if td > 0 min
d
is set in “General settings”).
19
Page 20
Complete Operating Menu / Description
Cycle
Solenoid
valve
on
off
on
off
The switching time of the DULCOMETER® D1C (solenoid valve) depends on the actuating variable and the
“min. time“ (smallest permitted operating factor of the connected device).
The actuating variable determines the ratio ton/cycle and thus the switching times (see fig. above).
The “min. time“ influences the switching times in two situations:
a) theoretical switching time < min. time:
Cycle
min. time
on
off
Cycle
on
off
t
on
t
on
Cycle
Cycle
Cycle
Cycle
Cycle
min. time
min. time
Actuating
variable: 80 %
t
on
Cycle
t
Actuating
variable: 50 %
t
on
Cycle
t
theoretical
t
actual
t
= 0.80
= 0.50
The DULCOMETER® D1C does not switch for a certain number of cycles until the sum of the theoretical
switching times exceeds the “min. time“. Then the DULCOMETER® D1C switches for the duration of this
total time.
b) theoretical switching time > (cycle - min. time) and calculated switching time < cycle
Cycle
min. time
on
off
Cycle
on
off
Cycle
Cycle
Cycle
Cycle
min. time
theoretical
t
actual
t
The DULCOMETER® D1C does not deactivate for a certain number of cycles until the differences between
cycle and theoretical switching time exceed the "min. time".
20
Page 21
Complete Operating Menu / Description
Parameter record
only with parameter record 2
parameter record
parameter record
setting ?
PAR1
Parameter recordPS1PS1
This setting menu can be used to temporarily enable additional parameters of the second parameter record
PAR2 to be edited (in addition to the parameters from the setting menu “measurement parameter
setting?“). I.e. the “limit value setting“ and “mA output 1 (2) setting?“ setting menus are enabled to allow
editing of PAR2 (designation PS2!) until you exit the series of setting menus from “parameter record“ to“general setting“.
It is possible to switch between the parameter records for operation using the “feed forward control“
contact input, e.g. during a flushing process (contact open = PAR1, contact closed = PAR2). When
switching between the parameter records, the current parameter record in the setting menus “meas.
parameter setting?”, “limit setting?“ and “mA output 1 (2) setting“ is replaced by the other parameter
record (designation PS1 or PS2).
There is no access to the calibration menus in the case of PAS2 (designation PS2!).
• The designation PAR1 indicates that the parameter record 1 is actively measuring (contact
input “feed forward control“ e.g. open).
The designation PS1 indicates that the parameter record 1 can be edited.
Exception: The parameter record 1 cannot be edited while it is actively measuring (designation
PAR1). If your controller switches the D1C to parameter record 1 while it is being edited, the
D1C returns to the permanent display.
• PAR2, no control, no standard signal outputs.
21
Page 22
Complete Operating Menu / Description
Limit values
Access to all setting menus can be blocked with an access code !
Checkout time limits t onoff1 s1 s/off9999 sResults in message and
alarm.
off = 0 s:
Function switched off,
no message, no alarm
Switch-on delay0 s1 s0 s9999 sOnly available in
∆ t oncomplete operating menu
Switch-off delay0 s1 s0 s9999 sOnly available in
∆ t offcomplete operating menu
Max. switch-on timeoff1 s0 s/off9999 sOnly available in
Limit value 1;complete operating menu
Limit value 2Function switched off
If the limit transgression is longer than the “delay time limit value“, an error message is triggered (requires
acknowledgement) and the alarm relay opens; if “control“ is also set to “off“, the control process will stop.
22
Page 23
Complete Operating Menu / Description
Servomotor
The operating range is given by the overall resistance range of the feedback potentiometer. Defining the
control range places a maximum limit on the control range used.
IMPORTANT
• The servomotor must be actuated with the same care as for the calibration of a measuring
sensor.
• To ensure correct function, the control duration of the servomotor must be at least
25 seconds for 0...100 % of the control range.
Access to all setting menus can be blocked with an access code !
Output range0…20 mA0…20 mANo setting in case of PAR2
4…20 mA
3.6/4-20 mAReduction to 3.6 mA
when alarm relay switches
(not limit value violation)
Range measured0–200 µS/cm0.1 µS/cm-10 µS/cm210 µS/cmat meas. range up to 200 µS/cm
value0–2000 µS/cm1 µS/cm-100 µS/cm2100 µS/cm at meas. range up to 2000 µS/cm
0–20 mS/cm0.01 mS/cm-1 mS/cm21 mS/cmat meas. range up to 20 mS/cm
0–200 mS/cm0.1 mS/cm-10 mS/cm210 mS/cmat meas. range up to 200 mS/cm
0-2000 mS/cm1 mS/cm-100 mS/cm2100 mS/cm at meas. range up to 2000 mScm
Range regulated0 %…+100 %1 %-100 %+100 %Minimum range 1 %
valueNo setting in case of PAR2
Range correction0…100,0 °C0,1 °C0,0 °C100 °CMinimum range 1 %
variable-199.9 °C+199.9 °CNo setting in case of PAR2
General setting
Access to all setting menus can be blocked with an access code !
general setting
information
ident-code: D1CA
DxL6xxxxxxxxx
software version
D1C-L0 FW-2.03
26
input HW:
rev. y
input SW:
xx.xx.xx
control input
– active closed
– parameter set 2
alarm relay
active
acces c.:
operating menu:
=
english
complete
=
5000
pause normal
– active...closed
– alarm off
– td: 0 min
Input controlactive closedactive closedSwitch status for
Inputactive open parameter record 2
Only with parameter
record switching
Control inputParameter set 2Parameter set 2off = parameter set
Functionoffswitching off
Only with parameter
record switching
Access code5000119999
Languageas peras per
identity codeidentity code
Operating menucompleterestricted
complete
Normal pause
If the pause switch is off, the DULCOMETER® D1C sets the operating outputs to “0“ for as long as the
pause switch is off or for a set time delay td (if td is set to > 0 min). Whilst the pause switch is off, the
DULCOMETER® D1C establishes the P-proportion in the background.
With PID-control (identity code characteristics “control characteristic“ : 2): the I-porportion is stored when
the pause is switched off (I-porportion then usually only present if Ti > 0 has been selected in the “control
setting?“ setting menu).
Exception: the standard signal outputs mA for the measured value or correction value are not affected by
the pause.
After pause is activated, the operating outputs remain at “0“ for the length of the time delay td. The time
delay td must be set up in such a way that in this time e.g. sample water (process-specific current
concentration) flows to the sensor.
With PID-control (identity code characteristics “control characteristic“: 2): The control variable output
resulting from the pause and the expiry of the time delay td is reconciled jointly with the current
P-component and (if Ti is set > 0) with the stored I-component.
Pause hold
If the pause switch is off, the DULCOMETER
®
D1C freezes the operating output at the most recent value
for as long as the pause switch is off or for a set time delay td (if td is set to > 0 min). Whilst the pause
switch is off, the DULCOMETER® D1C establishes the P-proportion in the background.
With PID-control (Identity code characteristics “control characteristic“ : 2):
Even the mA standard signal outputs for measured value or correction value are frozen.
After pause is activated, the operating outputs remain frozen for the length of the time delay td. The time
delay td must be set up in such a way that, in this time e.g. sample water (process-specific current
concentration) flows to the sensor.
With PID-control (identity code characteristics “control characteristic“: 2): The control variable output
resulting from the pause and the expiry of the time delay td is reconciled jointly with the current
P-proportion and (if Ti is set > 0) with the newly established I-proportion.
27
Page 28
9Fault / Remarks / Troubleshooting
Calibration
Calibration
with error
During servomotor setting
Position feedback
Upper position <40 % max. value
Lower position >30 % range
Stop button
Check measurement
Final value high
Check direction
Final value low
range
Stop
Basic loadStop––Repeat calibration
StopStopnoRelay drops out–
Operating stepNote textSymbolEffect
Pause contact
Control input
Parameter set 2
Pause/Hold
Pause
on meteringon control
StopStopno/yesNo further fault check–
StopStopno
Electronic error
position not reached
amplifier defective,
LF cable defective,
check SPI
System error
defective
Basic loadStopyes
StopStopyesElectronic data defectiveCall customer service