Promax PROMAX-10 Premium User Manual

Page 1
USER’S MANUAL. PROMAX-10 Premium
TABLE OF CONTENTS
1 GENERAL...............................................................................................................1
1.1 Description........................................................................................................ 1
1.2 Specifications....................................................................................................2
2 SAFETY RULES.....................................................................................................9
2.1 Generals ..........................................................................................................9
3 INSTALLATION.....................................................................................................13
3.1 Power supply ..................................................................................................13
3.1.1 Battery charge ..........................................................................................13
3.1.2 Recommendations for using the battery....................................................14
3.2 Installation and start-up...................................................................................14
3.2.1 Contrast adjustment ..................................................................................14
4 OPERATING INSTRUCTIONS..............................................................................15
4.1 Description of the controls and elements.........................................................15
4.2 Operating instructions.....................................................................................17
4.2.1 Global configuration menu ........................................................................19
4.2.2 SCAN operating mode..............................................................................22
4.2.2.1 SCAN mode configuration...................................................................23
4.2.3 CHANNEL-FREQUENCY operating mode................................................24
4.2.3.1 Channel tuning....................................................................................24
4.2.3.1.1 Video carrier + V/A + C/N measurement.......................................24
4.2.3.1.2 Audio carrier measurement and demodulation..............................25
4.2.3.1.3 CSO-CTB distortion products measurement.................................25
4.2.3.1.4 Power and C/N ratio measurement of digital channels (DVB-C /
DVB-T / DAB)...............................................................................27
4.2.3.1.5 Constellation Diagram representation, Bit Error Rate (BER) and Modulation Error Ratio (MER) measurement in digital channels. ..29
4.2.3.1.6 Configuration of QAM signal measurement using edit channel plan....29
4.2.3.2 Frequency tuning and configuration....................................................30
4.2.3.3 CHANNEL-FREQUENCY mode configuration....................................33
4.2.4 SPECTRUM ANALYSER operating mode................................................34
4.2.4.1 SPECTRUM operating mode. .............................................................34
4.2.4.2 MAX operating mode..........................................................................35
4.2.4.3 MIN operating mode...........................................................................36
4.2.4.4 TRANSIENT DETECTOR operating mode..........................................36
4.2.4.5 SPECTRUM ANALYSER mode configuration.....................................37
4.2.5 TILT operating mode.................................................................................38
4.2.5.1 TILT mode configuration.....................................................................38
4.2.6 DATALOGGER operating mode................................................................39
4.2.6.1 Datalogger configuration.....................................................................41
4.3 Connection to a Computer or Printer............................................................... 42
Englis h
Page 2
USERS MANUAL. PROMAX-10 Premium
CABLE TV ANALYSER
PROMAX-10 Premium
1 GENERAL
1.1 Description
The
Datalogger, Scan, Tilt, Spectrum Analyser, Transient Detector Analyser analogue and digital 862 MHz
(CATV) and
in the whole frequency band, very useful to check a possible input s aturation of some broadband demodulators.
measurements:
and stored, each with carrier levels, C/N and V/A ratios, channel power or MER of up t o 140 channels. The measurements obtained may be checked, transferred to a PC or printed at any time.
PROMAX-10
which makes it an excellent tool for the installation and maintenance of
range, which includes
wireless cable TV
The
PROMAX-10
As a
Level Meter
Analogue channels
Digital channels DVB-C, DVB-T and DAB
The
Datalogger
Premium
television signal reception/distribution syst ems working in the
Premium
- Video carrier level measurement
- Carrier / Noise ratio measurement (C/N)
- Video / Audio ratio measurement (V/A)
- CSO and CTB Intermodulation distortion measurement
- Channel power measurement by integration
- Carrier / Noise ratio measurement (C/N)
- Bit Error Rate (BER) measurement
- Modulation Error Ratio (MER) measurement
- Graphical representation of DVB-QAM signals Constellation
Diagram.
function allows up to 55 loggers or measurements t o be taken
is seven functions in one instrument,
FM radio, community TV systems
(MMDS) including the
incorporates the power level measurem ent f unction
, the
PROMAX-10
:
sub-band
Premium
:
, and
(MATV),
(return path).
enables the following
Level Meter
Cable Digital
5 to
cable TV
,
Englis h
In
operating mode, the channels present on the band in a bar-graph display. The span and reference level are user definable. A moving marker shows the exact level of each specific channel.
difference in levels between any four previously defined pilot frequency channels, in order to obtain a qualitative measurement on band equalisation.
05/2004 Page 1
Scan
In
operating mode, the screen shows, both graphically and numerically, the
Tilt
PROMAX-10
Premium
indicates the level of all
Page 3
USERS MANUAL. PROMAX-10 Premium
As a
Spectrum Analyser
it provides the analysis of the entire band, with a span defined by the user from 1 to 100 MHz. Furthermore, it is possible to alter the reference level, and to detect and maintain the
INGRESS
measurements.
The
Transient Detector
mode, permits to count the number of impulsive
maximum
and
minimum
values for
interferences in the return path and with a level higher than a threshold defined by the user. The maximum frequency range is from 5 to 100 MHz.
In designing the
PROMAX-10 Premium
, special attention was given to building a multi-purpose, accurate, yet easy-to-use instrument. A simple key board gives direct access to the various operating modes and, once in them, any measurement parameter can be easily changed using the rotary selector/button.
In addition, it has an
RS-232C
terminal for connecting to a printer or computer
for producing reports on the measurements obtained.
The instrument is powered by means of an internal rechargeable battery. All these functions have been brought together in one instrument weighing only
half a kilo. The ergonomic, sturdy design makes the
PROMAX-10 Premium
a worki ng
tool without parallel.
1.2 Specifications
TUNING Tuning range Tuning mode Channel plan
Resolution Indication Channel frequency offset
From 5 to 862 MHz. By channels or by frequency. 10 channel plans, each one with a maximum of 140 channels. Factory start-up channel plans: CCIR, EIA, OIRL, JAP2, UK, FCC, STD2L, AUST
(1)
. 10 kHz. Graphic LCD with back lighting.
2 MHz (10 kHz resolution).
±
POWER LEVEL MEASUREMENT Measuring range Bandwidth Resolution Accuracy
From 70 to 120 dBµV (From 10 dBmV to 60 dBmV From 5 to 862 MHz 1 dB
3 dB (From 5 to 40 °C)
±
(2)
LEVEL MEASUREMENT Measurement
Analogue channels Digital channels
Video carrier level measurement. Power measurement in the channel bandwidth by integration method.
Measuring range
Page 2 05/2004
From 25 to 120 dBµV. (From -35 dBmV to 60
(2)
dBmV)
.
).
Page 4
USERS MANUAL. PROMAX-10 Premium
Maximum input level
From 5 to 862 MHz DC to 60 Hz
120 dBµV. (60 dBmV 60 V DC or RMS
(2)
).
Reduction of the measurement range depending on the number of channels.
Up to 10 channels From 11 to 20 channels From 21 to 50 channels From 51 to 80 channels
Readout
110 dBµV. 107 dBµV. 103 dBµV. 101 dBµV. Digital in dBµV, dBmV or dBm and analogue through a graphic bar. 1 dB resolution
IF bandwidth Input impedance Audible indicator
230 kHz ± 50kHz 75
Tone which varies with the signal level.
Accuracy
Analogue channels
Digital channels
2 dB (from 0 to 40 °C) for negative video
±
modulation
3 dB (from 0 to 40 °C) for 8 MHz bandwidth
±
(3)
.
channels.
DIGITAL SIGNALS MEASUREMENT MER (Modulation error ratio)
Measurement range Accuracy
22 dB to 34 dB for 64 QAM.
2 dB
±
BER (Bit error rate) Measured before RS decoding
Measurement range
Constellation Diagram
10 E-2 to 10 E-8 DVB-QAM signals (Annex A/B/C) and DOCSIS / Euro-DOCSIS
Lock range Symbol rate
Measurement range
Datalogger
-10 dBmV to 60 dBmV
1000 to 7200 Msym/s
*
for 16/64/256 QAM For each digital channel, the level and the MER can be stored. (BER for data dumping to printer or transfer to PC)
Modulation type
16/32/64/128/256 QAM ITU J1 annex A/C and 64/256 QAM ITU J1 annex B.
Bandwidth Frequency tuner
8 MHz
62.5 kHz.
Englis h
VIDEO / AUDIO RATIO MEASUREMENT Measurement Measurement range
(ANALOGUE CHANNELS) Ratio of video to audio carrier levels. From 0 to 40 dB
Audio subcarrier frequency
Variable
Accuracy
*
Including the OP-010-E option, otherwise the margin is from 1000 to 7000.
4-9 MHz. ± 2 dB (from 5 to 40 °C) for FM audio carrier
(4)
.
Page 5
USERS MANUAL. PROMAX-10 Premium
CARRIER / NOISE RATIO MEASUREME NT Measurement
Analogue channels
Ratio between carrier level and the channel's noise level.
Digital channels
Ratio between the channel power and the noise level. The frequency where noise is measured is user definable in absolute or relative value. In the relative mode, the unit takes as defaul t frequency
Measurement range
Analogue channels
(5)
offset the value BW/2 + 0.5 MHz. 40-50 dB for input level between 60 and 70 dBµV.
> 50 dB for input level > 70 dBµV.
Digital channels
Accuracy
> 30 dB for input level > 60 dBµV.
2 dB (45 – 862 MHz) ± 3 dB (5 – 45 MHz)
±
CSO & CTB INTERMODULATION PRODUCTS MEASUREMENT
(ANALOGUE CHANNELS)
CSO
Ratio of the peak level of the video carrier to the peak of the distortion products of second order beat. Measured at four frequencies user definable.
Measuring frequencies
User definable from -2.50 to -0.50 MHz and from
0.50 to 2.50 MHz. (Default values -1.50, -0.50, 0.50 and 1.50 MHz).
CTB
Ratio of the peak level of the video carrier to the peak of the distortion products of third order beat. Measured at the carrier frequency or, approximately mode, in a channel previously defined by the user.
DATALOGGER FUNCTION Max. number of loggers Number of channels/logger
55 140
Measurements
Analogue channels Digital channels
Level, C/N and V/A. Channel power and MER. (BER for data dumping to printer or transfer to PC).
SCAN Span
Variable: 10, 30, 100, 300 MHz and full band (from 5 to 862 MHz, according to the channel plan).
Power detection Reference level
Forward band (45 to 862 MHz). Variable from 60 to 120 dBµV in 10 dB steps.
TILT Indication Analysed band
Numerical and by level bar. Forward (45 to 862 MHz) or Return path (5 t o 100 MHz).
Number of pilots Pilots frequency Pilots resolution
Page 4 05/2004
4 per band. From 5 to 862 MHz. 10 kHz.
Page 6
USERS MANUAL. PROMAX-10 Premium
SPECTRUM ANALYSER Span Reference level Analysed band
Detector Bandwidth Resolution
Peak detector
Span 100 MHz Span 50 MHz Span 30 MHz Span 15 MHz Span 5 MHz Span 1 MHz
Average detector
Span 30 MHz Span 15 MHz Span 5 MHz Span 1 MHz
TRANSIENT DETECTOR
Detection threshold Detection range Readout
AUDIO Demodulation
Output
From 1 to 100 MHz (1, 5, 15, 30, 50, 100 MHz). Variable from 60 to 120 dBµV in 10 dB steps. Forward (45 to 862 MHz) or Return path (5 to 100 MHz). Peak or average. 230 kHz.
900 kHz. 450 kHz. 280 kHz. 140 kHz. 50 kHz. 10 kHz.
280 kHz. 140 kHz. 50 kHz. 10 kHz.
From 20 to 60 dBµV in steps of 1 dB. From 5 to 100 MHz maximum. Number of detected transitory in the measuring time. Present detected level and maximum detected level in the time of the measurement.
AM / FM / LEVEL (variable tone according to the signal level). Internal speaker / external headphones.
Englis h
POWER SUPPLY LNB Input Output Voltage Current Protection
POWER SUPPLY NiMh battery
Low battery indication Autonomy
Automatic power-off Battery charge Equipment consumption
By external power supply. By input signal connector. 24 V nominal (25 V max.). 500 mA max. Current limiter.
7.2 V – 1.4 Ah. Graphic indication on the display:
Approximately 3 hours (30 % on / off excepting measurements of MER / BER). Power-off after approximately 10 minutes of non-use. By fast internal charger.
13.5 W.
Page 7
USERS MANUAL. PROMAX-10 Premium
Mains to charger adapter
230 V / 50-60 Hz / 13.5 V DC minimum (EUROPE and other countries).
ENVIRONMENTAL CONDITIONS
This equipment could be used on the following environmental conditions, in these conditions the specifications could also be applied:
Altitude Temperature range Maximum relative humidity
Up to 2000 metres. From 5 °C to 40 °C. 80 % (up to 31 °C), decreasing lineally up to 50% at 40ºC.
MECHANICAL FEATURES Dimensions Weight
70 W (90 on the display) x 218 H x 50 D mm. 825 g. (including battery and holster).
(1) Under request carried out at the factory. (See option OP-010-61) (2) Because of safety reasons, the maximum input power over the entire band is
limited up to 120 dBµV. The equivalent power level for a group of channels of similar levels is related with the input power level over the entire band according t o the following expression:
LT = L+10 log N (LT: total level , L: mean level of one channel, N: number of channels present).
For higher input power levels, the use of an external attenuator of 20 dB is recommended. There may be certain frequencies where the symbol "<" appears at levels higher than 25 dBµV (maximum 28 dB). This is due to the aut omatic correction of the frequency response. The value measured remains correct, although the accuracy becomes ±3 dB
(3) For the positive video modulation (L standard) it can vary from 0 to -2 dB among
white and black image.
(4) For the AM audio carrier (L standard), it can vary from 0 to -3 dB below the V/A
value.
(5) See appendix E
INCLUDED ACCESSORIES
AL-012 EUROPE and other countries 230 V / 50-60 Hz mains adapter
(basic version only).
AL-022 USA and CANADA 120 V / 50-60 Hz mains adapter (only with
the OPT-010-1). AL-032 UK 230 V / 50-60 Hz mains adapter (only with the OP-010-2). AL-042 Australia 240 V/50-60 Hz mains adapter (only with the OP-010-3) AL-052 Japan 100 V / 50-60 Hz mains adapter (only with the OP-010-4). AA-012 Car cigarette lighter adapter cable. DC-239
PROMAX-10 Premium
Carrying case. DC-286 Carrier belt. AD-057 F/female - F/female input adapter. AD-058 F/male - F/female rapid adapter. CC-030 F/male - F/male (1m) coaxial cable. DC-284 Rubber holster.
Page 6 05/2004
Page 8
USERS MANUAL. PROMAX-10 Premium
OPTIONS
OP-010-1 Substitute mains adapter by AL-022. OP-010-2 Substitute mains adapter by AL-032. OP-010-3 Substitute mains adapter by AL-042. OP-010-4 Substitute mains adapter by AL-052. OP-010-61 Change channel plan. (Carried out under request in the
OP-010-E Extension of SYMBOL RATE measurement range.
OPTIONAL ACCESSORIES
AD-055 F/female - BNC/female adapt er. AD-056 F/female - IEC/female adapter. CI-023 Portable serial printer. CC-208 Data transfer cable to PC or printer. RM-010 Remote control software for AT-20C 20 dB attenuator.
factory).
PROMAX-10 Premium
.
Englis h
Page 9
USERS MANUAL. PROMAX-10 Premium
2 SAFETY RULES
2.1 Generals
Use this equipment connected
*
at ground potential
.
only to devices or systems with their common
This equipment can be used in
*
Pollution Degree 2
Use the mains adapter in
Degree 1
When using some of the following accessories
*
ensure safety.
Observe all
*
Use this instrument under the
*
The user is not authorised to manipulate inside the instrument:
*
On the Maintenance paragraph the proper instructions are given.
Follow the
*
environments. It is for
Power adapter Car cigarette lighter adapter
Any other change on the equipment should be carried out by qualified personnel.
cleaning instructions
environments.
Over-Voltage Category II
specified ratings
Over-Voltage Category I
INDOOR USE
both of supply and measurement.
specified environmental conditions
described in the Maintenance paragraph.
.
use only the specified ones
installations and
installations and
.
Pollution
to
Englis h
Page 10
Symbols related with safety:
*
USERS MANUAL. PROMAX-10 Premium
DIRECT CURRENT ALTERNATING CURRENT DIRECT AND ALTERNATING GROUND TERMINAL
PROTECTIVE CONDUCTOR
FRAME TERMINAL
EQUIPOTENTIALITY
ON (Supply)
OFF (Supply)
DOUBLE INSULATION (Class II protection)
CAUTION (Risk of electric shock)
CAUTION REFER TO MANUAL FUSE
Page 10 05/2004
Page 11
USERS MANUAL. PROMAX-10 Premium
2.2 Descriptive Examples of Over-Voltage Categories
Cat. I Cat. II Cat. III Cat. IV
Low voltage installations isolated from the mains. Portable domestic installations. Fixed domestic installations. Industrial installations.
Englis h
Page 12
USERS MANUAL. PROMAX-10 Premium
3 INSTALLATION
3.1 Power supply
The NiMh rechargable battery. Before taking any measurement, the user should make sure that the battery is fully charged (use the mains adapter supplied with the instrument).
PROMAX-10
Premium
is a portable instrument powered by a built-in 7.2 V
3.1.1 Battery charge
The instrument has a 230 V / 50-60 Hz, mains adapter, for use in EUROPE and other countries, to power or to charge the instrument. (See accessories to request other types of adapters).
There are two different situations on battery charge:
1) With the instrument stopped, on having connected the external power adapter,
begins a cycle of fast load which duration will depend on the battery state. For an empty battery this time will be of 2.5 h. approximately. The charge indicator of frontal panel will remain illuminated during this period. On having finished the battery charge the indicator will be illuminated blinking to indicate that a maintenance charge takes place.
2) With the instrument working, when connecting the power adapter, this one feeds
the instrument and provides a maintenance charge, remaining illuminated the charge indicator. The MER / BER measurements produce a light discharge of the battery in spite of the external power adapter being connected.
ATTENTION When stopping the instrument is reinitiated a charge
process. It is for that reason recommendable to discharge the
battery using in strumen t without ex ternal power ad apter to make complet e the ch arge / discha rge pr oces s.
Englis h
ATTENTION The charge system of PROMAX-10 Premium
incorporates a security system that does not allow the process charge from cer tain temperature limit, be gining the maintenance charge instead.
CAUTION Before using the power adapter, make sure that the
adapter is suitable for the mains voltage.
Figure 1.- PROMAX-10
Premium
adapter connection.
and mains
Page 13
USERS MANUAL. PROMAX-10 Premium
3.1.2 Recommendations for using the battery
In case of anticipating a long period of inactivity for the equipment it is recommended to store it with the battery discharged.
When the equipment starts the operation after a long period of inactivity to follow next steps:
Put to charge the instrument with the external adapter.
If it starts on the maintenance mode (indicator flashing) wait for the pass to the
charging mode (indicator lighting permanently). Start up the equipment, assuring that the function AUTO POWER OFF is OFF, the
indicator will stand permanently lighted and it will stay in this condition during a period of time about 10 to 14 hours.
Use the equipment in normal way, reason why after one to three whole cycles of charge/discharge, depending on the time and temperature of inactivity, the battery will have been reconditioned.
3.2 Installation and start-up.
The
PROMAX-10
A fully charged battery can power the equipment for more than three hours. When the low battery indicator appears on the display (
recharged.
When a fully discharged battery is installed, it is possible that, due to residual charges, the automatically disconnect before the low battery indicator appears on the display.
PROMAX-10
Premium
has been designed for use as portable equipment.
), the battery must be
Premium
may start up. In this case, the instrument will
3.2.1 Contrast adjustment
When turning on the pushed, on the display will appear the mess age '
the rotary selector
display contrast to get the best viewing according to working conditions. The new contrast value is held when the unit is turned off and when battery is removed.
Page 14 05/2004
', then turning the rotary selector [9] it is possible to adjust the
PROMAX-10
Premium
, if the on/off key [16] is held
CONTRAST ADJUSTMENT - Turn
Page 14
USERS MANUAL. PROMAX-10 Premium
4 OPERATING INSTRUCTIONS
4.1 Description of the controls and elements
Front panel
Englis h
Figure 2.-
[1] F-F (or F-BNC or F-IEC) adapter.
Maximum input voltage level 60 VAC rms /50-60 Hz. [2] "F" male base connector. [3] Graphic display with back lighting. [4] DC power adapter external input. [5] Volume control. [6] Keyboard. 7 keys for function selection.
Front panel.
Page 15
[7] Loudspeaker. [8] Connection to computer or printer.
specific connection cable.
CC-208
Do not connect any cable other than that supplied by the manufacturer, otherwise the instrument may suffer serious damage.
[9] Rotary selector / Pu sh button.
USERS MANUAL. PROMAX-10 Premium
keyboard.
and the
TRANSIENT DETECTOR
[10]
Selects the
[11]
Selects the
[12]
Selects the
[13]
Selects the
Figure 3.- PROMAX-10 Premium
CHANNEL - FREQUENCY
operating mode.
SCAN
operating mode.
TILT
operating mode.
SPECTRUM ANALYSER
operating modes.
Page 16 05/2004
Page 16
USERS MANUAL. PROMAX-10 Premium
[14]
Access to global configuration menu of the unit.
[15]
Selects the be taken, visualised, printed or transferred to a PC automatically.
[16]
On/Off key. [17] LNB exterrnal supply indicator. [18] Battery charge indicator.
CONFIGURATION
DATALOGGER
menus specific to each operating mode and to the
operating mode. Enables multipl e measurements to
4.2 Operating instructions
The
PROMAX-10 Premium
The
CHANNEL-FREQUENCY
, the Carrier/Noise ratio (
level
audio carrier demodulation for channel and the Modulation Error ratio (
Diagram
intermodulation distortion. The
DATALOGGER
performed and memorised for subsequent checking, transfer to PC or printing. It can perform and store up to 55 obtained measurements or loggers in the memory. Each logger carries out level, C/N, V/A, channel power or MER measurements on the channels activated in the channel plan (up to a maximum of 140 channels).
The
SCAN
the chosen frequency band in a bar-graph display. The span and the reference level may be selected through the rot ary selector. In addition, a moving marker shows the numeric level of any specific channel . This mode also permits to define the pilot channels, used for the TILT measurement (only in the forward band).
This key permits to accede to 2 operating modes: The
SPECTRUM ANALYSER
entire band in two parts: return path or sub-band (5 to 100 MHz) and forward band (5 to 862 MHz). The span is user definable between 1 and 100 MHz. In addition, it is possible to change the reference level, and
minimum
and the Carrier/Noise ratio (
power
for
digital
operating mode shows the signal level of all channels present on
levels may be detected and held for INGRESS measurements.
has 6 independent operating modes:
operating mode measures the video carrier
), the Video/Audio ratio (
C/N
analogue
channels. It also permits to measure the
operating mode enables multi ple measurements to be
mode provides a spectrum analysis over the
channels; as well as measuring the
), the Bit Error rate (
C/N
) and represent the
MER
) and activates
V/A
Constellation
CSO
maximum
and
BER
CTB
and
)
Englis h
Page 17
USERS MANUAL. PROMAX-10 Premium
In the
TRANSIENT DETECTOR
mode, the
PROMAX-10 Premium
operates as a transitory counter in the return path. The level detection threshold and the frequency margin are user definable.
The
operating mode shows on the display, both graphically and
TILT
numerically, the level difference between any four channels, previously defined as pilot channels, in order to obtain information about band equalisation. This function can be applied to the forward band and to the return path, independently.
To access any operating mode, simply press the corresponding key. The parameters relative to a particular operating mode can be modified through
the
configuration menu associated to the mode
. In order to accede to the
configuration menu associated to a particular operating mode, simply press the
CONFIG
the second page it is necessary to press the
key [14]. Some modes have more than one configuration page, to accede to
CONFIG
key again. The general parameters of configuration (selecting/editing the channel plan, measurement units, language, etc.) can be changed through the acceded by pressing again the
CONFIG
key [14]. To leave a configuration menu, just
Global Configuration Menu
, to which it is
press the key of the operating mode you wish to accede.
Figure 4.- PROMAX-10 Premium
Page 18 05/2004
configuration menus.
Page 18
USERS MANUAL. PROMAX-10 Premium
4.2.1 Global configuration menu
In order to accede to the global configuration menu, from any operating mode (SCAN, CH-FR, SPECT, etc.), you must press the
CONFIG
time that presses this key accedes to the configuration menu relative to the mode in use, which can have more than one page). The global configuration menu consists of two pages (attached figure shows the first one), to switch from one to the other simply press the
CONFIG
In order to modify the state of a given parameter, you must turn the rotary selector [9] until this one appears shaded and next press it, then the value of the parameter will appear shaded and turning the selector a new value will be able to be defined. Finally, to validate the new state, press the rotary selector [9] again.
The first page of the global configuration menu permits to modify the following parameters:
a) CHANNEL PLAN
It allows you to choose the active channel plan between the 10 channel plans that unit can store (CCIR, EIA, FCC, etc.).
b) EDIT CHANNEL PLAN.
When selecting this field and pressing the rotary selector, the unit accedes to the active
CHANNEL PLAN
[14] key repeatedly (the first
[14] key.
.
Figure 5.-
Global configuration. 1/2.
Englis h
The attached figure shows an example of channel plan. The first line shows the name of the channel plan ( offset (
5.50 MHz
units of measurement (
in the figure), the audio carrier
CCIR
), the modulation type (FM) and the
). Next line shows the
dBµV
head of the columns that constitute the channel plan: the first column ( channel and the second one (
) shows the name of each
CHAN
) the associated
FREQ
Figura 6.-
Editing a channel plan.
frequency, in MHz. The third column (BW) defines channel bandwidth, in MHz. The fourth column (ON) activates or deactivates the channel and the fifth ones (
DIG
) defines
if the channel is analogue or digital.
In the case of defining the channel as digital, the configuration page of parameters relative to QAM measurement will be accessed directly (see section
4.2.3.1.6).
The maximum number of channels in a channel plan is of 140.
The activation (ON) / deactivation (OFF) of the channels affects to the followi ng operation modes:
CH-FR, SCAN
and
DATALOGGER
. When a channel has been deactivated, this one will not be able to be tuned nor to be measured. This property allows to make agile the operation of the
PROMAX-10
Premium
, because it allows to
activate only those channels in which we are interested.
Page 19
USERS MANUAL. PROMAX-10 Premium
The Edit Channel Plan function allows to automatically activate all the channels defined in the channel plan by means of the right of select
we select ON, all the channels of the plan will be activated, otherwise, if we
ALL
all channels will be deactivated. In order to activate / deactivate a particular
OFF
field on the third line. When to the
ALL
channel, turn the rotary selector [9] until this one appears s haded and then press it, t he cursor will jump to the ON column, then turning the selector it will be possible to activate it (it appears a cross) or to deactivate it (it does not appear a cross).
By means of the Edit Channel Plan function also it is possible to define channels as
analogue
or
l. To do this, turn the rotary selector [9] until the channel
digita
that you wish to modify appears shaded and then press it twice, the cursor will jump to the
column, then turning the selector it will be possible to define it as digital
DIG
(appears a cross, channel C04 in the example of the previous figure) or as analogue (it does not appear a cross). In the case of selecting it as digital it is possible to accede directly to the digital channel configuration. In order to return to the Edit Channel Plan function press the
CONFIG
[14] key.
To modify the rest of characteristics of the channel plans it is necessary to have the
software (optional accessory).
RM-010
c) UNITS
The
PROMAX-10 Premium
and
dBm
.
permits to select the level units between
dBmV, dBµV
d) TIME
To enter the time select the TIME field and press the rotary selector. Firstly, turn the rotary selector to change the minutes field. Next press the select or again to modify the hours and finally press it once again to confirm the new time.
e) DATE
To enter the date select the DATE field and press the rot ary selector. First c hange the year field, then the month and finally the day.
The parameters which may be modified on the second page of the global configuration menu of the unit are as follows:
Figure 7.-
Global configuration. 2/2.
f) AUTO POWER OFF
This field permits to activate (ON) or deactivate (OFF) the POWER-OFF function. When this function is ON, the unit automatically turns off when it has remained inactive for a period of 10 minutes.
g) BEEP
This function permits to activate (ON) or deactivate (OFF) the
Premium
beeper. When it is on, it sounds on pressing any key or when turning the
PROMAX-10
rotary selector in order to alert the user.
Page 20 05/2004
Page 20
USERS MANUAL. PROMAX-10 Premium
h) LANGUAGE
This field permits to select the language between: ESPANOL, FRANÇAIS, ENGLISH and DEUTSCH.
i) EXTERNAL UNITS POWER SUPPLY (VLNB)
By means of the necessary to feed external units (MMDS antennas Multichannel Multipoint Distribution Service in case of terrestrial television without cable the signal input connector.
This voltage will have to be provided externally by the power input [4] and to be included in the margin from 21 to 25 V, in opposite case the instrument will not al low to activate this function, displaying the message " configuration.
PROMAX-10 Premium
it is possible to provide the voltage
) by
Error VEXT
wireless cable
" at the moment of the
The activation or deactivation of process next:
- Press repeatedly the key the configuration menu.
- Turn the rotary selector until appears shaded line
- Press the rotary selector to activate the selection. Turning the rotary selector, this one goes successively of
- Finally, to press the rotary selector to activate the configuration changes.
When surpassing the maximum LNB current consumption the output feeding will be deactivated and reactivate again the LNB voltage, after solving the cause of malfuction, it must be access again to the instrument configuration according to the sequence previously described.
Before using the LNB feeding function make sure of absence of other power supplies in the measurement/power supply connector like DC / AC voltages.
The entrance of external voltages when trying to feed from the PROMAX-10
Premium, can produce damages in the implied equipment.
To leave the global configuration menu, just press the key of the operating mode you wish to accede.
The lower line of the screen shows the (2.32 in previous figure).
LED
(V
CONFIG
) from frontal will blink during a minute. In order to
LNB
takes place according to the described
VLNB
until acceding to the second screen of
.
VLNB
to ON.
OFF
of the unit
version
control software
ATTENTION
IMPORTANT
Englis h
Page 21
4.2.2 SCAN operating mode
USERS MANUAL. PROMAX-10 Premium
The
operating mode shows, in a single screen, the signal level of all the
SCAN
active channels in the channel plan by means of a bar-graph. In addit i on, the exact level of any particular channel may be measured by simply placing the marker over it (only analogue channels).
To access to this operating mode, press the
key [11]. The top line of the
SCAN
screen will display the channel name (C44), the tuned frequency (655.25 MHz) and the level (65 dBµV) of the channel the m arker is over. The marker can be moved by turning the rotary selector [9].
Figure 8.-
SCAN mode,
full span
The channel is selected when it is shaded.
To change the
span
(shown bandwidth), press the rotary select or [9] (the span field is activated) and then turn it. The span is variable from the entire band (full span) to 10 MHz in the following steps:
10, 30, 100, 300 MHz y
full span
.
Figure 9.-
SCAN mode,
span 100 MHz
By pressing the rotary selector [9] again, you can change the reference level, in other words, on turning the rotary selecter [9] the graph is displaced vertically.
To modify the tuned channel, press the rotary selector [9] again, the channel field will be activited and then turning the rotary selector you will be able to
Figure 10.-
Ref. 90 dB
SCAN mode,
V.
µ
change it.
.
.
The attached figure shows an horizontal line at 45 dBµV. This line allows to est ablish judgements of channel level acceptance easily. To activate and define this reference line see the
menu
(section 4.2.2.1).
Page 22 05/2004
Scan configuration
Figure 11.-
reference line at 45 dBµV.
SCAN mode with
Page 22
USERS MANUAL. PROMAX-10 Premium
In the case of having activated the broadband power detection function (
MEAS -> ∑PWR
mode (See section 4.2.2.1), on the display
SCAN
of
PROMAX-10
) in the configuration menu of
Premium
will appear the total
POWER
power level on the instrument RF input. See the attached figure.
This parameter indicates the total power that is being transmitted by the connec ted coaxial cable to the
PROMAX-10
Premium
in the band from 5 to 862 MHz.
Figure 12.-
detection function.
SCAN mode, power
pilots in
The
TILT
operating mode also allows channels to be programmed for use as
SCAN
operating mode, in the forward band (see sect ion 4.2.5 TILT operating
mode).
IMPORTANT
In the SCAN operating mode, all channels are considered as analogue, as a consequence digital channels power must be measured always from the CHANNEL-FREQUENCY operating mode.
4.2.2.1 SCAN mode configuration
In the SCAN operating mode, pressing the
CONFIG
for this mode.
THRESHOLD, REFERENCE LINE MEASUREMENT
the rotary selector [9] until it appears shaded and next press the selector, then the value of the parameter will appear shaded and turning the selector you will be able to modify its value. Finally, to validate the new value, press the rotary selector [9] again.
a) THRESHOLD
b) REFERENCE LINE
c) POWER MEASUREMENT
key [14] will access the configuration menu
This menu permits to change three parameters:
and
. To accede to any of them, turn
POWER
Figure 13.-
SCAN configuration menu.
It defines the minimum level to show channel on the display (OFF or between 21 and 120 dBµV). If the threshold is deactivated (OFF), on the SCAN representation will appear all the active channels of the channel plan with a level higher than 20 dBµV. Otherwise, if we define a level f or the THRESHOLD parameter, channels with a level lower than the threshold value will not be displayed.
It permits to activate or to deactivate (OFF) a reference line on the SCAN graph (between 21 and 120 dBµV). This reference line allows to establish judgements of channel level acceptance just with a glance at the SCAN screen.
It allows to define if the measurement of the power is carried out on the channel bandwidth ( 862 MHz (
) indicates by the MARKER or over the full band between 5 and
CHAN
).
PWR
Englis h
Page 23
USERS MANUAL. PROMAX-10 Premium
IMPORTANT
To leave the SCAN operating mode configuration menu, just press the key of the operating mode you wish to access.
4.2.3 CHANNEL-FREQUENCY operating mode
The
CHANNEL-FREQUENCY
operating mode provides the following
measurements:
Analogue channels
: - Video carrier level
- Carrier / Noise (C/N) ratio
- Video / Audio (V/A) ratio
- CSO-CTB intermodulation distortion measurement
Digital channels
: - Channel power by measurements integrati on
- Carrier / Noise (C/N) ratio
- Bit error rate (BER)
- Modulation error ratio (MER)
- Constellation Diagram
To access to this mode of operation, press the
tuning modes
: by
channel
or by
frequency
. When pressing the
key [10]. There exist two
CH-FR
CH-FR
key [10], if the
instrument is in this mode allows the switching from channel tuning to frequency tuning.
4.2.3.1 Channel tuning
4.2.3.1.1 Video carrier + V/A + C/N measurement
If the tuned channel has been defined as analogue by means of the Edit Channel Plan function (see section 4.2.1 Global Configuration Menu), the
PROMAX-10
Premium
one shown in the attached figure.
The tuned channel (C69 in the figure) appears in the higher left hand corner; t his may be changed by turning the rotary selector [ 9]. At its right appears the frequency (855.25 MHz) and the active channel plan (CCIR in the figure).
will display a screen as the
Figure 14.-
measurement
V/A and C/N Level
Below is shown the
video carrier level, VIDEO Carrier
(80 dBµV in the example). The units of measurement may be changed using the Global Configuration Menu (section 4.2.1). At the bottom, a bar-graph displays the level with a resolution of 1 dB.
Also the
ratios between the video and the audio carriers (V/A)
video carrier and noise (C/N)
are shown. The example in previous figure shows a
and
the
channel with a Video/Audio ratio of 16 dB and a C/N ratio of 54 dB.
Page 24
USERS MANUAL. PROMAX-10 Premium
4.2.3.1.2 Audio carrier measurement and demodulation
For the
demodul a te an audi o s i gn al an d sho w it s characteristics (level and frequency
, press the rotary selector [9] again.
offset)
In the attached figure, the audio carrier level (
) is 64 dBµV and the frequency offset is 5.50
Carrier
MHz ( F). It also shows if the audio (whether Level, FM or AM) is activated (a speaker appears) or not (no icon appears). To change the audio carrier offs et ( F) and the audio modulation (Level, FM or AM) accede to the Channel-Frequency configuration menu (section 4.2.3.3).
PROMAX-10
Premium
to
AUDIO
Figure 15.-
Audio demodulation
and measurement.
4.2.3.1.3 CSO-CTB distortion products measurement
The active devices present in the distribution systems, when working in its nonlinear zone, generate interferent signals, called distortion products, that can fall within the video bandwidth. From a certain level this interference becomes visible on t he TV image.
The distortion products of greater level and which usually fall within the channel bandwidth are those of second and third order beats.
The level of the video carrier to the peak of the distortion products of second order beat, produced by the rest of channels. The ratio is expressed in dB and can be resembled to a C/N measurement as it is desired to be maximum.
Similarly, the of the video carrier to the peak of distortion products of t hird order beat, which s how up at the same frequency as the video carrier.
distortion (Composite Second Order) is defined, as a ratio of the peak
CSO
, Composite Triple Beat, is defined as a ratio of the peak level
CTB
Englis h
In order to obtain these measurements, the carriers of the channel plan that is desired to transmit, must be present in the network and the measurements must be carried out for each one of the channels of the system.
Second order beats show up within the channel, around the video carrier, but as the relative position of the video carrier in all the c hannels can be di fferent, it is difficult to determine where they are going to appear, as a consequence we would hav e to be made a sweep within all the channel. The measurement automatically at four frequencies around the video carrier (-1.5, -0. 5, +0. 5 and +1.5 MHz). These frequencies can be modified by the user, see Channel- Frequency mode configuration menu (sect i on 4.2. 3. 3).
PROMAX-10
Premium
performs this
Page 25
USERS MANUAL. PROMAX-10 Premium
Figure 16.-
The
PROMAX-10
(that is to say, the
Interpretation of the CSO and CTB measures.
Premium
ratio that has minor value) accompanied of the frequency offset
CSO
displays as
value the worst measurement
CSO
for which it has been obtained (for example, in the figure previous at +0.5 MHz).
The
measurement is performed equivalently to the
CTB
. If all the channels
CSO
that are transmitted have the video carrier at the same position within the channel, t hen the reason, that the
beat will show up at the same frequency as the video carrier. It is for that
CTB
PROMAX-10
Premium
makes this measurement at the same frequency as the video carrier and, as consequence, to make this measurement it is necessary to possible to turn off the carrier at the head-end, the
the video carrier of the channel we want to analyse. If it is not
turn of f
PROMAX-10
Premium
allows, like
approach, to make this measurement at one of the free adjacent channels (see section
4.2.3.3. Channel-Frequency mode configuration).
MEASURING METHOD
When pressing the rotary selector
[9] again, we will accede to the
CSO-CTB
distortion products measuring screen. First of all it will appear the message
CARRIER. PRESS TO MEASURE
REMOVE
. That is to say, once the video carrier level on which it is desired to make the measurement appears on the screen, you must press the rotary selector [9] in order that the unit retains the carrier level and comes to
Figure 17.-
CSO and CTB
measurement.
calculate the CSO and CTB ratios; next you must turn off the video carrier of the channel you want to make the measurement (at the right of the CTB measurement will appear the message
Carrier OFF
).
The display will show the level of the video carrier (80 dBµV in the previous figure) and the ratio has been obtained) and the
according to the equipment has detected the absence or not of carrier).
ON
ratio (accompanied by the frequency offset for which the minimum
CSO
ratio (with the message
CTB
Carrier OFF
or
Carrier
The lower part of the screen shows the 4 measures made for the estimation of the CSO value.
Page 26
USERS MANUAL. PROMAX-10 Premium
If the rotary selector [9] is pressed again, the unit will return t o the measurement screen of the video carrier + V/A +C/N (paragraph 4.2.3.1.1).
4.2.3.1.4 Power and C/N ratio measurement of digital channels (DVB-
C / DVB-T / DAB).
When the tuned channel has been defined as digital by means of the Edit Channel Plan f unction (see section 4.2.1 Global Configuration Menu) a screen as shown in the attached figure will appear. This screen shows the digital example) together with its related channel bandwidth (
ratio (C/N
in the case of analogue channels, a bar-graph represents the channel power, with a resolution of 1 dB.
In order to measure digital channels correctly, previously, the channel must be defined as digital (see Edit Channel Plan function at section '4.2.1 Unit Global Configuration) and, if necessary, the CHANNEL BW parameter should be redefined by means of the Channel-Frequency configuration menu. When a channel has been defined as digital, the PROMAX-10 Premium tunes it at its central frequency.
Digital channels power measurement is carried out by means of an integration method. The
) in sections of 230 kHz (4 by each MHz approximately) and measures the
BW
contribution of each one to the total power of the channel. This way, a very exact measurement is obtained, specially in the case of degraded channels, because the channel flatness is had in consideration.
CHANNEL POWER
= 8.0 MHz) and the
BW
=17 dB). At the bottom of the screen, as
PROMAX-10
Premium
(40 dBµV in the
Carrier/Noise
divides the bandwidth of the c hannel (
Figure 18.-
Measuring a digital
channel.
VERY IMPORTANT
CHANNEL
Englis h
Figure 19.-
The and the noise power. The user can define the frequency where the noise power must be measured. Two possibilities exist:
C/N ratio
Measuring the power of a digital channel by integration.
measurement shows the ratio between the digital channel power
Page 27
I. Absolute method
USERS MANUAL. PROMAX-10 Premium
Selecting in the Channel-Frequency configuration menu the parameter as
, the equipment will interpret the
FREQ
NOISE FREQ.
NOISE MODE
value as the frequency where the noise power measurement must be done. Naturally the user must be sure that the
NOISE FREQ.
value corresponds with a free channel.
II. Relative method
Selecting in the Channel-Frequency configuration menu the
NOISE MODE
parameter as F, the unit will make the measurement of noise at the frequency obtained to add to the tuning frequency (channel central frequency) the value defined for the
NOISE FREQ.
parameter. The
= BW/2 + 0.5 MHz
, where BW is the channel bandwidth defined in the channel plan,
PROMAX-10 Premium
takes by default
NOISE FREQ.
thus for example, if it is desired to measure the C/N ratio of a digital channel with a bandwidth (BW) of 8 MHz,
NOISE FREQ.
will take as value 4.5 MHz. The Channel­Frequency configuration menu (paragraph 4.2.3.3) allows the user to redefine this parameter; when tuning a new channel the equipment will return to assign to
the default value.
FREQ.
NOISE
The following figure shows that when making C/N measurements in t he relative method (∆F), it is essential to bear in mind the presence of adjacent channels; otherwise, the noise could be confused with the signal of another channel.
Figure 20.-
Example of digital channels spectrum.
Taking as example the previous figure, if we suppose that for the channels that appear in the figure it has defined a bandwidth of 8 MHz, in the relative mode of measurement the
. Note that this method can cause erroneous measures in the measurement of the
MHz
channels
D24
For
PROMAX-10 Premium
and
D24
.
D25
it is advised to configure
will take as value for
NOISE FREQ = -4.5 MHz
NOISE FREQ. = 4.5
and to verify if the measurement increases. In case a channel has two adjacent channels (for example channel
NOISE FREQ.
between
Page 28 05/2004
) it is recommended to select the absolute mode and to assign to the
D25
parameter a frequency corresponding to a free channel (for example
D26
and
D27
).
Page 28
USERS MANUAL. PROMAX-10 Premium
4.2.3.1.5 Constellation Diagram representation, Bit Error Rate (BER)
and Modulation Error Ratio (MER) measurement in digital channels.
Once obtained the measurement on tuned channel as well as the representation for the
. (See appendix H 'Principle of QAM modulation. The Constellation Diagram').
The screen like the one of the attached figure. On it besides the also appears the type of for the digital signal appendix G 'Measurement of the modulation digital ratio (MER)'), the represented quadrants (
carries out a new measurement. If the selector rotary is pressed t wo times, it is acceded again to the screen of
quadrants), goes to only a quadrant e.g.: Q1, pressing again CH-FR [10] obtains an extended representation (zooming x4) e.g.: Q1 Z. Pres s CH-FR [10] to return afresh to the
Q1-4 mode.
PROMAX-10
) and the type of detected codification (
Q1-4
When pressing again the rotary selector
Figure 21.- Constellation Diagram representation and
BER and MER measurement in a digital channel.
When pressing the CH-FR [10] from Q1-4 (graphical representation in four
Power and C/N ratio measurement
DVB-QAM
(BER)
Power and C/N ratio measurement
digital signal can be obtained pressing the rotary selector
Premium
QAM
, after some seconds for calculation, will show a
modulation, the symbol rate (SR), the error rate obtained
, the modulation error ratio measurement (
Constellation Diagram
MPEG2
, the
BER
Constellation diagram
).
, the
PROMAX-10
(section 4.2.3.1.4).
and
graph
) (see
MER
Premium
MER
,
Englis h
In order to change the quadrant, from Q1-4 press and keep pressed the CH-FR [10] key until a quadrant frame selec tor will appear and slide it by means of the rotary selector [9], locate it in the wished quadra nt and finally loose the CH-FR [ 10] to set the new quadrant for representing.
4.2.3.1.6 Configuration of QAM signal measurement using edit
channel plan.
From the edit channel plan its possible to access directly to the followi ng page of the configuration menu of the parameters relative to this mode, for the tuned channel.
Page 29
USERS MANUAL. PROMAX-10 Premium
The parameters which may be changed and their margins are shown in the following table. To change a given parameter turn the rotary selector [9] until the field i s activated (it will appear shadowed) and then press. The value of the parameter will be activated and may be changed by turning the rotary selector. When the value you want appears, press the rotary selector [9] to validate it.
Figure 22.-
Configuration of digital
signal measurement.
DESCRIPTION PARAMETERS VALUES
System (according to country) QAM Modulation Symbol Rate
Attenuation Channel frequency Offset
(*) Remark:
Including the OP-0 10- E opt ion , ot her wi se the marg in is fr om 10 00 to 700 0.
ANNEX ITU-T
MODULATION
SYMBOL RATE
ATTN.
FCH
DVB, B, C
16, 32, 64, 128, 256
1.000 to 7.200
(*)
Auto or from 0 to 60 dB.
(Steps of 10 in 10)
From - 2.00 to 2.00 MHz
IMPORTANT
To leave the configuration menu, just press the key of the operating mode you wish to access.
4.2.3.2 Frequency tuning and configuration
When in operating mode, pressing again the [10] key switches
tuning by frequency by frequency mode
receiver with a resolution of 10 kHz in the 5 to 862 MHz band. In this mode you may tune to any
signal within the band, including leakage pi lots, telephone and communications signals. To change tuning simply press the rotary selector [9] until the digit you want to change is underlined and then turn the rotary selector [9]. The channel number will appear to the left of the tuning frequency, as long as it corresponds to an active channel.
a) Measurement functions
When the measure the signal level being represented on a graphical bar. Simultaneously, this i s indicated by the loudspeaker which emits a tone that varies with the measured level.
The access to the measurem ent functions from the operation mode when the instrument operates in for the
analogue
CHANNEL-FREQUENCY
CH-FR
tuning by channel
and vice versa. In the
the instrument becomes a
PROMAX-10
or
digital channels
Premium
, is described in the following figure.
to
tuning
Figure 23.-
Level signal measurement.
works in frequency tuning, it allows to
CHANNEL-FREQUENCY
channel tuning
or
frequency tuning
Page 30
USERS MANUAL. PROMAX-10 Premium
Figure 24. -
Access to the measurement functions, CH-FR operation mode.
The measurements when operating in frequency tuning are identical to the described ones for the channel tuning operation (See section “4.2.3.1 Channel tuni ng”).
b) Configuration
When the
PROMAX-10
Premium
works in the FREQUENCY-TUNING operating mode, pressing the
CONFIG
key [14] will access the first page of the configuration menu of the parameters relative to this mode.
Figure 25.-
CH-FR configuration 1/2.
Page 31
USERS MANUAL. PROMAX-10 Premium
The parameters, which may be changed, and their margins are shown in t he following tables. To change a given parameter turn the rotary select or [9] until the field is activated (it will appear shadowed) and then press. The value of the parameter will be activated and may be changed by turning the rotary selector. When the value you want appears, press the rotary selector [9] to validate it.
DESCRIPTION PARAMETER VALUES
Carrier type
MODE FREQ
ANL: Analogue
DIG: Digital Frequency steps (
Only digital channels
FREQ STEP
)
CHANNEL BW
From 0.01 to 10.00 MHz
From 0.3 to 9.9 MHz Digital channel bandwidth
(
Only digital channels
) Frequency where noise is measured in the C/N measurement.
(
Only digital channels
) Noise measuring mode.
NOISE FREQ.
NOISE MODE FREQ
± 99.9 MHz (relative mode).
5.00 MHz to 862.00 MHz absolute mode.
(Absolute): Noise level is measured at the frequency defined by
NOISE
FREQ.
(Relative): The value
BW/2
defined by
NOISE FREQ.
is
added to the tuning frequen cy.
(
Only ANL mode
CTB measurement mode (
Only ANL mode
FCS01
(
Only ANL mode
FCS02
(
Only ANL mode
FCS03
(
Only ANL mode
FCS04
System (according to country) QAM Modulation Symbol Rate
Attenuation
)
)
)
)
)
CTB SHOWED IN CH
FCS01
FCS02
FCS03
FCS04
ANNEX ITU-T
MODULATION
SYMBOL RATE
ATTN.
or free channel
From ± 0.5 to ± 2.50 MHz
From ± 0.5 to ± 2.50 MHz
From ± 0.5 to ± 2.50 MHz
From ± 0.5 to ± 2.50 MHz DVB, B, C
16, 32, 64, 128, 256
1.000 to 7.000 Auto or from 0 to 60 dB.
(Steps of 10 in 10)
The parameters
SYMBOL RATE
Page 32 05/2004
are changed digit by digit, starting at the lowest weight.
FREQ. STEP, CHANNEL BW, NOISE FREQ., ∆FCSO
y
Page 32
USERS MANUAL. PROMAX-10 Premium
REMARK
: When switching from frequency to channel, if the tuned frequency does not
correspond to any channel, the
PROMAX-10 Premium
will search for the channel nearest to this frequency and will stay tuned to this channel. This operation may take a few moments.
4.2.3.3 CHANNEL-FREQUENCY mode configuration
In
CHANNEL-FREQUENCY
when the instrument works in channel tuning, pressing the
CONFIG
key [14] will access the first page of the confi guration menu of the parameters relative to this mode.
The parameters which may be changed and their margins are shown in the following table. To change a given parameter turn the rotary selector [9] until the field is activated (it will appear shadowed) and then press. The value of the parameter will be activated and may be changed by turning the rotary selector. When the value you want appears , press the rotary selector [9] to validate it.
DESCRIPTION PARAMETER VALUES
(
Only analogue channels
)
Audio modulation
(
Only analogue channels
)
Audio carrier offset (
Only digital channels
)
Digital channel bandwidth (
Only digital channels
) Frequency where noise is measured in the C/N measurement.
(
Only digital channels
) Noise
measuring mode.
Channel frequency Offset
mode,
Figure 26.-
AUDIO MODE
AUDIO FREQ.
CHANNEL BW
NOISE FREQ.
NOISE MODE FREQ
FCH
CH-FR configuration 1/2.
FM: Audio FM AM: Audio AM LV: The loudspeak er emits
a tone whose frequency varies according to the received signal level.
OFF: Audio deactivation. From 4.00 to 9.00 MHz
From 0.3 to 9.9 MHz
± 99.9 MHz (relative mode).
5.00 MHz to 862.00 MHz absolute mode.
(Absolute): Noise level is measured at the frequency defined by
NOISE
FREQ.
(Relative): The value
∆F
defined by
NOISE FREQ.
is added to the tuning frequency.
From - 2.00 to 2.00 MHz
Page 33
USERS MANUAL. PROMAX-10 Premium
The parameters
AUDIO FREQ., CHANNEL BW
and
NOISE FREQ.
are
changed digit by digit, starting at the lowest weight.
To accede to the second page of the configuration menu, press
CONFIG
[14]
key again.
The
CTB SHOWED
parameter allows to define the method of measurement for the CTB.
IN CH
(within the channel) is the suitable method whenever the carrier of the channel in study can be turned off. If it is not possible, as approach, any other channel (free) can be defined for the measurement of the CTB.
Parameters
and
∆FCS04
PROMAX-10
∆FCS01, ∆FCS02, ∆FCS03
Figure 27.-
allow to modify the frequencies where the CSO is measured (the
Premium
admits values from -0.5 to -2.50 and from 0.5 to 2.5 MHz).
CH-FR configuration 2/2.
These parameters are modified digit to digit, beginning by the one of smaller weight.
tuning (
Finally,
PROMAX-10
parameter allows to set the frequency offset for the channel
FCH
Premium
admits values from –2.00 to 2.00 MHz).
IMPORTANT
To leave the configuration menu of the CHANNEL-FREQUENCY mode, just press the key of the operating mode you wish to access.
4.2.4 SPECTRUM ANALYSER operating mode.
Pressing the key the unit switches to the SPECTRUM ANALYSER mode. This function has 4 different operating modes (SPECT, MAX, MIN and TRANS), which are selectable through the spectrum mode configuration menu (see section 4.2.4.5).
4.2.4.1 SPECTRUM operating mode.
In the
SPECTRUM
analysis of the band; the span and the reference level are variable. The spectral analysis can be done in the spectrum configuration menu (section 4.2.4.5); this menu also allows to select the
detector
used for the representation of the spectrum between
The marker frequency is displayed in the higher left hand corner of the screen (650.00 MHz) and, to its right, the frequency signal level (34 dBµV). To alter the marker frequency turn the rotary selector [9].
mode, the
forward band
PROMAX-10
or in the
return path
Premium
peak
Figure 28.-
provides a spectral
as it is selected in the
and
average
SPECT mode.
.
Page 34
USERS MANUAL. PROMAX-10 Premium
By pressing the rotary selector [9] you can alter the
between 1 MHz and
span
100 MHz.
Frequency tuning resolution varies according to the span selected, as shown in the following table.
Figure 29.-
SPECT mode, span reduction.
DETECTOR SPAN
PEAK
100 MHz (full span in the
return path)
50 MHz 450 kHz 30 MHz 275 kHz 15 MHz 135 kHz
5 MHz 45 kHz 1 MHz 10 kHz
AVERAGE
30 MHz 280 kHz 15 MHz 140 kHz
5 MHz 50 kHz 1 MHz 10 kHz
By pressing the rotary selector [9] again the reference level may be modified.
Pressing the rotary selector [9] a second time will activate the tuning frequency field of the marker, being possible to tune new frequencies.
The attached figure shows an horizontal line at 45 dBµV. This line allows to identify levels over a reference of our interest easily. The activation and definition of this
reference line
is carried out on the
spectrum analyser configuration menu.
FREQUENCY
RESOLUTION
Figure 30.-
modification.
Figure 31.-
900 kHz
Reference level
Reference line.
4.2.4.2 MAX operating mode.
The SPECTRUM operating mode also permits to be configured as (MAXIMUM INGRESS). This option is selected on the Spectrum Configuration menu (section 4.2.4.5). In the
mode, the
MAX
PROMAX-10
screen the maximum measured level through a dotted line.
maximum hold
Premium
holds on the
Figure 32.-
MAX mode.
Page 35
USERS MANUAL. PROMAX-10 Premium
This measurement is used to be carried out on the
path and permits to
return
detect intermittent interferences.
Since on this operation mode the
PROMAX-10 Premium
holds on the screen the maximum measured value (through a dotted line), after making several measurements over the band, it will be possible to detect impulsive type interferences. It is advised to previously define a reference line which actuates as a maximum noise threshold (30 dBµV on previous figure).
4.2.4.3 MIN operating mode.
This measurement permits to detect permanent channel interferences that in an other way could remain masked because of the variable nature of the signal. It is interesting in analogue channel measurements as well as in digital channel.
To select this representation mode you must select the INGRESS (see section 4.2.4.5. Spectrum mode configuration menu).
In this operating mode it is advised to select the
Average detector
4.2.4.5. Spectrum mode configuration menu).
Figure 33.-
MIN mode. First scan.
Figure 34.-
Previous figures show an example of this measurement over a digital channel. Left figure corresponds to the first sweep. After several sweeps (right figure) an interference is seen at the centre of the channel, which previously remain masked by the noisy nature of the signal.
mode, MINIMUM
MIN
(see section
After several scans.
4.2.4.4 TRANSIENT DETECTOR operating mode
IMPORTANT
The Transient detector mode is only operative in the return path.
The TRANSIENT DETECTOR mode allows to count the number of transitory with a level higher than a certain threshold. The threshold is defined by the user (between 20 and 60 dBµV) as well as the detection frequency range.
The frequencies on which the detector acts (the higher frequency can be reduced by means of the
parameter on the configuration menu).
FREQ.
The lower line ( the detector is active.
Page 36 05/2004
field shows the margin of
SCAN
) shows the time since
TIME
STOP.
Figure 35.-
Transient detector mode.
Page 36
USERS MANUAL. PROMAX-10 Premium
The level defined by the mode (section 4.2.5.5).
dotted line shows the detection threshold (REF. LINE) whereas the continuous line indicates the maximum detected level.
COUNTER
To initialize the TIME and COUNTER fields, press the
In the lower part of the screen it appears a bar that shows present level, a
field shows the number of impulses that have exceeded the
REF. LINE
parameter in the configuration menu of the SPECT
SPECT
[13] key.
4.2.4.5 SPECTRUM ANALYSER mode configuration
Pressing the access the parameter configuration menu for the
SPECTRUM
To change the value assigned to one parameter, turn the rotary selector [9] until its field is activated (this will appear shadowed) and t hen press. The value of the parameter will be activated and may be changed by turning the rotary selector. When the value you want appears, press the rotary selector [9] to validate it.
The parameters that can be modified through this menu are:
a) BAND
It selects the analyzed band between:
b) MODE
It selects the representation mode:
CONFIG
function.
RETURN PATH FORWARD
SPECT MIN MAX TRANS
key [14] will
Figure 36.-
Sub-band spectrum (5 to 100 MHz). Band from 45 to 862 MHz.
Instantaneous value. Minimum values held (MIN INGRESS). Maximum values held (MAX INGRESS). Transient detector. Only operative in the return path.
SPECT mode configuration.
Englis h
c) DETECTOR (Only for the SPECT, MAX and MIN modes)
It selects the used detector between:
PEAK AVERAGE
In the
d) REF. LINE
It permits to activate / deactivate and to define the reference line level in 1 dB
steps from 20 to 120 dBµV (dBµV units). This line appears on the spectrum
representation and also it is the threshold value for the impulses detection in the
TRANSIENT mode (only if its value is under 60 dBµV).
e) STOP. FREQ
It defines the maximum frequency for the transient detection.
mode the detector used is always
TRANS
PEAK
.
Page 37
USERS MANUAL. PROMAX-10 Premium
F
IMPORTANT
To leave the configuration menu of the SPECT mode, just press the key of the operating mode you wish to access.
4.2.5 TILT operating mode.
The
operating mode displays on-screen, graphically and numerically, the
TILT
difference in level between any four frequencies previously defined as pilots. This function provides a quantitative measurement about band equalisation.
This function can be applied to forward and to the return paths, according to is defined in the TILT mode configuration menu (section 4.2.5.1 Tilt mode configuration).
To access this mode of operation press the
key [12]. The screen will
TILT
show a bar-graph of the four pilot chann els and the difference in level (TILT) between the upper pilot and the lower pilot (- 5 dB in the example). When the pilots are not present or it s level is lower than 20 dBµV, the message
NO PILOTS
rotary selector [9] the reference level may be modified.
will appear. Turning the
igure 37.-
TILT mode.
The pilots may be defined in two ways:
a) By frequency
Using the TILT configuration menu. See section 4.2.5.1 Tilt mode configuration.
b) By channel
From the
(only pilots in the forward band). mode. In order to do this:
SCAN
1. Place the marker over the channel you want as low pilot.
2. Press the
key [11], and the message
SCAN
PILOT 1: PRESS TILT
will appear at the bottom of the screen.
3. Keeping the message
PILOT 1 ENTERED
key [11] held down, press the
SCAN
will appear as confirmation.
key [12], and a
TILT
Repeat the steps 1 to 3 for the 3 following pilots.
4.2.5.1 TILT mode configuration.
Pressing the
CONFIG
access to the first page of the configuration menu.
To change a given parameter turn the rotary selector [9] until the field is activated (this will appear shadowed) and then press. The value of the parameter will be activated and may be changed by turning the rotary selector. When the value you want appears , press the rotary selector [9] to validate it.
Page 38 05/2004
[14] key will
Figure 38.-
TILT configuration. 1/2
Page 38
USERS MANUAL. PROMAX-10 Premium
a) BAND
It permits to select between the
RETURN PATH
mode (5 to 100 MHz).
FORWARD
mode (45 to 862 MHz) and the
b) F. PILOT 1
It defines a frequency belongi ng to the forwar d band (from 45 to 862 MHz) where the first measurement of level will be done. This paramet er, and the res t of pilots, is defined digit by digit, by pressing repeatedly and turning the rotary selector.
c) F. PILOT 2
It defines the second pilot frequency belonging to the forward band.
d) F. PILOT 3
It defines the third pilot frequency belonging to the forward band.
e) F. PILOT 4
It defines the fourth pilot frequency belonging to the forward band. To accede to the second page of the
TILT configuration menu, press the
CONFIG
key again. This screen permits to define the frequencies of the pilots belonging to the return path (from 5 to 100 MHz). As in previous screen, the frequencies are defined digit by digit.
Figure 39.-
TILT configuration 2/2
IMPORTANT
To leave the configuration menu of the TILT mode, just press the key of the operating mode you wish to access.
4.2.6 DATALOGGER operating mode.
The
LOGGER
Carrier/Noise ratio and the Video/Audio ratio (the latter only in the case of analogue channels) as well as the power channel and MER measurements of each active channel in the channel plan (see paragraph Channel Plan editor at section 4.2.1. Unit Global Configuration). These measurements are stored in the memory so that they can be subsequently viewed, printed or transferred to a PC. The store up to 55 obtained channels or loggers in the memory, with a maximum of up to 140 analysed channels in each.
When pressing the key, a screen similar to the one of the attached figure will appear. The first line shows the logger number (i.e. 53 in the attached fi gure) followed by the date it was acquired (only if MEASURE funct ion was executed on that logger previously). Below, the different functions you can perform from this operating
mode are shown:
function permits to automatically measure the level, the
PROMAX-10 Premium
and
[16]
PRINT
.
Figure 40.-
LOGGER function. Initial menu.
LOGGER
VIEW, MEASURE
will
Englis h
Page 39
USERS MANUAL. PROMAX-10 Premium
On the bottom of the display appears present time and date. To modify them see section 4.2.1 Global Configurat i on Menu.
To accede to any of the di fferen t f ields of the initial screen of the LOGGER funct ion , turn the rotary selector [9] unt il it i s act i vated (i t appe ars shadowed) and next press it .
First of all you must select the logger on which you want to perform any function: turn the rotary selector unt il the
LOGGER NUMBER
field is shadowed and press it. Then turning the rotary selector, select the desired logger (from 0 to 54) and finally press it again to validate. The acquisition date appears under the logger number (only if you have taken measurements on this logger).
To carry out a logger acquisiti on you must select the
MEASURE
function, for this purpose turn the rotary selector [9] and when this field appears s hadowed press it , in this way you will accede to the logger. Next press again the rotary selec tor [9] in order t hat t he
PROMAX-10 Premium
makes all the measurements defined in the LOGGER configuration menu over all the ac tive channels of the channel plan (s ee Edit Channel Plan function at section 4.2.1 Global configuration menu).
IMPORTANT
The processing of channels as analogue or digital and the parameters to make the measures (the frequency of the audio carrier for the analogue channels and the frequency offset for the noise measure of the digital channels C/N ratio) will correspond with the configuration of the equipment at the time of carrying out the measurement.
To return to the initial menu of the LOGGER function press the
LOGGER
To check the measurements stored in a specific logger select the
VIEW
key.
option:
The first line shows the channel plan (CCIR in the example), the audio carrier offset (5.50 MHz), the audio demodulation (FM) and the units of measurement (dBµV). The second line shows the logger number (53 in the attached figure) and the headings of the measurements (V,
and
V/A
shown in the following format: first column s hows the
). The measurements taken are
C/N
Figure 41.-
Viewing a logger.
channel, the D indication on its right means it has been defined as digital (see Edit Channel Plan function at section 4.2.1 Global configuration menu), the second column shows the level (analogue channels) or the channel power (digital c hannels), the third the V/A ratio (analogue channels) an the fourth the C/N ratio (analogue c hannels) or MER (digital channels). Turning the rotary selector [9] you can see the rest of channels.
The time and date when measurements were taken appear at the end of the list of measurements, in the following order: hour:minute:second and month:day:year. To return to the initial menu of the LOGGER FUNCTION press
To print the stored measurement select the
PRINT
LOGGER
key.
field, previously consult
section 4.3 Connecting to a computer or a printer.
Page 40 05/2004
Page 40
USERS MANUAL. PROMAX-10 Premium
LOGGER NUMBER 10 DATE: 08:55 21-03-2004 CHANNEL PLAN: CCIR
AUDIO: 5.50 MHz (FM) UNITS: dBuV THRESHOLD: OFF NOISE FREQ.: BW/2
CHAN FREQ V V/A C/N C23 487.25 76 14 52 C25 503.25 53 9 33 C27 519.25 81 16 55 C29 535.25 59 13 39 C31 551.25 78 16 44 C34 575.25 67 11 45 C37 599.25 54 7 34 C38 607.25 57 14 >37 C39 615.25 62 18 >42 C41 631.25 71 14 43 C43 647.25D 43 13 C14 70 MER = 33
PROMAX-10 Premium
BER = 8.8E-6
Figure 42.-
Example of print.
4.2.6.1 Datalogger configuration.
From the LOGGER mode, when pressing the
CONFIG
to the configuration menu. This menu permits to modify the parameters. To accede to them turn the rotary selector and once the parameter we want to change is appears shadowed, press it; then turning it its value will be modified. Finally press it again to validate the new value.
a) THRESHOLD: It activates / deactivates the measurement threshold.
This parameter allows to carry out the logger function in an agile way, by activating only those measures we consider signific ative. In the higher than 20 dBµV). On the other hand, when a level is defined, only those channels with a level higher than the measured.
b) MEASURES: It defines the types of measurements to be made.
This parameter allows to select the measurements that are desired to made between:
key [14] we will accede
THRESHOLD
mode (deactivated) all measurements are taken (with a level
OFF
and
MEASURES
Figure 4.3.-
Logger configuration menu.
THRESHOLD
Englis h
value will be
Page 41
USERS MANUAL. PROMAX-10 Premium
ANALOGUE CHANNELS DIGITAL CHANNELS
Carrier /
Noise
Ratio
Power MER BER
Yes
LEVEL / POWER
AV - C/N / MER
AV-CN / MER-
BER
(*) Note:
The
Level
Yes - - Yes - -
Yes Yes Yes Yes Yes -
Yes Yes Yes Yes Yes
measurements represent an increase of the data acquisition time.
BER
Audio /
Video Ratio
IMPORTANT
To leave the configuration menu of the DATALOGGER mode, just press the key of the operating mode you wish to access.
4.3 Connection to a Computer or Printer.
The
PROMAX-10
Premium
for data transfer, by means of the connection cable model
Do not connect any cable other than that supplied by the manufacturer, otherwise serious damage may be caused to the equipment.
1) Prior to connecting the equipment to a PC, disconnect both from their respective power supplies.
2) Connect the end of the connection cable corresponding t o t he
Premium
to connection [8] and the other end to the serial port of your
computer or printer.
permits the connection to a PC or to a serial printer
.
CC-208
PROMAX-10
*)
(
Once the computer or printer have been connected, select on the
Premium
the
LOGGER
operation mode. If
function is selected data will be
PRINT
PROMAX-10
dumped to the remote unit through the serial port.
The communication parameters used by the
PROMAX-10
Premium
, and which
therefore must be defined on the remote unit (PC or printer) are the following:
Rate Data bits Parity Stop bits
Page 42 05/2004
19200 bauds 8 bits None 1
Page 42
USERS MANUAL. PROMAX-10 Premium
The remote control software
from a computer the following options:
1) CHANNELS PLAN PROCESSOR: Modify, add or delete channel plans.
2) CONFIGURATION: Permits to modify all the configuration parameters.
3) DATALOGGER: Permits to edit and to save all the measurement contained in a logger.
4) UPDATE: Allows to update the
PROMAX-10 Premium
(optional accessory) permits to carry out
RM-010
software version.
Englis h
Page 43
USERS MANUAL. PROMAX-10 Premium
5 MAINTENANCE
This part of the manual describes the maintenance procedures and the location
of faults.
5.1 Instructions for returning by mail
Instruments returned for repair or calibration, either within or outwit the guarantee period, should be forwarded with the following information: Name of the Company, name of the contact person, address, telephone num ber, receipt (i n the c ase of coverage under guarantee) and a description of the problem or the service required.
5.2 Method of maintenance
The method of maintenance to be carried out by the user consists of cl eaning the cover and changing the battery. All other operations should be carried out by authorised agents or by personnel qualified in the servicing of instruments.
5.2.1 Cleaning the cover.
CAUTION
Do not use scented hydrocarbons or chlorized solvents. Such products may attack the plastics used in the construction of the cover.
The cover should be cleaned by means of a light solution of detergent and water applied with a soft cloth. Dry thoroughly before using the system again.
CAUTION
To clean the contacts, use a dry cloth. Do not use a wet or damp cloth.
CAUTION
Do not use for the cleaning of the front panel and particularly t he viewf inders, alcohol or its derivatives, these products can attack the mechanical properties of the materials and diminish their useful time of life.
5.3 Components which user can not replace
5.3.1 Not replaceable fuses by user
To be replaced by qualified personnel. Its position identifier and characteristics are:
F001: FUS 0.5 A T 125 V F002: FUS 2.5 A T 125 V
Englis h
Page 44
USERS MANUAL. PROMAX-10 Premium
APPENDIXES
APPENDIX A.- MEASUREMENT OF THE VIDEO CARRIER LEVEL (CL)
-ANALOGUE CHANNELS-
A) Negative Video Modulation (PAL/NTSC)
The measurement of the video carrier level is carried out taking the modulation peak as the measurement value, this being the maximum value of the signal during the line synchronism. The system requires a minimum length of time in order t o make this measurement, since it has to detect the peak of the modulated signal.
Figure 44.-
The typical values between which the video carrier level fluctuates are:
- In the transmission line: From 70 to 100 dBµV (From 10 to 40 dBmV)
- In the user’s terminal: From 60 to 80 dBµV (From 0 to 20 dBmV)
B) Positive Video Modulation (SECAM)
On this type of modulation, the line synchronism is defined by a minimum carrier level. The maximum signal level (measurement point) is variable in time, and it is a function of the picture that is being transmitted. It could vary from 10 dB among white and black image; nevertheless white signals, Video I nsertion Test (VIT), are t ransmitted in the sweep pulses, which reduce this margin to 4 dB approximately.
Due to this fact, and the small duration of the VIT, when we measure levels of SECAM signals, it is advisable to add 2 dB to the quantity showed on the displ ay, in order to obtain a more accurate measurement in its average value.
Measurement of the video carrier level.
Englis h
Page 45
USERS MANUAL. PROMAX-10 Premium
APPENDIX B.- MEASUREMENT OF THE ADJACENT CHANNEL LEVEL
-ANALOGUE CHANNELS-
The user can obtain the ratio between the video carrier amplitudes of two adjacent channels.
)dB(CC
2LV1LV
Figure 45.-
Measurement of the adjacent channel level.
Differences of more than 3 dB between carriers of adjacent channels may cause problems of interference in reception.
Page 48 05/2004
Page 46
USERS MANUAL. PROMAX-10 Premium
APPENDIX C.- MEASUREMENT OF THE VIDEO / AUDIO RATIO
-ANALOGUE CHANNELS-
)dB(AVA/V
=
LL
It permits to measure the ratio between the video carrier and the audio carrier amplitudes.
Although this process depends on the standard used, it is usual to consider that a properly transmitted PAL channel should have a sound subcarrier 13 dB below the video carrier.
Englis h
Figure 46.-
These specifications ensure that there is no interference in the same or the adjacent channel.
Video / Audio ratio measurement.
Page 47
USERS MANUAL. PROMAX-10 Premium
APPENDIX D.- CARRIER TO NOISE RATIO (C/N)
-ANALOGUE CHANNELS-
The Carrier to Noise ratio is a signal quality measurement. The noise power depends on the measurement filter bandwidth, in analogue TV it is usual to refer the noise level to a bandwidth of 5 MHz (this is the case of the
PROMAX-10
Premium
is desired to know the C/N measurement for a different bandwidth (BWx), the following correction should be applied.
Starting from the C/N definition:
NCN/C
=
BWBW
XX
where C: Video carrier level N
: Noise power referred to a BWx measurement bandwidth
BWX
and as the noise power referred to BWx could be related with the noi se power referred to 5 MHz according to:
BW
X
log(10NN
+=
MHz5BW
X
)
MHz5
the C/N ratio referred to a noise measurement bandwidth BWx should be expressed as:
). If it
BWx
XX
log(10NCNCN/C
MHz5
===
10PROMAXMHz5BWBW
BWx
log(10N/C)
)
mhZ5
That is to say, the factor 10 log (BWx / 5 MHz) should be subtracted to the
PROMAX-10
Premium
readout.
There exist some standards that determine the minimum C/N value for an installation. In general, greater values than 40 dB are considered as good quality signals. Under 40 dB snow should appear on the TV screen.
Page 50 05/2004
Page 48
USERS MANUAL. PROMAX-10 Premium
APPENDIX E.- CARRIER / NOISE RATIO MEASUREMENT (C/N)
Englis h
Figure 47.-
Example
Suppose that the video carrier level for an analogue channel, is of 10 dBmV,
from the diagram we can deduce that the measurement range is of 48 dB.
Thus, if when we carry out the measurement the C/N ratio is 54 dB the display
will show C/N > 48 dB.
On the other hand, if C/N=42 dB we will see on the display C/N = 42 dB.
C/N ratio measuring range.
Page 49
USERS MANUAL. PROMAX-10 Premium
APPENDIX F.- CARRIER TO NOISE RATIO FOR 64-QAM DIGITAL
CHANNELS (C/N).
For a 64-QAM DBV-C digital signal with a bandwidth of 8 MHz, the C/N ratio must be better than 20 dB. Measurements lower than this value will correspond to signals with a no acceptable quality.
Justification and example
The basic parameter which describes the quality of a digital signal i s the ratio of the number of erroneous bits to the total number of bits transmitted. This parameter is known as BER (Bit Error Rate).
The DBV Group (Digital Video Broadcasting) defines a 'quasi error free transmission' (ETR290 Measurements Group ETR 290) as that with less than 1 error event per transmission hour. For a 64 QAM DBV-C transmission, the pre-FEC error (Forward Error Correction) must be <1.E-4.
In digital transmissions, as there is not one carrier, it is better t o speak in Eb/ No terms. The relation between Eb/No and the C/N is given by the following equation:
E
b
)dB(N/C
+=
N
o
s
)M(log
2
*Rlog(*10)dB(
BW
)
Where:
E b = Energy per bit N o = Noise power in a 1 Hz bandwidth R s = Symbol rate M = Number of points of the constellation
BW = Bandwidth
For DVB-C, 64 QAM, Rs=6.875 Mbaud, BW=8 MHz,
C/N (dB) =Eb / No (dB) + 7.12.
For a BER of 1E-4, Eb/No ≈ 16 dB. Therefore C/N ≈ 23 dB.
Page 52 05/2004
Page 50
USERS MANUAL. PROMAX-10 Premium
APPENDIX G.- MEASUREMENT OF THE MO DULATION ERROR RATIO
(MER).
Analogue and digital carriers are very different i n terms of the signal c ontent and the power distribution over the channel. Therefore need to be measured differently.
The amount of distortion in a system is related to the total power of all of the carriers making accurate power measurements critical for optimum performance.
Instruments such as signal level meters that are designed to measure only analogue carriers will not accurately measure digital carriers.
Modulation error ratio ( Signal/Noise ( of the error power to the average power in an ideal at least 4 or 5 dB of margin from the system degradation. (noise) impairments, such as system noise and the second and third order beats ( and noise.
failure of the digital system. Unlike analogue systems where you can see degradations in Carrier/Noise ( to the point of system failure.
). This measurement takes in account not only amplitude noise, but also phase
CTB
Determining the
MER
) measurement used in analogue systems.
S/N
measurements are useful for early detection of non-transient
MER
of a signal is a key part of determining the margin from
MER
) performance, a poor
C/N
is defined as follows, expressed in dB:
log20
), used in digital systems is analogous to
MER
signal. Ideally you should have
QAM
where significant errors occur, to allow for
MER
is not noticeable on the picture right up
MER
magnitudeerrorRMS
represents the ratio
MER
)dB(
CSO
magnitudesymbolaverage
Englis h
05/2004 Page 53
Page 51
USERS MANUAL. PROMAX-10 Premium
RMS Error Magnitud e
Ideal Symb ol
Average Symbol Magnitude
64 QAM
Figure 48.-
Modulation error ratio (
set decoders require better than
23 dB MER
)
MER
to operate. To allow for system degradation, a margin of at least 3 or 4 dB is preferred. However, decoders require better than Typically the maximum
was chosen as the preferred measurement for cable TV because of it’s
MER
28 dB MER
displayed on portable analysers is about
MER
similar to the analogue Carrier/Noise (
to operate with margins of
34 dB
) measurement expressed in dB that most
C/N
people from cable industry are familiar with.
QAM 256
at least.
3 dB
.
set
Page 52
USERS MANUAL. PROMAX-10 Premium
APPENDIX H.- PRINCIPLE OF QAM MODULATION.
THE CONSTELLATION DIAGRAM
The modulation process implies to transfer the informat ion cont ained in a s ignal to a high frequency carrier. Modulation consisting of two carriers each one of the same frequency , one called I (in phase) and another out of phase 90° called Q (quadrature).
Each one of them is modulated in amplitude and phase by a portion of the digital input signal. The two modulated signals are combined then and they are transmitted as a single waveform. The receiver only needs to invert the process for generating a digital output that can be processed to produce images or anot her useful information also.
The number of levels used in the modulation of each carrier determines the number of possible symbols and, consequently, the number of bits that can be transmitted in a certain bandwidth. The
16 QAM, 32 QAM, 64 QAM, 128 QAM
For example, if four amplitude levels are applied to each one of the carriers, each signal could reach the value of -3.0, -1.0, +1.0, +3.0 at a given instant in time, so we have 16 possible combinations. This is known as exposition to four amplitudes, it allow us to generate 8 states for each carrier and 64 possible combinations (
64 QAM
).
, in concrete uses modulation in quadrature
QAM
standard allows 5 types of modulation:
DVB-C
and
256 QAM
.
16 QAM
. Extending the previous
These digital signals can be visualized graphically by means of the
Constellation Diagram
carrier (signal I or signal in phase) we would obtain the image in figure 49.
. If one imagines on an axis the possible states of the first
Figure 49.-
Figure 50.-
Signal I states
Signal Q states
Englis h
Page 53
USERS MANUAL. PROMAX-10 Premium
Figure 50 shows the other signal (Q in quadrature) on a vertical axis to consider the 90° phase change.
Figure 51.- I-Q
states combination
Figure 51 shows the combination of these two signals. This image constitutes the Constellation Diagram of the modulated digital signal.
Page 56 05/2004
Page 54
USERS MANUAL. PROMAX-10 Premium
APPENDIX I.- FREQUENCY OFFSET ADJUSTMENT FOR TUNING
ANALOGUE AND DIGITAL CHANNELS.
It is possible to introduce an offset of the tuning central frequency for the channels defined in each channel plan using the it is possible to adapt the measures to irregular tunings, due t o small displacements on channel central frequencies that are defined by the standard channel plan.
parameter (± 2 MHz). Therefore,
FCH
Figure 52.-
Taking as example the previous figure, if we suppose that in the configuration parameters of the digital channel (CH-33) a frequency offset equals to 1 MHz has been defined for the channel tuning frequency (∆FCH = -1MHz), although the central frequency corresponding to channel 30 (569 MHz) does not match with the defined one for the current CCIR channel plan (CH-33, 570 MHz), it will be possible to take measures corresponding to the Docsis normative for the standard FCC channel plan (CH-30, 569 MHz) with no need to modify the current channel plan or to ent er a new one.
Frequency offset on the tuning central frequency.
Englis h
Loading...