Features:
• 1500 Vdc Opto Isolation between HART and RS232 circuits
• Integral standard RS-232 female connector
• Integral heavy duty HART Minigrabber connectors
• Designed to meet HART Physical Layer Spec HCF_SPEC-54
• Self powered, no external power supply or batteries needed
• 48” (1.2m) cable length
• No driver installation necessary
• Compact size 2.75” x 2.0” x 0.8” (70mm x 51mm x 20mm)
• Rugged ABS plastic enclosure
• Custom cable lengths available
• HART Modem Test Program included
• HART Foundation Registered, Registration Number L2-06-1000-305
• CE certified
• Cost effective
• Each unit 100% tested
• 1 Year warranty
HM-RS232-ISO
HART® Modem - RS232
Electrically Isolated
MK-1038, 05/04/2016 Our Quality Management System is
ISO 9001:2008 Certified

Specifications:
Enclosure
Material High strength ABS Plastic
Dimensions 2.75” x 2.0” x 0.8” (70mm x 51mm x 20mm)
Total Unit Weight 4.8 oz (136 g)
Cables
HART Cable Length 4’ (1.2 m)
HART Cable Connectors Mini-Grabber
RS232 Cable Length 18” (0.5 m)
Cable Connector RS232 Female
HART
Interface Capacitive and transformer coupling
Output 600 mVpp
Leakage < 10 uA
Isolation Voltage 1500 Vdc
Connections Across loop resistor or HART field device
Specifications HART 4, HART 5, HART 6, HART 7,
HART Physical Layer Spec HCF_SPEC-54
HART Foundation Registration Number L2-06-1000-305
Computer
Operating System Windows XP, Vista (32/64), 7 (32/64), 8 (32/64), 10 (32/64)
RS232
Power RS232 port provides power to unit
Current Draw 17mA
Environmental
Operating Temperature 0 degC to 50 degC (32 degF to 122 degF)
Storage Temperature -40 degC to 85 degC (-40 degF to 185 degF)
Humidity 0% to 99% (non-condensing)
Compliances
CE Certification EN 61326 (EMC), Directive 2011/65/EU (RoHS)
ProComSol, Ltd
Process Communications Solutions
Phone: 216.221.1550
Toll Free: 877.221.1551
Fax: 216.221.1554
Email: sales@procomsol.com
Web: www.procomsol.com
MK-1038, 05/04/2016 Our Quality Management System is
ISO 9001:2008 Certified