© 2016 by Proceq SA 5
2. Tutorial
2.1. Principle
During measurement with Piccolo 2 / Bambino 2, an impact body with a hard metal ball indenter is
launched by spring energy against the sample to be measured, and then rebounds. Before and after
the impact, a permanent magnet inside the impact body passes through a coil in which a voltage
signal V is induced by the forwards and backwards movement. This induction signal behaves proportionally to the velocities. The ratio of the rebound velocity v
r
to the impact velocity vi multiplied by
1000 yields the hardness value HL (Leeb hardness). HL is a direct measure of the hardness. While
HLD refers to testing with the D device version (impact body 27 mm long, flat support ring), HLDL
refers to testing with the slim front section of the DL Accessory Kit (impact body 82 mm long, support
ring with slim long front section).
2.2. Sample Preparations
Keep the sample free of vibrations during the test. Light and thin parts must be specially fastened
(see “Testing light samples”). Ensure that the surface of the work piece is clean, smooth and dry. If
required, use appropriate cleaning agents for cleaning, such as acetone or isopropanol (no water!).
• Minimum sample weight
With a compact shape 5 kg (11 lbs)
On a solid base 2 kg (4.5 lbs)
Coupled to solid base 0.1 kg (0.25 lbs)
• Minimum sample thickness
Uncoupled 25 mm (1 inch)
Coupled 3 mm (1/8 inch)
Layer thickness 0.8 mm (1/32 inch)
• Sample surface condition
Roughness class ISO N7
Maximum roughness depth Rt 10 μm
Average roughness depth Ra 2 μm
• Maximum sample hardness 890 HLD (~ 950 HLDL, ~ 955 HV, ~ 68 HRC)
2.3. Testing Procedure
• Select a hardness unit (see section “Settings”).
• Select automatic compensation for impact directions other than vertical
down ( ), when testing in HLD / HLDL.
• Press to select the averaging mode AVE ( ).
• Conduct testing impacts by cycling through the “1. load – 2. position – 3, trigger” mechanism.
1. The impact device – when not in contact with the test piece – is held firmly with one hand and
the charging tube is depressed with the other hand until contact is felt. The charging tube is
allowed to slowly return to the starting position. Alternatively, this loading movement can be
performed in position after an impact (before the device is moved to a new position), or on any
other location that shall not be tested later on.
2. Position the support ring on the test piece. Take particular care to wholly seat the ring on the
test piece, but not coinciding with a previous test indentation.
3. To trigger an impact, push the actuation tube in the direction of the housing. After approx. 1
mm (0.04 inch), the impact body is released.
• After the last impact of a series, press the key repeatedly to display the hardness average and,
in case of the Equotip Piccolo 2, further statistics of the measurement.