Proceq Equotip 550 Operating Instructions Manual

Operating Instructions
60 Years of Innovation60 Years of Innovation
Swiss Precision since 1954

Table of Contents

1. Safety and Liability ....................................................4
1.1 General Information ............................................................. 4
1.2 Liability .................................................................................. 4
1.3 Safety Instructions ............................................................... 4
1.4 Correct Usage ...................................................................... 4
1.5 Optimizing Performance of the Battery System ............... 4
2. Getting Started ...........................................................5
2.1 Installation ............................................................................ 5
2.2 Main Menu ............................................................................ 6
3. Measurement .............................................................7
3.1 Performing Measurements ................................................. 7
3.2 Measurements Screen ...................................................... 10
3.3 Measuring Methods .......................................................... 12
3.4 Instrument Verification / Daily Performance Check ....... 24
4. Settings ....................................................................25
4.1 Measurements ................................................................... 25
4.2 Verification (Performance & Uncertainty Check) ............ 27
4.3 Conversions (Hardness Conversions) ............................. 28
4.4 Reporting ............................................................................ 28
5. Data (Explorer) .........................................................29
5.1 Measurements ................................................................... 29
5.2 Verifications ........................................................................ 31
6. Wizards .....................................................................31
6.1 Measurement Wizard ........................................................ 32
6.2 Device Verification ............................................................. 32
6.3 Impact Direction Calibration (Leeb only) ......................... 32
6.4 Conversion Curve Creation............................................... 32
6.5 Combined Method ............................................................. 34
6.6 Mapping Wizard (Coming Soon) ...................................... 34
7. Information ...............................................................35
7.1 Documents ......................................................................... 35
7.2 Upload PDF-Files from an USB-stick .............................. 35
8. System ......................................................................36
8.1 Features .............................................................................. 36
8.2 Probes................................................................................. 37
8.3 Hardware ............................................................................ 37
8.4 Date & Time ........................................................................ 37
8.5 Language ............................................................................ 37
8.6 Device information ............................................................. 37
9. Maintenance and Support .......................................38
9.1 Maintenance....................................................................... 38
9.2 Support Concept ............................................................... 39
9.3 Standard Warranty and Extended Warranty ................... 39
9.4 Disposal .............................................................................. 39
© 2016 Proceq SA 2
10. Troubleshooting ......................................................... 40
10.3 Battery ................................................................................... 42
11. Equotip Link Software ............................................... 42
11.1 Starting Equotip Link ............................................................ 42
12. Technical Specifications ............................................ 45
12.1 Instrument ............................................................................. 45
12.4 Equotip Portable Rockwell Probe ....................................... 47
13. Standards & Guidelines ............................................. 47
14. Ordering Information ...............................................48
14.1 Units .................................................................................... 48
14.2 Impact Devices & Probes ................................................. 48
14.3 Parts and Accessories ...................................................... 49
14.4 Test Blocks ..............................................................................
© 2016 Proceq SA 3

1. Safety and Liability

Table of Contents

1.1 General Information

This manual contains important information on the safety, use and main­tenance of the Equotip 550. Read through the manual carefully before the first use of the instrument.

1.2 Liability

Our “General Terms and Conditions of Sales and Delivery” apply in all cases. Warranty and liability claims arising from personal injury and dam­age to property cannot be upheld if they are due to one or more of the following causes:
• Failure to use the instrument in accordance with its designated use as described in this manual.
• Incorrect performance check for operation and maintenance of the in­strument and its components.
• Failure to adhere to the sections of the manual dealing with the per­formance check, operation and maintenance of the instrument and its components.
• Unauthorised modifications to the instrument and its components.
• Serious damage resulting from the effects of foreign bodies, accidents, vandalism and force majeure.
All information contained in this documentation is presented in good faith and believed to be correct. Proceq SA makes no warranties and excludes all liability as to the completeness and/or accuracy of the information.

1.3 Safety Instructions

The equipment is not allowed to be operated by children or anyone under the influence of alcohol, drugs or pharmaceutical preparations. Anyone who is not familiar with this manual must be supervised when using the equipment.
• Carry out the stipulated maintenance properly and at the correct time.
• Following completion of the maintenance tasks, perform a functional check.

1.4 Correct Usage

The instrument is only to be used for its designated purpose as described herein.
• Replace faulty components only with original replacement parts from Proceq.
• Accessories should only be installed or connected to the instrument if they are expressly authorized by Proceq. If other accessories are installed or connected to the instrument then Proceq will accept no liability and the product guarantee is forfeited.

1.5 Optimizing Performance of the Battery System

To increase the performance of the battery, it is recommended to first completely discharge and then completely charge it.
4 © 2016 Proceq SA

2. Getting Started

Table of Contents
The Equotip 550 is typically used for testing the hardness of metallic sur­faces. The user has a choice to select either Leeb rebound, the Portable Rockwell or UCI principle, see chapter “3.1 Performing Measurements”.
In combination with the Equotip Leeb Impact Device U the instrument is used to test the roll hardness of paper, film or foil rolls.

2.1 Installation

To install the Battery into the Equotip 550 Touchscreen Unit, lift the stand as shown, insert the battery and fasten it in place with the screw.
Buttons
On the upper right of the unit there are three buttons:
Power On/Off – Press to power on or to return to the home screen. Press and hold to power off.
Soft Key – Switches in and out of full screen view or toggles between the actual screen and the last viewed pdf document (eg. Operating Instructions).
Back Button – Returns to previous screen.
Energy Saving
Energy saving may be programmed as desired under System/Power set­tings, see chapter “8.3 Hardware”.
Connections
Figure 1: Insert Battery
There are three status LEDs on the right side of the display. The middle light is the power indicator which is red when charging and turns to green when battery is fully charged. The lower LED is used for application spe­cific notification.
A complete charge requires < 9 h (Instrument not operating)
Charging time is much longer if the instrument is in use.
An optional Quick Charger (Part No. 327 01 053) can be used to
© 2016 Proceq SA 5
NOTE! Only use the battery charger provided for charging.
charge a spare battery or to charge the battery outside of the instru­ment. In this case it takes < 5.5 h for a complete charge.
1 2
Snap-in connectors
Figure 2: Connections
For Leeb Impact Devices
use the Snap-in connector 1.
For UCI Probe
use the Snap-in connectors 1 or 2.
For the Portable Rockwell Probe use the USB Host connector.
USB Host
USB Device
Ethernet
Power Supply
USB Host:
Additionally connect a mouse, keyboard or USB stick.
USB Device:
Connect to PC.
Ethernet:
Connection to network.
Power Supply:
Connect the power supply through this connection.

2.2 Main Menu

Table of Contents
On start up the main menu is displayed. All functions may be accessed directly via the Touchscreen. Return to the previous menu by pressing the back button or the return icon (arrow) at the top left corner of the Touchscreen.
Settings: For application specific settings, see chapter “4. Settings”.
Measurement: Measurement display screen, see chapter “3. Measure­ment”.
Figure 3: Main Menu
6 © 2016 Proceq SA
Wizards: Task related workflows, see chapter “6. Wizards”.
Data (Explorer): File manager for reviewing measurement data on the instrument, see chapter “5. Data (Explorer)”.
System: System settings, e. g. language, display options etc, see chapter “8. System”.
Information: Operating instructions and other reference documents, see chapter “7. Information”.

3. Measurement

Table of Contents

3.1 Performing Measurements

3.1.1 Leeb Testing Procedure (except Leeb U)
Select automatic compensation for impact direction “Automatic”, see chapter “3.2.1 Controls”. If “Automatic” is not allowed, set the impact direction (         ). The Equotip Leeb Impact Devices DL doesn’t sup­port automatic mode. The impact direction must be selected manually.
Select the appropriate material group, hardness scales and number of impacts per measurement series. For more information see chapter “4. Settings”. Conduct impacts by cycling through “load, position and trig­ger” mechanism:
1. Load the impact device – while not in contact with the test piece – by hold­ing it firmly with one hand and sliding the loading tube with the other hand until contact is grabbed by the clutch.
2. Position the support ring on the test piece. Take particular care to fully position the support ring on the test piece, but not coinciding with a previ­ous test indentation.
3. To trigger an impact, press the trigger button to release the impact body. To perform another impact, repeat this cycle.
Figure 4: Leeb Testing Procedure
After the last of the impacts is performed, the hardness average and fur­ther statistics of the measurement series are displayed.
NOTE! Make sure the loading tube is allowed to slowly return back to the starting position. Do take care so the loading tube does not spring back uncontrolled, which may result in permanent device damage.
NOTE! If possible, follow the standard practice of Leeb re­bound hardness testing as described in the standards DIN 50156-1 (metallic materials), ASTM A956 (steel, cast steel and cast iron only), or other applicable standards. If these are not available, the user is recommended to average a minimum of n = 3 impacts at an indentation distance of 3 to 5 mm (0.12 to 0.20”) for each location of the sample that shall be tested.
NOTE! Do not carry out an impact in an area that has already been deformed by another impact. Also, do not load the device when it is already positioned in the new test location, since the material under the device may be affected through prior stress, and the catch chuck of the device may get damaged.
© 2016 Proceq SA 7
3.1.2 Portable Rockwell Testing Procedure
Table of Contents
1. Place the probe on the sample to test. For flat surfaces the standard foot is most suitable. For cylindrical objects it is recommended to use one of the special feet. For locations difficult to access a tripod foot can be used. See chapter “14. Ordering Information” for more details.
2. Press the probe slowly but firmly against the surface to perform the measurement. Suppress vibrations as much as possible, and follow the instructions on the screen.
3. Release the probe when the instru­ment says so. Again, this movement has to be done in a controlled man­ner. If the probe is released too fast, a warning will be shown and the measurement should be repeated.
Figure 5: Portable Rockwell Testing Procedure
3.1.3 UCI Testing Procedure
1. Place the probe on the sample to test. The probe must be perpen­dicular to the surface (± 5°). The special foot can be used to increase repeatability and reduce distor­tion of the results, see chapter “14. Ordering Information”.
2. Press the probe slowly but firmly against the surface until the se­lected measuring force is reached. The instrument will indicate when to release the probe with an on-screen prompt and audible sound.
3. Release the probe from the mate­rial. It is important to remove the probe completely from the test object. Otherwise the results can be biased.
Figure 6: UCI Testing Procedure
NOTE! A warning will be shown if the user applied too much load when pressing the probe against the surface. Please avoid frequent overloading, as this could seriously damage the probe.
8 © 2016 Proceq SA
3.1.4 Leeb U Testing Procedure
Table of Contents
The Equotip 550 Leeb U enables the user to quickly and precisely diag­nose roll imperfections, hardness inconsistencies and uneven winding, thereby preventing problems for printing and converting operations.
With Equotip Impact Device Leeb U automatic impact direction mode is not supported and the user must select the appropriate impact direction manually (90° down, 45° down, 0°).
As for roll hardness testing no conversion curves are used, no material group has to be selected.
Conduct impacts by cycling through “position and trigger”.
1. Position the probe on the roll to be tested. Make sure to fully posi­tion the support ring on the roll to ensure an impact perpendicular to the test surface.
2. While holding the impact device firmly with two hands, slide the loading tube smoothly down to load and trigger an impact.
Move the impact device to the next point on the roll and repeat.
Figure 7: Leeb U Testing Procedure
NOTE! Some features mentioned in this Operating Instruc­tions are specifically addressing metal hardness testing ap­plications and therefore not available for Equotip Leeb U.
© 2016 Proceq SA 9

3.2 Measurements Screen

Table of Contents
3.2.1 Controls
File Name: Enter the file name and tap return. Saved measure­ments will be stored automati­cally. If filename management is activated, this function is locked.
Impact Direction: To manually set the impact direction if required (Leeb only, by default this is automatic).
Material: Select the material group to use for conversions (not available for Leeb U).
Figure 8: Measurement Screen
10 © 2016 Proceq SA
Measurement Scale: Select the hardness scale (primary and secondary) to be displayed (not available for Leeb U).
Measurement Mode:
Switch between measurement and conversion.
Wizards: Direct access to wizards.
Redo: Restart the measurement series or a single measurement.
Time and battery status
Settings: direct shortcut to
settings menu. (only applicable to the actual measurement series)
Save: Store measurement data.
Delete: Delete the last measurement.
3.2.2 Measurement Views
HL
ID
+
Table of Contents
Equotip 550 is fully customizable as a device and can display three dif­ferent measurement views simultaneously. Each view can be switched to meet the user’s requirements by simply clicking on the icon related to the particular display at the top right corner of each screen.
Signal View: Display the probe signal from the last active measurement. This may be useful for advanced evaluations.
Statistic View: View statistics for the active measurement se­ries. Number of impacts (n), Average (x), Standard deviation (σ), Minimum/Maximum () and Range () are displayed in the primary scale.
Table View: Display the measurements for the active series in a table format.
Conversion View: Display the actual value on the conversion curve.
Bar View: Display the measurements of the series as a histo­gram.
Profile View: Display the measurement results as a profile.
Info: Display the measurement settings e.g. series length,
probe type, material group etc. User’s View: The user can choose between probe angle, mini-
mum, maximum, range and probe type for the field contents. To change, tap on each individual box.
Single Record View: Display the last or selected measurement result in both the primary and secondary hardness scales.
Sample ID’s: Defines the custom field.
Figure 9: Measurement Views
© 2016 Proceq SA 11
NOTE! Screen views cannot be duplicated.

3.3 Measuring Methods

Table of Contents
Equotip 550 family of instruments is capable of accepting three diffrent test methods using a single indicating unit.
3.3.1 Equotip Leeb
3.3.1.1 Test Principle
Release button
Loading tube
Impact Spring
Catch chuck
Connection cable – 4 pole
Sensor of impact device with ID ROM
Support ring
Figure 10: Schematic View of a Leeb Impact Device
Guide tube
Impact Body
During measurement with Equotip 550 impact devices (D, DL, DC, C, G, S, and E); an impact body with a ball indenter is launched by spring energy against the sample to be measured, and then rebounds. Before and after the impact, a permanent magnet inside the impact body passes through a coil in which a voltage signal is induced by the forwards and backwards movement. This induction signal behaves proportionally to the velocities. The ratio of the rebound velocity vr to the impact velocity vi multiplied by 1000 yields the hardness value HL (Leeb hardness). HL is a direct measure of the hardness. The third resp. fourth letter of the HL unit refers to the impact device HLD D impact device.
v
r
HL =
·1000
v
i
Equotip Leeb U
Although the Equotip Leeb Impact Device U is constructed differently to simplify the measurement process, the underlying principle is the same.
Figure 11: Equotip Leeb U Impact Device
Existing Parotester impact devices type U are fully supported by the Equotip 550. Typ P and PG impact devices can be still used, but the unit is shown as HLU although it would be actually LP resp. LPG.
NOTE! HLU values can be directly compared to LU on exist­ing Parotester instruments.
12 © 2016 Proceq SA
3.3.1.2 Sample Preparations
Table of Contents
Keep the sample free of vibrations during the test. Light and thin parts must be specially fastened, see chapter “3.3.1.6 Testing Light Samples”. Ensure that the surface of the work piece is clean, smooth and dry. If re­quired, use appropriate cleaning agents for cleaning, such as acetone or isopropanol. Do not use water or any other detergent fluids.
NOTE! Please use the surface roughness comparator plate provided to estimate the average roughness of the test piece prior to testing.
Figure 12: Surface Roughness Comparator Plate
3.3.1.3 Standards
Brief descriptions of referenced standards:
DIN 50156
ASTM A956
ASTM A370
ASTM E140
ISO 18265
ISO 16859
Leeb hardness testing of metallic materials Standard test method for Leeb hardness testing of steel
products Test methods and definitions for mechanical testing of
steel products Standard hardness conversion tables for metals rela-
tionship among Brinell, Vickers, Rockwell, Superficial, Knoop, Scleroscope and Leeb hardness
Metallic materials – Conversion of hardness values Leeb hardness testing of metallic materials
© 2016 Proceq SA 13
3.3.1.4 Test Conditions
Table of Contents
To ensure proper hardness readings, the following conditions must be fulfilled. If one or more conditions are not met, the measurement result may be significantly false.
Impact device type D/DC/DL/S/E G C
Surface preparation
Roughness grade class ISO 1302 Max. roughness depth R Average roughness R
Minimum sample mass
Of compact shape (kg / lbs) On solid support (kg / lbs) Coupled on plate (kg / lbs)
Minimum sample thickness
Uncoupled (mm / inch) Coupled (mm / inch) Surface layer thickness (mm / inch)
Minimum space
Between indentation and sample edge (mm/inch)
Between indentations (mm/inch) 3 / 0.12 4 / 0.16 2 / 0.08
Indentation size on test surface
With 300 HV, 30 HRC
With 600 HV, 55 HRC
With 800 HV, 63 HRC
Table 1: Leeb Test Piece Requirements
(μm / μinch)
t
(μm / μinch)
a
Diameter (mm / inch) Depth (μm / μinch) Diameter (mm / inch) Depth (μm / μinch) Diameter (mm / inch) Depth (μm / μinch)
N7 N9 N5
10 / 400 30 / 1200 2.5 / 100
2 / 80 7 / 275 0.4 / 16
5 / 11 15 / 33 1.5 / 3.3
2 / 4.5 5 / 11 0.5 / 1.1
0.05 / 0.2 0.5 / 1.1 0.02 / 0.045
25 / 0.98 70 / 2.73 15 / 0.59
3 / 0.12 10 / 0.4 1 / 0.04
0.8 / 0.03 0.2 / 0.008
5 / 0.2 8 / 0.3 4 / 0.16
0.54 / 0.021 1.03 / 0.04 0.38 / 0.015
24 / 960 53 / 2120 12 / 480
0.45 / 0.017 0.9 / 0.035 0.32 / 0.012
17 / 680 41 / 1640 8 / 2560
0.35 / 0.013 0.30 / 0.011
10 / 400 7 / 280
14 © 2016 Proceq SA
3.3.1.5 Selecting the Equotip Leeb Impact Device
Table of Contents
For optimized testing of diverse metallic materials and sample geometries, a range of impact devices are available as per “Table 1: Leeb Test Piece Requirements”.
Type G: Increased impact energy. For solid and inhomogenous components, e.g. heavy castings and forgings.
Impact energy: 90 Nmm
Type DC: Short device. For use in very
confined spaces, e.g. in holes, cylinders or for internal measurements on assembled machines. Impact energy: 11 Nmm
Type D: Universal unit. For the majority of your hardness testing requirements.
Impact energy: 11 Nmm
Type S: Si3N4 ball indenter. For testing
especially in the very high hardness range (in excess of 50 HRC / 650 HV): Tool steels with high carbide content inclusions.
Impact energy: 11 Nmm
Figure 13: Equotip Leeb Impact Devices
© 2016 Proceq SA 15
Type E: Diamond ball indenter. For testing in the very high hardness range (in excess of 50 HRC / 650 HV): Tool steels with high carbide content inclusions. More durable than type S.
Impact energy: 11 Nmm
Type C: Reduced impact energy. Surface
hardened components, coatings, thin walled or impact sensitive components (small measuring indentation).
Impact energy: 3 Nmm
Type DL: Slim front section.
For measurements in confined spaces, at the base of grooves or on recessed surfaces. Impact energy: 11 Nmm
3.3.1.6 Testing Light Samples
Table of Contents
If the samples are lighter than specified in chapter “3.3.1.4 Test Condi­tions” or sample sections have unfavorable mass distribution they can vibrate as the impact body hits the test point. This results in unwanted energy absorption. Such samples shall be supported by solid worktops. If the mass falls below the specific requirements but still exceeds the coupling amount then coupling it to a larger mass can help prevent vibra­tions.
The following requirements must be met for coupling:
• The contact surface of the sample and the surface of the solid support must be level, flat and ground smooth.
• The sample must exceed the minimum sample thickness for coupling. Follow the coupling procedure.
• Apply a thin layer of coupling paste to the contact surface of the sample.
• Press the sample firmly against the support.
• Push the sample in a circular motion and carry out the impact as usual, perpendicular to the coupled surface.
NOTE! Clamping may strain the sample, which can affect the hardness readings.
3.3.1.7 Testing Curved Surfaces
The instrument works properly only when the ball indenter at the front of the impact body is precisely at the end of the tube at the time of the im­pact. When concave or convex surfaces are tested, the ball indenter either does not entirely leave the test tube or comes out too far. In such cases, replace the standard support ring by a specially suited ring, see chapter “14. Ordering Information” or contact your local Proceq representative.
Figure 14: Leeb Testing on Curved Surfaces
16 © 2016 Proceq SA
3.3.1.8 Testing Thin Samples
Table of Contents
Pipes and tubes sometimes have mass distributions that can affect the result of the Leeb hardness test due to vibration. During on-site testing of pipelines, for example the test locations cannot be supported by solid worktops or clamped.
To benefit from the convenience and speed of the Leeb test, the user can make use of a custom conversion after conducting the following calibra­tion procedure, for example:
• Data pairs are measured on reference samples. For the Leeb HLDL ref­erence measurements, it is crucial that they are done on parts that are installed in the same way as those to be tested on-site. For example, two pipe samples “Pipe type 5 mm Duplex soft” (730 HLDL / 255 HB) and “Pipe type 5 mm Duplex hard” (770 HLDL / 310 HB) are measured using the Equotip Leeb impact device DL and a Brinell tester, respectively.
• The original HLDL-HB conversion curve for “1 Steel and cast steel” is now adapted using the two data points. The detailed procedure on how to create custom conversion curves in the Equotip 550 is given in chapter “6.4 Conversion Curve Creation”.
• To measure “Pipe type 5 mm Duplex” in future, it can be selected via “Material” – “Pipe type 5 mm Duplex”, using the hardness scale “HB Brinell” also see chapter “6.4.3 Example of a Custom Conversion (Two­Point Method)”.
NOTE! The user needs to determine and qualify the adapta­tion of conversion curves for each tube diameter and wall thickness. Guides to the procedure are provided in Nordtest Technical Report Series 424, Reports 99.12/13 and ASME Final Report CRTD-91.
3.3.1.9 Material Groups
No need to select any material when measuring in the native Leeb re­bound scale HL as no conversion is applied. In contrast, hardness scale conversions are correct only when the appropriate material group is se­lected. Free online material databases and the Equotip 550 on-board ref­erence documents can be useful to assign your materials to one of the default material groups. Suitability of conversions should be qualified on calibrated samples before use. For further information, please consult a Proceq representative.
NOTE! For a given test principle (native scale), the dropdown menu only lists the material groups for which conversions are available.
NOTE! If there is no conversion curve available, the user has the possibility to create its their own, see chapter “6.4 Con­version Curve Creation”.
© 2016 Proceq SA 17
NOTE! It is important to include all the critical information about the geometry of the test sample.
D/DC DL S E G C
Table of Contents
Steel and cast steel
Cold work tool steel
Stainless steel
Cast iron lamellar graphite GG
Cast iron, nodular graphite GGG
Cast aluminium alloys
Copper/zinc alloys (brass)
CuAI/CuSn-alloys (bronze) Wrought copper alloys, low alloyed
Table 2: Overview of Available Conversions
Vickers Brinell Rockwell
HV HB
HRB HRC HRA
Shore Rm N/mm²
HS
σ1 σ2 σ3
Vickers RockwellHVHRC Vickers Brinell Rockwell
HV HB HRB
HRC Brinell Vickers Rockwell Brinell Vickers Rockwell Brinell Vickers Rockwell Brinell Rockwell
HB
HV
HRC
HB
HV
HRC
HB
HV
HRB
HB
HRB
81-955 81-654 38-100 20-68
30-99 275-2194 616-1480 449-847 80-900 21-67 85-802 85-655 46-102 20-62 90-664 90-698 21-59 95-686 96-724 21-60 19-164 22-193 24-85 40-173 14-95
80-950 81-646
101-964
101-640 37-100 21-68
22-70
61-88 31-97 275-2297 614-1485 449-849 80-905 21-67
28-104
340-2194
615-1480
450-846
104-924
22-68 * 119-934
105-656
70-104
21-64 * * * 92-326 *
* * * 127-364
20-187 21-191
20-184
22-196
* * * * *
84-1211 83-686
20-72 61-88 29-103 283-2195 616-1479 448-849 82-1009 23-70 88-668 87-661 49-102 20-64
23-176 22-198
90-646 48-100
305-2194 618-1478 450-847 * 98-942
* *
19-37 19-168
24-86
Brinell HB 60-290 * * * * * Brinell HB 45-315 * * * * *
*Custom conversion curve / correlation
81-1012 81-694
20-70
30-102 275-2194 615-1479 450-846
20-67
*
21-167
23-85
18 © 2016 Proceq SA
3.3.2 Equotip Portable Rockwell
Table of Contents
3.3.2.1 Test Principle
During measurement with Equotip 550 Portable Rockwell probe, a diamond indenter is forced into the test piece, and then released back out of the ma­terial. The indentation depth is measured continuously during this process. Indentation depth is calculated after decreasing the total load to preload.
10N
Figure 15: Portable Rockwell Test Principle
50N
10N
3.3.2.2 Sample Preparations
Ensure that the surface of the work piece is clean, smooth and dry. If required, use appropriate cleaning agents for cleaning, such as acetone or isopropanol. Do not use water or any other detergent fluids.
3.3.2.3 Measurement on Basis of DIN 50157
Both depth measurements d1 and d2 are taken at preload, first during application (d1) then after release of the total load (d2). The difference be­tween d1 and d2 originates from the deformation response of the material to penetration.
NOTE! By calculating the penetration depth between the preload and total load, surface roughness discrepancies are significantly disregarded.
NOTE! The hardness testing principle in Portable Rockwell follows the Rockwell stationary test. As for the Rockwell test, no adjustment for the test direction is required. How­ever, there are three main differences to traditional stationary Rockwell tests:
• The test loads are lower.
• The Portable Rockwell indenter is sharper.
• The dwell times during the test are shorter.
NOTE! “MM” stands for “mobile mechanical” measurement, an ancillary that is required by the German standard DIN 50157 to explicitly denote the lower applied loads, sharper indenter shape and shorter loading times during a measure­ment. The different denomination is formal, i.e. the HMMRC results should be very close if not equal to stationary HRC readings.
© 2016 Proceq SA 19
3.3.2.4 Test Conditions
Table of Contents
To ensure proper hardness readings, the following conditions must be fulfilled. If one or more conditions are not met, the result may be signifi­cantly false.
3.3.2.5 Installing the Measuring Clamp
The clamp is designed to facilitate the hardness testing of very thin or small samples.
Probe setup 50 N
probe with clamp
50 N probe with round standard foot (ø = 42 mm)
Minimum test piece thickness
Maximum test piece thickness
Test piece surface condition
40 mm N/A
recommended mean surface roughness
< 2 μm to minimize data scatter
R
a
1 mm at ~20 HB
130 μm at ~70 HRC
foot to be
Surface curvature
used for plane surfaces
Maximum test piece hardness
Minimum spacing
Table 3: Portable Rockwell Test Piece Requirements
three times the diameter of a test indentation
70 HRC
50 N probe with tripod
very small cur­vatures accepta­ble
50 N probe with spe­cial feet
18 – 70 mm radius of curvature or 70 mm –
Cantilever
Cantilever screw
Knurled screw
Probe holder
Figure 16: Portable Rockwell Clamp
• Use the 3 mm Allen key setup tool to release the cantilever. Turn it by 90°.
• Take the probe and remove the foot. The diamond indenter remains mounted.
• Screw the probe into the probe holder of the clamp clockwise (hand-tight).
• Turn the cantilever so its tip is centred over the probe; tighten the can­tilever screw securely using the 3 mm Allen key setup tool.
• The recommended clearance between the bottom of the probe holder and the sample surface should be between 2 and 5 mm. Adjust the height with the two knurled screws.
20 © 2016 Proceq SA
NOTE! In case the probe connector is in an inconvenient
Table of Contents
position, release the set screw. Ensure that the springs in the mechanism do not get lost. Turn the mechanism into a convenient position, aligning the set screw with the guide channel. Lock the set screw so that the probe holder will still slide up and down without rubbing on the set screw.
3.3.2.6 Considerations
• When measuring cylindrical samples with adapters Z4 or Z4+28, make sure, the sample is not twisted on the clamp support. This is best ensured when the back part of the clamp rests on a table and only the sample support of the clamp sticks out over the table’s edge.
• When applying the load, slowly squeeze the leavers and allow the sample to adjust to the support. During the measurement, do not touch the sample, if possible. When releasing, grab the sample again.
• Whenever the sample geometry (i.e. the wall thickness) allows it, free­hand measurements usually offer better measuring performance. This applies particularly to measurements on cylinders.
• For small diameter rods (or stiff enough pipes), the V-notch clamp adapter Z2 has been designed. When installing the Z2 support ensure that the centre of the V-notch is centred underneath the probe holder.
3.3.2.7 Installation of Standard Foot or Tripod
The round standard foot permits meas­urements on test objects that are only accessible from one side, such as large metal sheets. The tripod is used when the flat foot cannot be placed on the test piece without wiggling.
1. The diamond indenter remains mounted.
2. Install the foot on the probe.
Figure 17: Portable Rockwell
with Tripod
3.3.2.8 Installing the Special Foot
Two special feet extend the Portable Rockwell application range to cy­lindrical test pieces.
1. The diamond indenter remains mounted.
2. Install the foot on the probe.
3. Place the foot on the test piece and release the set screw on the foot. Then press down the probe onto the test piece and lock the set screw.
Figure 18: Portable Rockwell
Special Feet
© 2016 Proceq SA 21
3.3.2.9 Conversion Standard
Table of Contents
Measurements in HV and HRC are direct correlations therefore no con­version is required. The user has the option of either ASTM E140 or ISO 18265 for conversion to any other scales.
3.3.3.2 Sample Preparations
Ensure that the surface of the work piece is clean, smooth and dry. If required, use appropriate cleaning agents for cleaning, such as acetone or isopropanol. Do not use water or any other detergent fluids.
3.3.2.10 Material Groups
Since Portable Rockwell is based on static indentation principle, the hardness conversions are less dependent on material specific proper­ties for the majority of times.
The user still has the possibility to apply customer conversion curves if required, see chapter “6.4 Conversion Curve Creation”.
3.3.3 Equotip Ultrasonic Contact Impedance (UCI)
3.3.3.1 Test Principle
The UCI method uses the same pyramid-shaped diamond as a conven­tional Vickers hardness tester. Unlike Vickers testing, no optical evalua­tion of the indentation is required, enabling fast and portable measure­ments. The UCI method excites a rod into an ultrasonic oscillation. The test load is applied by a spring and typically ranges from 1 to 5 kg of force (HV1 – HV5). As the diamond is forced into the material, the frequency of the rod oscillation changes in response to the contact area between the diamond and the material under test. The instrument detects the shift in frequency, converts it to a hardness value which is immediately displayed on the screen.
3.3.3.3 Standards for UCI measurements
There are two standards which describe the UCI measurements, resp. the instrument:
DIN 50159 Hardness testing with the UCI method ASTM A1038 Standard test method for portable hardness testing by
the Ultrasonic Contact Impedance method
For conversions from one hardness unit to another, the user can chose between the following standards:
ASTM E140 Standard hardness conversion tables for metals rela-
tionship among Brinell, Vickers, Rockwell, Superficial, Knoop, Scleroscope and Leeb hardness
ISO 18265 Conversion of hardness values
22 © 2016 Proceq SA
3.3.3.4 Test Conditions
Table of Contents
To ensure proper hardness readings, the following conditions must be fulfilled. If one or more conditions are not met, the result may be misleading.
Probe setup HV1 (~10 N) HV5 (~50 N)
Minimal required thickness
Minimal required weight
Required surface roughness
Acceptable surface curvature
Minimum space
Indentation size on test surface
300 HV, 30 HRC
600 HV, 55 HRC
800 HV, 63 HRC
Table 4: UCI Test Piece Requirements
Grade class N8 N10
Maximum roughness 15 μm / 600 μinch 60 μm / 2400 μinch
Average roughness 3.2 μm / 125 μinch 12.5 μm / 500 μinch
Indentation to edge 5 mm / 0.2 inch
Between indentations 3 mm / 0.12 inch
Depth 11.3 μm / 445 μinch 25.3 μm / 996 μinch
Diagonal 79.1 μm / 3114 μinch 177.1 μm / 6972 μinch
Depth 8 μm / 315 μinch 17.9 μm / 705 μinch
Diagonal 56 μm / 2205 μinch 125.3 μm / 4933 μinch
Depth 6.9 μm / 272 μinch 15.5 μm / 610 μinch
Diagonal 48.3 μm / 1900 μinch 108.5 μm / 4272 μinch
5 mm / 0.2 inch
0.3 kg / 0.66 lbs
Radius > 3 mm
© 2016 Proceq SA 23
3.3.3.5 Installation of the Special Foot
Table of Contents
5 º
The default foot allows the measurement to be performed on every surface. The probe must be perpendicular to the sur­face (± 5°). The special foot can be used to increase the re­peatability and avoid the distortion of the results, see chapter “14. Ordering Information”.
1. Unscrew the standard foot and remove it
2. Screw the special foot to the probe tightly
NOTE! To measure in places with limited accessibility, the probe can be used without any foot. If doing so, the side of the rod of the probe must not touch any surface or be han­dled, as this leads to biased readings.
3.3.3.6 Conversions into Other Units
The frequency shift measured by the UCI probe is not only influenced by the hardness, but also by its elastic properties. The default conversion curve from the frequency shift to Vickers is valid for low-alloyed steel with an e­modulus of 210 ± 10 GPa. As soon as a material has to be tested with a different e-modulus, this existing conversion curve must be adapted. The best way to do so is to calibrate the instrument on the material to be tested. The Equotip 550 offers for this a fast and easy way. Once the hardness value is converted to Vickers, it can be further converted to any other available hardness unit according to either the ASTM E140 or the ISO 18265. Another option is to adjust the default conversion based on the measurement of Portable Rockwell or Leeb. To do so, see chapter “6.5 Combined Method”.
3.4 Instrument Verification / Daily Performance Check
See chapter “6.2 Device Verification” and follow the on-screen proce­dure. After the verification process your instrument is fully operational and you can now continue with your measurements.
NOTE! The performance check should be done regularly be­fore using the instrument to verify the mechanical and elec­tronic functions of the probe and the indicating device. This requirement is also included in the relevant hardness stand­ards, see chapter “13. Standards & Guidelines”.
24 © 2016 Proceq SA

4. Settings

Table of Contents
Figure 19: Settings Menu

4.1 Measurements

If changes are made here, they are set for all the upcoming measure­ment series. If the settings menu is accessed through the measurement screen, changes are only valid for the actual measurement series.
4.1.1 Probe Type
Probe types are automatically recognised by the device. Default settings can be set and will be used for each measurement device. If the meas­urement settings are accessed from the measurement screen, the active probe can be selected.
4.1.2 Test piece management (Coming Soon)
Complete sets of settings can be managed in this menu. It allows the user to save, edit, recall or delete these sets. This can be used to have a quick access to different measurement settings, i.e. for different test pieces or applications.
4.1.3 Measurement Parameters
Material
The desired material group can be selected from the default list, in addition you may predefine custom material groups which will be displayed here. For custom material/curves please refer to chapter “6.4 Conversion Curve Crea­tion”. For more information on material groups related to Leeb please refer to chapter “3.3.1.9 Material Groups”, for Portable Rockwell chapter “3.3.2.10 Material Groups” and for UCI “3.3.3.6 Conversions into Other Units”.
Primary and Secondary scales
The user has the possibility to select two different scales in which the measurement results are displayed.
NOTE! Conversion for HLD to HV, HB and HRC are standard­ized as per ASTM E140. Conversion for Portable Rockwell (μm) and UCI can be switched between ASTM E140 and ISO
18265).
NOTE! Equotip Leeb Impact Device U does not support con­version curves, therefore these settings are not available.
Conversion Standards – Leeb
The conversion standard for Shore hardness HS switched between the default conversion according to ASTM E448 or the Japanese conversion according to JIS B7731.
NOTE! Measurements for certain types of Steel can be con­verted to tensile strength according to DIN EN ISO 18265.
Conversion Standards – Portable Rockwell
The default measuring method DIN 50157, is applicable for testing all metallic material, and it generally yields more consistency. For conver­sions, user has the choice to select either ISO or ASTM.
© 2016 Proceq SA 25
Conversion Standards – UCI
Table of Contents
The default measuring method is according to ASTM A1038 and DIN 50159. For conversions, the user has the choice to select either ISO or ASTM.
Use Advanced Algorithm (Portable Rockwell only)
Advanced Algorithm provides faster measurements. This is especially useful when testing softer material.
Impact direction (Leeb only)
With exception to DL and U devices all Leeb impact devices have auto­matic direction compensation. You may override this and set the impact direction manually. For more information on impact direction please refer to chapter “3.1.1 Leeb Testing Procedure (except Leeb U)”. Impact direc­tion is not relevant to the Portable Rockwell and the UCI devices.
Trigger Load (UCI only)
For the Equotip UCI probe the load where the measurement will be trig­gered can be chosen in the range from HV1 to HV5 (10 – 50 N). The trig­ger load cannot be changed once the measurment series was started.
Units (Portable Rockwell only)
For Portable Rockwell probe, choose to display the indentation depth in metric or imperial units.
4.1.4 Sample IDs
After Measurement
Use this setting to define if the current sample ID’s should be kept for the next measurement series or deleted.
Edit Entries
The entries of the different sample ID fields can be deleted or edited here. For easy increasing or decreasing, use the up- and down- arrows. For add­ing or removing entry fields, please see chapter “8.1.2 General Features”.
4.1.5 Workflow
Activate User Guidance
Select to display on screen instructions and messages when taking a measurement.
Auto Close Series
Automatically end a series after a set number of measurements. The user can set the series from 1 to 1000 measurements.
Measurement Comment Handling
Use this setting to allow or disallow the user to enter a comment at the completion of a measurement series. When set to “free” this enables the user inputing a comment.
Measurement Series Filename
Enter the file name for the measurement series will be stored. This pos­sibility is disabled, if filename management is activated.
Save to Folder
Set the folder location where the measurement series file to be stored. This option is disabled, if folder management is activated.
Store Signal Data (Leeb only)
Select to store the raw waveform for Leeb measurements. For Portable Rockwell the signal form will be stored for each measurement automatically, for UCI this option is not available..
NOTE! Storing signal data will cause measurement files to take up more memory.
Enable Warnings
Choose to enable warning display signals and sounds to indicate false measurements.
26 © 2016 Proceq SA
Use report templates
Table of Contents
Here a template for the report can be selected. By default, the default template will be used. This default template can be selected in the tem­plate manager.
Operator
Here the test operator can be edited. This operator name is stored for the following measurements, but not for the verifications.
4.1.6 Limits
Enable Upper and Lower Limits
Select to enable the display of the upper and lower tolerance limits for the measurements. Specific color coding is adopted to differentiate between upper and lower limits.
4.2 Verification (Performance & Uncertainty Check)
To see how a verification can be performed please refer to chapter “6.2 Device Verification”.
4.2.1 Test Block Management
It is important to verify the correct functionality of the instrument on a testblock calibrated in the genuine native scale of the probe being used. In the test block management section, different test blocks information can be stored. Test blocks that are listed here can then be used during the verification process.
4.2.2 Workflow
Standard
Select the standard according to which the verification should be performed. It can be selected between ISO, ASTM or a customer defined standard.
Minimum series count
The minimum required number of measurements can be selected here. If a standard was selected before, this setting is fixed.
Maximum series count
The maximum allowed number of measurements can be selected here. If a standard was selected before, this setting is fixed.
Operator reference
Here the reference operator name can be entered if required. This name will be used for the verification processes. If no name is entered here, the user can still enter it during the verification process.
NOTE! First time user: It is recommend to complete a “Tu­torial on Leeb and Portable Rockwell Hardness Testing” as appropriate or watch a demonstration by a qualified Proceq representative.
NOTE! The performance check should be done regularly be­fore each use of the instrument to verify the mechanical and electronic functions of the impact device and the indicat­ing device. This requirement is also included in the DIN and ASTM Leeb hardness standards, see chapter “8.1 Features”.
4.2.3 Verification Standards and Extended Uncertainty
It is recommended for the instrument to be verified prior to testing. This gives the user the extra assurance that the device is working correctly and the measurement data are accurate. Although the verification pro­cess is similar on all Leeb, Portable Rockwell (mechanical penetration depth) and UCI standards, user has the option to comply with the pre­ferred standard/verification procedure.
© 2016 Proceq SA 27
DIN 50156
Table of Contents
DIN 50157
DIN 50159 ASTM A956
ASTM A1038
ISO 16859 Extended Uncertainty
(Combined Uncertainty)
Leeb hardness testing of metallic materials. Will be re­placed by ISO 16859. Hardness testing of metallic materials with portable measuring instruments operating with mechanical pen­etration depth. Hardness testing with the UCI method. Standard test method for Leeb hardness testing of steel products. Standard test method for portable hardness testing by the Ultrasonic Contact Impedance method Hardness testing of metalic materials by Leeb. Measurement uncertainty analysis is applied to under­stand the differences in test results and to determine sources of error. The uncertainty of an Equotip Leeb, Equotip Portable Rockwell or Equotip UCI hardness testing system consists of a statistical component, a component inherent to the measurement device and a component arising from the metrological chain between national standard and the user device (traceability). Although uncertainty could be a complicated topic, Equotip 550 automatically calculates the combined un­certainty of the system. All the required information is al­ready available in the calibration certificates provided by Proceq. Therefore the device only requires adding these values in the specified fields and following the simple steps on the display in order to complete the process.

4.3 Conversions (Hardness Conversions)

There is no direct correlation between any two hardness scales. There­fore conversions must be determined by comparison testing for any given alloy.
4.3.1 Standard Conversions
Proceq has developed correlations to convert the Leeb hardness meas­urements to other commonly used hardness scales based on groups of alloys that have a close relationship. The conversions for HLD and Material Group 1 (Carbon Steels) are standardized according to ASTM E140-12b.
4.3.2 Custom Conversion Curves
See chapter “6.4 Conversion Curve Creation”.
4.3.2.1 Custom Compensation
In some cases a user must measure hardness on many samples with identi­cal size and shape that is below the ideal limits for accuracy. Studies have been published by ASME and Nordtest that have identified and confirmed the validity of the strategy to apply a compensation factor to correct for the inaccuracies induced by the non-ideal geometry. The methods outlined in chapters “6.4 Conversion Curve Creation” can be applied to create this compensation factor to be applied automatically to the Equotip test result.

4.4 Reporting

The content of the measurement reports can be adjusted here.
4.4.1 Images Explorer
Images, i.e. company logos can be loaded from an USB stick on the de­vice, for use in reports. Pictures must be in the *.png or *.jpg format, ideally come with 72dpi and a maximum resolution of 496x652 pixel.
Upload Images from an USB-stick
To do so follow the steps below.
• Create the folder “PQ-Import” in the main directory of the USB-stick
(not as a subfolder in another folder) and fill it with all the picture-files to be uploaded to the Equotip Touchscreen
28 © 2016 Proceq SA
• Connect the USB-stick to the USB Device plug on the left side of the
Table of Contents
Equotip Touchscreen
• Click on
• The uploaded images appear in the Images Explorer
and confirm with click on
NOTE! The USB stick must be either formatted in FAT or FAT32. NTFS is not supported.
4.4.2 Report template explorer
The report templates can be managed here. Either the default template can be used, or a fully customized template can be created and edited. Templates can also be copied or exported to an USB stick.
4.4.3 Reporting via PDF
It’s possible to create reports directly on the instrument as PDF and store them on USB stick. Choose the measurement files in the data explorer of which a report should be created, mark them by ticking the box. Tap the
button to create the reports. The report will be created with the se­lected report template. Repeat this for each file. For each measurement series there will be a separate PDF created.
NOTE! The report option is only visible if an USB stick is con­nected to the instrument. The USB stick must be either for­matted in FAT or FAT32. NTFS is not supported.
Optionally, also the project file can be exported to the USB stick. Here all files will be included in one file.
4.4.4 Reporting via Equotip Link
Alternatively to create reports the PC-Software Equotip Link can be used. For more details please refer to chapter “11. Equotip Link Software”.

5. Data (Explorer)

Figure 20: Data Explorer Menu

5.1 Measurements

5.1.1 Storing Measurements
If the auto close option is disabled or chosen number of impacts is not reached, the series can be closed and saved manually by tapping the store button .
If the auto close option is enabled, the measurement series will be auto­matically stored as soon as the chosen number of impacts is reached.
The name under which the series will be stored can be edited in the up­per left corner.
© 2016 Proceq SA 29
NOTE! If the file name already exists, the name will be ex-
Table of Contents
tended by a number and increased with every additional file.
Tap on a saved file to open it and return to the Data Explorer list by press­ing the back button.
The stored measurements can be organized in folders, by tapping the Data Explorer button on the new folder option
Enter the name of the new folder and confirm by tapping on the back button in the upper left corner.
The folder in which new measurements are stored can be selected under Settings Measurement Save to Folder.
.
5.1.2 Data Explorer
From the main menu select Data Measurements to review and manage saved measurements data.
Each folder and measurement series is shown as one line in the explorer view.
For each series the probe used, mean value of the series, series name, date & time of the measurement can be seen.
The list can be sorted by tapping on the corresponding header. The small arrow indicates which list is sorted.
Figure 21: Measurement View in Data Explorer
5.1.3 Review of Data
In the detailed view of a measurement series, all information can be seen and the settings are editable.
All the different views can be switched according to the users needs.
For more details about the different views, please refer to chapter “3.2.2 Measurement Views”.
5.1.4 Delete Files
From stored measurement files, single impacts can be eliminated after­wards. To do so, open the measurement series, tap the value to eliminate and tap delete button .
Whole measurement files can be deleted in the Data Explorer. To do so, tap on the box of the appropriate files to select and all selected files can be deleted by tapping the delete button
To erase all data stored on the instrument, in the root folder tap on the box on the left end of the header row then tap the delete button
.
.
30 © 2016 Proceq SA
5.1.5 Copy Files
Table of Contents
To copy measurement series, select the file and click on the icon. Go to the folder where the copy should be created and tap the icon to paste the file. When copying a file all attributes will be duplicated
NOTE! The file cannot be added in the same folder!
5.1.6 Cut & Paste Files
To move existing measurement series from one place to another, set the tick of the corresponding file and tap on the icon. Go to the folder where the file should be moved in and tap the icon to paste the file
5.2 Verifications
From the main menu select “Data”, and then “Verifications” to review and manage saved verification data “6.2 Device Verification”.
The verification data is stored and managed in the same manner as the measurement data. Except no deletion is allowed.
Each folder and measurement series is shown as one line in the explorer view.
Additionally the result for each verification data series, either a “passed” or a “failed” is displayed.

6. Wizards

Figure 22: Wizards Menu
Wizards are a unique feature of Equotip 550. These simple step-by-step instructions are for the majority of users, no matter how experienced they are. Interactive wizards help speed up the workflow and improve the measurement’s reliability.
All settings concerning wizards can be edited in the System User Set­tings. Also see chapter “8.1 Features”.
NOTE! For Equotip Leeb Impact Device U only the “Device Verification” wizard is available.
© 2016 Proceq SA 31

6.1 Measurement Wizard

Table of Contents
This particular wizard helps define the best measurement principle e.g. impact device to fit the application, simply based on sample geometries and surface conditions. To start, some basic information must be provid­ed to define the test piece. When information is evaluated by the device a series of recommendations are displayed in order of their relevance to the application in question.
After the initial process is complete, the device recommends the appro­priate probe, scope of application and preparation information. Settings will then be adopted and the intelligent measurement process begins.
NOTE! Please verify that correct series number, impact direc­tion, material group, scales and limits are defined as well as sub file and folder name.

6.3 Impact Direction Calibration (Leeb only)

Each Leeb impact device requires a calibration in order to automatically compensate for impact direction. This is easily achieved using this wizard.
NOTE! All impact devices are already factory calibrated when shipped, however based on the usage and application it is recommended to recalibrate the impact direction prior to verification process chapter “6.2 Device Verification”. If this process is not completed currently inaccurate data may be obtained.
NOTE! This wizard can also be started from the menu Sys­tem Probes.
6.2 Device Verification
During the verification process, the user will be guided through the en­tire procedure. At the end of the procedure the instrument is considered verified and the data is stored in the device memory. Verification data is also stored as the verification is performed, therefore any discrepancies occurring over time will be noticeable.
32 © 2016 Proceq SA
NOTE! This wizard can also be started from the menu Sys­tem Probes.
NOTE! A Proceq reference test block is required to complete this wizard successfully.

6.4 Conversion Curve Creation

When default conversions are not suitable for the material being tested, it is recommended to create a customized conversion/correlation. This wizard guides the user through the complete process in a simple man­ner and provides all the necessary information on comparative measure­ments.
This creates a brand new conversion curve which is used for any future measurements on the material in question.
6.4.1 Minimizing Conversion Errors
Conversion errors will not normally exceed ±2 HR for Rockwell scales and ±10 % for Brinell and Vickers provided the material group is se­lected correctly. In most cases, the conversion error is significantly lower. If higher accuracy is required or if the alloy under test is not covered by one of the standard conversions, the Equotip 550 provides a variety of methods to define material-specific conversions.
6.4.2 Methods for Setting Up Custom Conversions
Table of Contents
The Equotip 550 provides three techniques to accomplish custom con­versions, each can be used for all of the three different measuring princi­ples (example HLD HRC):
One-point method: The Leeb hardness HLD and the hardness in the desired scale (i.e. HRC) are determined for a reference work piece. A standard conversion function HLD-HRC is then adapted through vertical offset until the measured reference data pair lies on the shifted curve.
Two-point method: Two reference samples are tested, one as soft and one as hard as possible to find two data pairs (i.e. HLD / HRC). A standard conversion function HLD-HRC is then adapted through adding a straight line until both of the measured reference data pairs lie on the tilted curve.
Conversion polynomial: If a custom conversion needs to be applied throughout a wide hardness range, several reference samples shall be tested to find a stable basis for interpolation. Up to 5th order polynomials can be programmed into the Equotip 550 indicating device by defining the polynomial coefficients A
HRC(HLD) = A
+ A1· HLD + A2· HLD2 + A
0
see Equotip Technical Guide under Information Documents or in the download section of the Proceq website.
NOTE! When using a polynomial conversion of a higher or­der, please make sure to have the coefficients with sufficient digits to avoid inaccuracies in calculations.
i
in
· HLD3 + A4· HLD4 + A5· HLD
3
5
6.4.3 Example of a Custom Conversion (Two-Point Method)
The data pairs (640 HLD / 41.5 HRC) and (770 HLD / 54.5 HRC) were measured on two reference samples made from “special steel”.
To measure “special steel” in the future using an adapted HLD-HRC con­version, the original HLD-HRC conversion curve for “1 Steel and cast steel” is tilted using the two data points. In this example, the special conversion is defined as valid for the range 41 to 55 HRC.
Once this curve has been created, it can be selected via material group “Customer defined” – “Special steel”, using the hardness scale “HRC Rockwell C” also see chapter “3.3.1.8 Testing Thin Samples”.
Figure 23: Two-Point Conversion
© 2016 Proceq SA 33
Figure 24: Custom Conversion Menu
Table of Contents
6.4.4 Measuring Reference Samples
Sample surfaces must be prepared very carefully and if possible, sample shall met the specific geometries to avoid coupling method.
The functioning of the Equotip 550 shall be verified against the Leeb test block prior to each measurement series.
The functioning of the static hardness testing machine (HMMRC, HV, HB, HRC etc.) shall be verified against respective test blocks of the corre­sponding measuring scale and range.
To obtain a pair of comparative values, the mean value from at least 10 HL measurements and at least 3 values from the static test shall be cal­culated. These values shall be obtained from proximate positions in a small measuring area depending on application.

6.5 Combined Method

The existing default hardness conversions in Equotip Leeb devices are based on specific sample geometries. A Portable Rockwell probe has almost no restriction with regard to thickness and mass. For samples that don’t meet the Leeb specification, a simple custom correlation based on the Portable Rockwell measurements enables the user to apply a cor­rection factor and create a new hardness conversion. This is one exam­ple where the combined method is used to fit one measuring method with the help of another one, for an application which is not covered by the default setup. But there are several other occasions where this help­ful tool offers great help. This can be achieved following the combined method wizard on the Equotip 550. This wizard allows the combination of the Leeb and Portable Rockwell, the UCI and Portable Rockwell and also the combination of UCI with Leeb method. In each combination the later mentioned method is the reference method.
This wizard guides the user in five simple steps through the whole pro­cess, and finally creates the conversion curve. For other applications it can be used accordingly. For more information please see the “Equotip Application Guide” on the Proceq homepage.

6.6 Mapping Wizard (Coming Soon)

The mapping wizard allows the user to create a 2-dimensional map with readings. This is used to ‘scan’ a whole area. This wizard guides the user in the whole process from defining the area, through the measurements up to the final measuring report.
34 © 2016 Proceq SA

7. Information

Table of Contents
Figure 25: Information Menu

7.1 Documents

All documentation files are stored in this section of the instrument and can be viewed directly when required.
Quick Start Guide: Gives an overview on the instrument including
scope of delivery.
Operating Instructions: This document.
Certificates: Certificates applicable to this product.
Application Booklet: In-depth technical information on the different
measurement principles, its standards, influence of elevated tempera­tures, heavy use instructions and more.
Platform Remote Control Package: Instructions are given on how the instrument can be used with the remote control, i.e. for automa­tion etc.
More documents may be added at a later date.
NOTE! The last viewed document can be quickly accessed by pressing the “Soft Key”. For more information see chapter “2.1 Installation”.

7.2 Upload PDF-Files from an USB-stick

Additional documents in the PDF format can be stored on the instrument. To do so follow the steps below.
• Create the folder “PQ-Import” in the main directory of the USB-stick (not as a subfolder in another folder) and fill it with all the pdf-files to be uploaded to the Equotip Touchscreen
• Go to Information/Documents
• Connect the USB-stick to the USB Device plug on the left side of the Equotip Touchscreen
• Click on
• The uploaded PDF-files appear on the bottom of the document list
and confirm with click on
NOTE! The USB stick must be either formatted in FAT or FAT32. NTFS is not supported.
© 2016 Proceq SA 35

8. System

Table of Contents
Figure 26: System Menu
Verification Notification: Verification of the instrument can be set to
forced, optional or disabled. When set to disabled the user will not be forced to perform an indirect verification. The setting “optional” is just a reminder. If “forced” or “optional” is selected, an entry appears where the verification interval may be chosen.
Custom Fields: The custom entry fields can be edited here. Beside the five default fields which cannot be deleted, additional 20 fields can be added.
8.1.3 Data Management
Use Folder Manager
Activate this option to use automatic folder management as configured in the Folder Manager.
Folder Manager
Here the desired path can be edited. A maximum of four subfolders can be created with selectable information. As soon as one of this information changes, a new folder will be created automatically.

8.1 Features

8.1.1 Device Lock Settings
Lock/Unlock: Select this to lock the instrument and protect it from un­intentional changes.
Password: A password can be set for the lock/unlock function. If this field is left empty, no password is required to unlock the user settings.
8.1.2 General Features
Measurement Wizard: There are three options available on how the measurements wizards are enforced.
36 © 2016 Proceq SA
Use File Manager
Activate this option to use automatic file naming as configured in the File Manager.
File Manager
An automatic name can be configured here consisting of four different information fields.
Long Filename Viewing
Choose here between a full view of the filename, or if only a selected range should be displayed in the measurement screen. This setting influ­ences only the name on the measurement screen, but not in the explorer or reports.
8.1.4 Probe Features
Table of Contents
For each probe type there is an option to protect its settings. Further­more, for each probe type the different features to protect can be se­lected.
Factory reset: Select the options to delete all relevant data from the device.
NOTE! This step cannot be undone, deleted items are per­manently destroyed!

8.2 Probes

Information about the connected probe can be viewed here.
Angle calibration (Leeb only): The angle calibration for this particular probe can be redone. This calibration can be done only for Equotip Leeb Impact Devices.
Verification: Verification measurement series can be started here.
To see information about other used probes, tap on the
Probe Serial Number (only for Leeb U): As the serial number can not be recognized automatically, the user has to enter it manually here.
button.

8.3 Hardware

General settings about the user interface and power options can be ed­ited here.
Sound: The volume of the audio notifications of the instrument can be adjusted respectively switched off.
Display: The user can adjust the brightness of the display backlight.
Power: The time after when the instrument dims the display, or shuts
down can be adjusted, for both battery and AC powered operation.

8.4 Date & Time

Date and time is set in this submenu. The format of these settings and the time zone can also be modified.

8.5 Language

The language setting of the instrument can be selected. Eleven different languages are available. The language of the help file is the same as for the rest of the menu.

8.6 Device information

Tap the info button in the upper right corner to view all the informa­tion related to the device e.g. name, version and serial number can be found in this section as well as battery status. On this information page the IP address (if ethernet is connected and a DHCP server is available) and the instruments MAC address will also be shown.
© 2016 Proceq SA 37

9. Maintenance and Support

Table of Contents

9.1 Maintenance

The instrument should be calibrated annually to ensure consistent, reli­able and accurate measurements. However, the service interval may be based on actual experience and usage. Consult the applicable standards for more guidance.
9.1.1 Regular Device Check
Performance checks, see chapter “4.2 Verification (Performance & Uncer­tainty Check)” of the instrument should be carried out at least once a day or at the latest after 1000 impacts. In the case of infrequent use, carry out the check before the beginning and at the end of a test series. In addition, have the device calibrated by an authorized Proceq Service Center once a year.
NOTE! The unit is working properly when the average is within the target range. Otherwise, please see chapter “10. Troubleshooting”.
9.1.2 Cleaning
Leeb Impact device: Unscrew the support ring. Remove the impact body from the guide tube. Clean the guide tube with the cleaning brush.
Reassemble.
Leeb Indenters: Clean the ball of the Leeb impact bodies and Portable Rockwell diamond indenter with acetone or similar solvent. (Do not use water or water based detergents!)
Portable Rockwell and UCI probes: Clean the probes and the dia­mond indenters with a clean, dry cloth.
Housing: Clean the display and housing with a clean, dry cloth after use. Also clean the connector sockets with a clean, dry brush.
NOTE! Never immerse the device in water. Do not use com­pressed air, abrasives, solvents or lubricants to clean the device.
9.1.3 Storage
Only store the Equotip 550 in the original packaging and in a dry room free of dust.
9.1.4 Re-calibration for Impact Direction (Leeb only)
For Leeb impact devices the compensation function relies on parameters specific to each impact device, which are stored in the device. Validity of the active calibration can be checked through System Probes Angle Calibration and then pressing the “Test” button. For each impact direction, the deviation from the curve shall be less than ±0.2 Leeb (HL).
The parameters may change with time or due to external influences. Re­calibration of the automatic compensation function in Equotip Leeb im­pact devices (except type DL) is recommended particularly when:
• Impact device has been cleaned, or
• Impact device has not been used for a long time, or
• Impact body has been replaced.
A re-calibration is done by consecutive selection of “0° (vertical down)”, “90° (horizontal)” and “180° (vertical up)”.
9.1.5 Updating the Equotip 550 Operating System and
Application
Connect the device to the computer. Updates can be done using Equotip Link as follows:
• Select update symbol
• Select “Express” and confirm with “Next”.
in Equotip Link
38 © 2016 Proceq SA
• Select the device type and confirm with “Next”.
Table of Contents
• In the “Choose Communication Type” dialog box, select the type of communication used between the Equotip and PC, and then click “Next”.
• In the “Device search result and selection” dialog box, make sure the serial number of the device in the drop-down field is the device to be updated, and then click “Next”.
• PqUpgrade will now search the Proceq servers for any available up­dates. To do so, a working internet connection is required.
• Follow the on-screen instructions to finish the update.
NOTE! Although saved data is not deleted during the update process, it is recommended to save the stored data in before updating the firmware.
NOTE! The “Custom” update is recommended for advanced users only.

9.2 Support Concept

Proceq is committed to providing a complete support service for this instrument by means of our global service and support facilities. It is recommended that the user register the product on www.proceq.com to obtain the latest on available updates and other valuable information.

9.3 Standard Warranty and Extended Warranty

The standard warranty covers the electronic portion of the instrument for 24 months from the date of purchase. The mechanical portion of the instrument is covered by warranty for 6 months. An extended warranty for an additional one, two or three years for the electronic portion of the instrument may be purchased up to 90 days from the date of purchase.

9.4 Disposal

Disposal of electric appliances together with household waste is not permissible. In observance of European Directives 2002/96/
EC, 2006/66/EC and 2012/19/EC on waste, electrical and elec­tronic equipment and its implementation, and in accordance with nation­al and local law, electric tools and batteries that have reached the end of their life must be collected separately and returned to an environmentally compatible recycling facility.
© 2016 Proceq SA 39

10. Troubleshooting

Table of Contents

10.1 Incorrect Measurements / Failed Performance Check

10.1.1 Leeb
During the performance check, if the average deviates from the set point value by more than ± 6 HL (for Leeb U ± 12 HLU):
• Check first that the test block is clean, smooth and dry. See chapter “3.3.1.2 Sample Preparations”. Replace the test block if there is insuf­ficient space for additional tests.
• Clean the impact body, paying close attention to the indenter ball at the bottom and to the catch pin at the top of the body. Replace the impact body if necessary.
• Clean the impact device.
• Check the mounting and wear of the support ring. Check for deposits. Clean or replace if necessary.
• The incorrect material group, hardness scale or a wrong setting for the impact direction may have been selected. Refer to chapter “4. Settings”.
• The selected hardness scale is not in the permissible range (no con­version). Select another scale.
• Check if individual values are scattered very widely or are continuously too low.
• The impact is triggered while the device is not held vertically on the surface. This may occur especially when using impact device DL. Try using the plexiglass sleeve DL for better alignment.
• Sample is insufficiently supported. Prepare the sample for the impact e.g. through use of the coupling method, see chapter “3.3.1.6 Testing Light Samples”.
• If the instrument still shows excessive deviations: return the device to an authorized Proceq Service Centre for recalibration / inspection.
NOTE! Do not resurface test blocks or try to refurbish impact bodies. This will impair accuracy and may also deteriorate functionality of the Equotip 550.
10.1.2 Portable Rockwell
During the performance check, if the average deviates from the set point value by more than ±2 HRC:
• Confirm the foot is seated securely on the probe or the probe securely in the clamp, respectively.
• Clean the indenter, paying close attention especially to the front part (diamond) and to the screw thread.
• Check that the test block is clean, smooth and dry. See chapter “3.3.2.2 Sample Preparations”. Replace the test block if there is insuf­ficient space for additional tests.
• Check the mounting and wear of the stand and clamp. Check for de­posits. Clean or replace if necessary.
• An incorrect conversion may have been selected. See chapter”4. Set­tings”.
• The selected conversion scale is not in the permissible range (“no con­version”). Select another hardness scale
• The test is conducted while the device is not held vertically on the sur­face, which will generally give a warning on the guidance dialog. This may occur especially when using tripod. Try using another foot, or take more care to align the probe vertically to the surface.
• Test piece is insufficiently supported. Prepare the test piece for the test e.g. through supporting it with a larger metal piece.
• Make sure the probe does not tilt / move on the surface. See chapter “3.3.1.7 Testing Curved Surfaces”.
• If the instrument still shows excessive deviations: return the device to an authorized Proceq Service Centre for recalibration / inspection.
40 © 2016 Proceq SA
NOTE! Do not resurface test blocks or try to use non-Proceq
Table of Contents
indenters. This will impair accuracy and may also deteriorate functionality of the Portable Rockwell.
10.1.3 UCI
The allowable tolerances for the UCI performance test differ depending on the selected standard. According to the DIN 50159 it should typically not differ more than 5% from the given value. This tolerance is getting wider for harder test blocks. The Equotip 550 considers these limits ac­cording to the standard.
According to the ASTM A1038 the values may not deviate by more than 3%, no matter what hardness is tested.
• Check to have the correct settings selected, i.e. no conversion acti­vated.
• Clean the indenter, paying close attention especially to the front part (diamond).
• Check that the test block is clean, smooth and dry. See chapter “3.3.3.2 Sample Preparations”. Replace the test block if there is insuf­ficient space for additional tests.
• Check the mounting and wear of the special foot. Check for deposits. Clean or replace if necessary.
• If the test was conducted while the device was not held vertically on the surface, the reading can be misleading. This may occur especially when using the standard foot. Try using the special foot, or take more care to align the probe vertically to the surface.
• Test piece is not fulfilling the geometry requirements or is insufficiently supported. Check chapter “3.3.3.4 Test Conditions” for the minimum requirement. Prepare the test piece for the test i.e. by supporting it with a larger metal piece.
• If the instrument still shows excessive deviations: Return the device to an authorized Proceq Service Centre for recalibration / inspection.
NOTE! Do not resurface test blocks.

10.2 No Reading Displayed

• Check the connection of the probe.
• Check if a genuine Equotip impact body (with the engraving “equo”) is inserted in the impact device by unscrewing the support ring.
• Check for tight seating of the support ring on the thread of the impact device.
• Check for tight seating of the support ring on the thread of the impact device.Check whether the impact body is armed and released when conducting the load – trigger procedure. If not, the catch chuck of the impact device may be broken or the impact body is inserted upside­down. Insert the impact body correctly or replace the impact device with a basic Equotip Leeb impact device.
© 2016 Proceq SA 41

10.3 Battery

Table of Contents
If the indicating device does not switch on, recharge the battery using the power supply, see chapter “2.1 Installation”.
The battery can be replaced with another Equotip Lithium-Ion battery.
NOTE! If the operation time of the battery is shortening no­ticeably, order a new battery. The battery lifetime has expired when the LED does not go off even though the battery has been charged for several days.
Danger: Only use the power supply (12 V, 5 A) to charge the Equotip 550.

10.4 Touchscreen Calibration

In very seldom cases, or when a protective screen foil is used, it can be necessary to re-calibrate the Equotip 550.
To do so, press and hold the middle hardware button (Fullscreen) for 10s. During the calibration process, don’t touch the display unit, as it can bias the calibration.

11. Equotip Link Software

11.1 Starting Equotip Link

Locate the file “Equotip Link Setup.exe” on your computer or on the DVD and click on it. Follow the instructions on the screen. Make sure that the “Launch USB Driver install” tick is selected.
Double click on the Equotip Link Icon on your desktop or start the program via the start menu.

11.2 Application Settings

The menu item “File – Application settings” allows the user to select the language and the date and time format to be used.
42 © 2016 Proceq SA

11.3 Connecting to an Equotip 550 Touchscreen Unit

Table of Contents
Connect the Equotip 550 Touchscreen Unit to an USB port or connect it to the Ethernet (DHCP server requried)., then select to download data from the Equotip 550 Touchscreen Unit.
The following window will be displayed: Select the appropriate communi­cation type. If connecting through Ethernet, please enter the instruments IP address in the appropriate field. Click on “Next >”.
When an Equotip 550 has been found its details will be displayed on screen. Click on the “Finish” button to establish the connection
Measurement files and folders stored on the Equotip Touchscreen will be displayed as shown on the left.
Click on the measurement file to be transferred. If multiple files have to be transferred, select them while holding the “shift”or “ctrl” button, or click on “Select All”.

11.4 Connection to Portable Rockwell Probe

• Connect the Portable Rockwell probe to a PC using the provided probe cable.
• Start the Equotip Link software, and click on the Portable Rockwell icon
to detect the Portable Rockwell probe. Click the “New” button
at the bottom of the screen.
• Select the hardness scale to be displayed (Hardness scales).
• Select the number of readings “n” per measurement series.
11.4.1 Viewing Data
The transferred measurements from your Equotip 550 will be displayed on the screen:
Click on the double arrow icon in the first column to see more details.

11.5 Adjusting the Settings

“All settings, such as material group, scale, impact direction and both limits can be changed afterwards in Equotip Link.
If the settings of multiple measurement series have to changed, select each of them while holding the “shift” or “ctrl” button.
© 2016 Proceq SA 43
11.5.1 Adjusting Date and Time
Table of Contents
Right click in the “Date & Time” column.
The time will be adjusted for the selected series only.

11.6 Exporting Data

Equotip Link allows you to export selected objects or the entire project. To use the data for further analysis in a third party program, such as e.g. Microsoft Excel, it can be exported as a comma separated file (CSV). To use it directly in a report, the series data can be exported as a graphic. As a third option, selected series data can be printed out directly on a printer.
Click on the “Export as CSV file(s)” icon. The data is then exported as a Microsoft Office Excel comma separated file. The export options may be chosen in the following window:
Click on the “Export as graphic” icon to open the follow­ing window which allows the various export options to be chosen.
Click on the printer icon to directly print a report of the selected measurement series.
44 © 2016 Proceq SA
11.7 Exporting and Importing of Setting Profiles
Table of Contents
To transfer all the selected settings from one instrument to another, or to back up, click on “Equotip – Download device application settings”. The actual settings will be stored in the specified folder on the PC as an archive.
To re-install stored settings, click on “Equotip – Upload device applica­tion settings”.

11.8 Exporting and Importing of Conversion Curves

Custom conversion curves created on the instrument can be download­ed to the PC by clicking on “Equotip - Download customer conversion”. All available customer conversions from the instrument will be stored on the PC in the folder ...\Proceq\EquotipLink\Conversions.
To upload a conversion curve from your PC, choose “Equotip - Upload customer conversions”. This is also possible for existing Equotip3 con­version curves.
12. Technical Specifications

12.1 Instrument

Display 7” color display 800x480 pixels
Memory Internal 8 GB Flash memory
(up to 1’000’000 measurements)
Regional settings Metric and Imperial units, multi-language
and timezone supported
Battery Lithium Polymer, 3.6 V, 14.0 Ah
Battery lifetime > 8h (in standard operating mode)
Power input 12 V +/-25 % / 1.5 A
Weight (of display device) About 1525 g (incl. Battery)
Dimensions 250 x 162 x 62 mm
Max. altitude 2’500 m above sea level
Humidity < 95 % RH, non condensing
Operating temperature 0 to 30°C (32 to 86°F)
(Charging, running instrument) 0 to 40°C (32 to 104°F) (Charging, instrument is off)
-10 to 50°C (14 to 122°F) (Non-charging)
Environment Suitable for indoor & outdoor use
IP classification IP 54
Pollution degree 2
Installation category 2
© 2016 Proceq SA 45
NOTE! Charging equipment is for indoor use only (no IP clas­sification).

12.2 Power Supply

Table of Contents
Model HK-AH-120A500-DH
Input 100-240 V / 1.6 A / 50/60 Hz
Output 12 V DC / 5 A
Max. altitude 2500 m above sea level
Humidity < 95%
Operating temperature 0 - 40°C (32 to 104°F)
Environment Indoor use only
Pollution degree 2
Installation category 2

12.3 Equotip Leeb Impact Devices

Measuring range 1-999 HL
Measuring accuracy ± 4 HL (0.5 % at 800 HL)
± 6 HLU (for Leeb U)
Resolution 1 HL; 1 HV; 1 HB; 0.1 HRA; 0.1 HRB; 0.1 HRC;
0.1 HS; 1 MPa (N/mm2)
Impact direction Automatic compensation (excl. DL/U probe)
Impact energy • 11.5 Nmm for D, DC, E, S probes
• 11.1 Nmm for DL probe
• 3.0 Nmm for C probe
• 90.0 Nmm for G probe
• 200.0 Nmm for U probe
Mass of impact body • 5.45 g (0.2 ounces) for D, DC, E, S probes
• 7.25 g (0.26 ounces) for DL probe
• 3.10 g (0.11 ounces) for C probe
• 20.0 g (0.71 ounces) for G probe
• 26.0 g (0.92 ounces) for U probe
Ball indenter • Tungsten carbide, 3.0 mm (0.12“) diameter
for C, D, DC probes
• Tungsten carbide, 2.78 mm (0.11“) diameter for DL probe
• Tungsten carbide, 5.0 mm (0.2“) diameter for G probe
• Ceramics, 3.0 mm (0.12”) diameter for S probe
• Polycrystalline diamond, 3.0 mm (0.12”) diameter for E probe
• Hardened Steel, 50.0 mm (1.97”) diameter for U probe
Operating temperature -10 to 50˚C (14 to 122°F)
46 © 2016 Proceq SA

12.4 Equotip Portable Rockwell Probe

Table of Contents
Dimensions 112 x ø 40 mm (4.4 x ø 1.57 inches
without foot)
Weight 260 g (9.17 oz)
Power supply Via USB (5 V, max. 100 mA)
Measuring range 0-100 μm; 19-70 HRC; 34-1080 HV
Measuring accuracy 1.5 HRC according to DIN 50157
Resolution 0.1 μm; 0.1 HRC; 1 HV
Test direction Any direction (no correction required)
Test loads 10 N / 50 N (probe 50 N)
Diamond indenter Angle 100.0° ± 0.5°
Operating temperature 0 to 50°C (32 to 122 °F)
Humidity Non-condensing, 90% max.

12.5 Equotip UCI Probe

Dimensions 155 x ø 40 mm (6.1 x ø 1.57 inches) with-
out foot
Weight 270 g (9.52 oz)
Power Supply via Proceq Interface
Measuring Range 20 – 2000 HV
Measuring accuracy ± 2 % (150 – 950 HV)
Resolution 1 HV; 0.1 HRC
Test direction Any direction (no correction required)
Trigger loads Selectable: HV1 (~10N), HV2 (~20N),
HV3 (~30N), HV4 (~40N), HV5 (~50N)
Diamond indenter Vickers diamond according to ISO 6507-2
Operating temperature 0 to 50°C (32 to 122°F)
Humidity Non-condensing, 90% max.

13. Standards & Guidelines

• ISO EN 16859
• ISO 18265
• DIN 50156 / 50157 / 50159
• ASTM A956 / E140 / A370 / A1038
• DGZfP Guideline MC 1
• VDI / VDE Guideline 2616 Paper 1
• Nordtest Technical Report Series 424, Reports 99.12 / 99.13 / 99.36
• ASME CRTD-91
• GB/T 17394
• JB/T 9378
• JJG 747
• JIS B7731
© 2016 Proceq SA 47

14. Ordering Information

Table of Contents

14.1 Units

Part No. Description
356 10 001 Equotip 550 consisting of Equotip Touchscreen incl.
Battery, Power Supply, USB-Cable, Surface Roughness Comparator Plate, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 002 Equotip 550 Leeb D consisting of Equotip Touchscreen
incl. Battery, Equotip Basic Leeb Impact Device D, Im­pact Body D, Support Rings (D6,D6a), Cleaning Brush, Impact Device Cable, Test Block ~775 HLD / ~56 HRC, Coupling Paste, Power Supply, USB-Cable, Surface Roughness Comparator Plate, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 003 Equotip 550 Leeb G consisting of Equotip Touchscreen
incl. Battery, Equotip Basic Leeb Impact Device G, Im­pact Body G, Support Rings (G6,G6a), Cleaning Brush, Impact Device Cable, Test Block ~570 HLG /~340 HB, Coupling Paste, Power Supply, USB-Cable, Surface Roughness Comparator Plate, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 004 Equotip 550 Portable Rockwell consisting of Equotip
Touchscreen incl. Battery, Equotip Portable Rockwell Probe 50 N, Protective Rubber Sleeve, Probe Cable, Test Block ~62HRC, Power Supply, USB-Cable, Sur­face Roughness Comparator Plate, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 005 Equotip 550 UCI consisting of Equotip Touchscreen
incl. Battery, Equotip UCI Probe HV1-HV5, UCI Probe Cable, UCI Test Block ~850 HV, Power Supply, USB­Cable, Surface Roughness Comparator Plate, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 006 Equotip 550 Leeb U (for paper, film and foils) consisting
of Equotip Touchscreen incl. Battery, Equotip Leeb Impact Device U, Cleaning Brush, Probe Cable, Power Supply, USB-Cable, DVD with Software, Documentation, Carrying Strap and Carrying Case
356 10 020 Equotip 550 Portable Rockwell & UCI Kit consisting
of Equotip 550 UCI (356 10 005) and Equotip Portable Rockwell Probe 50 N (356 00 600)
356 10 021 Equotip 550 Portable Rockwell & Leeb D Kit consist-
ing of Equotip 550 Leeb D (356 10 002) and Equotip Portable Rockwell Probe 50 N (356 00 600)
356 10 022 Equotip 550 Leeb D & UCI Kit consisting of Equotip
550 Leeb D (356 10 002), Equotip UCI Probe HV1-HV5 (356 00 700) and Equotip UCI Test Block ~850 HV, ISO 6507-3 HV5 Calibration (357 54 100)

14.2 Impact Devices & Probes

Part No.
356 00 500 Equotip Leeb Impact Device C
356 00 100 Equotip Leeb Impact Device D
356 00 110 Equotip Leeb Impact Device DC
Description
Impact Device incl. Support Ring, Impact Body, Cable
48 © 2016 Proceq SA
356 00 120 Equotip Leeb Impact Device DL
Table of Contents
356 00 400 Equotip Leeb Impact Device E
356 00 300 Equotip Leeb Impact Device G
356 00 200 Equotip Leeb Impact Device S
360 04 600 Equotip Leeb Impact Device U
Impact Device only
353 00 501 Equotip Basic Leeb Impact Device C
353 00 101 Equotip Basic Leeb Impact Device D
353 00 111 Equotip Basic Leeb Impact Device DC
353 00 121 Equotip Basic Leeb Impact Device DL
353 00 401 Equotip Basic Leeb Impact Device E
353 00 301 Equotip Basic Leeb Impact Device G
353 00 201 Equotip Basic Leeb Impact Device S
360 04 032 Equotip Basic Leeb Impact Device U
356 00 600 Equotip Portable Rockwell Probe 50N
(for Equotip 550 or PC)
356 00 700 Equotip UCI Probe HV1-HV5

14.3 Parts and Accessories

Part No. Description
327 01 043 Carrying Strap complete
327 01 033 Battery complete
351 90 018 USB-Cable 1.8 m (6 ft)
327 01 061 Power Supply
711 10 013 Power Supply Cable USA 0.5 m (1.7 ft)
711 10 014 Power Supply Cable UK 0.5 m (1.7 ft)
711 10 015 Power Supply Cable EU 0.5 m (1.7 ft)
327 01 053 Quick Charger
356 00 081 Equotip Surface Roughness Plate
350 01 015 Equotip Coupling Paste
356 00 082 Display Antiglare Protection Film for Touchscreen Unit
356 00 080 Equotip Leeb Impact Device Cable 1.5 m (5 ft)
353 00 086 Equotip Leeb Impact Device Extension Cable 5 m (15 ft)
356 00 083 Equotip Leeb Impact Device U Cable 1.5 m (5 ft)
350 01 004 Equotip Impact Body D/DC
350 71 311 Equotip Impact Body DL
350 71 413 Equotip Impact Body S
350 08 002 Equotip Impact Body G
350 07 002 Equotip Impact Body E
350 05 003 Equotip Impact Body C
360 04 504 Equotip Impact Body U
350 01 009 Equotip Support Ring D6
350 01 010 Equotip Support Ring D6a
350 08 004 Equotip Support Ring G6
© 2016 Proceq SA 49
350 08 005 Equotip Support Ring G6a
Table of Contents
350 71 314 Equotip Support Ring DL
360 04 531 Equotip Support Ring U
353 03 000
350 01 008
350 08 006 Equotip Leeb Impact Device Cleaning Brush G
360 04 502 Equotip Leeb Impact Device Cleaning Brush U
350 01 007 Equotip Leeb Impact Device DC Loading Stick
350 71 316 Equotip Leeb Impact Device DL Plexiglas Sleeve
360 04 530 Assembly Gauge for Equotip Leeb Impact Device U
354 01 139 Equotip Portable Rockwell Probe Cable 2 m (6 ft)
354 01 200 Equotip Portable Rockwell Measuring Clamp
354 01 130 Equotip Portable Rockwell Tripod
354 01 250 Equotip Portable Rockwell Special Foot RZ 18 - 70
354 01 253 Equotip Portable Rockwell Special Foot RZ 70 -
354 01 137 Equotip Portable Rockwell Protective Rubber Sleeve
354 01 243
354 01 229
Set of Support Rings (12pcs) for Equotip Leeb Impact Device D/DC/C/E/S
Equotip Leeb Impact Device Cleaning Brush D/DC/C/E/S
Equotip Portable Rockwell support Z2 for measuring clamp
Equotip Portable Rockwell Support Z4+28 for measuring clamp (for tubes and pipes over Ø 28 mm)
354 01 228
356 00 720 Equotip UCI Special Foot
Equotip Portable Rockwell support Z4 for measuring clamp (for tubes and pipes up to Ø 28 mm)
14.4 Test Blocks
Part No. Description
357 11 500 Equotip Test Block C, ~565 HLC / <220 HB,
Proceq Factory Calibration
357 12 500 Equotip Test Block C, ~665 HLC / ~325 HB,
Proceq Factory Calibration
357 13 500 Equotip Test Block C, ~835 HLC / ~56 HRC,
Proceq Factory Calibration
357 11 100 Equotip Test Block D/DC, <500 HLD / <220 HB,
Proceq Factory Calibration
357 12 100 Equotip Test Block D/DC, ~600 HLD / ~325 HB,
Proceq Factory Calibration
357 13 100 Equotip Test Block D/DC, ~775 HLD / ~56 HRC,
Proceq Factory Calibration
357 13 105 Equotip Test Block D/DC, ~775 HLD, one side,
Proceq Factory Calibration
357 11 120 Equotip Test Block DL, <710 HLDL / <220 HB,
Proceq Factory Calibration
357 12 120 Equotip Test Block DL, ~780 HLDL /~325 HB,
Proceq Factory Calibration
357 13 120 Equotip Test Block DL, ~890 HLDL / ~56 HRC,
Proceq Factory Calibration
50 © 2016 Proceq SA
357 13 400 Equotip Test Block E, ~740 HLE / ~56 HRC,
Table of Contents
Proceq Factory Calibration
357 14 400 Equotip Test Block E, ~810 HLE / ~63 HRC,
Proceq Factory Calibration
357 31 300 Equotip Test Block G, <450 HLG / <200 HB,
Proceq Factory Calibration
357 32 300 Equotip Test Block G, ~570 HLG / ~340 HB,
Proceq Factory Calibration
357 13 200 Equotip Test Block S, ~815 HLS / ~56 HRC,
Proceq Factory Calibration
357 14 200 Equotip Test Block S, ~875 HLS / ~63 HRC,
Proceq Factory Calibration
360 04 503 Equotip Test Block U, ~560 HLU,
Proceq Factory Calibration
357 41 100 Equotip Portable Rockwell Test Block ~20 HRC,
ISO 6508-3 HRC Calibration
357 42 100 Equotip Portable Rockwell Test Block ~45 HRC,
ISO 6508-3 HRC Calibration
357 44 100 Equotip Portable Rockwell Test Block ~62 HRC,
ISO 6508-3 HRC Calibration
357 51 100 Equotip UCI Test Block ~300 HV, ISO 6507-3 HV5
Calibration
357 52 100 Equotip UCI Test Block ~550 HV, ISO 6507-3 HV5
Calibration
357 54 100 Equotip UCI Test Block ~850 HV, ISO 6507-3 HV5
Calibration
Test Blocks Calibrations
Part No. Description
357 10 109
357 10 129 Equotip Leeb Test Block Additional Calibration HLDL
357 10 209 Equotip Leeb Test Block Additional Calibration HLS
357 10 409 Equotip Leeb Test Block Additional Calibration HLE
357 10 509 Equotip Leeb Test Block Additional Calibration HLC
357 30 309 Equotip Leeb Test Block Additional Calibration HLG
357 90 909
357 90 919
357 90 929
357 90 939
357 90 918
357 90 928
357 90 940
357 90 941
357 90 942
Equotip Leeb Test Block Additional Calibration HLD/ HLDC
Equotip Leeb Test Block Additional Calibration HL, DIN 50156-3
Equotip Leeb Test Block Additional Calibration HB, ISO 6506-3
Equotip Leeb Test Block Additional Calibration HV, ISO 6507-3
Equotip Leeb Test Block Additional Calibration HR, ISO 6508-3
Equotip Portable Rockwell Test Block Additional Calibration HB, ISO 6506-3
Equotip Portable Rockwell Test Block Additional Calibration HV, ISO 6507-3
Equotip UCI Test Block Additional Calibration HB, ISO 6506-3
Equotip UCI Test Block Additional Calibration HR, ISO 6508-3
Equotip UCI Test Block Additional Calibration HV1, ISO 6507-3
© 2016 Proceq SA 51
Subject to change. Copyright © 2017 by Proceq SA, Schwerzenbach. All rights reserved. 82035601E ver 05 2017
Loading...