This manual contains important information on the safety, use and maintenance of the Equotip 550. Read through the manual carefully before the
first use of the instrument.
1.2 Liability
Our “General Terms and Conditions of Sales and Delivery” apply in all
cases. Warranty and liability claims arising from personal injury and damage to property cannot be upheld if they are due to one or more of the
following causes:
• Failure to use the instrument in accordance with its designated use as
described in this manual.
• Incorrect performance check for operation and maintenance of the instrument and its components.
• Failure to adhere to the sections of the manual dealing with the performance check, operation and maintenance of the instrument and its
components.
• Unauthorised modifications to the instrument and its components.
• Serious damage resulting from the effects of foreign bodies, accidents,
vandalism and force majeure.
All information contained in this documentation is presented in good faith
and believed to be correct. Proceq SA makes no warranties and excludes
all liability as to the completeness and/or accuracy of the information.
1.3 Safety Instructions
The equipment is not allowed to be operated by children or anyone under
the influence of alcohol, drugs or pharmaceutical preparations. Anyone
who is not familiar with this manual must be supervised when using the
equipment.
• Carry out the stipulated maintenance properly and at the correct time.
• Following completion of the maintenance tasks, perform a functional
check.
1.4 Correct Usage
The instrument is only to be used for its designated purpose as described
herein.
• Replace faulty components only with original replacement parts from
Proceq.
• Accessories should only be installed or connected to the instrument
if they are expressly authorized by Proceq. If other accessories are
installed or connected to the instrument then Proceq will accept no
liability and the product guarantee is forfeited.
1.5 Optimizing Performance of the Battery System
To increase the performance of the battery, it is recommended to first
completely discharge and then completely charge it.
The Equotip 550 is typically used for testing the hardness of metallic surfaces. The user has a choice to select either Leeb rebound, the Portable
Rockwell or UCI principle, see chapter “3.1 Performing Measurements”.
In combination with the Equotip Leeb Impact Device U the instrument is
used to test the roll hardness of paper, film or foil rolls.
2.1 Installation
To install the Battery into the Equotip 550 Touchscreen Unit, lift the stand
as shown, insert the battery and fasten it in place with the screw.
Buttons
On the upper right of the unit there are three buttons:
Power On/Off – Press to power on or to return to the home
screen. Press and hold to power off.
Soft Key – Switches in and out of full screen view or toggles
between the actual screen and the last viewed pdf document (eg.
Operating Instructions).
Back Button – Returns to previous screen.
Energy Saving
Energy saving may be programmed as desired under System/Power settings, see chapter “8.3 Hardware”.
Connections
Figure 1: Insert Battery
There are three status LEDs on the right side of the display. The middle
light is the power indicator which is red when charging and turns to green
when battery is fully charged. The lower LED is used for application specific notification.
• A complete charge requires < 9 h (Instrument not operating)
• Charging time is much longer if the instrument is in use.
• An optional Quick Charger (Part No. 327 01 053) can be used to
NOTE! Only use the battery charger provided for charging.
charge a spare battery or to charge the battery outside of the instrument. In this case it takes < 5.5 h for a complete charge.
1 2
Snap-in connectors
Figure 2: Connections
For Leeb Impact Devices
use the Snap-in connector 1.
For UCI Probe
use the Snap-in connectors 1 or 2.
For the Portable RockwellProbe
use the USB Host connector.
USB Host
USB Device
Ethernet
Power Supply
USB Host:
Additionally connect a mouse,
keyboard or USB stick.
USB Device:
Connect to PC.
Ethernet:
Connection to network.
Power Supply:
Connect the power supply
through this connection.
2.2 Main Menu
Table of Contents
On start up the main menu is displayed. All functions may be accessed directly via the Touchscreen. Return to the previous menu by pressing the
back button or the return icon (arrow) at the top left corner of the Touchscreen.
Settings: For application
specific settings, see chapter
“4. Settings”.
Measurement:
Measurement display screen,
see chapter “3. Measurement”.
Wizards: Task
related workflows, see
chapter “6. Wizards”.
Data (Explorer): File manager for
reviewing measurement data on the
instrument, see chapter “5. Data
(Explorer)”.
System: System settings,
e. g. language, display
options etc, see chapter
“8. System”.
Information: Operating
instructions and other
reference documents, see chapter
“7. Information”.
3. Measurement
Table of Contents
3.1 Performing Measurements
3.1.1 Leeb Testing Procedure (except Leeb U)
Select automatic compensation for impact direction “Automatic”, see
chapter “3.2.1 Controls”. If “Automatic” is not allowed, set the impact
direction ( ). The Equotip Leeb Impact Devices DL doesn’t support automatic mode. The impact direction must be selected manually.
Select the appropriate material group, hardness scales and number of
impacts per measurement series. For more information see chapter “4.
Settings”. Conduct impacts by cycling through “load, position and trigger” mechanism:
1. Load the impact device – while not in
contact with the test piece – by holding it firmly with one hand and sliding
the loading tube with the other hand
until contact is grabbed by the clutch.
2. Position the support ring on the test
piece. Take particular care to fully
position the support ring on the test
piece, but not coinciding with a previous test indentation.
3. To trigger an impact, press the
trigger button to release the impact
body. To perform another impact,
repeat this cycle.
Figure 4: Leeb Testing Procedure
After the last of the impacts is performed, the hardness average and further statistics of the measurement series are displayed.
NOTE! Make sure the loading tube is allowed to slowly return
back to the starting position. Do take care so the loading
tube does not spring back uncontrolled, which may result in
permanent device damage.
NOTE! If possible, follow the standard practice of Leeb rebound hardness testing as described in the standards DIN
50156-1 (metallic materials), ASTM A956 (steel, cast steel and
cast iron only), or other applicable standards. If these are not
available, the user is recommended to average a minimum of
n = 3 impacts at an indentation distance of 3 to 5 mm (0.12
to 0.20”) for each location of the sample that shall be tested.
NOTE! Do not carry out an impact in an area that has already
been deformed by another impact. Also, do not load the device
when it is already positioned in the new test location, since the
material under the device may be affected through prior stress,
and the catch chuck of the device may get damaged.
1. Place the probe on the sample to
test. For flat surfaces the standard
foot is most suitable. For cylindrical
objects it is recommended to use
one of the special feet. For locations
difficult to access a tripod foot can
be used. See chapter “14. Ordering
Information” for more details.
2. Press the probe slowly but firmly
against the surface to perform the
measurement. Suppress vibrations
as much as possible, and follow the
instructions on the screen.
3. Release the probe when the instrument says so. Again, this movement
has to be done in a controlled manner. If the probe is released too fast,
a warning will be shown and the
measurement should be repeated.
Figure 5: Portable Rockwell Testing Procedure
3.1.3 UCI Testing Procedure
1. Place the probe on the sample to
test. The probe must be perpendicular to the surface (± 5°). The
special foot can be used to increase
repeatability and reduce distortion of the results, see chapter “14.
Ordering Information”.
2. Press the probe slowly but firmly
against the surface until the selected measuring force is reached.
The instrument will indicate when to
release the probe with an on-screen
prompt and audible sound.
3. Release the probe from the material. It is important to remove the
probe completely from the test
object. Otherwise the results can be
biased.
Figure 6: UCI Testing Procedure
NOTE! A warning will be shown if the user applied too much
load when pressing the probe against the surface. Please avoid
frequent overloading, as this could seriously damage the probe.
The Equotip 550 Leeb U enables the user to quickly and precisely diagnose roll imperfections, hardness inconsistencies and uneven winding,
thereby preventing problems for printing and converting operations.
With Equotip Impact Device Leeb U automatic impact direction mode is
not supported and the user must select the appropriate impact direction
manually (90° down, 45° down, 0°).
As for roll hardness testing no conversion curves are used, no material
group has to be selected.
Conduct impacts by cycling through “position and trigger”.
1. Position the probe on the roll to be
tested. Make sure to fully position the support ring on the roll to
ensure an impact perpendicular to
the test surface.
2. While holding the impact device
firmly with two hands, slide the
loading tube smoothly down to
load and trigger an impact.
Move the impact device to the next
point on the roll and repeat.
Figure 7: Leeb U Testing Procedure
NOTE! Some features mentioned in this Operating Instructions are specifically addressing metal hardness testing applications and therefore not available for Equotip Leeb U.
File Name: Enter the file name
and tap return. Saved measurements will be stored automatically. If filename management
is activated, this function is
locked.
Impact Direction: To
manually set the impact
direction if required
(Leeb only, by default
this is automatic).
Material: Select the
material group to use
for conversions (not
available for Leeb U).
Measurement Scale: Select the hardness
scale (primary and secondary) to be displayed
(not available for Leeb U).
Measurement Mode:
Switch between
measurement and conversion.
Wizards: Direct
access to wizards.
Redo: Restart the measurement series
or a single measurement.
Time and battery status
Settings: direct shortcut to
settings menu.
(only applicable to the
actual measurement series)
Save: Store measurement data.
Delete: Delete the last measurement.
3.2.2 Measurement Views
HL
ID
+
Table of Contents
Equotip 550 is fully customizable as a device and can display three different measurement views simultaneously. Each view can be switched to
meet the user’s requirements by simply clicking on the icon related to the
particular display at the top right corner of each screen.
Signal View: Display the probe signal from the last active
measurement. This may be useful for advanced evaluations.
Statistic View: View statistics for the active measurement series. Number of impacts (n), Average (x), Standard deviation
(σ), Minimum/Maximum () and Range () are displayed in the
primary scale.
Table View: Display the measurements for the active series in
a table format.
Conversion View: Display the actual value on the conversion
curve.
Bar View: Display the measurements of the series as a histogram.
Profile View: Display the measurement results as a profile.
Info: Display the measurement settings e.g. series length,
probe type, material group etc.
User’s View: The user can choose between probe angle, mini-
mum, maximum, range and probe type for the field contents. To
change, tap on each individual box.
Single Record View: Display the last or selected measurement
result in both the primary and secondary hardness scales.
Equotip 550 family of instruments is capable of accepting three diffrent
test methods using a single indicating unit.
3.3.1 Equotip Leeb
3.3.1.1 Test Principle
Release button
Loading tube
Impact Spring
Catch chuck
Connection cable – 4 pole
Sensor of impact device with
ID ROM
Support ring
Figure 10: Schematic View of a Leeb Impact Device
Guide tube
Impact Body
During measurement with Equotip 550 impact devices (D, DL, DC, C,
G, S, and E); an impact body with a ball indenter is launched by spring
energy against the sample to be measured, and then rebounds. Before
and after the impact, a permanent magnet inside the impact body passes
through a coil in which a voltage signal is induced by the forwards and
backwards movement. This induction signal behaves proportionally to
the velocities. The ratio of the rebound velocity vr to the impact velocity vi
multiplied by 1000 yields the hardness value HL (Leeb hardness). HL is a
direct measure of the hardness. The third resp. fourth letter of the HL unit
refers to the impact device HLD D impact device.
v
r
HL =
·1000
v
i
Equotip Leeb U
Although the Equotip Leeb Impact Device U is constructed differently to
simplify the measurement process, the underlying principle is the same.
Figure 11: Equotip Leeb U Impact Device
Existing Parotester impact devices type U are fully supported by the
Equotip 550. Typ P and PG impact devices can be still used, but the unit
is shown as HLU although it would be actually LP resp. LPG.
NOTE! HLU values can be directly compared to LU on existing Parotester instruments.
Keep the sample free of vibrations during the test. Light and thin parts
must be specially fastened, see chapter “3.3.1.6 Testing Light Samples”.
Ensure that the surface of the work piece is clean, smooth and dry. If required, use appropriate cleaning agents for cleaning, such as acetone or
isopropanol. Do not use water or any other detergent fluids.
NOTE! Please use the surface roughness comparator plate
provided to estimate the average roughness of the test piece
prior to testing.
Figure 12: Surface Roughness Comparator Plate
3.3.1.3 Standards
Brief descriptions of referenced standards:
DIN 50156
ASTM A956
ASTM A370
ASTM E140
ISO 18265
ISO 16859
Leeb hardness testing of metallic materials
Standard test method for Leeb hardness testing of steel
products
Test methods and definitions for mechanical testing of
steel products
Standard hardness conversion tables for metals rela-
tionship among Brinell, Vickers, Rockwell, Superficial,
Knoop, Scleroscope and Leeb hardness
Metallic materials – Conversion of hardness values
Leeb hardness testing of metallic materials
To ensure proper hardness readings, the following conditions must be fulfilled. If one or more conditions are not met, the measurement result may be
significantly false.
Impact device typeD/DC/DL/S/EGC
Surface
preparation
Roughness grade class ISO 1302
Max. roughness depth R
Average roughness R
Minimum sample
mass
Of compact shape (kg / lbs)
On solid support (kg / lbs)
Coupled on plate (kg / lbs)
For optimized testing of diverse metallic materials and sample geometries, a range of impact devices are available as per “Table 1: Leeb Test Piece
Requirements”.
Type G: Increased impact energy. For solid and inhomogenous
components, e.g. heavy castings and forgings.
Impact energy: 90 Nmm
Type DC: Short device. For use in very
confined spaces, e.g. in holes, cylinders or
for internal measurements on assembled
machines. Impact energy: 11 Nmm
Type D: Universal unit. For the majority of
your hardness testing requirements.
Impact energy: 11 Nmm
Type S: Si3N4 ball indenter. For testing
especially in the very high hardness range
(in excess of 50 HRC / 650 HV): Tool steels
with high carbide content inclusions.
Type E: Diamond ball indenter.
For testing in the very high hardness
range (in excess of 50 HRC / 650 HV):
Tool steels with high carbide content
inclusions. More durable than type S.
For measurements in confined spaces, at the
base of grooves or on recessed surfaces.
Impact energy: 11 Nmm
3.3.1.6 Testing Light Samples
Table of Contents
If the samples are lighter than specified in chapter “3.3.1.4 Test Conditions” or sample sections have unfavorable mass distribution they can
vibrate as the impact body hits the test point. This results in unwanted
energy absorption. Such samples shall be supported by solid worktops.
If the mass falls below the specific requirements but still exceeds the
coupling amount then coupling it to a larger mass can help prevent vibrations.
The following requirements must be met for coupling:
• The contact surface of the sample and the surface of the solid support
must be level, flat and ground smooth.
• The sample must exceed the minimum sample thickness for coupling.
Follow the coupling procedure.
• Apply a thin layer of coupling paste to the contact surface of the sample.
• Press the sample firmly against the support.
• Push the sample in a circular motion and carry out the impact as usual,
perpendicular to the coupled surface.
NOTE! Clamping may strain the sample, which can affect the
hardness readings.
3.3.1.7 Testing Curved Surfaces
The instrument works properly only when the ball indenter at the front of
the impact body is precisely at the end of the tube at the time of the impact. When concave or convex surfaces are tested, the ball indenter either
does not entirely leave the test tube or comes out too far. In such cases,
replace the standard support ring by a specially suited ring, see chapter
“14. Ordering Information” or contact your local Proceq representative.