Proceq Equotip 550 Operating Instructions Manual

Operating Instructions
60 Years of Innovation60 Years of Innovation
Swiss Precision since 1954

Table of Contents

1. Safety and Liability ....................................................4
1.1 General Information ............................................................. 4
1.2 Liability .................................................................................. 4
1.3 Safety Instructions ............................................................... 4
1.4 Correct Usage ...................................................................... 4
1.5 Optimizing Performance of the Battery System ............... 4
2. Getting Started ...........................................................5
2.1 Installation ............................................................................ 5
2.2 Main Menu ............................................................................ 6
3. Measurement .............................................................7
3.1 Performing Measurements ................................................. 7
3.2 Measurements Screen ...................................................... 10
3.3 Measuring Methods .......................................................... 12
3.4 Instrument Verification / Daily Performance Check ....... 24
4. Settings ....................................................................25
4.1 Measurements ................................................................... 25
4.2 Verification (Performance & Uncertainty Check) ............ 27
4.3 Conversions (Hardness Conversions) ............................. 28
4.4 Reporting ............................................................................ 28
5. Data (Explorer) .........................................................29
5.1 Measurements ................................................................... 29
5.2 Verifications ........................................................................ 31
6. Wizards .....................................................................31
6.1 Measurement Wizard ........................................................ 32
6.2 Device Verification ............................................................. 32
6.3 Impact Direction Calibration (Leeb only) ......................... 32
6.4 Conversion Curve Creation............................................... 32
6.5 Combined Method ............................................................. 34
6.6 Mapping Wizard (Coming Soon) ...................................... 34
7. Information ...............................................................35
7.1 Documents ......................................................................... 35
7.2 Upload PDF-Files from an USB-stick .............................. 35
8. System ......................................................................36
8.1 Features .............................................................................. 36
8.2 Probes................................................................................. 37
8.3 Hardware ............................................................................ 37
8.4 Date & Time ........................................................................ 37
8.5 Language ............................................................................ 37
8.6 Device information ............................................................. 37
9. Maintenance and Support .......................................38
9.1 Maintenance....................................................................... 38
9.2 Support Concept ............................................................... 39
9.3 Standard Warranty and Extended Warranty ................... 39
9.4 Disposal .............................................................................. 39
© 2016 Proceq SA 2
10. Troubleshooting ......................................................... 40
10.3 Battery ................................................................................... 42
11. Equotip Link Software ............................................... 42
11.1 Starting Equotip Link ............................................................ 42
12. Technical Specifications ............................................ 45
12.1 Instrument ............................................................................. 45
12.4 Equotip Portable Rockwell Probe ....................................... 47
13. Standards & Guidelines ............................................. 47
14. Ordering Information ...............................................48
14.1 Units .................................................................................... 48
14.2 Impact Devices & Probes ................................................. 48
14.3 Parts and Accessories ...................................................... 49
14.4 Test Blocks ..............................................................................
© 2016 Proceq SA 3

1. Safety and Liability

Table of Contents

1.1 General Information

This manual contains important information on the safety, use and main­tenance of the Equotip 550. Read through the manual carefully before the first use of the instrument.

1.2 Liability

Our “General Terms and Conditions of Sales and Delivery” apply in all cases. Warranty and liability claims arising from personal injury and dam­age to property cannot be upheld if they are due to one or more of the following causes:
• Failure to use the instrument in accordance with its designated use as described in this manual.
• Incorrect performance check for operation and maintenance of the in­strument and its components.
• Failure to adhere to the sections of the manual dealing with the per­formance check, operation and maintenance of the instrument and its components.
• Unauthorised modifications to the instrument and its components.
• Serious damage resulting from the effects of foreign bodies, accidents, vandalism and force majeure.
All information contained in this documentation is presented in good faith and believed to be correct. Proceq SA makes no warranties and excludes all liability as to the completeness and/or accuracy of the information.

1.3 Safety Instructions

The equipment is not allowed to be operated by children or anyone under the influence of alcohol, drugs or pharmaceutical preparations. Anyone who is not familiar with this manual must be supervised when using the equipment.
• Carry out the stipulated maintenance properly and at the correct time.
• Following completion of the maintenance tasks, perform a functional check.

1.4 Correct Usage

The instrument is only to be used for its designated purpose as described herein.
• Replace faulty components only with original replacement parts from Proceq.
• Accessories should only be installed or connected to the instrument if they are expressly authorized by Proceq. If other accessories are installed or connected to the instrument then Proceq will accept no liability and the product guarantee is forfeited.

1.5 Optimizing Performance of the Battery System

To increase the performance of the battery, it is recommended to first completely discharge and then completely charge it.
4 © 2016 Proceq SA

2. Getting Started

Table of Contents
The Equotip 550 is typically used for testing the hardness of metallic sur­faces. The user has a choice to select either Leeb rebound, the Portable Rockwell or UCI principle, see chapter “3.1 Performing Measurements”.
In combination with the Equotip Leeb Impact Device U the instrument is used to test the roll hardness of paper, film or foil rolls.

2.1 Installation

To install the Battery into the Equotip 550 Touchscreen Unit, lift the stand as shown, insert the battery and fasten it in place with the screw.
Buttons
On the upper right of the unit there are three buttons:
Power On/Off – Press to power on or to return to the home screen. Press and hold to power off.
Soft Key – Switches in and out of full screen view or toggles between the actual screen and the last viewed pdf document (eg. Operating Instructions).
Back Button – Returns to previous screen.
Energy Saving
Energy saving may be programmed as desired under System/Power set­tings, see chapter “8.3 Hardware”.
Connections
Figure 1: Insert Battery
There are three status LEDs on the right side of the display. The middle light is the power indicator which is red when charging and turns to green when battery is fully charged. The lower LED is used for application spe­cific notification.
A complete charge requires < 9 h (Instrument not operating)
Charging time is much longer if the instrument is in use.
An optional Quick Charger (Part No. 327 01 053) can be used to
© 2016 Proceq SA 5
NOTE! Only use the battery charger provided for charging.
charge a spare battery or to charge the battery outside of the instru­ment. In this case it takes < 5.5 h for a complete charge.
1 2
Snap-in connectors
Figure 2: Connections
For Leeb Impact Devices
use the Snap-in connector 1.
For UCI Probe
use the Snap-in connectors 1 or 2.
For the Portable Rockwell Probe use the USB Host connector.
USB Host
USB Device
Ethernet
Power Supply
USB Host:
Additionally connect a mouse, keyboard or USB stick.
USB Device:
Connect to PC.
Ethernet:
Connection to network.
Power Supply:
Connect the power supply through this connection.

2.2 Main Menu

Table of Contents
On start up the main menu is displayed. All functions may be accessed directly via the Touchscreen. Return to the previous menu by pressing the back button or the return icon (arrow) at the top left corner of the Touchscreen.
Settings: For application specific settings, see chapter “4. Settings”.
Measurement: Measurement display screen, see chapter “3. Measure­ment”.
Figure 3: Main Menu
6 © 2016 Proceq SA
Wizards: Task related workflows, see chapter “6. Wizards”.
Data (Explorer): File manager for reviewing measurement data on the instrument, see chapter “5. Data (Explorer)”.
System: System settings, e. g. language, display options etc, see chapter “8. System”.
Information: Operating instructions and other reference documents, see chapter “7. Information”.

3. Measurement

Table of Contents

3.1 Performing Measurements

3.1.1 Leeb Testing Procedure (except Leeb U)
Select automatic compensation for impact direction “Automatic”, see chapter “3.2.1 Controls”. If “Automatic” is not allowed, set the impact direction (         ). The Equotip Leeb Impact Devices DL doesn’t sup­port automatic mode. The impact direction must be selected manually.
Select the appropriate material group, hardness scales and number of impacts per measurement series. For more information see chapter “4. Settings”. Conduct impacts by cycling through “load, position and trig­ger” mechanism:
1. Load the impact device – while not in contact with the test piece – by hold­ing it firmly with one hand and sliding the loading tube with the other hand until contact is grabbed by the clutch.
2. Position the support ring on the test piece. Take particular care to fully position the support ring on the test piece, but not coinciding with a previ­ous test indentation.
3. To trigger an impact, press the trigger button to release the impact body. To perform another impact, repeat this cycle.
Figure 4: Leeb Testing Procedure
After the last of the impacts is performed, the hardness average and fur­ther statistics of the measurement series are displayed.
NOTE! Make sure the loading tube is allowed to slowly return back to the starting position. Do take care so the loading tube does not spring back uncontrolled, which may result in permanent device damage.
NOTE! If possible, follow the standard practice of Leeb re­bound hardness testing as described in the standards DIN 50156-1 (metallic materials), ASTM A956 (steel, cast steel and cast iron only), or other applicable standards. If these are not available, the user is recommended to average a minimum of n = 3 impacts at an indentation distance of 3 to 5 mm (0.12 to 0.20”) for each location of the sample that shall be tested.
NOTE! Do not carry out an impact in an area that has already been deformed by another impact. Also, do not load the device when it is already positioned in the new test location, since the material under the device may be affected through prior stress, and the catch chuck of the device may get damaged.
© 2016 Proceq SA 7
3.1.2 Portable Rockwell Testing Procedure
Table of Contents
1. Place the probe on the sample to test. For flat surfaces the standard foot is most suitable. For cylindrical objects it is recommended to use one of the special feet. For locations difficult to access a tripod foot can be used. See chapter “14. Ordering Information” for more details.
2. Press the probe slowly but firmly against the surface to perform the measurement. Suppress vibrations as much as possible, and follow the instructions on the screen.
3. Release the probe when the instru­ment says so. Again, this movement has to be done in a controlled man­ner. If the probe is released too fast, a warning will be shown and the measurement should be repeated.
Figure 5: Portable Rockwell Testing Procedure
3.1.3 UCI Testing Procedure
1. Place the probe on the sample to test. The probe must be perpen­dicular to the surface (± 5°). The special foot can be used to increase repeatability and reduce distor­tion of the results, see chapter “14. Ordering Information”.
2. Press the probe slowly but firmly against the surface until the se­lected measuring force is reached. The instrument will indicate when to release the probe with an on-screen prompt and audible sound.
3. Release the probe from the mate­rial. It is important to remove the probe completely from the test object. Otherwise the results can be biased.
Figure 6: UCI Testing Procedure
NOTE! A warning will be shown if the user applied too much load when pressing the probe against the surface. Please avoid frequent overloading, as this could seriously damage the probe.
8 © 2016 Proceq SA
3.1.4 Leeb U Testing Procedure
Table of Contents
The Equotip 550 Leeb U enables the user to quickly and precisely diag­nose roll imperfections, hardness inconsistencies and uneven winding, thereby preventing problems for printing and converting operations.
With Equotip Impact Device Leeb U automatic impact direction mode is not supported and the user must select the appropriate impact direction manually (90° down, 45° down, 0°).
As for roll hardness testing no conversion curves are used, no material group has to be selected.
Conduct impacts by cycling through “position and trigger”.
1. Position the probe on the roll to be tested. Make sure to fully posi­tion the support ring on the roll to ensure an impact perpendicular to the test surface.
2. While holding the impact device firmly with two hands, slide the loading tube smoothly down to load and trigger an impact.
Move the impact device to the next point on the roll and repeat.
Figure 7: Leeb U Testing Procedure
NOTE! Some features mentioned in this Operating Instruc­tions are specifically addressing metal hardness testing ap­plications and therefore not available for Equotip Leeb U.
© 2016 Proceq SA 9

3.2 Measurements Screen

Table of Contents
3.2.1 Controls
File Name: Enter the file name and tap return. Saved measure­ments will be stored automati­cally. If filename management is activated, this function is locked.
Impact Direction: To manually set the impact direction if required (Leeb only, by default this is automatic).
Material: Select the material group to use for conversions (not available for Leeb U).
Figure 8: Measurement Screen
10 © 2016 Proceq SA
Measurement Scale: Select the hardness scale (primary and secondary) to be displayed (not available for Leeb U).
Measurement Mode:
Switch between measurement and conversion.
Wizards: Direct access to wizards.
Redo: Restart the measurement series or a single measurement.
Time and battery status
Settings: direct shortcut to
settings menu. (only applicable to the actual measurement series)
Save: Store measurement data.
Delete: Delete the last measurement.
3.2.2 Measurement Views
HL
ID
+
Table of Contents
Equotip 550 is fully customizable as a device and can display three dif­ferent measurement views simultaneously. Each view can be switched to meet the user’s requirements by simply clicking on the icon related to the particular display at the top right corner of each screen.
Signal View: Display the probe signal from the last active measurement. This may be useful for advanced evaluations.
Statistic View: View statistics for the active measurement se­ries. Number of impacts (n), Average (x), Standard deviation (σ), Minimum/Maximum () and Range () are displayed in the primary scale.
Table View: Display the measurements for the active series in a table format.
Conversion View: Display the actual value on the conversion curve.
Bar View: Display the measurements of the series as a histo­gram.
Profile View: Display the measurement results as a profile.
Info: Display the measurement settings e.g. series length,
probe type, material group etc. User’s View: The user can choose between probe angle, mini-
mum, maximum, range and probe type for the field contents. To change, tap on each individual box.
Single Record View: Display the last or selected measurement result in both the primary and secondary hardness scales.
Sample ID’s: Defines the custom field.
Figure 9: Measurement Views
© 2016 Proceq SA 11
NOTE! Screen views cannot be duplicated.

3.3 Measuring Methods

Table of Contents
Equotip 550 family of instruments is capable of accepting three diffrent test methods using a single indicating unit.
3.3.1 Equotip Leeb
3.3.1.1 Test Principle
Release button
Loading tube
Impact Spring
Catch chuck
Connection cable – 4 pole
Sensor of impact device with ID ROM
Support ring
Figure 10: Schematic View of a Leeb Impact Device
Guide tube
Impact Body
During measurement with Equotip 550 impact devices (D, DL, DC, C, G, S, and E); an impact body with a ball indenter is launched by spring energy against the sample to be measured, and then rebounds. Before and after the impact, a permanent magnet inside the impact body passes through a coil in which a voltage signal is induced by the forwards and backwards movement. This induction signal behaves proportionally to the velocities. The ratio of the rebound velocity vr to the impact velocity vi multiplied by 1000 yields the hardness value HL (Leeb hardness). HL is a direct measure of the hardness. The third resp. fourth letter of the HL unit refers to the impact device HLD D impact device.
v
r
HL =
·1000
v
i
Equotip Leeb U
Although the Equotip Leeb Impact Device U is constructed differently to simplify the measurement process, the underlying principle is the same.
Figure 11: Equotip Leeb U Impact Device
Existing Parotester impact devices type U are fully supported by the Equotip 550. Typ P and PG impact devices can be still used, but the unit is shown as HLU although it would be actually LP resp. LPG.
NOTE! HLU values can be directly compared to LU on exist­ing Parotester instruments.
12 © 2016 Proceq SA
3.3.1.2 Sample Preparations
Table of Contents
Keep the sample free of vibrations during the test. Light and thin parts must be specially fastened, see chapter “3.3.1.6 Testing Light Samples”. Ensure that the surface of the work piece is clean, smooth and dry. If re­quired, use appropriate cleaning agents for cleaning, such as acetone or isopropanol. Do not use water or any other detergent fluids.
NOTE! Please use the surface roughness comparator plate provided to estimate the average roughness of the test piece prior to testing.
Figure 12: Surface Roughness Comparator Plate
3.3.1.3 Standards
Brief descriptions of referenced standards:
DIN 50156
ASTM A956
ASTM A370
ASTM E140
ISO 18265
ISO 16859
Leeb hardness testing of metallic materials Standard test method for Leeb hardness testing of steel
products Test methods and definitions for mechanical testing of
steel products Standard hardness conversion tables for metals rela-
tionship among Brinell, Vickers, Rockwell, Superficial, Knoop, Scleroscope and Leeb hardness
Metallic materials – Conversion of hardness values Leeb hardness testing of metallic materials
© 2016 Proceq SA 13
3.3.1.4 Test Conditions
Table of Contents
To ensure proper hardness readings, the following conditions must be fulfilled. If one or more conditions are not met, the measurement result may be significantly false.
Impact device type D/DC/DL/S/E G C
Surface preparation
Roughness grade class ISO 1302 Max. roughness depth R Average roughness R
Minimum sample mass
Of compact shape (kg / lbs) On solid support (kg / lbs) Coupled on plate (kg / lbs)
Minimum sample thickness
Uncoupled (mm / inch) Coupled (mm / inch) Surface layer thickness (mm / inch)
Minimum space
Between indentation and sample edge (mm/inch)
Between indentations (mm/inch) 3 / 0.12 4 / 0.16 2 / 0.08
Indentation size on test surface
With 300 HV, 30 HRC
With 600 HV, 55 HRC
With 800 HV, 63 HRC
Table 1: Leeb Test Piece Requirements
(μm / μinch)
t
(μm / μinch)
a
Diameter (mm / inch) Depth (μm / μinch) Diameter (mm / inch) Depth (μm / μinch) Diameter (mm / inch) Depth (μm / μinch)
N7 N9 N5
10 / 400 30 / 1200 2.5 / 100
2 / 80 7 / 275 0.4 / 16
5 / 11 15 / 33 1.5 / 3.3
2 / 4.5 5 / 11 0.5 / 1.1
0.05 / 0.2 0.5 / 1.1 0.02 / 0.045
25 / 0.98 70 / 2.73 15 / 0.59
3 / 0.12 10 / 0.4 1 / 0.04
0.8 / 0.03 0.2 / 0.008
5 / 0.2 8 / 0.3 4 / 0.16
0.54 / 0.021 1.03 / 0.04 0.38 / 0.015
24 / 960 53 / 2120 12 / 480
0.45 / 0.017 0.9 / 0.035 0.32 / 0.012
17 / 680 41 / 1640 8 / 2560
0.35 / 0.013 0.30 / 0.011
10 / 400 7 / 280
14 © 2016 Proceq SA
3.3.1.5 Selecting the Equotip Leeb Impact Device
Table of Contents
For optimized testing of diverse metallic materials and sample geometries, a range of impact devices are available as per “Table 1: Leeb Test Piece Requirements”.
Type G: Increased impact energy. For solid and inhomogenous components, e.g. heavy castings and forgings.
Impact energy: 90 Nmm
Type DC: Short device. For use in very
confined spaces, e.g. in holes, cylinders or for internal measurements on assembled machines. Impact energy: 11 Nmm
Type D: Universal unit. For the majority of your hardness testing requirements.
Impact energy: 11 Nmm
Type S: Si3N4 ball indenter. For testing
especially in the very high hardness range (in excess of 50 HRC / 650 HV): Tool steels with high carbide content inclusions.
Impact energy: 11 Nmm
Figure 13: Equotip Leeb Impact Devices
© 2016 Proceq SA 15
Type E: Diamond ball indenter. For testing in the very high hardness range (in excess of 50 HRC / 650 HV): Tool steels with high carbide content inclusions. More durable than type S.
Impact energy: 11 Nmm
Type C: Reduced impact energy. Surface
hardened components, coatings, thin walled or impact sensitive components (small measuring indentation).
Impact energy: 3 Nmm
Type DL: Slim front section.
For measurements in confined spaces, at the base of grooves or on recessed surfaces. Impact energy: 11 Nmm
3.3.1.6 Testing Light Samples
Table of Contents
If the samples are lighter than specified in chapter “3.3.1.4 Test Condi­tions” or sample sections have unfavorable mass distribution they can vibrate as the impact body hits the test point. This results in unwanted energy absorption. Such samples shall be supported by solid worktops. If the mass falls below the specific requirements but still exceeds the coupling amount then coupling it to a larger mass can help prevent vibra­tions.
The following requirements must be met for coupling:
• The contact surface of the sample and the surface of the solid support must be level, flat and ground smooth.
• The sample must exceed the minimum sample thickness for coupling. Follow the coupling procedure.
• Apply a thin layer of coupling paste to the contact surface of the sample.
• Press the sample firmly against the support.
• Push the sample in a circular motion and carry out the impact as usual, perpendicular to the coupled surface.
NOTE! Clamping may strain the sample, which can affect the hardness readings.
3.3.1.7 Testing Curved Surfaces
The instrument works properly only when the ball indenter at the front of the impact body is precisely at the end of the tube at the time of the im­pact. When concave or convex surfaces are tested, the ball indenter either does not entirely leave the test tube or comes out too far. In such cases, replace the standard support ring by a specially suited ring, see chapter “14. Ordering Information” or contact your local Proceq representative.
Figure 14: Leeb Testing on Curved Surfaces
16 © 2016 Proceq SA
Loading...
+ 36 hidden pages