Princeton SP-300i, SP-500i User Manual

15 Discovery Way, Acton, MA 01720
Phone: (978)263-3584, Fax: (978)263-5086
Web Site:
www.piacton.com
Operating Instructions
for
SP-300i/500i
Ver 2.0
Supplemental Instructions for Acton Models
With Kinematically Mounted Triple Grating Turret
This Acton series monochromator/spectrograph is supplied with two kinematically mounted triple grating turret assemblies that are interchangeable in the monochromator/spectrograph without re-alignment.
To interchange the triple grating turrets in the monochromator/spectrograph, refer to the section below labeled “Grating Turret Interchange”. For general operation of the Acton series
monochromator/spectrograph, refer to the enclosed standard Acton monochromator/ spectrograph manual.
CAUTION: THE OPTICAL SURFACES IN THE MONOCHROMATOR/SEPCTROGRAPH ARE EXTREMELY DELICATE AND CAN BE PERMANENTLY DAMAGED BY CONTACT WITH ANYTHING. DO NOT TOUCH, TALK OR BREATHE ON THE OPTICAL SURFACES.
Grating Turret Interchange A) Changing Grating Turrets:
The following procedure is recommended for interchanging grating turrets. Observe proper optical handling procedures.
1. Locate the storage container with the grating turret to be installed. The turrets are marked
Turret I contains the following 3 gratings: Position 1: G/mm @
Position 2: G/mm @ Position 3: G/mm @
Turret Position 1: G/mm @
Position 2: G/mm @ Position 3: G/mm @
Turret Position 1: G/mm @
Position 2: G/mm @ Position 3: G/mm @
I, II & III as indicated in Figure 1 in this section.
II contains the following 3 gratings:
III contains the following 3 gratings: