Copyright 2003-2011 Princeton Instruments, a division of Roper Scientific, Inc.
3660 Quakerbridge Rd
Trenton, NJ 08619
TEL: 800-874-9789 / 609-587-9797
FAX: 609-587-1970
All rights reserved. No part of this publication may be reproduced by any means without the written
permission of Princeton Instruments, a division of Roper Scientific, Inc. ("Princeton Instruments").
Printed in the United States of America.
IPLab is a trademark of Scanalytics, Inc.
Logitech is a registered trademark of Logitech International SA.
Macintosh is a registered trademark of Apple Computer, Inc.
Microsoft, Windows and Vista are registered trademarks of Microsoft Corporation.
Pentium is a registered trademark of Intel Corporation.
PVCAM is a registered trademark of Photometrics, Ltd. Corporation.
Radio Shack is a registered trademark of TRS Quality, Inc.
Scientific Imaging ToolKit and SITK are trademarks of R Cubed Software Consultants, LLC.
TAXI is a registered trademark of AMD Corporation.
The information in this publication is believed to be accurate as of the publication release date. However,
Princeton Instruments does not assume any responsibility for any consequences including any damages
resulting from the use thereof. The information contained herein is subject to change without notice.
Revision of this publication may be issued to incorporate such change.
Table 6. I/O Address & Interrupt Assignments before Installing Serial Card ................. 77
Table 7. I/O Address & Interrupt Assignments after Installing Serial Card .................... 77
Table 8. Camera vs. Approximate Cooldown Temperature ............................................ 87
Table 9. WinView/WinSpec Features Supported under USB 2.0 ................................. 101
Chapter 1
Introduction
When Flexibility Counts
Princeton Instrumentsknows that a camera designed to detect x-ray photons for scientific
research needs to be flexible. X-ray scattering from complex crystals such as proteins
requires high dynamic range and low-noise readout, whereas scattering from materials
like polymers, fibers, and powders requires deep cooling and rapid imaging to capture
phase or morphological changes during thermal or mechanical processing. Shutterless
operation, low noise, and the ability to select a fiber ratio up to 5:1 make the Princeton
Instruments PI·SCXsystem an ideal choice for both of these applications.
In fact, PI-SCX high-performance systems are designed specifically for applications like
x-ray tomography, x-ray topography, medical x-ray imaging, nondestructive testing
(NDT), streak tube readout, and image intensifier readout.
When a 1:1 fiberoptic-coupling configuration is chosen, these cameras are well suited for
streak tube and image intensifier readout or low x-ray flux imaging. A softwareprogrammable high-sensitivity or high-capacity amplifier means that the PI-SCX system
can also provide x-ray photon-counting capability with 16-bit dynamic range.
Fiberoptic Expertise
The PI-SCX camera system utilizes Princeton Instruments' patented fiberoptic-coupling
technology (US Patent 5,134,680) to ensure the highest possible resolution. This process
eliminates any intermediate fiberoptic faceplate or unreliable oil layer between surfaces.
An optional phosphor screen (GdOS:Tb) that absorbs x-rays and emits visible light (~550
nm) can also be coupled to the CCD with a fiberoptic in order to detect x-rays. When the
highest sensitivity is required, such as when detecting low flux or lower x-ray energies
(~5 keV), smaller taper ratios and back-illuminated CCDs are offered. For distortion-free
imaging, a 1:1 fiberoptic system is recommended, while for maximum field of view –
ideal for high-brilliance sources like third-generation synchrotrons – tapers with
diameters as large as 165 mm are recommended.
Shuttering and Synchronization
Electrical timing signals are available from the camera controller for synchronization
with a customer-provided x-ray shutter. The CCD can be continuously cleared of dark
charge while waiting for an exposure to begin. Exposures that are much longer than the
CCD readout time can be performed without a shutter, as the amount of smearing (due to
continued exposure during readout) will be low.
9
10 PI-SCX System Manual Version 2.E
WARNING!
Low-Noise Configurations
The PI-SCX system provides either a forced-air or water-cooling option to reduce dark
current. The thermoelectric (Peltier) design and precision electronics provide very high
precision as well as the capability to set the required temperature.
For higher performance, models with the phosphor, fiberoptic, and CCD all entirely
contained within a vacuum are also available. This configuration allows deeper cooling
while eliminating condensation on the phosphor surface. Another configuration, one in
which the fiberoptic extends outside the vacuum, allows the flexibility of changing
scintillators/phosphors for specific x-ray energy without breaking vacuum or coupling the
camera system to streak tubes or image intensifiers.
Software Solutions
The PI-SCX camera runs under WinView and WinSpec, Princeton Instruments' 32-bit
Windows
provide comprehensive image or spectra capture and display functions, so you can
perform data acquisition without having to rely on third-party software. The packages
also facilitate snap-ins to permit easy user customization of any function or sequence.
Windows DLLs are available, which allows you to write your own software. This makes
integrating the system into larger experiments or instruments a straightforward endeavor.
®
software packages designed for imaging and spectroscopy, respectively. They
Custom Design
On a special-order basis, custom fiberoptic-taper configurations can be provided,
including designs that support electron backscatter imaging, simultaneous wide- and
small-angle x-ray scattering imaging, or nondestructive testing. Configurations with
multiple CCD / taper assemblies can also be provided.
Contact Princeton Instruments for the latest information concerning these and other
options for fiberoptically-coupled camera needs.
PI-SCX System Components
The Princeton Instruments PI-SCX system normally contains the major components
shown in Figure 1 and described below. Actual system components will depend on the
system configuration ordered.
Fiberoptic Taper CCD Camera: X-rays pass through the Beryllium window and
are absorbed by a phosphor screen that emits visible light. This light is coupled to the
CCD by a reducing fused fiber optic taper. The CCD detects one or more visible
photons per X-ray photon absorbed. Detected photons are converted to electrical
signals. These are then transferred from the camera to the ST-133 Controller.
Do not touch the Beryllium window at the front of the camera.
Avoid abrasion of the Beryllium window since powdered Beryllium is highly toxic if
ingested. Beryllium should never be handled in a way that places personnel at risk.
Units are shipped with a protective label over the Beryllium window. This label should
be removed and discarded. Damage to the Beryllium window may result in loss of
vacuum and improper camera operation.
Chapter 1 Introduction 11
ST-133 Controller: Controls the camera and experiment, including the temperature set
point. Collects the data and transfers it to the computer for further processing.
Camera-Controller Cable: This 10' cable (6050-0484) connects from the
DETECTOR connector (25 pins) on the back of the ST-133 to the connector
(25 pins) on the back of the camera. Note that the end that connects to the camera has
a right-angle connector and is secured by a slide lock.
User Manuals: PI-SCX System and WinView/32 Imaging Software. The PI-SCX
manual contains the basic setup and operation information for PI-SCX camera
systems. For information about WinView/32, refer to its manual and to the on-line
help supplied with the program.
Note: The system and software manuals may be provided on the software CD
supplied with your system.
WinView/32 CD-ROM: This CD contains the WinView/32 imaging software and
related manuals in PDF format.
Interface Card: Princeton Instruments (RSPI) High-Speed Serial PCI Card or USB
2.0 Card to be installed in the computer. If the computer is purchased from Princeton
Instruments, the interface card will be installed at the factory.
Interface Cable:
TAXI cable: DB9 to DB9 cable (6050-0148-CE is standard) or
USB cable: Five (5) meter cable (6050-0494) is standard
Figure 1. Standard System Components
12 PI-SCX System Manual Version 2.E
About this Manual
Manual Organization
This manual provides the user with all the information needed to install a PI-SCX camera
and place it in operation. Topics covered include a detailed description of the camera,
installation, cleaning, specifications and more.
Notes:
1. The general identifier "ST-133" is used for both the ST-133A Controller and the
ST-133B Controller. Where there is a difference, the specific identifier is used.
2. "WinX" is a generic term for WinView, WinSpec, and WinXTest application
software.
Chapter 1, Introduction briefly describes the PI-SCX family of cameras; details
the structure of this manual; and documents environmental, storage, and cleaning
requirements.
Chapter 2, Component Descriptions provides descriptions of each system
component.
Chapter 3, Installation Overview cross-references system setup actions with the
relevant manuals and/or manual pages. It also contains system layout diagrams.
Chapter 4, System Setup provides detailed directions for interconnecting the
system components.
Chapter 5, Operation includes a simple procedure for verifying system operation
and discusses operational considerations associated with exposure, readout, and
digitization.
Chapter 6, Advanced Topics discusses standard timing modes (Free Run,
External Sync, and Continuous Cleans), Fast and Safe, and TTL control.
Chapter 7, Troubleshooting provides courses of action to take if you should
have problems with your system.
Appendix A, Specifications includes PI-SCX camera and ST-133 specifications.
Appendix B, Outline Drawings includes outline drawings of the PI-SCX cameras
and the ST-133A and ST-133B Controllers.
Appendix C, Repumping the Vacuum explains how to restore the camera's
vacuum if that vacuum has deteriorated over time.
Appendix D, USB 2.0 Limitations covers the currently known limitations
associated with operating under the USB 2.0 interface.
Declaration of Conformity contains the Declaration of Conformity for PI-SCX
systems.
Warranty & Service provides warranty and customer support contact information.
Chapter 1 Introduction 13
Caution! The use of this symbol on equipment indicates that one or more nearby
items should not be operated without first consulting the manual. The same
symbol appears in the manual adjacent to the text that discusses the hardware
item(s) in question.
Warning! Risk of electric shock! The use of this symbol on equipment
indicates that one or more nearby items pose an electric shock hazard and should
be regarded as potentially dangerous. This same symbol appears in the manual
adjacent to the text that discusses the hardware item(s) in question.
Safety Related Symbols Used in this Manual
Environmental Conditions
Storage temperature < 50°C
Operating environment 0°C < T< 30°C
Operating temperature range over which specifications can be met is 18 C to 23 C
Relative humidity 50%, noncondensing.
Repairs
Repairs must be done by Princeton Instruments. If your system hardware needs repair,
contact Princeton Instruments Customer Service. Please save the original packing
material so you can safely ship the system to another location or return it for repairs.
Warnings
To prevent injury, please observe the following warnings:
Beryllium
Do not touch the Beryllium window at the front of the camera.
Avoid abrasion of the Beryllium window since powdered Beryllium is highly
toxic if ingested. Beryllium should never be handled in a way that places
personnel at risk.
Units are shipped with protective label over the Beryllium window. This label
should be removed and discarded. Damage to the Beryllium window may result
in loss of vacuum and improper camera operation.
For additional information on safe handling practices or technical data on
Beryllium, contact Brush Wellman Inc. at www.brushwellman.com
Coolant
COOLANT IS HARMFUL IF SWALLOWED. Store coolant securely and keep
it out of the reach of children.
14 PI-SCX System Manual Version 2.E
WARNING!
Precautions
To prevent permanently damaging the system, please observe the following precautions:
Always switch off and unplug the ST-133 Controller before changing your system
configuration in any way.
The CCD array is very sensitive to static electricity. Touching the CCD can destroy
it. Operations requiring contact with the device can only be performed at the factory.
Never operate the camera cooled without proper evacuation or backfill. This could
damage the CCD!
Never connect or disconnect any cable while the PI-SCX system is powered on.
Reconnecting a charged cable may damage the CCD.
Never prevent the free flow of air through the equipment by blocking the air vents.
Never operate a water-assisted or water-cooled-only PI-SCX:1300 camera with
coolant at a temperature below that specified for it.
Cleaning
Turn off all power to the equipment and secure all covers before cleaning the units.
Otherwise, damage to the equipment or personal injury could occur.
Camera and Controller
Although there is no periodic maintenance that must be performed on the camera or the
ST-133 Controller, you may clean these components from time to time by wiping them
down with a clean damp cloth. This operation should only be done on the external
surfaces and with all covers secured. In dampening the cloth, use clean water only. No
soap, solvents or abrasives should be used. Not only are they not required, but they could
damage the finish of the surfaces on which they are used.
Beryllium Window
Cleaning may be necessary to remove oil or other contaminants from the surface of the
window. Because a fingerprint left on the surface will disrupt the effectiveness of the
final etch or coating and because of the potential toxicity, protective gloves should be
worn when cleaning the window. To clean the window, wipe it down with isopropanol
and a lintless cloth. DO NOT use water. Beryllium is highly susceptible to localized
pitting when in contact with the chloride and sulfate ions contained in ordinary water.
Princeton Instruments Customer Service
Refer to the contact information located on page 108 of this manual.
Chapter 2
System Component Descriptions
PI-SCX Camera
CCD Array: PI-SCX currently offers three CCD formats: the 1340x1300 array with
20x20 µm pixels, the 2084x2084 array with 24x24 µm pixels, and the 4096x4096 array
with 15x15 µm pixels. All are scientific grade full-frame arrays with MPP. The
application of indium tin oxide (ITO) technology to the PI-SCX:4300 array provides a
QE of approximately 65% to 550 nm. When used with an x-ray scintillator screen and a
software-programmable, high-capacity or high-sensitivity amplifier, these systems can
effectively provide x-ray photon-counting capability with up to 16-bit dynamic range.
CCD Chamber: The vacuum-sealed CCD chamber protects the CCD from
contamination as well as insulates it from the warmer air in the camera body. The
inherent low humidity prevents condensation on the cooled surface of the array. The
thermal barrier provided by the fiberoptic prevents the condensation from forming on the
front of the fiberoptic .
PI-SCX cameras are normally shipped with a vacuum level of ~10 mTorr or better.
Because this vacuum may deteriorate over time due to outgassing of electrical
components, some PI-SCX cameras are designed with a built-in vacuum port that can be
used to restore the vacuum to its original level. Instructions for repumping the vacuum
are provided in Appendix C.
Fiber Optic: The PI-SCX fiberoptic tapers are bonded to the face of the CCD arrays
with Princeton Instruments' patented fiberoptic-coupling technology. The direct bonding
to the face of the array eliminates the need for an intermediate fiberoptic faceplate or an
oil layer between surfaces, thereby increasing sensitivity. The PI-SCX:1300 is bonded to
a 1:1 or 1.5:1 fiberoptic taper which extends outside the vacuum for a resolution of
25 lp/mm for the 1:1 taper or 16 lp/mm for the 1.5:1 taper. The PI-SCX:4096 is available
with either a 1:1 fiberoptic taper (31 lp/mm, 61x61 mm field of view) or a 1.9:1
fiberoptic taper (120x120 mm field of view). The PI-SCX:4300 is available with either a
1:1 fiberoptic taper (20 lp/mm) or a 2.4:1 taper (120x120 mm field of view).
Thermoelectric Cooler: While the CCD accumulates charge, thermal activity releases
electrons, generating dark current. Cooling the CCD enhances the low-light sensitivity by
reducing thermally generated charge. With forced-air assistance, the PI-SCX:1300
camera’s thermoelectric cooler is capable of cooling the CCD to -30°C with ±0.05°C
stability at temperature lock. With 25°C water for water-assisted cooling, the
PI-SCX:1300 camera can reach -35°C. With +5°C water, the PI-SCX:4096 (≤1.5:1 fiber
ratio) and the PI-SCX:4300 are designed to reach -50°C. With +0°C water, the
PI-SCX:4096 (>1.5:1 fiber ratio) is designed to reach -45°C
Cooling is accomplished by mounting the CCD on a cold finger, which in turn is seated
on a thermoelectric (Peltier-effect) cooler. With air-cooling, the heat is transferred
through the Peltier stages to internal fins and is removed by forced air. With watercooling, a 50:50 ethylene glycol-water coolant is circulated through a heatsink thermally
15
16 PI-SCX System Manual Version 2.E
connected to the Peltier. CCD temperature is controlled and monitored by the host
computer and the ST-133 Controller.
Electronics: The camera electronics enclosure contains the preamplifier and array
driver board. This design keeps all signal leads to the preamplifier as short as possible
and also provides complete RF shielding.
Speed of data acquisition and dynamic range is determined primarily by the A/D converter
used (binning on the array is also a factor). The PI-SCX is available with a 1 MHz, 16-bit
readout. The PI-SCX:1300 is available with dual-speed digitization (100 kHz and 1 MHz).
These 16-bit digitizers give you the choice of the 100 kHz for the better signal-to-noise ratio
or the 1 MHz for increased data acquisition speed. Both the PI-SCX:1300 and the
PI-SCX:4300 have dual-amplifier circuits (low-noise or high-capacity). The PI-SCX:4096
has a single readout-amplifier.
Connectors: Power, control signals, and data are transmitted between the ST-133 and the
PI-SCX camera via the 25-pin D connector located on the rear of the camera.
Mounting Holes: The PI-SCX:1300 camera can be ordered with an optional tripod
mount kit with three different mounting holes (1/4"-20, 3/8"-16, and M6). The
PI-SCX:4096 and the PI-SCX:4300 cameras have three threaded holes on the underside
of the camera body (two are M8 x 1.25 tap, 25 mm deep; and one is 1/4"-20 tap, .50"
deep).
Fan: PI-SCX:1300 only. There may be a fan located inside the camera's back panel. Its
purpose is:
to remove heat from the Peltier device that cools the CCD array
to cool the electronics.
An internal Peltier device directly cools the cold finger on which the CCD is
mounted. The heat produced by the Peltier device is then removed by the air
drawn into the camera by the internal fan and exhausted through the back panel.
The fan is always in operation and air cooling of both the Peltier and the internal
electronics takes place continuously. The fan is designed for low-vibration and
does not adversely affect the image. For the fan to function properly, free
circulation must be maintained between the rear of the camera and the laboratory
atmosphere.
Coolant Ports: Depending on the camera, there may be two coolant ports for watercooling. The Inlet and Outlet ports for the PI-SCX:4096 and the PI-SCX:4300 are
interchangeable, are barbed (straight or right-angled), and will accommodate 3/8" ID
Tygon tubing . Optimal cooling for the PI-SCX:1300 requires that the correct ports be
used for the Inlet and the Outlet. These quick-disconnect ports require 1/4" thin-wall
plastic tubing.
Chapter 2 System Component Descriptions 17
POWER Switch and Indicator: The power
switch location and characteristics depend on
the version of ST-133 Controller that was
shipped with your system. In some versions,
the power switch (when located on the front
panel as shown in Figure 3), has an integral
indicator LED that lights whenever the ST-133
is powered. In other versions, the power switch
is located on the back of the ST-133 and does
not include an indicator LED.
Rear Panel Connectors: There are three
controller board slots. Two are occupied by the
plug-in cards that provide various controller
Figure 2. Power Switch Location
(ST-133A and ST-133B)
Pixel A/D conversion
Timing and synchronization of readouts
CCD scan control
Temperature control
Exposure control
Video output control
WARNING!
ST-133 Controller
Electronics: The Model ST-133 is a compact, high performance CCD Camera
Controller for operation with Princeton Instruments cameras. Designed for high speed
and high performance image acquisition, the ST-133 offers data transfer at speeds up to 5
megapixel per second, standard video output for focusing and alignment. A variety of
A/D converters are available to meet different speed and resolution requirements.
In addition to containing the power supply, the controller contains the analog and digital
electronics, scan control and exposure timing hardware, and controller I/O connectors, all
mounted on user-accessible plug-in modules. This highly modularized design gives
flexibility and allows for convenient servicing.
functions. The third, covered with a blank panel, is reserved for future development. The leftmost plug-in card is the Analog/Control module. Adjacent to it is the Interface Control
module. Both modules align with top and bottom tracks and mate with a passive backplane
via a 64-pin DIN connector. For proper operation, the location of the modules should not be
changed. Each board is secured by two screws that also ground each module’s front panel.
Removing and inserting boards is described in Chapter 7, pages 80-81.
To minimize the risk of equipment damage, a module should never be removed or
installed when the system is powered.
The Analog/Control Module, which should always be located in the left-most slot,
provides the following functions.
The Interface Control Module, which should always be located in the center slot,
provides the following functions.
TTL in/out Programmable Interface
Communications Control (TAXI or USB 2.0 protocol)
18 PI-SCX System Manual Version 2.E
WARNING!
Always turn the power off at the Controller before connecting or disconnecting any cable
that interconnects the camera and controller or serious damage to the CCD may result.
This damage is NOT covered by the manufacturer’s warranty.
Figure 3. ST-133 Rear Panel
Chapter 2 System Component Descriptions 19
# Feature
1. Temperature Lock LED: Indicates that the temperature control loop has locked and that
the temperature of the CCD array will be stable to within 0.05C.
2.Video Output: If labeled Video, composite video output is provided at this connector. The
amplitude is 1 V pk-pk and the source impedance is 75 . Note that video output is not
currently supported under USB 2.0. If labeled Aux, this output is reserved for future use.
3.External Sync Input: TTL input that has a 10 k pullup resistor. Allows data acquisition and
readout to be synchronized with external events. Through software, either positive or negative
(default) edge triggering can be selected.
4. Output WinView/32 (ver. 2.4 and higher) software selectable NOT SCAN or
SHUTTER signal. Default is SHUTTER. , reports when the controller is finished
reading out the CCD array. is high when the CCD array is not being scanned, then
drops low when readout begins, returning to high when the process is finished. The second
signal, SHUTTER, reports when the shutter is opened and can be used to synchronize
external shutters. SHUTTER is low when the shutter is closed and goes high when the
shutter is activated, dropping low again after the shutter closes. See Figure 4 for timing
diagram.
5. Output: Initially HIGH. After a Start Acquisition command, this output changes
state on completion of the array cleaning cycles that precede the first exposure. Initially
high, it goes low to mark the beginning of the first exposure. In free run operation it remains
low until the system is halted. If a specific number of frames have been programmed, it
remains low until all have been taken, then returns high.
6.Zero Adjustment: Control the offset values of the Fast (F) and Slow (S) A/D converters.
Preadjusted at factory. The offset is a voltage that is added to the signal to bring the A/D
output to a non-zero value, typically 50-100 counts. This offset value ensures that all the
true variation in the signal can really be seen and not lost below the A/D "0" value. Since
the offset is added to the signal, these counts only minimally reduce the range of the signal
from 65535 (16-bit A/D) to a value in the range of 50-100 counts lower.
Caution: Do not adjust the offset values to zero, or some low-level data will be missed.
7. Detector Connector: Transmits control information to the camera and receives data back from the camera via the Detector-Controller cable.
8. TTL In/Out: User-programmable interface with eight input bits and eight output bits that
can be written to or polled for additional control or functionality. Under USB 2.0, output is
not currently supported in WinView. See Chapter 6.
9. AUX Output: Reserved for future use.
10. Serial COM Connector: Provides two-way serial communication between the controller and
the host computer. Contact the factory if an application requires use of the optional fiberoptic
data link to increase the maximum allowable distance between the Camera and the
computer.
11. Fan: Cools the controller electronics. Runs continuously when the controller is turned on. Do not
block the side vents or the fan exhaust port.
Rear Panel Features: The descriptions of the rear panel connectors are keyed to Figure 3.
Depending on your system, either the TAXI or the USB 2.0 Interface Control Module will be installed
in the second from the left slot (as you face the rear of the ST-133). The TAXI module is shown in
that position.
20 PI-SCX System Manual Version 2.E
# Feature
12. Shutter Setting Selector: Sets the shutter hold voltage. Contact factory for the correct
shutter setting for a remote x-ray shutter.
13.Remote Shutter Connector: Provides shutter-drive and hold voltages for a Princeton
Instruments 25 mm external shutter (typically, an entrance slit shutter). May be used for a remote
x-ray shutter.
WARNING:Dangerous live potentials are present at the Remote Shutter Power connector.
To avoid shock hazard, the Controller power should be OFF when connecting or
disconnecting a remote shutter.
14. Power Input Module: Contains the power cord socket and two fuses. Depending on the
ST-133 version, the power switch may be located directly above the power module.
15. Fuse/Voltage Label: Displays the controller’s power and fuse requirements. This label may
appear below the power module.
16. USB 2.0 Connector: Provides two-way serial communication between the controller and the
host computer. Uses USB 2.0 protocol.
Detector-Controller: 1 MHz or 100kHz/1MHz systems. The standard 10'
cable (6050-0484) has DB-25 Male connectors with slide-latch locking
hardware. This cable interconnects the Detector connector on the rear of the
ST-133 with the 25-pin D connector on the back of the PI-SCX camera. The
Detector-Controller cable is also available in 6', 15', 20', and 30' lengths. Note
that a longer cable may degrade camera performance.
Computer Interface Cable: Depending on the system configuration, either a
USB or a TAXI cable will be shipped.
Figure 4. NOT SCAN and SHUTTER Signals
Cables
Chapter 2 System Component Descriptions 21
TAXI: The standard 25' (7.6 m) cable (6050-0148-CE) has DB-9 Male
connectors with screw-down locking hardware. The TAXI (Serial
communication) cable interconnects the Serial Com connector on the rear of the
ST-133 with the PCI card installed in the host computer. In addition to the
standard length, this cable is available in 10', 50', 100', and 165' lengths. Also
available are fiber optic adapters with fiber optic cables in 100, 300, and 1000
meter lengths.
USB 2.0: The standard 16.4' (5 m) cable (6050-0494) has USB connectors
that interconnect the USB 2.0 connector on the rear of the ST-133 with a
USB card installed in the host computer.
Interface Card
PCI Card: This interface card is required when the system interface uses the
TAXI protocol rather than USB 2.0. The PCI card plugs-into the host computer's
motherboard and provides the serial communication interface between the host
computer and the ST-133. Through WinView/32 or WinSpec/32, the card can be
used in either High Speed PCI or PCI(Timer) mode. High Speed PCI allows data
transfer to be interrupt-driven and can give higher performance in some situations.
PCI(Timer) allows data transfer to be controlled by a polling timer.
USB 2.0 Card: This interface card is required when the system interface uses the
USB 2.0 protocol rather the TAXI protocol and the computer does not have native
USB 2.0 support. The USB 2.0 card plugs-into the host computer's motherboard and
provides the communication interface between the host computer and the ST-133. The
USB 2.0 PCI card (70USB90011) by Orange Micro is recommended for desktop
computers; the SIIG, Inc. USB 2.0 PC Card, Model US2246 is recommended for
laptop computers. See www.orangemicro.com or www.siig.com, respectively, for
more information.
Application Software
The Princeton Instruments WinView/32 software package provides comprehensive
image acquisition, display, processing, and archiving functions so you can perform
complete data acquisition and analysis without having to rely upon third-party
software. WinView/32 provides reliable control over all Princeton Instruments
cameras, regardless of array format and architecture, via an exclusive universal
programming interface (PVCAM®). WinView/32 also features snap-ins and macro
record functions to permit easy user customization of any function or sequence.
PVCAM is the standard software interface for cooled CCD cameras from
Princeton Instruments. It is a library of functions that can be used to control and
acquire data from the camera when a custom application is being written. For
example, in the case of Windows, PVCAM is a dynamic link library (DLL). Also,
it should be understood that PVCAM is solely for camera control and image
acquisition, not for image processing. PVCAM places acquired images into a
buffer, where they can then be manipulated using either custom written code or by
extensions to other commercially available image processing packages.
Scientific Imaging ToolKit™ (SITK™) is a collection of LabVIEW
scientific cameras and spectrographs. This third party software can be purchased
from Princeton Instruments.
®
VIs for
22 PI-SCX System Manual Version 2.E
User Manuals
PI-SCX System User Manual: This manual describes how to install and use
the PI-SCX system components.
WinView/32 User Manual: This manual describes how to install and use the
application program. A PDF version of this manual is provided on the installation
CD. Additional information is available in the program's on-line help.
Note: You can download current versions of Princeton Instruments manuals at
1. If the system components have not already been unpacked, unpack
them and inspect their carton(s) and the system components for intransit damage. Store the packing materials.
Chapter 4 System Setup,
page 25
2. Verify that all system components have been received.
Chapter 4 System Setup,
page 25
3. If the components show no signs of damage, verify that the
appropriate voltage settings have been selected for the Controller.
Chapter 4 System Setup,
page 28
4. If the WinView/32 software is not already installed in the host
computer, install it. This will install the appropriate drivers for the
interface card.
WinView/32 manual
5. If the appropriate interface card is not already installed in the host
computer, install it.
Chapter 4 System Setup,
PCI Interface, page 31 or
USB 2.0 Interface, page 32
6. With the Controller and computer power turned OFF, connect the
interface cable (TAXI or USB) to the Controller and the interface
card in the host computer. Then tighten down the locking hardware.
Chapter 4 System Setup,
page 35
7. With the Controller power turned OFF, connect the Detector-Controller
cable to the appropriate connector on the rear of the Controller and the
other end to the appropriate connector on the rear of the Camera.
Adjust the slide latches so the cable connections are locked.
Chapter 4 System Setup,
page 35
Chapter 7, Troubleshooting,
page 82
8. With the Controller power turned OFF, connect the Controller
power cable to the rear of the controller and to the power source.
9. If the camera has water-assisted cooling or is water-cooled, make the
tubing connections between the circulator and the camera. Fill the
circulator with the required mixture, turn on the circulator, turn on
the refrigeration, and set the coolant temperature.
Chapter 4 System Setup,
page 37
Chapter 5 Operation,
page 43
10. Turn the Controller ON.
Installation Overview
The list and diagrams below briefly describe the sequence of actions required to
hookup your system and prepare to gather data. Refer to the indicated references
for more detailed information. This list assumes that the application software is
Princeton Instruments WinView/32.
23
24 PI-SCX System Manual Version 2.E
Action
Reference
11. Turn on the computer and begin running WinView/32. When the
computer boots, you may be asked for the location of the interface
drivers.
Chapter 4 System Setup,
page 31 (PCI drivers) or
page 32 (USB drivers)
WinView/32 manual
12. Enter the hardware setup information or load the defaults from the
controller.
Chapter 5 Operation,
page 43
13. Set the target array temperature.
Chapter 5 Operation,
page 43
14. When the system reaches temperature lock, begin acquiring data in
focus mode.
Chapter 5 Operation,
page 45
Figure 5. PI-SCX:1300 System Diagram
Figure 6. PI-SCX:4096 and PI-SCX:4300 System Diagram
Chapter 4
System Setup
Unpacking the System
During the unpacking, check the system components for possible signs of shipping
damage. If there are any, notify Princeton Instruments and file a claim with the carrier. If
damage is not apparent but camera or controller specifications cannot be achieved,
internal damage may have occurred in shipment. Please save the original packing
materials so you can safely ship the camera system to another location or return it to
Princeton Instruments for repairs if necessary.
Checking the Equipment and Parts Inventory
Confirm that you have all of the equipment and parts required to set up the system. A
complete PI-SCX system consists of
Camera: PI-SCX:1300, PI-SCX:4096, or PI-SCX:4300.
ST-133 Controller: Do not substitute any other controller for the controller supplied
with your system.
Camera to Controller cable: DB25 to DB25, 10 ft (6050-0484).
Controller to Computer cable:
TAXI cable: DB9 to DB9 cable (6050-0148-CE is standard) or
USB cable: Five (5) meter cable (6050-0494) is standard
Interface Board: PCI or USB 2.0
Proprietary Princeton Instruments (RSPI) High Speed PCI Interface board for
TAXI interface.
USB 2.0 board for USB 2.0 interface is user-provided: native USB 2.0 support on
the motherboard or USB 2.0 Interface Card (Orange Micro 70USB90011 USB2.0
PCI is recommended for desktop; SIIG, Inc. USB 2.0 PC Card, Model US2246
for laptop).
WinView/32 CD-ROM: This CD contains the WinView/32 imaging software and
related manuals in PDF format.
User Manuals: PI-SCX System and WinView/32 Imaging Software. These manuals
are supplied on the CD shipped with your system.
Host Computer: Typically, the computer is user-supplied.
Coolant Circulator: Not required by some systems. Typically, the coolant circulator
and hoses are user-supplied.
25
26 PI-SCX System Manual Version 2.E
WARNING!
System Requirements
Environmental
Operating temperature: 0ºC to +30ºC;
Operating temperature range over which system specifications can be met: +18ºC to
+23ºC
Relative humidity <50% noncondensing.
Ventilation
Camera: Allow at least one inch of clearance for side and rear air vents.
ST-133: There is an internal fan located at the right side of the rear panel behind an
exhaust opening. Its purpose is simply to cool the controller electronics. This fan
runs continuously whenever the controller is powered. Air enters the unit through
ventilation openings on the side panels, flows past the warm electronic
components as it rises, and is drawn out the rear of the controller by the fan. It is
important that there be an adequate airflow for proper functioning. As long as
both the controller’s intake ventilation openings and the fan exhaust opening
aren’t obstructed, the controller will remain quite cool.
Vacuum
The CCD is housed in the vacuum with the fiberoptic taper extending beyond the vacuum
chamber with a thin Beryllium window to block visible light and to admit X-rays. The
camera is shipped with a vacuum quality superior to the minimum required to assure
proper operation of the system for an extended period of time without concern for the
quality of the vacuum. See Appendix C for additional information.
Coolant
COOLANT IS HARMFUL IF SWALLOWED.
KEEP OUT OF REACH OF CHILDREN.
PI-SCX cameras with water-assisted cooling or water-only cooling require circulating
coolant (50:50 mixture of ethylene glycol and water) for proper operation. All hose
connections should be secured with good quality hose clamps.
Flow Rate: Users are advised to install a flow meter to monitor the rate.
PI-SCX:1300 Two quick-disconnect coolant ports are supplied with a water-
cooled system and are mounted at the sides of the camera. For the best cooling, make
the hose connections to the Inlet and Outlet ports as shown in the outline drawing on
page 38. Use 1/4" thin-wall tubing.
Chapter 4 System Setup 27
Caution
The plug on the line cord supplied with the system should be compatible with the linevoltage outlets in common use in the region to which the system is shipped. If the line
cord plug is incompatible, a compatible plug should be installed, taking care to maintain
the proper polarity to protect the equipment and assure user safety.
Caution
PI-SCX:4096 and PI-SCX:4300 Two barbed 3/8 O.D. coolant ports are located
at the back of the camera. Either port can be the inlet or the outlet. 3/8" ID Tygon
tubing is suitable.
Coolant Temperature: Coolant should be no colder than the following temperatures:
PI-SCX:1300 +25°C. Operating a PI-SCX:1300 camera with coolant at a colder
temperature could cause induced condensation in the electronics enclosure and
possible catastrophic damage to the camera. Damage resulting from this type of
operation may void the warranty.
PI-SCX:4096+5°C for tapers with ≤ 1.5:1 fiber ratio; +0°C for tapers with
>1.5:1 fiber ratio.
PI-SCX:4300+5°C.
Power
Camera: The PI-SCX camera receives its power from the controller, which in turn plugs
into a source of AC power.
ST-133: The ST-133 Controller can operate from any one of four different nominal line
voltages: 100, 120, 220, or 240 V AC. Refer to the Fuse/Voltage label on the
back of the ST-133 for fuse, voltage, and power consumption information.
Host Computer
Note: Computers and operating systems all undergo frequent revision. The following
information is only intended to give an approximate indication of the computer
requirements. Please contact the factory to determine your specific needs.
Requirements for the host computer depend on the type of interface, TAXI or USB 2.0,
that will be used for communication between the ST-133 and the host computer. Those
requirements are a listed below according to protocol.
TAXI Protocol:
AT-compatible computer with 200 MHz Pentium
Windows
®
XP (32-bit, SP3 or later) or Vista
Princeton Instruments (RSPI) High speed serial PCI card (or an unused PCI card
slot). Computers purchased from Princeton Instruments are shipped with the PCI
card installed if High speed PCI was ordered.
Minimum of 256 Mbytes of RAM.
CD-ROM drive.
Hard disk with a minimum of 80 Mbytes available. A complete installation of the
program files takes about 17 Mbytes and the remainder is required for data
storage, depending on the number and size of spectra collected. Disk level
compression programs are not recommended.
®
II (or better).
®
(32-bit) operating system.
28 PI-SCX System Manual Version 2.E
The Power Input Module on the rear of the Controller
contains the voltage selector drum, fuses and the power cord
connector. The appropriate voltage setting is set at the
factory and can be seen on the power input module.
Each setting actually defines a range and the setting that is
closest to the actual line voltage should have been selected.
The fuse and power requirements are printed on the panel
above the power input module. The correct fuses for the
country where the ST-133 is to be shipped are installed at
the factory.
Note: On ST-133s, the voltage ranges and fuse ratings may
be printed above or below the power module (Figure 7).
Figure 7. Controller
Power Input Module
Super VGA monitor and graphics card supporting at least 256 colors with at least
1 Mbyte of memory. Memory requirement is dependent on desired display
resolution.
Two-button Microsoft
®
-compatible serial mouse or Logitech® three-button
serial/bus mouse.
USB 2.0 Protocol:
AT-compatible computer with Pentium 3 or better processor and runs at 1 GHz or
better.
Windows
®
XP (32-bit, SP3 or later) or Vista® (32-bit).
Native USB 2.0 support on the mother board or USB Interface Card (Orange
Micro 70USB90011 USB2.0 PCI is recommended for desktop; SIIG, Inc. USB
2.0 PC Card, Model US2246 for laptop)
Minimum of 256 Mbytes of RAM.
CD-ROM drive.
Hard disk with a minimum of 80 Mbytes available. A complete installation of the
program files takes about 17 Mbytes and the remainder is required for data
storage, depending on the number and size of spectra collected. Disk level
compression programs are not recommended.
Super VGA monitor and graphics card supporting at least 256 colors with at least
1 Mbyte of memory. Memory requirement is dependent on desired display
resolution.
Two-button Microsoft compatible serial mouse or Logitech three-button
serial/bus mouse.
Verifying Controller Voltage Setting
Chapter 4 System Setup 29
To Check the Controller's Voltage Setting:
1. Look at the lower righthand corner on the rear of the Controller. The current voltage
setting (100, 120, 220, or 240 VAC) is displayed on the Power Input Module.
2. If the setting is correct, continue with the installation. If it is not correct, follow the
instructions on page 67 for changing the ST-133 Controller's voltage setting and
fuses.
Mounting the Camera
PI-SCX:1300 This camera head style is available with a tripod adapter option with three
mounting holes (1/4"-20, 3/8"-16, and M6). Figure 8 shows the adapter mounted to the
bottom of the camera.
Figure 8. Mounting Holes on Tripod Adapter
30 PI-SCX System Manual Version 2.E
Installation is performed via the
WinView/32 installation process. If
you are installing WinView for the
first time, you should run the
installation before the interface card
is installed in the host computer. On
the Select Components dialog
(see Figure 10), click on the AUTO
PCI button to install the interface
card drivers (the Princeton
Instruments PCI and the USB
drivers) and the most commonly
installed program files. Select the
Custom button if you would like to
choose among the available
program files or do not want to
Figure 10. WinView Installation: Interface Card
Driver Selection
install the PCI driver.
Note: WinView/32 and WinSpec/32 (versions 2.6.0 and higher) do not support the ISA
interface.
PI-SCX:4300 On the bottom of the camera towards the rear are three threaded holes that
can be used to mount the camera. Two of the holes are M8 × 1.25 tap (25 mmdeep) and
the third one is 1/420 tap (0.5deep). Figure 9 shows the location of the mounting
holes and their distance from the back of the camera.
Figure 9. Mounting Holes on Bottom of Camera
Installing the Application Software
Chapter 4 System Setup 31
A Princeton Instruments (RSPI) PCI card must be installed in the host computer if
the communication between computer and controller uses the TAXI protocol (i.e., the
Interface Control Module installed in the ST-133 has a 9-pin SERIAL COM
connector as shown in the figure at right). With TAXI protocol, the standard cable
provided with an ST-133 is 7.6 meters (25 feet). Cable lengths up to 50 meters
(164 feet) are available and the digitization rate may be as high as 2 MHz.
A computer purchased from Princeton Instruments will be shipped with the PCI card
already installed. Otherwise, a PCI card will be shipped with the system and you will
have to install it in the host computer at your location.
Note: The PCI card can be installed and operated in any Macintosh having a PCI
bus, allowing the ST-133 to be controlled from the Macintosh via IPLab™
software and the PI Extension.
Caution
Setting up the Communication Interface
PI-SCX camera systems require either an installed Princeton Instruments (RSPI) PCI
card or an installed USB2.0 interface card in the host computer. The type of interface
card is dictated by the Interface Control Module installed in the ST-133 controller.
Setting up a PCI Interface
Administrator privileges are required under Windows® XP and Windows Vista®
(32-bit) to install software and hardware.
If using WinX software, select either High Speed PCI or PCI(Timer) as the Interface
type. This selection is accessed on the Hardware Setup|Interface tab. High Speed
PCI allows data transfer to be interrupt-driven and gives the highest performance in some
situations. PCI(Timer) allows data transfer to be controlled by a polling timer. This
selection is recommended when there are multiple devices sharing the same interrupt.
To Install a PCI Serial Buffer Card in the Host Computer:
1. Review the documentation for your computer and PCI card before continuing
with this installation.
2. To avoid risk of dangerous electrical shock and damage to the computer, verify
that the computer power is OFF.
3. Remove the computer cover and verify that there is an available PCI slot.
4. Install the PCI card in the slot.
5. Make sure that the card is firmly seated and secure it.
6. Replace and secure the computer cover and turn on the computer only. If an error
occurs at bootup, either the PCI card was not installed properly or there is an address
or interrupt conflict. Refer to Chapter 9 "Troubleshooting", page 76 for instructions.
Note: The PCI card has no user-changeable jumpers or switches.
32 PI-SCX System Manual Version 2.E
Windows Version
PCI INF Filename
Located in "Windows"\Inf
directory*
PCI Device Driver Name
Located in "Windows"\System32\Drivers
directory
Windows® XP and
Windows Vista®
(32-bit)
rspi.inf (in Windows\inf, for
example)
rspipci.sys (in Windows\System32\Drivers,
for example)
* The INF directory may be hidden.
Administrator privileges are required under Windows® XP and Windows
Vista® (32-bit) to install software and hardware.
Your system has been configured to use the USB communication protocol if the
Interface Control Module installed in the ST-133 has a USB 2.0 connector as
shown in the figure at right). The advantages to the USB 2.0 interface are that it
uses a much higher data transfer rate than many common serial data formats (such
as the TAXI protocol) and it simplifies the connection to external devices. USB
supports "plug and play" -- you do not need to be heavily involved in the setup
process.
To Install the PCI Card Driver
The following information assumes that you have already installed the WinView/32 or
WinSpec/32 software.
1. After you have secured the PCI card in the computer and replaced the cover, turn
the computer on.
2. At bootup, Windows will try to install the new hardware. If it cannot locate the
driver, you will be prompted to enter the directory path, either by keyboard entry
or by using the browse function.
If you selected AUTO PCI during the application software installation, WinView/32
or WinSpec/32 automatically put the required INF file into the Windows\Inf
directory and the PCI card driver file is copied into the Windows\System32\
Drivers directory. Refer to Table 1 for the appropriate file names and locations.
Table 1. PCI Driver Files and Locations
Setting up a USB 2.0 Interface
USB 2.0 Limitations
Maximum cable length is 5 meters (16.4 feet)
1 MHz is currently the upper digitization rate limit for the ST-133
Controller. Large data sets and/or long acquisition times may be subject
to data overrun because of host computer interrupts during data acquisition.
USB 2.0 is not supported by the Princeton Instruments PC Interface Library (Easy
DLLS).
Some WinView and WinSpec 2.5.X features are not fully supported with
USB 2.0. Refer to Appendix D, page 101, for more information.
Note: If you are installing the USB 2.0 interface on a laptop, you will need to perform all
of the operations described in this section. In addition, if you are using the recommended
USB Interface Card (SIIG, Inc. USB 2.0 PC Card, Model US2246), you must replace the
OrangeUSB USB 2.0 Host Controller driver installed for that card with the appropriate
Chapter 4 System Setup 33
Microsoft driver. Instructions for making the replacement are included in "To Update the
OrangeUSB USB 2.0 Driver".
To Update the OrangeUSB USB 2.0 Driver:
This procedure is highly recommended when a laptop computer will be used to
communicate with the ST-133. As stated before, we recommend the SIIG, Inc. USB 2.0
PC Card, Model US2246 if USB 2.0 is not native to the laptop's motherboard. To reduce
the instances of data overruns and serial violations, the OrangeUSB USB 2.0 Host
Controller installed for the SIIG card, should be replaced by the appropriate Microsoft
driver (Windows XP or Windows Vista, depending on the laptop's operating system.)
Note: This procedure may also be performed for desktop computers that use the
Orange Micro 70USB90011 USB2.0 PCI.
1. Download and install Microsoft Service Pack 3 (for Windows XP) if theservice pack
has not been installed.
2. From the Windows Start menu, select Settings|Control Panel.
3. Select System and then System Properties.
4. Select the Hardware tab and click on Device Manager button.
5. Expand Universal Serial Bus Controllers.
6. Right-mouse click on OrangeUSB USB 2.0 Host Controller and select
Properties.
7. On the Driver tab, click on the Update Driver… button. You may have to wait a
minute or so before you will be allowed to click on the button.
8. When the Upgrade Device Driver wizard appears, click on Next. Select the
Search for a suitable driver … radio button.
9. On the next screen select the Specify a location checkbox.
10. Browse and select the location. Click on OK.
11. In the Driver Files Search Results window, check the Install one of the other
drivers check box.
12. Select the NEC PCI to USB Enhanced Host Controller B1 driver. Click on
Next and the installation will take place. When the Completing the Upgrade
Device Driver wizard window appears, click on Finish. You will then be given
the choice of restarting the computer now or later. According to the window text, the
hardware associated with the driver will not work until you restart the computer.
34 PI-SCX System Manual Version 2.E
The following information assumes that:
You have verified that the host computer meets the required specifications
for USB 2.0 communication with the PI-SCX system (see page 28).
A USB 2.0 board and its driver are installed in the host computer.
The ST-133 has an installed USB 2.0 Interface Control module.
You have already installed the WinX software (versions 2.5.15 and higher).
Versions 2.5.15 and higher automatically install the driver and INF files
required to support the USB 2.0 Interface Control module.
Windows
Version
USB INF
Filename
Located in
"Windows"\Inf
directory*
USB Properties DLL
Located in
"Windows"\System32
directory
USB Device Driver Name
Located in
"Windows”\System32\Drivers
directory
Windows® XP
and Windows
Vista® (32-bit)
rsusb2k.inf (in
Windows.inf, for
example)
apausbprop.dll (in
Windows\System32,
for example)
apausb.sys (in
Windows\System32\Drivers,
for example)
* The INF directory may be hidden.
To Install the Princeton Instruments USB2 Interface:
1. Before installing the Princeton Instruments USB2 Interface, we recommend that
you defragment the host computer's hard disk. This operation reduces the time the
computer spends locating files. Typically, the "defrag" utility "Disk
Defragmenter" can be accessed from the Windows® Start menu and can usually
accessed from the Programs/Accessories/System Tools subdirectory.
2. After defragmenting the hard disk, turn off the computer and make the USB cable
connections between the host computer and the ST-133. Then, turn the ST-133 on
before turning on the host computer.
3. At bootup, Windows will detect the Princeton Instruments USB2 Interface
hardware (i.e., the USB 2.0 Interface Control module). You may be prompted to
enter the directory path(s) for the apausbprop.dll and/or the apausb.sys file(s),
either by keyboard entry or by using the browse function.
If you selected AUTO PCI during the application software installation, WinX
automatically put the required INF, DLL, and USB driver file in the "Windows"
directories shown below. Refer to the Table 2 for the file locations.
Table 2. USB Driver Files and Locations
Chapter 4 System Setup 35
Caution
Caution
Caution
Connecting the Interface (Controller-Computer) Cable
TAXI® Cable (6050-0148-CE)
Turn the Controller power OFF (OFF = 0, ON = |) and the Computer power OFF before
connecting or disconnecting the Controller-Computer (TAXI) cable.
To Connect the TAXI Cable:
1. Verify that the Controller power is OFF.
2. Verify that the Computer power is OFF.
3. Connect one end of the TAXI
host computer.
4. Tighten down the screws to lock the connector in place.
5. Connect the other end of the cable to the "Serial Com" port on the rear of the
Controller.
6. Tighten down the screws to lock the connector in place.
USB 2.0 Cable (6050-0494)
Turn the Controller power OFF (OFF = 0, ON = |) and the Computer power OFF before
connecting or disconnecting the Controller-Computer (TAXI) cable.
cable to the 9-pin port on the Interface card in the
To Connect the USB 2.0 Cable:
1. Verify that the Controller power is OFF.
2. Verify that the Computer power is OFF.
3. Connect one end of the USB cable to the USB port on the host computer.
4. Connect the other end of the cable to the USB 2.0 port on the rear of the
Controller.
Connecting the Camera-Controller Cable
Turn the Controller power OFF (OFF = 0, ON = |) before connecting or disconnecting the
Camera-Controller cable.
To Connect the Camera-Controller Cable (6050-0484):
1. Verify that the Controller power is OFF.
2. Connect male end of the Camera-Controller cable to the "Detector" port on the
back of the Controller.
3. Move the slide latch over to lock the connector in place.
4. Connect the female end of the cable to the Camera.
5. Move the slide latch over to lock the connector in place.
36 PI-SCX System Manual Version 2.E
Entering the Default Camera System Parameters into WinX
(WinView/32, WinSpec/32, or WinXTest/32)
Software changes implemented in WinX version 2.15.9.6 affected the way in which
default parameters were entered for camera systems. Therefore, two sets of instructions
are included. Follow the instructions appropriate to the software version that you
installed. Note that these instructions assume that you have performed the computer
interface installation.
WinX Versions 2.5.19.6 and later
1. Make sure the ST-133 is connected to the host computer and that it is turned on.
2. Run the WinX application. The Camera Detection wizard will automatically run if
this is the first time you have installed a Princeton Instruments WinX application
(WinView/32, WinSpec/32, or WinXTest/32) and a supported camera. Otherwise, if
you installing a new camera type, click on the Launch Camera Detection Wizard… button on the Controller/CCD tab to start the wizard.
3. On the Welcome dialog (Figure 11), leave the checkbox unselected and click on
Next.
Figure 11. Camera Detection Wizard - Welcome dialog
4. Follow the instructions on the dialogs to perform the initial hardware setup: this
wizard enters default parameters on the Hardware Setup dialog tabs and gives you an
opportunity to acquire a test image to confirm the system is working.
WinX Versions before 2.5.19.6: Run RSConfig.exe
1. Make sure the ST-133 is connected to the host computer and that it is turned on.
2. Run RSConfig from the Windows|Start|Programs|PI Acton menu or from the
directory where you installed WinView, WinSpec, or WinXTest.
3. When the RSConfig dialog (Figure 12) appears, you can change the camera name
to one that is more specific or you can keep the default name "Camera1". When
you have finished, click on the Done button.
Note: If the first camera in the list is not the "Princeton Style (USB2)", you will
need to edit the PVCAM.INI file created by RSConfig. See the instructions in
"Demo, High Speed PCI, and PCI(Timer) are Choices on Hardware
Wizard:Interface dialog (Versions 2.5.19.0 and earlier)", page 73.
Chapter 4 System Setup 37
Figure 12. RSConfig dialog
4. Open the WinX application and, from Setup|Hardware…, run the Hardware
Setup wizard.
5. When the PVCAM dialog (Figure 13) is displayed, click in the Yes radio button,
click on Next and continue through the wizard. After the wizard is finished, the
Controller/Camera tab card will be displayed with the Use PVCAM checkbox
selected. You should now be able to set up experiments and acquire data.
Figure 13. Hardware Setup wizard: PVCAM dialog
6. Run the software in focus mode to verify communication between the ST-133
and the host computer.
Making the Circulator-Camera Connections
1. Set up the coolant circulator according to the directions in the user manual for that
equipment. Do not apply power to the circulator until directed to do so.
2. Make the hose connections between the circulator and the camera. For best cooling
performance, the tubing should be no longer than necessary.
PI-SCX:1300 Use 1/4", thin-wall plastic tubing. Be sure the tubing is properly
secured at both ends. Note that the ports on this camera use a ferrule-less quickdisconnect method of securing the tubing and that both the camera's valve body
and the fitting insert include automatic shutoff to prevent coolant leaks when
disconnected. For best cooling, connect the inflow and outflow tubing to the ports as indicated in Figure 14.
38 PI-SCX System Manual Version 2.E
To Secure the Tubing:
a. Remove the retaining nuts for the fitting and slide them over the outside of
the plastic tubing.
b. Slide the tubing over the barb on the fitting.
c. Slide the retaining nuts to the end of the tubing and tighten them to the
threads of the fitting.
d. Insert the fitting into the appropriate valve body (Inlet or Outlet) until it
clicks.
Because of the automatic shutoffs, disconnecting the coolant supply is done by
simply depressing the release tabs (Figure 14) and removing the fittings.
Reconnecting the supply is done by reinserting each fitting into the appropriate
valve body until you hear a click.
Figure 14. PI-SCX:1300 Camera Back Panel
PI-SCX:4300 Use 3/8 I.D. plastic tubing. Be sure the tubing is properly
secured with hose clamps at both ends. Note that either of the camera’s coolant
ports can function as the inlet or outlet. We advise installing an in-line flow
meter so that the flow rate can be monitored.
PI-SCX:4096 Use 3/8 I.D. plastic tubing. Be sure the tubing is properly
secured with hose clamps at both ends. Note that either of the camera’s coolant
ports can function as the inlet or outlet. We advise installing an in-line flow
meter so that the flow rate can be monitored.
Chapter 4 System Setup 39
Caution
Do not disturb the Vacuum Pumping fitting or the Indium-tipped Vacuum Seal on the
PI-SCX:4300. Damage to the vacuum fittings and seals could void your warranty. Refer
servicing to qualified personnel and contact the factory Customer Support for guidance.
Detailed information about the vacuum system is provided in Appendix C.
3. Fill the circulator with a mixture of 50% ethylene glycol and 50% water.
Service Notes
There are no user-serviceable components inside the ST-133 or the fiberoptic taper camera.
Refer servicing to qualified personnel and contact the factory Customer Support for guidance.
Contact information is provided in the Warranty & Service section of the manual.
Note: Systems are now shipped with fittings that allow you to select
either a right angle or straight port configuration.
Figure 15. PI-SCX:4300 and PI-SCX:4096 Camera Back Panel
40 PI-SCX System Manual Version 2.E
This page intentionally left blank.
Chapter 5
Figure 16. Block Diagram
of Light Path in System
Operation
Introduction
Once the PI-SCX camera has been installed, camera operation is basically
straightforward. In most applications you simply establish optimum performance using
the Focus mode (WinView/32 or WinSpec/32), set the target detector temperature, wait
until the temperature has stabilized at the set temperature (see the "Setting the
Temperature" section in this chapter), and then do actual data acquisition in the Acquire
mode. Additional considerations regarding experiment setup and equipment
configuration are addressed in the software manual.
During data acquisition,
the CCD array is exposed
to a source and charge
accumulates in the pixels.
After the defined
exposure time, the
accumulated signal
is readout of the
array, digitized, and
then transferred to
the host computer.
Upon data transfer,
the data is data is
displayed and/or stored via the application
software. This sequence is illustrated by the block
diagram shown in Figure 16.
Whether or not the data is displayed and/or stored
depends on the data collection operation (Focus or
Acquire) that has been selected in the application
software. In WinView and WinSpec, these
operations use the Experiment Setup parameters to
establish the exposure time (the period when signal of interest is allowed to accumulate
on the CCD). As might be inferred from the names, Focus is more likely to be used in
setting up the system (see the "First Light" discussions) and Acquire is then used for the
collection and storage of data. Briefly:
In Focus mode, the number of frames and accumulations settings are ignored. A
single frame is acquired and displayed, another frame is acquired and overwrites the
currently displayed data, and so on until Stop is selected. Only the last frame
acquired before Stop is selected can be stored. When Stop is selected, the File Save
function can be used to save the currently displayed data. This mode is particularly
convenient for familiarization and setting up. For ease in focusing, the screen refresh
rate should be as rapid as possible, achieved by operating with axes and cross-
sections off, and with Zoom 1:1 selected.
41
42 PI-SCX System Manual Version 2.E
In Acquire mode, every frame of data collected can be automatically stored (the
completed dataset may include multiple frames with one or more accumulations).
This mode would ordinarily be selected during actual data collection. One limitation
of Acquire mode operation is that if data acquisition continues at too fast a rate for it
to be stored, data overflow will eventually occur. This could only happen in Fast
Mode operation.
The remainder of this chapter is organized to first talk about the system on/off sequences.
Then "First Light" procedures for imaging and spectroscopy applications follow: these
procedures provide step-by-step instruction on how to initially verify system operation.
The last three sections discuss factors that affect exposure, readout, and digitization of the
incoming signal. By understanding these factors and making adjustments to software
settings you can maximize signal-to-noise ratio. For information about synchronizing
data acquisition with external devices, please refer to Chapter 6,Advanced Topics.
System On/Off Sequences
If your system is configured for the USB 2.0 communication interface, you must follow
the system on/off sequences as stated below. These sequences ensure that communication
is established and maintained between the camera and the host computer:
1. The PI-SCX camera must be powered ON before WinView/32 or WinSpec/32 is
opened to ensure communication between the camera and the computer. If WinView
or WinSpec is opened and the PI-SCX is not powered ON, many of the functions will
be disabled and you will only be able to retrieve and examine previously acquired
and stored data. You must close WinView or WinSpec, power the camera ON, and
reopen WinView or WinSpec before you can set up experiments and acquire new
data.
2. WinView/32 or WinSpec/32 must be closed before powering the camera OFF. If you
power the camera OFF before closing WinView or WinSpec, the communication link
with the camera will be broken. You can operate the program in a playback mode
(i.e., examine previously acquired data) but will be unable to acquire new data until
you have closed WinView or WinSpec, powered the camera ON, and then re-opened
WinView or WinSpec.
Chapter 5 Operation 43
First Light
The following paragraphs provide step-by-step instructions for verifying the operation of
your PI-SCX system. The intent of this simple procedure is to help you gain basic
familiarity with the operation of your system and to show that it is functioning properly.
The procedure does not require an X-ray source. Once basic familiarity has been
established, then operation with other operating configurations, ones with more complex
timing modes, can be performed.
Once the PI-SCX system has been installed, operation of the camera is basically
straightforward. In most applications you simply establish optimum performance using
the Focus mode (WinView/32), select full frame, set the target camera temperature, and
watch the dark charge decrease as the CCD temperature approaches the set temperature.
Cabling
If the system cables haven’t been installed, connect them as instructed in Chapter 4.
Getting Started
1. If the system is water-cooled, double check that the circulator is filled with a
50:50 mixture of ethylene glycol and water and that the hose connections are
secure. When satisfied that these requirements are met, do the following.
Turn on the circulator. The circulator will power up and begin pumping
coolant through the camera. See the circulator instruction manual for detailed information.
Inspect the coolant hose connections to be sure there are no leaks.
Turn on the refrigeration, if this feature is available, and set the coolant
temperature (+5° for PI-SCX:4096 with ≤1.5:1 ratio and PI-SCX:4300, 0°
for PI-SCX:4096 with >1.5:1 ratio, and +25°C for other PI-SCX models,
unless otherwise specified). The compressor will start and cooldown will
begin.
2. Verify that the line-voltage setting of the ST-133 is correct for the available
power and switch ON the ST-133.
3. Turn on the power at the computer and start the application software
(WinView/32, for example).
Setting the Parameters
Note: The following procedure is based on WinView/32: you will need to modify it if
you are using a different application. Basic familiarity with the WinView/32 software is
assumed. If this is not the case, you may want to review the software manual or have it
available while performing this procedure.
Set the software parameters as follows:
Environment dialog (Setup|Environment): Set the DMA Buffer size to
32 Mbytes (96 Mbytes for PI-SCX:4096), reboot the computer for this memory
allocation to be activated, and then restart WinView.
44 PI-SCX System Manual Version 2.E
Controller/Camera tab (Setup|Hardware): Controller and Camera parameters
should be set automatically to the proper values for your system. However, you
can click on the Load Defaults From Controller button on this tab to load the
default settings.
Use PVCAM: If you are using the USB 2.0 interface, verify that the box
is checked.
Note: This check box is not present on software versions 2.5.19.6 and
higher.
Controller type: ST-133
Controller version: 4 or higher
Camera type: The array for your camera will be
Interface tab (Setup|Hardware): Verify that the High Speed PCI is selected.
Note: This tab is not available if you are using the USB 2.0 interface.
Cleans and Skips tab (Setup|Hardware): Default
Camera Temperature (Setup|Detector Temperature…):
-30C for air-cooled PI-SCX:1300
-35°C for PI-SCX:1300 with +25° water
-50°C for PI-SCX:4300 and PI-SCX:4096 (≤ 1.5:1 fiber ratio) with +5°
water
-45°C for PI-SCX:4096 (>1.5:1 fiber ratio) with +0° water
The temperature should drop steadily, reaching the set temperature in about
60 minutes (typical). At that point the green Temp Lock LED on the rear of
the ST-133 will light and there will be a locked indication at the computer
monitor, indicating that temperature lock has been established. Some
overshoot may occur that may cause temperature lock to be briefly lost and
then quickly re-established. If you are reading the actual temperature
reported by the application software, there may be a small difference
between the set and reported temperature when lock is established. This is
normal and does not indicate a system malfunction. Once lock is established,
the temperature will be stable to within ±0.05°C.
Note: If you are using the USB 2.0 interface, the Detector Temperature
dialog will not display temperature information while you are acquiring data.
Experiment Setup Main tab (Acquisition|Experiment Setup…):
Exposure Time: 100 ms
Accumulations & Number of Images: 1
CCD Readout: Use Full Chip
Timing Mode: Free Run
Shutter Control: Normal
Safe Mode vs. Fast Mode: Fast Mode
General tab (Display|Layout…): Select Horizontal and Vertical Cross
Sections.
Acquiring Data
If you are using WinView/32 and the computer monitor for focusing, select Focus from
the Acquisition menu. Successive images will be sent to the monitor as quickly as they
are acquired. Since no X-ray source is being used, the acquired images will be of the
camera's dark charge. As the CCD temperature decreases, the image will get darker as the
dark charge decreases. Signal changes may be more easily seen on the cross-section
views.
Figure 17 shows the kind of image data you might see in WinView. The horizontal and
vertical cross sections have been turned on via the General tab card (Display|Layout).
Figure 17. Example of WinView Data Acquired from First Light Procedure
Because the time to acquire and read out an image varies directly with the size of the
CCD, the observed frame rate will vary greatly depending on the CCD installed. With a
short exposure time, it is not uncommon for the frame readout time to be significantly
longer than the exposure time: for a Fairchild 4096x4096 array, a full frame readout at
1 MHz takes approximately 17 seconds.
This completes First Light. If the system functioned as described, you can be reasonably sure
it has arrived in good working order. In addition, you should have a basic understanding of
how the system hardware is used. A recommended procedure for powering down the camera
is provided in the next section. Other topics, which could be quite important in certain
situations, are discussed in the following chapters. See the appropriate application software
manual for information on using the software to control the system.
46 PI-SCX System Manual Version 2.E
Power Down Procedure
It is recommended that the camera warm-up be carefully controlled. Proper warm-up will
be achieved by proceeding as follows:
1. While running WinView, gradually bring the camera temperature up in 10°C
steps from its current lock temperature to ambient temperature. To accomplish
this, open the Detector Temperature dialog, set the target temperature to 10°C
above the current lock temperature, wait until lock is reached, and then raise it
again in 10°C steps until ambient temperature is reached.
2. Set the ST-133 Controller’s front-panel Power switch to OFF.
3. Set the circulator’s refrigeration On/Off switch to OFF.
4. Set the circulator’s main power switch to OFF. This completes the warm-up
procedure.
Exposure and Signal
Introduction
The following topics address factors that can affect the signal acquired on the CCD array.
These factors include array architecture, exposure time, CCD temperature, dark charge,
and saturation.
CCD Array Architecture
Charge coupled devices (CCDs) can be roughly thought of as a two-dimensional grid of
individual photodiodes (called pixels), each connected to its own charge storage “well.”
Each pixel senses the intensity of light falling on its collection area, and stores a
proportional amount of charge in its associated “well.” Once charge accumulates for the
specified exposure time (set in the software), the pixels are read out serially.
CCDs are rugged and compact: unintensified, uncoated CCDs can withstand direct
exposure to relatively high light levels, magnetic fields and RF radiation. They are easily
cooled and can be precisely thermostated to within a few tens of millidegrees.
Because CCD arrays, like film and other media, are always sensitive to light, light must
not be allowed to fall on the array during readout (with a few exceptions). Unintensified
full-frame CCD cameras like the PI-SCX cameras may require a customer-provided
X-ray shutter to prevent light from reaching the CCD during readout.
Chapter 5 Operation 47
Setting the Temperature
The temperature of the CCD array is set
through software. With WinView/32, you
enter and set the target temperature after
selecting Detector Temperature from the
Setupmenu.
Exposure Time
Exposure time (set on the Experiment
Setup|Main tab) is the time between Start
Acquisition and Stop Acquisition commands sent
by the application software to the camera. In
combination with triggers, these commands
control when continuous cleaning of the CCD
stops and when the accumulated signal will be
readout. The continuous cleaning prevents
buildup of dark current and unwanted signal prior
to the X-ray pulse. At the end of the exposure
time, the CCD is readout and cleaning starts
again.
Because PI-SCX cameras do not incorporate an
internal shutter, some signal may accumulate on
the array while it is being readout. This
continuous exposure of the array during readout
may result in some smearing. However,
exposures that are significantly longer than the readout time can be performed without a
shutter, as the amount of smearing will be low.
If smearing or other factors require a shutter, the NOT SCAN or the SHUTTER signal at
the ST-133's output can be used to control a customer-supplied external X-ray
shutter. By using one of the signals to synchronize the shutter operation with exposure,
the CCD can be read out in darkness. Alternatively, the X-ray source can be interrupted
elsewhere in the system while readout is taking place.
CCD Temperature
Each PI-SCX camera contains a Peltier-effect thermoelectric cooler that cools the CCD.
This cooling, which may be assisted by an internal fan or by liquid coolant circulation,
reduces the amount of dark charge that is generated on the array. A thermal sensing diode
attached to the cooling block of the camera monitors the array temperature and the
current temperature is reported back to the operating software. The operating (or target)
temperature is settable in software. Ambient temperature, cooling type, and CCD array
size all affect the time required to reach and stabilize at the appropriate operating
temperature (see Table 8, page 87, for typical cooldown temperatures).
Temperature Stabilization
After the system begins cooling, it takes about 60 minutes for the CCD to reach its preset
temperature. Because the control loop is designed to achieve temperature lock as quickly as
possible, overshoot may occur. If this happens, temperature lock will be briefly indicated and
then discontinue during the overshoot. However, the lock indication will be quickly restored
48 PI-SCX System Manual Version 2.E
WARNING!
as stable control is re-established. This is normal behavior and should not be a cause for
concern. Once temperature lock is established, the temperature is thermostated to within
±0.05°C. The controller is equipped with an LED that indicates temperature lock: this
indicator will change to green to indicate lock.
Notes:
1. The time to reach temperature lock is affected by CCD array size and the ambient
temperature. Typically, the larger the array or the warmer the ambient temperature,
the longer the time to reach lock. If the ambient temperature is above +23°C,
temperature lock may not be achievable at the lowest specified temperature for your
camera. Temperature regulation does not reach its ultimate stability for at least 30
minutes after lock is established.
2. If you are using the USB 2.0 interface, the Detector Temperature dialog will not
display temperature information while you are acquiring data.
Dark Charge
Dark charge (or dark current) is the thermally induced buildup of charge in the CCD over
time. Even with the light into the camera completely blocked, the CCD will collect a dark
charge pattern, dependent on the exposure time and camera temperature. The longer the
exposure time and the warmer the camera, the larger and less uniform this background
will appear. Thus, to minimize dark-charge effects, you should set the camera
temperature at the lowest CCD temperature within the recommended range for your
camera (see Table 8, on page 87).
The statistical noise associated with dark charge is known as dark noise. Dark charge
values vary widely from one CCD array to another and are exponentially temperature
dependent. At the typical operating temperature of a thermoelectrically-cooled camera,
dark charge is reduced by a factor of ~2 for every 6-7 degree reduction in temperature. In
the case of cameras such as the PI-SCX cameras, which have MPP type arrays, the
average dark charge is extremely small. However, the dark-charge distribution is such
that a significant number of pixels may exhibit a much higher dark charge, limiting the
maximum practical exposure.
Notes:
1. Do not be concerned about either the DC level of this background or its shape unless
it is very high, i.e., > 1000 counts with 16-bit ADC. What you see is not noise. It is a
fully subtractable readout pattern. Each CCD has its own dark charge pattern, unique
to that particular device. Simply acquire and save a dark charge “background image”
under conditions identical to those used to acquire the “actual” image. Subtracting the
background image from the actual image will significantly reduce dark-charge
effects.
2. Offset and excess noise problems are more likely to occur if the controller and
camera weren’t calibrated and tested as a system at the factory.
If you observe a sudden change in the baseline signal you may have excessive humidity
in the camera's vacuum enclosure. Immediately turn off the system. Refer to Appendix C
and contact Princeton Instruments Customer Support for information on how to refresh
the vacuum. See page 108 for contact information.
Chapter 5 Operation 49
Introduction
After the exposure time has elapsed, the charge
accumulated in the array pixels needs to be read
out of the array, converted from electrons to
digital format, and transmitted to the application
software where it can be displayed and/or
stored. Readout begins by moving charge from
the CCD image area to the shift register. The
charge in the shift register pixels, which
typically have twice the capacity of the image
pixels, is then shifted into the output node and
then to the output amplifier where the electrons
are grouped as electrons/count. This result
leaves the CCD and goes to the preamplifier
where gain is applied.
WinView and WinSpec allow you to specify the
Figure 18. Array Terms for a
CCD with a Single Output
Amplifier
Saturation
When signal levels in some part of the image are very high, charge generated in one pixel
may exceed the "well capacity" of the pixel, spilling over into adjacent pixels in a process
called "blooming." In this case a more frequent readout is advisable, with signal
averaging to enhance S/N (Signal-to-Noise ratio) accomplished through the software.
For signal levels low enough to be readout-noise limited, longer exposure times, and
therefore longer signal accumulation in the CCD, will improve the S/N ratio
approximately linearly with the length of exposure time. There is, however, a maximum
time limit for on-chip averaging, determined by either the saturation of the CCD pixels
by the signal or the loss of dynamic range due to the buildup of dark charge in the pixels
Readout
type of readout (full frame or binned), the output amplifier (if dual amplifiers are
available), and the gain (the number of electrons required to generate an ADU).
Full Frame Readout
In this section, a simple 6 4 pixel CCD is used to demonstrate how charge is shifted and
digitized. Full frame readout, for full frame CCDs, reads out the entire CCD surface at
the same time.
50 PI-SCX System Manual Version 2.E
The upper left drawing in Figure 19
represents a CCD after exposure but
before the beginning of readout. The
capital letters represent different
amounts of charge, including both
signal and dark charge. This section
explains readout at full resolution,
where every pixel is digitized
separately.
Readout of the CCD begins with the
simultaneous shifting of all pixels
one row toward the “shift register,”
in this case the row at the top. The
shift register is a single line of
pixels along one edge of the CCD,
not sensitive to light and used for
readout only. Typically the shift
register pixels hold twice as much
charge as the pixels in the imaging
area of the CCD.
After the first row is moved into the
shift register, the charge now in the
shift register is shifted toward the
Figure 19. Full Frame at Full Resolution
output node, located at one end of the shift register. As each value is “emptied” into the
output it is digitized. Only after all pixels in the first row are digitized is the second row
moved into the shift register. The order of shifting in our example is therefore A1, B1,
C1, D1, A2, B2, C2, D2, A3 ....
After charge is shifted out of each pixel the remaining charge is zero, meaning that the
array is immediately ready for the next exposure.
A subsection of the CCD can be read out at full resolution, sometimes dramatically increasing
the readout rate while retaining the highest resolution in the region of interest (ROI). Note that
some overhead time is required to shift out and discard the unwanted pixels.
Binning
Binning is the process of adding the data from adjacent pixels together to form a single
pixel (sometimes called a super pixel), and it can be accomplished in either hardware or
software. Rectangular groups of pixels of any size may be binned together, subject to
some hardware and software limitations.
Hardware Binning
Hardware binning is performed on the CCD array before the signal is read out of the
output amplifier. For signal levels that are readout noise limited this method improves
S/N ratio linearly with the number of pixels grouped together. For signals large enough to
render the camera photon shot noise limited, the S/N ratio improvement is roughly
proportional to the square root of the number of pixels binned.
Binning also reduces readout time and the burden on computer memory, but at the
expense of resolution. Since shift register pixels typically hold only twice as much charge
Chapter 5 Operation 51
as image pixels, the binning of large sections may result in saturation and “blooming”, or
spilling of charge back into the image area.
Figure 20 shows an example of 2 2 binning. Each pixel of the image displayed by the
software represents 4 pixels of the CCD array. Rectangular bins of any size are possible.
Figure 20. 2 × 2 Binning
Software Binning
One limitation of hardware binning is that the shift register pixels and the output node are
typically only 2-3 times the size of imaging pixels. Consequently, if the total charge binned
together exceeds the capacity of the shift register or output node, the data will be lost.
This restriction strongly limits the number of pixels that may be binned in cases where there
is a small signal superimposed on a large background, such as signals with a large
fluorescence. Ideally, one would like to bin many pixels to increase the S/N ratio of the weak
peaks but this cannot be done because the fluorescence would quickly saturate the CCD.
The solution is to perform the binning in software. Limited hardware binning may be used
when reading out the CCD. Additional binning is accomplished in software, producing a
result that represents many more photons than was possible using hardware binning.
Software averaging can improve the S/N ratio by as much as the square root of the
number of scans. Unfortunately, with a high number of scans, i.e., above 100, camera 1/f
noise may reduce the actual S/N ratio to slightly below this theoretical value. Also, if the
light source used is photon-flicker (1/f noise) limited rather than photon shot-noise
limited, this theoretical signal improvement cannot be fully realized. Again, background
subtraction from the raw data is necessary.
52 PI-SCX System Manual Version 2.E
This technique is also useful in high light level experiments, where the camera is again
photon shot-noise limited. Summing multiple pixels in software corresponds to collecting
more photons, and results in a better S/N ratio in the measurement.
Output Amplifier Selection
The output amplifier amplifies the collected charge from the output node and outputs it as
electrons/count. Although Figure 18 shows an array with a single output node and
amplifier, some PI-SCX systems are available with dual output nodes and amplifiers (one
set at each end of the shift register). If your system has dual output amplifiers, you can
choose the output amplifier to be used (High Capacity or Low Noise) via WinView/32 or
WinSpec/32 on the Acquisition|Experiment Setup…|Main tab:
High Capacity amplifier: Provides a spectrometric well capacity that is
approximately 3 times the well capacity for the Low Noise amplifier selection.
High Capacity is suitable when you have intense light signals or signals with
high dynamic range.
Low Noise amplifier: Provides the highest sensitivity performance and is
suitable when you have weak signals.
Note: The choice of output amplifier and analog gain setting should be considered
together for the best signal capture.
Analog Gain Control
Note: A gain control switch is located on the rear of the PI-SCX:1300 camera but has
been deactivated. However, analog gain is changeable via the WinView/WinSpec
software (on the Acquisition| Experiment Setup…|ADC tab card). If your camera
is not designed for analog gain selection, these settings will not be accessible in the
software.
Analog gain control (a function of the preamplifier) is used to change the relationship
between the number of electrons acquired on the CCD and the Analog-to-Digital Units
(ADUs or counts) generated. In WinView/32 and WinSpec/32, the analog gain choices
vary depending on the CCD array and the number of output amplifiers:
Single Amplifier: Typically, three settings are available: Low 2x, Medium 1x,
and High 1/2x.
Dual Amplifier: Typically, three settings are available: Low 2x, Medium 1x, and
High 1/2x (Low-noise mode) or Low 4x, Medium 2x, and High 1x (Highcapacity mode).
The analog gain of the detector should generally be set so that the overall noise is ~1
count RMS. In most instances this will occur with the switch set to Medium. If the array
is a 1340 × 1300 configured with the low-noise output, Low will probably be a more
suitable Analog Gain setting. In situations where the A/D range exceeds that of the array,
it will generally be better to set the Analog Gain to High so that the signal can be spread
over as much of the A/D range as possible. This is a particularly important consideration
in absorbance measurements. Users who consistently measure low-level signals may
wish to select High, which reduces some sources of noise. Users who measure high-level
signals may wish to select Low to allow digitization of larger signals. Customized values
of gain can be provided. Contact Princeton Instruments for additional information.
Chapter 5 Operation 53
Example: The following descriptions assume that the actual incoming light level is
identical in all three instances. The numbers used illustrate the effect of changing an
analog gain setting and do not reflect actual performance: gain at the Low, Medium,
and High settings depends on the CCD installed and the amplifier selected.
Low requires four electrons to generate one ADU. Strong signals can be acquired
without flooding the CCD array. If the gain is set to Low and the images or spectra
appear weak, you may want to change the gain setting to Medium or High.
Medium requires two electrons to generate one ADU. If the gain is set to Medium
and the images or spectra do not appear to take up the full dynamic range of the CCD
array, you may want to change the gain setting to High. If the CCD array appears to
be flooded with light, you may want to change the setting to Low.
High requires one electron to generate one ADU and some noise sources are reduced.
Because fewer electrons are needed to generate an ADU, weaker signals can be more
readily detected. Lower noise further enhances the ability to acquire weak signals. If
the CCD array appears to be flooded with light, you may want to change the setting
to Medium or Low.
Note: The baseline level may require adjustment if you change the analog gain. See the
"ADC Offset" section on page 54 for more information.
Digitization
Introduction
After gain has been applied to the signal, the Analog-to-Digital Converter (ADC)
converts that analog information (continuous amplitudes) into a digital data (quantified,
discrete steps) that can be read, displayed, and stored by the application software. The
number of bits per pixel is based on both the hardware and the settings programmed into
the camera through the software (see "Readout", page 49).
Factors associated with digitization include the digitization rate and baseline signal.
Depending on the camera model, you may be able change the speed at which digitization
occurs and/or offset the baseline. These factors are discussed in the following paragraphs.
Digitization Rate
Depending on the PI-SCX system, the camera may contain single (1 MHz) or dual
digitizers (100 kHz/1 MHz or 50kHz/1 MHz). When a camera has a single digitizer, the
data can only be digitized at that speed. However, for cameras with dual digitization, you
have a choice of how quickly the data will be digitized (with optimum signal-to-noise
ratios at both readout speeds). Because the readout noise of CCD arrays increases with
the readout rate, it is sometimes necessary to trade off readout speed for high dynamic
range. The 1 MHz conversion speed is used for the fastest possible data collection and the
100 kHz or 50 kHz conversion speed is used where noise performance is the paramount
concern. Switching between the conversion speeds is completely under software control
for total experiment automation.
Note: If the camera has dual digitization, the ADC rate can be changed on the
Experiment Setup|ADC tab.
54 PI-SCX System Manual Version 2.E
WARNING!
ADC Offset
ADC offset (also known as baseline offset) provides another way of dealing with dark
charge (see "Dark Charge", page 48). By offsetting the baseline signal, much of the
background is ignored during conversion.
Offsetting the baseline is accomplished by adding a voltage to the signal to bring the A/D
output to a non-zero value, typically 50-100 counts. This offset value ensures that all the
true variation in the signal can really be seen and not lost below the A/D “0” value. Since
the offset is added to the signal, these counts only minimally reduce the range of the
signal from 65535 (16-bit A/D) to a value in the range of 50-100 counts lower.
Notes:
1. Do not be concerned about either the DC level of the baseline signal or its shape
unless it is very high (i.e., > 1000 counts with 16-bit ADC).
2. The ADC Offset can be adjusted by using the F and S Zero pots located on the rear
panel of the controller. If these pots are not present, the ADC Offset may be
software-adjustable on the Experiment Setup|ADC tab.
3. Do not adjust the offset values to zero or some low-level data will be missed.
If you observe a sudden change in the baseline signal you may have excessive humidity
in the camera's vacuum enclosure. TURN OFF THE SYSTEM IMMEDIATELY.
Contact Princeton Instruments Customer Support for information on how to refresh the
vacuum. See page 108 for contact information.
Chapter 6
Previous chapters have discussed setting up the
hardware and the software for basic operation.
This chapter discusses topics associated with
experiment synchronization (set up on the
Experiment Setup|Timing tab in WinView and
WinSpec).With the exception of Edge Trigger, the
topics are addressed in order of their appearance
on the Timing tab (see Figure 21).
"Timing Modes", the first topic, discusses
standard Timing Modes, Shutter Control, and
Edge Trigger. Also included under this topic is a
discussion of the EXT SYNC connector, the input
connector for a trigger pulse.
"Fast Mode and Safe Mode", the second topic,
discusses the Fast and the Safe speed modes. Fast
is used for real-time data acquisition and Safe is
used when coordinating acquisition with external
devices or when the computer speed is not fast
Figure 21. Timing tab
The chart to the right lists the timing mode
combinations (selected on the Experiment
Setup|Timingtab). Use this chart in
combination with the detailed descriptions in
this chapter to determine the optimal timing
configuration.
The basic timing modes are Free Run,
External Sync, and External Sync with
Continuous Cleans. These modes are
combined with the Shutter options to provide
the widest variety of timing modes for
precision experiment synchronization.
Mode
Shutter
Free Run
Normal
External Sync
Normal
External Sync
PreOpen
External Sync with
Continuous Cleans
Normal
External Sync with
Continuous Cleans
PreOpen
Table 3. Camera Timing Modes
Advanced Topics
Introduction
enough to keep pace with the acquisition rate.
"TTL Control", the final topic, discusses the TTL IN/Out connector on the rear of the
ST-133 and how commands can be sent via TTL levels to and from the ST-133.
Timing Modes
The shutter options available include Normal, PreOpen, Disable Opened or Disable
Closed. Disable simply means that the shutter will not operate during the experiment.
Disable closed is useful for making dark charge measurements or when no shutter is
present. PreOpen, available in the External Sync and External Sync with Continuous
55
56 PI-SCX System Manual Version 2.E
Free Run
In the Free Run mode the controller does not
synchronize with the experiment in any way. The
shutter opens as soon as the previous readout is
complete, and remains open for the exposure time, t
exp
.
Any External Sync signals are ignored. This mode is
useful for experiments with a constant light source.
Other experiments that can utilize this mode are high
repetition studies, where the number of shots that occur
during a single shutter cycle is so large that it appears to
be continuous illumination.
Other experimental equipment can be synchronized to
the ST-133 controller by using the output signal
(software-selectable SHUTTER or NOT SCAN on the
Hardware Setup|Controller Camera tab) from the
connector. Shutter operation and the NOT
SCAN output signal are shown in Figure 23.
Figure 22. Free Run Timing Chart,
part of the chart in Figure 28
Cleans modes, opens the shutter as soon as the ST-133 is ready to receive an External
Sync pulse. This is required if the time between the External Sync pulse and the event is
less than a few milliseconds, the time it takes the shutter to open.
The shutter timing is shown in the timing diagrams that follow. Except for Free Run,
where the modes of shutter operation are identical, both Normal and PreOpen lines are
shown in the timing diagrams and flowchart.
The timing diagrams are labeled indicating the exposure time (t
), shutter compensation
exp
time (tc), and readout time (tR). Note that if there is no shutter selected in the software, the
shutter compensation time (the time required to close a mechanical shutter) will be
approximately 0 ms.
External Sync
In this mode all exposures are synchronized to an external source via signal input to the
Ext Sync BNC on the back of the ST-133. To ensure synchronization, the trigger edge
(negative- or positive-going) of the Ext Sync signal must be identified in the application
software (in WinView and WinSpec, this is done on the Experiment Setup|Timing
tab). As shown in the flowchart, Figure 24, External Sync mode can be used in
combination with Normal or PreOpen Shutter operation. In Normal Shutter mode, the
controller waits for an External Sync pulse and then opens the shutter for the
Figure 23. Free Run Timing Diagram
Chapter 6 Advanced Topics 57
programmed exposure period. As soon as the exposure is complete, the shutter closes and
the CCD array is read out.
Because the external shutter requires a finite amount of time to open completely (shutter
open time may be 5-28 ms depending on the shutter), the External Sync pulse trigger
edge provided by the experiment should precede the actual signal by at least that much
time. If not, the shutter will not be open for the duration of the entire signal, or the signal
may be missed completely.
Also, since the amount of time from initialization of the experiment to the first External
Sync pulse trigger edge is not fixed, an accurate background subtraction may not be
possible for the first readout. In multiple-shot experiments this is easily overcome by
simply discarding the first frame.
In the PreOpen Shutter mode, on the other hand, shutter operation is only partially
synchronized to the experiment. As soon as the controller is ready to collect data, the
shutter opens. Upon arrival of the first External Sync pulse trigger edge at the ST-133, the
shutter remains open for the specified exposure period, closes, and the CCD is read out.
As soon as readout is complete, the shutter reopens and waits for the next frame.
Figure 24. Chart Showing Two External Sync Timing Options
The PreOpen mode is useful in cases where an External Sync pulse trigger edge cannot
be provided 5-28 ms (shutter open time) before the actual signal occurs. Its main
drawback is that the CCD is exposed to any ambient light while the shutter is open
between frames. If this ambient light is constant and the triggers occur at regular
intervals, this background can also be subtracted, providing that it does not saturate the
CCD. As with the Normal Shutter mode, accurate background subtraction may not be
possible for the first frame.
Also note that, in addition to signal from ambient light, dark charge accumulates during
the “wait” time (tw). Any variation in the external sync frequency also affects the amount
of dark charge, even if light is not falling on the CCD during this time.
58 PI-SCX System Manual Version 2.E
Note: If EXT SYNC is still active (in Figure 25, this means that if it is still LOW) at the
end of the readout, the hardware may interpret this as a second sync pulse, and so on.
Figure 25. Timing Diagram for the External Sync Mode (- edge trigger)
External Sync with Continuous Cleans
Another timing mode available with the ST-133 controller is called Continuous Cleans.
In addition to the standard "cleaning" of the array, which occurs after the controller is
enabled, Continuous Cleans will remove any charge from the array until the moment the
External Sync pulse trigger edge is received.
Figure 26. Continuous Cleans Operation Flowchart
Once the External Sync pulse trigger edge is received, cleaning of the array stops as soon
as the current row is shifted, and frame collection begins: a delay time of up to one row
shift can be expected. With Normal Shutter operation the shutter is opened for the set
exposure time. With PreOpen Shutter operation the shutter is open during the continuous
cleaning, and once the External Sync pulse trigger edge is received, the shutter remains
open for the set exposure time, then closes. If the vertical rows are shifted midway when
the External Sync pulse trigger edge arrives, the pulse is saved until the row shifting is
Chapter 6 Advanced Topics 59
completed, to prevent the CCD from getting "out of step." As expected, the response
latency is on the order of one vertical shift time, from 1-30 µs depending on the array.
This latency does not prevent the incoming signal from being detected, since photogenerated electrons are still collected over the entire active area. However, if the signal
arrival is coincident with the vertical shifting, image smearing of up to one pixel is
possible. The amount of smearing is a function of the signal duration compared to the
single vertical shift time.
Note: If EXT SYNC is still active (in Figure 27, this means that if it is still LOW) at the
end of the readout, the hardware may interpret this as a second sync pulse, and so on.
The Experiment Setup|Timing tab allows you to choose Fast Mode or Safe Mode.
Figure 28 is a flowchart comparing the two modes. In Fast Mode operation, the ST-133 runs
according to the timing of the experiment, with no interruptions from the computer. In Safe
Mode operation, the computer processes each frame as it is received. The ST-133 cannot
collect the next frame until the previous frame has been completely processed.
Fast Mode operation is primarily for collecting "real-time" sequences of experimental data,
where timing is critical and events cannot be missed. Once the ST-133 is sent the Start
Acquisition command by the computer, all frames are collected without further intervention
from the computer. The advantage of this timing mode is that timing is controlled
completely through hardware. A drawback to this mode is that the computer will only
display frames when it is not performing other tasks. Image display has a lower priority, so
the image on the screen may lag several images behind. A video monitor connected to the VIDEO output will always display the current image. A second drawback is that a data
overrun may occur if the number of images collected exceeds the amount of allocated
RAM or if the computer cannot keep up with the data rate.
Safe Mode operation is primarily useful for experiment setup, including alignment and
focusing, when it is necessary to have the most current image displayed on the screen. It is
also useful when data collection must be coordinated through software with external devices
such as external shutters and filter wheels. As seen in Figure 28, in Safe Mode operation, the
computer controls when each frame is taken. After each frame is received, the camera sends
the Stop Acquisition command to the camera, instructing it to stop acquisition. Once that
frame is completely processed and displayed, another Start Acquisition command is sent from
the computer to the camera, allowing it to take the next frame. Display is therefore, at most,
only one frame behind the actual data collection.
60 PI-SCX System Manual Version 2.E
One disadvantage of the Safe Mode is that events may be missed during the experiment,
since the ST-133 is disabled for a short time after each frame.
Figure 28. Chart of Safe Mode and Fast Mode Operation
Chapter 6 Advanced Topics 61
TTL IN
Value
TTL IN
Value
1
1 5
16
2
2 6
32
3
4 7
64 4 8 8
128
TTL IN
Value
TTL IN
Value
1
High (1)
5 Low (0)
2
High (2)
6 Low (0)
3
Low (0)
7 Low (0)
4
Low (0)
8 Low (0)
TTL Control
Fully supported by WinView/WinSpec Version 2.5 when the communication
protocol is TAXI (PCI), this feature is not supported when the protocol is USB 2.0.
Introduction
Princeton Instruments' WinView and WinSpec software packages incorporates WinX32
Automation, a programming language that can be used to automate performing a variety
of data acquisition and data processing functions, including use of the TTL In/Out
functions. WinX32 Automation can be implemented in programs written in Visual Basic
or Visual C++. See the WinX32 documentation for more detailed information.
The TTL lines are made available through the TTL In/Out connector on the rear of the
ST-133 Controller. This connector provides 8 TTL lines in, 8 TTL lines out and an
input control line. Figure 29 illustrates the connector and Table 5 lists the signal/pin
assignments.
TTL In
The user controls the 8 TTL Input lines, setting them high (+5 V; TTL 1) or low (0 V;
TTL 0). When the lines are read, the combination of highs and lows read defines a
decimal number which the computer can use to make a decision and initiate actions as
specified in the user’s program. If a TTL IN line is low, its numeric value is 0. If a TTL
IN line is high, its numeric value is as follows.
This coding allows any decimal value from 0 to 255 to be defined. Thus, as many as 256
different sets of conditions can be specified, at the user’s discretion, using the TTL IN
lines. Any unused lines will default to TTL high (+5 V). For example, to define the
number three, the user would simply set the lines TTL IN 1 and TTL IN 2 both high
(+5 V). It would be necessary to apply TTL low to the remaining six lines because they
would otherwise default to TTL high as well.
Table 4 illustrates this coding for decimal values 0 through 7. Obviously this table could
easily be extended to show the coding for values all the way to 255.
Table 4. Bit Values with Decimal Equivalents:
1 = High,
0 = Low
Buffered vs. Latched Inputs
In controlling the TTL IN lines, users also have the choice of two input-line states,
buffered or latched. In the buffered state, the line levels must remain at the intended
levels until they are read. With reference to the preceding example, the high level at TTL
IN 1 and TTL IN 2 would have to be maintained until the lines are read. In the latched
state, the applied levels continue to be available until read, even if they should change at
the TTL In/Out connector.
This control is accomplished using the EN/CLK TTL input (pin 6). If EN/CLK is open or
high, buffered operation is established and the levels reported to the macro will be those
in effect when the READ is made. With reference to our example, if pin 6 were left
unconnected or a TTL high applied, TTL IN 1 and TTL IN 2 would have to be held high
until read. If, on the other hand, EN/CLK were made to go low while TTL IN 1 and TTL
IN 2 were high, those values would be latched for as long as EN/CLK remained low. The
levels actually present at TTL IN 1 and TTL IN 2 could then change without changing
the value that would be read by software.
TTL Out
The state of the TTL OUT lines is set from WinView/32 (or WinSpec/32). Typically, a
program (for example, a macro) monitoring the experiment sets one or more of the TTL
Outputs. Apparatus external to the PI-SCX system interrogates the lines and, on detecting
the specified logic levels, takes the action appropriate to the detected condition. The
coding is the same as for the input lines. There are eight output lines, each of which can
be set low (0) or high (1). The combination of states defines a decimal number as
previously described for the TTL IN lines.
Chapter 6 Advanced Topics 63
Pin #
Assignment
Pin #
Assignment
1
IN 1
14
IN 2
2
IN 3
15
IN 4
3
IN 5
16
IN 6
4
IN 7
17
IN 8
5
GND
18
GND
6
EN/CLK
19
Reserved
7
(future use)
20
GND
8
GND
21
OUT 2
9
OUT 1
22
OUT 4
10
OUT 3
23
OUT 6
11
OUT 5
24
OUT 8
12
OUT 7
25
GND
13
Reserved
Figure 29. TTL In/Out
Connector
Table 5. TTL In/Out Connector Pinout
Figure 30. TTL Diagnostics dialog
TTL Diagnostics Screen
Note that WinView/32 and WinSpec/32 provide a
TTL Diagnostics screen (located in WinView/32
under Hardware Setup|Diagnostics) that can
be used to test and analyze the TTL In/Out lines.
Hardware Interface
A cable will be needed to connect the TTL In/Out
connector to the experiment. The design will vary
widely according to each user’s needs, but a
standard 25-pin female type D-subminiature
connector will be needed to mate with the TTL
In/Out connector on the back of the ST-133. The
hardware at the other end of the cable will depend
entirely on the user’s requirements. If the
individual connections are made using coaxial
cable for maximum noise immunity
(recommended), the center conductor of the coax should connect to the proper signal pin
and the cable shield should connect to the nearest available ground (grounds are
conveniently provided at pins 5, 8, 18 and 20). Connector hardware and cables of many
different types are widely available and can often be obtained locally, such as at a nearby
Radio Shack® store. A list of possibly useful items follows. Note that, although the items
listed may be appropriate in many situations, they might not meet your specific needs.
25-pin female type D-subminiature solder type connector (Radio Shack
276-1548B).
®
part number
RG/58U coaxial cable.
64 PI-SCX System Manual Version 2.E
Shielded Metalized hood (Radio Shack part no 276-1536A).
BNC connector(s) type UG-88 Male BNC connector (Radio Shack part no 278-103).
Example
Suppose you needed to build a cable to monitor the line TTL OUT 1. One approach
would be to build a cable assembly as described in the following paragraphs. This
procedure could easily be adapted to other situations.
1. Begin with a 25-pin female type D-subminiature solder type connector (Radio Shack
part no 276-1548B). This connector has 25 solder points open on the back.
2. Referring to Table 5, note that pin 8 = GND and pin 9 = TTL OUT 1.
3. Using coaxial cable type RG/58U (6 feet length), strip out the end and solder the
outer sheath to pin 8 (GND) and the inner line to pin 9 (TTL OUT 1). Then apply
shielding to the lines to insulate them.
4. Mount the connector in a Shielded Metalized hood (Radio Shack part no
276-1536A).
5. Build up the cable (you can use electrical tape) to where the strain relief clamp holds.
6. Connect a BNC connector (UG-88 Male BNC connector) to the free end of the cable
following the instructions supplied by Radio Shack on the box (Radio Shack part no
278-103).
7. To use this cable, connect the DB-25 to the TTL In/Out connector on the back of the
ST-133.
8. To check the cable, start WinView/32 and open the TTL Diagnostics screen (located
in WinView under Hardware Setup|Diagnostics). Click the Write radio button.
Then click the Output Line 1 box. Next click the OK button to actually set TTL
OUT 1 high. Once you set the voltage, it stays until you send a new command.
9. Measure the voltage at the BNC connector with a standard voltmeter (red on the
central pin, black on the surrounding shielding). Before clicking OK at the TTL
Diagnostics screen you should read 0 V. After clicking OK you should read +5 V.
Note that adding a second length of coaxial cable and another BNC connector would be
straightforward. However, as you increase the number of lines to be monitored, it
becomes more convenient to consider using a multiple conductor shielded cable rather
than individual coaxial cables.
Chapter 7
Baseline Signal Suddenly Changes
Page 66
Camera Stops Working
Page 66
Camera1 (or similar name) in Camera Name field
Page 66
Changing the ST-133's Line Voltage and Fuses
Page 67
Controller Is Not Responding
Page 68
Cooling Troubleshooting
Page 69
Data Loss or Serial Violation
Page 70
Data Overrun Due to Hardware Conflict message
Page 70
Data Overrun Has Occurred message
Page 71
Demo is only Choice on Hardware Wizard:Interface dialog (Versions 2.5.19.0 and
earlier)
Page 71
Demo, High Speed PCI, and PCI(Timer) are Choices on Hardware
Wizard:Interface dialog (Versions 2.5.19.0 and earlier)
Page 73
Detector Temperature, Acquire, and Focus are Grayed Out (Versions 2.5.19.0 and
earlier)
Page 75
Error Creating Controller message
Page 76
Error Occurs at Computer Powerup
Page 76
WARNING!
WARNING!
Troubleshooting
Do not attach or remove any cables while the PI-SCX system is powered on.
Introduction
The PI-SCX system is designed to be highly reliable. Nevertheless, problems could
develop. In most instances, the best recourse will be to contact the factory for guidance.
Most often the system will be most reliably and quickly restored to service by identifying
the faulty component and returning it to the factory for service. The information in this
chapter is provided as an aid for gathering information that will assist in identifying the
source of the problem. The most likely problem with the camera would be gradual
deterioration of the vacuum to where temperature lock can no longer be achieved.
Do NOT attempt any disassembly of the camera. The camera sections have Indium
vacuum seals where they join. Attempting to disassemble the camera would damage
these seals. Once the integrity of a seal is compromised, there would be no alternative but
to return the camera to the factory for repair. Such a repair would not be covered by the
equipment Warranty.
The following issues have corresponding troubleshooting sections in this chapter.
65
66 PI-SCX System Manual Version 2.E
No CCD Named in the Hardware Wizard:CCD dialog
(Versions 2.5.19.0 and earlier)
Page 79
Program Error message
Page 79
Removing/Installing a Plug-In Module
Page 80
Securing the Detector-Controller Cable Slide Latch
Page 82
Serial violations have occurred. Check interface cable.
Page 83
Baseline Signal Suddenly Changes
If you observe a sudden change in the baseline signal, you may have excessive humidity
in the vacuum enclosure of the camera. Turn off the controller and have the camera
repumped before resuming normal operation. Contact the factory Customer Support
Dept. for information on how to refresh the vacuum. See page 108 for contact
information.
Camera Stops Working
Problems with the host computer system or software may have side effects that appear to
be hardware problems. If you are sure the problem is in the PI-SCX system hardware,
begin with these simple checks:
Turn off all AC power.
Verify that all cables are securely fastened and that all locking screws are in place.
Check for a burned-out fuse in the Controller power input module. For
information about changing a fuse, see "Changing the ST-133's Line Voltage
and Fuses" on page 67.
Correct any apparent problems and turn the system on.
If the system still does not respond, contact Customer Support.
Camera1 (or similar name) in Camera Name field
Figure 31. Camera1 in Camera Name Field
When a PVCAM-based camera is detected/selected during the Camera Detection wizard
(formerly the Hardware Setup wizard), a default name such as Camera1 will be shown in
the Detected Hardware table and will be entered in the Camera Name field on the
Setup|Hardware|Controller/CCD tab. Because this name is not particularly
descriptive, you may want to change it. Such a change is made by editing the
PVCAM.INI file that is generated by Camera Detection wizard (or by the RSConfig.exe
if you have a software version 2.5.19.0 or earlier).
Chapter 7 Troubleshooting 67
1. As shown in Figure 32, place the flat
side of a flat bladed screwdriver
parallel to the back of the Controller
and behind the small tab at the top of
the power module, and twist the
screwdriver slowly but firmly to pop
the module open.
Figure 32. Power Input Module
2. To change the voltage setting, roll the
selector drum until the setting that is
closest to the actual line voltage is
facing outwards.
WARNING!
WARNING!
To change the default Camera Name:
1. Using Notepad or a similar text editor, open PVCAM.INI, which is located in
the Windows directory (C:\Windows, for example). You should see entries like
the ones below.
[Camera_1]
Type=1
Name=Camera1
Driver=apausb.sys
Port=0
ID=523459
2. Change the "Name=" entry to something more meaningful for you (for example,
ST133USB - to indicate that this is a PVCAM-based system using an ST-133
with a USB 2.0 interface) and save the edited file.
[Camera_1]
Type=1
Name=ST133USB
Driver=apausb.sys
Port=0
ID=523459
3. The new camera name will now appear in the Camera Name field.
Changing the ST-133's Line Voltage and Fuses
The appropriate voltage setting for your country is set at the factory and can be seen on
the power input module. If your voltage source changes, you will need to change the
voltage setting and you may need to change the fuse configuration.
Use proper fuse values and types for the controller and camera to be properly protected.
To Change Voltage and Fuse Configuration:
Turn the Controller OFF and unplug the line cord before opening the power
input module.
68 PI-SCX System Manual Version 2.E
3. Confirm the fuse ratings by removing the
two white fuse holders. To do so, simply
insert the flat blade of the screwdriver
behind the front tab of each fuse holder
and gently pry the assembly out.
Figure 33. Fuse Holder
4. Refer to the Fuse/Voltage label (above or below the Power Module) to see which
fuses are required by the selected voltage. If Controller power switch is on the back
of the ST-133, the Fuse/Voltage label is located below the Power Module.
5. After inspecting and if necessary, changing the fuses to those required by the
selected voltage, reinstall the holders with the arrow facing to the right.
6. Close the power module and verify that the correct voltage setting is displayed.
7. Verify that the Controller power switch is in the OFF position and then plug the
power cord back into the power module.
Controller Is Not Responding
If this message pops up when you click on OK after selecting the "Interface Type" during
Hardware Setup (under the WinView/32 Setup menu), the system has not been able to
communicate with the Controller. Check to see if Controller has been turned ON and if
the interface card, its driver, and the interface cable have been installed.
If the Controller is ON, the problem may be with the interface card, its driver,
interrupt or address conflicts, or the cable connections.
If the interface card is not installed, close the application program and turn the
Controller OFF. Follow the interface card installation instructions in provided
with your interface card and cable the card to the SERIAL COM port on the rear
of the Controller. Then do a "Custom" installation of WinView/32 with the
appropriate interface component selected: "PCI Interface" or "ISA Interface",
depending on the interface card type. Be sure to deselect the interface component
that does not apply to your system.
Note: WinView/32 (versions 2.6.0 and higher) does not support the ISA
interface.
If the interface card is installed in the computer and is cabled to the SERIAL
COM port on the rear of the Controller, close the application program and turn
the Controller OFF. Check the cable connections and tighten the locking screws
if the connections are loose.
If the interface card was installed after WinView/32 has been installed, close that
application and do a "Custom" installation of it with the appropriate interface
component selected: "PCI Interface" or "ISA Interface", depending on the
interface card type. Be sure to deselect the interface component that does not
apply to your system.
Note: WinView/32 (versions 2.6.0 and higher) does not support the ISA
interface.
Chapter 7 Troubleshooting 69
Caution
Cooling Troubleshooting
Temperature Lock Cannot be Achieved or Maintained.
Operating a PI-SCX:1300 camera with coolant at a temperature colder than specified
could cause induced condensation in the electronics enclosure and possible catastrophic
damage to the camera. Damage resulting from this type of operation may void the
warranty.
Possible causes could include:
Ambient temperature greater than +23°C. If it is greater than +23°C, you will need to
cool the camera environment or raise the set temperature.
The vacuum has deteriorated and needs to be refreshed.
The connectors of the cable that interconnects the controller and the camera need to
be secured.
The target array temperature is not appropriate for your particular camera and CCD
array.
For cameras that are liquid-cooled or have liquid-assisted cooling, the coolant flow
rate may be insufficient due to a pinched coolant line, blockages, circulator power
problem, or pump failure. Check the flow rate and coolant temperature.
For a TE-cooled camera, the camera's internal temperature may be too high, such as
might occur if the operating environment is particularly warm or if you are
attempting to operate at a temperature colder than the specified limit. TE-cooled
cameras are equipped with a thermal-protection switch that shuts the cooler circuits
down if the internal temperature exceeds a preset limit. Typically, camera operation
is restored automatically in about ten minutes. Although the thermo-protection switch
will protect the camera, you are nevertheless advised to power down and correct the
operating conditions that caused the thermal-overload to occur.
Camera loses Temperature Lock
The internal temperature of the camera is too high. This might occur if the operating
environment is particularly warm or if you are trying to operate at a temperature colder
than the specified limit. If this happens, an internal thermal overload switch will disable
the cooler circuits to protect them. Typically, camera operation is restored in about ten
minutes. Although the thermal overload switch will protect the camera, users are advised
to power down and correct the operating conditions that caused the thermal overload to
occur. With some versions of the software, the indicated temperature when the camera is in thermal overload (thermal switch is in the cut-out state) is -120° C.
Gradual Deterioration of Cooling Capability
With time, there may be a gradual deterioration of the camera’s vacuum. This can affect
temperature performance such that it may be impossible to achieve temperature lock at
the lowest temperatures. In the kind of applications for which cooled CCD cameras are so
well suited, it is highly desirable to maintain the system’s lowest temperature
performance because lower temperatures result in lower thermal noise and better the
signal-to-noise ratio.
Vacuum deterioration occurs primarily as a result of material outgassing in the vacuum
chamber. Because outgassing normally diminishes with time, the rate of vacuum
70 PI-SCX System Manual Version 2.E
deterioration in new cameras may be faster than in old ones. For example, a camera that
has to be repumped after perhaps a year of operation, may not have to be pumped again
for several years. In any case you may notice a gradual deterioration in temperature
control performance indicative of vacuum deterioration. If this happens, the camera will
have to be repumped. Appendix C provides directions for repumping the vacuum. If you
have the appropriate equipment and personnel with the necessary expertise available, you
may wish to pump down the camera at your facility by following the procedure outlined
in Appendix C. If the necessary equipment and expertise isn’t available, simply contact
the factory to make arrangements for returning the camera to the support facility nearest
to you to have the vacuum restored.
Data Loss or Serial Violation
You may experience either or both of these conditions if the host computer has been set
up with Power Saving features enabled. This is particularly true for power saving with
regard to the hard drive. Make sure that Power Saving features are disabled while you are
running WinView/32.
Data Overrun Due to Hardware Conflict message
Figure 34. Data Overrun Due to Hardware Conflict dialog
If this dialog appears when you try to acquire a test image, acquire data, or run in focus
mode, check the CCD array size and then check the DMA buffer size. A large array (for
example, a 2048x2048 array), requires a larger DMA buffer larger setting than that for a
smaller array (for example, a 512x512 array).
To change the DMA buffer setting:
1. Note the array size (on the Setup|Hardware|Controller/CCD tab or the
Acquisition|Experiment Setup|Main tab Full Chip dimensions).
2. Open Setup|Environment|Environment dialog.
3. Increase the DMA buffer size to a minimum of 32 Mb (64 Mb if it is currently
32 Mb or 128 Mb if it is currently 64 Mb), click on OK, and close the WinX
application.
4. Reboot your computer.
5. Restart the WinX application and begin acquiring data or focusing. If you see
the message again, increase the DMA buffer size.
Chapter 7 Troubleshooting 71
Data Overrun Has Occurred message
Because of memory constraints and the way that USB transfers data, a "Data overrun has
occurred" message may be displayed during data acquisition. If this message is displayed,
take one or more of the following actions:
1. Minimize the number of programs running in the background while you are
acquiring data with WinView/32.
2. Run data acquisition in Safe Mode.
3. Add memory.
4. Use binning.
5. Increase the exposure time.
6. Defragment the hard disk.
7. Update the Orange Micro USB2 driver. See "To Update the OrangeUSB USB
2.0 Driver:", page 33.
If the problem persists, your application may be USB 2.0 bus limited. Since the host
computer controls the USB 2.0 bus, there may be situations where the host computer
interrupts the USB 2.0 port. In most cases, the interrupt will go unnoticed by the user.
However, there are some instances when the data overrun cannot be overcome because
USB 2.0 bus limitations combined with long data acquisition times and/or large data sets
increase the possibility of an interrupt while data is being acquired. If your experiment
requirements include long data acquisition times and/or large data sets, your application may
not be suitable for the USB 2.0 interface. Therefore, we recommend replacement of the USB
2.0 interface module with our TAXI interface module and Princeton Instruments (RSPI) PCI
card. If this is not the case and data overruns continue to occur, contact Customer Support
(see page 108 for contact information).
Demo is only Choice on Hardware Wizard:Interface dialog
(Versions 2.5.19.0 and earlier)
If RSConfig.exe has not been run and there is not an installed Princeton Instruments (RSPI)
high speed PCI card, the Hardware Wizard will only present the choice "Demo" in the
Interface dialog (Figure 35). Clicking on Next presents an "Error Creating Controller.
Error=129." message, clicking on OK presents "The Wizard Can Not Continue Without a
Valid Selection!" message, clicking on OK presents the Interface dialog again.
Figure 35. Hardware Wizard: Interface dialog
At this point, you will need to exit WinView and run the RSConfig.exe program, which
creates a file called PVCAM.INI. This file contains information required to identify the
72 PI-SCX System Manual Version 2.E
interface/camera and is referenced by the Hardware Wizard when you are setting up
WinView with USB for the first time:
1. If you have not already done so, close WinView/32.
2. Make sure the ST-133 is connected to the host computer and that it is turned on.
3. Run RSConfig from the Windows|Start|Programs|PI Acton menu or from the
directory where you installed WinView.
4. When the RSConfig dialog (Figure 36) appears, you can change the camera name
to one that is more specific or you can keep the default name "Camera1". When
you have finished, click on the Done button.
Figure 36. RSConfig dialog
5. You should now be able to open WinView and, from Setup|Hardware…, run
the Hardware Wizard.
6. When the PVCAM dialog (Figure 37) is displayed, click in the Yes radio button,
click on Next and continue through the Wizard. After the Wizard is finished, the
Controller/Camera tab card will be displayed with the Use PVCAM checkbox
selected. You should now be able to set up experiments and acquire data.
Figure 37. Hardware Wizard: PVCAM dialog
Chapter 7 Troubleshooting 73
Demo, High Speed PCI, and PCI(Timer) are Choices on Hardware
Wizard:Interface dialog (Versions 2.5.19.0 and earlier)
If there is an installed Princeton Instruments (RSPI) high speed card in the host computer
and you want to operate a camera using the USB 2.0 interface, the PVCAM.INI file
(created by RSConfig.exe) must exist and the USB 2.0 supported camera must be
[Camera_1]. PVCAM.INI, which contains information required to identify the
interface/camera, is referenced by the Hardware Wizard when you are setting up
WinView/32 with USB for the first time. If the Wizard did not find a PVCAM.INI file or
if RSConfig.exe was run but the USB 2.0 camera is [Camera_2] in the PVCAM.INI file,
"Demo", "High Speed PCI", and "PCI(Timer)" will be selectable from the Wizard's
Interface dialog.
Figure 38. Hardware Wizard: Interface dialog
At this point, you will need to run the RSConfig.exe program:
1. If you have not already done so, close WinView/32.
2. Make sure the ST-133 is connected to the host computer and that it is turned on.
3. Run RSConfig from the Windows|Start|Programs|PI Acton menu or from
the directory where you installed WinView.
4. When the RSConfig dialog (Figure 39) appears, you can change the camera
name to one that is more specific or you can keep the default name "Camera2".
When you have finished, click on the Done button. You will next edit the
generated PVCAM.INI file.
Figure 39. RSConfig dialog: Two Camera Styles
74 PI-SCX System Manual Version 2.E
If the contents of the file look like:
Change the headings so the contents now look like:
5. Using Notepad or a similar text editor, open PVCAM.INI, which is located in
the Windows directory (C:\Windows, for example).
Note: The [Camera_#] must be changed so the camera supported by the USB
interface will be recognized (the USB driver is "apausb.sys"). For consistency,
you may also want to change the camera names.
6. Save the file. With the ST-133 connected and on, open WinView/32.
7. Run the Hardware Wizard.
8. When the PVCAM dialog (Figure 40) is displayed, click in the Yes radio button,
click on Next and continue through the Wizard. After the Wizard is finished, the
Controller/Camera tab card will be displayed with the Use PVCAM checkbox
selected. You should now be able to acquire data.
Figure 40. Hardware Wizard: PVCAM dialog
Chapter 7 Troubleshooting 75
If the contents of the file look like:
Change the headings so the contents now look like:
Detector Temperature, Acquire, and Focus are Grayed Out
(Versions 2.5.19.0 and earlier)
These functions and others will be deactivated if you have installed a camera being run
under USB 2.0 and have opened WinView/32 without having first turned on the ST-133.
They will also be deactivated if you have installed a camera being run under USB 2.0 and
a Princeton Instruments high speed PCI card was also detected when RSConfig.exe was
run.
1. Check to see if the ST-133 is connected to the host computer and is turned on. If
it is not connected or is connected but not turned on, go to Step 2. If it is
connected and on, go to Step 3.
2. Close WinView, verify that the ST-133 is connected to the host computer, turn on
the ST-133, and reopen WinView. The formerly grayed out functions should now
be available.
3. If the ST-133 is connected and on, the USB 2.0 camera may not be listed as
Camera 1 in the PVCAM.INI file.
4. Run RSConfig.exe (accessible from the Windows|Start|Programs|PI Acton
menu). If the USB 2.0 camera is listed as Camera 2 (Princeton Style (USB2) in
Figure 41), you will have to edit the PVCAM.INI file.
5. Using Notepad or a similar text editor, open PVCAM.INI, which is located in
the Windows directory (C:\Windows, for example).
Figure 41. RSConfig dialog: Two Camera Styles
76 PI-SCX System Manual Version 2.E
Note: The [Camera_#] must be changed so the camera supported by the USB
interface will be recognized (the USB driver is "apausb.sys"). For consistency,
you may also want to change the camera names.
6. Save the file. With the ST-133 connected and on, open WinView/32. The
formerly grayed out functions should now be available.
Error Creating Controller message
This message may be displayed if you are using the USB 2.0 interface and have not run
the RSConfig.exe program (see previous topic), if the PVCAM.INI file has been
corrupted, or if the ST-133 was not turned on before you started WinView/32 and began
running the Hardware Wizard.
Figure 42. Error Creating Controller dialog
Error 129: Indicates that the problem is with the PVCAM.INI file. Close WinView/32,
run RSConfig, make sure the ST-133 is on, reopen WinView, and begin running the
Hardware Wizard.
Error 183: Indicates that the ST-133 is off. If you are running the Hardware Wizard
when this message appears, click on OK, turn on the ST-133, and, on the PVCAM
dialog, make sure Yes is selected and then click on Next. The Hardware Wizard
should continue to the Controller Type dialog.
Error Occurs at Computer Powerup
If an error occurs at boot up, either the interface card is not installed properly or there is
an address or interrupt conflict. Turn off the computer, reinstall the interface card (make
sure it is firmly seated), and reboot the system.
If an error occurs while you are using the WinView/32 program, check the interface
selection on the Hardware Setup|Interface tab (WinView/32). If the current choice is
"High Speed PCI", change the selection to "PCI(Timer)". If the problem goes away, you
can either correct the interrupt conflict or you can continue using PCI(Timer) for data
transfer (data transfer is controlled by a polling timer rather than interrupts). Note that
data transfer can be slower in PCI(Timer) mode on slower computers.
Conflicts
One of the many advantages that PCI offers over ISA is that the whole issue of address and
interrupt assignments is user transparent and under BIOS control. As a result, users
typically do not have to be concerned about jumpers or switches when installing a PCI card.
Nothing more should be required than to plug in the card, make the connections, and
operate the system. As it turns out, however, in certain situations conflicts may nevertheless
occur and user intervention will be required to resolve them.
Typical PCI motherboards have both ISA and PCI slots and will have both PCI and ISA
cards installed. In the case of the ISA cards, the I/O address and Interrupt assignments
will have been made by the user and the BIOS will not know which addresses and
Chapter 7 Troubleshooting 77
Slot Type
Status
I/O Address
Interrupt
1 (ISA)
ISA Network Card
200-210
11
2 (PCI)
PCI Video Card
FF00-FFFF
15
3 (ISA)
ISA Sound Card
300-304
9
4 (PCI)
Empty
N/A
N/A
Slot Type
Status
I/O Address(s)
Interrupt
1 (ISA)
ISA Network Card
200-210
11
2 (PCI)
PCI Video Card
FE00-FEFF
11
3 (ISA)
ISA Sound Card
300-304
9
4 (PCI)
Princeton Instruments PCI
Serial Card
FF80-FFFF
15
interrupts have been user assigned. When a PCI card is installed, the BIOS checks for
available addresses and interrupt levels and automatically assigns them so that there are
no PCI address or interrupt conflicts. However, because the BIOS doesn't know about the
user-assigned ISA I/O address and interrupt level assignments, it is possible that a PCI
card will be assigned an address or interrupt that is already assigned to an ISA card. If
this happens, improper operation will result. Specifically, the problems could range from
erratic operation under specific conditions to complete system failure. If such a conflict
occurs, because the user has no control over the PCI address and interrupt assignments,
there will be no recourse but to examine the .ISA assignments and change them to values
which do not conflict. Most (but by no means all) ISA cards make provision for selecting
alternative I/O addresses and interrupt levels so that conflicts can be resolved. Software is
available to help identify specific conflicts.
The following example may serve to illustrate the problem. Suppose you had a system with
an ISA network card, a PCI video card and an ISA sound card. Further suppose that you
were then going to install a PCI Serial Buffer card. Before installing the PCI Serial card, the
I/O address and interrupt assignments for the installed cards might be as follows.
Table 6. I/O Address & Interrupt Assignments
before Installing Serial Card
As shown, there are no conflicts, allowing the three peripheral cards to operate properly.
If the PCI Serial card were then installed, the BIOS would interrogate the PCI cards and
may reassign them new address and interrupt values as follows.
Table 7. I/O Address & Interrupt Assignments
after Installing Serial Card
As indicated, there is now an interrupt conflict between the ISA Network Card and the
PCI Video card (both cards have been assigned Interrupt 11), causing the computer to no
longer function normally. This doesn't mean that the PCI Serial card is defective because
the computer stops functioning properly when the Serial card is installed. What it does
mean is that there is an interrupt conflict that can be resolved by changing the interrupt
level on the conflicting Network card in this example. It is up to the user to consult the
documentation for any ISA cards to determine how to make the necessary change.
78 PI-SCX System Manual Version 2.E
Note: Changing the order of the PCI cards, that is, plugging them into different slots,
could change the address and interrupt assignments and possibly resolve the conflict.
However, this would be a trial and error process with no guarantee of success.
Diagnostics Software
Many diagnostics programs, both shareware and commercial, are available to help
resolve conflicts. Most often, all that's required is a program that will read and report the
address and interrupt assignments for each PCI device in the computer. One such
program available from Princeton Instruments' Customer Support department is called
PCICHECK. When the program is run, it reports the address and interrupt assignments
for the first PCI device it finds. Each time the spacebar is pressed, it moves on to the next
one and reports the address and interrupt assignments for that one as well. In a few
moments this information can be obtained for every PCI device in the computer. Note
that, even though there are generally only three PCI slots, the number of PCI devices
reported may be larger because some PCI devices may be built onto the motherboard. A
good strategy for using the program would be to run it before installing the PCI Serial
card. Then run it again after installing the card and note any address or interrupt
assignments that may have changed. This will allow you to easily focus on the ones that
may be in conflict with address or interrupt assignments on ISA cards. It might be noted
that there are many programs, such as the MSD program supplied by Microsoft, that are
designed to read and report address and interrupt assignments, including those on ISA
cards. Many users have had mixed results at best using these programs.
Operation
There are no operating considerations that are unique to the PCI Serial card. The card can
easily accept data as fast as any Princeton Instruments System now available can send it.
The incoming data is temporarily stored in the card’s memory, and then transferred to the
main computer memory when the card gains access to the bus. The PCI bus arbitration
scheme assures that, as long as every PCI card conforms to the PCI guidelines, the onboard memory will never overflow.
Unfortunately, there are some PCI peripheral cards that do not fully conform to the PCI
guidelines and that take control of the bus for longer periods than the PCI specification
allows. Certain video cards (particularly those that use the S3 video chip) are notorious in
this respect. Usually you will be able to recognize when memory overflow occurs
because the displayed video will assume a split-screen appearance and/or the message
Hardware Conflict will be displayed (WinView/32). At the same time, the LED on the
upper edge of the PCI Serial card will light.
Users are thus advised not to take any actions that would worsen the possibility of
memory overflow occurring when taking data. In that regard, avoid multitasking while
taking data. Specific operations to avoid include multitasking (pressing ALT TAB or
ALT ESC to start another program), or running a screensaver program.
Chapter 7 Troubleshooting 79
No CCD Named in the Hardware Wizard:CCD dialog
(Versions 2.5.19.0 and earlier)
If you have installed a USB 2.0 Interface Module in your ST-133, a blank field may be
displayed in the Detector/Camera/CCD dialog (Figure 43) if the ST-133 controller was
made before January 2001. Earlier versions of the ST-133 did not contain non-volatile
RAM (NVRAM), which is programmed with information about the controller and the
camera. PVCAM, the program under which the Princeton Instruments USB works,
retrieves the information stored in NVRAM so it can enter specific camera characteristics
into WinView/32.
Check the serial label on underside of your controller. If the first five characters are
D1200 (December 2000) or earlier (J0797 or July 1997, for example), contact Customer
Support to find out about an NVRAM controller upgrade.
Program Error message
This dialog may appear if you have tried to acquire a test image, acquire data, or run in
focusing mode and the DMA buffer size is too small. A large array (for example, a
2048x2048 array), requires a larger setting than that for a smaller array (for example, a
512x512 array).
To correct the problem:
1. Click on OK.
2. Reboot the WinX application.
3. Note the array size (on the Setup|Hardware|Controller/CCD tab or the
Acquisition|Experiment Setup|Main tab Full Chip dimensions). If your
camera contains a large array (such as a 2048x2048 array), and the DMA buffer
size is too small, there will not be enough space in memory for the data set.
4. Open Setup|Environment|Environment dialog.
5. Increase the DMA buffer size to a minimum of 32 Mb (64 Mb if it is currently
32 Mb or 128 Mb if it is currently 64 Mb), click on OK, and close the WinX
application.
Figure 44. Program Error dialog
80 PI-SCX System Manual Version 2.E
WARNINGS!
6. Reboot your computer.
7. Restart the WinX application and begin acquiring data or focusing. If you see the
message again, increase the DMA buffer size.
Removing/Installing a Plug-In Module
The ST-133 Controller has three plug-in slots. The Analog/Control module (leftmost slot
when the controller is viewed from the rear) and the Interface Control module (either a
TAXI or a USB 2.0 compatible module in the middle slot) are always provided. The third
slot, however, is covered with a blank panel unless a PTG module has been installed in
the ST-133.
If a module is ever removed for any reason, internal settings should not be disturbed.
Changing a setting could radically alter the controller’s performance. Restoring normal
operation again without proper equipment and guidance would be very difficult, and it
might be necessary to return the unit to the factory for recalibration.
Note: If you have an installed PTG and are changing from one interface type to another
(TAXI to USB 2.0 or USB 2.0 to TAXI), contact Customer Support for specific
changeover instructions. See page 108 for contact information.
1. Always turn the Controller OFF before removing or installing a module. If a
module is removed or installed when the controller is powered, permanent
equipment damage could occur which would not be covered by the warranty.
2. Before handling any boards, take precautions to prevent electrostatic discharge
(ESD). The modules are susceptible to ESD damage. Damage caused by
improper handling is not covered by the Warranty.
Figure 45. Module Installation
Chapter 7 Troubleshooting 81
WARNING!
To Remove a Module:
1. Verify that the Controller has been turned OFF.
2. Rotate the two locking screws (one at the top of the module and one at the bottom)
counterclockwise until they release from the chassis.
3. Then, grasp the module and pull it straight out.
4. Set the module aside in a safe place. If you are replacing it with another module, as in
the case of exchanging a TAXI module for a USB 2.0 module, you may be able to
use the packaging from the new module to store the old module. This packaging is
usually an antistatic bag that will protect the module components from electrostatic
discharge.
To Install a Module:
Installing a module is a bit more complex because you first have to be sure the locking
screws are aligned correctly. The following procedure is suggested.
1. Verify that the Controller has been turned OFF.
2. Remove the replacement module from its antistatic packaging. This packaging is
designed to protect the module components from electrostatic discharge.
3. Rotate the two locking screws counterclockwise until the threads on the screws
engage those of the module panel. See Figure 45. By doing this, the screws will be
perfectly perpendicular to the module panel and will align perfectly when the module
is inserted.
4. Insert the module so the top and bottom edges of the board are riding in the proper
guides.
5. Gently but firmly push the module in until the 64-pin DIN connector at the back of
the module mates with the corresponding connector on the backplane, leaving the
module panel resting against the controller back panel.
6. Rotate the two locking screws clockwise. As the screws are rotated, they will first
disengage from the module panel threads, and then begin to engage those of the
bracket behind the controller panel.
Tighten the screws to where they are just snug. Do not tighten them any further because
you could easily bend the mating bracket.
82 PI-SCX System Manual Version 2.E
Securing the Detector-Controller Cable Slide Latch
Some Princeton Instruments Detector-Controller cables use a slide latch to secure the
Detector-Controller cable to the DETECTOR connector on the back of the ST-133.
Incorrectly plugging this cable into the connector and improperly securing the slide latch
may prevent communication with the PI-SCX (the camera may appear to stop working).
1. Before trying to plug in the cable, slide the latch up (toward Pin 1). Then, plug the
cable into the DETECTOR connector on the ST-133.
2. Slide the latch down. You may hear a click when the latch locks.
3. Verify that the connector is fully secured.
If you are having trouble sliding the latch, slightly pull the connector out and then slide
the latch into its locked position.
Chapter 7 Troubleshooting 83
Serial violations have occurred. Check interface cable.
Figure 46. Serial Violations Have Occurred dialog
This error message dialog will appear if you try to acquire an image or focus the camera
and either (or both) of the following conditions exists:
The camera system is not turned ON.
There is no communication between the camera and the host computer.
To correct the problem:
1. Turn OFF the camera system (if it is not already OFF).
2. Make sure the Detector-Controller cable is secured at both ends and that the
computer interface cable is secured at both ends.
3. After making sure that the cables are connected, turn the camera system power
ON.
4. Click OK on the error message dialog and retry acquiring an image or running in
focus mode.
Note: This error message will also be displayed if you turn the camera system OFF or a
cable comes loose while the application software is running in Focus mode.
84 PI-SCX System Manual Version 2.E
This page intentionally left blank.
Appendix A
Model
CCD (WinView/32
Name)
Pixel
Format
Pixel Size
CCD Type
1300
EEV CCD36-40
(EEV 1300x1340F)
1300 × 1340
20 × 20 µm
100 kHz/1 MHz, Full
Frame
4300
Kodak KAF 4300E
(KAF 2084x2084 SCX)
2084 × 2084
24 × 24 µm
1 MHz, Full Frame
4096
Fairchild CCD486
(FCD 4096x4096F)
4096 × 4091
15 × 15 µm
1 MHz, Full Frame
Specifications
CCD Arrays
Types
The following list is not necessarily current. Other chips may also be available. Contact
the factory for up-to-date information.
PI-SCX:1300
Array Size: 1340x1300
Pixel Size: 20x20 µm
Spectral Range (typical): 400-1100 nm
Cooling: Air or Water
Output Amplifiers: Low-noise (~ 250,000 e-) and High-capacity (500,000 e-)
Parallel Shift Rate: 30 µs/row
Dual A/D: 100 kHz and 1 MHz
Readout Rates:
Full Frame: ~ 1.8 s @ 1 MHz, ~18.0 s @ 100 kHz
2x2 Binning: ~ 0.65 s @ 1 MHz, ~4.8 s @ 100 kHz
8x8 Binning: ~ 0.15 s @ 1 MHz, ~ 0.5 s @ 100 kHz
Quantum Efficiency:
85
86 PI-SCX System Manual Version 2.E
PI-SCX:4096
Array Size: 4096x4096
Pixel Size: 15x15 µm
Spectral Range (typical): 400-1000 nm
Cooling: Water
Output Amplifier: Single port
Parallel Shift Rate: 200 µs/row
Pixel Well: 70 Ke- min, 85 Ke- typical
Quantum Efficiency:
PI-SCX:4300
Array Size: 2084x2084
Pixel Size: 24x24 µm
Spectral Range (typical): 350-1000 nm
Cooling: Water
-50°C with +5°C circulating water for fiber ratio ≤ 1.5:1
-45°C with +0°C circulating water for fiber ratio >1.5:1
PI-SCX:4300
-50°C with +5°C circulating water
Camera Cooling
PI-SCX:1300: -30°C (forced air) to -35°C (water cooled at +25°C) thermoelectric (Peltier)
PI-SCX:4096: -50°C with +5°C water (≤1.5:1 fiber ratio); -45°C with +0°C water
(>1.5:1 fiber ratio)
PI-SCX:4300: -50°C with +5°C water
Passive Cooling: CCD array cooled by Peltier device. Heat is radiated away by
cooling fins on body of the round head camera.
Supplemental Air Cooling: The PI-SCX:1300 camera may have an internal fan that
draws air in from the vents in the camera shell, circulates it past the internal cooling fins,
and then exhausting the warmed air back into the atmosphere.
Water Cooling: Mixture of 50% ethylene glycol and 50% water pumped through
coolant ports at the side of the camera (PI-SCX:1300) or the rear of the camera
(PI-SCX:4096 and PI-SCX:4300).
Temperature Control
Setting Mechanism: Temperature is set by the application software.
Display: The actual temperature can be displayed at the computer by the application
software.
Gain
Control Precision: ±0.05°C; closed-loop stabilized-temperature control
Cooldown Temperature (typical): As long as the ambient temperature is between
+18°C and +23°C and the humidity is <50%, PI-SCX CCDs can typically be cooled
down to the following temperatures:
Table 8. Camera vs. Approximate Cooldown Temperature
If the ambient temperature is greater than +23°C, temperature lock may not be achievable
at the above temperatures
Software selectable (high, medium, low)
88 PI-SCX System Manual Version 2.E
PI-SCX:4096 ≤ 1.5:1
PI-SCX:4096 >1.5:1
6.00 in (15.24 cm) width;
8.50 in (21.59 cm) width;
6.00 in (15.24 cm) height;
8.50 in (21.59 cm) height;
7.94 in (20.17 cm) length;
9.20 in (23.37 cm) length;
35 lb (15.89 kg) weight
Dimensions
All dimensions are approximate and subject to change. Refer also to the outline drawings
in Appendix B.
PI-SCX:1300
4.63 in (11.8 cm) width;
4.63 in (11.8 cm) height;
8.71 in (22.15 cm) length;
10 lb (4.54 kg) weight
PI-SCX:4300
6.00 in (15.24 cm) width;
6.00 in (15.24 cm) height;
8.41 in (21.36 cm) length;
35 lb (15.89 kg) weight
Female, D-subminiature 9-pin connector for RS232 serial communication
Controller Power Input
Refer to the Fuse/Voltage label on the back of the ST-133 for fuse, voltage, and power
consumption information.
Controller Dimensions
Dimensions: See Appendix B.
Controller Weight: ST-133A: 13 lb (5.9 kg); ST-133B: 12.5 lb (5.7 kg)
Mounting
PI-SCX:1300 The optional Tripod mount provides one each 1/4"-20, 3/8"-16, and M6
mounting hole.
PI-SCX:4096 and PI-SCX:4300 There are two M8 × 1.25 tap (25 mmdeep) and one
1/420 tap (0.5deep).
Appendix B
Outline Drawings
Note: All dimensions are in inches unless otherwise noted.
Figure 47. ST-133A Controller
Figure 48. ST-133B Controller
89
90 PI-SCX System Manual Version 2.E
Figure 49. PI-SCX:1300
Appendix B Outline Drawings 91
Figure 50. PI-SCX:1300/PW
92 PI-SCX System Manual Version 2.E
Figure 51. PI-SCX:1300/1.5PW
Appendix B Outline Drawings 93
Figure 52. PI-SCX:4096 (100 mm Fiber Optic)
94 PI-SCX System Manual Version 2.E
Figure 53. PI-SCX:4096 (165 mm Fiber Optic)
Appendix B Outline Drawings 95
Figure 54. PI-SCX:4300 (75 mm Fiber Optic)
96 PI-SCX System Manual Version 2.E
Figure 55. PI-SCX:4300 (165 mm Fiber Optic)
Appendix C
Repumping the Vacuum
Introduction
Fiberoptic taper CCD cameras are shipped with a vacuum level below 10-5 Torr to assure
proper cooling performance and to prevent condensation from collecting on the CCD. In
time, the vacuum level could deteriorate to where achieving temperature lock will no
longer be possible. If this happens, it will be necessary to restore the vacuum.
A ¼ Vacuum Pumping fitting (Figure 56) is supplied with the camera but is not factory
installed. The access port, located at the top of the camera, is capped. The cap can easily
be removed, allowing the vacuum pumping fitting to be installed.
Figure 56. Vacuum Pumping Fitting
A special Indium-tipped Vacuum Seal (Figure 57) is installed in the side of the camera
(see Figure 15 on page 39). At installation, the Vacuum Seal’s Indium tip is crushed to
provide a metal-to-metal seal in the vacuum chamber. The number of times this can be
done before a new Vacuum Seal is required is limited. Typically, an Indium tip can be
crushed up to ten times before the Vacuum Seal must be replaced to assure a good
vacuum seal. Because the tip is crushed only once at the factory, it can be successfully
used again numerous times in the field. With repeated repumpings, however, it will
eventually be necessary to replace the Indium-tipped Vacuum Seal (PI #2550-0352).
Note: The required seating torque for successful vacuum sealing is 28 in. lb. (3.2 Nm).
Figure 57. Indium-tipped Vacuum Valve
97
98 PI-SCX System Manual Version 2.E
WARNING!
Caution
Requirements
A laboratory-type vacuum pump capable of achieving absolute pressures below
10 mTorr.
Your vacuum system must have a trap (ideally cryogenic) placed between the camera and
the pump to prevent contamination due to oil backstreaming from the pump.
5/16 hex torque wrench for tightening the Indium-tipped Vacuum Seal.
A new Indium-tipped Vacuum Seal (PI #2550-0352). A new seal will only be
required if the camera has been repumped a number of times and the original seal is
no longer able to maintain vacuum.
Pumping Procedure
Wait at least two hours after the camera was last operated before proceeding. This is
necessary so that the CCD can warm to ambient before being opened to the atmosphere.
If opened when still cold, potentially harmful condensation could form on the CCD.
Preliminary Steps
1. Turn off the system power and disconnect all cables.
2. Apply a light coating of high-quality vacuum grease such as Apiezon M to the O-
ring on the Vacuum Pumping Fitting supplied with the system.
3. Remove the cap from the vacuum access port at the top of the camera.
4. Insert the end of the Vacuum Pumping Fitting into the port at the top of the
camera to where the threads can be engaged. Then tighten it down. Finger tight is sufficient. The greased O-ring makes the seal.
5. Connect the vacuum hose from the pump to the ¼ fitting at the end of the
Vacuum Pumping fitting and clamp the hose connection securely.
6. If you are going to reuse the currently installed Indium-tipped Vacuum Seal,
turn on the vacuum pump. Then:
Using the 5/16 hex torque wrench, rotate the Indium-tipped Vacuum Seal
two full revolutions counterclockwise, just enough to unseal the vacuum.
Perform the steps outlined in the "Pumping Down" section, located on the
following page.
7. If you are going to install a new Indium-tipped Vacuum Seal, do not turn on the
pump. Instead, perform the steps outlined in the section titled "Installing a New Vacuum Seal", which follows the "Pumping Down" section.
Appendix C Repumping the Vacuum 99
Caution
Pumping Down
1. Do not bake during pumping. Maintain ambient temperature (≈25°C).
2. Continue pumping for 48 hr. The final vacuum must be below 10 mTorr. At the
factory, a much higher initial vacuum level is generated using a turbo pump.
Although this is desirable, it is not necessary and may be cost prohibitive for the
user to purchase similar equipment; 10 mTorr is sufficient to assure normal
operation of the unit.
3. When the required vacuum level has been achieved, tighten down the Indium-
tipped Vacuum Seal with the 5/16 torque wrench to crush the Indium tip and
reseal the vacuum. A sealing torque of 28 in. lb. or 3.2 Nm is required.
4. Shut down the pump and remove the vacuum hose from the Vacuum Pumping
fitting. Although it shouldn’t be required, a 7/16 nut driver may prove helpful in
removing the fitting.
5. Replace the vacuum port cap.
Installing a New Vacuum Seal
Preventing atmospheric contamination of the vacuum chamber is imperative to the
operation and longevity of the camera.
Placing the camera inside a plastic bag filled with dry nitrogen gas prior to removing
the Vacuum Seal will backfill the camera when the Vacuum Seal is removed and
help prevent contamination. Consult the factory.
Alternatively, a dry nitrogen pressure may be applied to the vacuum pumping fitting
before removing the vacuum valve. Pressure should be 3 psig or less. Maintain a flow
of nitrogen until the valve is replaced in Step 3. Consult the factory.
1. Using the 5/16" hex torque wrench, rotate the Indium-tipped Vacuum Seal
counterclockwise to where the threads completely disengage and then remove it
from the vacuum block.
2. Lightly grease the Viton O-ring on the new Indium-tipped Vacuum Seal. Use a
high-quality grade of vacuum grease such as Apiezon M.
3. Carefully insert the new Indium-tipped Vacuum Seal into the corresponding
opening on the camera vacuum block to where the threads can be engaged – a
turn or two is sufficient. Do not rotate the plunger to where it bottoms out. The Indium tip should not be disturbed until the new vacuum has been established.
4. Turn on the pump and follow the "Pumping Down" instructions.
100 PI-SCX System Manual Version 2.E
This page intentionally left blank.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.