PRIMETECH UCD300 Users Manual

SMP/UCD-300
This quick guide show you how to use iPRECIO management software.
®
The iPRECIO
use.
is for use in Laboratory Animal Research ONLY. Not for huma
®
, programmable infusion pump for Lab animal for
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Read me
®
pump and operate iPRECIO
®
Please obey the Guide for the Care and use of laboratory animal by your
®
organization and or institute when use iPRECIO
NOTICE TO THE USER One or more of the following statements may be applicable for this equipment.
Notice (For Model UCD-300 only)
The provided LAN Cable with a ferrite core shall be used when connect UCD-300 to a PC and do not replace the LAN cable with other LAN cable.
FCC WARNING
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
For more information : PRIMETCH CORPORATION TEL:+81-3-3816-0851 E-mail : iprecio@primetech.co.jp
URL :www.iprecio.com
.
2
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
INFORMATION TO THE DIGITAL DEVICE USER REQUIREMENT by FCC (For Model UCD-300 only)
This e quipment has been tested and found to com ply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver. —Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. —Consult the dealer or an experienced radio/TV technician for help.
3
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Contents
READ ME ..................................................................................................................................... 2
CONTENTS .................................................................................................................................. 4
ABOUT IPRECIO ...................................................................................................................... 5
®
THE WORLD FIRST WIRELESS PROGRAMMABLE INFUSION PUMP ......................................................... 5
ACCESSORIES WHICH INCLUDED IN IPRECIO STD SYSTEM ........................................................... 5
IPRECIO SYSTEM CONSTRUCTION ................................................................................................ 6
WORK FLOW ..............................................................................................................................7
PRODUCT WARRANTY............................................................................................................. 8
BATTERY LIFE............................................................................................................................ 9
INSTALL MANAGEMENT APPLICATION............................................................................. 10
PREPARATION BEFORE START......................................................................................................... 10
INSTALL IPRECIO MANAGEMENT SOFTWARE................................................................................ 10
CABLE CONNECTION OF DATA COMMUNICATION UNIT UCD-300 .................................................... 12
TCP/IP CONNECTION BETWEEN DATA COM UNIT AND CONTROL PC................................................ 13
PREPARATION BEFORE SOFTWARE USE .......................................................................... 14
FOR THE FIRST TIME EXECUTE MANAGEMENT SOFTWARE ................................................................14
USER REGISTRATION .................................................................................................................... 14
ACCESS CONTROL OF SOFTWARE................................................................................................... 15
CHANGE LANGUAGE ...................................................................................................................... 15
NETWORK SETTING ...................................................................................................................... 16
CREATE INFUSION PROFILE........................................................................................................... 16
CUSTOMIZE INFORMATION............................................................................................................. 20
START STUDY........................................................................................................................... 21
®
®
®
START NEW STUDY....................................................................................................................... 21
CATHETER CONNECTION, PLOMB ADMIN LIQUID AND ACTIVATION................................................... 23
PUMP DETECTION.......................................................................................................................... 25
KVO SETTING, START AND COMPLETE ............................................................................................ 25
INFUSION PROTOCOL SETTING ......................................................................................................27
START INFUSION PROTOCOL.......................................................................................................... 27
MONITOR THE RUNNING PUMP STATUS........................................................................................... 28
REPLACE AND/OR REFILL THE MEDICAL SOLUTION ......................................................................... 29
END OF STUDY ............................................................................................................................. 30
SAVE STUDY FILE.......................................................................................................................... 30
EXPORT STUDY FILE...................................................................................................................... 30
PRODUCT SPECIFICATION ................................................................................................... 31
www.iprecio.com
4
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
About iPRECIO®
The world first wireless programmable infusion pump
iPRECIO implantable into a laboratory mouse. Utilize high precision mechanical pump technology that was unique developed and controlled by built in microprocessor. An administration protocol can be programmed with dedicated management application software and pump will be work by the programmed protocol with accurate low flow rate (L/H order) medical solution infusion. In addition the medical solution in the reservoir can be complimented and or replace through septum by transcutaneous even after implanted in an animal.
Accessories which included in iPRECIO® STD System
®
SMP-300 model is the world first programmable pump which is complete
iPRECIO
(L) Pump in pre-sterilized pack and 5pcs box. (R)Packaged data UCD-
®
system configured with the following accessories.
Small Lab animal micro infusion pump iPRECIO iPRECIO Data communication unit UCD-300 (AC Adapter, LAN cable included) Management Software iPRECIO Quick guide (this guide)
®
management system IMS-300
®
SMP-300
300
5
www.iprecio.com
iPRECIO® System construction
Top sur f ace Bottom surface
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Number
Refill septum ② Outlet tube ③ Auxiliary Wire antenna for radio communication ④ Reservoir ⑤ Micro cam ⑥ Pump ID label ⑦ Power switch ⑧ LAN cable connect port ⑨ AC Adaptor connect port ⑩ Antenna (The antenna is permanently fixed)
name
6
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Work flow
A study performed with iPRECIO®を, normally follow work flow below.
7
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Product Warranty
PRIMETECH CORPORATION (hereinafter PRIMETECH) guarantee the product and it’s
material are free from defect within the below period which was distributed through
the official sales channel. PRIMETECH do not responsible with any problem that caused by using with other company product. Also the product performance is not guaranteed when a customer uses it beyond the original specification.
The defect part will be replaced and or repair by free of charge except, in-transit accident, customer’s fault, breakage caused by abnormal/improper use and damnification cause by
PRIMETECH recommendation of use. Any defect parts back to PRIMETCH for replace, always need the RMA format and ship back through original sales channel.
Duration of guarantee
iPRECIO
iPRECIO
This warrantee represent any other warrantee and include an implied consent of merchantability and dedicated fitness for purpose, but not limited to.
Liability Limitation Except the warrantee describe in above, PRIMTECH do not responsible any direct, indirect and special loss caused by violence of use condition within the maximum LAW protection. This includes any return loss and or opportunity loss (includes profit loss by contract) with direct, indirect, and damage. This warrantee will be no longer valid when customer (include stuff, member and any other bystander) use iPRECIO conditions.
®
Micro Infusion Pump SMP-300: 24 months after manufacturer
®
Management System IMS-300: 12 months after shipping date
®
for any other purpose against PRIMETECH sales
8
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Battery life
iPRECIO® SMP-300 can continuous operate time below table during product warranty period (After 24 months from manufacturing date)
Flow rate (μl/hr) Battery life (hr)
0.1 408
0.5 384
1.0 360
5.0 216
10.0 144
Please note that below condition makes shorter battery life. :
In case of kept certain period of time without use after power on. (After
power on, internal circuit becomes active and consume battery power by standby power requirement)
In case of not storage under proper circumstances. (Under frequent
temperature swing or exposed in over storage temp range in spec may cause increasing self-discharge of battery)
9
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Install management application
Preparation before start
iPRECIO Windows 7(32bit OS) installed and LAN port ready PC Data communication unit UCD-300 (data com unit)
UCD-300 main body AC adapter LAN cable (Cross over cable with Ferrite core)
Install iPRECIO® management software
®
Management software install CD
PC setting should be:
Off the screen saver Get into the control panel Power option of “Change when PC sleep” and set
“Put the computer to sleep” = “Never”
1. Run the setup.exe” in a install CD.
2. When the startup window appeared, click “Next>”. Then setup wizard window comes up,
click “Next>” again.
3. Chose the program files store folder and click “Next>”. Subsequently confirmation
screen open, then click “Next>”. For this step the installation will be started.
www.iprecio.com
10
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
4. When install complete windows appears, click “Close”.
5. Check start menu that the following software installed.
iPRECIO
®
Management Software IMS-300
Utility Base Station Manager Utility Pump Database Manager
6. Three of those software has been installed, then the management software install
is successful.
11
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Cable connection of data communication unit UCD-300
1. Plug in the AC Adapter to data communication unit and connect with the PC which
the management software installed by cross-over type LAN cable.
Connect to LAN port of Control PC
Plug in the AC Adapter to outlet (100-240V, 50/60Hz)
Ferrite core attached end shall be connected to UCD-300.
12
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
TCP/IP connection between data com unit and control PC
1. Connect data com unit and control PC with LAN cable
2. Open Network and Sharing Center from control panel.
3. Click Change adapter setting, then right click on the data com Local Area Control
and open Property.
4. Open Internet Protocol version 4(TCP/IPv4) property and input IP address below
IP Address: 192.168.1.11 Subnet mask: 255.255.255.0
Then press OK to finalize.
The Data communication unit factory default IP address configuration is 192.168.1.10. If IP address doesn’t work by usage environment with default configuration, please change IP address in accordance with Appendix A: How to change IP address of the Data communication unit
5. With that the Data communication unit and control PC is connected by Internet
Protocol. Subsequently network connection should be made in the management software.
13
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Preparation before software use
For the first time execute management software
For the first time run the management software, the user registration window will open automatically. To register user information, user management and access c o n tr o l function will be available. The language default setting is in English and after user registration the language selectable function will be active to change.
User Registration
1. Input required information in the user registration window. The asterisk (*) marked
item below are mandatory.
Institution & Department Administrator
2. In addition, at least one operator registration required at this time. Input operator’s User
Name, Login ID and Password (Confirmation required), then click on Add tab.
The operator will be listed in right table.
3. To use Access control, check the “Enable Access Control” check box. Detail of Access
Control function, please refer the nest page “Access control of software”.
4. After all fill in required column, click OK to finalize.
14
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Access Control of software
To utilize Access Control function, administrator can confine and or control of user who can reach the management software. To permit to access the software, check the “Enable
Access Control” in the user registration window.
Once use control function is active, every time Login ID and Password is required whenever the software executed.
Please note Login ID and Password. When forgot password, then software access will be denied.
Change Language
15
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
1. Click Language menu
2. Select language from pull down menu, then click Change.
3. Now the language has been changed. (Below example shows Japanese)
Network Setting
1. Click network setting menu in tab.
2. Input default IP address(192.168.1.10) in the Host IP address column for the
data com unit.
3. Click “connect”
4. After confirm “Status: connected”, then close the window.
5. If the status column shows “disconnected”, check the IP address setting. Please refer
the “TCP/IP connection between data com unit and control PC”.
Create Infusion Profile
Administration protocol that perform in vivo pump of a subject animal of experiments, it is required create in the management software in advance and store as a “administration protocol”. The administration profile should be determined by each group of dosing
preparation and concentrations and assign administration schedule.
1. Open Administration profile menu by clicking.
2. Administration window comes up.
16
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
3. Add new administration profile, click Add” button.
4. Group profile creating window will be opened.
17
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
5. Input and or select basic information. When chose the injected amount unit as
μg/kg/hr(Dose) or mg/kg/hr(Dose), amount set by dose. It is therefore the animal weight range and concentration input are mandatory.
6. Input dose amount and duration from Step1 to create the infusion schedule. Input
the value to the each boxes and press Enter to finalize dose. The starting time and finish time will be shown in elapsed time that accumulate the injection period. The battery life shown in the bottom of window indicate the expected battery life which calculated as a new battery. It may help to create an administration schedule.
Weight range should be input by maximum to minimum. Please do not forget to finalize the data by pressing “Enter” key. Do not refer the battery life when the pump was re-use and or
re-programming. The indicated battery life was calculated from the battery capacity as a new one.
7. When ditto the same schedule, the use the “Repeat Setting”. Check the Repeat Start” box for start of repeated schedule step then check the Repeat End” box for termination ditto schedule. For number of cycle, input “Times ” of “Repeat Start” column and finalize to press Enter key.
8. After all information input, press OK then new administration profile will be
registered in the Infusion Profile list.
18
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
9. When to change the existing Infusion Profile, select intended Infusion Profile and click Edit”button.
If want to eliminate the existing Infusion Profile on the list, select the intended
Infusion profile to delete and click the “Delete” button.
10. Close the Infusion profile window by clicking “Close” button.
19
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Customize information
Pull down menu registration and elimination are possible for Study Setting. And KVO setting of flow rate and duration also set in advance.
1. Open customize window from “Customize(Z)” menu in the tool bar.
2. Input a desire information in the Animal Species, Animal Strain and
Administration Route, then click Add button in the below of each boxes.
3. If would like to delete existing information, select information and click Del button
in the below of each boxes.
4. Input the numerical value of KVO Flow Rate and Duration.
5. To complete the input value, click OK button in the left bottom of the Window.
20
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Start Study
Start New Study
iPRECIO® Management software has a function to abort the study during the KVO and Administration with pumps which implanted in under study animals. For example, to be able to except individual animal from study that is slow-healing than others during KVO. Also even during Administration to be able to deselect individual animal that is difficult to continue Administration by some trouble.
KVO
Change medical solution
The Data access Radio system of iPRECIO®is not Broadband type in use. For that reason, the management system can control only one pump in the same time, then it may require long time to complete the following data communication.
Pump detection Start KVO Abort KVO Complete KVO Start Administration protocol Abort Administration protocol Finish Administration protocol Extract executed study log
Administration
21
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
1. Click New from file menu bar.
2. Appear Study Header window. (*) marked items are mandatory.
3. Click Next after fill the information input boxes.
4. Pumps and Animals setting window comes up. Input intend animal number in the
Number of Animals and click Set button. Default Number of Animal is 20. The information input column will be active in answer to input animal number.
5. Input and select by pull down for except Pump ID and Calibration Factor, such as
individual information. (Animal ID, weight[g] etc.)
6. Can copy and paste of the column which already input individual information to
another column.
22
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Catheter connection, plomb admin liquid and activation.
Activate the pump by switch on to communication enable. If required connect additional catheter as necessary.
To fill administration liquid till the end of catheter before power on, because once a pump switched on, the internal cam will start moving to the initial position to block the inner tube.
1. Open pump pre-sterilization pack in clean environment.
2. Cut, replace tube and/or catheter connection of pump as necessary.
Each pump dead volume can be calculated in KVO setting, so keep final tube length and inner diameter or measured dead volume need to be recorded.
3. Fill the admin into the end of tube with 27G syringe needle through upper port.
23
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
4. Press down the back side power button tenaciously with pointed top.
5. When activation successfully, inner cam rotates about 100 seconds and most rear
pin block the tube. (When a bland new pump in use)
Factory-configured
Block tube mode
6. After activate with switch on, a pump is ready to implant to an animal under test.
Once activate an infusion pump, it is impossible to turn off power of the pump.
24
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Pump detection
1. Place pumps by recommended position with data com unit UCD-300.
2. Click Detect button. During Communication progress bar is shown, the
communication enable pump will be detected and assign individual animal ID.
If there is no response from pump, please check the followings:
It may take long term to detect pump due to radio wave back
ground environment.
Verify PC and UCD-300 connect properly. For detail, please refer
“TCP/IP connection between data com unit and control PC” and “Network Setting”.
Verify pump activation. For detail, please refer “Catheter connection,
plomb admin liquid and activation”.
3. After all pump detected, then click Next button.
KVO setting, start and complete
KVO is abbreviate of K continuous liquid dosing by low flow rate. Be normally administrate biocompatible liquid such as saline as KVO by low flow rate after implant surgery and prevent clog of catheter end during animal healing period. After animal comeback form the surgery, the KVO liquid in reservoir can be replaced by medicine liquid from the septum. And acknowledge user the timing that when the medicine liquid will reach the catheter end by pump and appurtenant catheter dead volume calculation.
eep Vein Open, semantics to prevent the catheter edge clog by
1. KVO setting window pops up. If KVO is not required, then check skip KVO and click
Next>> button.
2. Click individual column dead volume setting.
3. Input the required information for dead volume calculation and click OK button.
25
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Even if there is no need to change dead volume setting, it need to open every animal dead volume setting and determine the dead volume choice. If this determine was lost, the dead volume and flash rate will not apply to the administration start schedule.
4. Check KVO flow rate and KVO duration and if it’s ok then click to KVO start. The pump
status will be KVO after start KVO. Pump status color will change in accordance with communication stats, red YellowGreen. When the color will be Green, then injection is started.
During KVO is running, please DO NOT STOP the management software.
5. When the pre-set KVO period will end, pumps will stop and Pump status changes
into Finished. Click Next>, the injection schedule setting. All of animal looks fully recovered before pre-setting KVO period, you can abort all KVO by clicking abort as a pack and start administration schedule. When some animal can not heal enough during KVO period and need to terminate the experiment of that animal, then click the remove button from individual animal column.
After KVO stopped, please check the Pump status as “Finish”. Demepnd on the communication condition, “Finish” status appears very slow after prompt to end KVO from management software due to communication condition.
26
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Infusion Protocol setting
1. Group information window appears.
2. Chose infusion group name (Profile) from Group ID pull down menu and set
Flushing Rate and Administration Start Time, then click Set button.
3. When cl ick St a gger, t he admin i stration w ill be sta g gered fro m set time in that value.
4. If the proper infusion group name does not found, then create new one. Please refer “Create Infusion Profile”.
5. After all animal are assigned to profile group, click Next>> button.
Start Infusion Protocol
1. Infusion Protocol window comes up.
2. Click Start all or click start for individual animal to start Infusion Protocol.
During Infusion Protocol is running, please DO NOT STOP the management software.
27
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
Monitor the running pump status
1. Open Monitor window by clicking Monitor (M) from menu bar.
2. Upper part of window shows Study Name, Group and Animal. Select Group ID
from pull down menu in Group column and Animal ID from Animal column.
3. There is Refill Log table in the Right side. At first input the medical solution refill
time and volume, then click Enter button. For de tail, please refer “Replace and/or refill drag solution”. (Nest chapter)
4. The estimated remain solution volume is shown by calculated from infusion
schedule and duration in the bottom of window. Also software alarm will be appear when the calculated volume reaches the Alarm Setting volume.
5. The Infusion Profile Graph locates in the left side of window. Clicking “Tim e Unit
button upper right, the Graph axis unit can be switched. Also clicking “Schedule” button bottom right, the Infusion schedule table will be shown. TO return, just click “Graph” button.
28
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Replace and/or Refill the medical solution
1. Open Monitor (M) window.
2. Verify the estimated remain volume in the bottom area.
3. Make pump implanted animal immobilization by weak anesthesia or retention.
4. When replace the solution, aspirate remain fluid in reservoir by 27G syringe needle
through the skin to port. Subsequently fill the replacing solution into syringe and refill it to reservoir. To restock the drag solution, just fill the solution in a loaded syringe through the skin to port.
5. Input Refill Time and Volume, click Enter button. When executed by dose mode,
input new weight and click Calculate button, recast New Concentration result will be displayed to keep dose condition.
6. After fix the Refill Log, the table record automatically the Refill time, volume and
comments (weight and density). To remove the refill log, select target refill log and click “Delete” button.
29
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
End of Study
After all infusion protocol finished, click “Study End” makes automatically record study date and finish study. If study should be abort during infusion schedule, click “Finish All” then click “Study End”
Save Study file
1. Click Save Study from the File (F) in menu bar. The data file stored in
C:\Users\XXX\AppData\Local\iPRECIO\study
2. The save data can be loaded from “Open” in the File (F) in the menu bar.
Export Study file
1. Chose “Export” from the File (F) in menu bar.
2. Select folder to export, save with file name.
3. The exported data format is “csv”. To view the data, simply import it by regular spreadsheet program.
30
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j
Product Specification
Micro infusion pump for small laboratory animal iPRECIO®(Model:SMP-300)
Administration method Rotary finger type (Patented: Peristaltic mechanism) Reservoir 130μL Built-in reservoir: (Material) Medical grade SEBS
The Maximum capacity is 130μL, but out of accuracy guarantee range. To keep accuracy, only within 100μL is guaranteed. Inner Tube: (Material) Medical grade SEBS Tube Outer Tube: (Material) Medical grade SEBS (3Fr)
(Size) ID=0.5mm, L=140mm
Outer tube is detachable.
Replace/Refill port 2mm size refill port stick out in the pump.
(Recommend below 27Ga size needle)
Data communication ISM band radio (Frequency: 916.1MHz)
Com unit:Within 3m from UCD-300
Antenna Wire antenna with pump (L: ca. 50mm)
Coating by Polyurethane(1.2Fr) Titanium wire(0.1mm) Programmable Flow Rate (FR) Infusion accuracy ±5 (Under pressure 0~80cmH Battery Life Continuous motion FR 0.1μL/hr.: 19days,
Power on Activation By pressing down of the power on nub switch on the bottom
Size 24.8(L) 15.0(W) 7.2 (H) mm
Weight 3.3g (Include accompanying catheter and wire antenna) Solid content 2.15cc Type of usage Disposable Package EOG Pre-Sterilization pack (5 package/box) Sterilization TTL 2yrs (Built in batter causes self-discharge during storage)
Data com unit (Model UCD-300)
Interface with control PC LAN ( Cross type cable with ferrite core)
Communication way with pump
One unit controllable 20~25 pumps (Theoretical value without time expense )
0.1~10.0μL/hr. (setting resolution:0.1μL/hr.)
Discharge flow by 360 degree roll of cam:3.4μL
2O)
FR 10.0μL/hr.: 1 week
of pump to pump starter. (Coinstantaneously activated)
Once activated, it’s not reversible to switch off.
(Except accompanying catheter and wire antenna)
Control through Ethernet by setting fixed IP address to the
Data com unit.
ISM band radio (Frequency: 916.1MHz)
Not broadband type. One –on one communication between
Data com unit and individual pump.
31
www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e
pump number
Power source
Size and Weight 120(L)  77 (W)  31(H) mm 167g STD accessory Anchorage to cage(Velcro) LAN cable with ferrite core
AC adapter cable included. [Weight] 73g [Cable length] 1.85m [I/P] 100-240V 50/60Hz 25VA  [O/P] DC 5V 2A
(3m) 1 cable
Management software
Supported OS Windows XP Windows 7 (32bit)/Japanese English
32
www.iprecio.com