19.3.2 The use of the measuring objectives 1:1 and 5:1 ......................................................62
19.3.3 Maximum Power ....................................................................................................... 62
19.4 Optical path in LQM+ HP (with IPG Collimator) ........................................................................ 63
20 Formula and algorithms for raw beam back calulation 64
Revision 02/2019 EN
5
35,0(
6
PRIMES - The Company
PRIMES manufactures measuring devices used to analyze laser beams. These devices are employed for
the diagnostics of high-power lasers ranging from CO
length range from infrared through to near UV is covered, offering a wide variety of measuring devices to
determine the following beam parameters:
• Laser power
• Beam dimensions and position of an unfocused beam
• Beam dimensions and position of a focused beam
• Beam quality factor M²
PRIMES is responsible for both the development, production, and calibration of the measuring devices. This
guarantees optimum quality, excellent service, and a short reaction time, providing the basis for us to meet all
of our customers’ requirements quickly and reliably.
lasers and solid-state lasers to diode lasers. A wave-
The LaserQualityMonitor LQM+ is exclusively intended for measurements which are carried out in or nearby
the optical path of high power lasers. Please mind and adhere to the specifications and limit values given in
chapter 17 „Technical data“ on page 49. Other forms of usage are improper. The information contained in
this operating manual must be strictly observed to ensure proper use of the device.
Using the device for unspecified use is strictly prohibited by the manufacturer. By usage other than intended
the device can be damaged or destroyed. This poses an increased health hazard up to fatal injuries. When
operating the device, it must be ensured that there are no potential hazards to human health.
The device itself does not emit any laser radiation. During the measurement, however, the laser beam is
guided onto the device which causes reflected radiation (laser class 4). That is why the applying safety regulations are to be observed and necessary protective measures need to be taken.
In measuring mode, the laser control’s safety interlock must be connected with the device.
Observing applicable safety regulations
Please observe valid national and international safety regulations as stipulated in ISO/CEN/TR standards as
well as in the IEC-60825-1 regulation, in ANSI Z 136 “Laser Safety Standards” and ANSI Z 136.1 “Safe Use
of Lasers”, published by the American National Standards Institute, and additional publications, such as the
“Laser Safety Basics”, the “LIA Laser Safety Guide”, the “Guide for the Selection of Laser Eye Protection”
and the “Laser Safety Bulletin”, published by the Laser Institute of America, as well as the “Guide of Control
of Laser Hazards” by ACGIH.
Necessary safety measures
DANGER
Serious eye or skin injury due to laser radiation
During the measurement the laser beam is guided on the device, which causes scattered or
directed reflection of the laser beam (laser class 4).
The LaserQualityMonitor LQM+ cannot be operated in any of the available configurations
without taking the following precautions. All precautions must be taken, even when the fiber
is in the collimator or fiber adapter.
Please take the following precautions.
X
If people are present within the danger zone of visible or invisible laser radiation, for example near laser
systems that are only partly covered, open beam guidance systems, or laser processing areas, the following
safety measures must be implemented:
• Connect the laser control’s safety interlock to the device. Check that the safety interlock will switch off
the laser properly in case of error.
• Please wear safety goggles adapted to the power, power density, laser wave length and operating
mode of the laser beam source in use.
• Depending on the laser source, it may be necessary to wear suitable protective clothing or protective gloves.
• Protect yourself from direct laser radiation, scattered radiation, and beams generated from laser radiation
(by using appropriate shielding walls, for example, or by weakening the radiation to a harmless level).
• Use beam guidance or beam absorber elements that do not emit any hazardous substances when they
come in to contact with laser radiation and that can withstand the beam sufficiently.
• Install safety switches and/or emergency safety mechanisms that enable immediate closure of the laser
shutter.
Revision 02/2019 EN
7
35,0(
6
• Ensure that the device is mounted securely to prevent any movement of the device relative to the beam
axis and thus reduce the risk of scattered radiation. This in the only way to ensure optimum performance
during the measurement.
Employing qualified personnel
The device may only be operated by qualified personnel. The qualified personnel must have been instructed
in the installation and operation of the device and must have a basic understanding of working with highpower lasers, beam guiding systems and focusing units.
Conversions and modifications
The device must not be modified, neither constructionally nor safety-related, without our explicit permission.
The device must not be opened e.g. to carry out unauthorized repairs. Modifications of any kind will result in
the exclusion of our liability for resulting damages.
Liability disclaimer
The manufacturer and the distributor of the measuring devices do not claim liability for damages or injuries
of any kind resulting from an improper use or handling of the devices or the associated software. Neither the
manufacturer nor the distributor can be held liable by the buyer or the user for damages to people, material
or financial losses due to a direct or indirect use of the measuring devices.
LaserQualityMonitor LQM+
8
Revision 02/2019 EN
LaserQualityMonitor LQM+
2 Symbol explanations
The following symbols and signal words indicate possible residual risks:
DANGER
Means that death or serious physical injuries will occur if necessary safety precautions are not
taken.
WARNING
Means that death or serious physical injuries may occur if necessary safety precautions are not
taken.
CAUTION
Means that minor physical injury may occur if necessary safety precautions are not taken.
NOTICE
35,0(6
Means that property damage may occur if necessary safety precautions are not taken.
The following symbols indicating requirements and possible dangers are used on the device:
Components susceptible to ESD
Read and observe the operating instructions and safety guidelines before startup!
Further symbols that are not safety-related:
Here you can find useful information and helpful tips.
With the CE designation, the manufacturer guarantees that its product meets the requirements of
the relevant EC guidelines.
Call for action
X
Revision 02/2019 EN
9
35,0(
6
LaserQualityMonitor LQM+
3 About this operating manual
This documentation describes the installation and configuration of the LaserQualityMonitor LQM+ and the
execution of measurements with the LaserDiagnosticsSoftware LDS.
The LaserDiagnosticsSoftware LDS must be installed on the PC for measuring operation of the LaserQualityMonitor LQM+. The basic version of the LaserDiagnosticsSoftware LDS is included in the scope of delivery
for the device.
For a detailed description of the software installation, file management and evaluation of the measured data,
please refer to the separate operating manual LaserDiagnosticsSoftware LDS.
4 Conditions at the installation site
• The device must not be operated in a condensing atmosphere.
• The ambient air must be free of organic gases.
• Protect the device from splashes of water and dust.
• Operate the device in closed rooms only.
5 Introduction
5.1 System description
The LaserQualityMonitor LQM+ is used to measure the beam properties of a beam source either by measuring the open beam across several meters or with a fiber placed in a fiber holder or collimator.
Cw lasers can be measured and pulsed lasers can also be measured using an internal trigger.
The electronics and all optical components shown are installed in the LQM+ 20 basic device. The LQM+ 20
basic device can be supplemented by a front-end module with a beam splitter, an absorber and an alignment unit, which enables the use of the device even in the multikilowatt range. Fiber retainers, collimators,
neutral density filters as well as additional measuring objectives are available upon request.
The laser power can also be measured with the integrated power meters in the 1st and 2nd attenuator (only
LQM+ 500 and LQM+ HP).
10
Revision 02/2019 EN
LaserQualityMonitor LQM+
Fig. 5.1: Components of the LQM+ 20 basic device
35,0(6
Fig. 5.2: Components of the LQM+ 200/500
Absorber with
power meter
1st Attenuator
Alignment unit
Power meter
(LQM+ 500 only)
1st Attenuator
Alignment unit
2nd Attenuator with beam splitter
Fig. 5.3: Components of the LQM+ HP (10/20 kW)
Revision 02/2019 EN
11
35,0(
6
LaserQualityMonitor LQM+
5.2 Measuring principle
Due to the very long measuring path of three to six Rayleigh lengths, a lot of measuring is needed to characterize a collimated laser beam. The LaserQualityMonitor LQM+ therefore generates a compact caustic
inside the device by focusing the irradiated collimated laser beam with a focussing optic. In the process, the
focused beam is weakened as it passes through an integrated prisms and OD filter and then appears on the
CCD sensor, enlarged by a measuring objective.
The LaserDiagnosticsSoftware LDS software determines the radius, location, and orientation of the laser
beam by moving the prisms inside the device and measuring the two-dimensional power density distribution
repeatedly at various positions. The electronic integration time control of the CCD sensor expands the system’s dynamic area. As a result of this, it generally isn’t necessary to adjust the filter during a measurement.
The following laser sources can be measured:
• All cw and pulsed laser sources in the wave length area of 340
1090nm with medium power from approx. 1mW up to multikilowatt ranges with an M² ranging from 1
to approx. 50.
• The beam diameter at the aperture of the measuring system can vary in the range of 1.5mm to 15mm
(22mm only for LQM+ HP 20 kW), with divergence below 10mrad.
Focusing opticLaser beamPrisms
Prisms
Trigger
Measuring objective 5:1
Position 1
(illustrated inserted)
Measuring objective 1:1
Position 2
(illustrated not inserted)
–
360nm/515
–
545nm/1030
CCD sensorOD filter
–
Absorber
Fig. 5.4: Optical assembly of the LQM+ 20 basic device
Scaning rangeRaw beam
Focusing opticMeasuring objective
Fig. 5.5: Measuring Principle
12
CCD sensor
Revision 02/2019 EN
LaserQualityMonitor LQM+
5.3 Short overview installation
35,0(6
1. Installing the LaserDiagnosticsSoftware LDS on the computer
• Software is part of the scope of delivery
2. Taking safety precautionsChapter 1 on page7
3. Prepare Installation (only during open beam measurement)
• Make preparations
• Set the installation position
• Align the device manually
4. Connect the water-cooling (LQM+ HP and LQM+ 500 only)
• Connection diameter
• Observe flow rate
5. Power supply
• Connect power supply
See separate Operating
Manual of the LaserDiagnosticsSoftware LDS
Chapter 7 on page14
Chapter 8 on page20
Chapter 9 on page24
• Connect safety interlock
6. Connection with the computer
• Via Ethernet or LAN
7. Complete Installation (only during open beam measurement)
• Align the device with the LaserDiagnosticsSoftware LDS according to
chapter 11.4 on page35.
• Mount the device firmly in accordance with chapter 7 on page14
8. Measure
• Follow the safety instructions
• Select and use the measuring objective and neutral-density filter
• Check the device‘s alignment with the LaserDiagnosticsSoftware LDS
• Perform a test measurement
Chapter 9.4 on page28
Chapter 11.4 on page35
Chapter 11 on page30
Revision 02/2019 EN
13
35,0(
6
6 Transport
WARNING
Risk of injury when lifting or dropping the device
Lifting and positioning heavy devices can, for example, stress intervertebral disks and cause
chronic changes to the lumbar or cervical spine. The device may fall.
Use a lifting device to lift and position the device.
X
NOTICE
Damaging/destroying the device
Optical components may be damaged if the device is subjected to hard shocks or is allowed
to fall.
Handle the device carefully when transporting or installing it.
X
To avoid contamination, cover the apertures with the provided lid or optical tape.
X
Only transport the device in the original PRIMES transport box.
X
LaserQualityMonitor LQM+
NOTICE
Damage/destruction of the device caused by leaking or freezing cooling water
Leaking cooling water can damage the device. Transporting the device at temperatures
near or below freezing and without emptying the cooling circuit completely can damage the
device.
Empty the lines of the cooling circuit completely.
X
Don’t use any compressed air to empty the cooling circuit.
X
Even when the lines of the cooling circuit have been emptied, a small amount of residual
X
water will remain in the device at all times. This may leak out and end up inside the device.
Close the connector plug of the cooling circuit with the included sealing plug.
7 Installation
7.1 Preparation and installation position
Check the space available before mounting the device, especially the required space for the connection
cables and hoses (please see chapter 18 „Dimensions“ on page 50). The device must be firmly assembled
and must be mounted with screws (see chapter 7.3 on page17).
The LaserQualityMonitor LQM+ can be mounted and operated in any position. During open beam measurement, the LaserQualityMonitor LQM+ must be aligned with the laser beam. The steps taken for installation
can be found in chapter 5.3 on page13.
14
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
7.2 Manually aligning the LaserQualityMonitor LQM+
For an open beam measurement, the LaserQualityMonitor LQM+ must be aligned with the alignment tool to
the laser beam. The inlet aperture must be aligned centrally and at a right angle to the laser beam:
1. Align the LaserQualityMonitor LQM+ manually with a pilot laser and the alignment tool.
• The LaserQualityMonitor LQM+ is properly aligned when the pilot beam is centered through the holes in
the alignment tool.
2. Check the alignment as specified in chapter 11.4 on page35 with the alignment mode of the
LaserDiagnosticsSoftware LDS.
7.2.1 Alignment tool
An alignment tool is provided for the various LaserQualityMonitor versions that can be used for all aperture
diameters. After inserting the alignment tool into the aperture, it is fastened to a metal half ring with magnets.
MagnetsMetal half ring
Fig. 7.1: Alignment tool for aligning the LQM+ versions with various apertures
7.2.2 Manual alignment of the LQM+ 20 basic device and the LQM+ 200/500 with 1st attenuator
LQM+ 20 basic deviceLQM+ 200/500 with 1st attenuator
Fig. 7.2: Alignment tool mounted on the LQM+ 20 basic device and the LQM+ 200/500 with 1st attenuator
Revision 02/2019 EN
15
35,0(
6
7.2.3 Fine adjustment at the 1st attenuator
LaserQualityMonitor LQM+
With the
angle with two integrated micrometer screws (see Fig. 7.3 on page16).
The laser beam can thus be aligned within the boundaries for the LaserQualityMonitor LQM+. The feedback
to change the beam position through fine adjustments is issued via the LaserDiagnosticsSoftware LDS (see
chapter 11.4 on page35).
1st attenuator, the angle of incidence of the laser beam can be finely adjusted to the appropriate
+
-
+
-
Fig. 7.3: Micrometer screws on the 1st attenuator module for fine adjustment (for the example of the LQM+ 500)
16
Revision 02/2019 EN
LaserQualityMonitor LQM+
7.3 Install the LaserQualityMonitor LQM+
DANGER
Serious eye or skin injury due to laser radiation
For an open beam measurement, the LaserQualityMonitor LQM+ must be aligned to the
laser beam.
If the device is moved from its calibrated position, increased scattered or directed reflection
of the laser beam occurs during measuring operation (laser class 4).
When mounting the device, please ensure that it cannot be moved, neither due to an unin-
X
tended push or a pull on the cables and hoses.
7.3.1 Install the LQM+ 20/200/500 without bottom plate
44.360
35,0(6
12060
285
150
Fig. 7.4: Fastening threaded holes LQM+ 20/200/500 without bottom plate
There are four M6 mounting threads on the underside of the device for fastening it onto a holder provided by
the customer. We recommend screws of the strength class 8.8.
4 threaded holes M6
Revision 02/2019 EN
17
35,0(
6
7.3.2 Install the LQM+ 200/500 with bottom plate
LaserQualityMonitor LQM+
60
M6
120
Fig. 7.5: Fastening threaded holes LQM+ 200/500 with bottom plate
For the connection of a customer specific part, there are four threads M6 in the bottom plate. We recommend screws of the strength class 8.8.
4 threaded holes M6
18
Revision 02/2019 EN
LaserQualityMonitor LQM+
7.3.3 Install the LQM+ HP with bottom plate
35,0(6
Fig. 7.6: Fastening bores and alignment threads LQM+ HP with bottom plate
4 mounting holes Ø 7.8mm
4 threaded holes M8 for the alignment
Revision 02/2019 EN
19
35,0(
6
LaserQualityMonitor LQM+
8 Connect cooling circuit (LQM+ HP and LQM+ 500only)
DANGER
Fire hazard; Damage/Destruction of the device due to overheating
If there is no water cooling or a water flow rate which is insufficient, there is a danger of
overheating, which can damage the device or set it on fire.
Operate the device with a connected water cooling and a sufficient water flow rate only.
X
8.1 Water quality
NOTICE
Damage/Destruction of the device due to different chemical potentials
The parts of the device which get in contact with cooling water consist of copper, brass or
stainless steel. Connecting the unit to a colling curcuit containing aluminum components
may cause corrosion of the aluminum due to the different chemical potentials.
Do not connect the device on a cooling circuit in which aluminum components are installed.
X
• The device can be operated with tap water as well as demineralized water.
• Do not operate the device on a cooling circuit containing additives such as anti-freeze.
• Do not operate the device on a cooling circuit in which aluminum components are installed. Especially
when it comes to the operation with high powers and power densities, it may otherwise lead to corrosion
in the cooling circuit. In the long term, this reduces the efficiency of the cooling circuit.
• Should the cooling fail, the device can withstand the laser radiation for a few seconds. In this case,
please check the device as well as the water connections for damages.
• Large dirt particles or teflon tape may block internal cooling circuits. Therefore, please thoroughly rinse
the system before connecting it.
8.2 Water pressure
Normally, 2 bar primary pressure at the entrance of the absorber are sufficient in case of an unpressurized
outflow.
NOTICE
Damage/Destruction of the device due to overpressure
The maximum permissible water inlet pressure must not exceed 4bar.
X
20
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
8.3 Humidity
• The device must not be operated in a condensing atmosphere. The humidity has to be considered in
order to prevent condensates within and outside the device.
• The temperature of the cooling water must not be lower than the dew point (see Tab. 8.1 on page21).
NOTICE
Damage/Destruction of the device due to condensing water
Condensation water inside of the objective will lead to damage.
Mind the dew-point in Tab. 8.1 on page21.
X
Do only cool the device during the measuring operation. We recommend starting the cooling approx. 2minutes before the measurement and terminating it approx. 1minute after the measurement.
40
35
30
25
20
15
10
Cooling water temperature in °C
5
0
Tab. 8.1: Dew point diagram
Example
Air temperature: 22°C
Relative humidity: 60%
0510152025303540
Air temperature in °C
100
80
70
60
50
40
30
20
Relative humidity in %
10
The cooling water temperature cannot fall below 14°C.
PE hoses 16mm16l/min – 20l/min (1l/(min · kW)Not lower than 14l/min
8.4.2 LQM+ 500
LQM+ 500
Connection diameterRecommended flow rate
PE hoses 6mm1.5l/min (1l/(min · kW)
Cooling Circuit LQM+ HP
(return flow)
Cooling Circuit LQM+ HP
(inlet flow)
Cooling Circuit LQM+ 500
(return flow)
Cooling Circuit LQM+ 500
Fig. 8.1: Water connections of the LQM+ HP and the LQM+ 500
(inlet flow)
22
Revision 02/2019 EN
LaserQualityMonitor LQM+
8.4.3 Remove the sealing plugs of the water connections
35,0(6
Release ring
1. Push
2. Pull
Fig. 8.2: Remove the sealing plugs of the water connections
1. Please push down the release ring of the connection
and pull out the plug with your free hand.
2. Remove the sealing plugs of the water connections and
keep it in a save place.
3. Close the flow line (Water In) and the return flow (Water
Out) of the device, by inserting the hose as far as possible (approx. 2cm deep).
Revision 02/2019 EN
23
35,0(
6
LaserQualityMonitor LQM+
9 Electrical connection
The LaserQualityMonitor LQM+ requires a supply voltage of 24V±5% (DC) for the operation. A suitable
power supply is included in the scope of delivery.
Please use only the provided PRIMES power supply and connection lines.
Please ensure that all electrical connections have been established and switch the device on before
starting the LaserDiagnosticsSoftware LDS.
The LaserQualityMonitor LQM+ serves as a dongle for the software on the computer in order to
enable certain software functions.
Tab. 9.1: Connection socket for the PRIMES power supply in the connection panel
1
PinFunction
1+24V
2Not assigned
3GND
4Not assigned
5FE (functional earth)
9.2.2 Inlet external trigger
BNC connector (view: connector side)
PinFunction
1+5 V (Trigger signal)
1
2
Fig. 9.2: Connection socket Inlet for an external trigger in the connection panel
2GND
9.2.3 Outlet internal trigger
BNC connector (view: connector side)
PinFunction
1+5 V (Trigger signal)
1
2
Fig. 9.3: Connection socket Outlet for the internal trigger in the connection panel
2GND
Revision 02/2019 EN
25
35,0(
6
LaserQualityMonitor LQM+
9.3 Safety facilities
9.3.1 Monitoring side access to the measuring objective and neutral-density filters
A safety switch monitors the side frame plates. When the frame plate is removed, the laser will switch off if
the safety interlock is connected.
DANGER
Serious eye or skin injury due to laser radiation
When the safety interlock is not connected, radiation (Laser class 4) reflected during measuring operation may escape from the device when the side frame plate is open.
Only operate the device with a mounted side frame plate.
X
Fig. 9.4: Device with side frame plate removed
9.3.2 Temperature control
st
and 2nd attentuators each have a temperature switch built into them that will trigger a warning mes-
The 1
sage in the LaserDiagnosticsSoftware LDS when a temperature of 75°C is exceeded. If the safety interlock
is connected, the laser will be turned off.
NOTICE
Damage/Destruction of the device
Without a safety interlock connected, the device may be damaged or destroyed when subjected to excess temperatures.
Operate the device only with a connected safety interlock.
X
The LaserQualityMonitor LQM+ should never be left unattended while taking measurements.
X
26
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
9.3.3 Safety interlock
st
The port for a safety interlock, which monitors the 1
and 2nd attenuator for excess temperature and whether
or not the side frame plates are opened, is integrated into the LaserQualityMonitor LQM+ 20 basic device.
DANGER
Serious eye or skin injury due to laser radiation; Damage/Destruction of the device
If the safety interlock is not connected, the safety equipment of the device will not be monitored.
Connect the laser control to terminals 1 to 4 so that the laser shuts off when these con-
X
nection is interrupted.
Pin assignment (view: plug side)
Safety interlock connector
2
1
8
73
Fig. 9.5: Safety interlock device plug in the connection panel
Binder M8; 86 6119 1100 00008
PinFunction
1Safety interlock (potential free).
2Safety interlock (potential free). Connected with pin 1 when ready for operation.
3Safety interlock (potential free). Connected with pin 1 when not ready for operation.
4Safety interlock (potential free).
5Safety interlock (potential free). Connected with pin 4 when ready for operation.
6Safety interlock (potential free). Connected with pin 4 when not ready for operation.
7
8Safety interlock GND
Tab. 9.2: Pin assignment safety interlock
Connected with pin 2 when ready for operation. Connected with pin 3 when not ready.
Connected with pin 5 when ready for operation. Connected with pin 6 when not ready.
+24V
64
5
When connecting the safety interlock with pin 1 and pin 3 or with with pin 4 and pin 5, it has to be ensured
that the laser is turned off when this connection is interrupted.
To be able to identify a connected safety interlock, pin 7 and pin 8 have to be bridged in the cable socket.
If the cable socket is not connected and the safety interlock is thus not active, a warning message in the
LaserDiagnosticsSoftware LDS is issued. A suitable 8-pole cable socket is part of the scope of delivery.
Revision 02/2019 EN
27
35,0(
6
LaserQualityMonitor LQM+
9.4 Connection to the PC and connect power supply
NOTICE
Damage/Destruction of the device
When the electrical cables are disconnected during operation (when the power supply is applied), voltage peaks occur which can destroy the communication components of the measuring device.
Please turn off the PRIMES power supply before disconnecting the cables.
X
1. Connect the device with the PC via a crossover cable or with the network via a patch cable.
2. Plug the power supply into the connector socket for the power supply.
PRIMES Power Supply
LQM+/LQM+ HP
Crossover Cable
Ethernet
PC
Fig. 9.6: Connection via Ethernet with a PC or a local network
or
Patch cable
Ethernet
LAN
28
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
10 Status display
The status display consists of an indicator field that indicates the various states of the LaserQualityMonitor
LQM+ with various colors.
ColorMeaning
WhiteThe device is switched on and operational
YellowMeasuring
RedLights up briefly while the measuring plane is being recorded.
Tab. 10.1: States of the status display
Tab. 10.2: Status display
Staying lit signals a device error. The device error is shown in the LaserDiagnosticsSoftware LDS.
Status display
Revision 02/2019 EN
29
35,0(
6
11 Measurement
11.1 Safety instructions
DANGER
Serious eye or skin injury due to laser radiation
During the measurement the laser beam is guided on the device, which causes scattered or
directed reflection of the laser beam (laser class 4).
The LaserQualityMonitor LQM+ must not be operated in any of the available configurations
without taking the following precautions. All precautions must be taken, even when the
fiber is in the collimator or fiber adapter.
Please wear safety goggles adapted to the power, power density, laser wave length and
X
operating mode of the laser beam source in use.
Wear suitable protective clothing and protective gloves.
X
Protect yourself from laser radiation by separating protective devices (e.g. by using appro-
X
priate shielding).
LaserQualityMonitor LQM+
DANGER
Serious eye or skin injury due to laser radiation
If the device is moved from its calibrated position, increased scattered or directed reflection
of the laser beam occurs during measuring operation (laser class 4).
When mounting the device, please ensure that it cannot be moved, neither due to an unin-
X
tended push or a pull on the cables and hoses.
DANGER
Serious eye or skin injury due to laser radiation
When the safety interlock is not connected, radiation (laser class 4) reflected during measuring operation may escape from the device when the side frame plate is open.
Connect the device’s safety interlock to the laser control. Check that the safety interlock will
X
turn off the laser properly in case of error.
NOTICE
Damage/Destruction of the device
Without safety interlock connected, the device may be damaged or destroyed when subjected to excess temperatures.
30
Connect the device’s safety interlock to the laser control. Check that the safety interlock will
X
turn off the laser properly in case of error.
The LaserQualityMonitor LQM+ should never be left unattended while taking measurements.
X
Revision 02/2019 EN
LaserQualityMonitor LQM+
4
f
rawbeam
⋅
λ
35,0(6
11.2 Selection and change of the measuring objective and the neutral density filter
11.2.1 Selection of the measuring objective
The selection of the correct measuring objective is of vital importance for the measurement quality. The limitation of use for the 1:1 or 5:1 measuring objectives is shown in the diagram Tab. 11.1 on page31.
14
12
10
8
6
4
Beam diameter in mm
2
0
110
M²
Tab. 11.1: Range of application of the LQM+ objectives
Blue area 1:1 measuring objective
Red area
Example
The following example explains how a measuring objective is selected based on the minimum beam diameter in the LaserQualityMonitor LQM+ and the number of lit pixels to be achieved depending on the measuring objective (MOB).
λ= 1064nm
M² = 1
d
= 7.3mm
rawbeam
d
foc
5:1 measuring objective
= 37.10µm (d
foc
mm
200
d
rawbeam
mm
⋅⋅
d
=
foc
= Focus diameter of the internal caustic)
π
M
2
100
For a fill factor of 35% and a resolution of 64pixel minimum 22illuminated pixels are needed.
d
Quantity of illuminated pixels at 4.4µm pixel pitch =
· β (Magnification of the measuring objective 1 or 5)
31
35,0(
6
11.2.2 Exchanging the measuring objective or the neutral density filter
For changing a measuring objective or the neutral-density filter, the plate on the side of the LaserQualityMonitor LQM+ has to be opened:
1. Turn off the laser.
2. Press down two locking bolts (see Fig. 11.1 on page32):
• The plate on the side will then jump out.
• The safety interlock is triggered (see chapter 9.3.1 on page26).
Optical components can be slid into the optical path of the LaserQualityMonitor LQM+ at three locations (see
Fig. 11.1 on page32, Positions 1, 2, 3).
SlotOptical Component
15:1 Measuring objective
21:1 Measuring objective
3Neutral density filter (OD-Filter)
4 5 6Storage slots for unused components
LaserQualityMonitor LQM+
Tab. 11.2: Slot positions of the optical components
Important:
In order to prevent measurement errors, there should only ever be one measuring objective in
the optical path at a time.
All insertions are coded differently via two dowels in order to keep the measuring objective and neutral-density filter from getting mixed up. The measuring objective and OD filter are also equipped with an electronic
memory chip (EEPROM). The EEPROM contains information on the measuring objectives and the OD filters
that is displayed in the metadata of the LaserDiagnosticsSoftware LDS.
Locking Bolts (2x)
123
4
Fig. 11.1: Slot positions for the optical components
32
Receptacles (3x)
65
Revision 02/2019 EN
LaserQualityMonitor LQM+
NOTICE
Damage of the device
Contamination in the device can damage the optical components.
Seal unused slots with the provided dummy inserts.
X
Fig. 11.2: Dummy insert
DANGER
Serious eye or skin injury due to laser radiation
35,0(6
When the safety interlock is not connected and the the side frame plate is opened, radiation
(Laser class 4) reflected during measuring operation may escape from the device.
Only operate the device with a mounted housing plate.
X
3. Place the housing plate in the receptacles (see Fig. 11.1 on page32).
4. Flip the housing plate up until the two locking bolts clip into place.
5. Check that the housing plate is plan on the casing.
11.2.3 Neutral density filter
Depending on laser beam source and configuration of the LaserQualityMonitor LQM+, it can be necessary to
reduce the power once more.
This is done with a neutral density filter which is able to go into the beam path in front of the CCD sensor.
The optical density of the filter can be varied between 1 (1:10) and 5 (1:100000).
Fig. 11.3: Neutral density filter
Revision 02/2019 EN
33
35,0(
6
LaserQualityMonitor LQM+
11.3 Connect the LaserQualityMonitor LQM+ with the LaserDiagnosticsSoftware LDS
11.3.1 Connect device
1. Turn on the LaserQualityMonitor
LQM+.
The operating mode is shown in the
status display (see chapter 10 on
page29).
2. Start the LaserDiagnosticsSoftware
LDS.
3. Click on the Devices tab.
4. Click on the + Connect to device
button under the tab.
The Connections window appears.
5. Click on the desired device.
6. Click on the Connect to device button.
11.3.2 Change the network address of a connected device
Proceed as follows in order to assign a different IP address to a connected device:
1. Right-click on the device and select
the Device settings menu point.
2. Enter in the desired IP address or
use the Use DHCP option and confirm the entry with OK.
3. Turn the LaserQualityMonitor LQM+
off and on again.
34
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
11.4 Set up the LaserQualityMonitor LQM+ with the LaserDiagnosticsSoftware LDS
After performing manual alignment with the alignment tool as described in chapter 7.2 on page15 (only
for an open beam measurement) or after inserting the fiber, you can check the alignment using a function of
the LaserDiagnosticsSoftware LDS.
11.4.1 Choose the beam pointing function
1. Connect the device according to
chapter 11.3 on page34 with the
LaserDiagnosticsSoftware LDS.
The LaserQualityMonitor LQM+ is
displayed as a connected device.
2. Click on CCD LQM+.
The menu Device control opens.
3. Click on the Beam pointing button
or
on the drop-down list measurement
mode and Beam pointing.
The corresponding Device control
opens.
The Beam pointing workbench
opens with the following tools:
• Beam pointing stability with the
windows: Position and Angle
• False-color image
Number of measuring points (planes)
Enter the number of measuring points
shown in the Beam position workbench.
It is preset to 30 planes.
Function CCD LQM+
Drop-down list Measuring mode
Revision 02/2019 EN
35
35,0(
6
11.4.2 Start beam search
1. Follow the safety instructions in chapter 11.1 on page30.
2. Turn on the laser.
3. Click on the Start button.
The progress of beam pointing
determination is indicated when Pre-caustic, Measuring beam pointing
and then Adjustment successful are
displayed in the notifications:
Pre-caustic
LaserQualityMonitor LQM+
During the pre-caustic, the optimal z1 and
z2 positions along the beam propagation
are identified in the False-color image
window (see also Fig. 11.4 on page39).
Measuring beam pointing
During the measurements, measuring
points are constantly recorded at the positions z1 and z2.
The current position of the beam is
marked with a large X, with previous positions being marked with a small x.
Align the laser beam so that the current measuring point X is centered in the
crosshairs as accurately as possible in the
Position and Angle windows.
Only for a 1
Use the adjusting screws on the 1
tenuator module.
st
attenuator:
36
st
at-
Revision 02/2019 EN
LaserQualityMonitor LQM+
The display of the beam position in the Position and Angle windows can be switched to a display of Beam
pointing stability by clicking on the gear icon
You can reset the values for beam position stability in the Position and Angle windows by pressing the Reset button.
You can log the Beam pointing stability as described in chapter 11.4.3 on page38.
.
35,0(6
Adjustment successful
As soon as the laser beam has been
properly aligned, the message Adjust-ment successful will appear
The display circle will change from red
to green.
4. Click on the Stop button.
5. Turn off the laser.
If the automated beam search is unable
to find the leaser beam, enter the positions of the measurements z1 and z2 for a
manual beam search in accordance with
chapter 11.4.4 on page39.
Revision 02/2019 EN
37
35,0(
6
11.4.3 Log beam pointing stability
You can log the beam pointing stability
under Device control > Settings:
1. Enter the value of 0 in the Plane quantity field.
2. Click on the Projects tab.
3. Click on the Start button.
• Continuous measurement data for
beam position stability is generated in
the project tree.
4. Click on the Stopbutton.
5. Click on the Save icon in order to
save the measurement data.
LaserQualityMonitor LQM+
The measuring points in the Position and Angle windows can be deleted by clicking the gear icon and the
Reset button.
The measured data in the project tree is not deleted in the process.
38
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
11.4.4 If the laser beam is not found
If the laser beam isn’t found during the automatic beam search or if the laser beam is too large for the measuring area (the laser beam exceeds the size of the CCD sensor), you can manually enter the position of the
measurements z1 and z2 into the LaserDiagnosticsSoftware LDS for the beam search. The positions z1 and
z2 define the scan range for the laser beam search.
Scaning rangeRaw beam
Focusing optics
2800
Fig. 11.4: Position of the measurements z1 and z2 for entry into the adjustment preset fields
You can enter the measuring positions for
the beam search under Device control > Settings:
1. Mark the checkbox Activate preset.
z1z2
z
0
CCD sensor
2. Enter the measurement positions for
z1 and z2 into the z-axis position (z1/z2) fields.
3. Start the beam search according to
chapter 11.4.2 on page36.
• If the laser beam still isn’t found, enter
the modified values into the z-axis position (z1/z2) fields.
z1z2
z
0
Revision 02/2019 EN
39
35,0(
6
LaserQualityMonitor LQM+
11.5 Perform an automatic caustic measurement
For the purpose of getting to know the LaserQualityMonitor LQM+, this chapter describes an automatic
caustic measurement with the LaserDiagnosticsSoftware LDS as an example.
For a detailed description of the software installation, file management and evaluation of the measured data,
please refer to the separate operating manual LaserDiagnosticsSoftware LDS.
11.5.1 Choose automatic caustic measurement mode
1. Connect the device according to
chapter 11.3 on page34 with the
LaserDiagnosticsSoftware LDS.
The LaserQualityMonitor LQM+ is
displayed as a connected device.
2. Click on CCD LQM+.
The Device control menu opens.
3. Click on the Automatic Caustic button or
on the drop-down list measurement
mode and Automatic Caustic.
The corresponding Device control
opens.
The Automatic Caustic toolbench
opens with the following tools:
• Caustic analysis
• Plane analysis
• Raw Beam Analysis
Function CCD LQM+
Drop-down list Measuring mode
40
Revision 02/2019 EN
LaserQualityMonitor LQM+
11.5.2 Configure settings (Device control > Settings)
1. Click on the Settings tab.
2. Select the calibrated wavelength of
your LQM+ in nm.
3. Enter the laser wavelength used in
nm.
4. Enter the laser output in W.
Only when measuring a time series
5. Enter the number of automatic caustic measurements.
6. Enter the measuring break between
automatic caustic measurements in s.
11.5.3 Configure advanced settings (Device control > Advanced)
1. Click on the Advanced tab.
2. Enter the number of pixels in the
x/y-direction in order to configure the
resolution on the CCD sensor.
35,0(6
3. Enter the number of planes.
Specifying 21 planes already generates a valid measurement.
Save preset
You can save information on the beam
position in a file with preset.
If the laser beam isn’t found
1. Enter the determined values for z1
and z2 from the z-axis position into
the entry fields in the Pre-caustic
area as described in chapter 11.4.4
on page39.
2. Click on the Save icon in order to
save the entry.
z1z2
z
0
Revision 02/2019 EN
41
35,0(
6
11.5.4 Starting an automatic caustic measurement
1. Follow the safety instructions in chapter 11.1 on page30.
2. Turn on the laser.
3. Click on the Start button.
The progress of the measurement is
shown in the Pre-caustic, Measure-
ment caustic displays and then in
Measurement ended:
Pre-caustic
As the indication is being displayed, the
optimal measuring parameters such as
measuring window position, measuring
window size, integration time, and the
z-range (measuring range along the beam
propagation) are determined automatically.
Measuring caustic
The actual measurement is carried out
during the indication.
LaserQualityMonitor LQM+
Measurement completed
The measurement has been completed.
4. Turn off the laser.
11.5.5 Measurement results display
The measurement results are shown in the opened tools after the measurement is completed (see below).
A detailed description of the tools and the assessment of the measuring results can be found in the separate
operating manual for the LaserDiagnosticsSoftware LDS.
42
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
11.6 Perform power measurement
With LaserQualityMonitor LQM+ devices, the power measurement can be performed at 500W with a 1st attenuator. When there is an additional 2nd attenuator, the power can be measured separately at the 2nd attenuator. The measurements at the 1st and 2nd attenuators are performed in the same way.
Separately from the power measurement described in this chapter, the power measurement values are
automatically recorded for a single plane, caustic, or automatic caustic measurement and shown under the
Details table in the workbenches.
For the purpose of getting to know the LaserQualityMonitor LQM+, this chapter describes a power measurement with the LaserDiagnosticsSoftware LDS as an example.
You can find more detailed information on LaserDiagnosticsSoftware LDS in the separate operating manual
for the LaserDiagnosticsSoftware LDS.
11.6.1 Choose power measurement mode
1. Connect the device according to
chapter 11.3 on page34 with the
LaserDiagnosticsSoftware LDS.
The LaserQualityMonitor LQM+ is
displayed as a connected device.
2. Click on Power measurement1.
• The measurement is performed for
st
the 1
attenuator.
The corresponding Device control
opens.
The Power Measurement toolbench
opens.
Revision 02/2019 EN
43
35,0(
6
11.6.2 Configure settings (Device control)
1. Start and end a measurement as
described in chapter 11.6.3 on page44.
Before the device offset can be determined, the device must go through a
thermalization period.
2. After a measurement has ended and
without the laser turned on, press the
Start button.
3. After the display in the Power Mea-
surement tool stabilizes, click on the
Stop button.
4. Click on the Set current value as
offset button.
• The value will show up in the Additional Offset in W entry window.
• The device offset can also be entered
manually.
5. Enter the measurement duration in
min.
• If a measurement duration is not
entered, the power will be measured
indefinitely.
6. Enter the Measurement frequency in Hz.
LaserQualityMonitor LQM+
11.6.3 Starting a power measurement
1. Follow the safety instructions in chap-
ter 11.1 on page30.
2. Turn on the laser.
3. Click on the Start button.
The progress of the measurement is
displayed.
When a measuring period is entered in the
Device control > Measurement duration in min. entry window, the measure-
ment is terminated automatically.
4. Click on the Stop button.
• The measurement is completed.
5. Turn off the laser.
44
Revision 02/2019 EN
LaserQualityMonitor LQM+
11.6.4 Measurement results display
The measurement results are shown in the opened tool Power Measurement after the measurement is com-
pleted (see below).
A detailed description of the tools and the assessment of the measuring results can be found in the separate
operating manual for the LaserDiagnosticsSoftware LDS.
35,0(6
Revision 02/2019 EN
45
35,0(
6
12 Troubleshooting
ErrorPossible causeRemedy
LaserQualityMonitor LQM+
Error during a
measurement
Apart from the ambient noise and zero
offset, no measuring signal is available.
The safety interlock has been triggered.The side frame plate is not
Error messages in
LaserDiagnosticsSoftware LDS
Attention, the casing is open! Please
close the casing again after changing the
measuring objectives.
No measuring objectiv or two measuring
objectives detected.
• Data transmission error
• Processor crash in the
measuring system
• Program execution error
The device is not aligned
correctly.
The power density in the
focus is too low.
closed.
The temperature sensor in
the device is measuring a
temperature >60°C.
Possible causeRemedy
The side frame plate is not
closed.
A measuring objectiv has not
been inserted or two measuring objectives have been
inserted into the device.
1. Restart the software.
2. Turn off the supply voltage, turn it on
again, and start another reset cycle.
3. Restart the computer.
Check the device alignment to the laser
beam.
Increase the laser power.
Close the side frame plate.
1. Turn off the laser immediately.
2. Check the cool water supply (only
LQM+ HP and LQM+ 500 (cooled by
water).
Close the side frame plate.
Insert only one measuring objective.
Check the position of the measuring
objective.
The measuring objectiv used is not calibrated for this device, the accuracy can
be increased considerably.
Measurement can’t be taken, the connection with the camera head has been
disrupted. Please turn the device off and
on again.
Error in saving the user-defined device
settings, the parameters must be reset.
Internal device error. Turn the device off
and on again. Device moves to the start
position during this process.
Tab. 12.1: Troubleshooting
A measuring objectiv from
a different device may have
been inserted.
The device wasn’t able to
initialize the camera head.
It wasn’t possible to save the
device settings in the device.
The device is experiencing
an internal error.
Check the inserted measuring objective.
PRIMES should calibrate the measuring
objectiv for higher accuracy.
Turn the device off and on again.
Save the device settings again.
Turn the device off and on again.
46
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
13 Maintenance and service
The operator is responsible for determining the maintenance intervals for the measuring device.
PRIMES recommends a maintenance interval of 12 months for inspection and validation or calibration.
If the device is used only sporadically, the maintenance interval can also be extended up to 24 months.
14 Storage
Please note before storing devices with water cooling circuit:
NOTICE
Damage/destruction of the device caused by leaking or freezing cooling water
Leaking cooling water can damage the device. Storing the device at temperatures near or
below freezing and without emptying the cooling circuit completely can damage the device.
Empty the lines of the cooling circuit completely.
X
Don’t use any compressed air to empty the cooling circuit.
X
To avoid contamination, please cover the apertures with the provided lid or optical tape.
X
Even when the lines of the cooling circuit have been emptied, a small amount of residual
X
water will remain in the device at all times. This may leak out and end up inside the device.
Close the connector plug of the cooling circuit with the included sealing plug.
Store the device in the original PRIMES transport box.
X
15 Measures for the product disposal
Due to the Electrical and Electronic Equipment Act (“Elektro-G“) PRIMES is obliged to dispose PRIMES measuring devices manufactured after August, 2005, free of charge.
PRIMES is a registered manufacturer in the German “Used Appliances Register“ (Elektro-Altgeräte-Register
“EAR“) with the number WEEE-reg.-no. DE65549202.
Provided that you are located in the EU, you are welcome to send your PRIMES devices to the following address, where they will be disposed free of charge (this service does not include shipping costs):
Measurement ParametersLQM+ 20LQM+ 200/500LQM+ HP (10 kW)LQM+ HP (20 kW)
Power range for 1064 nm20 W200W
Pulse duration100 fs - cw
Wavelength range
Beam dimensions1.5–9mmSingle mode:
Beam quality factor M²1 - 50
Max. beam divergence10 mrad
Determined parameters
Power density distribution2D, 3D
Supply Data
Power supply24V DC ±5%, max. 1.8A
Cooling Air coolingAir cooling (opt.
Cooling water pressure ---
340 – 360 nm (UV)/
515 – 545 nm (Green)/
1030–1090 nm (NIR)
(opt. 500W)
Water cooling)
2 bar primary pressure with an
unpressurized outflow, max. 4 bar
M² < 5 = 3kW
M² > 5 = 10kW
532nm
on request/
1030–1090 nm
(NIR)
Water cooling Water cooling
M² < 5 = 5kW
M² > 5 = 20kW
1 030–1 090 nm
(NIR)
14–16mm
Multi mode:
18–22mm
min. 4 bar
Min. Cooling water flow rate ---1,5l/min7l/min18l/min
Cooling water temperature T
1)
Please consult with PRIMES before doing anything that does not comply with this specification.
Communication
InterfaceEthernet
Dimensions and Weight
Dimensions (L x W x H)285 x 190 x 180 mm350 x 230 x 190 mm480 x 300 x 190 mm495 x 320 x 190 mm
Weightapprox. 10 kgapprox. 18 kgapprox. 35 kgapprox. 35 kg
Environmental Conditions
Operating temperature range10 – 40 °C
Storage temperature range5 – 50 °C
Reference temperature22 °C
Permissible relative humidity
(non-condensing)
1)
in
–-Dew point temperatur < Tin < 30 °C
10 – 80 %
Revision 02/2019 EN
49
35,0(
6
18 Dimensions
18.1 LQM+ 20
LaserQualityMonitor LQM+
150
Ansicht A
View A
285
A
A
175
170
All dimensions in mm (general tolerance ISO2768-v)
50
Revision 02/2019 EN
LaserQualityMonitor LQM+
A
18.2 LQM+ 200/500 (without bottom plate)
31075
35,0(6
A
285
385
299.3
413.7
View A
Ansicht A
177
150
ca. 5
169.5
129.5
89.9
All dimensions in mm (general tolerance ISO2768-v)
Revision 02/2019 EN
51
35,0(
6
18.3 LQM+ HP
104.9
109.9
101.5
320
185
LaserQualityMonitor LQM+
190
480
494.5
All dimensions in mm (general tolerance ISO2768-v)
52
Revision 02/2019 EN
LaserQualityMonitor LQM+
35,0(6
19 Appendix
19.1 Changing the factory setting on the LQM+ 20/200UV
NOTICE
Damage/Destruction of optical components
Changing the factory settings may lead to damages to the device during the measuring
operation.
Changing the factory settings for positioning range is only allowed by well trained and expe-
X
rienced personnel. In case of any doubt, please contact PRIMES service.
The length of the internal beam path is influenced by the movement of prisms inside the device. Depending
on the optical properties and the pulse parameters of a laser source, damages to the internal optical elements can occur. Especially in case of short pulsed UV-lasers the damages are foreseen.
To prevent the device from getting damaged and to avoid a positioning of a prism close to the internal focus,
a shortened positioning range can be chosen.
The LaserQualityMonitor LQM+ 20/200UV is typically delivered with the shortened positioning range as a
factory setting. Factory settings for standard and shortened positioning range are preset by a jumper, see
picture below.
You can change the positioning range by replugging a jumper in the device.
DANGER
Serious eye or skin injury due to laser radiation
If a safety interlock is not connected, a reflected beam (Class 4 Laser) may escape from the
device when the laser is turned on and the side frame plate is open.
Turn off the laser before removing the housing plate.
X
1. Turn off the laser.
2. Turn off the power supply of the
LQM+.
3. Remove the housing plate on
the side of the device by pressing down the two locking bolts
(see red arrows).
The housing plate will then jump
out.
Jumper
Revision 02/2019 EN
53
35,0(
6
NOTICE
Component susceptible to electrostatic discharge
The circuit board could be destroyed by electrostatic discharge.
Before reconnecting the jumper, put on an ESD armband.
X
4. Before switching the jumpers,
put on an ESD armband.
5. On the board you can find a
jumper which has to be brought
into the desired position.
6. Place the housing plate in the
receptacles (see Fig. 11.1 on
page32) in the casing.
7. Flip the housing plate up until the
two locking bolts clip into place.
8. Check that the housing plate is
plan on the casing.
9. Turn on the LaserQualityMonitor
LQM+.
• The device moves the prism
carriage to a reference point and
stops.
• After referencing is initiated, the
prism carriage will be located
within the allowable range.
LaserQualityMonitor LQM+
Position Shortened positioning range
(factory setting for ultra short pulse laser)
Position B: Complete positioning range
(factory setting for continuous wave laser)
54
Revision 02/2019 EN
LaserQualityMonitor LQM+
19.2 HighYAG collimation module up to 6 kW
DANGER
Serious eye or skin injury due to laser radiation
During the measurement the laser beam is guided on the device, which causes scattered or
directed reflection of the laser beam (laser class 4).
The LaserQualityMonitor LQM+ cannot be operated in any of the available configurations
without taking the following precautions. All precautions must be taken, even when the fiber
is in the collimator or fiber adapter.
Please wear safety goggles adapted to the power, power density, laser wave length and
X
operating mode of the laser beam source in use.
Wear suitable protective clothing and protective gloves.
X
Protect yourself from laser radiation by separating protective devices (e.g. by using appro-
X
priate shielding).
35,0(6
As an option PRIMES offers a 67mm HighYAG collimation module which can be mounted directly to the 2
attenuator of the LQM+ HP.
Specifications
Max. power6kW (multi mode)
3kW (single mode)
Focal length67mm
Acceptable divergence160mrad (half angle)*
Wavelength range 1025 – 1080nm
Design wave length 1064nm
Fibre core diameter10 – 1000 µm
Fibre socketLLK-D & QBH (LLK-B on request)
* If the 67mm collimation module is operated with LQM+ HP the divergence (half angle) cannot be more
than 110mrad.
Tab. 19.1: Specifications of the 67mm HighYAG collimation module up to 6kW
nd
Revision 02/2019 EN
55
35,0(
6
19.2.1 Key data for the cooling system for the HighYAG collimation module
Specifications
Max. cooling water pressure6bar
Min. cooling water flow rate2l/min
Cooling water qualityDeionized water with corrosion inhibitor
Filter mesh<100µm
Tab. 19.2: Specifications of the cooling system
Cooling circuit
LQM+ HP (return flow)
Cooling circuit
LQM+ HP (inlet flow)
LaserQualityMonitor LQM+
Fig. 19.1: Water connections
19.2.2 Schematic of the cooling system for the HighYAG collimation module
Important: The collimator cannot be connected to the cooling circuit of the LQM+ HP, but rather
to the cooling circuit of the fiber.
The two examples demonstrates the calculation of beam parameters based on a given collimator focal
length. The calculation can also be done using other collimator focal lengths.
Revision 02/2019 EN
59
35,0(
6
19.3 IPG collimation module up to 20 kW
DANGER
Serious eye or skin injury due to laser radiation
During the measurement the laser beam is guided on the device, which causes scattered or
directed reflection of the laser beam (laser class 4).
The LaserQualityMonitor LQM+ cannot be operated in any of the available configurations
without taking the following precautions. All precautions must be taken, even when the fiber
is in the collimator or fiber adapter.
Please wear safety goggles adapted to the power, power density, laser wave length and
X
operating mode of the laser beam source in use.
Wear suitable protective clothing and protective gloves.
X
Protect yourself from laser radiation by separating protective devices (e.g. by using appro-
X
priate shielding).
LaserQualityMonitor LQM+
As an option PRIMES offers a IPG collimation module which can be mounted directly to the 2
the LaserQualityMonitor LQM+.
Tab. 19.3: Specifications of the IPG collimation module up to 20kW
For the measurement of a divergent beam emerging from a fiber, the user is free to combine the LQM+ HP
(20 kW) with any collimator that is suitable for the required power range and produces a minimum of artifacts.
At the date of issue (August 2018), we are only aware of one supplier (IPG Photonics) delivering collimator
units with a specification of 20kW. As an option, PRIMES can combine the LQM+ HP with a collimator from
that supplier.
The following table lists the available collimators from that source and the nummerical aperture that can be
measured if these collimators are combined with the LQM+ HP (20kW).
nd
attenuator of
60
Revision 02/2019 EN
LaserQualityMonitor LQM+
Table2:Possible combinations of the LQM 20kW with IPG collimators and resulting NA for a measurement.Color code iscorresponding to theresulting diameter of the collimated light. Light green:diameter suitable for single modemeasurement(< 2mm*mrad),dark green: diameter suitable for multi mode measurement(> 2mm*mrad).
(full angle)f (collimator) / mm
NAdivergence / rad100120160180200
0,030,06006,007,209,6010,8012,00
0,0350,07007,008,4011,2012,6014,00
0,040,08008,009,6012,8114,4116,01
0,0450,09009,0110,8114,4116,2118,01
0,050,100010,0112,0116,0218,0220,02
0,0550,110011,0113,2217,6219,8222,03
0,060,120012,0214,4219,2321,6324,04
0,0650,130113,0215,6320,8423,4426,05
0,070,140114,0316,8422,4525,2528,06
0,0750,150115,0418,0524,0627,0730,08
0,080,160116,0519,2625,6728,8832,09
0,0850,170217,0620,4727,2930,7034,11
0,090,180218,0721,6828,9132,5236,14
0,0950,190219,0822,9030,5334,3538,16
0,10,200320,0924,1132,1536,1740,19
35,0(6
19.3.1 Possible combinations of the LaserQualityMonitor LQM+ HP (20 kW) with IPG collimators
Possible combinations of the LQM+ HP (20kW) with IPG collimators and resulting nummerical aperture (NA)
for a measurement. Color code is corresponding to the resulting diameter of the collimated light.
Light green: diameter suitable for single mode measurement (<2mm ∙ mrad),
Dark green: diameter suitable for multi mode measurement (>2mm ∙ mrad).
(Full angle)Focus (Collimator) / mm
Divergence / rad
Tab. 19.4: Possible combinations of the LaserQualityMonitor LQM+ HP (20kW) with IPG collimators
Revision 02/2019 EN
61
35,0(
6
Table2:Possible combinations of the LQM 20kW with IPG collimators and resulting NA for a measurement.Color code iscorresponding to theresulting diameter of the collimated light. Light green:diameter suitable for single modemeasurement(< 2mm*mrad),dark green: diameter suitable for multi mode measurement(> 2mm*mrad).
(full angle)f (collimator) / mm
NAdivergence / rad100120160180200
0,030,06006,007,209,6010,8012,00
0,0350,07007,008,4011,2012,6014,00
0,040,08008,009,6012,8114,4116,01
0,0450,09009,0110,8114,4116,2118,01
0,050,100010,0112,0116,0218,0220,02
0,0550,110011,0113,2217,6219,8222,03
0,060,120012,0214,4219,2321,6324,04
0,0650,130113,0215,6320,8423,4426,05
0,070,140114,0316,8422,4525,2528,06
0,0750,150115,0418,0524,0627,0730,08
0,080,160116,0519,2625,6728,8832,09
0,0850,170217,0620,4727,2930,7034,11
0,090,180218,0721,6828,9132,5236,14
0,0950,190219,0822,9030,5334,3538,16
0,10,200320,0924,1132,1536,1740,19
The use of themeasuring objectives1:1and 5:1
Using measuring objectives of 2 different magnifications, the range of M² values that can be measured is extended. The following table lists the combination of beam diameters andM² values that are accessible, andwhich measuring objective must be used for this combination.
Table3: The use of the 5:1 (dark green)and 1:1 (light green)measuring objective.Dark grey: SM beam, diameter too large.Light grey: MMbeam, spot diameter in internalbeam path too large. The numbers correspond to the spot diameter(unitµm)in the internal beam path, which must be below700 µm. Up to 1.2mm internal spot size, a measurement is possible withhigher uncertaintyof measurement yellow,(<10% diameter deviation,less than 3 Rayleigh-lengths of measurement range in zwill be available).
19.3.2 The use of the measuring objectives 1:1 and 5:1
LaserQualityMonitor LQM+
Using measuring objectives of two different magnifications, the range of
measured is extended. The following table lists the combination of beam diameters and beam quality M² values that are accessible, and which measuring objective must be used for this combination.
Dark grey: Single mode beam, diameter too large.
Light grey: Multi mode beam, spot diameter in internal beam path too large.
Beam quality M²
Beam diameter / mm
(second moment method)
beam quality M² values that can be
Tab. 19.5: The use of the 5:1 (dark green) and 1:1 (light green) measuring objective
The numbers correspond to the spot diameter (unit μm) in the internal beam path, which must be below
700μm. Up to 1.2mm internal spot size, a measurement is possible with higher uncertainty of measurement
yellow, (<10% diameter deviation, less than 3 Rayleigh-lengths of measurement range in z will be available).
19.3.3 Maximum Power
This table is a summary of the tables above and lists the maximum power the LQM+ HP can measure as a
function of beam quality M² and beam diameter.
Beam quality M²
Tab. 19.6: Maximum power
62
Beam diameter / mm
(second moment method)
Revision 02/2019 EN
LaserQualityMonitor LQM+
19.4 Optical path in LQM+ HP (with IPG Collimator)
61
f = 67
2nd Attenuator
1st Beamsplitter
Fiber
Collimator
Beam entrance LQM+
35,0(6
110.8
69.3
25
Adjustable prism
st
1
Attenuator
Fig. 19.5: Optical Path in LQM+ HP (with IPG Collimator)
20 Formula and algorithms for raw beam back calulation
To calculate the raw parameters from the measured beam parameters of the focus, the formula given in
ISO11146 is used. Index F describes the beam parameters of the focus.
z
R
Fig. 20.1: Raw beam back calculation
Beam waist radius within the raw beam
Far field divergence
H1H
23
x
1
z
0
ff
2
z
RF
1
x
2
z
0F
Position of beam waist within the raw beam
Rayleigh length within the raw beam
With focus position difference
Beam diameter onto the focusing lens
64
Revision 02/2019 EN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.