The Model SD220 Shell Detector for conversion press applications is a rugged but
sensitive dual channel instrument designed specifically to detect and report missing
blanks or double blanks at the infeed to a conversion press and to detect and report
ends that exit the press without tabs.
A complete system for one lane of a press comprises a control module housed in an
aluminum enclosure and four probes. One transmitting and receiving probe pair
senses doubles at the press input while the second transmitting and receiving pair
senses the presence or absence of tabs at the press output.
INSTALLATION
Installing the Probes ……………………………………………….….… Page 5
Mounting the Control Module ………………………….………….…….Page 6
Wiring the Contro l Mod ule ………………………………………..……..Page 6
Setting Switch Options ………………………………………….………. Page 7
Sensing Aluminum Ends ……………………………………….. Page 8
Sensing Steel Shells ……………………………………………. Page 8
Sensing Al Tabs on Al Ends ……………………………….……Page 8
Sensing Al Tabs on Steel Ends ………………………………. . . Page 8
Sensing Steel Tabs on Stee l Ends ………………………….… Page 8
Set Sinking or Sourcing Outputs ………………………………. Page 8
Set Outputs for Failsafe or Compatibility …………………….. Page 8
Enable or Disable Tab Profiling ………………………………...Page 9
Setting Options Through the Front Panel…………………………..…. Page 9
Adjusting the Tolerance……………………………………….… Page 9
Set Display Direction………………………………………….…. Page 10
Parameter Value Ranges……………………………………….. Page 11
DOCUMENT APPLICABILITY…………………………………………….……. Page 12
ELECTRICAL SPECIFICATIONS……………………………………………… Page 13
CONTROL MODULE DIMENSIONS…………………………………………... Page 14
WIRING RECOMMENDATI O NS………………………………………………. . Page 15
OUTPUT SIGNAL DIAGRAMS………………………………………………….Page 16
SUMMARY OF DIFFERNCES BETWEEN SD200 AND SD220…………… Page 17
SD200 TO SD220 MIGRATION…………………………………………………Pa ge 1 8
LIMITATION AND EXCLUSION OF WARRANTIES………………………….Page 19
3
Control Module
The control module allows for fast and easy setup and for quick diagnosis of system
errors or problems. Setup is achieved through the simple press of a push-button switch
or an external contact closure. Faults are reported on digital displays on the control
module and through FAULT outputs that may be connected to a PLC or system
controller.
The DOUBLE, SINGLE, NO TAB, TAB and FLT outputs may be switch selected as
sinking or sourcing drivers. An eight position DIP switch, accessible on the front panel,
allows the installer to select a number of operational options as described in the
installation section of this document.
The AUX inputs provide a means to switch between two different setups, such as
sensing steel shells and sensing aluminum shells, without recalibrating or changing the
DIP switch settings.
When the unit first powers up, if probes are properly connected, the digital displays
indicate the operational mode of the unit, i.e. whether set up for sensing Aluminum,
Steel, or Aluminum on Steel by displaying the characters "Al", "St" or “AS” for a period
of approximately two seconds. Which mode comes up is determined by the setup of
the unit through the DIP switches and the state of the AUX input. See setup later in this
document.
Calibrate Pushbutton Operation
The push-button switches on the front panel of the control module serve to initiate the
calibration process and to make adjustments for system setup. The pushbuttons may
be used to adjust the gauge tolerances (tL) and set the display option to invert the
digital display (do).
To initiate calibration, simply press the appropriate pushbutton and release it within 3
seconds.
To view the current value of a parameter, press the pushbutton and hold it for more
than 3 seconds until the appropriate parameter identifier (tL or do) appears on the
digital display. After the parameter identifier appears, release the pushbutton and the
current value of the parameter displays for 5 seconds. To retain the current value of
the parameter, simply allow the 5 second display interval to elapse. The display
reverts to displaying the gauge signal.
To change the value of a parameter, press the pushbutton and hold it for more than 3
seconds until the appropriate parameter identifier appears on the digital display. After
the parameter identifier appears, release the pushbutton and the current value of the
parameter displays. Press the pushbutton while the parameter is displaying and the
value increments, first slowly then more rapidly. For more precise control of the value
adjustment, simply tap the pushbutton repeatedly until the desired value is displayed.
All parameters roll back to their minimum values after reaching the maximum value.
To retain the adjusted value of the parameter, simply allow the 5 second display
interval to elapse. The display reverts to displaying signal strength.
4
Probes
The SD220 may be used with any P15 or P70 series two wire probes or with older
model three wire probes including AV, AY, AZ, AZA, and CB probes. When used with
older 3 wire probes, the white wire in the probe cable is not used and must remain
disconnected.
OPERATION
Operation of the SD220 Shell/Tab Detector involves only two processes, calibration and
fault message interpretation. These processes are described below.
Calibration
Calibration requires the following steps:
1. Stop the press in the dwell portion of the cycle with a single end between the
infeed probes and a good tabbed end between the discharge probes.
2. Observe that no error conditions are being reported on either display of the
SD220 control module.
3. Press and release the CALIBRATE pushbutton on channel 1 of the SD220
control module and observe the SINGLE and DOUBLE LEDs flashing alternately
and “CAL” on the digital display. If the calibration is successful, the LED flashing
stops in less than three seconds and the digital display shows a number between
48 and 52.
If the display flashes alternately “CAL” and “Lo”, the calibration was
unsuccessful due to insufficient signal at the receiving probe. This can be an
indication that the probes are too far apart or that the unit is not set to the
appropriate frequency for the material being sensed, e.g. frequency set to
aluminum when steel ends are present. If the display flashes alternately “CAL”
and “Hi”, the receiver signal is too strong indicating nothing between the probes.
If the calibration problem is not resolved and the calibration switch is not pressed
again within 30 seconds, calibration mode is aborted and the previous calibration
values are reinstated.
4. Press the CALIBRATE pushbutton on channel 2 of the SD220 control module
and observe the TAB and NO TAB LEDs flashing alternately and ”CAL” on the
digital display. If the calibration is successful, the LED flashing stops in less than
3 seconds and a number between 28 and 33 is displayed.
If the display begins to alternately flash “CAL” and “Lo”, the calibration was
unsuccessful due to insufficient signal at the receiving probe. This can be an
indication that the probes are too far apart or that the unit is not set to the
appropriate frequency for the material being sensed e.g. frequency set to
aluminum when steel ends are present.
5
If the calibration problem is not resolved and the calibration switch pressed again
within 30 seconds, calibration mode is aborted and the previous calibration
values are reinstated.
If both channels calibrate successfully, calibration is complete. The calibration process
may also be initiated through an external switch or signal controlling the CAL+ and
CAL- inputs to the SD220. Activation of this input performs the same function as
pressing the CALIBRATE push-button switch for the corresponding channel.
Fault Message Interpretation
The SD220 monitors the probe connections on a continuous basis and reports what it
detects to be disconnected or malfunctioning probes. The probe faults are reported as
follows:
Alternately flashing "PR" and "1" - transmitter probe disconnected or failing.
Alternately flashing "PR" and "2" - receiver probe disconnected or failing.
Alternately flashing "PR" and "3" - both probes disconnected or failing
During calibration:
Alternately flashing “CAL” and “Lo” indicates a low signal at the receiving probe.
Alternately flashing “CAL” and “Hi” indicates the signal at the receiver is too
strong, likely no end between the probes.
The SD220 performs extensive self diagnostics at power up and more limited
diagnostics while running. Most fatal faults, if not involving the display subsystem, are
reported on the digital displays through the alternate flashing of "Err" and "nnn" where
"nnn" is a one to three digit number indicating the source of the fault. These faults are
not field repairable and require the change-out of the control module.
Any detected fault causes the FLT output to be turned OFF until the fault is cleared.
INSTALLATION
Installation comprises four basic steps: 1) Installing the probes, 2) Mounting the
control module, 3) Wiring the unit, and 4) Setting system options. Each of these steps
is further expanded below.
Installing the Probes
1. Mount the double detecting infeed probes, one above and one below the
centerline of the track carrying the ends into the press. The probes must be
positioned such that they are centered on the can end during the dwell portion of
the press cycle. In the vertical, the track should run midway through a gap of
approximately 5/8 inch (16mm) between the probes.
6
2. Positioning of the tab detecting probes is more critical than the positioning of the
infeed probes. For most tabbed ends, the optimum position for the probes is
above and below the ring of the tab, not centered on the rivet. Centering on the
rivet only works on some older tab designs where the tab formed a ring around
the rivet. If positioning of the tab sensing probes is uncertain, contact Prime
Controls for help.
Mount the tab detecting probes, one above and one below the tabbed ends as
they are carried on the track from the press. The probes must be positioned
such that they are centered on the tab ring during the dwell portion of the press
cycle. In the vertical, the track should run midway through a gap of
approximately 5/8 inch (16mm) between the probes.
3. Run the probe cabling through conduit back to the cabinet housing the control
module. Do not run the sensor cables through conduit carrying high level or noisy signals.
Mounting the Control Module
Mount the control module on the back panel of an industrial enclosure. The footprint is
8.25 inches (210 mm) by 6.25 inches (159 mm) with mounting slot locations on a
rectangle 7.625 inches (194 mm) in the horizontal and 4.0 inches (102 mm) in the
vertical. Insure that the mounting screws make good electrical contact between the
module housing and the control enclosure back panel. See drawing at the end of this
document.
Wiring the Control Module
1. Connect 24 volt dc power between both sets of the +24V and COM terminals of
the control module. The left and right side modules are electrically independent
and must be powered independently. The two +24V terminals are NOT
internally connected. The supply must be capable of delivering 0.250 amps
continuously with a startup surge of 0.5 amps for 2 milliseconds.
2. Connect the double detecting transmitter probe wires to the two TMTR terminals
on the channel 1 side of the control module and the double detecting receiving
probe wires to the two RCVR terminals on the channel 1 side of the control
module. The probe connections are not polarized Connect the shield wires to
the terminal labeled SHLD.
Though both the transmitting and receiving probes are identical, it is preferred
practice to choose the transmitting probe as the one that will remain farthest
from the track as it moves and stretches.
Note: Prime Controls assures compatibility with PN: CBL101-XX cables and the
cabling pre-attached to certain probes. Other types of cabling, and also, cable
lengths above 30 meters (approx 100 ft) are not advised.
On retrofit installations where older three wire probes are installed, cut back and
do not connect the third (white) wire. If in doubt about which wires to use,
measure the resistance between the wires in pairs, and then use the pair that
produces the highest resistance reading (typically 24 ohms).
Loading...
+ 13 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.