19 20
the test being attempted. Alternatively, re-code the instrument for the respective
test strip.
E05 error indicate s data- chip error.
Solution: remove the chip and re-insert properly. If the E05 error remains, call
Customer care.
E06 error appears when insucient blood sample was applied to the test strip
application area.
Solution: remove the used test strip and repeat the measurement applying a
large drop of blood.
E07 error indicates that the data-chip is not inserted.
Solution: Insert a data-chip for proper coding of the instrument. L0 error indi-
cates that the result falls outside of the lower limit of the instrument sensitivity
(e.g. too low values).
Solution: Perform a control function check. If control passes the check, repeat
the test applying a large drop of blood and ensuring that the rst drop was wiped
out and only the second drop was te sted. If the control check fails, your instrument is not functioning properly. Contact the Customer care.
HI error appears when the result falls outside of the highe r limit of the instru-
ment’s sensitivity (e.g. too high values).
Solution: Perform a control function check. If control passes the check, repeat
the test. If the repeated result is still high make a visit to your doctor. If the control
check fails, your instrument is not functioning properly. Contact the Customer
care.
Flashing results are indicative that the test strip is too old
and may be expired. Note that ashing results are not
stored in the memory.
Battery is dying and needs replacement.
CARE AND MAINTENANCE
A clean optical measuring system is a basic prerequisite
for obtaining accurate and precise values for cholesterol and triglycerides. Therefore, regular cleans are highly
recommended.
• Always power o the meter before cleaning.
• Use ordinary lint-free cotton balls or gauze or cotton
clean tissue or cloth.
• Mild soap as well as 70% alcohol or isopropanol alcohol are suitable to moisten the cotton cloth to clean
plastic surfaces.
• Lift the front cover and gently clean glass sur faces of
the optic with a dry gauze or cotton ball.
• Clean plastic cover support of the meter with moistened and then dr y gauze or cloth.
• Do not use any disinfectant sprays or cloth/cotton balls which are dripping wet
as the liquid may penetrate and damage the instrument.
ERROR MESSAGES
In certain circumstances error messages may appear on your display.
E01 error means that the instrument or ambient temperature does not fall within
the acceptable range for the selected test.
Solution: move the instrument to an environment which has an appropriate tem-
perature (18°-35°C) and wait for the meter to equilibrate before repeating meas-
urement. Do not articially heat or cool the meter by any means.
E02 error can imply several things. First, the test strip is compromised because
of coloration of the reactive area. This might happen when a used or dirty strip is
inser ted or when a new strip was stored inappropriately. It can also mean that the
blood sample was applied too soon, before the meter displayed ashing symbol
of a blood drop.
Solution: remove and discard the used strip, open the plastic lid covering the test
strip clot and clean plastic window with a slightly moistened with water gauze.
Take a new test strip, insert it into the designate d slot and apply a blood sample
only after a blood symbol is displayed on the screen.
E03 error indicates that the strip was removed prematurely, without applying a
blood sample.
Solution: power o the meter and repeat the test after a few seconds.
E04 error appears when the test strip belongs to a dierent lot than the container
which was last coded for this parameter, or when the meter was coded for cholesterol test but a strip for triglycerides was inserted for testing.
Solution: remove the strip and repeat using a test strip matching the last code for