Vetlab Supplies Ltd Unit 13 Broomers Hill Park Broomers Hill Lane Pulborough RH20 2RY
Telephone 01798 874567 Fax 01798 874787 Email info@vetlabsupplies.co.uk
www.vetlabsupplies.co.uk
OPERATING INSTRUCTIONS
Set light dimmer control to lowest intensity. Switch on the power at the mains
and then at the back of the microscope.
Begin by placing a microscope slide containing a specimen to be observed on the
object stage and clamp it carefully with the movable spring clip. Ensure the
specimen is on the top side.
Adjust the light level using the dimmer control. Adjust the interpupillary distance
between the two eyepieces to the most comfortable position for the user. The
left eyepiece is equipped with a diopter adjustment to enable the operator to
focus his/her left eye on the specimen.
Using the X and Y Direction Knobs located just below the stage (on the right
hand side) the specimen may be moved to the center of the stage for
observation. Once the specimen is in focus, the X and Y Direction Knobs may be
adjusted to observe different sections of the specimen.
Begin observation of your specimen with the 4x Objective. Turn the coarse
focusing knob until a clear image is obtained, then use the fine focus knob to
enhance the observation of the specimen to its clearest image. On the coarse
focus knob is a tension control ring (located on the right side of the microscope
against the stand) which can be turned to tighten or loosen the tension of the
focusing knob to the user’s desired setting.
When the desired view is obtained under the lowest power (4x), rotate the
nosepiece to the next higher magnification (10x). The nosepiece should “click”
into position. It should only be necessary to adjust the fine focus knob slightly to
have a clear view of the specimen. As the magnification is increased, the user
may also find adjustments in the various methods of regulating the brightness
(see below) will improve the view of the specimen.
It is important to remember when adjusting the focus that the objective should
never touch the specimen. When increasing to 100x magnification, the objective
will appear to be very close to the slide. Because the 100x is an oil immersion
objective, a drop of immersion oil should be applied in the gap between the
objective and the specimen. The 100x oil immersion objective should be wiped
off with a piece of soft clean cloth or lens tissue to remove the immersion oil
immediately after using.
This microscope is equipped with a “locking nut” (located behind the stage
against the arm of the stand) to help prevent the stage from moving up far
enough to collide with the objectives. This locking nut is pre-set in position at
the factory and should not require adjustment. If you find that the stage is not
moving up far enough to get your specimen in focus, try turning the locking
knob slightly to allow the stage to travel farther.