premiere MRJ-01, MRJ-03T, MRJ-03D, MRJ-03 User Manual

PROFESSIONAL MICROSCOPE

Model:

MRJ-01 - Monocular

MRJ-03 - Binocular MRJ-03D - Dual View MRJ-03T - Trinocular
1
Contents MRJ
1. Before use…………………………………..………………..……3
2. Diagram…………………………………….……………...………4
3. Assemble………………………………………………..…………6
4. Operation……………………………………………...…………...6
5. General specification………………………...………………..…10
6. Configuration drawing……………………………….…………..11
7. Specification………………………………………..……………12
8. Troubleshooting…………………………………...……………..13
9. Trinocular / Dual View microscope……………...………………15
10. Warranty………………………………………………………….15
2
1. Before use MRJ
1-1. Operation
(1) As the microscope is a precision instrument, always use it with care. Avoid dropping or shaking during
operation. Make sure the microscope is placed on a flat surface.
(2) Do not put the microscope in direct contact with the sun. Do not expose the microscope to high temperature,
damp, or dusty environment.
(3) When moving the microscope, you should use both hands, typically one hand under the base and one hand on
the arm (stand) of the microscope and set it down carefully.
It will damage the microscope if you hold the microscope either by the stage, focusing knob or head. (4) Make sure the microscope is unplugged before replacing the bulb or fuse and wait until the bulb has cooled
down.
Lamp selected only 6V/20W Halogen Bulb
(5 ) This microscope has a built in dual voltage transformer and can be used with wide voltage range from 90 to
240V.
(6) Use only electrical plug supplied with the microscope.
1-2 Maintenance
(1) Wipe the lens gently with a soft lens tissue. After every use with oil or fingerprints clean the lens surfaces with
lens paper or lens paper moistened with a little rubbing alcohol. If you do not have lens paper make sure to
use a lint free towel or cloth.
Alcohol is flammable. Do not place this chemical near fire or fire causing source. You should use this
chemical in a ventilated place.
(2) Don’t use organic solution to wipe the surfaces of the other components. (3) If the microscope comes in contact with a liquid, then you should turn off the power immediately and wipe it
dry.
(4) Never disassemble or service the microscope yourself. (5) After using, cover the microscope with a dust cover.
3
2. Diagram MRJ
MRJ Intermediate Biological Microscope Diagr am
Arm
Stage Stop
Course Focus
Fine Focus
Head
Head lock screw
Eyepieces
Nosepiece
Objectives
Stage
Condenser
Condenser
up-down
knob
Variable light
control knob
Light
4
2. Diagram MRJ
MRJ Biological Microscope Diagr am
Course Focus
Aperture Diaphragm
Condenser
Condenser center
adjustment screw
Fine Focus
Condenser
up-down
knob
Mechanical Stage
Y-Axis knob
5
X-Axis knob
3. Assembly MRJ
3-1 Install the Objectives
These have been factory installed.
3-2 Insert the Eyepieces
(1) Remove the eyepiece tube covers. (2) Insert the eyepiece into the tube completely.
4. Operation MRJ
4-1 Turn on
(1) Insert the plug into an outlet and turn on. (2) Adjust the light adjustment knob until you get the suitable brightness. (See fig.1)
move the specimen to the center (the center of the
objective).
Fig.1
4-2 Set the specimen slide
(1) Push the arm of the specimen holder back.
(2) Insert the slide with the cover glass facing up
and allow the arm of the specimen holder forward. (See fig.2) (3) Rotate the X and Y-axis knobs on the mechanical stage to
6
4.Operation MRJ
4-3 Adjust focus
Course Focus
knob
Fine-focusing
-knob
3
Fig.3
.
Eye distance indicate board
Fig.4
Rotate Diopters
(1) Move the objective 4X into the light path. (2) Rotate the coarse-focusing-knob until the image
appears in focus.
(3) Turn the fine-focusing-knob to make the specimen clearer. (See fig.3)
When using the 4X and 10X objective, open the aperture diaphragm and field diaphragm
to the widest position.
4-4 Adjust the interpupillary distance
When observing with both eyes hold the bases of the base
of the eyepieces and rotate them around the axis until there
is only one field of view. (See fig.4). On the left and right
eyepiece base should point to the same number. The
number is the interpupillary distance.
4-5 Adjust diopters
Rotate one eyepiece to the “0” position using the white
thread. This should be done to your dominate eye, then
using that same (one side only) eyepiece focus on an
specimen clearly. Now observe the other eyepiece and
rotate its diopter adjusting ring until the image is clear in
that eyepiece. (See fig.5)
Fig.5
7
4. Operation MRJ
① ② ③
4-6 Center the condenser
(1) Rotate the condenser up-down knob to raise it to the
highest position. (See fig.6)
(2) Rotate the 4X objective into the light path.
Fig.6
Image of aperture diaphragm (inside)
Outside edge of objective
Fig 7
Fig 8
Fig.10
(3) Rotate the Aperture Diaphragm direction and move to the smallest setting. and you will see the
image of field diaphragm from the eyepieces.
(4) You want to see the circle of light closest to the center. If
is centered go to section 4-7, if not go to the next step
(5) Adjust the center adjustment screw
to the center of the field of view [Make sure the aperture diaphragm is
(See fig.7). When the shadow around the field of view of the
eyepiece is symmetrical, it shows the condenser has been
centered correctly. (See fig.7)
(6) Now, you can enlarge the aperture diaphragm
remove the shadow.
smaller than the field of view of the eyepiece
4-7 Adjust aperture diaphragm
The aperture diaphragm decides the numerical aperture of the illumination. If the N.A. of illumination is matching with N.A. of the objective contrast.
If the contrast is low, you can adjust the N.A. of the condenser to 70%-80% of the N.A. of the objective when using. Adjust
the upper diaphragm ring until you see the image like above (See fig.8). Use of diaphragm and put the N.A. of condenser
to 80% of the objective. e.g.: objective 40X (N.A. =0.65),
, you can obtain better resolution and the
put the aperture diaphragm to about 0.65X0.8=0.52. (See
fig.9)
in a counterclockwise
and put the light
it
.
and
Fig 9
Switch to open and close
Aperture
Diaphragm
8
4. Operation MRJ
4-8 Use the oil objective
(1) Use the objective 40X to focus the specimen. (2) Put a drop of oil on the specimen. (See fig.10) (3) Rotate the nosepiece to place the 100X objective (oil) into
the light path. Then use the fine focusing knob to focus.
Oil position
Fig.10
(4) After using oil, clean the 100X lens with clean lens paper
4-9 Use the filter
Filter holder
F
Fig.11
4-10 Replace the halogen bulb
(1)
door to the lamp holder.
(3) Wait until the bulb cools down, and then insert the new
bulb into the bulb holder on the lamp base.
(4) Close the door of the lamp holder and retighten the screw (See fig.12)
Fig.12
4-11 Replace the fuse
bottom screw off the fuse holder from the base with a “flat head”
screwdriver. Remove the damaged fuse and install the new fuse.
Fuse base
Replacement Fuse - Rating: 250V3.15A.
Fig.13
Finally screw the fuse base back to the base. (See fig.13)
Make sure there is no air bubble in the oil.
A. Take out the eyepieces to examine the air bubble. Open the
aperture diaphragm and field diaphragm fully and observe the edge of the objective from the tube without the eyepiece.
B. You can rotate nosepiece slightly and swing the oil
objective some times to remove the air bubble.
using rubbing alcohol if the oil does not come off easily.
ilter can make the background softer. (fig.11)
There is a filters: blue.
(You also place the Polarizer in this position when using the
optional simple polarizing set)
Turn off the power supply before replacing the bulb.
(2) Loosen the screw of the lamp holder base, and open the
Turn off the power supply before replacing the fuse. On the
9
5.General specification MRJ
j
r
MRJ Biological Microscope general specifications (For optional upgrade please call your authorized dealer)
Parts fications MRJ-01
Eyepiece
(pair)
Objective
Plan achromatic objective 10X
Plan achromatic objective 40X
Plan achromatic objective 100X
Head
Nosepiece
Stand
Stage
Power
Accessory
Dark field
condense
Polarizing
set
140 x 135mm mechanical stage
6V/20W halogen bulb (built in)
Speci
10X/18mm eyepiece 1 pcs
PL16X plan eyepiece
Center eyepiece
Achromatic objective 4X
Achromatic objective 10X
Achromatic objective 40X
Ac ctive 100X
hromatic obje
Plan achromatic objective 4X
45” Monocular Head
30 ocular head
°Bin
30 lar head
°Trinocu
30°Dual Binocular head
Rev ple nosepiece
ersed quadru
N.A.1.25 abbe condenser
N.A.1.25 five position phase
Si ase contrast
ngle special ph
Single special phase contrast
Single special phase contrast
V-240Vw
90 ide voltage
Variable ontrol ad
c
Blue filter(φ32
Spare bulb
Replacement fuse
Dry dark filed condenser
Oil dark field condenser
Polarizer(φ32)
Analyzer
Rotating stage (center)
C
ustment
Install
Model
MRJ-03 MRJ-03T MRJ
ir
• •
• •
-03D
2 pa
10
6. Configuration drawing MRJ
r
y
r
r
j j
MRJ Biological Microscope Configuration drawing
0.35X TV
adapter
Trinocula
Bod
Center
Eyepiece
15X
Eyepiece
Binocular
Blue
Green
Polarize
10X
Eyepiece
Filte
N.A.1.25
Oil dark field
Mechanical
moving
stage for left
condenser
Dry dark field
N.A.0.9
condenser
N.A.1.25
Abbe
achromatic
Phase
contrast
condenser
condenser
ective
ective
Phase contrast ob
Achromatic ob
11
N.A. 1.25
7. Specification MRJ
7-1 MRJ Biological Microscope general specifications
Optical Achromatic Optical system
Binocular, 30°inclined.
Head
Eyepiece 10X eyepiece, line field of view 18mm
Nosepiece Reversed quadruple nosepiece
Trinocular, 30°inclined. 45°Dual Binocular head
Objective
Focus
Stage
Condenser
Illumination
Achromatic objective(4X、10X、40X、100X)
Plan objective(4X, 10X, 40X, 100X) (not included)
Coaxial coarse & fine focusing system with limit-stopper and tension adjustment. Travel range: 20mm. Fine focusing precision:0.002mm
Mechanical stage 140x135mm; moving range 75mmx50mm; built in low position coaxial adjustment, left and right hand for option.
N.A.1.25 achromatic condenser N.A.1.25 five position phase contrast condenser (not included) N.A. 0.9 dry dark field condenser (not included) N.A. 1.25 oil dark field condenser (not included)
90240V wide voltage. Built-in Koehler illuminator system. 6V/20W halogen bulb. Pre-centered, a continuous adjustment of brightness.
7-2 Parameter of objective
Type Magnification
4x 0.1 17.912 195 45 0.17
Plan objective
10x 0.25 2.043 195 45 0.17
40x 0.65 0.65 195 45 0.17
Numerical
aperture
N.A.
Working
distance
mm
Conjugate
distance
mm
Parfocal distance
mm
Thickness of
the cover
glass
100x 1.25 0.16 195 45 0.17
4x 0.1 17.912 195 45 0.17
10x 0.25 2.043 195 45 0.17
Plan objective
40x 0.65 0.65 195 45 0.17
100x 1.25 0.16 195 45 0.17
12
8. Trouble shooting MRJ
Symptom Cause
Optics
(1)The lamp is bright, but the light in the field of view is not bright enough.
(2)The side of the field of view is dark or not even.
(3)Stain or dust is observed in the field of view.
(4)Unclear image
Field diaphragm is not large enough.
Condenser is too low.
Condenser is not centered.
The nosepiece is not in the right position.
Stain or dust has accumulated on the condenser, objective, eyepieces, and base lens.
Condenser adjustment needed (see step 3-7) Center the condenser.
Stains have accumulated on the specimen. Clean the specimen.
Stains have accumulated on the lens. Clean the lens.
No cover glass on the specimen slide. Add the cover glass.
The cover glass is not standard.
The cover glass faces down. Put the cover glass to face up.
The immersion oil has accumulated on the dry objective. The immersion oil is not used for oil objective 100XR.
Air bubble in the immersion. Get rid of the air bubble.

Remedy

Enlarge the field diaphragm.
Adjust the condenser.
Center the condenser.
Turn the nosepiece into the right position.
Clean the lens.
Use a standard cover glass with thickness 0.17mm.
Clean thoroughly.
Use immersion oil.
(5)One side of the field of view is dark or the image moves while focusing.
(6) The eyes feel tired easily. The right field of view doesn’t superpose with the left.
Use wrong immersion oil. Use correct one.
The aperture is not opened correctly. Adjust the iris diaphragm.
Stain or dust has accumulated on the lens in the inlet of the head.
The condenser is not in the right position. Adjust the condenser.
The specimen slide is not fixed. Fix with clips.
The nosepiece is not in the right position.
Condenser is centered incorrectly. Center the condenser.
Interpupillary distance is wrong. Adjust the interpupillary distance
Diopter adjustment is wrong. Adjust the diopter.
The eyepiece for the right is different from the left.
13
Clean the lens.
Turn the nosepiece into the right position.
Use the same eyepieces.
8. Trouble shooting MRJ
Symptom Cause
Mechanics
(1)Can not get the objective focused.
(2) The objective touches the cover glass while turning the nosepiece. (3)Coarse focusing knob is too tight. (4)Stage declines itself and can’t stay in the focus plane. (5)Coarse focusing knob can’t rise. (6)Coarse focusing knob can’t decline.
(7)Can not move the slide smoothly.
(8)The image moves obviously when touching the stage.
The cover glass faces down. Put the cover glass to face up.
The cover glass is not standard.
The cover glass faces down. Put the cover glass to face up.
The cover glass is not standard.
Tension knob is too tight. Loosen a little.
Tension knob is too loose. Tighten a little.
The limit stop knob is locked. Loosen the knob.
The base of the condenser is too low. Raise the base.
The slide is not fixed correctly. Adjust it correctly.
The movable specimen holder is not fixed properly.
The stage is fastened incorrectly. Fasten the stage correctly.
Electrical
No power supply.
(1)The bulb does not work.
(2)The bulb burnt out usually.
(3)The field of view is not bright enough.
The bulb is not inserted correctly. Insert it correctly.
The bulb burnt out. Replace it.
Use a wrong bulb. Replace with a correct one.
Use a wrong bulb. Replace with a correct one.
The use of light adjustment knob is wrong. Adjust correctly.

Remedy

Use a standard cover glass with thickness 0.17mm.
Use a standard cover glass with thickness 0.17mm.
Adjust it correctly.
Check the connection of the power cable.
(4)The bulb flickers or the brightness is not stable.
The bulb will burn out soon. Replace with a new one.
The wire doesn’t connect all right. Connect correctly.
14
MONOCULAR / TRINOCULAR / DUAL VIEW
MICROSCOPE
Monocular – Can be fitted with a monocular for a economical price
Trinocular - The Trinocular head fits on the base in place of binocular head, third 10X eyepiece
(not included) may be used or camera eyepiece (sold separately) can be inserted in the vertical tube for video monitor use or digital images with your computer. (Ask for MA87 or MA88).
Dual View - MRJ microscopes may be adapted to function as a Dual View microscope with front
and back viewing.
MRJ-01 MRJ-03D MRJ-03T
Warranty
The manufacturer warrants this instrument to be free from defects in material and workmanship under normal use for five years from the date of purchase (one year for electrical components). It does not cover damage resulting from abuse or misuse, repairs or alterations performed by other than authorized repair technicians, or damage occurring in transit. If you have questions concerning this product or warranty, contact the dealer from whom it was purchased. For warranty service, microscope should be well packed to avoid damage in transit, preferably in original box and packing. Include your complete return address and telephone number as well as a description of the difficulty, date and place of purchase, and ship to the address below. It will be repaired or replaced at no charge and returned. If misuse, alterations, accident or abnormal conditions of operation caused failure, an estimate for repairs will be provided for your approval prior to work being performed.
(8/3)
C & A Scientific Co., Inc.
7241 Gabe Ct Manassas Virginia 20109
15