Precision Rated Optics HCC-CW-201A Operation Manual

HCC-CW-201A
CWDM Analyzer
HCC-CW-201A
2 888-545-1254 | www.PrecisionRatedOptics.com
Table of Contents
Description ................................................................................................................................................................ 3
Features ................................................................................................................................................................... 3
Specifications ............................................................................................................................................................. 4
Safety Information .................................................................................................................................................... 5
Preparing for Operation ........................................................................................................................................... 6
Unpacking the instrument ....................................................................................................................................... 6
Front keys definition ............................................................................................................................................... 6
Rear panel definition ............................................................................................................................................... 7
Function Modules ................................................................................................................................................. 10
Description of functions ........................................................................................................................................ 11
Menu Mode .................................................................................................................................................................. 12
CWDM Parameters / CWDM Parameter setting............................................................................................... 12
Back light setting .............................................................................................................................................. 13
Time setting ...................................................................................................................................................... 13
OPM Parameters/OPM setting .......................................................................................................................... 14
View History ..................................................................................................................................................... 14
Auto off ............................................................................................................................................................ 15
Name File Name ...................................................................................................................................................... 15
Open File ..........................................................................................................................................................
About ................................................................................................................................................................ 16
Warranty ................................................................................................................................................................. 17
15
HCC-CW-201A
www.PrecisionRatedOptics.com | 888-545-1254 3
Description
HCC-CW-201A Handheld CWDM Analyzer is specially designed for CWDM installation, maintenance and troubleshooting, which is able to measure and monitor power values of 10 CWDM channels. HCC-CW-201A can replace high-cost Spectrometers and conduct quick and reliable measurements in all environments. Thanks for its light, compact and sturdy design, HCC-CW-201A is the ideal tool for CWDM installation and maintenance technicians.
Features:
Hand-held, more compact and easy to operate
TFT LCD display
Inner timing system
Threshold configuration: can preset threshold value
10CH wavelength testing: 1431nm-1611nm
Data Transfer to PC via USB
Histogram or list data display modes
Back light function
Can work as the normal power meter
10 hours continuous work data storage & download
Result display in histogram and list straight forward result display for easy understanding
Histogram mode X axis represents the channels 1-10
from 1431nm to 1611 nm. Y axis represents the power value (dBm). The red value of channel number represents cannot get through the threshold value.
List mode 10 Channels list 3 kinds of unit:
mW / uW / dBm / dB (REF). The red value represents failure to pass.
General Power Meter There are six wavelengths 850nm,
1300nm, 1310nm, 1490nm, 1550nm, 1625nm.
HCC-CW-201A
4 888-545-1254 | www.PrecisionRatedOptics.com
Specifications
Wavelength (nm) 1431 ~ 1611
Power Range (dBm) -40 ~ +10
Channel Number 10
Channel Spacing 20nm
Central Wavelength ITU
Channel Band-pass ITU ±6.5nm
Channel Power Resolution ±0.01 dB
Channel Power Accuracy ±0.5 dB
Maximum Input Power 13 dBm
Return Loss >45 dB
Close Channels Insulation 25 dB
Non-close Channels Insulation 30 dB
Units of Measurement mW/uW/dB (REF)
Data Storage 1,000 Operations
Connector FC/PC (Interchangeable SC, ST)
Power Meter
Calibrated Wavelengths (nm) 850, 1300, 1310, 1490, 1550, 1625
Power Range (dBm) -60 ~ +10
Unit mW/uW/nW/dBm/dB (REF)
Resolution ±0.25 dB (5%) @ 25 ºC
General
Power Supply NiMH Rech argeable Batter y / AC Adaptor
Battery Capacity Continuous Operation ≥10 Hours
Operating Temperature
Environmental Conditions 0 ~ 95% (Humidity)
Weight & Dimensions(HxWxD) 7.5” x 4” x 2” / 1.5 lbs. (191 x 102 x 51 mm / 680 g)
32º F ~ 122º F (0º C ~ 50º C)
HCC-CW-201A
www.PrecisionRatedOptics.com | 888-545-1254 5
Safety Information
Warnings!
Never look directly into optical outputs or a fiber while the equipment is on because invisible laser
radiation may damage your eyes.
Do not short-circuit the terminal of AC adapter, charger or the batteries. Excessive electrical current
may cause personal injury due to fumes, electric shock or equipment damage.
Connect AC power cord with the equipment and wall socket properly. While inserting the AC plug, make
sure there is no dust or dirt on the terminals and both plugs are fully seated. Incomplete engagement may cause fuming, electric shock or equipment damage and may result in personal injury.
Do not operate the equipment near hot objects, in hot environments, in dusty/ humid atmosphere or when
condensation is present on the equipment. This may result in electric shock, product malfunction or poor performance.
Discharged batteries
Remarks:
1) When the battery power is almost out, there will be a warning of indicator keeps blinking, then please
replace the batteries or plug in AC adapter to charge batteries.
2) Please make sure that you have turned the instrument on before charge the batteries, unplug the AC
adapter when the batteries are fully charged.
3) Please make sure the batteries are well placed before charging them.
4) To eliminate the possibility of acid leakage, please take out the batteries if the unit will not be used for a
long time.
AC operation
If the instrument is mainly used at one location, e.g. in a laboratory or test department, the AC adapter can be used to power it instead of batteries. There is a DC input jack on the left side of the HCC-CW-201A instrument casing into which the output cable of the AC adapter is plugged. And when the AC adapter is plugged in, the AC Indicator on the LCD will be displayed.
Note:
1) Power is supplied by the AC adapter even if battery is fitted. And the battery indicator is not displayed on
the screen when AC adapter is plugged.
2) Make sure that the operating voltage of the AC Adapter/Charger is the same as the local AC line
voltage.
HCC-CW-201A
6 888-545-1254 | www.PrecisionRatedOptics.com
Preparing for Operation
Unpacking the instrument
Packing material
We suggest that you keep the original packing material. Using the original packing material is your guarantee of protecting the instrument during transit.
Checking the package contents
The standard accessories of HCC-CW-201A are as follows:
Main unit
Quality Check Report
Carrying Case
User's Guide
4*Ni-MH Batteries
Checking for damage in transit
After unpacking the instrument, check to see whether it was damaged in transit. This is particularly likely if the outer casing is clearly damaged. If there is damage, do not attempt to operate the instrument or to repair it without authorization. Doing so can cause further damage and you may lose your warranty qualification.
Front keys definition
ON/OFF key ; Long press for 2-5 seconds to turn on the meter
Menu key; Testing mode , press to go into the menu; Menu mode, Press to get back to the former menu
Units switch key ;CWDM mode , press to switch the units dBm/dB/mW ; OPM mode , press to switch the wavelengths
Save key; Save the important data of CWDM values
Direction selecting key
Insertion key
Loading...
+ 12 hidden pages