PreciChek Cloudia User Manual

PRECICHEK Glucose Monitoring System
Cloudia
User Guide
Contents
About PRECICHEK 4 About the System 5 Characteristics 6 Blood Glucose Meter 8 Meter LCD Window 10 Blood Glucose Test Strip 12 GL™ Control Solution 15 Check Strip 18 Setting Meter Parameters 20 Prepare for Blood Sampling 23 Performing Blood Test 25 Understanding Your Test Result 31 Memory Recall 33 Data Transmission 35 Taking Care of Your Meter 36 How to Clean and Disinfect 38 Disinfect Procedures 41 References 44 Storage and Handling 46 Warranty 47 Display Messages and Problem-Solving Guide 48
System Specication 54
Performance Evaluations 56
About PRECICHEK About the System
The PRECICHEK System are designed to pursuit the accuracy in blood glucose monitoring to provide you with easy and comfortable testing.The system is next generation blood glucose monitoring system due to the new cloud technology. Through the GSM wireless transmission, healthcare professionals and distant family member will be able to access immediate information of the end user. The system requires
0.5μL of blood sample and 5 seconds for the test
to complete.
The PRECICHEK System consists of: PRECICHEK Blood Glucose Meter, PRECICHEK Blood Glucose Test Strips (Model No. : ACH), PRECICHEK Check Strip, GL™ Control Solution, Lancing Device, Lancet (optional)
These products are intended to be used together to get accurate blood glucose test results. Do not use other test strips or control solutions with your meter.
Indication for Use
PRECICHEK Cloudia Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (In vitro diagnostic use). It is indicated to be used for single patient testing blood glucose, as an aid to monitoring levels in Diabetic Mellitus. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
Principle of the Test
The PRECICHEK System is an electrochemical biosensor system that measures the amount of electric current produced and displays the result as a blood glucose level.
4 5
Characteristics
76
Important Information
The PRECICHEK System is intended for in vitro diagnostic use with capillary whole blood. The system should not be used for diagnosis of diabetes or for testing newborn infant (neonatal testing).
CAUTION
1. The user should not take any medical relevance
decision based on test result without first consulting his or her medical practitioner.
2. Call your doctor immediately if you
experience symptoms that are not consistent with your blood glucose test results.
3. Severe dehydration or excessive water loss
may cause false, high results. Call your doctor right away if you suspect you are suffering from dehydration.
4. For ACH Strip, A sample with large amount
of reducing substances such as triglycerides
(>1000mg/dL), ascorbic acid (>2.25mg/dL), uric acid (>15mg/dL) and xylose (> 50mg/dL).
5. A red blood cell count (hematocrit) that is either very high (over 55%) or very low (under 30%) can cause false result.
6. High altitudes above than 2,750 meter (8,800 feet) may affect the test results.
7. Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test results.
8. The meter and lancing device are for single patient use. Do not share them with anyone including other family members. Do not use on multiple patients.
9. All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after you have performed cleaning and disinfection.
10. Do not use PRECICHEK system to test on
critically ill patient.
NOTE: PLEASE MAKE SURE THAT ALL PRODUCTS LISTED ON THE “CONTENTS” OF THE BOX ARE CONTAINED IN THE PACKAGE BEFORE USING THIS SYSTEM. IF YOU FIND ANY IMPERFECTION IN OUR PRODUCTS, PLEASE RETURN THE
WHOLE SYSTEM TO THE PLACE OF PURCHASE.
98
Blood Glucose Meter
10
For Self-testing
Battery:AAA,6V
HMD BioMedical Inc. Hsinchu , Taiwan
3L-02-XXXXXX - Sep. 2011
IG1-0000001
8901204043780370392
Blood Glucose Monitoring System
Cloudia
Cloudia
DISPLAY: The large, easy to read display shows the test results, mes­sages, blood glucose re­sults stored in memory, time and date.
Mem BUTTON: Press Mem but­ton to enter memory mode to recall the information stored in meter’s memory.
Set BUTTON: Press Set button to enter date and time setting.
button: For data transmission
STRIP SLOT: Holds a Blood Glucose Test Strip or Check Strip in place when you perform blood glucose test or perform check test.
BATTERY COMPARTMENT: Holds four AAA batteries. Bat­teries are not yet installed into meter when new purchasing. Before using the meter, please install the batteries rst.
Meter SN Label: The label shows the meter serial number.
11
Meter LCD Window
12
DATE: Month/Year/Day and the temperature at the moment of the test.
THERMOMETER SYMBOL: Appears when ambient temperature is above or be­low the acceptable range needed for testing.
CODE: Code appears to indicate code NO. when ready to test.
appears to indicate
memory mode when reading previous results.
Symbol:The test can be adjusted and stored into 3 types of test modes - Before Meal Test (
), After Meal Test(
) and Control Solution Test(
).
Symbol: Ap­pears when data transmission.
Alarm Symbol: Appears when alarm is set.
13
Test Result
Time: 12 and 24 hour period format.
mg/dL,mmol/L: The units of measurement (mg/dL or mmol/L), mg/dL will be default unit.
Memory capacity is 999 entries. The display shows to
ensure that all digits are working properly.
Strip Symbol: Appears when the strip is inserted and meter is ready for blood testing. Replace vial cap after removing test strips.
Battery Symbol: Appears when battery is low.
:
Appears when the meter is in the memory mode while
recalling 7/14/21/28 day test averages
.
Blood Glucose Test Strip
14
ACH
Aperture: ap­ply the blood from the front strip.
Black Electrode End: Insert this end of test strip into meter.
Blood Reaction Zone: It must be filled with blood entirely.
IMPORTANT TEST STRIP INFORMATION
1. PRECICHEK System measures the amount of glucose in capillary whole blood. Blood can be applied to the front of the test strip and is
15
automatically drawn to the test strip through capillary action.
2. PRECICHEK Blood Glucose Test Strips are in­tended for in vitro diagnostic use with capil­lary whole blood or GL™ Control Solution. Re­sults will not be accurate if used with plasma or serum samples.
3. Do not use test strips beyond the expiration date indicated on the strip vial label.
4. The discard date for test strips is 90 days af-
ter rst opening the vial. Record the discard
date on the vial when you open a new vial of test strips.
5. Blood Glucose Test Strip can be damaged by heat and light. Keep them sealed in the original vial.
6.
Store the vial in a cool, dry place be­low to 40°C/104°F and above
10°C/50°F. Do not refrigerate.
16
7. Do not use damaged test strips in any way. Use test strip immediately after taking it out from the vial or foil packet; replace the vial cap and close it tightly.
8. Do not transfer test strips to a new vial. Always carry test strips in their original vial.
9.
Do not place in direct heat or sunlight.
10. Do not carry loose test strips in your carrying case.
11.
Test strips are for single use only.
GL™ Control Solution
17
GL™ Control Solution is used to check if the PRECICHEK Blood Glucose Meter and Test Strip are working correctly as a system.
It can be used in two ways:
1. To practice the test procedure
2. To make sure that the PRECICHEK Blood Glucose Meter and Test Strips are working together p r o p e r l y.
IMPORTANT INFORMATION
1. Do not use control solution beyond the expira­tion date indicated on the bottle label.
2. The discard date for control solution is 90 days
after rst opening. Record the discard date on
the bottle when you open a new bottle of con­trol solution.
3.
30
86
Store the control solution closed at temperatures between 10ºC (50ºF) and
18
30ºC (86ºF).
4. Please carefully read the label before use.
Perform Control Test
1. Insert a new test strip into the strip slot, the me­ter will activate.
2. The last code number will appear on the screen. Compare the code number shown on the screen against the code number on the test strip vial. If the two numbers match, you may begin test, otherwise consult your local distributor.
3. When shows up, press Set button and choose control solution test.
4. Gently shake the control solution and apply a drop to the aperture of strip. Make sure that the control solution has saturated the test con-
rmation window.
5. Test result will show up in 5 seconds. The re­sult should correspond to the range printed on the label of strip vial used.
19
NOTE: REPEAT TEST IF THE RESULT FALLS OUTSIDE THE CONTROL RANGE. IF YOU CONTINUE TO GET THE RESULT FALLING OUTSIDE THE CONTROL RANGE, YOUR METER AND STRIP MAY NOT BE WORKING PROPERLY. DO NOT USE THE SYSTEM TO TEST YOUR BLOOD UNTIL YOU GET A TEST RESULT FALLS WITHIN THE CONTROL RANGE. CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP.
20
Check Strip
The Check Strip can be used in 2 ways:
1. To test only the function of the meter and not complete BGM test system. Please check the complete test system with Contrl Solution
2. To delete all test memories.
CHECK
How to check meter by check strip
1. Insert the check strip into strip slot with label side up as above.
2. You should obtain an “OK“ reading within 10 seconds, which means your meter is working properly.
3. Remove the check strip to exit. Meter will automatically turn off.
NOTE: IF YOU DO NOT GET “OK” READING BUT APPEAR OTHER ERROR MESSAGE, TURN OFF THE METER BY REMOVE CHECK STRIP FROM THE METER. THEN CHECK THE BATTERY AND REPEAT THE TEST. IF THE SECOND
21
RESULT PERSISTS, CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP.
How to delete memory
1. Insert check strip into strip slot with label side up.
2. After “OK” displayed, press and hold the Set button until flashing “dEL” shows up with a beep sound.
3. Press and hold the Mem button until you hear a beep sound, Meter will display “OK” before turning off and all the memory has been deleted successfully.
4. Remove the check strip from the meter.
22
Setting Meter Parameters
Setting Time & Date
When you rst time installing the battery into the
meter or when replacing a battery, the meter will
connect with GSM system. If it succeeds to connect, the time and date will be updated automatically. If
the connection is failed, the “Err“ message will show up. Please refer to chapter “Display Messages and Problem-Solving Guide “
1. Press Mem button, the meter will enter to the setting of “year”. Press Set button to select the desired year.
2. Press Mem button to conrm and shift to the next setting.
3. Repeat above steps to set the month, day, hour
and minute by Set and Mem buttons.
4. After minute is set, the meter will display “OK”
before turning off.
NOTE: WHILE SETTING THE TIME AND DATE, YOU CAN EXIT THE SETTING MODE ANYTIME BY PRESSING Mem BUTTON FOR 3 SECONDS.
23
Setting Temperature Unit/ 12h,24h period/ Alarm
Celsius and Fahrenheit can be set according to your preference.
1. Press Set and Mem buttons at same time for 2 seconds to turn on the meter.
2. Then Press Set button to choose between “C” (for Celsius) and “F” (for Fahrenheit).
3. When Temperature Unit preference is set, press Mem button to enter the 12h/24h period setting.
4. Press Set button to choose your preference of 12h/24h period.
5. The PRECICHEK has 5 alarms to keep your blood glucose monitoring on time. Press Set button to turn on or turn off the alarm.
When “ON“ is choosen, press Mem button
to congure the alarm hour. Press Set but-
ton to change hour. When alarm hour is set, Press Mem button to configure the alarm minutes. Press Set button to change minutes
24
and press Mem button to enter next setting.
When “OFF“ is chosen, press Mem button to enter next setting.
6. To congure the other 4 alarms, please repeat step 5.
7. After setting, press and hold Mem button until the meter turning off.
Setting Auto GSM Transmission
1. Press Set button for 2 seconds to turn on the meter. When “ON“ show up, it means auto transmission..
2. Press Set button to choose between “ON“ (auto) and “OFF“ (manual). When the preference is set, press Mem button, meter will display “OK“ be­fore turning off.
25
Prepare for Blood Sampling
Adjustable Lancing Device
Lancet
5
Protective Cap
Depth Selector
Cocking Control
Release Button
Lancet Holder
Depth Adjustment Cap
Ejection Button
IMPORTANT TEST STRIP INFORMATION
1.
Lancet is for single use only.
2. Be cautious when removing the used lancet from the device and when disposing the used lancet.
3. The lancing device and lancets are in confor­mity with MDD 93/42/EEC. Refer to product labels for contact information of manufacturer and CE marking.
26
Set Lancing Device
1. Remove the depth adjust­ment cap. Insert a lancet into the lancet holder and push down until it is fully seated.
2. Twist off the protective cap un­til it separates from the lancet.
3. Replace the depth adjust­ment cap and set the puncture depth to the desired number.
NOTE: THE SHORTER THE SCALE THE SHALLOWER THE PUNCTURE.
4. Pull back the cocking control until it makes a click, press release button for penetra­tion. If it does not click, the device may have been cocked when the lancet was inserted. Please contact lo­cal distributor.
27
Performing Blood Test
1. Wash your hands in warm, soap water. Rinse and dry completely. Warm your fin-
gers to increase blood ow.
2. Remove new test strip from
vial. Be sure to tightly re­place vial cap after removing test strips. Insert test strip immediately into strip slot as illustrated. The meter turns on automatically.
Set
Mem
3. Check if the code number on
the meter matches the code on the vial. If the 2 numbers match, you may begin blood testing. Otherwise insert an­other new one. If the code number still doesn’t match, please contact your local dis­tributor.
9
28
4. When the symbol flashes, you are ready to perform a test.
5. If the test is done within 2 hours of meal, press Set but­ton to change the setting from
to . If the test is a control solution test, press Set button again to change the setting from PC
to .
NOTE: IF A TEST IS SET FOR “ ”, THE RESULT WILL NOT BE CALCU­LATED INTO AVERAGES.
29
6. Hold the prepared lancing
device rmly against the side
of your finger. Press the re­lease button. Gently massage your finger to obtain the required blood volume. To perform the test, you need only 0.5uL of blood sample
resting on your nger.
NOTE: FOR LANCING DEVICE SET­TING, PLEASE REFER TO SECTION “SET LANCING DEVICE“.
7. Apply the blood sample to the front aperture of test strip in a way that is comfort­able for you. Make sure that the blood drop has saturated the blood reaction zone.
Mem
8. Test result will show up in 5 seconds.
30
9. Before removing the strip from the meter, you can press Set button to set the test as BE­FORE MEAL TEST
, or AFTER
MEAL TEST
, or a CONTROL
SOLUTION TEST
. If not set, the default of the meter is BE­FORE MEAL TEST
.
Set
Mem
10. After setting complete, re­move the strip from meter and discard the used strip and lancet safely in a punc­ture resistant container.
Set
Mem
31
11. After removing the strip, the result will be transmit­ted to database through GSM automatically. If OK appears on screen, the transmission succeeds. If Err appears, please see Display Messages and problem solving guide to eliminate error.
12. The testing range of the me­ter is from 20 to 600 mg/dL (1.1 to 0 33.3 mmol/L). If HI is displayed, your blood glucose result may be higher than
33.3 mmol/L or 600 mg/dL. If LO is displayed, your blood glucose result may be lower than 1.1 mmol/L or 20 mg/dL.
Set
Mem
Set
Mem
32 33
Understanding Your Test Result
Normal Blood Glucose Range
According to ADA current information, glucose values are listed for an healthy adult.
Time of Day Expected Range, Non-Diabetes
Before Meals Less than 100 mg/dL
After Meals Less than 140 mg/dL
Reference: American Diabetes Association Standards of Med­ical Care in Diabetes.
Consult your healthcare professional to find out your target blood glucose value.
If your blood glucose result seems unusually high or low, or inconsistent with your previous results, check the following:
1. Does the code number on the test strip vial match the code number on the meter?
2. Was the blood sample applied to the test strip
immediately after the strip was removed from the vial?
3. Was the size of the blood sample sufcient?
4. Was the test strip vial cap tightly sealed?
5. Was the test strip used before the expiration date?
6. Were the test strips stored away from extreme temperatures or from areas of high humidity?
Then run a quality control check with your GL™ Control Solution and a new test strip. If the control test result is within the acceptable range, review testing procedure and repeat your blood glucose test with a new test strip. If your blood glucose value is still inconsistent with your previous results, contact your doctor immediately for help.
34 35
Memory Recall
The PRECICHEK Blood Glucose Meter can automati­cally store up to 999 results and auto-calculate your average results of 7-, 14-, 21-, 28- days on both before and after meal.
When recalling the results, each single result ap­pears from the latest to the earliest with time and date.
To recall Results Stored in Memory
1. Turn meter on by pressing Mem but­ton till you hear a beep sound. The first result displayed on the screen is your latest test result. 12 rep­resents the 12
th
recorded of the
overall results on this meter.
2. By pressing Mem button, you will see your test record from the latest to the oldest.
3. Press the Set button to view the average results. “19” represents within the past 7 days there were 19 -test results. Press the Set button continuously to view the average results sequentially for 14, 21 and 28 days.
4. Please press the Mem button for 2 seconds to EXIT or put it aside for 3 minutes for auto shutdown.
NOTE: ANYTIME IN MEMORY MODE, YOU CAN PRESS Mem BUTTON UNTIL YOU HEAR THE BEEP TO EXIT AND TURN OFF THE METER.
36 37
Data Transmission
The test result could be transmitted to database for further management.
1. When the meter is set “auto transmission” (refer to section “Setting auto GSM Transmission”), the test result will be transmitted to database throught GSM system after every time you test.
2. When the meter is set “manual transmission”, press “
” button for 2 seconds to turn on the
meter.
If the meter without the test record, it will dis-
play “
0”, and then “OK“ before turning off.
During transmission, it will display “OnL“. Af-
ter nishing, it will display “OK“. Press Mem
button to turn off or put it aside for 3 minutes for auto shut down.
3. When the meter display “Err”, press Mem but­ton to turn off or see the display message and problem solving guide section.
Taking Care of Your Meter
Replacing the battery
The PRECICHEK Blood Glucose Meter comes with 4 AAA battery (not yet be installed when new purchasing). Please install your batteries before starting.
Battery life will vary depending on usage, so al­ways keep a spare on hand.
The batteries should last about 24 months when testing 3 times a day.
When the battery symbol appears on the meter display, batteries level is getting low. You will still be able to test with low battery, but you should replace them as soon as possible. When battery
symbol appears ashing on the display, the meter
will no longer give results and you must replace the batteries immediately.
To Replace the Batteries
1. Make sure the meter is turned off.
2. Let the front of meter rest in the palm of your
38 39
hand.
3. Slide battery compartment door open.
4. Remove the old batteries and insert 4 new AAA batteries into the battery compartment. Be sure to align the plus (+) and minus (-) signs correctly.
5. Close the battery compartment door. Check to see that your meter is working. If it fails to turn on, the batteries may have been inserted incor­rectly. Remove the batteries and reinsert them correctly.
NOTE: EVERY TIME WHEN YOU REPLACE THE BATTERIES, THE METER WILL TURN ON AUTOMATICALLY AND ENTER TO THE TIME/DATE SETTING. PLEASE SET CORRECT TIME AND
DATE BEFORE TESTING.
How to Clean and Disinfect
The following disinfactant product have been shown to be safe for use with Cloudia meter, but any disinfectant product with the EPA registration number of 67619-12 may be used on this device.
6”
15”
40 41
Differences between Cleaning and Disinfection
According to the Merriam-Webster Medical
dictionary, the denitions of both terms are
below:
Clean - to free from the dirt, or pollution to free from disease or infectious agents.
Disinfect - to free from infection especially by destroying harmful microorganisms.
Cleaning Procedures
1. Place device and wipes on a smooth surface. Be sure there is enough light.
Set
Mem
2. Wash hand with soap and water then pat dry.
3. Take a piece of premoistened wipe out of canister. Remember to recap the canister.
4. Wipe entire device and lancing device until visibly clean.
Set
Mem
5. Throw away the used wipe. Please do not reuse the wipe.
CAUTION
1. Always clean before disinfect
2. All parts of devices are considered biohazard. Disinfect the whole parts thoroughly.
42
Disinfect Procedures
43
1. Prepare wipes and meter.
Set
Mem
2. Take a piece of pre moistened wipe out of canister. Remember to recap the canister.
3. Put the moistened wipe on a smooth surface.
Set
Mem
4. LCD side up, wipe the device from left to right 3 times slowly.
Set
Mem
5. Repeat step 4 for Front surface.
6. Repeat step 4 for left sides surface.
7. Repeat step 4 for right sides surface.
8. Repeat step 4 for Bottom surface.
9. Repeat step 4 for Top surface.
44 45
10. Make sure the meter stays moisted for 1 min.
Set
Mem
11. Discard the used wipes.
Warning
1. Please use disinfectant with caution.
2. For your own safety, please do not mix disinfectant with other liquid.
3. Please read the instructions carefully on the cainesiter before Starting cleaning and disinfection.
4. Please don’t use the wipes on any human part.
5. All parts of the glucose monitoring system should consider potentially infectious and are capable of transmitting blood­born pathogens even after cleaning and
disinfection.
6. Please do not cloud up LCD by over-moist the meter.
7. Please call us if a tear or a crack appeared on the meter.
For Lancing Device
1. For the lancing device take another new piece of pre moistened wipe out.
2. Repeat the same steps as meter.
3. Cover the device in the wipes to keep it disinfected for 1 min.
46 47
4. Discard the used wipe according to the local regulation.
NOTE
1. The Clorox Germicidal Wipes can be purchased on www. cloroxprofessional.com or call the service number at 1-800-234-7700.
2. If you have any question about the cleaning and disinfection or need more information, please contact your distributor or HMD costumer service
at 1-855-692-3511.
References
1. “FDA Public Health Notification: Use of Fingerstick Devices on More than one Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevice/
Safety/AlertsandNotices/ucm224025.htm
2. “CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/ Fingerstick-DeviceBGM.html
3. “Letter to Manufacturers of Blood Glucose Monitoring Systems Listed With the FDA” (2010)
http://www.fad.gov/MedicalDevice/ ProductsandMedical Procedures/ InVitroDiagnostic/
4. “ISO 15197 In Vitro Diagnostic Test System- Requirements for Blood- Glucose Monitoring Systems for Self-Testing in Managing Diabetes Mellitus”
48
Storage and Handling
Keep your meter free of dust or water. Protect it from
extreme temperature and humidity.
CAUTION
If the product is not used for longer periods of time, please remove the inserted battery in order to prevent damage caused by leaking battery. Please follow all the instructions on the User Guide while operating PRECICHEK Blood Glucose Meter. HMD BioMedical Inc. will not be responsible for any impairment occurred from NOT following the instructions.
49
Warranty
HMD BioMedical Inc. warrants to the original pur­chaser that this instrument will be free of defects in material and workmanship for 5 years from the date of original purchase. During the stated 5-year period, HMD shall, at no charge, replace a unit found to be defective with an equivalent or cur­rent version of the owner’s model.
This warranty is limited to replacement due to defects in parts or workmanship. HMD shall not be required to replace any units which malfunction or are damaged due to abuse, accidents, altera­tion, misuse, neglect, maintenance by other than HMD, or failure to operate the instrument accord­ing to instructions.
POST- SALE SERVICE
Questions? Please call toll free number at 1-855-692-3511
Mon-Fri 9:00am-4:30pm (Pacic Time)
50
Display Messages and Problem-Solving Guide
When any of the following messages appears, there is a problem with your PRECICHEK Blood Glucose Meter or perform wrong steps during a test. These messages will help you to identify cer­tain problems. If error messages appear that are not listed on the following pages, contact with your local distributor for help.
Display Description Action to Take
Display check
If some parts of the display are not work­ing. Contact your local distributor for help.
Blood glu­cose result may be higher than 600 mg/ dL or 33.3 mmol/L.
Review proper testing procedure and perform a quality check with control solution. Re­peat blood test, if “HI” still appears, call your doctor immediately.
51
Blood glu­cose result may be lower than 20 mg/dL or
1.1 mmol/L.
Review proper testing procedure and perform a quality check with control solution. Re­peat test, if “LO” still appears, call your doc­tor immediately.
Tempera­ture is above or below the operating range of test strips.
The result you have obtained may not be accurate. Move to an area with temperature between 10ºC to 40ºC (50ºF - 104ºF). Do not
articially heat or cool
the meter.
Battery is lo w.
Change battery imme­diately.
52
Test strip unknown or may be damaged.
Perform the test with new test strip.
Test strip is used or test was not performed correctly.
Perform the test with a new test strip and fol­low the test procedure correctly.
No re­sponses when s t r i p i s inser ted into the m e t e r.
Maybe:
1. Battery is dead.
2. W r o n g strip was inserted.
3. M e t e r i s defective.
You have to:
1. Replace battery.
2. Insert the test strip correctly.
3. Contact us for help.
53
No re-
sponses when
b l o o d
sample is a p p l i e d t o t h e strip.
Maybe:
1. B l o o d sample is not suf-
cient.
2. Me t e r i s defective.
You have to:
1. Repeat test with suf-
cient sample.
2. Perform Meter Check by inserting check strip.
Communi­cation be­tween me­ter and GSM Module­failure.
Please return the de­vice to distributor if err code 1 appears.
Device can not be rec­ognized by base.
Please return the de­vice to distributor if err code 2 appears.
54
Device can not be syn­chronous with local date/time.
Please switch to man­ual transmission and change location then try again.
Device can not con­nect to the server.
Please switch to man­ual transmission and change location then try again.
Database fail to re­turn signal.
Please switch to man­ual transmission and change location then try again.
55
System Specication
1. Assay Method: Electrochemical biosensor
2. Test Sample: Fresh Capillary Whole Blood
3. Sample Size: 0.5 μL
4. Measuring Time: 5 seconds
5. Measuring Range: 20 – 600 mg/dL (1.1 – 33.3 mmol/L)
6. Acceptable Hematocrit Range: 30%-55%
7. Operating Temp. Range: 10°C~40°C (50°F~104°F)
8. Operating Relative Humidity: 20% - 80% RH
9. Memory Capacity: 999 test results with time and date
10. Result Setting: Before meal, After meal and Control solution test setting
11. Average Display: 7, 14, 21, 28-days AC/PC average
12. Power Supply: AAA Alkaline Batteries x 4
13. Battery Life: Approximately 1000 tests
14. Automatic shut-off: In 3 minutes
15. Meter Dimension: 108.9mm (L) x 59.9mm (W) x 16.5mm (H)
16. Weight: Approximate 100g (with batteries)
58
Precision
Tests were carried by trained technicians in the laboratory setting. The venous whole blood from one subject was adjusted to 6 different levels. Strips out of a single lot were tested. The results are shown in the following table.
ACH
Level
No. of
test
Mean (mg/dL) Within-Run C.V. (%)
I
II
III
IV
V
VI
100
100
100
100
100
100
41.6
100.1
124.8
215.8
340.0
537.4
4.6
4.1
2.5
3.6
3.5
2.6
Accuracy
Tests were performed at hospital by healthcare professionals and diabetic patients. Fresh capil-
lary nger whole blood samples were tested with
Performance Evaluations
59
the PRECICHEK System; plasma samples from the subjects were tested with YSI Model 2300 Glucose Analyzer as reference. The results are shown in the following table.
Slope 0.94
Y-Intercept 13.9 mg/dL
Correlation Coefcient 0.99
Number of test 118
Range Tested 25-580 mg/dL
Reference
1. American Diabetes Association Position Statement, Diabetes Care Vol. 34 (Suppl.1), p. S13 (2011)
2. American Diabetes Association Position Statement, Diabetes Care Vol. 34 (Suppl.1), p.S21 (2011).
NOTE: THE PRODUCT IS TESTIFIED BY A THIRD PARTY THAT IT DOESN’T INTERFACE WITH OTHER POSSIBLE ELECTROMAG­NETIC WAVE.
60
Batch code Please consult instructions for use
Temperature limitation/ Store at
In vitro diagnostic medical device
Caution Use by /Expiry date
Do not reuse Serial number
Keep away from sunlight
This product fullls the require­ment of directive 98/79/EC in vitro diagnostic medical device
Biological risks Manufacturer
Authorised representative in the European community
The symbol indicating seperate collectionfor electrical and electronic equipment consists of the crossed-out wheeled bin. When the end-user wishes to discard this product, it must be sent to seperate collection facilities for recovery and recycling. By separating this product from other household-type waste, the volume of waste sent to incinerators or land-
lls will be reduced and natural resources will thus be conserved.
The content of this user guide pass the Flesch-kincaid grade level at 8th grade.
FCC Regulations:
15.19(a)(3):
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
61
15.105(b):
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
—Consult the dealer or an experienced radio/ TV technician for help.
RF Exposure Information (SAR)
This device meets the government’s requirements for exposure to radio waves.
This device is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the Federal Communications Commission of the U.S. Government.
The exposure standard for wireless mobile devices employs a unit of
measurement known as the Specic Absorption Rate, or SAR. The SAR limit
set by the FCC is 1.6W/kg. *Tests for SAR are conducted using standard operating positions accepted by the FCC with the device transmitting at its
highest certied power level in all tested frequency bands. Although the
SAR is determined at the highest certied power level, the actual SAR level
of the device while operating can be well below the maximum value. This is because the device is designed to operate at multiple power levels so as to use only the poser required to reach the network. In general, the closer you are to a wireless base station antenna, the lower the power output.
The highest SAR value for the device as reported to the FCC when tested for use when worn on the body, as described in this user guide, is 1.26 W/ kg. (Body-worn measurements differ among device models, depending upon available accessories and FCC requirements.)
While there may be differences between the SAR levels of various devices and at various positions, they all meet the government requirement.
The FCC has granted an Equipment Authorization for this device with all reported SAR levels evaluated as in compliance with the FCC RF exposure
guidelines. SAR information on this device is on le with the FCC and can
be found under the Display Grant section of www.fcc.gov/oet/ea/fccid/ after searching on FCC ID: AYY0000001.
This device is compliance with SAR for general population /uncontrolled exposure limits in ANSI/IEEE C95.1-1999 and had been tested in accordance
with the measurement methods and procedures specied in OET Bulletin 65
Supplement C.
For body worn operation, this device has been tested and meets the FCC RF exposure guidelines for use with an accessory that contains no metal and the positions the handset a minimum of 1.0 cm from the body. Use of other accessories may not ensure compliance with FCC RF exposure guidelines.
Zengqing Jian Maire-Baum-Str.3, D-76137 Karlsruhe, Germany
HMD BioMedical Inc.
Questions? Please call toll free number
at 1-855-692-3511
Mon-Fri 9:00am-4:30pm (Pacic Time)
62 63