Warranty & Additional Information ........................................ 45
General Information
Get more tools in a smaller calibrator
Carry eight single function calibrators
plus
a 4-20/10-50 milliamp calibrator with loop supply
plus
a loop troubleshooter in the palm of your hand!
Milliamp • Voltage • Frequency • pH • Ohms
Thermocouples • RTDs • Pressure
Loop Diagnostics • Transmitter Supply
Detect ‘hidden’ loop problems
Quickly diagnose troublesome ground faults & current leakage with patented Loop Diagnostic technology. These prob-
1
lems are undetectable with other instruments!
Check all loop parameters at once with LoopScope™
Simultaneously displays current, voltage and resistance
to let you know the condition of a live loop. Finds prob-
2
lems with power supplies & loops with too many loads.
Patented by PIE!
Troubleshoot wiring problems without a multimeter
Built in continuity checker with ‘beeper’ quickly finds broken
wires or shorts in instrumentation wiring. Also handy for
3
checking operation of relays and controller outputs.
Page 1
General Information
The PIE Model 850 is more than a multifunction
calibrator. It is also a loop detective that is able to
diagnose common problems that other test
equipment just can’t find. Have a flooded junction
box or unknown ground faults? Our Loop Diagnostic
technology will detect it. Or use the LoopScope to
see at a glance all the parameters - milliamps,
voltage and resistance - in the loop.
features are covered by
US Patent #7,248,058.
Diagnostic
Become a troubleshooting technician with
Patented Diagnostic Technology - Available only
with PIE Calibrators!
Page 2
Accessories
INCLUDED:
Four “AA” Lithium batteries, Certificate of Calibration
Evolution Hands Free Carrying Case Part No. 020-0211
(Evolution Hands Free Carrying Case with
Pressure Module Pocket, Part No. 020-0233,
will replace the 020-0211 when the Model 850
is ordered with at least one pressure module)
Blue Rubber Boot Part No. 020-0213
Test Leads - one pair with Part No. 020-0207
banana plug & alligator clips
Evolution RTD Wire Kit Part No. 020-0208
2 Red & 2 Black Leads with
Banana Plugs & Spade Lugs
OPTIONAL:
T/C Wire Kit 1* for Types J, K, T & E Part No. 020-0202
T/C Wire Kit 2* for Types B, R/S & N Part No. 020-0203
*Three feet (1 meter) of T/C extension wire, stripped on one
end with a miniature T/C male connector on the other end.
Page 3
Operating Instructions
FIELD & BENCH USE
PIE 850 comes with a carrying case designed for hands-free
operation and a rubber boot with a built-in tilt stand. The PIE 850
is held in the case by elastic straps for use with the carrying
case open. The tilt stand is easily raised by pulling the stand until
it locks into place.
CHANGING BATTERIES
Low battery is indicated by a battery symbol on the display.
Approximately one to four hours of typical operation remain before
the PIE 850 will automatically turn off. To change the batteries
remove the rubber boot and remove the battery door from the
back of the unit by sliding the door downward. This allows access
to the battery compartment. Replace with four (4) “AA” 1.5V
batteries being careful to check the polarity. Replace the battery
door and replace the boot. All stored configuration options (T/C
Type, EZ-CHECK Memories, etc.) are reset to factory settings
when the batteries are removed.
Note: Lithium batteries are supplied and recommended for typical
battery life and performance.
Page 4
Operating Instructions
STORING HI and LO EZ-CHECK Source Outputs
Speed up your calibration by storing Span & Zero output setting
for instant recall with the EZ-CHECK switch.
1) Store your high (SPAN) output temperature by moving the
EZ-CHECK switch to the HI position and turning the EZ-Dial
knob until the desired output value is on the display. Press
and hold the EZ-Dial knob until STORED appears to store
the value. Release the EZ-Dial knob.
2) Store your low (ZERO) output value by moving the EZ-CHECK
switch to the LO position and turning the EZ-Dial knob until
the desired output value is on the display. Press and hold the
EZ-Dial knob until STORED appears to store the value.
Release the EZ-Dial knob.
3) Instantly output your SPAN and ZERO output settings by
moving the EZ-CHECK switch between HI and LO. You may
also select any third output setting (such as mid-range) using
the SET position on the EZ-CHECK switch.
Connections
PIE 850 has banana jacks compatible with unshielded or
retractable banana plugs. Included with your calibrator are a pair
of test leads with alligator clips for mA, V, pH & Hz connections.
Four test leads with spade lugs are also included for 2, 3 and 4
Wire RTD connections. Thermocouple connections are made
through a miniature thermocouple socket.
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
Page 5
4
1
Ω
RTD
+
-
T/C
Operating Instructions
Basic Operation
PWRM
+
-
5 6
18.300 mA
850.0°C
TYPE K
SOURCE
MAX
HI
READ
SET
MIN
LO
READ
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset settings by moving the
EZ-CHECK™ switch to the “LO” position or “HI” position.
three point checks select the “SET” position. The PIE 850 will
remember the last “SET” value, even with the power off.
These values can easily be changed to suit the calibration requirements.
The values stored in the HI and LO positions are also used for Auto
Stepping.
For fast
READ: Slide the switch to the SET position. The PIE 850 will display
the current reading from the sensor or device being measured. Slide
the switch to MAX and the highest value measured since turn-on or
reset will be displayed; slide the switch to MIN and the lowest value
measured since turn-on or reset will be displayed.
Page 6
Operating Instructions
Basic Operation
w SOURCE/OFF/READ Switch
Select “SOURCE” to output mA, V, pH, T/C, Ω, RTD or frequency.
Select “READ” to read mA, V, T/C, Ω, RTD, pressure or frequency.
Select “OFF” to turn off the 850.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output level. Turn clockwise
to increase the output, counter clockwise to decrease the output in
one least significant digit step at a time. Push down and turn the
EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store desired
EZ-Check™ HI/LO points in SOURCE mode. Continue to press and
hold the knob for two more seconds to start the automatic ramping.
READ: Press and hold to transfer the current temperature
into the EZ-Check™ MIN/MAX points. This clears the MIN/
MAX readings which will update as the input value changes.
SELECTING FUNCTIONS
The EZ-DIAL knob is used to setup the PIE 850 to match the
instrument to be calibrated or signal to be measured. Each time
you turn on the PIE 850 the LCD displays the screen below for
about 1 second followed by operating in the function used the
last time it was operated.
PIE 850
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
Double Click the EZ-DIAL knob to change the function of the
calibrator and to select ranges, units and other user settings.
Each function (mA, V, T/C, Ohms, RTD, Frequency) has up to
three pages of menus. The first menu page has settings for the
function and the last menu page has settings for STEPPING,
AUTO OFF and BACKLIGHT. Settings are remembered even
with the power off but are reset when the batteries are changed.
Page 7
Operating Instructions
Double Click Menus - MAIN Page
Double click the EZ-DIAL knob to access the Double Click
Menus. Shown are the MAIN menus for each function. Turn the
knob to scroll thru the menus and press the knob to select. Default
settings are represented in black and other choices in grey.
Source mA & Simulate 2 Wire Transmitters
MAIN
> EXIT
FUNCTION mA
MODE SOURCE 2W SIM
UNITS mA %
HART 250Ω* ON OFF
Read mA, Power/Measure Transmitters
& Leak Detect
> EXIT
FUNCTION mA
MODE READ PWR MEAS
UNITS mA %
HART 250Ω* ON OFF
Source V & mV
>EXIT
FUNCTION V
RANGE 10V 100mV 1V
Read V & mV
>EXIT
FUNCTION V
RANGE 10V 1V 60V100mV
Source pH
> EXIT
FUNCTION pH
Source & Read Frequency
> EXIT
FUNCTION FREQ
RANGE 20KHZ 10000HZ 1000HZ 2000CPM
* Hart 250Ω only appears
in 4-20 mA operation
Page 8
Operating Instructions
Double Click Menus - MAIN Page
Source & Read Thermocouples
> EXIT
FUNCTION T/C
UNITS °C °F
T/C TYPE J K E T R S B N L U G C D P
COLD JUNC ON OFF
Source RTD
> EXIT
FUNCTION RTD
UNITS °C °F
RTD Pt 100 a=3850 [*RTD Types - See Read RTD]
WIRE MODE 234W
Read RTD
> EXIT
FUNCTION RTD
UNITS °C °F
RTD Pt 100 a=3850,
Pt 1000 a=3850; Cu 10 a=4274, Cu 50 a=4280
Ni 120 a=6720, Ni 110 a=5801WIRE MODE 2W 3W 4W
Source Ohms
>EXIT
FUNCTION OHMS
RANGE 40 0Ω 4000Ω
WIRE MODE 234W
a=3902, a=3916, a=3926
Read Ohms
>EXIT
FUNCTION OHMS
RANGE 400Ω 4000Ω CONT
WIRE MODE 2W 3W 4W
Read Pressure
>EXIT
FUNCTION PRESSURE
UNITS psi inH2O* ftH2O* mmH2O* cmH2O* mH2O* inHg mHg
*Engineering unit available at 4°C, 20°C and 60°F.
lb/ft
Page 9
Operating Instructions
Double Click Menu - FEATURES
To change the Automatic Stepping settings
Double click the e DIAL KNOB at any time the unit is on and
the following typical display (will be different for each FUNCTION)
will appear for 15 seconds:
MAIN
> EXIT (1/3)
FUNCTION mA
MODE SOURCE
UNITS mA
HART 250Ω* ON
Turn the
page so the word FEATURES appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e
DIAL KNOB to toggle between OFF and ON or to change the
STEPS/RAMP and the STEP/RAMP TIME settings. These
settings are remembered even with the power off.
EXIT MENU - exits this menu immediately and saves any
changes. Menu will automatically exit after 15 seconds of
inactivity.
AUTO OFF - If AUTO OFF is ON, the unit will turn off after 30
minutes of inactivity to save battery life. If AUTO OFF is OFF the
unit will stay on until the POWER SWITCH is moved to the off
position.
* Hart 250Ω only appears in 4-20 mA operation
e DIAL KNOB to move to the second or third menu
FEATURES
> EXIT (3/3)
AUTO OFF ON
BACKLIGHT ON
STEPS/RAMP 3
STEP/RAMP TIME 5
Page 10
Operating Instructions
Double Click Menu - FEATURES
STEPS/RAMP - pressing the knob will cycle through 2, 3, 5, 11
and RAMP. The endpoints of the steps or ramp are based on the
values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch bet ween the values stored in
the HI & LO EZ-CHECK (0 & 100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50
& 100%).
5 steps between the HI and LO EZ-CHECK in 25%
increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10%
increments (0, 10, 20...80, 90 &100%).
RAMP continuously ramps up and down between the HI and
LO EZ-CHECK outputs.
STEP/RAMP TIME - pressing the knob will cycle through 5, 6, 7,
8, 9, 10, 15, 20, 25, 30 and 60 seconds.
To start the Automatic Stepping
Start automatic stepping or ramping by placing the EZ-CHECK
Switch into the HI or LO position then press and hold the e
DIAL KNOB for 6 seconds (the word STORE will appear on the
display after 3 seconds and continue to press the EZ-DIAL
KNOB) until the word STEPPING appears on the display. The
word STEPPING will appear on the display anytime the selected
automatic function is running. Stop the stepping by again
pressing and holding the
e DIAL KNOB for 3 seconds.
BACKLIGHT - If BACKLIGHT is ON the backlight will light all the
time the unit is powered up. For maximum battery life turn the
backlight off when using the calibrator in areas with enough
ambient light to read the display.
Page 11
Operating Instructions
Choose 4-20 mA or 10-50 mA operation.
Works with SOURCE mA, Read mA and Simulate 2-Wire
Transmitters.
Move the power switch
click the e DIAL KNOB and the MAIN menu for the function in
use will appear for 15 seconds:
MAIN
> EXIT (1/3)
FUNCTION mA
MODE SOURCE
UNITS mA
HART 250Ω* ON
Turn the e DIAL KNOB to move the indicator to FUNCTION and
press the DIAL KNOB until mA appears. Turn the e DIAL KNOB
to move to the second menu page so the word mA RANGE
appears at the top of the menu.
2-WIRE
SIMULATE
MENU
w to SOURCE or READ and Double
mA RANGE
> EXIT (2/3)
mA RANGE 4-20mA 10-50mA
LOOPSCOPE OFF ON
POWER
MEASURE
MENU
Turn the
DIAL KNOB to toggle between 4-20mA and 10-50mA or to turn
on LOOPSCOPE (2-Wire Transmitter only) or LEAK DETECT
(POWER MEASURE only).
* Hart 250Ω only appears
in 4-20 mA operation
mA RANGE
> EXIT (2/3)
mA RANGE 4-20mA 10-50mA
LEAK DETECT OFF ON
e DIAL KNOB to move through the menu. Press the e
Page 12
Operating Instructions
10-50 mA Operation - The 10-50 mA control signal predated the
now common 4-20 mA signal. Many loops with 10-50 signal
have power supplies that run higher than 24 V DC.
SOURCE supplies from 0.000 to 52.000 mA with up to 40 V
DC compliance which has enough power to stroke valves.
2-Wire Transmitter Simulation limits the loop current passing
through it in the same manner as a 2-Wire transmitter with
loop power supplies up to 60 V DC.
READ measures from 0.000 to 52.000 mA from mA outputs
of controllers or to monitor the signal in a loop.
PWRM is POWER MEASURE which uses the internal loop
supply of the 850 to power up the transmitter while indicating
the current passing through the loop proportional to the input
of transmitter which is controlled by the output of the 850.
When simulating a 10-50 mA
2-Wire Transmitter in loops
with high voltage this section
of the calibrator may get
warm.
Page 13
2 Wire SIM mA, 2 Wire SIM % (Percent of 4-20 or 10-50 mA)
Choose this function to simulate a 2 Wire Transmitter output from
0.000 to 24.000/0.000 to 52.000 milliamps. Operates in loops
with power supply voltages from 2 to 60 VDC.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select mA for the FUNCTION and 2W SIM for the MODE.
Choose either mA or % and whether you need the 250Ω HART
resistor active in the loop (250Ω resistor is only available in
4-20mA operation).
If you would also like to see the condition of the loop turn the
DIAL KNOB to move to the second menu page so the word mA RANGE appears at the top of the menu.
> EXIT
mA RANGE 4-20mA 10-50mA
LOOPSCOPE OFF ON
LOOPSCOPE turns on the Loop Diagnostic display which
simultaneously indicates the current, voltage and resistance of
the loop. When enabled the LOOPSCOPE also indicates errors
by flashing the LED and writing diagnostic messages on the
display. Messages include LOW LOOP VOLT, HIGH LOOP VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 850
occasionally performs a test which affects the loop mA signal.
w to SOURCE then Double Click the
e
mA RANGE
2-WIRE
12.000 mA
Transmitter
2-WIRE
SUPPLY 25.4 V
LOAD 485 Ω
12.000 mA
850 with LOOPSCOPE850 Simulating a
Page 14
2 Wire SIM mA, 2 Wire SIM % (Percent of 4-20 or 10-50 mA)
Connect the output leads of the PIE 850 to the inputs of the
device being calibrated, making sure to check polarity. Red lead
to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO (defaults to 20 & 4 or
50 & 10 mA). You may also select any third output setting (such
as mid-range) using the SET position on the EZ-CHECK switch.
The output is adjusted in 0.001 mA (0.01%) increments by
turning the knob
with 0.100 mA (1.00%) increments.
e. Press and turn the knob for faster dialing
Receiver
Powers External
2-Wire Transmitter
Power Supply
(2 to 60 VDC)
All milliamp connections
2-WIRE
+
-
5 6
4.000mA
LO
All mA∙ Read V
RTD
+
23
Ω
-
65
To
Sensor
-
Source & Read Hz∙ mV
Source V∙ pH
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
(Disconnected)
4
1
Ω
RTD
+
-
T/C
Note: With LOOPSCOPE on the 850 occasionally performs a
test on the loop. During this short test the loop mA signal will be
affected.
Page 15
mA SOURCE/ % SOURCE (Percent of 4-20 or 10-50 mA)
Choose this function to provide an output from 0.000 to
24.000/0.000 to 52.000 milliamps. The compliance voltage is a
nominal 24 V DC below 29 mA and 40V DC above 30 mA to
provide the driving power to your milliamp receivers.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select mA for the FUNCTION and SOURCE for the MODE.
Choose either mA or % and whether you need the 250Ω HART
resistor active in the loop (250Ω resistor is only available in
4-20mA operation).
Connect the output leads of the PIE 850 to the inputs of the
device being calibrated, making sure to check polarity. Red lead
to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO (defaults to 20 & 4
mA or 50 & 10 mA). You may also select any third output setting
(such as mid-range) using the SET position on the EZ-CHECK
switch. The output is adjusted in 0.001 mA (0.01%) increments
by turning the knob
with 0.100 mA (1.00%) increments.
SOURCE
+
-
5 6
20.000mA
HI
w to SOURCE then Double Click the
e. Press and turn the knob for faster dialing
Milliamp Receiver Input
Controller
Transmitter
Computer
Logger
I/P
DCS
Page 16
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
All milliamp connections
+
Source V∙ pH
4
1
Ω
RTD
-
T/C
READ mA, READ % (Percent of 4-20 or 10-50 mA)
Choose this function to measure from 0.000 to 24.000/0.000 to
52.000 milliamps or -25.00 to 125.00%.
Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob
e to scroll
through the settings and press the knob to make your selection.
Select mA for the FUNCTION and READ for the MODE. Choose
either mA or % and whether you need the 250Ω HART resistor
active in the loop (250Ω resistor is only available in 4-20mA
operation).
Connect the red input lead (+) of the PIE 850 to the more
positive point of the break and the black input lead (-) to the
more negative point.
Signals below 0 mA or open circuits are indicated by 0.000 mA
(-25.00%) on the display. Signals above 24 or 52 mA are current
limited by protection circuitry with the red OVERLOAD LED lit.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob
MIN readings.
e to clear the MAX and
READ
+
-
5 6
19.999mA
MAX
All milliamp connections
Milliamp Output Signal
Controller
Transmitter
P/I
DCS
23
Ω
RTD
All mA∙ Read V
-
+
65
Page 17
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Power/Measure mA, Power/Measure % (% of 4-20/10-50 mA)
Choose this function to simultaneously supply power to a 2 Wire
Transmitter while displaying the 4.000 to 20.000/10.000 to
50.000 mA output of the transmitter.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob e to scroll
through the settings and press the knob to make your selection.
Select mA for the FUNCTION and PWR MEAS for the MODE.
Choose either mA or % and whether you need the 250Ω HART
resistor active in the loop (250Ω resistor is only available in
4-20mA operation).
Disconnect one or both input wires from the device to be
calibrated. Connect the red source lead of the PIE 850 to the
plus (+) input of the device and the black source lead to the
minus (-).
The PIE 850 supplies a nominal 24 volts DC at 24 mA or 40 volts
DC at 52 mA to the 2 Wire Transmitter. The current passed by
the transmitter will be accurately displayed by the PIE 850.
Calibrate the transmitter in the usual manner and disconnect the
PIE 850. Signals above 24 or 52 mA are current limited by
protection circuitry with the red OVERLOAD LED lit.
Transmitter Input
Sensor
Process Signal
Simulated Input
w to READ then Double Click the
PWRM
+
-
5 6
11.999mA
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
SOURCE
MAX
HI
READ
SET
MIN
LO
READ
Page 18
Using Ground Leak Detection
mA OUT, % OUT (Percent of 4-20 or 10-50 mA)
Find current leaks in loops caused by ground faults, moisture or
corrosion. The 850 simultaneously supplies power to a 2 Wire
Transmitter (or loop with a transmitter) while displaying the 4 to 20
or 10-50 mA output and the amount of current leaking in the loop.
1) Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob e to scroll
through the settings and press the knob to make your selection. Select mA for the FUNCTION and PWR MEAS for the
MODE. Choose either mA or %.
2) Turn the knob
mA RANGE
> EXIT (2/3)
mA RANGE 4-20mA 10-50mA
LEAK DETECT OFF ON
3) Turn the knob
knob to make your selection. Turn on the LEAK DETECT.
4) Connect the red source lead from the mA (+) jack of the 850
to the plus (+) input of the device and the black source lead
from the mA (-) to the minus (-).
The PIE 850 supplies a nominal 24 volts DC in 4-20 mA range or
40 volts DC in 10-50 mA range to the 2 Wire Transmitter or loop.
The current passed by the transmitter will be accurately displayed
by the 850 along with an indication of leakage current at the top
of the display. If there is an uncontrolled loop, a transmitter with
upscale burnout and bad or missing sensor or a short the red
OVERLOAD LED lights.
e until the following menu appears.
e to scroll through the settings and press the
23
Ω
RTD
All milliamp connections
Page 19
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
+
Source V∙ pH
4
1
Ω
RTD
-
T/C
Using Ground Leak Detection
12.280
PROCESS
INDICATOR
Transmitter Input
Sensor
Process Signal
Simulated Input
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
Typical Error Conditions
PWRM LEAK
+
-
5 6
00.28 mA
12.280mA
PWRM LEAK
00.28 mA
12.280 mA
The PIE 850 is supplying
the loop voltage. A
calibrated transmitter is
limiting the loop current to
12.00 mA. An additional
0.28 mA is not controlled
by the transmitter and is
leaking somewhere in the
loop.
PWRM LEAK mA
24.999 mA
The PIE 850 is supplying
the loop voltage. There is a
control loop error. This may
be a transmitter (set for
upscale burnout) with a bad
or missing sensor, or a short
in the loop. For these errors
the numbers blank and the
red OVERLOAD LED lights.
Note: Many installed transmitters will normally indicate 0.01 to
0.02 mA leakage without significant control problem. Unstable
readings may indicate loose connections or the presence of
moisture.
Page 20
mV/V SOURCE
Choose this function to provide an output from -20.000 to 99.999
mV, -500.00 TO 999.99 mV or from 0.000 to 10.250 V. The
source current is a nominal 20 mA to provide the driving power
to your voltage receivers.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select V for the FUNCTION and 1 V, 10V or 100 mV for the
RANGE.
Connect the output leads of the PIE 850 to the inputs of the
device being calibrated, making sure to check polarity. Red lead
to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in
0.001 mV, 0.01 mV or 0.001 V increments by turning the knob
Press and turn the knob for faster dialing with 0.100 mV, 1.00 mV
or 0.100 V increments.
+
-
2 1
5.000V
HI
w to SOURCE then Double Click the
e.
Voltage Receiver Input
Controller
Transmitter
Computer
Logger
DCS
V & mV Source
Page 21
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Read mV/V
Choose this function to measure from -99.999 to 99.999
millivolts, -999.99 to +999.99 mV, 0.000 to 10.250 V dc or 0.00
to 60.00 V dc.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select V for the FUNCTION and 1V, 10V, 60V or 100 mV for the
RANGE.
Connect the red input lead (+) of the PIE 850 to the more
positive point of the break and the black input to the more
negative point.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the red
OVERLOAD LED lit.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and
MIN readings.
Voltage Output Signal
+
-
2 1
8.300mV
MAX
Controller
Transmitter
Power Supply
All mA∙ Read V
Page 22
23
Ω
RTD
-
-
+
Source & Read Hz∙ mV
65
Connections for Read
10 V & 60V ranges
Connections for
Read mV ranges
4
1
Ω
RTD
+
-
T/C
Source V∙ pH
pH SOURCE
Choose this function to provide an output from 0.000 to 14.000
pH @ 25°C (77°F) which corresponds to 414.12 to -414.12 mV.
The source current is a nominal 20 mA to provide the driving
power to your pH receivers.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select pH for the FUNCTION.
Connect the output leads of the PIE 850 to the inputs of the
device being calibrated, making sure to check polarity. Red lead
from the mV (+) jack of the 850 to the plus (+) input and black
lead from the mV (-) jack to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the DIAL
position on the EZ-CHECK switch. The output is adjusted in
0.001 pH increments by turning the knob
knob for faster dialing with 0.100 pH increments.
+
-
2 1
4.010pH
ACID
w to SOURCE then Double Click the
e. Press and turn the
pH Receiver Input
Controller
Transmitter
Analyzer
pH Source
Page 23
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Simulate pH probes into transmitters & analyzers
Use the pH simulator to verify proper operation of pH devices
before you place a probe into a calibrated buffer. Adjusting the
pH transmitter or analyzer without a probe allows you to make
sure the device is calibrated and operating correctly. The 850
simulates 0.000 to 14.000 pH @ 25°C corresponding to 414.12
to -414.12 mV.
Once the pH instrument has been adjusted against the PIE 850
reconnect the pH probe and check it against the proper buffer
(typically 7 pH). If the instrument zero point requires more than
the manufacturer’s recommendations (typically within 0.5 pH) it
is time to clean or replace the probe.
Page 24
Thermocouple Source
Choose this function to provide a simulated thermocouple signal
into controllers, temperature transmitters, indicators or any input
devices that measure thermocouple sensors.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select T/C for the FUNCTION, °F or °C for the UNITS, T/C Type
(J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or P (Platinel
II)) and internal COLD JUNC ON or OFF (ON is the default).
Connect the PIE 850 to the inputs of the device being calibrated
using the proper type of thermocouple wire via the miniature
thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob
faster dialing with 10.0° increments.
-
T/C
1083.0°C
HI
TYPE K
w to SOURCE then Double Click the
e. Press and turn the knob for
Thermocouple Wire
Instrument with T/C Input
Controller
Temperature Transmitter
+
Temperature Indicator
Temperature Trip or Alarm
23
Ω
RTD
All mA∙ Read V
All thermocouple connections
-
-
+
65
Page 25
1
RTD
+
-
Source & Read Hz∙ mV
T/C
Source V∙ pH
4
Ω
Read Thermocouple Sensors
Choose this function to measure temperatures with a
thermocouple probe, sensor or any device that output a
thermocouple signal.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select T/C for the FUNCTION, °F or °C for the UNITS,
T/C Type (J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or
P (Platinel II)) and COLD JUNC ON or OFF (ON is the default).
Connect the PIE 850 to the inputs of the device being calibrated
using the proper type of thermocouple wire via the miniature
thermocouple socket. If no sensor is connected, a wire is broken
or the sensor is burned out, OPEN TC will appear on the display.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” on the display and the OVERLOAD
LED blinks red.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob
MIN readings.
830.0 °C
MIN
Thermocouple Wire
+
-
T/C
TYPE K
e to clear the MAX and
Page 26
Resistance Source
Choose this function to provide a simulated resistance into any
device that measures resistance.
Move the power switch w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select OHMS for the FUNCTION, 400Ω or 4000Ω for the
RANGE.
Disconnect all sensor wires from the devices to be calibrated
and connect the PIE 850 to the inputs of the device using 2, 3
or 4 wires.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in
0.01Ω/0.1Ω increments by turning the knob
the knob for faster dialing with 1.00Ω/10.0Ω increments.
Instrument with
Resistance Input
+-+
-
2 1
3 4
250.00 Ω
LO
Controller
Transmitter
PLC
2, 3 & 4 Wire Resistance Connections
e. Press and turn
Page 27
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Read Resistance & Check Continuity
Choose this function to measure resistance or check continuity.
Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select OHMS for the FUNCTION, 400Ω, 4000Ω or
Continuity for the RANGE. You must also select the WIRE
MODE for 2W, 3W or 4W to match the 2, 3 or 4 wires being used
to measure resistance. For continuity the only choice is 2W.
If continuity is selected, resistance is measured up to 400.0Ω.
The beeper will sound and
)))) appears on the display when
resistance below 100.0Ω is measured.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the
OVERLOAD LED blinks red.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and
MIN readings.
+
-
2 1
830.00 Ω
Page 28
RTD Source
Choose this function to provide a simulated RTD signal into
controllers, temperature transmitters, indicators or any input
devices that measure RTD sensors.
Move the power switch
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select RTD for the FUNCTION, °F or °C for the UNITS and RTD
(Choose from one of Platinum 100Ω, or 1000Ω, Copper 10Ω or
50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850 is the most common RTD type.
Disconnect all sensor wires from the devices to be calibrated
and connect the PIE 850 to the inputs of the device using 2, 3
or 4 wires.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob
faster dialing with 10.0° increments.
+-+
-
2 1
3 RTD 4
830.0 °C
HI
Pt 100 α=3850
w to SOURCE then Double Click the
e. Press and turn the knob for
Instrument with RTD Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
2, 3 & 4 Wire RTD Connections
Page 29
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Read RTD Sensors
Choose this function to measure temperatures with an RTD
probe, sensor or any device that output an RTD signal.
Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select RTD for the FUNCTION, °F or °C for the UNITS
and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper
10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850 is the most common RTD type. You must also select the WIRE
MODE for 2W, 3W or 4W to match the 2, 3 or 4 wires on the RTD
sensor.
Disconnect all wires from the sensor and connect the PIE
820-ELITE to the inputs of the device using 2, 3 or 4 wires.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the red
OVERLOAD LED lit.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and
MIN readings.
+
-
-
2 1
3 RTD
83.0 °C
Pt 100 α=3850
Page 30
Frequency Source
Choose this function to provide a frequency signal into any input
devices that measure frequency.
Move the power switch w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select FREQ for the FUNCTION and 20KHZ, 10000HZ, 1000HZ
or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and
connect the PIE 850 to the input of the device matching polarity.
The green HZ SYNC LED pulses in synch with the output pulses
and may be used to calibrate optical pickups. The output signal
is a zero crossing square wave with a fixed amplitude of 6 V
peak-to-peak from -1 and + 5 V.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 1
count increments by turning the knob
knob for faster dialing with 100 count increments.
Frequency Receiver Input
+
-
2 1
8300.0Hz
HI
Transmitter
e. Press and turn the
Flowmeter
Computer
Logger
Controller
DCS
All Frequency
Connections
Page 31
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Read Frequency
Choose this function to count frequency.
Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select FREQ for the FUNCTION and 20KHZ, 10000HZ, 1000HZ
or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and
connect the PIE 850 to the output of the device matching
polarity.
The green HZ SYNC LED pulses in synch with the input
frequency.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the red
OVERLOAD LED lit.
The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and
MIN readings.
Frequency Output Signal
+
-
2 1
830.00 Hz
Flowmeter
Flow Sensor
Variable Speed Drive
Controller
Transmitter
All Frequency
Connections
Page 32
23
Ω
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
Ω
RTD
+
-
T/C
Read Pressure
Choose this function to measure pressure in one of 32 different
engineering units using a PIE Pressure Module.
1) Move the power switch w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select PRESSURE for the FUNCTION and make your choice of
UNITS to match the pressure instrument to be checked.
2) Remove the covers from the ends of the connector on the
pressure module cable and the pressure connector of the 850.
Align the white arrows and plug the cable into the 850.
3) Connect pressure hoses, fittings & pumps (if required) to the
pressure instrument to be checked.
4) Press and hold the
e E-Z DIAL KNOB for 2 seconds (after
MAX/MIN RESET appears) to 'Zero' or 'Tare' the pressure. The
display will briefly display 'TARE ON' then a '
0' appears on the
display indicating that all measurements are relative to the
pressure measured when the calibrator was zeroed. Press and
hold the e E-Z DIAL KNOB for 2 seconds (after MAX/MIN
RESET appears) when you want to turn the 'Tare' off.
The PIE 850 measures the pressure and constantly updates the
display with the current reading. Move the EZ-CHECK switch
q
to MAX to see the highest reading and to MIN to see the lowest
reading. Press and hold the knob e for 1 second to clear the
MAX and MIN readings.
To disconnect the pressure module gently pull up on the groved
area just above the white arrow.
23
Ω
RTD
All mA∙ Read V
-
-
+
65
Port for Pressure Module
1
RTD
+
-
Source & Read Hz∙ mV
T/C
Source V∙ pH
4
Ω
Page 33
Read Pressure
PIE 850 with Pressure Module,
Pressure/Vacuum Pump & Hose
Hands free carrying case with pockets for the PIE 850 and
the pressure module. Back of the case has a zippered pocket
for the manual, test leads, hoses and pressure fittings.
Page 34
Optional Pressure Modules
Sensor CodeApplication
DNxxxxDifferential, Non-isolated
0 to 0010*, 0028, 0200, 0415, 2000” H2O
DIxxxxDifferential, Isolated
0 to 0001, 0005, 0015, 0030, 0100, 0300, 0500 PSID
Non-isolated DN sensors: clean, dry, non-corrosive,
non-condensing gases only
Isolated DI sensors: any media compatible with 316L SS &
Viton®
Isolated GI, CI & AI sensors: any media compatible with 316L SS
Pressure Module Accuracy
±0.025% of full scale including all effects of linearity,
repeatability and hysteresis from -20° to +50°C (-4° to +122°F)
* The DN0010 sensor accuracy is ±0.050% of full scale
32 Engineering Units:
PSI • inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F
inches, meter, cm and mm of Hg @ 0°C
torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar
ATM • oz/in2 • lb/ft2
Page 35
Specifications
General
Operating Temp Range
Storage Temp Range-30 to 60 °C (-22 to 140 °F)
Temperature effect≤ ± 0.01 %/°C of Full Scale; Cold
Source mA: ≥ 14 hours @ 12 mA into 250Ω
≥ 9.5 hours @ 30 mA into 250Ω
Pwr/Meas: ≥ 12 hours @ 20 mA, ≥ 4.5
hours @ 50 mA
Source V, Ω, T/C, pH, RTD & Hz: ≥ 20 hours
hour of operation left
Over-voltage protection to 60 vrms
(rated for 30 seconds)
Red LED indicates OVERLOAD or out
of range conditions
High contrast graphic liquid crystal display,
0.315” (8.0 mm) high digits, backlighting
Page 36
Specifications
Read mA
Ranges and Resolution
Accuracy≤ ± (0.02 % of Reading + 0.003 mA)
Voltage burden≤ 2V at 24 mA or 52 mA.
Overload/Current
limit protection
Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution
Accuracy≤ ± (0.02 % of Reading + 0.003 mA)
Loop compliance voltage
Loop drive capability 1200 Ω at 20 mA for 15 hours nominal;
mA 2-Wire Transmitter Simulation
AccuracySame as Source/Power & Measure
Voltage burden≤ 2V at 20 mA or 52 mA
Overload/Current limit 24 mA nominal in 20 mA range, 53 mA
Loop voltage limits2 to 60 VDC (fuse-less protected)
0.000 to 24.000 mA or -25.00 to 125.00%
of 4-20 mA, 0.000 to 52.000 mA or -25.00
to 105.00% of 10-50 mA
25 mA nominal in 20 mA range
52.5 mA nominal in 50 mA range
Same as Read mA
≥ 24 DCV @ 20 mA, ≥ 40 DCV @ 52 mA
800 Ω at 50 mA for 4.5 hours nominal
Subtract 250 Ω with Hart Resistor
enabled (4-20 only)
nominal in 50 mA range
DC Voltage Read
Range and Resolution
Accuracy
Input resistance≥ 1 MΩ
±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00
to 60.00 V DC
≤ ± (0.02 % of Reading + 0.01% Full Scale)
Page 37
Specifications
Source V dc
Ranges and Resolution
Accuracy
Source Current
Sink Current> 16 mA
Output Impedance< 1 Ohm
Short Circuit Duration Infinite
pH Source
Range and Resolution
Accuracy in mV
Accuracy in pH≤ ± 0.003 pH @ 25°C
Thermocouple Source
Accuracy≤ ± (0.02 % of Reading + 0.01 mV)
Cold Junction Compensation
Output Impedance< 1 Ohm
Source Current> 20 mA (drives 80 mV into 10 Ohms)
Thermocouple Read
Accuracy & Cold
Junction Compensation
Input Impedance
Open TC Threshold;
Pulse
-20.000 to 99.999 mV, -500.00 to 999.99 mV,
0.000 to 10.250V
≤ ± (0.02 % of Reading + 0.01% Full Scale)
≥ 20 mA
0.000 to 14.000 pH
≤ ± (0.02 % of Reading in mV + 0.1 mV)
± 0.05°C; Thermistor traceable to
NIST for 11 years
Same as Thermocouple Source
> 1 Megohms
10K Ohms; <5 µamp pulse for 300
milliseconds (nominal)
Page 38
Specifications
RTD, OHMS and Continuity Read
Resistance Ranges0.00 to 401.00, 0.0 to 4010.0 Ohms
Accuracy±(0.025% of Reading + 0.075 Ohms)
Excitation Current
Continuity0.0 to 401.0 Ohms; Beeps from 0.0 to
RTD and OHMS Source
3 Wire & 4 Wire
Accuracy
F ro m 1 t o 1 0. 2 m A
External Excitation Current
B e l o w 1 m A o f E x t e r n a l
Excitation Current
2 Wire Accuracy
Resistance Ranges0.00 to 401.00, 0.0 to 4010.0 Ohms
Allowable Excitation
Current Range
Pulsed Excitation
Current Compatibility
1.0 mA to 401 Ohms, 0.5 mA to 4010
Ohms (nominal)
100.0 Ohms
±(0.025% of Full Scale + 0.075 Ohms)
±(0.025% of Full Scale+0.075 Ohms +
0.025 mV
mA Excitation Current
Add 0.1 Ohm to specifications listed above
<401 Ohm:10.2 mA max; steady or pulsed/
intermittent
401 to 4000 Ohms: 1 mA max; steady or
pulsed/intermittent
DC to 0.01 second pulse width
)
Page 39
Specifications
Frequency Source
Ranges1 to 2000 CPM, 0.01 to 999.99 Hz, 0.1
Accuracy
Output Waveform
Risetime (10 to 90% of
amplitude)
Output Impedance< 1 Ohm
Source Current> 1 mA rms at 20 kHz
Short Circuit Duration Infinite
Optical CouplingGreen LED (HZ SYNC) flashes at
Frequency Read
Ranges & AccuracySame as Frequency Source
Accuracy
Trigger Level1 V rms, dc coupled
Input Impedance> 1 Meg Ohm + 60 pF
to 9999.9 Hz, 0.001 to 20.000 kHz
± (0.02 % of Reading + 0.01% Full Scale)
Square Wave, Zero Crossing -1.0 to +5 V
peak-to-peak ±10%
< 10 microseconds
output frequency
± (0.02 % of Reading + 0.01% Full Scale)
Page 40
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/CDegrees C
Range
J-200.0 to -50.0±0.5°-328.0 to -58.0±1.0°
-50.0 to 300.0±0.2°-58.0 to 572.0±0.4°
300.0 to 900.0±0.3°572.0 to 1652.0±0.6°
900.0 to 1200.0±0.4°1652.0 to 2192.0±0.8°
K-230.0 to -50.0±1.2°-382.0 to -58.0±2.2°
-50.0 to 550.0±0.3°-58.0 to 1022.0±0.6°
550.0 to 1000.0±0.5°1022.0 to 1832.0±0.8°
1000.0 to 1371.1±0.6°1832.0 to 2500.0±1.1°
T-260.0 to -230.0±2.9°-436.0 to -382.0±5.2°
-230.0 to -210.0±1.0°-382.0 to -346.0±1.9°
-210.0 to -50.0±0.8°-346.0 to -58.0±1.4°
-58.0 to 50.0±0.3°-58.0 to 122.0±0.6°
50.0 to 400.0±0.2°122.0 to 752.0±0.4°
E-240.0 to -200.0±0.9°-400.0 to -328.0±1.7°
-200.0 to 0.0±0.5°-328.0 to 32.0±0.8°
0.0 to 350.0±0.2°32.0 to 662.0±0.3°
350.0 to 1000.0±0.3°662.0 to 1832.0±0.6°
R-18.3 to 100.0±2.1°-1.0 to 212.0±3.8°
100.0 to 500.0±1.3°212.0 to 932.0±2.4°
500.0 to 1400.0±1.0°932.0 to 2552.0±1.8°
1400.0 to 1767.8±1.2°2552.0 to 3214.0±2.0°
Note: Doesn’t include cold junction error of ±0.05°C
°CDegrees F
Range
°F
Page 41
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/CDegrees C
Range
S-18.3 to 100.0±2.0°-1.0 to 212.0±3.7°
100.0 to 350.0±1.4°212.0 to 662.0±2.5°
350.0 to 1600.0±1.1°662.0 to 2912.0±2.0°
1600.0 to 1767.8±1.3°2912.0 to 3214.0±2.4°
B315.6 to 600.0±3.2°600.0 to 1122.0±5.7°
600.0 to 850.0±1.7°1122.0 to 1562.0±3.1°
850.0 to 1100.0±1.3°1562.0 to 2012.0±2.4°
1100.0 to 1820.0±1.1°2012.0 to 3308.0±2.0°
N-230.0 to -150.0±1.9°-382.0 to -238.0±3.4°
-150.0 to -50.0±0.7°-238.0 to -58.0±1.2°
-50.0 to 950.0±0.4°-58.0 to 1742.0±0.8°
950.0 to 1300.0±0.5°1742.0 to 2372.0±1.0°
G
100.0 to 350.0±1.7°212.0 to 662.0±3.0°
(W)
350.0 to 1700.0±0.8°662.0 to 3092.0±1.5°
1700.0 to 2000.0±1.0°3092.0 to 3632.0±1.8°
2000.0 to 2320.0±1.1°3632.0 to 4208.0±2.1°
C
-1.1 to 100.0±0.8°30.1 to 212.0±1.4°
(W5)
100.0 to 1000.0±0.7°212.0 to 1832.0±1.3°
1000.0 to 1750.0±1.2°1832.0 to 3182.0±2.1°
1750.0 to 2320.0±2.0°3182.0 to 4208.0±3.5°
Note: Doesn’t include cold junction error of ±0.05°C
°CDegrees F
Range
°F
Page 42
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/CDegrees C
Range
D-1.1 to 150.0±1.0°30.1 to 302.0±1.8°
150.0 to 1100.0±0.7°302.0 to 2012.0±1.3°
1100.0 to 1750.0±1.0°2012.0 to 3182.0±1.8°
1750.0 to 2320.0±2.0°3182.0 to 4208.0±3.6°
P0.0 to 600.0±0.3°32.0 to 1112.0±0.6°
600.0 to 900.0±0.4°1112.0 to 1652.0±0.8°
900.0 to 1200.0±0.6°1652.0 to 2192.0±1.1°
1200.0 to 1395.0±0.7°2192.0 to 2543.0±1.2°
L
-200.0 to -50.0±0.4°-328.0 to -58.0±0.7°
J-DIN
T-DIN
Note: Doesn’t include cold junction error of ±0.05°C
-50.0 to 300.0±0.2°-58.0 to 572.0±0.4°
300.0 to 900.0±0.3°572.0 to 1652.0±0.5°
U
-200.0 to -50.0±0.6°-328.0 to -58.0±1.1°
-50.0 to 50.0±0.3°-58.0 to 122.0±0.5°
50.0 to 550.0±0.2°122.0 to 1022.0±0.4°
550.0 to 600.0±0.3°1022.0 to 1112.0±0.5°
°CDegrees F
Range
°F
Page 43
RTD Ranges & Accuracies
RTD Accuracy Based on
RTD
Type
Pt 100 Ohm
DIN/IEC/JIS 1989
1.3850 (ITS-90)
Pt 100 Ohm
(Burns)
1.3902
Pt 100 Ohm
(Old JIS 1981)
1.3916
Pt 100 Ohm
(US Lab)
1.3926
Pt 1000 Ohm
DIN/IEC/JIS 1989
1.3850
Copper 10 Ohm
1.4274 (Minco)
Copper 50
1.4280
Ni 120 Ohm
1.6720 (Pure)
Ni 110 Ohm
Bristol 7 NA 1.5801
Degrees C
-200.0 to 0.0
340.0 to 640.0
640.0 to 850.0
-200.0 to 10.0
10.0 to 350.0
350.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 660.0
660.0 to 850.0
-200.0 to 0.0
340.0 to 640.0
640.0 to 850.0
-200.0 to 260.0
-50.0 to 150.0
-80.0 to 260.0
-100.0 to 260.0
±(0.025% of Reading in Ohms + 0.075 Ohms)
Range
0.0 to 340.0
0.0 to 340.0
°C
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±2.0°
±0.4°
±0.1°
±0.2°
Degrees F
Range
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 50.0
50.0 to 662.0
662.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1220.0
1220.0 to 1562.0
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 500.0
-58.0 to 302.0
-112.0 to 500.0
-148.0 to 500.0
°F
±0.4°
±0.6°
±0.8°
±1.0°
±0.4°
±0.6°
±0.8°
±0.9°
±0.4°
±0.6°
±0.8°
±0.9°
±0.4°
±0.6°
±0.8°
±0.9°
±0.4°
±0.6°
±0.8°
±1.0°
±3.6°
±0.8°
±0.3°
±0.4°
Page 44
Standard Warranty
Our equipment is warranted against defective material and
workmanship (excluding batteries) for a period of three years
from the date of shipment. Claims under warranty can be made
by returning the equipment prepaid to our factory. The equipment
will be repaired, replaced or adjusted at our option. The liability
of Practical Instrument Electronics (PIE) is restricted to that
given under our warranty. No responsibility is accepted for
damage, loss or other expense incurred through sale or use of
our equipment. Under no condition shall Practical Instrument
Electronics, Inc. be liable for any special, incidental or
consequential damage.
Pressure sensors that have been damaged by over
pressurization or contaminated by process chemicals are not
covered by our warranty. Pneumatic pumps that are contaminated
with process chemicals are also not covered by our warranty.
Optional Repair/Replacement Warranty
Under our Repair/Replacement Warranty (RP-WAR-B), our
equipment is warranted against ANY damage or malfunction
that may cause the unit to fail for a period of three (3) years from
the date of shipment.
This warranty is limited to one complete replacement against
any damage or malfunction during the warranty period. If
replaced, the new calibrator will carry our Standard Warranty for
the remainder of the three (3) years or a minimum of one (1)
year from the date of shipment.
Additional Information
PIE Calibrators are manufactured in the USA.
calibrated on equipment traceable to NIST and includes a
Certificate of Calibration. Test Data is available for an additional
charge.
Practical Instrument Electronics recommends a calibration
interval of one year. Contact your local representative for
recalibration and repair services.